Manual Verifier HR FSX 090
Manual Verifier HR FSX 090
Manual Verifier HR FSX 090
Operations &
Reference Manual
Model: FSX‐090
High‐Resolution X‐Ray Inspection System
Portable, X-Ray Inspection
The Verifier Series are affordable x-ray systems
suitable for virtually any quality inspection
application. Electronics and material analysis,
quality inspection, proto design, failure analysis,
manufacturing process validation, rework
verification and more…
FocalSpot, Inc.
9915 Businesspark Avenue, Suite A
San Diego, California 92131
Tel: (858)-536-5050
Fax: (858)-536-5054
Email: [email protected]
Verifier Series, Operation & Maintenance Manual
Legal Notice
Verifier Operations & Reference Manual
Copyright 2009 by FocalSpot, Inc.
The information in this publication is provided for reference only. FocalSpot, Inc. shall not be liable for
errors or omissions contained herein nor for incidental or consequential damages in connection with
the furnishing, performance, or use of this material. This publication may contain or refer to
information and products protected by copyrights or patents not held by FocalSpot, Inc. and does not
convey any license under the patent rights of FocalSpot, Inc nor the rights of others.
Indemnification
This manual is shipped with the equipment and is considered as is, without warranty of any kind,
either expressed or implied, respecting the contents of this manual, including but not limited to implied
warranties for the manual’s quality, performance, merchantability, or fitness for any particular purpose.
Neither FocalSpot, Inc nor its dealers, agents, representatives, distributors, or the authors shall be
liable to the purchaser or any other person or entity with respect to any liability, loss or damage
caused or alleged to have been caused directly or indirectly by this manual.
Revision History
Legal Notice 2
Verifier Operations & Reference Manual 2
Copyright 2009 by FocalSpot, Inc. 2
Intellectual Property Rights 2
Indemnification 2
Introduction 8
System Components 8
User Responsibility 9
Receiving & Inspection 9
Potential, Current, and Duty Cycle Rating 9
X-Ray Warning Labels 10
Radiation Safety Regulation References 14
Safety Precautions 14
Terms Used in this Manual 14
Installation Note 14
Installation 17
System Requirements 17
AC Power 17
Clearance 17
Environment 17
Safety 17
Inspection 17
Cabling 18
Main Power Line Cord 18
Computer, Monitor, and Printer 18
Initial Setup 19
The Rear Machine Panel 19
Front Control Panel 19
Before You Begin 21
Major Machine Components Location, Description, and Key Items 21
Rear Panel 21
Front Panel 22
System Controls and Indicators 22
Image Intensifier (I.I.) 24
4”/2” Function Switch 24
Imaging Control Switches 25
Chassis Compartments 26
X-Ray ON Indicators (Locations) 28
Starting the System 30
Cold Start Procedures 30
X-Ray Source Warm-Up 30
Chapter 1
Introduction
System Components
User Responsibility
Receiving and Inspection
Potential, Current, and Duty Cycle Rating
X-Ray Warning Label
Radiation Safety Regulation Reference
Safety Precautions
Terms Used in this Manual
Installation Note
Introduction
FocalSpot’s Verifier HR, model number FSX-090, X-ray System is used for microfocus real-time
inspection of electronic components, cables, multi-layer circuit boards, sealed components, and
electronic SMT assemblies.
The Verifier Series are stand-alone self enclosed cabinet x-ray systems which are fully radiation
shielded, interlocked and key switch controlled.
System Components
The standard Verifier HR system consists of the following:
Lead Lined cabinet that houses the x-ray tube, detector, and positioner.
90 kV, 5 micron x-ray tube with integrated power supply
Image Intensifier with Zoom Lens and CCD Camera
Control Panel with Adjustment controls, Key switch, E-Stop and indicators for x-ray generation
and detection.
X-Y manual positioner including: adjustable board width fixture and variable board height
positioning.
Standard PC with Microsoft Operating System
19 inch Flat Panel Monitor
Printer
Verifier Imaging Software.
User Responsibility
Your Verifier HR X-Ray System will perform in conformity with the description contained in this
operating manual and the accompanying labels and/or inserts when assembled, operated, maintained,
and repaired in accordance with the instruction provided.
The system must be checked periodically. A defective system should not be used. Parts that are
broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should
such repair or replacement become necessary, FocalSpot Inc. recommends that you telephone, email,
or Fax a request for service from FocalSpot Inc. or its nearest affiliates.
Your Verifier HR system or any of its parts should not be repaired other than in accordance with
written instructions provided by FocalSpot Inc., and FocalSpot Inc. trained personnel.
The system must not be altered without the prior written approval of FocalSpot, Inc.
The user of this product shall have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or alternation by anyone other than
FocalSpot, Inc.
You are required to register the use of this X-Ray System with the state in which the product is
installed; contact FocalSpot Inc for additional information.
The Verifier HR system is carefully inspected mechanically, electrically, and for radiation safety before
shipment and should be free from damage. As a normal part of receiving, please do the following:
1. When the system is delivered by commercial carrier, check it for damage with the carrier.
Contact the carrier’s office and FocalSpot Customer Service if any damage is found.
2. When unloading from commercial carrier, DO NOT use any hooks.
3. Following the uncrating instructions attached to the shipping crate. If you have any questions,
contact FocalSpot, Inc. Customer Service at the number provided on the inside title page of
this manual.
4. Compare the packing list with your order invoice. If you find any discrepancies, contact
FocalSpot, Inc or local affiliates.
5. Save all packing material for the system in case it should ever have to be moved or shipped
again.
System Rear
X-Ray Interlock
Inspection Load Door, Control Panel and Leaded Glass Window Labels
Leaded Glass Inspection Area Viewing Window Leaded Glass Warning Label
(Bottom Center of Window)
FDA U.S. Food and Drug Administration 21 CFR Ch I (4-1-91 Edition), Subchapter J –
Radiological Health, Part 1000 – General, paragraph 1020.40, Cabinet X-Ray Systems.
http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?FR=1020.40
U.S. Department of Commerce/National Bureau of Standards NBS Handbook 114, General Safety
Standards for Installations Using Non-medical X-Ray and Sealed Gamma Ray Sources, Energies
up to 10MeV. http://epswww.unm.edu/xrd/nbs111.pdf
Health Canada: Radiation Emitting Devices Regulations (REDA), Schedule II, Part XV, Cabinet X-
Ray Equipment. http://www.canlii.org/en/ca/laws/regu/crc-c-1370/latest/crc-c-1370.html
Safety Precautions
Specific notations are used in this manual to call attention to conditions which could result in injury,
damage to the system, or require special attention.
Operation: Means the use of the FSX-090-DCC system over the full range of its functions.
Maintenance: Means the performance of those adjustments or procedures specified in this
manual which are to be performed by you.
Service: Means the performance of those adjustments or procedures which are to be performed
only by individuals certified by FocalSpot, Inc.
Installation Note
The installation procedures described in Chapter 2 of this manual are provided for informational
purposes only. Your system must be installed by a qualified FocalSpot Service Representative, x-ray
certified and documented prior to use.
Chapter 2
Installation
The installation procedures described in this chapter are provided for pre-installation planning
purposes only. Your FSX-090-DCC FocalSpot X-Ray System must be installed by a FocalSpot, Inc.
Service Representative or Authorized Affiliate.
System Requirements
AC Power
Your FSX-090-DCC system is equipped with a three-conductor line cord in accordance with NEMA
requirements. When plugged into the appropriate receptacle, the power cord is designed to ground
the equipment cabinet.
Clearance
Allow at least 30 inches in front of the cabinet for operating the system
Allow at least 4 inches between the rear of the system and wall for air circulation.
Environment
The environment for your FSX-090-DCC system should be free from excess dust and dirt, and the
floor should not vibrate. The system is designed to tolerate the following temperatures and humidity
ranges:
Safety
This system is designed to conform to Federal Regulations (CFR) 21, subchapter J, paragraph
1020.40, “Cabinet X-Ray Systems.” Various localities may have different rules and regulations that
must be complied with prior to installation. Make an inquiry as to these rules and regulations with the
installation facility’s Safety Department and with the local Department of Health. Notify these officials
the installation is taking place and inquire about any special plant requirements that may apply.
Requirements may involve wiring, notices, or special safety procedures unique to that particular
company. Take whatever steps necessary to comply with these requirements.
Inspection
Carefully inspect the FSX-090-DCC for any physical damage. Look for:
Dislodged components
Loose cables
Dents, Broken Glass, Damaged casters, bent frame members.
NOTE: Do not discard the packing material. If you find damage, the carrier may want to examine the
material. Also, in the event the system needs to be returned to FocalSpot, it should be shipped in the
original containers.
Record the system serial number on the Service Installation form. The serial number appears on the
back of the cabinet.
Compare the shipped items to those listed on the packing slip.
Report any discrepancies to FocalSpot Customer Service. Contact FocalSpot Customer Service
directly at: 1-858-536-5050.
Cabling
These instructions tell you how to route and connect all cables for the system and its components.
Before you begin, locate all cables.
Initial Setup
Familiarize yourself with the Rear Machine and Front Control panels.
1. Set the x-ray voltages (x-ray Voltage, kilovoltage, kV) to the lowest setting by turning the kV
adjust knob all the way to the left (counter clockwise)
2. Set the x-ray current (x-ray current, milliamp, mA) to the lowest setting by turning the mA
adjust knob all the way to the left (counter clockwise)
3. Turn the Computer On, Turn the Monitor On.
4. Insert the key into the Power key switch on the front control panel.
Chapter 3
WARNING Never operate this system if either of these conditions exists. The viewing glass
window is cracked or broken; the lead lining is separating from the inside of the
sample compartment.
WARNING This equipment produces x-rays when energized. Follow all safety procedures set by
the installation site.
Rear Panel
The main machine power switch and fuse block is located on the rear machine panel. The Main
Power Switch provides AC power to the machine and internal AC distribution blocks. Additionally, the
main power switch, upon activation, starts the cooling fan and turns on the light in the interior of the
machine.
Removing the top panel will provide partial access to the upper electronics controls and the motion
components (pulley and belt, tensioner) of the x-y table in the sample compartment. Top panel
access is interlocked with a low voltage (24V DC) switch to disable x-ray generation.
Key Items:
1. Main AC Power Switch with Integrated Fuse Block
2. Machine Nameplate, including Serial Number
3. Warning Labels
4. Fan Guard
Note: Always refer to the machine serial number when contacting FocalSpot Customer Service.
Front Panel
The front control panel is located on the front of the
machine to the left of the load door. The front panel
is the main location for control over the x-ray and
image generation process.
Key Items
1. Main Power ON/OFF key-switch
2. Emergency Stop button
3. Power, X-ray On/Off, and Interlock LED
indicators
4. X-ray On/Off buttons
5. X-ray Voltage (kV) control knob
6. X-ray Current (mA) control knob
7. Camera Gain control switch
8. I.I. (2”/4”) rocker switch
9. Camera Lens Zoom control rocker switch
10. ESD Ground terminal
Camera Gain Switch A Camera Gain Switch is used to increase or decrease the gain of the
CCD camera. Increasing the gain of the CCD camera increases camera
sensitivity (brightness). This is most often used when viewing very dense
materials. However, increasing camera gain too much can introduce noise
making the image appear grainy and overly contrasted.
Peripheral Controls * Located to the right and/or left side of the keyboard tray
I.I. (2” / 4”) rocker switch The I.I. (2” / 4”) rocker switch is used to select either 2” or 4” mode
operation of the dual-mode image intensifier.
2” Mode — Small FOV mode (High Magnification)
4” Mode — large FOV mode (Low Magnification)
Camera Lens Zoom rocker switch The Camera Control (Zoom) rocker switch is used to adjust the camera
lens optical magnification (zoom-in / zoom-out). Pressing either the plus
side of the rocker switch will increase optical magnification making the
object larger to provide improved diagnostic capabilities. Pressing the
minus side of the rocker switch will decrease magnification thus providing
a larger overall view of the entire assembly. Lower magnification provides
QuickCheck™ capability to quickly and easily locate bridging, missing
solder, or other gross solder defects.
Indicators
Main Power ON LED Indicator The GREEN LED Power Indicator signifies that both AC and DC power
are energized
X-Ray ON LED Indicator The RED LED X-Ray ON Indicator signifies production of x-ray radiation
Interlocks OK LED Indicator The YELLOW LED Interlock Indicator signifies that all High Voltage and
Low Voltage interlock safety switches are satisfied. Interlocks are located
on every access panel or door.
Image Intensifier (I.I.)
Definition: An X-ray Image Intensifier converts low intensity x-rays into light. Through this process,
several thousand light photons are produced for each x-ray photon, which are captured by a CCD
Video Camera. This process produces a visible real-time image and the resulting video signal is then
interfaced with an Image Processor for enhancement, display and analysis.
4”/2” Function Switch
A 2”/4” or “Dual-Mode” Image Intensifier offer two settings:
2” Mode: Low Magnification, Large Field-of-View (FOV), this is used to display as much of the
inspection area at one time to facilitate fast inspection.
4” Mode: High Magnification, Small Field-of-View (FOV), this is used to provide greater detail
and presents more information to reveal subtle abnormalities.
As a rule, a good balance between sufficient magnification (inspection detail) and sufficient FOV (to
view as much of the inspection area at one time to facilitate fast inspection). With high-magnification,
the field-of-view (FOV) becomes smaller thereby reducing the speed at which a board can be
completely inspected. Therefore, throughput and magnification are opposing objectives and the best
balance of each depends upon your application.
The 2/4” toggle has been added to the Verifier X-Ray Control Panel to provide the end-user with this
capability and convenient access to this control.
Imaging Control Switches
Chassis Compartments
Chassis Diagram Compartment Description
Upper Electronics
The upper electronics compartment
Compartment houses the major electronic
components, including the x-ray tube,
power supplies, relays, AC power
distribution, and an x-ray control /
safety PCB.
Sample
The sample compartment contains
Compartment the manual x-y table. Samples
(boards, trays, cables, etc) can be
loaded onto the x-y table at several
different height positions. To
accommodate variable board sizes, the
width of the fixture can be adjusted by
unlocking the key lock mechanism and
sliding the board holder apart.
Lower Electronics
The lower electronics compartment
Compartment houses the Image intensifier, zoom
lens, and camera assembly.
Upper Electronics
Compartment
Sample
Compartment
Lower Electronics
Compartment
The upper
electronics
compartment tower
contains an x-ray
RED ON indicator
LED which
illuminates when x-
rays are ON and the
potential for x-ray
Tower X-Ray ON Indicator LED radiation is present.
The Control Panel
contains a redundant
x-ray RED ON LED
indicator which
illuminates when x-
rays are ON and the
potential for x-ray
radiation is present.
Chapter 4
X-Ray Operation
Starting the System
Cold Start Procedures
X-ray Source Warm-up
Turning X-ray production ON
Turning X-ray production OFF
2. Provide DC Power to the machine by turning on the Power ON/OFF switch, located on the Front
Panel.
Your x-ray tube has been fully processed during manufacturing. As such, it does not require any
special start-up procedures. However, as the generation of x-rays involves high voltages, it is
recommended that you follow a stepped pattern to reach the full power rating. This will ensure the
maximum life of your x-ray tube. Below is an example of a common stepped pattern procedure.
X‐Ray Source Warm‐Up
A burn-in period must always be implemented when making a cold start (power up of a machine after
power has been removed). The Verifier FSX-090-DCC is equipped with the latest integrated x-ray
tube and power supply and does not require special attention. Follow the cold start procedures
outlined above.
Turning X‐Ray Production ON
The x-ray tube can be activated as long as the safety interlocks are satisfied (YELLOW INDICATOR
LED) is ON. To activate x-ray energy; depress the RED x-ray ON Button (A buzzer will sound). If the
system has not been warmed-up, and to ensure a long lasting tube, follow the warm up procedure.
When the x-ray tube has been warmed up, increase the kV slowly. Note: Upon initial power up, the x-
ray on button will not activate for 10 seconds.
Turning X‐Ray Production OFF
The x-ray tube can be de-activated by depressing the GREEN X-ray OFF button. Prior to shutting
OFF the x-ray system, turn the x-ray kV and mA control knobs fully counterclockwise to the minimum
kV and minimum mA positions.
2 Turn the key switch located on the front of the control panel to “OFF”.
3 Flip the ‘main power’ switch located at the upper right rear of machine
to “OFF”.
4 Unplug the main power cord from the electrical outlet.
Chapter 5
Applications
X-Ray Generating Operation
Controlling the Image
Loading the Board
Adjusting the Board Width
Positioning the Board for Magnification
Position the Board for Oblique Viewing
Manipulating the Board
Inspecting the Board
Verifier Software
NOTE: The Emergency Power OFF (EPO) button should not be used to turn the system OFF other
than in emergency situations. Using the EPO button for other than emergency shut off may shorten
the life of critical hardware components, including the Image Processor.
3 Camera Gain The camera gain is used to adjust the internal gain of the camera
allow the camera signal to be amplified. The gain is used together
with the kV, mA, and camera aperture to lighten, darken, and control
contrast of the object and the surrounding area.
4 Camera Focus The camera focus rocker switch is used to focus the image and is
most useful when the camera zoom is used to magnify the image.
5 Camera Zoom The camera zoom rocker switch control the zoom lens of the
camera. The zoom lens is used to provide magnification without
moving the object.
6 Dual 2” / 4” Image The image intensifier switch is used to change from 4” to 2”
Intensifier Settings intensifier output screen. This change over increases/decreases
overall image magnification by 2X, depending on the current settings.
This feature provides additional magnification when needed by the
operator/user.
Sample Handling
Loading the Board
The Verifier HR has a unique sample loading fixture to allow the PCBA
to be held on rails and therefore ensure it is flat. This fixture is
adjustable in width to accommodate small to large boards. And, by
using the adjustable width feature, a board can be loaded at an angle
to the x-ray imaging system. Shown w/ Delrin Tray
Adjusting the Board Width
The PCBA sample fixture can be adjusted by unlocking the lock down
knob. Turn the knob counterclockwise to disengage the key locking
mechanism. This allows the fixture to slide to allow for different sides
boards to be held
Load the Board onto the Fixture and into the Imaging Zone
Load the sample with the object of interest toward the left hand side of
the fixture. The Verifier HR was designed to accommodate a 16 x 18
inch sample with a scan area of greater than 8 x 9. Therefore, review
the board and object you wish to inspect and load the sample such
that the area of interest is toward the left hand side and to the rear.
Adjusted to Maximum Width
You may need to rotate or flip the board to position the board into the
inspection area of interest & imaging zone.
Position the Board for Magnification
The fixture has rails which can be secured into 11 different positions.
If the rails are placed in the highest slots, this will deliver the maximum
magnification. Conversely, if the rails are placed into the lowest slots,
this will deliver the lowest magnification and largest viewing area.
To move the rails from one slot to another, use the hex key to unscrew PCB Loaded Angular on Rails
the mounting screws. Move the rails and tighten the mounting screws.
Manipulating the Sample
Use the motion control knob to move the board from front to back and
right to left (x-y). Pull the handle out to translate the board toward the
front of the machine (y). Push the handle in to translate the board
toward the back of the machine. Rotate the handle counterclockwise
Multi-Axis Manipulation w/
to move the board to the left and clockwise to move the board to the
Motorized Tilt & Rotate
right.
Motorized, Tilt/Rotate (Oblique Viewing Fixture)
The Motorized Tilt/Rotation Fixture is a remotely actuated multi-
position sample holder designed to mount directly to the XY Table.
This fixture provides 360° rotation and +/- 90° tilt, operated by the
Rocker Switches mounted on the system cabinet on either side of the
keyboard.
Focus Control
The Right Micro Switch adjusts the focus of the displayed image.
Camera Aperture Control
The Left Micro Switch adjusts the camera aperture and provides
Gain (brightness) control to the displayed image.
Motorized Rotation Control
The Right Rocker Switch [Optional] controls the rotate operation of
the motorized PCB fixture.
Motorized Tilt Control
The Left Rocker Switch [Option] controls the tilt operation of the
optional motorized PCB fixture.
Bridge Detection
Use lower kV and mA to find bridges or shorts. Using the kV control knob, turn the knob slightly
clockwise and maintain it in the “bridge” zone.
Void Detection
Use higher kV and mA to find voids inside the soldered connection. Use the kV control knob and turn
the knob clockwise into the “void” zone.
Metal Lids
Use higher kV and mA to properly penetrate BGA devices with metal or ceramic lids. Use the kV
control knob and turn the knob clockwise into the “void” zone.
Shorted / Poor Reflow / Poor Shape Metal Lid Good BGA connection
Verifier HR Software
X-Ray Imaging Workstations are used to perform x-ray inspection, nondestructive test (NDT),
diagnostic radiology, failure analysis, industrial metrology, manufacturing process validation and
rework verification. The VIP Series provides advanced image-processing algorithms to assist with
inspection analysis. The (3) three workstation / Image Processor options (below) were developed to
address a broad spectrum of applications and budgetary requirements from: cost-effective essential
tools to precision measurement, advanced analysis and automation tools.
The Verifier HR comes complete with Standard x-ray imaging software which is most frequently used
by x-ray inspection operators.
VIP Standard
VIP is Windows™ based Standard
Image Processing software that
provides fundamental tools for x-ray
image capture, file saving and printing,
line measurement and sharpen filtering.
See the VIP and VIP Plus Software
Operations Manual for additional
information.
VIP Plus
VIP Plus is a Windows™ based
Enhanced Image Processing software
that provides additional capabilities for
BGA and void measurements, and
analysis reporting. See the VIP and
VIP Plus Software Operations Manual
for additional information.
VIPx
VIPx provides a rich Windows™ based
toolbox for image enhancement,
filtering and analysis of PCBs,
components and mechanical parts.
VIPx delivers analytical (pass/fail)
reporting based on user defined
inspection criteria. See the VIPx
Software Operations Manual for
additional information.
Chapter 6
Maintenance
Introduction
Recommended Intervals
Ordering Information
Warranty
Maintenance
Introduction
In order to keep this FocalSpot unit operating properly, it must be inspected and checked in
accordance with the recommended intervals depending on the age and use of the equipment. If there
is a problem, call an authorized serviceman.
Warning
DO NOT OPERATE THE EQUIPMENT UNTIL REPAIRS ARE COMPLETED!
Recommended Intervals
Periodic preventive maintenance includes cleaning, and electrical testing, and any necessary
calibration adjustments. These maintenance procedures are to be at intervals not to exceed 12
months. The maintenance schedule is required to keep the unit in proper and accurate working
condition.
Ordering Information
To order replacement parts contact the FocalSpot service department at:
(858) 536-5050 between the hours of 6:30 AM and 6:00 PM Monday through Friday.
If replacement parts are for a unit still under warranty, be prepared to submit model numbers and
serial numbers to expedite your order.
Do not return your instrument without return authorization from the factory or authorized service
center. Always include the serial number, which is located at the rear of the instrument.
DELL™ Service
Computer Service is handled directly through DELL’s Client System. All DELL computers used on
FocalSpot, Inc. inspection systems are covered under the DELL Pro Support “Next Business Day On-
Site Service.” On-Site Service is designed to deliver convenient and timely warranty repair service at
your location.
NOTE: Computer maintenance, security and backup are the responsibility of the end user. Should
service be required for software setup, repair or reinstallation; these activities are not covered by
DELL Service.
Instructed and tested with regards to the potential hazards of exposure to radiation and
likewise have been instructed in the protocols and procedures necessary to safely ensure
against potential hazards, as well as the responsibilities to follow all safety guidelines.
Instructed and tested with regards to the importance of personal monitoring devices, and have
been issued such devices “Dosimeters”
See also FocalSpot X-Ray Safety Training documentation (on file and available by request).
Reference: Radiation safety certification classes http://www.radiationprograms.com/index.html
http://www.focalspot.com/BREEZE/XrayTheory/index.htm
This presentation provides an overview of applications, techniques and examples to assist with the
understanding of faults as they appear in x-ray images. See online:
http://www.focalspot.com/BREEZE/XraySignature/index.htm
The entire machine is surveyed. Each test point of the survey is measured and recorded on the
acceptance test form / radiation survey form. The test is performed once at the factory after testing the
machine in conformity to published performance capabilities and then once more at the customer's
site during at installation. The customer receives a copy of the test results.
Records of the test are maintained in the customer file and the customer is provided with a copy of the
test. A spreadsheet is maintained with the original baseline test and a sample of subsequent tests
recorded annually.
The Radiation Survey test performed at the factory is performed under worst case scenario at max
kVp and mA and with a scatter block. For the Insight 90, 90kVp and 0.200 mA are the maximum
output limits and are used during radiation safety testing.
Testing of the Insight Cabinet x-ray systems manufactured by FocalSpot, Inc is accomplished during
the final assembly process and factory acceptance testing and during the customer installation.
FocalSpot uses a both a Geiger Muller (GM) type and Ion Chamber Radiation meters to validate
compliance. The GM meter is used to identify any leakage or All survey testing is performed on
calibrated meters (annual calibration cycle). FocalSpot testing, both at the factory and at the
customer’s site is performed with the meter directly on the surface of the machine.
Note further that CDRH and REDA specify acceptable norms as: 0.5 milliroentgen per hour (mR/hr).
Naturally occurring radioactivity commonly referred to as “background”, can vary greatly from one
area of the world to another due to elevation and geographical properties. People living in Denver
receive two to four times as much radiation from cosmic rays as do tine people living in San Diego or
Los Angeles, and people living in some parts of India and China receive 10 to 20 times as much
background radiation as do the people in other parts of the world. The ill effects to those people who
are exposed to these higher doses of radiation appear to be quite small.
Types of Radiation
Radioactive materials consist of atoms that have more energy than they can retain and in order to get
down to a stable state, they emit this energy as radiation. Radiation can actually take many forms,
both particulate and electromagnetic. Particulate radiation is small Particles of matter that generally
have an electric charge. Electromagnetic radiation is electric waves similar to light waves or radar
waves.
Electromagnetic Radiation
GAMMA RAYS AND X-RAYS are electromagnetic waves similar to radio waves, heat waves and
microwaves. The difference between these is wavelength; the gamma rays and x-ray have very short
wavelengths compared to the other waves. In fact the gamma and x-ray wavelengths are so short,
that they act like small particles and are called photons.
Ionizing Radiation
Gamma rays and x-rays, and radiation particles, are frequently, called ionizing radiation; while infrared
radiation, radio waves and microwaves are called non-ionizing radiation. The difference between the
two types is that gamma rays, x-rays and radiation particles can dislodge an orbital electron from its
position around the atom leaving behind a positively charge atom. This reaction is called ionization.
The other waves, radar and infrared, cause ionization because the energy that they contain is too
spread out to dislodge the electron.
The unit of absorbed dose is the RAD (Radiation Absorbed Dose). Radiation is measured in many
different units, depending on the type of radiation and the material it is interacting with. The original
radiation dose unit used was the Roentgen, and was defined as a specified electrical charge
produced in a volume of dry air by X-ray or gamma radiation. The Radiation Absorbed Dose (RAD) is
used for all types of radiation being absorbed in all types of matter, and is defined as the deposition of
100 ergs of energy per gram of matter. Since different types of radiation would produce different
effects in tissue, the Relative Biological Effectiveness (RBE) was developed, and each type of
radiation was tested to establish a Quality Factor (QF) number used in calculations from RAD to the
Roentgen Equivalent Man (REM). The unit used to measure the radiation absorbed dose, taking into
account the relative biological effectiveness of the different types of radiation, is called the REM or the
Sievert (Sv). REM is the old dose name and the Sv is the new dose term. Note: The federally
allowable dosage for an occupational worker to the whole body is 5,000 mR/year.
Radiation Detectors
A number of radiation detectors have been designed and they include Geiger counters, ion chambers,
and proportional counters which are all different types of detectors that depend on the ionization of
molecules in a gas to measure the passage of a radiation particle or photon. There are solid state
detectors such as dosimeter, scintillation detectors, and other electronic detectors that will detect and
measure radiation by various means. All of these detectors and monitors are used to identify the
particles or photons or to measure the radiation dose rate.
Exposure limits
The amount of radiation exposure a person receives is defined under the governing regulations of the
United States. Some of the agencies that establish these levels are the NRC (Nuclear Regulatory
Commission), CDRH (Center for Devices and Radiological Health [part of the FDA]), as well as state
departments of health and safety. These agencies have agreed on the following radiation exposure
limits, based on an occupational worker, 18 years or older, being exposed to these levels over their
working lifetime and suffering no observable ill effects.
• Whole body, blood forming organs, gonads 5,000 mRem/year
• Lens of eye 15,000 mRem/year
• Extremities and skin 50,000 mRem/year
• General public 100 mRem/year
• Pregnant Females 500 mRem/term
To give some understanding of these numbers, the following are some values of background readings,
as mentioned earlier, that are naturally occurring all around us:
• Cosmic radiation at sea level = 40 mRem/year
• Living in a wood house = 35 mRem/year
• Living in a concrete house = 50 mRem/year
• Working outdoors for eight hours/day = 15 mRem/year
• Food and water = 25 mRem/year
• Each 1500 mile airline flight = 1 mRem/year
• Each chest X-ray = 50-100 mRem
• Each dental X-ray = 20 mRem
As mentioned earlier, people occupationally exposed to radiation may receive up to 5,000 mRem/year,
with no observable effects. For comparison, some doses (in a 24 hour period) and their effects are as
follows:
Dose in Rems
• No obvious effect, except in minor blood changes.
• Vomiting and nausea for about one day, but no serious disability.
• Vomiting and nausea on first day, followed by loss of hair, fever, hemorrhage, diarrhea.
• About 20 percent deaths within two to six weeks.
• Vomiting and nausea within four hours of exposure. Up to 100 percent deaths, due to
secondary infection and damage to the bone marrow, blood system, and lymph system.
• 10,000 No survivors, all people die within two days due to severe internal damage.
Dosimeter Reports
The RSO shall review all radiation exposure reports to ensure occupational exposure limits are not
exceeded and conduct formal and informal investigations as needed. The RSO shall initial and date
each exposure report to indicate the review was conducted.
Record Keeping
Records related to occupational exposure to radiation must be retained indefinitely. Records related to
internal audits, surveys, inventories, training, and other safety program documentation must be kept
for a period of three (3) years. Radiation program records are stored FocalSpot, Inc. corporate office.
Radiation Protection
Time, distance and shielding are vital components to radiation safety, and should be kept in mind
when working with the machines.
Restrict the length of time you are exposed to a source of radiation, keep it as short as
possible.
Maintain the greatest possible distance between yourself and the source of radiation as is
practical.
Place as much shielding material between yourself and the source of radiation that will allow
the job to proceed without hindering the operation. Proper shielding is especially useful for
storage of radioactive materials.
The radiation symbol must be included on all radiation warning signs. The symbol is the
conventional three-bladed design, with the blades being colored magenta or purple on a
yellow background.
A Radiation sign must be posted in each area in which an individual could receive in one
hour body dose in excess of 5 mRem/hr.
A High Radiation sign must be posted in each area in which an individual could receive in
any one hour a dose to the whole body in excess of 100 mRem/hr.
• Techniques used in x-ray inspection applications rely heavily upon a users knowledge of the
inspected samples, as well as practical experience with x-ray inspection system operations.
• NOTE: Interpretations can be accomplished with or without the aid of X-Ray Image Processor
software tools.
There are several types of solder joint defects, each caused by unique conditions during the
manufacturing process. Excess or insufficient solder, uneven solder deposition and misalignment can
result in low joint strength and poor electrical connection, or defects such as:
• Bridging or a short is a connection of solder across conductors or lands that should not be
joined.
• Solder bridging is one of the most common and potentially destructive manufacturing faults.
Shorts present an obvious x-ray signature even in the smallest package types.
• Also shown in the image below is insufficient solder in some of the BGA joints. This condition
can range in size from visually obvious to unperceivable differences that can only be
calculated by image processing software measurement tools.
• Voiding is an open area caused by trapped gas inside a solder joint while cooling, aka “out
gassing.” If the some of the gas escapes during the solder process this can cause tiny “pin
holes” or “blow holes.”
• Voiding is also a common structural artifact in reflowed solder joints which can cause thermal
performance problems, especially in high-power devices. Otherwise, this is generally an
allowable condition as long as minimum soldering requirements have been met.
• To display voids clearly under x-ray requires using a slightly higher power setting which is why
the background of the image appears mostly white.
Voiding, Off-Axis BGA X-Ray Image Analysis, of 2D BGA X-Ray Voiding, BGA 3D Rendering
• Internal delamination of components, also called “pop corning,” due to rapid moisture out
gassing in the connector housing resins.
• Higher liquidus temperatures required for lead-free solders may have a profound effect on the
connector insulator housings. Some materials currently used may deform (melt) or show signs
of delamination on the side walls, referred to as "blistering."
Delamination, Leadless Chip Carrier (LLC) Voiding, in Gullwing Lead to Pad Connections
• Tombstoning is a fault where one end of the component termination is completely lifted off the
land (pad).
• This can be caused by component placement error or a “disturbed joint” and results in the
component orientation being other than intended.
X-Ray Inspection is becoming a manufacturing process requirement, for the following reasons:
Good BGA X-Ray Image Example Faulty BGA X-Ray Image Example
An overview with a large field-of-view showing An overview with a large field-of-view showing gross
apparently good ball alignment and consistent defects with bridging and inconsistent ball sizes.
information and size.
Using the proper Kv level is the key to detecting and Over penetration (Kv too high) will result in the
measuring voids. The setting should be high enough outside diameter of the ball to erode. It is important
to detect a change in grayscale level without over to note that the system is capable of distinguishing
penetrating the balls/objects. greater levels of grayscale than the human eye.
The “inner ring” is the result of greater density in the By rotating the sample an operator can determine
sphere and the underlying pad. The outside edges the location of the voids (towards the top, middle or
will start to “wash out” as more Kv is used. pad area of the ball). Voids in the top or pad area
are more likely to cause cracks or poor joints.
X-Ray Magnification
Using a larger field-of-view (lower magnification) it is Using higher magnification an operator can now see
easier to inspect the whole component for gross more clearly defect details. In this example, bridging
defects. In this rotated view an operator can see (shorted) solder joints are shown.
defects in the center region of the component.
Even edges surrounding the boundary of a BGA Rough edges (as seen in the image below) indicate
indicate good reflow of the paste to the solder balls that solder paste did not fully reflow.
(as seen on the image to the below).
The primary criteria in BGA voids is the percentage of voids within each ball (IPC states is a maximum
of 20%). However, the location of the voids can be equally important. A large void on the pad area or
component attach area can lead to a fractured solder joint.
A flat view (below) is used to measuring void A rotated view (below) is used to show the location
percentage. of the voids within the solder joint.
Even edges surrounding the boundary of a BGA Rough edges indicate that solder paste did not fully
indicate good reflow of the paste to the solder balls. reflow.
The solder joints in both images below show good collapse and complete wetting of the joint to the
pad. To achieve this type of image the board is tilted and rotated 45 degrees or mounted in a fixture
at an angle and rotated 45 degrees. The joint-to-pad area (meniscus) will appear as a flat spot on the
ball with a slightly lighter grayscale then the rest of the ball.
The image (above) clearly shows the paste has not The image on the right shows the beginning of ball
yet interacted with the solder ball (a large void in a and paste interaction but the column joints and
ball can also be seen). some opens confirm reflow is still incomplete.
The solder joints in both of the images below show opens due to incomplete reflow. On a couple of
the balls a gap can actually be seen but in most cases the rounded shape of the ball at the pad and
exposed pad area are key indicators of possible opens.
Column shaped solder joints are shown in both images below. Column joints can be the result of
either incomplete reflow (paste and ball interaction has started but collapse is not complete) or curling
of the component corners. Curling corners are typically due to moisture in the component due to
storage or lack of “baking” the components.
The image on the left shows good solder joints or The image (below) displays missing solder
‘bumps” on this package. Look for even solder over deposition from several pads. Subtle faults could
the entire pad area. easily be overlooked at low magnification when
inspecting CSP components.
Both of these images are taken from the same The image (above) shows the right side of the
package. The image (above) shows the left side of package has opens in the column joints and
the package is lifted and some of the bumps are collapsed joints. (viewing from left-to-right) Uneven
open, not touching the pads. heating or disturbing the component during reflow
could cause these results.
Gullwing Bridging – Top Down View Gullwing Bridging – Off Axis View
A flat viewing perspective of the Quad Flat Package By rotating the QFP an operator can see the bridge
(above) indicates a bridge between two leads. from another angle and verify that this is not a
bottom side component.
Open & Bent Peripheral Leads – QFP Gullwing Bent Peripheral Lead – QFP Gullwing
The image (above) shows five gullwing leads that The image (above) shows Coplanarity, where one
have no solder and two leads that are bent inward lead or series of leads on a component is out of
toward the component body. alignment and fails to make contact with the land.
The image (above) shows good J-lead solder joints The image (above) shows fractured (open) solder
with good solder fill on both toe and heel. joints. This could be the result of contamination on
the leads preventing proper reflow “wetting” or a
“disturbed joint”.