V-Sub B22 Capture
V-Sub B22 Capture
V-Sub B22 Capture
263
01 ARTICLE 22
RADIOGRAPHIC STANDARDS
SE-1030 Standard Test Method for Radiographic Examination of Metallic Castings.. 314
(ASTM E 1030.95)
SE-1114 Standard Test Method for Determining the Focal Size of Iridium-
[ASTM E 1114.92 192 Industrial Radiographic Sources, 328
(K1997)]
SE-1815 Standard Test Method for Classification of Film Systems for Industrial
(ASTM E 1X15-96) Radiography 370
264
STANDARD GUIDE FOR RADIOGRAPHIC TESTING
SE-94
265
SIG94 2001 SECTION V
266
ARTICLE 22 RADlOGRAPHlC STANDARDS SE-94
TABLE 1
TYPICAL STEEL HVL THICKNESS IN INCHES (MM)
FOR COMMON ENERGIES
Energy Thickness, in. (mm)
120 kV 0.10 (2.5)
150 kV 0.14 (3.6)
200 kV 0.20 (5.1)
250 kV 0.25 (6.4)
400 kV (II 192) 0.35 (8.9)
1 Mv 0.57 (14.5)
2 Mv (Co 60) 0.80 (20.3)
4 Mv 1.00 (25.4)
6 Mv 1.15 (29.2)
10 Mv 1.25 (31.8)
16 Mv and higher 1.30 (33.0)
TABLE 2
APPROXIMATE RADIOGRAPHIC EQUIVALENCE FACTORS FOR SEVERAL METALS
(RELATIVE TO STEEL)
Energy Level
261
SE-94 2001 SECTION V
1.1.2 To determine exposure factors for one metal 9.4.3 Variation of thickness of the material radio-
from exposure techniques for other metals. graphed.
9.4.4 Energy spectrum of the radiation used.
9.1 Definition-Filters
are uniform layers of material 11.1 Effects of back-scattered radiation can be re-
placed between the radiation source and the film. duced by confining the radiation beam to the smallest
practical cross section and by placing lead behind the
9.2 Purpose - The purpose of filters is to absorb film. In some cases either or both the back lead screen
the softer components of the primary radiation, thus and the lead contained in the back of the cassette or
resulting in one or several of the following practical film holder will furnish adequate protection against
advantages: back-scattered radiation. In other instances, this must
9.2.1 Decreasing scattered radiation, thus increasing be supplemented by additional lead shielding behind
contrast. the cassette or film holder.
9.2.2 Decreasing undercutting, thus increasing con- 11.2 If there is any question about the adequacy of
trast. protection from back-scattered radiation, a characteristic
symbol [frequently a ‘/& (3.2~mm) thick letter R]
9.2.3 Decreasing contrast of parts of varying should be attached to the back of the cassette or film
thickness. holder, and a radiograph made in the normal manner.
If the image of this symbol appears on the radiograph
9.3 Location - Usually the filter will be placed in
as a lighter density than background, it is an indication
one of the following two locations:
that protection against back-scattered radiation is insuf-
9.3.1 As close as possible to the radiation source, ficient and that additional precautions must be taken.
which minimizes the size of the filter and also the
contribution of the filter itself to scattered radiation to
the film. 12. Scrcem
9.3.2 Between the specimen and the film in order
to absorb preferentially the scattered radiation from the
specimen. It should be noted that lead foil and other 12.1.1 Lead foil screens are commonly used in
direct contact with the films, and, depending upon their
metallic screens (see 12.1) fulfil1 this function.
thickness, and composition of the specimen material,
9.4 Thickness and Filrer Material - The thickness will exhibit and intensifying action at as low as 90
and material of the filter will vary depending upon the kV. In addition, any screen used in front of the film
following: acts as a filter (Section 9) to preferentially absorb
scattered radiation arising from the specimen, thus
9.4.1 The material radiographed.
improving radiographic quality. The selection of lead
9.4.2 Thickness of the material radiographed. Screen thickness, or for that matter, any metallic screen
7.68
ARTICLE 22 RADIOGMPHlC STANDARDS SE-94
thickness, is subject to the same considerations as from lead screens with a solvent. Fluorescent screens
outlined in 9.4. Lead screens lessen the scatter reaching should be cleaned in accordance with the recommenda-
the film regardless of whether the screens permit n tions of the manufacturer. Screens showing evidence
decrease or necessitate an increase in the radiographic of physical damage should be discarded.
exposure. To avoid image unsharpness due to screens,
there should be intimate contact between the lead screen
and the film during exposure. 13. Radiographic Contrast
12.1.2 Lead foil screens of appropriate thickness 13.1 The various radiation intensities that penetrate
should be used whenever they improve radiographic an object are rendered as different photographic densities
quality or penetrameter sensitivity or both. The thickness in a radiograph. Using transmitted or reflected light to
of the front lead screens should be selected with care view a radiograph, an observed change in film density
to avoid excessive filtration in the radiography of over a background is defined as contrast. Radiographic
thin or light alloy materials, particularly at the lower contrast depends mostly upon subject contrast and film
kilovoltages. In general, there is no exposure advantage gradient.
to the use of 0.005 in. in front and back lead screens 13.2 Subject contrast is the ratio of radiation intensi-
below 125 kV in the radiography of ‘lcin. (6.35-mm) ties transmitted by two selected portions of a specimen.
or lesser thickness steel. As the kilovoltage is increased
to penetrate thicker sections of steel, however, there 13.3 The film gradient is the value of the slope of
is a significant exposure advantage. In addition to the tangent line drawn to a particular density point on
intensifying action, the back lead screens are used as the characteristic curve to the abscissa. Film mannfactnr-
protection against back-scattered radiation (see Section ers can furnish characteristic curves of their products.
II) and their thickness is only important for this 13.4 The quality of radiography is influenced by
function. As exposure energy is increased to penetrate many variables; the effects of changes in some of these
greater thicknesses of a given subject material, it is variables are illustrated in Fig. 1.
customary to increase lead screen thickness. For radiog-
raphy using radioactive sources, the minimum thickness
of the front lead screen should be 0.005 in. (0.13 mm) 14. Geometry
for iridium-192, and 0.010 in. (0.25 mm) for cobalt-60.
14.1 The focus-film distance necessary to reduce
12.2 Other Mefallic Screen Materials: geometric nnsharpness to a negligible amount depends
12.2.1 Lead oxide screens perform in a similar upon the film or film-screen combinations, focal-spot
manner to lead foil screens except that their equivalence size, and object-film distance. Geometric unsharpness
in lead foil thickness approximates 0.0005 in. (0.013 is given [see Fig. 2(a)] by the equation:
“IllI).
U, = R/d,
12.2.2 Copper screens have somewhat less absorp-
tion and intensification than lead screens, but may where:
provide somewhat better radiographic sensitivity with ug = geometric unsharpness,
higher energy above 1 MV. F= size of the radiation source,
f= specimen thickness, when in contact with the
12.2.3 Gold, tantalum, or other heavy metal screens
film, and
may be used in cases where lead cannot be used.
do = source+hject distance.
12.3 Fluorescent Screens - In general, for a given
NOTE 3
source of radiation, fluorescent screens should be used units of U,
only when the exposure necessary without them would NOTE 4 -
be prohibitively long. In any event, if fluorescent screens
must be used, they should be proven capable of achiev-
ing the required quality level. Good screen-film contact Example:
is essential for the successful use of fluorescent screens.
12.4 Screen Care - All screens should be handled
Sourcefilm distance (d,) = 40 in.,
carefully to avoid dents and scratches, dirt, or grease Source size (FJ = 500 mils, and
on active surfaces. Grease and lint may be removed Source side of specimen to film distance (1) = 1.5 in
269
SE-94 2001 SECIION V
Radiation Duality
.n
Soft - tend toward high contrast
Hard - tend toward low contrast
Scettered Rxlietion
Small proportion -tend toward high contrast
Large proportion -tend toward low contra81
Degree of Development
Adequate - tend toward high contmt
Under or over - tend to low contrsn
Density
Low - tend toward low contrast
High -tend toward high contrast
GENERAL NOTE: The maximum usable density on Class 1, 2, and 3 film depends on the illuminator available.
270
ARTIC1.F. 22 RADlOGRAPHIC STANDARDS SE-94
Film
l---+--I
(a) Geometric Unsharpners (b) Radiographic Enlargement (cl Radiographic Distortion
LEGEND: LEGEND:
do = source-to-object distance Li = dimension of undistorted image
I = object-to-film distance L, = dimension of distorted image
L, = dimension of object L, - L[ = AL
L, = dimension of image Percentage distortion = (AL/L,) x 100
U, = R/do
Li - L, = AL
^ ^ = 2t x tan l/28
ALIL, x I”” = percentage enlargement
271
SE-94 2001 SECTION V
“1”
SDWX
Side
“I=” Pivot Line Specimen-
Focal
““i’
to.Film GeometrIcal
spot. Distance. Upaharpnsas,
mik in. mlls
- 800 s: - SO
- 600 6- - 60
- 50
- 40
- 30
- 20
1.0 - - 10
0.6 : - 8
0.6 - - 6
0.5 - - 5
0.4 : - 4
0.3 - - 3
0.2 - - 2
!2
40
- 10 0.10 - . 10
- 8 30 0.06 z . 06
- 6 0.06 - - 06
- 5 0.05 - . 05
- 4 20 0.04 - . 04
- 3 0.03 - . 03
FIG. 3
1 10
0.01
-
-
. 02
212
AKTICLE 22 - KADlOCKAPHlC STANDARDS SE-94
“I”
source
Pivot Line Specimen.
“‘F-
FOE?8
SP% Distance.
nlm cm
1 10 1 100 to-film
Side
100 7 - 1.0
900 8 : 0.6
600 6 - 0.6
500 5 - 0.5
400 4'. - 0.4
300 3 - 0.3
200 2 - 0.2
60 0.6 - 60 64 - 0.06
50 0.5 5-j - 0.05
40 0.4 - 0.04
30 0.3 - 0.03
20 0.2 - 0.02
- 0.010
- 0.009
- 0.004
- 0.003
- 0.002
- 0.001
213
SE.94 2001 SECTION V
274
ARTICLE 22 RADIOGRAPHIC STANDARDS SE-94
162.11 Penetrameter and shim thickness, 18.1.2 The minimum identification should at least
include the following: the radiographic facility’s name,
16.2.12 Special masking or filters, the date, part number and serial number, if used,
16.2.13 Collimator or field limitation device, and for unmistakable identification of radiographs with the
specimen. The letter R should be used to designate a
16.2.14 Processing method. radiograph of a repair area, and may include -1 , -2,
16.3 The recommendations of 16.2 are not mandatory, etc., for the number of repair.
but are essential in reducing the overall cost of radiogra- 18.2 Location Markers:
phy, and serve as a communication link between the
radiographic interpreter and the radiographic operator. 18.2.1 Location markers (that is, lead or high-
atomic number metals or letters that are to appear as
images on the radiographic film) should be placed on
17. Penetrameters (Image Quality Indicators) the pm? being examined, whenever practical, and not
on the cassette. Their exact locations should also be
17.1 The selection and placement of penetrameters marked on the surface of the part being radiographed,
shall be in accordance with applicable standards Method thus permitting the area of interest to be located accu-
E 142, Test Method E 747, and Practices E 801 and rately on the part, and they should remain on the part
E 1025. during radiographic inspection. Their exact location
may be permanently marked in accordance with the
17.2 Another image quality indicator (IQI) may be
customer’s requirements.
found in Test Method E 746 for determining relative
image quality response of industrial X-ray films at 18.2.2 Location markers are also used in assisting
200 KEV. the radiographic interpreter in marking off defective
areas of components, castings, or defects in weldments;
17.2.1 However, the E 746 Equivalent Penetrame-
also, sorting good and rejectable items when more than
ter Sensitivity (EPS) plaque may be used to study the
one item is radiographed on the same film.
effects of various radiographic variables on radiographic
system EPS performance. 18.2.3 Sufficient markers most be used to provide
evidence on the radiograph that the required coverage
17.2.2 For example, multiple X-ray machines may
of the object being examined has been obtained, and
be observed for their effect on EPS by holding film
that overlap is evident, especially during radiography
and processing constant and taking image quality expo-
of weldments and casting.
sures with all the X-ray machines. The machines would
be set for the given exposure condition in Test Method 18.2.4 Parts that must be identified permanently
E 746 and film density equalized. By reading the may have the serial numbers or section numbers, or
resultant films, the relative EPS variations between both, stamped or written upon them with a marking
machines may be determined. pen with a special indelible ink, engraved, die stamped,
or etched. In any case, the part should be marked in
17.2.3 Exposure condition variables may also be
an area not to be removed in subsequent fabrication.
studied using this plaque.
If die stamps are used, caution is required to prevent
17.2.4 While Test Method E 746 plaque can be breakage or future fatigue failure. The lowest stressed
useful in quantifying relative radiographic image quality, surface of the part should be used for this stamping.
these other applications of the plaque may he useful. Where marking or stamping of the pars is not permitted
for some reason, a marked reference drawing or shooting
sketch is recommended.
18. Identification of and Location Markers on
Radiographs
PART II - PROTECTION AND’ CARE OF
18.1 Identification of Radiographs:
UNPROCESSED FILM
18.1.1 Each radiograph must he identified uniquely 19. Storage of Film
so that there is a permanent correlation between the
part radiographed and the film. The type of identification 19.1 Unexposed films should be stored in such a
and method by which identification is achieved shall manner that they are protected from the effects of light,
be as agreed upon between the customer and inspector. pressure, excessive heat, excessive humidity, damaging
215
SE-94 2001SECTlONV
276
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-94
emulsion should be neutralized by an acid stop bath humidity and temperature are likely to be excessive,
or, if this is not possible, by rinsing with vigorous as is the case in tropical and subtropical climates. The
agitation in clear water. Follow the tilm manufacturer’s method of determining residual fixer concentrations may
recommendation of stop bath composition (or length be ascertained by reference to ANSI PH4.8, PH1.28, and
of alternative rinse), time immersed, and life of bath. PHl.41.
24.7 Fixing - The films must not touch one another 24.12 Drying - Drying is a function of (I) film
in the fixer. Agitate the hangers vertically for about
(base and emulsion); (2) processing (hardness of emul-
10 s and again at the end of the first minute, to ensure sion after washing, use of wetting agent); and (3)
uniform and rapid fixation. Keep them in the fixer until drying air (temperature, humidity, flow). Manual drying
fixation is complete (that is, at least twice the clearing can vary from still air drying at ambient temperature
time), but not more than 15 min in relatively fresh
to as high as 140°F (60°C) with air circulated by a
fixer. Frequent agitation will shorten the time of fixation. fan. Film manufacturers should again be contacted for
24.8 Fixer Neutralizing - The use of a hypo elimina- recommended drying conditions. Take precaution to
tor or fixer neutralizer between fixation and washing may tighten film on hangers, so that it cannot touch in the
be advantageous. These materials permit a reduction of dryer. Too hot a drying temperature at low humidity
both time and amount of water necessary for adequate can result in uneven drying and should be avoided.
washing. The recommendations of the manufacturers
as to preparation, use, and useful life of the baths
should be observed rigorously.
24.9 Washing -The washing efficiency is a function 25. Testing Developer
of wash water, its temperature, and flow, and the film
being washed. Generally, washing is very slow below 25.1 It is desirable to monitor the activity of the
60°F (16°C). When washing at temperatures above radiographic developing solution. This can be done by
85°F (3o”C), care should be exercised not to leave periodic development of film strips exposed under
films in the water too long. The films should be washed carefully controlled conditions, to a graded series of
in batches without contamination from new film brought radiation intensities or time, or by using a commercially
over from the fixer. If pressed for capacity, as more available strip carefully controlled for film speed and
films are put in the wash, partially washed film should latent image fading.
he moved in the direction of the inlet.
24.9.1 The cascade method of washing uses less
water and gives better washing for the same length of
time. Divide the wash tank into two sections (may be 26. Viewing Radiographs
two tanks). Put the films from the fixer in the outlet
section, After partial washing, move the batch of film 26.1 Transmission - The illuminator must provide
to the inlet section. This completes the wash in fresh light of an intensity that will illuminate the average
water. density areas of the radiographs without glare and it
must diffuse the light evenly over the viewing area.
24.9.2 For specific washing recommendations, con- Commercial fluorescent illuminators are satisfactory for
sult the film manufacturer. radiographs of moderate density; however, high light
intensity illuminators are available for densities up to
24.10 Wefting Agem - Dip the film for approxi-
3.5 or 4.0. Masks should be available to exclude any
mately 30 s in a wetting agent. This makes water drain
extraneous light from the eyes of the viewer when
evenly off film, which facilitates quick, even drying.
viewing radiographs smaller than the viewing port or
24.11 Residual Fixer Concentrations - If the fixing to cover low-density areas.
chemicals are not removed adequately from the film,
they will in time cause staining or fading of the 26.2 Reflection - Radiographs on a translucent or
developed image. Residual fixer concentrations permis- opaque backing may be viewed by reflected light. It
sible depend upon whether the films are to be kept is recommended that the radiograph be viewed under
for commercial purposes (3 to 10 years) or must diffuse lighting conditions to prevent excess glare.
be of archival quality. Archival quality processing is Optical magnification can be used in certain instances
desirable for all radiographs whenever average relative to enhance the interpretation of the image.
211
SE-94 2001 SECTION V
278
AKTICLE 22 RADIOGRAPHIC STANDARDS SE-94
APPENDIX
(Nonmandatory Information)
Xl. USE OF FLUORESCENT SCREENS layer thickness, is minimized by using screens having
small, evenly spaced crystals in a thin crystalline layer.
X1.1 Description - Fluorescent intensifying screens
Fluorescent screens are highly sensitive to longer wave-
have a cardboard or plastic support coated with a
length scattered radiation. Consequently, to maxim&
uniform layer of inorganic phosphor (crystalline sub-
contrast when this non-image forming radiation is exces-
stance). The support and phosphor are held together by
sive, fluorometallic intensifying screens or fluorescent
a radiotransparent binding material. Fluorescent screens
screens backed by lead screens of appropriate thickness
derive their name from the fact that their phosphor are recommended. Screen technology has seen signifi-
crystals “fluoresce” (emit visible light) when struck by
cant advances in recent years, and today’s fluorescent
X or gamma radiation. Some phosphors like calcium
screens have smaller crystal size, more uniform crystal
tungstate (CaWOJ give off blue light while others
packing, and reduced phosphor thickness. This translates
known as rare earth emit light green.
into greater screen/film speed with reduced unsharpness
Xl.2 Purpose and Film Types - Fluorescent screen and mottle. These improvements can represent some
exposures are usually much shorter than those made meaningful benefits for industrial radiography, as indi-
without screens or with lead intensifying screens, be- cated by the three examples as follows:
cause radiographic films generally are more responsive
to visible light than to direct X-radiation, gamma radia- X1.3.1 Reduced Exposure (Increased Productiv-
tion, and electrons. ity) - There arc instances where prohibitively long
exposure times make conventional radiography impracti-
X1.2.1 Films fall into one of two categories: non- cal. An example is the inspection of thick, high atomic
screen type film having moderate light response, and
number materials with low curie isotopes. Depending
screen type film specifically sensitized to have a very on many variables, exposure time may be reduced by
high blue or green light response. Fluorescent screens factors ranging from 2x to 105x when the appropriate
can reduce conventional exposures by as much as 150
fluorescent screen/film combination is used.
times, depending on film type.
X1.3 Image Qua&y and Use - The image quality X1.3.2 Improved Safety Conditions (Field Sires) -
associated with fluorescent screen exposures is a func- Because fluorescent screens provide reduced exposure,
tion of sharpness, mottle, and contrast. Screen sharpness the length of time that non-radiation workers must
depends on phosphor crystal size, thickness of the evacuate a radiographic inspection site can be reduced
crystal layer, and the reflective base coating. Each significantly.
crystal emits light relative to its size and in all directions
thus producing a relative degree of image ansharpness. X1.3.3 Extended Equipment Capability - Utiliz-
To minimize this onsharpness, screen to film contact ing the speed advantage of fluorescent screens by
should be as intimate as possible. Mottle adversely translating it into reduced energy level. An example
affects image quality in two ways. First, a “quantum” is that a 150 kV X-ray tube may do the job of a 300
mottle is dependent upon the amount of X or gamma kV tube, or that iridium 192 may be used in applications
radiation actually absorbed by the fluorescent screen, normally requiring cobalt 60. It is possible for overall
that is, faster screen/film systems lead to greater mottle image quality to be better at the lower kV with fluores-
and poorer image quality. A “structural” mottle, which cent screens than at a higher energy level using lead
is a function of crystal size, crystal uniformity, and screens.
279
OI STANDARD METHOD FOR CONTROLLING QUALITY
OF RADIOGRAPHIC TESTING
SE-142
DELETED
STANDARD REFERENCE RADIOGRAPHS 01
DELETED
7.81
01 STANDARD REFERENCE RADIOGRAPHS FOR
APPEARANCES OF RADIOGRAPHIC IMAGES
AS CERTAIN PARAMETERS ARE CHANGED
SE-242
DELETED
282
STANDARD REFERENCE RADIOGRAPHS FOR OI
DELETED
283
01 STANDARD REFERENCE RADIOGRAPHS
FOR STEEL CASTINGS UP TO 2 in. (51 mm)
IN THICKNESS
SE-446
DELETED
284
STANDARD PRACTICE FOR DESIGN,
MANUFACTURE, AND MATERIAL GROUPING
CLASSIFICATION OF WIRE IMAGE QUALITY
INDICATORS (IQI) USED FOR RADIOLOGY
SE-141
(IdenticalwithASTM Specification
E 747-97)
285
SE-747 2001 SECTION V
TABLE 1 TABLE 2
WIRE IQ1 SIZES AND WIRE IDENTITY NUMBERS WIRE DIAMETER TOLERANCES (mm)
SET A SET B Wire Diameter cd), mm Tolevance,mm
Wire Diameter, Wire Diameter 0.000 < d10.125 f0.0025
in. (mm) Wire Identity in. (mm) Wire Identity 0.125 < di 0.25 f0.005
0.25 < di 0.5 f0.01
0.0032 (0.08)~ 1 0.010 (0.25) b
0.50 < d< 1.6 to.02
0.004 (0.1) 2 0.013 (0.33) 7
1.6 <d<4 to.03
0.005 (0.13) 3 0.016 (0.4) 8
4.0 <di8 f0.05
0.0063 (0.16) 4 0.020 IO.51) 9
0.008 (0.2) 5 0.025 (0.64) 10
0.010 (0.25) b 0.032 (0.81) 11
SET C SET D
286
ARTICLE 22 RADIOGRAPHIC STANDARDS
SE-141
c-3
ii \
. SEE TABLE /
281
SE.747 2001 SECTION V
289
SE-747 2001 SECTION V
istics: groups 03, 02, and 01 for light metals and groups 7.2 Materials Groups:
1 through 5 for heavy metals.
7.2.1 Mawids Group 01:
7.1.2 The light metal groups, magnesium (Mg),
7.2.1.1 Image quality indicators (IQIs) shall be
aluminum (Al), and titanium (Ti), are identified 03,
made of titanium or titanium shall be the predominant
02, and 01 respectively, for their predominant alloying
alloying constituent.
constituent. The materials are listed in order of increas-
ing radiation absorption. 7.2.1.2 Use on all alloys of which titanium is
7.1.3 The heavy metal groups, steel, copper-base, the predominant alloying constituent.
nickel-base, and kindred alloys, are identified 1 through
1.2.2 Materials Group 02:
5. The materials increase in radiation absorption with
increasing numerical designation. 7.2.2.1 Image quality indicators (IQIs) shall be
made of aluminum or aluminum shall be the predomi-
7.1.4 Common trade names or alloy designations
nant alloying constituent.
have been used for clarification of the pertinent mate-
rials. 7.2.2.2 Use on all alloys of which aluminum
7.1.5 The materials from which the IQ1 for the is the predominant alloying constituent.
group are to be made are designated in each case and
1.2.3 Materials Group 03:
these IQIs are applicable for all materials listed in that
group. In addition, any group IQ1 may be used for 7.2.3.1 Image quality indicators (IQls) shall be
any material with a higher group number, provided the made of magnesium or magnesium shall be the predomi-
applicable quality level is maintained. nant alloying constituent.
290
AKTICLE 22 RADlOGKAPHlC STANDAKDS SE-747
7.2.3.2 Use on all alloys of which magnesium 7.2.6.2 Use on nickel-chromium-iron alloy and
is the predominant alloying constituent. 18% nickel-maraging steel.
7.2.5.2 Use on all aluminum bronzes and all 7.2.8 Materials Group 5:
nickel-aluminum bronzes.
7.2.8.1 Image quality indicators (IQIs) shall be
1.2.6 Materials Group 3: made of tin bronze (Alloy D of Specification B 139).
7.2.6.1 Image quality indicators (IQIs) shall be 7.2.8.2 Use on tin bronzes including gun-metal
made of nickel-chromium-iron alloy (UNS No. N06600) and valve bronze, or leaded-tin bronze of higher lead
(Inconel). (See Specification B 166.) content than valve bronze. Group 5 IQIs may include
291
SE-747 2001 SECTlON V
TABLE 4
WIRE SIZES EQUIVALENT TO CORRESPONDING 1T. ZT, AND 4T HOLES IN VARIOUS HOLE TYPE PLAQUES
PlaqueThickness, PlaqueIQIIdentification Diameter of wire with EPS of hole in plaque, in. (mm)
in. (mm) Number 1T 2T 4T
0.005 (0.13) 5 0.0038 (0.09) 0.006 (0.15)
0.006 (0.161 6 0.004 (0.101 0.0067 (0.18)
0.008 (0.20) 8 0.0032 (0.08) 0.005 (0.13) 0.008 (0.20)
0.009 (0.23) 9 0.0035 (0.09) 0.0056 (0.14) 0.009 (0.23)
0.010 (0.25) 10 0.004 (0.10) 0.006 (0.151 0.010 (0.25)
0.012 (0.30) 12 0.005 (0.13) 0.008 (0.20) 0.012 (0.281
0.015 (0.381 15 0.0065 (0.161 0.010 (0.25) 0.016 (0.41)
0.017 (0.43) 17 0.0076 (0.191 0.012 10.28) 0.02d (0.51)
0.020 (0.51) 20 0.010 (0.25) 0.015 (0.38) 0.025 (0.63)
0.025 (O.b4) 25 0.013 (0.33) 0.020 (0.51) 0.032 (0.81)
0.030 (0.76) 30 0.016 (0.41) 0.025 (0.63) 0.040 (1.021
0.035 (0.891 35 0.020 (0.51) 0.032 (0.81) 0.050 (1.27)
0.040 (1.02) 40 0.024 (0.63) 0.040 (1.02) 0.063 (1.57)
0.050 (1.27) 50 0.032 (0.81) 0.050 (1.27) 0.080 (2.03)
0.060 (1.52) 60 0.040 (1.02) 0.063 (1.571 0.100 (2.54)
0.070 (1.78) 70 0.050 (1.27) 0.080 (2.03) 0.126 (3.20)
0.080 (2.03) 80 0.063 (1.57) 0.100 (2.54) 0.160 (4.06)
0.100 (2.5) 100 0.080 12.03) 0.126 (3.20) 0.200 (5.08)
0.120 (3.05) 120 0.100 (2.54) 0.160 (4.06) 0.250 (6.35)
0.140 (3.561 140 0.126 (3.20) 0.200 (5.08) 0.320 (8.131
0.160 (4.06) 160 0.160 (4.06) 0.250 (6.35)
0.200 (5.081 200 0.200 (5.081 0.320 03.13)
0.240 (6.10) 240 0.250 (6.35)
0.280 (7.11) 280 0.320 (8.13)
bronze of higher lead content since leadedbronze 1.3 Method for Other Materials:
increasesin attenuationwith increasein lead content. 7.3.1 For materialsnot herein covered,IQIs of
This would be equivalentto using a lower groupIQI. the samematerials,or any othermaterial,may be used
NOTE 1- Indeveloping
theeightlistedmaterials
groups.
anumber if the following requirementsare met. Two blocks of
of other trade names or other nominal alloy designations were equal thickness,one of the material to be examined
evaluated. For the purposeof makingthispractice asusefulas
possible, thesemaW’ialsarelisted and cakgorized, by group, as (productionmaterial)andoneof the IQ1material,shall
follows: be radiographedon one film by one exposureat the
lowestenergylevel to beusedfor production.Transmis-
(I) Group 2 - HaynesAlloy IN-100. sion densitometermeasurements of the radiographic
(2) Group 3 - HaynesAlloy No. 713C,Hastelloy image of eachmaterial shall be made.The densityof
D, G. E. Alloy SEL, Haynes Stellite Alloy No. 21,
eachimage shall be between2.0 and4.0. If the image
GMR-235 Alloy, HaynesAlloy No. 93, Inconel X, densityof the IQ1 mater&lis within 1.00to 1.15times
Inconel 718, and HaynesStellite Alloy No. S-816. (-0% to +15%) the image densityof the production
(3) Group 4 - HastelloyAlloy F, HastelloyAlloy material,IQIs madeof thatIQ1materialmaybe usedin
X, and Multimeter Alloy Rene41.
radiographyof that productionmaterial.The percentage
(4) Group5 - Alloys in orderof increasingattenua-
figure is basedon the radiographicdensityof the IQ1
tion: HastelloyAlloy B, Hastelloy Alloy C, Haynes
material.
Stellite Alloy No. 31, Thetaloy,HaynesStelliteNo. 3,
HaynesAlloy No. 25. Image quality indicators(IQIs) 7.3.2 It shall always be permissibleto use IQIs
of any of thesematerialsare consideredapplicablefor of similar compositionas the materialbeingexamined.
the materialsthat follow it.
NOTE 2 - The committee formulating these recommendations 8. Image Quality Indicator (IQI) Certilication
recommends other materials may be added to the materials groups
listed as the need arisesor as more information is gained. or that 8.1 Documentsshallbe providedby theIQ1manufac-
additional
materials
groups
mayheadded. turer attestingto the following:
ARTICI>F: 22 RADIOGRAPI-IIC STANDARDS SE-747
10. Keywords
8.1.4 ASTM standard designation, for example,
ASTM E 747 - (year designation) used for manufac- 10.1 density; image quality level; IQI; radiologic;
turing. radiology; x-ray and gamma radiation
ANNEX
(Mandatory Information)
293
SE-747 2001 SECTION V
ASTM z
P-----b
\
: /
294
ARTICLE 22 RADIOGRAPHIC STANDARDS
295
SE-747 2001 SECTION V
.n c 1
/
tl
L t J
296
ARTICLE 22 RADIOGRAPHIC STANDARDS
ASTM-
1
/
/
-ID 32’
I,
i:
Ll
I
J
FIG. Al.9 ALTERNATE IDENTlFICATION LOCATIONS AND LETTER, NUI,lBER SIZE-TYPICAL ALL SETS (A,
B, c, D)
297
SE-747 2001 SECTION V
TABLE Al.1
PENETRAMETER SIZES WIRE DIAMETER, in. hm)
SET A SFT B
SETC SETD
298
ARTICLE 22 RADlOGRAPHlC STANDARDS SE-747
APPENDIX
299
SE-747 2001 SECYION V
Specimen thickness
l/2 in.
I in.
2 in.
4 in.
1 2 3 4 5 6
CONVERSION CHART FOR 2-T QUALITY LEVEL HOLES TO PERCENTAGE WIRE SENSITIVITY
300
ARTICLE 22 RADIOGRAPHIC STANDARDS SE-747
Specimen thickness
l/2 in. 1 in. 2 in.
1 2 3 4 5 6
FIG. Xl.2 CONVERSION CHART FOR 1-T QUALITY LEVEL HOLES TO PERCENTAGE WIRE SENSITIVITY
301
STANDARD GUIDE FOR
CONTROLLING THE QUALITY OF INDUSTRIAL
RADIOGRAPHIC FILM
PROCESSING
SE-999
1.3 The necessityof applyingspecificcontrolproce- 3.1 Definitions - For definitionsof terms usedin
duressuchasthosedescribedin this guideis dependent, this guide,see TerminologyE 1316.
to a certainextent,on the degreeto which a facility
adheresto good processingpracticesas a mattesof 4. Significance and Use
routineprocedure.
4.1 The provisionsin this guide are intendedto
controlthe reliability or quality of the imagedevelop-
1.4 If a nondestructive
testingagency,as described ment processonly and arenot intendedfor controlling
in PracticeE 543,is usedto performthe examination, the acceptabilityor quality of industrial radiographic
the testingagencyshall meetthe requirementsof Prac- films or of the materialsor productsradiographed.It
tice E 543. is furtherintendedthatthis guidebe usedas an adjunct
to and not a replacementfor Guide E 94.
1.5 This standard does not purport to address all
of the safety concerns, if any, associated with its use.
5. Chemical Mixing for Manual and Automatic
It is the responsibility of the user of this standard to
Processes
establish appropriate safety and health practices and
determine the applicability of regulatory limitations 5.1 Any equipmentthat comesin contactwith pro-
prior to use. For more specific safety precautionary cessingsolutionsshouldbe madeof glass,hardmbber,
statements,see42.3, 4.3.1, 6.1.2,and 6.1.9. polyethylene,PVC, enameledsteel, stainlesssteel, or
302
ARTICLE 22 - RADlOGRAPl~lIC STANDARDS SE-999
other chemically inert materials. This includes materials 5.4.2 In addition to cleaning equipment immedi-
such as plumbing, mixing impellers, and the cores of ately after use, wash any mixing apparatus that has
filter cartridges. Do not allow materials such as tin, been idle for a long period of time to eliminate dust
copper, steel, brass, aluminum, or zinc to come into and dirt that may have accumulated.
contact with processing solutions. These materials can
cause solution contamination that may result in film 5.4.3 Processing hangers and tanks should be free
fogging or rapid oxidation. of corrosion and chemical deposits. Encrusted deposits
that accumulate in tanks, trays, and processing equip-
5.2 Mixing Chemicals: ment and that are difficult to remove by conventional
5.2.1 Do not mix powdered chemicals in processor cleaning can be removed by using the specially formn-
tanks, since undissolved particles may be left in the lated cleaning agents recommended by the chemical or
square corners of the tank. Mix solutions in separate equipment manufacturer.
containers made from materials specified in 4.1.
5.3.2 The water supply should either be distilled 6.3 Temperature - Store all solutions at normal
or filtered so that it is clean and sediment-free.
room temperature, between 40 to 80°F (4 to 27°C).
5.3.3 If large tanks are used for mixing, carefully Storing solutions, particularly developer, at elevated
mark the volume levels to be certain that volumes are temperatures can produce rapid oxidation resulting in
correct. loss of activity and a tendency to stain the film. Storage
at too low a temperature can cause some solution to
5.3.4 Use of impeller-type mixers provides rapid, crystallize, and the crystals may not redissolve even
thorough mixing but take care to position the impeller with heating and stirring.
at such an angle and depth that the minimum amount
of air will be drawn into the solution. Over-mixing 6.4 Deterioration - Photographic chemicals can’
of the solutions can cause oxidation, especially with deteriorate either with age or with usage. Carefully
developers, and should be avoided. Rinse the shaft, follow the manufacturer’s recommendations for storage
impeller, and mounting clamp with water after use. life and useful capacity. Discard processing solutions
when the recommended number of films have been
5.4 Maintaining Equipment:
processed or the recommended storage life of the
5.4.1 Immediately clean all mixing equipment prepared solution has been reached, whichever oc-
after use. curs first.
303
SE-999 2001 SECTlON \i
304
ARTICIB 22 - RADIOGRAPHIC STANDARDS SE-999
7.2.1 Immersion time and solution temperature 7.2.7.1Make sure the dryer is clean and that
relationshipscan be morecloselycontrolledwith auto- no foreignmaterialhassettledon the rollers.Routinely
matic processingsincethe equipmentprovidesexternal examinethe ventilationsystemto ensurethat air paths
gagesfor monitoringpurposes.As a generalguideline, are not blocked and that films are uniformly dried.
follow the manufacturer’srecommendations for indus- The heat setting usedfor air temperatureshouldbe
trial processingmaterials.However,theactualprocedure compatiblewith the film manufacturer’srecommenda-
used should be basedon the variablesencountered tions.
by the user and his particularneeds.Check solution
temperatures daily with a thermometerto ensurethatthe 7.2.7.2The dryer efficiency can be testedby
processor’sthermometersare accurate[*l”F (0.X)]. processingsix consecutive14 by 17-in.(35 by 43-cm)
productionfilms, or equivalent,and examining them
7.2.2Checkthe machinespeedby measuringthe immediately after the drying cycle is complete.If
time it takes for a given length of film to pass a dampor undriedareasareobserved,increasethe dryer
specificpoint. (For example,if the indicatedmachine temperature. Shouldan increasein temperaturenot dry
speedis 2 ft/min, placetwo markson a lengthof film the film, retestthe fixer activity.
1 ft apart.The secondmark should passa specific
location,suchas the entranceto the processor,exactly 7.3 Dry Processing:
30 s after the first mark has passedthe samepoint.) 7.3.1 Follow manufacturer’srecommendations
for
An optionalmethodfor measuringprocessorspeedis thermalprocessorwarm-uprequirements.
to install a tachometeron the main drive motor and
determinedesiredRPM/processingspeedrelationships. 7.3.2 Follow time-temperaturerecommendations
from the manufacturer.
7.2.3 Agitation is providedby the action of the
processorrollers,recirculationpumps,washwaterflow,
and no externalagitationis needed. 8. Activity Testing of Solutions for Manual and
7.2.4For processorswith replenishmentsystems, Automatic Processing
use the replenishmentratesrecommendedby the film 8.1 Developer - A suggestedmethod of testing
or solution manufacturer. developeractivity is by processinga productionradio-
7.2.4.1Accuratereplenishment
increases
theuse- graphywith a knownaim densityin the areaof interest,
andthenmeasuringthe actualdensityin thatarea.The
ful life of solutionsto a great extent by replacing
ingredientsthat aredepletedandmaintainsthe process actual densityshouldbe within kO.15log E units of
at a constant,efficient level. the expectedaim density.
NOTE 2 - The term known aim density refers to an area on a
7.2.4.2 Check replenishmentsystemsat least routinely radiographed part (such as a flange or a step wedge) whose
daily to ensurethat correctvolumesarebeinginjected thickness or composition is consistent. This area,provides a benchmark
into the solutions.For installationsprocessingvery density for the process, and can verve the function nomally associated
with a process control strip.
largeamountsof film (in excessof two tank turnovers
of replenisherper week),checkson replenishmentrates 8.2 Fixer:
shouldbemademorefrequently.Processormanufactur-
er’s recommendations will generallyprovide an ade- 8.2.1 Fixer solution activity can be tested by
quateprocedurefor checkingreplenishmentvolumes. determiningthetime requiredto removethesilverfrom
an unexposedfilm. The film should be periodically
7.2.5For seasoningfreshly mixed developersolu- agitatedduring processing.An active fixer solution
tion, refer to the provisionsin 7.1S. shouldremovethe silver (thecloudy appearance in the
fixer bath shoulddisappear)in approximatelyone-half
7.2.6 Always fill the fixer tank first, following of the time recommendedfor tbe process.
the manufacturer’sinstructions,then rinse and fill the
developertank. This minimizesthe possibilityof fixer 8.2.2If physicalexaminationshowsunfixedspots
accidentallysplashinginto thedevelopersolution.When or areas,the fixer shouldbe discarded.Unfixed areas
replacing or removing processorracks, always use may appearas dull, nonreflectiveareasthat may be
a splash guard to further reduce the possibility of yellowish in color dependingon the actual lack of
contamination. fixer activity.
305
SE-999 ‘2001 SECTION V
10. Mainlenance
8.3.3 If physical examination of the films after 10.1 Maintenance schedules provided by the manufac-
washing shows dirt or scum that was not present before turer for preventive maintenance should be adhered to
washing, the wash tanks should be drained and cleaned. in order to assure consistent chemical and mechanical
Drain wash tanks whenever they are not being used. operation as set forth by the manufacturer.
In order to minimize washing artifacts it is recommended
that scavenger jlms be processed at start up to clear
11. Keywords
out scum and foreign material; the use of algaecides
is also recommended to retard the growth of organisms 11.1 automatic processing; film; manual processing;
within the wash bath. processing; radiographic; solutions
306
STANDARD PRACTICE FOR DESIGN,
MANUFACTURE, AND MATERIAL GROUPING
CLASSIFICATION OF HOLE-TYPE IMAGE QUALITY
INDICATORS (IQI) USED FOR RADIOLOGY
SE-1025
(Identical
withASTM Specification
E 1025.98)
1. scope 3. Terminology
1.1 This practicecoversthedesign,materialgrouping 3.1 Definitions - The definitionsof terms relating
classification,andmanufactureof hole-typeimagequal- to gamma and x-radiologyin TerminologyE 1316,
ity indicators(IQI) used to indicate the quality of SectionD, shall applyto thetermsusedin thispractice.
radiologicimages.
1.2 This practiceis applicableto X-ray andgamma- 4. Hole-Type IQ1 Requirements
ray radiology.
4.1 Imagequality indicators(IQIs) usedto determine
1.3 The valuesstatedin inch-poundunits are to be radiologic-imagequality levels shall conform to the
regardedas standard. following requirements.
1.4 This standard does not purport to address all 4.1.1 Standard Hole-Type IQls:
of the safety concerns, if any, associated with its use.
It is the responsibility of the user of this standard to 4.1.1.1Imagequality indicators(IQIs) shall be
establish appropriate safety and health practices and fabricatedfrom materialsor alloys identifiedor listed
determine the applicability of regulatory limitations in accordancewith 7.3. Other materialsmay be used
prior to use. in accordancewith 7.4.
4.1.1.2 Image quality indicators (IQIs) shall
2. Referenced Documents dimensionallyconform to the requirementsof Fig. 1.
2.1 ASTM Standards: 4.1.1.3Both the rectangularandthecircularIQ1
B 139Specificationfor PhosphorBronzeRod,Bar, and shall be identifiedwith number(s)madeof lead or a
Shapes materialof similar radiationopacity.The numbershall
B 150Specificationfor AluminumBronzeRod,Bar,and be bondedto the rectangularIQIs andshall be placed
Shapes adjacentto circular IQIs to provide identificationof
B 161Specificationfor Nickel SeamlessPipe andTube the IQ1 on the image.The identificationnumbersshall
B 164 Specificationfor Nickel-CopperAlloy Rod,Bar, indicatethe thicknessof the IQ1 in thousandths of an
andWire inch, that is, a number 10 IQ1 is 0.010 in. thick, a
B 166 Specificationfor Nickel-Chromium-IronAlloys number 100 IQ1 is 0.100 in. thick, etc. Additional
(UNS N06600,NO6601,and NO6690)and Nickel- identificationrequirementsare providedin 7.2.
Chromium-Cobalt-Molybdenum Alloy (UNSN06617)
Rod,Bar, andWire 4.1.1.4 Alloy-group identificationshall be in
E 1316Terminologyfor Nondestructive Examinations accordancewith Fig. 2. RectangularIQls shall be
307
SE-1025 2001 SECTION V
Place identification
numbers here -
-
f F i- + 101 thickness
I+ Ad
(a) Design for it% to and Including 160
+- IO1 thickness
lb) Design for IDIs Over 160
Identification
Number T Tolerances
(Note 3) A B C D E F (Note 2)
308
AR’l’lCl>E 22 RAUIOC?RAPHIC STANDARDS SE-1025
5. IQ1 Procurement
5.1 When selecting IQIs for procurement, the follow-
ing factors should be considered:
309
SE-1025 2001 SECTION V
TABLE 1
TYPICAL IMAGE QUALITY LEVELS
“Far Level 2.2T Radiologic - The 2 r hole in an 141, & (2%) of 5. The materials increase in radiation absorption with
the specimen thickness, is visible. increasing numerical designation.
‘Equivalent IQI sensitivity is that thickness of the IQ,, expressed as
a percentage of the part thickness, in which the 2 T hole would be NOTE 3 - Thesegroups were establishedexperimentallyat 180
visible under the wrw conditions. kV on ‘/,-in. (19.mm) thick specimens.They apply from 125 kV
to the multivolt range.
310
ARTICLE 22 RADIOGRAPI-IIC STANDARDS SE-1025
311
SE-1025 2001 SECTION V
and dimensional tolerances of the IQls specified by of radiographs since the results merely state whether
this practice. there is confonnancc to the criteria for success specified
in this practice.
9.1 Precision and Bias - No statement is made 10.1 density; image quality level; IQI; radiologic;
about the precision or bias for indicating the quality radiology; X-ray and gamma radiation
APPENDIX
(Nonmandatory Information)
312
ARTICLE 22 RADIOGRAPbIIC STANDARDS SF,-1025
313
STANDARD TEST METHOD FOR RADIOGRAPHIC
EXAMINATION OF METALLIC CASTINGS
SE-1030
(Identical
withASTM Specification
E 1030-95)
314
ARTICLE 22 ~ RADIOGRAPHIC STANDARDS SE.1030
E 999 Guide for Controlling the Quality of industrial 5.1.4 Procedure Requirements (see 9.1, 9.1.1, 9.3,
Radiographic Film Processing and 9.7.7) shall be specified.
E 1025 Practice for Design, Manufacture, and Material
5.1.5 Records - Record retention (see 12.1) shall
Grouping Classitication of Hole-Type Image Quality
be specified.
Indicators (IQI) Used for Radiology
E 1254 Guide for Storage of Radiographs and Unexposed
Industrial Radiographic Film 6. Apparatus
E 13 16 Terminology for Nondestructive Examination
6.1 Radiation Sources:
2.2 ASNT/ANSI Standards:
Recommended Practice No. SNT-TC-IA “Personnel 6.1.1 X Radiation Sources - Selection of appro-
Qualification and Certification in Nondestructive priate X-ray voltage and current levels is dependent
Testing” upon variables regarding the specimen being examined
CP-189 Qualification and Certification of Nondestructive (material type and thickness) and economically permissi-
Testing Personnel ble exposure time. The suitability of these X-ray parame-
ters shall be demonstrated by attainment of required
2.3 Military Standard: penetrameter (IQI) sensitivity and compliance with all
MIL-STD-410 Nondestructive Testing Personnel Quali- other requirements stipulated herein. Guide E 94 con-
fication and Certification tains provisions concerning exposure calculations and
charts for the use of X-ray som’ces.
3. Terminology 6.1.2 Gamma Radiation Sources - Isotope
3.1 neJinitions - For definitions of terms used in sources, when used, shall be capable of demonstrating
this test method, see Terminology E 1316. the required radiographic sensitivity.
6.2 Film Holders and Cassettes - Film holders and
4. Significance and Use cassettes shall be light-tight and shall be handled prop-
erly to reduce the likelihood that they may be damaged.
4.1 The requirements expressed in this test method They may be flexible vinyl, plastic, or any durable
are intended to control the quality of the radiographic material; or, they may be made from metallic materials.
images, to produce satisfactory and consistent results, In the event that light leaks into the film holder and
and are not intended for controlling the acceptability produces images on the film extending into the area
or quality of materials or products. of interest, the film shall be rejected. If the film holder
exhibits light leaks, it shall be repaired before reuse
or discarded. Film holders and cassettes should he
5. Basis of Application
routinely examined to minimize the likelihood of
5.1 The following items shall be agreed upon by light leaks.
the purchaser and supplier:
6.3 Intensifying Screens:
5.1.1 Nondestructive Testing Agency Evaluation -
If specified in the contractual agreement, nondestructive 6.3.1 Lead-Foil Screens:
testing (NDT) agencies shall be qualified and evaluated 6.3.1.1 Intensifying screens of the lead-foil type
in accordance with Practice E 543. The applicable are generally used for all production radiography. Lead-
version of Practice E 543 shall be specified in the foil screens shall be of the same approximate area
contractual agreement. dimensions as the film being used and they shall be
5.1.2 Personnel Qualijcation - NDT personnel in direct contact with the film during exposure.
shall be qualified in accordance with a nationally recog-
6.3.1.2 For X-ray voltages between 200 kV and
nized NDT personnel qualification practice or standard
1 MeV, front and rear screen thicknesses shall be a
such as ANSIIASNT-CP-189, SNT-TC-lA, MIL-STD-
minimum of 0.005 in. (0.13 mm) thick. Below 200
410 or a similar document. The practice or standard
kV, front screen thicknesses up to 0.005 in. (0.13 mm)
used and its applicable revision shall be specified in and rear screen thicknesses of at least 0.005 in.
the contractual agreement between the using parties.
(0.13 mm) should be used if they improve radiographic
5.1.3 Requirements -- General requirements (see quality. For isotope and high-voltage X-radiography
8.1, 8.2, 8.5, and 8.7.4) shall be specified. (greater than 1 MeV) increased thicknesses may be
315
SE-1030 2001 SECTlON V
appropriate for improvements in radiographic quality energy radiography it may be necessary to use markers
and should be used accordingly. Intermediate screens that are l/s in. (3.17 mm) or more thick.
(between multiloaded film) may be used if desired.
6.9 Radiographic Density A4easurement Apparatus -
6.3.1.3 Sheet lead, with or without backing, Either a transmission densitometer or a step-wedge
used for screens should be visually examined for dust, comparison film shall be used for judging film density
dirt, oxidation, cracking or creasing, foreign material requirements. Step wedge comparison films or densitom-
or other conditions that could render undesirable nonrel- eter calibration, or both, shall be verified by comparison
evant images on the film. with a calibrated step-wedge film traceable to the
National Institute of Standards and Technology.
6.3.1 Fluorescent or Fluorometallic Screens:
316
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-1030
with the image of any discontinuityin the area of may be performedprior to heattreatmentandin theas-
intereston the radiograph.If any doubt exists as to cast, rough-machined,
or finished-machined condition.
the true natureof an indicationexhibitedby the film,
9.1.1Penerrameter (@I) Selection -Unless other-
the radiographshall be retakenor rejected.
wise specifiedin the applicablejob order or contract,
8.4 Radiographic @aliry Level - The applicable penetrameter(IQI) selectionshall be basedon the
job orderor contractshall dictatethe requirementsfor following: if the thicknessto be radiographed exceeds
radiographicquality level. (See PracticeE 1025 or thedesignthicknessof thefinishedpiece,thepenetrame-
PracticeE 747 for guidancein selectionof quality ter (IQI) sizeshall be basedon a thicknesswhich does
level.) not exceedthe designthicknessof the finishedpiece
by more than 20% or ‘/4 in. (6.35mm), whicheveris
8.5 Acceptance Level .- Radiographicacceptance
greater.In no caseshall the penetrameter(IQI) size
levelsandassociatedseve:ritylevels shall be stipulated
be basedon a thicknessgreaterthan the thicknessto
by the applicablecontract,job order,drawing,or other
be radiographed.
purchaserand supplieragreement.
9.2 Suface Preparation - The castingsurfacesshall
8.6 Radiographic - Radio-
Density Limitations
graphicdensityin the areaof interestshall be within be preparedas necessaryto remove any conditions
1.5 to 4.0 for either single or superimposed
viewing. that could mask or be confusedwith internal casting
discontinuities.
8.7 Film Handling:
9.3 Source-to-FilmDistance - Unless otherwise
8.7.1 Darkroom Facilities -
Darkroomfacilities specifiedin theapplicablejob orderor contract,geomet-
should be kept clean and as dust-freeas practical. ric onsharpness(Up) shall not be greaterthan one
Safelightsshouldbe thoserecommended by film mam- percentof themaximumpartthicknessbeinginterpreted
facturersfor the radiographicmaterialsusedandshould on the radiograph,or 0.070 in. (1.8 mm), whichever
be positionedin accordancewith the manufacturer’s is less. Geometriconsharpness valuesshall be deter-
recommendations. All darkroomequipmentandmateri- mined as specifiedin Guide E 94.
als shouldbe capableof producingradiographsthat
are suitablefor interpretat,ion. 9.4 Direction of Radiation -The directionof radia-
tion shall be governedby the geometryof the casting
8.7.2 Film Processing - Radiographicfilm pro- andthe radiographiccoverageandqualityrequirements
cessingshall be controlledin accordancewith Guide stipulatedby the applicablejob order or contract.
E 999. Wheneverpracticable,place the central beam of the
8.7.3Film Viewing Facilities - Viewing facilities radiationperpendicular
to thesurfaceof thefilm. Appen-
shallprovidesubduedbackgroundlighting of an inten- dix X2 providesexamplesof preferredsourceandfilm
sity thatwill not causetroublesome reflections,shadows, orientationsand examplesof casting geometriesand
or glareon the radiograph.The viewing light shall be configurationson which radiographyis impracticalor
of sufficientintensityto reviewdensitiesup to 4.0 and very difficult.
beappropriately controlledsothattheoptimumintensity 9.5 Back-Scattered Radiation Protection:
for singleor superimposed viewing of radiographsmay
be selected. 9.5.1 Back-Scattered Radiation - (secondary
radi-
ation emanatingfrom surfacesbehindthe film, that is,
8.7.4 Storage of Radiographs - Whenstorageis walls,floors,etc.)servesto reduceradiographiccontrast
requiredby the applicablejob order or contract,the and may produceundesirableeffects on radiographic
radiographsshouldbe storedin an areawith sufficient quality. A I/s-in. (3.17mm) lead sheetplacedbehind
environmentalcontrol to precludeimage deterioration the film generallyfurnishesadequateprotectionagainst
or otherdamage.The radiographstoragedurationand back-scattered radiation.
locationafter castingdeliveryshall be as agreedupon
betweenpurchaserand supplier.(See Guide E 1254 9.5.2 To detectback-scattered radiation,position
for storageinformation.) a leadletter‘5” [approximately‘/8in. (3.17mm) thick
by ‘/z in. (12.7mm) high] on the rear sideof the film
holder. If a light image (lower density) of the lead
9. Procedure letter “B” appearson the radiograph,it indicatesthat
9.1 Time of Examination - Unlessotherwisespeci- moreback-scatterprotectionis necessary.The appear-
fied by the applicablejob orderor contract,radiography anceof a dark imageof the lead letter “B” shouldbe
317
SE-1030 2001 SECTION V
disregarded unless the dark image could mask or he block conforming to the requirements of 6.7 may
confused with rejectablc casting defects. be used.
9.6 Penerrameter (IQIJ Placement - Place all pene- Y.7.6.1 The penetrameter (IQI) shall be no closer
trameters (IQI) being radiographed on the source side to the film than the source side of that part of the
of the casting. Place penetrameters (IQIs) in the radio- casting being radiographed in the current view.
graphic area of interest, unless the use of a shim or
9.7.6.2 The radiographic density measured adja-
separate block is necessary, as specified in 9.7.6.
cent to the penetrameter (IQI) through the body of the
9.1 Number of Penammeters (IQls): shim or separate block shall not exceed the density
measured in the area of interest by more than 15%.
9.7.1 One penetrameter (IQI) shall represent an The density may be lighter than the area of interest
area within which radiographic densities do not vary density, provided acceptable quality level is obtained
more than +30% to -15% from the density measured and the density requirements of 8.6 are met.
through the body of the penetrameter (IQI),
9.7.6.3 The shim or separate block shall be
9.7.2 When the film density varies more than
placed at the corner of the film holder or close to that
-15% to +30%, two penetrameters (IQIs) used as part of the area of interest that is furthest from the
follows will be acceptable: if one penetrameter (IQI) central beam. This is the worst case position from a
shows acceptable sensitivity representing the most dense
beam angle standpoint that a discontinuity would be in.
portion of the exposure, and the second penetrameter
(IQI) shows acceptable sensitivity representing the least 9.7.6.4 The shim or separate block dimensions
dense portion of the exposure, then these two penetrame- shall exceed the penetrameter (IQI) dimensions such
ters (IQIs) shall qualify the exposure location within that the outline of at least three sides of the penetrameter
these densities, provided the density requirements stipu- (IQI) image shall be visible on the radiograph.
lated in 8.6 are met.
9.7.7 Film Side Penetrameter (IQIJ - In the case
9.7.3 For cylindrical or flat castings where more where the penetrameter (IQI) cannot be physically
than one film holder is used for an exposure, at least placed on the source side and the use of a separate
one penetrameter (IQI) image shall appear on each block technique is not practical, penetrameters (IQIs)
radiograph. For cylindrical shapes, where a panoramic placed on the film side may be used. The applicable
type source of radiation is placed in the center of the job order or contract shall dictate the requirements for
cylinder and a complete or partial circumference is film side radiographic quality level (see 8.4).
radiographed using at least four overlapped film holders,
at least three penetrameters (IQIs) shall be used. On 9.8 Location Markers - The radiographic image of
partial circumference exposures, a penetrameter (IQI) the location markers for the coordination of the casting
shall be placed at each end of the length of the image with the film shall appear on the film, without interfering
to be evaluated on the radiograph with the intermediate with the interpretation, in such an arrangement that it
penetrameters (IQIs) placed at equal divisions of the is evident that the required coverage was obtained.
length covered. For full circumferential coverage, three These marker positions shall be marked on the casting
penetrameters (IQIs) spaced 120” apart shall be used, and the position of the markers shall be maintained
even when using a single length of roll film, on the part during the complete radiographic cycle.
The RSS shall show all marker locations.
9.7.4 When an array of individual castings in a
circle is radiographed, the requirements of 9.7.1 or 9.9 Radiographic Identification - A system of posi-
9.7.2, or both, shall prevail for each casting. tive identification of the film shall be provided. As a
minimum, the following shall appear on the radiograph:
9.7.5 If the required pettetrameter (IQI) sensitivity the name or symbol of the inspecting laboratory, the
does not show on any one film in a multiple film date, the casting identification number, and whether it
technique (see 9.11), but does show in composite is an original or subsequent exposure.
(superimposed) film viewing, interpretation shall be
9.10 Subsequent Exposure Identijication - All repair
permitted only by composite film viewing for the
radiographs after the original (initial) shall have an
respective area.
inspection status designation that indicates the reason.
9.7.6 When it is not practicable to place the Subsequent radiographs made by reason of a repaired
penetrameter(s) (IQI) on the casting, a shim or separate area shall be identified with the letter “R” followed
318
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-IO30
by the respective repair cycle (that is, R-1 for the first 11. ReferenceRadiographs
repair, R-2 for the second repair, etc.). Subsequent 11.1 Reference Radiographs E 155, E 186, E 192,
radiographs that are necessary as a result of additional E 272, E 280, E 310, E 446, E 505, E 689, and E
surface preparation should be identified by the letters 802 are graded radiographic illustrations of various
“REG.” casting discontinuities. These reference radiographs may
be used to help establish acceptance criteria and may
9.11Multiple Film Techniques - Two or more films also be useful as radiographic interpretation training
of equal or different speeds in the same cassette are aids.
allowed, provided prescribed quality level and density
requirements are met (see 9.7.2 and 9.7.5). 12. Report
14. Keywords
10.2 Film Evaluation .- Determine the acceptance
or rejection of the casting by comparing the radiographic 14.1 castings; gamma-ray; nondestructive testing; ra-
image to the. agreed upon acceptance criteria (see 8.5). diographic; radiography; x-ray
319
SE-1030 2001 SECTION V
APPENDICES
(Nonmandatory Information)
320
ARTICLE22 - RADIOGRAPHIC
STANDARDS SE.1030
321
SE-1030 2001 SECTION V
REVElON PP.GE 3 OF J
is No. xxx-WY-22 A
FIG. X1.2(b)
(bl VIEWS ILLUSTRATING LAYOUT AND EXTENT OF COVERAGE
X1.2.1.4 Spaces for approval (as applicable). (ij Required penetrameter (IQI) quality level,
(jj Radiographic acceptance standard, and
X1.2.1.5 Radiographic Technique Parameters
(k) Applicable radiographic severity level.
for Each Location:
(a) Radiographic location designation,
X1.2.2 The sketch(es) should provide the fol-
(bj Source type and size,
lowing:
(c) Finished thickness,
(dj Thickness when radiographed,
(e) Penetrameters, X1.2.2.1 Location marker placement.
(f) Source to film distance,
(6) Film type and quantity, X1.2.2.2 Location of foundry’s identification pad
(hj Film size, or symbol on the casting.
322
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-1030
X1.2.2.3 Designation of areas that require radiog- pilot radiographs to the purchaser for review. Purchaser
raphy (as applicable). approval of the technique shall be addressed in the
applicable job order or contract.
X1.2.2.4 Designation of areas that are considered
impractical or very difficult to radiograph (see 1.2
and 8.2). x2. PREFERRED SOURCE AND FILM
X1.2.2.5 Radiographic source and film arrange- ALIGNMENT FOR FLANGE
RADIOGRAPHY AND EXAMPLES OF
ment and radiation beam direction for each location.
AREAS THAT ARE CONSIDERED
NOTE XI.1 - The RSS should designate the involved locations IMPRACTICAL TO RADIOGRAPH
and stipulate that the technique for those locations is typical, for
sections of the casting on which a continuing series of locations are X2.1 Preferred Source and Film Alignment for Flange
to be radiographed with the fame basic source and film arrangement
for each location. Radiography - The effective use of radiography for
assessing material soundness in casting areas where a
flange joins a body is somewhat limited by the source
X1.2.3 Figure X1.1 of this appendix provides a
and film alignment that the geometric configuration of
sample RSS that has been developed for a typical
these areas requires. The following figures describe
production application, and Fig. Xl .2(a) and (b) provide
source and film alignments that can be employed and
sample RSS sketches that have been developed for a discusses the limits and benefits of each.
typical production application.
X1.2.4 The RSS may not provide what is consid-
ered to be the most effective means of technique control X3. EXAMPLES OF AREAS THAT ARE
for all radiographic activities, but, in any event, some CONSIDERED TO BE IMPRACTICAL TO
means of technique standardization should be employed. RADIOGRAPH
As a general rule, it is a beneficial practice for the X3.1 Certain casting geometry configuration are inac-
supplier to solicit purchaser approval of the radiographic
cessible for conventional source and film arrangements
methodology prior to performing production radiogra-
that will provide meaningful radiographic results. These
phy. This generally entails the demonstration of the
areas generally involve the juncture of two casting
adequacy of the methodology by submitting the pro- sections. The following illustrations provide typical
posed technique parameters and a corresponding set of
examples of such areas.
323
SE-11130 2001 SECTION V
FIG. X2.2 PERMISSIBLE SOURCE AND FILM ALIGNMENT WHEN FIG. X2.1 CANNOT BE APPLIED DUE TO
CASTING GEOMETRY
324
ARTICLE 22 - RADIOGRAI’I~IIC STANDARDS SE-1030
(IdenticalwithASTM Specification
E 1079.97)
326
ARTICLE 22 - KADlOGRAPNlC STANDARDS SE-1079
tally manufactured for the purpose of measuring the tion was performed, and the identification of the individ-
range of tilm densities described in 5.1.1 may be used. ual performing the calibration, may be applied to the
densitometer to verify the calibration reference check
5.1.3 Manufacturer’s Operating Instructions for recorded in the calibration log.
Appropriate Transmission Densitometer.
7.2 An alternative calibration documentation system
may be used provided the calibration traceability re-
6. Calibration quirements identified in 7.1 can be satisfied and docu-
mented properly,
6.1 Full-scale linearity calibration should be per-
formed at least every 90 days during use as follows:
8. Periodic Verification
6.1.1 Allow a minimum of 30.min “warmup”
time (or the manufacturer’s recommended warm-up 8.1 Periodic calibration verification checks using the
time) to stabilize circuitry before starting the calibration procedure described in Section 6 should be performed
procedure or the periodic verification checks described at the beginning of each shift, after 8 h of continuous
in Section 8. Adjust the “0” reading of the densitometer operation, or change of apertures, whichever occurs first.
after the warm-up period.
8.1.1 If the verification reading is within ~~0.05
6.1.2 Select and position for reading the neutral of the density values recorded in the calibration log
density closest to 0.3, 3.0, and 3.9 on the calibrated required by 7.1, the densitometer is ready for continued
step tablet. Read and record the density for each step. use. It is not necessary to record density values when
they are within the acceptable tolerance. If the density
6.1.3 Compare the densities recorded with the
values are not within the tolerance, recalibration is
actual density values on the calibrated step tablet or required and it shall be performed in accordance with
the density values listed on the calibration certificate.
Section 6.
If the densitometer has been calibrated properly, the
densities at 0.3, 3.0, and 3.9 steps should not vary 8.1.2 If the calibration verification check shows a
more than iO.05 density units. If any of the recorded variation greater than -tO.O5, then all radiographs exam-
density values vary more than kO.05 density units from ined since the last acceptable daily density check should
the density values on the calibrated step tablet, the be subject to a reverification for density after the
linearity of the densitometer is out of tolerance and densitometer has been recalibrated.
should be corrected. 8.2 Consult the Manufacturer’s Technical Manual
for troubleshooting information.
327
STANDARD TEST METHOD FOR DETERMINING
THE FOCAL SIZE OF IRIDIUM-192 INDUSTRIAL
RADIOGRAPHIC SOURCES
SE-1114
1. Sc0pe focal size of the sonme, the distance between the source
and the object to be radiographed, and the distance
1.1 This test method covers the determination of the between the object to be radiographed and the film.
focal size of an iridium-192 radiographic source. The
This test method allows the user to determine the focal
determination is based upon measurement of the image size of the sonme and to use this result to establish
of the iridium metal sonme in a projection radiograph source to object and object to film distances appropriate
of the sonme assembly and comparison to the measure- for maintaining the desired degree of geometric un-
ment of the image of a reference sample in the same sharpness.
radiograph.
1.2 This standard does not purport to address all
of the safety concerns, if any, associated with its use.
It is the responsibility of the user qf this standard to 5. Apparatus
establish appropriate safety and health practices and
5.1 Subject Iridium-192 Source, the focal size of
determine the applicability of regulatory limitations
which is to be determined. The appropriate apparatus
prior to use.
and equipment for the safe storage, handling, and
manipulation of the subject source, such as a radio-
graphic exposure device (also referred to as a gamma
2. ReferencedDocuments ray projector or camera), remote control, source guide
2.1 ASTM Standards: tube, and source stop am also required.
E 999 Guide for Controlling the Quality of Industrial
5.2 Reference Sample (see Figs. 1,2, and 3) - The
Radiographic Film Processing
reference sample shall be of material which is not
E 1316 Terminology for Nondestructive Testing
radioactive. The recommended material is iridium. How-
ever, substitutes such as platinum, tungsten, or other
material of similar radiopacity may be used. The sample
3. Terminology should be of the same geometric shape as the subject
3.1 For definitions of terms relating to this test sonme, should he approximately the same size as the
method, refer to Terminology E 1316. subject soume, and should be positioned on or within
a shim or envelope to simulate the soume capsule wall.
The resulting radiographic contrast, with reference to
adjacent background density of the image of the refer-
4. Significanceand Use ence sample, should be approximately the same as that
4.1 One of the factors affecting the quality of a of the subject sonme. The actual dimensions of the
radiographic image is geometric nnsharpness. The de- reference sample should be determined to the nearest
gree of geometric nnsharpness is dependent upon the 0.025 mm (0.001 in.).
328
ARTICLE 22 - KAI~IOGRAPHIC STANDAKIX SE-1114
3
DIAMETER OF
SOURCE ENCAPSULATION
t
REFERENCE SAMPLE
IRIDIUM, PLATINUM, TUNGSTEN
OR SIMILAR MAT’L (DIAMETERS APROX
THE SAME SIZE AS TYPICAL SOURCES)
I ENCAPSULATION
(USUALLY ST. STL.)
I
IRIDIUM, PLATINUM. TUNGSTEN
OR SIMILAR MAT’L
329
SE-1114 2001 SECTION V
TABLE 1
EXAMPLES OF TYPICAL X-RAY GENERATOR
OUTPUT REQUIREMENTS FOR RELATED IRIDIUM’92
SOURCE ACTIVITIES
Subject Ividium’g2 Source Typical X-ray Generator
Radiation Output Requirements
Activity Output
(Curie) (Rh at 1 m) Potential Curvent
6. Procedure
6.1 Set up the exposure arrangement as shown in
ray of the X-ray beam. The source stop and reference
Figs. 4, 5, 6, and 7. Position the X-ray tube directly
sample should be 0.15 m (6 in.) from the film. The
over the center of the film. The film plane must be
source stop should be connected to the radiographic
normal to the central ray of the X-ray beam. The X-
exposure device by the shortest source guide tube
ray focal spot should be 0.90 m (36 in.) from the film,
practicable in order to mini&e fogging of the film
Position the reference sample and apparatus used to
during source transit.
locate the subject source (source stop) as close together
as possible and directly over the center of the film. 6.2 Place identification markers to be imaged on the
The plane of the source stop and reference sample film to identify, as a minimum, the identification (serial
must be parallel to the film and normal to the central number) of the subject source, the size of the reference
330
ARTICLE 22 ~ RADlOGRAl’HlC STANDAKOS s1s-1114
,-DRIVE CABLE r POSITIONING source and the reference sample. Take measurements
FIXTURE from the perceptible edges of the image. When per-
REFERENCE forming the physical measurements with the optical
SAMPLE comparator, the actual measured values shall be to the
(SEE FIG 1) nearest graduation on the graticule scale being used.
I
7.2 The focal size for a given technique is the
maximum projected dimension of the source in the
plane perpendicular to a line drawn from the source
to the object being radiographed. Therefore, sufficient
measurements of the image of the iridium must be
made to determine the focal size of the source in
any orientation. Sections 7.2.1 through 7.2.4 serve as
FIG. 6 TYPICAL ARRANGEMENT USING A
examples.
STANDARD GUIDE TUBE AND SPECIAL
POSITIONING FIXTURE 7.2.1 UnifomzRight Circular Cylinder (seeFig.
8) - Determine the focal size of a uniform right
circular cylindrical source by measuring the diameter,
d, the height, h, and the diagonal, m, as illustrated in
sample, the identification of the organ&ion performing Fig. 8 and computing the actual dimensions as described
in 8.1.
the determination, and the date of the determination.
Care should be taken to ensure that the images of 7.2.2 Sphere (see Fig. 9) - Determine the focal
the subject source and reference sample will not be size of a spherical source by measuring the diameter,
superimposed on the image of the identification markers. d, as illustrated in Fig. 9 and computing the actual
6.3 Exposure - Select the X-ray tube potential (kV), dimension as described in 8.1.
X-ray tube current (mA), and exposure time such that
7.2.3 Nonuniform Stack of Right Circular Cylinders
the density in the image of the envelope surrounding
(see Fig. 10) - Determine the focal size of a nonuni-
the reference sample does not exceed 3.0 and that the
form stack of right circular cylindrical components of
density difference between the image of the reference
a source by measuring the intrinsic diameter, d, the
sample and the image of the envelope surrounding the
height, h, and the effective maximum dimension, m,
reference sample is at least 0.10.
as illustrated in Fig. IO and computing the actual
NOTE - The actual parameters that will produce acceptable results dimensions as described in 8.1.
may vary between X-ray units. and trial exposures may be necessary.
7.2.4 Separated Stack of Right Circular Cylinders
6.3.1 Energize the X-ray generator and, at the (see Fig. 11) - Determine the focal size of a separated
same time, manipulate the subject source into the stack of right circular cylindrical components of a
exposure position in the source stop. It is important source by measuring the intrinsic diameter, d, the
that this be performed as quickly as possible to minimize effective height, h, and the effective maximum dimen-
fogging of the film. sion, m, as illustrated in Fig. 11 and computing the
actual dimensions as described in 8.1.
6.3.2 At the conclusion of the exposure time,
deenergize the X-ray generator and, at the same time,
return the subject source to the proper shielded storage
position. 8. Calculation and Evaluation
6.3.3 Process the film 8.1 Measure the linear dimension of interest in the
subject source image and measure the same linear
dimension in the reference sample image (that is, the
I. Measurementof Focal Dimensions diameter of each). The actual dimension of the subject
7.1 View the radiograph with sufficient light intensity source is computed from the following:
for adequate viewing. Using an optical comparator with
built-in graticule as described in 5.5, measure the linear
dimensions of the image of the focal spot of the subject
331
SE-1114 2001 SECTION V
SUBJECT SOURCE
REFERENCE SAMPLE
POSITIONING DEVICE
SAMPLE ARRANGEMENT
(SEE FIG. 1)
CYLINDRICAL DISC
CYilNDRlCAL
DlSC
FIG. 10 NONUNIFORM CYLINDRICAL STACK FIG. 11 SEPARATED CYLINDRICAL STACK
332
AI(I‘IC1.B 22 ~ KADIOGRAPI-IIC S’l‘ANDARIX SE-1114
9.1 A report of the focal size of an iridium-192 10.1 cylinder(s); focal size; iridium 192; radiographic
source should indicate the model number and serial source; reference sample; sphere
333
STANDARD TEST METHOD FOR MEASUREMENT OF
FOCAL SPOTS OF INDUSTRIAL X-RAY TUBES BY
PINHOLE IMAGING
SE-1165
NOTE I - Line focal spots are associated with vacuum X-ray 3.1 Definitions of Terms Specific to This Standard
lubes whose maXimum voltage rating does not generally exceed
500 LV. 3.1.1 actual focal spot - the X-ray producing area
of the target as viewed from a position perpendicular to
1.2 This test method may not yield meaningful results the target surface (see Fig. 1).
on focal spots whose nominal size is less than 0.3 mm
(0.011 in.). (See Note 2.) 3.1.2 effective focal spot - the X-ray producing
area of the target as viewed from a position perpendicu-
NOTE 2 - The X-ray tube manufacturer may he contacted for lar to the tube axis in the center of the X-ray beam
nominal focal spot dimensions. (see Fig. 1).
1.3 This test method may also be used to determine 3.1.3 linefocal spot-a focal spot whose projected
the presence or extent of focal spot damage or deteriora- pinehole image consists primarily of two curved lines
tion that may have occurred due to tube age, tube (see Fig. 2).
overloading, and the like. This would entail the produc-
tion of a focal spot radiograph (with the pinhole method)
and an evaluation of the resultant image for pitting, 4. Significance and Use
cracking, and the like.
4.1 One of the factors affecting the quality of a
1.4 Values stated in SI units are to be regarded radiographic image is geometric ansharpness. The de-
as the standard. Inch-pound units are provided for gree of geometric ansharpness is dependent upon the
information only. focal size of the radiation source, the distance between
the source and the object to be radiographed, and the
1.5 This standard does not purport to address all distance between the object to be radiographed and the
of the safety concerns. if any, associated with its use. film. This test method allows the user to determine
It is the responsibility of the user of this standard to the focal size of the X-ray source and to use this result
establish appropriate safetyand health practices and to establish source to object and object to film distances
determine the applicability of regulatory limitations appropriate for maintaining the desired degree of geo-
prior to use. metric ansharpness.
334
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-1165
SKETCH 2
335
SE-1165 2001 SECTION V
TABLE 1
PINHOLE DIAPHRAGM DESIGN REQUIREMENTS (DIMENSION)’
NOTE - The pinhole diaphragm shall be made from one of the following materials:
(1) An alloy of 90% gold and 10% platinum,
(21 Tungsten,
(3) Tungsten carbide,
(4) Tungsten alloy,
(5) Platinum and 10% Iridium alloy, or
(6) Tantalum.
Required “D” and “L” Dimensions, mm tin.1
Nominal Dimension of Focal Nominal Diameter of Diaphragm
Spot, mm ti0.Y Opening, mm (in.1 0 L
>0.3 to 1.2 (0.011 to 0.0461 inch 0.030 (0.00111 0.030 * 0.005 0.075 * 0.010
(0.0011 * 0.00021 (0.0029 * 0.00041
>1.2 to 2.5 (0.046 to 0.097) inch 0.075 (0.0029) 0.075 * 0.005 0.350 * 0.010
(0.0029 f 0.0002) to.014 * 0.0004)
>2.5 to.0971 0.100 (0.0039) 0.100 * 0.005 0.500 * 0.010
(0.0039 * 0.0002) to.02 * 0.0004)
‘See Fig. 3.
’ Nominal focal spot dimensions may be obtained from the X-ray tube manufacturer.
rl.Jmm(.059 IN.)
6. Procedure
6.1 If possible,use a standard91.44 cm (36 in.)
focal spotto film planedistance(FFD) for all exposures.
If machinegeometryor accessibilitylimitations will
FIG. 3 PINHOLE DIAPHRAGM DESIGN not permit the use of a 91.44 cm (36 in.) FFD, use
the maximumattainableFFD (in theseinstancesadjust
the relativedistancesbetweenfocal spot,pinhole,and
film accordinglyto suit the imageenlargementfactors
specifiedin Table 2). The distancebetweenthe focal
5. Apparatus spot and the pinholeis basedon the nominal size of
the focal spotbeing measuredand the desireddegree
5.1 Pinhole Diaphragm- The pinhole diaphragm of imageenlargement(seeFig. 5). The specifiedfocal
shall conformto the designand materialrequirements spotto pinholedistance(FHD) for the differentnominal
of Table 1 and Fig. 3. focal spot sizerangesis providedin Table 2. Position
5.2 Camera-The pinholecameraassemblyconsists the pinhole such that it is within +l” of the central
of the pinhole diaphragm,the shielding material to axis of the X-ray beam. Fig. 6 illustrates a typical
which it is affixed, and any mechanismthat is used focal spot exposurearrangement.
to hold the shield/diaphragmin position(jigs, fixtures,
NOTE 3 - The accuracy of the pinhole system is highly dependent
brackets,and the like; seeFig. 4). upon the relative dista.nces between (and alignment ofI the focal
spot. the pinhole. and the film. Accordingly, specially designed
5.3 Film - Industrial type extra fine grain. No apparatus may be necessary in order to assure
compliance
withthe
intensifyingscreensare to be used.The film shall be above requirements. Figure 7 provides an example of a special
collimator that can he used to ensue conformance with the al”
processedin accordancewith GuideE 999. alignment tolerance. Some other standards imposeverystringent
alignment requirements and express these requirements in terms of
5.4ImageMeasurement Apparatus- This apparatus radial tolerances. These documents do not, however. address any
is usedto measurethe size of the image of the focal means for assuring compliance with such tolerances. In order to
spot.The apparatus
shallbe an opticalcomparatorwith simplify the focal spot radiography technique and to improve the
336
SE-1165
IMAGE
TABLE 2
ENLARGEMENT
I:
FIG. 4 PINHOLE
FACTORS
CAMERA
U
(TYPICAL)
331
SE-1165 2001 SECTION V
POSlllON OF FOCAL SP
PINHOLE CAMERA
6.6.3 Measure the focal spot image in two direc- 7.2 If an image enlargement technique was used,
tions [see Fig. 9(b)]: divide the “A” and “B” direction measurements by the
pertinent enlargement factor to obtain actual focal spot
6.6.3.1 Direction A - Parallel to the axis of
size (see Fig. 5 and Table 2).
the tube.
7. Calculation of Results
338
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE.1165
10. Keywords
10.1 focal spots; pinhole camera; pinhole imaging;
x-ray; x-ray tube
339
SE-1165 2001 SECTION V
TABLE 4
DENSITY RANGE FORDARKESTAREAOFFOCAL
SPOT IMAGE (See Fig. 7)
340
ARTICLE 22 - RADIOGRAPHIC STANDARDS
341
STANDARD PRACTICE FOR RADIOSCOPY
SE-1255
342
ARTICLE 22 ~ RADIOGRAPHIC STANDARDS SE-1255
sary to not only operate the system components, but 5.1.1 The minimum radioscopic examination sys-
to make accept/reject decisions as well. Systems having tem configuration will include an appropriate source
a wide range of capabilities between these extremes of penetrating radiation, a means for positioning the
can be assembled using available components. Guide test object within the radiation beam, in the case of
E 1000 lists many different system configurations. dynamic radioscopy, and a detection system. The system
may be as simple as a directly viewed fluorescent
3.2 This practice provides details for applying radio- screen with suitable radiation shielding for personnel
scopic examination techniques, however, supplemental protection that meets applicable radiation safety codes.
requirements are necessary to address areas that are
application and performance specific. Annexes Al and 5.1.2 A more complex system might include the
A2 provide the detailed supplemental requirements for following components:
government contracts (Annex Al) and nongovernment
5.1.2.1 A microfocus X-ray system to facilitate
contracts (Annex A2).
high-resolution projection imaging;
343
SE-1255 2001 SECTION V
344
AR’I’ICLI: 22 ~ RADIOGRAPI-IIC STANDARDS SC-1255
meters) and are useful for geometric magnifications of as certain feature orientations are emphasized while
more than 100x. others are suppressed. While many digital image pro-
cessing operations occur sufficiently fast to follow time-
61.2 Manipularion System for Dynamic Radios- dependent radioscopic system variables, others do not.
copy - The test part manipulation system has the Some image processing operations require significant
function of holding the test object and providing the image acquisition and processing time, so as to limit
necessary degrees of freedom, ranges of motion, and the dynamic response of the radioscopic exam, in
speeds of travel to position the test object areas of dynamic radioscopic systems.
interest in the radiation beam in such a way so as to
maximize the radioscopic examination system’s re- 6.1.5 Automatic Evaluation System - Some radio-
sponse. In some applications it may he desirable to scopic examination applications can be fully automated
manipulate the radiation source and detection system including the accept/reject decision through computer
instead of, or in addition to, the test object. The techniques. The automatic evaluation system’s response
manipulation system must be capable of smooth well- to various test object conditions must be carefully
controlled motion, especially so for high-magnification determined under actual operating conditions. The po-
microfocus techniques, to take full advantage of the tential for rejecting good test objects and accepting
dynamic aspects of the radioscopic examination. defective test objects must he considered. Automatic
evaluation system performance criteria should be mutu-
6.1.3 Detection System - The detection system
ally determined by the provider and user of radioscopic
is a key element. It has the function of converting the
examination services.
radiation input signal containing test part information,
into a corresponding optical or electronic output signal 6.1.6 Image Display:
while preserving the maximum amount of test object
information. The detector may be of one-dimensional 6.1.6.1 The function of the image display is to
design, providing test part information one line at a time, convey radioscopic information about the test object
or may be a two-dimensional area detector providing an to the system operator. For manual evaluation systems,
area field of view. the displayed image is used as the basis for accepting
or rejecting the test object, subject to the operator’s
6.1.4 Informakm Processing of System: interpretation of the radioscopic image. The image
6.1.4.1 The function of the information pro- display performance, size, and placement are important
cessing system is to take the output of the detection radioscopic system considerations.
system and present a useful image for display and
operator interpretation, or for automatic evaluation. The 6.1.6.2 When employing a television image pre-
sentation, vertical and horizontal resolution are often
information processing system may take many different
forms, and may process analog or digital information, not the same. Therefore, the effect of raster orientation
or a combination of the two. upon the radioscopic examination system’s ability to
detect line detail, regardless of orientation, must he
6.1.4.2 The information processing system in- taken into account.
cludes all of the optics, electronics, and interfaces after
the detection system to and including the image display 6.1.1 Radioscopic Examinarion Record Archiving
and automatic evaluation system. Information system System - Many radioscopic examination applications
components include such devices as lenses, fiber optic require an archival quality examination record of the
couplings, light amplifiers, video cameras, image proces- radioscopic examination. The archiving system may
sors, and in general any device that processes radio- take many forms, a few of which are listed in 6.1.7.1
scopic examination information after the detection through 6.1.7.1 I. Each archiving system has its own
system. peculiarities as to image quality, archival storage proper-
ties, equipment, and media cost. The examination record
6.1.4.3 The digital image processing system archiving system should be chosen on the basis of
warrants special attention, since it is the means by these and other pertinent parameters, as agreed upon
which radioscopic examination information may he by the provider and user of radioscopic examination
enhanced. Great care must be exercised in determining services. The reproduction quality of the archival method
which image processing techniques are most beneficial should be sufficient to demonstrate the same image
for the particular application. Directional spatial filtering quality as was used to qualify the radioscopic examina-
operations, for example, must he given special attention tion system.
345
SE-1255 ‘2,001SECTION V
346
ARTICLE 22 ~ RADIOGRAPHIC STANDARDS SE-1255
into account the fact that the radioscopic examination 6.2.4.1 A calibrated line pair test pattern and
system may exhibit asymmetrical sensitivity, in which step wedge may be used, if so desired, to determine
case the wire diameter axis shall be oriented along the and track radioscopic system performance in terms of
system’s axis of least sensitivity. Selection of IQ1 spatial resolution and contrast sensitivity. The line pair
thickness should be consistent with the test part radiation test pattern is used without an additional absorber to
path length thickness. evaluate system spatial resolution. The step wedge is
used to evaluate system contrast sensitivity.
6.2.3 Measurement With a Calibration Block -
The calibration block may be an actual test object with 6.2.4.2 The step wedge must be made of the
known features that are representative of the range of same material as the test part with steps representing
features to be detected, or may be fabricated to simulate lOO%, 99%, 98%, and 97% of both the thickest and
the test object with a suitable range of representative the thinnest material sections to be examined. The
features. Alternatively, the calibration block may be a thinner steps shall be contiguous to their respective
one-of-a-kind or few-of-a-kind reference test object 100% section thicknesses in order to facilitate discerning
containing known imperfections that have been verified the minimum visible thickness step. Other thickness
independently. Calibration blocks containing known, steps are permissible upon agreement between the pro-
natural defects are useful on a single-task basis, but vider and the user of radioscopic services.
are not universally applicable. Where standardization
among two or more radioscopic examination systems 6.2.4.3 The line pair test pattern and the step
is required, a duplicate manufactured calibration block wedge tests shall be conducted in a manner similar
should be used. The calibration blocks should approxi- to the performance measurements for the IQ1 or the
mate the test object as closely as is practical, being calibration block set forth in 6.2.2 and 6.2.3. It is
made of the same material with similar dimensions permissible to adjust the X-ray energy and intensity to
and features in the radioscopic examination region of obtain a usable line pair test pattern image brightness.
interest. Manufactured calibration blocks should include In the case of a radioisotope or X-ray generating system
features at least as small as those that must be reliably where the energy or intensity may not be adjusted,
detected in the actual test objects in locations where additional filtration may be added at the radiation source
they are expected to occur in the actual test object. to reduce the brightness to a useful level. Contrast
Where features are internal to the test object, it is sensitivity shall be evaluated at the same energy and
permissible to produce the calibration block in sections. intensity levels as are used for the radioscopic technique.
Calibration block details are a matter of agreement
6.2.4.4 A system that exhibits a spatial resolution
between the user and supplier of radioscopic examina-
of 3 line pairs/mm, a thin section contrast sensitivity
tion services.
of 3%, and a thick section contrast sensitivity of 2%
may be said to have an equivalent performance level
6.2.3.1 Use of a Calibration Block - The
of 3%-2%-3 lplmm.
calibration block should be placed into the radioscopic
examination system in the same position as the actual 6.2.4.5 The line pair test pattern and the step
test object and may be manipulated through the same wedge may be used to make more frequent periodic
range of motions through a given exposure for dynamic system performance checks than required in accordance
radioscopic systems as are available for the actual test with 6.2.1. Resolution and contrast sensitivity checks
object, so as to maximize the radioscopic examination must be correlated with IQ1 or calibration block perform-
system’s response to the simulated imperfection. ance measurements. This may be done by first evaluating
system measurement in accordance with 6.2.2 or 6.2.3
6.2.3.2 Radioscopic Examination Techniques - and immediately thereafter determining the equivalent
(radiation beam energy, intensity, focal spot size, en- spatial resolution and contrast sensitivity values.
largement, digital image processing parameters, manipu-
lation scan plan for dynamic radioscopic systems, scan- 6.2.5 Importance of Proper Environmental Condi-
ning speed, and other system variables) utilized for the tions - Environmental conditions conducive to human
calibration block shall be identical to those used for comfort and concentration will promote examination
the actual examination of the test object. efficiency and reliability, and must be considered in
the performance of manual evaluation radioscopic exam-
6.2.4 Use of Calibrated Line Pair Test Pattern ination systems. A proper examination environment will
and Step Wedge: take into account temperature, humidity, dust, lighting,
341
SE-1255 2001 SECTION V
access, and noise level factors. Proper reduced lighting provider and the user of radioscopic examination ser-
intensity is extremely important to provide for high- vices.
contrast glare-free viewing of radioscopic examination
Images.
8. Records, Reports, and Identification of
Accepted Material
8.1 Records and reports are a matter of agreement
I. Radioscopic Examination Interpretation and
between the supplier and the user. If an examination
Acceptance Criteria
record archiving requirement exists, refer to 6.1.8, which
7.1 Interpretation - Interpretation may be done outlines the necessary information that should be a part
either by an operator in a manual evaluation radioscopic of an archival examination record.
environment, or by means of a computer and appropriate
software in the case of an automated radioscopic exami-
nation system. A hybrid environment may also be 9. Safety Conditions
utilized whereby the computer and software present to 9.1 Radioscopic examination procedures shall be
the operator a recommended interpretation, which is carried out under protective conditions so that personnel
then subject to the operator’s final disposition. will not receive radiation dose levels exceeding that
permitted by company, city, state, or national regula-
7.2 Operator - The supplier and user should reach
tions. The recommendations of the National Committee
an agreement as to operator qualifications including
on Radiation Protection should he the guide to radiation
duty and rest periods. Recommended Practice SNT- safety.
TC-IA sets forth three levels of nondestructive testing
personnel qualifications that the radioscopic examination
practitioner may find useful. 10. Keywords
7.3 Accept/Reject Criteria - Accept/reject criteria 10.1 analog; detector: digital; display; examination;
are a matter of contractual agreement between the image; manipulator; processor; radioscopy; source
ANNEXES
(Mandatory Information)
348
ARTICLE 22 ~ RADIOGRAPI-IIC STANDARDS SE-1255
E 1411 Practice for Qualification of Radiographic A1.3.7 prime contractor - a contractor having
Systems responsibility for the design control and delivery to
E 1453 Guide for Storage of Media That Contains Analog the department of defense for system/equipment such
or Digital Radiographic Data as aircraft, engines, ships, tanks, vehicles, guns and
missiles, ground communications and electronic sys-
A1.2.3 Military Standards: tems, ground support, and test equipment.
MIL-STD-410 Nondestructive Testing Personnel QuaIi-
fication and Certification A1.3.8 ted object - the material, component
MIL-STD-453 Inspection, Radiographic or assembly that is the subject of the radioscopic
DOD-STD-2167 Defense System Software Development examination.
A1.2.4 American Welding Society Standard: A1.3.9 written procedure - in radioscopy, a series
ANSIlAWS 3.0 Welding Terms and Definitions of steps that are to be followed in a regular definite
order. The radioscopic system operator follows the
A1.2.5 Government Standards - Unless otherwise written procedure to consistently obtain the desired
stated, the issues of these documents are those listed results and image quality level when performing radio-
in the Defense Index of Specifications and Standards scopic examination. The development of a radioscopic
(DODISS) and supplement thereto, cited in the contract technique usually precedes the preparation of a written
document. procedure.
A1.2.6 Order of Preference - In the event of A1.3.10 Other definitions not given herein shall
conflict between the text of this document and the be as specified in Terminology E 13 16.
references listed in A1.2.2, this document shall take Al.4 General Requirements
precedence. However, nothing in this document shall
supersede applicable laws and regulations unless a A1.4.1 Equipment Qualification - Radioscopic
specific exemption has been obtained from the cognizant system qualification shall be in accordance with Practice
authorities. E 1411 and can best be evaluated with IQIs similar
to the flaw type being investigated. A common IQ1 is
Al.3 Terminology described in MIL-STD-453.
A1.3.1 component - the test part or parts de- A1.4.2 Personnel Quali$cation - Radioscopic
scribed, assembled, or processed to the extent specified personnel shall be qualified and certified in accordance
by the drawing. with the general requirements of MII-STD-410, until
specific requirements for radioscopy are included. Ra-
A1.3.2 contracting agency - a prime contractor,
dioscopic system qualification, the development of M-
subcontractor, or government agency that procures ra-
dioscopic examination test techniques, scan plans, and
dioscopic examination services.
the overall implementation of radioscopic examination
A1.3.3 contract documents - the procuring con- in accordance with this annex, shall be under the control
tract and all drawings, specifications, standards, and and supervision of a qualified MIL-STD-410 Level III
other information included with or referred to by the with additional radioscopy training and experience or
procuring contract. in conjunction with an individual having the necessary
training and experience in radioscopic examination.
A1.3.4 mandatory radioscopic examination -
those radioscopic examinations which are a part of A1.4.3 Safety - The performance or radioscopic
the required radiographic examinations specified in the examination shall present no hazards to the safety of
contract documents. personnel or property. Applicable Federal, state, and
local radiation safety codes shall be adhered to. All
A1.3.5 NDT facility - the organization that is radioscopic procedures shall be performed in a safe
responsible for the providing of nondestructive examina- manner, such that personnel in that area are not exposed
tion services. to any radiation dosage and shall in no case exceed
Federal, state, and local limits.
A1.3.6 optional radioscopic examination - those
radioscopic examinations which are conducted for pro- Al.4.4 Archival Recording of Mandatory Radio-
cess verification or information only and are not a part scopic Emmination - When required by contractual
of the required radiographic examination specified in agreement, the radioscopic examination record shall
the contract documents. contain the results of mandatory radioscopic examina-
349
SE-1255 2001SECTIONV
tions. The radioscopicexaminationrecord shall be tion block may not be required for the following
suitably archivedfor a period of time not less than radioscopicexaminations:
live yearsfrom theexaminationdateor asmayotherwise A1.4.5.1When conductingradioscopyto check
be requiredin the contractdocuments.Efficientradio- for adequatedefect removal or grind-out, the final
scopicexaminationrecordrecallshallbeavailableat any
acceptance
radioscopicexaminationshallincludeanIQI;
time over the recordretentionperiod.The radioscopic
examinationrecordshallbe traceableto the testobject A1.4.5.2Examinationsto show materialdetails
(by serial numberor othermeans)or to the batchor or contrastbetweentwo or more dissimilar materials,
lot number,if testedin groups.Mandatoryradioscopic in componentpartsor assemblies,includinghoneycomb
examinationsshall be specifiedin the contractdocu- areasfor the detectionof fabricationirregularitiesor
ments.The optionalradioscopicexaminationsare not the presenceor absenceof material;
specifiedin the contractdocuments.
A1.4.5.3 Examinationsof electroniccomponents
A1.4.4.1 Radioscopic Examination Record - for contamination,loose or missing elements,solder
Therecordedradioscopicexaminationrecordfor manda- balls, broken or misplacedwires or connectors,and
tory examinationsshall include the written resultsof potted assembliesfor brokeninternal componentsor
theradioscopicexaminationandthe radioscopicimage, missing potting compound;
if an image is utilized in the accept/rejectdecision-
makingprocess.The recordedradioscopicimage shall A1.4.5.4Optionalradioscopicexaminations;and
be providedwith such additionalinformationas may
A1.4.5.5 Wherethe useof an IQ1is impractical
be requiredto allow the subsequent off-line review of
the radioscopicexaminationresultsand,if necessary, or ineffective,an alternatemethodmay he used,subject
the repeatingof the radioscopicexamination. to the approvalof the contractingagency.
A1.4.6 ClassiJcation of Test Object Zones for
A1.4.4.2 Image Recording Media - The radio- Radioscopy - The classificationof test objectsinto
scopic image shall be recordedon a media that is zonesfor variousaccept/rejectcriteria shall be deter-
appropriateto the radioscopicexaminationrequirement.
mined from the contractdocuments.
The recordedimage shall referencethe examination
zonesin such a way that the reviewercan confirm Al.5 Detailed Requirements
that all zoneshavebeencovered.The recordedradio-
Al.51 Application Qualijication:
scopic image shall provide an image quality at least
equal to that for which the radioscopicsystem is A1.5.1.1 New Applications - Radioscopymay
qualified. The recording media shall be capableof be usedwhereappropriatefor newexaminationrequire-
maintainingthe requiredimagequalityfor the required ments, provided the requiredperformance,including
recordstorageperiodor not less than five yearsfrom image quality, can be met.
the recordingdate.The radioscopicimagerecordshall
be maintainedin an operableconditionfor the duration A1.5.1.2 Replacement of Existing Radiographic
of the recordstorageperiod,measuredfrom the date Applications - When agreedto by the contracting
when the last radioscopicimage was recorded. officer, radioscopymay be usedto replaceor augment
existing radiographicapplications,provided that the
A1.4.4.3 Recording Media Storage Condi- radioscopicresultscorrelatefavorablywith the results
tions - Media storageandhandlingshallbe in accord- obtainedwith X-ray film producedin accordancewith
ancewith GuideE 1453. MIL-STD-453. Favorablecorrelationmeansthat the
radioscopicand film imagesshow similar sensitivity
A1.4.5 Image Quality Indicators - Imagequality to test object featuresthat are of interest.
indicatorsmust be chosenwith careto demonstrate the
radioscopicsystem’sability to detectdiscontinuitiesor A1.5.2 Written Procedure-It shallbetherespon-
otherfeaturesthat areof interest.MIL-STD-453,Prac- sibility of the NDE facility to developa written radio-
tice E 1025plaque-type,andPracticeE 747 wire-type scopicexaminationprocedureto ensuretheeffectiveand
IQIs and calibration blocks with real or simulated repeatableradioscopicexaminationof the testobject.A
defects,to match the application,are all acceptable test objectscanplan for dynamicradioscopicsystems,
unless a particular IQ1 is specifiedin the contract meeting the requirementsof Practice E 1255 (see
documents.The selectedIQ1 or calibrationblock shall 5.2.1.2), shall be included in the written procedure.
be detailedin the written procedure.An IQ1or calibra- Those portionsof the contractdocumentthat specify
350
ARTICLE 22 ~ RADIOGRAI'HIC STANDARDS SE-125s
and detail radioscopic examination shall become an Al.53 Test Object Examination - The number
appendix to the written procedure. The written procedure of test objects to be examined and the coverage required
must be approved by the Level III of the NDE facility. for each test object shall be specified in the contract
Where required, the written procedure shall be approved documents. If not specified, all test objects shall receive
by the contracting agency prior to use. The written 100% radioscopic coverage as detailed in the written
procedure shall include as a minimum the following procedure.
information:
A1.5.4 Image Quality-Unless otherwise specified
A1.5.2.1 A drawing, sketch, or photograph of in the contract documents, the required image quality
the component that shows the radiation beam axis, level is 2.2T. Image quality assessment shall be per-
positions(s) of the detector, and applicable IQ1 for each formed using the same system parameters as in the
and all variations of the test object orientation and inspection and as documented in the written procedure.
beam energy. This requirement may be expressed in
coordinates for automated systems having calibrated A1.5.4.1 The IQ1 may be placed on the test
manipulation systems; object or on a mounting block, at or near the test
object location, following the requirements of MlL-
A1.5.2.2 A physical description of the test object, STD-453. In the case of small radioscopic fields of
including size, thickness, weight, and composition; view or other situations where it is not practical to
place the IQ1 in the field of view with the test object
A1.5.2.3 Classification of the test object into and maintain it normal to the X-ray beam, the IQ1
zones for radioscopy; may be imaged immediately before and after the test
object examination. Batch quantities of similar parts
A1.5.2.4 Test part masking, if used, for each
need not have IQ1 images made between each part, at the
required view;
discretion of the Level III. The radioscopic examination
Al.S.2.5 Added radiation source collimation, results shall be invalid, if the before and after IQ1
expressed in terms of the radiation field dimensions at images fail to demonstrate the required sensitivity. The
the test object source side, for each required view; before and after IQ1 images shall be considered a
part of the test object image for radioscopic image
A1.5.2.6 Detector field of view for each re- interpretation and archiving purposes.
quired view;
A1.5.4.2 With written permission from the con-
A1.5.2.7 Detector diaphragm settings, expressed tracting agency, other IQIs or a calibration block with
in terms of field of view at the detector, for each natural or artificial flaws may be used instead of the
required view; specified IQI.
A1.5.2.8 The allowable range of radiation energy A1.5.S Radioscopic System Qualification - The
and beam current or source intensity and the focal spot radioscopic system, including mandatory radioscopic
or source size for each required view; image archiving devices, shall be qualified to the image
quality level required for test object examination. Radio-
A1.5.2.9 Added beam filtration, if used, for each
scopic system initial qualification shall be in accordance
required view:
with Practice E 1411.
A1.5.2.10 The inspection geometry and coverage
A1.5.6 Radioscopic System Requal$cation - The
for each required view;
radioscopic system, including mandatory image archiv-
A1.5.2.11 Type of IQ1 or calibration block used ing devices, shall be periodically requalified at intervals
and the required quality level; frequent enough to ensure the required level of radio-
scopic system performance. Each requalification shall
A1.5.2.12 All hardware and software settings be carried out in accordance with Practice E 1411.
that can be changed by the operator to affect the
outcome of the radioscopic examination. Such settings A1.5.7 hpecfion Image Control - The radio-
include, but are not limited to, video camera and display scopic system shall be checked for performance before
settings and image processor variables; and each day’s production usage using the method and
devices that were initially used to qualify the written
A.1.5.2.13 The recording media and storage procedure. A log shall be maintained to document any
image format for mandatory radioscopic image storage. changes in system performance requiring changes in
351
SE.1255 2001 SECTION V
operating parameters and listing all equipment mainte- A1.5.10.2 Feature Measurement by Conzpari-
nance. System requalification shall be required whenever son - A second method involves comparing the test
image quality requirements can no longer be met. obiect feature with a known, observable dimension
which must be wholly within the radioscopic field of
A1.5.8 Repair of Radioscopic System - Repair view. Many digital image processors facilitate this type
or replacement of key radioscopic system components of measurement by counting pixels over the feature
including, but not limited to, the radiation source, image length. The pixel number is often converted to engi-
forming, image transmission, image processing, and neering units by comparison with a known length.
image display subsystems shall be cause for system However, the orientation and position along the X-
requalification. In no case shall the interval between ray beam (magnification) of both the feature and the
qualification tests exceed one year. The qualification calibrating reference length affect the accuracy of such
statement shall be posted on the radioscopic system. measurements.
The results of the qualification tests shall be maintained
in the radioscopic system equipment file until the Al.511 Gray Scale Range - The gray scale
completion of the next qualification procedure or the range required to meet initial qualification contrast
expiration of the archival image retention period, which- sensitivity requirements for image quality shall be rec-
ever is longer. orded and monitored. For systems using human image
assessment, it is particularly important that the gray
A1.5.9 Image Interpretation: scale range and the number of gray scale steps be
closely matched to the response of the human eye.
A1.5.9.1 Static Imaging - Radioscopic system The written procedure shall include a means for monitor-
qualification in accordance with Practice E 1411 applies ing the required gray scale range using a contrast
to static imaging conditions only where the test part sensitivity gage, step wedge, or similar device made
is stationary with respect to the X-ray beam. Therefore, of the test object or IQ1 material.
all performance measurements are based upon static
image quality. All mandatory radioscopic examination A1.5.12 Timing of Radioscopic Examination -
accept/reject decisions shall be based upon the assess- Radioscopic examination shall be performed at the time
ment of static images. of manufacturing, assembly, or rework as required by
the contract documents.
A1.5.9.2 Dynamic Imaging - Dynamic or in-
Al.513 Identification - A means shall be pro-
motion imaging may be used to gain useful information
vided for the positive identification of the test object
about the test object. However, unless dynamic imaging
to the archival radioscopic inspection record. Archived
is specified, the final assessment of image formation
radioscopic images shall be annotated to agree with
for mandatory radioscopic examinations shall be made in
the test object identification.
the static mode. When the contracting agency specifies
dynamic inspection, all aspects of the procedure must A1.5.14 Locating the Radioscopic Examination
be approved by MIL-STD-410 Level III personnel. For Areas - Whenever more than one image is required
dynamic inspection, the image quality shall be measured for a weldment or other test object, location markers
under the same procedure as the inspection. shall be placed on the test object in order that the
orientation of the test object and the location of test
Al.510 Feature Size Determination - Where object features relative to the radioscopic field of view
feature measurement from the radioscopic image is may be established. This requirement shall not apply
required, the written procedure shall include methodol- to automated systems having programmed radioscopic
ogy for determining and maintaining the accuracy of examination sequences where coverage has been proven
the selected measurement method. during the development of the scan plan. Also, this
requirement does not apply to the radioscopic examina-
A1.5.10.1 Feature Measurement by Test Object tion of simple or small shapes where the test part
Displacement - For those radioscopic systems with orientation is obvious and coverage is not in question.
calibrated manipulation systems, the more accurate,
and therefore preferred, method of measurement is to A1.5.15 Surface Preparation - Test objects may
manipulate the extremities of the feature to be measured be inspected without surface preparation, except when
to a common central reference point within the radio- required to remove surface conditions that may interfere
scopic image field of view. The dimension may then with proper interpretation of the radioscopic image or
be read from the manipulation system position display. that may create a safety hazard.
3.52
ARTICLE 22 ~ RADIOGRAPHIC STANDARDS SE-1255
Al.516 Detailed Data - The provider of radio- Al.521 Test Object Marking - The marking of
scopic examination services shall keep the written proce- test objects shall be as specified in MIL-STD-453.
dure, qualification documentation, and the signed inspec-
tion reports or tabulated results, or both, for five years Al.6 Notes
from the radioscopic examination date, unless otherwise
specified in the contract documents. For software-based A1.6.1 This section contains information of a
automated radioscopic systems using custom software, general or explanatory nature and is not mandatory.
a copy of the sonrce code and the related inspection
parameters shall also be maintained on file for a like A1.6.1.1 Caution - Active electronic compo-
period of time. This requirement shall not apply to nents and some materials, such as tetrafluoroethylene,
standard commercially available software packages or ax subject to radiation damage if exposed to large
to traceable software documentation which complies doses of radiation. While normal radioscopic examina-
with DOD-STD-2167 where a separate copy of the tions should cause no problem, extended periods of
software is maintained. radiation exposure should be avoided.
353
SIS-1255 2001 SECTION v
A1.6.1.5 Feature Size Determination - As with the required radiographic examinations specified in the
radiography, great care must be exercised in trying to contract documents.
assess test part feature dimensions from a two-dimen-
A2.2.1 provider of radioscopic services - a con-
sional projected view.
tractor, subcontractor, or other entity that provides
radioscopic examination services.
A2.2.8 test object - the material, component,
A2. NONGOVERNMENT CONTRACT
or assembly that is the subject of the radioscopic
SUPPLEMENTAL REQUIREMENTS examination.
A2.1 Scope
A2.2.9 user of radioscopic services - a contractor,
A2.1.1 Purpose - This annex is to be used in subcontractor, or other entity that procures radioscopic
conjunction with Practice E 1255. This annex includes examination services. The provider and user of radio-
application-specific details as may be agreed upon by the scopic examination services may be a part of the same
purchaser and the supplier of radioscopic examination organization or different organizations.
services. A2.2.10 written procedure - in radioscopy, a
A2.1.2 Application - This document satisfies the series of steps that are to be followed in a regular
requirements of 3.2 and 5.2.1 of Practice E 1255. definite order. The radioscopic system operator follows
Should this annex conflict with any other requirements the written procedure to consistently obtain the desired
of Practice E 1255, this annex shall prevail. The results and image quality level when performing radio-
requirements of this annex are intended to control the scopic examination. The development of a radioscopic
quality of the radioscopic examination and not to specify technique usually precedes the preparation of a written
the accept/reject criteria for the test object. procedure.
Accept/reject criteria are provided in other contract A2.2.11 Other definitions not given herein shall
documents. be as specified in Terminology E 1316.
A2.2 Terminology A2.3 General Requirements
A2.2.1 component - the test part or parts de- A2.3.1 Equipment Qualijcation - Radioscopic
scribed, assembled, or processed to the extent specified system qualification shall be in accordance with Practice
by the drawing. E 1411, using Practice E 747 and Practice E 1025
image quality indicators or a calibration block containing
A2.2.2 contract documents - the procuring con-
actual or simulated defects.
tract and all drawings, specifications, standards, and
other information included with or referred to by the A2.3.2 Personnel Qualijication - Radioscopic
procuring contract. personnel shall be qualified and certified in accordance
with the requirements of SNT-TC-1A or ANSYASNT
A2.2.3 contractor - a contractor having first level CP-189. Radioscopic system qualification, the develop-
responsibility for the design, manufacture, and delivery ment of radioscopic examination test techniques, scan
of an end item. When radioscopic examination is re- plans, and the overall implementation of radioscopic
quired, the contractor is the user of radioscopic examina- examination in accordance with this annex shall be
tion services. under the control and supervision of a qualified Level
A2.2.4 mandatory radioscopic examination - III with additional radioscopy training and experience,
those radioscopic examinations which are a part of or in conjunction with an individual having the necessary
the required radiographic examinations specified in the training and experience in radioscopic examination.
contract documents. Operation of the radioscopic system, including interpre-
tation of the radioscopic image, shall be made by
A2.2.5 NDE faciliry - the organization that is qualified Level II personnel.
responsible for providing nondestructive examination
services. A2.3.3 Safety- The performance of radioscopic
examination shall present no hazards to the safety of
A2.2.6 optional radioscopic examination - those personnel or property. Applicable Federal, state, and
radioscopic examinations that are conducted for process local radiation safety codes shall be adhered to. All
verification or information only and are not a part of radioscopic procedures shall be performed so that per-
354
ARTICLE 22 ~ RADIOGRAPHIC STANDARDS SE-1255
sonnelshall receivethe minimum dosageand in no A2.3.5 Image Quality Indicators - An IQI must
caseexceedFederal,state,and local limits. be chosenwith care to demonstratethe radioscopic
system’sability to detectdiscontinuities,or otherfea-
A2.3.4 Archival Recording of Mandatory Radio- tures of interest. Practice E 1025 plaque-typeand
scopic Examinations - The radioscopicexamination PracticeE 747 wire-typeIQIs and calibrationblocks
recordshall containthe resultsfor mandatoryradio- with realor simulateddefectsthatmatchtheapplication
scopicexaminations.The radioscopicexaminationrec- are all acceptableunlessa specificIQ1 is specifiedin
ord shall be suitablyarchivedfor a periodof one year the contractdocuments.The selectedIQ1 or calibration
afterthedateof radioscopicexaminationor for a longer block shall be detailedin the written procedure.An
time if specifiedin the contractdocuments.Efficient IQ1 or calibrationblock may not be requiredfor the
radioscopicexaminationrecordrecall shall be available following radioscopicexaminations:
at any time over the record retention period. The
radioscopicexaminationrecord shall be traceableto A2.3.5.1 Examining assembliesfor debris or
the test objectby serial numberor other means.This foreign objects.
requirementwill not applyto optionalradioscopicexam- A2.3.5.2 Conductingradioscopyfor adequate
inationsthatarenot specifiedin the contractdocuments. defectremovalor grind-out.However,the final accept-
anceradioscopicexaminationshah includean IQI.
A2.3.4.1 Radioscopic Examination Record -
Therecordedradioscopicexaminationrecordfor manda- A2.3.5.3Examinationsto showmaterialdetails
tory examinationsshall include the written resultsof or contrastbetweentwo or more dissimilar materials
theradioscopicexaminationandtheradioscopicimage, in componentpartsor assembliesincludinghoneycomb
if an image is utilized in the accept/rejectdecision- areasfor the detectionof fabricationirregularitiesor
makingprocess.The recordedradioscopicimage shall the presenceor absenceof material.
be providedwith such additionalinformationas may
be requiredto allow the subsequent off-line review of A2.3.5.4Examining electroniccomponentsfor
the radioscopicexaminationresultsand, if necessary, contamination,looseor missingelements,solderballs,
the repeatingof the radioscopicexamination. brokenor misplacedwires, or connectorsand potted
assembliesfor brokeninternalcomponentsor missing
A2.3.4.2 Image Recording Media - The radio- potting compound.
scopic image shall be recordedon a media that is
appropriate to theradioscopicexaminationrequirement. A2.3.5.5Optionalradioscopicexaminations.
The recordedimage shall referencethe examination A2.3.5.6Wherethe useof anIQ1 is impractical
zonesin such a way that the reviewer can confirm or ineffective,an alternatemethodmay beused,subject
that all zcmeshavebeencovered.The recordedradio- to the approvalof the contractingagency.
scopic image shall providean image quality at least
equal to that for which the radioscopicsystem is A2.3.6 Classijicationof Test Object Zones for
qualified. The recordingmedia shall be capableof Radioscopy - The classificationof test objectsinto
maintainingthe requiredimagequality for the required zonesfor variousaccept/rejectcriteria shall be deter-
recordstorageperiodor not less than five yearsfrom mined from the contract documents.In caseswhere
the recordingdate. The recordedradioscopicimage no accept/rejectcriteria are specified,the Level III
playbackshall be maintainedin an operablecondition of the NDE facility shall documentthoseanomalies
for the durationof the recordstorageperiodmeasured consideredcritical andindicatein writing thatno formal
from the date when the last radioscopicimage was accept/rejectcriteria were provided.
recorded.
A2.4 Detailed Requirements
A2.3.4.3 Recording Media Storage Condi- A2.4.1 ApplicationQualification- Radioscopy
tions - Media storageandhandlingshall be in accord- may be used where appropriatefor new as well as
ancewith Guide E 1453. existing radiographicexaminationrequirementspro-
vided that the requiredperformance,including image
A2.3.4.4 Other Recording - Where the rec- quality,canbemet.Whereradioscopyis usedto replace
ordingof the radioscopicexaminationrecordis not in or augmentexistingradiographicapplications,theradio-
fulfillment of mandatoryarchival recordingrequire- scopicresultsshouldcorrelatefavorablywith theresults
ments,otherrecordingmethodsandmediamaybe used. obtainedwith radiographicfilm-producedtechniques.
SE-1255 2001SECTIONV
356
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-1255
changes in system performance requiring changes in of measurement by counting pixels over the feature
operating parameters and listing all equipment mainte- length. The pixel number is often converted to engi-
nance. System requalification shall he required whenever neering units by comparison with a known length.
image quality requirements can no longer be met. However, the orientation and position along the X-
ray beam (magnification) of both the feature and the
A2.4.8 Repair of Radioscopic System - Repair
calibrating reference length affect the accuracy of such
or replacement of key radioscopic system components
measurements.
including but not limited to the radiation source, image
forming, image transmission, image processing, and A2.4.11 Gray Scale Range - The gray scale
image display subsystems shall be cause for system range required to meet initial qualification contrast
requalification. In no case shall the interval between sensitivity requirements for image quality shall be rec-
qualification tests exceed one year. The qualification orded and monitored. For systems using human image
statement shall be posted on the radioscopic system. assessment, it is particularly important that the gray
The results of the qualification tests shall be maintained scale range and the number of gray scale steps be
in the radioscopic system equipment file at least until closely matched to the response of the human eye.
completion of the next qualification procedure or the The written procedure shall include a means for monitor-
expiration of the archival image retention period, which- ing the required gray scale range using a contrast
ever is longer. sensitivity gage, step wedge, or similar device made
A2.4.9 Image Interpretation: of the test object or IQ1 material.
358
STANDARD TEST METHOD FOR
RADIOSCOPIC EXAMINATION OF WELDMENTS
SE-1416
359
SE-1416 ?OOl SECTION V
360
ARTICLE22 - RADIOGRAPHIC
STANDARDS SE-1416
6.3 Timeof Examination - Thetime of examination 8.1.4 Image quality indicator designationand
shall be in accordance with 9.1 unlessotherwisespec- placement,
ified. 8.1.5 Test-objectscan plan, indicatingthe range
6.4 Procedures and Techniques - The procedures of motionsandmanipulationspeedsthroughwhich the
and techniquesto be utilized shall be as describedin test object shall be manipulatedin order to ensure
this test methodunless otherwisespecified.Specific satisfactoryresults(seedescriptionin 5.2.1.2of Practice
techniques may be specified in the contractual E 1255),
agreement. 8.1.6Image-processing
parameters,
6.5 EnrentofExamination- The extentof examina- 8.1.7Image-displayparameters,and
tion shall be in accordancewilh 8.3 unlessotherwise
specified. 8.1.8Image storage.
6.6 Reporting Criteria/Acceptance Re-
Criteria - 8.2 Radioscopic Coverage- Unlessotherwisespeci-
porting criteriafor the examinationresultsshall be in fied by purchaserand supplieragreement,the extent
accordance with Section10 unlessotherwisespecified. of radioscopiccoverageshall include 100% of the
Acceptancecriteriashall be specifiedin the contractual volumeof the weld and the adjacentbasemetal.
agreement. 8.3 Examination Speed - For dynamicexamination,
6.7 Reexamination of Repaired/Reworked Items - the speedof object motion relative to the radiation
Reexaminationof repaired/reworked items is not ad- sourceand detectorshall be controlledto ensurethat
dressedin this test methodand if requiredshall he the requiredradioscopicquality level is achieved.
specifiedin the contractualagreement. 8.4 Radioscopic Image Quality -
All imagesshall
be free of marks or otherblemishesthat could mask
or be confusedwith the imageof any discontinuityin
I. Safety the area of interest.It may be possible to prevent
blemishesfrom maskingdiscontinnitiesor being con-
7.1 Radioscopicproceduresshall comply with appli- fusedwith discontinuitiesby moving the object being
cable city, state,and federalsafetyregulations. examinedrelativeto the imaging device.If any doubt
existsas to the true natureof an indicationexhibited
in the image,the image shall be rejectedand a new
8. Requirements imageof the areashall be made.
8.1 Procedure Requirement -- Unlessotherwisespec- 8.5 Radioscopic QualityLevel- Radioscopicquality
ified by theapplicablejob orderor contract,radioscopic level shall he determinedupon agreementbetweenthe
examinationshall be performedin accordancewith a purchaserand supplier and shall he specifiedin the
written procedure.Specificrequirementsregardingthe applicablejob order or contract.Radioscopicquality
preparationandapprovalof the writtenproceduresshall shall be specifiedin terms of equivalentpenetrameter
be as agreedby purchaserandsupplier.The production (IQI) sensitivity and shall he measuredusing image
procedureshall addressall applicableportionsof this quality indicatorsconformingto PracticesE 747 or
test methodand shall be availablefor review during E 1025.
interpretationof theimages.The writtenprocedureshall
include the following: 8.6 Acceptance Level - Accept and reject levels
shallbe stipulatedby theapplicablecontract,job order,
8.1.1Materialandthicknessrangeto be examined, drawing,or other purchaserand supplieragreement.
8.1.2Equipmentto beused,includingspecifications 8.7 Image-Viewing Facilities - Viewing facilities
of sourceparameters(such as tube voltage, current, shall providesubduedbackgroundlighting of an inten-
focal spotsize)andimagingequipmentparameters (such sity thatwill not causetroublesomereflection,shadows,
as detectorsize,field of view, electronicmagnification, or glareon the image.
camerablack level, gain), 8.8 Storage of Images- When storageis required
8.1.3 Examinationgeometry,including source-to- by the applicablejob order or contract,the images
object distance,object-to-detector
distanceand orien- shouldbestoredin a formatstipulatedby the applicable
tation, contract,job order, drawing, or other purchaserand
361
SE-1416 2001 SECTlON V
supplier agreement.The image-storageduration and 9.1 Image Quality Indicator Selection -For selection
location shall be as agreedbetweenpurchaserand of the imagequality indicator,the thicknesson which
supplier. the imagequality indicatoris basedis the single-wall
thicknessplus the lesserof the actual or allowable
reinforcement.
Backingstripsor ringsarenotconsidered
9. Procedure as part of the weld or reinforcementthicknessfor
image quality indicator selection.For any thickness,
9.1 Time of Examination - Unlessotherwisespeci- an imagequality indicatoracceptablefor thinnermateri-
fied by the applicablejob order or contract, perform als may be used,providedall other requirementsfor
radioscopyprior to heat treatment. radioscopyare met.
9.2 Su$~cePreparation - Unlessotherwiseagreed 9.8 Number of Image Quality Indicators:
upon, removethe weld head ripple or weld-surface
irregularitieson both the inside and outside (where 9.8.1 Place at least one image quality indicator
accessible)by any suitableprocessso that the image (PracticesE 747 or E 1025)in the area of interest
of the irregularitiescannotmask,or be confusedwith, representingan areain whichthe brightnessis relatively
the image of any discontinuity.Interpretationcan be uniform. The degreeof brightnessuniformity shall he
optimized if surfaceirregularitiesare removedsuch agreedupon betweenpurchaserand supplier.If the
that the imageof the irregularitiesis not discernible. image brightnessin an area of interest differs by
more than the agreedamount,use two image quality
9.3Sourcefo Detecfor Distance - Unlessotherwise indicators.Use one imagequality indicatorto demon-
specifiedin theapplicablejob orderor contract,geomet- strateacceptableimage quality in the darkestportion
ric unsharpness (Us) shall not exceedthe following: of the image and use one image quality indicator to
Material Thickness Us max, in. (mm) demonstrateacceptableimage quality in the lightest
under 2 in. (50 nun) 0.020 (0.50) portion of the image.
2 through 3 in.@0 through 75 mm) 0.030 (0.75)
over 3 through 4 in. (75 through 0.040 (1.00) 9.8.2 When a seriesof imagesare made under
100 mm, identical conditions,it is permissiblefor the image
greater than 4 in. (100 mm) 0.070 (1.75) quality indicatorsto be usedonly on the first andlast
images in the series,provided this is agreedupon
Determinegeometricunsharpness
valuesas specifiedin betweenthe purchaserand supplier.In this case,it is
GuideE 94. not necessary for theimagequality indicatorsto appear
9.4 Examination Speed -
For dynamicexamination, in eachimage.
determinethe speedof object motion relative to the 9.8.3 Always retain qualifying images,on which
radiationsomceanddetectorupon agreementbetween one or more image quality indicatorswere imaged
the purchaserand supplier. Base this determination during exposure,as part of the record to validate
upontheachievement of therequiredradioscopicquality the required image quality indicator sensitivity and
level at that examinationspeed. placement.
- Direct the central
9.5 Direction of he Radiation 9.9 Image Quality Indicator Placement:
beamof radiationperpendicularlytowardthe centerof
the effectiveareaof the detectoror to a planetangent 9.9.1 Place the image quality indicator on the
to the centerof the image, to the maximum extent sourcesideadjacentto theweld beingexamined.Where
possible,exceptfor double-wallexposure-double-wall the weld metal is not radioscopicallysimilar to the
viewing elliptical projection techniques,as described basematerialor wheregeometryprecludesplacement
in 9.14.2. adjacentto the weld, placethe imagequality indicator
over the weld or on a separateblock, as described
9.6 Scattered Radiation - Scatteredradiation(radia- in 9.10.
tion scattered from thetestobjectandfrom surrounding
structures)reducesradioscopiccontrastand may pro- 9.9.2 Detector-Side Image Qua&y Indicators -
duce undesirableeffects on radioscopicquality. Use In those caseswhere the physical placementof the
precautionssuchas collimation of the source,collima- image quality indicators on the sourceside is not
tion of the detector,and additionalshieldingas appro- possible,place the image quality indicatorson the
priateto minimize the detrimentaleffectsof this scat- detectorside.The applicablejob orderor contractshall
teredradiation. specify the applicabledetector-sidequality level. The
362
ARTICLE 22 - RADIOGRAPHIC STANDARDS SE-1416
accompanying documents shall clearly indicate that the location markers for the identification of the part location
image quality indicators were located on the detector with the image shall appear on the image without
side. interfering with the interpretation and with such an
arrangement that it is evident that complete coverage
9.10 Separate Block - When configuration or size
was obtained.
prevents placing the image quality indicators on the
object being examined, use a shim, separate block or 9.12.1 Double-Wall Technique - When using a
like section conforming to the requirements of 5.8 technique in which radiation passes through two walls
provided the following conditions are met: and the welds in both walls are simultaneously viewed
for acceptance, and the entire image of the object being
9.10.1 The image quality indicator is no closer to examined is displayed, only one location marker is
the detector than the sonrce side of the object being
required in the image.
examined (unless otherwise specified).
9.12.2 Series of Images - For welds that require
9.10.2 The radioscopic brightness in the area of
a series of images to cover the full length or circumfer-
the image quality indicator including the shim, separate
ence of the weld, apply the complete set of location
block, or like section and IQ1 where applicable are markers at one time, wherever possible. A reference
similar to the brightness in the area of interest.
or zero position for each series must be identified on
9.10.3 The shim, separate block, or like section the component. A known feature on the object (for
is placed as close as possible to the object being example, keyway, nozzle, and axis line) may also be
examined. used for establishment of a reference position. Indicate
this feature on the radioscopic record.
9.10.4 When bole-type image quality indicators
are used, the shim, separate block, or like section 9.12.3 Similar Welds - On similar type welds
dimensions shall exceed the image quality indicator on a single component, the sequence and spacing of
dimensions such that the outline of at least three sides the location markers must conform to a uniform system
of the image quality indicator image is visible on the that shall be positively identified in the radioscopic
image. procedure or interpretation records. In addition, refer-
ence points on the component will be shown on the
9.11 Shim Utilization - When a weld reinforcement sketch to indicate the direction of the numbering system.
or backing ring and strip is not removed, place a shim
of material that is radioscopically similar to the backing 9.13 Image ldentijication - Provide a system of
ring and strip under the image quality indicators to positive identification of the image. As a minimum,
provide approximately the same thickness of material the following shall appear on the image: the name or
under the image quality indicator as the average thick- symbol of the company performing radioscopy, the
ness of the weld reinforcement plus the wall thickness, date, and the weld identification number traceable to
backing ring and strip. part and contract. Identify subsequent images made of
a repaired area with the letter “R.”
9.11.1 Shim Dimensions and Location - When
hole-type image quality indicators are used, the shim 9.14 Radioscopic Techniques:
dimensions and location shall exceed the image quality
9.14.1 Single-Wall Technique - Except as pro-
indicator dimensions by at least 0.12 in. (3 mm) on
vided in 9.14.2, 9.14.3, and 9.14.4, perform radioscopy
at least three sides. At least three sides of the image
using a technique in which the radiation passes through
quality indicator shall be discernible in accordance with
only one wall.
9.10.4 except that only the two ends of the image
quality indicator need to be discernible when located 9.14.2 Double-Wall Technique for Circumferential
on piping less than 1 in. (25 mm) nominal pipe size. Welds - For circumferential welds 4 in. (100 mm)
Place the shim so as not to overlap the weld image outside diameter (3.5 in. nominal pipe size) or less,
including the backing strip or ring. use a technique in which the radiation passes through
both walls and both walls are viewed for acceptance
9.11.2 Shim Image Brightness - The image bright-
on the same image. Unless otherwise specified, either
ness of the shim image shall be similar to the image
elliptical or superimposed projections may be used. A
brightness of the area of interest.
sufficient number of views should be taken to examine
9.12 Location Markers - Place location markers the entire weld. Where design or access restricts a
outside the weld area. The radioscopic image of the practical technique from examining the entire weld,
363
SE-1416 2001 SKTION V
agreement between contracting parties must specify 10.1.4 Image quality indicator (and shim, il’ used)
necessary weld coverage. placement,
9.14.3 For circumferential welds greater than 4 10.1.5 Test-object scan plan, including ranges of
in. (100 mm) outside diameter (3.5 in. nominal pipe motion and manipulation speeds,
size), use a technique in which only single-wall viewing 10.1.6 Image processing parameters,
is performed. A sufficient number of views should be
taken to examine the entire weld. Where design or 10.1.7 Image-storage data,
access restricts a practical technique from examining 10.1.8 Weld repair documentation, and
the entire weld, agreement between contracting parties
10.1.9 Image - Interpretation record shall contain
must specify necessary weld coverage.
as a minimum the following information:
9.14.4 For radioscopic techniques that prevent sin-
10.1.9.1 Disposition of each image (acceptable
gle-wall exposures due to restricted access, such as
or rejectable),
jacketed pipe or ship hull, the technique should be
agreed upon in advance between the purchaser and 10.1.9.2 If [ejectable, cause for rejection (slag,
supplier. It should be cecognized that image quality crack, porosity, etc.),
indicator sensitivities based on single-wall thickness
may not be obtainable under some conditions. 10.1.9.3 Surface indication verified by visual
examination (grinding marks, weld ripple, spatter,
etc.), and
364
STANDARD PRACTICE FOR DETERMINING 01
365
SE-1647 2001SECTIONV
TABLE 2
5. Significanceand Use
CONTRAST SENSITIVITY GAGE DIMENSIONS
5.1 The contrastsensitivitygagemeasurescontrast Gage
sensitivityindependentof the imaging systemspatial Size. B DIM. C DIM. D DIM. E DIM. F,G DIM.
resolutionlimitations.The thicknessrecessdimensions 1 0.750 in. 3.000 in. 0.250 in. 0.625 in. 0.250 in.
of the contrastsensitivitygageare large with respect 19.05 mm 76.20 mm 6.35 mm 15.88 mm 6.35 mm
to the spatialresolutionlimitations of most imaging 2 1.500 in. 6.000 in. 0.500 in. 1.250 in. 0.500 in.
systems.Fourlevelsof contrastsensitivityaremeasured: 38.10 mm 152.40 mm 12.70 mm 31.75 mm 12.7 mm
4%, 3%, 2%, and 1%. 3 2.250 in. 9.000 in. 0.750 in. 1.875 in. 0.750 in.
57.15 mm 228.60 mm 19.05 mm 47.63 mm 19.05 mm
5.2 The contrastsensitivitygageis intendedfor use 4 3.000 in. 12.000 in. 1.000 in. 2.500 in. 1.000 in.
in conjunctionwith a high-contrast
resolutionmeasuring 76.20 mm 304.80 mm 25.40 mm 63.50 mm 25.4 mm
gage,suchastheEN 462-5DuplexWire ImageQuality
Indicator.Suchgagesmeasurespatialresolutionessen-
tially independentof the imaging system’s contrast
sensitivity.Such measurements are appropriatefor the
TABLE 3
qualification and performancemonitoring of radio-
CONTRAST SENSITIVITY GAGE APPLICATION
graphicand radioscopicimaging systems.
Gaue Sire Use on Thicknesses
5.3 Radioscopicsystemperformance maybe specified
1 Up to 1.5 in. (38.1 mm)
by combining the measuredcontrast sensitivity ex- 2 Over 1.5 in. (38.1 mm) to 3.0 in. (76.2 mm1
pressedas a percentagewith the spatial resolution 3 Over 3.0 in. (76.2 mm1 to 6.0 in. (152.4 mm)
expressedin millimeters of unsharpness. For the EN 4 Over 6.0 in. 1152.4 mm)
462-5spatialresolutiongage,the unsharpness is equal
to twice the wire diameter.For the line pair gage,the
unsharpness is equalto the reciprocalof the line-pair/
mm value. As an example,an imaging system that
exhibits2% contrastsensitivityandimagesthe 0.1 mm characteristics:groups03, 02, and 01 for light metals
EN 462-5pairedwires (equivalentto imaging 5 line- and groupsI through5 for heavy metals.
pairs/millimeterresolutionon a line-pairgage)performs 6.2.2 The light metal groups,magnesium(Mg),
at a 2%-0.2 mm sensitivitylevel. A standardmethod aluminnm (Al), and titanium (Ti), are identified03,
of evaluatingoverall radioscopicsystemperformance 02, and01, respectively,for their predominantconstit-
is given in PracticeE 1411. uent. The materialsare listed in order of increasing
radiationabsorption.
6. Contrast Sensitivity Gage Construction and 6.2.3The heavymetalsgroup,steel,copperbase,
Material Selection nickel base,and other alloys, are identified1 through
6.1 Contrastsensitivitygagesshall be fabricatedin 5. The materialsincreasein radiationabsorptionwith
accordance with Fig. 1, using the dimensionsgiven in increasingnumericaldesignation.
Tables 1, 2, and 3. 6.2.4 Common tradenamesor alloy designations
6.2 The gageshall preferablybe fabricatedfrom the have beenusedfor clarificationof pertinentmaterials.
test objectmaterial.Othenuise,the following material 6.3 The materialsfrom which the contrastsensitivity
selectionguidelinesare to be used: gage is to be made is designatedby groupnumber.
6.2.1 Materialsare designatedin eight groupings, The gageis applicableto all materialsin that group.
in accordancewith their penetratingradiationabsorption Material groupingsare as follows:
366
ARTICLE 22 - RADIOGRAPI-llC STANDARDS SE-1647
6.3.1 Material Group 03: 6.3.5.1 The gage shall be made of aluminum
bronze (Alloy No. 623 of Specification B 150) or
6.3.1.1 The gage shall be made of magnesium
equivalent OT nickel-aluminum bronze (Alloy No. 630
or a magnesium alloy, provided it is no more radio-
of Specification B 150) or equivalent.
opaque than unalloyed magnesium, as determined by
the method outlined in 6.4. 6.3.5.2 Use for all aluminum bronzes and all
nickel aluminum bronzes.
6.3.1.2 Use for all alloys where magnesium is
the predominant alloying constituent. 6.3.6 Materials Group 3:
6.3.2 Materials Group 02: 6.3.6.1 The gage shall be made of nickel-chro-
6.3.2.1 The gage shall be made of aluminum mium-iron alloy (UNS No. N06600) (Inconel). See
or an aluminum alloy, provided it is no more radio- Specification B 166.
opaque than unalloyed aluminum, as determined by the 6.3.6.2 Use for nickel-chromium-iron alloy and
method outlined in 6.4. 18% nickel-maraging steel.
6.3.2.2 Use for all alloys where aluminum is
6.3.1 Materials Group 4:
the predominant alloying constituent.
6.3.7.1 The gage shall be made of 70 to 30
6.3.3 Materials Group 01: nickel-copper alloy (Monel) (Class A or B of Specifica-
6.3.3.1 The gage shall be made of titanium or tion B 164) or equivalent, or 70 to 30 copper-nickel
a titanium alloy, provided it is no more radio-opaque alloy (Alloy G of Specification B 161) or equivalent.
than unalloyed titanium, as determined by the method
outlined in 6.4. 6.3.7.2 Use for nickel, copper, all nickel-copper
series or copper-nickel series of alloys and all brasses
6.3.3.2 Use for all alloys where titanium is the (copper-zinc alloys) and all leaded brasses.
predominant alloying constituent.
6.3.8 Materials Group 5:
6.3.4 Materials Group 1:
6.3.8.1 The gage shall be made of tin-bronze
6.3.4.1 The gage shall be made of carbon steel (Alloy D of Specification B 139).
or Type 300 series stainless steel.
6.3.8.2 Use for tin bronzes including gun-metal
6.3.4.2 Use for all carbon steel, low-alloy steels, and valve bronze and leaded-tin bronzes.
stainless steels, and magnesium-nickel-aluminum bronze
(Superston). 6.4 Where the material to be examined is a composite
ceramic, or other nonmetallic material, or for some
6.3.5 Materials Group 2: reason cannot be obtained to fabricate a gage, an
367
SE-1647 2001 SECTION V
equivalent material may be utilized, provided it is no 8.3 The gage recess depth tolerance shall be within
more radio-opaque than the test object under comparable *IO% of the design value for the shallowest recess or
penetrating radiation energy conditions. To determine 0.001 in. (0.02 mm), whichever is greater.
the suitability of a substitute material, radiograph identi-
8.4 The gage recess inside and outside corner radius
cal thicknesses of both materials on one film using the
shall not exceed 0.062 in. (1.80 mm). To facilitate
lowest penetrating radiation energy to be used in the
fabrication, the gage may be assembled from three
actual examination. Transmission densitometer readings
individually machined components: (I) the machined
for both materials shall be in the range from 2.0 to
center section containing the 1% r, 2% K 3% i”, and
4.0. If the radiographic density of the substitute material
4% T milled slots; (2) the front rail, and (3) the rear
is within +15% to -0% of the test material, the substitute
rail. The assemblage of the three components forms
material is acceptable.
the complete gage similar to that shown in Appendix Xl.
6.4.1 All contrast sensitivity gages shall be suitably 8.5 The gage dimensional tolerances shall be held
marked by vibro-engraving or etching. The gage thick- to within *O.OlO in. (0.25 mm) of the dimensions
ness and material type shall be clearly marked. specified in Table 2.
368
ARTICLE 22 - KADIOGRAPHIC STANDARDS SE-1647
11.1 No statementis made aboutthe precisionor contrastsensitivity gage: gamma ray; image
12.1
bias for indicatingthe contrastsensitivityof a radio- formation;image processing;image quality indicator;
scopic systemusing the contrastsensitivitygagede- line-pairsper millimeter; penetratingradiation;spatial
scribedby this test method. resolution;X-ray
APPENDIX
(Nonmandatory Information)
-‘IFICATION
I
I I \ I I LI ,y+,,
I I I I I I,, , I 3.‘75
F@ --’ ---IL--tI---’@&$q50
FIG. X1.1 CONTRAST SENSITIVITY GAGE
369
STANDARD TEST METHOD FOR CLASSIFICATION
OF FILM SYSTEMS FOR INDUSTRIAL
RADIOGRAPHY
SE-1815
2.2 ANSI Standards: 3.2.4 IS0 speed S - determined by the dose K,?,
PH 2.18 Photography (Sensitometry) - Density Mea- measured in gray at a specified optical density, D, in
surements, Spectral Conditions the radiograph.
370
ARTICLE 22 - RADIOGRAPIIIC STANDAKDS SE-1815
TABLE 1
TYPICAL FILM SYSTEM CLASSIFICATION
4. Significanceand Use film systems may also perform the tests and measure-
ments outlined in this test method, provided that the
4.1 This test method provides a relative means for
required test equipment is used and the methodology
classification of film systems used for industrial radiog-
is followed strictly.
raphy. The film system consists of the film and associ-
ated processing system (the type of processing and
processing chemistry). Section 6 describes specific pa- 4.4 The publication of classes for industrial radiogra-
rameters used for this test method. In general, the phy film systems will enable specifying bodies and
classification for hard X rays, as described in Section contracting parties to agree to particular system classes,
6, can be transferred to other radiation energies and which are capable of providing known image qualities.
metallic screen types, as well as screens without films. See 1.2.
The usage of film system parameters outside the energy
ranges specified may result in changes to a film/system
performance classification.
4.1.1 The film performance is described by signal
and noise parameters. The signal is represented by 5. Sampling and Storage
gradient and the noise by granularity. 5.1 For determination of the gradient and granularity
4.1.2 A film is assigned a particular class if it of a film system, it is important that the samples
meets all four of the minimum performance parameters: evaluated yield the average results obtained by users.
for Gradient C at D = 2.0 and D = 4.0, granularity This will require evaluating several different batches
uD at D = 2.0, and gradient/noise ratio at D = 2.0. periodically, under the conditions specified in this test
method. Prior to evaluation, the samples shall he stored
4.2 This test method describes how the parameters according to the manufacturer’s recommendations for
shall be measured and demonstrates how a classification a length of time to simulate the average age at which
table can be constructed. the product is normally used. Several independent evalu-
4.3 Manufacturers of industrial radiographic iilm ations shall he made to ensure the proper calibration
systems will be the users of this test method. The of equipment and processes. The basic objective in
result is a classification table as shown by the example selecting and storing samples as described above is to
given in Table 1. This table also includes speed data ensure that the film characteristics are representative
for user information. Users of industrial radiographic of those obtained by a consumer at the time of use.
371
SE-1815 2001 SECTION V
6.2.1 Use the same X-ray spectral quality for 6.5.1 The plane of the film shall be normal to
determining both the film gradient and its root mean the central ray of the X-ray beam. Use a diaphragm
square granularity. Make the film exposures with an at the tube to limit the field of radiation to the film
g-mm (0.32.in.) copper filter at the X-ray tube and the area. The X-ray utbe target to film distance shall be
kilovoltage set such that the half value layer in copper adequate to ensure that the exposure over the useful
is 3.5 mm (0.14 in.). The kilovoltage setting will be area of each exposure step is uniform to within 3%.
approximately 220 kV. 6.5.2 To minimize the effects of backscattered
6.2.2 Determine the required kilovoltage setting radiation, use a 6.3 ;t 0.8 mm (‘/4 f 1/3ain.) thick lead
by making an exposure (or exposure rate) measurement shielding behind the cassette. The shielding lead shall
with the detector placed at a distance of at least 750 extend at least 25 mm (1 in.) beyond each edge of
mm (29.5 in.) from the tube target and an g-mm (0.32- the cassette. Alternatively, the shielding lead may be
in.) copper filter at the tube. Then make a second omitted, provided that the cassette is supported such
measurement with a total of II.5 mm (0.45 in.) of that the X-ray beam strikes no scattering material, other
copper at the tube. These filters should be made of than air, for a distance of at least 2 m (78.7 in.) behind
99.9% pure copper. the cassette.
6.2.3 Calculate the ratio of the first and second 6.5.3 Modulation of the X-ray exposure may be
readings. If this ratio is not 2, adjust the kilovoltage accomplished by changing the exposure time or tube
up or down and repeat the measurements until a ratio target to film distance. Changing the tube current is
of 2 (within 5%) is obtained. Record the machine not recommended but may be done, provided it is
setting of the kilovoltage for use with the film tests. verified by measurement (see 6.2) that the X-ray spectral
quality does not change.
6.3 Film Cassette and Screens:
6.5.4 Measure exposures with an air-ionization
6.3.1 The film cassette (holder) shall provide a chamber, or other types of X-ray detectors, having
means of ensuring good film screen contact. A vacuum linear response over the range of X-ray intensities and
cassette may be used. exposure times used for the film exposures.
6.3.2 Lead-foil screens shall be used with the front 6.5.5 During and after exposure, prior to pro-
screen thickness being 0.130 f 0.013 mm (0.005 f cessing, keep the film at a temperature of 23 * 5°C
0.05 in.) and the back screen thickness being 0.250 f (5.97 f 5°F) and a relative humidity of 50 f 20%.
0.025 mm (0.010 * 0.001 in.). Start processing of the film between 30 min and 8 h
after exposure. Process an unexposed specimen of the
NOTE-These thicknessesreflectcommerciallyavailabletolerances film sample with the X-ray-exposed specimen in order
in lead foil for use as radiographicscreens.
to determine the base plus fog density.
6.3.3 It is especially important that the exposure 6.5.6 Measure the visual diffuse transmission den-
to the film specimen for the granularity measurements sity of the processed films with a densitometer comply-
be spatially uniform. Any nonuniformities in X-ray ing with the requirements of ANSI PH 2.19 and IS0
372
AKTICLE 22 - RADIOGRAPHIC STANDARDS SE-1815
5-2 and calibrated by the method of Practice E 1079. the range from 1.80 to 2.20, may be measured, and
Use a minimum aperture of 7 mm (0.275 in.). the granularity value at a diffuse density of 2.00, above
base plus fog, shall be taken from a smooth curve
6.6 Measurement of Gradient G:
drawn through a plot of the data points. The granularity
6.6.1 Gradient G relates to a D versus log K value shall be in terms of diffuse density.
curve. In the scope of this test method, G is calculated
6.7.4.1 The microdensitometer scanner output is
from the slope of a D versus K curve at density (D
measured as projection density. Thus to obtain the
- D,), as follows:
desired diffuse density, convert the data using the slope
of the curve of diffuse density versus projection density
G=&K=LXdD at the mean density value of the granularity film
log e dK
specimen. Determine this curve using a film having a
stepped series of densities, which is prepared using the
where: same type film, exposure, and processing techniques
K = dose required for density D - D, and as used for the granularity film specimen. Measure the
D, = fog and base density.
diffuse density of each step with a microdensitometer.
6.6.2 The D versus K cnrve is approximated by The specimen film shall be scanned using identical
a polynomial of the third order. To obtain a regular microdensitometer settings. A limited range of densities
and reliable shape of this curve, make a series of can typically be measured for a given microdensitometer
exposures to obtain at least 12 uniformly distributed gain setting. The stepped series of densities shall lie
measuring points between density 1.O and 5.0 above D,. within that range. Choose the number of steps such
that the slope of the curve, at the mean density of the
6.6.3 Average the Gradient G measurements, with granularity film specimen, is determined to an accuracy
a maximum inaccuracy of *5%. of *5%.
6.1 Root Mean Square (rms) Granularity, Q:
6.7.5 Determine the granularity of the film speci-
6.7.1 Determine the rms granularity of the film men by evaluating no fewer than three samples of the
in accordance with ANSI PH 2.40, with the following specimen and determining their mean so that a maximal
exceptions: uncertainty of 10% is achieved.
6.7.2 The procedure is limited to the measurement 6.7.6 Adjust the optical system of the microdensi-
of continuous tone black-and-white industrial X-ray tometer so that both emulsions, or the one emulsion
films viewed by transmitted light. The film may have in the case of a single-coated film, are in focus at all
emulsion coated on one side or both sides of the film points in the scan.
support.
6.7.7 Scan the film specimen along three different
6.7.3 Expose the film specimen with X rays having paths within the test area. Take the median of the
the spectral quality described in 6.2. The cassette and three granularity readings as the granularity of the film
lead-foil screens shall be as specified in 6.3. Expose specimen at the mean measured density.
the film specimen in accordance with the exposure
conditions of 6.5. Exercise care to ensure that the film 6.7.8 Microdensitometer Requirements:
specimen does not contain density variations arising
from the exposing equipment (such as nonuniform beam 6.7.8.1 The intlux aperture of the microdensitom-
filters or damaged or defective lead screens). During eter shall be approximately circular in shape, with a
and after exposure, prior to processing, maintain the diameter (referred to the plane of the specimen) not
film specimen at the temperature and relative humidity less than 1.2x or more than 2x the diameter of the
conditions specified in 6.5.5. The film processing chemi- efflux aperture.
cals and procedures shall be the same as those used
for determining gradient, and they shall be described 6.7.8.2 Both the influx objective and the efflux
completely as specified in 6.4. objective shall be a high-quality microscope objective
having a numerical aperture no greater than 0.10.
6.7.4 The film specimen for granularity measnre-
merit shall have a diffuse density of 2.00 * 0.05 above 6.7.8.3 The reduction of the influx aperture by
base plus fog. As an alternative, three or more samples the influx optics and the magnification of the specimen
of the tilm specimen at different density levels, within onto the efflux aperture by the efflux optics shall lie
313
SE-IRIS 2001 SECTION V
TABLE2 TABLE3
DETERMINATIONOF IS0 SPEEDSFROM DOSE,&, LIMITINGVALUES FORGRADIENT,GRADIENT/
NEEDED FOR A FILM DENS1TY.D = 2.0.ABOVE Dn GRANULARITY RATIO.ANDGRANULARlTY
Minimum
Minimum Gradient G at Gradient/
FK%l TO IS0 Speed S” Granularity Maximum
-3.05 -2.96 loll0 ASTM Ratio, G/v* Granularity,
-2.95 -2.86 800 System D = 2.0 D = 4.0 at D = 2.0 ~0, at D = 2.0
-2.85 -2.76 640 class Above DoA Above 0, Above Do Above 0,
-2.75 -2.66 500 Special 4.5 7.5 300 0.018
-2.65 -2.56 400 I 4.1 6.8 150 0.028
-2.55 -2.46 320
II 3.8 6.4 120 0.032
-2.45 -2.36 250 III 3.5 5.0 100 0.039
-2.35 -2.26 200 W-A 3.6 5.7 135 0.027
-2.25 -2.16 I60 W-B 3.5 5.0 110 0.032
-2.15 -2.06 125 w-c <3.5 6.0 80 0.039
-2.05 -1.96 100
-1.95 -1.86 SO A 0, = density of an unexposed and processed film including base
-1.85 -1.76 64 (fag and base density).
-1.75 -1.66 50
-1.65 -1.56 40
-1.55 -1.46 32
-1.45 -1.36 25
-1.35 -1.26 20
-1.25 -1.16 16 ponding to the first zero in the spatial frequency response
-1.15 -1.06 12 of the circular aperture. Mathematical procedures that
-1.05 -0.96 10 can be shown to produce equivalent reductions in the
-0.95 -0.86 8
effects of system artifacts arc acceptable alternatives
-0.85 -0.76 6
-0.75 -0.66 5
to the use of this filter.
-0.65 -0.56 4
6.8 Measurement of IS0 Speed S - The IS0 Speed
* see IS0 7004 S is evaluated for an optical density, D = 2.0, above
fog and base, D,. Use Table 2 for determination of
the IS0 speed.
in the range from 20 to 100x. The two magnifications 7. Range of Classificationand Limiting Values
need not be equal.
7.1 There are film system classes that differ by their
6.7.8.4 The efflux (or measuring aperture) shall gradients and granularities. The limiting values are
be circular in shape. Its effective diameter referred to assigned to the film classes whose observance must be
the specimen plane shall be 100 f 2 pm. proved by the measuring methods in 6.6 and 6.7.
6.7.8.5 The scan path of the microdensitometer 7.1.1 In order to assign a film system to a system
may he linear or circular. If circular, the radius of the class, it must meet all four limiting values of the
path shall not be less than 16 mm. In either case, the gradient (at D = 2.0 and D = 4.0), the granularity
total scan length shall not be less than 100 mm (3.94 in.). (at D = 2.0), and the gradient/granularity parameter
of the system class. The classification is valid only for
6.7.8.6 The spectral response of the microdensi-
the complete film system.
tometer system shall be visual, as specified by ANSI
PH 2.18 and IS0 5-3. 7.2 Film system manufacturers will provide their
classification table, upon request, with a classification
6.7.8.7 The electronic band-pass filter, used to table that contains full data on. the four parameters
reduce the unwanted signal caused by system artifacts, according to Table 3. In addition, the following two
shall have its low-frequency boundary set so the system parameters (see Table 2 for data) will be listed with
response is 3 dB down at a temporal frequency corres-
the classification table: IS0 speed S, and dose, K,.
ponding to a spatial frequency of 0.1 cycles/mm. Its
high-frequency boundary shall be set so that the system 7.2.1 The classification table will additionally con-
response is 3 dB down at a temporal frequency corres- tain the following information on the processing system:
374
ARTICLE 22 - RADIOGRAPHJC STANL)ARDS SE.1815
manual or automatic, type of chemistry, developer classification of industrial radiographic tihn since the
immersion time, and developer temperature. results state merely whether there is conformance to
the criteria for success specified in the procedure.
7.3 For examples of a classification table, see Table 1.
9. Keywords
8. Precision and Bias
9.1 ASTM system class; film system; film system
8.1 No statement is made about either the precision classification; gradient; gradient/noise; granularity; in-
or bias of this test method for measuring a film system dustrial radiographic film; speed.
APPENDIX
(Nonmandatory Information)
TABLE X1.1
CLASSIFICATION COMPARISON OF TEST METHOD E 1815 AND GUIDE E 94-83 FOR HIGH-CONTRAST FILM
SYSTEMS
Minimum
Gradient/
Test Method Minimum GradientG at Granularity Maximum Guide Description
SE-1815 Ratio at Granularityat E 94.83
system Class D = 2.0 D = 4.0 D = 2.0 D = 2.6 Film Type Speed Contrast Graininess
376