IDS ISYS User Manual V14 Revision M1
IDS ISYS User Manual V14 Revision M1
User Manual
Software version V 14
Revision M1 of the User Manual was produced on the 21st December 2015, for the software
version:
V 14.04
The information contained in this document is applicable to any subsequent software version identified as «V
14.XX». Changes with the decimals of the version identifier are used to account for minor software
enhancements, concerning neither the features nor the use of the system.
The manual comprises the following sections and appendices:
If the equipment is used in a manner not specified by the manufacturer, the protection provided
Section 2
• Modification in 2-3-1-FILE Menu, page 25: adding of a new feature accessible from this menu.
• Modification in 2-3-4-MAINTENANCE Menu, page 28: adding of a new automatic maintenance accessible
from this menu
• Modification in 2-3-5-SETUP Menu, page 30: adding of a new feature accessible from this menu.
Section 3
• Modification in 3-3-1-Installation of internal ancillary reagents, page 47 and in 3-3-2-Installation of
external ancillary reagents, page 48: adding of automatic priming of ancillary reagents.
• Modification in 3-4-System Performance Checks, pages 42 and following: programming the automatic
qualification profile now in section 7.
• Adding of 3-5-1-Programming calibrations, page 55: new display options in the calibrations/controls
window.
• Modification in 3-7-Programming Samples, pages 69 and following: use of the sample type «Other» in
immunoassays.
• Adding of 3-8-6-Performing reflex tests, page 76.
• Adding of 3-8-7-Performing dilutions, page 77: post dilutions carried out on user request.
• Modification in 3-10-1-Result of a calibration, page 84: the validation criteria «calibrator CV in RLU» is
modified.
• Modification in 3-10-3-Sample results, page 87: change in the reporting of results outside the measuring
range; adding of reflex test launch.
• Modification in 3-10-4-Work list results, pages 87 and 88: adding of reflex test launch.
• Modification in 3-11-Messages associated with results, pages 89 and following: adding of the messages
RFX and AIN; modification of messages OMR- and OMR+.
• Modification in 3-12-Results Storage, pages 93 and 94: storage of incomplete profiles and removal of
stored profiles
• Modification in 3-13-1-Cumulative analysis, pages 95 and following: modification in Levey-Jennings
access.
• Modification in 3-14-1-Emptying the solid waste, page 101: use of disposable container.
Section 4
• Modification in 4-3-Messages associated with results, pages 108 and following: adding of the messages
RFX and AIN; modification of messages OMR- and OMR+.
Section 5
• Adding of 5-4-13-2-Cleaning the probe, page 145.
Section 7
• Adding of Section 7: System Configuration, pages 151 and following.
Appendices
• Modification in A-6- Analyser long stoppage period, pages A-8 and following.
Manufacturer.
High temperature.
Appendices A1
• Biochemistry.
• Substrates.
• Enzymes.
• Electrolytes.
• Specific Proteins.
The analyser is intended for professional use and must only be used by trained personnel
working in compliance with the precautions for use set out in this manual (see Section 1-6-1, page 15) and
good laboratory practice (GLP).
With all modules fitted the analyser comprises the following key components:
3 7
1 5 4
The barcode reader integrated into the reagent compartment identifies reagent cartridges supplied by IDS.
A second barcode reader, located on the front of the analyser identifies ancillary reagents.
Samples, calibrators and controls are identified or by the reader located on the front for analysers with a fixed
sample tray, either by a reader integrated into the sample compartment for analysers with removable sample
trays.
For the fixed-tray configuration, each position is equipped with a detection sensor. Samples are identified by
the barcode reader (on the front) prior to loading the sample, or manually programmed after sample loading.
For the removable-tray configuration, the tray and each of its positions are identified when loading by the
integrated barcode reader.
The reagent cartridges are stored in the refrigerated reagent compartment between 12 and 15°C while the
analyser is operating and between either 8 and 10°C or 12 and 15°C in standby mode (depending on the
analyser configuration).
Depending on the configuration of the analyser, up to 15 Immunoassay parameters can be programmed
simultaneously or up to 36 Biochemistry parameters or a combination of both, up to the maximum number of
positions available.
For Immunoassays, the analyser only uses reagents supplied by IDS and its partners. For Biochemistry, it is
possible to use reagents available from suppliers other than IDS, however these reagents will not be managed
using the barcode system, and the information required for traceability must be manually programmed by the
operator.
When Biochemistry regents available from third-party suppliers are used, IDS takes no responsibility for the
validity of results obtained. It is the responsibility of the operator to ensure any non-IDS reagents are optimised
and validated for use on the analyser.
IDS Specification
Site de POUILLY Rack Sample M
Samples and reagents are aspirated using a probe equipped to detect liquid level using capacitance. The
sample and reagents are pipetted in accordance with the validated parameters for each assay and transferred
into a cuvette where the reaction takes place. Between each sample, the probe is rinsed internally and
externally in order to eliminate any risk of contamination.
Immunoassays:
The cuvettes are transferred to the washer module where magnets hold the solid phase (magnetic particles)
whilst washing takes place.
When the reaction is over, depending on the type of the assay, the cuvette is either transferred to the
luminometer for chemiluminescence assays or kept in the carousel for enzymatic assays.
For chemiluminescence assays, trigger reagents are added to produce luminescence.
For enzymatic assays, substrate is added and the photometric changes are measured continuously by the
spectrophotometer.
When measurement is complete, cuvettes are automatically ejected into a re-usable solid waste tray.
Biochemistry assays:
Photometric changes taking place in the reaction are measured continuously by the spectrophotometer.
The analyser is connected to a computer via an Ethernet link.
The software is able to program the analyser workload and carry out the following functions:
• Management of lots of reagents, calibrators and controls.
• Quality controls (Levey-Jennings and Westgard).
• Management of ancillary reagents.
• Operator traceability.
• Transfer of results to a centralised computer system.
• Printing, storing and traceability of results.
1-3- Characteristics
Analyser
System • Multidiscipline Immunoassay and Biochemistry analyser.
• Continuous loading.
• Configuration:
• Immunoassay (4 washers) + Biochemistry.
Barcode readers • Reader on front face for barcode identification of cuvettes and ancillary
reagents. Used for sample identification in fixed tray configuration.
• Integrated reader for reagents barcode
• For configuration with removable trays, reader integrated into the
sample compartment.
Waste collection • 10 liters container for liquid waste posing a biological risk.
• Solid waste (cuvettes) disposed of in re-usable container.
Power supply • Voltage: 100 - 240 V.
• Frequency: 50 - 60 Hertz.
• Maximum power consumed: 750 VA.
1-4- Installation
The packaging of the has been designed to prevent any damage occurring during
transportation.
The analyser can be stored in its original packaging under the following conditions:
• Storage temperature
10-40°C.
• Relative humidity 80% non condensing.
In the event of prolonged storage, the analyser performance must be checked. The unpacking, installation and
initial qualification of the analyser must only be carried out by a qualified IDS representative.
After installation by Technical Services or an IDS representative, do not handle or move the
! analyser.
If the analyser must be moved, contact Technical Services or IDS representative.
1-4-1- Environment
A satisfactory installation site is essential for the analyser to function correctly.
The user must ensure compliance with the conditions required in terms of environment and electricity supply in
order to maintain the performance of the analyser and to guarantee safe use for the operator.
Environmental conditions required:
• The analyser must not be exposed to direct light.
• A clean and ventilated air environment.
• The analyser must be placed on a flat work surface, capable of supporting its weight (103 kg) without
significant vibration.
• The surrounding temperature must be between 15°C and 30°C.
• Relative humidity must be within 20 and 80% (non-condensing).
• The analyser must not be installed under an air-conditioning unit.
• Clearance of at least 15 cm must be provided at the rear, front, left and right of the analyser to allow
evacuation of heat produced by the apparatus.
• The analyser must be installed in such as way as to allow the user to easily access the ON/OFF button and
the main cable inlet.
• The analyser must not be installed near strong sources of electromagnetic radiation and electrical
interference (e.g. refrigerators).
The IVD medical device complies with the emission and immunity requirements described in the standard
IEC61326-2-6.
Physical characteristics of the analyser:
• Dimensions of the analyser L 105 x H 70 x W 75 cm.
• Overall dimensions L 124 x H 110 x W 71 cm.
• Weight 103 kg.
• Basis weight 130 kg/m2 *
• Computer system L 60 x H 40 x W 50 cm.
* Analyser only. Basis weight of analyser and its special table: 278 kg/m2
If necessary, the installation of a regulated electrical supply may be required by IDS Technical Services.
In order to ensure the analyser’s electrical safety (in accordance with standards), it is essential
! to check that the analyser and its associated peripheral computer equipment (external printer
and computer) are properly earthed.
1-4-3- Connections
Fluidic connections
The fluidic connections are located on the right hand side of the analyser. The tubing is identified by a colour
code.
Description Colour code Field
Electrical connections
The electrical connections are located on the left hand side of the analyser.
• Connect the Ethernet cable (RJ 45, 8 pins) between the analyser and the PC (local network).
• Connect the mains supply cable.
Depending on the type of the assay, Immunoassays using the chemiluminescence method are measured by
the luminometer and by the spectrophotometer for enzymatic detection assays.
BIochemistry assays are measured by the spectrophometer.
Absorbancies are measured with an optical pathway of 0.8 cm and corrected for an optical pathway of 1 cm.
1-5-3- Immunoassays
Assays are carried out using either a one-site or two-site method.
One-site (competitive) method assay
The assay is based on competition between an unknown quantity of analyte in a sample with the labelled
analyte in the kit.
In a sample where no analyte is present, maximum binding of the labelled analyte is possible. With the
increasing analyte concentrations, decreasing binding of labelled analyte is observed.
The signal generated by the labelled analyte in the luminometer is therefore inversely proportional to the
concentration of analyte in the sample.
Calculation of results
The results are calculated in comparison to a calibration curve.
The specific reference curve (master curve) for a reagent lot is in the file containing all the data for that lot, that
is provided on the CD accompanying the cartridge. This information is registered in the analyser’s database
when the CD is introduced on the controlling computer.
If a new reagent lot is used, this curve must be registered on the database and then adjusted by a 2 point
calibration before sample results can be calculated.
2 point calibration
Analyser-to-analyser variation and different reagent lots will require adjustment of the master curve
(calibration). This is done by assaying two calibrators supplied with the reagent cartridge. The analyser’s
software will automatically perform the data processing to generate a new curve by defining new values for
two of the four parameters considered to be critical.
In practice, the 4 parameters of the reference curve specific to the reagent cartridge lot are loaded in the user’s
analyser via a CD provided with the kit. Calibration of the test must then be requested prior to use. The
calibration is performed and then verified by assaying one or more controls. Subsequent calibrations must be
repeated regularly in accordance with each assay’s instructions for use or as prompted by the analyser.
Where:
• VT = Total volume.
• VE = Sample volume.
• l = Optical pathway (1 cm).
• ε =
Molar extinction coefficient of analyte (in L.mol -1.cm -1).
The calibrations and controls can be programmed on demand or automatically managed in terms of frequency
by the analyser.
A request for calibration is automatically accompanied by a request to perform QC control.
Requests for calibrations and controls can be made at any time. If the analyser is in the process of carrying
out the assay, the calibrations and controls take place prior to the analyses requested on the samples.
The analyser is intended for professional use and must only be used by trained personnel in compliance with
the following safety precautions and good laboratory practice (GLP).
• Maintenance operations must be carried out at the frequency stipulated for each type of maintenance
activity. As some parts of the analyser are in contact with the biological samples, they must be considered to
pose a potential risk of infection.
• Validation of biochemistry reagents is the responsibility of the user and IDS takes no responsibility for the
validity of results.
• In order to guarantee the thermo-regulation of the carousel, ambient temperature must be lower than 30°C.
If necessary, provide air conditioning for the site.
!
• Precaution which MUST be respected
WARNING:
Menu bar
Provides access to the different pull-down menus.
Sample compartment
The illustration of the sample compartment depends on the analyser configuration. For the two
configurations (fixed sample tray or removable trays), the occupied positions are shown.
Each position can be occupied by:
• Calibrators.
• Controls.
• Samples.
• Stat.
An icon identifies the type of sample installed in a position (see Section 3, page 67).
For each occupied position the following options are available:
• Entry of the identifier and the type of product installed (without a barcode).
• Programming of the sample analyses to be carried out.
• Results display for controls or samples.
A colour coded analysis status is associated with the center of the position (see Section 3, page 80).
A colour code is associated to the contour of the positions, to inform whether the system is using the installed
product (see Section 3, page 68). During a run cycle:
• a blue contour colour indicates that the installed product is scheduled in the workload and can still be
removed; remove this product and the planned workload will be modified.
• a green contour colour indicates that all the tests programmed for the sample are completed. It is
possible to configure the software in LOCAL SYSTEM SETTINGS menu to change the contour colour at
the end of sampling rather than completion of assays. The same colour code is applied on each tray for
the removable-tray configuration.
Removable trays
Fixed tray
Status bar
Number of incomplete profiles in the work list with at least one assay
programmed but not completed.
2-3- Menus
Other functions are accessed by using the pull-down menus on the interface.
Allows the user to update the analytical configuration from the CD-
ROM Master Database provided by Immunodiagnostic Systems.
Allows the user to exit the software.
Allows the user to start up the analyser. All modules will be initialised
and automatically primed. The reading modules are automatically
controlled.
Once start-up is complete, the option becomes inactive (grey).
Allows the user to start the run cycle: (see page 22).
Once the analyser is in assay mode, the option becomes inactive.
This option is active when the analyser is in run cycle.
Allows the user to stop the assay process.
This option is active when the analyser is in run cycle.
Displays information concerning the tests in process. The software
displays the time of processing and when each assay’s results will be
available.
Allows the user to view the available volumes and the number of
tests for each on-board reagent, before or during the run cycle.
Programming and processing of tests for the reagent Cartridge
Check System (CCS).
Simplifies programming of the work list by applying the same profile
to each occupied position.
Active only when removable-tray configuration (use in fail mode)
Simplifies programming of the work list by applying the same profile
to each occupied position of a removable sample tray.
Management of user access: when an access code is entered, a
specific session is opened. The user’s identifier and the level of
authorisation are displayed at the top of the title bar. Access codes
and identifiers are programmed in the SETTINGS menu.
Keep a record of your access codes. If lost, Operator sessions can be
opened by using the code HELP.
Each access will be recorded.
Allows access to all the calibrations currently valid for each test in the
personal library. Displays the active calibration of a reagent lot for
each test of the personal library
Allows access to the stored results. When opened, the results are
displayed. Various filters can be applied to display results for a single
date, between two dates or for the entire storage period.
User maintenances:
• XYZ adjustment: adjustment of the probe reference position.
• Daily maintenance: table of daily maintenance.
• Weekly maintenance: table of weekly maintenance.
• Monthly maintenance: table of monthly maintenance.
• Maintenance history: allows user to display and print the
maintenance carried out.
• Trace machine: used at the request of IDS Technical Services &
Support; allows the exporting of data from the device for Service
diagnostics.
Allows access to the list of all products and suppliers stored in the
system.
F2 Work list.
F4 Storage.
F5 Calibrations.
F8 Status of analyser.
• Insert the lid support tool onto the piston rod until the clips is fully inserted.
3-1- Start up
3-1-1- Initial Start up
• Switch the analyser on by pressing the switch located on the left-hand side to position “1”.
• Switch the computer on, then open the software by double-clicking on the IDS-iSYS icon.
• Once the software is open, enter your access code to open a session.
• Start the analyser by selecting Start up from the session menu:
All modules are initialised and the ancillary reagents are automatically primed.
! If a new version of the software is detected on the analyser after opening, a message about
downloading this new version to the analyser will appear.
Take care when opening and closing the reagent compartment: rough handling when opening
! and closing the drawer may cause the internal reagents to spill.
During the run cycle, if a reagent is to be aspirated within 2 minutes, the drawer light flashes
red and green.
! When opening the reagent compartment during the run cycle, assays which are under way may
be stopped if a reagent was scheduled to be added when the compartment was open. In the
event of this, the rescheduled assays are automatically added to the end of the work list.
Access to the reagent compartment is controlled. The light located on the front face of the drawer indicates
whether access is authorised (green indicator) or if the drawer is locked (red indicator).
For cartridge handling, refer to the reagent instructions for use (IFU).
! Magnetic particles in certain Immunoassay cartridges require a particular mixing step before
loading the cartridge into the rack. Follow the instructions for mixing described in the reagent
instructions for use.
Rack 6 x 20 ml Rack 3 x 50 ml
If the reagents have been stored on-board and the analyser has been put into standby mode, the identification
of these reagents will be restored upon start up.
Reagent cartridges can also be installed during the run cycle.
The magnetic particle vial of Immunoassay cartridges is continuously stirred when in the reagent compartment.
The pictogram below appears on the reagent cover, and describes the precautions to be taken when loading a
reagent rack:
The reagent cartridges are automatically identified by the barcode reader as the rack is inserted in the rail.
On the interface, the reagents are identified and displayed in green with the corresponding lot number.
The countdown sequence until use begins automatically.
Each time a cartridge (or a reagent bottle) already stabilized on-board is removed then
! installed again in the compartment, a countdown before use begins, even for an immediate
reloading.
Handle the reagent racks only when necessary.
! If positive identification has failed, the reagent cartridge position is displayed in orange with
“???” instead of the name and lot number.
It is imperative, when installing a cartridge or a reagent bottle, that each occupied position of
• Should identification fail, repeat the procedure for putting the reagent rack in position.
• If the position is still not identified, click on the position.
• Select the type of reagent rack installed.
• To select an Immunoassay rack, click on “Cartridge” and use the keyboard to manually enter the barcode
identifier for the cartridge.
The information relating to the Immunoassay reagent cartridge (composition, reagent lots, expiry date,
master curve) is automatically associated with the barcode reading and the position is displayed in green.
This information is also contained in the CD supplied with the kits.
• If this information is not available, the position of the reagent is displayed in white with red stripes.
• In this case, install the CD supplied with the kit and use the keyboard to re-enter the barcode identifier.
• To select another type of rack, click on “Rack”.
• Then select the type of rack installed:
- 3 x 50 mL, for Biochemistry reagents.
- 6 x 20 mL, for Biochemistry reagents.
3 x 50 ml rack
6 x 20 ml rack
• Then click on the position and use the keyboard to enter the barcode identifier for the bottle, and select the
reagent installed.
• The information about the bottle (type of reagent, lot, expiry date) is automatically associated with the
barcode reading. This information is also contained in the CD supplied with the reagents. If the information is
not available, the position of the reagent is displayed in white with red stripes.
• In this case, install the CD supplied with the kit.
A countdown may be stopped in certain clearly identified cases (e.g., quick unloading of a reagent previously
on board):
The reagents must be stabilised at the temperature of the refrigerated compartment before
use. Magnetic particles in Immunoassay cartridges must be sufficiently stirred.
! Wait for 40 minutes after installing reagent cartridges before starting assays (20 min for
Biochemistry).
Any interruption of countdown before use may impair the analytical quality of results.
3-2-1- Colour codes associated with reagent positions in the reagent compartment
Rail free.
Reagent correct.
Immunoassay cartridges
Deadline for using the cartridge on the system. Date is replaced by expiry time
when the day of expiry date begins.
A calibration will be requested in the next 24 hours for tests using a calibration
frequency managed by the system.
The calibration was recalculated after excluding at least one calibrator replicate
Status of the current calibration for the reagent lot (see page 44).
Calibration status
The icon on the cartridge shows the status of the current calibration.
Not calibrated
Valid Calibration
Biochemistry reagents
By using these stability data, the system manages the expiry of each individual cartridge or reagent bottle
installed in the refrigerated compartment. This expiry date is established with the two following dates:
• lot expiry date,
• on board expiry date, defined as the use-by date for using the cartridge or the reagent bottle.
The on board expiry date d’expiration is established by a dynamic calculation based on the following data:
• cumulative on board time,
• on board stability,
• in-use stability,
The on board expiry date (in days and hours) is initially calculated from the first installation date of the reagent
and from the on board stability supplied with the lot data. The expiry time is set at 23:59 for all reagents.
Then this date is constantly updated taking into account the cumulative on board time. This date is finally
compared with the in-use expiry date and with the lot expiry date.
The cartridge or bottle can be used up to the date and time displayed as its on board expiry date.
The date is replaced by expiry time (23:59 ) when the day of expiry date begins.
!
If the analyser is carrying out an assay, only the IDS-iSYS Cuvettes cube and the IDS-iSYS
System Liquid can be replaced. The other ancillary reagents cannot be replaced until the
analyser has completed the cycle.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open and the lid support tool must always be installed.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Remove the cap from the new ancillary reagent container.
• If necessary, take out the bung and remove the empty container.
• Scan the ancillary with the barcode reader.
The analyser will beep for the first time.
• Within 10 seconds, install the new bottle in its position identified by colour code.
• Install the supply line into the bottle.
The analyser will beep a second time.
• If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the identifier.
• Remove the lid support tool and close the lid (see page 36).
• Once the new bottles are installed, a message proposing an automatic priming is displayed:
! Refer to the catalogue for the product references of the ancillary reagents to be ordered.
! Refer to the catalogue for the product references of the ancillary reagents to be ordered.
Installation of AP Substrate
• Remove the cap from the new ancillary reagent bottle.
• If necessary, take out the bung and remove the empty bottle.
• Scan the ancillary with the barcode reader.
The analyser will beep for the first time.
• Within 10 seconds, install the new bottle in its position.
The analyser will beep a second time.
• Install the supply line into the bottle.
• If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the identifier.
• Once a new bottle is installed, a message is displayed:
• Remove the protective plastic strip: pull forwards to break the seal. Gently pull upwards to remove the
protective strip. Place the protective strip over the cube to protect the window from dust.
• If the barcode is not read by the barcode reader, click on the ancillary reagent then use the keyboard to enter
the barcode identifier.
! Refer to the catalogue for the product references of the ancillary reagents to be ordered.
• If results from a previous profile are present in the list of current results, click on .
The results are sent to storage and the list is released for a new profile.
• From the drop-down list, select the necessary test and programme the required number of replicates (see
CCS IFU) then click on .
• Repeat the operation for each test required.
• Click on .
• In Results storage menu select the CCS results you wish to export, then click on .
• Validate the selection corresponding to the results you wish to export.
• Enter the file name [Date CCS] and save the txt file in the assigned folder.
This will save the raw CCS data in a text file (.txt).
• Minimize GUI software then open assigned CCS folder.
• Right click on desired .txt file. Choose "Open With".
• Select «OpenOffice Scalc».
• On the "Text Import" screen, click on "OK".
• In the OpenOffice Scalc worksheet, highlight line 1 to line 18 (results for CCS).
Many important data are not visible on the screen as some columns in the file are not used
and the ancillary lot numbers are hidden on the right side of the file (See sign in red).
! All data including those not visible on the screen should be copied and pasted into the “QC
outil client CCS” file
All complete lines should be highlighted to be copied.
• With arrow on highlighted area, right click mouse and choose "Copy" (or press CTR+C).
• Minimize OpenOffice Scalc worksheet.
• Open the “QC outil client CCS” (EXL-CQR-001V2) with OpenOffice ‘Scalc” from CD drive (CCS mini CD-
ROM).
• Select the cell A2 then Right click on mouse with the arrow on the cell A2 and choose "Paste" (or press CTR
+V) to paste data into template.
• Verify that the data are paste in the correct order and all in triplicates.
• The data are automatically calculated and all necessary information is transmitted into the CCS result form.
• Validate the analyser performance using this data as directed by IDS Service and Support Personnel.
Do not perform immunoassays if the analyser performance level is not satisfactory (valid).
! Refer to Section 6 of this manual Troubleshooting (see Section 6, page 143).
• Complete the form with the laboratory name, the Operator name, the instrument S/N and GUI version. ( time
can be saved by saving this information in a template file in the assigned folder).
• Select File (from the toolbar) "Save As" (from the drop down menu) and rename the file using "date CCS» to
save in the assigned folder, then click on “Save”.
• Select "Keep current format.
• If a printed copy is required, print the CCS results.
The appearance of automatic actions are indicated with the icon flashing in the WARNING button.
• From the main screen, click on the WARNING button.
• All the automatic requests for calibration generated by the system are shown in the lower part of the window,
under the label «Automatic Action». The reagent lot is displayed along with the demand.
• A calibration request is accepted by clicking on .
Comment: an automatically generated request for calibration can be cancelled by clicking on . In this
case, a record is made in the event log. If the current calibration has expired, results will not be calculated.
Programming a calibration
• From the main screen, click on
The list of calibrations for each reagent lot is shown according to the selected display option:
! When calibrating a new Immunoassay reagent lot, the calibrators associated with this lot must
be placed into the sample rack before programming the request.
• Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
• Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
Comment: The request can be cancelled by clicking on .
• Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
Comment: The request can be cancelled by clicking on .
! Calibrations for several lots of the same Biochemistry reagent can be programmed
simultaneously.
! The current calibration can be displayed for a reagent lot by double-clicking on the lot number
or on the display area (light green area).
The appearance of automatic actions are indicated with the icon flashing in the WARNING button.
• From the main screen, click on the WARNING button.
All the automatic requests for controls generated by the system are shown in the lower part of the window,
under the label «Automatic Action». The reagent lot is displayed with the request.
• A control request is accepted by clicking on .
Comment: a request for control generated automatically can be cancelled by clicking on . In this case, a
record is made in the event log.
! Controls programmed by accepting automatic actions are always run in the number of
replicates defined in the assay setup.
• Select the relevant number of replicates for the control when applicable.
When appropriate, the number of replicates programmed is shown in grey.
• Click on to confirm the control programming.
Comment : Control request can be cancelled by clicking on .
• Load the necessary calibrators and controls onto the analyser.
If the value of a calibrator or control is not defined for the active lot, a window automatically
Access to the sample compartment is controlled. if the light is green, access is available. When assays are
under way, the sample compartment is locked (red light).
When the sample compartment is opened during the run cycle, new tests will be delayed.
The reagents will continue to be aspirated for the tests which are already running.
! Once the compartment is closed, the analyser can resume the sample workload, including
the new tests.
Close the sample compartment as soon as the loading is finished in order to maintain the
performance of the analyser.
Any position of the sample compartment may be used to install samples, calibrators and controls. Each of
these have an associated icon shown in the table below:
Calibrator
Serum, Plasma
Sample
Control
Urine
Sample identifiers (barcodes or manual entries) must not contain the character «%». This
When using the 64 position sample tray, sample cups and paediatric tubes must be placed in
! the tube adapters provided with the system to guarantee the quality of sampling.
When not on-board, removable trays should be placed on a flat solid surface, to avoid any risk
of spillage.
! During loading/unloading operations in the sample compartment, handle the removable trays
with care.
Rough handling may cause the samples to spill.
• Then identify the position(s) not identified by the barcode reader (see Identify the positions of the tray
without barcode reading, page 63).
! If positive identification has failed, the rail position is displayed in orange with “???” instead of
the removable tray representation.
• Should identification fail, repeat the procedure for putting the sample tray in position.
• If the rail is still not identified, the tray must be identified via the user interface.
If a barcode of one of the samples installed on the removable tray contains a %character the
tray is not identified and the rail position is displayed in orange with «???».
! In this case, remove the relevant sample from the tray, then place it again at its position with its
barcode label facing inwards. Once the removable tray is installed into the compartment, this
sample must be identified using the keyboard.
Configure a tray
• Click on the tray position:
• Select in the corresponding dialog boxes the type and number of the tray.
• Click on .
The corresponding tray is displayed on the interface, with each position considered to be occupied but not
identified. The software opens a window which allows the positions to be identified:
• Next, identify the occupied positions (see Identify the positions of the tray without barcode reading, page
63).
When the tray has been identified by the barcode reader, and when the window is
displayed, the tray can be removed to check identifier(s).
! In this case, information can be entered, but can only be validated only after replacing the
tray in the same position, with a new automatic identification.
Should automatic identification fail, the information will not be saved.
In this case, click on .
For the calibrators and controls supplied by IDS, enter the 12 figures from the barcode: the
! product, the lot number, the expiry date as well as the table of corresponding values are
automatically entered.
Take care when opening and closing the sample drawer: rough handling when opening and
! closing the drawer may cause the internal samples to spill.
For the calibrators and controls supplied by IDS, enter the 12 figures from the barcode: the
! product, the lot number, the expiry date as well as the table of corresponding values are
automatically entered.
• Click on .
Samples
• Use the keyboard to enter the identifier (1 to 32 alphanumeric characters).
• Select “type of product” as Sample and click on .
For samples whose profiles have already been programmed or sent through via a centralised computer, the
profile in the memory is automatically associated with the sample position.
• If the profile is not in the memory, the profile programming window opens (see Section 3-7, page 69).
Sample identifiers (barcodes or manual entries) must contain only alpha-numeric and blank
characters.
Removable trays
If the analyser is initialised but not in the run cycle, the compartment remains unlocked and the assays will be
carried out as soon as the run cycle is started. Each on-board tray will be assayed whether defined for
automatic start or not.
If the analyser is in the run cycle, the compartment locks automatically.
For the trays defined for automatic start (in Rack Configuration menu), any new assays waiting to be carried
out are added to the analyser’s workload.
For the others, the new assays will be added to the analyser’s workload after clicking on the corresponding
button.
Fixed tray
If the analyser is initialised but not in the run cycle, the drawer remains unlocked and the assays will be carried
out as soon as the run cycle is started.
If the analyser is in the run cycle, the drawer locks automatically and any new assays waiting to be carried out
are added to the analyser’s workload.
Position free.
Position identified.
Rail free
During the run cycle, the tray displays in green when all corresponding assays are completed, or if the option is
selected in SETTINGS LOCAL SYSTEM, when all corresponding assays are pipetted.
Lights are placed in front of each rail of the compartment:
Tray in-use
Sample identifiers (barcodes or manual entries) must not contain the character «%». This
The validity of results obtained depends on the correct programming of the sample type.
! When samples other than serum, plasma or urine must be assayed, refer to the reagent IFU or
contact Technical Services.
! If the patient profile is programmed during the run cycle, the sample status may immediately
be displayed as loaded.
Do not open the analyser lid during the run cycle, as this will cause the moving parts to stop
! immediately for safety purposes and all assays underway will be lost.
The solid waste can contain 400 cuvettes. Empty the solid waste before to start the run cycle.
3-8- Assays
• Click on .
• If all the items required for carrying out the programmed workload are available, the analyser initialises and
begins aspirating samples.
• If any of the items required to carry out the programmed workload are missing, a window is opened which
lists them all.
• Install the missing items then click on . The analyser will begin aspirating samples.
• The run cycle monitoring window displays. After 30 seconds an estimated time for the end of processing all
the on-board samples is displayed.
The run cycle will start even if some items are missing.
! The analyser will then carry out all the assays for which all required items are available.
The missing items can be installed at any time during the assay by following the procedures
described below.
Comment:
For the removable-tray configuration, each on-board tray will be carried out, whether or not defined for
automatic start.
At the run start, samples are tested in the order of the sample compartment positions.
A predictive list including all the tests that will be sent to the analyser is established when starting the run for all
the products installed onto the sample compartment, according to the priority order mentioned above. The
contour colour of each relevant position turns to blue.
Tests are sent to the analyser 5 at a time, up to a maximum of 90.
During a run cycle, a test may be terminated by the analyser as a result of certain faults, details of which can
be viewed by clicking the ‘Warning’ messages button. Faults may include:
• Calibration fails to comply with the criteria for automatic validation.
• The result of the control violates one of the Westgard rules (if activated).
• One of the reagents or a diluent required for the test has been detected as missing.
• One of the reagents or a diluent is empty.
When issues preventing test completion have been resolved, the test will be automatically restarted.
Removable trays
• Slide the removable tray containing the new samples into a rail in the sample compartment until the
positioning pin is fully inserted.
• The positions and the barcoded tubes are automatically identified by the integrated barcode reader as the
tray is inserted in the rail.
• On the interface, the tray is displayed with the free and occupied positions. Each identified position is
labelled with the corresponding icon.
For tubes from patients whose profile has already been sent through via a centralised computer or manually
programmed, the profile in the memory is automatically associated with the position.
• If the barcode is not read by the barcode reader or is missing for a position, identify the products installed
(see page 63).
• Proceed in the same way for other trays to be added.
• Close the compartment.
• The trays defined for automatic start are automatically added to the analyser’s workload.
• For the others, the new assays will be added to the analyser’s workload after clicking on the corresponding
button.
Fixed tray
• Scan the barcode with the reader located on the front face. The analyser will beep when the barcode has
been correctly read.
• Within 10 seconds, place the sample in a free position on the rack. If the sample is recognised, the analyser
will beep a second time. On the interface, the display will indicate that the position has been identified. If no
second beep is heard, the position is considered to be occupied but not identified. In this case, re-start the
barcode reading and installation process.
• If the barcode label is illegible, click on .
• Use the keyboard to enter the barcode identifier.
• Select ‘Vial association with samples tray’ and click ‘OK’.
• Select the appropriate identifier (Sample).
• Click on .
• Within 10 seconds, place the sample in a free position on the tray.
• Proceed in the same way for other samples to be added.
• Close the drawer.
• The analyser will resume sample aspiration once the drawer is locked.
After loading a reagent into the compartment, the lag time before use is automatically
managed by the system.
The reagents must be stabilised at the temperature of the refrigerated compartment before
When the reagent compartment is opened, assays which are under way may be cancelled if
reagent aspiration was scheduled.
! It is essential for Immunoassay reagent cartridges to remain on the analyser until all the results
are obtained.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
Clot detected
When a clot is detected during sample pipetting, the arm stops over the position where the fault is detected
and the probe waits over the sample concerned.
• Select the fault then click on . The sampling probe moves up.
The following message displays:
• Open the lid and insert the lid support tool (see pages 35 and 36). The sampling arm stays at its position.
• Install the specific protective tray under the probe while holding the sampling head.
• Remove the clot under the probe.
• Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
• Remove the protective tray.
• Remove the lid support tool and close the lid (see page 36), then click on . The cycle
resumes at the end of the rinse.
• Click on .
A message confirming the request displays:
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Install the specific protective tray under the probe while holding the sampling head.
• Unscrew the probe.
• Unclog the probe (see Section 5, 5-4-11- Unclogging the sample probe, page 141).
• Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
• Click on . The following message displays:
• Reassemble the probe by screwing it in by hand as far as possible to avoid risk of leakage.
• Remove the protective tray.
• Remove the lid support tool and close the lid (see page 36), then click on .
• Then, adjust the probe reference position (see Section 5, 5-4-10- Adjustment of probe reference position,
page 132).
• Click on to continue the run cycle.
The option indicating the end of sampling does not take into account the dilution, if applicable,
when a result is outside the field of measurement.
! If a sample cannot be diluted, the corresponding identifier will be displayed as in error in the
work list.
The number of incomplete profiles is indicated in the status bar.
Removable trays
• Check that the light of the corresponding rail is green.
• Remove the tray(s).
• If necessary, install new tray(s).
• Close the sample compartment.
A completed tray can contain one or more positions in error (position code ).
Fixed tray
• Remove each completed sample.
• If necessary, install new sample(s).
• Close the sample compartment.
Option
As soon as the result of the analyte concerned by a reflex test meets the defined conditions, additional assays
are automatically added to the profile. The relevant assays are included in the list of assays and scheduled
after calibrations, controls and STAT samples according to the order of the sample compartment positions.
Option
An automatic action is generated by the system as soon as reflex tests must be carried out for a patient. The
appearance of this automatic action is indicated with the icon flashing in the WARNING button.
Reflex tests can be requested either by accepting the automatic action or from the relevant profile edition.
• Then click on .
Option
Post dilution is automatically performed.
Option
When a patient must be diluted, an automatic action is generated by the system. The appearance of this
automatic action is indicated with the icon flashing in the WARNING button. A dilution can be requested
either by accepting the automatic action or by programming in the profile edition.
Accepting the automatic action
• From the main screen, click on the WARNING button.
• Then click on .
• Click on :
• Then click on .
Sample identifiers (barcodes or manual entries) must not contain the character «%». This
Removable trays
• Install the tube on a removable tray. Slide the tray into a rail in the sample compartment until the positioning
pin is fully inserted.
• The positions and the barcoded tubes are automatically identified by the integrated barcode reader as the
tray is inserted in the rail.
• On the interface, the tray is displayed with the free and occupied positions. Each identified position is
labelled with the corresponding icon.
• If the barcode label is illegible, identify the products installed (see page 69).
• Proceed in the same way for other trays to be added.
• Close the compartment.
• The trays defined for automatic start are automatically added to the analyser’s workload.
• For the others, the new assays will be added to the analyser’s workload after clicking on the corresponding
button.
Fixed tray
• Scan the barcode with the reader located on the front face. The analyser will beep when the barcode has
been correctly read.
• Within 10 seconds, place the sample in a free position on the rack. If the sample is recognised, the analyser
will beep a second time. On the interface, the display will indicate that the position has been identified.
• If no second beep is heard, the position is considered to be occupied but not identified. In this case, re-start
the barcode reading and installation process.
• If the barcode label is illegible, click on , then select ‘Vial association with samples tray’.
• Use the keyboard to enter the barcode identifier, then select the appropriate identifier (Sample).
• Click on .
• Within 10 seconds, place the sample in a free position on the tray. The position is configured.
• Close the drawer again.
The analyser starts initiating tests again once the drawer is locked and deals with the ‘STAT’ sample
immediately.
3-10- Results
As the run cycle progresses, the status of the sample tray position changes when all the tests associated with
this position have been completed.
The colour code associated with the output of results is as follows:
Samples: the results are stored in accordance with the criteria defined for
automatic filing and received by a centralised computer system (if the
connection is activated).
Calibration validated.
• To display the calibration curve for a test, double-click on the reagent lot number.
• The software displays the date and the status of the calibration and controls. The value calculated for each
control associated with the calibration (QC1 for Biochemistry; QC1 to QC4 for Immunoassay), is also given.
• If the calibration is validated, the message OK is displayed.
• In the event of an invalid calibration, a message is displayed at the top of the screen indicating the cause
(see Section 4-1, page 105).
• When it is not possible to automatically validate a calibration, the button is displayed when
the user has a Supervisor level of access (or above).
When a calibration is displayed before each associated control has been calculated,
! and buttons are not displayed.
The general tab shows whether a calibration has passed or not, type of
curve fit used, analyte it refers to and date of calibration. It also displays
data for the controls including target values, mean obtained values, CV%
and status (i.e. pass/fail).
Displays the calibration curve and each calibrator:
• Raw data obtained.
• Target value.
• Value calculated using the mathematic model of the calibration.
• Calibrator lot.
Displays value of coefficient of determination (r2).
Allows calculation of multiple-point calibrations after removing a calibrator.
Invalid Calibration:
If the calibration is invalid, a message is displayed at the top of the screen indicating the cause
(see Section 4-1, page 105).
• Request a new calibration after having replaced failed calibrator/control.
The software allows the user to validate a calibration when not automatically validated.
The validation or modification of an invalid calibration requires a Supervisor level of access (or
! above).
The button is not displayed when user has an Operator level of access.
! If the reason for validation failure has been clearly identified and deemed to have no impact
on results, the calibration may be accepted at the discretion of the user.
• When points have been excluded, the icon is displayed with the calibration date.
• When the recalculated calibration is valid, with the control(s) within acceptable limits, the message OK is
displayed.
• Click on to save the modification.
• When points have been excluded, the icon is displayed with the calibration date.
• When the recalculated calibration is valid, with the control(s) within acceptable limits, the message OK is
displayed.
• Click on to save the modification.
Control
data
No
12s In control -> Accept run
Yes No
No No No No
13s R 4s 22s 41s 10x
The violation of certain rules will generate automatic requests. In this event, the icon flashes in the
WARNING button.
• The violation of 2 2S, 4 1S, 10 X rules will automatically generate a request of calibration.
• Click on the WARNING button to open the window.
• Click on to request the calibration and the controls. If necessary, install the required calibrators and
controls on the analyser.
• To cancel the calibration request, click on . This action will be detailed in the Records.
• Then release to restart the test.
• All the results are identified by the message WE!.
• If another rule is violated, the test is stopped.
• Perform the necessary actions to correct the problem and request a calibration or a single control.
• Each time a violation occurs, a message displayed in the WARNING window allows the release of the
violation, whatever rule is violated.
• To release a test, click . This action will be detailed in the Records. All the results are identified by the
message WE!.
If the analyser is connected to a centralised computer system, the information sent for results
outside the field of measurement depends on the selected connection protocol:
• ASTM Compatible: results are always sent in the format * * * * * ..
• ASTM V2 Compatible and ASTM V3 Compatible: results are sent in the following format:
! - OMR-: < Low limit of the sample measurement range, even when the measuring range
starts at 0.
- OMR+: > High limit of the sample measurement range.
Limits are expressed in working units.
• To print out all the results of the work list, select “Print results” and click on .
Programming reflex tests
• Select the relevant profile.
• Double-click on the line to display the profile.
For each result, the display provides the calculated result ( column), the raw RLU measurement
( column), with their relevant units and the messages. A comma displayed after the last message
indicates that all the messages cannot be shown in the column. A double-click on the line will open a window
containing the full list of messages. For assays performed in replicates ( column), the mean of the
measurements and of the results are displayed in bold, accompanied by each corresponding CV.
Results are displayed in the order of positions in the sample compartment. Other sorting criteria may be
selected for the display by using one of the following options:
• To print out all the results of the work list, select “Print results” and click on .
Message Meaning
AIN Spectrophotometry
An absorbance is below the limit of reaction defined in the analytical configuration.
* * * * * is displayed instead of the result.
ASU Spectrophotometry
An absorbance exceeds the limit of reaction defined in the analytical configuration.
* * * * * is displayed instead of the result. The sample will be automatically diluted if a
dilution rate is defined in the analytical configuration.
If the ASU message accompanies the message “RED”, one of the absorbances is still
outside the limits of reaction defined after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
BLR Biochemistry
Absorbance of blank outside the limits defined in the analytical configuration.
CDE Spectrophotometry
Slope of the kinetics has been calculated on a non-linear section of the reaction
(for KINETIC type of assay).
The coefficient of determination is lower than the limit value defined in the analytical
configuration.
CE All fields
Error when calculating a parameter calculated.
CLU Luminescence
The RLU measurement was corrected.
One measurement of the integrated signal, out of the scope of the luminometer, was
excluded and replaced by a point calculated by smoothing.
CT All fields
Temperature of the carousel outside limits or not stable during the assay.
Message Meaning
EDE! The calibration used for calculating the result was obtained with at least one calibrator
excluded.
KIN! Spectrophotometry
The absorbances measured while monitoring the reaction are not strictly increasing or
decreasing (non-monotony of the reaction).
* * * * * is displayed instead of the result.
LS Spectrophotometry
The result was calculated from absorbencies measured with an unstable lamp.
MAR Luminescence
This message is applicable to semi-quantitative assays.
The luminescence signal measured is below the minimal acceptable signal.
In this case, the sample must be tested again.
* * * * * is displayed instead of the result.
MES Luminescence
Error when recovering the measurements.
Spectrophotometry
Error when recovering the measurements, or the result can not be calculated because the
number of measurements insufficient.
MRE! Luminescence
The calibration used for calculating the result was obtained with at least one relative light
unit (RLU) measurement excluded.
OMR It has not been possible to calculate the result because the calibration is not valid.
* * * * * is displayed instead of the result.
OMR- The result calculated is below the lower limit of the field of measurement.
«< Low limit of the sample measurement range» is displayed instead of the result.
Message Meaning
OMR+ The result calculated is above the upper limit of the field of measurement. The sample will
be automatically diluted if a dilution rate is defined in the analytical configuration.
«> High limit of the sample measurement range» is displayed instead of the result.
If the message OMR+ accompanies the message “RED”, the result is still higher than the
upper limit of the field of measurement after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
ORA Luminescence
The result cannot be calculated as the measurement is outside the scope of the
measurements defined by the master curve. * * * * * is displayed instead of the result.
PMS Luminescence
Problem with luminometer stability.
RDE! Luminescence
The calibration used for calculating the result was obtained with one of the replicates for
calibrator excluded.
RDS Immunoassay
Problem with agitation of magnetic particles on the reagent rack.
RED The result was calculated after a dilution. The value given takes into account the dilution
factor.
SUB Spectrophotometry
Detection of enzymatic hyperactivity. The change in absorbance per minute (∆A/min)
calculated between the injection of the sample and the first measurement used in the
calculation exceeds the programmed limit.
* * * * * is displayed instead of the result.
The sample will be automatically diluted if a dilution rate is defined in the analytical
configuration.
If the SUB message accompanies the message “RED”, enzymatic hyperactivity is still
detected after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
SLU Luminescence
At least two measurements of the integrated signal are out of the scope of the
luminometer, * * * * * is displayed instead of the result.
WE! One of the Westgard rules applied for this analyte has been violated.
Message Meaning
• Select the relevant profile option to be stored (all profiles, selection, profiles without message),
then click .
The incomplete profiles can be stored when the analyser is not carrying out an assay.
This option automatically removes the analysis requests remaining in the profiles when storing the profiles.
• The stored profiles change to a yellow colour code.
• As soon as a position for which the profile has been stored is released, the profile will be available in results
storage.
! It is possible to request repeat analyses for a stored profile as long as the tube is still in the
sample compartment.
! In the STORAGE menu, results can be dispayed either in standard mode or in tabular mode.
The sorting options available in the work list menu can be applied to stored results.
• It is also possible to delete all the stored profiles from the work list.
• To do this, click on :
! been stored prior to remove them (display the work list in «standard mode», stored profiles are
with a yellow colour code).
Otherwise, profiles that have not been previously stored are deleted from the work list and
definitively lost.
In tabular mode display, when a result is selected, all the profile is removed.
• The system displays the active lot of the selected control. Another lot can be selected in the relevant
dropdown list:
• The software displays the control values for the selected time period along with the calculated mean,
standard deviation and CV.
The rules to apply are selected for each test in the 6 rules defined as follows:
Rules Inspection:
The rules defined in the set-up test will be tested in the following order:
Control
data
No
12s In control -> Accept run
Yes No
No No No No
13s R 4s 22s 41s 10x
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
• Check that the work list is empty. If necessary, delete profiles that are no longer required.
• Check the time and date of “wake-up” programmed in the SYSTEM SETTINGS menu if applicable.
• From the Session menu, select SHUT DOWN to put the analyser into standby mode.
! For systems using DSORB solution, a maintenance procedure for the four washers will take
place automatically during the SHUT DOWN procedure.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
• Open the compartment by pushing on the cover, then remove the waste tray.
• Discard the disposable solid waste container and its content into a container reserved for disposal of
biological waste.
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
• Install a new disposable container into the solid waste tray.
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
• Remove the level sensor from the container collecting the liquid waste.
• Treat liquid waste prior to disposal in accordance with current local regulations (see APPENDIX I: Waste
disposal, page A2).
• Decontaminate the liquid waste container (see Section 5, 5-5-1- Decontaminating the containers of solid
and liquid wastes, page 147).
• Replace the level sensor in the empty waste container.
• Then release the solid waste alarm by clicking on .
• Click on the liquid waste icon.
A confirmation message displays.
• Click on to confirm the emptying of the waste.
• Exit the ancillary management window.
The container provided by IDS is intended to be used for liquid waste collection and is not
• From the File menu, click on EXIT, the software will close.
• Then turn off the computer following the procedure for shutting down WINDOWS.
• Finally, press the switch located on the left-hand side of the analyser into position ‘0’.
In order to maintain the performance of the operating system, the computer must be turned off
! one time per week.
When the analyser has been inactive for more than two days, a specific procedure must be
! applied depending on duration of the stoppage (see Appendix A-6- Analyser long stop period,
page A-8).
Messages 104
Message Meaning
When the controls carried out with the calibration do not meet the criteria for automatic validation, the status of
each control is represented by a red circle in the appropriate column(s) (e.g. QC3) and by a red half-circle in the
calibration column.
For Biochemistry assays, only QC1 is carried out with the calibration.
In this case, the corresponding assays will not be processed by the analyser until a new calibration has been
carried out and is correct, or until the calibration is forced. The messages associated with the calibration
control are displayed with the calibration curve in the control section.
Message Meaning
Message Meaning
AIN Spectrophotometry
An absorbance is below the limit of reaction defined in the analytical configuration.
* * * * * is displayed instead of the result.
ASU Spectrophotometry
An absorbance exceeds the limit of reaction defined in the analytical configuration.
* * * * * is displayed instead of the result. The sample will be automatically diluted if a
dilution rate is defined in the analytical configuration.
If the ASU message accompanies the message “RED”, one of the absorbances is still
outside the limits of reaction defined after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
BLR Biochemistry
Absorbance of blank outside the limits defined in the analytical configuration.
CDE Spectrophotometry
Slope of the kinetics has been calculated on a non-linear section of the reaction
(for KINETIC type of assay).
The coefficient of determination is lower than the limit value defined in the analytical
configuration.
CE All fields
Error when calculating a parameter calculated.
CLU Luminescence
The RLU measurement was corrected.
One measurement of the integrated signal, out of the scope of the luminometer, was
excluded and replaced by a point calculated by smoothing.
CT All fields
Temperature of the carousel outside limits or not stable during the assay.
Message Meaning
EDE! The calibration used for calculating the result was obtained with at least one calibrator
excluded.
KIN! Spectrophotometry
The absorbances measured while monitoring the reaction are not strictly increasing or
decreasing (non-monotony of the reaction).
* * * * * is displayed instead of the result.
LS Spectrophotometry
The result was calculated from absorbencies measured with an unstable lamp.
MAR Luminescence
This message is applicable to semi-quantitative assays.
The luminescence signal measured is below the minimal acceptable signal.
In this case, the sample must be tested again.
* * * * * is displayed instead of the result.
MES Luminescence
Error when recovering the measurements.
Spectrophotometry
Error when recovering the measurements, or the result can not be calculated because the
number of measurements insufficient.
MRE! Luminescence
The calibration used for calculating the result was obtained with at least one relative light
unit (RLU) measurement excluded.
OMR It has not been possible to calculate the result because the calibration is not valid.
«> High limit of the sample measurement range» is displayed instead of the result.
OMR- The result calculated is below the lower limit of the field of measurement.
«< Low limit of the sample measurement range» is displayed instead of the result.
Message Meaning
OMR+ The result calculated is above the upper limit of the field of measurement. The sample will
be automatically diluted if a dilution rate is defined in the analytical configuration.
«> High limit of the sample measurement range» is displayed instead of the result.
If the message OMR+ accompanies the message “RED”, the result is still higher than the
upper limit of the field of measurement after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
ORA Luminescence
The result cannot be calculated as the measurement is outside the scope of the
measurements defined by the master curve. * * * * * is displayed instead of the result.
PMS Luminescence
Problem with luminometer stability.
RDE! Luminescence
The calibration used for calculating the result was obtained with one of the replicates for
calibrator excluded.
RDS Immunoassay
Problem with agitation of magnetic particles on the reagent rack.
RED The result was calculated after a dilution. The value given takes into account the dilution
factor.
SUB Spectrophotometry
Detection of enzymatic hyperactivity. The change in absorbance per minute (∆A/min)
calculated between the injection of the sample and the first measurement used in the
calculation exceeds the programmed limit.
* * * * * is displayed instead of the result.
The sample will be automatically diluted if a dilution rate is defined in the analytical
configuration.
If the SUB message accompanies the message “RED”, enzymatic hyperactivity is still
detected after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
SLU Luminescence
At least two measurements of the integrated signal are out of the scope of the
luminometer, * * * * * is displayed instead of the result.
WE! One of the Westgard rules applied for this analyte has been violated.
Message Meaning
•
A request for calibration, QC1, QC2 or QC3 has been generated automatically.
These requests are generated in the following cases:
• No calibration in the memory for the analytes requested.
• Reagent lot present in the compartment is out of date.
• For Biochemistry if certain Westgard rules have been violated, a request for calibration or control is
generated depending on the rule in question.
• The calibration and/or control(s) will be carried out after acceptance by the user (by clicking ): If
the user does not want to accept these actions, it is possible to cancel the request by clicking .
This action will be detailed in the Records.
•
Daily, weekly or monthly maintenance has not been carried out.
Messages are displayed by clicking on the WARNING button. If items are missing, the installation of
the item automatically deletes the message.
Depending on the nature of the message, certain tests may stop. In this case, the analyser will
! not be able to complete the entire workload. To carry out the entire workload without affecting
the deadline for producing the results, it is important to act upon the messages as soon as
possible.
•
The temperatures of the carousel and the reagent compartment are continually monitored. If, during
operations, one of the temperatures is outside the acceptable limits, the analyser will continue to carry
out assays and the icon will be displayed in the SYSTEM STATUS button.
Daily maintenance activities are presented under a single tab. Certain maintenance activities are carried out
automatically by the system whilst others must be carried out by the operator. In both cases, the maintenance
activities are recorded.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open and the lid support tool must always be
installed.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
If the analyser is equipped with a locking system, unlock lid before selecting the
! MAINTENANCE menu (see page 35).
Close the lid when all maintenance procedures are complete (see page 36).
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
These parts of the analyser are in contact with biological samples. There is, therefore,
a potential risk of infection.
Wear disposable gloves for all handling procedures.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Gently clean the probe with a cloth soaked in deionized water to remove traced on the needle.
• Then use absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl) to decontaminate the needle.
• When the cleaning procedure is finished, verify the tightness of the needle.
• Remove the lid support tool and close the lid (see page 36).
! Handle the probe with care. Do not twist or bend the probe during cleaning.
! This maintenance activity is managed only for systems using DSORB solution.
For other systems, this maintenance activity is not required.
• Click . The probe will automatically aspirate the IDS-iSYS D-Sorb solution.
• When maintenance is completed without any error, the date and time it is carried out are recorded.
The weekly maintenance procedures are presented under two tabs: General and Immunology. Certain
maintenance activities are carried out automatically by the system whilst others must be carried out by the
operator. In both cases, the maintenance activities are recorded.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open and the lid support tool must always be
installed.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
If the analyser is equipped with a locking system, unlock lid before selecting the
! MAINTENANCE menu (see page 35).
Close the lid when all maintenance procedures are complete (see page 36).
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
WARNING:
RISK OF INJURY
If a position of the fixed tray has to be cleaned, take care not to push the wipe into
the position with your finger.
Risk of cuts from the detection sensors.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Pour alcohol at 70% v/v (ethyl or isopropyl) into the rinsing well.
• Use a cotton bud soaked in alcohol at 70% v/v (ethyl or isopropyl) to clean the well.
! This maintenance activity can be carried out with the lid opened.
When this maintenance is completed, proceed to the weekly shutdown of the computer:
• From the File menu, click on EXIT, the software will close.
• Then turn off the computer following the procedure for shutting down WINDOWS.
• Switch the computer on, then open the software by double-clicking on the IDS-iSYS icon.
• Once the software is open, enter your access code to open a session.
The monthly maintenance activities are presented under two tabs: General and Immunology. Certain
maintenance activities are carried out automatically by the system whilst others must be carried out by the
operator. In both cases, the maintenance activities are recorded.
! If the analyser is equipped with a locking system, unlock lid before selecting the
MAINTENANCE menu (see page 35).
5-3-1-3- Cleaning the IDS-iSYS D-Sorb pump and level sensor shafts
! This maintenance activity is managed only for systems using DSORB solution.
For other systems, this maintenance activity is not required.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open and the lid support tool must always be
installed.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Remove the cover located at the rear.
• Click on the corresponding button in front of this item in the maintenance list.
• Follow the instructions displayed on screen.
• Release the pump body from its shaft by pinching the two clips at either side and gently pulling, ensuring the
tubes remain connected.
• Clean the pump shaft with a cloth soaked in alcohol at 70% v/v (ethyl or isopropyl).
• Return the pump body to its shaft.
• Put the cover back in place.
• Remove the lid support tool and close the lid (see page 36).
• To clean the level sensor, use absorbent paper soaked in de-ionised water.
• When maintenance is completed, click on . The date and time it is carried out are recorded.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
• Click on the corresponding button in front of this item in this maintenance list.
• Remove the lid from the liquid waste bottle.
• Clean the level sensor with bleach (commercial preparation).
• Rinse the level sensor with de-ionised water.
• Replace the lid.
• When maintenance is completed, click on . The date and time it is carried out are recorded.
• Remove the de-ionised water bottle and replace the original AP Substrate bottle.
• Click on .
• When maintenance is completed without any error, the date and time are recorded.
WARNING:
RISK OF BURNS
Do not handle the lamp to be replaced immediately after turning off the analyser.
Before handling, allow the lamp to cool for approximately 10 minutes after switching
the analyser off.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Switch the analyser off and unplug the power cable.
• Remove the bottom partition at the rear of the rinsing well.
• Disconnect the lamp from its supply.
• Unscrew the securing screws located on both sides of the lamp in order to remove the lamp-support unit.
• Remove the old lamp.
WARNING:
RISK OF BURNS
• Put the new lamp and its base in place, with the guide pin positioned downwards.
• Tighten the securing screws using a screwdriver.
• Connect the lamp to its power supply.
• Replace the partition.
• Remove the lid support tool and close the lid (see page 36).
• Switch the analyser back on and start it up.
• Wait for 10 minutes, then request a measurement of the intensity values by the following pathway:
• From the list of elements, select ABS READER then click on . The analyser carries out a
measurement for each of its filters. When the procedure is finished, exit this menu.
• Click on the SYSTEM STATUS button and check the intensity values for each of the filters are between 900
and 3100.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe must be disposed of in accordance with current local regulations.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• When the lid of the analyser is equipped with a locking system, the sampling arm moved to its home
position.
• Install the specific protective tray under the probe.
• Unscrew the probe requiring replacement.
• Install the new probe by screwing it in by hand as far as possible to avoid risk of leakage. Take care not to
twist or bend the probe, always handling it by the threaded screw connector.
• Remove the protective tray.
• Remove the lid support tool and close the lid (see page 36).
• Adjust the probe reference position (see page 132).
• When the adjustment on the reference position is completed, exit this menu.
• Prime the fluidics circuit as follows:
WARNING:
RISK OF ELECTRIC SHOCK
The primary fuses are located in the mains plug located on the left-hand side of the analyser.
• Switch the analyser off and unplug the power cable.
• Using a screwdriver, remove the fuse-holder from its housing.
• Replace the faulty fuse, ensuring it is of the same value and the same type (time-delay fuse).
• Return the fuse holder to its housing.
• Switch the analyser back on and perform the start-up procedure.
WARNING:
RISK OF ELECTRIC SHOCK
• Insert a flat head screwdriver into the fuse-holder and turn anti-clockwise to unscrew.
• Take out the fuse-holder and replace the faulty fuse with one of the same value and the same type (time-
delay fuse).
• Return fuse-holder to its housing and screw firmly into place, taking care not to over-tighten.
• Replace the fuse protection plate.
• Switch the analyser back on again and perform the start-up procedure.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
Perform this operation only at the request of IDS Technical Service & Support personnel.
! Slightly lift the box to introduce the plate and then put if back before lifting the cube.
WARNING:
RISK DUE TO ARM MOVEMENT
The reference position of the probe must be adjusted each time the needle is replaced, or when requested by
the Technical Services.
This operation is considered as a particular use of the analyser and must be performed only by a user specially
trained to this process.
The different probe positions are defined in relation to this reference position , located to the left of the rinsing
well.
Adjustment must be defined for the three dimensions of the analyser’s functional plane (XYZ adjustments).
Depending on whether the locking system is present or not, the procedure is different. The operations to be
performed in each case are described in two different procedures.
If the lid is not equipped with a locking system, refer to the procedure described in chapter 5-4-10-1-
Adjustment procedure for analyser without lid locking system, page 133.
If the lid is equipped with the locking system, the activation key provided with the analyser’s accessories will
be used in this procedure. Refer to the procedure described in chapter 5-4-10-2- Adjustment procedure for
analyser with lid locking system, page 136.
WARNING:
USE OF THE ACTIVATION KEY
The activation key allows a particular use of the analyser, without user’s protection
against the pipetting arm movements. The pipetting arm will move with the lid
opened.
Contact with the moving pipetting arm may cause severe injuries.
The use of the activation key is strictly restricted to this adjustment procedure.
WARNING:
OPENING/CLOSING THE LID
The lid must be opened for this operation.
Always handle the lid carefully during opening and closing.
The lid must always be fully open and the lid support tool must always be
installed.
Take care not to knock the lid during any analyser intervention.
When the lid is not fully open there is a risk of it falling.
During closing, maintain the lid open when removing the lid support tool, to avoid any
risk of accidental falling.
• Click on .
• The probe will be placed over the reference position, located to the left of the rinsing well.
The probe is moved using the arrow buttons on the screen :
• Adjust first the XY positions of the probe, using the corresponding movement arrows (Left, Right, Forward,
Backward). The probe must be centered on the pin of the target.
• Then adjust the Z position: click on the Down (1 mm) button twice.
• Adjust the probe position using repeatedly the Down (0,1 mm) button. The probe must be adjusted just
above the center pin of the target.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Gently touch the probe: the probe must slightly resist to the lateral movement.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe can be moved using Up and Down buttons when the lid is open.
! The Left, Right, Forward, Backward buttons can only move once to each direction when the
lid is open.
• When the adjustment is completed, remove the lid support tool and close the lid (see page 36).
• Click on .
• The pipetting arm will move to the rinsing well before returning to the new adjusted position :
• Check the probe position, then remove the lid support tool and close the lid (see page 36).
• Click on .
The pipetting arm will move to the rinsing well.
A message confirming completion of the procedure is displayed :
• Click on .
The main position adjustment window is displayed.
• Click on to exit this menu
Comment:
The button is active in two windows. Clicking on this button will cancel changes in the position
adjustment and the current adjustment will be kept.
To keep the previous adjustment :
• Click on .
The pipetting arm will move to the rinsing well.
A message confirming cancellation of the procedure is displayed :
• Click on .
The pipetting arm moves to its home position and the lid is unlocked.
The message displays:
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Insert firmly the activation key into place.
WARNING:
USE OF THE ACTIVATION KEY
The activation key allows a particular use of the analyser, without user’s protection
against the pipetting arm movements. The pipetting arm will move with the lid
opened.
Contact with the moving pipetting arm may cause severe injuries.
The use of the activation key is strictly restricted to this adjustment procedure.
• Click on .
The pipetting arm moves to the probe reference position, located to the left of the rinsing well.
WARNING:
PIPETTING ARM MOVEMENT
During this operation, the pipetting arm will move with the lid opened.
Contact with the moving pipetting arm may cause severe injuries.
• The probe will be placed over the reference position, located to the left of the rinsing well.
The probe is moved using the arrow buttons on the screen :
• Adjust first the XY positions of the probe, using the corresponding movement arrows (Left, Right, Forward,
Backward). The probe must be centered on the pin of the target.
• Then adjust the Z position: click on the Down (1 mm) button twice.
• Adjust the probe position using repeatedly the Down (0,1 mm) button. The probe must be adjusted just
above the center pin of the target.
• Gently touch the probe: the probe must slightly resist to the lateral movement.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
• The pipetting arm will move to the rinsing well before returning to the new adjusted position :
• Remove the activation key. The removal is detected by the analyser and the message is displayed:
• Remove the lid support tool and close the lid (see page 36). The lid closing is detected by the analyser and a
window containing the settings values is displayed:
• Click on .
The main position adjustment window is displayed.
• Click on to exit this menu.
Comment:
The button is active in two windows. Clicking on this button will cancel changes in the position
adjustment and the current adjustment will be kept.
To keep the previous adjustment :
• Click on .
The pipetting arm will move to the rinsing well.
A message confirming cancellation of the procedure is displayed:
• Remove the activation key. The removal is detected by the analyser and the message is displayed:
• Remove the lid support tool and close the lid (see page 36). The lid closing is detected by the analyser and
the main position adjustment window is displayed.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe must be disposed of in accordance with current local regulations.
Use only the nylon mandrel provided for this purpose with the analyser.
! Do not use a metallic mandrel that will damage the internal surface of the probe.
• Open the lid and insert the lid support tool (see pages 35 and 36).
• If the lid of the analyser is equipped with a locking system, the sampling arm moves to its home position.
• Install the specific protective tray under the probe.
• Unscrew the sampling probe.
• Insert the mandrel into the bottom of the probe, and pass it through the entire length of the probe, until the
probe becomes unblocked.
• Outside of the analyser, verify that the probe is fully unblocked by passing deionized water through the top of
the probe with a syringe:
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The probe is in contact with biological samples. There is, therefore, a potential risk of
infection.
Wear disposable gloves for all handling procedures.
The probe must be disposed of in accordance with current local regulations.
• If the probe is properly unblocked, the water will exit the probe in a straight jet and the syringe plunger can
be pushed freely.
• If the syringe plunger must be strongly pushed and if water drips when exiting the probe, repeat this
procedure.
• Then clean the probe with an absorbent paper soaked in alcohol at 70% v/v (ethyl or isopropyl).
• Reassemble the probe by screwing it in by hand as far as possible to avoid risk of leakage.
• Remove the protective tray.
• Remove the lid support tool and close the lid (see page 36).
! Perform this procedure only at the request of IDS Technical Service & Support personnel.
If a module remains unusable once the error message is released, the faulty element is no longer operational
and the error message will remain. When this fault involves either one of the washers or a particular wavelength
of the spectrophotometer, the relevant element can be temporarily deactivated using the SYSTEM
CONFIGURATION menu. This should only be used when following guidance by IDS Technical Service &
Support personnel.
This operation requires a Supervisor level of access.
• In the MAINTENANCE menu, select SYSTEM CONFIGURATION:
Deactivate a washer
Deactivate a wavelength
As soon as the menu option is selected, the maintenance is carried out by the system.
• When maintenance is completed without any error, the date and time are recorded in the journal of events (in
the SYSTEM RECORDS tab).
As soon as the menu option is selected, the maintenance is carried out by the system.
The cleaning and decontamination process for the parts of the analyser protected by the lid is described in
5-2- Weekly Maintenance, page 118 and following.
The analyser bodywork can be cleaned with absorbent paper soaked in de-ionised water. If necessary, use a
neutral, phosphate free detergent suitable for medical devices, diluted as recommended by the supplier, to
clean traces on the bodywork.
The external side of the lid can be cleaned with absorbent paper soaked in de-ionised water.
Decontaminate the bodywork and the lid (internal and external sides) with a disinfectant wipe/solution suitable
for medical devices.
For the internal side of the lid:
• Open the lid and insert the lid support tool (see pages 35 and 36).
• Switch off the analyser.
• Decontaminate the internal side of the lid with a disinfectant wipe/solution suitable for medical devices.
! A slight touch on this spring releases the locking mechanism and prevents the lid closing.
In this case, switch off the analyser and push the lock axis back into its housing.
In case of hazardous material accidental spillage into or onto the analyser, decontaminate with a disinfectant
wipe/solution suitable for medical devices.
The following products cannot be used for cleaning or decontamination of the analyser:
• Strong or weak acids
• Bases
• Pure alcohol
• Solvents (ether, White Spirit, gasoline)
• Ammonia
• Chlorine
• Bleach
• Benzene
! If there is any doubt about using a particular product, contact Technical Services.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
IDS-iSYS User Manual - Revision M1
Problems & Corrective Action 6-1 149
Software version V 14
Problems & Corrective Action
6-1- Resolving Errors in Cartridge Check System (CCS) (continued)
CCS1/CCS2 Ratio outside 1 - Presence of air bubbles in the sampling fluidic circuit. Check for air bubbles
limits in the tubing circuit. Carry out a full priming of the arm.
2 - Check for the presence of bubbles in the reagents: there should be none.
3 - Check that the probe is not blocked or bent. Carry out a decontamination
of the probe (see Section 5-1-1-4, page 117, if applicable. If the probe is bent,
replace it and adjust the XYZ reference position (see Section 5-4-2, page 127).
4 - Check for air bubbles in the IDS-iSYS Triggers A and B tubing.
5 - Repeat the CCS1 & 2 tests.
6 - If the problem persists, contact IDS Service & Support personnel.
IDS-iSYS User Manual - Revision M1
Problems & Corrective Action 6-1 150
Software version V 14
SECTION 7:
System Configuration
Section 7
Provided a valid CCS reagent cartridge is present, along with sufficient levels of ancillary
! reagents, the analyser will automatically perform the relevant priming and QC profile at the
scheduled ‘wake-up’ time.
• Click on .
The rule is created with its number in the list and is immediately applied to the references Man, Woman,
Child and Other:
• Two other conditions can be defined as described above by clicking in the relevant column.
• The application can be restricted to certain patient categories by clicking in the cell of the Reference column:
Depending upon the selected option, additional tests to perform can be:
- automatically added to the relevant profile,
- added to the profile after accepting the automatic action or by using a specific button in the profile
edition.
• Tick or untick the launch option of tests according to the desired behavior.
• Exit the menu by clicking on to save the modification.
• The options of validating and transferring the results are defined in the TRANSFER/VALIDATION OF
RESULTS tab:
• Click in the entry field of the name then enter the name using the keyboard (maximal length: 20
alphanumeric characters).
• Click in the entry field of the access code then enter the password sing the keyboard (maximal length:8
alphanumeric characters).
Remember the defined password and its cases (combination of characters in capital and lower
! cases).
Modify a user
• Select the user to modified then click on .
• Enter the relevant modification(s) then click on .
Remove a user
• Select the user to remove then click on .
• The user is removed from the list.
When the analytical configurations must be updated, a "Master Database & Update Tool" CD is provided by
IDS.
The update is done from the user interface.
• Insert the CD provided by IDS in the CD drive.
• In the FILE menu, select:
• Open the file «config.acu» from the CD by indicating the relevant pathway.
• When the update starts, a window is displayed with the list of the modified analytes and with the result of
each modification:
In case of error message or failure when updating one or several analytes, contact IDS
! Technical Services.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or which has been in contact with, biological specimens must
be considered to be a potential risk of infection.
Always wear disposable gloves for all handling procedures.
Waste must be disposed of in accordance with current local regulations.
Liquid waste
Liquid waste is collected in a 10 liters container.
• Before installing the liquid waste container container, put 250 mL of commercial bleach ready-to-use (2,6%
of active Chloride) into the empty container.
Liquid waste should be considered potentially infectious and must, therefore, be disposed of in accordance
with current local regulations.
Except when an automated system of liquid waste treatment is used, liquid waste must be processed prior to
disposal in order to eliminate a biological risk. This procedure must be carried out outside of the analyser:
• When the container is full, remove the plunger and remove the container from the analyser.
• Add 500 mL of concentrated bleach (9,6% of active Chloride).
• Leave on overnight.
• Eliminate the processed waste in accordance with current local procedures.
Solid waste
• Probes.
• IDS-iSYS Cuvettes.
Solid waste should be considered potentially infectious and must, therefore, be disposed of in accordance with
current local regulations.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
The analyser must be decontaminated after carrying out the regular maintenance procedures described in
Section 5 of this user manual. It is essential that decontamination is carried out:
• Prior to any intervention by the Technical Services & Support personnel.
• Prior to any transportation of the system.
NAME OF ANALYSER:!
SERIAL NUMBER:
LABORATORY!
NAME
ADDRESS
I declare that I have carried out all stages of cleaning and disinfecting described in the user manual.
NAME
Position (optional)
Since 13/08/2005, the disposal of electrical and electronic waste has been governed by Directive 2002/96/
E.C. of 27 January 2003, revised by Directive 2012/19/UE of 4th July 2012.
This Regulation is applied in each European country. The OWNER of the equipment must contact its VENDOR
in order to obtain information about current national and local laws.
In case where the equipment is removed from the OWNER’s site, the OWNER must return the device
decontaminated, and must apply the disinfection procedure described in Appendix 2.
Furthermore, containers that have been in contact with solid waste must be disinfected. Solid and liquid waste,
as well as liquid waste tubing must be disposed of according to the current local laws.
In the event of resale to a third party, the first OWNER must notify the VENDOR of the name and address of the
new OWNER of the device in order to guarantee traceability of the equipment and for its subsequent disposal,
and must inform the new OWNER that the equipment must be disposed of according to the current local laws.
Failing this, the first OWNER will have to pay all the costs and all the fines the government may enforce upon
the SUPPLIER (the VENDOR) for breach of its obligation to ensure traceability of the disposal of its equipment
in accordance with regulations.
Manufacturer.
Keep dry.
Fragile
Single use.
Catalogue number.
Lot number.
Expiry date.
Storage temperature.
Quantity.
EAN/UPC yes no no
Recommendations
• The barcode should contain at least 5 digits for the removable rack configuration (sample tubes identified by
the barcode reader integrated into the sample compartment) or at least 4 digits (except if Interleaved 2of 5)
for the fixed tray configuration (sample tubes identified by the barcode reader in the front of the analyser).
• The barcode must not contain more than 32 digits. The character «%» must not be used .
• The height of the barcode should not be less than 10 mm
• The barcode symbol should be placed at the center of the area between 20 mm above the bottom of the
tube and 14 mm below the top of the tube, excluding the cap.
• The label should be placed with the bars perpendicular to the axis of the tube.
TRIGGER A and TRIGGER B solutions must not be kept on-board during long periods of
Description Quantity
IDS-iSYS WASH S 35 mL
As the solid waste will be full after 14 days, the automatic startup will no longer be performed.
! When the spectrophotometer is deactivated, the solid waste will be full after 3 days and the
automatic startup will no longer be performed.
Description Quantity
IDS-iSYS WASH S 24 mL
IDS-iSYS D-SORB 3 mL
• Click on the corresponding button in front of this item in the maintenance list.
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
• Click on .
• Click on the liquid waste icon.
Then validate the message confirming the liquid waste emptying.
! For systems using DSORB solution, a maintenance procedure for the four washers will take
place automatically during the SHUT DOWN procedure.
• Click on the corresponding button in front of this item in the maintenance list.
• For emptying the D-SORB fluidic circuit, remove the D-SORB level sensor from the bottle and wipe it.
• Invert the D-SORB level sensor and keep it in the air.
• Identify the position by using the barcode label of the D-SORB bottle that just has been removed from the
system.
• Request three times the priming SPECIFIC DECONTAMINATING SOLUTION, from the menu
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
• Click on .
• Click on the liquid waste icon.
Then validate the message confirming the liquid waste emptying.
• Click on the solid waste icon.
Then validate the message confirming the solid waste emptying.
• Check the current volume of each ancillary reagent is adequate with the stop period.
• From the File menu, click on EXIT, the software will close.
• Then turn off the computer following the procedure for shutting down WINDOWS.
• Finally, press the switch located on the left-hand side of the analyser into position ‘0’.
Description
IDS-iSYS Trigger A
IDS-iSYS Trigger B
IDS-iSYS SYST L
IDS-iSYS WASH S
IDS-iSYS D-SORB
• For systems using AP Substrate, install a new bottle of AP Substrate in its corresponding position.
• Once the new bottles are installed, a message proposing an automatic priming is displayed:
Wait for 40 minutes after installing CCS reagent cartridge before starting qualification profile, as
Contents 1
Contents (continued)
7- Important comments 67
The ASTM Compatible version is the original connection protocol version, existing in the IDS-iSYS software
versions prior to V10.00.
The ASTM Compatible V2 version offers additional functionalities for the transfers to the LIS:
• transfer of the traceability associated with a result,
• transfer of messages associated with a result in a Comment block,
• transfer of quality control results (if the option is selected).
The ASTM Compatible V3 version offers, in addition to the features of ASTM Compatible V2 version:
• transfer of the operator identifiers,
• transfer of the traceability associated with a result including information about AP Substrate and
pretreatment washer,
• results are resent when a clot is detected on a sample,
• transfer ofTanner stage.
This document describes the structure and construction of ASTM messages used in each version of the
connection protocol.
The ASTM message has a hierarchical structure, each block from the lower level refers to the block from the
level immediately above, except for the comments blocks.
The structure of the ASTM message is the following for this version of protocol:
Header
Patient 1
Order 1
Result 1
…….
Result n
……..
Order n
Result 1
…….
Result 2
Message Terminator
H|\^_|||IDS-iSYS B300B0378|||||||P|2|20111124135800
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
R|1|^^^25OHD|25,2|ng/mL|10,0 to 100,0|N||F||||20111124133726|IDS-iSYS
B300B0378|
L|1|N
H|\^_|||IDS-iSYS B300B0378|||||||P|2|20111124135800
This block is the first in the message, and is the one which defines its characteristics.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type H Fixed value H to indicate a Header block.
Delimiters Definition |\^_ Definitions of the delimiters that will be used in the
message. These are the 4 characters which follow
the H, with the following being initialized:
• The field delimiter ( I ) which separates the various
fields.
• The repeat delimiter ( \ ) which separates a
repetition in a field.
• The component delimiter ( ^ ) which separates
the various components of the field.
• The escape delimiter ( _ ) which indicates an
escape sequence for dealing with special
characters.
N.B. The record delimiter is fixed and is always the
character ASCII 13 (carriage return).
Message Control ID Not used.
Access Password Not used.
Sender Name or ID IDS-iSYS B300B0378 Name or identifier of the issuer:
•Not used in the direction LIS => analyser.
•Fixed at IDS-iSYS + Serial number in the direction
device to LIS.
Sender Street Address Not used.
Reserved Field
Sender Telephone Number Not used.
Characteristics of Sender Not used.
Receiver ID Not used.
Comment or Special Not used.
Instruction
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
This block represents the information relating to a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type P Fixed value P to indicate a Patient block.
Sequence Number 1 Number of the Patient block in the message starts
at 1 and is increased for each block in the
message.
Practice Assigned Patient ID Not used.
Laboratory Assigned Patient 20105863 PID.
ID
ASTM Name of the field Value* Description and use in IDS-iSYS system
Patient Telephone Number Not used.
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
This block represents the information about one or several test requests for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID O Fixed value O for indicating an Order block.
Sequence Number 1 Number of the order within the Patient block starts
at 1 and is increased for each Order block in the
message.
Container and Specimen ID 01^7148345 This field consists of 2 components:
• The type of container used for the sample
(2 digits).
• 01 PT_11_4.5 mL.
• 02 PT_13_6 mL.
• 03 ST_13_5 mL H.
• 05 Cup 500 µL.
• 06 Cup 2 mL.
• 07 PT_16_10 ml.
• 08 to 17 for configurable containers.
PT for Primary Tube; ST for Secondary Tube
For correct functioning of the apparatus, it is
important to allocate the correct type of container
used for the sample.
If this component is null, the default container type
defined for the tray will be used.
• The SID of the profile concerned by the request.
Instrument Specimen ID Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Priority R This field may contain the values:
• S = STAT.
• A = ASAP.
• R = Routine.
• C = Callback.
• P = Preoperative.
For the device, only S has any influence, the
passage of the associated profile into Urgent, and
the other values are ignored.
Requested/Ordered Not used.
Date and Time
Specimen Collection Not used.
Date and Time
Collection End Time Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Specimen Descriptor S This field consists of two sections:
• The sample type which can take for the IDS-
iSYS the following values:
• U = Urine.
• S = Serum.
• O = Other.
• The sample source (place where the sample
was taken) is not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID R Fixed value R for indicating a Result block.
Sequence Number 1 The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
Universal Test ID ^^^25OHD This ASTM field consists of 4 components:
• Universal test ID.
• Universal test ID name.
• Universal test ID type.
• Manufacturer’s code.
Only the last component is used and will be
provided with information with the unique IDS-
iSYS identifier of the analyte.
Data or Measurement Value 25,2 The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
• * * * * * when the result is below the measuring
range
• * * * * * when the result is above the measuring
range
(For 25OHD: <5 or >140)
Units ng/mL Blocks used
Reference Range 10,0 to 100,0 Normal reference values used for the patient sent
through in the form of an ASCII chain.
Result abnormal Flags N This field may take the values:
• L = Below low normal.
• H = Above high normal.
• LL = Below alarm low.
• HH = Above alarm high.
• < of absolute low (Field of measurement).
• > of absolute high (Field of measurement).
• N = Normal.
• A = Abnormal.
• U = Significant change Up.
• D = Significant change Down.
• B = Better.
• W = Worse.
Only A and N are managed by IDS-iSYS.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Nature of Abnormality Testing S This field forces S or A for the IDS-iSYS can take
the values:
• A = Age.
• S = Sex.
• R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
Result Status F This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
•C = Correction.
•P = Preliminary results.
•F = Final results.
•X = Request not honored.
•I = Results pending.
•S = Partial results.
Date of change in Instrument Not used.
Normative Values
Operator ID Not used.
Date and Time Test Started Not used.
Date and Time Test 20111124133726 Date and time of result of test on the GUI.
Completed
Instrument Identification IDS-iSYS B300B0378 Name of device + Serial number.
Q|1|^7148344||ALL||||||||O
This block represents a request for information whose origin now can only be the IDS-iSYS.
This request will be used by the IDS-iSYS after discovering a tube for a patient whose profile is
known on the work list in order to request the tests to be carried out on this tube.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID Q Fixed value Q to indicate a Request Information block.
Starting Range ID Number ^7148344 This field consists of 3 elements; only the first 2 are
used by the IDS-iSYS in order to specify the PID
(optional) and the SID (obligatory) allowing the profile
quoted to be identified.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Request Information O Potential values:
Status Code
• C
• P
• F
• X
• I
• S
• M
• R
• A
• N
• O
• D
Only the values O and A are managed in the
direction IDS-iSYS to LIS.
A is used when the LIS does not respond to a
request. In such cases, the request is cancelled
(this field takes the value A). The request is
transmitted 3 times, after which the transmission is
aborted.
L|1|N
This block is used in order to indicate the end of an ASTM block. It consists of the following fields:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID L Fixed value L for indicating a Terminator block.
Sequence Number 1 Always at 1 as there is only one sequence per
message.
Termination code N This code designates of end of the message with
the following values, depending on the situation:
• N = Normal.
• T = Abandon issuer.
• R = Abandon receiver.
• E = Unknown Error.
• Q = Error in last request.
• I = No Information available for last query.
• F = Last request for information processed. of
comment.
2-8-1-Equipment configuration
Communication with the centralized computer system is connected to the PC associated with the analyser and
can be made by:
• RS232C link. DB9 standard pinouts.
• TCP/IP link. Ethernet cable (RJ45,8 pins)
• Configurable options:
The sample container is set with the type sent in the field
«Container and Specimen ID» (see Order Block, page 9).
The user can modify this container type until the sample is
processed by the analyser (blue colour code).
The quality control results are transferred to the LIS when the option is selected (see 3-10-3- Transferring
quality control results, page 37).
The ASTM message has a hierarchical structure, each block from the lower level refers to the block from the
level immediately above, except for the comments blocks.
The structure of the ASTM message is the following for this version of protocol:
Header
Patient 1
Order 1
Result 1
Comment 1
Manufacturer 1
…….
Result n
Comment 1
Manufacturer 1
……..
Order n
Result 1
Comment 1
Manufacturer 1
…….
Result n
Comment 1
Manufacturer 1
Message Terminator
H|\^_|||IDS-iSYS B300B0378|||||||P|4|20111124135800
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
R|1|^^^25OHD^04^F^1|63,9|ng/mL|10,0 to 100,0|N||F||||20111123171733|IDS-
iSYS B300B0378
C|1|I|FOR, MRE!|N
M|1|^^^25OHD^04^F^1|51||25OHD^0952^209525602987^20120229|System
Liquid^72634^00162^20121209\Cuvettes^18121^02177^20120506|
TRIGB^70245^00049^20130203\TRIGA^70244^00200^20130203\Wash
S^72508^00322^20121104\Wash1^2\Wash2\Wash3\DSORB
L|1|N
! A blue italic font identifies the differences with Compatible ASTM in the description of the
various fields of each block.
H|\^_|||IDS-iSYS B300B0378|||||||P|4|20111124135800
This block is the first in the message, and is the one which defines its characteristics.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type H Fixed value H to indicate a Header block.
Delimiters Definition |\^_ Definitions of the delimiters that will be used in the
message. These are the 4 characters which follow
the H, with the following being initialized:
• The field delimiter ( I ) which separates the various
fields.
• The repeat delimiter ( \ ) which separates a
repetition in a field.
• The component delimiter ( ^ ) which separates
the various components of the field.
• The escape delimiter ( _ ) which indicates an
escape sequence for dealing with special
characters.
N.B. The record delimiter is fixed and is always the
character ASCII 13 (carriage return).
Message Control ID Not used.
Access Password Not used.
Sender Name or ID IDS-iSYS B300B0378 Name or identifier of the issuer:
•Not used in the direction LIS => analyser.
•Fixed at IDS-iSYS + Serial number in the direction
device to LIS.
Sender Street Address Not used.
Reserved Field
Sender Telephone Number Not used.
Characteristics of Sender Not used.
Receiver ID Not used.
Comment or Special Not used.
Instruction
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
This block represents the information relating to a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type P Fixed value P to indicate a Patient block.
Sequence Number 1 Number of the Patient block in the message starts
at 1 and is increased for each block in the
message.
Practice Assigned Patient ID Not used.
Laboratory Assigned Patient 20105863 PID.
ID
ASTM Name of the field Value* Description and use in IDS-iSYS system
Patient Telephone Number Not used.
Attending Physician ID DUPONT Name of doctor associated with profile.
Special Field 1 This field allows to define the profile type sent.
This field can take the values:
• NULL for a patient
• 02 for a control
Special Field 2 Not used.
Patient Height Not used.
Patient Weight Not used.
Patient Known or Suspected Comment Comment associated with profile.
Diagnosis
Patient Active Medication Not used.
Patient Diet Not used.
Practice Field 1 Not used.
Practice Field 2 Not used.
Admission and Discharge Not used.
Dates
Admission Status Not used.
Location Not used.
Nature of Alternative Not used.
Diagnostic Code
Alternative Diagnostic Code Not used.
Patient Religion Not used.
Marital Status Not used.
Isolation Status Not used.
Language Not used.
Hospital Service Radiology Department associated with the profile.
Hospital Institution Not used.
Dosage Category Not used.
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
This block represents the information about one or several test requests for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID O Fixed value O for indicating an Order block.
Sequence Number 1 Number of the order within the Patient block starts
at 1 and is increased for each Order block in the
message.
Container and Specimen ID 01^7148345 This field consists of 2 components:
• The type of container used for the sample
(2 digits).
• 01 PT_11_4.5 mL.
• 02 PT_13_6 mL.
• 03 ST_13_5 mL H.
• 05 Cup 500 µL.
• 06 Cup 2 mL.
• 07 PT_16_10 ml.
• 08 to 17 for configurable containers.
PT for Primary Tube; ST for Secondary Tube
For correct functioning of the apparatus, it is
important to allocate the correct type of container
used for the sample.
If this component is null, the default container type
defined for the tray will be used.
• The SID of the profile concerned by the request.
Instrument Specimen ID Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Priority R This field may contain the values:
• S = STAT.
• A = ASAP.
• R = Routine.
• C = Callback.
• P = Preoperative.
For the device, only S has any influence, the
passage of the associated profile into Urgent, and
the other values are ignored.
Requested/Ordered Not used.
Date and Time
Specimen Collection Not used.
Date and Time
Collection End Time Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Specimen Descriptor S This field consists of two sections:
• The sample type which can take for the IDS-
iSYS the following values:
• U = Urine.
• S = Serum.
• O = Other.
• The sample source (place where the sample
was taken) is not used.
R|1|^^^25OHD^04^F^1|63,9|ng/mL|10,0 to 100,0|N||F||||20111123171733|IDS-iSYS
B300B0378
This block represents the information sent through by the IDS-iSYS to the LIS after a test has been carried out
for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID R Fixed value R for indicating a Result block.
Sequence Number 1 The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
Universal Test ID ^^^25OHD^04^F^1 This ASTM field consists of 8 components:
- Universal test Id : forces NULL
- Universal test Id Name: forces NULL.
- Universal test Id Type: forces NULL.
-Manufacturer’s Code: contains the unique
IDS-iSYS identifier for the analyte.
-Test type: 02 for a control, 04 for a patient
-Result type: F for final, P for partial.
-Replicate number: 1, 2,up to 10
-Control identifier, defined as follow:
• 21 for a control Level 1
• 22 for a control Level 2
• 23 for a control Level 3
• 24 for a control Level 4
For a patient,this component is NULL.
-Dilution rate: NULL if not applicable.
Data or Measurement Value 63,9 The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
• < Low limit of the sample measurement range,
when the result is below the measuring range
• > High limit of the sample measurement range,
when the result is above the measuring range
(For 25OHD: <5 or >140)
ASTM Name of the field Value* Description and use in IDS-iSYS system
Result abnormal Flags N This field may take the values:
• L = Below low normal.
• H = Above high normal.
• LL = Below alarm low.
• HH = Above alarm high.
• < of absolute low (Field of measurement).
• > of absolute high (Field of measurement).
• N = Normal.
• A = Abnormal.
• U = Significant change Up.
• D = Significant change Down.
• B = Better.
• W = Worse.
L, H, <, >, A and N are managed by IDS-iSYS.
All associated messages are sent in a Comment
block.
Nature of Abnormality Testing S This field forces S or A for the IDS-iSYS can take
the values:
• A = Age.
• S = Sex.
• R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
Result Status F This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
•C = Correction.
•P = Preliminary results.
•F = Final results.
•X = Request not honored.
•I = Results pending.
•S = Partial results.
Date of change in Instrument Not used.
Normative Values
Operator ID Not used.
Date and Time Test Started Not used.
Date and Time Test 20111123171733 Date and time of result of test on the GUI.
Completed
Instrument Identification IDS-iSYS B300B0378 Name of device + Serial number.
C|1|I|FOR, MRE!|N
This block represents a comment.
This type of block can be inserted after a Result Block. Depending on configuration, several Comment blocks
can be sent.
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID C Fixed value C for indicating a Comment block.
Sequence Number 1 Number of the comment begins at 1 and is
increased for each comment block in the
message.
Comment Source I Issuer can take the following values:
• P = Practice.
• L = computer system in the direction LIS.
• I = Instrument.
Comment Text FOR, MRE! If the field Comment Type takes the value N, this
field contains the message associated with a
result.
Comment Type N Type of comment. This field can take the values:
• G = Generic comment (any permitted).
• T = Test name comment.
• P = Positive test comment.
• N = Negative test comment.
• I = Instrument flags comment.
M|1|^^^25OHD^04^F^1|51||25OHD^0952^209525602987^20120229|System
Liquid^72634^00162^20121209\Cuvettes^18121^02177^20120506|
TRIGB^70245^00049^20130203\TRIGA^70244^00200^20130203\Wash
S^72508^00322^20121104\Wash1^2\Wash2\Wash3\DSORB
This block contains traceability information sent by IDS-iSYS to the LIS, and follows a Result block.
Here are the various fields of this block used in the protocol version:
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID M Fixed value M for indicating a Manufacturer
Information block.
Sequence Number 1 Number of the block associated with a result,
begins at 1 and is increased for each
Manufacturer Information block in the message.
Universal Test ID ^^^25OHD^04^F^1 • Universal test ID: NULL.
• Universal test ID name: NULL.
• Universal test ID type: NULL
• Manufacturer’s code
• Test type: 02 for a control, 04 for a patient.
• Result type: always F for final.
Depending on the test, a final result can be a
mean value.
• Replicate number: 1, 2, up to 10
• 52 for Biochemistry.
Component information This field will be used only for control transfers.
For a patient, this field is NULL.
For a control, this field consists of:
• target value,
• low value of acceptable range,
• high value of acceptable range
• lot number
• bottle number,
• expiry date.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Reagent traceability 25OHD^0952^209525602 This field contains infomation for each reagent
987^20120229 used in the assay:
Reagent 1^Lot^Bottle Number ^Expiry date\
Reagent 2^Lot^Bottle Number ^Expiry date\…….
Q|1|^7148344||ALL||||||||O
This block represents a request for information whose origin now can only be the IDS-iSYS.
This request will be used by the IDS-iSYS after discovering a tube for a patient whose profile is
known on the work list in order to request the tests to be carried out on this tube.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID Q Fixed value Q to indicate a Request Information block.
Starting Range ID Number ^7148344 This field consists of 3 elements; only the first 2 are
used by the IDS-iSYS in order to specify the PID
(optional) and the SID (obligatory) allowing the profile
quoted to be identified.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Request Information O Potential values:
Status Code
• C
• P
• F
• X
• I
• S
• M
• R
• A
• N
• O
• D
Only the values O and X are managed in the
direction IDS-iSYS to LIS.
A is used when the LIS does not respond to a
request. In such cases, the request is cancelled
(this field takes the value A). The request is
transmitted 3 times, after which the transmission is
aborted.
L|1|N
This block is used in order to indicate the end of an ASTM block. It consists of the following fields:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID L Fixed value L for indicating a Terminator block.
Sequence Number 1 Always at 1 as there is only one sequence per
message.
Termination code N This code designates of end of the message with
the following values, depending on the situation:
• N = Normal.
• T = Abandon issuer.
• R = Abandon receiver.
• E = Unknown Error.
• Q = Error in last request.
• I = No Information available for last query.
• F = Last request for information processed. of
comment.
3-10-1-Equipment configuration
Communication with the centralized computer system is connected to the PC associated with the analyser and
can be made by:
• RS232C link. DB9 standard pinouts.
• TCP/IP link. Ethernet cable (RJ45,8 pins)
• Configurable options:
The sample container is set with the type sent in the field
«Container and Specimen ID» (see Order Block, page 24).
The user can modify this container type until the sample is
processed by the analyser (blue colour code).
The quality control results are transferred to the LIS when the option is selected (see 4-10-3- Transferring
quality control results, page 56).
The ASTM message has a hierarchical structure, each block from the lower level refers to the block from the
level immediately above, except for the comments blocks.
The structure of the ASTM message is the following for this version of protocol:
Header
Patient 1
Order 1
Result 1
Comment 1
Manufacturer 1
…….
Result n
Comment 1
Manufacturer 1
……..
Order n
Result 1
Comment 1
Manufacturer 1
…….
Result n
Comment 1
Manufacturer 1
Message Terminator
H|\^_|||IDS-iSYS B300B0378|||||||P|5|20111124135800
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
R|1|^^^25OHD^04^F^1|63,9|ng/mL|10,0 to 100,0|N||F||||20111123171733|IDS-
iSYS B300B0378
C|1|I|FOR, MRE!|N
M|1|^^^25OHD^04^F^1|51||25OHD^0952^209525602987^20120229|System
Liquid^72634^00162^20121209\Cuvettes^18121^02177^20120506|
TRIGB^70245^00049^20130203\TRIGA^70244^00200^20130203\Wash
S^72508^00322^20121104\Wash1^2\Wash2\Wash3\DSORB
L|1|N
! A blue italic font identifies the differences with Compatible ASTM V2 in the description of the
various fields of each block.
H|\^_|||IDS-iSYS B300B0378|||||||P|5|20111124135800
This block is the first in the message, and is the one which defines its characteristics.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type H Fixed value H to indicate a Header block.
Delimiters Definition |\^_ Definitions of the delimiters that will be used in the
message. These are the 4 characters which follow
the H, with the following being initialized:
• The field delimiter ( I ) which separates the various
fields.
• The repeat delimiter ( \ ) which separates a
repetition in a field.
• The component delimiter ( ^ ) which separates
the various components of the field.
• The escape delimiter ( _ ) which indicates an
escape sequence for dealing with special
characters.
N.B. The record delimiter is fixed and is always the
character ASCII 13 (carriage return).
Message Control ID Not used.
Access Password Not used.
Sender Name or ID IDS-iSYS B300B0378 Name or identifier of the issuer:
•Not used in the direction LIS => analyser.
•Fixed at IDS-iSYS + Serial number in the direction
device to LIS.
Sender Street Address Not used.
Reserved Field
Sender Telephone Number Not used.
Characteristics of Sender Not used.
Receiver ID Not used.
Comment or Special Not used.
Instruction
P|1||20105863||BONTEMPS^Robert||19500226|M|||||DUPONT|||||Comment||||||||||||||
Radiology||TANNER STAGE^5
This block represents the information relating to a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type P Fixed value P to indicate a Patient block.
Sequence Number 1 Number of the Patient block in the message starts
at 1 and is increased for each block in the
message.
Practice Assigned Patient ID Not used.
Laboratory Assigned Patient 20105863 PID.
ID
ASTM Name of the field Value* Description and use in IDS-iSYS system
Patient Telephone Number Not used.
Attending Physician ID DUPONT Name of doctor associated with profile.
Special Field 1 This field allows to define the profile type sent.
This field can take the values:
• NULL for a patient
• 02 for a control
Special Field 2 Not used.
Patient Height Not used.
Patient Weight Not used.
Patient Known or Suspected Comment Comment associated with profile.
Diagnosis
Patient Active Medication Not used.
Patient Diet Not used.
Practice Field 1 Not used.
Practice Field 2 Not used.
Admission and Discharge Not used.
Dates
Admission Status Not used.
Location Not used.
Nature of Alternative Not used.
Diagnostic Code
Alternative Diagnostic Code Not used.
Patient Religion Not used.
Marital Status Not used.
Isolation Status Not used.
Language Not used.
Hospital Service Radiology Department associated with the profile.
Hospital Institution Not used.
Dosage Category TANNER STAGE^5 Allows defining the Tanner stage.
The second component of the field can take the
values from 1 to 5 (integral values).
O|1|01^7148345||^^^25OHD|R||||||A||||S||||||||||O
This block represents the information about one or several test requests for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID O Fixed value O for indicating an Order block.
Sequence Number 1 Number of the order within the Patient block starts
at 1 and is increased for each Order block in the
message.
Container and Specimen ID 01^7148345 This field consists of 2 components:
• The type of container used for the sample
(2 digits).
• 01 PT_11_4.5 mL.
• 02 PT_13_6 mL.
• 03 ST_13_5 mL H.
• 05 Cup 500 µL.
• 06 Cup 2 mL.
• 07 PT_16_10 ml.
• 08 to 17 for configurable containers.
PT for Primary Tube; ST for Secondary Tube
For correct functioning of the apparatus, it is
important to allocate the correct type of container
used for the sample.
If this component is null, the default container type
defined for the tray will be used.
• The SID of the profile concerned by the request.
Instrument Specimen ID Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Priority R This field may contain the values:
• S = STAT.
• A = ASAP.
• R = Routine.
• C = Callback.
• P = Preoperative.
For the device, only S has any influence, the
passage of the associated profile into Urgent, and
the other values are ignored.
Requested/Ordered Not used.
Date and Time
Specimen Collection Not used.
Date and Time
Collection End Time Not used.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Specimen Descriptor S This field consists of two sections:
• The sample type which can take for the IDS-
iSYS the following values:
• U = Urine.
• S = Serum.
• O = Other.
• The sample source (place where the sample
was taken) is not used.
R|1|^^^25OHD^04^F^1|63,9|ng/mL|10,0 to 100,0|N||F|||SYSTEM^SYSTEM|
20111123171733|IDS-iSYS B300B0378
This block represents the information sent through by the IDS-iSYS to the LIS after a test has been carried out
for a patient.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID R Fixed value R for indicating a Result block.
Sequence Number 1 The number of the result within the Patient block
begins at 1 and is increased for each Result block
in the message.
Universal Test ID ^^^25OHD^04^F^1 This ASTM field consists of 8 components:
- Universal test Id : forces NULL
- Universal test Id Name: forces NULL.
- Universal test Id Type: forces NULL.
-Manufacturer’s Code: contains the unique
IDS-iSYS identifier for the analyte.
-Test type: 02 for a control, 04 for a patient
-Result type: F for final, P for partial.
-Replicate number: 1, 2,up to 10
-Control identifier, defined as follow:
• 21 for a control Level 1
• 22 for a control Level 2
• 23 for a control Level 3
• 24 for a control Level 4
For a patient,this component is NULL.
-Dilution rate: NULL if not applicable.
Data or Measurement Value 63,9 The value in the form of ASCII chain.
Results which cannot be calculated are sent in the
format * * * * * . Each character * is separated by a
space.If the value is outside the measuring range,
is transmitted in this field:
• < Low limit of the sample measurement range,
when the result is below the measuring range
• > High limit of the sample measurement range,
when the result is above the measuring range
(For 25OHD: <5 or >140)
ASTM Name of the field Value* Description and use in IDS-iSYS system
Result abnormal Flags N This field may take the values:
• L = Below low normal.
• H = Above high normal.
• LL = Below alarm low.
• HH = Above alarm high.
• < of absolute low (Field of measurement).
• > of absolute high (Field of measurement).
• N = Normal.
• A = Abnormal.
• U = Significant change Up.
• D = Significant change Down.
• B = Better.
• W = Worse.
L, H, <, >, A and N are managed by IDS-iSYS.
All associated messages are sent in a Comment
block.
Nature of Abnormality Testing S This field forces S or A for the IDS-iSYS can take
the values:
• A = Age.
• S = Sex.
• R = Race.
And is able to provide an indication about the
bases on which the values of normality were
determined.
Result Status F This field forces to F or P (replicate) on the IDS-
iSYS can take the following values:
•C = Correction.
•P = Preliminary results.
•F = Final results.
•X = Request not honored.
•I = Results pending.
•S = Partial results.
Date of change in Instrument Not used.
Normative Values
Operator ID SYSTEM^SYSTEM Identifying the operators who handled the profile.
This field consists of 2 components:
- Identifier of the operator who creates the
profile; if the profile was transferred by the LIS,
the identifier SYSTEM is sent.
- Identifier of the operator who stores/tranfers
the profile; if the profile was automatically
stored/transferred, the identifier SYSTEM is
sent.
Date and Time Test Started Not used.
Date and Time Test 20111123171733 Date and time of result of test on the GUI.
Completed
Instrument Identification IDS-iSYS B300B0378 Name of device + Serial number.
C|1|I|FOR, MRE!|N
This block represents a comment.
This type of block can be inserted after a Result Block. Depending on configuration, several Comment blocks
can be sent.
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID C Fixed value C for indicating a Comment block.
Sequence Number 1 Number of the comment begins at 1 and is
increased for each comment block in the
message.
Comment Source I Issuer can take the following values:
• P = Practice.
• L = computer system in the direction LIS.
• I = Instrument.
Comment Text FOR, MRE! If the field Comment Type takes the value N, this
field contains the message associated with a
result.
Comment Type N Type of comment. This field can take the values:
• G = Generic comment (any permitted).
• T = Test name comment.
• P = Positive test comment.
• N = Negative test comment.
• I = Instrument flags comment.
M|1|^^^25OHD^04^F^1|51||25OHD^0952^209525602987^20120229|System
Liquid^72634^00162^20121209\Cuvettes^18121^02177^20120506|
TRIGB^70245^00049^20130203\TRIGA^70244^00200^20130203\Wash
S^72508^00322^20121104\Wash1^2\Wash2\Wash3\DSORB
This block contains traceability information sent by IDS-iSYS to the LIS, and follows a Result block.
Here are the various fields of this block used in the protocol version:
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID M Fixed value M for indicating a Manufacturer
Information block.
Sequence Number 1 Number of the block associated with a result,
begins at 1 and is increased for each
Manufacturer Information block in the message.
Universal Test ID ^^^25OHD^04^F^1 • Universal test ID: NULL.
• Universal test ID name: NULL.
• Universal test ID type: NULL
• Manufacturer’s code
• Test type: 02 for a control, 04 for a patient.
• Result type: always F for final.
Depending on the test, a final result can be a
mean value.
• Replicate number: 1, 2, up to 10
• Control identifier, defined as follow:
• 21 for a control Level 1
• 22 for a control Level 2
• 23 for a control Level 3
• 24 for a control Level 4
For a patient, this component is NULL.
• Dilution rate: NULL if not applicable.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Reagent traceability 25OHD^0952^209525602 This field contains infomation for each reagent
987^20120229 used in the assay:
Reagent 1^Lot^Bottle Number ^Expiry date\
Reagent 2^Lot^Bottle Number ^Expiry date\…….
Q|1|^7148344||ALL||||||||O
This block represents a request for information whose origin now can only be the IDS-iSYS.
This request will be used by the IDS-iSYS after discovering a tube for a patient whose profile is
known on the work list in order to request the tests to be carried out on this tube.
Here are the various fields of this block used in the protocol version:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID Q Fixed value Q to indicate a Request Information block.
Starting Range ID Number ^7148344 This field consists of 3 elements; only the first 2 are
used by the IDS-iSYS in order to specify the PID
(optional) and the SID (obligatory) allowing the profile
quoted to be identified.
ASTM Name of the field Value* Description and use in IDS-iSYS system
Request Information O Potential values:
Status Code
• C
• P
• F
• X
• I
• S
• M
• R
• A
• N
• O
• D
Only the values O and X are managed in the
direction IDS-iSYS to LIS.
A is used when the LIS does not respond to a
request. In such cases, the request is cancelled
(this field takes the value A). The request is
transmitted 3 times, after which the transmission is
aborted.
L|1|N
This block is used in order to indicate the end of an ASTM block. It consists of the following fields:
* The field value corresponds to the above block
ASTM Name of the field Value* Description and use in IDS-iSYS system
Record Type ID L Fixed value L for indicating a Terminator block.
Sequence Number 1 Always at 1 as there is only one sequence per
message.
Termination code N This code designates of end of the message with
the following values, depending on the situation:
• N = Normal.
• T = Abandon issuer.
• R = Abandon receiver.
• E = Unknown Error.
• Q = Error in last request.
• I = No Information available for last query.
• F = Last request for information processed. of
comment.
4-10-1-Equipment configuration
Communication with the centralized computer system is connected to the PC associated with the analyser and
can be made by:
• RS232C link. DB9 standard pinouts.
• TCP/IP link. Ethernet cable (RJ45,8 pins)
• Configurable options:
The sample container is set with the type sent in the field
«Container and Specimen ID» (see Order Block, page 43).
The user can modify this container type until the sample is
processed by the analyser (blue colour code).
! Data sent by the LIS to IDS-iSYS are the same for the protocol versions ASTM Compatible
and ASTM Compatible V2.
• Profiles can be sent by the LIS by downloading of the work list and/or in answer to a request sent by IDS-
iSYS (the two modes are managed interchangeably).
• When a new SID is detected on IDS-iSYS (barcode reading or keyboard entry), IDS-iSYS examines the work
list first of all.
• If the associated profile is not in the work list, IDS-iSYS sends a request to the LIS. The LIS must send a
message in answer to each request.
• If the LIS does not answer to a request, IDS-iSYS will not send or accept a new message.
• When no profile is associated to a request sent by IDS-iSYS, the LIS must send a message containing a Test
Order block with the Report Type field fixed at the value Z.
• In answer to a request, the LIS must send a message containing a Test Order block with the profile and with
the Report Type field imperatively fixed at the value Q.
• The «Specimen type» in the field «Specimen Descriptor» (see Order Block, page 9) must be sent with the
value corresponding to the sample type: S for Serum, U for Urine (in this case, verify that the assay protocol
is defined for this sample type). The type «Other» (value =O) must not be used for Immunoassays.
• In the case of a field containing multiple components, if the last component(s) takes a NULL value, the last
component delimiter(s) can be not sent.
• The exchanges between IDS-iSYS and LIS can be recorded by selecting the option available in the tab
under SET-UP/SETTINGS/SYSTEM menu:
The transfers between both systems will be recorded in a log file, in the interface software folder (folder GUI).
The log must be activated only for a sort period of time, to solve transfer problems.
Activating the log permanently may cause interface software dysfunctions.
• The barcode should contain at least 5 digits for the removable rack configuration (sample tubes identified by
the barcode reader integrated into the sample compartment) or at least 4 digits (except if Interleaved 2/5) for
the fixed tray configuration (sample tubes identified by the barcode reader in the front of the analyser).
• If data sent by the LIS cannot be processed, the message is accepted but the corresponding profile will not
be created in the worklist.
• For system equipped with the option «Clot detection», when a clot is detected in a sample, all the results
previously transferred to the LIS will be resent with a field Report Type set at the value C (correction) in the
Test Order block.
In the Result block, the field Result abnormal Flags is then forced to A.
With the versions ASTM Compatible V2 and ASTM Compatible V3, a message FIB is sent in the Comment
block.
7- Important notes
IDS-iSYS Connection Protocol - Revision N1
67
Software version V 14