X-Ray Powder Diffraction
X-Ray Powder Diffraction
Martin Ermrich
Martin Ermrich
Detlef Opper
XRD for the analyst
PANalytical GmbH
Nürnberger Str. 113
34123 Kassel
+49 (0) 561 5742 0
[email protected]
www.panalytical.de
PANalytical B.V.
Lelyweg 1, 7602 EA Almelo
P.O. Box 13, 7600 AA Almelo, The Netherlands
Tel: +31 (0)546 534 444
Fax: +31 (0)546 534 598
[email protected]
www.panalytical.com
ISBN: 978-90-809086-0-4
We encourage any feedback about the content of this booklet. Please send to the
address above, referring to XRD_for_the_analyst.
2
XRD for the analyst
Contents
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. What is XRD? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3. Basics of XRD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1 What are X-rays? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Interaction of X-rays with matter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Generation of X-rays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.2 The sealed X-ray tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.3 White radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3.4 Production of characteristic X-rays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4 A short introduction to crystallographic terminology . . . . . . . . . . . . . . . 17
3.5 The powder diffractogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.2 The position of the reflections: Bragg’s law . . . . . . . . . . . . . . . . . . . . . . . 22
3.5.3 The intensity of the reflections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.4 The shape of the reflections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.6 Overview of X-ray scattering techniques . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.1 Single crystal XRD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.2 Powder diffraction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.3 Special X-ray techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7 General terms used in X-ray powder diffraction . . . . . . . . . . . . . . . . . . . 27
3
XRD for the analyst
8. Crystallographic analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.1 Indexing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.1.1 Lattice parameter determination and refinement . . . . . . . . . . . . . . . . . 71
8.1.2 Structure refinement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.1.3 Structure solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
9. Microstructural analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.1 Residual stress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.1.1 Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.1.2 sin2 ψ method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.2 Texture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.2.1 Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.2.2 The Lotgering factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
9.2.3 Omega scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
9.3 Crystallite size determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4
XRD for the analyst
9.3.1 Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
9.3.2 Crystallite size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
9.3.3 Micro-strain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
9.3.4 The Williamson-Hall plot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
9.3.5 Discussion of the peak broadening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
12. Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5
XRD for the analyst
Dr. Martin Ermrich studied Physics at the Technical University in Dresden. In 1997 he
founded his own X-ray laboratory with a special focus on X-ray diffraction.
6
XRD for the analyst
1. Introduction
As a (future) user of a PANalytical X-ray diffractometer you can perform almost
all types of diffraction applications, depending on the configuration of your
system. To get the best possible results out of your system, this book gives you an
introduction to X-ray powder diffraction (XRPD or, mostly used, XRD) and provides
helpful information. It gives a simple explanation of how a diffractometer works
and how XRD analysis is done. This book is intended both for people new to the
field of XRD analysis and for more experienced users to find new applications
which might be helpful in daily work. This book avoids complex mathematical
equations and understanding its contents only requires a basic knowledge
of crystallography, mathematics and physics. It is not dedicated to specific
diffractometer configurations or application areas, but aims to give a global
overview of the various possibilities in XRD.
Chapter 2 briefly explains XRD and its benefits. General application fields are
presented. Chapter 3 gives an introduction to the physics and crystallography
of XRD, while Chapter 4 describes how this is used for diffraction experiments.
Chapter 5 explains the XRD data collection. This includes presentation of sample
preparation and measurement procedures to get the best possible results. In
Chapters 6-9 the procedures to perform qualitative and quantitative phase analysis
are presented. Chapter 10 deals with non-ambient XRD and describes its methods
and applications together with other techniques. Chapter 11 lists recommended
literature for further information on XRD analysis.
7
XRD for the analyst
2. What is XRD?
X-ray diffraction (XRD) is a versatile, non-destructive analytical method to analyze
material properties like phase composition, structure, texture and many more of
powder samples, solid samples or even liquid samples.
Many of these techniques can also be used for poly-crystalline layered materials
such as coatings.
X-ray powder diffraction is used in a wide variety of research and process control
environments. For example:
• Characterization of (new) materials at universities and research centers
• Process control in several industries like building materials, chemicals,
pharmaceuticals, for instance phase composition and content
• Determination of polymorphism, API (active pharmaceutical ingredient)
concentration determination, API stability studies in the pharmaceutical industry
• Phase identification of minerals in geological samples
• Optimization of fabrication parameters for wear-resistant ceramics and
biomaterials
• Determination of the crystallinity of a phase
• Determination of amorphous phase contents in mixtures
8
XRD for the analyst
10000
2500
0
20 30 40 50 60 70 80
Figure 1. Typical powder pattern showing the presence of a crystalline phase and an amorphous
phase
9
XRD for the analyst
3. Basics of XRD
3.1 What are X-rays?
X-rays are electromagnetic waves with associated wavelengths, or beams of
photons with associated energies. Both views are correct, and the view that you
use at a particular time usually depends on the specific phenomena that you are
interested in. Other electromagnetic waves include visible light, radio waves and
γ-rays. Figure 2 shows that X-rays have wavelengths and energies between those
of γ-rays and ultraviolet light. The wavelengths of X-rays are in the range from
0.01 nm to 10 nm, which corresponds to energies in the range from 0.125 to
125 keV.
c
E = hν = h
λ
where h is the Planck’s constant, ν is the frequency, and c the velocity of light.
The unit Å (Ångström), which is a unit length of 10-10 m, or 0.1 nm, is historically
used in X-ray diffraction because it simplifies the notation of wavelengths, atomic
distances and lattice parameters.
10
XRD for the analyst
I = I 0e − µ d
The attenuation coefficient (μ ) of the material depends on
• atomic number (Z )
• density (ρ)
• packing fraction ( p) (a typical value is between 0.6 to 0.8)
• wavelength (λ)
• Coherent or Rayleigh scatter is the most important effect for X-ray diffraction:
it is the elastic scattering of the incoming photon upon collision with inner-shell
electrons. The wavelength of the photon is not changed, i.e. the photon energy
remains constant.
• Incoherent scatter can be divided into Compton scatter and fluorescence. In
both cases, the wavelength of the photon increases by the scattering process,
i.e. the photon energy decreases.
–– Compton scatter: an electron is pushed out of its shell or is excited to a
higher energy state. The incoming X-ray photon looses energy. This type of
scatter can be ignored in laboratory X-ray powder diffraction because of the
low energy of the incoming photons.
–– Fluorescence: the incoming photon ejects an inner-shell electron from the
atom. The vacancy is then filled by an electron from one of the atom’s outer
shells. The action of this electron moving from one shell to another creates an
X-ray photon with the energy difference of the two shells. The energy of this
photon is dependent on the atomic number and is therefore characteristic
for the atom itself. This effect is used in XRF (X-ray fluorescence) analysis
for determination of elemental concentrations. In XRD it is generally
considered as an unwanted radiation because it raises the background of the
diffractogram and reduces the peak/background ratio.
11
XRD for the analyst
3.3.1 Introduction
There are several physical methods of obtaining X-rays of sufficient intensity for
X-ray diffraction. The most common methods are:
• Bombarding a target of a suitable material (anode) with a focused electron
beam. The maximum achievable X-ray intensity is limited by the maximum
power, which is restricted by the cooling system of the stationary anode.
These X-ray sources are called sealed X-ray tube. Modern designs use ceramic
insulators instead of glass bodies to improve stability and lifetime. These tubes
are widely used in modern laboratory diffraction systems, like the X’Pert PRO
and Empyrean.
• Deflection of high-energy electrons by electromagnetic fields, yielding emission
of X-rays. This principle is used in synchrotrons. These instruments yield a very
strong X-ray intensity but are large and costly to operate so that they are
available only in huge research centers, mainly used by scientists.
Water out
HT connection
contacts
Automatic
tube recognition
Water in
X-rays are emitted from the tube anode through thin beryllium windows. The
windows keep the tube sealed and allow X-rays to pass through. Optics are
’looking’ to the anode under a grazing angle, which results in a smaller apparent
source spot size. The windows are located at parallel or perpendicular positions
with respect to the focal line to form the point or line focus. In a typical long fine
12
XRD for the analyst
focus tube the line focus has a dimension of 0.04 mm × 12 mm, point focus of
1.2 mm × 0.4 mm. Both focal positions have their typical applications, which are:
• Line focus:
–– High-resolution phase analysis
–– Omega stress
• Point focus:
–– Micro-diffraction
–– Texture
–– Psi stress
The intensity of the X-ray tube decreases with time. This is mainly caused by the
evaporation of the tungsten filament (the cathode) leading to deposition of
tungsten on the anode and the Be windows. A deposition of tungsten on the
anode leads to additional lines in the diffractogram.
When the electrons hit the anode, two types of radiations are emitted: white
radiation (bremsstrahlung) and the characteristic X-ray radiation.
c
E = eU = hν = h
λ
e elementary charge
U used high voltage of the tube
c velocity of light
h Planck constant
n frequency
hc 12.4
It therefore follows: λ0 = ⇒ λ0 (Å) =
eU U (kV)
The Dauvillier law allows the estimation of the wavelength λmax of the maximum of
the radiation wavelength:
13
XRD for the analyst
I wr ~ ZU 2i
With Z atomic number
U tube voltage
i used emission current of the tube
White radiation is unwanted for most X-ray diffraction experiments, the majority
of diffraction experiments require a monochromatic radiation.
The emission of an electron produces a void in a shell (in the example a void in
the K-shell). This puts the atom in an excited state with a higher energy. The atom
wants to restore the original configuration by releasing the excess energy. This is
14
XRD for the analyst
done by transferring an electron from an outer shell - in this case from the L-shell
- to the void in the K-shell. An L-shell electron has a higher energy than a K-shell
electron. Thus when an L-shell electron is transferred to the K-shell, the energy
excess is emitted as an X-ray photon. The emission energy of the photon is seen as
a certain line in the spectrum. The energy of the emitted X-ray photon depends
on the difference in energy of the shell with the initial void and the energy of the
electron that fills the void (in the example, the difference between the energy
of the K- and the L-shell). Each atom has its specific energy levels, so the emitted
radiation is characteristic for that atom (in this case that of the anode material).
To expel an electron from an atom, the X-rays must have a higher energy than
the binding energy of the electron. If an electron is expelled, the incoming X-ray
photon is absorbed, leading to scattering. If on the other hand the energy is too
high, many photons will ‘pass through’ the atom and interaction with atoms will
only occur occasionally. Figure 5 shows that high energies are hardly absorbed.
If the energy is reduced the absorption increases and the scattering yield goes
up. The highest yield is reached when the energy of the photon is just above
the binding energy of the electron to be expelled. If the energy becomes lower
than the binding energy, a jump or edge can be seen: the energy is too low to
expel electrons from that shell, but is too high to expel electrons from the lower
energetic shells. The figures show the K-edge corresponding to the K-shell, and
three L-edges corresponding with the LI-, LII-and LIII-shells.
Based on quantum mechanics not all transitions are allowed. For instance a
transition from the LI- to the K-shell. Figure 6 gives an overview of the most
important emission lines with their transitions.
15
XRD for the analyst
K-lines L-lines
Figure 6. Major lines and their transitions
Because each atom has its own energy levels, the emitted radiation is specific to
that atom: the so-called characteristic X-ray radiation. This procedure is the basis of
X-ray fluorescence analysis (XRF). The wavelength of the photon depends on the
atomic number Z following the Moseley law:
1
~ Z2
λ
An X-ray tube delivers the characteristic wavelengths of its anode material (most
commonly Cu, Co, or Mo). The characteristic radiation is superimposed on the
white radiation.
6
Kα
5
25kV λmax
4
White radiation or bremsstrahlung
X-ray intensity
(relative units)
3 20
Kβ
Characteristic radiation
2 15
1
10
0 5
0 1.0 2.0 3.0
λ0 Wavelength (Å)
Figure 7. The total X-ray spectrum for various acceleration voltages (Mo X-ray tube)
16
XRD for the analyst
Figure 7 shows the X-ray spectrum for various acceleration voltages, in this case
for a Mo X-ray tube. One can see that if the voltage is too low, the amount of
characteristic radiation is very small. If the excitation voltage is too high, the
amount of white radiation increases over the amount of characteristic radiation.
This leads to a high background in the diffractogram. The photon yields are
described in Chapter 3.3.3.
Each anode material has its own optimal voltage for the production of
characteristic X-rays. This is approximately four times the energy of the Ka line of
the anode material (Table 1).
Anode material Mo Cu Co Cr
Kα energy (keV) 17.4 8.0 6.9 5.4
Uopt (kV) 60-50 40 30 25-20
Table 2. Wavelengths used for XRD and attenuation filters for Kβ radiation (source: ICDD,
International Centre of Diffraction Data)
17
XRD for the analyst
Unit cell
The unit cell is a spatial arrangement of atoms which is tiled in three-dimensional
space to describe the crystal. The unit cell is given by its lattice parameters. These
are the lengths of the cell edges and the angles between them. The positions of
the atoms inside the unit cell are described by the set of atomic positions (xi,yi,zi)
measured from a selected lattice point.
For each crystal structure there is a conventional unit cell, which is the smallest
unit that has the full symmetry of the crystal. However, the conventional unit cell is
not always the smallest possible choice. A primitive unit cell of a particular crystal
structure is the smallest possible unit cell one can construct such that, when tiled,
it completely fills space. This primitive unit cell does not, however, display all the
symmetries inherent in the crystal.
Crystalline state
The atoms are periodically and regularly arranged in three dimensions. Ideal
crystals are described completely by only one unit cell and its infinite three-
dimensional-periodic repetition (translation symmetry), delivering a long-range
order. Crystalline material leads to pronounced peaks in powder diffractometry.
Amorphous state
In contrast to the crystalline state, in the amorphous state the structure has only a
short-range order of the atoms (or unit cells). This leads to very broad humps in the
diffraction pattern instead of clear diffraction peaks.
Reciprocal lattice
The reciprocal lattice is a mathematical construction that helps to deliver a
description of X-ray diffraction.
Each point (hkl) in the reciprocal lattice corresponds to a set of lattice planes (hkl)
in the real space lattice. The direction of the reciprocal lattice vector corresponds
to the normal to the real space planes, and the magnitude of the reciprocal lattice
vector is equal to the reciprocal of the interplanar spacing of the real space planes.
18
XRD for the analyst
Bravais lattices
The Bravais lattices (see Table 3) are the 14 possible lattices given by translation
symmetry. These simple lattices allow to describe all crystal structures, even
complicated ones (Figure 9).
C I F
19
XRD for the analyst
Crystal classes
A further classification of the contents of a unit cell can be made by classifying
possible crystal structures into those that obey different symmetry operations:
rotation, glide reflection, inversion and combinations of them. There are 32 crystal
classes (also known as point symmetry groups).
Space groups
If all possible Bravais lattices and crystal classes are taken into consideration, one
ends up with 230 unique space groups. Every crystal has a structure which can be
described by one of the 230 space groups.
Miller indices
The Miller indices are the designation of interplanar spacings, which cut the
axes lengths a, b, c or the axis intercepts in the reciprocal space a/h, b/k and c/l,
respectively. Numbers are reciprocals of the intercepts (see Figure 10). For the (102)
plane for example, you have to go one step in ‘a’ direction and half a step in ‘c’.
20
XRD for the analyst
Ewald sphere
The Ewald sphere contains all of the components that are needed to visualize the
diffraction process geometrically. It is a geometric construction which demonstrates
the relationship between
(a) the wavelength of the incident beams of light,
(b) the angle of diffraction for a given reflection,
(c) the unit cell and reciprocal unit cell of the crystal, and
(d) the distance between the crystal and the receiving slit/detector.
A reflection hkl fulfills the Bragg equation, if the lattice point coincides with the
Ewald sphere. The radius of the Ewald sphere is 1/λ (Figure 11).
kh
d k0
θ
3.5.1 Introduction
A powder diffractogram displays the scattered intensity versus the Bragg angle
(2θ ). It contains a number of peaks (reflections). The peaks are characterized by
their position, intensity and profile. The peaks and the background are the source
of all information of the X-ray powder diffraction technique. Table 4 shows an
overview.
21
XRD for the analyst
The origins of these three fundamental parameters are discussed in more detail in
the following paragraphs.
Intensity (counts)
15000
10000
5000
0
20 25 30 35 40 45 50 55 60 65 70 75
2Theta (°)
22
XRD for the analyst
In 1913 W.H. Bragg and W.L. Bragg described diffraction and interference of X-rays
in a crystal as reflections at the atomic planes of the crystal lattice. The positions of
the reflections are calculated using the optical path difference 2s, with s = d sin θ,
between two reflected rays at neighboring interplanar spacings. As in visible light
optics, maxima are produced for integer multiples of λ.
23
XRD for the analyst
2. The order n of interferences is described by higher (hkl ), i.e. the second order
of the (111) reflection gets the indices (222). Therefore n = 1 is valid.
5. Varying of l causes:
–– the number of measurable reflections to vary, and
–– the angular position of the reflections to vary; also,
–– the separation between two neighboring reflections is changed.
1 h2 + k 2 + l 2
cubic system =
d2 a2
1 h2 + k 2 l 2
tetragonal system = + 2
d2 a2 c
7. Sometimes it is confusing that different terms are used having the same
meaning:
–– Diffraction in the sense of the interaction of waves at a lattice, yielding
interferences.
–– Reflection in the sense of a reflection at the interplanar spacings.
24
XRD for the analyst
All factors that determine the absolute intensity are discussed in Chapter 7.1.
25
XRD for the analyst
26
XRD for the analyst
High-resolution XRD
High-resolution X-ray diffraction (HR XRD) is widely used to investigate the match
of heteroepitaxial layers to a single crystal substrate. These layers have very small
differences in their d-spacing compared with the substrate. Thus the reflections
of substrate and layer have a very close distance in 2θ. This requires a highly
monochromatic X-ray beam with very low divergence in combination with a high-
resolution goniometer.
Resolution
In powder diffraction, resolution is defined as the ability to separate two (or more)
neighboring reflections. In single crystal diffraction it is defined as the lowest
possible d-value, which can be measured with an instrument. In small-angle X-ray
scattering it is the highest possible d-value.
27
XRD for the analyst
Crystallinity
Crystallinity describes the percentage of the crystalline part in a mixture of
crystalline and amorphous materials (known as the crystallinity index or %
crystallinity). Note that crystallinity is often synonymously used to describe the
quality of the crystal itself in terms of lattice defects, micro-strain etc.
28
XRD for the analyst
29
XRD for the analyst
Depending on the diffractometer geometry and the type of sample, the angle
between the incident beam and the sample is either fixed or variable, normally
coupled to the diffracted beam angle.
4.2 Geometries
There are two main methods to investigate a powder sample with X-rays:
• Reflection geometry: the X-rays are scattered by a flat sample surface,
• Transmission geometry: the X-rays pass through an X-ray transparent sample,
which can be prepared as:
–– a sample between foils,
–– a thin solid sample, transparent for X-rays,
–– a sample in a glass capillary.
The preferable method depends on the sample type as discussed in the following
sections.
30
XRD for the analyst
Divergence slit
Beta-filter
Soller slits Soller slits
Anti-scatter slit
Sample
Figure 17. Classical powder diffractometer setup in Bragg-Brentano geometry with components for
reflection measurements
In the Bragg-Brentano geometry, the diffracted angle is always twice the incident
angle. In a traditional arrangement the diffractometer works in the θ/2θ-mode, i.e.
the tube is fixed and a 1:2 coupled movement of sample and detector is performed.
Most current vertical diffractometers use the θ/θ-mode, where the sample always
stays horizontal and tube and detector perform a 1:1 movement. This setup has
some advantages:
• no powder spillage at high angles, even when spinning is used
• possibility to measure liquid or molten samples (non-ambient XRD)
• ability to measure very large samples on corresponding sample platforms or
positioned not fixed to the goniometer
31
XRD for the analyst
Figure 19 shows a typical transmission beam path. The focus on the detector circle
results in a very high resolution.
Figure 19. Sketch of typical transmission setup for capillary measurements, using focusing X-ray
mirror and line detector
32
XRD for the analyst
Further optical options for specific applications with other than focusing geometry
can be:
–– multi-layer X-ray mirrors
–– monocapillaries
–– polycapillary lenses
–– hybrid monochromators
–– parallel plate collimators
beam, defined by a divergence slit. The divergence is substantial for the focusing
principle, so that the reflected beam from the lattice planes is focused on the
receiving slit in front of the 0D or 1D detector. The distance between the focus and
the sample is the same as between sample and receiving slit/1D detector.
Diffracted beam
The anti-scatter slit assures that just the reflected beam and no air scatter reach the
detector. The slit size corresponds to the divergence slit size. The anti-scatter slit
is followed by the receiving slit. The receiving slit is placed at the focal point. The
receiving slit size defines the resolution and the intensity. Smaller slits give higher
resolution and lower intensity and vice versa. The size of the anti scatter slit has
to match the size of the divergence slit to observe just the illuminated length. The
receiving slit has to be set according to the resolution or intensity needed for the
application.
Programmable slits
Programmable divergence and anti-scatter slits can be used both in fixed and in
automatic/variable mode. The difference between the modes is:
–– In fixed mode the length of the irradiated (and observed) area on the sample
changes during a scan. The irradiated volume remains constant, the so-called
measurement with constant sample volume.
–– In automatic (or variable) mode the length of the irradiated (and observed) area
on the sample does not change during a scan. The irradiated length remains
constant, the so-called measurement with constant sample area.
The automatic divergence and anti-scatter slits of the X´Pert PRO and Empyrean
diffractometers use two independently controlled knifes and are controlled to
assure symmetrical sample illumination even at very low angles.
The automatic anti-scatter slit on X’Pert PRO and Empyrean diffractometers
provides additional functionality: together with a motorized receiving slit
or a position-sensitive detector it can define a beam tunnel for reflectivity
measurements, small-angle X-ray scattering or simple parallel-beam path
applications.
34
XRD for the analyst
Fixed mode:
for phase analysis and Rietveld (structure) refinement, because the diffractograms
match the data of the ICDD database (see Chapter 6.2). The constant sample
volume is a prerequisite for Rietveld analysis.
Automatic mode:
–– for measurements at low angles to reduce the influence of primary beam scatter
–– to improve the counting statistics at higher diffraction angles
In general, fixed slit intensities can be converted to variable slit intensities and vice
versa with algorithms available in e.g. the HighScore phase analysis software.
Figure 21 demonstrates the change of the irradiated length with the 2θ angle for a
range of slit sizes (from 1/32° to 2°) and a measuring circle with R = 240 mm.
Note that at low angles the illuminated length increases drastically.
35
XRD for the analyst
0.02 and 0.01 rad) are available. A smaller axial divergence improves the symmetry
of the peaks, especially at low angles, as well as the resolution (smaller ‘full width
at half maximum’, FWHM), but leads to lower intensities. The optimum choice
depends on the required resolution and intensity.
Kβ filters are thin metal foils, which can be mounted in the primary or secondary
beam path. They are usually used in the diffracted beam path to reduce the
influence of sample fluorescence. The filter suppresses the beta emission line
by utilizing the material-specific absorption edge. As shown in Figure 22, a Ni
filter, used with Cu radiation absorbs >99% of the Kβ part but also lowers the Kα
radiation by ~50%. Tungsten (W) L-lines, mainly present when using older tubes,
are also suppressed.
36
XRD for the analyst
The materials used as Kβ filters are normally one atomic number below that of the
anode material (Table 5).
When using X-ray mirrors or primary monochromators, the use of a β filter is not
necessary, since Kβ is completely suppressed by these optics.
Primary monochromators consist of one or more single crystals. They are positioned
in the beam path in such a way that the incident beam on the lattice plane fulfills
the Bragg equation (see Chapter 3.5.2). Since the Bragg equation is only valid for
specific wavelengths at a given angle, the diffracted beam from the crystal is highly
monochromatic. The use of a monochromator always leads to a lower primary
beam intensity, because only a part of the wavelengths is used and intensity is
absorbed in the monochromator crystal. The combination of several adjacent
crystals improves the spectral purity of the incident beam but leads to additional
attenuation of the intensity.
37
XRD for the analyst
Figure 23 shows the beam path when using an incident beam monochromator in
reflection mode. The curved incident beam monochromator is aligned in a way
that just the Kα1 wavelength passes the divergence slit. Fluorescent radiation of the
sample is not eliminated, but the monochromator reduces this effect, because now
only the Kα1 component can excite the fluorescence. Kβ and white radiation are
not present any more. If necessary, an additional secondary monochromator can be
used for further suppression of fluorescent radiation.
4.4 Detectors
Point or 0D detector All incoming photons are counted, their position on the
active element is not relevant.
Line or 1D detector Also called position-sensitive detector, i.e. photons are
counted relative to their position on the detector window
in 2θ direction.
Area or 2D detector Similar to the line detectors, but intensities are also
detected perpendicular to 2θ axis.
38
XRD for the analyst
Application-specific requirements
The table below lists all the properties that are of importance for a detector used
in X-ray diffraction:
Property Description
Angular resolution The ability of the detection system to
distinguish between different d-spacings (the
ability to separate neighboring reflections).
Mainly defined by slits in front of a point
detector or the width of the channels of a
position-sensitive detector.
Energy resolution The ability of the detection system to detect
only Kα or Kα1 radiation and suppress
unwanted radiation (for example: Kβ or
sample fluorescence). The point and line
detectors incorporate electronics that allow
using a (variable) pulse height discrimination,
which acts like a band pass filter. The energy
resolution of such a filtering is usually
moderate, so that additional means of energy
filtering such as Kβ filters or diffracted beam
monochromators have to be used.
Count rate linearity The intensity at which the recorded intensity
starts to deviate more than 5% from the
incident intensity, above this value, the
deviation will become larger until the detector
finally saturates. In extreme cases this leads to
untypical peak shapes with a minimum at the
peak tip position.
Detector noise The noise level of a detector, in most cases
<< 1 cps.
Dynamic range The counting range of a detector, defined as
the difference between noise level and the
highest intensity possible without saturation.
Measurement time The time that the detector needs to measure a
certain angular range.
Possibility for in-situ Determination of changes in the sample as a
measurements function of an external parameter or time, for
example: temperature-controlled experiments,
humidity, crystallization and so on.
Active length or active area The angular range that is covered by a line
detector or the size of the detecting element
of a point detector (important for grazing
incidence or parallel beam experiments).
39
XRD for the analyst
Point detectors
Two types of point detectors are widely used for almost all diffraction applications:
gas-filled proportional counters and scintillation counters. The selection of the
appropriate detector heavily depends on the choice of tube anode material.
40
XRD for the analyst
gas. The ionization electrons are accelerated to the wire and are ionizing further
atoms (charge avalanche). At the counting wire an impulse with an amplitude
proportional to the energy of the absorbed X-ray photon is generated and
recorded by a pulse-height discriminator.
This detector type is the optimum choice for standard wavelengths as Cu Kα, Co
Kα, and Cr Kα radiation. The detector has a very low background and a good
energy resolution.
Scintillation counter
Scintillation counters consist of a crystal and a photomultiplier. The crystal is
capable to convert an X-ray photon into a UV or visible photon. The incoming X-ray
photon is absorbed in the crystal (for example: NaI, doped with Tl). The resulting
light impulse is recorded by the photomultiplier as an amplitude proportional
to the energy of the absorbed X-ray photon. Scintillation detectors have a lower
energy resolution than gas proportional counters for Cu Kα radiation. However,
they are preferred for short wavelengths (e.g. Mo Kα or Ag Kα).
Line detectors
Line detectors cover a certain angular range in 2θ. They resemble a large number
of adjacent single detectors. Line detectors are thus much faster than point
detectors. Different types are on the market. The first commercial line detectors
used a wire and a counting gas. Since 2001 semiconductor detectors define the new
standard. The PANalytical X'Celerator with its unique semiconductor RMTS (Real
Time Multiple Strip) technology was the first detector of that kind on the market.
The X'Celerator overcame the problems of the gas-filled detectors, which have a
small dynamic range and need regular maintenance because of aging of counting
gas and detector wires. The current state-of-the art PANalytical PIXcel3D detector
consists of smaller strips than the X’Celerator. The strips are subdivided in segments
so that the PIXcel resembles a solid-state area detector. It has a better resolution
and the highest linearity of all detectors available for laboratory XRD equipment.
The detector can be used in point detection mode, in line detection mode and
as an area detector. This makes it also the ideal detector for measurements of
epitaxial layers, reflectometry and small-angle X-ray scattering, which require a
very large dynamic range. Secondary monochromators are exclusively available for
these detectors, to improve energy resolution and suppress sample fluorescence.
Most line detectors have an active length of a few degrees so that the deviations
from the ideal focusing geometry are not too large. The use as a scanning detector
further reduces this defocusing effect. Line detectors are primarily designed for
focusing geometry, but can also be used with parallel beam geometry in reflection
41
XRD for the analyst
Area detectors
Area detectors can record larger parts of the Debye rings or even complete rings
at low angles. They are suited for special applications like micro-diffraction and
texture analysis. They are also widely used for single crystal diffraction. Different
designs are on the market.
42
XRD for the analyst
43
XRD for the analyst
If the powder is too coarse this results in single points on the Debye-Scherrer ring.
Therefore point or line detectors do not recognize all reflections which leads to
incorrect intensity ratios.
If crystallites are much smaller than 1 μm, reflections are broadened. This effect can
be used for crystallite size determination, see Chapter 9.3.
Table 6 shows some of the possibilities and risks involved in the milling process.
44
XRD for the analyst
45
XRD for the analyst
Especially for small-angle diffraction a knife edge (sometimes called a beam knife)
close to the surface of the sample is useful to reduce the background produced by
air scattering.
Powders
Powdered samples are placed between two thin foils (e.g. Kapton or Mylar).
Gels / liquids
A special sample holder is needed to fix the material in a special small chamber.
5.1.5 Capillaries
Powder samples sensitive to moisture or texture effects are ideally analyzed in
capillaries. The powder is loosely filled into the capillary without application of
pressure.
The X-ray transparency in the capillary may be improved by diluting the sample
with amorphous substances, for example cork flour, glass and/or lycopodium.
46
XRD for the analyst
Powders
The material is filled in 0.1 to 2 mm thick capillaries. The capillary diameter is
selected depending on the grain size and the X-ray transparency.
Gels / liquids
These materials can be prepared into a thicker capillary. Alternatively, the material
can be spread onto the surface of the capillary.
5.2 Measurements
Reflection mode
In general a coupled movement of the scan axes (θ /2θ or θ /θ ) is used (Bragg-
Brentano geometry). Other movement modes are possible, depending on the
application (e.g. fixed ω for grazing incidence measurements of thin layers
or ω=0 for SAXS experiments). Variable or fixed slit settings can be used (see
Chapter 4.3.2). Sample spinning improves the particle (crystallite) statistics.
Foil transmission
The maximum sample thickness depends on the sample material. For successful
transmission experiments the sample should have a low mass. When investigating
thick samples (e.g. a 3 mm pharmaceutical tablet) the sample is normally rotated
(spun) to improve the particle statistics and the center of the sample should be at
the center of the rotation in ω. In general sample spinning improves the particle
(crystallite) statistics.
Capillary
The capillary is fixed on a goniometer head and adjusted using a microscope or
CCD camera (Figure 27).
The incident beam is collimated so that it matches the diameter of the capillary
as good as possible. This will reduce air scattering to a minimum and will lead to
a low background at low angles. The ω angle is fixed at zero and the detector
moves on the 2θ circle. The capillary rotation decreases the influence of preferred
orientations.
5.2.2 Resolution
The resolution of a diffractometer is defined by its ability to differentiate neighbor-
ing reflections (see Chapter 3.7). In order to obtain an optimal resolution, d ifferent
geometries with certain properties can be used. The maximum achievable is the
47
XRD for the analyst
sum of the properties of the sample and the diffractometer, i.e. X-ray tube, goni-
ometer radius, incident and diffracted beam optics and detector.
General remarks
Assuming that the sample is fine-grained (see Chapter 4.3.1), unstressed and
without lattice dislocations, the resolution will be improved by
–– increasing the radius of the measuring circle
–– using pure Kα1 radiation instead of the Kα1,2 doublet
–– using longer wavelengths
–– using smaller Soller slits
–– using narrower slit widths:
The divergence slit and Soller slit have a substantial effect on the attainable
half widths of reflections. This is particularly true for the Soller slits when you
consider how they determine the cutout from the Debye-Scherrer cone in the
plane of measurement.
–– making the sample as thin as possible. For
reflection fine powder on zero background holder
foil transmission low amount of powder between the foils
capillary smallest as possible diameter.
Small step sizes in 2θ are necessary when measuring peaks with a small FWHM (at
least five measurement points over the FWHM are required).
In general:
1. A compromise is often necessary, because the possibilities shown here usually
result in a decrease of the measured intensity.
For best results the incident beam should illuminate as many sample particles as
possible, so careful selection and setup of primary and secondary optics is necessary.
It may be helpful to perform the measurement without spinning the sample, when
particles do not stick to the sample holder.
Reflection mode
The substances should be fixed on a on a 'zero-background holder' (e.g. silicon
(510) or (511) oriented or quartz (6° offcut to c axis)). Also an amorphous
substrate such as glass can be used, but it leads to higher undesired background.
48
XRD for the analyst
Foil transmission
A small amount of the sample material should be carefully placed between the two
foils. Using a smaller beam mask helps to reduce air scattering.
Capillary
The amount of material in the capillary should be sufficiently illuminated by the
primary beam. If the amount of sample material is insufficient to entirely fill the
capillary one can partially fill it with a supporting substance and subsequently with
the sample so that the sample material is placed in the X-ray beam. Alternatively
the material itself can be diluted with a reflection-free powder (cork powder, glass
and/or lycopodium) to increase the total sample volume.
Reflection mode
The sample is prepared in a special sample holder that can be covered with a foil.
Foil transmission
Prepare the sample between foils, or, if it is thick enough, in a closed sample holder
(only for low absorbing materials).
Capillary
The use of a sealed capillary is usually the best choice when investigating air-
sensitive or hygroscopic materials.
Rotating (spinning) the sample only improves the particle statistics. The influence
of preferred orientations may be reduced by oscillating the sample in an additional
direction, for example: ± Δω (wobbled scan).
49
XRD for the analyst
Figure 28. Texture effect through different sample preparation, resulting in missing peaks
Reflection mode
Texture problems are mostly visible when using reflection geometry.
Preferred orientation caused by the preparation of powder measurements can be
minimized by
–– avoiding smoothing of the surface in one direction
–– back-loading of the sample into the sample holder
–– filling the powder against a lateral glass plate
–– roughing the surface
–– spray-drying the sample on a plate
Note: for clay mineral analysis, texture is sometimes desired to study the swelling of
minerals (determination of moisture absorption) or to simplify the phase analysis.
A special preparation method is used to maximize the preferred orientation of the
platy clay minerals. This method pronounces the (00l) reflection and allows a better
peak position determination (see also Figure 28).
Transmission mode
To enhance the (hk0) reflections, an additional movement of the sample in ± Δω
can help to decrease a texture regarding the (00l) reflections. The influence of
preferred orientation is significantly reduced.
Capillary
The investigation of loosely filled powders in capillaries is clearly advantageous to
overcome the texture problem. Both the rotation of the capillary and the possible
movement of the particles inside the capillary reduce texture effects.
50
XRD for the analyst
In thin film experiments the intensity of the layer peaks is not comparable to the
intensity we get from powder samples, because the number of reflecting lattice
planes is much less. Besides that the sample is not infinitely thick. This requires a
special beam path to optimize the intensities. Two options are widely used:
Table 7 shows how the penetration depth changes by varying the incident angle ω
and the wavelength (λ). A brass sample (layer) is used as an example. A penetration
depth d86,5 (delivering 86.5% of the information) is calculated. It can be clearly seen
that the penetration depth can vary between 9.67 µm and 0.25 µm by changing
from Mo Kα, radiation with a ‘normal’ incident angle of 25° to Cr Kα with an ω of
1.0°.
Table 7. Influence of the used wavelength and the variation of the incident angle ω on penetration
depth
Sample: Brass (30 at.% Zn), for effective penetration depth d86,5 (in µm)
Radiation µ/cm-1 w: 25° 10° 5° 3° 2° 1°
Cr Ka 1351 3.11 1.58 1.01 0.66 0.46 0.25
Co Ka 705 6.00 3.05 1.93 1.27 0.90 0.47
Mo Ka 434 9.67 4.93 3.13 2.07 1.43 0.78
51
XRD for the analyst
The most regularly used standards come from the US National Institute for
Standards and Technology (NIST). Their Standard Reference Materials (SRMs) are
listed in Table 8.
Table 8. Standard materials for XRD as supplied by NIST (main application field marked (x)).
SRM no. Type Application
Quantitative
parameters
Intensities
Reflection
position
analysis
Lattice
Profile
phase
640d Si X X X X
660b LaB6 X X X X
675 Mica low angles
656 Si3N4 (mixture a / b ) X
674b Mixture Al2O3 , CeO2 , X
Cr2O3 , TiO2 , ZnO
676a Al2O3 powder X spiking method
1878a Respirable α quartz X spiking method
1879a Respirable cristobalite spiking method
2910a Ca-hydroxyapatite X X X
1976b Al2O3 disk X X X
52
XRD for the analyst
Currently, there is no NIST certified standard for powder diffraction at low angles.
Nevertheless silver behenate (CH3(CH2)20COOAg) (reference to Powder Diffr. 10
(1995) 91) is widely used for this purpose.
53
XRD for the analyst
2. Start of a peak search algorithm to get the positions and both absolute
and relative maximum intensities of the reflections. The strongest line is
normalized to 100% (see Table 9).
54
XRD for the analyst
Such a peak table can be used for the search/match procedure. Better intensities
and peak positions are obtained after a profile fitting step has been applied, but
this is not crucial for a correct phase analysis.
Different software routines like for example available in the HighScore software
package allow automatic processing of data, using user-defined parameter
sets, even including reporting to either a printer or a laboratory information
management system (LIMS), as well as trend analysis.
55
XRD for the analyst
Since 1998 the database also contains calculated stick patterns, taken from
databases containing crystal structure information, for example: the ICSD
(Inorganic Crystal Structure Data Base) and the LPF (Linus Pauling File). In
some cases, not only the stick patterns but also the structures themselves are
incorporated in the PDF database. This extends the possibilities to perform certain
types of quantitative analyses directly, such as the Rietveld analysis.
Recently also alternative databases have been developed, ranging from freely
downloadable ones calculated from crystal structures, to specialty databases for
niche fields, such as narcotics.
Cluster analysis is basically a three step process, but can also include an optional
fourth step:
1. Comparison of all scans with each other. The result is a correlation matrix
representing the similarity of any given pair of scans.
56
XRD for the analyst
57
XRD for the analyst
7.1 Basics
The fundamental condition of the quantitative phase analysis is that the intensity
of the X-rays diffracted by a certain phase is proportional to its amount in the
phase mixture (H.P. KLUG and L.E. ALEXANDER).
The following equations refer to the investigation of crystalline phases. The
samples are powder mixtures in the sense of the statistical demands. The conditions
for the application of the kinematic theory must be fulfilled.
provided that Σ vk = 1.
58
XRD for the analyst
The absorption coefficient Ak can be set ½µk , whereby the attenuation coefficient
µk has to be possibly modified for a phase k in the mixture with other phases (then
named µk’).
General remarks
Quantitative phase analysis is usually implemented automatically using calibrations
or the Rietveld method. Some fundamental requirements should be met in order to
get the most reliable results.
–– Use the integral intensities of the reflections.
–– Use as many reflections as possible, but with different indices. Reflections with
the same indices may falsify the result, if they present a preferred orientation
(e.g. 220 and 440 reflection).
–– Take into account a possible influence of the absorption contrast, also called
micro-absorption, especially if a spiking material has to be selected.
–– Take care of the influences of texture through careful milling and/or sample
preparation and sample movements (at least spinning) during the measurement
in an optimal measuring geometry.
–– Check the reproducibility of the quantitative result by repeating the experiment
a couple of times, preferably with a complete new preparation and handling, or
even the use of automatic sample preparation equipment in order to eliminate
operator dependencies.
–– Double-check critical results by validation with an alternative technique, for
example: X-ray fluorescence, cathodoluminescence or another microscopic
method.
Preparation
The sample is milled until the ideal particle sizes of 1 to 5 µm are achieved. Because
the preparation effect is especially important for reflection mode measurements,
the preparation problems are described in Chapter 5.1.
59
XRD for the analyst
Measurement
Figure 29 shows the measurements of a mica sample in three measurement modes.
With regard to the reflection intensities it becomes clear that
–– the reflection geometry enhances the (00l)-reflections
–– the foil transmission reduces the (00l)-reflections in favor of the (hk0)-reflections
–– the capillary measurement delivers the optimum result.
The rotation of the sample (spinning) may be one way to reduce the preferred
orientations, but in reflection and foil transmission mode it mainly increases the
particle statistics. A more effective way is an additional small oscillation of the
sample holder, i.e. adding up different scans performed with different ω offsets
(wobbled scan).
60
XRD for the analyst
Micro-absorption
If a sample contains particles of different sizes, there is a significant effect on the
resulting intensities, because the size of the particle corresponds to the absorption
factor (the bigger the particles the higher the absorption). This effect strongly
depends on the distribution of the different particles in the sample.
Critical thickness
Figure 30. Effect of different particle sizes on resulting intensities (micro-absorption)
Figure 30 demonstrates the influence of both the differences in particle size and/or
the absorption in a mixture of two phases:
Part A shows 1. The effect of the crystallite size. The particles have different
crystallite sizes.
2. Micro-absorption. The particles differ in their absorption
coefficients µ. In this figure the white circles represent
particles with a higher absorption. Absorption contrasts (the
ratio between the two µ’s) larger than about 5 means that
the intensity ratios will be significantly wrong and therefore
the quantification will be incorrect
The effect on the intensity is the same for both examples. The beam either passes
the black particles and stops in a white particle, or it stops in a white particle
without having ‘seen’ any black particles. So the white particles form the ‘preferred
phase’, delivering a higher granularity quantity.
Extinction
Extinction is another effect that decreases intensity. The kinematic theory is not
applicable as the crystallites have a too high degree of perfection. A reduction
of intensity is due to an interference effect from secondary reflections of the
diffracted beam from the undersides of the atomic planes, destructively interfering
with the incident beam (phase shift 180° is given).
61
XRD for the analyst
σ= N
N 1
σ rel = =
N N
A usual criterion for the limit of detection (LOD) of a particular reflection, is that
Nreflection > Nbackground + 3σbackground
Note: Imagine we have a background of 100 counts and a small hump of 120
counts. Clearly this cannot be classified as a reflection because 3σbackground = 30 is
obviously higher than 20. The only solution is to increase the measurement time to
improve the peak-background ratio.
Limit of quantification
On the other hand, the limit of quantification (LOQ) depends on the sum of
the possible influences on the intensity: particle size, preferred orientation, line
overlap, crystal symmetry, matrix effects, and amorphous amount. So in many cases
the LOQ of a phase must be carefully determined by a calibration curve.
I net
%C = 100 ⋅ ∑
∑ tot − ∑ Iscat
I
62
XRD for the analyst
If the pure amorphous phase is available, the ratio C between the crystalline and
the amorphous part of a material can be determined by using diffraction patterns
at each 2θ position of the pure amorphous (a) and the pure crystalline (c) sample in
combination with the mixture:
I ( 2θ ) = m ( c I c ( 2θ ) + (1 − C ) I a ( 2θ ) )
I(2θ ) intensity at positions 2θ of the actual sample and of both the pure
amorphous (a) and the pure crystalline (c) sample
m sample mass
All measurements have to be carried out using the same instrument settings and
the same scan range parameters.
63
XRD for the analyst
Full pattern
Multi-phase Structures Is there an
Type of fitting method
or hkl files amorphous
application? quantification (e.g. Rietveld)
available? component?
based on
theoretical
Single phase FULLPAT/Le data
quantification Bail method; Quantitative
based on method
experimental
data
Quantitative Are the
method Can the
External constituents
amorphous
standard available as pure
component be
applicable? phases (incl.
modelled?
amorphous)?
RIR method
Semiquantitative
method
Add a material with a
similar absorption, and
Calibration Method: Internal
minimum peak overlap
curve standard or standard
of a known quantity
addition method
There are some aspects which should be considered when selecting a standard
material:
It should:
–– be available over a long period of time to enable investigations to be compared
with each other.
–– be possibly an internationally recognized standard material.
–– be chemically stable (no oxidation or hydration).
–– be mechanically stable.
–– have a sufficiently small crystallite size.
–– produce reflections with high intensities, because the quantitative results also
depend on the counting statistics.
–– not have fluorescent radiation.
–– show as few peak overlaps as possible.
–– be a substance with an absorption coefficient similar to that of the mixture to
prevent micro-absorption effects.
64
XRD for the analyst
Note that the stability of the incident beam intensity has to be monitored. This
is done by repeated measurements of the maximum intensity of a reflection of
a standard sample. The intensity of the standard is used to calculate a correction
factor for the intensity of the sample measurements at a given time.
Addition methods relate the intensity of the phase of interest in the original
sample with the intensity of the sample after the addition of a known amount of
the same phase.
65
XRD for the analyst
7.4 Examples
For the intensity ratio of the sum of (all) integral intensities I(hkl)A and I(hkl)R the
following applies:
I ( hkl )R R( hkl )R ρ A mR
=
I ( hkl)A R( hkl )A ρ B mA
Per gram of the mixture the quantity zR of rutile is added. Then applies
I ( hkl)R R( hkl)R ρ A mR + zR
= = S ( mR + zR )
I ( hkl)A R( hkl )A ρ B mA
S is the slope in the representation I(hkl)R / I(hkl)A as f (zR). The intersection of the
straight line with the x axis delivers the amount mR (note: y = Sx + SzR = 0).
In the following example (Figure 33) for each 1 g of the original mixture pure
rutile was added in amounts of 50 mg, 125 mg, 200 mg, 280 mg and 330 mg.
(Note: sample is different from the one shown in Figure 32).
66
XRD for the analyst
Figure 33. Spiking method for an example mixture, mrutil = 30.2 ± 5.2 wt.% (confidence interval 95%)
Using the Rietveld method for the same sample, which is well established for this
problem, the rutile content is determined to be 28.5 ± 0.5 wt.%.
Semi-quantitative methods are used if calibration standards and pure phases are
not available - but the reference intensity ratios (RIR) or I/Ic- values are published.
The I/Ic - values are given by the ICDD (for Cu radiation only) and correspond to
a 50:50 mixture of the phase and corundum. Analysis is automatically performed
when RIR values of the accepted candidates are available (see Figure 34, HighScore
software).
22500
10000
red. rutile
blue: anatase
2500
20 30 40 50 60 70
67
XRD for the analyst
Full pattern methods (especially the Rietveld method) compare the complete
measured diffractogram with calculated diffractograms of all phases based on their
crystal structure data.
The method is based on the minimization of the sum of the weighted squares
of the deviations between the observed and the theoretical intensities of the
diffractogram. The theoretical diffractogram is calculated from the crystallographic
and structure data of the phases.
The Rietveld method was first introduced for structure refinement of neutron
diffraction experiments (H.M. RIETVELD, 1969). It can also be used for structure
refinement of X-ray data, but is preferably used as a standardless quantification
method in many fields (cement, mineralogy and so on).
The major disadvantage of the method is that the structure data of all phases
in the sample have to be known. The currently available structure databases
incorporate most of the common phases and can easily be extended by
experimentally determined structure data, given for example as CIF or hkl files.
68
XRD for the analyst
The external standard method is not applicable because neither pure austenite nor
pure martensite phases can be produced. Nevertheless, determining the retained
austenite is an uncomplicated routine procedure. It is straightforward, because
the R values for the reflections and different wavelengths are well known (see
Chapter 7.1, FANINGER & HARTMANN) and µk’ = µk is valid (see Chapter 7.3):
2
F
R = hkl H P L
Vuc
The accuracy of the result decreases with increasing concentrations of alloy
elements. This requires a modification of the R values. Since as many reflections
as possible should be used in the calculation, the use of Mo radiation is strongly
recommended.
Note that additional international norms exist for this type of quantification.
69
XRD for the analyst
Counts
C:\Program Files\PANalytical\X'Pert HighScore Plus\Tutorial\Clinker.RD
10000 C3S - Alite, Nishi et al 72.00 %
C2S - beta - Belite (Mumme) 9.07 %
Calcium oxide - Lime 0.80 %
Brownmillerite (2/1.52/0.48/5) 10.31 %
C3A - Aluminate cubic 6.79 %
C3A - Na-Aluminate ortho, NIST 0.00 %
Magnesium oxide - Periclase 0.00 %
Potassium sulfate, beta - Arcanite 1.03 %
5000
0
20 30 40 50
400
300
200
100
0
-100
-200
-300
-400
Figure 35. A clinker diffractogram showing the experimental pattern as well as the result of the
Rietveld refinement, together with the difference plot
70
XRD for the analyst
8. Crystallographic analysis
This type of analysis is usually carried out on single phase samples. The aim is to
gain a better understanding of the crystallographic structure, ranging from unit
cell refinement to structure solution.
8.1 Indexing
In the indexing or unit cell search procedure, hkl values are assigned to the
reflections in the powder diffractogram. There are several routines described in the
literature, for example:
DICVOL (04)
ITO
TREOR
McMaille
Once possible unit cells have been found, the next step is to refine a selected unit
cell and to decide on a lattice and possible space group by recalculating the pattern
and looking for systematic extinctions.
Indexing can be very helpful to determine whether a sample consists of one phase
or more (for example: to detect crystalline impurities).
71
XRD for the analyst
72
XRD for the analyst
9. Microstructural analysis
9.1 Residual stress
9.1.1 Basics
Residual stress is mechanical stress, which is present in a material independently
of outside forces and/or moments. X-ray diffraction is an important method
to determine the residual stress by analyzing the interplanar spacings d(hkl) in
different directions.
If a material has a tensile state (or a material is stressed by an axial tensile state),
then the distances of the interplanar spacings, which are perpendicular to the
tension, are increased, while lattice planes parallel to the tension experience a
compression (due to the Poisson contraction).
Residual stresses change the interplanar distances and thus the diffraction angles
2θ (derivation of Bragg's law). By measuring the change of the diffraction angle
(∆ 2θ ) the lattice extension ε can be determined:
d − d0
ε=
d0
Figure 36 shows the definition of the angles used in stress analysis.
Now, if the coordinate system is put in such a way into the surface that the ε3
direction is the direction of the surface normal, then the extension εϕ,ψ of the
sample surface follows from the main equations of the elasticity theory for the
two axial tensile state delivering the basic equation of the X-ray residual stress
determination:
d φ,ψ − d 0 1
ε φ,ψ = = s1röσ 1 + σ 2 + s2röσ φ sin 2 ψ
d0 2
73
XRD for the analyst
X-ray measurements are always carried out along the normal of the diffracting
lattice plane (hkl). Due to the small penetration depth of the X-rays σ3 can be set to
0 and it follows
σ φ = σ 1 cos 2 ϕ + σ 2 sin 2 ϕ
The indicated X-ray elastic constants s1 and ½ s2 of σ1 and σ2 are connected by
the Voigt abbreviations with the material parameters n and E (valid for isotropic
materials):
−ν 1 ν +1
s1 = s2 =
E 2 E
E Young’s modulus (modulus of elasticity)
ν Poisson ratio (transverse contraction ratio)
74
XRD for the analyst
The Δ2θ increases with larger diffraction angles 2θ. Therefore reflections with
higher angles are preferred for residual stress analysis. This might require longer
X-ray wavelengths like Cr or Co radiation. However, when changing to longer
wavelength, the information depth will be decreased. Also, different line overlaps
can occur.
The irradiated area of the sample should be as small as possible, since during
sample tilting the absorption depth in the sample changes. Hence the tube point
focus should be used together with point focus collimators or X-ray lenses.
The different methods of determining the reflection maxima/d spacings (e.g. cross
correlation method, center of gravity, parabola etc.) are available in special analysis
software like PANalytical's Stress software. All methods have their advantages and
disadvantages, so you should choose the one that meets your requirements best.
9.2 Texture
9.2.1 Basics
In a crystal, atoms form a three-dimensional periodic arrangement (crystal lattice).
This periodicity is constant within a crystallite and ends at the grain boundaries.
Since the results of the qualitative and quantitative phase analyses are substantially
affected by textures, this has to be taken into consideration or needs to be
minimised as much as possible, see Chapter 7.2.
Note that textures are not necessarily disadvantageous for the property of
materials. They are highly desired in some cases, for example in electrical, magnetic
or also mechanical components, since the anisotropy is the basis for certain physical
properties such as superconductivity. In these cases texture analysis is used to
characterize and quantify the structure / property relation.
Measurement
A texture measurement is performed by rotating the sample around the surface
normal (∆j) and tilting it perpendicular to the beam direction (Δχ) while setting
the goniometer to a fixed 2θ - position corresponding to the selected lattice plane/
reflection (i.e. d value). A point detector is used together with a large receiving slit.
A line detector in receiving slit (0D) mode can be used as well.
75
XRD for the analyst
Pole figure
The result of a texture analysis is displayed as the so-called pole figure, which is a
two-dimensional graphical representation (e.g. stereographic projection) of the
intensity distribution of a particular Bragg reflection relative to the sample surface,
see Figure 38.
ϕ
χ
Figure 38. Representation of the intensities of a certain set of {hkl} planes with respect to the sample
reference frame (pole figure)
P − P0
F=
1 − P0
For example for a (001) texture, the value P is the sum of the integral intensities of
all (00l) directions divided by the sum of all intensity (hkl) directions in the textured
sample. P0 is the equivalent factor for the irregularly oriented sample.
P=
∑I 00l
∑I hkl
The factor F varies between 0 for a completely random orientation in the sample
and 1 for a completely oriented sample.
76
XRD for the analyst
9.3.1 Basics
The full width at half maximum of a reflection depends both on the diffractometer
geometry and the setup (slits, optics, 2θ angle, detector, etc.) and on the
sample itself. Under ideal conditions (for example: smallest divergence of the
monochromatic beam and a ‘perfect’ sample with crystallite sizes between 0.5 and
5 µm without crystal imperfections) a standard powder diffractometer can produce
diffractograms with peak widths (FWHM) between 0.03° and 0.1° 2θ over the
whole angular range.
Peak broadening caused by the sample can be explained by the presence of very
small crystallites and/or micro-strain.
Kλ
τ=
β τ cosθ
βτ Broadening of the reflection due to small crystallite sizes
with βτ = (B - b) B FWHM of the reflection of the real sample
b FWHM of the standard reflection
(crystallite size approx. 0.5 to 5 µm)
Note that βτ is indicated in rad!
K approx. 0.9 (0.89 for spherical, 0.94 for cubic crystallites)
λ wavelength
θ diffraction angle
Figure 39 describes the line broadening effect given by decreasing crystallite sizes.
Its influence will be stronger than instrumental broadening for crystallite sizes
much smaller than 1 µm.
77
XRD for the analyst
9.3.3 Micro-strain
The peak broadening βε is related to micro-strains ε, for example those caused by
defects or atom substitution, as given in the Wilson equation:
βε
ε=
4 tan θ
βε Broadening of the reflection due to lattice strain
with βε = (B2 -b2)½ B FWHM of the reflection of the real sample
b FWHM of the standard reflection
(crystallite size approx. 0.5 to 5 µm)
Note that βτ is indicated in rad!
θ diffraction angle
78
XRD for the analyst
0.7
0.65
0.6
Struct. B * cos(theta)
0.55
0.5
0.45
0.4
0.35
0.3
0.25
0.2
0.2 0.25 0.3 0.35 0.4 0.45 0.5 0.55 0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95
Size [Å]: 613.9(7) sin(theta)
Strain [%]: 0.15(1)
79
XRD for the analyst
Figure 41. Influence of strains both on position Figure 42. Influence of thermal treatment on the
and FWHM of the reflections healing process
a) unstrained
b) uniform strain
c) non uniform strain
(from JENKINS, SNYDER, 1996)
80
XRD for the analyst
10.1 Temperature
The most-widely used non-ambient variable is temperature. Low- and high-
temperature chambers are available for both reflection geometries and capillaries
for transmission. The possible temperatures range from 4 K up to more than
2000 K, using different types of non-ambient chambers. The chambers can be built
as a reactor for high pressures, for aggressive gas mixtures or for investigations
under computer-controlled humidity.
81
XRD for the analyst
There are three additional issues to be taken into account when performing non-
ambient experiments:
• Correct adjustment and possible changes in position of the sample, causing
shifts of the reflections
• Additional requirements for sample preparation, especially for high
temperatures
• Determination of the actual sample temperature
The right choice is especially important when working at low angles of incidence,
for example when investigating thin films.
The problem can be ignored if only phase transitions shall be observed.
The sample amount and the reaction volume can be minimized in the capillary
geometry. By sealing the capillary, a controlled environment can be achieved, so
that for example oxygen or humidity from the ambient air cannot influence the
experiment.
The use of capillaries is also advantageous for samples that react with metallic
heating strips. Since only indirect heating is available, the maximum temperature is
limited.
Reflection geometry
Bulk materials or powders are put on a heating strip or a special sample cup.
The direct or indirect heating has to assure a uniform heating with a minimal
temperature gradient. Small sample amounts (and samples with high thermal
82
XRD for the analyst
conductivity) are favored. In this setup, very high temperatures above 2000°C are
possible.
Possible reactions between the sample and the sample carriers (e.g. glass, Pt, Ta,
W, steel, corundum) and/or the inert gasses or an oxidative atmosphere have to be
considered. Also, the emission of reactive gasses during the temperature treatment
could damage the interior of the temperature chamber.
83
XRD for the analyst
400
380
Counts
40,588.617
39,322.087
360 38,055.556
36,789.025
35,522.495
340 34,255.964
32,989.433
31,722.903
30,456.372
320 29,189.841
27,923.31
26,656.78
300 25,390.249
24,123.718
22,857.188
Temperature [°C]
21,590.657
280 20,324.126
19,057.596
17,791.065
260 16,524.534
15,258.004
13,991.473
12,724.942
240 11,458.412
10,191.881
8,925.35
220 7,658.82
6,392.289
5,125.758
3,859.228
200 2,592.697
1,326.166
180
160
140
120
100
8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51
Position [°2theta] (copper (Cu))
84
XRD for the analyst
Table 10. Some temperatures for calibration using phase transitions / melting points
substance phase transition melting point temperature / (°C)
NH4NO3 orth. ==> tetr.; moist 84.2
tetr. ==> cubic; moist, dry 125.2
TlNO3 orth. ==> trig. 79
trig. ==> cubic 143
KNO3 orth. ==> trig. 128
CsNO3 trig. ==> cubic 160
RbNO3 trig. ==> cubic 164
cubic ==> tetr. 219
tetr. ==> cubic 291
AgNO3 orth. ==> trig. 165
Sn 231.9
KClO4 orth. ==> cubic 295.7
Pb 327.5
KNO3 333.6
Zn 419.6
Ag2SO4 orth. ==> hex. 427
CuCl 430.0
Quartz trig. ==> hex. ( ==> ) 573.0
Sb 630.5
Al 660.2
KCl 776
NaCl 804 ± 3
Bi2O3 820
Ag 962.0
NaF 988.0
Au 1064.4
K2SO4 1069
Cu 1083.0
CaF2 1360
Ca2SiO4 ==> ’ 1425 ± 10
Ni 1453
Co 1495
Fe 1535
Ti 1675
Zr 1852
Cr 1890
Rh 1966
Use of expansion coefficients:
BN: for T = 20...900°C : c / Å = 6.6516 Å + 2.74 · 10-4 · T
MgO: a /Å = 4.2100 Å · (1 + 11.39 · 10-6 T + 2.46 · 10-9 T 2)
(Taylor; Br. Ceram. Trans. J. 83,1984)
85
XRD for the analyst
10.2 Pressure
The operation of high-pressure devices requires a lot of safety precautions.
There are two basic types of high pressure chambers: conventional, commercially
available high-temperature reactor chambers that allow the investigation of
powders or solid materials under certain atmospheres and medium pressure of
up to about 10 MPa (e.g. Anton Paar XRK 900), and special non-commercial high-
pressure cells like diamond anvil cells.
The pressure cell is fully closed; it can either be used as a self-pressurizing system or
can be pressurized by using a gas tube or compressor.
The diamond anvil cell consists of two diamond anvils with a metal foil between
them. A hole (approx. 0.1 mm) in the metal foil takes the sample which is
embedded in a liquid. The high hydrostatic pressure results from pressing the
diamond anvils together, whereby the metal foil becomes smaller by plastic
deformation, and thereby the hole, too. Static pressures up to 100 GPa are
reachable.
In a different type of anvil cell (up to 10 GPa) from ALLAN, MILETICH & ANGEL
the force is generated by four screws which, when tightened, draw two halves of
the cell together. This force is then transmitted through the steel body of the cell,
through the beryllium plates to the diamond anvils. The two anvils are brilliant-
cut diamonds, almost the same as found in wedding rings except that small flat
surfaces are ground on their tips. High pressure is generated because the area of
the diamond tips or culets is much smaller than the area of the screw threads.
10.3 Humidity
Humidity chambers (e.g. Anton Paar CHC) allow the investigation of phase
transitions and stabilities under a controlled humidity. It is normally combined
with a heating and cooling facility to extend the applications to low- and high-
temperature work. The relative humidity range from 5% to 95% is computer-
controlled. The setup is ideal for investigating humidity and/or temperature-
dependent structural changes of pharmaceuticals, fine chemicals, clays or zeolites.
86
XRD for the analyst
M.E. Bowden and M.J. Ryan, Comparison of Intensities from Fixed and Variable
Divergence X-Ray Diffraction Experiments, Powder. Diffr. (1991), 6, 78
G.W. Brindley, The effects of grain and particle size on X-ray reflections from mixed
powders and alloys, considered in relation to the quantitative determination of
crystalline substances by X-ray methods, Phil. Mag. (1945), 36, 347 - 369
G. Caglioti, A. Paoletti & F.P. Ricci, Choice of collimators for a crystal spectrometer
for neutron diffraction, Nucl. Inst. (1958), 3, 223 - 228
S.J. Chipera & D.L. Bish, Fullpatt: a full-pattern quantitative analysis program for
X-ray powder diffraction using measured and calculated patterns, J. Appl. Cryst.
(2002), 35, 744 - 749
L.W. Finger, D.E. Cox & A.P. Jephcoat, A correction for powder diffraction peak
asymmetry due to axial divergence, J. Appl. Cryst. (1994), 27, 892 - 900
J.D. Hanawalt, see: Appendix in: ‘Search Manuals for the Powder Diffraction File’,
International Centre for Diffraction Data, Swarthmore PA, USA, (1988)
R.J. Hill and C.J. Howard, Quantitative Phase Analysis from Powder Diffraction Data
using the Rietveld Method, J. Appl. Cryst. (1987), 20, 467
87
XRD for the analyst
T.H. de Keijser, J.L. Langford, E.J. Mittemeijer and A.B.P. Vogels, Use of the
Voigt function in a single line method for the analysis of X-ray diffraction line
broadening, J. Appl. Cryst. (1982), 15, 308-314
L.A. Kelley, S.P. Gardner and M.J. Sutcliffe, An automated approach for clustering
an ensemble of NMR-derived protein structures into conformationally-related
subfamilies, Protein Engineering (1996), 9 1063-1065
H.P. Klug and L.E. Alexander, X-ray Diffraction Procedures, Wiley, New York (1974),
661 pages
J.L. Langford and A.J.C. Wilson, Scherrer after Sixty Years: A Survey and Some New
Results in the Determination of Crystallite Size, J. Appl. Cryst. (1978), 11, 102-113.
G.S. Pawley, Unit-cell refinement from powder diffraction scans, J. Appl. Cryst.
(1981), 14, 357-361
M. Sato, Y. Sato and L.C. Jain, Fuzzy Clustering Models and Applications, Studies in
Fuzziness and Soft Computing vol. 9, Springer group (1997), New York, 122 pages
A. Savitsky and M.J.E. Golay, Smoothing & differentiation of data by simplified least
squares procedures, Anal. Chem. (1964) 36, 8, 1627-1639
P. Scherrer, Bestimmung der Grösse und der inneren Struktur von Kolloidteilchen
mittels Röntgenstrahlen, Nachr. Ges. Wiss. Göttingen, (1918), 2, 96-100
D.K. Smith, G.G. Jr. Johnson, A. Scheible, A.M. Wims, J.L. Johnson and G. Ullmann,
Quantitative X-ray powder diffraction method using the full diffraction pattern,
Powder. Diffr. (1987), 2, 73-77
G.S. Smith and R.L. Snyder, a criterion for rating powder diffraction patterns and
evaluating the reliability of powder indexing, J. Appl. Cryst. (1979), 12, 60-65
R.L. Snyder, The Use of Reference Intensity Ratios in X-Ray Quantitative Analysis,
Powder. Diffr. (1992), 7, 186-193
88
XRD for the analyst
J.W. Visser, A fully automatic program for finding the unit cell from powder data,
J. App. Cryst (1969), 2, 89
D.B. Wiles and R.A. Young, A new computer program for Rietveld analysis of X-ray
powder diffraction patterns, J. Appl. Cryst. (1981), 14, 149-151
R.A. Young (editor), The Rietveld Method, Oxford University Press, Oxford (1993),
paperback (1995, 2000)
89
XRD for the analyst
12. Symbols
The symbols used in formulas in this book are:
90
XRD for the analyst
91
XRD for the analyst
1. Index
Symbols D
0D detector – 38 Dauvillier law – 13
1D detector – 38 Debye-Scherrer ‘cones’ – 40
2D detector – 38 DICVOL – 71
Diffractogram – 27
A d-I tables – 55
Absorption – 60 Duane-Hunt law – 13
Amorphous amounts – 62
Amorphous state – 18 E
Anatase – 66 Electromagnetic waves – 10
Anode – 12 Epitaxy – 75
Area detectors – 42 Ewald sphere – 21
ASTM – 56 Extinction – 61
Attenuation coefficient (μ ) – 11
F
B Fluorescence – 11
Bragg-Brentano – 30 Foil transmission – 47, 49
Bragg-Brentano geometry – 33 FOM (figure of merit) – 71
Bragg’s law – 22 Full pattern methods – 68
Bravais lattices – 19 Full width at half maximum – 36
Bremsstrahlung – 13, 16 FWHM – 36, 77
C G
Calibration – 83 Gas proportional counter – 40
Capillary – 47, 49, 50 Gels, liquids – 46, 47
Capillary geometry – 82
Capillary measurements – 45 H
Cement – 69 High-pressure devices – 86
Characteristic radiation – 16 Humidity chambers – 86
Characteristic X-ray radiation – 13
Characteristic X-rays – 14 I
Clinker – 69 ICDD (International Centre for
Cluster analysis – 56 Diffraction Data) – 56
Coherent or Rayleigh scatter – 11 Incoherent scatter – 11
Compton scatter – 11 Intensity I0 – 11
Crystal classes – 20 Interplanar spacings – 20
Crystalline state – 18 ITO – 71
Crystallinity – 28
Crystallite size – 28 J
Crystal structure – 17 JCPDS – 56
92
XRD for the analyst
K Peak table – 55
K-edge – 15 Percentage crystallinity
K-shell – 14, 15 determination – 62
Kβ filter – 36 PIXcel3D – 41
Point detectors – 40
L Poisson ratio – 74
Lattice parameters – 71 Pole figure – 76
Lattice planes – 20 Preferred orientation – 59
L-edges – 15 Pultrusion – 75
Limit of detection – 62
Limit of quantification – 62 R
Line detectors – 41 Rapid solidification – 75
LOD – 48 Reciprocal lattice – 18
LOQ – 48 Reflection – 45
Low angles – 28 Reflection geometry – 30, 82
L-shell – 14, 15 Reflection mode – 47, 48, 49, 50
Reflection profile – 78
M Residual stress – 73
McMaille – 71 Resolution – 27
Micro-absorption – 61 Retained austenite – 69
Micro-strains – 78 Rietveld method – 67, 68
Miller indices – 20 Rolling – 75
Monochromator – 36 Rough surfaces – 52
Monochromators – 37 Rutile – 66
Moseley law – 16
S
N Scherrer equation – 77
Non-ambient – 81 Scintillation counter – 41
Search/match – 55
O Semi-quantitative methods – 67
ODF – 76 Soller slit – 78
Omega scan – 77 Space groups – 20
Orientation distribution function – 76 Spiking method – 66
Synchrotrons – 12
P
Particle size – 28 T
PCA (principal components analysis) – 57 Thin layers – 51
PDF-2 – 56 Transmission – 45
PDF-4+ – 56 Transmission geometry – 32
PDF-4/Minerals – 56 Transmission mode – 50
PDF-4/Organics – 56 TREOR – 71
PDF database
(Powder Diffraction File) – 56 U
Peak profile – 78 Unit cell – 18, 71
93
XRD for the analyst
V
Voigt abbreviations – 74
W
Wavelength – 10, 11
Wilson equation – 78
X
X'Celerator – 41
X-ray tube – 12
Y
Young’s modulus – 74
94