Dig Exam
Dig Exam
Contents
Contents ...........................................................................................................................................................1
2 Safety ..........................................................................................................................................................11
2.1 Explanation of safety symbols ............................................................................................................... 11
2.1.1 Hazard warning symbol .................................................................................................................. 11
2.1.2 Explanation of different levels of hazard ........................................................................................ 11
2.1.3 Structure .........................................................................................................................................11
2.2 Intended purpose ................................................................................................................................... 12
2.2.1 General ..........................................................................................................................................12
2.2.2 Product-specific .............................................................................................................................. 12
2.3 Safety symbols .......................................................................................................................................14
2.4 Prerequisite for proper technical operation ............................................................................................ 16
2.4.1 Data backup ................................................................................................................................... 16
2.4.2 Sensor handling ............................................................................................................................. 17
2.4.3 Standards for IT-components ......................................................................................................... 18
2.4.4 Standard conformity ....................................................................................................................... 18
1/60
Instructions for use KaVo Dig eXam 3520, 3530
Contents
4 Operation .....................................................................................................................................................40
4.1 Configuration of the Dig eXam driver software ...................................................................................... 40
4.1.1 Selection Information ..................................................................................................................... 40
4.1.2 Selection configuration ................................................................................................................... 41
4.1.3 Selection parameters ..................................................................................................................... 43
4.2 Taking a digital X-ray ............................................................................................................................. 44
4.2.1 Overview ........................................................................................................................................44
4.2.2 Activate programme and select patient .......................................................................................... 44
4.2.3 Prepare and position sensor .......................................................................................................... 45
4.2.4 X-ray ...............................................................................................................................................47
4.2.5 Conclude X-ray procedure ............................................................................................................. 49
4.2.6 Create a diagnosis based on the radiograph ................................................................................. 49
8 Troubleshooting ...........................................................................................................................................58
8.1 Symptoms, causes and solution ............................................................................................................ 59
2/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.1 Intended users
1 User notes
This document is intended for use by dentists and other practice or clinic employees
who are conversant with the taking, processing and archiving of computer-based
intraoral X-rays on the Microsoft® Windows operating system .
Prior to using the equipment, it assumed that users have been instructed in the use
of KaVo Dig eXam-Sensors in accordance with the stipulations of the country in
which they are practising, on the safe and proper use of X-ray radiation for the taking
of intraoral dental X-rays.
3/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.2 User guidelines
Requirement
Read these instructions before the initial startup to prevent misuse and damage.
Note
Microsoft and Windows are registered trademarks of Microsoft Corporation and its
registered subsidiaries.
VixWin PRO is a registered trademark of Gendex Dental Systems.
XCP-DS is a registered trademark of DENTSPLY RINN.
KaVo Dig eXam is a protected trademark of KaVo.
The utilisation of company logos or trademarks not named here does not indicate
that they are subject to unrestricted use. Every company logo or trademark men‐
tioned here is the sole property of the company mentioned. Brands, names and
prices are given for information purposes only.
1.2.2 Abbreviations
Abbrevia‐ Meaning
tion
ALARA "As low as reasonably achievable": The guiding principle behind radiation protection during the
taking of X-rays, meaning that radiation doses for both workers and the public are typically kept
lower than their regulatory limits.
EMC Electromagnetic compatibility
UI User instructions
Hub A central point: a distribution device that makes available a number of USB interfaces.
IEC International Electrotechnical Commission
AI Assembly instructions
SQL "Structured Query Language": Standardised computer language for querying relational databases
SFC Safety checks
SW Software
STI Service Technician's instructions
USB Universal Serial Bus. Standardised means of connecting peripherals to personal computers.
4/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.2 User guidelines
Action request
5/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.3 Service
1.3 Service
Service hotline:
+ 49 (0) 7351 56-2900
[email protected]
Please indicate the version number of the program.
Additional information can be obtained at: www.kavo.com
KaVo Dental
KaVo Technologies / Gendex Imaging
019 West Oakton Street
Des Plaines, IL 60018-1884
USA
+1 (847) 640-48 00
6/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.4 Implied warranty
The KaVo end-user customer guarantee for the product named in the completion
certificate guarantees that the product functions correctly and that there are no faults
in the material or workmanship for a duration of 12 months following the purchase
date, according to the following conditions:
Following a reasonable complaint relating to defects or short delivery, KaVo will
provide a replacement or perform repairs. KaVo reserves the right to perform re‐
pairs.
Claims of any other nature, damages in particular, are excluded. In case of default
and gross negligence or intent, the latter only applies if there are no compelling legal
provisions opposing it.
KaVo shall not be liable for defects and their consequences, which occur as a result
of normal normal wear and tear, or of improper cleaning and maintenance, non-
observance of the operating, maintenance or connection regulations; calcination or
corrosion; a contaminated air or water supply; or chemical or electrical effects, which
are non-standard or not permitted according to company regulations.
As a general rule, this guarantee does not apply to lamps, glassware, rubber parts
or the colour durability of synthetic materials.
KaVo shall not be liable for defects or their consequences if they are likely to be a
direct result of actions or modifications by a customer or third party.
Any claims arising from this guarantee can only be lodged if the completion certifi‐
cate (carbon copy) has been sent in to KaVo and the operator/user is able to produce
the original.
7/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.5 Limitation of implied warranty
KaVo shall assume responsibility for the safety, reliability and performance of all
KaVo system components provided that:
▪ Assembly, upgrades, new settings, modifications or repairs are all carried out
by Service Technicians, who have been trained by KaVo, or by KaVo authorised
dealer personnel
▪ The IT components provided by the operator correspond to the technical hard‐
ware and software requirements included in these user instructions.
▪ The system has been installed and operated in accordance with the User in‐
structions and Assembly instructions.
▪ All servicing is performed in full compliance with DIN VDE-075-1.
▪ In Germany, the location in whichKaVo Dig eXam is installed, must comply with
the regulationsDIN VDE 0100-710: 2002-11 and VDE 0100-560:1995 in its de‐
sign. In other countries, the corresponding national regulations must be respec‐
ted.
▪ The VixWin PRO software has been installed in accordance with the VixWin Pro
user instructions
8/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.6 Transportation and storage
Outside of Germany
Note
KaVo is not liable for damage arising from transportation.
Immediately inspect the delivery after receipt!
If external damage to the packaging is visible upon delivery, follow the procedure
below:
1. The recipient must record the loss or damage in the notice of delivery. The re‐
cipient and employee of the transportation firm must sign the notice of delivery.
The recipient can only assert damages against the transportation company ba‐
sed on these records.
2. Leave the product and packaging unchanged.
3. Do not use the product.
Note
If the recipient does not follow one of the above instructions, the damage will be
held to have occurred after the delivery (according to . CMR law , section 5, Art.
30).
Note
Keep the packaging for returning the product for service or repairs .
The symbols printed on the outside are for transportation and storage, and have the
following meaning:
Keep dry
9/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.6 Transportation and storage
Temperature range
Humidity
Air pressure
10/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.1 Explanation of safety symbols
2 Safety
Warning symbol
Safety instructions with three hazard levels are used in this document for avoiding
personal and property damage.
CAUTION
indicates a hazardous situation that can lead to property damage or minor to mo‐
derate injury.
CAUTION
WARNING
indicates a hazardous situation that can lead to serious injury or death.
WARNING
DANGER
indicates a maximum hazardous situation that can directly cause serious injury or
death.
DANGER
2.1.3 Structure
11/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.2 Intended purpose
2.2.1 General
This KaVo product is intended only for use in the field of dentistry. It is impermissible
to use the product for a purpose for which it was not intended.
"Proper use" includes following all the instructions for use and ensuring that all in‐
spections and service tasks are performed.
Apply and meet the overarching guidelines and/or national laws, national regulati‐
ons and the rules of technology for medical devices applicable for startup and use
of the KaVo product for the intended purpose.
The user must ensure that that the device works properly and is in a satisfactory
condition before each use.
Note
The product must be cleaned and serviced according to instructions if it is not to
be used for a long period.
2.2.2 Product-specific
Dig eXamtogether with VixWin PRO is intended for processing dental, intraoral X-
rays.
The radiographs are displayed and archived using the software VixWin PRO.
Prior to using the equipment, users of the KaVo Dig eXam-Sensors must have un‐
dergone full instruction, in accordance with the stipulations of the country in which
they are practising, on the safe and proper use of X-ray radiation for the taking of
intraoral dental X-rays.
KaVo Dig eXam has electronic components in the Trigger and Sensor that are to
be disposed of in accordance with the relevant country specific regulations.
12/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.2 Intended purpose
Disposal of electronics
Note
According to the EC Directive 2002/96 concerning electrical and electronic used
devices, this product is subject to the cited directive and must be disposed accor‐
dingly within Europe.
Before disassembling and disposing of the product, it must be completely proces‐
sed (disinfected, sterilised) according to the section "Preparation methods".
Additional information can be obtained from KaVo (www.kavo.com) or your dental
supplier.
13/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.3 Safety symbols
Ionising radiation
Overexposure to X-ray radiation.
▶ The handler can and must take measures, in accordance with country-specific
CAUTION
legal stipulations, to minimise their personal dose of radiation!
14/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.3 Safety symbols
Note
When positioning the screen, keep it out of direct sunlight and prevent light reflec‐
ting off it to enable the medical data to be read properly.
15/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation
Responsibility is assumed for the safety, reliability and performance of the compo‐
nent supplied by KaVo when:
▪ Installation, expansions, adjustments, changes or repairs is done by technicians
trained by KaVo or third parties authorised by KaVo for the Dig eXam, or by the
personnel of authorised distributors.
▪ The unit is operated according to the instructions for use and installation.
▪ The IT components supplied by the operator meet the technical requirements in
these instruction for use for hardware and software, and they are installed and
set up according to the descriptions of these components.
▪ KaVo Dig eXam is an electronic medical device, and is subject to the special
precautions necessary for EMC. KaVo Dig eXam may only be operated when it
has been installed according to the EMC instructions in the instructions for use
and installation.
▪ When setting up the unit, follow all the requirements of VDE 0751-1, "Repeated
tests and test before startup of electronic medical devices and systems - general
guidelines".
▪ In Germany ,the room in which the Dig eXam is installed must be designed ac‐
cording to the specifications DIN VDE 0100-710:2002-11 ad VDE
0100-560:1995. In other countries, follow the corresponding national specifica‐
tions.
Note
The installation and fitting of IT components for Dig eXam are the responsibility of
the practice or clinic.
Unauthorised accessories
Only original replacement parts and accessories should be used, to ensure that
there is no increase in the emissions and/or a reduction in the stability in terms of
CAUTION
the electromagnetic compatibility.
▶ Only use authorised or recommended spare parts and accessories.
16/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation
Data loss
Establish a data backup strategy.
▶ VixWin PROuses the file system supplied with the Microsoft operating system
CAUTION
for saving image data. Correspondingly, this data must also be included within
the backup strategy of your practice or clinic.
Note
The image data produced using KaVo Dig eXam replaces the conventional intraoral
radiographs taken on X-ray film and, as such, is subject to the same country-spe‐
cific requirements governing this; such as the mandatory length of time for storing
X-rays etc.
These country-specific stipulations and requirements may also include the recor‐
ding of dosage levels and information about the patient's indication.
The user should regularly make back-up copies If a hard disk in the PC fails, this
protects from data loss and increases the safety of the system data. Such a copy is
also termed a backup. It should not be confused with the export of data in which the
data remain in the PC.
Store backup copies in a safe place.
Note
The computer expert or system administrator at your practice will have the most
up-to-date technical information as regards keeping and running backup copies.
Never:
▪ Sterilise the sensors or subject them to high-pressure sterilisation (autoclave).
▪ Drop the sensors.
▪ Twist or tread on the sensor wires.
▪ Let patients bite on the sensor or its wire.
▪ Remove the sensors when in use or while the X-ray is being taken.
▪ Pull or bend the wire..
▪ Immerse the KaVo trigger in the cleaning bath.
17/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation
Electrostatic discharge
▪ When the sensors are not in use, it is recommended that they be stored in the
storage containers provided.
▪ Never touch the USB connection contact points!
The KaVo Dig eXam Sensors and the KaVo Dig eXam Interface comply with the
following standards
▪ IEC 60601-1:1988 +A1:1991 + A2:1995 [For the USA UL 60601-1:2003. For
Canada CAN/CSA 22.2 N0.: 601-1]
▪ IEC 60601-1-2:2001
Note
The delivered Vix Win Pro software also meets this standard.
KaVo Dig eXam and the IT components together make up a medical, electrical sys‐
tem.
The standards applicable to these IT components depend on their specific location
within the practice or clinic. In general, these IT components must comply with IEC
60950-1:2001.
The system that is formed by these IT components andKaVo Dig eXam when in‐
stalled in accordance with the assembly instructions for KaVo Dig eXam and meets
the requirements stipulated in this section, conforms to standard
- IEC 60601-1-1:2000
The room where KaVo Dig eXam Sensors are operated, must be designed in ac‐
cordance with regulationDIN VDE 0100-710:2002-11. In other countries, the corre‐
sponding national regulations must be respected.
18/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation
patient environment
Medical system
Monitors
Note
The dental radiographs created byKaVo Dig eXam are examined and diagnosed
using the dental practice's own image representation device (monitor). There are
country-specific regulations in force governing (regular) check-ups of the greyscale
display for these image representation devices.
In Germany, at least one monitor at the dental practice must be tested for constancy
on a regular basis in accordance with DIN-V 6868-57:2001.
19/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.1 Entire system
3 Product description
Dig eXam has two additional software programs for customers in Germany:
1. A tool for supporting the required acceptance and constancy test in Germany.
2. A program for listing the existing images (x-ray folder)
< 5m
3
< 5m
HUB
HUB
1
HUB
HUB
20/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels
3.2 Where to affix the nameplates, power rating plates and caution
labels
3.2.1 Labeling
SN :
1
X.XXX.XXXX
am
XXXXX
X
eX
ig
D
4
o
aV
K
3
2
A
④ KaVo Dig eXam sensor rating plate (C) USB symbol on connector
21/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels
22/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels
23/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors
3.3 Sensors
The receiver side is marked with a dot that is also shown on the digital X-ray.
24/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors
KaVo Trigger
The KaVo trigger connects the X-ray image detector to the computer system via a
USB 2.0 data path. This connection takes place in the X-ray device
KaVo In eXamsimply by inserting the interface in the X-ray head and in the
KaVo Dig eXamfitted in the machine.
When using the Dig eXam sensors with other X-ray machines it is possible to con‐
nect the KaVo Trigger to the so-called "active sensor holder" or alternatively to
connect it directly to the computer using the accessory cable supplied. In this case
a Dig eXAM sensor of one size can be exchanged with a sensor of another size by
changing the trigger.
Note
The accessory line is "Pull/Push“-Connection. It must not be twisted when pulled
out and plugged in.
when changing the Dig eXam Sensor from machine to machine or for another sensor
size, the detector can be laid down in the sensor holder fitted on the KaVo Trigger.
This sensor holder can be removed for cleaning the KaVo Trigger..
When the sensor holder is opened and the detector is removed, a button is acces‐
sible.
This button directly activates the system to take x-rays when the Dig eXam sensor
is not connected to the KaVo In eXam or is being operated in mixed-mode.
The sensor is connected to the machine via a USB 2.0 connection. The components
are physically connected to each other via simple plug-fit connections. Anchoring
the KaVo trigger into the device on the X-ray generator establishes the physical
connection between the machine and KaVo trigger or sensors. In accordance with
USB standards, the USB data connection may comprise up to four 5m sections.
Hubs that are compliant with USB 2.0 must be used to connect up the individual
sections
25/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors
Note
When using 5 metre long USB-cables the the USB-connection must plugged di‐
rectly into the so-called motherboard. Insertion in the USB ports that are within the
PC on further cables, is not permitted.
Note
Passive USB extension cords may not be used. The distances between two HUBs
may only be bridged with a cord 5 m or less.
KaVo Dig eXam is can be transported from one location to another, in order to
enable operation of the sensors in different treatment rooms.
KaVo Dig eXam is thus excellently suited for a multi-user installation in dental clinics
or group practices.
The handy size provides optimum flexibility and offers you independence when
using the sensors on a day-to-day basis.
The positions in which the sensor can be placed in the mouth are the same as those
for classic radiology when using X-ray film.
Note
The use of the retention system is shown on the poster that comes with the retention
system.
See also:
7 Accessories and compatibility, Page 53
26/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.4 Integration in KaVo In eXam
KaVo Dig eXam then enables the sensor to be removed without switching off the
machine, simply by pulling gently and placing in another room.
A Dig eXam sensor can be stored in the provided RWT holder on the In eXam.
The sensor must have hygienic protection and be inserted in the desired variation
of the RWT holding set.
27/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.4 Integration in KaVo In eXam
The prepared RWT holder can be inserted with the guide rod into the provide clip
on the In eXam.
Note
KaVo recommends that this option only be used for temporary storage to prevent
damage from bumps, etc.
Technical background:
KaVo X-ray machines are equipped with an internal USB line that leads directly to
theKaVo Dig eXaminterface in the X-ray machine via the scissor arm.
28/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.5 Sensor holder
With the optionally obtainable Dig eXam interface Mat. no. 1.003.2363 the distance
to the PC or hub can be increased to a max. 5 m line length from the USB 2.0, for
example to place the IT components outside of the patient's environment.
The interface in this case is mounted covered by the plastic part of the sensor holder,
and the accessory line to the KaVo trigger is tightly connected to the arising active
sensor holder.
29/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.5 Sensor holder
In addition, it is possible in this case to run the line under the wall surface from the
sensor cover to the IT component.
30/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.6 Status icon and sensor status on the PC
Two icons are visible in the Windows system area (cut-out in the so-called taskbar
next to the system clock)
The blue -Icon displays that the software interfaces between Dig eXam-Hard‐
ware and VixWin-Software is loaded.
The colour of the -Icon shows the availability of the Dig eXam-Systems connected.
Note
Other icons can be displayed dependent on the set up of the X-ray system. These
icons refer to further KaVo or Gendex equipment and are described in the relevant
documentation.
Note
This area of the Windows desktop in which the status icons are show is a part of
the so-called windows taskbar. Windows should be configured so that the status
icons are always visible when taking an X-ray. The configuration of the taskbar is
described in Windows Help.
The following is a list of all the possible statuses and their icons
31/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.6 Status icon and sensor status on the PC
When the icon is green, the electronics, sensor and connections are operating cor‐
rectly.
The sensor is ready to detect the x-rays.
When the icon is YELLOW, the electronics, sensor and connections are operating
correctly.
The sensor is waiting for the exposure to be activated.
Note
If you are using a system that is not connected to KaVo In eXam, the KaVo trigger
must be used to activate the process and must be pressed before the X-ray is taken.
If you are using a system that is connected to KaVo In eXam, activation occurs
automatically once you start taking the X-ray.
The icon in RED status: Dig eXam is not connected to the PC and/or not ready to
take an X-ray.
32/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
Note
Further information in theuser instructions VixWin PRO.
33/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
3.7.3 Ambient conditions for KaVo Dig eXam sensors and interface
[compliant with IEC601-1]
Operation
▪ Class I
– in accordance with EC Directive 93/42/EEC of June 14, 1993
▪ Application part type BF (IEC 601-1)
– Not protected from the penetration of water [IP20].
– Not suitable for sterilisation.
– Suitable for wipe or surface disinfection using the cleansers described in the
instructions for use.
– Not suitable for operation in environments with combustible mixtures.
34/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
Note
In Europe, the manufacturer's CE marking on the power supply unit confirms that
relevant power supply standards have been complied with.
Note
This recommended dosage is based on thresholds which are applicable for the
Federal Republic of Germany and may need to be adjusted in accordance with
relevant country-specific regulations and/or individual diagnostic circumstances.
35/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
Stability tests IEC 60601 - Test level Compliance level Electromagnetic environment - gui‐
de
Discharge of static +/- 6 kV Contact di‐ +/- 6 kV Contact di‐ Floors should be made of wood or
electricity in accor‐ scharge scharge concrete or be covered with cera‐
dance with IEC +/- 8 kV Air discharge +/- 8 kV Air discharge mic tiles If the floor is covered with
61000-4-2 a synthetic material, the relative air
humidity must be at least 30 %.
Rapid transient electri‐ +/- 2 kV +/- 2 kV for +/- 2 kV for mains ca‐ The quality of the supply voltage
cal disturbances / burst mains cables +/- 1 kV bles should be equivalent to that of a ty‐
in accordance with IEC for input and output ca‐ +/- 1 kV for input and pical business or hospital environ‐
61000-4-4 bles output cables ment.
Surges in accordance +/- 1 kV Normal mode +/- 1 kV Common mode The quality of the supply voltage
with IEC 61000-4-5 voltage voltage should be equivalent to that of a ty‐
+/- 2 kV Common mode +/- 2 kV Common mode pical business or hospital environ‐
voltage voltage ment.
36/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
Stability tests IEC 60601 - Test level Compliance level Electromagnetic environment - gui‐
de
Voltage drops, shortli‐ < 5 % UT (> 95 % drop < 5 % UT (> 95 % drop The quality of the supply voltage
ved interruptions und in UT) for 1/2 period) in UT) for 1/2 period) should be equivalent to that of a ty‐
fluctuations in the sup‐ 40 % UT (60 % drop in 40 % UT (60 % drop in pical business or hospital environ‐
ply voltage in accor‐ UT) for 5 periods, UT) for 5 periods, ment. When the user of Dig eXam
dance with IEC demands continuing function of the
70 % UT (30 % drop in 70 % UT (30 % drop in
61000-4-11 device even when the power supply
UT) for 25 periods, UT) for 25 periods, is interrupted, it is recommended
< 5 % UT (> 95 % drop < 5 % UT (> 95 % drop that Dig eXam be powered via an
in UT) for 1/2 periods in UT) for 1/2 period) uninterruptible power supply or a
battery.
Magnetic field at the 3 A/m 3 A/m Magnetic fields at the mains fre‐
supply frequency quency should correspond to the
(50/60 Hz) in accor‐ typical values found in business
dance with IEC and hospital environments.
61000-4-8
Dig eXam is intended for use in the electromagnetic environment stipulated below.
The customer or user of In eXam should ensure that it is operated in an environment
of this type.
Stability tests IEC 60601 - test level Compliance level Electromagnetic environment - gui‐
de
Directed HF disturban‐ 3 Veff 3V Portable and mobile radio equip‐
ces in accordance with 150 kHz to 80 MHz 3 V/m ment should not be used near to In
IEC 61000-4-6 3 V/m eXam or its cables, as the recom‐
Directed HF disturban‐ 80 MHz to 2.5 GHz mended safe distance has been
ces in accordance with calculated in accordance with the
IEC 61000-4-3 equation applicable for the trans‐
mitting frequency.
Recommended safe distance :
d = 1,2 * P up to 800 MHz
d = 2,3 * P for 800 MHz to 2.5 GHz
where PA is the power rating of the
transmitter in watts (WA) according
to information from the manufactu‐
rer of the transmitter, and d is the
recommended safe distance in me‐
tes (m).
The field intensity of stationary ra‐
dio transmitters should, at all fre‐
quencies according to local investi‐
gations,a be lower than the
compliance levelB.
Near to devices bearing the symbol
, interference is possible.
Comment:
The higher frequency range is applicable at 80 MHz und 800 MHz.
37/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data
a
The field intensity of stationary transmitters such as the base stations of radio
telephones and mobile land radio devices, amateur radio stations, AM and FM
broadcasting services or television transmitters cannot, in theory, be determined
precisely in advance. To determine the electromagnetic environment in respect of
stationary transmitters, a study of the location should be considered.
If the measured field intensity at the location in which Dig eXam is being used ex‐
ceeds the compliance level above, care should be taken with Dig eXam to ensure
that it is functioning in accordance with regulations. If unusual performance features
are noted, additional measures may be required, such as changing the alignment
or moving In eXam somewhere else.
b
The field intensity should be less than 3 V/m over the frequency range 150 kHz to
80 MHz.
The table shows the necessary safe distance, as a function of the transmission
frequency, in m:
Rated power of the trans‐ 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
mitter in W D=1.2 * P D=1.2 * P D= 2.3 * P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters for which the rated power is not shown in the table above, the dis‐
tance can be determined using the equation belonging to the relevant column, where
P is the rated power of the transmitter in watts (W) according to information from
the manufacturer of the transmitter.
Note
These guidelines may not be applicable in all situations The propagation of elect‐
romagnetic waves is affected by absorption and reflection due to buildings, objects
and people.
38/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.8 Dig eXam inspection service (only available in Germany)
The function and calibration of connected devices, such as the monitor and X-ray
generator can be inspected and/or reconfigured via a separate service application
(KaVo Dig eXam inspection service).
39/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software
4 Operation
The configuration dialogue covers three areas that are activated via the selection
tab:
▪ Information ("Information“)
▪ Configuration ("Configure“)
▪ Parameters ("Treatments“)
This dialog shows the software version of the interface between Dig eXam and Vix‐
Win in the top part.
The status of the connected sensors is shown in the middle area. When the Dig
eXam interface is activated in the Configure dialog, the status of the interface is also
shown. By clicking the Update button, this information can be reentered by the
Windows operating system.
When the Dig eXam Interface is not used, click the button "Prepare Acquisition" in
the bottom area (instead of using the button on the KaVo trigger) or in mixed mode
to prepare the device for taking an x-ray.
Note
The button is only displayed in the "Use In eXam" mode when "Mixed mode" is also
active.
40/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software
The upper part of the dialogue makes it possible to set the X-ray parameters that
are transferred automatically to VixWin.
The system events (e.g. "Sensor covered", "In eXam switched off") to be displayed
by the operating system in their own windows can be set in the bottom area of the
dialogue window.
Note
These data are used by the separate X-ray folder support program.
Without In eXam
By selecting the mode "Custom“ means that the user can specify standard values
for tube voltage (kV), the tube current (mA) and the exposure time (ms). These
parameters are added to every X-ray produced using Dig eXam and can be dis‐
played later in the VixWin-Software and if necessary adjusted to the parameters
selected for the individual X-ray image.
Using In eXam
When using In eXam the exposure time of the current X-ray image is automatically
recognised by the system and added to the X-ray image taken using Dig eXam. The
possible standard values can be preset for the values for the tube voltage (kV) and
the tube current (mA), in order to add these to the X-ray image as well. All three
parameters can be displayed in VixWin later and if necessary adjusted to the para‐
meters selected for the individual X-ray image.
41/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software
Requirement
Each time before a sensor image is taken with Dig eXam, the system must assume
a ready state by pressing the trigger button or the button "Prepare for exposure" to
obtain a sensor image.
42/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software
This dialogue enables the setting of the details that are used for displaying the X-
ray image taken.
Note
It is recommended that the parameters are not altered without consulting the tech‐
nician and or KaVo Customer Services.
43/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
4.2.1 Overview
Taking an X-ray using KaVo Dig eXam together with VixWin PRO takes place in a
few steps that are explained in detail in the following:
▪ Start the program or toggle to it if already open
▪ Select patient
▪ Prepare and position sensor
▪ Take X-ray
▪ Form diagnosis from image
▪ Accept X-ray
44/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
or
▶ Activate the program from the start menu
or
▶ Call up VixWin Pro via the interface administration software used in the practice /
clinic.
▶ Select patient.
Note
It is possible, as an alternative to the procedure described here, to take the X-ray
first and then after taking the X-ray to allocate the X-ray to the patient.
Hygiene is a key issue during the preparation and taking of X-rays and must be
considered alongside the issue of radiation protection.
Hygiene
Risk of infection
Contaminated sensors can transmit infections.
▶ Your practice should have a set of hygiene procedures which must be adhered
CAUTION
to.
▶ Disinfect sensors after contamination.
▶ Disinfect the used parts of the right-angle holder ( XCP-DS) after contamination.
▶ Always use hygienic covers Mat. no. 1.003.2690.
45/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
The XCP DS holding system is supplied with an operating instruction poster. For
further explanations on X-raying techniques, please see related textbooks.
See also:
Pasler, Friedrich A.: Zahnärztliche Radiologie (Dental Radiology), Thieme, 4th edi‐
tion 2003, ISBN 3136046048
Whaites, Eric: Essentials of Dental Radiography and Radiology (Dental Series),
Churchill Livingstone 1996, ISBN 0443053499
TheKaVo Dig eXamdetector is positioned at the X-ray tube head during use of
theKaVo In eXam so that it is always in close proximity to the patient's head.
If you are usingKaVo Dig eXam with X-ray sources from other manufacturers, the
patient should either be given the KaVo trigger to hold or have it placed on their
chest.
46/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
4.2.4 X-ray
▶ Position the X-ray source in accordance with the X-ray machine's operating in‐
structions for the type of X-ray you are about to take.
Requirement
When using third-party x-ray equipment or when the mixed mode setting is activa‐
ted to support the alternating use of other image receptors, activate the status
"Ready to take x-rays".
KaVo Trigger
or
47/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
▶ Select button "Repair Acquisition“ in the information area of the Dig eXam-driver
software.
Note
This step is not necessary for KaVo In eXam with activatedKaVo Dig eXam Inter‐
face. KaVo Dig eXamis always ready to take an X-ray if the sensor is inserted on
account of its permanent connection to the interface.
Visual inspection
Visual inspection
Before you take the X-ray, have a quick look to make sure that
▪ The green icon is showing in the Windows System area
▪ The patient has not moved at all
▪ When using Dig eXam without In eXam-connection the ready dialogue is visible.
48/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray
▶ Take X-ray.
49/60
Instructions for use KaVo Dig eXam 3520, 3530
5 Maintenance/safety inspections | 5.1 Checks
5 Maintenance/safety inspections
5.1 Checks
50/60
Instructions for use KaVo Dig eXam 3520, 3530
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning and disinfecting
Note
KaVo Dig eXam does not include any parts requiring sterilisation as part of their
intended use.
The sensor holder can be removed from the KaVo trigger for cleaning and disin‐
fection.
Note
The detector including the connecting line should be provided with a hygiene sheet
to prevent direct contamination by the detector during use.
The sensors used within intraoral dentistry are highly sophisticated electronic devi‐
ces. And as such should be handled with extreme care.
▪ Before you begin cleaning and disinfecting, detach the sensors from the USB
connection.
▪ Do not exert pressure on the sensors during the cleaning and disinfection pro‐
cedure; in particular, do not pull or bend their wires.
▪ Do not entangle the KaVo trigger and USB line when connecting or disconnec‐
ting.
▪ Do not allow moisture to penetrate the KaVo trigger and USB line plug-fit con‐
nections and the line's USB connector.
6.1.1 Cleaning
All external surfaces of the KaVo Dig eXam can be wiped with a soft cloth impreg‐
nated with a mild cleaning fluid.
The sensor holders and separate parts of the right-angle holder set can also be
machine-cleaned.
6.1.2 Disinfection
All external surfaces of the KaVo Dig eXam can be wiped with a soft cloth impreg‐
nated with a mild cleaning fluid.
51/60
Instructions for use KaVo Dig eXam 3520, 3530
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning and disinfecting
▪ MIKROZID Pump-Spray
Manufacturer: Schülke & Mayr
▪ Dürr System-Hygiene FD 322 Sprühdesinfektion
Manufacturer: DÜRR Dental GmbH
▪ Incidin liquid
Manufacturer: Ecolab
▪ USA: Asepticare Aerosol Spray
Manufacturer: Ecolab
Prohibited disinfectants:
▪ Phenol-based glutaraldehyde
Heat disinfection
When circumstances dictate, the sensor holder and separate parts of the right-angle
holder set may be disinfected at 135°C in an autoclave in accordance with the ma‐
nufacturer's instructions. The plastic parts must not be allowed to come into contact
with the autoclave walls.
Thermodisinfection/Miele
When circumstances dictate, the sensor holder and separate parts of the right-angle
holder set may be heat disinfected (up to a maximum of 50 times) in a thermodi‐
sinfector system, such as Miele's G 7781, in accordance with the manufacturer's
instructions.
52/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family
KaVo Dig eXam belongs to the KaVo eXam intraoral X-ray system. The X-ray ma‐
chine KaVo In eXam is available under no. Mat. no. 1.002.4800 .
53/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family
54/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family
55/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family
See also: the spare part sheet for the separate parts of the holding system.
56/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.2 Compatibility
7.2 Compatibility
KaVo Dig eXam is compatible with all X-ray machines whose intended purpose is
to take dental intraoral radiographs, provided that these machines do not have do‐
sage level settings which exceed film sensitivity class D.
7.2.2 IT components
▪ Image viewers/monitors
– The computer monitors must be suitable for the examination of radiographs.
Observe the relevant federal regulations. The provided test images can also
be used for monitors that are only for showing radiographs.
▪ Printing out radiographs
– The achievable quality of printouts greatly depends on the printer technology.
Printouts are generally unsuitable for diagnostics. X-ray images should ge‐
nerally be diagnosed on suitable monitors.
See also:
3.7.2 IT component requirements , Page 33
2.4.4 Standard conformity, Page 18
57/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 7.2 Compatibility
8 Troubleshooting
The following table is a troubleshooting guide to help solve typical problems without
requiring any special technical training.
The only precondition is that the KaVo Dig eXam has been installed correctly.
If you are unable to remedy your problem by following the solution methods pre‐
sented here, please contact your responsible KaVo partner.
58/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 8.1 Symptoms, causes and solution
59/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 8.1 Symptoms, causes and solution
60/60
1.003.4073 · KF · 20050801 - 03 · en