ISVR Offical Report On Calmer
ISVR Offical Report On Calmer
Ref: 10903-R01
Dr Mike Lower
Principal Consultant
June 2020
1. Introduction .................................................................................................................... 1
5. Conclusions .................................................................................................................. 28
6. Appendix ...................................................................................................................... 30
6.1 Equipment list 30
6.2 Narrow band frequency response data 31
6.3 TD measurement results 34
6.4 A-Weighting values 36
7. References .................................................................................................................... 37
This report shall not be reproduced except in full, without the prior written consent of ISVR Consulting
ISVR Consulting was engaged by Flare Audio to perform acoustic tests on their new
Calmer technology prototypes. These measurements aimed to characterise the acoustic
response of this technology and to objectively assess its typical effect on the hearing
profile of the user of this device.
The devices can also be worn under circumaural and supra-aural headphones. In this
case, the technology aims to improve the headphone listening experience by controlling
the acoustic characteristics of the human ear coupled to the headphone.
During the time of this measurement project, the acoustic facilities of ISVR Consulting
were out of service, due to the COVID-19 lockdown period in the United Kingdom.
The test procedures were designed specifically to ensure that measurements do not rely
on free-field (anechoic) acoustic conditions.
The primary aim of the project therefore was to evaluate the acoustic effect of Calmer
devices under headphones using an Acoustic Head and Torso Simulator (HATS).
Frequency response functions (FRF) of 2 headphone specimens on the HATS were
measured with and without the Calmer devices fitted. FRF results were analysed
comparatively to demonstrate and quantify the acoustic effect of wearing the devices.
Any alteration in the transfer characteristics of the ear that Calmer achieves could also
influence the level of distortion perceived by the user of the technology. Distortion
measurements with and without Calmer were assessed using headphones. Whilst the
technology is not expected to control the amount of distortion generated by the
headphone itself, some distortion components falling in the modal frequency range
could be reduced at the ear drum, if the resonant behaviour of the system is attenuated
by the technology. In order to test the total perceived distortion with and without
Calmer, measurements were taken with a multi-tone stimulus signal.
The project aimed to quantify the effectiveness of the Calmer technology using
frequency response and distortion measurements. This report details the measurement
configuration and procedure, presents the results and provides an objective assessment
of the acoustic features of this technology. The report specifically does not discuss the
subjective experience of wearing the device and does not attempt to link any of these
with the test results. Detailed analysis and identification of the exact physical-acoustic
mechanisms of the device also fell outside the scope of this investigation.
All tests were performed by ISVR Consulting on the 29th and 30th April. This section
of the report details the equipment, test conditions and data processing in the project.
All tests were carried out in a furnished, ordinary room with dimensions of
3.5 m x 3.0 m x 2.2 m. The side and top boundaries are plastered walls/ceiling, the floor
is covered by a thin layer of carpet. Besides a regular wooden door, the room is built
with double glazing (1.6 m x 1.1 m) on one of the walls. During all tests, windows and
the door were closed in order to minimise background noise levels. Furthermore, the
amplitudes of test signals were selected to ensure high signal-to-noise ratios. During
testing, external noise levels were observed and tests were halted if background levels
were higher than normal.
Tests were carried out using ISVR Consulting’s ‘Kemar’1 acoustic head and torso
simulator (HATS). This device is a manikin with a realistic head and pinnae and
incorporates ear simulators with ear canals and ‘eardrum’ microphones. It is
representative of a median human adult and is designed to permit acoustic
measurements of wearable acoustic devices. The shape of the manikin and the nature
of the ear simulators ensure that the obtained recordings include the acoustic effect of
the human body exposed to external sound sources or wearing headphones. For
headphones the manikin also creates a realistic acoustic load on the headphone drivers.
This platform is therefore ideal to investigate the changes that Calmer brings to this
system.
The ear simulators and microphones within the manikin are tuned to imitate the transfer
impedance of a typical human ear. ISVR Consulting’s ear simulators conform to
BS EN 60318-4:2010 2. The calibration of the HATS’s two microphones was checked
The ear simulators accurately imitate the human ear’s (standardised) acoustic transfer
impedance up to the frequency of 10,000 Hz in compliance with BS EN 60318-4:2010.
Results above this frequency fall outside the range of the standard and may not be an
accurate human ear simulation, however comparative conclusions are made in this
report up to 16,000 Hz by relating recordings with and without Calmer in the exact
same conditions.
The primary aim of this test was to measure and characterise how wearing Calmer
influences the sound pressure reaching the eardrum when used under headphones.
These measurements were performed by determining the impulse response between the
headphone excitation signal and the corresponding microphone recordings of the ear
simulator, with and without the Calmer prototypes ‘worn’ by the HATS under
headphones.
To account for any imperfections imposed by the USB measurement interface used with
this software, Dirac’s sound device calibration routine was performed prior to the
measurements. This routine also ensured the input and output level calibrations are
obtained by the Dirac and the output levels therefore were automatically adjusted by
the software.
The exact calculation technique of this proprietary software is not detailed in its
reference manual, but it is understood that the Dirac software estimates the impulse
response by deconvolution of the ear simulator microphone signals and the
corresponding loudspeaker driving signal. This technique for estimating the impulse
response was established by A.Farina 3 at the 2000 AES convention in Paris.
The obtained impulse responses were processed using a 131072-point FFT analysis to
obtain the corresponding frequency response functions (FRF).
The total signal distortion present at the ear drum microphones produced by the
headphones was measured with and without the Calmer devices using a Prism Audio
dScope M1 measurement system. A synchronous multi-tone measurement technique
was chosen specifically to perform distortion measurements in test conditions that can
emulate the complexity of real music and other programme material, whilst allowing
simultaneous distortion measurements in the whole audible range.
- The test signal consists of 31 tones with frequencies close to, but not exactly
corresponding to 1/3rd octave centre frequencies as defined in BS EN 61260-
1:20145.
- The tones are identical in magnitude and their relative phases are varied randomly
to ensure that signal clipping is avoided, and that the crest factor of the signal is
controlled
- The frequencies of the tones are defined to avoid spectral overlaps of fundamentals
and harmonics. Furthermore, all frequency components of the input signal also
correspond to the exact centre frequency of an even FFT bin, allowing the use of
rectangular windowing without spectral leakage.
- The system in this configuration can detect the signal in the predefined FFT bins,
distortion in other even-numbered FFT bins and noise in all odd-numbered FFT
bins simultaneously
Exploiting these features of the multi-tone signals, this test can maximise the production
and detection of harmonic, and intermodulation distortion as well as other non-
linearities, which are more difficult to evaluate with more traditional swept-sine or
twin-tone tests. The complexity of the multi-tone test signal enables realistic test
conditions by better approximating the dynamic characteristics and tonal content of
music and other programme material on the whole audible range. Similar to real-life
listening, the headphones playing the multi-tones are expected to produce significantly
higher amounts of distortion products compared to those detectible in the FRF
measurements. Results are presented in this report using the following quantity:
† calculated using spectral power magnitudes in even-numbered FFT bins that do not contain any components of the
input signal
∗ Reference used in TD calculation is the measured signal magnitude of the tone component closest to 1,000 Hz
N.B. TD results are not directly comparable with total harmonic distortion measurements
Since these tests aimed to establish the distortion characteristics of the signals measured
at the simulated eardrum rather than distortion generated by the headphones, sound
levels were not adjusted to give the equivalent undisturbed field sound levels. This test
relies on the assumption that the ear simulators used in this project provide a good
approximation to the behaviour of the human ear, both for the main (high magnitude)
signal components and the (low magnitude) distortion products. It is also assumed that
the instrumentation itself does not introduce a significant amount of its own distortion
that are typically not present in the human ear.
The reference power used in the distortion calculations may vary comparing cases with
and without Calmer if the corresponding frequency responses are changed by the
presence of the device at 1,000 Hz. Results were recorded with reference powers
calculated by the audio analyser, however individual correction factors were used in
calculations to ensure that the distortion readings are directly comparable.
Total Distortion results with and without Calmer were analysed comparatively. For
each test case, these metrics were logged for 30 seconds in the measurement system
and results in this report are calculated using linear averaging. The FFT results exported
from the analyser were also utilised for further calculations.
The Prism Audio multi-tone measurement system and calculation software complies
with requirements regarding multi-tone based distortion measurements set in BS EN
IEC 60268-21:2018 4.
In order to characterise the acoustic effect of Calmer when worn without any other
acoustic devices around the ear, the impulse response measurements detailed in section
2.3 were performed with a studio-grade loudspeaker as an excitation source instead of
headphones. This test is designed to measure and characterise how wearing Calmer
influences the sound pressure reaching the eardrums in a certain sound field due to
external sources.
In order to ensure that such results are not affected by the acoustic response of a test
venue, these measurements would need to be performed under free-field acoustic
conditions. During the time of this project, it was not possible to utilise ISVR
Consulting’s large anechoic chamber. Measurements were carried out in the same room
as used for the headphone measurements and the approximate (simulated) free-field
conditions were achieved using digital signal processing.
The acoustic centre of the excitation source was placed 1.2 m away from the centre of
the HATS (midway between the two ears). The direction of the HATS was aligned so
both ears were the same distance from the centre of the loudspeaker.
The following steps were taken in order to minimise the influence of the room on the
measurement results:
1. Each measured impulse response was analysed individually to detect the ‘time-of-
arrival’ of the first distinct reflection from the boundaries of the room reaching the
simulated eardrums. Knowing the time-of-arrival of the direct sound from the
loudspeaker and that of the first reflection, a weighting window was constructed
using a sinusoidal function. The length of the window function was adjusted to be
shorter than twice the difference of the detected timings. The peak/centre of this
Besides on-axis sound incidence described above, tests were also performed at 45o and
90o azimuth angles.
The aim of this test was to measure and characterise how wearing Calmer influences
the sound pressure reaching the eardrum. These measurements were performed by
determining the impulse response between the excitation source (loudspeaker or
headphone) and the corresponding microphone recordings of the ear simulator. Using
headphones, Total Distortion (TD) was measured with a multi-tone stimulus by
analysing the corresponding ear simulator signals on the frequency domain. In all
measured data pairs (with and without Calmer), every measurement conditions were
consistent, except the presence of Calmer in the artificial ears.
The tests were carried out using the following Headphone test specimen:
1. BOSE Soundlink AroundEar II Wireless
2. Nubwo N16 circumaural gaming headphone
Headphone FRF results were recorded using 20 mV rms signal level. Input levels in
TD tests were varied.
Loudspeaker measurements were carried out using a Genelec 8030C studio monitor.
Playback level was adjusted to be 75 dB at 1,000 Hz.
Any variability in the fitting of both the Calmer prototypes and/or the headphones on
the HATS can significantly influence the measurement results in all tests, particularly
at low and high frequencies. In order to minimise measurement errors related to the
fitting of the devices and the headphones, the following actions were taken:
1. The devices were examined and worn by the experimenter to determine the
properties of the typical fit the user of this device would achieve
2. Fit and seal of Calmer was visually examined and confirmed in the HATS’s ears
without the microphones in place
Once various fit properties were observed and reasonable consistency was achieved, 3
sets of FRF and 3 sets of TD measurements using headphones were performed with a
refit between every consecutive test. Using the left and right units individually,
presented results are based on the average of 6 (refitted) FRF measurements and 6
(refitted) TD measurements.
In the case of measurements with external sources, on-axis, 45o and 90o azimuth
(incidence of sound), measurements were performed consecutively. For each
positioning of the HATS, the reference HRTF was measured once (after preliminary
measurements to verify the setup), followed by 5 measurements with Calmer fitted in
the ears. Amongst these 5 measurements, the devices were refitted each time.
Figure 2: Typical fit of BOSE (left) and N16 (right) headphones on HATS
Figure 3 presents the magnitude of the measured frequency response function of the
BOSE headphones with and without Calmer technology applied.
Figure 4 presents the magnitude of the measured frequency response function of the
N16 headphones with and without Calmer technology applied.
Figure 5 shows the effect of wearing Calmer normalised to the measurement result
without the devices. These results quantify Calmer’s influence on the sound pressure
levels reaching the eardrums for these two headphones.
At frequencies below 1,000 Hz, Calmer does not significantly affect the measured
response. This may be explained by the fact that at these frequencies, the wavelengths
of the sound components are significantly longer than the dimensions of these devices.
Using the BOSE headphones with Calmer, no change in sound pressure levels were
observed up to 2,000 Hz. In case of the N16 headphones, Calmer had a minor boosting
effect on frequencies between 500 Hz and 2,000 Hz. The peak of 2.5 dB of this boosted
region was observed at 1,600 Hz.
Considering the measured results of the two headphones without Calmer, both
headphones show elevated responses in the 2,000 Hz to 8,000 Hz range. In this
frequency region the various dimensions of the headphones and outer ear become
ISVR Consulting Report 10903-R01 / June 2020 Page 16
comparable with the wavelengths of the sound waves. Therefore, various wave
reflections from the headphones’ body, parts of pinnae and the eardrum lead to
modal/resonating acoustic characteristics in this frequency region.
Resonant mechanisms of the open/unaided ear are well established. Shaw and
Teranishi6 (1968) investigated the various components of head-related acoustic transfer
functions. Amongst various mechanisms related to the presence of head and shoulders,
for a free-field sound incidence of 45o azimuth, they identified resonances of
approximately 11 dB at 2,700 Hz and 8 dB at 5,800 Hz. Their research linked these
resonances to the geometry of the ear canal and concha respectively. When using
headphones and depending on the listeners exact ear geometry, the magnitude and
frequency of these resonances may change, however they will be present, influencing
the sound pressure reaching the eardrum. Results in this frequency region may also be
influenced by the frequency response characteristics of the headphone drivers
themselves.
According to test results on Figure 3 and Figure 4, Calmer attenuated these naturally
occurring acoustic phenomena and it was observed that this technology significantly
modified most of these effects. The responses of both headphones showed their highest
peak around 3,000 Hz (most likely due to simulated ear canal resonances). Calmer
attenuated these features, reducing levels by approximately 6.6 dB and 7.8 dB around
3,000 Hz respectively with the BOSE and N16 headphones.
The responses of both headphones showed further resonances in the 4,000 Hz to 8,000
Hz range. These are most likely linked to the geometry of the simulated concha and
other parts of the outer ear. Calmer achieved 6.4 dB and 6.7 dB of average attenuation
in this range with the BOSE and N16 headphones respectively.
Results in Figure 3 and 4 show that responses recorded with Calmer were lower than
those without this technology above 8,000 Hz. It is important to note that the response
of the ear simulators in the HATS are not defined above 10,000 Hz in BS EN 60318-
The audio analyser used in the tests calculated total distortion during the tests as
detailed in section 2.4. Results for both headphones were recorded at 3 different
Table 1 shows the average reduction in Total Distortion that Calmer achieved.
Individual distortion readings (with appropriate corrections applied) are shown in
Table A2 in the Appendix.
Table 1: A-weighted average reduction in Total Distortion (TD) by Calmer
Total reduction in
Total reduction in TD Total reduction in distortion by Calmer
Input Overall A-weighted
by Calmer A- signal power by A-weighted ref'd to
(mV SPL at eardrums
weighted (ΔTD) Calmer A-weighted reduction in signal
rms) without Calmer [dBA]
[dBA] (ΔSIG) [dBA] power (ΔTD/ΔSIG)
[dBA]
The A-weighted total distortion results show an average reduction of 5.4 dB and 6.0 dB
with the BOSE and N16 headphones respectively. Reductions in corresponding signal
powers of 3.3 dB and 4.0 dB were recorded. In both cases, the reduction in distortion
power was greater than the reduction in signal power.
- Apart from distortion detected below 300 Hz, the frequency of the highest distortion
peaks without Calmer corresponded to the frequency of the resonant peaks observed
in Figure 3 and 4.
- Highest reductions in distortion components were found in the 2,000 Hz to 8,000
Hz range
- Analysis of distortion components in the 2,000 Hz to 8,000 Hz range, that fell close
to the corresponding signal tones showed comparable reduction to results observed
at the nearby tones
- Distortion components were present in a large number of even FFT bins across the
whole measured frequency range. A relatively small number of even FFT bins were
observed with levels comparable to odd frequency bins (containing noise only)
Using these observations, it is speculated that the primary acoustic mechanism behind
these results, is Calmer’s capability to attenuate the naturally occurring resonances in
the ear (as observed in the FRF results). In other words, Calmer reduces amplification
of distortion components at these resonant frequencies.
This test assumes that most real playback signals would behave similarly to the test
signal used, in terms of excitation and corresponding broadband distortion generation
mechanisms. However, the actual reduction in distortion relative to the signal
components experienced by the user will depend on the headphones used, the fit of the
devices and properties of the music or programme material reproduced.
It is important to note that this analysis is not exhaustive and is based purely on the data
collected in this project.
Figure 6 shows the magnitude of measured FRFs of Calmer technology applied to the
HATS with an external source place at 0o azimuth. Results without these devices are
also shown for comparison.
Figure 7 and Figure 8 show the magnitude of measured FRFs of Calmer technology
applied to the HATS with and external source place at 45o and 90o azimuth respectively.
The corresponding results without these devices are also shown for comparison.
These measurements are analogous to sound pressure levels reaching the eardrum with
and without Calmer from a point excitation source with a flat frequency response placed
in front of the listener (at 0°, 45° and 90° positions), in free-field conditions, assuming
that all frequencies are excited equally.
At frequencies below 1,000 Hz, Calmer does not significantly affect the measured
response. Frequencies between 500 Hz and 1,600 Hz are slightly boosted with the
devices. The highest amplification of 2.5 dB was observed at 1,250 Hz when the
loudspeaker was placed 90o off-axis relative to the HATS.
Resonant mechanisms of the human ear were briefly discussed in section 4.1 of this
report. HRTFs measured without the devices in Figures 6, 7 and 8 show the frequency
response of a typical, unaided human ear in the sound field established in this
experiment. It can be observed that the 3,000 Hz region is significantly enhanced by
ISVR Consulting Report 10903-R01 / June 2020 Page 25
the presence of the human body and ear as well as that lower ranges show an increasing
trend towards 1,000 Hz. This observation is understood to be the combined effect of
numerous acoustic factors, of which the most significant are the natural resonances
occurring in the simulated ear canal, diffraction around the head, and reflections from
the pinnae and shoulders of the HATS, though shoulder reflections are minimised by
placing a tee-shirt on the HATS, as is recommended.
These results revealed that fitting Calmer to the HATS influenced these naturally
occurring acoustic phenomena. In the 2,000 Hz to 8,000 Hz range, the devices reduced
sound pressure levels measured at the eardrums. Comparative analysis of results with
and without Calmer were also performed in the 10,000 Hz, 12,500 Hz and 16,000 Hz
third octave bands. These high frequency bands also showed a reduction in sound
pressure levels compared to the corresponding HRTFs. Table 2 shows average
reductions observed at high frequencies.
Results shown in Table 2 indicate comparable average attenuation figures at the 3 tested
source locations up to 8,000 Hz. It can be concluded that up to this frequency, the
acoustic effect of Calmer is reasonably uniform for ipsilateral sources in front of the
listener. At higher frequencies however, Calmer behaves more directionally with
moderately higher attenuation at wider source angles.
1. The devices do not influence the magnitude responses measured below 1,000 Hz
(below the resonant region of the outer ear) significantly.
2. The devices attenuate sound pressure levels reaching the ear drum above 1,000 Hz.
Peak SPLs due to natural resonant mechanisms observed in HRTFs in this
frequency region are significantly reduced.
3. The devices showed a comparable effect with near-field sources (headphones) and
far-field sources (external sounds and noises). Flare’s design target is met in both
cases. However, it was observed that the attenuation achieved with loudspeaker
sources were moderately higher compared to those measured with headphones.
4. With external sources, the devices did not significantly alter the directivity of
hearing (in the measured span of source locations)
Acoustic tests were performed on Flare Audio’s new Calmer technology prototypes.
Measurements were carried out using an acoustic head and torso simulator using two
different headphones and an external sound source. The acoustic effect of Calmer fitted
in the open ear and placed under the headphones was evaluated.
Frequency response measurements of the headphones and external source with and
without the technology prototypes revealed:
The actual auditory experience and the effectiveness of wearing Calmer may vary
depending on the external sound field, type of headphone, fit and properties of the
programme material played.
N16
Calmer Input, mV TD Left TD Right
off 10 -26.1 -26.8
off 10 -25.5 -26.6
off 10 -26.5 -26.0
on 10 -31.8 -32.8
on 10 -31.7 -32.2
on 10 -30.7 -33.7
1/1 Octave band centre frequencies 1/3 Octave band centre frequencies A-weighting values
Hz Hz dB
50 -30.2
63 63 -26.2
80 -22.5
100 -19.1
125 125 -16.1
160 -13.4
200 -10.9
250 250 -8.6
315 -6.6
400 -4.8
500 500 -3.2
630 -1.9
800 -0.8
1000 1000 0
1250 0.6
1600 1.0
2000 2000 1.2
2500 1.3
3150 1.2
4000 4000 1.0
5000 0.5
6300 -0.1
8000 8000 -1.1
10000 -2.5