Aquis 500 CR
Aquis 500 CR
B 202565.0
Operating Instructions
V2.00/EN/00476321
WARNING:
A sudden malfunction of the devicedevice, or one of the sensors connected to it,
could potentially result in dangerous, imprecise dosing! Suitable preventive
measures must be in place to prevent this from happening.
Note:
Please read these Operating Instructions before placing the device in operation.
Keep the manual in a place which is accessible to all users at all times.
2 Description ................................................................................ 6
4 Mounting .................................................................................. 11
4.1 General .......................................................................................................11
4.2 Surface mounting .......................................................................................11
4.3 Pipe installation set / weather protection roof ...........................................12
4.4 DIN rail installation set ...............................................................................12
4.5 Mounting in a panel ....................................................................................13
6 Operation ................................................................................. 24
6.1 Displays and controls .................................................................................24
6.2 LC display ..................................................................................................25
6.3 Principle of operation .................................................................................26
6.4 Measurement mode ...................................................................................29
6.5 Input/output information ............................................................................29
6.6 MANUAL mode / simulation mode ............................................................31
6.7 HOLD mode ...............................................................................................35
6.8 Operator level .............................................................................................36
Contents
6.9 Administrator level .....................................................................................36
6.10 Device info .................................................................................................43
6.11 Controller functions ....................................................................................44
7 Commissioning ....................................................................... 45
7.1 Fast start ....................................................................................................45
7.2 Setup examples .........................................................................................46
8 Calibration ............................................................................... 66
8.1 General .......................................................................................................66
8.2 Calibrating the relative cell constant ..........................................................66
8.3 Cell constants ............................................................................................68
8.4 Calibrating the temp. coefficient of the sample solution ...........................68
8.5 Calibration logbook ....................................................................................71
11 Appendix .................................................................................. 75
11.1 Operator level parameters .........................................................................75
11.2 Parameter explanations .............................................................................81
11.3 Glossary .....................................................................................................85
13 Index ......................................................................................... 97
1 Typographical conventions
1.1 Warning signs
Danger
This symbol is used when there may be danger to personnel if the
instructions are disregarded or not followed accurately!
Caution
This symbol is used when there may be damage to equipment or data if the
instructions are ignored or not followed accurately!
Caution
This symbol is used where special care is required when handling components
liable to damage through electrostatic discharge.
abc1 Footnote
Footnotes are remarks that refer to specific points in the text.
Footnotes consist of two parts:
A marker in the text, and the footnote text.
The markers in the text are arranged as continuous superscript numbers.
✱ Action instruction
This symbol indicates that an action to be performed is described.
The individual steps are marked by this asterisk.
Example:
✱ Remove crosspoint screws.
5
2 Description
Block diagram
2 analog inputs Analog outputs (option)
Input 1: Output 1 + 2:
Conductivity / resistivity 0(4) 20 mA or 0 10 V
Setup interface
6
3 Identifying the device version
3.1 Nameplate
on the
transmitter JUMO AQUIS 500 CR TN: 00491200
Typ: 202565/10-888-000-000-000-23/000 Fulda, Germany
www.jumo.net
F-Nr.: 0168122901016010001
AC 110..240V -15/+10% 48..63Hz ≤ 14VA
The date of manufacture is coded in the 12th to 15th position (from the left) of
the “F-Nr.” (serial number):
1601 means manufactured in year 2016 / week 01
7
3 Identifying the device version
3.2 Type designation
(1) Basic type
202565 JUMO AQUIS 500 CR
Transmitter/controller for conductivity, TDS, resistivity and temperature
(2) Basic type extensions
10 for panel mounting
20 in surface-mountable housing
8
3 Identifying the device version
3.1 Accessories (in delivery package)
Contents Designation
4 x fixings
1 x cable cover
9
3 Identifying the device version
3.4 Accessories (optional)
Type Part no.
Protection cover for JUMO AQUIS 500 1
00398161
2
Pole-mounting kit for JUMO AQUIS 500 00483664
Support pillar with pedestal base, arm and chain 00398163
PC setup software 00483602
PC interface, including USB/TTL converter and 00456352
adapter (USB connecting cable)
Fixing for suspension fitting 00453191
1
The pole-mounting kit is needed for mounting the protection cover.
2 By using the pole-mounting kit, the JUMO AQUIS 500 can be fitted to a pole
(e.g. support pillar or railing).
10
4 Mounting
4.1 General
Mounting Find a location that ensures easy accessibility for the later calibration.
location The fastening must be secure and must ensure low vibration for the device.
Avoid direct sunlight!
Permissible ambient temperature at the installation location: -10 to 55°C with
max. 95% rel. humidity, no condensation.
Installation The device can be mounted in any position.
position
30
60
(1)
149
134 77 94
23
94
140
161
120
11
4 Mounting
4.3 Pipe installation set / weather protection roof
The pipe installation set for JUMO AQUIS 500 (part no.: 00483664) can be
used to fasten the device (and optionally the protective roof for JUMO AQUIS
500, part no.: 00398161) onto pipes or railings with a diameter from 30 to 50
mm.
ø30-50
(1)
(2)
(1)
12
4 Mounting
4.5 Mounting in a panel
Drilling template See section 12.2 "Panel cut-out", page
95.
The panel must be sufficiently thick to achieve the
specified IP65 enclosure protection!
(1)
(5)
(4)
(3)
(2)
✱ Prepare the panel cut-out and holes based on the drill template.
✱ Place the control panel (1) with gasket (2) in the panel cut-out and fasten it
with screws (2) spacing rollers (4) and nuts (5).
13
4 Mounting
(1)
(2)
(3)
Depth behind
panel 44
14
5 Electrical connection
5.1 Installation notes
The electrical connection must only be carried out
by qualified professional persons.
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of
Power Circuits with Nominal Voltages below 1000 V” or the appropriate local
regulations. Only flexible cables and wires shall be used!
If contact with live parts is possible while working on the device, it must be
completely disconnected from the electrical supply.
Load circuits must be fused for the maximum relay current in each case, in
order to prevent welding of the relay contacts in the event of a short circuit.
The electromagnetic compatibility conforms to EN 61326.
Run input, output and supply cables separately and not parallel to one
another.
Use shielded sensor cables with twisted conductors. Do not run these cables
close to current-carrying components or cables. Ground shielding at one end.
Sensor leads should be implemented as uninterrupted cables (not routed via
terminal blocks etc.).
Do not connect any additional loads to the supply terminals of the device.
The device is not suitable for use in areas with an explosion hazard (Ex areas).
Apart from faulty installation, incorrect settings on the device may also affect
the proper functioning of the subsequent process or lead to damage. Safety
devices independent of the device should therefore always be provided and
should only be capable of adjustment by specialist personnel.
The IP67 enclosure protection for the device will only be achieved if not more
than one cable per cable fitting is led into the device.
15
5 Electrical connection
5.2 Electrical isolation
Setup interface
3700 V AC
30 V AC
50 V DC
30 V AC
50 V DC
1
3700 V AC
Supply voltage
16
5 Electrical connection
5.3 Preparatory work
(1)
(2)
Opening the ✱ Prior to opening, loosen all cable fittings (2) so that the cables are
device moveable.
✱ Push connection cable a little into the case so that enough cable reserve is
available for opening.
✱ Loosen the 4 front-panel screws (1) of the case lid and pull them out as
much as possible.
✱ Pull the lid to the front and then fold to the front. The user needs to be able
to easily open the lid. Do not use force while opening!
Closing the ✱ When closing the device, pull the connecting cables to the outside while
device the cable fittings are in a released state and make sure that the lines in the
inside of the device run properly. Pay attention to the corresponding
sheathing measurement to ensure strain relief and protection type (IP67) of
the cable fitting.
✱ The user must be able to close the lid with the 4 screws without a high
degree of pressure.
✱ Tighten cable fittings.
17
5 Electrical connection
5.4 Connection of conductivity cells
Items needed
from the
accessories
bag1:
3
4 1 1 2
1 1
2 2
Fabricating the
connecting The connecting cable between sensor and transmitter must be a
cable shielded cable with a maximum diameter of 8 mm.
60 8
55
18
5 Electrical connection
✱
~ 20
✱ Insulate the screen at the cable end with heat shrink tube.
✱ Apply core ferrules to the ends of the conductors.
Core ferrule dimensions see Chapter 5.1 “Installation notes”, page 15.
(1)
The connection cable between sensor and transmitter must be a shielded cable
with a diameter of 8 mm max.
The device contains a guide plate that ensures an optimum cable path. After
laying the cables, the cable cover (1) must be attached until it clicks, like
shown above. This is important to ensure the electrical safety!
The sensor cables are run to the plug-in screw terminals and must have a strain
relief.
The cable clip (3) must only be screwed down (see next page) by a 3.5x6.5 pan
head screw! A longer screw may cause a hazardous voltage to be contacted to
the shielding!
19
5 Electrical connection
5.4.1 Conductivity cell with 2-electrode system
(3)
(1)
(2)
Sensor
connection
Row 1
Row 2
SC = shielding (screen)
ϑ
Resistance thermometer
Conductivity cell
20
5 Electrical connection
5.4.2 Conductivity cell with 4-electrode system
(3)
(1)
(2)
Sensor
connection
Row 1
Row 2
SC = shielding (screen)
ϑ
Resistance thermometer
Conductivity cell
21
5 Electrical connection
5.5 Terminal assignments
Connection Screw Row
terminals
Supply voltage
Supply voltage (23): 1 N (L-)
110 — 240 V AC -15/+10%, 48 — 63 Hz 2 L1 (L+)
Supply voltage (25):
20 — 30 V AC/DC, 48 — 63 Hz
1
Supply voltage (30):
12 — 24 V DC +/-15% (permissible only for connection to
SELV/PELV circuits)
NC 3
Inputs
Conductivity cell (2-electrode system) 1 1
Terminals 1+2 and 3+4 are linked in the device; 2
2
2-wire cable routed to the head of the conductivity cell. 3
3 4
For concentric cells, terminal 1 is connected to the outer
4
electrode.
Conductivity cell (2-electrode system) 1 1
Wiring for the highest accuracy; 2
2
4-wire cable routed to the head of the conductivity cell. 3
For concentric cells, terminal 1 is connected to the outer 3 4
electrode. 4
10
Resistance thermometer in 3-wire circuit 9 8
9
ϑ
8 10
10
22
5 Electrical connection
Connection Screw Row
terminals
Binary input 11 11
12
2
12
Outputs
Analog output 1 + 13
0 — 20 mA resp. 20 — 0 mA, or - 14
4 — 20 mA resp. 20 — 4 mA,
or
0 — 10 V resp. 10 — 0V
(electrically isolated)
2
Analog output 2 + 15
0 — 20 mA resp. 20 — 0 mA, or - 16
4 — 20 mA resp. 20 — 4 mA,
or
0 — 10 V resp. 10 — 0V
(electrically isolated)
Switching output K1 5
Pole 4
(floating) Break
4
(SPST-NC) 5
Make
(SPST-NO) 6
6
1
Switching output K2 9
Pole 8
(floating) Break
8
(SPST-NC) 9
Make
10
(SPST-NO) 10
23
6 Operation
6.1 Displays and controls
(1)
(2)
(3)
(1) LC display
(2) Control panel with 5 keys
(3) Five cable glands (max.)
24
6 Operation
6.2 LC display
6.2.1 Measurement mode (normal display)
(10)
(7)
(9) (8)
(11)
25
6 Operation
6.3 Principle of operation
Meas. mode
or
PGM
> 3 sec
PGM EXIT
CTRL. SETPOINTS
< 2 sec SETPOINT 1
SETPOINT 2 or timeout
(adjustable)
MIN/MAX VALUES
PGM
< 2 sec
EXIT
OUTPUT LEVEL
Controller 1
Controller 2
PGM
<2s
EXIT
MANUAL MODE
OVERVIEW
PGM
< 3 sec
Operator EXIT
level
or timeout (adjustable)
Administrator EXIT
level
or timeout (adjustable)
Calibration EXIT
level
or timeout (adjustable)
EXIT
Calibration
logbook or timeout (adjustable)
EXIT
Device info
or timeout (adjustable)
26
6 Operation
6.3.1 Operation in levels
Measurement mode (normal display); see Chapter 6.4 “Measurement mode”, page 29
CTRL. SETPOINTS
MIN/MAX values
see Chapter 6.5.1 “MIN/MAX
values”, page 29
Password
PARAMETER LEVEL, see Chapter 6.9.2 “Parameter level”, page 38
CONDUCT. INPUT
TEMPERATURE INPUT
BINARY INPUT
CTRL. CHAN. 1
CTRL. CHAN. 2
CTRL.SPEC.FUNCT.
SWITCH OUTPUT 1
SWITCH OUTPUT 2
ANALOG OUTPUT 1
ANALOG OUTPUT 2
DISPLAY
27
6 Operation
Measurement mode ADMINISTRATOR LEVEL
BASIC SETTINGS, see Chapter 6.9.4 “Basic settings”, page 41
CELL TYPE
CELL CONSTANT
PROBE BREAK DETECT
OPERATING MODE
TEMP. COMP.
TEMP. COEFF.
RANGE 1 UNIT
RANGE 1 DEC. POINT
AUTORANGE
RANGE 2 UNIT
RANGE 2 DEC. POINT
NEW DEVICE
INITIALIZE
CALIB. ENABLE
LINEAR TEMP.CO.
ENABLE
REL. CELL CONSTANT
ENABLE
DELETE LOGBOOK
REALLY DELETE LOGBOOK?
CALIB. LEVEL
LINEAR TEMP.CO.
REL. CELL CONSTANT
CALIB. LOGBOOK
DEVICE INFO
CELL TYPE
CELL CONSTANT
PROBE BREAK DETECT
OPERATING MODE
TEMP. COMP.
TEMP. COEFF.
RANGE 1 UNIT
RANGE 1 DEC. POINT
AUTORANGE
RANGE 2 UNIT
RANGE 2 DEC. POINT
28
6 Operation
6.4 Measurement mode
6.4.1 Normal display
Presentation The normal display shows either the conductivity (compensated for the
reference temperature) or the concentration and temperature of the medium
being measured.
(1)
(3)
(2)
In measurement mode, the display types “Trend display” and “Bar graph” can
also be selected, see “MEAS. DISPLAY TYPE”, page 82
29
6 Operation
30
6 Operation
6.6 MANUAL mode / simulation mode
These functions can be used to set the switching outputs and analog outputs
of the device manually to a defined state. This facilitates dry startup,
troubleshooting and customer service.
"Higher order"
controller
Select manual
mode
In the factory setting of the device the MANUAL mode parameter is disabled
and can only be activated by the administrator!
This parameter must first be enabled for other users, see Chapter 6.9.3
“Enable level”, page 38.
31
6 Operation
Locked = No manual mode, control via JUMO AQUIS 500.
Momentary =the outputs are active as long as the or key is pressed.
action
Switching = the outputs are active if the or key is pressed. If the
corresponding key is pressed again, the output becomes inactive
again.
If the EXIT and keys are pressed for longer than 3 seconds, the device goes
into HOLD mode.
Then the outputs of the device respond according to the default settings.
To exit HOLD mode, press the EXIT and keys for longer than 3 seconds.
Control is no longer through the JUMO AQUIS 500. The output level of the
controller channels is 0%.
Controller channel 1 is activated by the key. In this case the output level of
controller channel 1 is 100%.
Controller channel 2 is activated by the key. In this case the output level of
controller channel 2 is 100%.
Overview of You can display which outputs and/or controllers are in MANUAL mode.
MANUAL/ The device is in "normal display" mode.
Simulation Press the PGM key several times for less than 2 seconds (the number of times
mode varies depending on the equipment and configuration of the device).
32
6 Operation
Activate
simulation
In the factory setting of the device the MANUAL mode parameter is set to "No
simulation" and can only be activated by the administrator!
This parameter must first be enabled for other users, see Chapter 6.9.3
“Enable level”, page 38.
33
6 Operation
6.6.3 Simulation of analog outputs via MANUAL mode
Enabling and ✱ Select activation of simulation of the actual value output:
activation Administrator level / Password / Parameter level / Analog output 1 or 2 /
Simulation / Off or On.
With "On" the output takes on the value of the "Simulation value" parameter.
When the JUMO AQUIS is in display mode, the word MANUAL appears in the
status line of the display.
34
6 Operation
6.7 HOLD mode
In HOLD status the outputs take on the states programmed in the relevant
parameter (controller channel, switching output or analog output).
This function can be used to "freeze" switching outputs and the analog
outputs of the device. This means the current status of the output will be
retained even when the measured value changes. Control is not via the device.
HOLD mode can be activated by pressing the key or by the binary input.
Activation by ✱ Press and hold the EXIT and keys longer than 3 seconds.
pressing key Then the outputs of the device respond according to the default settings.
The word HOLD appears in the status line of the display.
If the EXIT and keys are pressed for less than 3 seconds, the device goes
into manual mode.
Then the outputs of the device respond according to the default settings.
Pressing a key ✱ Press the EXIT and keys for longer than 3 seconds.
to deactivate
HOLD mode
If the EXIT and keys are pressed for less than 3 seconds, the device goes
into Manual mode.
Then the outputs of the device respond according to the default settings.
Control is through the outputs of the device again. The word MANUAL
disappears from the status line of the display.
35
6 Operation
6.8 Operator level
All the parameters that have been enabled by the administrator (Administrator
level, see “Administrator level”, page 36) can be edited in this level. All other
parameters (marked by a key ) can only be read.
✱ Press the PGM key for at least 3 seconds.
✱ Select OPERATOR LEVEL.
For operator level parameters and their explanations, see Chapter 11.1
“Operator level parameters”, page 75 ff.
36
6 Operation
6.9.1 Administrator levels
ADMIN. LEVEL
PASSWORD
0
300 PGM
EXIT
PARAMETER
LEVEL or
timeout
(adjustable)
EXIT
ENABLE
LEVEL or
timeout
EXIT
BASIC
SETTINGS or
timeout
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
37
6 Operation
6.9.2 Parameter level
Here you can make the same settings as at the operator level. Since the user
has administrator rights in this case, parameters can also be altered that
would be locked at the operator level.
For the list of adjustable parameters, see Chapter 6.8 “Operator level”, page
36ff.
TEMPERATURE IN
Sensor type
Unit
Manual temperature
Filter time constant
Offset
BINARY INPUT
No function
Key inhibit
Hold operation
Alarm Stop
38
6 Operation
Reset time
Derivative time
Pulse period
Min. ON time
Output level limit
Max. pulse frequency
Hysteresis
Pull-in delay
Drop-out delay
Controller alarm
In Hold mode
In event of error
MAX process value
MIN process value
CNTRL.SPEC.FUNCT. (Reglersonderfunktion)
I switch-off
separate controllers
Manual mode
39
6 Operation
DISPLAY
Language
Lighting
LCD inverse
Meas. display type
Upper display
Lower display
Bar graph calibration start
Bar graph calibration end
MIN/MAX reset
Operator timeout
Contrast
40
6 Operation
6.9.4 Basic settings
The basic settings for the device are defined at this level. The parameters are
altered by and keys. Use the PGM key to select the next parameter.
Cell type
2-wire 4-wire
Contamination
dectection
Off / On
Off / On
Operating mode
Unit Range 1
Autorange 1 Decimal places Range 1
Customized / / ppm /
µS/cm / mS/cm / XXXX / XXX.x /
Off On kOhm*cm / MOhm*cm XX.xx / X.xxx
Display format
Range 2
XXXX / XXX.x /
XX.xx / X.xxx
Re-initialize instrument
No Yes
41
6 Operation
If you leave the "Basic settings" level with EXIT, all changes will be discarded
and the previous settings will be restored.
see Chapter 8.4 “Calibrating the temp. coefficient of the sample solution”,
page 68.
see Chapter 8.2 “Calibrating the relative cell constant”, page 66.
42
6 Operation
6.9.7 Delete logbook
REALLY DELETE LOGBOOK ?
YES / NO
43
6 Operation
6.11 Controller functions
Simple In the JUMO AQUIS 500, simple switching functions, such as alarm contacts
switching and limit comparators or the signal from the calibration timer, are configured at
functions the parameter level, through the parameters for “Switching output 1 or 2”.
The parameters for the controller channel 1 or 2 respectively must then be set
to “Off”.
Higher-level Higher-level control functions are configured at the parameter level, through
control the parameters for “Controller channel 1 or 2”.
functions The controller channel parameter must then be set to “Controller 1 or 2”.
Operator level
parameters Switching output 1 / 2 Explanation
no switching function and
none
no control function required
Controller 1 the device should have “higher-level” control
Controller 2 the device should have “higher-level” control
Controller alarm 1 / 2
Controller alarm
main variable
main variable
main variable
main variable
Temperature
Temperature “simple” switching functions
Temperature
Temperature
Sensor fault
Calibration timer
Autorange
USP
USP pre-alarm
PH.-EUR
PH.-EUR pre-alarm
Controller channel 1 / 2
Limit value
Pulse width
Pulse frequency “higher-level” control functions
Continuous
Modulating
must be selected if “simple” switching
Off
functions are required
44
7 Commissioning
7.1 Fast start
This is a recommendation for configuring the device reliably in a
short time.
If you check the setting options from this list before starting the
configuration, you can avoid timeouts during configuration.
45
7 Commissioning
7.2 Setup examples
7.2.1 Measuring drinking water with a 2-electrode cell
Measurement range: 0 — 1000 µS/cm
Display: No decimal place
Cell constant K: 1.0 1/cm
Output signal: 4 — 20 mA
Temperature compensation: Linear
Temperature measurement: By Pt100
Controller function: Limit controller, MAX function
Limit: 600 µS/cm
Call up
administrator
level ✔ Meas. mode
or
PGM
>3s
PGM EXIT
CTRL. SETPOINTS
< 2 sec SETPOINT 1
SETPOINT 2 or timeout
(adjustable)
MIN/MAX VALUES
PGM
< 2 sec
EXIT
OUTPUT LEVEL
Controller 1
Controller 2
PGM
< 2 sec
EXIT
MANUAL MODE
OVERVIEW
PGM
< 3 sec
Operator EXIT
✔ level
or timeout (adjustable)
✔ Administrator EXIT
PGM
level
or timeout (adjustable)
continue on
next page
Calibration EXIT
level
or timeout (adjustable)
EXIT
Calibration
logbook or timeout (adjustable)
EXIT
Device info
or timeout (adjustable)
46
7 Commissioning
Call up basic
settings
ADMIN. LEVEL
✔ PASSWORD
0
300 PGM
EXIT
PARAMETER
✔ LEVEL or
timeout
(adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
✔ BASIC EXIT
PGM
SETTINGS or
continue on timeout
next page
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
47
7 Commissioning
Basic settings
for main input:
✔ Cell type
Contamination detection
Off / On
Off / On
✔ Operating mode
✔
None
Temp. comp. Temp. coef. Input variable Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm / Customized / ppm 0.0 – 2.20 – 8.0%/°C 0.0 – 2.20 – 8.0%/°C
✔
Decimal places Range 1 Decimal places Range 1 Factor Unit Range 1
1
Customized / ppm /
XXXX / XXX.x / XXXX / XXX.x / µS/cm / mS/cm /
XX.xx / X.xxx XX.xx / X.xxx 0.01 — 0.67 — 2.00 kOhm*cm / MOhm*cm
✔ Autorange
Unit Range 1
1
Customized / ppm /
Decimal places Range 1
µS/cm / mS/cm / XXXX / XXX.x /
Off On kOhm*cm / MOhm*cm XX.xx / X.xxx
Display format
Range 2
XXXX / XXX.x /
XX.xx / X.xxx
✔ Re-initialize instrument
No Yes
48
7 Commissioning
Call up the
parameter level
ADMIN. LEVEL
PASSWORD
0
300 PGM
PGM ✔ PARAMETER
EXIT
LEVEL or
continue on
timeout
next page (adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
EXIT
BASIC
✔ SETTINGS or
timeout
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
49
7 Commissioning
Concluding device settings
50
7 Commissioning
7.2.2 Measuring highly-purified water with a 2-electrode cell
Measurement range: 0 — 2 µS/cm
Display: 2 decimal places
Cell constant K: 0.01 1/cm
Output signal: 0 — 20 mA
Temperature compensation: Off
Temperature measurement: By Pt100
Controller function: USP contact, as per USP <645>
Limit: As per table, see “Temperature compensation”,
page 88
Call up
administrator
level ✔ Meas. mode
or
PGM
>3s
PGM EXIT
CTRL. SETPOINTS
< 2 sec SETPOINT 1
SETPOINT 2 or timeout
(adjustable)
MIN/MAX VALUES
PGM
< 2 sec
EXIT
OUTPUT LEVEL
Controller 1
Controller 2
PGM
< 2 sec
EXIT
MANUAL MODE
OVERVIEW
PGM
< 3 sec
Operator EXIT
✔ level
or timeout (adjustable)
✔ Administrator EXIT
PGM
level
or timeout (adjustable)
continue on
next page
Calibration EXIT
level
or timeout (adjustable)
EXIT
Calibration
logbook or timeout (adjustable)
EXIT
Device info
or timeout (adjustable)
51
7 Commissioning
Call up basic
settings
ADMIN. LEVEL
✔ PASSWORD
0
300 PGM
EXIT
PARAMETER
✔ LEVEL or
timeout
(adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
✔ BASIC EXIT
PGM
SETTINGS or
continue on timeout
next page
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
52
7 Commissioning
Basic settings ✔ Cell type
for the
main input: 2-wire 4-wire
procedure
✔ Cell constant Cell constant
0.01 / 0.1 / 0.5 /
1.0 / 3.0 / 10.0 0.5 /1.0
Contamination detection
Off / On
Off / On
✔ Operating mode
✔
None
Temp. comp. Temp. coef. Input variable Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm / Customized / ppm 0.0 2.20 8.0%/°C 0.0 2.20 8.0%/°C
✔
Decimal places Range 1 Decimal places Range 1 Factor Unit Range 1
1
Customized / ppm /
XXXX / XXX.x / XXXX / XXX.x / µS/cm / mS/cm /
XX.xx / X.xxx XX.xx / X.xxx 0.01 0.67 2.00 kOhm*cm / MOhm*cm
✔ Autorange
Unit Range 1
1
Customized / ppm /
Decimal places Range 1
µS/cm / mS/cm / XXXX / XXX.x /
Off On kOhm*cm / MOhm*cm XX.xx / X.xxx
Display format
Range 2
XXXX / XXX.x /
XX.xx / X.xxx
✔ Re-initialize instrument
No Yes
53
7 Commissioning
Call up
parameter
level ADMIN. LEVEL
PASSWORD
0
300 PGM
PGM ✔ PARAMETER
EXIT
LEVEL or
continue on
timeout
next page (adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
EXIT
BASIC
✔ SETTINGS or
timeout
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
54
7 Commissioning
Concluding device settings
55
7 Commissioning
7.2.3 Measuring highly-purified water with a 2-electrode cell
Measurement range: 0 — 20 MΩ x cm
Display: 2 decimal places
Cell constant K: 0.01 1/cm
Relative cell constant
as per test report for the cell: 102.5%
Output signal: 4 — 20 mA
Temperature compensation: ASTM
Temperature measurement: By Pt100
Controller function: Off
Limit: Not applicable
Call up
administrator
level ✔ Meas. mode
or
PGM
>3s
PGM EXIT
CTRL. SETPOINTS
< 2 sec SETPOINT 1
SETPOINT 2 or timeout
(adjustable)
MIN/MAX VALUES
PGM
< 2 sec
EXIT
OUTPUT LEVEL
Controller 1
Controller 2
PGM
< 2 sec
EXIT
MANUAL MODE
OVERVIEW
PGM
< 3 sec
Operator EXIT
✔ level
or timeout (adjustable)
✔ Administrator EXIT
PGM
level
or timeout (adjustable)
continue on
next page
Calibration EXIT
level
or timeout (adjustable)
EXIT
Calibration
logbook or timeout (adjustable)
EXIT
Device info
or timeout (adjustable)
56
7 Commissioning
Call up basic
settings
ADMIN. LEVEL
✔ PASSWORD
0
300 PGM
EXIT
PARAMETER
✔ LEVEL or
timeout
(adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
✔ BASIC EXIT
PGM
SETTINGS or
continue on timeout
next page
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
57
7 Commissioning
Basic settings ✔ Cell type
for main input:
procedure 2-wire 4-wire
Contamination detection
Off / On
Off / On
✔ Operating mode
✔
None
Temp. comp. Temp. coef. Input variable Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm / Customized / ppm 0.0 – 2.20 – 8.0%/°C 0.0 – 2.20 – 8.0%/°C
✔
Decimal places Range 1 Decimal places Range 1 Factor Unit Range 1
1
Customized / ppm /
XXXX / XXX.x / XXXX / XXX.x / µS/cm / mS/cm /
XX.xx / X.xxx XX.xx / X.xxx 0.01 — 0.67 — 2.00 kOhm*cm / MOhm*cm
✔ Autorange
Unit Range 1
1
Customized / ppm /
Decimal places Range 1
µS/cm / mS/cm / XXXX / XXX.x /
Off On kOhm*cm / MOhm*cm XX.xx / X.xxx
Display format
Range 2
XXXX / XXX.x /
XX.xx / X.xxx
✔ Re-initialize instrument
No Yes
58
7 Commissioning
Call up
parameter
level ADMIN. LEVEL
PASSWORD
0
300 PGM
PGM ✔ PARAMETER
EXIT
LEVEL or
continue on
timeout
next page (adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
EXIT
BASIC
✔ SETTINGS or
timeout
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
59
7 Commissioning
Concluding device settings
60
7 Commissioning
7.2.4 Measuring with autorange and a 2-electrode cell
Call up
administrator
level ✔ Meas. mode
or
PGM
>3s
PGM EXIT
CTRL. SETPOINTS
< 2 sec SETPOINT 1
SETPOINT 2 or timeout
(adjustable)
MIN/MAX VALUES
PGM
< 2 sec
EXIT
OUTPUT LEVEL
Controller 1
Controller 2
PGM
< 2 sec
EXIT
MANUAL MODE
OVERVIEW
PGM
< 3 sec
Operator EXIT
✔ level
or timeout (adjustable)
✔ Administrator EXIT
PGM
level
or timeout (adjustable)
continue on
next page
Calibration EXIT
level
or timeout (adjustable)
EXIT
Calibration
logbook or timeout (adjustable)
EXIT
Device info
or timeout (adjustable)
61
7 Commissioning
Call up basic
settings
ADMIN. LEVEL
✔ PASSWORD
0
300 PGM
EXIT
PARAMETER
✔ LEVEL or
timeout
(adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
✔ BASIC EXIT
PGM
SETTINGS or
continue on timeout
next page
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
62
7 Commissioning
Basic settings ✔ Cell type
for main input:
procedure 2-wire 4-wire
Contamination detection
Off / On
Off / On
✔ Operating mode
✔
None
Temp. comp. Temp. coef. Input variable Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm / Customized / ppm 0.0 – 2.20 – 8.0%/°C 0.0 – 2.20 – 8.0%/°C
✔
Decimal places Range 1 Decimal places Range 1 Factor Unit Range 1
1
Customized / ppm /
XXXX / XXX.x / XXXX / XXX.x / µS/cm / mS/cm /
XX.xx / X.xxx XX.xx / X.xxx 0.01 — 0.67 — 2.00 kOhm*cm / MOhm*cm
✔ Autorange
Unit Range 1
1
Customized / ppm /
Decimal places Range 1
µS/cm / mS/cm / XXXX / XXX.x /
Off On kOhm*cm / MOhm*cm XX.xx / X.xxx
Display format
Range 2
XXXX / XXX.x /
XX.xx / X.xxx
✔ Re-initialize instrument
No Yes
63
7 Commissioning
Call up
parameter
level ADMIN. LEVEL
PASSWORD
0
300 PGM
PGM ✔ PARAMETER
EXIT
LEVEL or
continue on
timeout
next page (adjustable)
EXIT
ENABLE
✔ LEVEL or
timeout
EXIT
BASIC
✔ SETTINGS or
timeout
CALIBRATION EXIT
LEVEL
or
timeout
CALIB. EXIT
ENABLE
or
timeout
DELETE EXIT
LOGBOOK
or
timeout
64
7 Commissioning
Concluding device settings
65
8 Calibration
8.1 General
The device offers several calibration options to increase the precision.
66
8 Calibration
- The transmitter must be in the measurement mode.
67
8 Calibration
✱ The present measured value and temperature will be displayed.
4-electrode
systems Cell constant Setting range for Resulting usable
[1/cm] relative cell constant range [1/cm]
0.5 20 — 150% 0.1 — 0.75
1.0 0.2 — 1.5
68
8 Calibration
✱ Immerse the conductivity sensor in a sample of the solution to be
measured.
(1)
(2)
69
8 Calibration
✱ Press the PGM key.
The LC display now shows the conductivity on the right (399µS/cm) at the
current temperature (24.3°C).
On the left, you can see the other temperatures – T1 (25°C) and
T2 (70.0°C) – that are to be used.
✱ Press the PGM key.
✱ Warm up the sample medium until it reaches the working temperature.
When the temperature of the medium has risen above T2 (73.0°C), the device
calculates the temperature coefficient.
The LC display now shows the derived temperature coefficient in %/°C.
70
8 Calibration
The transmitter is now in the measurement mode, and shows the
compensated conductivity of the solution.
71
9 Setup program
9.1 Function
Configurable The setup program (available as an option) can be used for easy adaptation of
parameters the device to the requirements.
- Setting the measurement range and the range limits.
- Setting the response of the outputs to an out-of-range signal.
- Setting the functions of the switching outputs K1 and K2.
- Setting the function of the binary input E1.
- Setting up special functions (e.g. tables for specific linearizations).
- etc.
Connection
The setup interface is not electrically isolated.
When connecting the PC interface cable with a TTL/RS232
converter and adapter (serial connection cable) (00350260) , it is
therefore absolutely essential to ensure that the supply for either
the transmitter or the PC is not electrically earthed (for instance:
use a battery-powered notebook).
The PC interface cable with USB/TTL converter (USB connection
cable) (00456352) is, however, electrically isolated.
72
10 Eliminating faults and malfunctions
10.1 Possible faults
t
73
10 Eliminating faults and malfunctions
10.2 Checking the device
General The device is calibrated at the factory and is maintenance-free. If,
nevertheless, measurement deviations appear with no apparent cause, the
transmitter can be tested as follows.
Test sequence ✱ Prepare the conductivity test solution in a container of adequate size.
✱ Connect up the device, see Chapter 5 “Electrical connection”, page 15.
✱ Connect up the device, see Chapter 5 “Electrical connection”, page 15.
✱ Carry out the basic settings, see Chapter 6.9.4 “Basic settings”, page 41.
- Set the temperature coefficient to 0%/°C.
- Select the range appropriate to the conductivity test solution.
- Conclude the basic settings with “Re-initialize the device”.
✱ Set the temperature coefficient for the reference device to 0%/°C as well
(see operating instructions for the reference device).
If this is not possible, then the sample liquid must be tempered to the
reference temperature for the reference device.
✱ Immerse the cell under test and the cell for the reference device in the
container, and do not move them during the measurement.
✱ The output and display of the device under test and the reference device
must match within the acceptable tolerance limits.
✱ If necessary, calibrate the relative cell constant, see “Calibrating the
relative cell constant”, page 66.
74
11 Appendix
11.1 Operator level parameters
If a number of device parameters have to be modified in the device, then it is
advisable to note them in the table below, and then modify these parameters
in the sequence given.
The following list shows the maximum number of parameters that
can be altered. Depending on the configuration, some of the
parameters will not be visible, i.e. not alterable (editable) for your
device.
75
11 Appendix
Parameter Selection / value range New
Factory setting setting
Proportional band 0 — 9999
Reset time 0 — 9999
Derivative time 0 — 999
Pulse period 2.5 — 20 — 999.5
Minimum ON time 0.5 — 999.5
Output level limit 0 — 100%
Maximum pulse frequency 0 — 60 1/min
Hysteresis (differential) 0 — 200 — 9999
Pull-in delay 0.00— 999.5 sec
Drop-out delay 0.00— 999.5 sec
Controller alarm OFF
ON
Alarm tolerance 0,00...1,00...16,00
Alarm delay 0...9999
In Hold mode FROZEN
0%
100%
In event of error FROZEN
0%
100%
MAX. setpoint 0 — 9999
MIN setpoint 0 — 9999
Controller channel 2
Controller type LIMIT
PULSE WIDTH
PULSE FREQ.
CONTINUOUS
OFF
1
Setpoint 0 — 9999
MIN / MAX contact MAX CONTACT
(increasing / decreasing MIN CONTACT
characteristic)
Proportional band 0 — 9999
Reset time 0 — 9999
Derivative time 0 — 999
Pulse period 2.5 — 20 — 999.5
Minimum ON time 0.5 — 999.5
Output level limit 0 — 100%
Maximum pulse frequency 0 — 60 1/min
Hysteresis (differential) 0 — 200 — 9999
Pull-in delay 0.00— 999.5 sec
Drop-out delay 0.00— 999.5 sec
Controller alarm OFF
ON
Alarm tolerance 0,00...1,00...16,00
Alarm delay 0...9999
76
11 Appendix
Parameter Selection / value range New
Factory setting setting
In Hold mode FROZEN
0%
100%
In event of error FROZEN
0%
100%
MAX setpoint 0 — 9999
MIN setpoint 0 — 9999
Controller special function
I switch-off INACTIVE
ACTIVE
Separate controllers OFF
YES
Manual mode LOCKED
PULSED
SWITCHED
Switching output 1
Function NO FUNCTION
CONTROLLER 1
CONTROLLER 2
CTRLR ALARM 1
CTRLR ALARM 2
CTRLR ALARM
AF1 MAIN VAR.
AF2 MAIN VAR.
AF7 MAIN VAR.
AF8 MAIN VAR.
AF1 TEMP.
AF2 TEMP.
AF7 TEMP.
AF8 TEMP.
SENSOR ERROR
CALIB. TIMER
AUTORANGE
USP
USP PRE-ALARM
PH EUR.
PH EUR.PRE-AL
Manual mode NO SIMULATION
INACTIVE
ACTIVE
77
11 Appendix
Parameter Selection / value range New
Factory setting setting
Switching output 2
Function NO FUNCTION
CONTROLLER 1
CONTROLLER 2
CTRLR ALARM 1
CTRLR ALARM 2
CTRLR ALARM
AF1 MAIN VAR.
AF2 MAIN VAR.
AF7 MAIN VAR.
AF8 MAIN VAR.
AF1 TEMP.
AF2 TEMP.
AF7 TEMP.
AF8 TEMP.
SENSOR ERROR
CALIB. TIMER
AUTORANGE
USP
USP PRE-ALARM
PH EUR.
PH EUR.PRE-AL
Manual mode NO SIMULATION
INACTIVE
ACTIVE
Analog output 1
Signal selector MAIN VARIABLE
CONTROLLER 1
CONTROLLER 2
Signal type 0 — 20 mA
20 — 0 A
4 — 20 mA
20 — 4 mA
0 — 10 V
10 — 0 V
Scaling start 0.00 — 89.99 = e.g. 0 mA
Scaling end 99.99 — 0.00 = e.g. 20 mA
During calibration MOVING
FROZEN
SAFE VALUE
In event of error LOW
HIGH
FROZEN
SAFE VALUE
In Hold mode LOW
HIGH
FROZEN
SAFE VALUE
MOVING
78
11 Appendix
Parameter Selection / value range New
Factory setting setting
Safe value 0 — 22 mA
Simulation OFF
ON
Simulation value 0 — 22 mA
Analog output 2
Signal selector TEMPERATURE
CONTROLLER 1
CONTROLLER 2
Signal type 0 — 20 mA
20 — 0 mA
4 — 20 mA
20 — 4 mA
0 — 10 V
10 — 0 V
Scaling start -50.0 to 220°C = e.g. 0 mA
Scaling end 250 to -20°C = e.g. 20 mA
During calibration MOVING
FROZEN
SAFE VALUE
In event of error LOW
HIGH
FROZEN
SAFE VALUE
In Hold mode LOW
HIGH
FROZEN
SAFE VALUE
MOVING
Safe value 0 — 22 mA
Simulation OFF
ON
Simulation value 0 — 22 mA
Display
Language GERMAN
ENGLISH
FRENCH
Lighting DURING OPERATION
OFF
LCD inverse OFF
ON
Meas. display type NORMAL
TREND
BAR GRAPH
79
11 Appendix
Parameter Selection / value range New
Factory setting setting
Lower display TEMPERATURE
OUTP. LEVEL 1
OUTP. LEVEL 2
SETPOINT 1
SETPOINT 2
NONE
COMPENSATED
UNCOMPENSATED
Upper display COMPENSATED
UNCOMPENSATED
TEMPERATURE
OUTP. LEVEL 1
OUTP. LEVEL 2
SETPOINT 1
SETPOINT 2
NONE
MIN/MAX reset NO
YES
Interface timeout 0 — 1— 10 min
Contrast 0 — 5 — 20
80
11 Appendix
11.2 Parameter explanations
TEMP. COMPENSATION
LINEAR
NATURAL WATER (permissible temperature range 0 to 36°C
as per EN 27 888)
ASTM D1125-95 (permissible temperature range 0 to 100°C)
FUNCTION
NO FUNCTION
Alarm window AF1 MAIN VAR.
Alarm window AF2 MAIN VAR.
Limit function AF7 MAIN VAR.
Limit function AF8 MAIN VAR.
Alarm window AF1 TEMP.
Alarm window AF2 TEMP.
Limit function AF7 TEMP.
Limit function AF8 TEMP.
SENSOR ERROR
CALIB. TIMER
HySt HySt
1 1
0 0
x x
AF AF
w w
Alarm window AF1 Alarm window AF2
HySt HySt
1 1
0 0
x x
w w
Limit function AF7 Limit function AF8
81
11 Appendix
1 1
0 0
t t
Pulse contact Pulse contact
1 1
0 0
t t
tP tP
0 Off t Time
1 On tP Pulse duration
AL Spacing w Setpoint / Limit
HySt Hysteresis x Actual value / Measurement
value
NORMAL In the normal display, the conductivity is shown (compensated for the
reference temperature) or the concentration and temperature of the medium
being measured.
(1)
(3)
(2)
82
11 Appendix
TREND The operator can quickly recognize in which direction the measurement is
changing.
rising falling
stable
strong medium slightly slightly medium strong
83
11 Appendix
✱ Confirm selection with PGM .
LOWER DISPLAY
.
(1)
(3)
(2)
This parameter is only available for the measurement display types NORMAL
and TREND.
TEMPERATURE
OUTP. LEVEL 1
OUTP. LEVEL 2
SETPOINT 1
SETPOINT 2
NONE
COMPENSATED
UNCOMPENSATED
UPPER DISPLAY
This parameter is only available for the measurement display types NORMAL
and TREND.
COMPENSATED
UNCOMPENSATED
TEMPERATURE
OUTP. LEVEL 1
OUTP. LEVEL 2
SETPOINT 1
SETPOINT 2
NONE
84
11 Appendix
11.3 Glossary
TDS
Display / control with ppm for the unit.
In this mode, the specific TDS factor can be entered in addition.
TDS (Total Dissolved Solids, i.e. filtrate residue).
This value is important, e.g. for ground water analysis and power station
engineering.
This value is also used as a measure of drinking water quality (e.g. in the USA,
Arab and Asian countries).
Various organizations have published limit values on this topic.
- WHO (World Health Organisation) <1000mg/liter
- USEPA (United States Environmental Protection Agency) <500mg/liter
The standard method of determination is gravimetric, i.e.
- filter a sample,
- evaporate the filtrate,
- weigh the residue.
Conductivity measurement is used for on-line measurement. The conversion
factor only has to be determined once. It corresponds to the ratio between the
conductivity value for the water and the value for the gravimetrically measured
filtrate residue. This factor varies from 0.55 to 1.0, and a typical value for
drinking water is about 0.67.
On modern devices, such as the JUMO AQUIS 500 CR, this factor can be
individually entered, thus enabling the most accurate measurement.
Customer-specific table
In this mode, the input value (conductivity or resistivity) can be displayed in
accordance with a table (up to 20 value pairs). Thanks to this function, it is
possible to implement simple concentration measurements, for example. The
values in the table can only be entered through the optional setup program.
Calibration The calibration timer indicates (if required) when the next routine calibration is
timer due. The calibration timer is activated by entering a number of days, after
which recalibration has to be carried out (plant or operator requirement).
85
11 Appendix
Detection of deposits
Deposit detection can be activated for 4-electrode cells.
During normal operation, it can happen that deposits form on electrodes. This
has the result that a lower concentration is displayed than actually present.
When the “deposit detection” function is activated, the device tells you when
the cell needs to be serviced.
Autorange
The availability of two measuring ranges can be advantageous for some
processes – rinsing or regeneration processes, for instance.
What is usually required here, is the precise acquisition of a low conductivity.
Rinsing/regeneration, however, involves a much higher conductivity, which
could lead to an out-of-range condition (error). This situation is not just
unsatisfactory, but may even be dangerous. Thanks to the autorange function,
two measuring ranges can be determined. The device then switches between
them in a defined manner.
Autorange can only be configured for the units mS/cm and µS/cm.
Range 1 must be smaller than range 2.
The controlling is only performed in range 1.
The process value output for range 2 is scaled to the full display span.
The changeover from range 1 to range 2 occurs when range 1 is exceeded.
The change back takes place when the process value goes below 90% of
range 1.
The change between the ranges can be signaled through the relay contact.
Pulse width controller (output is active for X > W and a P control structure)
10%
100% 90%
90%
Output level y
Switching period
50% 50%
50%
tOn tOff
10%
90%
10%
0%
Proportional band XP Process value X
0 X-W 1
XP
Setpoint W
If the process value X exceeds the setpoint W, the P controller will control
proportionally to the control deviation. On going outside the proportional
band, the controller operates with an output level of 100% (100% duty cycle).
86
11 Appendix
Pulse frequency controller (output is active for X > W and a P control structure)
100%
Output level y
Maximum pulse frequency
50%
50% of pulse frequency
0%
Proportional band XP No pulses Process value X
0 X-W 1
XP
Setpoint W
If the process value X exceeds setpoint W, the P controller will control
proportionally to the control deviation. On going outside the proportional
band, the controller operates with an output level of 100% (maximum
switching frequency).
Separate controllers
This function is normally deactivated (factory setting or "No" selection).
In the deactivated state, the software prevents the two controller outputs from
being able to work "against each other". So, for example, it is not possible to
dose acid and lye at the same time.
If the controllers are separate ("Yes" selection), each controller can be freely
configured.
I-component switch-off
This function is normally deactivated (factory setting or "No" selection).
In the deactivated state, the controller works in accordance with general con-
troller theory.
When I-component switch-off is activated ("Yes" selection), the part of the out-
put level that can be traced back to the I-component is set to zero when the
setpoint is reached.
This can be useful with mutual neutralization (acid and lye dosing both pos-
sible) in one treatment tank.
87
11 Appendix
Wash timer
The wash timer can be used to implement automated sensor cleaning. This
function is assigned to a switching output (1 or 2) for that purpose.
The cycle duration (cleaning interval) can be adjusted in the range from 1 to
240 hours. The wash duration (cleaning duration) is adjustable from 1 to 1800
seconds. During the wash duration the controller goes into the HOLD state,
which is maintained for 10 seconds after completion of the wash duration. A
sensor calibration within the cycle duration restarts the wash timer.
The wash timer is deactivated with the "0" cycle duration.
88
11 Appendix
USP table, then the USP contact of the JUMO AQUIS 500 CR will switch.
The limits are defined in steps. For example: at 8°C a value of 5°C is applied.
Note:
Temperature compensation must be switched off (temperature coefficient = 0)
during monitoring.
To do this, select Administrator level / Basic settings / Temperature
compensation / None.
Extract from USP <645>
Temperature Max. conductivity Temperature Max. conductivity
°C µS/cm (uncompensated) °C µS/cm (uncompensated)
0 0.6 55 2.1
5 0.8 60 2.2
10 0.9 65 2.4
15 1.0 70 2.5
20 1.1 75 2.7
25 1.3 80 2.7
30 1.4 85 2.7
35 1.5 90 2.7
40 1.7 95 2.9
45 1.8 100 3.1
50 1.9
If the conductivity exceeds the value for the corresponding temperature, the
configured contact will switch.
USP pre-alarm
The USP pre-alarm switches before the water quality reaches the set limit.
The parameter: SWITCH OUTPUT 1 / FUNCTION / PRE-ALARM (0 — 100) is
used to define a margin between pre-alarm and USP limit, as a percentage
value referred to the active limit.
89
11 Appendix
90
12 device description
12.1 Technical data
12.1.1 Inputs
Main input Indication range Accuracy Temperature error
µS/cm 0.000 — 9.999 ≤ 0.6% of range + 0.3 µS x cell 0.2%/10°C
00.00 — 99.99 constant (K)
000.0 — 999.9
0000 — 9999
mS/cm 0.000 — 9.999 ≤ 0.6% of range + 0.3 µS x cell 0.2%/10°C
00.00 — 99.99 constant (K)
000.0 — 999.9
0000 — 9999
kΩ x cm 0.000 — 9.999 ≤ 0.6% of range + 0.3 µS x cell 0.2%/10°C
00.00 — 99.99 constant (K)
000.0 — 999.9
0000 — 9999
MΩ x cm 0.000 — 9.999 ≤ 0.6% of range + 0.3 µS x cell 0.2%/10°C
00.00 — 99.99 constant (K)
000.0 — 999.9
0000 — 9999
91
12 device description
12.1.5 4-electrode systems
Cell constant Setting range of Resulting usable
[1/cm] relative cell constant range [1/cm]
0.5 0.1 — 0.75
20 — 150%
1.0 0.2 — 1.5
12.1.7 Controller
Controller type limit comparators, limit controller, pulse width controller, pulse frequency controller,
modulating controller, continuous controller
Controller action P / PI / PD / PID
A/D converter dynamic resolution up to 14-bit
Sampling time 500 msec
12.1.12 Display
Graphics LC display 120 x 32 pixels
Background lighting programmable:
- off
- on for 60 seconds during operation
92
12 device description
12.1.13 Housing
Material ABS
Cable entry cable glands, 3xM16 and 2xM12 max.
Special feature venting device to prevent condensation
Ambient temperature range -10 to 50°C
(the specified accuracy is adhered
to within this range)
Operating temperature range -15 to 65°C
(device is operational)
Storage temperature range -30 to 70°C
Climatic conditions rel. humidity ≤ 90% annual mean, no condensation
(following EN 60721 3-3 3K3)
Enclosure protection in surface-mountable housing: IP67
as per EN 60529 for panel mounting: IP65 front, IP20 rear
The specified enclosure protection ratings will only be achieved if each cable is fed into the device
through a cable gland (exception: special sealing supplied for two cables).
For panel mounting, the control panel must be of sufficient thickness.
Vibration strength as per EN 60068-2-6
Weight surface-mountable housing: approx. 900 g
for panel mounting: approx. 480 g
Dimensions see dimensioned drawings on page 9.
Cable glands
Internal mounting material
Operating instructions
93
12 device description
94
12 device description
12.2 Panel cut-out
100.5
ø1
0
ø4
.5
Note:
1. Fix template to panel.
2. Drill holes ( 4.5 mm and 10 mm dia.)
121.6
120.5
108.6
95
12 device description
96
13 Index
A MIN/MAX values 29
Accessories 9–10 Mounting location 11
Administrator 36
Analog output 39 N
Nameplate 7
B Normal display 25, 29
Binary input 38
Block diagram 6 O
Operation in levels 27
C Operator 36
Cable routing 19 Output 30
Calibration enable 42 Output level display 30
Conductivity input 38 Outputs 23
Conductor 15 Overview of MANUAL mode 32
Conductor cross-sections 15
Connection of conductivity cell (2-electrode P
system) 20 Panel mounting 13
Connection of conductivity cell (4-electrode Parameter 38
system) 21 Password 36
Controller channel 38 Pipe 12
Controller functions 44 Pipe installation 12
Controller special function 39 Principle of operation 26
Controls 24
R
D Relative cell constant 42
Date of manufacture 7
Delete 43 S
Display 40 Scope of delivery 8
Sensor connection 20–21
E Setup examples 46
Electrical connection 15 Setup interface 72
Electrical isolation 16, 72 Simulation mode 31
Enable 38 Simulation of switching outputs 33
Sunlight 11
F Supply 22
Fast start 45 Surface 11
Fixing brackets 11 Surface mounting 11
Switching output 39
H
Highly-purified water 66
T
HOLD mode 35 Temperature compensation 88, 91
Temperature input 38
I Terminal assignments 22
Inputs 22 Type designation 8
Installation position 11
W
M Wash timer 88
MANUAL mode 31 Weather protection roof 12
MANUAL mode for analog outputs 34
MANUAL mode for switching outputs 31
Measurement mode 25, 29
97
13 Index
98
JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.
Street address: JUMO House 6733 Myers Road
Moritz-Juchheim-Straße 1 Temple Bank, Riverway East Syracuse, NY 13057, USA
36039 Fulda, Germany Harlow - Essex CM20 2DY, UK Phone: 315-437-5866
Delivery address: Phone: +44 1279 63 55 33 1-800-554-5866
Mackenrodtstraße 14 Fax: +44 1279 63 52 62 Fax: 315-437-5860
36039 Fulda, Germany Email: [email protected] Email: [email protected]
Postal address: Internet: www.jumo.co.uk Internet: www.jumousa.com
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Email: [email protected]
Internet: www.jumo.net