RG1 User Manual v3
RG1 User Manual v3
Raingauge
type RG1
RG1-UM-3
DELTA-T DEVICES
ABOUT THIS MANUAL
This manual describes how to install and use the raingauge type
RGL
COPYRIGHT
Page ii
Raingauge
CONTENTS
INTRODUCTION
SITING THE RAINGAUGE
WIRING CONNECTIONS
CONFIGURING THE LOGGER
USE OF CONVERSION & CORRECTION
FACTORS
SPECIFICATION
CALIBRATION
REFERENCES
GLOSSARY
GUARANTEE, REPAIRS & SPARES
Page 3
INTRODUCTION
Page 4
SITING THE RAINGAUGE
Page 5
WIRING CONNECTIONS
RAINGAUGE . LOGGER
- - a - -- --
I
----------- - L
1 I
1 l
I I e i t h e r wire I
I I I
Page 6
USE OF CONVERSION AND CORRECTION FACTORS
For example,
Given 1000 tips during a recording intcwal.
The calibration for a particular raingauge is 0.198 mm per tip,
ie 5.051 tips per mm.
The logger calculates 100015 = 200 mm of rain
The true value is 100015.051
or 1000 X 0.198 = 198 mm of rain
The correction factor to be applied is
Page 7
SPECIFICATIONS
Page 8
CALIBRATION PROCEDURES
Static Calibration
The raingauge is calibrated during manufacture as described
below. This may be repeated by the user if desired.
Install the raingauge, as illustrated in Figure 5.1, over a sink ensur-
ing that it is level. Using a burette or pipette, adjust the two cali-
bration screws under the tipping bucket until it tips with the correct
amount of water ( 10.13 cc for a 0.2 mm sensitivity). It will not
be possible to set the screws with very high precision in this way
but it should be done with as much care as possible. ( This stage
may be omitted if rechecking the calibration ). Only by means of
a dynamic test can the exact calibration of a tipping bucket ( of any
type ) be established. This test is described below. Many manu-
facturers and users of tipping bucket raingauges aim to adjust the
bucket settings until EXACTLY the correct calibration is obtained.
To achieve this, it would be necessary to repeat the dynamic test
many times over, and this is very time consuming. In any case, it
is almost impossible to get the adjustment absolutely correct.
Therefore, it isconsidered better practice by the Institute of Hy-
drology to adjust the settings as closely as is reasonably practical
and then to derive a Calibration Factor for each gauge.
The Calibration Factor can be found using the dynamic calibration
procedure described below. .
Dynamic Calibration
When the static burette test described above has been done, the
calibration of the gauge should be checked as follows (most users
will Start from here). This procedure was carried out prior to
dispatch, and the resulting Calibration Factor written inside the
raingauge and at the front of this chapter.
Set up the calibration test as shown in Figure 5.1. Ensure that the
gauge is level. Connect the output wires to channel 61 or 62 of
the logger, or any convenient electronic or electromechanical
counter. Fill the container with 810.4 cc of water. This is done
most precisely by weighing the water on scales capable of measur-
ing to 0.1 gm. If this is not practical, it should be done using a
good quality graduated measuring cylinder. Allow the water to drip
slowly into the gauge, taking at least 30 minutes to empty the
container. At the end of this period there should have been 80 tips.
It is unlikely, however, that there will be exactly this number.
Note how many tips actually occurred and also estimate as precise-
ly as possible what fraction of a tip is left in the bucket. Only if the
adjustment has been very precise indeed will the bucket tip at the
moment the last drop passes into it. The Calibration Factor can
now be calculated.
Page 9
FIGURE 5.1 :DIAGRAM OF RAINGAUGE DYNAMIC CALIBRATION SET-UP
CONTAINER
_ -. HOLDING 1OOOcc WATER
ENSURE BUCKET
IS LEVELLED
PRECISELY
RAINGAUGE
WITI-1 COUNT
SINK
Provided the Conversion Factor lies between 0.195 and 0.205 mm,
or 78.0 and 82.1 tips, respectively, then it is acceptable for most
purposes. The gauge is calibrated at manufacture such that each
tip will lie within the above limits.
There is no reason, however, why users should not, if they so wish,
aim for closer agreement with the nominal tip of 0.2 mm by re-
peated adjustments of the calibration screws followed by further
dynamic tests. Only time limitations will set a limit to this.
Page 11
REFERENCES
GLOSSARY OF TERMS
Page 12
GUARANTEE, REPAIRS AND SPARES
Our Conditions of Sale ref COND/91/11 set out Delta-T's legal obligations on these matters.
For your information the following paragraphs summarise Delta-T's position but reference
should always be made to our Conditions of Sale which prevail over the following
explanation.
Instruments manufactured by Delta-T are guaranteed for one year against defects in
manufacture or materials used. The guarantee does not cover damage through misuse or
inexpert servicing, or other circumstances beyond our control.
For the U.K. this means that no charges are made for labour, materials or return carriage for
guarantee repairs.
For other countries, the guarantee covers free exchange of faulty parts during the guarantee
period.
Alternatively, if the equipment is returned to us for guarantee repair, we make no charge for
labour or materials but we do charge for carriage and U.K. customs clearance.
We strongly prefer to have such repairs discussed with us first, and if we agree that the
equipment does need to be geturned, we may at our discretion waive these charges.
We recognise that some users of our instruments may not have easy access to technically
specialised backup.
Spare parts for our own instruments can be supplied from our works. These can normally be
despatched within 1 working day of receiving an order.
Spare parts and accessories for sensors not manufactured by Delta T, but supplied as part of
the weather station, may be obtained from the original manufacturer. We will endeavour to
obtain parts if requested, but a certain amount of additional delay is inevitable.
Should it prove necessary, instruments may be returned to our works for servicing. We
normally expect to complete repairs of our own instruments within 2 days of receiving the
equipment. Other manufacturers' sensors supplied by us and returned for servicing will take
longer. They will have to be returned to the original manufacturer for servicing, and may be
subject to additional delays of two to four weeks.
Users in countries which have a Delta-T Agent or Technical Representative should contact
them in the first instance.