Software Nvgate Oros
Software Nvgate Oros
TECHNICAL
®
NVGate SPECIFICATIONS
M002-012-7 1
NVGate® TECHNICAL SPECIFICATIONS
Table of Content
Table of Content ................................................................................................................................. 2
General description ................................................................................................................................. 3
Modules .............................................................................................................................................. 3
Signal sources ......................................................................................................................................... 5
Front-end ............................................................................................................................................ 5
Recorder / Player................................................................................................................................ 7
Recorder ............................................................................................................................................. 7
Player ................................................................................................................................................. 8
Plug-ins analyzers ................................................................................................................................. 10
Time Domain Analysis - TDA ........................................................................................................... 10
Fast Fourier Transform analysis - FFT ............................................................................................. 11
Constant Band Order Tracking Add-on (CBT) ................................................................................. 12
Diagnostic Add-on (FFTDiag)........................................................................................................... 13
Synchronous Order Analysis (SOA) ................................................................................................. 13
1/n octave constant percentage band filter analysis - OCT ............................................................. 14
Overall acoustics sound level meter - OVA ...................................................................................... 15
Options .................................................................................................................................................. 16
Instantaneous angular Velocity Converter (IVC) .............................................................................. 16
Combined tachometer editor (CTE) ................................................................................................. 16
Virtual inputs (VIn) ............................................................................................................................ 17
Virtual parameter (VDC) ................................................................................................................... 18
General purpose analysis ...................................................................................................................... 19
Monitor .............................................................................................................................................. 19
Waterfall ........................................................................................................................................... 19
Shared ressources ................................................................................................................................ 21
Triggering: ........................................................................................................................................ 21
Output signals:.................................................................................................................................. 22
Filters: ............................................................................................................................................... 23
Notes ..................................................................................................................................................... 24
M002-012-7 2
NVGate® TECHNICAL SPECIFICATIONS
General description
® ®
The following specifications concern NVGate software for OR34, OR35, OR36, OR38 and Mobi-Pack multi-
analyzer instruments. These specifications apply for on-line analysis and post analysis with or without the
instrument connected to the PC. The detailed features of analyzer hardware and metrological information are
available in OR3X multi-analyzer/recorder instruments specifications documents.
Modules
NVGate® software is based on a general-purpose platform where optional analysis modules called plug-in
analyzers are added.
The FFT plug-in features specific add-on for specialized analysis.
®
NVGate Platform
NVGate® platform provides a comprehensive set of tools for noise and vibration acquisition, recording and
analysis.
• Signal sources that condition, provide and store time domain signals.
• Shared resources that apply simultaneously or independently on distributed signal to plug-in analyzers.
• Operational tools that increase measurement efficiency and reliability.
• General-purpose analysis for monitoring and result tracking.
Masks Mask editor for spectra, profiles, 1/n octave, order spectra
Operational tools Report On-line report editor, automated report generation, word
files.
M002-012-7 3
NVGate® TECHNICAL SPECIFICATIONS
®
NVGate Plug-ins analyzers
®
In addition to the standard functions featured in the software platform, NVGate can receive additional plug-in
analyzers that offer flexible configuration of independent analysis modes.
®
NVGate features the following Plug-ins analyzers:
®
NVGate Options
Beside the plug-ins analyzers and their add-on, the 3-Series analyzers software features options:
PC requirements
Pentium 4/ 2 GHz / 2561 MB RAM with Windows XP or 5121 MB with Windows Vista/ Graphics video with at
Minimum least 32 MB dedicated (not shared) memory / 100 MB free on HD + storage for measurements and signals,
CD ROM drive, 1024 x 768 display (XGA), DirectX 8.0
Intel Core 2 Duo / 2 GHz / 1 GB of RAM with XP/7, 2 GB of RAM with Vista / Graphics video with 256 MB
Recommended
dedicated (not shared) memory / 100 MB free on HD + storage for measurements and signals, CD or DVD
(for laptop)
ROM drive, 1400 x 1024 display (SXGA+), DirectX 10
Intel Core 2 Duo / 2.6 GHz or AMD Athlon 64 X2 Dual-Core 6000+/ 3 GB of RAM / Graphics video with 512
Recommended
MB dedicated (not shared) memory / 100 MB free on HD + storage for measurements and signals, CD or
(for desktop)
DVD ROM drive, 1600 x 1200 display (UXGA), DirectX 10
Type: Ethernet 100base TX, 100 Mbit/s - Connector: RJ45
Connection
For removable disk: USB 2.0 - At least one USB port for dongle key.
Operating systems Windows XP Pro Service Pack 3 (recommended), Windows Vista Business Service Pack 2, Windows 7
1) Waterfall depth depends on available memory. Minimum configuration does not allow waterfall storage.
M002-012-7 4
NVGate® TECHNICAL SPECIFICATIONS
Signal sources
NVGate®/OROS 3-Series multi-analyzers platform can process signal coming from different sources. These
sources are detailed in this section as modules.
Front-end
The front-end module gathers the different input and output settings available in the instrument unit. In addition,
the front-end module generates virtual signals (synthesized) when running the office mode (no unit connected to
the PC) which is useful for set-up.
Front-end settings
Sampling clock 102.4 kS/s to 3,200 S/s or 65.536 kS/s to 2,048 S/s
LEMO1 polarization 200 V on/off per block of 8 dynamic input
2,3 AC (0.35 Hz) - DC - ICP - ICP + TEDS - AC float - DC float -
Coupling
Grounded
Label - component - node - direction (± X ,Y,Z) - type
Node information2
(translation, rotation)
Any physical quantity can independently be associated to
Physical quantity2,3
Dynamic inputs inputs.
Sensitivity2,3 User defined in V/unit.
Pk to pk expressed in the associated unit, (taking in account
Range2
sensitivity) - linear or dB
External conditioner compensation2 Gain - polarity - offset.
2 Any NVGate filter can independently be applied to each
Filter
dynamic input (See Filters §)
Auto-range Enable/disable auto-ranging independently on any input
Sampling over-sampled 64 time the dynamic inputs sampling
Coupling2 AC (0.35 Hz) - DC
Information2 Label
Physical quantity2,3 Any physical quantity can be associated to inputs.
Sensitivity2,4 User defined in V/unit.
Pk to pk expressed in the associated unit, (taking in account
Range2
sensitivity) - linear or dB
External conditioner compensation2 Gain, offset.
Threshold (in associated unit) - Edge (rise, fall) - Hystersis -
Edge detection
hold off (sec)
Ext. sync inputs Pre-divider Hardware - 1 to 1024 pulses – accept up to 375 kpulses/sec
1 to 50 pulses/pulse - Generate higher frequencies on low
Post-multiplier
rate triggers. (ex. GPS)
Pulse per rev (0 to 1024) - average (n rev) - min (detect
Tachometer mode stopped shaft) and max RPM speed - hold off (in % of
revolution) - rotation (clockwise/counterclockwise)
Pulse per rev (1 to 1024) - min (detect stopped shaft) and
1 max RPM speed - sliding average (1 to 20 samples) -
Torsional mode : Instantaneous angular
Missing pulse number (0 to 3), Missing pulses detection
velocity converter
(delay of 0 to ±20% of previous pulses duration) - Filters (dt,
dt², 1/dt)
Sampling Same as dynamic inputs
2 Any outputs signals, see §outputs - play-back of signal file
Generated signals
Outputs tracks - DC levels - Inputs playback (Delay > 256 samples)
Any NVGate filter can independently be applied to each
Filters2
dynamic output (See Filters §)
1
Optional features
2
Independent for each input
3
Linked with the transducer database
M002-012-7 5
NVGate® TECHNICAL SPECIFICATIONS
Front-end (continued)
Ext. Sync inputs To any: start or stop averaging of plug-in analyzers, recorder
start or stop recording, waterfall start or stop acquisition,
Connection2
new block trigger for FFT and SOA and new slice for
waterfall acquisition.
Profiles versus time (160 ms to 163 sec user selectable), and
Monitoring
Parametric1 (DC) digital/analog view meter.
inputs To any: recorder track, waterfall reference (Z/X axis) and
Connection2
profiles, level and delta level detectors.
1
Optional features
2
Independent for each input or output
3
Comes automatically from transducer database
4
Available in the remote controller display
5
Not supported on OR34 & OR35
M002-012-7 6
NVGate® TECHNICAL SPECIFICATIONS
Recorder / Player
OROS 3-Series/NVGate instruments feature a recorder and a player module that allows users to:
• Record time domain signal, trigger and parametric input.
• Export or import time domain files.
• Playback time domain files on analyzers outputs during analysis
• Listen recorded tracks on the PC speakers.
• Post-analyze time domain files in the available plugs-in.
Recorder
TM
The recorder module memorizes time domain signal into files located on 3-Series analyzer HD, Mobi-Disc or
PC HD. This is available on 2 different modes:
• On-line record, the chosen inputs (Dynamic, ext. sync, parametric) are recorded at selected sampling
frequency during acquisition. Real-time analysis is available simultaneously.
• Time and tracks split, the chosen recorded tracks (from signal file) are played back in the recorder
using the post-analysis mode. The sampling frequency, duration, and/or track arrangement can be
changed.
Recorder settings
5
For OR34, OR35 and Office-8 licenses.
6
For OR36, OR38 and Office licenses.
M002-012-7 7
NVGate® TECHNICAL SPECIFICATIONS
Management Upload Load PC HDD records on the analyzer HDD - Rate 1.4 MB/s
Location All records accessible from the project manager tree
Hard drive management Format HDD - Selective delete - Sorting tools
TM
Mobi-Disk
OR36 and OR38 multi analyzers/recorders feature a removable hard drive with 2 ports:
• Parallel High speed port for direct data throughput to the Mobi-DiskTM.
• USB 2.0 port for post processing and data management without the instrument.
Recorder monitoring
The following displays are available during real-time or post-analysis.
Player
The player module plays the recorded signal files. This is achieved following 2 different ways:
• Post analysis, the player tracks take place of the corresponding inputs (Dynamics, Ext. Sync and
parametric) to be processed by the plug-in analyzers, recorder, trigger and tachometer
• Playbacks, the recorded dynamic input are generated on the instrument outputs simultaneously with
standard analysis of the inputs.
M002-012-7 8
NVGate® TECHNICAL SPECIFICATIONS
Player settings
NVGate® recorded files - Imported files - located on
File selection
instrument HDD, Mobi-DsicTM or on PC HDD.
Record selection Record number selection (for multi record files)
User selectable Start and stop offsets (in sec.) - available in
Played section Section definition
the file overview window (marker and slider)
Duration Play backed duration (User information)
On/off - repeat continuously the selection (available only for
Repeat
playback on analyzer output)
Swap loaded files with same tracks number - applies
File swap
identical post-analysis setup to multiple files.
Mode Playback Continuous or step by step (5 ms to 360,000 s)
All tracks at lowest frequency (compatible with all plugs-in) or
Bandwidth
initial record frequencies (no track mix in plugs-in)
Number 38 tracks (32 + 6 ext. Sync) for all licenses.
Dynamic input record (2 sampling frequencies), parametric
Type
(DC) input record and ext. sync record.
Label, Coupling, external gain, input range, sampling
Fixed setting (information) / track
frequency and signal bandwidth
Tracks
Modifiable measurement point
information settings (apply for post Component - node number - direction - type
analysis or re-recording) / track
Modifiable settings (apply for post Associated transducer - unit - sensitivity - offset
analysis or re-recording) / track compensation
Analyzer HD - Mobi-DiscTM connected with USB or inside the
File location
Listening analyzer - PC HD
Playback One track - from cursor position - repeat displayed signal
Display Recorded markers are available in the file overview
Markers
Use Set the start and stop playback offset (user selectable)
Player connections
The following table describes the available connections to the analysis modes (plug-in analyzers and recorder)
during post-analysis operation:
7
Independently on each track.
M002-012-7 9
NVGate® TECHNICAL SPECIFICATIONS
Plug-ins analyzers
®
The following plug-in analyzers are available as options of NVGate software platform.
TDA Settings
800 mHz to 40 kHz
Oscillator 1
(sampling oscillator 102.4 kS/s to 2,048 S/s)
Bandwidths
512 mHz to 25.6 kHz
Oscillator 2
(sampling oscillator 65.536 kS/s to 3,277 S/s)
7200/ABW8 sec to 100 000 sec (27 hours) - Independent on
Depth
Time views each channel
Time base (resolution) Depth/2048- Independent on each channel
Time span for Continuous sliding with refresh each 256 samples (Exponential)
Type
extraction - One shot (Linear) - Repeated shot (Repeated linear)
(averaging) Duration 1/ABW to 2 106/ABW (ex: 50 ms to 1000 sec @ 2 kHz)
OR36 & OR38 analyzers 32 channels per plug-in for on line and post-analysis
Channels
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
Any NVGate filter can independently be applied to each dynamic
Filters On each channel
input (See Filters §)
Start analysis Any NVGate® event plus manual and free run
Triggering
End analysis Any NVGate® event plus manual and free run
Repeat analysis On new start or end of averaging
Trigger delay Positive = unlimited - negative = 32k Samples
TDA Results
The following results are available for real time or post-analysis display, report and saving.
Time Views Overview Signal envelope: Min/Max line for each time step -
DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor -
Statistical values Skewness - Kurtosys - computed on a user defined area of the
time view
DC - RMS - Min- Max - Peak - Peak to Peak - Crest factor -
Extractions Statistical values Kurtosys - computed on the Time span – All values available for
Viewmeter and/or Waterfall profiles
8
ABW: the plug-in Analysis Bandwidth.
M002-012-7 10
NVGate® TECHNICAL SPECIFICATIONS
FFT Settings
800 mHz to 40 kHz
Oscillator 1
(sampling oscillator 102.4 kS/s to 2,048 S/s)
Bandwidths
512 mHz to 25.6 kHz
Oscillator 2
(sampling oscillator 65.536 kS/s to 3,277 S/s)
Lines 101, 201, 401, 801, 1601, 3201 and 6401 lines
Resolution 80 µHz (512 mHz/6401 lines)
Frequency resolution
to 400 Hz (40 kHz/101 lines)
Spectral (power) - time (with phase) - FDSA (Synchronized with
Domain
one frequency)
Exponential - linear - repeated linear - referenced peak hold
Averaging Type (eq. to tracking filter locked on reference channel frequency) -
peak hold
Size Duration or number of blocks, unlimited size
Overlap 0% to 99.9% depending on block size - retrigger
OR36 & OR38 analyzers 32 channels per plug-in for on line and post-analysis
Channels
OR34 & OR35 analyzers 8 channels per plug-in for on line and post-analysis
General Uniform - Hanning - Hamming - Kaiser Bessel
Weighting Modal acquisition User defined Force (rect.) and Response (Rect. + exp)
Assignment Independently on any channels
Any NVGate filter can independently be applied to each dynamic
Filters On each channel
input (See Filters §)
Cross functions Reference selection Full or partial matrix available (all channels can be references)
Magnification factor 2 to 128 by step power of 2
Zoom
Min., max and center frequencies Graphically selectable on wide band spectrum
Triggering Start averaging Any NVGate® event plus manual and free run
End averaging Any NVGate® event plus manual and free run
New block Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
Positive = unlimited - negative = 32k Samples
Trigger delay
M002-012-7 11
NVGate® TECHNICAL SPECIFICATIONS
FFT Results
The following results are available for real time or post-analysis display, report and saving.
CBT Settings
Number 8 per channels
Max 0.001 to 800
Tracked Order
User selectable per channel - minimum depend on weighting
Constant bandwidth
windows
Associated tachometer Any front end, recorded or virtual tachometer
Computation Peak tracking On/Off center bandwidth on nearest peak
Order amplitude Weighting windows eq. noise bandwidth correction
CBT Results
The following results are available for real time or post-analysis display, report and saving.
9
These result cannot be saved
M002-012-7 12
NVGate® TECHNICAL SPECIFICATIONS
FFTDiag Settings
Activation Enable/disable on all channels - requires zoom activation
Envelop
Averaging Spectral domain - FDSA
demodulation
Bandwidth ½ of zoom span - ¼ of FFT bandwidth
Enable/disable on all channels - All other FFT results are
Activation
Correlation affected
Weighting window Uniform - Left zero padding - Centred zero padding
FFTDiag Results
The following results are available for real time or post-analysis display, report and saving.
SOA Settings
M002-012-7 13
NVGate® TECHNICAL SPECIFICATIONS
SOA Results
The following results are available for real time or post-analysis display, report and saving.
OCT Settings
Sampling oscillator 1 Center of highest 1/3rd band = 40 kHz
(102.4 kS/s to 2,048 S/s) Center of lowest 1/3rd band = 100 mHz
Bandwidths Sampling oscillator 2 Center of highest 1/3rd band = 25 kHz
(65.536 kS/s to 3,277 S/s) Center of lowest 1/3rd band = 100 mHz
Frequency span Highest Band / lowest band < 2000 (ex. 10Hz - 20kHz)
Resolution 1/n Octave 1, 1/3 rd, 1/12th, 1/24th
Linear & repeated linear (20 ms to 60,000 s) exponential (20
Basic
ms to 60s)
Averaging
Acoustics Short LEQ 1 s, Short LEQ 1/8 s, Fast, Slow, Impulse
Constant Bandwidth * Time B*T = 0.2dB, 0.5 dB, 1 dB, 2 dB
Detectors Complies with IEC 651, IEC 804, IEC 61-672 class 1
Standards
CPB filters Complies with IEC 1260 class 1
M002-012-7 14
NVGate® TECHNICAL SPECIFICATIONS
OCT Results
The following results are available for real time or post-analysis display, report and saving.
OVA Settings
Standard IEC 61-672-1 - IEC 651 - IEC 804
Type Class 1
Bandwidths 10 Hz to 40 kHz - Adjustable
OR36 & OR38 analyzers 32 channels for on line and post-analysis
Channels
OR34 & OR35 analyzers 8 channels for on line and post-analysis
Linear User selectable duration - repeat mode
Averaging
Short leq 1/8 s - 1 s - User selectable duration
Start averaging Any NVGate® event plus manual and free run
Triggering End averaging Any NVGate® event plus manual and free run
Repeat averaging On new start or end of averaging
1 peak detector/Ch. - A, C or Z (none) weighting -Independent
Peak
for each channel
Detectors 3 overall detectors/Ch. - A, C or Z (none) weighting - Fast,
Weighting Slow, Impulse and linear time weighting - Independent for
each channel
OVA Results
The following results are available for real time or post-analysis display, report and saving. The following levels
can be tracked in profiles up to 100 000 pt each.
M002-012-7 15
NVGate® TECHNICAL SPECIFICATIONS
Options
IVC Results
The following results are available for real time or post-analysis.
M002-012-7 16
NVGate® TECHNICAL SPECIFICATIONS
CTE Settings
Number 2
Sources
Type Any NVGate tachometer – excepted another CTE
Type New item in the active tachometers list
Output
Synchronization 0° phase shift with sources and analyzed signal
Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Editor Predefined Rmp1, Rpm2 (sources speed), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg
Rotation Clock wise or counter clock wise
CTE Results
The following results are available for real time or post-analysis
Number 4
Signals Use Same as on-line Tachometer or recorded tachometer
Magnitude Angular velocity (RPM)
View Types View meter - 1 to 100 s rolling profile
VIn Settings
Number 12 Operators
Type New item in the active inputs/tracks list
Output
Synchronization 0° phase shift with sources and analyzed signal
Magnitude Automatic or voltage if unresolved
Number 1 to 12 channels
Sources
Type Analyzer dynamic inputs or recorded dynamic inputs
Type (Input * Coeff + Offset ) ^power - independent on each output
M002-012-7 17
NVGate® TECHNICAL SPECIFICATIONS
VDC Settings
Number 12 Operators
Type New item in the DC inputs/track list
Output
Synchronization 0° phase shift with sources and analyzed signal
Magnitude User defined
Number 1 to 12 channels
Sources
Type Analyzer DC inputs or recorded DC inputs
Type Text
Variables Real values
Operators +, -, x, /, ^, = (affectation)
Editor Predefined Ch x (Channels signals), Pi, e
Trigonometric functions Sine, Cosine, Tangent - with Arc and Hyperbolic combination
Logarithmic functions Log2, Log10, Ln, Exp
Miscellaneous Sqrt, Abs, Min, Max, Sum, Avg
M002-012-7 18
NVGate® TECHNICAL SPECIFICATIONS
Monitor
The monitor is an independent (dedicated processor) module that processes continuously FFT analysis on 4
channels at the maximum available bandwidth. The aim of this module is to monitor in both domains (time and
spectral) 4 inputs and to compute basic indicators for monitoring and triggering purpose.
Monitor settings
Bandwidth ABW = Sampling freq / 2.56
Resolution 401 lines
Average domain Spectral (power)
Fixed setup Overlap 0%
Average type Exponential
Weighting windows Hanning
Trigger Free run
Number 4
Channels
Swap Between any active dynamic input - hot swap capable
Average Duration Instantaneous (20 ms) to unlimited
Pass band filter Butterworth order 2 to 10 - IIR type - Bypass function
Upper & lower frequency User defined - can be graphically modified
Overall analysis DC - Max - Min - RMS - Skew - Kurtosis - Into defined pass
Computed indicators
band - overall if bypass is on
Average User defined duration - independent from spectral averaging
Waterfall
The waterfall module operates as a stack for plug-in analyzers results. Waterfall module features advanced
graphics for 3D and profiles, including extraction tools. It can synchronize results coming from multiple plug-in
analyzers and sources in one result.
Waterfall settings
Continuous Circular buffer of results
Mode
One shot Fill stack and stop waterfall acquisition
Start acquisition Any NVGate® event plus manual and free run
Stop acquisition Any NVGate® event plus manual and free run
Triggering
Any NVGate® event - manual - free run - availability of
New slice (point or spectra)
connected results - periodic
M002-012-7 19
NVGate® TECHNICAL SPECIFICATIONS
Waterfall results
The following results are available for real time or post-analysis display, report and saving. All stacked result can
be saved.
Waterfall Tools
The following results are available for real time or post-analysis display, report and saving.
YZ Sections (Profiles vs. Ref) Any order/frequency - Power - Peak - on selected bandwidth
Any position in the current ref. - Min - Max - Average - on
XY Sections (Spectra)
selected range or all slices
Sections
Order/Frequency extraction (profiles Any order/frequency extraction - user selectable tach. - Power -
vs. Ref.) Peak - on selectable bandwidth - Max order
Number of sections Unlimited - available on result or real-time waterfalls
General Single or dual cursor in each view - linked with other graphs
Linked by value Linked cursors track the same X value in different windows.
Linked cursors
Linked cursors track the same acquisition slice in different
Linked by reference
waterfall windows with different X or Z-axis.
References are: Time - Slice number - DC channels (expressed in physical quantity) - TDA scalars - Monitor
levels- any Tachometer
M002-012-7 20
NVGate® TECHNICAL SPECIFICATIONS
Shared ressources
The following resources are available for each plug-in analyzer (when it is compatible). For example once a
trigger is activated it can be applied to FFTx and SOA: Then modifying the trigger setting in the resources will
apply on all corresponding plug-in.
Triggering:
Each plug in analyzer (and recorder) can be started, stopped or triggered (new block) with events defined in the
Event module.
Event settings
Source Any dynamic input - Any recorded dynamic input
String for event identification (Ex. “Impact” for a hammer impact
Label
detection)
Between min and max range - use source unit (ex. g for an
Threshold
accelerometer)
M002-012-7 21
NVGate® TECHNICAL SPECIFICATIONS
Output signals:
®
NVGate proposes a set of generator signals.
Output signals
The following output signals are available as standard and can be independently applied to the front-end outputs.
Sine Sweep speed Linear: 300 mHz/s to 20 kH/s - Log: 5 mOct/s to 330 Oct/s
Step control Synchronized with analysis end - free run
-15 dB to +60 dB - Independent for each output
Gain control - Amplitude variation controlled by settling time (1 ms to 1000
s)
Offset ±360° - Independent for each output - phase variation
Phase control
controlled by phase speed (1.5°/s to 360°/s)
2 to 6 sine generated simultaneously - with independent
Number
phases and amplitudes - synchronized frequencies
Frequency span From 125 mHz to 40 kHz
Amplitude 0 to 2.5 Vrms - Settling time 100 µs to 10 s
Resolution 125 mHz to 400 Hz - 101 lines to 6401 lines
Multi-sine
Phase Fixed (all sine have same phase) - Random
Burst 0 to 100% - Step 1%
Number 2 to 4 independents multi-sine
M002-012-7 22
NVGate® TECHNICAL SPECIFICATIONS
Filters:
The following filters apply on front-end inputs, plug-ins analyzers and output generators.
10
ABW : the analysis bandwidth, of the plug-in or front end where the filter is in use.
M002-012-7 23
NVGate® TECHNICAL SPECIFICATIONS
Notes
The above specifications describe all the guaranteed capacities and performances of the software installed in a
PC running Windows XP Pro SP3 (recommended configuration). Functionalities may change depending on mode
operation (connected to a 3-Series unit or office). Plug-in analyzers, options and channel number availability
depend on purchased options.
The instruments hardware are described separately in the “Instruments specifications” data sheet for OR34 &
OR35 and OR36 & OR38.
M002-012-7 24
NVGate® TECHNICAL SPECIFICATIONS
Our Philosophy
Reliability and efficiency are our ambition everyday. We know you require the same for your measurement instruments:
comprehensive solutions providing performance and assurance, designed to fit the challenges of your demanding world.
Our Emphasis
Continuously paying attention to your needs, OROS collaborates with a network of proven scientific affiliates to offer the
latest of the technology, always based on innovation.
Worldwide Presence
OROS products are marketed in more than 35 countries, through our authorized network of representatives, offices and
accredited maintenance centers.
M002-012-7 25