MRI Basics
MRI Basics
MRI Basics
Initial concepts MRI basics include formation of magnetic fields from electric currents in loops
of wire, the resonance phenomenon,the hydrogen proton and its frequency of precession, and
absorption of radiofrequency energy. These concepts can then be applied to learn about T1 and
T2 relaxation and contrast and how the acquisition parameters of echo time and repetition time
can be used to achieve these image contrasts. Basic pulse sequences include the spinecho,
multiecho spin-echo, turbo spin-echo, inversion-recovery, and gradient-recalled-echo sequences
Artifacts in magnetic resonance (MR) imaging result from the complex interaction of
contemporary imager subsystems, including the mainmagnet, gradient coils, radiofrequency (RF)
transmitter and receiver,and reconstruction algorithm used.We no longer look to MR imaging
only structural information,but also functional information of various kinds information about
blood flow,cardiac function,biochemical processes,tumor kinetics, and blood oxygen levels for
mapping of brain function. We have gone from the single receiver to the quadrature coil, to
multielement fast receivers and the RF chain has been completely revamped with phased-array
coil imaging and parallel imaging. The high magnetic field strengths and the highperformance
gradients have brought a new awareness of the issue of safety for both clinicians and patients.
The effects of exposure to magnetic fields and the compatibility of the many so-called MR-
compatible or MR-safe surgical implants and other tools are being reinvestigated at high field
strengths. The use of rapid imaging techniques such as echo-planar imaging and half-Fourier
rapid acquisition with relaxation enhancement, or RARE, along with the use of ultra-high-speed
gradients, has raised awareness and concern for the biologic effects of exposure. The main
subsystems are the magnet, gradient coil, RF generator, and computer, the last of which controls
the interplay between subsystems and the reconstruction, storage, and display of the images.
There are many sources of artifacts on MR images. These could broadly be classified as image
reconstruction–related, system-related, and physiology-related sources. Typically, image
reconstruction–related artifacts occur because of limitations intrinsicto the reconstruction
algorithm used by particular vendor. Filling of k-space is accomplished by acquiring frequency
encoded data samples for a given phase encoding step. Sampling is much quicker in the
frequency encoding direction than in the phase encoding direction,where the time between
adjacent samples is greater than or equal to the repetition time of the particular sequence (except
for fast SE and echoplanar imaging sequences). Low spatial frequencies are encoded in the
center of k-space and provide contrast resolution to the image, whereas high spatial frequencies
are encoded toward the edges of k-space and contribute to the spatial
resolution of the image. Fourier transform of k-space is then applied to convert the data into an
image. Each pixel in the resultant image is the weighted sum of all the individual
points in k-space. Therefore, the information in each pixel is derived from a fraction of every
point in k-space. On the basis of these facts, we can conclude that any disruption of k-space,
whether by motion, extraneous frequencies, or frequency spikes, has the potential to corrupt the
entire image.
Abbreviations: ADC-analog-to-digital converter, RF _ radiofrequency, TE _ echo time, TR _
repetition time
Initial Concepts
When an electron travels along a wire, a magnetic field is produced around the electron. When
an electric current flows in a wire that is formed into a loop, a large magnetic field will be
formed perpendicular to the loop.
Resonance
Hydrogen Protons
It is necessary to have a source of hydrogen protons (protons in the nuclei of hydrogen atoms,
which are associated with fat and water molecules) in order to form our MR signal. The
hydrogen proton is positively charged and spins about its axis. The hydrogen protons in our
body thus act like many tiny magnets. When the protons are placed in a strong magnetic field
(B0 ), a net magnetization will be produced parallel to the main magnetic field. This net
magnetization becomes the source of our MR signal and is used to produce MR images.
For typical 1.5-T superconducting cylindrical-bore magnets, the z direction is horizontal and
corresponds to the head-to-foot (or foot-to-head) direction. The plane perpendicular to this
direction is called the transverse plane or the x-y plane. For a patient who is headfirst and
supine in a superconducting magnet, the x direction is often chosen to be the left-right
direction of the patient and the y directionis often chosen to be the anterior-posterior
direction. Interestingly, the transverse plane matches the axial plane for typical 1.5-T magnets.
Precession
The proton precessional frequency is determined from the Larmor equation, in which the
frequency of precession, f, is equal to a constant times the main magnetic field strength .The
constant is called the gyromagnetic ratio and is a characteristic of each type of nuclei. For
hydrogen protons, the gyromagnetic ratio is equal to 42.6 MHz/T (megahertz per tesla). The
main magnetic field strength, B0, depends on the magnet design. For a typical superconducting
MR system, the magnetic field strength may be 1.5 T.The frequency of precession then will qual
2.6 MHz/T 1.5 T or about 64 MHz (64 million times per second).
Radiofrequency Energy
Radiofrequency (RF) energy comes in the form of rapidly changing magnetic and
electric fields generated by electrons traveling through loops of wire with the
direction of current flow rapidly changing back and forth at “radio frequencies.”
The magnetic field (generated by the flow of electrons) will also rapidly change
directions. For the MR system, this RF energy is transmitted by an RF transmit coil
(eg, body coil, head coil, knee coil). Typically, the RF is transmitted for a short
period of time; this is called an RF pulse. This transmitted RF pulse must be at the
precessional frequency of the protons (calculated via the Larmor equation) in order
for resonance to occur and for efficient transfer of energy from the RF coil to the
protons.
White matter has a very short T1 time and relaxes rapidly. Cerebrospinal fluid
(CSF) has a long T1 and relaxes slowly. Gray matter has an intermediate T1 and
relaxes at an intermediate rate.
TE and TR
TE has already been described as the time between the peak of the 90° RF pulse
and the peak of the echo that is formed. Note that the 180° RF pulse occurs at
half of the echo time TE. A parameter not yet discussed is the repetition time or
TR. TR is the time that it takes to run through the pulse sequence one time.
The raw data have the same number of rows and columns as the reconstructed
image. For basic pulse sequences, one time through the pulse sequence
provides one row of raw data. We must repeat the pulse sequence as many times
as necessary to provide as many rows of data as are needed to reconstruct
the image. TR is the time it takes to go through the pulse sequence one time. In
order to acquire all rows of data, it will take a time equal to TR times ...
Spin Echo
The spin-echo pulse sequence can produce proton density weighting, T1
weighting, and T2 weighting. Typical values of TE and TR for T1 weighting (at
1.5 T) are TE _ 20 msec and TR _ 500 msec; the typical values for T2 weighting
are TE _ 80 msec and TR _ 2,000 msec. The 90° RF pulse produces an initial
signal (free induction decay), which is not used. The 180° RF pulse occurs at half
the TE time, and the echo is centered at TE. The ADC (analog-to-digital converter)
line indicates that the echo is digitized and stored in the computer as raw data.
However, instead of each echo forming a different image data set, all the
echoes are used to create a single image data set at a faster rate. A new
acquisition parameter will be introduced called the echo train length, which
is the number of echoes that are formed. Recall that the echo is digitized and the
data from this echo are used for one row of raw data. Recall also that the pulse
sequence must be repeated as many times as is needed to acquire all
the rows of raw data. In the turbo spin-echo sequence, if four echoes are
produced (each time through the pulse sequence), the digitized data
from these four echoes can be used for four different rows of raw data. If 256
rows of raw data are needed, and four rows of raw data are acquired
each time through the pulse sequence, then the sequence must be repeated only
64 times rather than 256 times. With TR the same as in a spinecho
sequence, this would result in a factor of four speed increase in data acquisition.
Likewise, an echo train length of eight or 16 will decrease
imaging time by a factor of eight or 16, respectively.The turbo spin-echo pulse
sequence can be used to produce T1 and T2 contrast weighting.
Each echo will still occur at a different TE and thus will really have a different
contrast weighting associated with it. However, there is a way that we
can use the echoes closest to our TE of interest to form the contrast weighting
that we desire.
Inversion Recovery
The inversion-recovery pulse sequence is useful for suppressing unwanted signals
in MR images (eg, signals from fat or fluid). Contrast weighting can still be
controlled through selection of TR and TE, as described earlier.
The difference between this and the spin-echo pulse sequence is the occurrence
of the 180° RF pulse prior to the regular spin-echo pulse sequence.
The 180° RF pulse causes an initial inversion of the longitudinal magnetization (so
that it is aligned in the z direction). The magnetization then begins to grow
back in the direction of the main magnetic field . The magnetization of different
tissues will grow back at different rates. When the signal from
the tissue to be suppressed crosses the zero axis, application of a 90° RF pulse will
rotate all other signals into the transverse plane. Since the signal
from the tissue at the zero point is zero, there is nothing to rotate into the
transverse plane. Thus, this tissue will not contribute any brightness to
the resulting image. The acquisition parameter TI (time of inversion)
is the time between the initial 180° RF pulse and the 90° RF pulse. Fat relaxes
relatively quickly, and a short TI of approximately 170 msec is used to suppress
signal from fat at a field strength of 1.5 T. This method can also be used
to suppress signal from other tissues that cross through the zero point by
appropriate application of TI for that tissue.
DWI
The complex structural organization of the white matter of the brain can be
depicted in vivo in great detail with advanced diffusion magnetic resonance (MR)
imaging schemes. Diffusion MR imaging techniques are increasingly varied, from
the simplest and most commonly used technique—the mapping of apparent
diffusion coefficient values— to the more complex, such as diffusion tensor
imaging, q-ball imaging, diffusion spectrum imaging, and tractography. Current
clinical applications are based on many different types of contrast, such
as contrast in relaxation times for T1- or T2- weighted MR imaging, in time of
flight for MR angiography, in blood oxygen level dependency for functional MR
imaging, and in diffusion for apparent diffusion coefficient (ADC) imaging.
The options consist of other radiofrequency pulses, gradients or variable reconstruction methods to:
Either modify the contrast (preparing magnetization by inversion-recovery, fat saturation, magnetization
transfer…)
or accelerate the sequence (partial Fourier plane filling, parallel acquisition, fast magnetization restoration…)
or to reduce artifacts (flow compensation, synchronisation, presaturation bands …)
Finally, the user must choose the sequence parameters (TR, TE, flip angle, turbo factor, field of view matrix) to find
the best compromise between contrast, spatial resolution and speed.
With the exception of inversion-recovery, the optional techniques to modify the contrast of a sequence will be
examined in a dedicated chapter. Likewise, the treatment of artifacts and parallel imaging methods will be dealt with
in separate chapters.
Sequence classification
There are two main sequence families, depending on the type of echo recorded:
Numerous variations have been developed within each of these families, mainly to increase acquisition speed:
Some sequences are hybrid, mixing spin echo and gradient echo (GRASE, SE-EPI).
Magnetic resonance angiography sequences (FBI, contrast-enhanced MRA, TOF, PC) perfusion imaging, diffusion
imaging (DW) and MR spectroscopy will be dealt with in separate chapters in the second part.
Sequences acronyms
Due to manufacturers each using their own terminology to denominate their sequences, there are no standard
denominations for each common type of sequence.
Here is a table of the equivalent manufacturers’ acronyms with the corresponding type of sequence.
MS
IR IR IR/IRM IR IR IR
FAME QUICK 3D
LAVA
FFE
TFE-EPI EPIFI
Spin echo
TR = Repetition time
NPy = Number of phase encoding steps
Nex = Number of excitations
Contrast
A spin echo sequence has two essential parameters: TR and TE.
TR is the time interval between two successive 90° RF waves. It conditions the longitudinal relaxation of the explored
tissues (depending on T1). The longer the TR, the more complete the longitudinal magnetization regrowth (Mz tends
to M0). Reducing TR will weight the image in T1 as the differences between the longitudinal relaxation of the tissues’
magnetization will be highlighted . In classic spin echo, after TR time, a single k-space line will be acquired. TR
repetition is thus responsible for the duration of the sequence.
TE is the time interval between the 90° flip and receipt of the echo, the signal being produced by transverse
magnetization. Transverse magnetization decreases according to the time constant T2 of each tissue (the field
heterogeneities [which give T2*] being compensated by the 180° flip applied at TE/2).
In the T2–weighted spin echo sequence the TR and TE parameters are optimized to reflect T2 relaxation.
When the TR is long (over 2000 milliseconds), longitudinal magnetization recovery is complete and on the following
flip, the influence of T1 on signal magnitude will be minimized. Associated with long TE (80 to 140 milliseconds), the
different tissues are better highlighted according to their T2.
Long T2 tissues will appear as a hypersignal, as opposed to short T2 structures, which will appear as a hyposignal.
The proton density weighted spin echo sequence has optimized TR and TE parameters to minimize the influence of
both T2 and T1. The contrast obtained will depend on the density of the hydrogen nuclei (i.e. protons).
A long TR (over 2000 milliseconds), associated with a short TE (10 to 20 milliseconds) will relatively suppress both
the influence of T1 and the effect of T2 on signal magnitude.
Interest
Historically, spin echo was the first sequence to be used. It has been a benchmark for all subsequent developments,
namely in terms of contrast. The 180° rephasing pulse gives a « true T2 » signal rather than a T2*signal.
Choosing the right sequence parameters (TR and TE) will produce images weighted in T1, T2 or proton density. The
major disadvantage with T2 weighted spin echo sequences is linked to long TR resulting in prohibitive acquisition
times.
While spin echo sequences can be used in clinical practice to obtain good quality anatomical T1-weighted images,
faster types of sequence are preferred to obtain T2-weighted images.
Multi-echo SE sequences
These sequences allow several images of the same slice position without increasing overall acquisition time. The
advantage is that the images are obtained with a different contrast, which is useful in characterizing certain lesions
(for example, highlighting contrast at long TE for hepatic angioma, which appears as a relative hypersignal).
After the first echo is obtained, there is a free interval until the next TR. By applying a new 180° pulse, a new echo is
received, with the same phase encoding, to build the second image . The echo time of the 2 images differs and the
second image will be more T2 weighted than the first.
Typically, these sequences are used to obtain simultaneously PD- and T2-weighted images.
Generic diagram
The echo train technique can be pushed to the limit to fill the entire Fourier plane with a single 90° pulse (TR is thus
infinite) . These so-called « single-shot » sequences require the successive application of as many 180° pulses as
there are k-space lines to fill.
The sequence can be further accelerated, avoiding the need to register the latest echoes (whose signal is much
reduced by T2 relaxation) by partial k-space acquisition. Just over half the k-space lines are actually acquired and the
missing lines are calculated using k-space symmetry properties. This reduces acquisition time by a factor close to 2,
but to the detriment of the signal to noise ratio of the image.
IR IR IR/IRM IR IR IR
IR TSE TurboIR/TIRM FSE-IR FIR Fast IR
Inversion-recovery is a magnetization preparation technique followed by an imaging sequence of the spin echo
type in its « standard » version .
The sequence starts with a 180° RF inversion wave which flips longitudinal magnetization Mz in the opposite
direction (negative). Due to longitudinal relaxation, longitudinal magnetization will increase to return to its
initial value, passing through null value.
To measure the signal, a 90° RF wave is applied to obtain transverse magnetization. The delay between the
180° RF inversion wave and the 90° RF excitation wave is referred to as the inversion time TI.
As longitudinal regrowth speed is characterized by relaxation time T1, these sequences are weighted in T1.
Inversion-recovery also increases weighting of the associated imaging sequence (spin echo or gradient echo of
varying speeds).
With this type of sequence, certain tissues have a negative signal. In terms of display, two possibilities exist :
Either signal magnitude (amplitude in relation to 0) used for gray scale display: the more absolute the value of the tissue
signal (positive or négative), the stronger it will be.
Or the gray levels will be distributed from the negative signal values to the positive values (with a null signal background
that will be gray rather than black): this is the « true » display type.
Another property of inversion-recovery sequences is linked to the choice of TI: if a TI is chosen such that the
longitudinal magnetization of a tissue is null, the latter cannot emit a signal (absence of transverse
magnetization due to the absence of longitudinal magnetization). The inversion-recovery technique thus allows
the signal of a given tissue to be suppressed by selecting a TI adapted to the T1 of this tissue .
Inversion-recovery can be combined with sequence types other than the standard spin echo. In particular, it can
be used with fast spin echo sequences, to save considerable time, as inversion-recovery requires relatively long
TR to allow magnetization the time to regrow. Iinversion-recovery also serves as magnetization preparation for
gradient echo sequences, to weight them in T1.
STIR sequences
In the standard STIR sequence, the spin echo sequence is completed by a previous 180° inversion pulse. Fat
has a short T1. Hence by choosing a short TI of 140 milliseconds, the fat signal can be suppressed . The
combination of short TI inversion-recovery and fast spin echo sequences reduces acquisition time to acceptable
limits for clinical practice.
The advantage of these sequences is that they offer a fat signal suppression technique with low sensitivity to
magnetic field heterogeneities or to the effects of magnetic susceptibility in the presence of metal (orthopedic
prostheses in osteoarticular imaging for example). They can be used with T1 or T2 weighting (particularly in spin
echo sequences where fat appears as a hypersignal).
This technique must not be used to suppress a fat signal gadolinium injection because gadolinium–enhanced
tissues have a shortened T1 and may be erased by short TI inversion-recovery (which is not specific to tissue but
to its relaxation time T1).
FLAIR sequences
The aim of a FLAIR sequence is to suppress liquid signals by inversion-recovery at an adapted TI. Water has a
long T1. Nulling of the water signal is seen at TI of 2000 milliseconds. . As in the case of the other inversion-
recovery sequences, an imaging sequence of the fast spin echo type is preferable to compensate the long
acquisition time linked to long TR.
These sequences are routinely used in cerebral MRI for edema imaging.
Gradient echo
Type of sequence Philips Siemens GE Hitachi Toshiba
Characteristics
The gradient echo sequence differs from the spin echo sequence in regard to:
A flip angle lower than 90° (partial flip angle) decreases the amount of magnetization tipped into the transverse plane.
The consequence of a low-flip angle excitation is a faster recovery of longitudinal magnetization that allows shorter
TR/TE and decreases scan time.
The advantages of low-flip angle excitations and gradient echo techniques are faster acquisitions, new contrasts
between tissues and a stronger MR signal in case of short TR.
The flip angle determines the fraction of magnetization tipped in the transverse plane (which will produce the NMR
signal) and the quantity of magnetization left on the longitudinal axis.
If the flip angle decreases, the residual longitudinal magnetization will be higher and the recovery of magnetization for
a given T1 and TR will be more complete. On the other hand, the result of a lower flip angle excitation is a lower
tipped magnetization.
As GE techniques use a single RF pulse and no 180° rephasing pulse, the relaxation due to fixed causes is not
reversed and the loss of signal results from T2* effects (pure T2 + static field inhomogeneities). The signal obtained is
thus T2*-weighted rather than T2-weighted. These sequences are thus more sensitive to magnetic susceptibility
artifacts than are spin echo sequences.
Gradient echo
As there is no 180° RF pulse, a bipolar readout gradient (which is the same as the frequency-encoding gradient) is
required to create an echo. The gradient echo formation results from applying a dephasing gradient before the
frequency-encoding or readout gradient.
The goal of this dephasing gradient is to obtain an echo when the readout gradient is applied and the data are
acquired. The dephasing stage of the readout gradient is in the inverse sign of the readout gradient during data
acquisition. Moreover, its dephasing effect is designed so that it corresponds to half of the dephasing effect of the
readout gradient during data acquisition. Consequently, during data acquisition, the readout gradient will rephase the
spins in the first half of the readout (by reversing the dephasing effect of the dephasing lobe), and the spins will
dephase in the second half (due to the dephasing effect of the readout gradient). The time during which the peak
signal is obtained is called Echo Time (TE).
Steady state
In gradient echo, TR reduction may cause permanent residual transverse magnetization in TR below T2: the
transverse magnetization will not have completely disappeared at the onset of the following repetition and will also be
submitted to the flip caused by the excitation pulse.
Two main classes of gradient echo sequence can be distinguished, depending on how residual transverse
magnetiztion is managed:
Ferromagnetic materials generally contain iron, nickel, or cobalt. These materials include magnets, and various objects
one might find in a patient, such as aneurysm clips, parts of pacemakers, shrapnel, etc.
These materials have a large positive magnetic susceptibility, i.e., when placed in a magnet field, the field strength is much
stronger inside the material than outside. Ferromagnetic materials are also characterized by being made up of clusters of
10^17 to 10^21 atoms called magnetic domains, that all have their magnetic moments pointing in the same direction. The
moments of the domains is random in unmagnetized materials, and point in the same direction in magnetized materials.
Figure 1 illustrates the effect of a ferromagnetic material (grey circle) on the magnetic field flux lines (blue).
The ability to remain magnetized when an external magnetic field is removed is a distinguishing factor compared to
paramagnetic, superparamagnetic, and diamagnetic materials.
On MR images, these materials cause susceptibility artifacts characterized by loss of signal and spatial distortion. This can
occur with even fragments to small to be seen on plain x-ray. This is a common finding in a cervical spine MRI post anterior
fusion.
Spoiled gradient echo MRI is an MRI technique which destroys residual transverse magnetization at the end of each
excitation cycle.
Paramagnetic materials include oxygen and ions of various metals like Fe, Mg, and Gd. These ions have unpaired
electrons, resulting in a positive magnetic susceptibility. The magnitude of this susceptibility is less than one one-thousands
of that of ferromagnetic materials.
The effect on MRI is an increase in the T1 and T2 relaxation rates (decrease in the T1 and T2 times). The figure illustrates
the effect of a paramagnetic material (grey circle) on the magnetic field flux lines (blue).
Gd is used as a in MR contrast agents. At the proper concentration, Gd contrast agents cause preferential T1 relaxation
enhancement, causing increase in signal on T1-weighted images. At high concentrations, as is sometimes seen in the
urinary bladder, loss of signal is seen instead, a result of the T2 relaxation effects dominating.
Superparamagnetic materials consist of individual domains of elements that have ferromagnetic properties in bulk. Their
magnetic susceptibility is between that of ferromagnetic and paramagnetic materials.
The figure illustrates the effect of a superparamagnetic material (grey circle) on the magnetic field flux lines (blue). Examples
of superparamagnetic materials include iron containing contrast agents for bowel, liver, and lymph node imaging.
Diamagnetism is the property of materials that have no intrinsic atomic magnetic moment, but when placed in a magnetic
field weakly repel the field, resulting in a small negative magnetic susceptibility. Materials like water, copper, nitrogen,
barium sulfate, and most tissues are diamagnetic.
The figure illustrates the effect of a diamagnetic material (grey circle) on the magnetic field flux lines (blue). The weak
negative magnetic susceptibility contributes to the loss of signal seen in bowel on MRI after administration of barium sulfate
suspensions.
The Larmor frequency and equation are named after the Irish Physicist and MathematicianJoseph Larmor (1857-1942).
Protons and neutrons pair up in nuclei causing the cancellation of their individual angular momentum. All nuclei also have a
spin; those with an odd number of protons and/or neutrons will have a property call magnetic (dipole) moment. Magnetic
moment is characterized by it's alignment with an external magnetic field analogous to a small bar magnet. These particular
nuclei are also called dipoles because they have two poles like the north and south pole of a bar magnet. H-1 and P-31 are
examples of nuclei with an unpaired proton. N-14 is an example of a nucleus with both unpaired proton and neutron.
In addition to alignment of nuclei with a magnetic moment, application of an external magnetic field will produce a secondary
spin or wobble (precession) of nuclei around the main or static magnetic field. The precessional path around the magnetic
field is circular like a spinning top.
The Larmor or precessional frequency in MRI refers to the rate of precession of the magnetic moment of the proton
around the external magnetic field. The frequency of precession is related to the strength of the magnetic field, B0.
The precessional frequency of nuclei of a substance placed in a static magnetic field B0 is calculated from Larmor
Equation :
ω = γB
where ω is the Larmor frequency in MHz , γ is the gyromagnetic ratio in MHz/Tesla and B is the strength of the static
magnetic field in Tesla. Note that the gyromagnetic ratio is define in different ways by different authors. See the article
on gyromagnetic ratio. In this case a useful, simplified version is shown representing the Larmor frequency when B 0=1.
The gyromagnetic ratio (MHz/T) for a few commonly measured or imaged isotopes are 1:
o H-1 42.58
o F-19 40.05
o Na-23 11.26
o P-31 17.24
Magnets
Dr Jeremy Jones and Dr J. Ray Ballinger et al.
Magnets used for MRI are of three types: permanent, resistive and superconductive.
Permanent MRI magnets use permanently magnetized iron like a large bar magnet that has been twisted into a "C" shape
where the two poles are close together and parallel. In the space between the poles the magnetic field is uniform enough for
imaging. Up to 30 tons of iron may be needed, restricting their placement to rooms with a strong enough floor. Their low-field
strength of about 0.15-4 T restrict their use to imaging; being impractical for spectroscopy, chemical shift and susceptibility
imaging such as function brain imaging. Their magnetic field homogeneity is also sensitive to ambient temperature so room
temperature must be controlled carefully. The initial purchase price and operating costs are low compared to
superconductive magnets. These magnets can also be made with alloys such as neodymium markedly reducing the weight
of the magnet but at significant additional cost.
Resistive (air core) MRI magnets operate at room temperature using standard conductors such as copper in the shape of
a solenoid or Helmholtz pair coil. A solenoid is a cylindrical shape coil of wire. The uniform magnetic field is found inside the
coil, especially in the center. These magnets are relatively inexpensive to make but require a large constant flow of current
while magnetized and imaging. The coil has electrical resistance that requires cooling of the magnet. The operating costs
are high because of the large power requirements of the magnetic coils and associated cooling system.
Both permanent and resistive MRI scanners are limited to low-field applications, primarily open MRI and extremity scanners.
These magnets are useful for claustrophobic patients.
Superconductive MRI magnets use a solenoid shaped coil made of alloys such as niobium/titanium or niobium/tin
surrounded by copper. These alloys have the property of zero resistance to electrical current when cooled down to about
10o Kelvin. The coil is kept below this temperature with liquid He. The power supply is connected on either side of a short
heated segment of the coil and the current to the coil is gradually increased over several hours until the desired magnet field
is reached. The heated segment is allowed to coil to superconducting temperature and the power supply removed and taken
away. The current continues in the closed loop of the coil for years without significant decline. A resulting property is that the
magnetic field is always present.
The surrounding copper acts as an insulator at low temperatures compare to the zero resistance of the alloy. The copper
also protects the alloy coil from being destroyed in case of a quench of the magnet. A quench can occur if the helium levels
drop too low or if a large ferromagnetic object is brought into the fringe field of the magnet. A quench results in loss of
superconductivity with a large amount of heat produced by the current and rapid boiling off of cryogens. The gas produced is
vented out of the room but can occasionally enter the scanner room with life-threatening consequences. Quenches and the
constant magnetic field are a couple of the safety issues that are discussed elsewhere.
The cost of cryogen replacement is reduced on modern magnets that incorporate a refrigeration system called a "cold head"
to condense the cryogen gas. Startup cost for the scanner can run up to about $1.5 millon for a 1.5 T MRI. Site preparation
can frequently run several $100,000's that includes room RF shielding, possible magnetic shielding, floor reinforcement,
vibration mitigation and adequate power supply.
Superconducting magnets at 1.5 T and above allow functional brain imaging, MR spectroscopy and superior SNR and/or
improved time and spatial resolution. Magnets above 1.5 T have addition challenges from RF heating of the subject, and
increased artifacts from susceptibility and RF penetration among others.
Quenching refers to rapid expulsion of liquid cryogen used to maintain the MR magnet in superconducting state.
Discussion
Modern MRI scanners contain superconducting magnets which have very low energy consumption, made possible by
maintaining internal subzero temperatures by way of a 'cryogen bath' of liquid helium. In the event that somebody pushes
the emergency stop button or there is equipment fault, the liquid helium boils off rapidly accompanied by a lound banging or
hissing sound, safely expelled from the building by means of vents.
RF coils
RF (radiofrequency) coils are the "antenna" of the MRI system that broadcasts the RF signal to the patient and/or receives
the return signal. RF coils can transmit and receive (transceiver) or receive-only, in which case the body coil is used as a
transmitter. The body coil supplies a more uniform RF field than many smaller coils but requires more energy to do so; a
detriment at higher field strengths (3T+).
Surface coils are the simplest design of coil. They are simply a loop of wire, either circular or rectangular, that is placed over
the region of interest (see figure). The depth of the image of a surface coil is generally limited to about one radius. Surface
coils may be used for spines, shoulders, TMJ's, and other relatively small body parts close to the skin surface.
Helmholtz pair coils consist of two circular coils parallel to each other. They are used as the z gradient coils in MRI
scanners allowing localization in the z direction (head to foot in a horizontal magnet). They are may also used occasionally
as RF coils for pelvis imaging and cervical spine imaging.
Paired saddle coils are commonly used for imaging of the extremities such as the knee. These coils provide better
homogeneity of the RF in the area of interest than a surface of Helmholtz pair and are used as volume coils, unlike surface
coils. Paired saddle coils are also used for the x and y gradient coils. By running current in opposite directions in the two
halves of the gradient coil, the magnetic field is made stronger near one and weaker near the other.
The bird cage coil provides the best RF homogeneity of all the RF coils. It has the appearance of a bird cage; hence, its
name. This coil is commonly used as a transceiver coil for imaging of the head. This type of coil is also used occasionally for
imaging of the extremities.
Gradient coils
Gradient coils are used to produce deliberate variations in the main magnetic field (Bo). There are three sets of gradient
coils, one for each direction. The variation in the magnetic field permits localization of image slices as well as phase
encoding and frequency encoding. The set of gradient coils for the z axis are Helmholtz pairs, and for the x and y axes,
paired saddle coils.
Gradient coils need to provide linear gradations of the magnetic field. When turned on, one end or side the bore of the
magnet has a lesser strength and the other a greater strength than the static main magnetic field. Drop off of the z gradient
near the bore of the magnet can cause misplacement of objects in the image if signal is detected beyond the gradients. This
should not happen with appropriate engineering.
Signal processing
Fourier transform is a mathematical operation which converts a time domain signal into a frequency domain signal.
Discussion
Fourier transform is integral to all modern imaging, and is particularly important in MRI. The signal received at the detector
(receiver coils in MRI, piezoelectric disc in ultrasound and detector array in CT) is a complex periodic signal made of a large
number of constituent frequencies (i.e., bandwidth). This can be visualized as multiple sine and or cosine waves along a
time-axis. Fourier transform represents the same data over a frequency-axis. A common example is the MR spectroscopy
image in which different molecules are at different frequencies along the x-axis.
K space is an abstract concept and refers to a data matrix containing the raw MRI data. This data is subjected to
mathematical function or formula called a transform to generate the final image. A discrete Fourier or fast Fourier
transform 1-3 is generally used though other transforms such as the Hartley 4 can also work.
Discussion
A single slice corresponds to a k space plane acquired in real-time. Each point on the k space contains specific frequency,
phase (x,y coordinates) and signal intensity information (brightness). Inverse FT is applied after k space acquisition to derive
the final image. Each pixel in the resultant image is the weighted sum of all the individual points in the k-space. Hence,
disruption of any point in the k-space translates into some form of final image distortion, determined by the frequency- and
phase-related data stored in that particular point. In general:
Relevance
Knowledge of the k space is essential as it relates to different techniques of image acquisition and explains several MRI
artifacts.
Signal to noise ratio (SNR) is a generic term, which in radiology is measure of how much true signal (e.g. reflecting actual
anatomy) versus how much noise (e.g. random quantum mottle etc...) a particular image has, which results in a grainy
appearance.
Each modality has its own source of noise and ways to maximise signal.
Radiographic interpretation
Plain film radiography
In X-rays, the larger number of photons absorbed, the greater the SNR, the less noisy the image. The use of high kV and
intensifying screens would reduce the number of photons and radiation dose to patients, therefore reducing the SNR.
Conversely, measures to increase the SNR, such as by increasing the mAs, would increase the patient dose.
CT
Content pending
MRI
The SNR is measured frequently by calculating the difference in signal intensity between the area of interest and the
background (usually chosen from the air surrounding the object). In air, any signal present should be noise. The difference
between the signal and the background noise is divided by the standard deviation of the signal from the background-- an
indication of the variability of the background noise.
SNR is proportional to the volume of the voxel and to the square root of the number or averages and phase steps (assuming
constant sized voxels). Since averaging and increasing the phase steps takes time, SNR is related closely to the acquisition
time.
Spatial resolution determines how "sharp" the image looks. Low resolution will give either fuzzy edges, or a pixelly
appearance to the image.
In MRI, spatial resolution is defined by the size of the imaging voxels. Since voxels are three dimensional rectangular solids,
the resolution is frequently different in the three different directions. The size of the voxel and therefore the resolution
depends on matrix size, the field-of-view (FOV), and the slice thickness. The matrix size is the number of frequency
encoding steps, in one direction; and the number of phase encoding steps, in the other direction of the image plane.
Assuming everything else is constant, increasing the number of frequency encodings or the number of phase steps results
in improved resolution. The frequency encoding depends of how rapidly the FID signal is sampled by the scanner.
Increasing the sampling rate results in no time penalty. Increasing the number of phase steps increases the time of the
acquisition proportionately. This is why images that have fewer phase encodings than frequency encodings, e.g., 128x256
or 192x256 will be used.
The FOV is the size of the area that the matrix of phase and frequency encoding cover. Dividing the FOV by the matrix size
gives you the in-plane voxel size; hence, increasing the FOV in either direction increases the size of the voxels and
decreases the resolution. Decreasing the FOV improves the resolution.
The depth of the voxel is determined by the slice thickness. This is almost always the largest dimension of the voxel in 2D
imaging. Therefore, the resolution perpendicular to the image plane is the poorest. This is related to the maximum strength
of the z-gradient coils as well as time restraints limiting the number of slices available. 3-D imaging utilizing phase encoding
in the z direction is capable of smaller slice thickness than 2-D imaging but carries a time penalty proportional to the number
of slices.
The time of acquisition for a conventional spin echo or gradient echo sequence is the product of the repetition time, phase
encoding steps, and number of averages (TR x phase steps x NEX). For example, with a one second TR, 128 phase steps,
and two averages we would get an acquisition time of about 1 x 128 x 2= 256 seconds or 4 minutes and 16 seconds. The
actual time will be slightly longer.
Fast spin echo and fast gradient echo sequences perform more than one phase encoding step per repetition time, resulting
in reduce acquisition time by the number of phase step per repetition time. (Needs more on other sequences)
Trade-offs exist when changing imaging parameters to obtain the best images possible. The SNR, resolution, and
acquisition time, are all interrelated. Changing one effects the others. An important job of the radiologist and MR
radiographer is to decide what factors are more important for an examination of a particular body part, patient and suspected
abnormality.
When looking at the pituitary or cranial nerves, for example, you may want to sacrifice some SNR or longer acquisition time
for improved spatial resolution. On the other hand, in a claustrophobic of patient in pain who may be moving around, you
may sacrifice both resolution and SNR for the shortest possible examination time. The table summarizes the trade offs
between SNR, resolution, time, maximum number of slices and distance covered.
Gap + - nc + nc
TE - nc nc nc -
Zero filling interpolation (ZIP) is the substitution of zeroes for unmeasured data points in order to increase the matrix size
of the new data prior to Fourier transformation of MR data. This results in pixels smaller than the actual resolution of the
image. The zero filling occurs in the periphery of k-space rather than the center where most of the signal comes from.
For example
Magnetic resonance imaging (MRI) is able to achieve its striking ability to image tissues in various ways by manipulating the
various electromagnetic fields it is able to generate. The pattern in these fields are changed is known as a pulse sequence.
There are many (hundreds - see MRI pulse sequence abbreviations) of various pulse sequences each changing the
appearances of tissue.
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
These can be grouped as follows:
Contrast agents
Contrast agents can be grouped in many different ways. One way is to consider them according to the compartments in
which they distribute.
Summary
TR: short
TE: short
fat: bright
fluid: dark
T2 weighted image (also referred to as T2WI) is one of the basic pulse sequences in MRI and demonstrates the differences
in the T2 relaxation time of tissues.
The T2WI relies upon the transverse relaxation of the net magnetisation vector (NMV). T2 weighting tend to have
long TE and TR times.
Paramagnetic contrast agents, e.g. gadolinium-containing compounds, do not have nearly as significant effect as they do
in T1WI.
Summary
TR: long
TE: long
fat: intermediate-bright
fluid: bright
Inversion recovery pulse sequences are used to give heavy T1-weighting. The basic part of an inversion recovery
sequence is a 180 degree RF pulse that inverts the magnetization followed by a 90 degree RF pulse that brings the residual
longitudinal magnetization into the x-y or transverse plane where it can be detected by an RF coil.
In imaging, the signal is usually refocused with a 180 degree pulse as in a spin echo sequence.The time between the initial
180 degree pulse and the 90 degree pulse is the inversion time (TI). A diagram of the sequence is shown to the right.
- Short tau inversion recovery (STIR ) also called short T1 inversion recovery is a fat suppression technique with an
inversion time TI = T1 ln2 where the signal of fat is zero. This equates to approximately 140 ms at 1.5 T.
To distinguish two tissue components with this technique, the T1 values must be different. FLAIRis a similar technique to
suppress water.
Inversion-recovery imaging allows homogeneous and global fat suppression and can be used with low-field-strength
magnets. However, this technique is not specific for fat, and the signal intensity of tissue with a long T1 and tissue with a
short T1 may be ambiguous.
- Fluid attenuation inversion recovery (FLAIR) is a special inversion recovery sequence with longT1 to remove the
effects of fluid from the resulting images.1
The T1 time of the FLAIR pulse sequence is adjusted to the relaxation time of the component that should be suppressed.
For fluid suppression the inversion time (long T1) is set to the zero crossing point of fluid, resulting in the signal being
'erased'.1
This type of sequence is particularly useful in the detection of subtle changes at the periphery of the hemispheres and in the
periventricular region close to CSF.
The usefulness of FLAIR sequences has been evaluated in diseases of the central nervous system such as 2-4 :
infarction
multiple sclerosis
subarachnoid haemorrhage
head injuries, and others.
- A TIRM or Turbo inversion recovery magnitude sequence is type of inversion recovery MRI pulse sequence.
It has be shown to be superior in the asseesment of osteomyelitis in bone 1 and in the asessment of head and neck
tumours 2.
Gradient echo sequences are an alternative technique to spin echo sequences,differing from it in two principle points :
utilization of gradient fields to generate transverse magnetization
flip angles of less than 90°
Compared to the spin echo and inversion recovery sequences, gradient echo sequences are more versatile. Not only is the
basic sequence varied by adding dephasing or rephasing gradients at the end of the sequence, but there is a significant
extra variable to specify in addition to the usual TR and TE. This variable is the flip or tip angle of the spins.
Flip angle
The flip angle is usually at or close to 90 degrees for a spin echo sequence but commonly varies over a range of about 10 to
80 degrees with gradient echo sequences. For the basic gradient echo sequence FLASH (figure 1) the larger tip angles give
more T1 weighting to the image and the smaller tip angle give more T2 or actually T2* weighting to the images.
Gradient echo
The gradient echo is generated by the frequency-encode gradient, except that it is used twice in succession, and in opposite
directions: it is used in reverse at first to enforce transverse dephasement of spinning protons and then right after, it is used
as a readout gradient (like in spin echo MRI) to re-align the dephased protons and hence acquire signal.
Because low flip angles are used, there is some retention of the original longitudinal magnetization as opposed to the 90°
pulse used in spin echo, which completely eliminates the longitudinal magnetization. As a result, the build up time for
longitudinal magneitzation is significantly reduced for the subsequent pulses , allowing faster image acquisition in GE.
Another important feature of GE is that the dephasement of spinning protons occurs as a result of T2* decay which is more
rapid than the T2 decay process underlying Spin Echo sequence (leading to shorter TE) and is susceptible to static field
inhomogeneities (leading to compounded influence of degraded blood products, and metal objects on the signal ).
Image characteristics
Images from other gradient echo sequences such as GRASS and FISP have less intuitive tissue contrast characteristics
than FLASH. The FLASH and SPGR sequences show better tissue contrast between white matter and grey matter in the
brain and spinal cord than GRASS or FISP and are preferred when the time of acquisition does not have to be very short.
GRASS and FISP maintain better SNR than FLASH at short TR times and are therefore preferred with breath-holding
techniques, for example.
A vector magnetization diagram of the gradient echo sequence is shown below. Note that the spins are refocused by
reversing the direction of the spins rather than flipping them over to the other side of the x-y plane as occurs with the spin
echo sequence. Gradient refocusing of the spins takes considerably less time than 180 degree RF pulse refocusing. One big
disadvantage of gradient echo sequences is the loss of signal from static magnetic field inhomogeneity. This occurs to a
lesser degree with spin echo sequences (and for a different reason). Magnetic susceptibility artifacts are therefore more
pronounced on gradient echo sequences that on spin echo sequences.
- Spoiled gradient echo MRI is an MRI technique which destroys residual transverse magnetization at the end of each
excitation cycle.
- Steady-state free precession MRI (SSFP) is a type of gradient echo MRI pulse sequence in which a steady, residual
transverse magnetization (Mxy) is maintained between successive cycles. The sequence is noted for its superiority in
dynamic / cine assessment of cardiac function.
Discussion
To understand SSFP (and all MR imaging for that matter), the first thing to bear in mind is that the magnetization vector has
two components: Mz and Mxy. Both are at 90° to eachother and excitation pulses which flip the vector result in conversion of
one component into the other. Added to this interconversion is the spontaneous regrowth and decay of M z and Mxy governed
by T1 and T2 intervals respectively. Over several sequences, a steady-state of equilibrium is achieved, with constant
magnitudes of Mz and Mxy at the beginning of the cycle.
A steady-state is achieved in tissues with a sufficiently long T2 interval, by keeping the TR shorter than T2.
Applications
cardiac imaging
fetal imaging
abdominal imaging
Fat suppression is commonly used in magnetic resonance (MR) imaging to suppress the signal from adipose tissue or
detect adipose tissue.1
Due to short relaxation times, fat has a high signal on magnetic resonance images (MRI). This high signal, easily recognized
on MRI, may be useful to characterize a lesion.2
However, small amounts of lipids are more difficult to detect on conventional MRI. In addition, the high signal due to fat may
be responsible for artifacts such as ghosting and chemical shift. The high signal can also mask subtle contrast difference in
non-fatty tissue by filling the dynamic range of the receiver with mostly fat signal. Lastly, a contrast enhancing tumor may be
hidden by the surrounding fat. These problems have prompted development of fat suppression techniques in MRI. 3
Fat may be suppressed on the basis of its difference in resonance frequency with water by means of frequency selective
pulses (CHESS) or phase contrast techniques (by same mechanism as black boundary or india ink artifacts), or on the basis
of its short T1 relaxation time by means of inversion recovery sequences (STIR technique) or the Dixon method. 5 Lastly,
hybrid techniques combining several of these fat suppression techniques such as SPIR (spectral presaturation with
inversion recovery) are also possible.3,6
Selection of a fat suppression technique should depend on the purpose of the fat suppression (contrast enhancement vs
tissue characterization) and the amount of fat in the tissue being studied, the field strength of the magnet and the
homogeneity of the main magnetic field.
Diffusion weighted imaging (DWI) is a form of MR imaging based upon the diffusion of water molecules within a voxel.
Diffusion weighted imaging is based upon the random Brownian motion of water molecules within the voxel. As opposed to
free diffusion of water kept inside a container, diffusion of water inside a voxel of brain tissue, for example, is hindered
primarily by cell membrane boundaries. The greater the cellularity, the greater the diffusion restriction, e.g. tumors.
Clinical application
DW imaging has a major role in the following clinical situations 3-5:
early identification of ischemic stroke
differentiation of acute from chronic stroke
differentiation of acute stroke from other stroke mimics
differentiation of epidermoid cyst from arachnoid cyst
differentiation of abscess from necrotic tumors
assessment of cortical lesions in CJD
differentiation of herpes encephalitis from diffuse temporal gliomas
assessment of the extent of diffuse axonal injury
grading of gliomas and meningiomas (need further study)
assessment of active MS plaque (old plaques will not be bright)
Physics
Figure 1 depicts a spin echo sequence with diffusion gradients added. The gradient coil used to produce the diffusion need
not be a separate gradient or gradients from those used for spatial encoding. The degree of diffusion weighting is dependent
primarily on the area under the diffusion gradients and on the interval between the gradients. Other factors include the effect
of the spatial localization gradients and the size of the voxels. Diffusion tensor imaging is an extension of diffusion
weighted imaging (DWI) which allows data profiling based upon white matter tract orientation
DWI is based on the measurement of brownian motion of water molecules. This motion is restricted by membranous
boundaries. In white matter, diffusion follows the 'pathway of least resistance' along the white matter tract; this direction of
maximum diffusivity along the white-matter fibers is projected into the final image.
Saturation recovery (SR) sequences are rarely used for imaging now. Their primary use at this time is as a technique to
measure T1 times more quickly than an inversion recovery pulse sequence. Saturation recovery sequences consist of
multiple 90 degree RF pulses at relatively short repetition times (TR). An example of a SR sequence is shown below.
Residual longitudinal magnetization after the first 90 degree RF pulse is dephased by a spoiling gradient (in this case with
the slice select gradient). Longitudinal magnetization that develops during the TR period after the dephasing gradient is
rotated into the transverse plane by another 90 degree pulse. A gradient echo is acquired immediately after this. The signal
will reflect T1 differences in tissues because of different amounts of longitudinal recovery during the TR period.
Echo planar imaging is performed using a pulse sequence in which multiple echoes of different phase steps are acquired
using rephasing gradients instead of repeated 180o RF pulses following the 90° / 180° in a spin-echo sequence. This is
accomplished by rapidly reversing the readout or frequency- encoding gradient. This switching or reversal may also be done
in a sinusoidal fashion. Echo planar sequences may use entirely gradient echos or may combine a spin echo with the train
of gradient echos as illustrated in the diagram to the right.
In a single-shot echo planar sequence, the entire range of phase encoding steps, usually up to 128, are acquired in one
TR. In multi-shot echo-planar imaging, the range of phase steps is equally divided into several "shots" or TR periods. For
example an image with 256 phase steps could be divided into 4 shots of 64 steps each.
As a result an image can be acquired in 20 - 100 msec, allowing excellent temporal resolutionsuch as that required in
cardiac imaging. Each subsequent echo results in a progressively T2-weighted signal.
Benefits
reduced imaging time
decreased motion artifact,
ability to image rapid physiologic processes of the human body.
Drawbacks
Sensitive to susceptibility effects
Sensitive to main magnetic field inhomogeneity
Long gradient echo train causes greater T2* weighting
Requires high-performance gradients
Applications
cardiac imaging
abdominal imaging, i.e., breath-hold sequences and 3D MR angiography
diffusion imaging
perfusion imaging
functional imaging
Metal artifact reduction sequence
Dr Jeremy Jones and Dr Jan Frank Gerstenmaier et al.
A metal artifact reduction sequence (MARS) is intended to reduce the size and intensity of susceptibility artifacts resulting
from magnetic field distortion.
A number of simple changes to the scan protocol can greatly reduce artifacts. Examples are:
STIR for fat suppression (spectral fat suppression performs better in a homogeneous field)
spin echo instead of gradient echo where possible
shorter echo spacing
smaller water-fat sift
thinner slices
maintain good SNR
Spiral scanning in MRI is unlike spiral scanning in CT where the x-ray tube is continuously rotating and data is continuously
being acquired. In MRI the word "spiral" refers to the pattern of sampling k-space. In conventional imaging sequences
including spin echo and gradient echo and in fast imaging sequences, a line or multiple lines of k-space in the frequency
direction are acquired consecutively. In spiral scanning, k-space is acquired in a spiral trajectory. The entire k-space can be
acquired during a single acquisition, or interleaved using more than one acquisition.
This sequence allows faster image acquisition than the fast echo sequences but is slower thanecho-planar imaging. Spiral
scanning tends to have fewer artifacts than echo-planar imaging since adjacent points in k-space are acquired in close
temporal proximity. The figures to the right show how the acquisition of data in k-space is done with conventional sequences
and with spiral scanning.
MR angiography
Dr Jeremy Jones and Dr Usman Bashir et al.
MR Angiography is an alternative to conventional angiography and CT angiography, eliminating the need for iodinated
contrast media and ionizing radiation. It has evolved into several techniques with different advantages and applications:
contrast enhanced MR angiography
non-contrast enhanced MR angiography
Contrast enhanced MR angiography is a technique involving 3D spoiled gradient-echo (GE) sequences, with
administration of Gd-based contrast. It can be utilised to assess vascular structures of almost any part of the body. It's key
features are as follows:
T1- weighted spoiled gradient-echo sequence (flip angle 25° - 50° allows T1-weighting)
central k-space acquisition corresponding to arterial phase of the study maximizes preferential visualization of
arteries
use of Gd-based contrast to shorten T1-interval of the blood which appears bright as a result
Non contrast enhanced MR angiography is performed in several ways including:
time of flight angiography
phase contrast angiography
three-dimensional (3D) electrocardiograph-triggered half-Fourier fast spin echo
Generally, these techniques are time-consuming as compared with contrast enhanced MR angiography.
With 3-D TOF, a volume of images is obtained simultaneously by phase-encoding in the slice-select direction. An
angiographic appearance can be generated using MIP, as is done with 2-D TOF. Several 3-D TOF volumes can be
combined to visualize longer segments of vessels. 3-D TOF angiography will allow greater spatial resolution in the slice-
select direction than 2-D TOF; however, with thick volumes and slow flowing blood, loss of signal is seen with the 3-D TOF
method.
Key points
short TR
image-plane kept perpendicular to flow-direction
Potential pitfalls
slow flow or flow from a vessel parallel to the scan-plane, may become de-saturated just like stationary tissue,
resulting in signal loss from the vessel
turbulent flow may undergo spin-dephasing and unexpectedly short T2 relaxation - again resulting in signal loss from
the vessel
acquisition times are relatively long.
retrograde arterial flow may be obscured if venous saturation bands have been applied.
artifacts : ghosting, susceptibility artifact
very T1 bright signal will be visible (e.g. haemorrhage)
Phase contrast imaging is an MRI technique that can be used to visualise moving fluid. It is typically used for MR
venography as a non-IV-contrast requiring technique.
Spins that are moving in the same direction as a magnetic field gradient develop a phase shift that is proportional to the
velocity of the spins. This is the basis of phase-contrast angiography. In the simplest phase-contrast pulse sequence, bipolar
gradients (two gradients with equal magnitude but opposite direction) are used to encode the velocity of the spins.
Stationary spins undergo no net change in phase after the two gradients are applied. Moving spins will experience a
different magnitude of the second gradient compared to the first, because of its different spatial position. This results in a net
phase shift. This information can be used directly to determine the velocity of the spins. Alternatively, the image can be
subtracted from one acquired without the velocity encoding gradients to obtain an angiogram.
+ TRICKS
MR spectroscopy
Dr Maxime St-Amant and Dr Frank Gaillard et al.
When we see spectra in general radiology practice, this is usually of protons, although phosphorus can also be targeted to
examine ATP.
If raw signal was processed then the spectra would be dominated by water, which would make all other spectra invisible.
Water suppression is therefore part of any MRS sequence, either via Inversion Recovery or Chemical shift selective
(CHESS). If water suppression is not successful then a general slope to the base line can be demonstrated, changing the
relative heights of peaks.
Magnetic resonance spectroscopy (MRS) is performed with a variety of pulse sequences. The simplest sequence consists of
a 90 degree RF pulse without any gradients with reception of the signal by the RF coil immediately after the single RF pulse.
Many sequences used for imaging can be used for spectroscopy also (such as the spin echo sequence). The important
difference between an imaging sequence and a spectroscopy sequence is that for spectroscopy, a read out gradient is not
used during the time the RF coil is receiving the signal from the person or object being examined. Instead of using the
frequency information (provided by the read out or frequency gradient) to provide spatial or positional information, the
frequency information is used to identify different chemical compounds. This is possible because the electron cloud
surrounding different chemical compounds shields the resonant atoms of spectroscopic interest to varying degrees
depending on the specific compound and the specific position in the compound. This electron shielding causes the observed
resonance frequency of the atoms to slightly different and therefore identifiable with MRS.
History
Magnetic resonance spectroscopy (MRS) of intact biological tissues was first reported by two groups: Moon and
Richards using P-31 MRS to examine intact red blood cells in 1973, and Hoult et al. using P-31 MRS to examine excised leg
muscle from the rat in 1974.
Peaks
[Editor: These should all gradually migrate to their own articles]
N-acetylaspartate (NAA)
resonates at 2.0 ppm chemical shift
NAA is an acetylated amino acid which is found in high concentrations in neurons and is a marker of neuronal viability. It is
therefore reduced in any process that destroys neurons.
Radiation effects
Distinguishing radiation change and tumour recurrence can be problematic. In recurrent tumour choline will be elevated,
whereas in radiation change, NAA, choline and creatine will all be low.
Infection
As in all processes which destroy normal brain tissue, NAA is absent. Within bacterial abscess cavities, lactate, alanine,
cytosolic acid and acetate are elevated / present.
Of note choline is low or absent in toxoplasmosis, whereas it is elevated in lymphoma, helping to distinguish the two.
White matter diseases
progressive multifocal leukoencephalopathy (PML) may demonstrate elevated myo-inositol.
Canavan disease characteristically demonstrates elevated NAA.
Hepatic encephalopathy
Markedly reduced myo-inositol, and to a lesser degree choline. Glutamine is increased.
Mitochondrial disorders
Leigh syndrome : elevated choline, reduced NAA and occasionally elevated lactate.
Mnemonic
My ChoCrNaaLa (think of a new chocolate energy bar or something)
My : Myoinisitol 3.3
Cho : Choline 3.2
Cr : Creatine 3.0
Naa : Naa 2.0
L : Lactate 1.3
Prostate MRS
In prostate MRS, a citrate peak is looked for at 2.6 ppm. For more information go to: MR spectroscopy in prostate cancer
+ Hunter's angle
Dr Jeremy Jones and Dr Umamaheswara Reddy V et al.
Hunter's angle is a term coined from a neurosurgeon, Hunter Sheldon, at Huntington Medical Research Institutes. He
placed his comb on the spectrum at approximately a 45š angle and connected several of the peaks. If the angle and peaks
roughly corresponded to the 45š angle, the curve was considered probably normal . If the peaks strayed off the comb's
angle, the curve was abnormal. This is a quick, useful method to read MRS and determine normal from abnormal. It is
important to remember, however, that this angle was used with STEAM spectra from the brain
Hunter's angle is the line formed by the metabolites in MR spectroscopy.
Metabolites Myoinositol ,Choline ,Creatine and N-Acetyl aspartate peaks are ascending in normal spectrum, any alteration in
the ascending nature of the peaks means spectrum is abnormal.
Hunters angle is alternative of doing complex ratios and analysis of the spectra
! MR perfusion[–]
o time to peak (TTP)
o mean transit time (MTT)
o cerebral blood volume (CBV)
o cerebral blood flow (CBF)
o negative enhancement integral (NEI)
Functional MRI
Dr J. Ray Ballinger and Dr Frank Gaillard et al.
Functional magnetic resonance imaging (fMRI) is a technique used to obtain functional information by visualising cortical
activity. fMRI detects subtle alteration in blood flow in response to stimuli or actions.
It is used in two broad ways:
1. clinical practice
o typically in pre-surgical patients
o aimed at localising eloquent areas (e.g. speech, motor)
2. research
o often cohort of patients (often normals)
o aimed at elucidating novel neural networks
fMRI is technically challenging to perform as the techniques used to visualise cortical activity (typically BOLD imaging) rely
on minute changes in a low signal to noise ratio (SNR)environment.
Technical requirements
1.5T or higher MRI
excellent quality assurance
appropriate software
appropriate paradigms and ability to deliver visual and auditory stimuli and record motor response
cooperative patients
Study design
There are two most commonly employed testing designs:
1. block design
2. event related design
Block design uses repeated blocks of activity (paradigm) separated by blocks of inactivity of alternative activity. This is by
far the most frequently used study design in clinical fMRI.
Event related design involves individual events rather than blocks, and can be randomly distributed during the study.
Paradigms
The activity performed or stimulus received by the patient is termed a paradigm, and each is designed to elicit a specific
cortical response. Numerous paradigms have been developed of various complexity. In the clinical setting four paradigms
(with modifications according to the clinical situation) suffice for most indications.
1. visual paradigm
2. motor paradigm
3. speech paradigm
4. memory paradigm
MR artefacts
MR hardware and room shielding:
Herringbone artifact (also called as crisscross artifact or corduroy artifact) is an MRI artifact , it appears as a fabric of
herring bone . Artefact is scattered all over the image in a single slice or multiple slices .
Causes
electromagnetic spikes by gradient coils
fluctuating power supply
RF pulse discrepencies
Moire fringes are an interference pattern most commonly seen when doing gradient echo images with the body coil.
Because of lack of perfect homogeneity of the main magnetic field from one side of the body to the other, aliasing of one
side of the body to the other results in superimposition of signals of different phases that alternatively add and cancel. This
causes the banding appearance and is similar to the effect of looking though two screen windows.
Zebra stripes / artifacts appear as alternating bright and dark bands in a MRI image. The term has been used to describe
several different kind of artifacts causing some confusion.
Artifacts that have been described as a zebra artifact include the following:
Moire fringes1,2
Zero-fill artifact3,4
Spike in k-space5
Zebra stripes have been described associated with susceptibility artifacts6
In CT there is also a zebra artifact from 3D reconstructions7 and a zebra sign from hemorrhage in the cerebellar sulci8.
It therefore seems prudent to use "zebra" with a term like "stripes" rather than "artifacts".
The central point artifact is a focal dot of increased signal in the center of an image. It is caused by a constant offset of the
DC voltage in the receiver. After Fourier transformation, this constant offset gives the bright dot in the center of the image as
shown in the diagram.
The axial MRI image of the head shows a central point artifact projecting in the pons in the center of the image.
RF overflow artefact cause a nonuniform, washed-out appearance to an image. This artifact occurs when the signal
received by the scanner from the patient is too intense to be accurately digitized by the analog-to-digital converter.
Autoprescanning usually adjusts the receiver gain to prevent this from occurring but if the artifact still occurs, the receiver
1 2
gain can be decreased manually . Post-processing methods also exist but may be time consuming .
inhomogeneity artifacts
MR software:
The slice-overlap artefact (also known as cross-talk artefact) is a name given to the loss of signal seen in an image from
a multi-angle, multi-slice acquisition, as is obtained commonly in the lumbar spine. It should not be confused with cross
excitation which although similar in causation, is not due to angled images.
If the slices obtained at different disk spaces are not parallel, then the slices may overlap. If two levels are done at the same
time, e.g., L4-5 and L5-S1, then the level acquired second will include spins that have already been saturated. This causes a
band of signal loss crossing horizontally in your image, usually worst posteriorly.The dark horizontal bands in the bottom of
the following axial image through the lumbar spine demonstrates this artifact.
As long as the saturated area stays posterior to the spinal canal it causes no harm.
Cross-excitation artifact is a type of MRI artifact and refers to loss of signal within a slice due to pre-excitation from RF
pulse meant for an adjacent slice.
The frequency profile of the RF pulse is imperfect; this means that during slice selection there is some degree of excitation
of the adjacent slices as well. If that adjacent slice is imaged during the same TR (i.e., multi-slice imaging) or soon after (i.e.,
imaging without leaving a gap), it will be partially saturated to begin with, and the resulting signal will be reduced. This
phenomenon is more conspicuous in inversion recovery (180°) sequences
Remedy
leaving a minimum gap of 1/3 slice thickness when imaging contiguous slices
interleaving between slices
employing 3D imaging if volume imaging is required
Using optimized pulse sequences that have a time penalty of a higher minimum TE and reduced number of slices for
a given TR.
Patient and physiologic motion:
Phase-encoded motion artefact is one of many MRI artefact, and occurs as a result of tissue / fluid moving during the scan
and manifests as ghosting in the direction of phase encoding, usually in the direction of short axis of the image (i.e left to
right on axial or coronal brains, and anterior to posterior on axial abdomen).
These artefacts may be seen from arterial pulsations, swallowing, breathing, peristalsis, and physical movement of a patient.
When projected over anatomy it can mimic pathology, and needs to be recognized. Motion that is random such as the
patient moving produces a smear in the phase direction. Periodic motion such as respiratory or cardiac/vascular pulsation
produces discrete, well defined ghosts. The spacing between these ghosts is related to the TR and frequency of the motion.
Motion artifacts can be distinguished from Gibbs or truncation artifacts because they extend across the entire FOV, unlike
truncation artifacts that diminish quickly away from the boundary causing them.
1. identifying known moving / flowing structures and noting that the artefact is in line with them (horizontal or vertical
depending on phase encoding orientation)
2. matching shape of ghost to that of flowing vessel (e.g round psuedolesion due to aorta ghost)
3. wide windowing to see repetitive ghost beyond confines of anatomy
4. they can be distinguished from Gibbs or truncation artifacts because they extend across the entire FOV, unlike
truncation artifacts that diminish quickly away from the boundary causing them
Solutions to phase mismapping include:
Entry slice phenomenon occurs when unsaturated spins in blood first enter into a slice or slices. It is characterized by
bright signal in a blood vessel (artery or vein) at the first slice that the vessel enters. Usually the signal is seen on more than
one slice, fading with distance. This mechanism is used in a positive fashion to generate flight MR angiograms.
This artifact has been confused with thrombosis with disastrous results. The characteristic location and if
necessary, the use of gradient echo flow techniques can be used to differentiate entry slice artifacts from
occlusions.
Spatial saturation bands place before the first slice and after the last can be used to eliminate this artifact.
This artifact occurs in gradient echo sequences as a result of selecting an echo time (TE) in which the fat and water spins
(located in the same pixel at an interface) are out of phase, cancelling each other. At 1.5 T, the 3.5 PPM difference in
frequency between water and saturated fat results in cancellation of spins at 4.5 ms multiples, starting at about 2.3 ms; for
example at 6.8 ms, 11.3 ms, and 15.9 ms. This artifact does not occur with spin echo sequence as the spins are rephased
by the 180o refocusing gradient.
To avoid this artifact, TE's close to 4.5 ms, 9 ms, 13.6 ms, should be chosen or fat suppression used. Using a SE sequence
instead of GE will also eliminate the artifact.
The magic angle is an MRI artefact which occurs on sequences with a short TE (less than 32ms;T1W sequences, PD
sequences and gradient echo sequences).
It is confined to regions of tightly bound collagen at 54.74° from the main magnetic field (Bo), and appears hyper intense,
thus potentially being mistaken for tendonopathy.
In tightly bound collagens, water molleules are restricted usually causing very short T2 times, accounting for the lack of
signal. When molecules lie at 54.74° there is lengthening of T2 times (don't understand why, but it involves 'bipolar coupling')
with corresponding increase in signal.
Typical sites include :
Tends to occur only on short TE sequences (e.g. T1, GRE, PD) - sequences with a longer TE (e.g. T2 including FSE T2)
can be used to avoid this artefact.
Other non-pathologic causes of high signal within tendons include near tendon insertions, and/or where the tendon
normally fans out or merges with other tendons.
Magnetic susceptibility artifact refers to a distortion in the MR image especially seen while imaging metallic orthopedic
hardware or dental work. This results from local magnetic field inhomogeneities introduced by the metallic object into the
otherwise homogeneous external magnetic field B0. These local magnetic field inhomogeneities are known as magnetic
susceptibility and are a property of the object being imaged.
In terms of magnetic susceptiblity, most materials can be classified as diamagnetic, paramagnetic, superparamagnetic, or
ferromagnetic.
In MRS the shift in Lamor frequency allows separation of different chemical peaks. The actual amount of chemical shift as
an absolute value is difficult to measure, so instead it is represented relative to a reference, and expressed in parts per
million (ppm).
In MRI, both spin echo sequences (SE) and gradient echo sequences (GE) may demonstrate chemical shift misregistration
or mismapping. The mismapping will occur in the frequency encoding direction, and show up as a bright band on one side
and a dark band on the other side of a fat-soft tissue interface.
In addition to mismapping, GE sequences can show another type of chemical shift induced artifact known as the black
boundary or india ink artifact. In the artifact a black line is seen in all directions at fat-water interfaces. In pixels with roughly
equal amounts of fat and water, the fat and water spins are 180o out of phase at certain echo times because of their
chemical shift or frequency difference causing cancellation of signal.
These effects can be used to confirm, for example, the presence of fat in a lesion.
The so-called india ink artifact also known as the black boundary artifact is an MRI feature seen on out-of-phase imaging
and represents signal drop out in voxels that contain both fat and non-fat components. As such it is seen surrounding
interfaces, lending the image an appearance as if someone has outlined these interfaces with ink.
Remedy
increasing the matrix size (i.e. sampling frequency for the frequency direction and number of phase encoding steps
for the phase direction)
use of smoothing filters (2-D Exponential filtering, Gegenbauer reconstruction etc.)
If fat is one of the boundaries, use of fat suppression
Zero fill artefact is one of many MRI artefacts and is due to data in the K-space array missing or set to zero during
scanning. The abrupt change from signal to no signal results in artifacts in the images showing alternating bands of shading
and darkness, often in an oblique direction.
A spike in k-space as from an electrostatic spark is another artifact that causes oblique stripes.
Aliasing / wrap around artifact in MRI (also known as wrap-around) is a common MRI artifact that occurs when the field of
view (FOV) is smaller than the body-part being imaged. The part of the body that lies beyond the edge of the FOV is
projected on to the other side of the image.
This can be corrected, if necessary, by oversampling the data. In the frequency direction, this is accomplished by sampling
the signal twice as fast. In the phase direction, the number of phase-encoding steps must be increased with a longer study
as a result.
Case 1 demonstrates axial T2-weighted images of the brain that demonstrate aliasing. The first image shows wrap-around
with the back of the head projected over the front because the phase-encoded direction is anterior-posterior and the FOV is
too small. The second image has the phase and frequency directions reversed resulting in absence of the aliasing artifact.
Oversampling was used in the frequency direction to eliminate the aliasing.
More detail
The basis of aliasing lies in "analog-to-digital conversion" wherein the continuous MR signal picked by the receiver coil is
converted into its digital counterpart for presentation as a grey-scale image. This ubiquitously involves sampling of the
continuous signal at pre-defined intervals. For greater fidelity in signal conversion, the sampling rate should be at least twice
the highest frequency within the signal (Nyquist rate). At lower sampling rates, high frequency signals become
indistinguishable from lower frequency signals, i.e., they become aliases.
In MRI, spatial localization within a single image depends upon the frequency signature of the MR signal originating from
that portion. Within a given bandwidth, higher frequency signals generally come from the periphery of the image and are
aliased over the lower frequency (relatively) central portion of the image. Aliasing in MRI can occur in both phase and
frequency axis.
Remedy
Aliasing in MRI can be compensated for by:
MRI contrast agents have become an indispensable part of modern magnetic resonance imaging. Although MRI was
initially hoped to provide a means of making definitive diagnoses noninvasively, it has been found that the addition of
contrast agents in many cases improves sensitivity and/or specificity.
History
Paul Lauterbur and his associates were the first to demonstrate the feasibility of using paramagnetic contrast agents to
improve tissue discrimination in MRI 1.
Contrast agents
Contrast agents can be grouped in many different ways. One way is to consider them according to the compartments in
which they distribute.
Paramagnetic contrast agents are use to enhance MRI images and display areas of hypervascularity and associated
pathology. They have their strongest effect in T1 weighted imaging because they predominantly alter the T1 relaxation
time in the tissues in which they have accumulated.
MRI imaging relies upon the signal generated from the behaviour of water protons. The relaxation of these water protons is
affected by surrounding tissues and, when a paramagnetic contrast agent has been absorbed in its vicinity, the relaxation of
water protons is enhanced.
The paramagnetic contrast agents generate a magnetic field 1000 times stronger than water protons. The interaction
between the contrast agent and the water proton is exactly the same as the corresponding interactions with other molecules
except that the magnitude of their magnetic interaction has a much greater effect on the relaxation time.
Commonly used contrast agents include:
Gadovist®
MultiHance®
Omniscan®
OptiMARK®.
There is an association between the use of paramagnetic contrast agents in patients with renal failure and nephrogenic
systemic fibrosis (NSF).
Though safer than the frequently used iodinated contrast agents used in x-ray and CT studies, there are safety issues with
MRI contrast agents as well. Paramagnetic metal ions suitable as MR contrast agents are all potentially toxic when injected
IV at or near doses needed for clinical imaging. With chelation of these ions, acute toxicity is reduced and elimination rate is
increased thereby reducing the chance of long term toxicity.
Intravenous gadolinium contrast agent safety
See also: nephrogenic systemic fibrosis
The most commonly reported reactions associated with the injection of Gd-DTPA are: headache (6.5%), injection site
coldness (3.6%), injection site pain or burning (2.5%), and nausea (1.9%). Recent adverse rates for Gd-DTPA are lower
than this and comparable to those of Gadodiamide and Gadoteridol (1.4%-3% for headache, nausea, and dizziness; <1% for
the others). The safety factor or ratio (ratio of the LD50 to the imaging dose) may be used to assess the relative acute
toxicity of contrast agents. The elimination half-life for the Gd containing contrast agents range from 1.25-1.6 hours.
Gadolinium containing contrast agents usually have no effect on blood chemistries and hematologic studies except transient
elevation of serum iron and bilirubin levels. These elevations peaked at 4 to 6 hours post injection and returned to baseline
values in 24 to 48 hours. The mechanism of these elevations is uncertain but may be related to mild hemolysis. A 10%-11%
increase in the activated partial thromboplastin time and thrombin time occurs in vitro with inhibition of platelet aggregation.
The platelet aggregation inhibition is less than that seen with iodinated ionic contrast material and no bleeding problems are
reported clinically.
Deoxygenated sickle erythrocytes align perpendicular to a magnetic field in in vitro studies raising the possibility of occlusive
complications in patients with sickle cell anemia. No clinical reports of this potential problem have been found 1.
Transient and mild drop blood pressure is reported in both animals and humans. A study of 1,068 patients reports
hypotension in 0.3% of the subjects and other symptoms such as syncope probably associated with hypotension in 0.8%.
Most of these symptoms occur 25-85 minutes after the injection.
Reports of several episodes of severe anaphylactoid reactions after IV injection of Gd- DTPA are published. The frequency
of these reactions is about 1 in 100,000 doses. Potential risk factors may include a history of asthma and significant reaction
to previously administered iodinated contrast material. It is suggested that the threshold for injecting Gd be raised, in those
patients, based on an individual risk/benefit ratio. Prophylactic pharmacotherapy with antihistamines and corticosteriods,
such as Greenberger's protocol, is suggested for high risk patients prior to contrast injections 2.
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
At low concentrations used for bowel opacification, the T1 shortening dominates the signal intensity. This results in high
intensity on T1-weighted, T2-weighted and gradient echo images. At high concentrations, T2 shortening causes decreased
signal in all but very short echo sequences. This resembles the effect seen with superparamagnetic iron oxide (see Negative
GI Contrast Agents). At intermediate concentrations, a mixture of T1 and T2 shortening results in increased signal on T1-
weighted images and decreased signal on T2-weighted images.
Ferric ammonium citrate and Gd-DTPA with mannitol are safe and effective in humans, but both have minor side effects.
Ferric iron can cause teeth staining, gastric irritation, nausea, diarrhea, and constipation. Mannitol can nausea, vomiting,
and diarrhea. Gd-DTPA without mannitol is well tolerated but usually fails in opacify the entire small bowel. It also needs to
be buffered when used orally since this chelate is not very stable at the low pH found in the stomach 1.
Short T1-relaxation Agents
Short T1 relaxation time GI contrast agents include mineral oil, oil emulsions, and sucrose polyester. In these materials,
protons contained in -CH2- groups relax at a faster rate than those in water resulting in a short T1 time. This gives a bright
signal in the bowel on T1-weighted sequences. Of these materials only oil emulsions have been used successfully in
humans. These are palatable and produce homogeneous opacification of the stomach and small bowel, but are absorbed in
the distal small bowel and fail to fill the colon. This is circumvented by using a contrast enema when the colon must be better
visualized.
A novel approach to retrograde opacification of the colon has been shown in rats with a nonabsorbable fat substitute,
sucrose polyester but no human trials for this use have been done 2.
Combination Contrast Agents
Combinations of oil emulsion and paramagnetic substances may be used as bowel contrast agents. These include an
emulsion containing corn oil and ferric ammonium citrate, and an emusion containing baby formula with ferrous sulfate.
These are palatable mixtures that distribute uniformly in the bowel, however signal is lost in the distal small bowel in adults,
because of absorption of both the oil and the iron. Unlike in adults, the faster transit through the small bowel in infants
delivers bright contrast to the colon. The advantage of this combination over oil emulsions alone is the enhancement of
signal on T1-weighted and especially T2-weighted images.
Perfluorochemicals are organic compounds in which the protons are replaced by fluorine. This results in an absence of
signal in the bowel. Perfluoroctylbromide(PFOB)(C8F17Br) is the only perfluorochemical that has been investigated for oral
use in humans to date. It is commercially available now as perflubron (Imagent GI, Alliance), but at high cost. Potential
advantages are a rapid transit through the small bowel because of its low surface tension, the lack of taste or odor making it
palatable, and the absence of any known side effects. PFOB is immiscible as are all perfluorochemicals that are in their pure
or "neat" state. This may be an advantage because PFOB cannot be diluted by bowel contents, however, miscible agents
that mix with fluid in the bowel may give more uniform filling of the GI tract. Emulsifying PFOB, as is done for intravascular
use of perfluorochemicals, may overcome this potential problem.
Positive vs Negative contrast agents
The question of which type of contrast enhancement of the bowel is the best, positive or negative, is sill debated. We may
find a positive or negative oral contrast agent better depending on the specific organ or disease suspected and the pulse
sequence used.
Two disadvantages of positive oral contrast agents are ghosting artifacts because of respiratory and peristaltic motion, and
loss of signal from dilution with secretions and retained fluid in the bowel. One method of reducing ghosting artifacts is to
use a pharmaceutical, such as glucagon or scopolamine, to reduce bowel motion. This increases the invasiveness of the
procedure. Other methods include the use of breath holding pulse sequences and first order flow compensation. Further
refinements of pulse techniques probably will make breath holding sequences more popular for abdominal MRI. This will
decrease artifacts from both peristalsis and breathing.
Dilution of positive contrast agents occurs in the upper GI tract if they are miscible with water because of gastrointestinal
secretions. This allows for the use of a small dose, but will cause loss of signal intensity as the concentration decreases.
Immiscible positive agents using oils, especially nonabsorbable ones, will not experience the loss of signal with dilution.
They will probably require a larger volume to replace any residual bowel contents.
Another disadvantage of a positive oral contrast agent is the possibility of residual material in the bowel simulating a mass
when surrounded by bright signal. The opposite is also true. A bright mass (such as a lipoma) might be obscured by the
contrast agent.
An advantage of positive oral contrast agents is the availability of several of these materials at this time. These include ferric
ammonium citrate, pediatric formula, and homemade oil emulsions. Positive agents are also inexpensive (except for
gadolinium solutions) and are safe to use.
Disadvantages of negative oral contrast materials include their high cost and lack of general availability (except for CO2 and
barium), and limited evaluations of safety on large number of patients. The expense may decrease with greater use of these
contrast materials and with competition between manufacturers.Metallic artifacts are seen when iron oxide concentrations,
ideal for spin echo sequences, are used with gradient echo sequences. This is because gradient echo sequences have
greater sensitivity to magnetic field inhomogeneity. Also there were some metallic artifacts seen in the colon on delayed (24
hour) imaging with the iron oxide preparations that probably can be eliminated as discussed above.
Lack of a fat plane between the negative contrast filled bowel and low signal intensity organs may make it difficult to
distinguish normal contours. An example of this is the plane between the stomach and the pancreas on T2-weighted
sequences. The majority of pathology appears bright on T2-weighted sequences and should be seen, however.
Advantages of negative oral contrast materials are several. The lack of signal in the bowel removes a source of ghosting
artifacts from spin echo sequences that may be present with positive agents. The loss of signal is fairly independent of
concentration of superparamagnetic iron oxide suspensions on spin echo sequences so that dilution should not be a
problem. The perfluorochemicals are immiscible with water and will not encounter dilution problems either.
Intravenous MRI contrast agents include chelates of paramagnetic ions, both ionic and nonionic. The particulates,
sequestered in the liver, spleen, and lymph nodes, the intravascular agents, confined to the blood pool, and tumour specific
agents are discusses separately (see bottom).
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
Paramagnetic metal ions suitable as MR contrast agents are all potentially toxic when injected IV at or near doses needed
for clinical imaging. With chelation of these ions, acute toxicity is reduced and elimination rate is increased thereby reducing
the chance of long term toxicity.
Ionics
Chelates of paramagnetic ions Cr and Gd with EDTA were first used however EDTA was of relatively low stability resulting
in toxicity in animals 1. Chelates with a higher stability constant have since been used successfully such as Gd-DTPA .
Gd-DTPA was the first intravenous MR contrast agent to be approved for human use (Magnevist, Berlex Labs). Gd has a
large magnetic moment, exceeded only by Dysprosium(III) and Holmium(III), explaining its paramagnetic properties at low
concentrations. This large magnetic moment is related to its seven unpaired orbital electrons. Gd-DTPA has similar
pharmacokinetics as iodinated contrast agents. It is distributed in the intravascular and extracellular fluid spaces, does not
cross an intact blood-brain-barrier, and is excreted rapidly by glomerular filtration 2.
Nonionics
The development of nonionic contrast agents for MRI has paralleled that for iodinated contrast materials. Ionic chelates are
also hyperosmolar and some of their side effects may be attributed to this property.
Gadodiamide
Gadodiamide (Omniscan, Winthrop Pharm.) is a nonionic complex with two-fifths of the osmolality of Gd-DTPA. It has a
median lethal dose of 34 mmol/kg resulting in a safety ratio of 2-3 times that of Gd-DOTA, and 3-4 times that of Gd-DTPA.
No abnormal serum bilirubin levels occur, however elevated serum iron levels occurred with an incidence of 8.2% in one
study of 73 patients.The efficacy of this contrast is similar to that of Gd-DTPA 3.
Gadoteridol
Gadoteridol (Prohance, Squibb) is the third intravenous contrast agent on the market. It is a low osmolar, nonionic contrast
as is Gadodiamide. Indications for use and efficacy are similar to the other agents 4.
Intravenous MRI contrast agent safety
Intravenous MRI contrast agent safety and Nephrogenic systemic fibrosis are discussed separately.
Gadolinium (Gd) chelates are paramagnetic agents that are injected intravenously during MR imaging.
Gd molecules shorten spin-lattice relaxation time (T1) of voxels in which they are present. As a result, on T1-weighted
images they have a brighter signal. This can have a number of utilities:
detection of focal lesions, i.e. tumour, abscess, metastasis
imaging of vessels in MR angiography
characterization of liver lesions, e.g. hepatoma, haemangioma, based on enhancement characteristics
Intravascular MRI contrast agents normally remain confined to the intravascular space, compared to Gd-DTPA which
distributes throughout the extracellular fluid space. This is a result of intravascular agents having a molecular weight of
approximately 70,000 and above, compared to a molecular weight of 590 for Gd-DTPA.
There are several advantages of intravascular agents:
they can assess perfusion in areas of ischemia and provide information about capillary permeability in areas of
reperfusion
they can show the extent of tumor neovascularity and associated permeability changes
they are useful in studies requiring prolonged imaging.
Three types of intravascular contrast agents will be discussed separately:
Gd-DTPA labeled albumin- Gd-DTPA labeled albumin is an intravascular MRI contrast agent. Gd-DTPA is covalently
bonded to albumin in ratios from 16:1 to 31:1 providing excellent enhancement of liver, spleen, myocardium, brain, and slow
moving blood of rats and rabbits. The albumin has a molecular weight of about 92,000 and a biological half-life of 88
minutes. The dose of Gd-DTPA required when bound to albumin is 0.062 mmol/kg compared to usual doses of the chelate
alone of 0.1-0.2 mmol/kg. No adverse reactions are reported, but in vivo retention of the gadolinium for several weeks in
liver and bone raises concerns of long term toxicity.
Gd-DTPA labeled dextran- Gd-DTPA labeled dextran is an intravascular MRI contrast agent. Dextran is a polysaccharide
consisting of a polymer of glucose molecules with a molecular weight between 75,000-100,000. Dextran has a high level of
safety and is broken down more rapidly than albumin. Approximately 15 Gd-DTPA molecules are attached to each dextran
molecule with an easily hydrolyzable bond. It is hoped that this will reduce the long term in vivo retention seen with Gd-
labeled albumin. The trade off is a shorter biological half-life of 43 minutes. In a rat model, satisfactory enhancement of liver,
spleen, kidneys, myocardium, and brain is seen for up to 1 hour. The dose of dextran Gd-DTPA in these studies was 0.01-
0.05 mmol/kg 1.
A variation of this contrast agent has been commercialized as Riovist (Schering) 2.
chromium-labeled red blood cells- Chromium labeled red blood cells is an intravascular MRI contrast agent. The
use of 51Cr-labeled RBC's in nuclear medicine suggested the use of paramagnetic Cr(III)-labeled RBC's as an
intravascular contrast agent for MRI. In dogs, significant enhancement of the liver and spleen is noted with minimal
enhancement of the kidneys. The blood volume to be replaced with labeled RBC's in these studies corresponding to
about one unit in humans. The survival half-life of labeled cells is 4.7 days compared to 16.6 days for unlabeled cells.
Free Cr may contribute to the enhanced relaxation rates of the liver and spleen. Short term toxicity appears to be
low, but further studies are necessary.
Tumor specific MRI contrast agents are pharmaceuticals that are targeted to tumors, either specifically or nonspecifically.
Monoclonal antibodies are targeted to specific tumors such as adenocarcinoma of the colon. Metalloporphyrins exhibit
affinity for many tumor types including carcinoma, sarcoma, neuroblastoma, melanoma and lymphoma.
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
Monoclonal antibodies
Monoclonal antibodies (McAb) are used successfully in nuclear medicine for localization of tumors but an initial attempt at
extending this use to MRI with paramagnetic (Gd3+) labeled antibodies was unsuccessful because of the estimated 800-fold
lesser sensitivity of MRI. This problem can be addressed in several ways: 1) increasing the number of paramagnetic ions
attached to the McAb; 2) attaching several paramagnetic ions to a macromolecule that in turn is attached to a McAb; 3)
using more antibodies or those with an affinity to many antigenic sites per cell or both; and 4) using a superparamagnetic
particle attached to the McAb. Implanted human colon carcinoma tumors in mice have been successfully imaged by using
monoclonal antibodies with a large number of Gd-DTPA molecules attached.
Additional studies report the use of very small magnetite particles coated with McAb. The magnetite cores are 10-20 nm in
diameter with a total particle diameter of 20-32 nm. The magnetic moment of these superparamagnetic particles is about
1000 times that of comparable paramagnetic particles. This allows the use of 1-10 nmol concentrations of the McAb coated
magnetite particles. Mixed success has been obtained in rodents with implanted neuroblastoma and human colon
carcinoma.
Metalloporphyrins
The metalloporphyrin most commonly used as a MRI contrast agent is Mn(III)TPPS4 (manganese(III) tetra-[4-
sulfanatophenyl] porphyrin) because of its low toxicity (compared to Fe(III)TPPS4 for example). A safety ratio of about 6:1 is
estimated in mice. This material appears to work best with tumors that are isointense to surrounding structures on T1-
weighted sequences. Incidentally, the fluorescent and tumor localizing characteristics of porphyrin derivatives have been
exploited in phototherapy of tumors.
Nitroxides
Nitroxide stable free radicals or nitroxyl spin labels as they may be called, are chemically stable organic compounds that
have an unpaired electron that results in paramagnetic properties. They generally consist of a six- member ring piperidine
derivative or a five-member ring pyrroxamide derivative. The pharmacokinetics of nitroxides are similar to iodinated contrast
agents and Gd- DTPA. They do not cross an intact blood brain barrier and undergo glomerular filtration as a dominant route
of elimination. Their ease of conjugation to various biomolecules makes them attractive for targeting to various organ
systems. Nitroxides are chemically stable and show limited in vivo metabolism. Their relaxation effects in vivo can be
eliminated almost immediately by IV injection of sodium ascorbate, a strong reducing agent. This will allow an unenhanced
MR study to be performed immediately after a contrast enhanced study, if the contrast study is not satisfactory alone.
The early ionic derivatives of piperidine have a 38 minute half-life and a safety ratio of between 8:1 and 100:1. Nonionic
pyrrolidine derivatives are formulated with a longer half-life of 45-50 minute in dogs, estimated to be about 2 hours in
humans. The LD50 in mice of this nonionic formulation is about 25 mmol/kg, making it twice as safe as earlier ionic
piperidinyl preparations. Mutation and toxicity studies show no evidence of genetic or other cellular damage in mammalian
cell preparations.
Larger molecular weight nitroxides exhibit increased relaxation rates as do paramagnetic ions attached to macromolecules.
This phenomenon occurs when attaching five-membered nitroxide rings to fatty acids. The fatty acids attach to human
serum albumin, either in vitro or in vivo, resulting in a significant increase in relaxation rate. Safety studies and clinical trials
need to be performed before nitroxides will be available for use.
Ferrioxamine methanesulfonate
Ferrioxamine methanesulfonate is a paramagnetic contrast agent that has undergone phase I and phase II clinical trials for
use as an IV and retrograde contrast agent for the kidneys, ureters and bladder. It is more stable than Gd-DTPA, though its
relaxivity is somewhat less, as expected from it having 5 unpaired electrons, vs 7 unpaired electrons for Gd-DTPA. 80% is
eliminated by renal excretion and 20% by hepatic excretion. Ferrioxamine undergoes renal excretion by glomerular filtration
but is actively reabsorbed in the tubules. This results in a longer plasma half-life than Gd-DTPA (128 min. vs 20 min. in rats).
In clinical imaging the long plasma half-life allows enhancement of the kidneys for 60 minutes with little change in intensity.
Significant improvement in detectability of lesions in the kidneys is demonstrated over unenhanced controls. Side effects
include epigastric distress and transient burning at the injection site. Increase in serum iron levels and a transient elevation
of serum liver enzymes (SGOT/SGPT) have been reported.
Hepatobiliary MRI contrast agents are desirable for several reasons: to detect mass lesions such as metastases within the
liver; to evaluate functional status of the liver in diffuse hepatocellular diseases such as cirrhosis; and to obtain high
resolution images of the gallbladder and biliary tree.
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
Three advantages of a hepatobiliary contrast agent over a particulate agent targeted for Kupffer cells exist. First of all, there
are many more hepatocytes than Kupffer cells (78% vs. 2% by volume) in the liver, improving uptake efficiency of contrast
material. Second, the biliary ducts are opacified by excreted contrast material, eliminating confusion of normal bile ducts
from focal abnormalities as may occur with particulate agent contrast materials. Third, the contrast agent is rapidly excreted
from the body reducing potential toxicity. In contrast, materials phagocytized by the reticuloendothelial system (including
Kupffer cells) remain in the body for a long period of time.
The uptake and excretion of contrast material allows visual assessment of basic hepatocyte function. The biliary
opacification will allow functional information to be obtained as with radionuclide hepatobiliary studies but with significantly
higher spatial resolution.
Manganese chloride
Manganese chloride (MnCl2) is a prototype hepatobiliary contrast agent. IV and oral administration in animals results in a
rapid decrease in the T1 relaxation time of the liver, spleen, kidneys, heart, and bile causing a bright signal on T1-weighted
images. In its ionic state, Mn+2, it is relatively toxic. In imaging doses of 0.2 mmol/kg, it caused severe hypotension and
ventricular fibrillation in dogs. Cerebral damage results from chronic manganese toxicity. Mn can be used in the form of a
chelate with diminished toxicity for hepatobiliary imaging in humans.
Hepatobiliary chelates
Chelates used as hepatobiliary contrast agents consist of a paramagnetic ion bound to an organic ligand, forming a
complex that shows affinity for hepatocytes. This type of complex is desirable to increase uptake of the contrast agent by the
hepatocytes and to reduce toxicity of the paramagnetic metal ion as is done with gadolinium. Possible chelates for
hepatobiliary imaging include Fe-EHPG and derivatives, Gd-HIDA, Cr-HIDA, B-19036, and Mn-DPDP.
Chromium diethyl HIDA meglumine (Cr-HIDA) is another analog of a hepatobiliary radiopharmaceutical that has been tested
in rats and rabbits. Excretion was 45%-77% after one hour. The dose required to give significant increased signal in the liver
(0.25 mmol/kg) results in a safety ratio in mice of 6:1, compared to 100:1 for Gadolinium-DTPA. This low ratio for Cr-HIDA
precludes clinical use.
An octadentate chelate of gadolinium coded B-19036 (Bracco Industria Chimica S.p.A., Milan, Italy) may be used as a
hepatobiliary contrast agent. It is a highly stable complex with an LD50 in mice comparable to Gd-DTPA. It has yet to be
tested in humans.
Manganese(II)-dipyridoxal diphosphate (Mn-DPDP) is a manganese chelate derived from vitamin B6, pyridoxal-5-
phosphate. It shows efficacy in detecting small liver metastases in rabbits and has undergone phase I clinical trials showing
it to be safe and effective in enhancing the signal intensity of the liver. It has a safety ratio of 200:1 in rats which is somewhat
better than that for Gd-DTPA. Unlike Fe-EHPG and its derivatives whose uptake by hepatocytes depend on their lipophilic
attraction to the cell membrane, Mn-DPDP is recognized by a vitamin B6 transport system in the cell membrane.
Reticuloendothelial MRI contrast agents can best be discussed in terms of those used for liver and spleen imaging and
those for lymph node imaging.
NOTE: This article has been transferred from mritutor.org and was last updated in March 5, 1996. Review and edit pending.
Two major cell types can be targeted for hepatic imaging. Hepatocytes comprise about 78% of the liver by volume, and
Kupffer cells of the reticuloendothelial system comprise about 2% by volume. Originally, particulate contrast agents were
targeted for the RES but recently ultra-small particles have been used that bind to a specific receptor site on the hepatocyte
cell membrane.
The following problems that can arise with detecting small lesions in the liver using SPIO:
1. small lesions may be indistinguishable from the flow void in small blood vessels seen in cross-section
2. aortic pulsation artifacts are more noticeable
3. the one hour delay between injection and imaging make it impractical to decide at the last minute to give contrast.
Liposomes
A liposome is a spherical vesicle consisting of one or more bilayer phospholipid membranes or lamella. Liposomes for
hepatic imaging range in size from about 20 nm to 400 nm diameter. Reasons to use liposomes as a carrier for
paramagnetic contrast materials include:
1. changing the interaction between water molecules and the contrast agent
2. changing the rate of removal of the contrast agent from the blood pool
3. targeting specific organ systems, e.g., liver, spleen, and bone marrow.
Paramagnetic materials can be incorporated into either the aqueous inner chamber or the bilayer membrane. Encapsulation
of superparamagnetic iron oxide particles into liposomes (Ferrosomes) has been reported. Both Gd-DTPA and MnCl2 can
be encapsulated into the aqueous inner chamber of liposomes 4.
Liposomes are taken up only by the Kupffer cells. Once in the Kupffer cells, Mn+2 or Gd-DTPA is slowly released and
diffuses into adjacent hepatocytes, resulting in enhancement of normal liver but not malignancies.
Stable nitroxide free radicals have been attached to phospatidylcholine, a common constituent of liposome lamellae. They
may also be attached to derivatives of the fatty acid, stearic acid, as have the DTPA chelates of Mn and Gd. This results in a
lipophilic side chain that allows incorporation into the liposome membrane.
Lymph nodes
Two clinical problems common to CT and MR imaging are: 1) distinguishing unenlarged metastatic lymph nodes from
normal lymph nodes; and 2) differentiating enlarged metastatic nodes from benign hyperplastic nodes. Differentiation of
metastases from fibrosis, lipomatosis and cysts is possible with resected lymph nodes in a 4.7T magnet using voxels of size
0.1 by 0.1 by 1.0 mm; however, gradient strength and switching capabilities are not adequate in clinical imagers to obtain
the necessary spacial resolution. This inadequacy of clinical imagers is circumvented by the use of USPIO.
USPIO particles with a mean diameter of 80 nm may be injected into the interstitium of the foot pad of rats. After a suitable
delay, marked loss of signal of normal lymph nodes is seen. Metastatic nodes show less uptake resulting in less decrease in
signal, allowing differentiation of normal-sized, metastatic nodes from uninvolved, normal nodes. From experience with
conventional lymphangiography, this route of injection is unlikely to opacify all the abdominal lymph nodes 5.
USPIO particles, with a median diameter less than 10 nm, will localize in lymph nodes following an IV injection. This material
does not undergo uptake by the RE system as rapidly as larger particles, resulting in a longer plasma half-life in rats (81
minutes, vs 6 minutes). This factor and its small size allow transcapillary passage either into the interstitium and then to the
lymph nodes or directly into the lymph nodes. In the rat model, IV injection of USPIO allows differentiation of normal lymph
nodes from normal size metastatic nodes based on differences in signal characteristics. MR microscopy of excised lymph
nodes, performed at 9.4T shows the USPIO to be associated with macrophages in the medullary sinuses.
MRI safety
Dr Jeremy Jones and Dr J. Ray Ballinger et al.
MRI scanners, although free from potentially cancer inducing ionising radiation found in plain radiography and CT, have a
host of safety issues which must be taken very seriously. These can be divided into:
The strong field also effects common devices such as pacemakers and watches. The magnetic reed switch in modern
pacemakers is disturbed by strong magnetic fields resulting in possible deleterious effects to the patient with one implanted.
Mechanic watches will "freeze up" in a strong field, sometimes permanently.
Many intracranial aneurysm clips are ferromagnetic and as a result experience a torque or twisting in a magnetic field. Not
everyone with an aneurysm clip experiences a fatal hemorrhage when placed in a magnet, but several cases have been
reported.
Some types of heart valves (e.g., Star-Edwards) are torqued in a magnetic field: however, this torque is less than the
stresses that occur normally as a result of blood flow. Therefore heart valves are now considered not to be an absolute
contraindication for MRI.
More of an annoyance than a safety problem is the ability of the magnetic field of a MRI machine to erase the information
contained on the magnetic strip on ATM and credit cards. This may occur a short distance inside of the scanner room of a
MRI machine.
Some metallic objects that are usually safe near an MRI machine are gold jewelry and eyeglass frames.
Radio-frequency
The radio frequency power that is capable of being produced matches that of many small radio stations (15-20 kW). As a
result there is the presence of heating effects from the RF. In most pulse sequences, the heating is insignificant and does
not exceed the FDA guidelines. New pulse sequences such as for echo planar imaging and some spectroscopy localization
techniques are capable of exceeding the FDA guidelines. Monitoring of the power deposition in patients is a requirement for
FDA approval of clinical MRI scanners.
Potential for electrical shock exists with RF coils so proper grounding and insulation of coils is necessary. Any damage to
coils or their cables needs prompt attention. Also looping of the cable to a coil can result in burns to patients that come into
contact with them. It is best to avoid all contact with the RF coil cables.
Neuroanatomy and Cortical / Landmarks of Functional Areas
Sensorimotor Cortex
Transverse Sections
The precentral gyrus fuses with the superior frontal gyrus at the very upper
convexity .The precentral gyrus is the most posterior part of the frontal lobe that
extends inferiorly to the Sylvian fi ssure . At the apex, the pre- and postcentral
gyri form the paracentral lobule as they fuse. Making a little detour to a lateral
view the cingulate sulcus ascends at the medial interhemispheric surface dorsal
to the paracentral lobule (pars marginalis) and thus separates it from the
precuneus . This intersection can be appreciated on axial sections as the
“bracket”-sign that borders the postcentral gyrus . Somatotopographically, the
apex harbours the cortical representation the lower extremity. Following its
course along the lower portion of the gyrus adjacent to the Sylvian superficial
convexity (from medial–posterior–superior
to lateral–anterior–inferior), the cortical surface of the
precentral gyrus increases at its posterior margin, building
the omega-shaped motor hand knob; Within this primary motor cortex (M1)
of the hand, there is an additional somatotopic order of
the individual digits .From medial to lateral, the hand is organized
beginning with digit 5 (D5), to the thumb representation
(D1) being the most lateral .The motor hand knob is another typical landmark of
the precentral gyrus ; however, as the CS and the
postcentral gyrus follow this course, there is also an
omega-shaped structure in the postcentral gyrus (harboring
the somatosensory hand area). However, as
described above, the ap-dimension of the postcentral
gyrus /is smaller compared to the precentral gyrus , thus often enabling a
differentiation/. Somatotopographically,the cortical somatosensory
representation follows the distribution of the precentral gyrus . Lateral to the
SFG , the medial frontal gyrus zigzags posteriorly and points towards the motor
hand knob . Beginning at this “junction” and lateral–inferior to this landmark, the
ap-diameter of the PreCG decreases, but it increases again along the lower
convexity.
This is the primary motor cortex (M1) of lip representation and tongue
movements. Previously, the anatomy of the frontal lobe has been
described partially. As the course of the medial frontal gyrus can be followed
nicely on axial sections, the lateral inferior aspect of the frontal lobe represents
the inferior frontal gyrus. Anterior to the preCG the prefrontal motor areas can
be found. The inferior frontal gyrus borders and overhangs the insula anteriorly.
This part is the frontal operculum harbouring the motor speech area of Broca.
The lateral ventricles with its anterior and posterior horn can easily be depicted
on axial sections due to its typical form and typical signal caused by cortico- spinal
fl uid. Their shape is formed through, the head of the caudate nucleus lateral to
the anterior horn, the thalamus lateral at its waist (III. ventricle) and posteriorly
by the fi bers of the anterior–posteriorly running optic radiation and left–right
running fibers of the splenium . Lateral to these structures, descending
corticospinal fibers pass the internal capsule and follow a certain somatotopic
organization. The internal capsule is framed medial by the head of the caudate
nucleus , the third ventricle and the thalamus (at the posterior aspect of the third
ventricle) and lateral by the globus pallidum. From medial to lateral towards the
insula the globus pallidus, putamen and claustrum within the lentiform nucleus
can be differentiated. In the anterior limb and the genu of the internal capsule,
corticospinal fi bers from the tongue, lip and face descend, whereas, in the
posterior limb, fi bers from the upper extremity, body and fi nally lower extremity
are found.
Sagittal Sections
The corpus callosum represents the biggest connection between the two
hemispheres. The frontal aspect is the genu , the medial part is the body and the
most rostral part is the splenium. The corpus callosum encases the lateral
ventricles. At the base the anterior commissure can be identifi ed as a roundish
structure. Sometimes, the posterior commissure (pc) can also be defi ned, which
represents a bundle of white fi bers crossing the midline, at the dorsal aspect of
the upper end of the cerebral aqueduct. Previously slice orientation of most fMRI
studies had been performed according to this ac-pc line in order to have a
reference system.
From the base to the apex, the corpus callosum is abutted by the callosal sulcus
and the cingulate gyrus.The gyrus abutting the cingulate sulcus is the medial
part of the SFG . In the region (at the medial cortical surface) framed by vertical
lines perpendicular to the ac (Vac) or pc (Vpc) the sup plementary motor area
(SMA) is harboured in the cigulate gyrus and superior frontal gyrus. As described
above, the cingulated sulcus ascends at the medial interhemispheric surface
dorsal to the paracentral lobule ;pars marginalis and thus separates it from the
precuneus .This intersection can be nicely appreciated on axial sections
as the “bracket”-sign that borders the postcentral gyrus .The postcentral gyrus is
already a part of the parietal lobe. The precuneus is located dorsal to the
postcentral sulcus. There is another important landmark that separates the
parietal lobe from the occipital lobe (cuneus , the parieto-occipital sulcus ). It can
be easy recognized in sagittal views, as the dorsal sulcus that follows an inferior–
anterior to superior–posterior course, posterior to the ascending part of the
cingulate sulcus . In mid-sagittal sections the motor hand knob [a] can again be
recognized as a “hook” that rises out of the parenchyma and points dorsally.
Further, laterally the sensorimotor cortex overhangs the insula . The Sylvian fi
ssue that separates the frontal lobe and the temporal lobe has an inferior–
anterior to superior–posterior course. At its anterior margin, it ascends into the
anterior horizontal ramus , and more dorsally into the anterior ascending
ramus of the frontal operculum, that also overhangs the anterior aspect of the
insula . The anterior horizontal ramus separates the pars orbitalis from the pars
triangularis , whereas the anterior ascending ramus separates the pars
triangularis from the pars opercularis of the frontal operculum of the inferior
frontal gyrus and thus form a “M”. The pars opercularis of the frontal operculum
of the inferior frontal lobe harbours Broca’s area. At its posterior margin, the
pars opercularis is delimited by the anterior subcentral sulcus. At the base of the
sensorimotor strip the precentral and postcentral gyrus fuse . This junction is
delimited dorsally by the posterior subcentral sulcus. Movement of
the lips or tongue induce an increase in BOLD signal at this portion . The
base of the sensorimotor area has, depending on anatomical
variations, a “K”- or “N”-shape that is built by the anterior subcentral sulcus and
inferior precentral sulcus, the precentral gyrus, posterior subcentral sulcus,
postcentral gyrus and postcentral sulcus that again borders the angular gyrus .The
posterior part of the Sylvian fi ssure separates – following its superior–posterior
course - ascends into the posterior ascending ramus flanked by the anterior
and posterior aspect of the supramarginal gyrus that has a horseshoe
appearance.
Fig. 2.2 Axial T2-weighted TSE MR images. 1 superior frontal gyrus; 2 medial frontal gyrus; 3
precentral gyrus; 4 postcentral gyrus; 5 “pars bracket,” cingulated sulcus; 6 precuneus, parietal
lobe; 7 intraparietal sulcus; 8 interhemispheric fi ssure; a hand knob; b paracentral lobule
Visual Cortex
Sagittal Sections
At the medial surface of the occipital lobe, there is a sulcus that zigzags anterior–
posteriorly called the calcarine sulcus , along which the visual cortex is located. The
calcarine sulcus separates the superior lip from the inferior lip of the visual cortex.
Напомним :
В случае T1 распад, если слишком много диполей протон начать указывают в Z-
направлении , суммаXY- компонентами отдельных диполей слишком малы, чтобы
быть измерено , даже если они выровнены. В случае T2 распад, если слишком
много диполей протона направлены в разные стороны , отдельные диполи будут
уравновешивать друг друга , даже если они по большей части сидят вдоль XY-
плоскости. В действительности , оба явления приводят к ослаблению сигнала .
11.Это слайд показывает, что протон не сидит спокойно в пространстве. Из-за
теплового взаимодействия , его ориентации и положения постоянно меняются.
Графическое не был произведен , имитируя явления , которые вызывают
случайные движения протонов, но просто , вызывая его ориентации на хаотически
меняются . Значение этого является то, что графики , показанные здесь, все , в
некоторой степени , неточны. Протоны не шаров с тонко разграниченных объем ,
магнитные поля не стрелы идвижений, описываемых происходить гораздо
быстрее и с разной пропорции друг к другу.
Цель этих графиков не показыватьточное описание физической реальности, но
эффективно передавать динамику концепций, описанных . Как и во всех таких
моделей , это всегда важно , чтобы думать о них как более качественный , чем
количественный .
12.The возврата дипольного к равновесию описывается математически как два
компонента: Т1 и Т2.
T1 распада описывает, каким образом параллельно - составляющей магнитного
поля возвращается к равновесию. Она определяется уравнением
B ( T) = B0 * (1- е ^ ( -t/T1 ) )
Важно только понять, из этого уравнения , что чем больше T1 , тем больше
времени требуется для дипольных , чтобы вернуться , чтобы перестроить с
магнитным полем .
В этом графические ( и во всех графика в этой статье ) ,параллельно - компонента
вектора такой же, какZ- компонента. Показанадипольных прецессирующим от + х -
х до+ Z- оси. XY- компоненты дипольного показана всем прецессии и
прослеживается в синий цвет.
Хотя T1 определяется по отношению к Z- компонент , важно помнить, что T1
распада влияет на XY- компонент , потому что, как было сказано ранее, это
прецессия XY- компонент, который поддается измерению .
Таким образом , чем больше T1 , тем большесигнал от данного протона остаться.
Протона с коротким T1 , однако, повернет вспять к Z- оси очень быстро, и сигнал
измеряется от такой протон распадается быстро.
13.Here ,отдельных диполей все были перемещены в начало координат. Обратите
внимание, что на 180 градусов импульс поворачивает диполей вокруг оси ординат
. Это не только восстанавливает Z- компонента на большее значение , но
изменяет порядок прецессии так , что эхо имеет самое высокое возможное XY-
компонента.
Ранее , графические показал протонов с разными частотами ларморовским из-за
градиента магнитного поля , но помните, что это десинхронизации происходит,
даже если мы попытаемся применить однородном магнитном поле , и которые мы
называем это явление T2 .
Одна большая разница между эхо градиента и спинового эха в том, что спин-эхо
меняет Z- компонент потери, в то время как градиент эхо может только
обратнаяXY- компонент десинхронизации . Интересно, что в результате
последовательности спинового эха может быть повторен несколько раз с
небольшим распада сигнала.
14.Most студентов учат, что протоны имеют заряд и что они также имеют угловой
момент -причудливый способ сказать , что они вращаются . Большинство
студентов также учат , что протоны имеют магнитный диполь. Эти два понятия
неразрывно связаны , но это , как правило, не рассматриваются в основные
классы химии.