F205
EN G L I S H User’s manual
Clamp Multimeter F205 English
CONTENTS
1 PRESENTATION ........................................................................................................ 8
1.1 THE SWITCH .....................................................................................................................9
1.2 THE KEYS OF THE KEYPAD .......................................................................................10
1.3 THE DISPLAY UNIT.......................................................................................................11
1.3.1 The symbols of the display unit .................................................................... 11
1.3.2 Measurement capacity exceeded (O.L) ........................................................ 13
1.4 THE TERMINALS ...........................................................................................................13
2 THE KEYS ................................................................................................................. 13
3
Clamp Multimeter F205 English
4
English Clamp Multimeter F205
You have just acquired an F205 clamp multimeter and we thank you.
9 V battery.
AC – Alternating current.
Earth.
5
Clamp Multimeter F205 English
This device complies with safety standards IEC-61010-1 and 61010-2-032 for
voltages of 1000V in category III or 600V in category IV at an altitude OF less than
2000m, indoors, with a degree of pollution not exceeding 2.
These safety instructions are intended to ensure the safety of persons and proper
operation of the device. If the tester is used other than as specified in this data
sheet, the protection provided by the device may be impaired.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If you use this instrument other than as specified, the protection it provides
may be compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of
flammable gases or fumes.
Do not use the instrument on networks of which the voltage or category
exceeds those mentioned.
Do not exceed the rated maximum voltages and currents between terminals
or with respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any element of which the insulation is deteriorated (even
partially) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal to
those of the instrument. If not, an accessory of a lower category lowers the
category of the combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use.
Do not modify the instrument and do not replace components with
"equivalents". Repairs and adjustments must be done by approved qualified
personnel.
Replace the battery as soon as the symbol appears on the display unit.
Disconnect all cords before opening the battery compartment cover.
Use personal protective equipment when conditions require.
Keep your hands away from the unused terminals of the instrument.
6
English Clamp Multimeter F205
When handling the test probes, crocodile clips, and clamp ammeters, keep
your fingers behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the
device, we recommend performing configuration operations only when the
device is disconnected from all dangerous voltages.
MEASUREMENT CATEGORIES
7
Clamp Multimeter F205 English
1 PRESENTATION
The F205 is a professional electrical measuring instrument that combines the
following functions:
Current measurement;
Measurement of inrush current / overcurrent (True-Inrush);
Voltage measurement;
Frequency measurement;
Measurement of level of harmonics (THD)
Continuity test with buzzer;
Resistance measurement;
Diode test;
Power measurements (W, VA, var and PF);
Indication of the order of the phases.
1
Item Designation See §
3 3 Switch 1.1
7
4 Function keys 2
8
English Clamp Multimeter F205
9
Clamp Multimeter F205 English
1 2 3
4 5 6
10
English Clamp Multimeter F205
7 6
1
3
4
Symbol Designation
11
Clamp Multimeter F205 English
V Volt
Hz Hertz
W Watt
A Ampere
% Percentage
Ω Ohm
m Milli- prefix
k Kilo- prefix
VA Apparent power
PF Power factor
Continuity test
Diode test
The display of “rdy”, for “ready”, indicates that the device is ready (“Indicator of
order of the phases” function).
12
English Clamp Multimeter F205
Item Function
1 Cold terminal (COM)
2 Hot terminal (+)
1 2
Figure 5 : the terminals
2 THE KEYS
The keys of the keypad respond differently to short, long, and sustained presses.
The , and keys provide new functions and allow the detection and
acquisition of parameters complementary to the usual elementary measurements.
- Each of these keys can be used independently of the others or in perfect
complementarity with them: this makes navigation simple and intuitive for
looking up all measurement results.
- It is possible, for example, to look up in turn the MAX, MIN, etc. values of the
RMS voltage alone, or else look up in turn all of the MAX (or MIN, or PEAK)
values of all power results (W, VA, var, etc.).
In this section, the icon represents the possible positions of the switch for which
the key concerned has some action.
2.1 KEY
This key is used to:
store and look up the last values acquired specific to each function (V, A, Ω,
T°) according to the specific modes previously activated (MAX/MIN); the
present display is then maintained while the detection and acquisition of new
values continues;
13
Clamp Multimeter F205 English
Successive
presses on … serve
See also § 2.4.2 and § 2.5.1 for the action key with the action of the key
and with the action of the key.
Remark: the key is invalid in the MAX/MIN/PEAK, HOLD and ΔREL modes.
Successive
… serve
presses on
short -to select AC, DC or AC+DC. Depending on your choice, the
screen displays AC, DC or AC+DC
14
English Clamp Multimeter F205
-to cycle through the Ω and diode test modes and to return
to the continuity test
- to reset the measurement process for the “indicator of order
of rotation of the phases” function.
Long (>2 sec) - to display the total three-phase power of a balanced system
( is displayed).
- by pressing again, to return to display of the single-phase
power ( is off)
2.3 KEY
This key is used to backlight the display unit.
Successive
… serve
presses on
2.4 KEY
This key activates detection of the MAX, MIN, PEAK+ and PEAK- values of the
measurements made.
Max and Min are the extreme mean values in DC and the extreme RMS values in AC.
Peak+ is the maximum instantaneous peak and Peak- the minimum instantaneous
peak.
R emark : in this mode, the "automatic switching off" function of the device is
automatically de-activated. The symbol is displayed on the screen.
15
Clamp Multimeter F205 English
Successive presses
… serve
on
-to activate detection of the MAX/MIN/PEAK values
-to display the MAX, MIN, PEAK+ or PEAK- value successively
-to return to display of the present measurement without exiting
from the mode (the values already detected are not erased)
R emark: the MAX, MIN, PEAK+, PEAK- symbols are both displayed, but
only the symbol of the quantity selected blinks.
short
Example: If MIN has been selected, MIN blinks and MAX, PEAK+, PEAK-
are lit steadily.
-to activate detection of the MAX/MIN values
-to display the MAX or MIN value successively
-to return to display of the present measurement without exiting
from the mode (the values already detected are not erased)
Remark : ΔREL function can be used with the functions of the MAX/MIN/PEAK
mode.
Note : the HOLD function does not interrupt the acquisition of new MAX, MIN,
PEAK values.
16
English Clamp Multimeter F205
Successive
…serves
presses on
long (>2 sec) to enter the True-INRUSH mode
-"Inrh" is displayed for 3s (the backlighting blinks)
-the triggering threshold is displayed for 5s (the
backlighting is steady);
-"------" is displayed and the "A" symbol flashes
-after detection and acquisition, the inrush
current measurement is displayed, after the
calculations stage "------" (backlighting off)
2.5 KEY
This key is used to display the frequency measurements of a signal, of the power,
and of the level of harmonics.
Successive
…serves
presses on
17
Clamp Multimeter F205 English
short to display:
-the frequency of the signal measured
-the present voltage (V) or current (A)
measurement
to display:
-the apparent power (VA)
- the reactive power (var)
- the power factor (PF)
- the frequency of the signal
- the active power (W)
long - to enter or exit from the level of harmonics
(THD) calculation and display mode
then short - to select THDf, THDr or the frequency of the
fundamental
Successive
…serves
presses on
short -to store the frequency
-to display successively the stored frequency,
then the voltage or the current
- to display in turn the stored values of THDf,
then of THDr, then of the frequency of the
fundamental
2.6 KEY
This key is used to display and store the reference value or to display the diferential
and relative value, in the unit of magnitude measured or in %.
18
English Clamp Multimeter F205
Successive
…serve
presses on
- to enter the ΔREL mode, to store then display the reference
value. The ΔRef symbol is displayed.
short - to display the diferential value:
- (current value – reference (∆))
The ΔREL symbol is displayed.
- to display the relative value in %
current value – reference (∆)
reference (∆)
The ΔREL and % symbols are displayed.
- to display the reference. The ΔRef symbol is displayed
- to display the current value. The ΔRef symbol blinks.
Remark : the “Relative mode ΔREL” function can be used with the functions of the
MAX/MIN/PEAK mode.
3 USE
3.1 COMMISSIONING
Insert the battery supplied with the device as follows:
1. Using a screwdriver, unscrew the screw of the battery compartment cover
(item 1) on the back of the housing and open it.
2. Place the battery in the compartment (item 2), taking care to get the
polarities right.
3. Close the battery compartment cover and screw it to the housing.
19
Clamp Multimeter F205 English
3.4 CONFIGURATION
As a safety measure, and to avoid repeated overloads on the inputs of the device,
we recommend performing configuration operations only when the device is
disconnected from all dangerous voltages.
20
English Clamp Multimeter F205
1. From the OFF position, hold the key down while turning the switch to
, until the "full screen" display ends and a beep is emitted, to enter the
configuration mode. The display unit indicates the value below which the
buzzer is activated and the symbol is displayed.
The value stored by default is 40Ω. The possible values lie between 1Ω
and 599Ω.
2. To change the threshold, press the key. The right-hand digit flashes:
each press on the key increments it. To shift to the next digit, apply a
long press (>2s) to the key.
To exit from the programming mode, turn the switch to another setting. The
detection threshold chosen is stored (emission of a double beep).
1. in the OFF position, hold the key down while turning the switch to
, until the "full screen" display ends and a beep is emitted, to enter the
configuration mode. The display unit indicates the percentage overshoot to
apply to the measured current to determine the measurement triggering
threshold.
The value stored by default is 10%, representing 110% of the established
current measured. The possible values are 5%, 10%, 20%, 50%, 70%,
100%, 150%, and 200%.
21
Clamp Multimeter F205 English
2. To change the threshold, press the key. The value flashes: each
press on the key displays the next value. To record the chosen
threshold, apply a long press (>2s) on the key. A confirmation beep is
emitted.
To exit from the programming mode, turn the switch to another setting. The chosen
threshold is stored (emission of a double beep).
Note: The starting current measurement triggering threshold is fixed at 1% of the
least sensitive range. This threshold is not adjustable.
22
English Clamp Multimeter F205
23
Clamp Multimeter F205 English
Warning : before the compensation is executed, the MAX/MIN and HOLD modes
must be de-activated.
To perform automatic compensation of the resistance of the leads, proceed as
follows:
1. Short-circuit the leads connected to the device.
2. Hold the key down until the display unit indicates the lowest
value. The device measures the resistance of the leads.
3. Release the key. The correction and the symbole are
displayed. The value displayed is stored.
Remark : the correction value is stored only if it is
≤ 2 Ω. Above 2 Ω, the value displayed blinks and is not stored.
Remark : to measure low resistance values, first carry out the compensation of the
resistance of the leads (see § 3.6.1).
24
English Clamp Multimeter F205
3.9.1 AC measurement
For an AC current measurement, proceed as follows:
1. Set the switch to and select AC by pressing the key. The AC
symbol is displayed.
2. Encircle only the conductor concerned with the clamp ;
25
Clamp Multimeter F205 English
26
English Clamp Multimeter F205
27
Clamp Multimeter F205 English
For the power factor (PF) and the powers VA and var, the measurement is possible
only in AC or AC+DC.
28
English Clamp Multimeter F205
Reminder : the arrow on the jaws of the clamp (see the diagram below) must point
in the presumed direction of flow of the current from the source (producer) to the
load (consumer)
Remark : You can also measure the three-phase power on a balanced 4-wire
network by proceeding in the same way, or by proceeding as for the measurement
on a single-phase network, then multiplying the value found by three.
29
Clamp Multimeter F205 English
30
English Clamp Multimeter F205
31
Clamp Multimeter F205 English
32
English Clamp Multimeter F205
33
Clamp Multimeter F205 English
4 CHARACTERISTICS
Note (1) - The display indicates "+OL" above +2000V and "-OL" below – 2000V in
REL mode
34
English Clamp Multimeter F205
Note (1) - The display indicates "OL" above 1000V (1400V in PEAK mode).
- Above 1000V, a repetitive beep indicates that the voltage being
measured is greater than the safety voltage for which the device is
guaranteed. The display indicates "OL".
- Bandwidth in AC = 3 kHz
Note (2) Any value between zero and the min. threshold of the measurement range
(0.15V) is forced to "----" on the display.
35
Clamp Multimeter F205 English
Note (1) - The display indicates "OL" above 1000V (1400V in PEAK mode).
- Above 1000V (DC or RMS), a repetitive beep indicates that the voltage
being measured is greater than the safety voltage for which the device is
guaranteed.
- Bandwidth in AC = 3 kHz
Note (2) - Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to "----" on the display.
36
English Clamp Multimeter F205
Note (1) -The display indicates “+OL" above 1800A and "-OL" below –1800A in
REL mode. The “-“ and “+” signs are managed (polarity).
Note (2) - The residual current at zero depends on the remanence; It can be
corrected by the “DC zero” function of the HOLD key.
Note (1) - The display indicates "OL" above 900A in PEAK mode. The "-" and "+"
signs are not managed.
- Bandwidth in AC = 3 kHz
Note (2) - Any value between zero and the min. threshold of the measurement
range (0.15V) is forced to “----“ on the display.
- Residual current at zero <150mA.
Note (1) - In DC, the display indicates "+OL" above +1800A and "+-OL" above -
1800A in REL mode. The "-" and "+" signs are managed (polarity).
- In AC and AC+DC, the display indicates "+OL" above 900A in
PEAK mode. The "-" and "+" signs are not managed.
- Bandwidth in AC = 3 kHz
Note (2) - In AC, any value between zero and the min. threshold of the
measurement range (0.15A) is forced to "----" on the display.
- Residual current at zero:
37
Clamp Multimeter F205 English
Note (1) - Above the maximum display value, the display unit indicates "OL".
- The "-" and "+" signs are not managed.
Note : The "-" sign is disabled for the diode test function.
39
Clamp Multimeter F205 English
Note (2) - Any applied voltage greater than 1000V causes the emission of an
intermittent alarm beep to report a dangerous overload.
Note (3) - The measurement result may be perturbed by an instability linked to the
current measurement (approximately 0.1A).
Example: for a power measurement made at 10A, the instability of the
measurement will be 0.1A/10A or 1%.
6.00 kW
Measurement 5 W to 60.0 kW to
to 600 kW (1)
range (2) (4) 5999 W 599.9 kW
59.99 kW
Specified 1 to 100% of the
measurement measurement 0 to 100% of the measurement range
range range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (7)
Resolution 1W 10W 100W 1000W
40
English Clamp Multimeter F205
Note (7) - In balanced three-phases, with deformed signals (THD and harmonics),
uncertainties are guaranted since Ф > 30°. Additionals errors are following ,
depending of THD :
Add +1% for 10% < THD < 20%
Add +3% for 20% < THD < 30%
Add +5% for 30% < THD < 40%
6.00 kW
Measurement 5 W to 60.0 kW to 600 kW to
to
range (2) (4) 5999 W 599.9 kW 900 kW (1)
59.99 kW
Specified
1 to 100% of the
measurement 0 to 100% of the measurement range
measurement range
range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (7)
Resolution 1W 10W 100W 1000W
41
Clamp Multimeter F205 English
42
English Clamp Multimeter F205
6.00 kvar
Measurement 5 var to 60.0 kvar to
to 600 kvar (1)
range (2) (4) 5999 var 599.9 kvar
59,99 kvar
Specified
1 to 100% of the
measurement 0 to 100% of the measurement range
measurement range
range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (8)
Resolution 1 var 10 var 100 var 1000 var
Note (8) - In single phase, with deformed signals (THD and harmonics),
uncertainties are guaranted since Ф > 30°. Additionals errors are following ,
depending of THD :
Add +1% for 10% < THD < 20%
Add +3% for 20% < THD < 30%
Add +5% for 30% < THD < 40%
4.2.17 Measurement of reactive AC+DC power
6.00 kvar 60.0 kvar
Measurement 5 var to 600 kvar to
to to
range (2) (4) 5999 var 900 kvar (1)
59.99 kvar 599.9 kvar
Specified
1 to 100% of the
measurement 0 to 100% of the measurement range
measurement range
range
Uncertainties
± (2% R +10 pt) ± (2% R +3 pt)
(3) (8)
Resolution 1 var 10 var 100 var 1000 var
43
Clamp Multimeter F205 English
Note (1) - If one of the terms in the calculation of the power factor is displayed as
"OL", or forced to zero, the display of the power factor is an indeterminate
value "----".
44
English Clamp Multimeter F205
Note : - The display is "----" if the input signal is too low (U<5V or I<6A) or if the
frequency is less than 5Hz.
46
English Clamp Multimeter F205
Batter: 1 x 9 V LF22
Mean life: >120 hours (without backlighting)
Duration of operation before After 10 minutes without action on the switch
automatic switching off: and/or keys
Mechanical
Free fall: 2m (in accordance with standard IEC-68-2-32)
strength:
Level of protection
IP40 (per standard IEC-60529)
of the housing:
47
Clamp Multimeter F205 English
48
English Clamp Multimeter F205
5 MAINTENANCE
The instrument has no parts that can be replaced by personnel who are not trained
and approved. Any non-approved repair or other work, or replacement of a part by
an "equivalent", may severely compromise safety.
5.1 CLEANING
• Disconnect everything connected to the device and set the switch to OFF.
• Use a soft cloth moistened with soapy water. Rinse with a damp cloth and
dry quickly using a dry cloth or forced air.
• Dry perfectly before putting back into use.
5.4 REPAIR
For all repairs before or after expiry of warranty, please return the device to your
distributor.
49
Clamp Multimeter F205 English
6 WARRANTY
Except as otherwise stipulated, our warranty is valid for three years starting from
the date on which the equipment was sold. Extract from our General Conditions of
Sale provided on request.
The warranty does not apply in the following cases:
• Inappropriate use of the equipment or use with incompatible equipment;
• Modifications made to the equipment without the explicit permission of the
manufacturer’s technical staff;
• Work done on the device by a person not approved by the manufacturer;
• Adaptation to a particular application not anticipated in the definition of the
equipment or not indicated in the user’s manual;
• Damage caused by shocks, falls, or floods.
7 DELIVERY CONDITION
50
04 - 2013
Code : 692884A02 - Ed. 3