Image Production
Image Production
Image Production
Presented by:
John Fleming, M.Ed., RT(R)(MR)(CT)
St. Petersburg College
Office: (727) 341-3758
E-mail: [email protected]
Lesson Objectives:
ARRT Content Specifications
Image Production
Image Acquisition & Technical Evaluation
Equipment Operation & Quality Assurance
Midterm & Final Exam
https://www.arrt.org/Handbooks
2017
January-December 2017
ARRT Content Specifications:
Erased Cassette
CR vs. Background Radiation:
One Week
CR vs. Scatter Radiation:
CR Collimation Considerations:
If the collimated borders are not recognized,
the image data analysis will include all of the
data outside the collimation.
The final image may not have the correct level
of brightness and contrast.
Tightly collimated images should be placed in
the center of the IR.
Exposure Indicators vs. mAs (x):
Symbol ½(x) Perfect (x) 2(x)
Fuji/Konica Minolta/
S# 400 200 100
Siemans/Philips
S# 397 S# 201 S# 98
Which of these images would be most ideal?
Direct Radiography
Unit
Direct Radiography (DR):
DR employs a flat panel image receptor (IR)
that is about the size of a CR cassette.
Two Types of Flat Panel IRs:
Indirect Capture
Direct Capture
Flat Panel IR:
Flat Panel IR:
Indirect Capture Flat Panel IR:
Image Forming Beam
Surface Reflector
Scintillation Detector Array that
contains CsI or GOS
Digital Frequency
Raw Data
Raw Image
Histograms: Luminance Values
Bone Area
Direct Exposure Area
Digital Frequency
Bone Area
Soft Tissue
Area
Raw Data
Raw Data
Look-up-table Algorithm
Look-up-tables:
2048
0
0 1024 2048
Pixel Values from the Rescaled Image
Look-up-tables:
Rescaled No Change from
Image Original Image
AP Hip
Look-up-Table
Algorithm
0
0 1024 2048
Pixel Values from the Rescaled Image
Look-up-tables:
Processed Pixels
Original Pixels
45 11
AP Hip
600 Look-up-Table 725
Algorithm
1420 2011
AP Hip
Look-up-Table
Algorithm
Histogram LUT
Image Quality vs. CR Image Processing:
kVp: 55 65 75
mAs: 5 2.3 1.2
S#: 111 109 106
Contrast 2.68 2.68 2.68
Note how image quality remained fairly constant.
Digital Optimum kVp Values:
Chest (Grid) 110 to 130
Chest (Non Grid) 80 to 90
* Be sure to follow
Abdomen 80 to 85
hospital protocol.
Abdomen (Iodine) 76 to 80 *
Extremities (Non Grid) 65 to 75 **
Extremities (Grid) 85 to 95 ** If possible, avoid
AP Spines 85 to 95 using less than 1
Lateral Spines 85 to 100 mAs during any
Ribs 80 to 90 exposure.
Skull 80 to 90
Barium Studies 110 to 120
Modified from Barry Burns
ALARA: Dose Creep
Konica S# Range: 100 to 400
kVp mAs S# Tube Output
80 75 107 321 mR
80 38 211 164 mR
80 20 406 87 mR
*92 10 409 60 mR
Total Reduction in Tube Output: 535%
Total Reduction in Patient Dose???
*Using the 15% Rule will decrease tube output by
approximately one third.
Konica S# Range 100 to 400
kVp 80 80 80 92
mAs 75 38 20 10
S# 107 211 406 409
mR 321 164 87 60
Note how image quality remained fairly constant.
Spatial Resolution:
Controlled by the following:
Size of the Matrix
Pixel Size
Field of View (FOV)
Detector Element Size for DR
Sampling Frequency for CR
Spatial Resolution:
Matrix
Made of columns and rows of cells (pixels).
Spatial Resolution:
Pixel (Picture Element)
They are generally square.
Voxel (Volume Element)
Pixel Voxel
Spatial Resolution:
Pixel Pitch
Refers to the distance between each pixel.
Pixel Size vs. Matrix Size:
Large Medium Small
24 x 30
35 x 43
Note how the pixel pitch has changed.
Which FOV has the highest spatial resolution?
Spatial Resolution vs. FOV:
Some manufactures will decrease the pixel
pitch as the FOV is reduced.
The same number of columns and rows are
packed into a smaller FOV.
The original matrix is maintained within the
smaller FOV and results in an increase in
spatial resolution.
Spatial Resolution vs. FOV:
24 x 30
35 x 43
Since the pixel pitch remains the same, the spatial
resolution is unchanged.
Spatial Resolution vs. FOV:
Other manufactures maintain the same pixel
pitch as the FOV is reduced.
The net result is a decrease in the size of the
matrix while the pixel size remains the same.
Bottom line: there is no change is spatial
resolution.
CR Spatial Resolution:
A thinner diameter laser
will create a higher
sampling frequency.
A higher sampling
frequency will decrease
pixel pitch and result in
higher spatial resolution.
CR Spatial Resolution:
CR Spatial Resolution:
F/S Latitude
Saturation
Density
Brightness
Quantum Noise
Exposure in mR
Quantum Noise:
1024
0
0 1024 2048
Inversion:
2048
1024
0
0 1024 2048
Inversion:
Inversion:
Equalization:
This is also referred to as Dynamic Range
Compression (DRC).
An algorithm is employed to reduce file size by
subtracting the extreme light and dark areas of
the image.
The remaining dynamic range will be modified
to improve the overall contrast of the image.
Equalization:
Dynamic Range
Compression (cropping)
Shape Size
Abdomen
* **
* IR **
10 mAs at 50 kVp 20 mAs at 50 kVp
15% Rule:
To maintain the RE, increase the kVp by 15% and
reduce the mAs by 50%.
Using the 15% Rule in this manner will reduce
tube output by approximately one third.
How will reducing the kVp by 15% and doubling
the mAs affect the RE?
How about the patient dose?
15% Rule:
The 15% Rule vs. RE and tube output:
kVp mAs Tube Output RE
115 5 40 mR 30 mR
100 10 60 mR 30 mR
85 20 80 mR 30 mR
What might be another clinical application for this
rule?
15% Rule:
How would you change the following technique if
you wanted to use kVp to help reduce motion but
maintain the original RE?
80 kVp 100 mA 200 ms (0.2 s)
80 kVp x 1.15 = 92 kVp
200 ms x 0.5 = 100 ms
New Technique: 92 kVp 100 mA 100 ms (0.1 s)
The Anode Heel Effect:
X-rays are emitted isotropically from the anode.
The intensity of the beam decreases as you travel
closer to the anode side of the tube.
The Anode Heel Effect:
Cathode
Stator
(-)
Target
Anode (+)
Filament
Focusing Cup
The Anode Heel Effect:
Stator Anode
(+) Focusing Cup
69.5
69 90
X-ray Photon Energy (keV)
The X-ray Emission Spectrum:
With 2.5 mm of
Filtration At 90 kVp
Quantity or
Intensity of
X-ray
Photons
30 69.5
69 90
X-ray Photon Energy (keV)
Added Filtration vs. the X-ray Emission Spectrum:
What happened
What happened to the beam intensity?
to the
average energy of the
How about
primary beam? the RE?
Quantity or
Intensity of 2.5 mm
X-ray
Photons
3.5 mm
30 69.5
69 90
X-ray Photon Energy (keV)
Collimation vs. RE:
Patient
IR
Image
Tight Collimation
Collimation vs. RE:
Patient
IR
Radiographic Image
* No Collimation *
Collimation vs. RE:
Patient
IR
Radiographic Image
* No Collimation *
Collimation vs. RE: Magnified
IR
Radiographic Image
* *
No Collimation
Collimation vs. RE: Magnified
IR
Radiographic Image
*
Bottom Line:
*
Lack of collimation adds
unwanted scatter (noise) to the IR.
Collimation vs. RE: Konica CR
24 x 30 cm Collimation No Collimation
S = 142 S = 103
new SID
The Direct Square Law:
This formula is employed to maintain beam
intensity at a new distance
I1 D12
I2 D22
old SID
Example #1:
If your original beam intensity was 5.0 mR at a 30” SID,
what would be your new beam intensity if the SID was
changed to 60”?
•
new mR = old mR old SID 2
new SID
60”
new mR = 1.25 mR
Example #2:
Your original SID was 40” and 75 mAs produced an
acceptable RE. What new mAs would you use at 60” if
you wanted to maintain the original RE?
•
new mAs = old mAs new SID 2
old SID
40”
Patient
* *
*
Grid
*
IR
Grid Construction:
Grid Frequency
Grid Ratio (GR)
The height of the lead strips is divided by the
distance between them.
GR = h/D
With all other grid construction factors
constant, the higher the GR, the greater the
scatter clean-up.
Higher GRs also require more accuracy in their
use and result in a higher patient dose.
Grid Ratio Calculation:
H
H == 24
4.0mm
mm
GRGR= =h/D
H/D
GRGR= =4.0/0.5
24/4
GRGR= =8:1
6:1
DD==0.5
4 mm
mm 6060Lines/Inch
Lines/Inch
H
H ==24
4.0mm
mm
GR
GR == 4.0/0.25
24/1
GR
GR == 16:1
12:1
120
120 Lines/Inch
Lines/Inch
DD = 2 mm
= 0.25 mm
Grid Construction:
Grid Radius or Focal Range
Parallel vs. Focused Grids
Parallel Linear Grid:
72”
SID
IR
Focused Linear Grid:
72”
SID
IR
Grid Cap:
Front Back
Grid Cap: Magnified
Front
Grid Cap: Magnified
Ratio: 8:1
Focus: 40 to 72
LPI: 132
The LPI range for digital IRs is between103 and 200
with 140 being the average.
Grid Cut-Off:
Patient Factors to Consider:
Additive Diseases
Require a 50% increase in mAs or a 7.5%
increase in kVp to maintain RE
Additive Disease: Ascites
Normal Ascites
Patient Factors to Consider:
Destructive Diseases
Require a 30% decrease in mAs or a 5%
decrease in kVp to maintain RE
Destructive Disease: Osteoporosis
Normal Osteoporosis
Patient Factors to Consider: Casts
Fiberglass: increase mAs by 30% or increase
kVp by 5%
Patient Factors to Consider: Casts
Contrast Agents
Patient Factors to Consider:
Body Habitus
Hypersthenic
Sthenic
Hyposthenic
Asthenic
The Contrast Unit
Subject Contrast:
The magnitude of the signal difference in the
remnant beam as a result of the different
absorption characteristics of the tissues and
structures making up that part.
The difference in the thickness and atomic
numbers of the structures that comprise the body
part of interest.
kVp is the primary controlling factor for subject
contrast.
This is the basis for optimum kVp.
Image (Radiographic) Contrast:
The visible difference between any two selected
brightness levels within a displayed image.
It is determined by algorithms that are applied to
raw data collected during image capture.
Image (Radiographic) Contrast:
Grayscale
The number of brightness levels or shades of
gray visible on an image.
Linked to the bit depth of the system which a
reference to the total number of shades of gray
available.
How would noise affect grayscale?
Image (Radiographic) Contrast:
Short Scale or High Contrast
Major differences between shades of gray.
Long Scale or Low Contrast
Slight differences between shades of gray.
kVp dictates the starting point for image contrast.
All other factors that impact image contrast
will cause it to increase or decrease from here.
Image (Radiographic) Contrast:
Note the
relationship
between kVp
and subject
contrast.
Patient
IR
Image
Tight Collimation
Collimation vs. Contrast:
Patient
IR
Radiographic Image
* No Collimation *
Collimation vs. Contrast:
Patient
IR
Radiographic Image
* No Collimation *
Collimation vs. Contrast:
IR
Radiographic Image
* *
No Collimation
Collimation vs. Contrast:
IR
Radiographic Image
*
Bottom Line:
*
Lack of collimation adds
No Collimation
unwanted scatter (noise) to the IR.
Collimation vs. Contrast: Konica CR
24 x 30 cm Collimation No Collimation
S = 142 S = 103
30 69.5
69 90
X-ray Photon Energy (keV)
Filtration vs. Contrast: Konica CR
2.5 mm 3.5 mm 4.5 mm 5.5 mm
Filtration vs. Contrast: Konica CR
2.5 mm 5.5 mm
Patient
What happens *
to contrast?
Patient 4” OID
*
1” OID
10 mAs at 75 kVp
IR 20 mAs at 75 kVp
OID vs. Contrast: Konica CR
4” OID 10” OID
*
8:1 GR
*
IR
Grids vs. Contrast: Konica CR
6:1 Grid Ratio 15:1 Grid Radio
S = 259 S = 257
Voluntary Involuntary
Geometric Factors:
1. Focal Spot (FS)
2. Object-to-Image-receptor Distance (OID)
3. Source-to-Image-receptor Distance (SID)
4. Source-to-Object Distance (SOD)
Geometric Factors:
Focal Spot
Primary controlling factor for SR.
Only variable that exclusively affects SR.
Generally, techniques > 200 mA will require a
large focal spot.
Penumbra vs. Umbra
--
Focal Spot Size vs. SR:
0.6 mm FS 2.0 mm FS
Patient
IR
--
FS Size vs. SR: Konica CR
Small FS Large FS
Geometric Factors:
OID
Of the geometric factors, this is the most
critical factor to consider.
What is the overall MOST critical factor to
consider regarding SR?
--
OID vs. SR:
2.0 mm FS 2.0 mm FS
Patient 4” OID
1” OID
IR
OID vs. SR: Konica CR
1” OID 4” OID 8” OID
--
40” SID
Patient
IR
SID vs. SR: Konica CR
40” SID 60” SID 72” SID
x 40
=
20 36
36x = 800