U50 Service Manual
U50 Service Manual
Urine Analyzer
Service Manual
1 Introduction 1
2 Trouble-shooting 1
2.1 Table for error due to installation environment 1
2.2 Error and remedy 1
3 Instrument malfunction and remedy 4
3.1 Power is on with on response 4
3.2 Abnormal display 4
3.3 The results cannot be printed out 4
3.4 Reagent strip detector is abnormal 5
3.5 Abnormal time display on the screen 6
3.6 TROUBLE-01 reference light (LED1) is too strong 6
TROUBLE-04 reference light (LED1) is too weak
3.7 TROUBLE-02 red light( LED2) is too strong 7
TROUBLE-05 red light( LED2) is too weak
3.8 TROUBLE-03 green light( LED3) is too strong 7
TROUBLE-06 green light (LED3) is too weak
3.9 TROUBLE-10 motherboard problem 7
3.10 TROUBLE-12 8
1 Introduction
Uritest-50 urine analyzer is manufactured under strict supervision. This is a kind of precise analyzer,
improper operation or any damaged component may affect the results or make damage to the instrument.
If an operational or instrument problem occurs, an error number may display on the Analyzer screen. This
section of the manual lists various errors and messages, along with probable causes and recommended
remedies to quickly and easily correct the problem. If the problem persists after following the
recommended remedy, records the error number displayed and contact your local distributor or URIT
MEDICAL ELECTRONIC SALES CO., LTD.
2 Trouble-shooting
2.1 Table for error due to installation environment
Table-1
1
Uritest-50 Service Manual
Table-2
2
Uritest-50 Service Manual
Table-3
TROUBLE-07 Improper position of reagent strip Check the Routine Test chaper to
place the strip correctly
TROUBLE-08 Improper type of reagent strip Set reagent strip type correctly
TROUBLE-11 Strip holder is not clean Clean the strip holder or take the dust
out
3
Uritest-50 Service Manual
3 Instrument malfunction and Remedy
3.1 Power is on with no response (Picture 1)
a) Check the voltage of TP6 (DC 12 0.4V).
b)Check whether the power adapter, power input plug, power switch, power input connector and
power input cable are normal, and replace the defective parts.
TP6
Picture 1
4
Uritest-50 Service Manual
The chip(LB1838)
of printed driver Printer board
Printer cable 10-core cable IC (4051)
TP1
The potentiometer
of contrast
Picture 2
5
Uritest-50 Service Manual
4Pin on the detector output connector
Picture 3
a) Adjust the LED 1 potentiometer to make sure the voltage of Tp3 is within the normal range (DC
3V~4.7V)
b) If the voltage is normal, while the same error persists, contact the distributor.
6
Uritest-50 Service Manual
The green light LED 3 The red light LED 2 The reference light LED 1
adjustment potentiometer adjustment potentiometer adjustment potentiometer
TP3
TP4
TP5
Light-detection module
Picture 4
a) Adjust the LED 2 potentiometer to make sure the voltage if Tp4 is within the normal range (DC
4.6V 0.1V)
b) If the voltage is normal, while the same error persists, contact the distributor.
a) Adjust the LED 3 potentiometer to make sure the voltage if Tp5 is within the normal range (DC
3.8V 0.1V)
b) If the voltage is normal, while the same error persists, contact the distributor.
7
Uritest-50 Service Manual
RAM CPLD MCU Battery
Picture 5
White block
Picture 6
8
Uritest-50 Service Manual