Manual - Vortex Flow Meter
Manual - Vortex Flow Meter
VORTEX
FLOW METER
ARTVT
Working principle
LUGB-type 2 vortex flow sensor measuring the flow of liquid of steam,gas and low viscosity
according to Carmen and Strouhal relevant spiral produce and on the theory of the flow
relationship. As shown in picture, In the meter body vertical insert a triangular prism root
namely the happening of the body, when eddies of medium flow through the table body, in
triangular prism behind the alternate produce in opposite directions regular karman swirl,its
spiral separation and the flow of the medium frequency F speed by sensing head is
proportional to the V detected the number of spiral,can measure the flow velocity,again
according tp the table body mouth
2
ARTVT
� F ······ The liquid flow through the vortex triangular prism produce the vortex frequency(unit:Hz)
� Sr ······ Strouhal number(unit:dimensionless)
� V ······ Internal of pipe fluid flow rate(unit:m/s)
� d ······ the width of Vortex meter internal triangular prism(unit:m)
� D ······ Vortex meter inner diameter(unit:m)
� Q ······ Instantaneous volume flow rate(unit:m³ /h)
� K ······ Vortex of meter coefficient(unit:pulse number/m ³ )
� M ······ Instantaneous quality flow rate(unit:kg/h)
� P ······ Fluid density(unit:kg/m ³ )
Working principle
� 5.Power supply:12~24VDC
� 6.EX-proof Class:ExiaIICT6
� 7.Output signal:Voltage pulse low level≤1V high voltage≥ 6V standard current signal 4~20mA
Flow range
Absolute pressure(MPa) 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.2 1.5 2.0
Temperature(℃) 120.23 133.54 143.62 151.84 158.94 164.96 170.41 175.36 179.88 187.96 19.84 212.37
Density(kh/m³) 1.129 1.651 2.163 42.669 3.170 3.667 4.162 4.655 5.147 6.127 7.602 1.005
DN50 standard low limit 52 63 73 81 88 95 101 107 112 122 136 157
DN50 standard high limit 520 630 730 810 880 950 1010 1070 1120 1220 1360 1570
DN65 standard low limit 90 106 121 134 146 158 168 178 187 204 227 261
DN65 standard high limit 900 1060 1210 1340 1460 1580 1680 1780 1870 2040 2270 2610
DN80standard low limit 122 148 170 188 205 221 235 249 261 285 318 365
DN80 standard high limit 1220 1480 1700 1880 2050 2210 2350 2490 2610 2850 3180 3650
DN100 standard low limit 175 212 242 269 293 315 336 355 374 408 454 522
DN100 standard high limit 1750 2120 2420 2690 2930 3150 3360 3550 3740 4080 4540 5220
DN125 standard low limit 262 318 363 404 440 473 504 533 561 612 681 783
DN125 standard high limit 2620 3180 3630 4040 4400 4730 5040 5330 5610 6120 6810 7830
DN150 standard low limit 350 423 484 538 586 631 672 711 747 815 908 1044
DN150 standard high limit 3500 4230 4840 5380 5860 6310 6720 7110 7470 8150 9080 10440
DN200 standard low limit 700 846 969 1076 1173 1261 1344 1421 1494 1630 1815 2088
DN200 standard high limit 7000 8460 9690 10760 11730 12610 13440 14210 14940 16300 18150 20880
DN250 standard low limit 1050 1269 1453 1641 1759 1892 2016 2132 2241 2445 2722 3132
DN250 standard high limit 10500 12690 14530 16410 17590 18920 20160 21320 22410 24450 27220 31320
Kaifeng Qingtianweiye
DN300 standard low limit Flow Instrument
1750 2116 Co.,Ltd
2422 2690 2932 3153 3359 3553 3736 4076 4536 5220
DN300 standard high limit 17500 21160 24220 26900 29320 31530 33590 35530 37360 40760 45360 52200
ARTVT
(3)the quality flow flange of superheated steam
Flange connection type vortex flowmeter external dimension picture(Flange card installed type)
6
ARTVT
Flange card installed type external dimension
Caliber L1 L2 D1 D2 H D3 N Caliber L1 L2 D1 D2 H D3 N
40 75 109 145 110 470 13 4 150 120 170 280 240 594 21 8
50 75 109 160 125 481 17 4 200 150 200 335 295 646 21 12
65 75 117 180 145 497 17 6 250 160 214 405 355 708 21 12
80 90 122 195 160 510 17 6 300 170 224 460 410 760 21 12
7
LUGB
� Insertion type vortex flowmeter are mainly used for various of industry,industrial
pipe large caliber gas,liquid,steam medium flow measurement,its characteristic
is structure simple,no move mechanical parts,low pressure loss,measuring
range rate is wide,range degrees of 10~15,high price-performance ratio.
� Pay attention to the installation;
� Mounting base insert pipe and the insert parts less than pipe inwall
� Base in pipelines should be correct position,not skewed.
� Deburring,wipe off welding slag
� The flange plane of base shoud be parallel with pipe axis
� Ensure that medium flow direction and Flow direction indicating rod consistent,
forbidden to pull the Flow direcitiong indicating rod.
Assenbling conditions
8
LUGB
� Near the Installation liquid sensor of pipeline should be filled with measuredliquid
� Sensor should avoid installed in pipeline which has strong mechanical vibration
� The inner diameter of the straight pipe section consistnet with sensor caliber
as far as possibl,if not consistent,it should use the pipeline which is slightly
bigger than sensor caliber
� When measuring medium which is need to corrected for temperature and
pressure,it should be take after the sensor 3 ~ 5 DN distance places as pressure
point, and take 5 ~ 8 DN distance places as temperature point
� The measured medium contains many impurities,it should be add install filters
which is in outside of the required length of the sensor straight pipe section upstream
� Sensor should avoid installed in the place which has a strong electromagnetic
interference,small spaces,and maintenance convenient..