IARP Lecture Notes RSO-RA 2021
IARP Lecture Notes RSO-RA 2021
on
Lecture Notes
ORGANISED BY
8.1
Contents
1. Radiation Physics
2. Interaction of Ionising Radiation with Matter
3. Radiation Quantities & Units
4. Principles of Radiation Detection and Monitoring Devices
5. Biological Effects of Ionising Radiation
6. Operational Limits
7. Radiation Hazards Evaluation and Control
8. Planning of Radioisotope Laboratories
9. Regulatory Aspects of Radiological Safety
10.Disposal of Radioactive Waste
11.Transportation of Radioisotopes
12.Production of Radioisotopes and Labelled Compounds
8.2
1. RADIATION PHYSICS
ATOMIC STRUCTURE
Matter is built up of individual entities called elements like hydrogen, oxygen,
carbon, etc. For example, water is made up of hydrogen and oxygen. An atom is the basic
component of each element. All atoms of a given element are identical, for instance, all
hydrogen atoms have the same properties.
An atom has a nucleus at its center. It is positively charged. The nucleus of
an atom is orbited by electrons, which are negatively charges. Electrons occupy one
or more orbits around the nucleus. The atomic nucleus is built up primarily of protons
(+vely charged particles) and neutrons (which carry no electric charge but have
almost the same mass as the protons). The positive charge of a proton is equal in
magnitude to the negative charge of an electron. The number of electrons orbiting around
the nucleus is always equal to the number of protons in the nucleus. Hence, the
atom is electrically neutral. The structure of an atom may be visualized as something
similar to that of our solar system with the sun at the center and the planets representing
the electrons orbiting around it.
8.3
increasing number of protons, neutrons, and electrons. Hydrogen is the simplest of the
elements. It has only one proton (and no neutron) in the nucleus and one electron in its
orbit. Hydrogen is followed by helium, lithium, beryllium, boron, etc.
The number of protons in a nucleus is called the atomic number ‘Z’ of
an element which determines its place in the periodic table and its chemical properties.
The total number of protons and neutrons in the nucleus constitutes the
mass number ‘A’ of the atom and it approximates its atomic weight.
An element is specified by using its chemical symbol, the mass number, and
atomic number,
A 1 6
Z X : For example 1 H (Hydrogen), 3 Li (Lithium)
This indicates that hydrogen atom has only one proton in its nucleus and no
neutron and lithium has three protons and three neutrons.
ISOTOPES
All atoms of the same element have the same number of protons and electrons and
thus the same atomic number. They can, however, have different numbers of neutrons
and are then called isotopes of that element. Thus isotopes of an element have the same
atomic number, but different mass numbers. Isotopes of an element are chemically
identical.
59 60
1 2
Example: 1 H , 1 H ,1 H
3 *
27Co, 27 Co * *
radioactive isotopes
Isotopes of Hydrogen Isotopes of Cobalt (radioisotope)
p
p 2n
p,n
8.4
RADIOACTIVITY
The stability of an element depends upon the ratio of neutrons to protons in
the nucleus. In elements of lower atomic number, except hydrogen, this ratio is generally
one. In heavier elements i.e., elements having higher atomic number, this ratio tends
to be more than unity, resulting in some degree of instability in the nucleus. An
unstable nucleus, which has excess energy, tries to attain stability by releasing this energy
in the form of radiation. Henry Becquerel, a French Physicist, discovered in 1996, that
a compound of uranium emitted some invisible radiant energy. Madame Curie studied
a large number of such substances and gave the name “Radioactivity” to this property.
Thus, radioactivity is a phenomenon in which an unstable nucleus of an element
disintegrates with the emission of energy and becomes a new element. It is a spontaneous
process unaffected by physical and chemical agents. Radioactivity exists in nature among
heavier elements. However, many lighter elements can be rendered radioactive by
bombardment with charged particles or neutrons. This is called artificial radioactivity.
Isotopes of an element having radioactive property are known as “Radioisotopes” (Refer
figure 1.2).
8.5
produced alpha emitters. The resulting daughter nucleus after alpha decay has an atomic
number less by two and a mass number less by four than that of the parent nucleus. Alpha
emission may be followed by gamma emission. All alpha particles emitted from a
particular decay scheme have the same energy.
A
Thus alpha () decay from an element Z X will be as follows:
A
Z X ZA42Y 24He e.g.
226
88 Ra 222
86 Rn 2 He
4
Gamma
Rays
Beta particles
Alpha particles
8.6
1
0 n11 p 10 e (-) decay when n/p ratio is high in a nucleus
3. Electron Capture
This is an alternative process to positron emission. In this mode of decay, the
proton is changed into a neutron after capture of an electron, generally from the
8.7
innermost k-orbit. Here, the atomic number of the product nucleus is reduced by one unit
from that of the parent. As it will have one electron missing from the orbit, an electron
from the outer orbit will subsequently fall into the inner orbit accompanied by the
emission of radiation (X-rays) characteristic of the daughter atom.
RADIOACTIVE DECAY
The decay of a radioactive sample is statistical in natural and it is impossible to
predict when a particular atom will disintegrate. The radioactivity in a sample decreases
constantly with time in a manner shown in figure 1.4 and follows an exponential law
which can be mathematically represented as
Ao Ai .e .t
Where A0 is the initial activity, At is the activity after a lapse of time ‘t’ and ‘’ is the
disintegrations or decay constant (fractional disintegration per unit time), which is a
characteristic of the radioisotope.
HALF-LIFE
In actual practice, the rate of decay of a radioisotope is usually in terms of its half-
life, the time required for one-half of the atoms originally present to decay. Physical half
life of some of the radio nuclides used in research laboratories are given in table-1.1. The
half-life is usually denoted by t1/2 and is expressed by relation:
8.8
0.693
t1 / 2
Radioactivity
Half-life can be read directly from decay graphs as shown in figure 1.4. It may be
noted that in a time equal to one half-life, the radioactivity will be reduced by a factor of
2; at the end of second half-life, the activity will be reduced by a factor of 4 and so on.
Theoretically, it requires infinite time for the complete decay of a radioisotope, although
in most cases, a time of about 10 half-lives will reduce the activity to a negligible value
(1/1024) compared to the initial activity. Like, , the decay constant, half-life is also
characteristic of a radioisotope. This is because there is a relationship between and half-
life, given by the relation
0.693
t1 / 2
For different radioisotopes, the half-lives are different varying from fraction of a
second to several thousand years. Measurement of the half-life of a radioactive sample
8.9
helps in the identifications of the isotope. It should be noted that half-life of
radioactive isotope does not depend upon the amount of radioactivity.
---------
8 . 10
2. INTERACTION OF IONISING RADIATION WITH
MATTER
INTRODUCTION
Radiation emitted during the radioactive decay is characterized by their kinetic
energy. Their energies are expressed in units of electron volts (eV). An electron volt
corresponds to the kinetic energy acquired by an electron when it is accelerated through a
potential difference of 1 volt. Electron volt is a small unit. The energy of different
radiations emitted during radioactive decay (and X-rays) will be much higher and hence
expressed in terms of kilo (103) electron volts (keV) or million (106) electron
volts (MeV). In comparison, the energy of visible light is in the range of 1-4 eV.
Radiation emitted by a radioisotope cannot be seen or felt by any human senses.
However, when they are incident upon matter they interact with the atoms to produce
excitations and ionizations. Excitation is a process in which the orbital
electron of an atom is raised to a higher energy state, while in ionization one or
more electrons are removed from an atom, forming an ion pair, viz., a positive ion
and negative ion. Both these processes lead to transfer of energy from radiation to matter.
Because of their ability to ionize matter, these radiations are called ionising
radiations.
Ionisation is finally responsible for the observed biological, chemical, and
physical effects of radiation. They are also the means of detecting and measuring these
radiations. In this chapter, the interaction of ionizing radiation with matter is briefly
discussed.
Ionising radiations can be divided in to electromagnetic and particulate radiations.
Particulate radiations can be further divided in to charged and uncharged particles.
8 . 11
leads to excitation and ionization along the path of a charged particle in the medium.
If the energy transferred to an electron in the medium is sufficient to remove it
completely out of the atom, the process is referred to as ionization. If the electron is just
raised to a higher energy state, the process is called excitation. Figure 2.1 shows
schematically the process of ionization by an alpha particle.
α- particle
Electron
α particle
The number of ion pairs produced per unit path length of charged
particle is called specific ionization. The specific ionization is directly
proportional to the mass and charge of the particle and inversely
proportional to its velocity.
Energy absorbed in the medium per unit path length of the particle is called its
Linear Energy Transfer (LET). It is usually expressed in keV/m. The concept of LET is
important as biological effects depend on the rate of energy absorption in the medium.
Since alpha particles are doubly charged and are of comparatively heavy mass,
they have a high specific ionization. Hence alpha particles lose energy in matter,
relatively rapidly, by these processes. As alpha rays from a given radionuclide are all
emitted with the same energy, they will have approximately the same range in a given
material. The range of an alpha particle is usually expressed in cm of air.
Beta particles, however, have only 1/7300 of the mass of alpha particle
and have only unit charge. Therefore, beta particles on entry in any material will be
scattered more and thus will have a tortuous path. They penetrate further into the
8 . 12
material. Beta particles have lower specific ionization and LET than alpha
particles. The range of beta particles in any medium is a function of the beta particle
energy and of the density of the medium. An empirical relation between the range (in
g/cm2 of aluminium) and energy (in MeV) of electrons of energies more than 0.6 MeV is
given by the relation.
Rmax(g/cm2) = 0.53 Emax – 0.106
Radiative Collision
When a fast moving charged particle passes close to a nucleus, it undergoes
deflection and loses energy in the form of electromagnetic radiation (Fig.2.2). Radiation
thus emitted is known as bremsstrahlung. Continuous X-rays from an X-ray equipment
are produced by this process. Production of bremsstrahlung is an important consideration
in the shielding of high energy beta particle. The intensity of bremsstrahlung
increases with the atomic number of the medium and decreases with
increase in the mass of the particle. Hence, energy loss by radiation is more
important in heavy elements than for light particles such as electrons.
Beta Particle
hν
Nucleus
Bremsstrahlung
8 . 13
is very small. An alpha particle of energy 5 MeV has a range of less than 5
cm in air. Even a thin sheet of paper can stop these particles and they cannot penetrate
the human skin. Since beta particles have less mass and lower charge, they lose much less
energy per collision than alpha particles of the same initial kinetic energy and hence they
have a larger range. A beta particle of energy 1 MeV has a range of nearly 40
cm in air and 0.5 cm in tissue.
ELECTROMAGNETIC RADIATIONS
These are wavelike disturbances, which arise in association with vibrating electric
charges. Radio-waves, infrared radiations, visible light, ultraviolet radiation, x-rays and
gamma rays, etc., belong to the class of electromagnetic radiations. They differ only in
their wavelength or frequency. Wave length and frequency are related to each other
by the relation.
C =
where is wavelength in cm, is frequency in hertz and C is the velocity of
electromagnetic radiation (3 x 108m/sec in vacuum). Energy of electromagnetic radiation
is given by the relation,
E = h
where h is a constant called Plank’s constant.
Among electromagnetic radiations only X-rays and gamma rays have sufficient
energy to ionise matter.
Gamma rays and X-rays (also called photons) interact with matter in a variety of
ways, the three main processes of which are: 1) Photoelectric absorption, 2)
Compton scattering, and 3) Pair production.
8 . 14
probability of photoelectric effect decreases with increase of the energy of the photon, but
increases with the atomic number of the medium. Hence for high atomic number
materials such as lead, at energies lower than 100 keV, the predominant interaction is
photoelectric effect. Figure 2.3(a) shows schematically the photoelectric process.
A photon of energy E will release an electron with kinetic energy Ee = E-,
where is the binding energy of the electron in that particular orbit.
Some salient features of photoelectric process
1. Photoelectric process involves bound electrons.
2. The probability of ejection of an electron is maximum when the photon energy is
just higher than the binding energy of the electron.
3. The photoelectric absorption coefficient varies with energy, approximately as
1/E3.
4. Photoelectric absorption coefficient varies approximately as the third power of the
atomic number of the absorber (Z3).
5. As the photon energy increases there is greater probability for photoelectron to be
ejected in the forward direction.
Compton scattering
Compton process involves transfer of a part of the energy of the incident
photon to a free electron. The outermost electrons of an atom, which have very low
binding energies, are considered free electrons. Since Compton scattering
involves these free electrons, the process is independent of the atomic number of
the medium in which the interaction takes place. Since many materials have
approximately equal number of electrons per gram (3 x 1023), absorption by this process
is nearly equal for equal masses of such materials. The photon transfers only a part of its
energy to the electron and gets scattered with reduced energy [Figure 2.3(b)]. The energy
given to the Compton electron is ultimately absorbed in the medium.
Some salient features of Compton interaction
1. This process involves a photon and a free electron.
2. Mass attenuation coefficient for the process is independent of Z of the
medium.
8 . 15
3. In this interaction, probability decreases with increase in energy of the incident
photon.
4. In this interaction, some energy is absorbed and the rest is scattered depending
upon the angle of scattering.
5. In soft tissues, in the energy range of 100 keV to 10 MeV, this interaction is
more predominant than the photoelectric or pair production
processes.
6. As the energy of the incident photon is increased, the electron will be ejected
more in the forward direction and it will carry larger portion of the energy.
Pair Production
In the intense electric field close to a nucleus an energetic photon may be
converted into a positron-electron pair. This is known as pair production and is an
example of conversion of energy into matter, which is shown schematically in figure
2.3(c). The minimum photon energy required for this process to occur is 1.02
MeV and the excess photon energy is shared as kinetic energy between the electron and
positron. The positron at the end of its track would encounter an electron. The two
particles annihilate and produce two photons each of energy 0.51 MeV.
Probability of pair production increases with increasing photon energy beyond
the threshold and also with the atomic number of the material. However, at energies less
than 10 MeV it is not an important mode of interaction with most materials.
Some salient features of pair production
1. It is an interaction between a photon and the nucleus.
2. Threshold energy for this process is 1.02 MeV.
3. Probability for this process increases with the square of the atomic number
(Z2) of the medium.
4. This process increases with increase in energy of the photon.
Thus all the three interactions result in the photon energy being transferred to
electrons which subsequently lose energy to the medium as described above.
8 . 16
When gamma radiation traverses matter it can undergo all the three interactions
with varying probabilities. Part of it may be absorbed, part scattered and part may be
transmitted without undergoing any interaction. This is shown schematically in figure
2.4. The dominating process depends on the energy of the radiation and the nature of the
medium. There is a certain probability for the occurrence of each process and this
probability is referred to as photoelectric, Compton and pair production
attenuation coefficients. The total attenuation coefficient is the sum of these three
coefficients.
Incident Photon
Compton
Electron
Incident
Photon Scattered
Photon
Photoelectron
(b)
(a)
Pair Production
h 0.51 MeV
Incident Photon
(c)
h 0.51 MeV
The amount of radiation transmitted through matter decreases with thickness and
can be described by the exponential relation,
8 . 17
I = I0 e-x
Incident photon Fluence Absorber
Transmitted Photons
Fluence
Scattered Photons
Collimator
8 . 18
2. In the medium energy region, the major interaction is Compton process.
3. The pair production starts at 1.02 MeV and increases with energy.
4. There is a dip in the curve, which corresponds to a transition from decreasing
predominance of Compton interaction and increasing predominance of pair
production process.
10
Mass Attenuation Coefficient (cm2/g)
Lead
1.0
0.1
Water
Fig. 2.5: Mass Attenuation Coefficient of water and Lead for different Photon Energy
8 . 19
3. RADIATION QUANTITIES AND UNITS
ACTIVITY, ‘A’
The activity, A of a radioactive material is a measure of its spontaneous
transformation. It is defined as the average number of spontaneous nuclear transformation
(or disintegration) taking place per unit time.
The special name of the unit of activity is Becquerel (Bq)
1 Bq = 1 disintegration per second = 1 dps
The old unit of activity is Curie (Ci)
1 Ci = 3.7 x 1010 disintegration per second
= 3.7 x 1010 Bq = 37 GBq
Both the indirectly ionising (photons and neutrons) and directly ionising (charged
particles) radiations transfer part or all of their energy when they interact with matter.
8 . 20
EXPOSURE, ‘X’
Exposure, ‘X’ is defined as the absolute value of the total charge of the ions of
single sign produced in air when all the secondary electrons (inclusive of positrons and
electrons) liberated by photons in air of mass m are completely stopped in air.
With the present technique, it is difficult to measure exposure when photon energies
involved lie above a few MeV or below a few KeV.
The unit of exposure in use is Roentgen, ‘R’, which is defined as that amount of x
or gamma radiation, which would liberate 1 esu of charge of either sign in 1 cc of air
at STP.
1R (Roentgen) = 1 esu of charge liberated per cc of air at STP
= 2.58 x 10-4 C Kg-1 (air)
Except at very high energies, the exposure defined above is the ionisation
equivalent of the air kerma. And by definition exposure is a quantity restricted to
photons and to air as the medium.
DOSE, ‘D’
The effects (physical, chemical and biological) of radiation depend not only on
the energy transferred to the medium, but also on the energy absorbed by it. The quantity
absorbed dose (or simply dose) is defined as the amount of energy absorbed per unit mass
8 . 21
EQUIVALENT DOSE, ‘HT’
Effects of radiation depend not only on the amount of energy absorbed, but also
on the spatial distribution of ion pairs. Hence, the biological damage caused by the same
dose of different radiations may be different if they have different rates of energy loss per
unit of path length, which in other terms referred as Linear Energy Transfer (LET).
Alpha particles, because of their high energy, charge and mass, cause greater ionisation
per unit path length than gamma radiations, which mediate through singly charged
electrons. One gray of alpha dose is found to be more effective than one gray of gamma
dose. Hence, in radiation protection, to account for this variation in the effectiveness of
different types of radiation, radiation-weighting factor (wR) is used to multiply the
absorbed dose due to each type of radiation. The weighted absorbed dose is called
equivalent dose HT.
i.e. HT = R DT,R wR
where, DT,R is the absorbed dose in tissue for radiation R of radiation weighting factor
wR.
Since wR is a dimensionless quantity, the unit of dose equivalent is also J
Kg-1 (J/Kg). Radiation weighting factor was formerly called quality factor (QF). The
special name for the unit of equivalent dose is Sievert (Sv).
For radiation protection purposes, 20 mGy of gamma dose, 1 mGy of alpha
dose and 2 mGy of fast neutron dose are equivalent.
It should be noted that this is true only for low equivalent dose levels such as
milli-sievert (mSv), centi-sievert (cSv) and applicable only for radiation protection
purposes.
Equivalent dose in Sv = (Dose in Gy) x (wR)
Formerly, the unit of dose equivalent was roentgen equivalent man (rem)
Dose equivalent in rem = (Dose in rad) x (wR)
In radiation protection 1 Sv is too large a quantity. Hence dose equivalents are
expressed in units of mSv (10-3 Sv). On the other hand, Bq is too small a quantity for
many applications. Hence, the amount of radioactivity is expressed in MBq (106 Bq),
GBq (109 Bq) etc.
8 . 22
Table - 3.1 shows wR values for different types of radiations. (The values of wR
reflect the Relative Biological Effectiveness, (RBE, a term used in radiobiology, for
different types and energies of radiation in production of stochastic effects)
8 . 23
Equiv. Dose rem Sv (Sievert) 1 Sv = 100 rems
Effective Dose rem Sv (Sievert) 1 Sv = 100 rems
If several tissues, T1, T2, T3, etc., individually receive equivalent dose HT1, HT2,
HT3, etc. then the total risk to the individual should not exceed that resulting from the
stipulated dose limit to uniform whole body irradiation. A number of organs are
considered on the basis or their sensitivity and the seriousness of the damage. Risk
factors are age and sex dependent. Depending on the extent to which the risk from
stochastic effects in a tissue/organ may contribute to the total risk from stochastic effects,
a weighting factor, called the tissue-weighting factor, wT is assigned to each
tissue/organ. Thus the effective dose (E) is defined as,
E = T HT . wT Sieverts
wT represents the contribution of tissue T to the total risk due to stochastic effects
resulting from uniform irradiation of whole body. (Table 3.3)
8 . 24
Tissue /Organ Tissue Weighting Factor, (wT)
Gonads 0.20
Bone marrow 0.12
Colon 0.12
Lung 0.12
Stomach 0.12
Bladder 0.05
Breast 0.05
Liver 0.05
Oesophagus 0.05
Thyroid 0.05
Skin 0.01
Bone surface 0.01
Remainder* 0.05
* Other tissue/organ which are not included in the table
------------
8 . 25
4. RADIATION DETECTORS & MONITORS
Radiation measuring instruments are needed to detect and quantify two types of
potential exposure: external exposure due to penetrating radiation emitted by the
radioactive sources lying nearby at the workplaces; and internal exposure from the
radioactive material that has got entry into the human body inadvertently while working
with it.
There are four basic types of radiation measuring (monitoring) instruments that
are used in a research lab handling radio-nuclides:
a. Dose rate meters: used to measure the potential external exposure rate
b. Dosimeters: used to measure cumulative external exposure
c. Surface contamination meters: used to measure level of radioactive
contamination on surface to evaluate potential internal exposure when
radioactive substance is distributed over a work surface
d. Airborne contamination meters: used to measure level of radioactive
contamination in air to evaluate potential internal exposure when a
radioactive substance is distributed within an atmosphere
8 . 26
All radiation monitoring instruments consist of following key components:
(a) Detector: The detector contains a medium, which absorbs radiation energy
and converts it into a signal. The signal can be electric charge, light,
chemical change etc. The medium generally used for radiation detection
are –
i. Gases (Ionisation Chambers, Proportional Counters, Geiger-
Muller Counters)
ii. Scintillation Media [NaI(Tl), Anthracene etc.]
iii. Photographic Emulsions [Film]
iv. Solid State Detectors [Semiconductors,
Thermoluminescent Phosphors]
(b) Amplifier: The signals from a detector need to be electronically
amplified
(c) Processor: According to the type of instrument, the processor is a
device to measure the size or number of signals produced in the detector.
It may also translate the quantity measured into appropriate radiological
units
(d) Display: The measurement is presented either in a digital format or an
analog display by a pointer on a graduated scale.
8 . 27
proportional counter, and Geiger-Mueller (GM) counter.
gas and size of the chamber depends on the purpose of use and radiation
intensity to be measured. The ionisation chambers used for radiation monitoring are
mostly filled with air at atmospheric pressure. These detectors are mostly used for
ionisation chambers are often used for personnel monitoring such as pocket dosimeter.
electrodes is higher than that applied in ionization counter. In this counter, the initial
Radiation
(+ve) Central
Electrode Displ
ay
1.56 mGy/h
Insulator High
Voltage
(-ve)Outer Electrode
Electro Positive
n causes further (secondary)
ionisation produced in the gas due to radiation ion ionisation
because of the existence of higher electric field across the electrodes. Therefore, larger
8 . 28
number of ion pairs are produced and make the counter more sensitive than ionisation
chamber (counter). In addition to its use as a dose-rate meter it can also be used for
energy discriminations.
advantages of this detector are : available in wide variety of shapes, large signal output
counter display in case of GM survey meter gives the exposure-rate, whereas the display
in case of contamination monitor gives the count-rate. These instruments are used
for measuring low radiation level in the range of 2 - 200 µSv/h. It is very useful for
radiation survey of the laboratories. These are not useful in high radiation field or to
SCINTILLATION DETECTORS
The gas filled detectors register the ionisation produced by radiation in gas. The
fluorescence (light flash). The light flashes (fluorescence) produced are used for
Scintillation detectors are usually in solid form. But in some applications the
scintillators are used in liquid form also. NaI(Tl), CsI(Tl), ZnS(Ag), Anthracene
8 . 29
and Plastic scintillators are some of the examples of solid scintillation detectors.
NaI(Tl) and CsI(Tl) detectors are used for detecting gamma radiation,
whereas ZnS(Ag) is used to detect alpha radiation and the Anthracene and
The light flashes (scintillation) produced are converted in electric pulses (using
Photo Multiplier Tubes) and then fed into suitable electronic circuitry where the
scintillations are discriminated to analyse different types of radiation and even the
8 . 30
Radiation interacts with the molecules of scintillators in the manners described in
chapter-2.
The organic scintillators used in LSC have very high absorption coefficient for
low energy beta, alpha and gamma radiation of energy <20 keV, but very low for
energetic (>20 keV) gamma radiation (refer chapter-2). If energy of gamma radiation is
>300 keV and mass number (Z) of interacting medium is low, as is the case with LS, the
probability of photoelectric effect is very small. Thus the probability of absorption of
gamma radiation in LS by Compton scattering process is very high. As a result the
detection efficiency of energetic gamma radiation will be very poor. For gamma radiation
of energy <20 keV, the energy absorption by photoelectric effect process in LS would be
very high, resulting in high detection efficiency. The counting efficiency for gamma
radiation emitted by 125I can be as high as 76% in a typical emulsifier type LSC.
SEMICONDUCTOR DETECTORS
The solid state detectors, such as silicon and germanium detectors, are mainly
used for gamma spectroscopy (energy spectrum or discrimination). These detectors have
good energy resolution (4 eV). This is because the average energy required to produce
an electron-hole pair in semiconductors is about 3.5 eV as against 35 eV energy to create
ion pair in air. As a result larger number of electron hole pairs would be produced in
semiconductor as compared to number of ion pairs produced in air for same gamma
radiation energy.
Where, the energy resolution of gamma radiation is high in (germanium)
semiconductors, the detection efficiency is low. These detectors are mainly used to
identify gamma emitting radio nuclides in unknown samples by resolving radiation
energy emitted.
8 . 31
material is heated may be measured and related to the absorbed dose in the material. The
material thus exposed and heated, can be reused after proper annealing process.
Many thermo-luminescent materials like LiF3, Al2O3, CaF2, CaSO4:Dy have been
studied. Of these CaSO4 : Dy has been found to be very useful for dosimetric purposes,
due to its high sensitivity, low fading. indigenous production and many other useful
characteristics.
Since the effective atomic number of Li F is comparable to that of tissue and air,
it is found to be almost energy independent. This phosphor is also widely used in both
personnel monitoring as well as in other dosimetric applications. Other TLDs such as Ca
S04 : Dy, Al2 O3, have higher effective atomic numbers and hence show significant
energy dependence to X and gamma rays. Hence metal filters have to be used to
compensate the dependence of response on photon energy. The use of filters also permits
the estimation of energy of photons below 200 keV.
CHEMICAL DETECTORS
Photographic Emulsions Films
8 . 32
metallic filters.
8 . 33
example, if the radiation being measured is gamma radiation, it would be exposing the
whole body. If the radiation is a beta radiation it would be exposing the external skin. In
earlier situation it would give whole body equivalent dose-rate whereas in the later one it
would be shallow dose rate and not a whole body equivalent dose-rate. The knowledge of
radiation would allow using appropriate radiation shield to eliminate hazard. If the
radiation is pure alpha radiation, there is no external hazard but one have to take
precautions to avoid ingesting alpha emitting radioactivity. Generally, in mixed radiation
fields, numbers of different instruments are used to measure the intensity of each
radiation present.
radiation field of known intensity. The calibration of survey meters may be done with the
help of a standard source of reasonably high activity (137Cs in MBq strength). The survey
object nearby and the exposure rate is recorded. The expected value of exposure rate is
calculated using specific gamma ray constant for that radionuclide (Appendix -4 table 1).
The ratio of the expected value to measured value gives calibration factor. Routine
calibration commonly involves one or more sources of a specific radiation type and of
sufficient activity to provide calibration on all ranges of concern. Before calibration the
broken parts, proper operation of switches and zero adjustment. Radiation Standards
Section (RSS), Radiation Safety System Division (RSSD), Bhabha Atomic Research
Centre (BARC), in our country, carries out calibration of survey meters. It is desirable
that all the survey instruments are calibrated periodically (at least once in three years or
8 . 34
In the interval between calibrations, however, the instrument user should validate
establish whether the instrument is functioning within specified limit (i.e. within 20% of
the expected value). This is done by measuring the response of the instrument to a
radioactive source of known activity (137Cs or 60Co of the order of a few kBq activities)
under a specified geometry. The initial performance check should be carried out
the same source and same geometrical condition should be employed so as to correlate
the earlier findings. The user of survey instrument should have adequate knowledge of
preventive maintenance of the instrument, which includes the removal of batteries and
-----------
8 . 35
5. BIOLOGICAL EFFECTS OF IONIZING RADIATION
INTRODUCTION
The interaction of ionizing radiation with human body, arising either from external
radioactive sources (outside the body) or internal contamination, leads to biological
effects which may later show up as clinical symptoms. Both radiobiological
investigations and human exposure to ionizing radiation have contributed to our present
knowledge in this area. Radiobiological data have been derived mostly from micro-
organism, cultured mammalian cells and whole animal systems. Human data have been
derived from the follow-up of the (a) survivors of atomic bomb explosions in
Hiroshima and Nagasaki, (b) inhabitants of Marshall islands, who were
exposed to fallout from thermonuclear devices, (c) uranium miners, (d)
radium dial painters, (e) pioneer X-ray technicians and radiologists, (f)
patients exposed to radiation for medical reasons; and (g) victims of
nuclear accidents. A careful analysis of these data has yielded reasonable quantitative
estimate of biological effects of radiation in man. Some of these are briefly described in
this chapter.
THE CELL
Cell is the basic unit of all living organisms. A living organism may be made up
of a single cell or many cells; e.g. the organisms which cause diseases like typhoid and
tuberculosis are single celled. Man is a multi-cellular organism having about 1014 cells.
The structure of a typical animal cell is shown in figure 5.1. It has an outer envelope
8 . 36
called cell membrane or plasma membrane. Inside the cell there is nucleus in the
centre. Outside the nucleus is a viscous liquid called cytoplasm. The cytoplasm
contains structures like mitochondria, golgi bodies, lysosomes, etc., which perform
important cellular functions.
The nucleus contains chromosomes which are tiny thread-like structures made
up of deoxyribonucleic acid (DNA) and protein. DNA molecules contain in-coded
form (as sequence of bases), all the information required for the cellular function and thus
control the growth, development and well-being of the individual. Sections of
chromosomes which contain information for specific functions are called genes.
The size and shape of the cells are different in different parts of the body,
depending upon their function. Cells of similar nature are organized together to form a
tissue. Different types of tissues join to form an organ. Different organs form a system;
e.g. respiratory system consisting of nose, wind pipe and lungs; digestive system
consisting of mouth, food pipe, stomach and intestine. The cell can be compared to a
factory. The nucleus is .the control room. So, if the control room is damaged, work in the'
cell factory' slows down or stops.
Cel1s can be grouped into two categories: (1) somatic cells and (2) germ
cells. Somatic cells constitute various tissues such as brain, kidney, skin, liver etc. Germ
cells, also called reproductive cells, are those which participate in the reproductive
process. They are sperms in males and ova in females. All somatic cells in human
body contain 46 chromosomes which occur in 23 pairs. The germ cells contain
only 23 chromosomes - single copy of each pair. Figure 5.2 shows the human
chromosomes (male).
8 . 37
directly by the other bio-molecules. The excited and ionized water molecules undergo a
series of reactions. Radiation on interactions with water produces the radiolytic
products of water such is H, OH, HO2, eraq, O2 and H2O2. These free radicals react
readily with bio-molecules in the cell and result in damage to important bio-
molecules such as DNA and proteins. Such damages may lead to (a) inhibition of cell
division, (b) chromosome aberrations, (c) genes mutation, and (d) cell
death. While the absorption of radiation energy in the organ
8 . 38
Fig. 5.3: Human chromosome (Lymphocyte) with aberrations
(d: dicentric, Ace: acentric fragments, r : ring)
b. Chromosome aberrations: Radiation can cause breaks in chromosomes. Majority of
the breaks may get repaired and the damage may not manifest. However, certain
breaks may lead to rearrangements of genetic material which can be seen under a
microscope. Such events are called chromosomal aberrations (Figure 5.3), The frequency
of various types of di-centric chromosome aberrations can be correlated to dose and
hence can serve as a biological dosimeter.
c. Gene mutation: Alterations in the information content of genes (DNA) are known as
gene mutation. Damage to chromosomes (Chromosome aberrations) may lead to
change the information content of DNA.
d. Cell death: Irradiation can lead to cell death as a result of any or all of the above
effects. Cell death is usually expressed as fraction of cells surviving after a given
exposure. The effect of dose on mammalian cell survival is shown in figure 5.4(a). At
low doses the survival curve has a broad shoulder where cell survival decreases slowly
with dose. This is attributed to the ability of cells repair radiation damage. However, at
higher doses, where repair capacity is saturated, the survival decreases rapidly with
increasing dose, in an exponential manner. Figure 5.4(b) shows the dose response of
human peripheral blood lymphocytes for the induction of chromosomal aberrations.
Depending upon the type of radiation it may increase linearly or nonlinearly with
increasing dose. A typical metaphase spread of human lymphocyte carrying
chromosomal aberration is shown in figure 5.3.
8 . 39
Fig. 5.4: (a) Effect of dose on cell survival. (b) Calibration curve for induction of
dicentric
BIOLOGICAL EFFECTS OF RADIATION IN MAN
The detrimental effects of radiation in human body are produced as a result of
damage to the individual cells. These may be divided into two classes, namely somatic
effects and hereditary effects. The somatic effects arise from damage to cells in a
particular irradiated tissue and affect the irradiated person only. The hereditary effects are
due to damage to germ cells which may manifest in the progeny of the irradiated
person.
8 . 40
higher than 3 Gy, irrecoverable damage occurs to
the blood forming organs (bone marrow, spleen,
lymph node etc., with the possibility of death in a few
weeks time. In the dose range of 3-5 Gy about 50% of
the exposed persons may die within 60 days
(LD50/60). Anemia, infection, and high fever are the main
symptoms. These symptoms are called Haematopoietic
Syndrome.
At higher doses, in the range of 7 to 10 Gy, cells in the gastrointestinal
system get severely damaged leading to diarrhoea, loss of appetite, dehydration,
electrolyte imbalance, weight loss and high fever. These symptoms are typical of
Gastrointestinal Syndrome (GIS). Death occurs in 7 to 14 days time.
As the dose is increased further, consequences of damage to central nervous
system manifest. At doses in the range of 25 Gy and above, the damage becomes
so severe that depression, fatigue, delirium and comma appear and death
occurs within a few hours to 2 days. These symptoms are known as Central Nervous
System Syndrome (CNS).
The quickness of occurring early symptoms such as nausea, vomiting etc and the
degree of severity is good indications of the level of exposure. Persons exposed to doses
in the range of LD50/60 exhibit these symptoms within one or two hours.
The nature and seriousness of the early effects also depend upon whether the
exposure is to the whole body or part of the body. Even though whole body irradiation to
3-5 Gy may kill 50% of the exposed persons, the same dose given to a part of the
body will cause only local effects. Some of the local effects are reddening of the
skin (erythema), hair falling off (epilation), temporary or permanent sterility (when
reproductive organs are exposed). Table 5.1 lists some of the early effects of whole body
and local acute exposures.
All the early effects exhibit a threshold dose below which the effects do
8 . 41
not occur. Beyond the threshold dose, severity of the effect increases with increase of
dose. Hence, these are referred to as deterministic effects. Because of the large
threshold doses, acute effects do not occur from exposures arising from
normal work with radiation. However, the lethal effects of radiation have been
exploited in a number of beneficial applications such as in industry and medical where
the radioactive sources in giga or tera Bequeral quantities are used for example:
sterilization of medical products, food preservation, treatment of cancer, and treatment of
effluent from municipal sewage.
Late effects: Persons who recover from early effects may still develop some other types
of effects later in life. Similarly exposure to low levels of radiation over long periods of
time, which cannot produce any early effects, may also lead to late effects. Late effects
are characterized by a long latent period, which can be as long as 30-40 years. The
important late effects are cancer, fibrosis in various tissues and cataract of the eye lens.
Cataract is progressive loss of vision. The transparent cells in the eye lens,
which facilitate vision, are killed by high doses of radiation. Since there is no blood flow
through lens, the dead cells are not removed. Accumulation of dead cells results in the
opacity of lens, which lead to the impairment of vision.
8 . 42
5-10 Gy Severity of above effects increases. Almost 100
percent death (at the higher dose).
Cancer results from viable cells which have received damage to their control
system in the form of gene mutations or chromosomal aberrations. Such damage may
lead to over proliferation of cells in an organ resulting in cancer. Cancer cells multiply
and grow in an uncontrolled manner. Animal experiments and human
epidemiological studies have shown that radiation, at high doses, is carcinogenic.
Cancer mortality data based on the follow-up of atom bomb survivors of Japan are
presented in tables 5.2 and 5.3. At high doses (> 0.2 Sv) there is a clear indication that the
leukemia incidence increases with bone marrow dose. However, the shape of the
dose-effect relationship at low dose levels is not well established. Because of
the high frequency of naturally occurring cancers, it is not possible to demonstrate
conclusively whether or not small doses of radiation are carcinogenic to man.
Epidemiological studies in high back-ground radiation. Areas around the world do not
reveal any increased risk of cancer to the population.
Similarly, among the survivors of atomic bomb explosion in Hiroshima and
Nagasaki, no statistically significant increase in leukemia incidence rate has been
reported in population groups exposed to doses below 0.2 Sv. However, to be on the
conservative side in radiation protection, the probability of radiation induced cancer is
assumed to increase proportionately with dose, without any threshold. Accordingly, any
dose, however small is taken to be associated with a certain level of risk.
8 . 43
This kind of effect is referred to as Stochastic which means ‘random or statistical
nature'.
A number of human population groups exposed to high levels of radiation have
been studied to estimate the risk of radiation induced cancer at low levels of exposure.
The current estimate, suggested by the International Commission on Radiological
Protection, is that there would be 50 excess cases of cancer per million populations if
exposed to 1 mSv of radiation dose. This implies that if one million persons (both men
and women) each receive a dose of 1 mSv, the number of fatal cancers attributable to
radiation would be 50 and these appear over a period of a few decades. To put these
estimates into perspective, it should be noted that in a typical population of 106 people,
there are 1 - 2 x 105 deaths from spontaneous cancers. .
8 . 44
500 types of human diseases are attributable to genetic
factors and nearly 10 percent of all new born children suffer
from spontaneous genetic disorders. It may be mentioned
that mutations induced by various agents like heat
and chemicals etc. are indistinguishable from that
induced by radiations. Table 5.4 gives a summary of the
life-time risk of radiation induced stochastic effects in man.
Table 5.2 : Observed and Expected Deaths for solid Cancers in Japanese A-bomb
survivors
1950-1990
Colon Subjects Observed Expected Excess
Dose (Sv) Deaths Background Deaths
0(<0.005) 36,459 3,013 3,055 -42
0.005 - 0.1 32,849 2,795 2,710 85
0.1 - 0.2 5,467 504 486 18
0.2 – 0.5 6,308 632 555 77
0.5 – 1.0 3,202 336 263 73
1.0 – 2.0 1,608 215 131 84
> 2.0 679 83 44 39
Total 86,572 7,578 7,244 334
Donald Pierce et.al (RERF update, vol 8, p.10)
Table 5.4 : Estimated Lifetime Risk of Radiation Induced Stochastic Effects in Man
8 . 45
(Chronic Exposure)
Effect Risk per million per mSv Spontaneous risk per
million
Cancer 50 1 x 105 – 2 x 105
Hereditary 10 10% of all live born
children
----------
6. OPERATIONAL LIMITS
INTRODUCTION
All human endeavors entail risk of some type or the other. In the same manner
ionizing radiation too pose a physical risk to people, who may be exposed by natural and
artificial means. The interaction of radiation with matter produces ionisation phenomena
capable of modifying the chemical behavior of molecules. If this occurs in live cells,
biological effects of varying degree of severity may result in. People who work with
sources of ionizing radiation, some members of the public and patients who undergo
radiological procedures are exposed to risk of ionizing radiation. It is not possible to
eliminate these risks totally, but it is feasible to control them within
acceptable limits by following the principles of radiation protection. Hence, it is
necessary to develop safety standards to work with radiation so that the risks are kept
minimum.
The International Congress of Radiology established the International
Commission on Radiological Protection (ICRP) in 1928. ICRP issued its recent
8 . 46
basic recommendations on Radiation Protection in 1991- 'ICRP Publication
No.60'. These recommendations led to the joint publication of the International Basic
Safety Standards for Protection against Ionizing Radiation and for the
Safety of Radiation Sources (BSS). The objective of protection from ionizing
radiation is to prevent from radiation induced deterministic effects and to reduce the
likelihood of stochastic effects by considering economic and social factors. There
are three basic principles that sum up the philosophy of radiation protection and these are
discussed below.
Limitations
JUSTIFICATION OF PRACTICE
OPTIMIZATION OF PROTECTION
Implementation of protection measures calls for value judgment. The greater the
level of protection, the higher the degree of safety achieved. At the same time, greater
8 . 47
levels of protection would involve expenditure, which may reduce the ultimate value of
practice. For example research laboratories carrying out in-vitro experiments with small
amount of radioactivity and week energetic radiation don't call for additional structural
shielding for assuring prescribed degree of safety. But the laboratories, where large
activities are handled/stored need adequate structural and local radiation shields.
Providing additional structural shielding against radiation, a greater degree safety may be
achieved but this measure would achieve a small increase in degree of safety for a large
increase in the expenditure. Hence, it is recommended that protective measures should be
optimized that the radiation exposures are kept As Low As Reasonably Achievable
(ALARA). This appropriate socio-economic consideration should be factored in
optimizing the level of protection.
DOSE LIMITATION
In order to reduce the magnitude of the risks associates with a justified practice,
limits on individual doses are established to prevent the occurrence of deterministic
effects and minimise the likelihood of stochastic effects. The
monitoring of these limits should take into account of doses generated by external
sources and those produced by the intake of the radio nuclides into body. The dose limits
applicable to workers and to members of the public are given in table 6.1 and discussed.
It must be noted that the exposures due to natural radiation background and
medical procedures are excluded from the dose limits.
8 . 48
body tissues and organs (since the equivalent doses in all cases are less than the threshold
for any deterministic effects) except the lens of the eye, which makes a negligible
contribution to the effective dose and the skin which may be subjected to localised
exposures. Separate equivalent dose limits are needed for these tissues. The annual limits
are 150 mSv for the lens and 500 mSv for the skin, averaged over any 1 cm 2 regardless of
the area exposed.
INTERNAL EXPOSURE
A person may receive radiation doses from external and/or internal sources. In
research application, where radio nuclides in unsealed forms are used, radio nuclides may
enter the body through injection, inhalation, wounds, or absorption by skin.
The dose limit is the same, irrespective of the exposure route-internal, external
or combination. Accumulation of internally deposited radio nuclides will depend on
radioactive contamination present in air and water, the rate of intake and metabolism.
Limits are therefore prescribed in terms of Annual Limits on Intake (ALI) for
different radio nuclides if the exposure is only due to intake of activity. ALI is based on a
committed effective dose of 20 mSv. Dose limits are only the part of a system of
protection aimed at achieving levels of dose that are as low as reasonably
achievable, economic and social factors being taken into account. It is not to be seen as
a target because; if received over a lifetime may lead to risk level verging on becoming
unacceptable. However, any isolated exposure exceeding the limit need not cause alarm,
but a thorough examination of the design and operational aspects of protection must be
conducted. If the dose is unknown, or is thought to be high, referral to a physician should
be considered.
If it is likely to receive radiation exposure from external source and from the
internally deposited radioactivity, both the exposures should satisfy the following
equation.
Intake E.E
1
ALI ADL
8 . 49
Where: Intake - means through both ingestion and inhalation
ALI – ALI both ingestion and inhalation
E.E – External exposure
ADL – Annual dose limit
The basis for the control of occupational exposure of women who are not
pregnant is the same as that for men and the commission recommends no special
dose limits for women in general. Once pregnancy has been declared, the conceptus
should be protected by applying a supplementary equivalent dose limit, to the surface
of the woman’s abdomen, (lower trunk) of 2 mSv for the remainder period of
pregnancy and by limiting the intake of radio nuclides to about 1/20 (0.05) of ALI.
8 . 50
Table 6.1: Dose limits
@
The limit prescribed by Atomic Energy Regulatory Board is 30mSv in a year
* Averaged over areas of no more than any 1 cm2 regardless of the area exposed. The
nominal depth is 7.0 mg cm-2.
** Averaged over areas of the skin not exceeding about 100 cm2.
risk and who might be exposed during the whole of their lifetime. In addition, they do not
make their own choice to be exposed and may receive no direct benefit from the
exposure. For planning purposes, it is considered appropriate to set limits for the
8 . 51
members of the public lower than those are for occupational workers. Hence, the limit for
provided the average over 5 years does not exceed 1 mSv per year. For preventing
deterministic effects in lens of the eye and skin, annual dose limits of 15 mSv and 50
achieved its twin objectives namely, eliminating the deterministic effects and reducing
Table 6.2 shows for some important deterministic effects, acute threshold doses
(single exposure) and the annual dose rates, if received in highly fractionated or
clinical symptoms would be observable. These levels can be compared with the presently
recommended annual dose limit for whole body as well as specific organs given in the
last two columns. It can be seen that the dose limits are below the ones required for
Further, the ALARA principle has assured that the exposures are generally kept
below the dose limits. As per the personnel monitoring data available with BARC, the
average annual exposures to all categories of radiation workers in India during the year
8 . 52
1998 is 1.49 mSv. This is about 1/30 of the maximum annual dose limit. As discussed
earlier, the risk of fatality from radiation induced cancer is 10-2 per Sv. On the basis, an
average annual exposure of 1.49 mSv would correspond to approximately 0.6 excess
fatality per year amongst the 40,000 radiation workers. This number is quite small
compared with, and indistinguishable from the total number of annual fatalities which is
nearly 800 (taking national annual fatality rate of 2%). Hence, it is evident that the
average exposure of a hypothetical member adds little to the hazards of every day living.
Radiation has always been a part of our daily life. We are constantly exposed to
radiation from both natural and artificial sources. Natural exposures come from the sun
and the stars (cosmic radiation), the earth’s minerals (uranium, radium, and rubidium)
and even from within our own body (40K, 14
C). As a result, the food we eat, the air we
breathe the water we drink and houses we live in, all contain traces of radioactivity.
Every inhabitant of the planet receives an average dose of 2.4 mSv per year (Table 6.3).
Table 6.2: Threshold doses for some deterministic effects of radiation for single (acute)
and protracted (chronic) exposures in comparison with annual dose-equivalent limits.
Tissue & Effects Total Equivalent Threshold Annual Dose Annual dose limit for
Dose received in rate if received yearly in
a single brief highly fractionated or Whole Single
exposure (Sv) protracted exposures Body organ
for many years (Sv/Y) Expo.(Sv) (Sv)
Testis :
0.15 0.4 0.02 0.1
Temp. Sterility 0.35 2.0 0.02 0.1
Perm. Sterility 2.5 - 6.0 >0.2 0.02 0.1
Ovaries: Sterility 5.0 >0.15 0.02 0.15
Lens : Cataract
Bone marrow: 0.5 >0.4 0.02 0.17
Depression of
Hematopoesis 1.5 >1 0.02 0.17
Fatal aplasia
8 . 53
At any given place, however, the individual contribution varies depending upon
its geology, latitude, altitude, etc. Some of the coastal areas in Kerala have natural
background radiation levels as high as 10 mSv or more. The dose limit for
background radiation. But the average occupational dose is of the same order as the
handled with care rather than with fear and its risks should be kept in perspective with
other risks. The procedures available to control exposures to ionising radiation are
sufficient, if used properly, to ensure the radiation remains a minor component of the
The most important source of artificial exposure (man made) is the use of
radiation in the medical field. More than 80% artificial radiation dose comes to people
from medical diagnostic procedures. The average background dose resulting from natural
radioactivity and man-made is estimated to be about 2-3 mSv. 87%of this is contributed
from natural radiation and remaining 13% from man-made sources. This is shown in
figure 6.1, a pie chart. The contribution from various components of the natural
radioactivity is shown in Table 3.
8 . 54
Table 6.3: Average World Wide Exposures to Natural Radiation Sources
Annual Effective Dose Annual Effective Dose
Source of Exposure mSv mSv
0.36
Cosmic rays 0.41
2.36 0.46
Total
Fig 6.1: Percentage Contribution of Natural and Man Made Radiation to Background
8 . 55
EVALUATION OF HAZARD DUE TO EXTERNAL RADIATION
External radiation (EH) hazards are caused when the radioactive source(s) or
waste lying near a work place or table (outside the human body). The chances of EH are
high from non-penetrating radiation such as X-rays and gamma rays on account of their
higher penetrating power. EH would less from beta and alpha radiation.
This k-factor is constant for that gamma radiation emitting radionuclide and it is
the Exposure rate in mGy/h at ‘1’ metre from a point source of activity ‘A’ GBq
emitting gamma radiation of energy ‘E’ MeV. If the activity A is expressed in old unit of
activity i.e. in ‘Ci’ the exposure rate would be = 0.52 A.E R/h
8 . 56
CONTROL OF EXTERNAL HAZARD
EH can easily be controlled by adopting the three fundamental methods namely
(1) Distance (2) Time and (3) Shielding, depending on the situation. Some times only
one of the three methods would be adequate to follow to control the EH. But in certain
situation one may have to adopt all the 3 methods referred above.
Distance: Radiation exposure varies as Inverse Square of the distance from the source.
The variation of exposure rate with distance is given by the simple relationship.
I 1 D22
I 2 D12
By knowing the exposure rate at a given distance D1, the exposure rate at any
other distance D2 can easily be calculated using equation referred above.
‘Larger the distance between radioactive source and radiation worker; lesser be
the radiation exposure rate’. It is, therefore, advisable to maintain optimum distance with
disturbing the main work with radiation source. To understand the application of inverse
square law following examples are discussed.
Example 1 : What would be the radiation level at 10 cm distance from a 10 MBq of
57
Co? Specific gamma ray constant for 57Co = 0.2 mGy/h (k-factor) at 1 cm distance.
Strength of the source (A) = 10 MBq
Exposure rate at 1 cm from to MBq source (I1) = 0.2 x 10 = 2mGy/h
Exposure rate at l0cm from the source = I2 = ?
I 1 D22
I 2 D12
Therefore
I 2 I1 D12 D22
8 . 57
Exposure rate constant for 60Co = 3.1 mGy/h/MBq at 1cm
Strength of the source = 20 MBq
Exposure rate at 1 cm from the source = I1 = 20 x 3.1 = 62 mGy/h
Required exposure rate = I2 = 0.025 mGy/h
I 1 D22
I 2 D12
Here, I1 = 62 mGy/h, I2 = 0.025 mGy/h, D1 = 1 cm, D2 = ?
Substituting these values in the equation, we get,
D2 = 49.8 cm
Example 3 : While handling 25 mCi 65Zn source with a 15 cm tongs, within how much
time the operator will receive the weekly permissible equivalent dose. (Given k for 65Zn
= 2.7 R/h/mCi at 1 cm Assume 1 R = 1 rad).
Exposure level at 15 cm from 25 mCi Zn-65 source = 2.7 25 15 15
= 0.3 R/h
= 300 mR/h
Weekly permissible exposure = 40 mrad
Allowable time = 40 300 h
= 8 min
Time: Radiation dose received during work is directly proportional to the time
spent near the source. The dose received by a person remaining near a source for 10
minutes would be only half of that received by him in 20 minutes. Therefore, all the
radiation work should be planned in advance and should be executed in optimum
minimum time so that the exposure received could be kept as low as reasonably
achievable (Chapter 6). In order to minimize the time of operation with the actual
radiation source, it is advisable to perform trial operations with a dummy source. Source
should be taken out from the storage only when required or to be used.
Example 4 : In example- 3, if the user takes 4 min. to work with the source, the weekly
permissible exposure limit will be received in two such operations. Instead, if a 30 cm
tongs is used them 8 operation can be carried out without exceeding the weekly exposure
limit.
8 . 58
Radiation Shield: Where it is impracticable to adopt the method of distance and time
to ensure acceptable low radiation level at work place; method of radiation
shield is adopted. In this situation the radiation source(s) is/are shielded locally with an
adequate thickness of proper radiation shielding material. The type of radiation shield, the
material, and its thickness will depend upon the nature of radiation, strength of the
radiation source and the energy of radiation emitted. Alpha sources do not pose any
external radiation hazard because of their limited range in medium (even a few
centimeters of air can cut off most of energetic alpha particles emitted from radioisotope).
Therefore the alpha emitting sources do not need additional shield. But the beta emitting
radioisotopes need additional external shield to minimize the radiation level. These
radiation can travel far more distances in air than those of alpha radiation. In case of beta
particles, a fraction of energy, absorbed in the shielding material, will be converted to
bremsstrahlung (Chapter 2), which is similar to X-rays. The fraction of energy converted
in to bremsstrahlung depends on the atomic number (directly proportional to Z2) of the
shielding material and the energy of the beta particle (directly proportional to E). Hence,
to minimize the production of bremsstrahlung radiation from shielding material, it is
preferred to use a material of low atomic number such as perspex, aluminum, etc. A
perspex shield of 1 cm thickness would absorb most of the beta rays emitted by any
radioisotope. However, when a beta emitting radioisotope of very high activity (say in
curie) is to be shielded from radiation, a secondary shield of (Pb) lead is required to
absorb the small amount of bremsstrahlung produced in primary perspex shield.
X-rays and gamma radiation are penetrating radiation and can travel far off
distances in air, reducing intensity by inversely square principle. The radiation exposure
due to gamma radiation is brought down using high Z (atomic number) material as shield
such as Pb. Those materials are used as shielding for gamma where the process of
photoelectric effect is highest. The gamma radiation interact with shielding material by a)
Photoelectric, b) Compton and c) Pair production processes (Chapter 2). For medium
energetic gamma radiation, the photo electric effect in high in lead (Pb). Material of
higher Z such as tungsten, depleted uranium will be better shield than Pb for higher
energetic gamma radiation. Due to economic reasons concrete, brick, steel (low z
8 . 59
materials) are also used as structural shielding material to control radiation from X-rays
and gamma rays. The reduction in radiation level (Intensity or flux) of X and gamma rays
using shielding material is expressed by the following exponential equation:
I = Io e- .x
Half Value Thickness (HVT) : The half value thickness is that thickness of the
shielding material which will reduce the incident intensity of radiation beam to half. The
reduction factor offered by one HVT of the shielding material is 2 and by 2
HVT is 2 x 2 or 22. The reduction factor offered by ‘n’ HVT of shielding
material is 2n.
Tenth Value Thickness (TVT): Another specific thickness which is convenient to use
in shielding problems is the Tenth Value Thickness (TVT). TVT is that thickness which
will reduce the incident radiation intensity to one tenth. The reduction factor offered
8 . 60
by one TVT is 10, by two TVT is 10 x 10 or 102, by three TVT is 103 and so
on. Similarly the reduction factor offered by ‘n’ TVT is 10n.
For a given material both HVT and TVT depend upon the energy of the incident
radiation. The HVT and TVT values can be directly estimated from the attenuation
graphs figures 7.1 2a and 2b. The HVT, and TVT values for gamma rays of different
energy originated from different radio nuclides are given in table 7.3.
Fig. 7.1: Exposure attenuation. T½ = 2mm. (a) Linear Plot, (b) Semi logarithmic plot
(i.e. the vertical scale is logarithmic and the horizontal scale linear)
AREA MONITORING
The assessment of radiation levels at different locations in the vicinity of radiation sources is generally known as area
monitoring. Normally, area monitoring systems should be able to monitor radiation levels in the range of 0.2 mR/h to 5 R/h (1.76
Gy/h to 44 mGy/h) full scale and also have audio indication. A low range in the instrument is useful in assessing radiation
levels at occupied areas and general radiation survey around installations. The most commonly used radiation monitoring
instruments (e.g. MINIRAD, MR4500) have miniature GM counters, making them useful over wide range of exposure rates.
MINIRAD monitor can measure up to 5 Rlh and MR4500 can measure up to 50 Rlh making it useful in radiation emergencies.
Both these instruments can be used for area monitoring around source containers. Other monitoring instruments in use are :
8 . 61
ionization chamber of size 400 cm3 and can measure X and gamma radiation exposure-
rate from 5 mR/h to 5 Rlh (corresponding to air kerma values from 0.044 mGy/h to 44
mGy/h) in three ranges. The chamber is provided with a window and by opening the
window beta radiations can also be monitored. This instrument is useful for general
purpose monitoring and for checking the radiation levels around radiation source
housings.
Radiation survey meter (MR-121): It is also marketed by M/s. ECIL, Hyderabad. It is a
GM type survey meter. It has a long glass walled GM counter and can measure X and
gamma exposure rates from 0.2 - 20 mR/h (corresponding to 0.88 μGy/h to 0.175 mGy/h)
in three ranges. It can also respond to high energy beta radiations. It is very useful for low
level area monitoring. This instrument however has a drawback that it does not respond
at high radiation levels.
Table 7.1 lists some of the radiation measuring instruments, their range, and application. Two or more of these instruments are to
be normally maintained so that in case of failure of one, the other may serve as backup. As most of these instruments are
expensive and are likely to be used only occasionally, preventive maintenance and periodic calibration checks are mandatory. In
the case of ion-chamber based instruments, preventive maintenance consists mainly of keeping the sensitive input circuitry free
from moisture by proper desiccation, removal of batteries when the instrument is not in use, etc. For GM counter based
instruments only removal of battery when not used for a long period is required.
8 . 62
Gun Monitor Ionization 0-5R/h - do - - do -
(Exposure Rate chamber
Meter) GM-125
Radiation Survey NaI 0-1 mR/h Background radiation - do -
Meter PRM01215 scintillation measurement: low
detector level radiation area
monitoring
GM Survey Meter GM counter 0-20 - do - ECIL
(MR-121) mR/h
DIGICON GM counter 0-1000 Beta/gamma Nucleonix
Contamination cps contamination systems Pvt.
Monitor CM 710 monitoring Ltd.
8 . 63
such as urine, stool, sweat, sputum etc. The two means are known as the physical and
biological decay (or physical and biological half life) of the radioisotope.
Radioisotopes may gain entry into the body through any one of the following
pathways: a) Inhalation - by breathing contaminated air, b) Ingestion - eating drinking
food through contaminated hands or taking in contaminated food or water or transferring
radioactivity to the mouth through contaminated hands, c) Absorption of contamination
through intact skin or through wounds.
8 . 64
in GB. The main function of GB is that it isolates the contaminant from the environment
by containing it to a enclosed volume. A negative pressure of at least 13 mm (0.5 inches)
water inside the glove box assures no leakage of air into the box.
2. Control of Environment: Environmental control includes proper design of
laboratory, evolving handling procedures for radioactive waste, isotope storage,
ventilation and direction of air flow, good housekeeping, decontamination of working
surfaces, floors and walls, regular monitoring of work surfaces and persons. The
following safety procedures are followed to handle unseal radioactive materials:
1. Work with unseal radionuclide should, as far as possible, be restricted to a
minimum number of rooms. This enables confinement and containment of
radioactivity.
2. Room surfaces should be smooth, easily cleanable without and sharp corners,
crevices, and angles.
3. Room ventilation should ensure that the air flow is from low to high activity
areas.
4. In some establishments, work with unsealed radioactive materials has to be
carries out in areas where non-radioactive work is also undertaken. In such
situations, the work with radioactive materials should be confined to designated
locations within the work area.
5. Protective devices such as hand gloves, lab coats and shoe covers (if necessary),
should be provided to the workers working in the lab. The main function of
protective devices is to intercept radioactivity that would otherwise contaminate
the worker's personal clothing and skin. Respiratory protection is required in
labs where there is likelihood of air contamination levels exceeding the
prescribed limits and is recommended only in special cases.
6. Barriers and change rooms should be maintained at locations (point of entry)
near radioactive area with facilities appropriate to each situation.
7. Sufficient space should be available since over crowding of working area
inevitably increases the potential for accidents.
8. No upholstered furniture should be allowed in the laboratory, metal furniture is
desirable as it is fire resistant and easy to decontaminate.
8 . 65
9. Operations should be planned to limit the spread or dispersal of radioactive
material. In general wet operations should be preferred to dry ones and transfer
of material should be minimised.
10. Equipment, glassware, tools, etc. should be identified for use in radioactive
area. All such items should be chosen with a view for easy decontamination or
for disposal. Special care should be taken to avoid contamination of costly
equipment for economic reasons.
11. Manipulations should be carried out over drip trays to minimise the effects of
breakage or spillage. Work surfaces should preferably be covered with an
impervious covering, which should in turn be covered with absorbent material
to contain spillage, such covers should be changed as frequently as needed to
prevent any accumulation of contamination.
12. Handling tools and equipment used should be placed in non-porous trays on
absorbent disposable paper which should be changed frequently. Pipettes,
stirring rods and similar equipment should never be placed directly on the work
surface.
13. After use, all vessels and tools should be set aside for special attention when
cleaning.
14. Mouth pipetting should not be done in laboratories handling unsealed
radioactive materials. Pro-pipettes may be used for dispensing radioactive
liquid. Smoking, eating, drinking and use of cosmetics should be strictly banned
inside the laboratory.
15. Procedures should be evolved for collection, segregation, storage, and disposal
of radioactive waste.
16. No person should leave the radioactive area without checking contamination on
hands, feet, personal clothing, and shoes. Suitable monitors (hand feet monitor)
may be made available at exit locations. Reference levels of contamination and
action to be taken in case the contamination found is more than the reference
level, should be displayed in the laboratory.
3. Contamination: In a laboratory, where unsealed isotopes are handled,
contamination of work surfaces or personnel may occur either from normal operations or
8 . 66
as a result of breakdown of protective measures. Contamination can be classified into two
categories transferable (loose) and non transferable (fixed). Fixed contamination
is the source of external hazard and is that which cannot be removed even after
repeated smearing. Loose or transferable contamination is that which can be removed by
easy smearing of surface and is the source of internal hazard. There are derived
permissible limits of contamination at specific surfaces for specific type of radiation.
These are listed in table 7.2. For surface contamination monitoring two methods are
usually followed. They are direct and indirect methods.
Direct method: In this method of monitoring, the counting instrument is directly placed
over the area for measurement of contamination. The contaminated surfaces are scanned
using portable contamination monitor to locate the area of contamination. By this
method, total contamination is both fixed and loose contamination is measured. This
method is useful at such locations where sampling by smear test is difficult. The
monitoring results obtained in this method have large degree of reliability, but is
unreliable when monitoring is done in an area having large background e.g. near an
isotope generator.
Indirect method: The indirect method is more suitable in case of loose or transferable
contamination. In indirect method of monitoring, the smear samples are taken and
counted separately in low background set up. For collection of smear approximately 100
cm2 surface area is marked and dry or wet swabs with filter paper or cotton is taken. By
measuring the radioactivity transferred to it, the level of loose contamination can be
evaluated. The reliability of monitoring results, in this method, is poor because the
quantity of radioactivity transferred to swab will depend on pressure applied, area
swabbed, moisture content in the swab, technique used etc. To evaluate the level of
contamination one should know before hand, the coefficient of removal of activity from
the surface. This coefficient is usually defined as the ratio of the radioactivity on the swab
to the actual radioactivity present on the surface. The dry swabs are less efficient than the
wet; the wet techniques cause partial penetration into deeper layers of the contaminated
surface. Indirect contamination monitoring is more suitable to detect contamination
caused by low energy beta radiation such as 3H and 14
C. It is also suitable to detect
contamination near high radiation area (e.g. storage of large amount of activity). The
8 . 67
detection limit of surface contamination by this method is usually one order of magnitude
lower than that the direct method. In many situations, both the methods are
complimentary and are used to obtain a complete picture of surface contamination. For
detection of alpha contamination ZnS (Ag) scintillation probe is used.
4. Air Contamination Monitoring: The assessment of air contamination is done by
using a static air sampler and filter paper as medium. Filter paper with good collection
efficiency (99.9%) is preferred. Glass fiber filter papers have about 99.9% collection
efficiency for 1micron size particulate. The volume of air to be sampled and sampling
131
time will depend upon the contamination levels in the laboratory. For assessment of I
contamination in air, charcoal impregnated filter paper or ground charcoal is used. The
filter paper is then counted for collected radioactivity with the help of a pre-calibrated
counting setup. For measurement of 3H contamination in air, a different technique known
as cold strip method is used. The limits for air contamination on are governed by Derived
Air Concentration (DAC).
5. Personnel Contamination Monitoring: In practice it is difficult to correlate
contamination in a working place with the exposure to working staff. Therefore, the air
and/or surface contamination monitoring results are inadequate to assess the radiation
hazards. The complete environmental monitoring programme includes personal
monitoring also. It involves the monitoring of external (skin contamination) and internal
contamination in radiation workers. External contamination monitoring is done with an
appropriate portable or table type contamination monitors of known counting efficiency
for emitted radiations. For internal contamination monitoring, special techniques are
used. Most commonly used techniques are whole-body counting (for radio nuclides
emitting high energy gamma radiation), bioassay (urine analysis for 3H contamination) or
131 125
sometimes organ counting (thyroid counting in case of I or I). Personnel/individual
monitoring is carried out on routine basis or whenever it is necessary.
6. Decontamination Procedures: Removal of contaminant is the decontamination.
The procedure of decontamination should be such that it does not spread the
contamination to other areas. All efforts should be to contain the contamination to the site
of the incident. Whenever there is a need to move items out of the restricted area, they
should always be wrapped, to prevent spread of contamination. These considerations
8 . 68
apply equally to personal contamination and to contamination of equipment and facilities.
It is essential that decontamination is undertaken by properly informed and trained staff
only, using procedures and facilities that restrict doses to ALARA.
Personnel decontamination: The objective of personnel decontamination is to minimise
the radiation dose to the body and the skin. Decontamination techniques should remove
or reduce the external or internal contamination. External decontamination: Prior to
attempts to decontaminate externally deposited radioactive materials, all contaminated
clothing should be removed. Any skin swabs, nasal blows or other biological samples
should be retained in case they are needed to provide information in support of an
assessment of the incident. Contaminated clothing and skin swabs often provide a means
for the most rapid identification of radionuclide(s) causing the contamination. For injured
persons, first aid or partial decontamination is advised; only if there is likely to be delay
in reaching the medical facility. If contamination is localised over a limited body area
(e.g. hands, feet, neck, etc.) these areas should be cleaned by gently washing with soap
solution and water, but avoiding hot water. For more generalized contamination, the
person can be subjected to shower bath followed by a complete radiological survey.
Special attention should be given to body folds, the hair. Finger nails, but medical advice
should be sought before other decontamination methods are attempted. Care should
always be taken to avoid spreading contamination over the skin surface. Decontamination
should start on the outer perimeter of a contaminated area and proceed towards the
central point. Special care must be exercised to avoid spreading contamination to the
eyes, nose, or lips. Clean towels must be provided following each washing. Soaps, mild
detergents and approved chelating agents should generally suffice for skin
decontamination and these materials should be freely available in the workplace. In all
cases, washing must be gentle so that mechanical and chemical irritation of the skin is
minimised. Items and materials that have been used for personnel decontamination
should subsequently be treated as radioactive waste and treated accordingly. Special
medical attention should be sought if washing appears to be ineffective and the skin
contamination is not reduced to acceptable levels after several mild washings. Internal
decontamination: In case of internal contamination, the extent and magnitude of
contamination is unknown. Thus, detailed history of the accident or incident is often
8 . 69
important and must be recorded. The initial measurements and any samples obtained can
provide important evidence. Contamination that is caused internally (i.e., by inhalation,
ingestion or translocation from a wound or from skin) requires medical attention. There
are conventional methods, to be used under medical supervision for removing or
accelerating elimination of internally deposited radioactive material. Decontamination
of Surfaces: If the contamination is caused by short lived radio nuclides (e.g. 99
Tc) the
decontamination by physical decay of activity is preferred. After a period of 8 to 10 half
lives the surface can be cleaned with moist swab. Water is most commonly used
decontaminating agent. If this attempt is unsuccessful, detergents like soap solution,
teepol, EDTA, etc. can be tried to remove the contaminant. These decontaminating agents
are very suitable for removing loose contamination from the smooth, hard, and non-
porous surfaces. It is equally effective for those radioactive compounds which are water
soluble. To remove contamination from glass wares, the soap solution or chromic acid is
attempted. For the metal surfaces, where removal of contamination using above
mentioned reagents is difficult, mild acids like 0.1 N HCI or HNO3 may be tried. In all
these procedures, it is necessary to give time to moist the surface before removing the
contaminant. Fixed contamination of surfaces can be removed by gentle abrasion. If the
fixed contaminant emits high energy gamma radiation and has long half-life, the
contaminated object may be treated as radioactive waste and be disposed off
permanently. In case of spillage of large liquid activity in the laboratory, the contaminant
can be contained with in the area by placing enough absorbents carefully over it.
Protective tools like hand gloves, working clothes, shoe covers should be used.
Decontamination should be continued till the level of contamination reduced to the
specified levels. The specified levels are given in table 7.2.
8 . 70
INGESTION
INHALATION WOUNDS
EXHALATION
RESPIRATORY
TRACT
GASTROINTESTINAL TRACT
PULMONARY BLOOD
LYMPH NODES
SWEAT
URINE
FECES
Fig. 7.2 : Schematic Representation of Routes of Entry, Metabolic Pathways and Possible
Bioassay Samples for internally deposited radio nuclides.
at 1 cm
8 . 71
22
Na 2.6 y 1.274 (100%) 2.8
12.0
24
Na 15.0 h 2.57 (100%) 1.37 (100%) 4.4
18.4
42
K 12.5 h 1.53 (18%) 0.33
01.4
46
Sc 84.0 d 1.119 (10%), 0.887 (100%) 2.6
10.9
52
Mn 5.7 d 1.43 (100%, 1.44 (6%), 1.34 (6%), 4.4
0.935 (84%), 0.744 (82%), 0.51 18.6
(58%)
57
Co 267 d 0.136 (9% + 1% IC), 0.122 (89% + 0.2
1% IC), 0.0144 (6% + 84% IC) 00.9
60
Co 5.27 y 1.33 (100%), 1.17 (100%) 3.1
13.2
65
Zn 245 d 1.14 (49%), 0.51 (3.4%) 0.64
02.7
125
I 60 d 0.035 (7% + 93 % IC) 0.17
00.7
131
I 8.06 d 0.36 (80% + 1% IC), 0.28 (5%), 0.08 0.52
(2% + 4 % IC), 0.72 (3%), 0.54 (9%) 02.2
137
Cs 30 y 0.662 (85% + 10% IC) 0.78
03.3
141
Ce 32.5 d 0.145 (49%) 0.083
00.35
192
Ir 74.0 d 0.317 (72%), 0.468 (47%), 0.308 1.1
(28%), 0.296 (26%), 0.604 (9%), 04.80
0.612 (6%), 0.589 (5%)
Emergency Preparedness:
1. Charts, which detail various steps to be taken by a radiation worker, in case of a
radiation emergency, should be conspicuously displayed in the laboratory.
2. All the radiation monitoring and measuring instruments should be checked
routinely and kept always in working condition.
8 . 72
3. The ventilation system of the radioisotope laboratory should be checked
periodically and maintained properly.
4. A kit comprising of accessories like tongs, forceps, waste receptacles etc., which
are required for the decontamination operation should be available readily to
handle an accidental spillage.
5. A proper inventory of radioisotopes received, used, and disposed should be
maintained.
Emergency Procedures: It would be difficult to stipulate hard and fast rules for meeting
wide variety of radiation accidents. However, considering spillage of radioactivity as the
most likely accident in a radioisotope laboratory, following steps are recommended in
dealing with such emergencies.
1. Confine the spill immediately by tissue paper or such absorbent materials. This is
to avoid further spreading of the spillage.
2. Evacuate the immediate surroundings so that the spread of contamination by
accidental walking over the spill by laboratory personnel could be avoided.
3. If the spilled material has splashed on to a person or his clothing, immediate steps
should be taken to remove the contaminated clothes and to leave them in the area
meant for the purpose. The contaminated areas on the body should be washed
thoroughly with soap and water. Care should be taken not to scrub heavily or
inflame skin surfaces for removing the contamination. If internal contamination
has taken place, immediate action should be taken to minimize the deposit of
radioactivity in the internal organs and to enhance the excretion of the ingested
radioactive materials, under expert medical supervision. In case facilities are
available, bioassay or whole-body counting should be carried out to confirm
internal contamination.
4. Contaminated area should be decontaminated by experienced persons wearing
protective clothing like surgical gloves, shoe covers and surgical face mask etc.
tongs or forceps should be used to remove the contamination, which is confined
by absorbent materials. The absorbent materials so collected should be kept in a
polythene bag to be treated as radioactive waste. After as much contamination as
possible have been removed in this way, the surface should be-mopped with damp
8 . 73
(not wet) cotton or tissue paper held by forceps, always working' towards the
centre of the contaminated area, rather than away from it.
5. A contamination monitor should be used to monitor the area as well as the
personnel during the procedure of decontamination. The contamination monitor
should be operated by a person other than the person who does the cleaning up, to
avoid the possible contamination of the instrument. The contaminated gloves,
shoes covers, etc., should be kept in a polythene bag for decay for ultimate
disposal as radioactive waste. The forceps/tongs should be kept separately
covered in polythene bag for decay of radioactivity present on it.
6. If the spilled radioactive material is of a very short half life, say few hours, the
closure of the laboratory (if possible) for a day or overnight is recommended.
Since the natural decay process reduces the activity, the method is preferred to the
immediate decontamination steps as mentioned above. In this case, cleaning up
operations should be carried out once the activity is decayed to the safe levels.
7. In the case of large release of radioactive powder or aerosol, the room must be
immediately isolated from the surrounding by shutting off mechanical ventilation
and by closing windows and doors. A room with heavy air contamination can be
decontaminated from within by drawing the air of the room through an
appropriate filter.
8. Complete records of the accident giving details of the radioisotope, activity
involved and follow up procedures, etc. should be maintained.
For effective dealing with any kind of radiation emergency, the Radiological
Safety Officer (RSO) of the institution should educate and familiarize all the radiation
workers with the steps to be taken to meet the emergency.
Use of radio nuclides in research applications may not normally be a source for
exposure of general public, provided certain basic precautions are taken. These
precautions mainly involve good security for the radio nuclides, good work practice, and
a well controlled programme for the disposal of radioactive waste.
-----------
8 . 74
8. PLANNING OF RADIOISOTOPE LABORATORIES
INTRODUCTION
Handling of open radioactive materials requires a special radioisotope laboratory
depending on the nature of work. The radioisotope laboratory is graded in different
categories on the basis of the activities handled.
8 . 75
required will be that of Type-II laboratory.
Further, if two or more radio nuclides from the same groups or different groups
are required, the quantities that can be handled in the laboratory should be adjusted so
that Al/A* 1, where Al is the quantity of radio nuclides of the same or different
groups and A* is the prescribed limits for the classified laboratory depending on the
operation.
GENERAL FEATURES
Site
The site of the radioisotope laboratory should be so chosen that there is no
increased background radiation level either due to operation of X-rays, tele-therapy units,
storages of radioisotopes, or due to handling of radioisotopes in the vicinity. It will be
advantageous to have all the rooms of the radioisotope laboratory grouped together,
preferably at one end of the building so that entry of persons not connected with that
work may be effectively prevented. Separate areas and separate facilities should be
provided for high and low level activity work to reduce the possibility of cross
contamination.
8 . 76
application of radio nuclides, (iv) segregation of subjects containing radio nuclides after
application, (v) sample counting and other measurements, (vi) temporary storage of
radioactive waste, (vii) decontamination and (vii) ground for disposal of radioactive
waste.
Other facilities will be required for non-radioactive operation such as utility and
st9rage room, a dark room, animal or plant room, offices and toilets, etc. Some typical
plans of radioisotope laboratories are given in Figures 8.1, 8.2 and 8.3.
Table 8.1 : classification of Isotopes
Basis – Relative Radio-toxicity per Unit Activity
GROUP I
210 210 223 226 228 227 227 228
Pb Po Ra Ra Ra Ac Th Th
230 231 230 232 233 234 237 238
Th Pa U U U U Np Pu
239 240 241 242 241 243 242 243
Pu Pu Pu Pu Am Am Cm Cm
244 245 246 249 250 252
Cm Cm CM C- C- Cf
GROUP II
22 36 45 46 54 56 60 89
Na Cl Ca Sc Mn Co Co Sr
90 91 95 106 110m 115m 114m 124
Sr Y Zr Ru Ag Cd In Sb
125 127m 129m 125 126 131 133 134
Sb Te Te I I I I Cs
137 140 144 152 154 260 170 181
Cs Ba Ce Eu Eu Tb Tm Hf
182 192 204 207 210 211 212 224
Ta Ir Tl Bi Bi At Pb Ra
228 230 234 236 249
Ac Pa Th U Bk
GROUP III
7 14 18 24 38 31 32 35
Be C F Na Cl Si P S
41 42 43 47 47 48 48 51
A K K Ca Sc Sc Vr Cr
52 56 52 55 59 57 58 63
Mn Mn Fe Fe Fe Co Co Ni
65 64 65 69m 72 73 74 76
Ni Cu Zn Zn Ga As As As
77 75 82 85m 87 87 85 91
As Se Br Kr Kr Kr Sr Sr
90 92 93 97 93m 95 99 96
Y Y Y Zr Nb Nb Mo Tc
97m 97 99 97 97 105 105 103
Tc Tc Tc Ru Ru Ru Rh Pd
109 105 111 109 109 115m 113 125
Pd Ag Ag Cd Cd In Sn Sn
122 125m 127 129 131m 132 130 132
Sb Te Te Te Te Te I I
134 135 135 131 136 131 140 141
I I Xe Cs Cs Ba La Ce
8 . 77
143 142 143 147 149 147 149 151
Ce Pr Pr Nd Nd Pm Pm Sm
153 152m 155 153 159 165 166
Sm Eu Eu Gd Gd Dy Dy
166 169 171 171 175 177 181 185
Ho Er Er Tm Yb Lu W W
187 183 186 188 185 191 193 190
W Re Re Re Os Os Os Ir
194 191 193 197 196 198 199 197
Ir Pt Pt Pt Au Au Au Hg
197m 203 200 201 202 203 206 122
Hg Hg Tl Tl Tl Pb Bi Bi
220 222 231 233 239
Rn Rm Th Pa Np
GROUP IV
3 15 37 58m 59 69 71 85
H O A Co Ni Zn Ge Kr
85m 87 91m 93 97 96m 99m 103
Sr Rb Y Zr Nb Tc Tc Rh
113m 129 131m 133 134m 135 147 187
In I Xe Xe Cs Cs Sm Re
191m 193m 197m 232 235 238
Os Pt Pt Th Th-Nat U U U-Nat
8 . 78
Table.8.2 : Criteria for Grading Laboratories using Unsealed Radioisotopes
8 . 79
Type-III (Stringent)
Large scale laboratory-multi room complex with clear segregation of area based
on use, scale and type of operation with the radioisotopes. A general list is given below.
10 .1
5. DRY AND DUSTY OPERATIONS (e.g. GRINDING) x 0.01
Ventilation
The flow of ventilating air in the radioisotope laboratory should be from area
having no activity to low activity area then to medium activity area and from there to area
of high activity. This flow of air pattern will prevent the spread. of radioactivity from
high activity to low activity area e.g. into the counting rooms or in the adjoining rooms in
the event of accidental spill.
The air in the radioisotope laboratory should not be re-circulated.
Surfaces
All surfaces in any active area should be smooth, unbroken and be made from
materials which are chemically inert, non-absorbent and water repellent. Consideration
should always be given to possible decontamination problems, which might arise, and so
materials must be chosen which are either easily decontaminated or which can be
conveniently removed or replaced.
Specifications
Areas/rooms where radioisotope laboratory is being planned, special design
consideration should be given for the following:
10 .2
be satisfactorily filled by heavy waxing of the surfaces. Further, floor should have at least
15 cm skirting on the walls.
Walls and ceiling are less likely to become contaminated than floor. Nevertheless
the surfaces should be smooth, crack free and non-porous. The porous wall material
should be coated with a non-porous, washable paint and preferably with a fine layer of
strippable coating.
Work Surfaces
All work surfaces should have hard non-porous finishes which have the necessary
heat and chemical resisting properties such as Sunmica or Formica. It should be bonded
to the backing material with resin glue to give necessary temperature resistance. Further,
to ensure that the surfaces are not spoiled, it is advisable to have it always covered with
polythene sheet and then with absorbent paper, so that if there is any liquid spill. it is
immediately absorbed by the paper, which can be easily disposed off.
Containment System
To prevent spread of contamination into working areas, all work with
radioisotopes is carried out in containment systems like fume hoods/fume cupboards,
glove boxes, etc. The general purpose and the required degree of containment for
different systems are given below.
Fume hood
A fume hood is a box with sliding transparent front panel opening where an
inward air flow of 30-40 linear metres per minute is maintained. Fume hood is generally
used when contamination and external radiation hazard are not too high and relatively
low levels of activities are handled or operations where gases or reactions at elevated
temperatures are involved. Figure 8.4 presents a schematic drawing of a radiochemical
fume hood. A fume hood should not be placed near door ways or windows or in the
vicinity of strong air currents, which may tend to draw fumes from the hood. All the fume
10 .3
hoods in anyone room should be controlled QY the same switch. This will avoid the
chance that air flow from an operating hood will bring contamination into the room from
a non-operating hood. The services generally required in a fume hood are water, gas,
vacuum and electricity. The controls of these services should be situated outside the hood
to minimize the number of movements through the front opening. The exhaust air should
be discharged so as not to contaminate surrounding facilities; a point of discharge 120-
150 cm above the roof or 120-150 cm above the tallest surrounding building is usually
satisfactory. Blower should be located near the top of the exhaust duct so as to minimize
the escape of active material caused by positive pressure within the duct. Usually it will
not be necessary to filter the hood output. If required, filters should be installed near the
start of the exhaust duct.
Glove Box
A glove box is generally installed in active laboratories to facilitate the handling
of hazardous materials. It consists of a leak-tight enclosure in which objects or materials
may be manipulated through gauntlet gloves attached to ports in the walls of the box (see
Fig. 8.5). Its aim is to provide containment for materials which are either radioactive or
chemically toxic or both. Usually it does not provide any shielding protection against
penetrating radiation and so it is used for alpha or beta emitters. When gamma-emitting
isotopes have to be handled, a wall of lead bricks is usually constructed between the
operator and the glove box.
Glove box is maintained at a pressure slightly below that of the outside laboratory.
This means that air will flow into the glove box, should a leak develop and this will
prevent the contamination escaping. Two filters are normally placed on the ventilation
system-one to remove dust from the air being drawn into -the glove box and the second to
remove radioisotope particles from the air being drawn out of the box.
10 .4
RADIOISOTOPE STORAGE AREA SOLID WASTE
STORAGE D STORAGE
D
D LIQUID WASTE
FUME HOOD
D STORAGE
SINK
WORK TABLE
CM
CHANGE
ROOM
WB WB
D
ISOTOPE
FUME HOOD STORAGE
COUNTING ROOM
SOURCE SEORAGE
AND
D
HANDLING
D
(a)
10 .5
(b)
D D
SOURCE STORAGE
AND
BETA GAMMA HANDLING
COUNTING COUNTING
FUME ISOTOPE
HOOD STORAGE
10 .6
HHOT FUME
HOOD GLOVE BOX
HOT ROOM
GAMMA LAB
D D
D D
D
BETA COUNTING
MASS SPECTROMETER
AUTO RADIOGRAPHY
D
OFFICE
D D
D
SOFT BETA LAB
WASH
D
10 .7
8
6
1
3 7
5
90 cm
165 cm
180 cm
10 .8
5
4
6
60 cm
2
90 cm
45 cm
SUMMARY
1. The radio nuclides/radioisotopes are classified into four groups according to their
relative radio-toxicity.
2. There are four Groups viz. Group I, Group II, Group III and Group IV. These
groups are called High toxicity, Upper medium toxicity, Medium toxicity and
Lower medium and low toxicity respectively.
10 .9
3. Depending on the facilities available for handling radio nuclides and the
quantities of radio nuclides of each group handled there are three types of
radioisotope laboratories. These are Type I, Type II and Type III.
4. Radioisotope laboratory should not be located in an increased background
radiation area i.e. dose to installation like X-ray, tele-therapy or accelerator, etc.
5. All the rooms of a radioisotope laboratory should be grouped together.
6. The flow of ventilating air in a radioisotope laboratory should be from areas
having no activity to low activity area; then to medium and from there to area of
high activity.
7. The air in the radioisotope laboratory should not be re-circulated.
8. All the surfaces of the radioisotope laboratory either working or any other should
be smooth, unbroken, non porous and of material which can be easily
decontaminated.
----------
10 .10
9. REGULATORY ASPECTS OF RADIOLOGICAL SAFETY
INTRODUCTION
Radiation sources are widely used for research purposes. Work with radiation
sources may involve exposure of workers to radiation. It is known that exposure levels
significantly greater than the limits prescribed by the ICRP may entail certain health
hazards. It is, hence, necessary to ensure that any work involving radiation exposure is
carried out in a manner that results in minimum radiation exposure to workers and public.
In order to have an effective control on the use of radiation and to ensure radiological
safety of the user as well as the public, the -Government of India has promulgated the
Radiation Protection Rules, 1971 under the Atomic Energy Act, 1962. The Act empowers
the Government to make rules relating to radiological safety. The responsibility of
enforcing the rules is assigned to the Competent Authority specifically appointed by the
Central Government. The Chairman, Atomic Energy Regulatory Board (AERB), is the
Competent Authority in India. All radiation protection requirements in respect of
institutions outside the Department of Atomic Energy are being enforced by Head,
Radiological Safety Division (RSD) of AERB.
In India, only persons who are duly authorized by AERB are permitted to procure
and handle radiation sources. If the sources are to be procured locally or to be imported
an Authorization/No Objection Certificate (NOC), should be obtained by the prospective
user from AERB.
CURRENT PROCEDURE
Application for authorization to handle radiation sources should be made to Head,
RSD, AERB. The applicant should furnish in the prescribed form details regarding the
type of sources, its activity, proposed use, names of the users, their qualification and
experience in the handling of radiation source, etc. (See APPENDIX -1).
RSD would scrutinize the application and evaluate the radiological safety status of
the laboratory and also would advise the applicant regarding the radiation safety
10 .11
requirements, specific to his needs.
The advice would relate to:
- safe storage of the sources,
- facility for handling the sources,
- laboratory fittings.
- personnel monitoring and area monitoring, - trained staff, and
- disposal of radioactive waste
- emergency procedures
Upon fulfilling the requisite safety requirements, the applicant will be authorized
by RPAD to procure and handle radioisotopes. . The above authorisation is liable to be
revoked in case of non-compliance of the safety regulations by the institution.
a) to take all necessary steps aimed at ensuring that the operational limits of
radiation exposure to personnel are not normally exceeded.
b) to educate the radiation workers under his charge on the hazards of radiation and
on suitable safety measures and work practices aimed at minimizing exposure to
radiation.
c) to regulate the safe movement of all radioactive materials (including wastes
containing radioactive materials) within the area under his charge.
d) to maintain a log book for inventories like the name and amount of radioisotope
received, date of receipt, the department in which it will be used, date of using,
name of the user, etc.
e) to carry out periodic monitoring of areas and contamination check of surfaces
(wherever applicable) arid maintain these records in the log book.
f) to investigate and initiate prompt and suitable remedial measures in respect of any
situation that could lead to radiation hazards.
10 .12
g) to ensure that reports of all hazardous situations along with details of any
immediate remedial measure that may have been initiated, are made available to
his employer for onward transmission to the Head, RSD, AERB, Niyamak
Bhawan, Anushaktinagar, Mumbai 400 094.
h) to ensure that the ultimate disposal of waste containing radioactive materials is
done in a manner approved by the Competent Authority.
No person below the age of 18 years should be appointed as radiation worker. All
radiation workers should undergo a routine medical examination prior to employment
and thereafter "as specified by the Competent Authority from time to time.
The institution should maintain the following records in respect of its radiation
workers.
- Occupational history,
- Medical history, and
- Cumulative dose records.
SUMMARY
1. Government of India has promulgated the Radiation Protection Rules, 1971 (RPR,
1971) under the Atomic Energy Act, 1962.
2. The competent authority in India is Chairman, Atomic Energy Regulatory Board.
3. On behalf of the Competent Authority, the Radiological Safety Division RSD) of
AERB tenders advice to the users on radiological safety for ensuring the
regulatory requirements:
4. R.S.O. stands for Radiological Safety Officer.
5. Authorization/NOC for obtaining radiation sources is issued by RSD.
-----------
10 .13
11. TRANSPORTATION OF RADIOISOTOPES
INTRODUCTION
On receiving a purchase order from an authorized user, radioisotopes are packed,
marked, labelled by the supplier and then sent to the consignee by rail, road, sea or air.
During transport it is necessary to ensure both the security of the consignment and further
the exposure to the transport operators and other members of the public should be kept to
a minimum. For this purpose transport of radioactive materials is permitted only in
packages of prescribed design as governed by national/international regulations.
DESCRIPTION OF PACKAGES
A package should have a containment system for the purpose of housing the
radioactive material during transport. It should be provided with a closure device (e.g.
lid) in order to prevent release of the source during transport. It should also incorporate
adequate shielding. The maximum permitted radiation level at the external surface of a
package is 2 mSv/h (200 mrem/h) and the corresponding limit at 1 m from the external
surface of the package is 0.1 mSv/h (10 mrem/h). Accordingly they are classified in terms
of Type of package (depending upon the sturdiness of the packaging) and Category of
package (depending upon the radiation level outside the package).
TYPES OF PACKAGES
The various types of packages are a) Excepted packages, b) Low specific activity
(LSA) packages, c) Surface contaminated objects (SCO) packages, d) ‘Type A’ packages,
e) ‘Type B’ packages. An excepted package is either an empty package having contained
radioactive material or one which contains very small quantities of radioactive material
prescribed in relevant regulations as accepted from the regulatory requirements. If the
activity of the radioactive material per unit mass or unit volume is low enough, or if the
material to be transported is a contaminated object such that the contamination per unit
surface area is within a specified, limit, such material is described as low specific activity
(LSA) material or surface contaminated object (SCO), as appropriate. These
consignments pose limited radiological hazard as could be imagined from the above
10 .14
description. ‘Type A’ and ‘Type B’ packages can contain higher amounts of radioactive
material. A ‘Type A’ package is one which is designed to withstand normal conditions of
transport. ‘Type B’ packages should be sturdier and have to be designed to withstand
severe accidents that may occur during transport. These include crash, fire, and
immersion in water. Each package has to be type tested before approval for transport. The
tests simulating normal and accident conditions of transport are specified in the
regulations for safe transport of radioactive material.
CATEGORIES OF PACKAGES
Packages are classified into three categories. Before going into categories of
packages let us first define an important quantity, viz., the transport index. Transport
index is a number expressing the maximum radiation level in mrem/h at 1m from the
external surface of the package.
Category I WHITE: If the radiation level at any location on the external surface of the
package does not exceed 0.5 mrem/h and if the transport index of the package is 0.0 then
the package belongs to Category I WHITE.
Category II YELLOW: If either of the limits for Category I WHITE is exceeded then
the package belongs to Category II YELLOW, provided, however, that the radiation level
at any location on the external surface of the package does not exceed 50 mrem/h and the
transport index of the package does not exceed 1.0.
Category III YELLOW: If either of the limits for Category II YELLOW is exceeded
then the package belongs to Category III YELLOW, provided, that the radiation level at
any location on the external surface of the package does not exceed 200 mrem/h and the
transport index does not exceed 10.0. However, if a package is transported under
exclusive use conditions (i.e. chartered vehicle) the radiation level at the external surface
of the package shall not exceed 1000 mrem/h.
On Receipt of Packages
The responsibility for safety in the transport of radioactive material rests with the
consignor. However the procedure to be observed while receiving radioactive
consignments is briefly described below:
1. Immediately upon receipt of intimation from the carrier regarding the arrival of
10 .15
the consignment, arrangements should be made to take delivery of the package.
2. Only persons who are familiar with radiological safety should be sent to collect
the package.
3. The person who collects the package should wear his personnel monitoring badge.
He should also carry with him a polythene bag in order to house the package in
case it is received in a damaged condition.
4. Before collecting the package it should be ensured that the package is indeed
addressed to the institution on whose behalf the package is being collected.
5. If the package is in a damaged condition the observed condition of the package
should be recorded in the documents of the carrier and the package should be
taken delivery of, carefully wrapped in the polythene bag brought for the purpose
and taken to the institution. The matter should be immediately reported to the
supplier and also to Head, RSD, AERB, Niyamak Bhawan, Anushaktinagar,
Mumbai - 400 094 by telephone, telegram or telex.
-----------
10 .16
12. PRODUCTION OF RADIOISOTOPES AND LABELLED
COMPOUNDS
Radioisotopes are used today in almost every field of human endeavour. Its use in
industry, medicine, agriculture, and research represent some of the beneficial applications
of atomic energy. The application of radioisotope can be of simple irradiation of samples
like in industrial radiography, sterilization of medical products, irradiation of food items,
radiotherapy etc. or as tracers like in nuclear medicine, hydrology, mining etc. A number
of radioisotopes occur naturally and a large variety of them are produced artificially in
nuclear reactors or by particle accelerators. About 200 radioisotopes are currently
available. In the production of radioisotopes by artificial means, a suitable projectile is
made to bombard a suitable target element (or compound). The projectiles generally used
are neutrons, deuterons, alpha particles and some light nuclei. All the presently
available radioisotopes in the country are produced by neutron bombardment in nuclear
reactor.
The most frequently occurring reactions during neutron bombardment of the
target are of the following types:
a. A (n, ) B reaction
59
Co + 1n 60
Co
The product nuclide of this type of reaction is an isotope of the target element.
b. A (n,) B C reaction
130
Te + 1n 131
I + - +
In this reaction the radioactive daughter element C is produced by the
disintegration of primary product B. This type of reaction is useful in the preparation of
carrier-free materials
c. A (n,p) B reaction
35
Cl + 1n 35
S + 1P
32
S + 1n 32
P + 1P
The product nuclide differs chemically from the target element.
10 .17
d. A (n,) B reaction
6
Li + 1n 3
H + 4He
In this type of reaction also, the product nuclide is, chemically different from the
target nuclide.
e. Fission Products
When Uranium-235 or Plutonium-239 absorbs a neutron, fission of the nucleus
takes place to a variety of fission product nuclides of which many are radioactive.
235
U (n,f) fission products like 137Cs, 99Mo, 147Pm, etc.
The amount of radioactive atoms (Activity) produced is dependent on the number
of atoms of target, neutron flux of the reactor, neutron cross section, time of irradiation
and decay constant of the product radionuclide.
10 .18
as a labelled molecule, a tagged compound, a radiotracer compound or simply a
radiochemical (generally the radiochemical term is used for simple chemical forms like
halides, oxides, phosphates etc.) A radio labelled compound can either be, a) specifically
labelled, b) uniformly labelled, c) randomly labelled or d) non-isotopic labelled or foreign
labelled. Some examples of such labelled compounds are given below:
a) Specifically labelled
CH3CH(CH3)CH2CH(NH2)COOH Leucine (1-14C)
here carbon at position 1 is 14C.
b. Uniformly labelled
CH3CH(CH3)CH2CT(NH2)COOH Leucine (U-14C)
here all the carbon atoms are 14C
c. Randomly labelled
CH2TCT(CH3)CH2CT(NH2)COOH Leucine T -(G)
here tritium label is randomly positioned
d. Non-isotopically or foreign labelled
Protein labelled with 125I; Nucleotide labelled with 35S.
(In these cases. the isotope used for labelling is not an isotope of any of the elements
naturally present in the molecule).
Carbon and hydrogen being present in almost every organic compound, Carbon-
14 and tritium labelled compounds are being used extensively. In recent years,
125
phosphorus-32, sulfur-35 and I labelled compounds are finding wide applications in
the frontier areas of research in genetic engineering. All these labelled compounds are
synthesised either by chemical or biochemical methods.
Specific Activity of Radio-labelled Compounds
The amount of radioactivity in a labelled compound is expressed in terms of
specific activity, which is the total activity divided by free quantity of the element,
chemical or labelled compound.
S = A/W , where A is activity and W, the weight in grams or moles.
This is an important parameter, which should essentially be known for any
application, which involves quantitative analysis. .
10 .19
Specific activity is expressed in units of Ci/g, mCi/mmol, mci/ng or MBq/mmol.
10 .20
Table 12.1 : Some important Radioisotopes
---------
10 .21
APPENDIX -1
Radioisotopes in India can be procured and handled only by users who have
either worked under an authorised person or undergone training in safety and are duly
authorised by Radiological Safety Division (RSD), Atomic Energy Regulatory Board
(AERB), Niyamak Bhawan, Anushaktinagar, Mumbai - 400 094. This authorisation is
based on the radiological safety status of the laboratory, where radiation sources are
handled. For this authorisation, it is mandatory that an application is submitted to RSD,
furnishing the details of the laboratory and qualifications and experience of personnel, in
the prescribed format available at web site of AERB (www.aerb.gov.in) along with the
plan of the radioisotope laboratory.
The plan of a radioisotope laboratory depends upon the type of radioactive
material to be handled, its physical form and activity and the technique of its application
in the study. In general the minimum requirement is of two rooms of suitable size
(adjacent to each other) one for storage, and handling of radioisotopes and the other for
counting of radioactive samples. A second counting room may be provided taking into
account future expansion. This will ensure separate rooms for ‘high activity’ and ‘low
activity’ counting. If experiments with animals are to be carried out, then a separate room
for segregating ‘active’ animals is to be provided. This will imply that the area will
depend on the activities planned to be handled.
Two copies of the plan of the radioisotope laboratory, drawn to scale (1:50)
should be sent to RSD, indicating in it the room for storage and handling of
radioisotopes, counting radioactive samples and keeping animals etc. The dimensions of
the rooms, the position of doors, windows, exhausts, fume hoods, work benches and other
fixtures etc. also should be indicated in the plan. Preferably a site plan (to scale) of the
building housing the radioisotope laboratory may also be sent along with the plan of the
laboratory indicating the nature of occupancies in the immediate surrounding of the
radioisotope laboratory including those above the ceiling and below the floor, if any.
The suitability of the plan is generally decided on the basis of the above
information and if found necessary requisite modifications are suggested by RSD before
approval. Otherwise one of the plans of the radioisotope laboratory is returned duly
stamped and approved.
Order for the requisite quantity of radionuclides is placed with Senior Manager,
Technical, Sales and Operation, Board of Radiation and Isotope Technology (BRIT),
V.N. Purav Marg, Deonar, Mumbai - 400 094 by the user. This will be referred to RSD
by BRIT for Authorisation/No Objection Certificate. Queries regarding approval and
classification of radioisotope laboratory, approval of qualified staff, nomination of RSO,
Authorisation/No Objection Certificate are to be referred to the Head, RSD, whereas
queries regarding the supply of radionuclides are referred to BRIT.
10 .22
APPENDIX - 2
During the area survey, radiation levels must be determined at the source storage,
waste container, preparation room, counting room, work place, etc. Since radiation levels
expected are normally low, the monitoring instrument should be properly selected (e.g.
GM survey meter). During the survey, the instrument should be kept at the highest
sensitivity range and held close to the surface being monitored and moved across slowly
to cover the required area. If the radiation level exceeds the range, the meter should be
switched to the next higher range. The radiation levels at various places should be
recorded on a layout plan of the room. It should be noted that a background radiation
level exists at all places and may range from 10-20 R/h. The radiation level at any place
occupied by radiation workers should not exceed 2.5 mR/h and one tenth of this for
places occupied by non-radiation workers. All attempts must be made to keep the levels
further down, as low as reasonably achievable. During the survey it should be ensured
that all areas such as storage facility, waste container, work place, etc. have labels
indicating radioactive area displayed at the appropriate place.
Contamination by low energy beta emitters such as 3H, 14C, 35S cannot be detected
by GM counters. For this purpose a swab test must be carried out as described in
Appendix 3.
10 .23
Proforma for Radiation Protection Survey of a Radioisotope
Laboratory
Telephone No. :
Telex No. :
10 .24
4. Persons who were contacted/ :
available during survey
3. Not enrolled
10 .25
V. Layout of radioisotope laboratory (a rough sketch may be attached)
Number of rooms :
Total area :
1. Separate rooms are provided for each operation such as storage, handling, counting,
etc. Yes/No
10. Work table surfaces are covered with smooth, non-absorbing material Yes/No
(example: Sunmica, Formica, Stainless Steel, etc.)
10 .26
i) Can Qpener Yes/No
Location :
Exposure rate :
3. Contamination levels
1. Foot operated waste bins with disposable polythene lining inside are Yes/No
(for solid waste collection)
10 .27
ii) Special delay tank facility available
X. Records
1. 2.
1. 2.
10 .28
10 .29
APPENDIX-3
2. REQUISITES :
a. GM counter
3. THEORY :
Radioactive contamination results basically by contact between radioactive
material and. any surface, whenever radioisotopes in open form are handled.
Direct monitoring of surfaces, for beta ai1d gamma contamination is done where
the instrument probe can be directly held just over the contaminated area. For this
purpose, a GM counter having proper window thickness can be used at fixed
geometry. For alpha contamination, a probe with ZnS scintillator or a thin
window GM counter at a distance of 5 mm from the contaminated area can be
used.
Indirect monitoring is done where direct monitoring is not feasible. For this
purpose, a swipe is taken from the contaminated area and counted using a suitable
detector. It is assumed that 10 to 20 per cent of the contamination is removed by a
smear.
The experiment consists of assessing the contamination by direct method and
checking the effectiveness of various decontaminating agents for removal of
contamination, due to beta source, from different surfaces.
4. PROCEDURE :
a. Mark an area of to cm diameter at the centre of the given surface.
b. Dispense 1 microcurie of phosphorous-32 and spread it uniformly within the
marked area. Place it under infrared lamp to dry.
10 .30
c. Take background counts.
d. Monitor the contaminated surface over a fixed geometry.
e. Take a dry swipe with a filter paper (2 cm dia.) using a pair of tongs and count
it.
f. Determine ,the fraction of the original contamination removed by the swipe.
Decontamination :
Wet the surface with water and scrub with a small amount of cotton using a pair of tongs.
Dry the surface and monitor. Repeat the procedure once more.
Repeat this procedure with Teepol(soap). Radiac wash, EDTA and 0.1 N HCI (For metal
surface, 0.1 N HNO3). Wash the surface with water after each step. Repeat the counting
with the same geometry after each step and take the average of 3 to 4 readings.
Table 1
G.M. Counter Sr. No. : ________________ Operating voltage: ________________
C0 - B
10 .31
Decontamination Factor (FD) = -----------------------
C-B
The residual contamination RR (in per cent) is calculated by the relation RR = l/FD X 100.
The decontamination factor, arrived at after the last decontamination stage is the only one
used for assessment of the sample. Assessment is made in accordance with Table 2.
Table 2
FD
Possibility of
Decontamination
100 very good
100-50 Good
50-25 moderate
25 Poor
It is possible to calculate the contamination level of the, surface in the following way :
a. Keep the standard phosphorous-32 source at a distance of 3 cm (same distance
used for the experiment) below the GM counter for 1 minute. Let the countrate be
X counts/min.
b. Determine the background countrate. Let this be Y counts/min.
c. Repeat steps (a) and (b) at least four times and take the average count rate.
d. Let Z be the dpm of the standard source.
(X-Y) x 100
E% = ----------------------
Z
10 .32
The derived working level for alpha contamination is 10-5 Ci/cm2 and for beta 10-4
Ci/cm2. However, it is recommended that contamination level is brought down to
background level.
10 .33
APPENDIX-4
1. AIM :
2. REQUISITES :
3. THEORY :
Radiation monitors are portable, battery operated and they generally measure
exposure or exposure rates. Calibration of these instruments is usually done by the
manufacturer. However, due to many factors such as changes in detector
insulation, climatic effect on components, etc. the original calibration ceases to be
valid. Therefore in order to get accurate readings, it is essential to recalibrate the
instruments from time to time. The calibration of radiation monitors consists of
checking the electronic circuits by feeding simulated pulses and checking outputs
at various stages with the help of an oscilloscope. Finally, the combined response
of detector and measuring circuit is seen with the help of standard radioactive
sources, whose exposure rate at various distances can be calculated. If the
readings are not correct, then the calibration is done either by adjusting the
potentiometer specially provided for this purpose or modifying it with correction
factor (ratio of correct value to the observed reading). These values can be used
later for determining the correct exposure or exposure rate. It is also necessary to
check the response of the instrument at various radiation energies expected to be
encountered.
In this experiment, the following instruments, whose brief description is given,
will be calibrated. Assuming that the electronic circuit is working properly, only
calibration will be done, with standard radioactive sources.
10 .34
(b) Geiger Muller Survey Meter :
This is a portable area monitor consisting of a Geiger Muller (GM) counter, a
preamplifier, a pulse shaper, a countrate meter and a meter display. It has three
ranges of 0.2,2 and 20 inR/h. The main advantages of this instrument are its
ruggedness and high sensitivity. The disadvantages are its energy dependence,
finite dead time and paralysis at high exposure rates.
The following formula can be used to calculate exposure rates at various distances:
KxS
E = -------------
d2
Specific gamma ray constants ‘K’ for some radioisotopes are given in Table 1.
Table 1
Source
Specific gamma ray
constant
Caesium-137 3.2
Cobalt-60 13.2
Iridium-192 4.8
4. PROCEDURE :
a. Place the standard radioactive source on the axis of the cylindrical chamber of
the gun monitor at a distance do from the end of the chamber as shown in Fig.
1 (a).
10 .35
b. Choose the distance ‘d’ such that the deflection in the meter in a particular
range is more then half the scale.
c. Observe the reading on the meter and compare it with the exposure rate
calculated by the above formula.
d. If the readings do not tally, adjust the calibration potentiometer, such that the
meter reads, the calculated value.
e. Check the calibration in other ranges also. This is done by placing sources of
appropriate activity and noting down the observed readings.
f. Follow the above procedure for calibration of GM survey meter [see Fig. l (b)
and pocket dosimeter [see Fig.1 (c)]. The distance‘d’ should be measured from
the central axis of the cylindrical detectors.
g. Tabulate the readings in the respective tables.
x = 4 cm d1 d2 d3
GM detector
S1 S2 S3
d1 d2 d3
S1, S2 & S3 are the source position at distance d1; (d1+d2); & (d1+d2+d3)
10 .36
5. OBSERVATIONS :
10 .37
In case of pocket dosimeter, repeat the readings with dosimeters of various ranges viz.
250 mR, 500 mR, 1 R, 5 R, and 10 R.
In order to study the energy dependence, use sources of different energies and
find the correction factor for each energy. Draw a graph of energy versus correction
factor for each instrument.
10 .38
which the radionuclide is put. These can be solid, liquid, gaseous, combustible or non
combustible,: aqueous or non aqueous. In chemical operations the waste may be in the
form of solutions, precipitates, contaminated equipment. In medicine or biological
research work the waste may consist of excreta, tissue specimen, foliage, or animal
carcasses. Here we will discuss the various types of wastes generated in a radioisotope
research laboratory with respect to their origin, treatment, and disposal
.
10 .39
Table 10.3 : Categories 01 Solid Wastes
Category Radiation dose on the surface of Remarks
wastes
D (mGy/h)
1. D 2 x 10-3 - emitters significant
2. 2 x 10-3 < D 2 x 10-2 emitters insignificant
3. 2 x 10-2 D
4. activity expressed in MBq/m3 -emitters dominant -
emitters insignificant
concentrate and contain (2) delay and discharge and (3) dilute
and discharge. In selecting a method for waste disposal, each radioisotope should be
separately evaluated for hazard and it should be borne in mind that radioisotopes
discharged from different places may result in exposure of the same individual.
1.2
should be disposed off in conformity with the recommendation given below.
Solid Waste
For the disposal of solid radioactive waste, segregation should begin at the
point of generation itself. Two waste containers for each work table must be used one
marked "active" and another "inactive". Solid radioactive waste should be accumulated
in the laboratory in suitable receptacles and subsequently buried locally in pits. Site for
such burial should be selected taking into account the topographical, hydrological, and
meteorological characteristics of the environment in order to minimize the impact of the
waste on the environment. Factors to be taken into account include also the nature and
location of other facilities in the vicinity, usage of ground and surface waters in the
surrounding areas and the anticipated risk of accidental dispersal of waste to nearby
vegetation and subsoil waters.
The size of the pit may be 120 cm x 120 cm. The depth of the burial pit should be
so chosen that. the radioactive wastes have a top layer of compact earth of minimum 120
cm thickness when the pit head is finally closed. Additional pits maybe made in the burial
site, in case the quantity of radioactive waste exceeds the amount permissible in one pit.
Successive pits should be separated by a distance of at least 180 cm. Not more than
12 pits should be made in one year. A log book should be maintained for recording the
location, identity, and quantity of each isotope buried in the pits. The burial site should be
fenced off and provided with a gate which should be kept locked. Placards should be
displayed prohibiting unauthorized entry.
It should be ensured that the total activity of the wastes containing radioisotopes
buried at any time and at anyone location (or a pit) of the burial site does not exceed the
limits specified in Table 10.4. However, when using more than one radioisotope at a time.
the total quantity of radioisotopes that can be buried should be calculated as follows:
Determine the ratio. Ai/Li, where Ai is the activity to 'be disposed and Li is the limit
specified in table 10.4. The above ratio summed over all the radioisotopes to be buried
should not exceed unity i.e. 1.
131
A mixture containing 185 kBq of I, 3.7 MBq of 3H and 37 MBq of 32
P is to be
disposed. Find out whether the mixture can be buried in a single pit at a time. If not, how it
1.3
should be disposed off?
Total quantity of radioisotopes (more than one) that can be buried at a time can be
found out from the equation Ai/Li 1
A1 for 131I = 0.185 MBq; L1 for 131I = 37.0 MBq; A2 for 3H = 3.7 MBq;
L2 for 3H= 9250.0 MBq; A3 for 32
P = 37.0 MBq; L3 for 32P = 370.0 MBq
Condition to be satisfied for a single pit buried is
A1 A2 A3
1
L1 L2 L3
Substituting the values
0.185 3.7 37
37.0 9250 370
This will work out to be <1, therefore the mixture can be disposed at a time in a
single pit.
Normally after a period of seven to ten half-lives the radioactivity would decay
sufficiently so as to permit the contents of a pit to be removed and disposed off into the
municipal dump as normal waste. The aforesaid pit may then the reused for further burial
of radioactive waste. However, the waste should be released only after careful monitoring.
1.4
If the identity of each radionuclide in the mixture is known but the activity of one or
more radio nuclides in the mixture is not known, the total activity limit for the mixture is
the limit specified in the table 10.5 for the radionuclide having the lowest limit. A log book
should be maintained for: recording the identity and quantity of each radioisotope and the
time of its disposal into the sanitary sewage system.
Used liquid scintillation cocktail should not be disposed off
in sanitary sewage system. It should be treated as a chemical
waste.
1.5
89 210
Sr 37 Po 3.7
90
Sr+ 90Y 3.7
94
Zr + 95Nb 370
INCINERATION
Incineration of combustible waste (such as foliage, crops carcasses, etc.) releases
part of the radioactivity to the atmosphere and the remaining retained in ash. However,
incineration should be done under controlled conditions to ensure that gaseous
radioactivity does not affect the immediate environment and the ashes collected for
separate disposal as solid waste. Since these are difficult to achieve in practice,
incineration is not generally recommended.
In case an incinerator facility is available, radioactive waste should be incinerated in
a separate batch and the ash collected by wet method separately for further pit-disposal.
1.6
36
CL 0.37 74
45
Ca 3.7 10.1
60
Co 0.37 37.0
89
Sr 0.37 11.1
90
Sr + 90Y 0.037 0.148
94
Zr + 95Nb 3.7 74
99
Mo + 99m Tc 3.7 185
106
Ru + 106Rh 0.37 14.8
124
Sb 0.37 25.9
125
I 3.7 22.2
131
I 3.7 22.2
137 137m
Cs + Ba 3.7 185
140
Ba + 140La 0.37 29.6
144
Ce 0.37 11.1
170
Tm 3.7 37.0
192
Ir 3.7 37.0
210
Po 0.037 0.74
Summary
1. For the safe disposal of radioactive wastes, the Central Government, under Atomic Energy
Act 1962 (33 of 1962, subsection of clause 30) has promulgated rules and these are called,
The Atomic Energy (safe disposal of radioactive wastes) Rules, 1987.
2. Solid and liquid wastes should be collected separately in suitable containers in the
laboratory itself for disposal.
3. Three guiding principles for the disposal of radioactive wastes are (i) concentrate and
contain, (ii) delay and discharge and (iii) dilute and discharge.
4. The basic objective of waste. management procedure is to ensure that it will not result in
exposure to man and his environment by more than the prescribed limits.
5. Short lived liquid radioactive waste, soluble or dispersible in water (including those
containing 3H and 14
C) should only be disposed off in sanitary sewage system. Organic
waste from liquid scintillation counting should-not be disposed off in sanitary sewage
system. It should be treated as chemical waste.
6. Short-lived solid radioactive wastes should be disposed off by burying in pits.
7. If separate incineration facility is used for the burning of combustible radioactive wastes,
the ash should be collected by wet method and disposed off as solid radioactive waste.
8. Any advice-regarding waste disposal should be obtained from the Head, RSD, AERB.
1.7
-----------
1.8