2200I 2200I (H) User Manual V1.0F
2200I 2200I (H) User Manual V1.0F
2200I 2200I (H) User Manual V1.0F
Preface
Dear users,
First of all, thank you for your trust to choose our colposcope digital
imaging system!
To help you get familiar with the product as soon as possible, we provide a
detailed user manual along with the instrument. When you first install and use
the instrument, please read all the attached data carefully!
To meet the market and customer demands, the Company will
continuously improve quality and performance of the product. We will try to
keep you informed of the latest improvements to the product in a timely
manner. Your corrections are eagerly welcomed if there is any error and
omission in the User Manual!
The contents of the Manual are protected by copyright law. All rights
reserved. No part of the Manual may be reproduced, photographed, copied or
translated into other languages without prior written permission of the
Company.
● Contact information:
Edition: V1.0F
Colposcope System user’s manual
Instructions to Users
In case of any problem or in need of help during use, please contact
Technical Service Center of the Company. We will offer solutions as soon as
possible or send professionals for service.
Proper use can extend service life of the instrument, so as to maximize
its use value.
While improper use may cause harm to the instrument or workers. The
Company will neither assume any liability for any harm to the people or
instrument due to violations against provisions of the Manual, nor for reduced
performance indexes therefrom, such as instrument performance, reliability
and safety, nor offer free maintenance for such faults!
Table of Content
CHAPTER I SAFETY REQUIREMENTS AND PRECAUTIONS .................................................. 1
1.1 Contraindications
1.2 Warnings
1
Colposcope System user’s manual
1.3 Precautions
2
Colposcope System user’s manual
3
Colposcope System user’s manual
1.4 Note
4
Colposcope System user’s manual
As the lens has surface coating, the time for every cleaning should not be more
than 30 s, and the lens should not be pressed or rubbed hard.
If the battery is subject to obvious damage or run out of capacity, it should be
replaced.
Only the battery with the specifications specified by manufacturer can be used
for replacement, i.e. L50 AC 193 A 7.2 V, 2.2 Ah.
The battery replacement, requiring no professional, can be completed by the
user according to the method specified in Section 6.5.
Before the battery replacement, it is required to first turn off power supply of the
camera device and disconnect the control line.
5
Colposcope System user’s manual
Chapter II Overview
Application scope
Use of high resolution and high definition color digital camera device
High resolution color digital camera device
KN-2200I(H) is subject to high definition digital video camera, which has
effective pixels up to 32,700, and applies high definition monitor and
capture card that yield higher image quality and thus more accurate clinical
diagnosis.
KN-2200I supports video output; (KN-2200I(H) supports HD-SDI output).
Fast auto and manual focus control.
Automatic white balance adjustment.
KN-2200I supports two power supply modes: external AC and built-in
batteries (KN-2200I(H) does not support power supply by batteries).
Unique light source design scheme
Use of energy saving and environmentally friendly LED light source for
lighting, so that the brightness can be adjusted freely.
Combined cart
6
Colposcope System user’s manual
7
Colposcope System user’s manual
It is optionally equipped with bar code gun function to facilitate fast creation
and search of cases.
Rear LCD
Controller
Controller
Support Cart
KN-2200Ⅰ KN-2200Ⅰ(H)
Figure 2-1
definition one.
8
Colposcope System user’s manual
Camera device:
KN-2200I: Horizontal resolution of 550 lines; color digital CCD, automatic
focus control, and minimum illumination: 0.8 Lux (F1.2); PLA standard pixel
of 440,000; NTSC standard pixel of 380,000; magnification of 216 times.
KN-2200I(H): Horizontal resolution of 1,000 lines; automatic focus control,
and minimum illumination: 1.7 Lux (F1.6, when ICR is closed); pixel of
3,270,000; magnification of 240 times.
The instrument has no application part.
The instrument has no signal input/output part.
Working environment:
Temperature:5~40℃
Relative humidity:≤85%
Atmospheric pressure:700hPa~1060hPa
Computer configuration:
Dual-core CPU of 2.4 GHz or higher
Memory of not less than 2 G
Space of the hard disk for software installation of not less than 100 G
The mainboard has PCI-E slot, and the high definition products should
have HDMI.
Monitor: with hanging holes
KN-2200I: with resolution of not less than 1,440*900 pixels;
KN-2200I(H): with resolution of not less than 1,440*900 pixels; 1080P,
with HDMI or DP high definition interface
Printer: Supporting color printing, with resolution of not less than 600×600
pixels.
Software: Supporting operating system higher than WINDOWS 7.
Movable support: The support can be adjusted up and down, and the upper
cradle head part can be manually adjusted for 360 degrees;
Light source:
Illumination: The light source luminance can be adjusted. Within the
working distance, the illumination is not less than 1,200 Lux.
The illuminated surface temperature is not above 41°C.
Working distance: The camera device is 200~400 mm apart from the part to
be examined, and its focus should be adjusted so that clear images can be
observed on the monitor.
9
Colposcope System user’s manual
Note! Refer to the attached files (user's manual, installation and operation
Production date
Enhance luminance of the LED light source. Every press will increase the
luminance by one level.
Reduce luminance of the LED light source. Every press will reduce the
luminance by one level.
White balance
Freeze
10
Colposcope System user’s manual
Take the instrument out of the packing case, place it at a safe and stable position
for observation, and fasten the universal wheel at the bottom of the cart.
Check the attached accessories according to the packing list, so as to ensure
completeness of the accessories.
3.2 Connection
11
Colposcope System user’s manual
12
Colposcope System user’s manual
5. Insert the attached "encryption dog" into USB interface of the computer
6. Enter the WINDOWS operating system to have the icon pop up on the
desktop, and then double-click it to enter the system interface.
Precautions: Closing the main machine in an improper way may affect normal
use of its software system, or even damage software functions of the system.
Note: The "encryption dog" should be properly kept, not damaged or lost. If the
"encryption dog" is not inserted, the system can only implement demonstration.
Figure 5-1
While using, open the lens dust cover, and after using, close it to avoid
contamination or damage to the lens.
13
Colposcope System user’s manual
The operation panel of the camera device is as shown in the figure below:
Charging
indicator
Image Image
magnification reduction
Power
switch
Figure 5-2
Charging indicator: Yellow means charging, and blue means charging
completed.
Image magnification and reduction: Control magnification and reduction of
images.
Power switch: Press the switch for 2 s to open or close the camera device. To
open the camera device, blue light will be emitted when pressing the button,
while to close it, no light will be emitted.
The controller is used to control the LED light source and the image capture of
camera device.
White balance
Red light reduction
Image reduction Image magnification
14
Colposcope System user’s manual
Enhance luminance of the LED light source. Every press will increase
the luminance by one level.
Reduce luminance of the LED light source. Every press will reduce
the luminance by one level.
Note: Do not pull the connecting line hard when using the control handle in order
to avoid damage.
15
Colposcope System user’s manual
Chapter XI Maintenance
To ensure normal operation of the instrument, and prolong its service life, it is
required to pay attention to maintenance of the instrument.
To ensure normal and safe work of the system, preventive inspection (including
performance inspection and safety inspection) and maintenance should be conducted
for the instrument and its accessories once every six months to confirm that the
instrument can work normally and is in good working condition, and that it is safe for the
medical staff and patients, with accuracy up to the level required for clinical applications.
1. Ensure that the instrument is reliably grounded.
2. Pay attention to the local grid voltage fluctuation, and if it falls beyond the allowable
range, add voltage stabilizing equipment as suggested.
3. It is not allowed to open shell of the instrument without permission, so as to avoid
affecting its normal use due to undue fault.
4. The accessories equipped for the instrument should be handled with care to avoid
drop, impact or pull, and they should not be wiped with corrosive chemicals!
5. The accessories should be stored together with the instrument, so as to decrease the
insertion and extraction times to extend its service life.
6. The camera device should be placed in a clean, dust-free and stable place. The
anti-shock of lens should be strictly managed. The Company will not guarantee repair
for fault of the lens due to drop.
7. When not in use, the camera device should be held in rigid packaging box to avoid
collision.
8. The equipment and accessories should be inspected regularly according to the check
cycle specified by the hospital. At the check cycle time, please contact the
manufacturer. It is advised that the user should inspect conditions of the equipment
and accessories during daily operation and use, so as to take corresponding actions,
and conduct comprehensive technical inspection on them every half a year, so as to
detect mechanical and cable damage.
9. The instrument shall be subject to regular maintenance in accordance with relevant
provisions of the hospital.
10. Upon expiration of the service life, the instrument and accessories should be
disposed correspondingly according to relevant provisions on waste electronic
products.
16
Colposcope System user’s manual
6.2.1 Equipment
Note: The power supply should be turn off first, and the AC power supply cut off
before cleaning of the instrument.
1. The most commonly used hospital cleaning solutions and non-corrosive detergents
can be used for cleaning the instrument, but it is important to note that many
detergents must be diluted prior to use. Please follow instructions of the detergent
manufacturer for use.
2. Avoid using ethanol, amine or acetone-based detergents.
3. The shell and screen of the instrument should be kept from dust pollution, and can be
wiped with a lint-free soft cloth or a sponge wet with detergents.
Note: Do not pour the liquid on the instrument while cleaning, and ensure no
ingression of any liquid into the instrument.
4. It is forbidden to use abrasives, such as steel wire brush or metal polishing agent, as
these materials may cause damage to the instrument panel and display screen.
5. When patch plug of the cable is occasionally wet, rinse it with distilled or deionized
water, and then dry it under 40-80°C for at least one hour.
1. When cleaning the camera device, please do not use corrosive or abrasive
detergents. In the presence of dust, please use a dry cloth for cleaning. In case of
severe dirt, use neutral detergent to wipe it gently.
2. When cleaning the lens face, please do not use powerful or abrasive detergent. First
use a soft brush and blowing balloons to remove dust particles, then drip a small drop
of lens cleaning solution on the lens cloth or paper, gently wipe the lens surface, and
then wipe the lens with a clean cotton cloth in the end.
Note:
- Do not drop the cleaning fluid directly on the lens.
- As the lens has surface coating, the time for every cleaning should not be more
than 30 s, and the lens should not be pressed or rubbed hard.
6.3 Storage
If the instrument will not be used for a long time, it should be wiped clean and
placed in the packing box, and then stored in a dry and dust-free place.
17
Colposcope System user’s manual
18
Colposcope System user’s manual
19
Colposcope System user’s manual
The computer shows a Check whether the memory chip is tightly inserted or
blank screen and gives damaged.
alarm after start. The memory or mainboard is damaged.
Imaging cannot be The foot switch is damaged. Replace the foot switch.
implemented. The control line is subject to connection error.
20
Colposcope System user’s manual
21
Colposcope System user’s manual
1. During the warranty period, for all faults due to product quality, the After-sales Service
Center of the Company will offer free repair or replacement of accessories. Please do
not disassemble the machine for repair without permission.
2. In case of any special requirements for the services, the user can consult relevant
issues with the After-sales Service Center of the Company
3. Relevant technical data of the instrument can be provided to the technical service
staff authorized by the Company as necessary.
4. The Company will not offer free warranty services for faults as a result of the following
causes:
Faults due to disassembly & assembly or modification of the instrument without
permission.
Faults due to accidental break or drop during use or handling.
Faults due to lack of reasonable maintenance or failure to meet the environment
requirements for use.
Faults due to failure to conduct proper operation as required in the user manual.
Damage to the instrument and accessories due to human causes.
Faults due to maintenance without permission of the Company.
Faults or damage due to force majeure such as fire or earthquake.
22
Colposcope System user’s manual
Put the support in the base, fit M8*25 socket head cap screws from the bottom in
alignment with the installation holes, tighten them with a wrench, then put the
cantilever in the support, and fasten it with screws around:
Cantilever
②
Fastening
screw
Support
①
Base
23
Colposcope System user’s manual
Cradle head
Cantilever
support
Figure 2
②Damping
adjusting screw
②Damping ①
adjusting screw
Figure3
24
Colposcope System user’s manual
装
⑤4 M6*10 hexagon socket ⑧4 cross recessed round
countersunk head screws 到 head screws
①Remove this
piece first
⑨4 M4*10 cross recessed
countersunk head screws
④ 8 M6*10 hexagon socket
countersunk head screws
②5 M6*16 hexagon
socket head cap screws
Figure 1
25
Colposcope System user’s manual
Display
Camera
device
Cradle
head
Computer
Place the
mouse
Fix the computer
with binding
bands
Printer
Wireless receiving
box (optional)
Figure 2
26
Colposcope System user’s manual
After installation of the display, if it feels too loose or tight when rotating, adjust
two adjusting screws with the attached wrench clockwise for tightening or
counterclockwise for loosening till the display can both be easily rotated and fixed at
any position.
Adjusting screw
Figure 3
②Damping
adjusting screw
②Damping ①
adjusting screw
Figure 4
27
Colposcope System user’s manual
Figure 1
2. Unscrew the installation plate locking knob, and remove the installation plate of
camera device on the cradle head:
Installation plate of
camera device
Installation plate
locking knob
Figure 2
3. Install the installation plate of camera device on the fixing plate at bottom of the
camera device as shown in Figure 3, tighten the screws as shown in Figure 4, and
then lay the pull ring flat after installation, of which the effect is as shown in Figure 5:
Installation
Pull ring
plate
Fixing plate
Figure 3 Figure 4 Figure 5
4. Reinstall the installation plate of camera device in the cradle head, and tighten the
installation plate locking knob.
Figure 6
28
Colposcope System user’s manual
29
Colposcope System user’s manual
30
Colposcope System user’s manual
7
d P
E1 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m). b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, ashould
be less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the KN-2200Ⅰ/KN-2200Ⅰ(H) Colposcope System is used exceeds the
applicable RF compliance level above, the KN-2200Ⅰ/KN-2200Ⅰ(H) Colposcope System should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the KN-2200Ⅰ/KN-2200Ⅰ(H) Colposcope System.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
31
Colposcope System user’s manual
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
32