0% found this document useful (0 votes)
223 views64 pages

Dig Exam

Uploaded by

Mad Michu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
223 views64 pages

Dig Exam

Uploaded by

Mad Michu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 64

Instructions for use

KaVo Dig eXam 3520, 3530

Always be on the safe side.


Sales: Manufacturer:
KaVo Dental GmbH Kaltenbach & Voigt GmbH
Bismarckring 39 Bismarckring 39
D-88400 Biberach D-88400 Biberach
Tel. +49 7351 56-0 www.kavo.com
Fax +49 7351 56-1488
Instructions for use KaVo Dig eXam 3520, 3530
Contents

Contents

Contents ...........................................................................................................................................................1

1 User notes .....................................................................................................................................................3


1.1 Intended users .........................................................................................................................................3
1.2 User guidelines ........................................................................................................................................4
1.2.1 Standard documentation .................................................................................................................. 4
1.2.2 Abbreviations ................................................................................................................................... 4
1.2.3 Symbols used ................................................................................................................................... 5
1.2.4 Symbols on nameplates ................................................................................................................... 5
1.3 Service .....................................................................................................................................................6
1.4 Implied warranty .......................................................................................................................................7
1.5 Limitation of implied warranty .................................................................................................................. 8
1.6 Transportation and storage ...................................................................................................................... 9
1.6.1 Transportation damage .................................................................................................................... 9
1.6.2 Information on the packaging: Storage and transport ...................................................................... 9

2 Safety ..........................................................................................................................................................11
2.1 Explanation of safety symbols ............................................................................................................... 11
2.1.1 Hazard warning symbol .................................................................................................................. 11
2.1.2 Explanation of different levels of hazard ........................................................................................ 11
2.1.3 Structure .........................................................................................................................................11
2.2 Intended purpose ................................................................................................................................... 12
2.2.1 General ..........................................................................................................................................12
2.2.2 Product-specific .............................................................................................................................. 12
2.3 Safety symbols .......................................................................................................................................14
2.4 Prerequisite for proper technical operation ............................................................................................ 16
2.4.1 Data backup ................................................................................................................................... 16
2.4.2 Sensor handling ............................................................................................................................. 17
2.4.3 Standards for IT-components ......................................................................................................... 18
2.4.4 Standard conformity ....................................................................................................................... 18

3 Product description ......................................................................................................................................20


3.1 Entire system .........................................................................................................................................20
3.2 Where to affix the nameplates, power rating plates and caution labels ................................................. 21
3.2.1 Labeling ..........................................................................................................................................21
3.2.2 Symbols on identification labels ..................................................................................................... 23
3.3 Sensors ..................................................................................................................................................24
3.3.1 Installation sizes ............................................................................................................................. 24
3.3.2 KaVo Trigger .................................................................................................................................. 25
3.3.3 Data connection via USB 2.0 ......................................................................................................... 25
3.3.4 Portability of KaVo Dig eXam ......................................................................................................... 26
3.3.5 Holder system for sensor ............................................................................................................... 26
3.4 Integration in KaVo In eXam .................................................................................................................. 27
3.4.1 Attaching sensors to KaVo In eXam .............................................................................................. 27
3.4.2 Internal USB connection in the scissor arm ................................................................................... 28
3.5 Sensor holder .........................................................................................................................................29
3.5.1 Active Sensor holder ...................................................................................................................... 29
3.6 Status icon and sensor status on the PC ............................................................................................... 31

1/60
Instructions for use KaVo Dig eXam 3520, 3530
Contents

3.7 Technical Data .......................................................................................................................................33


3.7.1 KaVo Dig eXam Sensors and Interface ......................................................................................... 33
3.7.2 IT component requirements ........................................................................................................... 33
3.7.3 Ambient conditions for KaVo Dig eXam sensors and interface [compliant with IEC601-1] ............ 34
3.7.4 Equipment classification ................................................................................................................. 34
3.7.5 Conformity with standards .............................................................................................................. 34
3.7.6 Electrical connection ...................................................................................................................... 35
3.7.7 Dimensions, mass (original packaging) ......................................................................................... 35
3.7.8 Exposure times .............................................................................................................................. 35
3.7.9 Information on electromagnetic compatibility ................................................................................. 36
3.8 Dig eXam inspection service (only available in Germany) ..................................................................... 39

4 Operation .....................................................................................................................................................40
4.1 Configuration of the Dig eXam driver software ...................................................................................... 40
4.1.1 Selection Information ..................................................................................................................... 40
4.1.2 Selection configuration ................................................................................................................... 41
4.1.3 Selection parameters ..................................................................................................................... 43
4.2 Taking a digital X-ray ............................................................................................................................. 44
4.2.1 Overview ........................................................................................................................................44
4.2.2 Activate programme and select patient .......................................................................................... 44
4.2.3 Prepare and position sensor .......................................................................................................... 45
4.2.4 X-ray ...............................................................................................................................................47
4.2.5 Conclude X-ray procedure ............................................................................................................. 49
4.2.6 Create a diagnosis based on the radiograph ................................................................................. 49

5 Maintenance/safety inspections .................................................................................................................. 50


5.1 Checks ...................................................................................................................................................50

6 Preparation methods DIN EN ISO 17664 .................................................................................................... 51


6.1 Cleaning and disinfecting ....................................................................................................................... 51
6.1.1 Cleaning .........................................................................................................................................51
6.1.2 Disinfection .....................................................................................................................................51

7 Accessories and compatibility ..................................................................................................................... 53


7.1 eXam Family ..........................................................................................................................................53
7.2 Compatibility ..........................................................................................................................................57
7.2.1 Non-KaVo X-ray machines ............................................................................................................. 57
7.2.2 IT components ............................................................................................................................... 57

8 Troubleshooting ...........................................................................................................................................58
8.1 Symptoms, causes and solution ............................................................................................................ 59

2/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.1 Intended users

1 User notes

1.1 Intended users

This document is intended for use by dentists and other practice or clinic employees
who are conversant with the taking, processing and archiving of computer-based
intraoral X-rays on the Microsoft® Windows operating system .
Prior to using the equipment, it assumed that users have been instructed in the use
of KaVo Dig eXam-Sensors in accordance with the stipulations of the country in
which they are practising, on the safe and proper use of X-ray radiation for the taking
of intraoral dental X-rays.

3/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.2 User guidelines

1.2 User guidelines

Requirement
Read these instructions before the initial startup to prevent misuse and damage.

This document has been translated from the German original.

Note
Microsoft and Windows are registered trademarks of Microsoft Corporation and its
registered subsidiaries.
VixWin PRO is a registered trademark of Gendex Dental Systems.
XCP-DS is a registered trademark of DENTSPLY RINN.
KaVo Dig eXam is a protected trademark of KaVo.
The utilisation of company logos or trademarks not named here does not indicate
that they are subject to unrestricted use. Every company logo or trademark men‐
tioned here is the sole property of the company mentioned. Brands, names and
prices are given for information purposes only.

1.2.1 Standard documentation

The product-related documentation consists of the following components. You can


order more of these from KaVo, if necessary:
▪ Customer documents
– User instructions
– Assembly instructions
▪ User instructions Software VixWin PRO.

1.2.2 Abbreviations

Abbrevia‐ Meaning
tion
ALARA "As low as reasonably achievable": The guiding principle behind radiation protection during the
taking of X-rays, meaning that radiation doses for both workers and the public are typically kept
lower than their regulatory limits.
EMC Electromagnetic compatibility
UI User instructions
Hub A central point: a distribution device that makes available a number of USB interfaces.
IEC International Electrotechnical Commission
AI Assembly instructions
SQL "Structured Query Language": Standardised computer language for querying relational databases
SFC Safety checks
SW Software
STI Service Technician's instructions
USB Universal Serial Bus. Standardised means of connecting peripherals to personal computers.

4/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.2 User guidelines

1.2.3 Symbols used

See the section Safety/Warning Symbol

Important information for users and technicians

Action request

1.2.4 Symbols on nameplates

Type BF application part [5333;IEC 417]

Consult user instructions

CE (Communauté Européenne) mark. Products bearing this mark conform


to the requirements of the pertinent EC directives (applicable European
standards).
CSA label

SN: Serial number


REF: Material number (for sensor size IMat. no. 1.003.4498, for sensor size
IIMat. no. 1.003.4499)
Type: device type (for sensor size I 3520, for sensor size II 3530)
For information about proper disposal, see intended purpose

5/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.3 Service

1.3 Service

Service hotline:
+ 49 (0) 7351 56-2900
[email protected]
Please indicate the version number of the program.
Additional information can be obtained at: www.kavo.com

For North America

KaVo Dental
KaVo Technologies / Gendex Imaging
019 West Oakton Street
Des Plaines, IL 60018-1884
USA
+1 (847) 640-48 00

6/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.4 Implied warranty

1.4 Implied warranty

The KaVo end-user customer guarantee for the product named in the completion
certificate guarantees that the product functions correctly and that there are no faults
in the material or workmanship for a duration of 12 months following the purchase
date, according to the following conditions:
Following a reasonable complaint relating to defects or short delivery, KaVo will
provide a replacement or perform repairs. KaVo reserves the right to perform re‐
pairs.
Claims of any other nature, damages in particular, are excluded. In case of default
and gross negligence or intent, the latter only applies if there are no compelling legal
provisions opposing it.
KaVo shall not be liable for defects and their consequences, which occur as a result
of normal normal wear and tear, or of improper cleaning and maintenance, non-
observance of the operating, maintenance or connection regulations; calcination or
corrosion; a contaminated air or water supply; or chemical or electrical effects, which
are non-standard or not permitted according to company regulations.
As a general rule, this guarantee does not apply to lamps, glassware, rubber parts
or the colour durability of synthetic materials.
KaVo shall not be liable for defects or their consequences if they are likely to be a
direct result of actions or modifications by a customer or third party.
Any claims arising from this guarantee can only be lodged if the completion certifi‐
cate (carbon copy) has been sent in to KaVo and the operator/user is able to produce
the original.

7/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.5 Limitation of implied warranty

1.5 Limitation of implied warranty

KaVo shall assume responsibility for the safety, reliability and performance of all
KaVo system components provided that:
▪ Assembly, upgrades, new settings, modifications or repairs are all carried out
by Service Technicians, who have been trained by KaVo, or by KaVo authorised
dealer personnel
▪ The IT components provided by the operator correspond to the technical hard‐
ware and software requirements included in these user instructions.
▪ The system has been installed and operated in accordance with the User in‐
structions and Assembly instructions.
▪ All servicing is performed in full compliance with DIN VDE-075-1.
▪ In Germany, the location in whichKaVo Dig eXam is installed, must comply with
the regulationsDIN VDE 0100-710: 2002-11 and VDE 0100-560:1995 in its de‐
sign. In other countries, the corresponding national regulations must be respec‐
ted.
▪ The VixWin PRO software has been installed in accordance with the VixWin Pro
user instructions

8/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.6 Transportation and storage

1.6 Transportation and storage

1.6.1 Transportation damage

Outside of Germany

Note
KaVo is not liable for damage arising from transportation.
Immediately inspect the delivery after receipt!

If external damage to the packaging is visible upon delivery, follow the procedure
below:
1. The recipient must record the loss or damage in the notice of delivery. The re‐
cipient and employee of the transportation firm must sign the notice of delivery.
The recipient can only assert damages against the transportation company ba‐
sed on these records.
2. Leave the product and packaging unchanged.
3. Do not use the product.

If the product is damaged and there is no discernable damage to the packaging


upon delivery, proceed as follows:
1. Report the damage immediately or at least 7 days after the delivery to the deli‐
very company .
2. Leave the product and packaging unchanged.
3. Do not use a damaged product.

Note
If the recipient does not follow one of the above instructions, the damage will be
held to have occurred after the delivery (according to . CMR law , section 5, Art.
30).

1.6.2 Information on the packaging: Storage and transport

Note
Keep the packaging for returning the product for service or repairs .

The symbols printed on the outside are for transportation and storage, and have the
following meaning:

Transport upright with the arrows pointing upwards

Fragile - protect against knocks

Keep dry

Maximum permitted stacking load

9/60
Instructions for use KaVo Dig eXam 3520, 3530
1 User notes | 1.6 Transportation and storage

Temperature range

Humidity

Air pressure

10/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.1 Explanation of safety symbols

2 Safety

2.1 Explanation of safety symbols

2.1.1 Hazard warning symbol

Warning symbol

2.1.2 Explanation of different levels of hazard

Safety instructions with three hazard levels are used in this document for avoiding
personal and property damage.

CAUTION
indicates a hazardous situation that can lead to property damage or minor to mo‐
derate injury.
CAUTION

WARNING
indicates a hazardous situation that can lead to serious injury or death.
WARNING

DANGER
indicates a maximum hazardous situation that can directly cause serious injury or
death.
DANGER

2.1.3 Structure

The introduction describes the type and source of the hazard.


This section describes the potential consequences of non-observance.
▶ The optional step contains necessary measures for avoiding hazards.
DANGER

11/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.2 Intended purpose

2.2 Intended purpose

2.2.1 General

This KaVo product is intended only for use in the field of dentistry. It is impermissible
to use the product for a purpose for which it was not intended.
"Proper use" includes following all the instructions for use and ensuring that all in‐
spections and service tasks are performed.

Apply and meet the overarching guidelines and/or national laws, national regulati‐
ons and the rules of technology for medical devices applicable for startup and use
of the KaVo product for the intended purpose.

The user must ensure that that the device works properly and is in a satisfactory
condition before each use.

During use, national legal regulations must be observed, in particular:


▪ the applicable health and safety regulations.
▪ the applicable accident prevention regulations.

The user must observe the following:


▪ - only use properly operating equipment.
▪ protect himself or herself and third parties from danger.
▪ avoid contamination from the product.

The safety checks (SFC) must be performed on an annual basis.

Note
The product must be cleaned and serviced according to instructions if it is not to
be used for a long period.

Observe the corresponding country-specific regulations when finally shutting down


a KaVo product. Please direct all questions regarding the proper disposal of KaVo
products to the nearest KaVo branch.

2.2.2 Product-specific

Dig eXamtogether with VixWin PRO is intended for processing dental, intraoral X-
rays.
The radiographs are displayed and archived using the software VixWin PRO.
Prior to using the equipment, users of the KaVo Dig eXam-Sensors must have un‐
dergone full instruction, in accordance with the stipulations of the country in which
they are practising, on the safe and proper use of X-ray radiation for the taking of
intraoral dental X-rays.

KaVo Dig eXam has electronic components in the Trigger and Sensor that are to
be disposed of in accordance with the relevant country specific regulations.

12/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.2 Intended purpose

Disposal of electronics

Note
According to the EC Directive 2002/96 concerning electrical and electronic used
devices, this product is subject to the cited directive and must be disposed accor‐
dingly within Europe.
Before disassembling and disposing of the product, it must be completely proces‐
sed (disinfected, sterilised) according to the section "Preparation methods".
Additional information can be obtained from KaVo (www.kavo.com) or your dental
supplier.

13/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.3 Safety symbols

2.3 Safety symbols

Ionising radiation
Overexposure to X-ray radiation.
▶ The handler can and must take measures, in accordance with country-specific
CAUTION
legal stipulations, to minimise their personal dose of radiation!

Risk of injury by using unsuitable electrical devices.


Increased risks within patient surroundings; e.g. sensing electrical current or re‐
ceiving electric shock through contact with parts
CAUTION
▶ Adhere to the requirements described within these user instructions
▶ Only use IT components that are fully compliant with the requirements descri‐
bed within these user instructions.
▶ If any of the system components are damaged, stop use immediately.
▶ Only use standard-compliant devices.

Injury caused by damaged sensors


Damaged sensors may cause personal injury.
▶ If sensors are damaged, stop using them.
CAUTION
▶ If the sensor wire is damaged, stop using it.

Ignition of ignitable mixtures due to electrical equipment in operation


Deflagration
▶ KaVo Dig eXam must not be operated when using ignitable mixtures.
CAUTION

Malfunctions from electromagnetic fields.


The product meets the applicable requirements regarding electromagnetic fields.
Given the complex interactions between equipment and cell phones, the product
CAUTION
may be influenced by a cell phone that is in use.
▶ Do not use cell phones in medical offices, hospitals, or laboratories.
▶ Turn off electronic devices such as computer storage media, hearing aids, etc.
during operation .

Risks from electromagnetic fields.


The functions of implanted systems (such as pacemakers) can be influenced by
electromagnetic fields.
CAUTION
▶ Ask patients before treatment.

Loss of data when taking x-rays


When the transmission of data from the KaVo Dig eXam sensor to the computer
does not meet the USB 2.0 standard, data may be lost when taking x-rays due to
CAUTION
the limited bandwidth of earlier USB versions.
▶ Each USB connection between the computer and sensor must completely meet
the USB 2.0 standard.
▶ Each port (computer or hub) to which the sensors are connected must offer the
full output of 500 mA specified fur USB.
▶ The operating system for the computer must support USB 2.0 in high-speed
mode. The most recent service pack by Microsoft must be installed.
▶ The protective conductor connections described in these instructions must be
installed.

14/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.3 Safety symbols

Monitor image interference preventing accurate diagnosis


The quality of the X-ray image displayed on screen is dependent both on the prac‐
tice or clinic's ambient conditions as well as the monitor settings.
CAUTION
▶ If you suspect interference on the image display, use test images to check
whether the problem lies with the monitor.
▶ Clean the screen on a regular basis.
▶ Make sure that you use correct screen settings only.

Note
When positioning the screen, keep it out of direct sunlight and prevent light reflec‐
ting off it to enable the medical data to be read properly.

15/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation

2.4 Prerequisite for proper technical operation

Responsibility is assumed for the safety, reliability and performance of the compo‐
nent supplied by KaVo when:
▪ Installation, expansions, adjustments, changes or repairs is done by technicians
trained by KaVo or third parties authorised by KaVo for the Dig eXam, or by the
personnel of authorised distributors.
▪ The unit is operated according to the instructions for use and installation.
▪ The IT components supplied by the operator meet the technical requirements in
these instruction for use for hardware and software, and they are installed and
set up according to the descriptions of these components.
▪ KaVo Dig eXam is an electronic medical device, and is subject to the special
precautions necessary for EMC. KaVo Dig eXam may only be operated when it
has been installed according to the EMC instructions in the instructions for use
and installation.
▪ When setting up the unit, follow all the requirements of VDE 0751-1, "Repeated
tests and test before startup of electronic medical devices and systems - general
guidelines".
▪ In Germany ,the room in which the Dig eXam is installed must be designed ac‐
cording to the specifications DIN VDE 0100-710:2002-11 ad VDE
0100-560:1995. In other countries, follow the corresponding national specifica‐
tions.

Note
The installation and fitting of IT components for Dig eXam are the responsibility of
the practice or clinic.

Unauthorised accessories
Only original replacement parts and accessories should be used, to ensure that
there is no increase in the emissions and/or a reduction in the stability in terms of
CAUTION
the electromagnetic compatibility.
▶ Only use authorised or recommended spare parts and accessories.

See also: 3.7 Technical Data, Page 33

Observe the corresponding country-specific regulations when finally shutting down


a KaVo product. Please direct all questions regarding the proper disposal of KaVo
products to the nearest KaVo branch.

See also: Technical data in user instructions.

2.4.1 Data backup

It is the user's responsibility alone to back up data.

KaVo recommends daily back-up.

16/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation

Data loss
Establish a data backup strategy.
▶ VixWin PROuses the file system supplied with the Microsoft operating system
CAUTION
for saving image data. Correspondingly, this data must also be included within
the backup strategy of your practice or clinic.

Note
The image data produced using KaVo Dig eXam replaces the conventional intraoral
radiographs taken on X-ray film and, as such, is subject to the same country-spe‐
cific requirements governing this; such as the mandatory length of time for storing
X-rays etc.
These country-specific stipulations and requirements may also include the recor‐
ding of dosage levels and information about the patient's indication.

The user should regularly make back-up copies If a hard disk in the PC fails, this
protects from data loss and increases the safety of the system data. Such a copy is
also termed a backup. It should not be confused with the export of data in which the
data remain in the PC.
Store backup copies in a safe place.

There are various backup options. You can choose between:


▪ A full backup, which always backs up every single piece of data on the hard drive
that needs backing up.
▪ An incremental backup, which only saves data that has been modified since the
last backup.

A full backup is a lengthy procedure and cannot be performed on a daily basis.


Therefore, practices nowadays usually decide to implement a hybrid of the two
backup options (full and incremental). whereby, at one point in time, the entire
quantity of stored data is backed up and, on later occasions, backups are run at
regular intervals for modified data only.
Please also note that a hard drive error may occur whilst a backup is being written.
For this reason, you should never overwrite the most recent backup copy.

Note
The computer expert or system administrator at your practice will have the most
up-to-date technical information as regards keeping and running backup copies.

2.4.2 Sensor handling

Precautionary instructions for handling the sensors

Never:
▪ Sterilise the sensors or subject them to high-pressure sterilisation (autoclave).
▪ Drop the sensors.
▪ Twist or tread on the sensor wires.
▪ Let patients bite on the sensor or its wire.
▪ Remove the sensors when in use or while the X-ray is being taken.
▪ Pull or bend the wire..
▪ Immerse the KaVo trigger in the cleaning bath.

17/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation

Electrostatic discharge

▪ When the sensors are not in use, it is recommended that they be stored in the
storage containers provided.
▪ Never touch the USB connection contact points!

2.4.3 Standards for IT-components

The IT components used in conjunction with Dig eXam, in particular USB-connec‐


tion between Dig eXam and the PC must correspond with the details given in these
User instructions.

See also: 2.4.4 Standard conformity, Page 18

See also: 3.3.3 Data connection via USB 2.0, Page 25

2.4.4 Standard conformity

The KaVo Dig eXam Sensors and the KaVo Dig eXam Interface comply with the
following standards
▪ IEC 60601-1:1988 +A1:1991 + A2:1995 [For the USA UL 60601-1:2003. For
Canada CAN/CSA 22.2 N0.: 601-1]
▪ IEC 60601-1-2:2001

The KaVo Dig eXamdriver software conforms to standards:


▪ IEC 60601-1-4: 1996 + A1: 1999

Note
The delivered Vix Win Pro software also meets this standard.

KaVo Dig eXam and the IT components together make up a medical, electrical sys‐
tem.
The standards applicable to these IT components depend on their specific location
within the practice or clinic. In general, these IT components must comply with IEC
60950-1:2001.
The system that is formed by these IT components andKaVo Dig eXam when in‐
stalled in accordance with the assembly instructions for KaVo Dig eXam and meets
the requirements stipulated in this section, conforms to standard
- IEC 60601-1-1:2000

Rooms used for medical purposes

The room where KaVo Dig eXam Sensors are operated, must be designed in ac‐
cordance with regulationDIN VDE 0100-710:2002-11. In other countries, the corre‐
sponding national regulations must be respected.

18/60
Instructions for use KaVo Dig eXam 3520, 3530
2 Safety | 2.4 Prerequisite for proper technical operation

patient environment

When the above-mentioned IT components are installed in the patient environment,


the requirements of basic specialist standard IEC 60601-1:1988 +A1:1991 +
A2:1995
defines the surroundings of the patient in accordance with IEC 60601-1-1:2000: „ ...
any volume whereby, intentionally or unintentionally, a connection may be created
between the PATIENT and parts of the SYSTEM or other persons who may come
into contact with parts of the SYSTEM. ... “

Medical system

The requirements of the applicable standards have been met if:


▪ The IT-components on the Dig eXam via USB are installed in the same room
and
▪ One of the following applies:
– The Dig eXam Sensor together with the Dig eXam Interface is connected
(integrated solution in In eXam or in connection with the active sensor holder
of Dig eXam). The IT device can be run in accordance with IEC 60950-1 if
arranged outside of the patient surroundings.
– The IT components are located within the patient environment: The device
on which the Dig eXam is connected directly, fulfills the standard IEC
60601-1.
– The IT components are located outside of the patient environment: The de‐
vice is on which the Dig eXamis directly connected, complies with the
standard IEC 60950-1 and has a second permanently connected equipo‐
tential bonding that complies with IEC 60601-1-1.

Monitors

Note
The dental radiographs created byKaVo Dig eXam are examined and diagnosed
using the dental practice's own image representation device (monitor). There are
country-specific regulations in force governing (regular) check-ups of the greyscale
display for these image representation devices.
In Germany, at least one monitor at the dental practice must be tested for constancy
on a regular basis in accordance with DIN-V 6868-57:2001.

19/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.1 Entire system

3 Product description

3.1 Entire system

The entire KaVo Dig eXam consists of:


- the ① Dig eXam Sensors (comprising detector and trigger, the KaVo Dig eXam-
Interface (optional), the ② Dig eXam-Driver-Software
- the ③ VixWin PRO-Software
- IT-components
- accessories
KaVo Dig eXam and the VixWin PRO-Software together form a complete system
which can easily be used with non-KaVo X-ray machines.
Used in conjunction with KaVo In eXam and the Dig eXam interface, you have a
fully integrated X-ray system with extra functions.

Dig eXam has two additional software programs for customers in Germany:
1. A tool for supporting the required acceptance and constancy test in Germany.
2. A program for listing the existing images (x-ray folder)

< 5m
3

< 5m

HUB

HUB
1
HUB

HUB

20/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels

3.2 Where to affix the nameplates, power rating plates and caution
labels

3.2.1 Labeling

Kaltenbach & Voigt Gmb H


Bismarckring 39
D-88400 Biberach / Riss

Dig eXam Sensor Size 1


Type : 3520
Ref : 1.003.4498

SN :

5 C Manufactored : Month, Yea r

USB 2.0 max. 500 mA

1
X.XXX.XXXX
am

XXXXX
X
eX
ig
D

4
o
aV
K

3
2
A

① Product name ⑥ Rating plate overview (on packaging)


② Consult user instructions (A) Dot to mark position. This is entered

automatically on the radiographs by


KaVo Dig eXam.
③ Type BF application part (B) Sensor label (sensor I or sensor II)

④ KaVo Dig eXam sensor rating plate (C) USB symbol on connector

⑤ Label for KaVo Trigger/USB lead

21/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels

Hardware driver for Vix-Win Mat. no. 1.002.8220

① Program version ③ Unambiguous ID


② ID and language edition

Installations CD Vix Win Pro. Mat. no. 1.002.4813

22/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.2 Where to affix the nameplates, power rating plates and caution labels

3.2.2 Symbols on identification labels

Type BF application part [5333;IEC 417]

Consult user instructions

CE (Communauté Européenne) mark. Products bearing this mark conform


to the requirements of the pertinent EC directives (applicable European
standards).
CSA label

SN: Serial number


REF: Material number (for sensor size IMat. no. 1.003.4498, for sensor size
IIMat. no. 1.003.4499)
Type: device type (for sensor size I 3520, for sensor size II 3530)
For information about proper disposal, see intended purpose

23/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors

3.3 Sensors

3.3.1 Installation sizes

Sensor I (left) and sensor II (right).

The receiver side is marked with a dot that is also shown on the digital X-ray.

24/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors

3.3.2 KaVo Trigger

KaVo Trigger

The KaVo trigger connects the X-ray image detector to the computer system via a
USB 2.0 data path. This connection takes place in the X-ray device
KaVo In eXamsimply by inserting the interface in the X-ray head and in the
KaVo Dig eXamfitted in the machine.
When using the Dig eXam sensors with other X-ray machines it is possible to con‐
nect the KaVo Trigger to the so-called "active sensor holder" or alternatively to
connect it directly to the computer using the accessory cable supplied. In this case
a Dig eXAM sensor of one size can be exchanged with a sensor of another size by
changing the trigger.

Note
The accessory line is "Pull/Push“-Connection. It must not be twisted when pulled
out and plugged in.

when changing the Dig eXam Sensor from machine to machine or for another sensor
size, the detector can be laid down in the sensor holder fitted on the KaVo Trigger.
This sensor holder can be removed for cleaning the KaVo Trigger..

When the sensor holder is opened and the detector is removed, a button is acces‐
sible.
This button directly activates the system to take x-rays when the Dig eXam sensor
is not connected to the KaVo In eXam or is being operated in mixed-mode.

See also: 3.4.1 Attaching sensors to KaVo In eXam, Page 27

3.3.3 Data connection via USB 2.0

The sensor is connected to the machine via a USB 2.0 connection. The components
are physically connected to each other via simple plug-fit connections. Anchoring
the KaVo trigger into the device on the X-ray generator establishes the physical
connection between the machine and KaVo trigger or sensors. In accordance with
USB standards, the USB data connection may comprise up to four 5m sections.
Hubs that are compliant with USB 2.0 must be used to connect up the individual
sections

25/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.3 Sensors

Every hub that is directly connected to a KaVo sensor, i.e. with


noKaVo Dig eXamInterface, must be capable of supplying the requisite current. As
a rule, this means that a transformer will need to be used (also for laptops if neces‐
sary).
The requirements listed in the chapter "Standard conformity" must be complied with.

See also: 2.4.4 Standard conformity, Page 18

Note
When using 5 metre long USB-cables the the USB-connection must plugged di‐
rectly into the so-called motherboard. Insertion in the USB ports that are within the
PC on further cables, is not permitted.

Note
Passive USB extension cords may not be used. The distances between two HUBs
may only be bridged with a cord 5 m or less.

3.3.4 Portability of KaVo Dig eXam

KaVo Dig eXam is can be transported from one location to another, in order to
enable operation of the sensors in different treatment rooms.
KaVo Dig eXam is thus excellently suited for a multi-user installation in dental clinics
or group practices.
The handy size provides optimum flexibility and offers you independence when
using the sensors on a day-to-day basis.

3.3.5 Holder system for sensor

The positions in which the sensor can be placed in the mouth are the same as those
for classic radiology when using X-ray film.

A colour-coded XCP-DS holding system is supplied as standard to assist in the


positioning of KaVo Dig eXam sensors for the parallel/right-angle technique.

This system includes the following:


▪ Blue holder for the anterior teeth
▪ Yellow holder for the molar area
▪ A red holder for bitewing radiographs
▪ A green holder for endodontic use

Note
The use of the retention system is shown on the poster that comes with the retention
system.

See also:
7 Accessories and compatibility, Page 53

26/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.4 Integration in KaVo In eXam

3.4 Integration in KaVo In eXam

3.4.1 Attaching sensors to KaVo In eXam

KaVo Dig eXamSensors are specifically intended to be integrated with


theKaVo In eXam. In this case, the optional USB lead is separated from the KaVo
trigger and the KaVo trigger is inserted directly into the prepared opening at the head
of the X-ray machine. The detector itself is placed in its special holder on the KaVo
trigger, meaning that it is always on hand for use by the dentist.

Position of KaVo Sensors on the device

KaVo Dig eXam then enables the sensor to be removed without switching off the
machine, simply by pulling gently and placing in another room.

Storage in the RWT holder.

A Dig eXam sensor can be stored in the provided RWT holder on the In eXam.
The sensor must have hygienic protection and be inserted in the desired variation
of the RWT holding set.

27/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.4 Integration in KaVo In eXam

The prepared RWT holder can be inserted with the guide rod into the provide clip
on the In eXam.

Note
KaVo recommends that this option only be used for temporary storage to prevent
damage from bumps, etc.

3.4.2 Internal USB connection in the scissor arm

If the KaVo trigger is operated in conjunction with a KaVo In eXam machine, it is


connected to the system via a plug-fit connection at the X-ray emitter. This establi‐
shes a USB connection to the machine or to KaVo Dig eXam, enabling the transfer
of data.

Technical background:
KaVo X-ray machines are equipped with an internal USB line that leads directly to
theKaVo Dig eXaminterface in the X-ray machine via the scissor arm.

28/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.5 Sensor holder

3.5 Sensor holder

A sensor holder belongs to the Dig eXam.


It is fitted on the wall and is used for holding a sensor between treatments when the
Dig eXam-Sensor is not used together with the KaVo In eXam.

The sensor is held in the holder by a magnet.


The sensor holder can be taken off the KaVo Trigger in this case and laid in the
sensor tray.

3.5.1 Active Sensor holder

With the optionally obtainable Dig eXam interface Mat. no. 1.003.2363 the distance
to the PC or hub can be increased to a max. 5 m line length from the USB 2.0, for
example to place the IT components outside of the patient's environment.
The interface in this case is mounted covered by the plastic part of the sensor holder,
and the accessory line to the KaVo trigger is tightly connected to the arising active
sensor holder.

29/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.5 Sensor holder

In addition, it is possible in this case to run the line under the wall surface from the
sensor cover to the IT component.

30/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.6 Status icon and sensor status on the PC

3.6 Status icon and sensor status on the PC

Two icons are visible in the Windows system area (cut-out in the so-called taskbar
next to the system clock)
The blue -Icon displays that the software interfaces between Dig eXam-Hard‐
ware and VixWin-Software is loaded.
The colour of the -Icon shows the availability of the Dig eXam-Systems connected.

Note
Other icons can be displayed dependent on the set up of the X-ray system. These
icons refer to further KaVo or Gendex equipment and are described in the relevant
documentation.

Note
This area of the Windows desktop in which the status icons are show is a part of
the so-called windows taskbar. Windows should be configured so that the status
icons are always visible when taking an X-ray. The configuration of the taskbar is
described in Windows Help.

The following is a list of all the possible statuses and their icons

31/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.6 Status icon and sensor status on the PC

When the icon is green, the electronics, sensor and connections are operating cor‐
rectly.
The sensor is ready to detect the x-rays.

When the icon is YELLOW, the electronics, sensor and connections are operating
correctly.
The sensor is waiting for the exposure to be activated.

Note
If you are using a system that is not connected to KaVo In eXam, the KaVo trigger
must be used to activate the process and must be pressed before the X-ray is taken.
If you are using a system that is connected to KaVo In eXam, activation occurs
automatically once you start taking the X-ray.

The icon in RED status: Dig eXam is not connected to the PC and/or not ready to
take an X-ray.

32/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

3.7 Technical Data

3.7.1 KaVo Dig eXam Sensors and Interface

Data connection USB 2.0


Pixel size 18,5 µm / theoretical resolution 27 LP/
mm
Pixel binning 1x1
System resolution typ. 18 LP/mm / min 16 LP/mm [at
200µGy, 60kV] > 50 µGy
Active surface sensor size (3520) 30 x 22 mm2
Active surface sensor size II(3530) 36 x 27 mm2
kV range 60-70 kV

3.7.2 IT component requirements

Compliance with standards when instal‐ IEC 60601-1:1988 +A1:1991 +A2:1995


ling IT components in patient surroun‐
dings
Compliance with standards when instal‐ IEC 60950-1:2001
ling IT components outside of patient
surroundings and using the Dig eXam
interface to connect the sensor.
Compliance with standards when instal‐ IEC 60950-1: 2001 and a second pro‐
ling IT components outside of patient tective conductor connection
surroundings and connecting sensor di‐
rectly without using the Dig eXam inter‐
face.
Data interface for sensors USB 2.0, 500 mA, High Speed
Data interface for Dig eXam Interface 100 mA
USB 2.0
(self-powered mode)
Data interface for Dig eXam Interface 500 mA
USB 2.0
(bus-powered mode)
Operating system Windows 2000 or Windows 2003 with
most up-to-date service pack, Windows
XP with most up-to-date service pack
Labelling European CE marking
Connection of IT components to mains Use of portable multiple sockets is pro‐
supply hibited.

Note
Further information in theuser instructions VixWin PRO.

33/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

3.7.3 Ambient conditions for KaVo Dig eXam sensors and interface
[compliant with IEC601-1]

Operation

Air humidity Condensation is not permitted to form on


the machine; relative atmospheric humi‐
dity 30-75%
Air pressure between 700 hPa and 1060 hPa
Ambient temperature 10° to 40° Centigrade

Storage and transport

Air humidity relative Air humidity between 10% and


95% (TBC uwezel)
Air pressure between 700 hPa and 1060 hPa
Unpacked 5° to 70° centigrade
In original packaging -20° to 70° centigrade (short term up to
2h: -40° C to 85° centigrade)

3.7.4 Equipment classification

KaVo Dig eXam Sensors

▪ Class I
– in accordance with EC Directive 93/42/EEC of June 14, 1993
▪ Application part type BF (IEC 601-1)
– Not protected from the penetration of water [IP20].
– Not suitable for sterilisation.
– Suitable for wipe or surface disinfection using the cleansers described in the
instructions for use.
– Not suitable for operation in environments with combustible mixtures.

KaVo active sensor holder

▪ Not protected from the penetration of water [IP20].


▪ Not suitable for sterilisation.
▪ Suitable for wipe or surface disinfection using the cleansers described in the
instructions for use.
▪ Not suitable for operation in environments with combustible mixtures.

3.7.5 Conformity with standards

See also: 2.4.4 Standard conformity, Page 18

34/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

3.7.6 Electrical connection

KaVo Dig eXam Sensors

DC voltage 5V [supply via USB 2.0]


Max. power 500 mA
Data connection USB 2.0

KaVo active sensor holder

DC voltage 12V ±10 %


Max. current 600 mA
Data connection USB 2.0 interface
Compliance with Must comply with IEC 60950-1:2001
power supply standards
Length of feeder < 20 m

Note
In Europe, the manufacturer's CE marking on the power supply unit confirms that
relevant power supply standards have been complied with.

3.7.7 Dimensions, mass (original packaging)

Size of packaging approx. 300 x 330 x 220 mm³


Weight < 5 kg

3.7.8 Exposure times

KaVo Dig eXam sensors in conjunction with KaVo In eXam

Preset dose level of 5 (60 kV is recommended) in digital mode.

Note
This recommended dosage is based on thresholds which are applicable for the
Federal Republic of Germany and may need to be adjusted in accordance with
relevant country-specific regulations and/or individual diagnostic circumstances.

See also: User instructions KaVo In eXam.

KaVo Dig eXam sensors in conjunction with non-KaVo intraoral X-ray


equipment

The dose must be preselected according to applicable national regulations (ALARA


principle). The sensor can be operated over a wide dose range corresponding to
ISO film speed class F-20% up to speed class D.

35/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

KaVo recommends that the sensors be operated at 60kV

3.7.9 Information on electromagnetic compatibility

Guidelines and manufacturer's declaration - electromagnetic emissi‐


ons

In eXam is intended for use in an environment as stipulated below. The customer


or user of In eXam should ensure that it is operated in an environment of this type.

Emissions measurements Conformity Electromagnetic environment - gui‐


de
HF-emissions in accordance within Group 1 Dig eXam uses HF energy solely for
CISPR 11 its internal functions. Its HF radiati‐
on is therefore very low, and it is
unlikely that nearby devices will be
subject to interference.
HF-emissions in accordance with Class B Dig eXam is suitable for use in all
CISPR 11 devices, including those used in do‐
mestic environments and similar
which are connected directly to a
public supply grid which also sup‐
plies buildings which are used for
residence purposes.
Emissions of harmonic oscillations Class A Dig eXam is suitable for use in all
in accordance with IEC61000-3-2 devices, including those used in do‐
mestic environments and similar
which are connected directly to a
public supply grid which also sup‐
plies buildings which are used for
residence purposes.
Emissions of voltage fluctuations / Compliant Dig eXam is suitable for use in all
flicker in accordance with devices, including those used in do‐
IEC61000-3-3 mestic environments and similar
which are connected directly to a
public supply grid which also sup‐
plies buildings which are used for
residence purposes.

Stability tests IEC 60601 - Test level Compliance level Electromagnetic environment - gui‐
de
Discharge of static +/- 6 kV Contact di‐ +/- 6 kV Contact di‐ Floors should be made of wood or
electricity in accor‐ scharge scharge concrete or be covered with cera‐
dance with IEC +/- 8 kV Air discharge +/- 8 kV Air discharge mic tiles If the floor is covered with
61000-4-2 a synthetic material, the relative air
humidity must be at least 30 %.
Rapid transient electri‐ +/- 2 kV +/- 2 kV for +/- 2 kV for mains ca‐ The quality of the supply voltage
cal disturbances / burst mains cables +/- 1 kV bles should be equivalent to that of a ty‐
in accordance with IEC for input and output ca‐ +/- 1 kV for input and pical business or hospital environ‐
61000-4-4 bles output cables ment.
Surges in accordance +/- 1 kV Normal mode +/- 1 kV Common mode The quality of the supply voltage
with IEC 61000-4-5 voltage voltage should be equivalent to that of a ty‐
+/- 2 kV Common mode +/- 2 kV Common mode pical business or hospital environ‐
voltage voltage ment.

36/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

Stability tests IEC 60601 - Test level Compliance level Electromagnetic environment - gui‐
de
Voltage drops, shortli‐ < 5 % UT (> 95 % drop < 5 % UT (> 95 % drop The quality of the supply voltage
ved interruptions und in UT) for 1/2 period) in UT) for 1/2 period) should be equivalent to that of a ty‐
fluctuations in the sup‐ 40 % UT (60 % drop in 40 % UT (60 % drop in pical business or hospital environ‐
ply voltage in accor‐ UT) for 5 periods, UT) for 5 periods, ment. When the user of Dig eXam
dance with IEC demands continuing function of the
70 % UT (30 % drop in 70 % UT (30 % drop in
61000-4-11 device even when the power supply
UT) for 25 periods, UT) for 25 periods, is interrupted, it is recommended
< 5 % UT (> 95 % drop < 5 % UT (> 95 % drop that Dig eXam be powered via an
in UT) for 1/2 periods in UT) for 1/2 period) uninterruptible power supply or a
battery.
Magnetic field at the 3 A/m 3 A/m Magnetic fields at the mains fre‐
supply frequency quency should correspond to the
(50/60 Hz) in accor‐ typical values found in business
dance with IEC and hospital environments.
61000-4-8

Guidelines and manufacturer's declaration - electromagnetic stability

Dig eXam is intended for use in the electromagnetic environment stipulated below.
The customer or user of In eXam should ensure that it is operated in an environment
of this type.

Stability tests IEC 60601 - test level Compliance level Electromagnetic environment - gui‐
de
Directed HF disturban‐ 3 Veff 3V Portable and mobile radio equip‐
ces in accordance with 150 kHz to 80 MHz 3 V/m ment should not be used near to In
IEC 61000-4-6 3 V/m eXam or its cables, as the recom‐
Directed HF disturban‐ 80 MHz to 2.5 GHz mended safe distance has been
ces in accordance with calculated in accordance with the
IEC 61000-4-3 equation applicable for the trans‐
mitting frequency.
Recommended safe distance :
d = 1,2 * P up to 800 MHz
d = 2,3 * P for 800 MHz to 2.5 GHz
where PA is the power rating of the
transmitter in watts (WA) according
to information from the manufactu‐
rer of the transmitter, and d is the
recommended safe distance in me‐
tes (m).
The field intensity of stationary ra‐
dio transmitters should, at all fre‐
quencies according to local investi‐
gations,a be lower than the
compliance levelB.
Near to devices bearing the symbol
, interference is possible.

Comment:
The higher frequency range is applicable at 80 MHz und 800 MHz.

37/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.7 Technical Data

a
The field intensity of stationary transmitters such as the base stations of radio
telephones and mobile land radio devices, amateur radio stations, AM and FM
broadcasting services or television transmitters cannot, in theory, be determined
precisely in advance. To determine the electromagnetic environment in respect of
stationary transmitters, a study of the location should be considered.
If the measured field intensity at the location in which Dig eXam is being used ex‐
ceeds the compliance level above, care should be taken with Dig eXam to ensure
that it is functioning in accordance with regulations. If unusual performance features
are noted, additional measures may be required, such as changing the alignment
or moving In eXam somewhere else.
b
The field intensity should be less than 3 V/m over the frequency range 150 kHz to
80 MHz.

Recommended safe distances between portable and mobile telecom‐


munications equipment and Dig eXam

Dig eXam is intended for use in an electromagnetic environment in which HF dis‐


turbances are checked The customer or user of In eXam can help to prevent
electromagnetic disturbances by complying with the minimum distance between
portable and mobile HF telecommunications equipment (transmitters) and In eXam,
depending on the output power of the communication device, as stipulated below.

The table shows the necessary safe distance, as a function of the transmission
frequency, in m:

Rated power of the trans‐ 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
mitter in W D=1.2 * P D=1.2 * P D= 2.3 * P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

For transmitters for which the rated power is not shown in the table above, the dis‐
tance can be determined using the equation belonging to the relevant column, where
P is the rated power of the transmitter in watts (W) according to information from
the manufacturer of the transmitter.

Note
These guidelines may not be applicable in all situations The propagation of elect‐
romagnetic waves is affected by absorption and reflection due to buildings, objects
and people.

38/60
Instructions for use KaVo Dig eXam 3520, 3530
3 Product description | 3.8 Dig eXam inspection service (only available in Germany)

3.8 Dig eXam inspection service (only available in Germany)

The function and calibration of connected devices, such as the monitor and X-ray
generator can be inspected and/or reconfigured via a separate service application
(KaVo Dig eXam inspection service).

39/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software

4 Operation

4.1 Configuration of the Dig eXam driver software

▶ Left click on the X-ray Icon, to open the configuration dialogue.

The configuration dialogue covers three areas that are activated via the selection
tab:
▪ Information ("Information“)
▪ Configuration ("Configure“)
▪ Parameters ("Treatments“)

4.1.1 Selection Information

This dialog shows the software version of the interface between Dig eXam and Vix‐
Win in the top part.
The status of the connected sensors is shown in the middle area. When the Dig
eXam interface is activated in the Configure dialog, the status of the interface is also
shown. By clicking the Update button, this information can be reentered by the
Windows operating system.
When the Dig eXam Interface is not used, click the button "Prepare Acquisition" in
the bottom area (instead of using the button on the KaVo trigger) or in mixed mode
to prepare the device for taking an x-ray.

Note
The button is only displayed in the "Use In eXam" mode when "Mixed mode" is also
active.

40/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software

4.1.2 Selection configuration

The upper part of the dialogue makes it possible to set the X-ray parameters that
are transferred automatically to VixWin.
The system events (e.g. "Sensor covered", "In eXam switched off") to be displayed
by the operating system in their own windows can be set in the bottom area of the
dialogue window.

Note
These data are used by the separate X-ray folder support program.

Without In eXam

By selecting the mode "Custom“ means that the user can specify standard values
for tube voltage (kV), the tube current (mA) and the exposure time (ms). These
parameters are added to every X-ray produced using Dig eXam and can be dis‐
played later in the VixWin-Software and if necessary adjusted to the parameters
selected for the individual X-ray image.

Using In eXam

When using In eXam the exposure time of the current X-ray image is automatically
recognised by the system and added to the X-ray image taken using Dig eXam. The
possible standard values can be preset for the values for the tube voltage (kV) and
the tube current (mA), in order to add these to the X-ray image as well. All three
parameters can be displayed in VixWin later and if necessary adjusted to the para‐
meters selected for the individual X-ray image.

41/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software

Dig eXam as the exclusive image receptor

▶ -Deactivate the "Mixed mode" button.


Each time In eXam is triggered, a sensor image is taken.

Dig eXam alternating with other image receptors

Requirement
Each time before a sensor image is taken with Dig eXam, the system must assume
a ready state by pressing the trigger button or the button "Prepare for exposure" to
obtain a sensor image.

▶ Activate the "Mixed mode" button.

42/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.1 Configuration of the Dig eXam driver software

4.1.3 Selection parameters

This dialogue enables the setting of the details that are used for displaying the X-
ray image taken.

Note
It is recommended that the parameters are not altered without consulting the tech‐
nician and or KaVo Customer Services.

43/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

4.2 Taking a digital X-ray

4.2.1 Overview

Taking an X-ray using KaVo Dig eXam together with VixWin PRO takes place in a
few steps that are explained in detail in the following:
▪ Start the program or toggle to it if already open
▪ Select patient
▪ Prepare and position sensor
▪ Take X-ray
▪ Form diagnosis from image
▪ Accept X-ray

4.2.2 Activate programme and select patient

▶ Click the VixWin Pro-Software icon.

44/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

or
▶ Activate the program from the start menu

or
▶ Call up VixWin Pro via the interface administration software used in the practice /
clinic.

VixWin PRO is active.

▶ Select patient.

See also: User instructionsVixWin PRO.

Note
It is possible, as an alternative to the procedure described here, to take the X-ray
first and then after taking the X-ray to allocate the X-ray to the patient.

4.2.3 Prepare and position sensor

Hygiene is a key issue during the preparation and taking of X-rays and must be
considered alongside the issue of radiation protection.

Hygiene

To avoid health risks,KaVo sensors must be placed in a disposable hygienic pro‐


tective cover before each use. So called hygienic covers are placed over the sensor
before each treatment to guarantee the optimum levels of hygiene.

▶ Cover sensor with hygienic covers.


▶ Use a new disposable hygienic sensor cover for every patient.

Risk of infection
Contaminated sensors can transmit infections.
▶ Your practice should have a set of hygiene procedures which must be adhered
CAUTION
to.
▶ Disinfect sensors after contamination.
▶ Disinfect the used parts of the right-angle holder ( XCP-DS) after contamination.
▶ Always use hygienic covers Mat. no. 1.003.2690.

45/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

Prepare the sensor with the hygiene sheet

See also: 6.1 Cleaning and disinfecting, Page 51

Right-angle holder set

Avoidable exposure to radiation.


Increased exposure to radiation through not using a radiation field limitation facility
during the right-angle technique.
CAUTION
▶ Always use a right-angle holder and radiation field limitation facility unless there
is a contra-indication in a particular case.

The XCP DS holding system is supplied with an operating instruction poster. For
further explanations on X-raying techniques, please see related textbooks.

See also:
Pasler, Friedrich A.: Zahnärztliche Radiologie (Dental Radiology), Thieme, 4th edi‐
tion 2003, ISBN 3136046048
Whaites, Eric: Essentials of Dental Radiography and Radiology (Dental Series),
Churchill Livingstone 1996, ISBN 0443053499

Position KaVo Dig eXam

TheKaVo Dig eXamdetector is positioned at the X-ray tube head during use of
theKaVo In eXam so that it is always in close proximity to the patient's head.

Place the sensor

If you are usingKaVo Dig eXam with X-ray sources from other manufacturers, the
patient should either be given the KaVo trigger to hold or have it placed on their
chest.

46/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

Faults as a result of disruption to the USB line.


Breaking the connection between the KaVo trigger and USB line during use can
lead to faults.
CAUTION
▶ During use, never disconnect the KaVo trigger from the USB line!

Faults as a result of patient action.


If patients are unsupervised during the X-ray procedure, they may do something
that results in a fault or disruption.
CAUTION
▶ Ensure that patients are never left unsupervised during the X-ray procedure.

4.2.4 X-ray

▶ Position the X-ray source in accordance with the X-ray machine's operating in‐
structions for the type of X-ray you are about to take.

X-ray radiation danger


Increased exposure to radiation for patient and/or operator.
▶ Adhere to the X-ray machine operating instructions and country-specific regu‐
CAUTION
lations governing patient protection.
▶ Adhere to the X-ray machine operating instructions and country-specific regu‐
lations governing operator protection by keeping a safe distance or using a
screen.

Activate sensors (use eXam with other image receptors)

Requirement
When using third-party x-ray equipment or when the mixed mode setting is activa‐
ted to support the alternating use of other image receptors, activate the status
"Ready to take x-rays".

▶ Press the KaVo trigger button.

KaVo Trigger

or

47/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

▶ Select button "Repair Acquisition“ in the information area of the Dig eXam-driver
software.

Note
This step is not necessary for KaVo In eXam with activatedKaVo Dig eXam Inter‐
face. KaVo Dig eXamis always ready to take an X-ray if the sensor is inserted on
account of its permanent connection to the interface.

Visual inspection

Visual inspection

Before you take the X-ray, have a quick look to make sure that
▪ The green icon is showing in the Windows System area
▪ The patient has not moved at all
▪ When using Dig eXam without In eXam-connection the ready dialogue is visible.

Sensor is activated (operation without In eXam connection)

48/60
Instructions for use KaVo Dig eXam 3520, 3530
4 Operation | 4.2 Taking a digital X-ray

Image lost during the taking of X-rays.


The X-ray will not take if the driver is not active.
▶ Before taking an X-ray, make sure the software is ready.
CAUTION

▶ Take X-ray.

4.2.5 Conclude X-ray procedure

After you have finished taking the X-ray


▪ Detach the holding system and sensor
▪ Remove hygienic cover
▪ Restore used holder
▪ Restore contaminated sensor
▪ When using non-KaVo X-ray sources, store the sensor in a safe place

4.2.6 Create a diagnosis based on the radiograph

See also: User instructions VixWin Pro.

49/60
Instructions for use KaVo Dig eXam 3520, 3530
5 Maintenance/safety inspections | 5.1 Checks

5 Maintenance/safety inspections

5.1 Checks

The following checks must be made at the times specified


▪ Monthly: Check the data backup function is running smoothly.
▪ Annually: Perform a visual inspection of connection leads, wires and casing to
check for damage.

50/60
Instructions for use KaVo Dig eXam 3520, 3530
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning and disinfecting

6 Preparation methods DIN EN ISO 17664

Damage resulting from improper equipment preparation


Improper cleaning and disinfection can lead to damage to the KaVo Dig eXam.
▶ Adhere closely to the instructions provided in this section.
CAUTION
▶ Use the correct substances and procedures for the job at hand.

6.1 Cleaning and disinfecting

Note
KaVo Dig eXam does not include any parts requiring sterilisation as part of their
intended use.

The sensor holder can be removed from the KaVo trigger for cleaning and disin‐
fection.

Note
The detector including the connecting line should be provided with a hygiene sheet
to prevent direct contamination by the detector during use.

The sensors used within intraoral dentistry are highly sophisticated electronic devi‐
ces. And as such should be handled with extreme care.
▪ Before you begin cleaning and disinfecting, detach the sensors from the USB
connection.
▪ Do not exert pressure on the sensors during the cleaning and disinfection pro‐
cedure; in particular, do not pull or bend their wires.
▪ Do not entangle the KaVo trigger and USB line when connecting or disconnec‐
ting.
▪ Do not allow moisture to penetrate the KaVo trigger and USB line plug-fit con‐
nections and the line's USB connector.

6.1.1 Cleaning

All external surfaces of the KaVo Dig eXam can be wiped with a soft cloth impreg‐
nated with a mild cleaning fluid.
The sensor holders and separate parts of the right-angle holder set can also be
machine-cleaned.

6.1.2 Disinfection

Disinfect contaminated parts of KaVo Dig eXam after use.

Wipe disinfection with disinfecting agents containing alcohol

All external surfaces of the KaVo Dig eXam can be wiped with a soft cloth impreg‐
nated with a mild cleaning fluid.

The following disinfectant is permitted:


▪ Elastoclean Spray 2
Manufacturer: KaVo Dental GmbH

51/60
Instructions for use KaVo Dig eXam 3520, 3530
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning and disinfecting

▪ MIKROZID Pump-Spray
Manufacturer: Schülke & Mayr
▪ Dürr System-Hygiene FD 322 Sprühdesinfektion
Manufacturer: DÜRR Dental GmbH
▪ Incidin liquid
Manufacturer: Ecolab
▪ USA: Asepticare Aerosol Spray
Manufacturer: Ecolab

Prohibited disinfectants:
▪ Phenol-based glutaraldehyde

Heat disinfection

When circumstances dictate, the sensor holder and separate parts of the right-angle
holder set may be disinfected at 135°C in an autoclave in accordance with the ma‐
nufacturer's instructions. The plastic parts must not be allowed to come into contact
with the autoclave walls.

Thermodisinfection/Miele

When circumstances dictate, the sensor holder and separate parts of the right-angle
holder set may be heat disinfected (up to a maximum of 50 times) in a thermodi‐
sinfector system, such as Miele's G 7781, in accordance with the manufacturer's
instructions.

52/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family

7 Accessories and compatibility

As a rule, KaVo recommends additional equipment from the eXam range.

7.1 eXam Family

KaVo Dig eXam belongs to the KaVo eXam intraoral X-ray system. The X-ray ma‐
chine KaVo In eXam is available under no. Mat. no. 1.002.4800 .

Presentation Material summary Mat. no.


Rectangular radiation field limi‐ 1.003.0341
tation, max. radiation field 35 *
45 mm²

Rectangular radiation field limi‐ 1.003.7200


tation, max. radiation field 35 *
45 mm²

53/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family

Presentation Material summary Mat. no.


Test object for constancy 1.003.0252
check, digital

Test object for constancy 1.003.0250


check, film

Cable CCU/USB plug 1.003.1828

Hygienic covers 1.003.2690

sensor holder 1.002.4806

54/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family

Presentation Material summary Mat. no.


eXam holder set for parallel 1.003.2676
technique

Cone extension 30 cm 1.003.1576

Sensor holder 1.003.7203

55/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.1 eXam Family

Presentation Material summary Mat. no.


Sensor holder active 1.003.2363
Kit for sensor holder on light 1.003.0008
mount pole

See also: the spare part sheet for the separate parts of the holding system.

56/60
Instructions for use KaVo Dig eXam 3520, 3530
7 Accessories and compatibility | 7.2 Compatibility

7.2 Compatibility

7.2.1 Non-KaVo X-ray machines

KaVo Dig eXam is compatible with all X-ray machines whose intended purpose is
to take dental intraoral radiographs, provided that these machines do not have do‐
sage level settings which exceed film sensitivity class D.

Avoidable radiation exposure


Greater radiation exposure from not observing the ALARA principle.
▶ Apart from the technical suitability of the KaVo Dig eXam sensors for the doses
CAUTION
of film sensitivity D, use the smallest possible dose for each indication.

7.2.2 IT components

▪ Image viewers/monitors
– The computer monitors must be suitable for the examination of radiographs.
Observe the relevant federal regulations. The provided test images can also
be used for monitors that are only for showing radiographs.
▪ Printing out radiographs
– The achievable quality of printouts greatly depends on the printer technology.
Printouts are generally unsuitable for diagnostics. X-ray images should ge‐
nerally be diagnosed on suitable monitors.

The technical compatibility requirements of the USB components used to create


radiographs have already been mentioned in the product description.

See also:
3.7.2 IT component requirements , Page 33
2.4.4 Standard conformity, Page 18

57/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 7.2 Compatibility

8 Troubleshooting

The following table is a troubleshooting guide to help solve typical problems without
requiring any special technical training.
The only precondition is that the KaVo Dig eXam has been installed correctly.

If you are unable to remedy your problem by following the solution methods pre‐
sented here, please contact your responsible KaVo partner.

58/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 8.1 Symptoms, causes and solution

8.1 Symptoms, causes and solution

Malfunction Cause Remedy


Everything has stopped working. Power supply interrupted. ▶ Make sure equipment is plugged
in at the mains.
No image. Insufficient X-ray radiation. ▶ Increase X-ray radiation in ac‐
cordance with user instructions.
See also: GA X-ray equipment
When using a non-KaVo X-ray ma‐ ▶ Make ready before triggering the
chine: KaVo Icon red (Dig eXam X-ray radiation by activating
was not ready for capturing digital (press the button on the KaVo
images). Trigger).
Image is too dark. The exposure time is too long or the ▶ Reduce exposure time.
entry mode selected is not appro‐ See also: GA X-ray equipment
priate for Dig eXam.
The entry mode selected does not
match the X-ray dose.
Radiograph is pale or contains in‐ The exposure time is too low. ▶ Increase exposure time.
terference The image entry mode selected ▶ Consult X-ray emitter operating
does not match the configured X- instructions.
ray dose.
The voltage set for the X-ray emitter ▶ Inspect X-ray emitter.
is too low (< 60 kV).
The X-ray tube head is too far away ▶ Check the settings and positio‐
from the patient. ning and change if necessary.
Monitor is not positioned optimally. ▶ Check how the monitor is positi‐
oned and alter its position if ne‐
cessary.
Monitor settings are incorrect. ▶ Use appropriate test images to
test your monitor settings.
The parameter settings for automa‐ See also: VixWin Pro-user Instructi‐
tic image display using VixWin PRO ons.
have been set to unsuitable values. See also: 4.1.3 Selection parame‐
ters, Page 43
Radiograph is blurred Patient moved whilst X-ray was ▶ Inform the patient of the problem
being taken. and take another X-ray.
X-ray tube head was not fastened ▶ Check that the X-ray tube head
down securely enough or was mo‐ is firmly in place and, if neces‐
ved whilst the X-ray was being ta‐ sary, use the software's filter
ken. function to improve image quali‐
ty.
The radiograph is white X-ray dosage too low ▶ Increase the X-ray dosage.
The active side of the sensor, which ▶ Direct X-ray tube head and sen‐
is sensitive to radiation, was not di‐ sor.
rected towards the radiation source.
Sensor not connected properly. ▶ Check connection.
▶ Check that the X-ray emitter is
working.
▶ If necessary, repeat the sensor
installation procedure.
▶ If you have no success after
trying the above, please call the
Service team.
KaVo Icon stays red. In eXam switched off. ▶ Switch In eXam on.

59/60
Instructions for use KaVo Dig eXam 3520, 3530
8 Troubleshooting | 8.1 Symptoms, causes and solution

Malfunction Cause Remedy


When using a non-KaVo X-ray ma‐ ▶ Make ready before triggering the
chine: KaVo Icon red (Dig eXam X-ray radiation by activating
was not ready for capturing digital (press the button on the KaVo
images). Trigger).
In eXam the configuration setting of ▶ Select In eXam in the configura‐
the Dig eXam-driver software is tion area of the Dig eXam-driver
deactivated. software.
See also: 4.1.2 Using In eXam, Pa‐
ge 41
The software connection between ▶ Reset the software connection
Dig eXam-driver and the Windows between Dig eXam and Wind‐
operating system is faulty. ows.
To do this press the Update but‐
ton in the configuration area of
the Dig eXam-driver software.
See also: 4.1.1 Selection Informati‐
on, Page 40
▶ Reset the software connection
by pulling and reinserting the
KaVo-Trigger.
▶ Shutdown the computer (switch
off completely) and then start
again.
The physical USB 2.0-connection ▶ Reinsert USB-plug.
via cables and or hubs between the or
KaVo Trigger and the computer is ▶ Insert the hub power supply
faulty. again.
or
▶ Have USB-2.0-cables checked
and if necessary replace dama‐
ged cables.
or
▶ Check the potential equalisation
between the PC and eXam.
or
▶ Call service technician/
The operating system asks for con‐ ▶ Confirm the relevant buttons for
firmation to activate the sensor dri‐ activating the drivers queried by
ver. the operating system.

60/60
1.003.4073 · KF · 20050801 - 03 · en

You might also like