Non Contact Infrared Body Thermometer: User Manual
Non Contact Infrared Body Thermometer: User Manual
Non Contact Infrared Body Thermometer: User Manual
ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer Page 1 / 18
Instructions for use
HeTaiDa
Model: HTD8823US
Foreword
The non contact Infrared body thermometer operating Instructions intend to provide
the necessary information for proper operation of HTD8823US thermometer model.
Only body mode was reviewed and certified by notified body.
General knowledge of Infrared thermometer and an understanding of the features and
functions of the HTD8823US thermometer model are prerequisites for proper use.
The HTD8823US is a medical device, and can be used repeatedly with using life is 5 years.
And the shelf life of HTD8823US is 5 years too.
Please read the manual first before using it, if not fully understand the usages, please stop
using the thermometer.
NOTICE
Purchase or possession of this device does not carry any express or implied license to use
with replacement parts which would, alone or in combination with this device, fall within the
scope of one of the relating patents.
2
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 3 / 18
Page
Instructions for use
Safety Information
This device may only be used for the purposes described in these instructions. The
manufacturer cannot be held liable for damage caused by incorrect application.
The non contact infrared body thermometer is designed to minimize the possibility of hazards
from errors in the software program by following sound and light engineering design
processes, Risk Analysis and Software Validation.
Warning
The device should be kept out of the reach of children/pets. When not in use, store the
3
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 4 / 18
Page
Instructions for use
device in a dry room and protect it against extreme moisture, heat, lint, dust and direct
sunlight. Never place any heavy objects on the storage case.
Do not use this device if you think it is damaged or notice anything unusual.
This device comprises sensitive components and must be treated with caution. Observe
the storage and operating conditions described in the ‘Technical Specifications’ section.
When using, shall not touch battery and the patient simultaneously.
Do not use the device if it is damaged/ degraded/loosened in any way. The continuous
use of a damaged unit may cause injury, improper results, or serious danger.
Based on the current science and technology, other potential allergic reactions are
unknown.
This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS.
4
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 5 / 18
Page
Instructions for use
Section 1- Overview
Indication for use:
The Non-contact Infrared Body Thermometer, Model: HTD8823US, is an electronic clinical
thermometer using an infrared sensor to detect body temperature from the forehead in people
of all ages for home setting use.
Description of Non Contact Infrared Body Thermometer
Device principle and introduction
The HeTaiDa Non contact infrared body thermometer are hand-held, reusable,
battery operated devices, which can measure human body temperature on forehead,
the skin temperature on one's forehead.
The operation principle is based on Infrared Sensor technology. The IR sensor can
output different signal when measuring different object temperature or in different
ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital
value and display it on the LCD.
Description on Controls, Indicators, and Symbols
5
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 6 / 18
Page
Instructions for use
Thermometer Applications
Equipment Symbols
Operating atmospheric
Warning
pressure
6
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 7 / 18
Page
Instructions for use
Technical Specifications
Measurement Unit °C /°F (HTD8823US)
Operating mode Adjusted mode(Body mode)
Reference Body Site Oral
Rated output range 34.0°C ~ 43.0°C / 93.2°F -109.4°F
Range 34.0°C ~ 43.0°C / 93.2°F -109.4°F
Body mode:
34.0°C ~ 34.9°C : ± 0.3°C / 93.2°F - 94.8°F : ± 0.5°F;
Accuracy
35.0°C ~ 42.0°C : ± 0.2°C / 95.0°F - 107.6°F : ± 0.4°F;
42.1°C ~ 43.0°C : ± 0.3°C 107.8°F - 109.4°F : ± 0.5°F;
Display Resolution 0.1°C /0.1°F (HTD8823US)
34.0°C - 35.7°C / 93.2°F - 96.3°F: no backlight;
Three-color Backlight 35.8°C - 37.4°C /96.2°-99.4°F: Green (Normal Temperature);
(Color Alarm) 37.5°C - 38.5°C / 99.5°F - 101.3°F: Yellow (Slight Fever);
38.6°C - 43.0°C / 101.4 - 109.4°F: Orange(High Fever)
Auto Power Off Time ≤18s
Forehead mode: ≤ 2s
Measuring Time
Forehead scan mode: 3 ~ 10s
Memories 1
Power Supply Requirements
Batteries 1.5V (AAA) Alkaline batteryX2 (IEC Type LR03)
Adaptable Range 2.6V~3.6V
Environmental
Operating Temperature: 5°C ~40°C (41°F ~104°F),
Operating Condition Relative Humidity ≤ 85% (no condensing)
Atmospheric pressure: 70 Kpa -106Kpa
Storage Temperature: - 20°C -55°C / -4 °F - 131°F,
Transport and Storage
Relative Humidity ≤ 95% (no condensing)
Condition
Atmospheric pressure: 70 Kpa -106 Kpa
Dimension and Weighting
Weight (without batteries) 47g
Size L: 122mm X W: 30mm X H: 43mm
Compliance
Item Compliant with
7
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 8 / 18
Page
Instructions for use
2- Operation
2.1 Battery installation
Caution: The Non Contact Infrared Body Thermometer does not operate with dead
batteries and does not input outer power. Install new batteries.
1) Pull the battery downward, toward the bottom of the Non Contact Infrared Body
Thermometer, and remove the battery access door;
2) Insert two pieces AAA size batteries according to the “+” and “-”;
3) Close the battery cover.
2.2 How to Operate
Before Applying the Thermometer
Be sure to read and understand all warnings listed of the instructions before use.
8
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 9 / 18
Page
Instructions for use
Forehead mode: press On/Scan button to open the device, the thermometer is
aligned with the middle of the forehead to measure body temperature (between the
eyebrows above) and keep the vertical distance, press the On/Scan button, the
position lamp indicates measurement site with light spot on the forehead,
temperature display immediately, see figure 3.
Forehead Scan mode: gently position the probe flush (flat) on the center of the
forehead, midway between the eyebrow and the hairline. Press and hold the
On/Scan button. Lightly slide the thermometer across the fore-head keeping the
sensor flat and in contact with the skin until you reach the hairline, release the
On/Scan button and remove the thermometer from the head, then the temperature
will display on the screen. The whole process takes 3~10 seconds.
When the ambient temperature of the thermometer changes too much, such as
moving the Thermometer from one place of lower temperature to another place of
higher temperature, Allow the thermometer to remain in a room for 30 minutes where
the temperature is between 5°C to 40°C.
The ambient temperature around the test person should be stable, should keep away
from the larger flow fan, air-conditioning vents and so on.
When people moving from one place of lower temperature to another place of
higher temperature, should at least remain in the test environment more than 5
minutes, to be consistent with the ambient temperature after the re-measurement.
Wait at least 1 second for the next measurement. If the continuous measurement of
five times, it is recommended to wait at least 30 seconds and then continue
measurement.
You cannot use the thermometer in place where the sun is strong.
9
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 10 / 18
Page
Instructions for use
2. And then show last measurement value (Figure 5), and back to ready display
for measurement displayed ‘- - - -‘. (Figure 6)
3. Forehead mode: Align staff forehead to keep the distance, and then press the
On/Scan button, the position lamp indicates measurement site with light spot
on the forehead, and start the measurement, read the data.
4. Forehead scan method: position the probe flush (flat) on the center of the
forehead, midway between the eyebrow and the hairline. Press and hold the
On/Scan button. Lightly slide the thermometer across the fore-head keeping
the sensor flat and in contact with the skin until you reach the hairline, release
the On/Scan button and remove the thermometer from the head, then the
temperature will display on the screen. The whole process takes 3~10 seconds.
Note: 1) After full display over, you will hear a rattle or "bibi" four times, which means that
the measurements have been completed, while the target value of the measured
temperature is displayed on the LCD, while backlit display according to the
appropriate setting among the three colors green, yellow, orange one of. And the
Green means ready for next measurement. When 37.5°C ~ 38.5°C, it's yellow,
means slight fever warning. Please pay attention to body temperature. When the
10
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 11 / 18
Page
Instructions for use
body temperature is 38.6°C or above, it's orange, means high fever. Please take
action to cool down or go for a doctor.
3- Troubleshooting
MESSAGE SITUATION SOLUTION
The screen flicker, automatic turn Replace battery or the product has
off. been damaged, needs repairs.
11
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 12 / 18
Page
Instructions for use
Warning
1.The typical service life of the new and unused batteries is 2000 measurements for the
operation time is 18s.
2. Only use the recommended batteries, do not recharge non-rechargeable batteries and
do not burn them.
3. Remove the batteries if the thermometer is not to be used for a long period.
12
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 13 / 18
Page
Instructions for use
This thermometer is for single patient reuse, you may use 70% isopropyl alcohol wipes to gently
clean thethermometer after each use.
Rubbing device (including markings) by hand without undue pressure with 70% isopropyl alcohol,
especially skin contact part, such as the probe, shell, LCD for 15s after each measurement, wiping
sensor of thermometer (which is in the center of probe surrounded by ABS case) gently for 3s
using a cotton swab wetted 70% isopropyl alcohol if dirty.
Always keep the thermometer a within the storage temperature and humidity range as
specified.
Always keep the thermometer within the storage temperature range (- 20°C to 55°C or – 4°F
to 131°F) and humidity range (≤ 95% non-condensing)
It is recommended to store the thermometer in a dry location free from dust. Do not expose
the thermometer to direct sunlight, high temperature/ humidity or any extreme environment,
otherwise the function will be reduced.
When the ambient temperature of the thermometer changes too much, such as moving the
thermometer from one place of lower temperature to another place of higher temperature,
allow the thermometer to remain in a room for 30 minutes where the temperature is between
5°C to 40°C.
6-Disposal
1) Used batteries should not be disposed of in the household rubbish. Used Batteries
should be deposited at a collection point.
2) At the end of its life, the appliance should not be disposed of in household rubbish.
Enquire about the options for environment-friendly and appropriate disposal. Take local
regulations into account.
13
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 14 / 18
Page
Instructions for use
7-Warranty
Our company warrants Non Contact Infrared Body Thermometer at the time of its original
purchase and for the subsequence time period of one year.
The warranty does not cover the followings:
The device series number label is torn off or cannot be recognized.
Damage to the device resulting from misconnection with other devices.
Damage to the device resulting from accidents.
Changes performed by users without the prior written authorization of the company.
Batteries and packaging are not covered under warranty
When asked to provide warranty service, you must have a purchase date and purchase
stamp dealers (including dealers name and address) of the warranty card. Be sure to ask the
dealer to purchase this product signature on the warranty card. When asked to provide
warranty service, please put the product to get our distribution points for repair. Products
outside the warranty expires, will be charged accordingly.
Note:
1. If you have any problems with this device, such as setting up, maintaining or using,
please contact with SERVICE PERSONNEL of HeTaiDa Technology Co., Ltd. Don’t open or
repair the device by yourself.
2. Please report to HeTaiDa Technology Co., Ltd. if any unexpected operation or events
occur.
3. Calibration needed to ensure proper function for every two years or after device impact.
4. The patient is an intended operator. The patient can measure and change battery.
Under normal circumstances and maintain the device and its accessories according to the
user manual.
14
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 15 / 18
Page
Instructions for use
EMC Declaration
1) This equipment needs to be installed and put into service in accordance with the information
provided in the ACCOMPANYING DOCUMENTS;
This product needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided, and this unit can be affected by portable
and mobile RF communications equipment.
2)* Caution: Do not use a mobile phone or other devices that emit electromagnetic fields, near
the unit. This may result in incorrect operation of the unit.
3) *Caution: This unit has been thoroughly tested and inspected to assure proper performance
and operation!
4) * Caution: this machine should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, this machine should be observed to verify normal
operation in the configuration in which it will be used.
Voltage fluctuations/
flicker emissions Not applicable
IEC 61000-3-3
15
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 16 / 18
Page
Instructions for use
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency 30A/m 30A/m Power frequency magnetic fields
(50Hz/60Hz) magnetic should be at levels characteristic of a
field IEC 61000-4-8 typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
16
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 17 / 18
Page
Instructions for use
Compliance
Immunity test IEC 60601 test level Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the Non
Contact Infrared Body Thermometer, including cables,
than the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Conducted RF 3 Vrms Not applicable
Recommended separation distance
IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 10V/m 10 V/m Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
IEC 61000-4-3 80 MHz to 2.5 GHz
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a should
be less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Non Contact Infrared Body
Thermometer is used exceeds the applicable RF compliance level above, the Non Contact Infrared Body
Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the Non Contact Infrared Body Thermometer.
17
File No. ZHTF-CE-01-015
Technical File Rev. No. A.1
Non Contact Infrared Body Rev. Date 2020-6-8
Thermometer 18 / 18
Page
Instructions for use
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
portable and mobile RF communications equipment and the Non Contact Infrared Body Thermometer.
The Non Contact Infrared Body Thermometer is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Non Contact Infrared Body Thermometer can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Non Contact Infrared Body Thermometer as recommended below,
according to the maximum output power of the communications equipment.
(m)
Rated maximum output
power of transmitter 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(W)
18