Handheld Vector Network Analyzer Incomplete Manual: Hangzhou Minghong Electronic Technology Co., LTD
Handheld Vector Network Analyzer Incomplete Manual: Hangzhou Minghong Electronic Technology Co., LTD
Table of Contents
Overview ...............................................................................................................................................................................3
Characteristic ......................................................................................................................................................................4
FAQs ..................................................................................................................................................................................... 14
Overview
NanoVNA-F "Handheld Vector Network Analyzer" is a product based on the Open Source Project of
NanoVNA (https://ttrf.tk/kit/nanovna).
We designed the NanoVNA-F hardware, which is based on the STM32F103 Cortex-M3 processor and has a
larger and brighter 4.3-inch high resolution IPS display, a large-capacity battery, a metal body, a full set of
accessories, and more.
At the same time, we have transplanted the open source software of edy555 to FreeRTOS system. The current
open source address of this project is: https://github.com/flyoob/NanoVNA-F. Every enthusiast can view and
learn to compile the firmware.
Characteristic
Menu Structure
Each NanoVNA-F machine comes with this paper quick start guide (menu structure chart), which is convenient
for users to learn and use the machine as soon as they get it.
The guide was inspired by AE5CZ, and we would like to thank it.
Due to the firmware update, the Quick Start Guide is constantly being improved and optimized, please click
the link below to get the latest PDF version.
Basic operations
Start a measurement:
1. Set the frequency range (STIMULUS→START/STOP or CENTER/SPAN)
2. Calibration (According to actual needs, refer to the next chapter)
3. Select display format and channel (DISPLAY→FORMAT/CHANNEL)
4. Save the parameters for the next measurement. (SAVE→0/1/2/3/4)
You can change the display format and channel selection at any time. In the normal test mode, tap the right
area of the screen or press the multifunctional switch to call up the menu. Tap the screen or turn the
multifunctional switch to select a menu item.
Usage 1:
Set the center frequency point of the current screen through the STIMULUS→CENTER of the menu, and
display the values of the center frequency and sweep span, respectively, to the left and right of the bottom of
the grid.
In the lower-right corner of the pop-up Settings value screen, click to eject the soft keyboard and enter the
frequency value via the soft keyboard.
Usage 3:
Set the start frequency through the STIMULUS→START of the menu and display the start frequency and stop
frequency values on the left and right sides of the grid, respectively. Click on the bottom right corner of the
pop-up settings screen to eject the soft keyboard and enter the frequency value through the soft
keyboard.
Set the stop frequency through the STIMULUS→STOP of the menu, and display the start frequency and stop
frequency values on the left and right sides of the grid, respectively, in the lower right corner of the pop-up
Settings screen, click to eject the soft keyboard and enter the frequency value through the soft
keyboard. Please pay attention to the following points: The span and center frequency vary with the stop
frequency. The change of the span will affect other system parameters. For more details, please refer to
“Span”.
In Zero Span, the start frequency, stop frequency and center frequency are always the same value.
The VNA Master is a field portable unit operating in the rigors of the test environment. In order to ensure
measurement accuracy, RF calibration (SOLT) must be performed prior to making a measurement in the field.
Calibrating with the specified mechanical calibration requires three calibration parts: open, short, and
match(load) and one RG316 RF cable. Calibration data is saved as user calibration data. It can be saved to
status 0/1/2/3/4, which will be automatically loaded at the next boot and can be also be loaded via the RECALL
menu.
The figure below shows the OPEN, SHORT, and LOAD calibration components that are standard with the
NanoVNA-F.
Note: If there is already saved calibration data, first press “Reset” to clear the calibration data and then
calibrate!
If you make a mistake, start again with RESET!
Once the calibration is complete, the three calibrators can be connected to PORT1 again, and the correct
calibration should be based on the S11 SMITH chart, which should be:
When connected to SHORT, the curves should be concentrated on the left most side of the S11 SMITH chart;
When connecting LOAD, the curves should all be concentrated in the middle of the S11 SMITH chart;
The instrument has been fully frequency-calibrated(50kHz~1GHz) before shipment and Store the data
in parameter 0, and the boot will load the status 0 data directly.
If you change the measurement conditions, such as adding a adapter and modifying the frequency range,
you may need to recalibrate and store the parameters in 0-4. All parameters can be called up by the “recall”
function.
NanoVNA-F charging port is a standard USB Type-C interface. You can use the USB interface to connect to
computer communications or connect standard 5V charger charging.(charger with an output current capacity
of at least 1A). After turning off the power, the blue light stays on for about 40 seconds, and then goes out.
At the same time, we also designed a USB Type-A 5V power output port for customers, using internal batteries
to boost to 5V. In an emergency, users can use NanoVNA-F as a mobile power supply.
When NanoVNA-F is charging and discharging, the status of the indicator is as follows:
You can use a common polymer lithium-ion battery as the battery for your machine, which size limit <
L(90mm)*W(54mm)*H(5.6mm).
You can solder the battery directly to the PCB. But the recommended method is to add a terminal, the type
of the power interface is PH2.0, the connector on the battery side needs to match it, and the cable length is
about 4CM.
Packaging details
From January 6, 2020, the following new boxes and complete accessories are provided.
FAQs
The NanoVNA-F text with artistic fonts is a trademark / Logo (under registration) registered by Deepelec in
China. From November 15, 2019, we only produce NanoVNA-F with this trademark / Logo, Please check
carefully before buying. If the silk screen of the machine you received is incomplete or the logo is inconsistent
with the picture below, it is most likely a counterfeit.
(2) Use a toothpick or other sharp object to click the center of the cross that appears in order.
(3) Done.
(2) Then at PC side, the device will act as a disk drive, which should appear in File Explorer. Copy you new
update.bin into the root directory of that disk.
(3) Power the device off and on, and the boot-loader will run automatically, and update your NanoVNA-F.
Re-power for normal operation.
(2) Then at PC side, the device will act as a disk drive, which should appear in File Explorer. Copy your
callsign.txt into the root directory of that disk, Re-power the device.