Instruction Manual

P/N 20001741, Rev. B


October 2004

Modbus Mapping
®

Assignments for
Micro Motion Transmitters
®

Instruction Manual

Micro Motion
TM
Modbus Mapping ®

Assignments for
Micro Motion Transmitters ®

Instruction Manual

For technical support, phone the support center nearest


you:
• In U.S.A., phone 1-800-522-MASS (1-800-522-6277)
• In Canada and Latin America, phone (303) 527-5200
• In Asia, phone (65) 6770-8155
• In the U.K., phone 0800 - 966 180 (toll-free)
• Outside the U.K., phone +31 (0) 318 495 670

©2004, Micro Motion, Inc. All rights reserved. Micro Motion is a registered trademark
of Micro Motion, Inc. The Micro Motion and Emerson logos are trademarks of
Emerson Electric Co. All other trademarks are property of their respective owners.
Contents

Chapter 1 Using the Modbus Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1


1.1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Transmitters that support Modbus protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3.1 Register addresses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3.2 Register types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3.3 Transmitter codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3.4 LF-Series transmitters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.5 Organization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 Customer service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Chapter 2 Modbus Mapping Assignments. . . . . . . . . . . . . . . . . . . . . . . . . . . . 5


2.1 Coil registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Table 2-1 Coils . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Floating-point register pairs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Table 2-2 Floating-point register pairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3 Integer registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Table 2-3 Integer registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4 Long integer registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Table 2-4 Long integer registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.5 ASCII registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Table 2-5 ASCII registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.6 Integer codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Table 2-6 Sensor information codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Table 2-7 Outputs option board codes (register 1138) . . . . . . . . . . . . . . . . . . . . . . 34
Table 2-8 Digital communication codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table 2-9 Byte order codes (register 0521) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table 2-10 100 Hz variable codes (register 1164) . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Table 2-11 Burst output codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Table 2-12 Quaternary variable codes (register 0015). . . . . . . . . . . . . . . . . . . . . . . 37
Table 2-13 Polling codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Table 2-14 Channel/terminal configuration codes . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Table 2-15 mA output process variable assignment codes
(registers 0012, 0013) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Table 2-16 Frequency/pulse input/output codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Table 2-17 Fault output codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Table 2-18 Control output assignment codes (register 0015). . . . . . . . . . . . . . . . . . 41
Table 2-19 Discrete output assignment codes (registers 1151, 1153, 1155) . . . . . . 41
Table 2-20 Discrete input assignment codes – Series 1000/2000 transmitters
(register 1176) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Table 2-21 Triggering conditions codes – Series 3000 transmitters . . . . . . . . . . . . 42
Table 2-22 Fieldbus AI function block codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Table 2-23 Fieldbus integrator function block codes (register 1511) . . . . . . . . . . . . 44
Table 2-24 Discrete output fixed state codes (registers 1182, 1183, 1184) . . . . . . . 44
Table 2-25 Mass flow codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Table 2-26 Volume flow codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Modbus Mapping Assignments for Micro Motion Transmitters i


Contents continued

Table 2-27 Density codes (register 0040) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46


Table 2-28 Temperature codes (register 0041) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Table 2-29 Pressure codes (register 0044) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Table 2-30 Flow direction codes (register 0017) . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Table 2-31 Special measurement unit codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Table 2-32 Display variable codes (registers 1117–1131, 1362–1366) . . . . . . . . . . 48
Table 2-33 API reference table codes (register 0351) . . . . . . . . . . . . . . . . . . . . . . . 49
Table 2-34 Batch/fill flow source codes (register 1251) . . . . . . . . . . . . . . . . . . . . . . 50
Table 2-35 Enhanced density application codes . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Table 2-36 Event codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Table 2-37 TBR event action codes (register 1706) . . . . . . . . . . . . . . . . . . . . . . . . . 51
Table 2-38 Process variable codes for slot addresses (registers 0751–0782). . . . . 51
Table 2-39 Pressure input receiving method codes (register 0302). . . . . . . . . . . . . 53
Table 2-40 Series 3000 display password codes (registers 1351–1358) . . . . . . . . . 53
Table 2-41 Language codes (register 1359) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Table 2-42 Printer type codes (register 1441) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Table 2-43 Print request codes (register 1442) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Table 2-44 Meter fingerprinting process variable index codes (register 0626). . . . . 54

ii Modbus Mapping Assignments for Micro Motion Transmitters


Chapter 1

Using the Modbus Map


Using the Modbus Map

1.1 About this manual


This manual lists Modbus® mapping assignments for Micro Motion® transmitters that support
Modbus protocol.
This manual does not explain how individual registers are used. For detailed information about using
the registers that are listed in this document, see Using Modbus Protocol with Micro Motion
Transmitters, available on the CD that is shipped with the transmitter, or on the Micro Motion web

Modbus Map
site (www.micromotion.com).
This manual does not explain terminology and procedures for using Modbus protocol, or how to use a
host controller to communicate with other devices in a Modbus-compatible multidrop network. For
detailed information about using Modbus protocol, see www.modbus.org.
This manual does not explain transmitter installation or flowmeter wiring. For information about
installation and wiring, see the appropriate sensor and transmitter manuals.

1.2 Transmitters that support Modbus protocol


Micro Motion transmitters that support Modbus protocol include:
• RFT9739 transmitters, Version 2 and higher
• Micro Motion MVD™ Direct Connect™, with or without the MVD Direct Connect I.S. barrier
• Series 1000 and 2000 transmitters, all versions
• LF-Series transmitters, all versions
• Series 3000 MVD (4-wire) transmitters
Note: MVD Direct Connect installations do not include a transmitter. However, a Modbus/RS-485
connection can be made to the core processor component, and Modbus commands can be used to
communicate with the core processor. See the MVD Direct Connect flowmeter installation manual for
information on setting up this connection.
In this manual, the term “MVD” refers to all of the above except the RFT9739 transmitters.

1.3 How to use this manual


This manual is designed for use as a reference manual. It lists the Modbus registers for transmitter
configuration, calibration, use, and diagnostic procedures. Some procedures are required; others are
optional.

Modbus Mapping Assignments for Micro Motion Transmitters 1


Using the Modbus Map continued

This manual also lists the Modbus registers for the following special applications:
• Petroleum measurement application (API feature)
• Enhanced density application
• Discrete batch application and filling and dosing application
• Net Oil Computer (NOC) application
• Custody transfer application
For each register, checkmarks indicate the transmitters for which the register is enabled.
In addition, this manual lists the integer codes that are used in configuration.

1.3.1 Register addresses


Each register is identified by its address (or starting address), which is a 4-digit number.
Depending on the PLC that will be used to communicate with the transmitter, you may need to
subtract 1 from the address or starting address of the register. Refer to your PLC documentation to
know if this applies to you.

Example This Modbus manual specifies 0042 as the address of the register that
contains the unit for process variables that measure volume flow.
Use 0041 in the data field of the Modbus frame.

1.3.2 Register types


The register types are listed and briefly described in Table 1-1.
Note: Micro Motion terminology is a simplified version of Modbus terminology. The term “coil” is
used for both Modbus coils (read-write) and Modbus discrete inputs (read-only), and the term
“register” is used for both Modbus holding registers (read-write) and Modbus input registers (read-
only).

Table 1-1 Modbus register types

Register type Size Data type Description


Coil 1 bit Binary Boolean (0 or 1, ON or OFF)
Float 16 bit Floating-point Used in pairs to store 32-bit floating-point values in single precision
IEEE 754 format
Integer 16 bit Integer Unsigned 16-bit integers ranging from 0 to 65535
Long integer 16 bit Integer Used in pairs to store unsigned 32-bit integers ranging from 0 to
4294967295 (0xFFFFFFFF)
ASCII 16 bit 8-bit ASCII One or more consecutive registers. Each register stores two 8-bit ASCII
characters (16 bits total).

1.3.3 Transmitter codes


If a register applies only to a subset of the model or series, codes are used to identify the models to
which the register applies. Codes are listed in Table 1-2.

2 Modbus Mapping Assignments for Micro Motion Transmitters


Using the Modbus Map continued

Table 1-2 Transmitter codes

Using the Modbus Map


Code Transmitter
A All transmitters with the analog outputs option board
I All transmitters with the intrinsically safe outputs option board
C All transmitters with the configurable input/outputs option board
F All transmitters with the FOUNDATION Fieldbus outputs option board
P All transmitters with the Profibus-PA outputs option board
N Net Oil Computer (Series 3000 transmitters with the NOC application)

1.3.4 LF-Series transmitters


The LF-Series transmitters are based on Series 1000 and Series 2000 transmitters, and use the same
Modbus interface. For this reason, they are not called out separately in this manual. All references to
Series 1000 or Series 2000 transmitter also apply to the equivalent LF-Series transmitter:
• Model 1500 includes the LF-Series: DIN Rail Flow Only

Modbus Map
• Model 1700 includes the LF-Series: Field Mount Flow Only
• Model 2500 includes the LF-Series: DIN Rail Multivariable
• Model 2700 A includes the LF-Series: Field Mount Multivariable 1mA/1FO
• Model 2700 C includes the LF-Series: Field Mount Multivariable 2mA/1FO configurable
• Model 2700 F includes the LF-Series: Field Mount FOUNDATION fieldbus
• Model 2700 P includes the LF-Series: Field Mount Profibus-PA

1.3.5 Organization
The reference section of this manual is organized as follows:
• All coils, listed numerically by address (Table 2-1)
• All floating-point registers, listed numerically by address (Table 2-2)
• All integer registers, listed numerically by address (Table 2-3)
• All long integer registers, listed numerically by address (Table 2-4)
• All ASCII registers, listed numerically by address (Table 2-5)
• Integer codes, listed by function (Tables 2-6 through 2-44)
Additional information for a specific register or transmitter is provided in a footnote. For each table,
the page on which the footnotes can be found is listed at the top of the page.

1.4 Customer service


For customer service, or to order manuals, contact Micro Motion Customer Service. Contact
information is provided on the title page of this manual.

Modbus Mapping Assignments for Micro Motion Transmitters 3


4 Modbus Mapping Assignments for Micro Motion Transmitters
Chapter 2

Using the Modbus Map


Modbus Mapping Assignments

2.1 Coil registers


In Table 2-1:
• RW = Read/write
- ON (1) = Start, Reset, Apply, Enable, Resume
- OFF (0) = Stop, Abort, Disable, No action taken

Modbus Map
• RO = Read-only
- ON (1) = Condition active
- OFF (0) = Condition inactive

Table 2-1 Coils Transmitter codes on page 3, footnotes on page 9

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Address RO/RW Description


0002 RW Start/stop totalizers √ √ √ √ √ √ √
0003 RW Reset all totalizers √ √ √ √ √ √ √
0004 RW Reset all inventories √ √ √ √ √ √ √
0005 RW Perform flowmeter zeroing √ √ √ √ √ √ √
0006 RW Trim primary mA output at 0 or 4 mA √ √ √ √ AIC √
0007 RW Trim primary mA output at 20 mA √ √ √ √ AIC √
0008 RW Trim secondary mA output at 0 or 4 mA √ √ IC √
0009 RW Trim secondary mA output at 20 mA √ √ IC √
0010 RW Fix/unfix current level from primary mA output √ √ √ √ AIC √
0011 RW Fix/unfix current level from secondary mA output √ √ IC √
0012 RW Fix/unfix frequency from frequency/pulse output √ √ √ √ AIC √
0013 RW Perform low-density calibration √ √ √ √ √ √ √
0014 RW Perform high-density calibration √1 √ √ √ √ √ √
0014 RW Perform flowing density (FD) calibration √1
0015 RW Perform temperature offset calibration √ √ √ √ √ √ √
0016 RW Perform temperature slope calibration √ √ √ √ √ √ √
0018 RW Perform flowing density (FD) calibration √ √ √ √ √ √
0018 RW Save non-volatile data √
0020 RW Perform transmitter self-test √ √ √ √ √ √ √
0021 RO (E)EPROM checksum failure √ √ √ √ √ √ √
0022 RO RAM diagnostic failure √ √ √ √ √ √ √
0023 RO Real-time interrupt failure √
0024 RO Sensor failure √ √ √ √ √ √ √
0025 RO Temperature sensor failure √ √ √ √ √ √ √
0026 RO Flowmeter zeroing failure √ √ √ √ √ √ √
0027 RO Other failure occurred √ √ √ √ √ √ √
0028 RO Transmitter initializing/warming up √ √ √ √ √ √ √
0029 RO Primary variable out of range √ √ √ √ √ √ √
0030 RO Non-primary variable out of range √ √ √ √ √ √ √
0031 RO Milliamp output(s) saturated √ √ √ √ AIC √
0032 RO Milliamp output(s) fixed √ √ √ √ AIC √
0033 RO Watchdog timer error √ √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 5


Modbus Mapping Assignments continued

Table 2-1 Coils continued Transmitter codes on page 3, footnotes on page 9

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0034 RO Power reset occurred √ √ √ √ √ √ √
0035 RO Transmitter configuration changed √ √
0036 RO Transmitter electronics failure √ √ √ √ √ √ √
0037 RO Event 1 status √ √ √ √ √ AIC
0038 RO Event 2 status √ √ √ √ √ AIC
0039 RW Reset configuration and calibration registers √
0039 RW Save “factory calibration” meter fingerprint data set √ √ √ √ √ √
0040 RW Reset meter fingerprint statistics for “current” data set √ √ √ √ √ √
0041 RW Perform reboot √ √ √ √ √ √ √
0044 RW Perform T-Series sensor D3/K3 calibration √ √ √ √ √ √
0045 RW Perform T-Series sensor D4/K4 calibration √ √ √ √ √ √
0046 RW Fix/set discrete output 1 √2 √ AIC √
0047 RW Fix/set discrete output 2 √2 √ C √
0048 RW Fix/set discrete output 3 √
0050 RW Backlight control (ON=Backlight on; OFF=Backlight off) √ 3, 4
AICF √
3, 4

0052 RW Screen test √ 3, 4


√3 √
0053 RW Reset alarm history √
0054 RW Reset audit trail event counter and record counter √5
0055 RW Reset frequency input total √
0056 RW Reset mass total √ √ √ √ √ √
0057 RW Reset line volume (gross volume) total √ √ √ √ √ √
0058 RW Reset API reference volume total (standard volume total) √6 √6 √6 √6
0059 RW Reset enhanced density reference volume total √7 √7 √7
0060 RW Reset enhanced density net mass total √7 √7 √7
0061 RW Reset enhanced density net volume total √7 √7 √7
0062 RW Reset lifetime temperature statistics √ √ √ √
0065 RO Event 1 or event 2 status √ √ √ AIC
0066 RO Flow direction switch status √ √ √ √ √
0067 RO Flow rate indicator status √ √ √ √ √
0068 RO Zero in progress status √ √ √ √ √
0069 RO High-severity alarm status √ √ √ √ √
0071 RW Enable/disable line volume flow calculations √ √ √ √ √ √
0074 RW Enable/disable meter fingerprinting √ √ √ √ √ √
0079 RW Enable/disable copper RTD √ √ √ √ √ √
0081 RW Enable/disable cryogenic modulus compensation √ √ √ √ √ √
0082 RW Enable/disable pressure compensation √ √ √ √ √ √
0083 RW Enable/disable burst mode √ √ √ AIC √
0084 RW Enable/disable fieldbus simulation mode F
0085 RW Enable/disable locked ED curves √ 7, 8
√ 7, 8 √ 7, 9

0086 RW Enable/disable use externally written temperature (floating-point √ √ √ √ √ √


register pair 0449–0450) for internal calculations
0088 RW Enable/disable inhibit slug flow event actions √ √ 7
√7 √7 √7 √7
0089 RW Enable/disable fault on unidentified K1/FCF combination √ √7 √7 √7 √7 √7
0090 RW Enable/disable auto print at batch end √
0091 RW Enable/disable totalizer start/stop from display √4 √4
0092 RW Enable/disable display security √
0094 RW Enable/disable totalizer reset from display √ √
0095 RW Enable/disable automatic scrolling of display variables √ √
0096 RW Enable/disable access to display offline menu √ √
0097 RW Enable/disable password for access to display offline menu √ √
0098 RW Enable/disable access to alarm menu from display √ √
0099 RW Enable/disable acknowledge all alarms function using display √ √
0100 RW Start/stop (pause) batch √ 10
0100 RW Begin/end fill √2
0101 RW Resume batch/fill √2 √ 10

6 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-1 Coils continued Transmitter codes on page 3, footnotes on page 9

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0102 RW Enable/disable sensor zero from display √5
0103 RW Enable/disable sensor zero via remote communication √5
0104 RW Enable/disable batch reset from display √5
0105 RW Enable/disable batch reset via remote communication √5
0106 RW Reset AOC flow rate √2
0107 RW Pause fill √2
0108 RW Reset fill statistics √2
0109 RW Start/stop cleaning √2
0110 RW Override blocked start √2
0111 RW Enable purge stage √2
0112 RW Enable 3-position analog valve √2
0113 RW Read protect calibration factors √
0114 RW Write protect output variables and units √
0115 RW Write protect scaled integers √

Modbus Map
0116 RW Write protect sensor and transmitter information √
0117 RW Write protect special units factors √
0118 RW Write protect control output variable √
0119 RW Write protect flow direction √
0120 RW Write protect fault code √
0121 RW Write protect fault limit √
0122 RW Write protect output variables √
0123 RW Write protect flowmeter zeroing and process variable limits √
0124 RW Write protect pressure variables √
0125 RW Write protect calibration factors √
0126 RW Write protect coil 0002 (start/stop totalizer) √
0127 RW Write protect coil 0003 (reset totals) √
0128 RW Write protect coil 0004 (reset inventories) √
0129 RW Write protect coil 0005 (perform flowmeter zeroing) √
0130 RW Write protect coil 0006 (trim primary mA output at 4 mA or 0 mA) √
0131 RW Write protect coil 0007 (trim primary mA output at 20 mA) √
0132 RW Write protect coil 0008 (trim secondary mA output at 4 mA or 0 mA) √
0133 RW Write protect coil 0009 (trim secondary mA output at 20 mA) √
0134 RW Write protect coil 0010 (fix primary mA output) √
0135 RW Write protect coil 0011 (fix secondary mA output) √
0136 RW Write protect coil 0012 (fix frequency output) √
0137 RW Write protect coil 0013 (perform low-density calibration) √
0138 RW Write protect coil 0014 (perform high-density calibration) √
0142 RW Write protect coil 0018 (save non-volatile data) √
0143 RW Read/write protect master reset defaults √
0144 RW Read/write protect coil 00020 (perform transmitter self-test) √
0145 RW Read protect coil 0021 (EEPROM checksum failure) √
0146 RW Read protect coil 0022 (RAM diagnostic failure) √
0147 RW Read protect coil 0023 (real-time interrupt failure) √
0148 RW Read protect coil 0024 (sensor failure) √
0149 RW Read protect coil 0025 (temperature sensor failure) √
0150 RW Read protect coil 0026 (flowmeter zeroing failure) √
0150 RW Enable Coriolis input from sensor √
0151 RW Treat sensor alarms as informational until alarm timeout (see register N
1150)

Modbus Mapping Assignments for Micro Motion Transmitters 7


Modbus Mapping Assignments continued

Table 2-1 Coils continued Transmitter codes on page 3, footnotes on page 9

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0151 RW Read protect coil 0027 (other failure occurred) √
0152 RW Read protect coil 0028 (transmitter initializing/warming up) √
0153 RW Read protect coil 0029 (primary variable out of range) √
0154 RW Read protect coil 0030 (non-primary variable out of range) √
0155 RW Read protect coil 0031 (milliamp output saturated) √
0156 RW Read protect coil 0032 (milliamp output fixed) √
0157 RW Read protect coil 0033 (watchdog timer error) √
0158 RW Read protect coil 0034 (power reset occurred) √
0159 RW Read protect coil 0035 (transmitter configuration changed) √
0160 RW Read protect coil 0036 (transmitter electronics failure) √
0198 RW End batch √ 10
0199 RW Start/stop purge √2
0200 RW Purge mode (0=Automatic, 1=Manual) √2
0201 RW Enable/disable batcher/filling applicatoin √2 √ 10
0202 RW Enable/disable batch reset on start √ 10
0203 RW Enable/disable batch/fill count up √2 √ 10
0204 RW Enable/disable batch end warning √ 10
0205 RW Enable/disable AOC √2 √ 10
0206 RW Enable/disable batch overrun √ 10
0207 RW Enable/disable batch lockout target √ 10
0208 RW Enable/disable batch ignore source alarms √ 10
0209 RW Start AOC calibration √2 √ 10
0210 RW Save AOC calibration √2 √ 10
0211 RW Reset batch/fill current total √2 √ 10
0212 RW Reset batch preset 1 inventory √ 10
0213 RW Reset batch preset 2 inventory √ 10
0214 RW Reset batch preset 3 inventory √ 10
0215 RW Reset batch preset 4 inventory √ 10
0216 RW Reset batch preset 5 inventory √ 10
0217 RW Reset batch preset 6 inventory √ 10
0218 RW Inhibit batch √ 10
0219 RW Inhibit batch totalizer √ 10
0220 RW Reset fill counter √2
0221 RO Batch/fill timeout √2 √ 10
0222 RO Batch/fill in progress √2 √ 10
0223 RO Batch end warning √ 10
0224 RO Batch overrun √ 10
0225 RO Batch pump √ 10
0226 RO Batch/fill primary valve √2 √ 10
0227 RO Batch/fill secondary valve √2 √ 10
0228 RO Start not okay √2 √ 10
0230 RO AOC training active √2 √ 10
0231 RO Batch/fill ready to start √2 √ 10
0259 RO NOC in continuous mode N
0260 RO NOC in pause mode N
0269 RW Pause production measurements N
0270 RW Resume production measurements N
0271 RW Reset all production measurements N
0272 RW Reset back flow N
0273 RW Reset density N
0274 RW Reset drive gain N
0275 RW Reset gross flow N
0276 RW Reset mass flow N

8 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-1 Coils continued Transmitter codes on page 3, footnotes on page 9

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0277 RW Reset net oil rate (min/max/avg) N
0278 RW Reset net water cut (min/max/avg) N
0279 RW Reset temperature N
0280 RW Reset uncorrected gross N
0281 RW Reset uncorrected oil N
0282 RW Reset uncorrected water N
0283 RW Reset uncorrected water cut N
0284 RW Reset water cut N
0285 RW Enable continuous run mode N
0286 RW Enable well test mode N
0287 RW Start purge N
0288 RW Stop purge N
0289 RW Start well test N
0290 RW Stop well test N
0297 RW Start determination of produced water density N

Modbus Map
0298 RW Stop determination of produced water density N
0300 RO Configuration memory initialized; NOC not configured N
0301 RO NOC configuration corrupted N
0302 RO NOC configuration mismatched N
0303 RO NOC configuration invalid N
0304 RO Temperature out of range N
0305 RO Density out of range (slug flow) warning N
0306 RO Density out of range (slug flow) alarm N
0307 RO Continuous-mode measurements paused for more than 15 minutes N
0308 RO Transient bubbles in sensor (TBR) warning N
0309 RO Transient bubbles in sensor (TBR) alarm N
0310 RO Density calculation completed N
0311 RO Purge in progress N
0320 RW Calculate water density from manually entered values N
0321 RW Save results from calculated water density N
0322 RW Reset density determination volume total N
0323 RW Start water density determination measurement N
0324 RW Abort water density determination measurement N
0325 RW Save results from water density determination measurement N
0326 RW Start oil density determination measurement N
0327 RW Abort oil density determination measurement N
0328 RW Save intermediate results from oil density determination measurement N
0329 RW Calculate actual oil density at configured reference tempterature from N
oil density measurement and user-supplied water cut
0330 RW Save actual oil density at configured reference temperature N
0331 RO Density operation complete flag (0 = In progress; 1 = Finished) N
(1) For RFT9739 transmitters, coil 14 initiates either high-density (D2) or flowing density calibration. If the flow rate is greater than
50% of full scale, coil 14 initiates a flowing density calibration; otherwise, coil 14 initiates a D2 calibration.
(2) Requires filling and dosing application.
(3) Requires appropriate hardware.
(4) Transmitter software v3.3 and higher.
(5) Requires custody transfer application.
(6) Requires petroleum measurement application (API feature).
(7) Requires enhanced density application.
(8) Transmitter software v4.1 and higher, or, for transmitters with the FOUNDATION fieldbus outputs option board, transmitter software
v3.0 and higher.
(9) Transmitter software v6.1 and higher.
(10) Requires discrete batch application.

Modbus Mapping Assignments for Micro Motion Transmitters 9


Modbus Mapping Assignments continued

2.2 Floating-point register pairs


In Table 2-2:
• RO = Read-only
• RW = Read/write

Table 2-2 Floating-point register pairs Transmitter codes on page 3, footnotes on page 16

Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0141 0142 RW Slug flow duration (seconds) √ √ √ √ √ √ √
0143 0144 RW Fixed current for primary mA output test (milliamps) √ √ AI √ AIC √
0145 0146 RW Fixed current for secondary mA output test (milliamps) √ √ IC √
0147 0148 RW Fixed frequency for frequency output test (Hz) √ √ AI √ AIC √
0149 0150 RW Cutoff for density √ √ √ √ AICF √
0151 0152 RW Temperature for temperature offset/slope calibrations √ √ √ √ √ √ √
• MVD transmitters: °C
• RFT9739 transmitters: configured temperature unit
(register 0041)
0155 0156 RW Density for low-density calibration (g/cm3 ) √ √ √ √ √ √ √
0157 0158 RW Density for high-density calibration (g/cm3 ) √ √ √ √ √ √ √
0157 0158 RW Density for flowing-density calibration (g/cm3 ) √1
0159 0160 RW Density calibration constant 1 (µsec) √ √ √ √ √ √ √
0161 0162 RW Density calibration constant 2 (µsec) √ √ √ √ √ √ √
0163 0164 RW Density temperature coefficient √ √ √ √ √ √ √
0165 0166 RO High mass flow limit of sensor √ √ √ √ √ √2 √
0167 0168 RO High temperature limit of sensor √ √ √ √ √ √2 √
0169 0170 RO High density limit of sensor (g/cm3 ) √ √ √ √ √ √2 √
0171 0172 RO High volume flow limit of sensor √ √ √ √ √ √2 √
0173 0174 RO Low mass flow limit of sensor √ √ √ √ √ √2 √
0175 0176 RO Low temperature limit of sensor √ √ √ √ √ √2 √
0177 0178 RO Low density limit of sensor (g/cm3 ) √ √ √ √ √ √2 √
0179 0180 RO Low volume flow limit of sensor √ √ √ √ √ √2 √
0181 0182 RO Mass flow minimum range √ √ √ √ √ √2 √
0183 0184 RO Temperature minimum range √ √ √ √ √ √2 √
0185 0186 RO Density minimum range √ √ √ √ √ √2 √
0187 0188 RO Volume flow minimum range √ √ √ √ √ √2 √
0189 0190 RW Flow rate internal damping (seconds) √ √ √ √ √ √ √
0191 0192 RW Temperature internal damping (seconds) √ √ √ √ √ √ √
0193 0194 RW Density internal damping (seconds) √ √ √ √ √ √ √
0195 0196 RW Mass flow cutoff √ √ √ √ √ √ √
0197 0198 RW Volume flow cutoff √ √ √ √ √ √ √
0199 0200 RW Slug flow high-density limit (g/cm3 ) √ √ √ √ √ √ √
0201 0202 RW Slug flow low-density limit (g/cm3 ) √ √ √ √ √ √ √
0203 0204 RO Primary mA output present current (milliamps) √ √ √ √ AIC √
0205 0206 RW Added damping on primary mA output (seconds) √ √ √ √ AIC √
0207 0208 RW Flow cutoff for primary mA output √ √ √ √ AIC √
0209 0210 RW Primary variable at 20 mA √ √ √ √ AIC √
0211 0212 RW Primary variable at 0 mA or 4 mA √ √ √ √ AIC √
0213 0214 RO Secondary mA output present current (milliamps) √ √ IC √
0215 0216 RW Added damping on secondary mA output (seconds) √ √ IC √
0217 0218 RW Flow cutoff for secondary mA output √ √ IC √
0219 0220 RW Secondary variable at 20 mA √ √ IC √
0221 0222 RW Secondary variable at 0 mA or 4 mA √ √ IC √
0223 0224 RW Frequency factor or number of pulses (Hz) √ √ √ √ AIC √
0225 0226 RW Flow rate or total represented by frequency or number of pulses √ √ √ √ AIC √
0227 0228 RW Frequency maximum pulse width (milliseconds; 0 = 50% duty √ √ √ √ AIC √
cycle)
0229 0230 RO Frequency/pulse output present frequency (Hz) √ √ √ √ AIC √
0231 0232 RO Flowmeter zeroing standard deviation (units in µsec) √ √ √ √ √ √ √
0233 0234 RW Present flow signal offset at zero flow (units in µsec) √ √ √ √ √ √ √

10 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Using the Modbus Map


Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0235 0236 RO Failed DensCal/TempCal/ZeroCal value (valid only if last √ √ √ √ √ √
calibration failed)
0235 0236 RW Flowmeter zeroing standard deviation limit √
0237 0238 RW Special mass unit conversion factor √ √ √ √ √ √ √
0239 0240 RW Special volume unit conversion factor √ √ √ √ √ √ √
0241 0242 RW Event 1 setpoint √ √ √ √ √ AIC
0243 0244 RW Event 2 setpoint √ √ √ √ √ AIC
0245 0246 RO 1 (E)EPROM checksum failure √ √ √ √ √ √ √
2 RAM diagnostic failure √ √ √ √ √ √ √
4 Sensor failure √ √ √ √ √ √ √
8 Temperature sensor failure √ √ √ √ √ √ √
16 Input overrange √ √ √ √ √ √ √
32 Frequency/pulse output saturated √ √ √ √ AIC √
64 Transmitter not configured √ √ √ √ √ √ √
128 Real-time interrupt failure √ √ √ √ √ √ √

Modbus Map
256 Primary mA output saturated √ √ √ √ AIC √
512 Secondary mA output saturated √ √ IC √
1024 Primary mA output fixed √ √ √ √ AIC √
2048 Secondary mA output fixed √ √ IC √
4096 Density overrange √ √ √ √ √ √ √
8192 Flowmeter zeroing failure √ √ √ √ √ √ √
16384 Zero value too low √ √ √ √ √ √ √
32768 Zero value too high √ √ √ √ √ √ √
65536 Transmitter electronics failure √ √ √ √ √ √ √
131072 Flowmeter zeroing in progress √ √ √ √ √ √ √
262144 Slug flow √ √ √ √ √ √ √
524288 Power reset occurred √ √ √ √ √ √ √
1048576 Transmitter configuration changed √ √
2097152 Transmitter initializing/warming up √ √ √ √ √ √ √
0247 0248 RO Mass flow rate √ √ √ √ √ √ √
0249 0250 RO Density √ √ √ √ √ √ √
0251 0252 RO Temperature √ √ √ √ √ √ √
0253 0254 RO Volume flow rate √ √ √ √ √ √ √
0257 0258 RO Pressure (internally derived) √ √
0259 0260 RO Mass total √ √ √ √ √ √ √
0261 0262 RO Volume total √ √ √ √ √ √ √
0263 0264 RO Mass inventory √ √ √ √ √ √ √
0265 0266 RO Volume inventory √ √ √ √ √ √ √
0267 0268 RW Pressure correction factor for flow √ √ √ √ √ √ √
0269 0270 RW Pressure correction factor for density √ √ √ √ √ √ √
0271 0272 RW Flow calibration pressure √ √ √ √ √ √ √
0273 0274 RW Pressure input at 4 mA √
0275 0276 RW Pressure input at 20 mA √
0277 0278 RW Density for flowing-density calibration √ √ √ √ √ √
0277 0278 RW Flowing density calibration constant √3
0279 0280 RW Mass flow rate meter factor √4 √ √ √ √ √ √
0281 0282 RW Volume flow rate meter factor √3 √ √ √ √ √ √
0283 0284 RW Density meter factor √3 √ √ √ √ √ √
0285 0286 RO Raw tube frequency (Hz) √3 √ √ √ √ √ √
0287 0288 RO Left pickoff voltage (millivolts) √3 √ √ √ √ √ √
0289 0290 RO Right pickoff voltage (millivolts) √3 √ √ √ √ √ √
0291 0292 RO Drive gain (% for MVD Direct Connect, Series 1000, Series 2000, √3 √ √ √ √ √ √
Series 3000; volts for RFT9739)
0293 0294 RO Mass flow live zero flow √5 √ √ √ √ √ √
0303 0304 RW Flowing-density calibration constant √6 √ √ √ √ √ √
0319 0320 RW API reference temperature √7 √7 √7 √7
0323 0324 RW API thermal expansion coefficient √7 √7 √7 √7
0325 0326 RO Temperature-corrected density √7 √7 √7 √7

Modbus Mapping Assignments for Micro Motion Transmitters 11


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0329 0330 RO CTL √7 √7 √7 √7
0331 0332 RO Temperature-corrected volumetric flow √7 √7 √7 √7
0333 0334 RO Temperature-corrected volumetric total √7 √7 √7 √7
0335 0336 RO Temperature-corrected volumetric inventory √7 √7 √7 √7
0337 0338 RO Weighted average batch observed density √7 √7 √7 √7
0339 0340 RO Weighted average batch observed temperature √7 √7 √7 √7
0367 0368 RO DeltaT (seconds) √ √ √ √ √ √
0369 0370 RO Tube period (seconds) √ √ √ √ √ √
0371 0372 RO Temperature: Line RTD (Std RTD on all sensors); (°C) √ √ √ √ √ √
0373 0374 RO Temperature: Meter RTD (T-Series sensors only); (°C) √ √ √ √ √ √
0379 0380 RO LPO amplitude (volts) √ √ √ √ √ √
0381 0382 RO RPO amplitude (volts) √ √ √ √ √ √
0383 0384 RO Board temperature (°C) √ √ √ √ √ √
0385 0386 RO Input voltage (volts) √ √ √ √ √ √
0407 0408 RW Altus-compatibility 8 flow calibration factor (FCF) √ √ √ √ √ √
0409 0410 RW Altus-compatibility 8 temperature coefficient for flow (FT) √ √ √ √ √ √
0411 0412 RW Altus-compatibility 9 tempcal slope √ √ √ √ √ √
0413 0414 RW Altus-compatibility 9 tempcal offset √ √ √ √ √ √
0435 0436 RO Sensor maximum recorded temperature √ √ √ √ √ √
0437 0438 RO Sensor minimum recorded temperature √ √ √ √ √ √
0439 0440 RO Sensor average recorded temperature √ √ √ √ √ √
0441 0442 RW Electronics flow offset compensation for mass flow (nom 0, units √ √ √ √ √ √
µsec)
0449 0450 RW External temperature input value √ √ √ √ √ √
0451 0452 RW External pressure input value √ √ √ √ √ √
0453 0454 RW Gas density used to calculate reference volume gas and flow totals √ √ 10 √ 10 √ 10 √ 10 √ 11
0455 0456 RO Reference volume gas flow rate (not valid when petroleum √ √ 10 √ 10 √ 10 √ 10 √ 11
measurement application or enhanced density application is
enabled)
0457 0458 RO Reference volume gas total (not valid when petroleum √ √ 10 √ 10 √ 10 √ 10 √ 11
measurement application or enhanced density application is
enabled)
0459 0460 RO Reference volume gas inventory (not valid when petroleum √ √ 10 √ 10 √ 10 √ 10 √ 11
measurement application or enhanced density application is
enabled)
0463 0464 RO Electronics maximum recorded temperature √ √ √ √ √ √
0465 0466 RO Electronics minimum recorded temperature √ √ √ √ √ √
0467 0468 RO Electronics average recorded temperature √ √ √ √ √ √
0469 0470 RO Wire (cable) RTD resistance (ohms) √ √ √ √ √ √
0471 0472 RO On-board thermistor resistance (ohms) √ √ √ √ √ √
0473 0474 RO Meter RTD resistance (ohms) √ √ √ √ √ √
0475 0476 RO Line RTD resistance (ohms) √ √ √ √ √ √
0503 0504 RW T-Series K3 density constant √ √ √ √ √ √
0505 0506 RW T-Series FTG value √ √ √ √ √ √
0507 0508 RW T-Series FFQ value √ √ √ √ √ √
0509 0510 RW Density for T-Series sensor D3 density calibration √ √ √ √ √ √
0511 0512 RW Density for T-Series sensor D4 density calibration √ √ √ √ √ √
0513 0514 RW T-Series DTG value √ √ √ √ √ √
0515 0516 RW T-Series DFQ1 value √ √ √ √ √ √
0517 0518 RW T-Series DFQ2 value √ √ √ √ √ √
0519 0520 RW T-Series K4 density constant √ √ √ √ √ √
0531 0532 RW Enhanced density: Curven (6x1) temperature isothermx value √ 12 √ 12 √ 12
(x-axis)
0533 0534 RW Enhanced density: Curven (6x5) density at √ 12 √ 12 √ 12
temperature isothermx, concentrationy
0535 0536 RW Enhanced density: Curven (6x5) coefficient at √ 12 √ 12 √ 12
temperature isothermx, concentrationy
0537 0538 RW Enhanced density: Curven (1x5) concentrationy value √ 12 √ 12 √ 12
(label for y-axis)
0539 0540 RW Enhanced density: Curven (1x6) density at concentrationy √ 12 √ 12 √ 12
(at reference temperature)

12 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Using the Modbus Map


Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0541 0542 RW Enhanced density: Curven (1x6) coefficient at concentrationy √ 12 √ 12 √ 12
(at reference temperature)
0543 0544 RW Enhanced density: Curven (1x6) concentrationy value (y-axis) √ 12 √ 12 √ 12
0545 0546 RW Enhanced density: Curven reference temperature √ 12 √ 12 √ 12
0547 0548 RW Enhanced density: Curven specific gravity water reference √ 12 √ 12 √ 12
temperature
0549 0550 RW Enhanced density: Curven specific gravity water reference density √ 12 √ 12 √ 12
0551 0552 RW Enhanced density: Curven curve trim: slope √ 12 √ 12 √ 12
0553 0554 RW Enhanced density: Curven curve trim: offset √ 12 √ 12 √ 12
0555 0556 RW Enhanced density: Curven extrapolation alarm limit: % √ 12 √ 12 √ 12
0571 0572 RW Enhanced density: Curven curve fit expected accuracy √ 12 √ 12 √ 12
0611 0612 RW Discrete event x setpoint A (For x, see register 0609) √ √ 10 √ 10 √ 10 AIC 10 √
0613 0614 RW Discrete event x setpoint B (For x, see register 0609) √ √ 10 √ 10 √ 10 AIC 10 √
0629 0630 RO Fingerprint data for process/transmitter variable x, instantaneous √ √ √ √ √ √
(valid only for “current” data set) (for x, see register 0626)

Modbus Map
0631 0632 RO Fingerprint data for process/transmitter variable, average (1-minute √ √ √ √ √ √
rolling) (for x, see register 0626)
0633 0634 RO Fingerprint data for process/transmitter variable, standard √ √ √ √ √ √
deviation (1-minute rolling) (for x, see register 0626)
0635 0636 RO Fingerprint data for process/transmitter variable, maximum (since √ √ √ √ √ √
last statistics reset) (for x, see register 0626)
0637 0638 RO Fingerprint data for process/transmitter variable, minimum (since √ √ √ √ √ √
last statistics reset) (for x, see register 0626)
0687 0688 RW Slot 0 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0689 0690 RW Slot 1 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0691 0692 RW Slot 2 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0693 0694 RW Slot 3 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0695 0696 RW Slot 4 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0697 0698 RW Slot 5 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0699 0700 RW Slot 6 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0701 0702 RW Slot 7 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0703 0704 RW Slot 8 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0705 0706 RW Slot 9 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0707 0708 RW Slot 10 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0709 0710 RW Slot 11 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0711 0712 RW Slot 12 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0713 0714 RW Slot 13 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0715 0716 RW Slot 14 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0717 0718 RW Slot 15 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0719 0720 RW Slot 16 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0721 0722 RW Slot 17 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0723 0724 RW Slot 18 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0725 0726 RW Slot 19 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0727 0728 RW Slot 20 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0729 0730 RW Slot 21 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0731 0732 RW Slot 22 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0733 0734 RW Slot 23 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0735 0736 RW Slot 24 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0737 0738 RW Slot 25 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0739 0740 RW Slot 26 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0741 0742 RW Slot 27 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0743 0744 RW Slot 28 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0745 0746 RW Slot 29 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0747 0748 RW Slot 30 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √
0749 0750 RW Slot 31 value register pair for mapped address slots (Slot Type 1) √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 13


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0783 0784 RO Slot 0 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0785 0786 RO Slot 1 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0787 0788 RO Slot 2 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0789 0790 RO Slot 3 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0791 0792 RO Slot 4 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0793 0794 RO Slot 5 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0795 0796 RO Slot 6 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0797 0798 RO Slot 7 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0799 0800 RO Slot 8 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0801 0802 RO Slot 9 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0803 0804 RO Slot 10 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0805 0806 RO Slot 11 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0807 0808 RO Slot 12 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0809 0810 RO Slot 13 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0811 0812 RO Slot 14 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0813 0814 RO Slot 15 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0815 0816 RO Slot 16 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0817 0818 RO Slot 17 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0819 0820 RO Slot 18 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0821 0822 RO Slot 19 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0823 0824 RO Slot 20 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0825 0826 RO Slot 21 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0827 0828 RO Slot 22 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0829 0830 RO Slot 23 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0831 0832 RO Slot 24 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0833 0834 RO Slot 25 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0835 0836 RO Slot 26 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0837 0838 RO Slot 27 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0839 0840 RO Slot 28 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0841 0842 RO Slot 29 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0843 0844 RO Slot 30 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
0845 0846 RO Slot 31 value register pair for process variable slots (Slot Type 2) √ √ √ √ √ √
1101 1102 RW Frequency output pulses per unit √ √ √ AIC √
1103 1104 RW Frequency output units per pulse √ √ √ AIC √
1105 1106 RW Frequency output fault setting √ √ √ AIC √
1109 1110 RW Primary mA output fault setting √ √ √ AIC √
1111 1112 RW Secondary mA output fault setting √ IC √
1159 1160 RW Flow switch setpoint √ √ √ √ √
1189 1190 RW mA output 1 trim constant – 4 mA trim √ √ √ AIC √
1191 1192 RW mA output 1 trim constant – 20 mA trim √ √ √ AIC √
1193 1194 RW mA output 2 trim constant – 4 mA trim √ IC √
1195 1196 RW mA output 2 trim constant – 20 mA trim √ IC √
1223 1224 RW Frequency input frequency factor (Hz) or number of pulses √
1225 1226 RW Frequency input flow rate or total √
1227 1228 RW Frequency input K-factor √
1231 1232 RO Frequency input measured flow rate √
1233 1234 RO Frequency input measured total √
1235 1236 RO Frequency input measured inventory √
1257 1258 RW Batch: No-flow timeout (seconds) √ 13
1259 1260 RW Batch: Alarm timeout (minutes) √ 13
1261 1262 RW Batch: Maximum target √ 13
1277 1278 RW Batch/fill: Preset x open primary √ 14, 15
√ 13
1279 1280 RW Batch/fill: Preset x open secondary √ 14, 15
√ 13
1281 1282 RW Batch/fill: Preset x close primary √ 14, 15
√ 13
1283 1284 RW Batch: Preset x end warning √ 13
1285 1286 RW Batch: Preset x target √ 13
1287 1288 RW Batch: Preset x overrun √ 13

14 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Using the Modbus Map


Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1289 1290 RW Batch/fill: Current target √ 14 √ 13
1291 1292 RO Batch/fill: Current total √ 14 √ 13
1293 1294 RO Batch: Preset 1 inventory √ 13
1295 1296 RO Batch: Preset 2 inventory √ 13
1297 1298 RO Batch: Preset 3 inventory √ 13
1299 1300 RO Batch: Preset 4 inventory √ 13
1301 1302 RO Batch: Preset 5 inventory √ 13
1303 1304 RO Batch: Preset 6 inventory √ 13
1305 1306 RW Maximum fill time (seconds) √ 14
1311 1312 RW Purge delay (seconds) √ 14
1313 1314 RW Purge length (seconds) √ 14
1453 1454 RW Frequency input pulses per unit √
1455 1456 RW Frequency input units per pulse √

Modbus Map
1457 1458 RO AOC average √ 14 √ 13
1459 1460 RO Fill time (seconds) √ 14
1537 1538 RO Actual back flow rate N
1539 1540 RO Actual density N
1541 1542 RO Actual drive gain N
1543 1544 RO Actual gross flow rate N
1545 1546 RO Actual mass flow rate N
1547 1548 RO Actual net oil flow rate N
1549 1550 RO Actual net water flow rate N
1551 1552 RO Actual temperature N
1553 1554 RO Actual uncorrected net oil flow rate N
1555 1556 RO Actual uncorrected net water cut N
1557 1558 RO Actual water cut N
1567 1568 RO Average back flow rate N
1569 1570 RO Average density N
1571 1572 RO Average drive gain N
1573 1574 RO Average gross flow rate N
1575 1576 RO Average mass flow rate N
1577 1578 RO Average net oil flow rate N
1579 1580 RO Average net water flow rate N
1581 1582 RO Average temperature N
1583 1584 RO Average uncorrected net oil flow rate N
1585 1586 RO Average uncorrected net water cut N
1587 1588 RO Average water cut N
1597 1598 RO Maximum back flow rate N
1599 1600 RO Maximum density N
1601 1602 RO Maximum drive gain N
1603 1604 RO Maximum gross flow rate N
1605 1606 RO Maximum mass flow rate N
1607 1608 RO Maximum net oil flow rate N
1609 1610 RO Maximum net water flow rate N
1611 1612 RO Maximum temperature N
1613 1614 RO Maximum uncorrected net oil flow rate N
1615 1616 RO Maximum uncorrected net water cut N
1617 1618 RO Maximum water cut N
1627 1628 RO Minimum back flow rate N
1629 1630 RO Minimum density N
1631 1632 RO Minimum drive gain N
1633 1634 RO Minimum gross flow rate N
1635 1636 RO Minimum mass flow rate N

Modbus Mapping Assignments for Micro Motion Transmitters 15


Modbus Mapping Assignments continued

Table 2-2 Floating-point register pairs continued Transmitter codes on page 3, footnotes on page 16

Modeo 1500

Series 3000
Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1637 1638 RO Minimum net oil flow rate N
1639 1640 RO Minimum net water flow rate N
1641 1642 RO Minimum temperature N
1643 1644 RO Minimum uncorrected net oil flow rate N
1645 1646 RO Minimum uncorrected net water cut N
1647 1648 RO Minimum water cut N
1657 1658 RO Gross volume total N
1659 1660 RO Mass total N
1661 1662 RO Net oil volume total N
1663 1664 RO Net water volume total N
1665 1666 RO Uncorrected oil volume total N
1675 1676 RO Density of oil at reference temperature N
1677 1678 RO Maximum deviation of oil density during density determination N
1679 1680 RO Density of water at reference conditions N
1681 1682 RO Maximum deviation of water density during density determination N
1683 1684 RO Drive gain threshold for TBR (volts) N
1831 1832 RO Density value used in or resulting from density determination N
1833 1834 RO Temperature value used in or resulting from density determination N
1835 1836 RO Water cut percentage used in density determination N
1839 1840 RO Results from calculated water density operation N
1843 1844 RO Density value used in or resulting from density determination N
1847 1848 RO Density of water at reference conditions N
1851 1852 RO Density of oil at reference temperature N
1855 1856 RO Results from calculation of oil density at reference N
1859 1860 RO Density determination volume total N
2509 2510 RW Analog valve setpoint √ 14
2511 2512 RW Analog valve open full √ 14
2513 2514 RW Analog valve close partial √ 14
2515 2516 RW Fixed AOC value √ 14
2517 2518 RW Close secondary √ 14
2519 2520 RO Average of fill totals √ 14
2521 2522 RO Variance of fill totals √ 14
2527 2528 RW Analog valve closed value (mA) √ 14
(1) For RFT9739 transmitters v2.0 and higher, register pair 0157–0158 is used for both the high-density calibration and the
flowing-density calibration.
(2) Read-only for transmitters with the FOUNDATION fieldbus or Profibus-PA outputs option board.
(3) RFT9739 transmitters v3.5 and lower.
(4) RFT9739 transmitters v3 and higher.
(5) RFT9739 transmitters v3.x only.
(6) RFT9739 transmitters v3.6 and higher.
(7) Requires petroleum measurement application (API feature).
(8) Format compatible with Series 3000 9-wire (ALTUS) transmitters. The flow calibration factor is also stored in ASCII registers
0072–0076; changing the data here will change the data in ASCII registers 0072–0076, and vice versa.
(9) Format compatible with Series 3000 9-wire (ALTUS) transmitters. The temperature calibration factor is also stored in ASCII
registers 0080–0086; changing the data here will change the data in ASCII registers 0080–0086, and vice versa.
(10) Transmitter software v4.1 and higher, or for transmitters with the FOUNDATION fieldbus outputs option board, transmitter software
v3.0 and higher.
(11) Transmitter software v6.1 and higher.
(12) Requires enhanced density application.
(13) Requires discrete batch application.
(14) Requires filling and dosing application.
(15) Preset indexing does not apply; Model 1500 filling transmitter supports only one batch preset.

16 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

2.3 Integer registers

Using the Modbus Map


In Table 2-3:
• RO = Read-only
• RW = Read/write

Table 2-3 Integer registers Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0001 RO Bit #0 (E)EPROM checksum failure √ √ √ √ √ √ √
Bit #1 Transmitter configuration changed √ √
Bit #2 Sensor failure √ √ √ √ √ √ √
Bit #3 Temperature sensor failure √ √ √ √ √ √ √
Bit #4 Input overrange √ √ √ √ √ √ √
Bit #5 Frequency/pulse output saturated √ √ √ √ AIC √
Bit #6 Transmitter not configured √ √ √ √ √ √ √

Modbus Map
Bit #7 Real-time interrupt failure √
Bit #8 Milliamp output(s) saturated √ √ √ √ AIC √
Bit #9 Milliamp output(s) fixed √ √ √ √ AIC √
Bit #10 Density overrange √ √ √ √ √ √ √
Bit #11 Flowmeter zeroing failure √ √ √ √ √ √ √
Bit #12 Transmitter electronics failure √ √ √ √ √ √ √
Bit #13 Slug flow √ √ √ √ √ √ √
Bit #14 Transmitter initializing/warming up √ √ √ √ √ √ √
Bit #15 Power reset occurred √ √ √ √ √ √ √
0002 RO Mass flow rate (scaled integer) √ √ √ √ √ √ √
0003 RO Density (scaled integer) √ √ √ √ √ √ √
0004 RO Temperature (scaled integer) √ √ √ √ √ √ √
0005 RO Volume flow rate (scaled integer) √ √ √ √ √ √ √
0007 RO Pressure (scaled integer) √
0008 RO Mass total (scaled integer) 1 √ √ √ √ √ √ √
0009 RO Volume total (scaled integer) 1 √ √ √ √ √ √ √
0010 RO Mass inventory (scaled integer) 1 √ √ √ √ √ √ √
0011 RO Volume inventory (scaled integer) 1 √ √ √ √ √ √ √
0012 RW Process variable assigned to primary mA output (HART PV) √ √2 √ √ √ √ √
0013 RW Process variable assigned to secondary mA output (HART SV) √ √2 √2 √2 √2 √2 √
0014 RW Process variable assigned to frequency output (HART TV) √ √2 √ √ √2 √2 √
0015 RW Process variable assigned to RFT9739 control output √
0015 RW Process variable assigned to quaternary variable (HART QV) √2 √2 √2 √2 √2 √2
0016 RO Transmitter software revision (xxxx.x format; 141=rev14.1) √ √ √ √ √ √ √
0017 RW Flow direction √ √ √ √ √ √ √
0018 RW Maximum integer √ √ √ √ √ √ √
0019 RW Mass flow offset 3 √ √ √ √ √ √ √
0020 RW Density offset 3 √ √ √ √ √ √ √
0021 RW Temperature offset 3 √ √ √ √ √ √ √
0022 RW Volume flow offset 3 √ √ √ √ √ √ √
0024 RW Pressure offset 3 √
0025 RW Mass total offset 3 √ √ √ √ √ √ √
0026 RW Volume total offset 3 √ √ √ √ √ √ √
0027 RW Mass inventory offset 3 √ √ √ √ √ √ √
0028 RW Volume inventory offset 3 √ √ √ √ √ √ √
0029 RW Mass flow scale factor √ √ √ √ √ √ √
0030 RW Density scale factor √ √ √ √ √ √ √
0031 RW Temperature scale factor √ √ √ √ √ √ √
0032 RW Volume flow scale factor √ √ √ √ √ √ √
0034 RW Pressure scale factor √
0035 RW Mass total scale factor √ √ √ √ √ √ √
0036 RW Volume total scale factor √ √ √ √ √ √ √
0037 RW Mass inventory scale factor √ √ √ √ √ √ √
0038 RW Volume inventory scale factor √ √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 17


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0039 RW Standard or special mass flow unit √ √ √ √ √ √ √
0040 RW Density unit √ √ √ √ √ √ √
0041 RW Temperature unit √ √ √ √ √ √ √
0042 RW Standard or special volume flow unit √ √ √ √ √ √ √
0044 RW Pressure unit √ √ √ √ √ √ √
0045 RO Standard or special mass total or inventory unit 4 √ √ √ √ √ √ √
0046 RO Standard or special volume total or inventory unit 5 √ √ √ √ √ √ √
0047 RW Polling address 6 √ √ √ √ √ √ √
0120 RO Device type code √ √ √ √ √ √ √
0121 RO Electronics manufacturer’s code identification number √ √ √ √ √ √ √
0124 RW RFT9739 fault code √7
0124 RW MVD digital output fault code √ √ √ √ √ √8
0125 RO Bit #0 Primary mA output saturated √ √ √ √ AIC √
Bit #1 Secondary mA output saturated √ √ IC √
Bit #2 Primary mA output fixed √ √ √ √ AIC √
Bit #3 Secondary mA output fixed √ √ IC √
Bit #4 Density overrange √ √ √ √ √ √ √
Bit #5 Drive gain overrange √ √ √ √ √ √ √
Bit #6 Not used
Bit #7 External input failure √ √ √ AIC √
Bit #7 Milliamp input error √
Bit #8 (E)EPROM checksum failure, core processor or RFT9739 √ √ √ √ √ √ √
Bit #9 RAM diagnostic failure, core processor or RFT9739 √ √ √ √ √ √ √
Bit #10 Sensor failure (no tube interrupt) √ √ √ √ √ √ √
Bit #11 Temperature sensor failure √ √ √ √ √ √ √
Bit #12 Input overrange √ √ √ √ √ √ √
Bit #13 Frequency/pulse output saturated √ √ √ √ AIC √
Bit #14 Transmitter not configured √ √ √ √ √ √ √
Bit #15 Real-time interrupt failure √
0126 RO Bit #0 Burst mode enabled √ √ √ √ AIC √
Bit #1 Power reset occurred √ √ √ √ √ √ √
Bit #2 Transmitter initializing/warming up √ √ √ √ √ √ √
Bit #3 Security breach √
Bit #4 Display readback error √
Bit #5 Event 2 ON √ √ √ √ √ AIC
Bit #6 Event 1 ON √ √ √ √ √ AIC
Bit #7 Not used
Bit #8 Flowmeter zeroing failure √ √ √ √ √ √ √
Bit #9 Zero value too low √ √ √ √ √ √ √
Bit #10 Zero value too high √ √ √ √ √ √ √
Bit #11 Zero too noisy √ √ √ √ √ √ √
Bit #12 Transmitter electronics failure √ √ √ √ √ √ √
Bit #13 Data loss possible √ √ √ √ √ √ √
Bit #14 Calibration in progress √ √ √ √ √ √ √
Bit #15 Slug flow √ √ √ √ √ √ √
0129 RW Sensor flange type √ √ √ √ √ √ √
0130 RW Sensor flow tube construction material √ √ √ √ √ √ √
0131 RW Sensor flow tube liner material √ √ √ √ √ √ √
0132 RW Base mass unit for special mass unit √ √ √ √ √ √ √
0133 RW Base time unit for special mass unit √ √ √ √ √ √ √
0134 RW Base volume unit for special volume unit √ √ √ √ √ √ √
0135 RW Base time unit for special volume unit √ √ √ √ √ √ √
0136 RW Maximum zeroing time 9 √ √ √ √ √ √ √
0137 RW Event 1 variable assignment √ √ √ √ √ AIC
0138 RW Event 2 variable assignment √ √ √ √ √ AIC
0139 RW Event 1 type (1=High, 2=Low) √ √ √ √ √ AIC
0140 RW Event 2 type (1=High, 2=Low) √ √ √ √ √ AIC
0295 RO Security event configuration register √

18 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0296 RO Security event calibration register √
0297 RO Mass flow live zero flow √ 10
0302 RW Polling control code #1 √ √ √ AIC √
0302 RW Pressure input method √ 10
0313 RW Modbus polling address √ 11 √ √ √ √ √ √
0314 RW Last measured value fault timeout (seconds) √ √ √ √ √ √
0351 RW API 2540 CTL table type √ 12 √ 12 √ 12 √ 12
0365 RO Timestamp √ √ √ √ √ √
0366 RW DSP calculation update rate (20 or 100, Hz units) √ √ √ √ AIC
0399 RW Sensor failure timeout time (1/16 sec units) √ √ √ √ √ √
0400 RW Sensor failure status bits √ √ √ √ √ √
Bit #0 Low amplitude √ √ √ √ √ √
Bit #1 Clip flag (A/D) √ √ √ √ √ √
Bit #2 Amplitude mismatch between pickoffs √ √ √ √ √ √
Bit #3 Phase OOR (one pickoff is 180° of phase) √ √ √ √ √ √

Modbus Map
Bit #4 Freq OOR (main signal component detected on pickoff is √ √ √ √ √ √
outside expected frequency range)
0419 RO Bit #0 (E)EPROM checksum error, core processor √ √ √ √ √ √
Bit #1 RAM test error, core processor √ √ √ √ √ √
Bit #2 Not used
Bit #3 Sensor not vibrating √ √ √ √ √ √
Bit #4 Temperature sensor out of range √ √ √ √ √ √
Bit #5 Calibration failure √ √ √ √ √ √
Bit #6 Other failure occurred √ √ √ √ √ √
Bit #7 Transmitter initializing/warming up √ √ √ √ √ √
Bit #8 Primary variable out of limits √
Bit #9 Non-primary variable out of limits √
Bit #10 Not used
Bit #11 Not used
Bit #12 Watchdog error √
Bit #13 Cold start occurred √
Bit #14 Transmitter configuration changed (HART bit) √
Bit #15 High-severity alarm(s) active √ √ √ √ √ √
0420 RO Bit #0 Primary mA output saturated √ √ √ AIC √
Bit #1 Secondary mA output saturated √ IC √
Bit #2 Primary mA output fixed √ √ √ AIC √
Bit #3 Secondary mA output fixed √ IC √
Bit #4 Density overrange √ √ √ √ √ √
Bit #5 Drive overrange √ √ √ √ √ √
Bit #6 Not used
Bit #7 External input failure √ √ √ AIC √
Bit #8 (E)EPROM checksum failure, core processor √ √ √ √ √ √
Bit #9 RAM diagnostic failure, core processor √ √ √ √ √ √
Bit #10 Sensor not responding (no tube interrupt) √ √ √ √ √ √
Bit #11 Temperature sensor out of range √ √ √ √ √ √
Bit #12 Input overrange √ √ √ √ √ √
Bit #13 Frequency/pulse output saturated √ √ √ AIC √
Bit #14 Transmitter not characterized (flow calibration factor or √ √ √ √ √ √
sensor type)
Bit #15 Not used

Modbus Mapping Assignments for Micro Motion Transmitters 19


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0421 RO Bit #0 Burst mode enabled √ √ √ AIC √
Bit #1 Power reset occurred √ √ √ √ √ √
Bit #2 Transmitter initializing/warming up √ √ √ √ √ √
Bit #3 Sensor/transmitter communication failure (A28) √ 13 √ 13 √ 13 √ 13 √
Bit #4 Paper out √
Bit #5 Event 2 ON √ √ √ √ AIC
Bit #6 Event 1 ON √ √ √ √ AIC
Bit #7 Sensor/transmitter communication failure (A26) √ √ √ √ √
Bit #8 Calibration failure √ √ √ √ √ √
Bit #9 Zero value too low √ √ √ √ √ √
Bit #10 Zero value too high √ √ √ √ √ √
Bit #11 Zero too noisy √ √ √ √ √ √
Bit #12 Transmitter electronics failure √ √ √ √ √ √
Bit #13 Data loss possible √ √ √ √ √ √
Bit #14 Calibration in progress √ √ √ √ √ √
Bit #15 Slug flow √ √ √ √ √ √
0422 RO Bit #0 API: Temperature outside standard range √ 12 √ 12 √ 12 √ 12
Bit #1 API: Density outside standard range √ 12 √ 12 √ 12 √ 12
Bit #2 “Line RTD” temperature out of range √ √ √ √ √ √
Bit #3 “Meter RTD” temperature out of range √ √ √ √ √ √
Bit #4 Flow direction (0=Forward or Zero flow, 1=Reverse) √ √ √ √ √ √
Bit #5 Not used
Bit #6 Enhanced density: Unable to fit curve data √ 14 √ 14 √ 14
Bit #7 Last measured value override active √ √ √ √ √
Bit #8 Enhanced density extrapolation alarm √ 14 √ 14 √ 14
Bit #9 Transmitter not config0ured (flow calibration factor) √ √ √ √ √ √
Bit #10 (E)EPROM checksum error √ √ √ √ √
Bit #11 RAM test error in transmitter √ √ √ √ √
Bit #12 Invalid/unrecognized sensor type (K1 value) √ √ √ √ √ √
Bit #13 (E)EPROM database corrupt in core processor √ √ √ √ √ √
Bit #14 (E)EPROM powerdown totals corrupt in core processor √ √ √ √ √ √
Bit #15 (E)EPROM program corrupt in core processor √ √ √ √ √ √
0423 RO Bit #0 Core processor boot sector fault √ √ √ √ √ √
Bit #1 Transmitter software upgrade recommended √ √ √ √ √
Bit #2 Frequency output fixed √ √ √ AIC √
Bit #3 Not used
Bit #4 DO1 status (0=OFF, 1=ON) √ 15 √ AIC √
Bit #5 DO2 status (0=OFF, 1=ON) √ 15 √ C √
Bit #6 T-Series D3 calibration in progress √ √ √ √ √ √
Bit #7 T-Series D4 calibration in progress √ √ √ √ √ √
Bit #8 DO3 status (0=OFF, 1=ON) √
Bit #9 Not used
Bit #10 Temperature slope calibration in progress √ √ √ √ √ √
Bit #11 Temperature offset calibration in progress √ √ √ √ √ √
Bit #12 Flowing density calibration in progress √ √ √ √ √ √
Bit #13 High-density calibration in progress √ √ √ √ √ √
Bit #14 Low-density calibration in progress √ √ √ √ √ √
Bit #15 Flowmeter zeroing in progress √ √ √ √ √ √

20 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0424 RO Bit #0 Discrete input 1 status (0=OFF, 1=ON) √ 15 √ C √
Bit #1 Discrete input 2 status (0=OFF, 1=ON) √
Bit #2 Discrete output 1 fixed √ 15 √ AIC √
Bit #3 Discrete output 2 fixed √ 15 √ C √
Bit #4 Discrete output 3 fixed √
Bit #5 Not used
Bit #6 Security breach √ √ √ √ √
Bit #7 Frequency input saturated √
Bit #8 Batch/fill timeout √ 15 √ 16
Bit #9 Batch/fill in progress √ 15 √ 16
Bit #10 Batch end warning √ 16
Bit #11 Batch overrun √ 16
Bit #12 Batch pump √ 16
Bit #13 Batch/fill primary valve √ 15 √ 16
Bit #14 Batch/fill secondary valve √ 15 √ 16

Modbus Map
Bit #15 Not used
0433 RO Bit #0 K1/FCF combination unrecognized √ √ 17 √ 17 √ 17 √ 17 √ 18
0521 RW Floating-point byte order √ √ √ √ √ √
0522 RW Additional delay to Modbus message response 19 √ √ √ √ √ √
0523 RW Enhanced density: Active calculation curve (a=0,1,2,3,4,5) √ 14 √ 14 √ 14
0524 RW Enhanced density: Derived variable √ 14 √ 14 √ 14
0527 RW Enhanced density: Curve configuration index (n=0,1,2,3,4,5) √ 14 √ 14 √ 14
0528 RW Enhanced density: Curven temperature isotherm index (x=0,1,2,3,4,5) √ 14 √ 14 √ 14
0529 RW Enhanced density: Curven concentration index (y=0,1,2,3,4,5) √ 14 √ 14 √ 14
0563 RW Enhanced density: Curven using coefficients direct (0=No, 1=Yes) √ 14 √ 14 √ 14
0564 RW Enhanced density: Maximum fit order for 5x5 curve (Order=2,3,4,5) √ 14 √ 14 √ 14
0569 RW Enhanced density: Curven curve fit results (0=Good, 1=Poor, 2=Failed, √ 14 √ 14 √ 14
3=Empty)
0570 RW Enhanced density: Curven concentration units label code √ 14 √ 14 √ 14
0609 RW Discrete event index (x=0,1,2,3,4) √ √ 17
√ 17
√ 17 AIC √17

0610 RW Discrete event x action (0=HI, 1=LO, 2=IN HI/LO, 3=OUT HI/LO) (For √ √ 17 √ 17 √ 17 AIC 17 √
x, see register 0609)
0615 RW Discrete event x process variable (For x, see register 0609) √ √ 17 √ 17 √ 17 AIC 17 √
0616 RO Discrete event x status (1=ON, 2=OFF) (Bit #0: x = 0; Bit #1: x = 1, ...) √ √ 17 √ 17 √ 17 AIC 17 √
0625 RW Meter fingerprint units (0=SI, 1=English) √ √ √ √ √ √
0626 RW Meter fingerprint transmitter variable index √ √ √ √ √ √
0627 RW Fingerprint data set (0=Current, 1=Factory, 2=Installation, 3=Last √ √ √ √ √ √
zero)
0655 RW Slot 0 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0656 RW Slot 1 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0657 RW Slot 2 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0658 RW Slot 3 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0659 RW Slot 4 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0660 RW Slot 5 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0661 RW Slot 6 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0662 RW Slot 7 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0663 RW Slot 8 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0664 RW Slot 9 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0665 RW Slot 10 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0666 RW Slot 11 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0667 RW Slot 12 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0668 RW Slot 13 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0669 RW Slot 14 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0670 RW Slot 15 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0671 RW Slot 16 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0672 RW Slot 17 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0673 RW Slot 18 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0674 RW Slot 19 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 21


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
0675 RW Slot 20 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0676 RW Slot 21 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0677 RW Slot 22 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0678 RW Slot 23 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0679 RW Slot 24 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0680 RW Slot 25 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0681 RW Slot 26 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0682 RW Slot 27 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0683 RW Slot 28 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0684 RW Slot 29 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0685 RW Slot 30 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0686 RW Slot 31 configuration register for mapped address slots (Slot Type 1) √ √ √ √ √ √
0751 RW Slot 0 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0752 RW Slot 1 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0753 RW Slot 2 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0754 RW Slot 3 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0755 RW Slot 4 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0756 RW Slot 5 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0757 RW Slot 6 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0758 RW Slot 7 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0759 RW Slot 8 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0760 RW Slot 9 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0761 RW Slot 10 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0762 RW Slot 11 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0763 RW Slot 12 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0764 RW Slot 13 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0765 RW Slot 14 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0766 RW Slot 15 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0767 RW Slot 16 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0768 RW Slot 17 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0769 RW Slot 18 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0770 RW Slot 19 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0771 RW Slot 20 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0772 RW Slot 21 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0773 RW Slot 22 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0774 RW Slot 23 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0775 RW Slot 24 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0776 RW Slot 25 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0777 RW Slot 26 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0778 RW Slot 27 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0779 RW Slot 28 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0780 RW Slot 29 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0781 RW Slot 30 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
0782 RW Slot 31 configuration register for process variable slots (Slot Type 2) √ √ √ √ √ √
1107 RW Frequency/pulse output fault code √ √ √ AIC √
1108 RW Frequency/pulse output scaling method √ √ √ AIC √
1113 RW Primary milliamp output fault code √ √ √ AIC √
1114 RW Secondary milliamp output fault code √ IC √
1115 RW Display offline password (0000 to 9999) √ √
1116 RW Display scroll rate (1 to 10 seconds) √ √
1117 RW 20 Display variable #1 √ √ √
1118 RW Display variable #2 √ √ √
1119 RW Display variable #3 √ √ √
1120 RW Display variable #4 √ √ √
1121 RW Display variable #5 √ √ √
1122 RW Display variable #6 √ √ √
1123 RW Display variable #7 √ √ √
1124 RW Display variable #8 √ √ √

22 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1125 RW Display variable #9 √ √ √
1126 RW Display variable #10 √ √ √
1127 RW Display variable #11 √ √ √
1128 RW Display variable #12 √ √ √
1129 RW Display variable #13 √ √ √
1130 RW Display variable #14 √ √ √
1131 RW Display variable #15 √ √ √
1132 RW RS-485 digital communication protocol setting √ A √ A √
1133 RW RS-485 digital communication baud rate √ A √ A √
1134 RW RS-485 digital communication parity setting √ A √ A √
1135 RW RS-485 digital communication stop bits setting √ A √ A √
1136 RW RS-485/service port switch (0=Service port, 1=RS-485) √ √ √
1137 RO Core processor software revision √ √ √ √ √
1138 RO Outputs option board √ √ √ √ √
1139 RW Sensor type code √ √ √ √ √

Modbus Map
1144 RW Polling control code #2 √ √ √ AIC √
1145 RW Polled variable #1 code √ √ √ AIC √
1146 RW Polled variable #2 code √ √ √ AIC √
1147 RW Polling type code √ 21 AIC 21
1148 RW Totalizer display precision √ AICF
1150 RW Coriolis and sensor alarm timeout (minutes) N
1151 RW Discrete output 1 assignment √ 15 √ AIC √
1152 RW Discrete output 1 polarity (0=Active low, 1=Active high) √ 15 √ AIC √
1153 RW Discrete output 2 assignment √ 15 √ C √
1154 RW Discrete output 2 polarity (0=Active low, 1=Active high) √ 15 √ C √
1155 RW Discrete output 3 assignment √
1156 RW Discrete output 3 polarity (0=Active low, 1=Active high) √
1164 RW 100 Hz update rate variable assignment √ √ √ AIC
1165 RW Burst command √ √ √ AIC √
1166 RW Output channel A type assignment (read-only) √ √ √ AIC
1167 RW Output channel B type assignment √ 22 √ 23 √ AIC 24
1168 RW Output channel C type assignment √ 22 √ 23 √ AIC 24
1169 RW Burst variable 1 √ √ √ AIC √
1170 RW Burst variable 2 √ √ √ AIC √
1171 RW Burst variable 3 √ √ √ AIC √
1172 RW Burst variable 4 √ √ √ AIC √
1173 RW Channel A power (read-only) √ C
1174 RW 25 Channel B power √ 22 √ 23 √ AIC 24
1175 RW Channel C power √ 22 √ 23 √ AIC 24
1176 RW Discrete input 1 assignment √ 15 √ C
1178 RW Discrete input 1 polarity (0=Active low, 1=Active high) √ 15 √ 26 C 26 √
1179 RW Discrete input 2 polarity (0=Active low, 1=Active high) √
1181 RW 27 Frequency output mode √ √ √ AIC
1182 RW Discrete output 1 fixed value √ 15 √ AIC √
1183 RW Discrete output 2 fixed value √ 15 √ C √
1184 RW Discrete output 3 fixed value √
1186 RW Profibus station address (0–126) P
1197 RW Frequency output polarity (0=Active low, 1=Active high) √ √ √ AIC √
1198 RW Frequency output power √
1199 RW Display test on/off duration (1–60 seconds) √ 17 √ 17
1200 RO Transmitter software version (xxx.xx: 1353=rev13.53) √ √ 28 √ √ 28 √
1215 RW Real-time clock: Seconds (0–59) √
1216 RW Real-time clock: Minutes (0–59) √
1217 RW Real-time clock: Hours (0–23) √
1218 RW Real-time clock: Day (1–31) √
1219 RW Real-time clock: Month (1–12) √
1220 RW Real-time clock: Year (1996–2095) √

Modbus Mapping Assignments for Micro Motion Transmitters 23


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1221 RW Frequency input measurement units √
1222 RW Frequency input scaling method √
1229 RO Frequency input measured frequency √
1237 RW Alarm index (1=A1, 2=A2, ... 39=A39, 40=A100, 41=A101, ... √ √ 17 √ 17 √ 17 √ 17 √
70=A130)
1238 RW Alarm x severity (0=Ignore, 1=Informational, 2=Fault) (For x, see √ √ 17 √ 17 √ 17 √ 17 √
register 1237)
1239 RO Alarm status Bit #0: 0=Inactive, 1=Active √ 17 √ 17 √ 17 √ 17 √
RW Alarm status Bit #1: 0=Unacknowledged, 1=Acknowledged 29 √ 17 √ 17 √ 17 √ 17 √
1240 RO Alarm count √
1245 RW Alarm type index (0=Electronics, 1=Sensor, 2=Process, √
3=Configuration)
1246 RW Alarm type status (0=Acknowledged, 1=Unacknowledged) √
1251 RW Batch/fill: Flow source √ 15 √ 16
1252 RW Batch: Current preset (0–5) √ 16
1253 RW Batch/fill: Type (1=1 stage, 2=2 stage, 3=Analog 15 ) √ 15 √ 16
1254 RW Batch: Precision of display values √ 16
1255 RW Batch/fill: Configure presets by (0=%target, 1=Quantity) √ 15 √ 16
1256 RO Batch/fill: State 30 √ 15 √ 16
1263 RW Batch: Preset configuration index (x=0,1,2,3,4,5) √ 16
1264 RW Batch: Enable/disable preset x (0=Disable, 1=Enable) √ 16
(For x, see register 1263)
1276 RW Batch: preset x density curve (For x, see register 1263) √ 16
1309 RW AOC compensation algorithm (0=Learn upward, 1=Learn downward, √ 15
2=Fixed)
1310 RW AOC learning cycles √ 15
1315 RW Discrete input assignment: Start sensor zero √
1316 RW Discrete input assignment: Reset mass total √
1317 RW Discrete input assignment: Reset volume total √
1318 RW Discrete input assignment: Reset API reference (standard) volume √ 12
total
1319 RW Discrete input assignment: Reset enhanced density reference √ 14
(corrected) volume total
1320 RW Discrete input assignment: Reset enhanced density net mass total √ 14
1321 RW Discrete input assignment: Reset enhanced density net volume total √ 14
1322 RW Discrete input assignment: Reset all totals √
1323 RW Discrete input assignment: Start/stop all totalization √
1324 RW Discrete input assignment: End batch √ 16
1325 RW Discrete input assignment: Inhibit batch √ 16
1326 RW Discrete input assignment: Inhibit batch totalizer √ 16
1327 RW Discrete input assignment: Reset batch √ 16
1328 RW Discrete input assignment: Resume batch √ 16
1329 RW Discrete input assignment: Start batch √ 16
1330 RW Discrete input assignment: Stop batch √ 16
1331 RW Discrete input assignment: Increment current batch preset √ 16
1332 RW Discrete input assignment: Increment current enhanced density curve √ 14, 16

1333 RW Discrete input assignment: Print screen √


1351 RW Maintenance password cursor press 1 √
1352 RW Maintenance password cursor press 2 √
1353 RW Maintenance password cursor press 3 √
1354 RW Maintenance password cursor press 4 √
1355 RW Configuration password cursor press 1 √
1356 RW Configuration password cursor press 2 √
1357 RW Configuration password cursor press 3 √
1358 RW Configuration password cursor press 4 √

24 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1359 RW Language selection √
1361 RO Display temperature (°C) √
1362 RW Display variable #16 √
1363 RW Display variable #17 √
1364 RW Display variable #18 √
1365 RW Display variable #19 √
1366 RW Display variable #20 √
1367 RW Process variable code (n=0...94; see process variable codes in √
Table 2-32)
1368 RW Display precision (0–5) for process variable n (For n, see register √
1367)
1382 RW Header line 1 line feeds √
1383 RW Header line 1 font size (0=Normal height, 1=Double height) √
1395 RW Header line 2 line feeds √
1396 RW Header line 2 font size (0=Normal height, 1=Double height) √

Modbus Map
1408 RW Header line 3 line feeds √ 31
1409 RW Header line 3 font size (0=Normal height, 1=Double height) √ 31
1421 RW Header line 4 line feeds √ 31
1422 RW Header line 4 font size (0=Normal height, 1=Double height) √ 31
1434 RW Footer line feeds √
1435 RW Footer font size (0=Normal height, 1=Double height) √
1436 RW Lines per page √
1437 RW Number of tickets √ 16
1438 RW Number of data bits √
1441 RW Printer type √
1442 RW Print request configuration √
1444 RW Disable paper out detection (0=Allow for Epson, 1=Disallow for Epson) √
1445 RW Printer characters per second √
1446 RW Printer buffer size √
1461 RW Fill count √ 15
1470 RW World area (0=Worldwide, 1=Europe) √ 31
1471 RO Audit trail event counter (event index) √ 31
1472 RO Audit trail index value for next record √ 31
1473 RW Audit trail record index (n=0...999) √ 31
1474 RO Audit trail recordn event index √ 31
1477 RO Audit trail recordn changed register (0xxxx for coil, 4xxxx for register) √ 31
1478 RO Audit trail recordn changed register array index (valid only if register √ 31
specified in 1477 refers to an array)
1479 RO Audit trail recordn units (valid only if register specfied in 1477 normally √ 31
has associated units)
1505 RW Fieldbus AI function block index (0–3) F
1506 RW Fieldbus AI target mode F
1507 RO Fieldbus AI actual mode F
1508 RW Fieldbus AI channel F
1509 RW Fieldbus AI XD scale units F
1510 RW Fieldbus AI linearization type F
1511 RW Fieldbus integrator block type F
1691 RW Time period over which oil density will be averaged during density N
determination (seconds)
1692 RW Time period over which produced water density will be averaged N
during density determination (seconds)
1693 RW Purge time to allow the separator contents of the previous test to N
purge (seconds)
1694 RW Look-back time interval for TBR (5–30 seconds) N

Modbus Mapping Assignments for Micro Motion Transmitters 25


Modbus Mapping Assignments continued

Table 2-3 Integer registers continued Transmitter codes on page 3, footnotes on page 26

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address RO/RW Description
1701 RW Well to be configured (1–48) N
1702 RW Well currently being tested (1–48) N
1703 RW Test results (0=Well currently under test, 1=Last test, 2=Next-to-last N
test, 3=Oldest test)
1704 RW Well within archive for test results (1–48) N
1705 RW Reference temperature (0=60 °F, 1=15 °C, 2=20 °C) N
1706 RW Action to be taken on TBR event (0=Hold last value, 1=Stop well test, N
2=Alarm only)
2305 RW Index of well name to configure (1–48) N
2525 RW Analog valve number of stages (1 or 2) √ 15
2526 RW Bit #0 Fill in progress √ 15
Bit #1 Primary valve √ 15
Bit #2 Secondary valve √ 15
Bit #3 Start not okay √ 15
Bit #4 AOC flow rate too high √ 15
Bit #5 Cleaning in progress √ 15
Bit #6 Purge delay phase (auto mode) √ 15
Bit #7 Purge phase √ 15
Bit #8 AOC training √ 15
Bit #9 Maximum fill time exceeded √ 15
Bit #10 Not used
Bit #11 Not used
Bit #12 Not used
Bit #13 Not used
Bit #14 Not used
Bit #15 Not used
5003 RW Custody transfer security switch (7070=Secured; all other √ 31
values=Unsecured)
(1) For RFT9739 transmitters, read-only. For MVD Direct Connect and Series 1000, 2000, and 3000 transmitters, read-only during
standard use; to reset, write 0 to register.
(2) Supported in software even if associated output is not available.
(3) Signed integer.
(4) Automatically derived from register 0039.
(5) Automatically derived from register 0042.
(6) Modbus or HART polling address (v3.6 or lower RFT9739 transmitter); HART polling address (v3.7 or higher RFT9739 transmitter,
MVD Direct Connect, and Series 1000, 2000, and 3000 transmitters).
(7) Read-only for transmitters v3 and higher.
(8) Transmitter software v6.0 and higher.
(9) For MVD Solo and Series 1000, 2000, and 3000 transmitters, unit is seconds. For RFT9739 transmitters, unit is M (measurement
cycles); M = N / 2, where N represents the number of tube periods.
(10) RFT9739 transmitters v3.x only.
(11) RFT9739 transmitters v3.7 and higher.
(12) Requires petroleum measurement application (API feature).
(13) Transmitter software v3.7 and higher.
(14) Requires enhanced density application.
(15) Requires filling and dosing application.
(16) Requires discrete batch application.
(17) Transmitter software v4.1 and higher, or for transmitters with the FOUNDATION fieldbus outputs option board, transmitter software
v3.0 and higher.
(18) Transmitter software v6.1 and higher.
(19) Units: 38,400 baud – 200 µsec; 19,200 baud – 400 µsec; 9600 baud – 800 µsec; 3800 baud – 1.6 µsec; 2400 baud – 3.2 µsec; 1200
baud – 6.4 µsec.
(20) Read-write for Series 3000 transmitters. Read-only for Series 1000/2000 transmitters; primary variable automatically assigned as
first display variable.
(21) Transmitter software v2.x and lower.
(22) Read-write for Model 1500 filling transmitters; read-only for other Model 1500 transmitters.
(23) Read-only.

26 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

(24) Read-only for Model 2700 with analog outputs option board; configurable for Model 2700 with configurable input/outputs option

Using the Modbus Map


board or instrinsically safe outputs option board.
(25) Read-only if Channel B is configured for mA output; read-write for other output types.
(26) Read-write for transmitter software v4.1 and higher; read-only for earlier transmitters.
(27) Read-write for transmitters with configurable input/outputs option board; read-only for all other transmitters.
(28) Transmitter software v3.6 and higher.
(29) Can be changed from 0 to 1; cannot be changed from 1 to 0.
(30) For Series 3000 transmitters: 0=Startup, 1=Idle, 2=Stopped, 3=Running. For Model 1500 transmitters: 0=Startup, 1=Idle,
2=Paused, 3=Running, 4=Purge delay, 5=Purge in progress, 6=Cleaning in progress.
(31) Requires custody transfer application.

2.4 Long integer registers


In Table 2-4:
• RO = Read-only
• RW = Read/write

Table 2-4 Long integer registers Transmitter codes on page 3, footnotes on page 29

Modbus Map
Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
RO/
Address RW Description
0048 0049 RW Final assembly number of transmitter; 3-byte integer. √ √ √ √ √ √ √
• First register: 00000000 in high-order byte; first 8 bits of value in
low-order byte.
• Second register: last 16 bits of value.
0050 0051 RW Date; 3-byte integer. √ √ √ √ √ √ √
• First register: 00000000 in high-order byte; day of month (1–31)
in low-order byte.
• Second register: high-order byte: month of year (1–12);
low-order byte: X, where X = current year - 1900.
0122 0123 RO 1 HART device ID 2; 3-byte integer. √ √ √ √ √ √ √
• First register: 00000000 in high-order byte; first 8 bits of value in
low-order byte.
• Second register: last 16 bits of value.
0127 0128 RW Sensor serial number: 3-byte integer. √ √ √ √ √ √ √
• First register: 00000000 in high-order byte; first 8 bits of value in
low-order byte.
• Second register: last 16 bits of value.
0305 0308 RO Binary mass total; 8-byte integer. √3
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
0309 0312 RO Binary volume total; 8-byte integer. √3
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
1187 1188 RO 1 Core processor 4 HART device ID; 3-byte integer. √ √ √ √ √
• First register: 00000000 in high-order byte; first 8 bits of value in
low-order byte.
• Second register: last 16 bits of value.
1241 1242 RO Time that alarm was last posted, in seconds since January 1,1996; √
4-byte integer.
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
1243 1244 RO Time that alarm was last cleared, in seconds since January 1, √
1996; 4-byte integer.
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
1247 1248 RO Time that alarm type was last posted, in seconds since January 1, √
1996; 4-byte integer.
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
1249 1250 RO Time that alarm type was last cleared, in seconds since January 1, √
1996; 4-byte integer.
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
1439 1440 RW Start number (BOL); 4-byte integer √5
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)

Modbus Mapping Assignments for Micro Motion Transmitters 27


Modbus Mapping Assignments continued

Table 2-4 Long integer registers continued Transmitter codes on page 3, footnotes on page 29

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
RO/
Address RW Description
1475 1476 RO Audit trail timestamp for recordn; 4-byte integer √5
• First register: MSW (high-order byte, low-order byte)
• Second register: LSW (high-order byte, low-order byte)
For n, see register 1473.
1711 1712 RO Well test elapsed time (seconds) N
1713 1714 RO Production measurement paused time (seconds) N
1715 1716 RO Well test purge time remaining (seconds) N
1727 1728 RO Timecode when last well test was stopped (seconds since N
January 1, 1998)
1729 1730 RO Timecode when last well test was started (seconds since N
January 1, 1998)
1741 1742 RO Timecode when maximum back flow was achieved (seconds since N
January 1, 1998)
1743 1744 RO Timecode when maximum density was achieved (seconds since N
January 1, 1998)
1745 1746 RO Timecode when maximum drive gain was achieved (seconds since N
January 1, 1998)
1747 1748 RO Timecode when maximum gross flow rate was achieved (seconds N
since January 1, 1998)
1749 1750 RO Timecode when maximum mass flow rate was achieved (seconds N
since January 1, 1998)
1751 1752 RO Timecode when maximum net oil flow rate rate was achieved N
(seconds since January 1, 1998)
1753 1754 RO Timecode when maximum net water flow rate rate was achieved N
(seconds since January 1, 1998)
1755 1756 RO Timecode when maximum temperature was achieved (seconds N
since January 1, 1998)
1757 1758 RO Timecode when maximum uncorrected net oil flow was achieved N
(seconds since January 1, 1998)
1759 1760 RO Timecode when maximum uncorrected water cut was achieved N
(seconds since January 1, 1998)
1761 1762 RO Timecode when maximum water cut was achieved (seconds since N
January 1, 1998)
1771 1772 RO Timecode when minimum back flow was achieved (seconds since N
January 1, 1998)
1773 1774 RO Timecode when minimum density was achieved (seconds since N
January 1, 1998)
1775 1776 RO Timecode when minimum drive gain was achieved (seconds since N
January 1, 1998)
1777 1778 RO Timecode when minimum gross flow rate was achieved (seconds N
since January 1, 1998)
1779 1780 RO Timecode when minimum mass flow rate was achieved (seconds N
since January 1, 1998)
1781 1782 RO Timecode when minimum net oil flow rate rate was achieved N
(seconds since January 1, 1998)
1783 1784 RO Timecode when minimum net water flow rate rate was achieved N
(seconds since January 1, 1998)
1785 1786 RO Timecode when minimum temperature was achieved (seconds N
since January 1, 1998)
1787 1788 RO Timecode when minimum uncorrected net oil flow was achieved N
(seconds since January 1, 1998)
1789 1790 RO Timecode when minimum uncorrected water cut was achieved N
(seconds since January 1, 1998)
1791 1792 RO Timecode when minimum water cut was achieved (seconds since N
January 1, 1998)

28 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-4 Long integer registers continued Transmitter codes on page 3, footnotes on page 29

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
RO/
Address RW Description
1801 1802 RO Timecode when back flow was reset (seconds since January 1, N
1998)
1803 1804 RO Timecode when density was reset (seconds since January 1, N
1998)
1805 1806 RO Timecode when drive gain was reset (seconds since January 1, N
1998)
1807 1808 RO Timecode when gross flow rate was reset (seconds since January N
1, 1998)
1809 1810 RO Timecode when mass flow rate was reset (seconds since January N
1, 1998)
1811 1812 RO Timecode when net oil flow rate rate was reset (seconds since N
January 1, 1998)
1813 1814 RO Timecode when net water flow rate rate was reset (seconds since N
January 1, 1998)
1815 1816 RO Timecode when temperature was reset (seconds since January 1, N
1998)
1817 1818 RO Timecode when uncorrected net oil flow was reset (seconds since N

Modbus Map
January 1, 1998)
1819 1820 RO Timecode when uncorrected water cut was reset (seconds since N
January 1, 1998)
1821 1822 RO Timecode when water cut was reset (seconds since January 1, N
1998)
1837 1838 RO Timecode when density determination performed (seconds since N
January 1, 1998)
1841 1842 RO Timecode for last calculated water density operation (seconds N
since January 1, 1998)
1845 1846 RO Timecode for last water density determination measurement N
(seconds since January 1, 1998)
1849 1850 RO Timecode for current water density at reference (seconds since N
January 1, 1998)
1853 1854 RO Timecode for current oil density at reference (seconds since N
January 1, 1998)
1857 1858 RO Timecode for last calculation of oil density at reference (seconds N
since January 1, 1998)
(1) For all Series 1000, 2000, and 3000 transmitters, if the register contains a non-zero value, it is read-only; if it contains 0, it can be
written to.
(2) If wiring is to transmitter, register contains transmitter device ID; if wiring is direct to core processor, register contains core
processor device ID.
(3) RFT9739 transmitters v3.7 and higher.
(4) Must be queried through transmitter. If wiring is direct to core processor, registers 1187 and 1188 do not exist.
(5) Requires custody transfer application.

Modbus Mapping Assignments for Micro Motion Transmitters 29


Modbus Mapping Assignments continued

2.5 ASCII registers


In Table 2-5, each ASCII register holds two characters.

Table 2-5 ASCII registers Transmitter codes on page 3, footnotes on page 33

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address Description

0052 Label for special mass flow unit


0053 Label for special mass flow unit – Single-write
√ √ √ √ √ √ √
0054 Label for special mass flow unit 1 or space character 2 multiple
0055 Label for special mass flow unit 1 or space character 2

0056 Label for special mass total or mass inventory unit


0057 Label for special mass total or mass inventory unit
0058 Label for special mass total or mass inventory unit 1 – Single-write
√ √ √ √ √ √ √
or space character 2 multiple
0059 Label for special mass total or mass inventory unit 1
or space character 2

0060 Label for special volume flow unit


0061 Label for special volume flow unit – Single-write
√ √ √ √ √ √ √
0062 Label for special volume flow unit 1 or space character 2 multiple
0063 Label for special volume flow unit 1 or space character 2

0064 Label for special volume total or volume inventory unit


0065 Label for special volume total or volume inventory unit
0066 Label for special volume total or volume inventory unit 1 – Single-write
√ √ √ √ √ √ √
or space character 2 multiple
0067 Label for special volume total or volume inventory unit 1
or space character 2

0068 HART tag (software tag)


0069 HART tag (software tag) – Single-write
√ √ √ √ √ √3 √
0070 HART tag (software tag) multiple
0071 HART tag (software tag)

0072 Flow calibration factor 4


0073 Flow calibration factor
0074 Flow calibration factor
0075 Flow temperature coefficient 4
– Single-write
√ √ √ √ √ √ √
0076 Flow temperature coefficient multiple
0077 Space character 5
0078 Space character 5
0079 Space character 5

6
0080 Temperature calibration slope
0081 Temperature calibration slope
0082 Temperature calibration slope
0083 Temperature calibration slope – Single-write
6 √ √ √ √ √ √ √
0084 Temperature calibration offset multiple
0085 Temperature calibration offset
0086 Temperature calibration offset
0087 Space character (optional)

0096 Transmitter description


0097 Transmitter description
0098 Transmitter description
0099 Transmitter description – Single-write
√ √ √ √ √ √ √
0100 Transmitter description multiple
0101 Transmitter description
0102 Transmitter description
0103 Transmitter description

30 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-5 ASCII registers continued Transmitter codes on page 3, footnotes on page 33

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address Description

0104 User message


0105 User message
0106 User message
0107 User message
0108 User message
0109 User message
0110 User message
0111 User message – Single-write
0112 User message √ √ √ √ √ √ √
multiple
0113 User message
0114 User message
0115 User message
0116 User message
0117 User message
0118 User message
0119 User message

Modbus Map
0298 Polling tag for external device #1 7
0299 Polling tag for external device #1 – Single-write
√ √ √ √ AIC √
0300 Polling tag for external device #1 multiple
0301 Polling tag for external device #1

0425 Sensor type


0426 Sensor type
0427 Sensor type
0428 Sensor type – Single-write
0429 Sensor type √ √ √ √ √ √
multiple
0430 Sensor type
0431 Sensor type
0432 Sensor type

0557 Enhanced density: Curven curve name – 12 characters max


0558 Enhanced density: Curven curve name – 12 characters max
0559 Enhanced density: Curven curve name – 12 characters max – Single-write
√8 √8 √8
0560 Enhanced density: Curven curve name – 12 characters max multiple
0561 Enhanced density: Curven curve name – 12 characters max
0562 Enhanced density: Curven curve name – 12 characters max

0573 Enhanced density: Curven concentration units label –


8 characters maximum
0574 Enhanced density: Curven concentration units label –
8 characters maximum – Single-write
√8 √8 √8
0575 Enhanced density: Curven concentration units label – multiple
8 characters maximum
0576 Enhanced density: Curven concentration units label –
8 characters maximum

1140 Polling tag for external device #2


1141 Polling tag for external device #2 – Single-write
√ √ √ AIC √
1142 Polling tag for external device #2 multiple
1143 Polling tag for external device #2

1265 Batch: Preset x name (For x, see register 1263)


1266 Batch: Preset x name
1267 Batch: Preset x name
1268 Batch: Preset x name
1269 Batch: Preset x name – Single-write
1270 Batch: Preset x name √9
1271 Batch: Preset x name multiple
1272 Batch: Preset x name
1273 Batch: Preset x name
1274 Batch: Preset x name
1275 Batch: Preset x name

Modbus Mapping Assignments for Micro Motion Transmitters 31


Modbus Mapping Assignments continued

Table 2-5 ASCII registers continued Transmitter codes on page 3, footnotes on page 33

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address Description

1371 Ticket header line 1


1372 Ticket header line 1
1373 Ticket header line 1
1374 Ticket header line 1
1375 Ticket header line 1 – Single-write
1376 Ticket header line 1 √
1377 Ticket header line 1 multiple
1378 Ticket header line 1
1379 Ticket header line 1
1380 Ticket header line 1
1381 Ticket header line 1

1384 Ticket header line 2


1385 Ticket header line 2
1386 Ticket header line 2
1387 Ticket header line 2
1388 Ticket header line 2 – Single-write
1389 Ticket header line 2 √
1390 Ticket header line 2 multiple
1391 Ticket header line 2
1392 Ticket header line 2
1393 Ticket header line 2
1394 Ticket header line 2

1397 Ticket header line 3 10


1398 Ticket header line 3
1399 Ticket header line 3
1400 Ticket header line 3
1401 Ticket header line 3 – Single-write
1402 Ticket header line 3 √
1403 Ticket header line 3 multiple
1404 Ticket header line 3
1405 Ticket header line 3
1406 Ticket header line 3
1407 Ticket header line 3

1410 Ticket header line 4 10


1411 Ticket header line 4
1412 Ticket header line 4
1413 Ticket header line 4
1414 Ticket header line 4 – Single-write
1415 Ticket header line 4 √
1416 Ticket header line 4 multiple
1417 Ticket header line 4
1418 Ticket header line 4
1419 Ticket header line 4
1420 Ticket header line 4

1423 Ticket footer


1424 Ticket footer
1425 Ticket footer
1426 Ticket footer
1427 Ticket footer – Single-write
1428 Ticket footer √
1429 Ticket footer multiple
1430 Ticket footer
1431 Ticket footer
1432 Ticket footer
1433 Ticket footer

1480 Audit trail recordn configuration value 10


1481 Audit trail recordn configuration value
1482 Audit trail recordn configuration value
1483 Audit trail recordn configuration value √
1484 Audit trail recordn configuration value – Read-only
1485 Audit trail recordn configuration value
1486 Audit trail recordn configuration value
1487 Audit trail recordn configuration value

32 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-5 ASCII registers continued Transmitter codes on page 3, footnotes on page 33

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Address Description

1489 Fieldbus tag


1490 Fieldbus tag
1491 Fieldbus tag
1492 Fieldbus tag
1493 Fieldbus tag
1494 Fieldbus tag
1495 Fieldbus tag
1496 Fieldbus tag – Single-write
1497 Fieldbus tag F 11
multiple
1498 Fieldbus tag
1499 Fieldbus tag
1500 Fieldbus tag
1501 Fieldbus tag
1502 Fieldbus tag
1503 Fieldbus tag
1504 Fieldbus tag

Modbus Map
2306 Well name – characters 1–2
2307 Well name – characters 3–4
2308 Well name – characters 5–6
2309 Well name – characters 7–8 – Single-write
N
2310 Well name – characters 9–10 multiple
2311 Well name – characters 11–12
2312 Well name – characters 13–14
2313 Well name – characters 15–16

(1) MVD Direct Connect or Series 1000, 2000, or 3000 transmitters.


(2) RFT9739 transmitters.
(3) Does not apply to v3.0 or higher transmitters with the FOUNDATION fieldbus outputs option board. If you have this transmitter, use
registers 1489–1504.
(4) The flow calibration factor is also stored in floating-point registers 0407–0410. Changing the data here will change the data in
registers 0407–0410, and vice versa.
(5) Required for RFT9739 transmitters; optional for all other transmitters.
(6) The temperature calibration factor is also stored in floating-point registers 0411–0414. Changing the data here will change the data
in registers 0411–0414, and vice versa.
(7) Polling tag for Series 1000, 2000, or 3000 transmitters; for RFT9739 transmitters, HART tag for pressure input host.
(8) Requires enhanced density application.
(9) Requires discrete batch application.
(10) Requires custody transfer application.
(11) Transmitter software v3.0 and higher.

Modbus Mapping Assignments for Micro Motion Transmitters 33


Modbus Mapping Assignments continued

2.6 Integer codes

Table 2-6 Sensor information codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Sensor type codes (register 1139)


0 Curved-tube sensor (all sensors except T-Series) √ √ √ √ √
1 Straight-tube sensor (T-Series) √ √ √ √ √

Sensor flange type codes (register 0129)


0 ANSI 150 √ √ √ √ √ √ √
1 ANSI 300 √ √ √ √ √ √ √
2 ANSI 600 √ √ √ √ √ √ √
5 PN 40 √ √ √ √ √ √ √
7 JIS 10K √ √ √ √ √ √ √
8 JIS 20K √ √ √ √ √ √ √
9 ANSI 900 √ √ √ √ √ √ √
10 Sanitary clamp √ √ √ √ √ √ √
11 Union √ √ √ √ √ √ √
12 PN 100 √ √ √ √ √ √
250 Reserved √
251 None √ √ √ √ √ √
252 Unknown √ √ √ √ √ √ √
253 Special √ √ √ √ √ √ √
254 Reserved √
255 Reserved √

Flow tube construction material codes (register 0130)


3 Hastelloy C-22 √ √ √ √ √ √ √
4 Monel √ √ √ √ √ √ √
5 Tantalum √ √ √ √ √ √ √
6 Titanium √ √ √ √ √ √
19 316L stainless steel √ √ √ √ √ √ √
23 Inconel √ √ √ √ √ √
252 Unknown √ √ √ √ √ √ √
253 Special √ √ √ √ √ √ √

Flow tube liner material codes (register 0131)


10 PTFE (Teflon) √ √ √ √ √ √ √
11 Halar √ √ √ √ √ √ √
16 Tefzel √ √ √ √ √ √
251 None √ √ √ √ √ √ √
252 Unknown √ √ √ √ √ √ √
253 Special √ √ √ √ √ √ √

Table 2-7 Outputs option board codes (register 1138)


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 None √
1 Analog I/O (mA/Frequency/RS-485) A A
2 FOUNDATION fieldbus (H1) or Profibus-PA FP
3 Intrinsically safe output I I
4 Configurable input/output √ √ C

34 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-8 Digital communication codes

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Digital communication protocol codes (register 1132)


0 None √ A √ A √
1 HART A A √
2 Modbus RTU √ A √ A √
3 Modbus ASCII √ A √ A √
4 Printer √

Digital communication baud rate codes (register 1133)


0 1200 baud √ A √ A √
1 2400 baud √ A √ A √
2 4800 baud √ A √ A √
3 9600 baud √ A √ A √
4 19,200 baud √ A √ A √

Modbus Map
5 38,400 baud √ A √ A √

Digital communication parity codes (register 1134)


0 None √ A √ A √
1 Odd parity √ A √ A √
2 Even parity √ A √ A √

Digital communication stop bits codes (register 1135)


1 1 stop bit √ A √ A √
2 2 stop bits √ A √ A √

Table 2-9 Byte order codes (register 0521)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 0-1-2-3 (or 1-2-3-4) √ √ √ √ √ √
1 2-3-0-1 (or 3-4-1-2) (Default) √ √ √ √ √ √
2 1-0-3-2 (or 2-1-4-3) √ √ √ √ √ √
3 3-2-1-0 (or 4-3-2-1) √ √ √ √ √ √

Table 2-10 100 Hz variable codes (register 1164)


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 Mass flow rate √ √ √ AIC
1 Temperature √ √ √ AIC
3 Density √ √ √ AIC
5 Volume flow rate √ √ √ AIC
10 Event 1 √ √ √ AIC
11 Event 2 √ √ √ AIC
15 API: Temperature corrected density √1 √1 AIC 1
16 API: Temperature corrected (standard) volume flow √1 √1 AIC 1
19 API: Batch weighted average density √1 √1 AIC 1
20 API: Batch weighted average temperature √1 √1 AIC 1
(1) Requires petroleum measurement application (API feature).

Modbus Mapping Assignments for Micro Motion Transmitters 35


Modbus Mapping Assignments continued

Table 2-11 Burst output codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Burst command option codes (register 1165)


1 Read primary variable √ √ √ AIC √
2 Read PV current and percent of range √ √ √ AIC √
3 Read dynamic variables and PV current √ √ √ AIC √
33 Read transmitter variables √ √ √ AIC √

Burst variable codes for command 33 (registers 1169–1172)


0 Mass flow rate √ √ √ √ √
1 Temperature √ √ √
2 Mass total √ √ √ √ √
3 Density √ √ √
4 Mass inventory √ √ √ √ √
5 Volume flow rate √ √ √ √ √
6 Volume total √ √ √ √ √
7 Volume inventory √ √ √ √ √
15 API: Temperature corrected density √1 √1 √1
16 API: Temperature corrected (standard) volume flow √1 √1 √1
17 API: Temperature-corrected (standard) volume total √1 √1 √1
18 API: Temperature-corrected (standard) volume inventory √1 √1 √1
19 API: Batch weighted average temperature √1 √1 √1
20 API: Batch weighted average temperature √1 √1 √1
21 Enhanced density: Density at reference temperature √2 √2 √2
22 Enhanced density: Density (fixed SG units) √2 √2 √2
23 Enhanced density: Standard volume flow rate √2 √2 √2
24 Enhanced density: Standard volume total √2 √2 √2
25 Enhanced density: Standard volume inventory √2 √2 √2
26 Enhanced density: Net mass flow rate √2 √2 √2
27 Enhanced density: Net mass total √2 √2 √2
28 Enhanced density: Net mass inventory √2 √2 √2
29 Enhanced density: Net volume flow rate √2 √2 √2
30 Enhanced density: Net volume total √2 √2 √2
31 Enhanced density: Net volume inventory √2 √2 √2
32 Enhanced density: Concentration √2 √2 √2
33 API: CTL √1 √1 √1
46 Tube frequency √ √ √
47 Drive gain √ √ √
48 Meter temperature (T-Series sensors only) √ √ √
49 Left pickoff √ √ √
50 Right pickoff √ √ √
51 Board temperature √ √ √
52 Input voltage √ √ √
53 External pressure √ √ √
55 External temperature √ √ √
63 Gas standard volume flow rate √3 √3 √3 √3 √4
64 Gas standard volume total √3 √3 √3 √3 √4
65 Gas standard volume inventory √3 √3 √3 √3 √4
91 Frequency input flow rate √
92 Frequency input total √
93 Frequency input inventory √
94 Frequency input √
(1) Requires petroleum measurement application (API feature).
(2) Requires enhanced density application.
(3) Transmitter software v4.1 and higher.
(4) Transmitter software v6.1 and higher.

36 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-12 Quaternary variable codes (register 0015)

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Mass flow rate √ √ √ √ √ √ √
1 Temperature √ √ √ √ √
2 Mass total √ √ √ √ √ √ √
3 Density √ √ √ √ √
4 Mass inventory √ √ √ √ √ √ √
5 Volume flow rate √ √ √ √ √ √ √
6 Volume total √ √ √ √ √ √ √
7 Volume inventory √ √ √ √ √ √ √
8 Viscosity √1
15 API: Temperature corrected density √2 √2 √2 √2
16 API: Temperature corrected (standard) volume flow √2 √2 √2 √2
17 API: Temperature-corrected (standard) volume total √2 √2 √2 √2
18 API: Temperature-corrected (standard) volume inventory √2 √2 √2 √2
19 API: Batch weighted average temperature √2 √2 √2 √2
√2 √2 √2 √2

Modbus Map
20 API: Batch weighted average temperature
21 Enhanced density: Density at reference temperature √3 √3 √3
22 Enhanced density: Density (fixed SG units) √3 √3 √3
23 Enhanced density: Standard volume flow rate √3 √3 √3
24 Enhanced density: Standard volume total √3 √3 √3
25 Enhanced density: Standard volume inventory √3 √3 √3
26 Enhanced density: Net mass flow rate √3 √3 √3
27 Enhanced density: Net mass total √3 √3 √3
28 Enhanced density: Net mass inventory √3 √3 √3
29 Enhanced density: Net volume flow rate √3 √3 √3
30 Enhanced density: Net volume total √3 √3 √3
31 Enhanced density: Net volume inventory √3 √3 √3
32 Enhanced density: Concentration √3 √3 √3
33 API: CTL √2 √2 √2 √2
46 Tube frequency √ √ √ √
47 Drive gain √ √ √ √
48 Meter temperature (T-Series sensors only) √ √ √ √
49 Left pickoff √ √ √ √
50 Right pickoff √ √ √ √
51 Board temperature √ √ √ √
52 Input voltage √ √ √ √
53 External pressure √ √ √ √
55 External temperature √ √ √ √
63 Gas standard volume flow rate √ √4 √4 √4 √4 √5
64 Gas standard volume total √ √4 √4 √4 √4 √5
65 Gas standard volume inventory √ √4 √4 √4 √4 √5
91 Frequency input flow rate √
92 Frequency input total √
93 Frequency input inventory √
94 Frequency input √
(1) RFT9739 transmitters v3.7 and lower.
(2) Requires petroleum measurement application (API feature).
(3) Requires enhanced density application.
(4) Transmitter software v4.1 and higher, or for transmitters with the FOUNDATION fieldbus outputs option board, transmitter software
v3.0 and higher.
(5) Transmitter software v6.1 and higher.

Modbus Mapping Assignments for Micro Motion Transmitters 37


Modbus Mapping Assignments continued

Table 2-13 Polling codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Polling control option codes (registers 0302 and 1144)


0 Do not poll √ √ √ √ √
1 Poll transmitter: HART primary √ √ √ √ √
2 Poll transmitter: HART secondary √ √ √ √ √

Polling type option codes (register 1147)


0 None √ √ √ √ √
1 Pressure compensation only √ √ √ √ √
2 Enhanced density temperature compensation √1 √1 √1
4 API temperature √2 √2 √2
8 API pressure compensation √2 √2 √2 √2

Polled variable codes (registers 1145 and 1146)


53 Externally read pressure √ √ √ √ √
55 Externally read temperature √ √ √ √ √
(1) Requires enhanced density application.
(2) Requires petroleum measurement application (API feature).

Table 2-14 Channel/terminal configuration codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description

Output type codes (registers 1166–1168)


0 Milliamp (primary) output √ √ √ √
1 Frequency output √ √ √ √
2 Digital communications A A
3 Milliamp (secondary) output √ IC
4 Discrete output √1 √ √
5 Discrete input √1 √ C
6 None I

Power source codes (registers 1174 and 1175)


0 External (passive) √ √ C √
1 Internal (active) √ √ C √

Polarity codes (registers 1152, 1154, 1156, 1178, 1179, 1197)


0 Active low √1 √ AIC √
1 Active high √1 √ AIC √

Frequency output mode codes (register 1181)


0 Single √ √ √ AIC √
1 Quadrature √ C
2 Dual with 0° phase shift √ C
3 Dual with 180° phase shift √ C
4 Dual with +90° phase shift √ C
5 Dual with –90° phase shift √ C
(1) Requires filling and dosing application.

38 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-15 mA output process variable assignment codes (registers 0012, 0013)

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Mass flow rate √ √ √ √ √ √ √
1 Temperature √ √ √ √ √
3 Density √ √ √ √ √
5 Volume flow rate √ √ √ √ √ √ √
9 Pressure √
10 Event 1 √
11 Event 2 √
15 API: Temperature-corrected density √1 √1 √1 √1
16 API: Temperature-corrected (standard) volume flow √1 √1 √1 √1
19 API: Batch weighted average density √1 √1 √1 √1
20 API: Batch weighted average temperature √1 √1 √1 √1
21 Enhanced density: Density at reference √2 √2 √2
22 Enhanced density: Density (fixed SG units) √2 √2 √2
23 Enhanced density: Standard volume flow rate √2 √2 √2
√2 √2 √2

Modbus Map
26 Enhanced density: Net mass flow rate
29 Enhanced density: Net volume flow rate √2 √2 √2
32 Enhanced density: Concentration √2 √2 √2
47 Drive gain √ √ √
56 Fixed Baume units √2 √2 √2
62 Gas standard volume flow rate √ √3 √3 √3 √3 √4
70 Drive gain N
73 Actual uncorrected oil flow N
74 Actual uncorrected water cut N
75 Actual uncorrected water flow N
76 Actual uncorrected gross flow N
77 Actual uncorrected back flow N
78 Net oil flow N
79 Water cut N
80 Gross flow N
81 Net water flow N
82 Average uncorrected oil flow N
83 Average uncorrected water cut N
84 Average uncorrected gross flow N
85 Average uncorrected water flow N
86 Average back flow N
87 Average net oil flow N
88 Average water cut N
89 Average gross flow N
90 Average net water flow N
91 Frequency input flow rate √
110 Batch/fill: Primary valve √5
111 Batch/fill: Secondary valve √5
(1) Requires petroleum measurement application (API feature).
(2) Requires enhanced density application.
(3) Transmitter software v4.1 and higher.
(4) Transmitter software v6.1 and higher.
(5) Requires filling and dosing application.

Modbus Mapping Assignments for Micro Motion Transmitters 39


Modbus Mapping Assignments continued

Table 2-16 Frequency/pulse input/output codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Frequency/pulse output variable codes (register 0014)


0 Mass flow rate √ √ √ √ √ √ √
2 Mass totalizer √
5 Volume flow rate √ √ √ √ √ √ √
6 Volume totalizer √
16 API: Temperature-corrected (standard) volume flow √1 √1 √1 √1
23 Enhanced density: Standard volume flow rate √2 √2 √2
26 Enhanced density: Net mass flow rate √2 √2 √2
29 Enhanced density: Net volume flow rate √2 √2 √2
57 Frequency input N
62 Gas standard volume flow rate √3 √3 √3 √3 √4
73 Actual uncorrected oil flow N
75 Actual uncorrected water flow N
77 Actual uncorrected back flow N
78 Net oil flow N
80 Gross flow N
81 Net water flow N
91 Frequency input flow rate √

Frequency/pulse input/output scaling method codes (registers 1108 and 1122)


0 Frequency=flow √ √ √ √ √
1 Pulses/unit √ √ √ √ √
2 Units/pulse √ √ √ √ √
(1) Requires petroleum measurement application (API feature).
(2) Requires enhanced density application.
(3) Transmitter software v4.1 and higher.
(4) Transmitter software v6.1 and higher.

Table 2-17 Fault output codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description

Version 2 RFT9739 fault output codes (register 0124) 1


0 Upscale √
1 Downscale √
2 Last measured value √
3 Internal zero √

Analog fault output codes (registers 1113, 1114, 1107)


0 Upscale √ √ √ √ √
1 Downscale √ √ √ √ √
2 Last measured value N
3 Internal zero √ √ √ √ √
4 None √ √ √ √ √

40 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-17 Fault output codes continued

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Digital communications fault output codes (register 0124)


0 Upscale (Hold at value greater than upper sensor limit, stop totalizing) √ √ √ √ √ √
1 Downscale (Hold at value less than lower sensor limit, stop totalizing) √ √ √ √ √ √
2 Internal zero (Drive outputs to zero values of process variables, stop √ √ √ √ √ √
totalizing)
3 NAN (Report not-a-number or maximum scaled integer, stop totalizing) √ √ √ √ √ √
4 Flow zero (Drive flow rate to zero value, other process variables remain √ √ √ √ √ √
unaffected)
5 None (default; use status bits for fault detection) √ √ √ √ √ √
(1) Read-only for transmitters v3 and higher. On these transmitters, the fault output must be configured using hardware switches or the
display.

Modbus Map
Table 2-18 Control output assignment codes (register 0015)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Forward/reverse flow √
1 Zero in progress √
2 Faults √
3 Event 1 √
4 Event 2 √

Table 2-19 Discrete output assignment codes (registers 1151, 1153, 1155)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
10 Event 1 √ AIC
11 Event 2 √ AIC
54 Discrete input 1 N
55 Discrete input 2 N
57 Discrete event 1 (x=0) (For x, see register 0609) √1 AIC 1 √
58 Discrete event 2 (x=1) (For x, see register 0609) √1 AIC 1 √
59 Discrete event 3 (x=2) (For x, see register 0609) √1 AIC 1 √
60 Discrete event 4 (x=3) (For x, see register 0609) √1 AIC 1 √
61 Discrete event 5 (x=4) (For x, see register 0609) √1 AIC 1 √
72 Fault alarm N
97 Transient bubble remediation N
98 Discrete input 1 √
99 Discrete input 2 √
100 Event 1 or event 2 √ AIC
101 Flow switch indication √ √ √
102 Forward/reverse indication √ √ √
103 Calibration in progress √ √ √
104 Fault condition indication √2 √ √ √
105 Batch/fill: Batch/fill timeout √3
106 Batch/fill: Batching/filling in progress √2 √3

Modbus Mapping Assignments for Micro Motion Transmitters 41


Modbus Mapping Assignments continued

Table 2-19 Discrete output assignment codes (registers 1151, 1153, 1155) continued

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
107 Batch: Batch end warning √3
108 Batch: Batch overrun √3
109 Batch: Batch pump √3
110 Batch/fill: Primary valve √2 √3
111 Batch/fill: secondary valve √2 √3
251 None √3
(1) Transmitter software v4.1 and higher.
(2) Requires filling and dosing application.
(3) Requires discrete batch application.

Table 2-20 Discrete input assignment codes – Series 1000/2000 transmitters (register 1176)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 None √1 √ C
1 Start sensor zero √ C
2 Reset mass total √1 √ C
3 Reset volume total √1 √ C
4 Reset API reference (standard) volume total √2 C2
5 Reset all totals √1 √ C
6 Reset enhanced density reference (corrected) volume total √3 C3
7 Reset enhanced density net mass total √3 C3
8 Reset enhanced density net volume total √3 C3
9 Begin fill √1
10 End fill √1
11 Pause fill √1
12 Resume fill √1
13 Reset fill total √1
(1) Requires filling and dosing application
(2) Requires petroleum measurement application (API feature).
(3) Requires enhanced density application.

Table 2-21 Triggering conditions codes – Series 3000 transmitters


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
57 Discrete event 1 (x=0) √
58 Discrete event 2 (x=1) √
59 Discrete event 3 (x=2) √
60 Discrete event 4 (x=3) √
61 Discrete event 5 (x=4) √
98 Discrete input 1 √
99 Discrete input 2 √
105 Batch: Batch timeout √1
106 Batch: Batching √1
107 Batch: Batch end warning √1

42 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-21 Triggering conditions codes – Series 3000 transmitters continued

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
108 Batch: Batch overrun √1
109 Batch: Batch pump √1
110 Batch: Batch primary valve √1
111 Batch: Batch secondary valve √1
251 None √1
(1) Requires discrete batch application.

Table 2-22 Fieldbus AI function block codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Modbus Map
AI mode codes (register 1506)
1 Out of service F
16 Auto F

AI channels codes (register 1508)


0 Uninitialized F
1 Mass flow F
2 Temperature F
3 Density F
4 Volume flow F
5 Drive gain F
6 Not used F
7 API: Temperature-corrected density F
8 API: Temperature-corrected (standard) volume flow F
9 API: Batch weighted average density F
10 API: Batch weighted average temperature F
11 API: CTL F
12 Enhanced density: Density at reference F
13 Enhanced density: Density (fixed SG units) F
14 Enhanced density: Standard volume flow rate F
15 Enhanced density: Net mass flow rate F
16 Enhanced density: Net volume flow rate F
17 Enhanced density: Concentration F
18 Enhanced density: Density (fixed Baume units) F
19 Gas standard volume F

AI linearization type codes (register 1510)


0 Initial value F
1 Direct (only value that can be written) F
2 Indirect F
3 Indirect square root F

Modbus Mapping Assignments for Micro Motion Transmitters 43


Modbus Mapping Assignments continued

Table 2-23 Fieldbus integrator function block codes (register 1511)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Standard F
1 Internal mass total mode F
2 Internal volume total mode F
3 Internal mass inventory mode F
4 Internal volume inventory mode F
5 Gas standard volume total F
6 Gas standard volume inventory F
7 API: Temperature-corrected (standard) volume total F
8 API: Temperature-corrected (standard) volume inventory F
9 Enhanced density: Standard volume total F
10 Enhanced density: Standard volume inventory F
11 Enhanced density: Net mass total F
12 Enhanced density: Net mass inventory F
13 Enhanced density: Net volume total F
14 Enhanced density: Net volume infentory F

Table 2-24 Discrete output fixed state codes (registers 1182, 1183, 1184)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

Code Description MVD DC


0 Off √1 √ √ √
1 On √1 √ √ √
255 Unfix discrete output √1 √ √ √
(1) Requires filling and dosing application.

Table 2-25 Mass flow codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description

Mass flow unit codes (register 0039)


70 Grams/second √ √ √ √ √ √ √
71 Grams/minute √ √ √ √ √ √ √
72 Grams/hour √ √ √ √ √ √ √
73 Kilograms/second √ √ √ √ √ √ √
74 Kilograms/minute √ √ √ √ √ √ √
75 Kilograms/hour √ √ √ √ √ √ √
76 Kilograms/day √ √ √ √ √ √ √
77 Metric tons/minute √ √ √ √ √ √ √
78 Metric tons/hour √ √ √ √ √ √ √
79 Metric tons/day √ √ √ √ √ √ √
80 Pounds/second √ √ √ √ √ √ √
81 Pounds/minute √ √ √ √ √ √ √
82 Pounds/hour √ √ √ √ √ √ √
83 Pounds/day √ √ √ √ √ √ √

44 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-25 Mass flow codes continued

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
84 Short tons (2000 pounds)/minute √ √ √ √ √ √ √
85 Short tons (2000 pounds)/hour √ √ √ √ √ √ √
86 Short tons (2000 pounds)/day √ √ √ √ √ √ √
87 Long tons (2240 pounds)/hour √ √ √ √ √ √
88 Long tons (2240 pounds)/day √ √ √ √ √ √
253 Special √ √ √ √ √ √ √

Mass totalizer or mass inventory unit codes (register 0045)


60 Grams √ √ √ √ √ √ √
61 Kilograms √ √ √ √ √ √ √
62 Metric tons √ √ √ √ √ √ √
63 Pounds √ √ √ √ √ √ √
64 Short tons (2000 pounds) √ √ √ √ √ √ √
65 Long tons (2240 pounds) √ √ √ √ √ √
253 Special √ √ √ √ √ √ √

Modbus Map
Table 2-26 Volume flow codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Volume flow unit codes (register 0042)


15 Cubic feet/minute √ √ √ √ √ √ √
16 U.S. gallons/minute √ √ √ √ √ √ √
17 Liters/minute √ √ √ √ √ √ √
18 Imperial gallons/minute √ √ √ √ √ √ √
19 Cubic meters/hour √ √ √ √ √ √ √
22 U.S. gallons/second √ √ √ √ √ √ √
23 Million U.S. gallons/day √ √ √ √ √ √
24 Liters/second √ √ √ √ √ √ √
25 Million liters/day √ √ √ √ √ √
26 Cubic feet/second √ √ √ √ √ √ √
27 Cubic feet/day √ √ √ √ √ √
28 Cubic meters/second √ √ √ √ √ √ √
29 Cubic meters/day √ √ √ √ √ √ √
30 Imperial gallons/hour √ √ √ √ √ √ √
31 Imperial gallons/day √ √ √ √ √ √ √
130 Cubic feet/hour √ √ √ √ √ √ √
131 Cubic meters/minute √ √ √ √ √ √ √
132 Barrels/second (42 U.S. gallons) √ √ √ √ √ √ √
133 Barrels/minute (42 U.S. gallons) √ √ √ √ √ √ √
134 Barrels/hour (42 U.S. gallons) √ √ √ √ √ √ √
135 Barrels/day (42 U.S. gallons) √ √ √ √ √ √ √
136 U.S. gallons/hour √1 √ √ √ √ √ √
137 Imperial gallons/second √1 √ √ √ √ √ √
138 Liters/hour √1 √ √ √ √ √ √
235 U.S. gallons/day √ √ √ √ √ √
253 Special √ √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 45


Modbus Mapping Assignments continued

Table 2-26 Volume flow codes continued

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Volume totalizer or volume inventory unit codes (register 0046)


40 U.S. gallons √ √ √ √ √ √ √
41 Liters √ √ √ √ √ √ √
42 Imperial gallons √ √ √ √ √ √ √
43 Cubic meters √ √ √ √ √ √ √
46 Barrels (42 U.S. gallons) √ √ √ √ √ √ √
112 Cubic feet √ √ √ √ √ √ √
253 Special √ √ √ √ √ √ √
(1) Version 3.x RFT9739 transmitters only.

Table 2-27 Density codes (register 0040)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
90 Specific gravity units √ √ √ √ √ √ √
91 Grams/cubic centimeter √ √ √ √ √ √ √
92 Kilograms/cubic meter √ √ √ √ √ √ √
93 Pounds/U.S. gallon √ √ √ √ √ √ √
94 Pounds/cubic foot √ √ √ √ √ √
95 Grams/milliliter √ √ √ √ √ √
96 Kilograms/liter √ √ √ √ √ √
97 Grams/liter √ √ √ √ √ √
98 Pounds/cubic inch √ √ √ √ √ √
99 Short tons (2000 pounds)/cubic yard √ √ √ √ √ √
102 Degrees Baume (heavy) √1 √1 √1
103 Degrees Baume (light) √1 √1 √1
104 Degrees API √ √ √ √ √
(1) Requires enhanced density application.

Table 2-28 Temperature codes (register 0041)


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
32 Degrees Celsius √ √ √ √ √ √ √
33 Degrees Fahrenheit √ √ √ √ √ √ √
34 Degrees Rankine √ √ √ √ √ √ √
35 Degrees Kelvin √ √ √ √ √ √ √

46 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-29 Pressure codes (register 0044)

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
1 Inches water at 68 °Fahrenheit √ √ √ √ √ √ √
2 Inches mercury at 0 °Celsius √ √ √ √ √ √ √
3 Feet water at 68 °Fahrenheit √ √ √ √ √ √ √
4 Millimeters water at 68 °Fahrenheit √ √ √ √ √ √ √
5 Millimeters mercury at 0 °Celsius √ √ √ √ √ √ √
6 Pounds/square inch √ √ √ √ √ √ √
7 Bar √ √ √ √ √ √ √
8 Millibar √ √ √ √ √ √ √
9 Grams/square centimeter √ √ √ √ √ √ √
10 Kilograms/square centimeter √ √ √ √ √ √ √
11 Pascals √ √ √ √ √ √ √
12 Kilopascals √ √ √ √ √ √ √
13 Torr at 0 °Celsius √ √ √ √ √ √ √
14 Atmospheres √ √ √ √ √ √ √

Modbus Map
Table 2-30 Flow direction codes (register 0017)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Forward flow only √ √ √ √ √ √ √
1 Reverse flow only √ √ √ √ √ √ √
2 Bidirectional flow √ √ √ √ √ √ √
3 Absolute forward/reverse √ √ √ √ √ √
4 Negate/Forward only √ √ √ √ √ √
5 Negate/Bidirectional √ √ √ √ √ √

Table 2-31 Special measurement unit codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description

Base mass unit codes for special mass units (register 0132)
60 Grams √ √ √ √ √ √ √
61 Kilograms √ √ √ √ √ √ √
62 Metric tons √ √ √ √ √ √ √
63 Pounds √ √ √ √ √ √ √
64 Short tons (2000 pounds) √ √ √ √ √ √ √
65 Long tons (2240 pounds) √ √ √ √ √ √

Base time unit codes for special mass units (register 0133)
50 Minutes √ √ √ √ √ √ √
51 Seconds √ √ √ √ √ √ √
52 Hours √ √ √ √ √ √ √
53 Days √ √ √ √ √ √ √

Base volume unit codes for special volume units (register 0134)
40 U.S. gallons √ √ √ √ √ √ √
41 Liters √ √ √ √ √ √ √
42 Imperial gallons √ √ √ √ √ √ √
43 Cubic meters √ √ √ √ √ √ √
46 Barrels (42 U.S. gallons) √ √ √ √ √ √ √
112 Cubic feet √ √ √ √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 47


Modbus Mapping Assignments continued

Table 2-31 Special measurement unit codes continued

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Base time unit codes for special volume units (register 0135)
50 Minutes √ √ √ √ √ √ √
51 Seconds √ √ √ √ √ √ √
52 Hours √ √ √ √ √ √ √
53 Days √ √ √ √ √ √ √

Table 2-32 Display variable codes (registers 1117–1131, 1362–1366)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Mass flow rate √ √ √
1 Temperature √ √ √
2 Mass totalizer √ √ √
3 Density √ √ √
4 Mass inventory √ √ √
5 Volume flow rate √ √ √
6 Volume totalizer √ √ √
7 Volume inventory √ √ √
15 API: Temperature-corrected density √1 √1
16 API: Temperature-corrected (standard) volume flow √1 √1
17 API: Temperature-corrected (standard) volume total √1 √1
18 API: Temperature-corrected (standard) volume inventory √1 √1
19 API: Batch-weighted average corrected density √1 √1
20 API: Batch-weighted average temperature √1 √1
21 Enhanced density: Density at reference √2 √2
22 Enhanced density: Density (fixed SG units) √2 √2
23 Enhanced density: Standard volume flow rate √2 √2
24 Enhanced density: Standard volume total √2 √2
25 Enhanced density: Standard volume inventory √2 √2
26 Enhanced density: Net mass flow rate √2 √2
27 Enhanced density: Net mass total √2 √2
28 Enhanced density: Net mass inventory √2 √2
29 Enhanced density: Net volume flow rate √2 √2
30 Enhanced density: Net volume total √2 √2
31 Enhanced density: Net volume inventory √2 √2
32 Enhanced density: Concentration √2 √2
33 API: CTL √1 √1
46 Raw tube frequency √ √ √
47 Drive gain √ √ √
48 Meter temperature (T-Series) √ √ √
49 Left pickoff amplitude √ √ √
50 Right pickoff amplitude √ √ √
51 Board temperature √ √ √
52 Input voltage √ √ √
53 Externally read pressure √ √ √
54 Mechanical zero √ √ √
55 Externally read temperature √ √ √
56 Enhanced density: Density (fixed Baume units) √2 √2
57 Discrete event 1 (x=0) (For x, see register 0609) √
58 Discrete event 2 (x=1) (For x, see register 0609) √
59 Discrete event 3 (x=2) (For x, see register 0609) √
60 Discrete event 4 (x=3) (For x, see register 0609) √

48 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-32 Display variable codes (registers 1117–1131, 1362–1366) continued

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
61 Discrete event 5 (x=4) (For x, see register 0609) √
63 Gas standard volume flow rate √3 √3 √4
64 Gas standard volume total √3 √3 √4
65 Gas standard volume inventory √3 √3 √4
69 Live zero √5 √5 √
91 Frequency input flow rate √
92 Frequency input total √
93 Frequency input inventory √
94 Raw frequency input frequency √
98 Discrete input 1 √
99 Discrete input 2 √
101 Flow switch indicator √
102 Forward/reverse indication √

Modbus Map
103 Calibration in progress
104 Fault condition indication √
105 Batch: Batch timeout √
106 Batch: Batching √
107 Batch: Batch end warning √
108 Batch: Batch overrun √
109 Batch: Batch pump √
110 Batch: Batch primary valve √
111 Batch: Batch secondary valve √
251 None √ √ √
(1) Requires petroleum measurement application (API feature).
(2) Requires enhanced density application.
(3) Transmitter software v4.1 and higher, or for transmitters with the FOUNDATION fieldbus outputs option board, transmitter software
v3.0 and higher.
(4) Transmitter software v6.1 and higher.
(5) Transmitter software v3.5 and higher.

Table 2-33 API reference table codes (register 0351)


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
17 Table 5A √ √ √ √
18 Table 5B √ √ √ √
19 Table 5D √ √ √ √
36 Table 6C √ √ √ √
49 Table 23A √ √ √ √
50 Table 23B √ √ √ √
51 Table 23D √ √ √ √
68 Table 24C √ √ √ √
81 Table 53A √ √ √ √
82 Table 53B √ √ √ √
83 Table 53D √ √ √ √
100 Table 54C √ √ √ √

Modbus Mapping Assignments for Micro Motion Transmitters 49


Modbus Mapping Assignments continued

Table 2-34 Batch/fill flow source codes (register 1251)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Mass flow rate √1 √
5 Volume flow fate √1 √
16 API: Temperature-corrected (standard) volume flow √2
23 Enhanced density: Standard volume flow rate √3
26 Enhanced density: Net mass flow rate √3
29 Enhanced density: Net volume flow rate √3
91 Frequency input flow rate √
251 None √1 √
(1) Requires filling and dosing application.
(2) Requires petroleum measurement application (API feature).
(3) Requires enhanced density application.

Table 2-35 Enhanced density application codes

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Derived variable codes (register 0524)


0 Standard √ √ √
1 Density at reference temperature √ √ √
2 Specific gravity √ √ √
3 Mass concentration (density) √ √ √
4 Mass concentration (SG) √ √ √
5 Volume concentration (density) √ √ √
6 Volume concentration (SG) √ √ √
7 Concentration (density) √ √ √
8 Concentration (SG) √ √ √

Concentration unit label codes (register 0570)


100 Degrees Twaddell √ √ √
101 Degrees Brix √ √ √
102 Degrees Baume (heavy) √ √ √
103 Degrees Baume (light) √ √ √
105 Percent solids per weight (%mass) √ √ √
106 Percent solids per volume (%volume) √ √ √
107 Degrees Balling √ √ √
108 Proof per volume √ √ √
109 Proof per mass √ √ √
160 Percent Plato √ √ √
255 Special √ √ √

50 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-36 Event codes

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description

Event variable codes (registers 0137, 0138, 0615)


0 Mass flow rate √ √ √ √ √ AIC
1 Temperature √ √ √ √ √ AIC
2 Mass totalizer √ √ √ √ √ AIC
3 Density √ √ √ √ √ AIC
4 Mass inventory √ √ √ √ √ AIC
5 Volume flow rate √ √ √ √ √ AIC
6 Volume totalizer √ √ √ √ √ AIC
7 Volume inventory √ √ √ √ √ AIC

Event type codes (downward compatibility) (registers 0139 and 0140)


1 HI √ √ √ √ √ AIC
2 LO √ √ √ √ √ AIC

Modbus Map
Event type codes (register 0610)
0 HI √1 √1 √1 √1 AIC 1 √
1 LO √1 √1 √1 √1 AIC 1 √
2 IN HI/LO √1 √1 √1 √1 AIC 1 √
3 OUT HI/LO √1 √1 √1 √1 AIC 1 √
(1) Transmitter software v4.1 and higher.

Table 2-37 TBR event action codes (register 1706)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 Hold last value N
1 Stop well test N
2 Alarm only N

Table 2-38 Process variable codes for slot addresses (registers 0751–0782) Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 Mass flow rate √ √ √ √ √ √
1 Temperature √ √ √ √ √ √
2 Mass totalizer √ √ √ √ √ √
3 Density √ √ √ √ √ √
4 Mass inventory √ √ √ √ √ √
5 Volume flow rate √ √ √ √ √ √
6 Volume totalizer √ √ √ √ √ √
7 Volume inventory √ √ √ √ √ √
10 Event 1 √ √ √ √ AIC
11 Event 2 √ √ √ √ AIC
12 Status word 1 (419/420) √ √ √ √ √ √
13 Status word 2 (421/422) √ √ √ √ √ √
14 Status word 3 (423/424) √ √ √ √ √ √
15 API: Temperature-corrected density √1 √1 √1 √1
16 API: Temperature-corrected (standard) volume flow √1 √1 √1 √1
17 API: Temperature-corrected (standard) volume total √1 √1 √1 √1
18 API: Temperature-corrected (standard) volume inventory √1 √1 √1 √1

Modbus Mapping Assignments for Micro Motion Transmitters 51


Modbus Mapping Assignments continued

Table 2-38 Process variable codes for slot addresses (registers 0751–0782) continued

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
19 API: Batch-weighted average corrected density √1 √1 √1 √1
20 API: Batch-weighted average temperature √1 √1 √1 √1
21 Enhanced density: Density at reference √2 √2 √2
22 Enhanced density: Density (fixed SG units) √2 √2 √2
23 Enhanced density: Standard volume flow rate √2 √2 √2
24 Enhanced density: Standard volume total √2 √2 √2
25 Enhanced density: Standard volume inventory √2 √2 √2
26 Enhanced density: Net mass flow rate √2 √2 √2
27 Enhanced density: Net mass total √2 √2 √2
28 Enhanced density: Net mass inventory √2 √2 √2
29 Enhanced density: Net volume flow rate √2 √2 √2
30 Enhanced density: Net volume total √2 √2 √2
31 Enhanced density: Net volume inventory √2 √2 √2
32 Enhanced density: Concentration √2 √2 √2
33 API: CTL √1 √1 √1 √1
34 High-order doubleword of binary mass total in grams √ √ √ √ √ √
35 Low-order doubleword of binary mass total in grams √ √ √ √ √ √
36 High-order doubleword of binary volume total in cubic centimeters √ √ √ √ √ √
37 Low-order doubleword of binary volume total in cubic centimeters √ √ √ √ √ √
38 Raw 3 API: Temperature-corrected volume total, high-order doubleword √1 √1 √1 √1
39 Raw 3 API: Temperature-corrected volume total, low-order doubleword √1 √1 √1 √1
40 Raw 3 enhanced density: Standard volume total, high-order doubleword √2 √2 √2 √2
41 Raw 3 enhanced density: Standard volume total, low-order doubleword √2 √2 √2 √2
42 Raw 3 enhanced density: Net mass total, high-order doubleword √2 √2 √2 √2
43 Raw 3 enhanced density: Net mass total, low-order doubleword √2 √2 √2 √2
44 Raw 3 enhanced density: Net volume total, high-order doubleword √2 √2 √2 √2
45 Raw 3 enhanced density: Net volume total, low-order doubleword √2 √2 √2 √2
46 Raw tube frequency √ √ √ √ √ √
47 Drive gain √ √ √ √ √ √
48 Meter temperature √ √ √ √ √ √
49 Left pickoff amplitude √ √ √ √ √ √
50 Right pickoff amplitude √ √ √ √ √ √
51 Board temperature √ √ √ √ √ √
52 Input voltage √ √ √ √ √ √
53 Externally read pressure √ √ √ √ √ √
55 Externally read temperature √ √ √ √ √ √
56 Enhanced density: Density (fixed Baume units) √2 √2 √2
57 Discrete event 1 (x=0) (For x, see register 0609) √
58 Discrete event 2 (x=1) (For x, see register 0609) √
59 Discrete event 3 (x=2) (For x, see register 0609) √
60 Discrete event 4 (x=3) (For x, see register 0609) √
61 Discrete event 5 (x=4) (For x, see register 0609) √
69 Live zero flow √ √ √ √ √
91 Frequency input flow rate √
92 Frequency input total √
93 Frequency input inventory √
94 Raw frequency input frequency √
98 Discrete input 1 √
99 Discrete input 2 √
100 Event 1 or event 2 4 √ √ √ AIC
101 Flow switch indicator 4 √ √ √ √ √
102 Forward/reverse indication 4 √ √ √ √ √
103 Calibration in progress 4 √ √ √ √ √
104 Fault condition indication 4 √ √ √ √ √
105 Batch: Batch timeout √
106 Batch: Batching √
107 Batch: Batch end warning √
108 Batch: Batch overrun √

52 Modbus Mapping Assignments for Micro Motion Transmitters


Modbus Mapping Assignments continued

Table 2-38 Process variable codes for slot addresses (registers 0751–0782) continued

Using the Modbus Map


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
109 Batch: Batch pump √
110 Batch: Batch primary valve √
111 Batch: Batch secondary valve √
(1) Requires petroleum measurement application (API feature).
(2) Requires enhanced density application.
(3) “Raw” for this register means that the value is not presented in IEEE floating-point format, thus providing more resolution.
(4) Available only when mapped to a discrete output.

Table 2-39 Pressure input receiving method codes (register 0302)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Modbus Map
Code Description
0 Do not poll √
1 Poll DP HART Primary √1
2 Poll DP HART Secondary √1
3 Poll pressure HART Primary √1
4 Poll pressure HART Secondary √1
5 Use analog as DP √
6 Use analog as Pressure √
7 Use Modbus DP √1
8 Use Modbus Pressure √1
(1) Version 3.x transmitters only.

Table 2-40 Series 3000 display password codes (registers 1351–1358)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 Left arrow √
1 Right arrow √
2 Up arrow √
3 Down arrow √

Table 2-41 Language codes (register 1359)


Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 English √
1 German √
2 French √
3 Katakana √1
(1) Transmitter software v6.1 and higher.

Modbus Mapping Assignments for Micro Motion Transmitters 53


Modbus Mapping Assignments continued

Table 2-42 Printer type codes (register 1441)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 Epson √
1 Digitec √
2 Generic √
3 Dumb terminal √

Table 2-43 Print request codes (register 1442)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC
Code Description
0 None √
1 Process monitor √
2 Batch ticket √
4 All configuration data √
8 Alarm history √
16 Active alarm log √
32 Audit trail event log √

Table 2-44 Meter fingerprinting process variable index codes (register 0626)

Series 3000
Model 1500

Model 1700

Model 2500

Model 2700
RFT9739

MVD DC

Code Description
0 Mass flow rate √ √ √ √ √ √
1 Temperature √ √ √ √
3 Density √ √ √ √
5 Volume flow rate √ √ √ √ √ √
46 Tube frequency √ √ √ √
47 Drive gain √ √ √ √
48 Meter temperature (T-Series sensors only) √ √ √ √
49 Left pickoff √ √ √ √
50 Right pickoff √ √ √ √
51 Board temperature √ √ √ √
52 Input voltage √ √ √ √
54 Mechanical zero √ √ √ √

54 Modbus Mapping Assignments for Micro Motion Transmitters


©2004, Micro Motion, Inc. All rights reserved. P/N 20001741, Rev. B

*20001741*
For the latest Micro Motion product specifications, view the
PRODUCTS section of our web site at www.micromotion.com

Micro Motion Inc. USA


Worldwide Headquarters
7070 Winchester Circle
Boulder, Colorado 80301
T (303) 527-5200
(800) 522-6277
F (303) 530-8459

Micro Motion Europe Micro Motion Asia


Emerson Process Management Emerson Process Management
Wiltonstraat 30 1 Pandan Crescent
3905 KW Veenendaal Singapore 128461
The Netherlands Republic of Singapore
T +31 (0) 318 495 670 T (65) 6777-8211
F +31 (0) 318 495 689 F (65) 6770-8003

Micro Motion United Kingdom Micro Motion Japan


Emerson Process Management Limited Emerson Process Management
Horsfield Way Shinagawa NF Bldg. 5F
Bredbury Industrial Estate 1-2-5, Higashi Shinagawa
Stockport SK6 2SU U.K. Shinagawa-ku
T 0800 966 180 Tokyo 140-0002 Japan
F 0800 966 181 T (81) 3 5769-6803
F (81) 3 5769-6843

Micro Motion
TM

You might also like