67% found this document useful (3 votes)
1K views11 pages

U50 Service Manual

This document is the service manual for the Uritest-50 urine analyzer. It contains troubleshooting information to help with common errors, including possible causes and recommended remedies. The manual includes tables listing common installation environment errors, general errors and remedies, and specific trouble codes. It also provides guidance on checking voltages and replacing parts to address issues like no power, display abnormalities, printing problems, detector abnormalities, and time display or LED strength problems. The service manual aims to help technicians quickly diagnose and fix issues with the Uritest-50 urine analyzer.

Uploaded by

Georgy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
67% found this document useful (3 votes)
1K views11 pages

U50 Service Manual

This document is the service manual for the Uritest-50 urine analyzer. It contains troubleshooting information to help with common errors, including possible causes and recommended remedies. The manual includes tables listing common installation environment errors, general errors and remedies, and specific trouble codes. It also provides guidance on checking voltages and replacing parts to address issues like no power, display abnormalities, printing problems, detector abnormalities, and time display or LED strength problems. The service manual aims to help technicians quickly diagnose and fix issues with the Uritest-50 urine analyzer.

Uploaded by

Georgy
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 11

Uritest- 50

Urine Analyzer
Service Manual

URIT MEDICAL ELECTRONIC CO.,LTD


Table of Contents

1 Introduction 1
2 Trouble-shooting 1
2.1 Table for error due to installation environment 1
2.2 Error and remedy 1
3 Instrument malfunction and remedy 4
3.1 Power is on with on response 4
3.2 Abnormal display 4
3.3 The results cannot be printed out 4
3.4 Reagent strip detector is abnormal 5
3.5 Abnormal time display on the screen 6
3.6 TROUBLE-01 reference light (LED1) is too strong 6
TROUBLE-04 reference light (LED1) is too weak
3.7 TROUBLE-02 red light( LED2) is too strong 7
TROUBLE-05 red light( LED2) is too weak
3.8 TROUBLE-03 green light( LED3) is too strong 7
TROUBLE-06 green light (LED3) is too weak
3.9 TROUBLE-10 motherboard problem 7
3.10 TROUBLE-12 8
1 Introduction

Uritest-50 urine analyzer is manufactured under strict supervision. This is a kind of precise analyzer,
improper operation or any damaged component may affect the results or make damage to the instrument.
If an operational or instrument problem occurs, an error number may display on the Analyzer screen. This
section of the manual lists various errors and messages, along with probable causes and recommended
remedies to quickly and easily correct the problem. If the problem persists after following the
recommended remedy, records the error number displayed and contact your local distributor or URIT
MEDICAL ELECTRONIC SALES CO., LTD.

NOTE: Be careful of the strong voltage inside.

2 Trouble-shooting
2.1 Table for error due to installation environment

Table-1

Number Environment Relative troubles


1 Direct sunlight Incorrect result
2 T emperature exceeding the recommended range Incorrect result
3 Electromagnetism interference Incorrect result
4 Dusty, moist or caustic gas Damage to the instrument
5 Out of horizon Incorrect result

2.2 Error and remedy


Error and remedy(table-2) Troubles and remedy (table-3).

1
Uritest-50 Service Manual
Table-2

Number Error Possible Cause Remedy

1 The voltage falls short of the adapter


input voltage requirement;
2 Power supply cable connection is
1 Power is on with no twisted or abnormal; 3.1
response 3 Receptacle abnormity;
4 Switch abnormity;
5 Power adaptor output abnormity.

1 LCD board 20-core cable connection


is twisted;
2 Abnormal display 2 LCD contrast abnormity; 3.2
3 LCD module abnormity;
4 LCD drive circuit problem.

1 The printer is OFF;


2 Printer board 10-core cable or cable
3 The results can not be connection is twisted; 3.3
printed out 3 Printer abnormity;
4 Printer drive circuit problem.

4 Reagent strip detector is 1 The Detector is set to be OFF ;


abnormal 2 Reagent s trip detector abnormity; 3.4
3 Detector circuit malfunction.

1 battery loose or failed;


5 Abnormal time display 2 Crystal Oscillator 32.768K 3.5
on the screen abnormity;
3 Clock chip ISL1208 abnormity.

2
Uritest-50 Service Manual
Table-3

Trouble Number Probable Cause Remedy

TROUBLE-01 Reference light 3.6


(LED1 ) is too strong

TROUBLE-02 Red light 3.7


(LED 2) is too strong

TROUBLE-03 Green light 3.8


(LED 3) is too strong

TROUBLE-04 Reference light 3.6


(LED 1 )is too weak

TROUBLE-05 Red light 3.7


(LED 2 ) is too weak

TROUBLE-06 Green light 3.8


(LED 3) is too weak

TROUBLE-07 Improper position of reagent strip Check the Routine Test chaper to
place the strip correctly

TROUBLE-08 Improper type of reagent strip Set reagent strip type correctly

TROUBLE-09 No reagent strip Place reagent strip

The instrument is not in good


TROUBLE-10 condition (Mainboard problem) 3.9

TROUBLE-11 Strip holder is not clean Clean the strip holder or take the dust
out

Transfer system problem


TROUBLE-12 Optical-electronic sensor system 3.10
problem or circuit problem

3
Uritest-50 Service Manual
3 Instrument malfunction and Remedy
3.1 Power is on with no response (Picture 1)
a) Check the voltage of TP6 (DC 12 0.4V).
b)Check whether the power adapter, power input plug, power switch, power input connector and
power input cable are normal, and replace the defective parts.

Power input Power switch


Power input plug connector

Power input cable

TP6

Picture 1

3.2 Abnormal display (Picture 2)


a) Check the voltage of TP1(DC-5.16 0.5V) ,when the Contrast has >>>> on the screen.
b) Adjust the potentiometer of contrast
c) Replace the Ic4051 or potentiometer

4
Uritest-50 Service Manual
The chip(LB1838)
of printed driver Printer board
Printer cable 10-core cable IC (4051)

TP1

The potentiometer
of contrast

LCD module Display board 20-core cable

Picture 2

3.3 The results can not be printed out (Picture 2)


a)Check whether the printer cable is connected correctly.
b) Replace the printer.
c)Replace the chip (Lb1838).

3.4 Reagent strip detector is abnormal (Picture 3)


a)Check the voltage of 4 pin on the detector output connector
b) If reagent strip is on the holder, the voltage is more than 3.5V
c) If reagent strip is not on the holder, the voltage is less than 1.5V

5
Uritest-50 Service Manual
4Pin on the detector output connector

Reagent strip detector

Picture 3

3.5 Abnormal time display on the screen (Picture 5)


a) Replace the battery.
b)Check wether the crystal oscillator (32.768K) is normal and replace the defective one..
c) Replace the clock chip(ISL1208).

3.6 TROUBLE-01(Reference light (LED 1) is too strong) ¡¢TROUBLE-04(Reference light


(LED 1) is too weak) (Picture 4)

a) Adjust the LED 1 potentiometer to make sure the voltage of Tp3 is within the normal range (DC
3V~4.7V)
b) If the voltage is normal, while the same error persists, contact the distributor.

6
Uritest-50 Service Manual
The green light LED 3 The red light LED 2 The reference light LED 1
adjustment potentiometer adjustment potentiometer adjustment potentiometer

TP3

TP4

TP5

Light-detection module

Picture 4

3.7 TROUBLE-02(Red light (LED 2) is too strong) ¡¢TROUBLE-05(Red light (LED2) is


too weak) (Picture 4)

a) Adjust the LED 2 potentiometer to make sure the voltage if Tp4 is within the normal range (DC
4.6V 0.1V)
b) If the voltage is normal, while the same error persists, contact the distributor.

3.8 TROUBLE-03(Green light (LED 3) is too strong) ¡¢TROUBLE-06(Green light (LED


3)is too weak) (Picture 4)

a) Adjust the LED 3 potentiometer to make sure the voltage if Tp5 is within the normal range (DC
3.8V 0.1V)
b) If the voltage is normal, while the same error persists, contact the distributor.

3.9 TROUBLE-10(Motherboard problem) (picture 5)


a)Check wether the IC, MCU, RAM and CPLD are normal.
b)Replace the IC, MCU, RAM and CPLD
c) If the same error persists, contact the distributor.
.

7
Uritest-50 Service Manual
RAM CPLD MCU Battery

Crystal Oscillator 32.768K

Clock chip ISL1208

Picture 5

3.10 TROUBLE-12 (picture 6)


a) Check whether the white block of the strip holder is clean.
b) When the white block of the strip holder is under the reflect electro-optical sensor the reagent strip
holder returns to the initial position , check whether the voltage of 3Pin of the detector output
connector is more than 1.5V.
c) If the same error persists, contact the distributor.

White block

3Pin of the detector


output connector

Reflected optical-electronic sensor (U1)

Picture 6

8
Uritest-50 Service Manual
URIT MEDICAL ELECTRONIC CO.,LTD
Add: Jiuhua Road , Guilin , P.R. China
Tel : +86GUILIN
773URITEST
2832799
MEDICAL Fax: +86CO.
ELECTRONIC 773
LTD2824559

URIT MEDICAL ELECTRONIC SALES CO.,LTD


Add: ZhongHUA Road , Guilin , P.R. China
Tel : +86 773 2822205 2836022 2823271
Fax: +86 773 2824559
Http://www.uritest.com E-mail:[email protected]

You might also like