Lightwave II User Manual Version 2 - 3
Lightwave II User Manual Version 2 - 3
USER MANUAL
Declaration of Conformity
+
This is to certify that the Lightwave II and Lightwave II UV/Visible Spectrophotometers part numbers:
80-3003-72 / -73 / -74 & 80-3004-60 / -61 / -62
manufactured by Biochrom Ltd. conform to the requirements of the following Directives-: 73/23/EEC &
89/336/EEC & IVD
EN 61010-1: 2001 Safety requirements for electrical equipment for measurement, control and laboratory
use.
EN 61326-2.3: 1998 Electromagnetic compatibility - generic emission standard Electrical equipment for
measurement, control and laboratory use.
BS EN 591:2001 Instruction for use for in vitro diagnostic instruments for professional use.
2002/96/EC This appliance is marked according to the European directive 2002/96/EC on Waste Electrical
and Electronic Equipment (WEEE). By ensuring this product is disposed of correctly, you will help prevent
potential negative consequences for the environment and human health, which could otherwise be caused
by inappropriate waste handling of this product.
The symbol on the product, or on the documents accompanying the product, indicates that this
appliance may not be treated as household waste. Instead it shall be handed over to the applicable
collection point for the recycling of electrical and electronic equipment. Disposal must be carried out in
accordance with local environmental regulations for waste disposal.
Signed:
David Parr
Managing Director
Biochrom Ltd
Biochrom Ltd
22 Cambridge Science Park
Milton Road
Cambridge CB4 0FJ
England
Telephone +44 (0)1223 423723
Telefax +44 (0)1223 420164
E mail: [email protected]
Website: www.biochrom.co.uk
Registered in England No: 3526954
INTRODUCTION.......................................................................................................................................................... 5
Your spectrophotometer .............................................................................................................................. 5
Sample handling tips .................................................................................................................................... 5
Keypad and display ...................................................................................................................................... 6
Software style................................................................................................................................................ 7
ACCESSORIES ......................................................................................................................................................... 38
After Sales Support .................................................................................................................................... 38
Lamp Replacement ..................................................................................................................................... 38
Cleaning and general care of the instrument........................................................................................... 38
There are a number of warning labels and symbols on your instrument. These are there to inform you where
potential danger exists or particular caution is required. Before commencing installation, please take time to
familiarise yourself with these symbols and their meaning.
• Check the contents of the pack against the packing list. If any shortages are discovered, inform your supplier
immediately.
• Inspect the instrument for any signs of damage caused in transit. If any damage is discovered, inform your
supplier immediately.
• Ensure your proposed installation site conforms to the environmental conditions for safe operation:
Indoor use only.
Temperature range 5°C to 35°C. Note that if you use the instrument in a room subjected to extremes of
temperature change during the day, it may be necessary to recalibrate (by switching off and then on again)
once thermal equilibrium has been established (2-3 hours).
Maximum relative humidity of 80% up to 31°C decreasing linearly to 50% at 40°C
• The instrument must be placed on a stable, level bench or table that can take its weight (< 4.5 kg) so that air
can circulate freely around the instrument.
• This equipment must be connected to the power supply with the power cord supplied. It can be used on 90 –
240 V, 50-60 Hz supplies.
• If the instrument has just been unpacked or has been stored in a cold environment, it should be allowed to
come to thermal equilibrium for 2-3 hours in the laboratory before switching. This will prevent calibration failure
as a result of internal condensation.
• Switch on the instrument via the keypad ( ) after it has been plugged in. The instrument will perform a
series of self-diagnostic checks.
• Please read through this user manual prior to use.
• Please contact your original supplier in the first instance if you experience technical or sample handling
difficulties.
If this equipment is used in a manner not specified or in environmental conditions not appropriate for safe
operation, the protection provided by the equipment may be impaired and instrument warranty withdrawn.
Your spectrophotometer
Your spectrophotometer is a simple-to-use UV/Visible instrument with a CCD array detector (1024 pixels). It has no
moving parts, which is the basis of the rapid scanning operating system. The look and operation of the Lightwave II
+
and II are identical; the only difference between them is the bandwidth. Throughout the rest of this manual the
term Lightwave II will be used to cover both instruments.
The user interface is built around folders which are displayed on the home page when the instrument is switched
on. After switch on and calibration, the default home page is "Lightwave II" offering the choice of
The instrument is supplied with a program PVC (Print via Computer) on the accompanying CD. When used with a
USB cable to connect to a PC onto which the software has been installed, it enables the user to “print through” the
PC directly to the printer that is connected to it. The data may also be stored as an Excel spreadsheet, as an EMF
graphics file, a comma delimited (csv) data file, a tab delimited (txt) data file or in native PVC format for later access
Alternatively, results may be sent to the PC via a Bluetooth accessory; this can either be supplied pre-installed or is
available as an optional accessory if the need for its use arises after installation of the product. PVC works in a
similar way.
A printer is available for the instrument; this may either be supplied pre-installed or is available as an optional
accessory if the need for its use arises after installation of the product.
• Note that the light beam is directed from RIGHT to LEFT through the cell chamber; therefore please ensure the
cell is inserted in the correct alignment.
• The cell holder supplied with the instrument accepts standard 10 - 40mm pathlength quartz, glass or plastic
cells.
• The optical height is 15 mm, and the minimum volume that can be used is approx. 70µl in a micro cell.
• 12 mm test tubes may be used (e.g. for cell cultures), however they are not recommended as higher quality
data is produced by using disposable cuvettes for the analysis. If used, align the indicator line on 12 mm test
tubes in the same direction to ensure reproducible positioning of the tube. Note that test tubes do not last
forever, and that the surface becomes scratched and blemished through repetitive use; if this is the case they
should be replaced.
The back-lit liquid crystal display is very easy to navigate around using the alphanumeric entry and navigation
arrow keys on the hard wearing, spill proof membrane keypad.
On/off key
Alphanumeric keys
Set reference
View options
Enter selection/take measurement
Key Action
On/off key Turns the instrument on/off
Arrow keys Use the four arrow keys to navigate around the display and select the required
setting from the active (highlighted) option.
View Options: ::; View options for that application mode. Some of these are common to all
applications and described below. Options unique to an application are
described in the relevant section.
Alphanumeric keys Use these to enter parameters and to write text descriptions where
appropriate, or required. Use repeated key presses to cycle through lower
case, number and upper case. Leave for 1 second before entering next
character. Use C button to backspace and 1 to enter a space.
Escape/Cancel: Escape from a selection and return to the previous folder. Stop making
measurements.
Set Reference: 0A/100%T Set reference to 0.000 A or 100%T on a reference solution at the current
wavelength in the mode selected. When in scan mode, do a reference scan.
Software style
The user interface is built around having folders of files which are displayed on the home page when the instrument
is switched on. Different folders are numbered and opened by using the associated number key on the keypad.
Summary
Function Keypad number Description
SUMMARY:
OPTIONS
Within each application the user has the possibility to select various options that define the way results are
treated. If not using a stored method, it is advisable to check that these Options have been appropriately set
for your experiment when coming to the instrument. Note that setting the “History” parameter to on (see
Preferences later) will cause the instrument to store it’s last settings. If the “History” parameter is turned off, all
parameters and options will return to their default settings when you leave that application. (Unless it has
been saved as a method).
This makes simple absorbance (A) and % transmission (%T) measurements on samples, measuring the amount of
light that has passed through a sample relative to a reference (this can be air).
The procedure is as follows:
Step 1
Set wavelength by using keypad numbers or left and right
arrows.
Press the down arrow key.
Step 2
Select the mode, Absorbance or %T, using the left and right
arrows.
Step 3
To enter the results screen with the selected parameters press
OK
OR
Cancel the selections and return to the Applications Folder by
pressing Cancel .
Step 4
Insert the reference. Press 0A/100% key. This will be used for all
subsequent samples until changed.
Step 5
Insert sample and press .
Results
The result at the selected wavelength is displayed on screen.
Use the left and right arrows to move the cursor and display the
value at the cursor position (+/- 15nm from set wavelength).
This makes simple concentration measurements on samples, by measuring the amount of light that has passed
through a sample relative to a reference (this can be air). Concentration is obtained by multiplying the measured
absorbance at a specific wavelength by a factor. The factor may be known in advance, or may be calculated by the
instrument by measuring a standard of known concentration.
Step 1
Set wavelength by using keypad numbers or left and right
arrows.
Press the down arrow key.
Step 2
Select the mode, Factor (user entered) or Standard (factor is
calculated from a calibration sample), using the left and right
arrows.
Press the down arrow key.
Step 3 (if Factor is selected)
Enter the Factor using the keypad numbers. Range 0.001 to
9999. Use the C button to delete the last digit entered.
Press the down arrow key.
Step 3 (if Standard is selected)
Enter the concentration using keypad numbers. Range 0.01-
9999. Use the C button to delete the last digit entered.
Press the down arrow key.
Step 4
Units: The user can enter a text string up to 8 characters long.
To access a list of pre-defined units press the Options key ::;
and then use the left/right arrows (µg/ml, µg/µl, pmol/µl, mg/dl,
mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or none).
These units can also be edited once OK is pressed.
This screen also allows the number of displayed decimal points
(DP) to be selected, from 0 to 2 Note that the result will always
be fixed to 5 significant figures regardless of how many decimal
points are selected (so 98768.2 will display as 98768 even with 1
decimal point selected). Press OK to store the chosen
parameters or Cancel .
Step 5
To enter the results screen with the selected parameters press
OK
OR
Cancel the selections and return to the Applications Folder by
pressing Cancel .
Step 6 (if using a Factor)
Insert the reference. Press 0A/100% key. This will be used for all
subsequent samples until changed.
Step 7
Insert sample and press .
Step 7
Insert the sample and press .
The concentration of the sample is displayed. Results shown as
---- indicate the concentration is out of range.
An absorption spectrum can be obtained from your instrument, enabling simple identification of peak height and
position. The procedure is as follows:
Step 1
Set start wavelength by using keypad numbers or left and right
arrows.
Press the down arrow key.
Step 2
Set end wavelength by using keypad numbers or left and right
arrows.
Press the down arrow key.
Step 3
Select the mode, Absorbance or %T, using the left and right
arrows.
Step 4
To enter the measurements screen with the selected parameters
press OK
OR
Cancel the selections and return to the Applications Folder by
pressing Cancel .
Step 5
Insert the reference. Press 0A/100% key. This will be used for all
subsequent samples until changed.
Step 6
Insert sample and press .
Results
A graph of the wavescan is displayed, along with a table of
Absorbance/%T at each peak. Use the left and right arrows to
move the cursor along the graph. When it reaches a peak the
peak height and width of the peak is displayed at the top of the
screen.
To zoom in on the wavelength scale, use the up arrow. This
auto-scales on the Absorbance/%T scale (dependent on the
Graph Scale option) and this is retained for subsequent
measurements.
To zoom out again, use the down arrow.
Note Storing a method at this stage will save these user defined
wavelengths, each time method is run Absorbance value at
these wavelengths is reported
Graph Scale…
This enables the user to set up a defined graph by defining the
limits in either or both of the x and y axes.
Zoom mode:
This sets up the operation of the Zoom keys (up and down
arrows). “x & y axes” expands the display around the cursor
measurement point, whilst the other options select the
absorbance or wavelength axes respectively. With x or y axis
limits set to on, zooming out will only be permitted to the set
limits.
Kinetics studies, where the change in absorbance needs to be followed as a function of time at a fixed wavelength,
can be readily performed.
Reagent test kits are routinely used for the enzymatic determination of compounds in food, beverage and clinical
laboratories by measuring NAD / NADH conversion at 340 nm. The change in absorbance over a specified time
period can be used to provide useful information when an appropriate factor, defined in the reagent kit protocol, is
applied. Reaction rate and enzyme activity can be calculated if the factor used takes account of the absorbance
difference per unit time, as opposed to the absorbance difference per se.
For this reason, the change in absorbance per minute (∆A/min), concentration (∆A/min x factor) and correlation
coefficient (calculated from a best fit of the data points) are displayed. They may not be relevant for simple
kinetics experiments.
Step 8
Set the Factor by which the result is multiplied to give the
amount in the chosen range using the left and right arrows.
Range of 0.01 to 9999.
Step 9
Press Next to enter the Results screen
OR
Press Cancel to return to the Parameters 1 screen.
Results
Insert the reference and press the 0A/100%T key.
Use the left and right arrows to move the cursor and display the
time and absorbance value at measured data points.
The construction of a multi-point calibration curve from standards of known concentration to quantify unknown
samples is a fundamental use of a spectrophotometer; this instrument has the advantage of being able to store this
curve as a method, using up to 9 standards.
To include a zero concentration standard, include this in the number of standards to be entered and enter 0.00 for
concentration; use a reagent blank when required to enter the zero standard.
Step 1
Select the wavelength using the keypad numbers or left and right
arrows.
Press the down arrow.
Step 2
Enter the number of standard concentration points to be used in
the curve (1-9).
Press the down arrow.
Step 3
Units: The user can enter a text string up to 8 characters long.
To access a list of pre-defined units press the Options key ::;
and then use the left/right arrows (µg/ml, µg/µl, pmol/µl, mg/dl,
mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or none).
These units can also be edited once OK is pressed.
This screen also allows the number of displayed decimal points
(DP) to be selected, from 0 to 2 Note that the result will always
be fixed to 5 significant figures regardless of how many decimal
points are selected (so 98768.2 will display as 98768 even with 1
decimal point selected). Press OK to store the chosen
parameters or Cancel .
Step 4
Select the type of curve fit using the left and right arrows.
Options: straight line regression, a zero regression (this forces
the straight line through the origin), interpolated or cubic spline.
Step 5
Select the calibration mode: either Standards (measure prepared
standards) or Manual (keypad data entry).
Press the down arrow.
Step 6 (if standards has been selected in step 5)
Select the number of standards to be measured and averaged at
each standard concentration point. Can be OFF (1), 2 or 3.
Step 7
Press Next to enter the Standards screen
OR
Press Cancel to cancel selections and return to the
Applications Folder.
A graph will display the results and the fitted curve as the
measurements are input.
Step 12
Press OK to accept the calibration and go to the Results
screen (see below)
OR
Press Back to return to the Standards screen.
A graph will display the results and the fitted curve as the
measurements are input.
Use the up and down arrows to select a standard to be repeated
if a poor reading has been obtained. Use C to clear the previous
reading.
Step 12
Press to accept the calibration and go to the Results screen
(see below)
OR
Press Back to return to the Standards screen.
Results screen
Step 13
Insert the reference and press the 0A/100%T key. This will be
used for all subsequent samples until changed.
Step 14
Insert the sample and press .
The concentration of the sample is taken and displayed.
Repeat step 14 for all samples.
Press to return to the Applications Folder.
This makes up to 5 absorbance measurements on the same sample. The procedure is as follows:
Step 1
Select the number of wavelengths.
Press the down arrow.
Step 2
Enter the first wavelength using either the number keys or the
left and right arrows.
Press the down arrow.
Enter the second wavelength as above and repeat for the
number of wavelengths selected (up to 5).
Step 3
Press OK to enter the results screen
OR
Press Cancel to return to the Applications Folder.
Step 4
Insert the reference. Press 0A/100% key. This will be used for all
subsequent samples until changed.
Step 5
Insert sample and press .
Results
A scan plot covering the range of wavelengths selected (with
cursors at the relevant wavelengths) and a table of values is
displayed.
This makes simple absorbance ratio measurements on samples, measuring the amount of light that has passed
through a sample relative to a blank (this can be air) at two wavelengths. The procedure is as follows:
Step 1
Enter the first wavelength by using the keypad numbers or the
left and right arrows.
Press the down arrow.
Step 2
Enter the second wavelength as above.
Press the down arrow.
Step 3
Select whether a background correction is applied to both
wavelengths 1 and 2 using the left and right arrows.
Step 4 (If background correction is On)
Enter the third wavelength, from which the background
correction will be obtained).
Step 5
Press Next to enter the Parameters screen
OR
Press Cancel to return to the Applications Folder.
These folders are the storage locations for any user modified Applications (Methods) that are saved in the Options
menu. Both are accessible from the home folders page.
Favourites:
This folder enables the user to quickly select any frequently used Methods. Up to 9 Methods may be stored in the
folder.
Methods:
These are further storage folders enclosed in the top level Methods folder. Up to 9 Methods may be stored in each
folder.
Operation is identical to the Favourites Folder.
Saved methods can be locked, unlocked and deleted using the Options menu. Select the method by pressing the
relevant key pad number and then press the ::; key.
Delete Method
Press 1 to select delete method.
Select the method to be deleted using the left and right arrows.
Press to delete the method
OR cancel to return to Favourites/Methods folder.
Lock Method
Press 2 to select lock method.
Select the method to be locked using the left and right arrows.
Press the down arrow.
Select a pass code using the keypad numbers or left and right
arrows.
Press to lock the method
OR cancel to return to the Favourites/Methods folder.
Unlock Method
Press 3 to select unlock method.
Select the method to be unlocked using the left and right arrows.
Press the down arrow.
Enter the pass code using the keypad numbers or left and right
arrows.
Press to unlock the method
OR cancel to return to the Favourites/Methods folder.
Summary
Function Keypad number Description
3 Printer/output options
6 Re-name folders
8 Spectro Blocks/Sudoku
Enter the day using the keypad numbers or left and right arrows.
Press the down arrow.
Enter the month as above.
Press the down arrow.
Enter the year.
Press the down arrow.
Enter the hour.
Press the down arrow
Enter the minute. Seconds are zeroed when OK is pressed.
Press OK to store the settings and return to the Utilities folder
OR
Press Cancel to return to the Utilities folder without storing
the time.
2: Regional
3: Printer
Ambient temperature can affect the display. This function can optimise the display for local conditions
The procedure is as follows:
6: Folder Names
Select the folder you wish to rename using the left and right
arrows.
Press the down arrow.
Input the new name for the folder.
Press OK to store the settings and return to the Utilities folder
OR
Press Cancel to return to the Utilities folder without storing
the settings.
8: Games
1: Spectroblocks
Can be set up as Computer mode (50 preset games) or User mode (enter your own pattern)
Use the cursors to select the square and the key pad to enter a number. Invalid numbers cannot be entered. Cells
can be locked (or unlocked) by using the decimal point. Unlocked cells can be cleared using the C key (see also
option key below)
Options
Press ::; to display the options menu
Printer installation
3. Paper gripped.
4. Cover replaced.
Installation
See the manual included on the PVC CDROM for installation and operating instructions.
MAINTENANCE
After Sales Support
Support agreements that help you to fulfil the demands of regulatory guidelines concerning GLP/GMP are
available.
• Calibration, certification using filters traceable to international standards
• Certificated engineers and calibrated test equipment
• Approved to ISO 9001 standard
Choice of agreement apart from break down coverage can include
• Preventative maintenance
• Certification
When using calibration standard filters, insert such that the flat surface is facing away from the spring end of the
cell holder.
Lamp Replacement
The xenon lamp should not need replacement until after several years of use. In the unlikely event that it does need
replacing, this should be undertaken by a service engineer from your supplier.
External cleaning
Switch off the instrument and disconnect the power cord.
Use a soft damp cloth.
Clean all external surfaces.
A mild liquid detergent may be used to remove stubborn marks.
Specifications are measured after the instrument has warmed up at a constant ambient temperature and are typical
of a production unit. As part of our policy of continuous development, we reserve the right to alter specifications
without notice.
Warranty
• Your supplier guarantees that the product supplied has been thoroughly tested to ensure that it meets its
published specification. The warranty included in the conditions of supply is valid for 12 months only if the
product has been used according to the instructions supplied. The supplier can accept no liability for loss or
damage, however caused, arising from the faulty or incorrect use of this product.
• This product has been designed and manufactured by Biochrom Ltd, 22 Cambridge Science Park, Milton
Road, Cambridge CB4 0FJ, UK. However, please contact your original supplier in the first instance if you
experience technical or sample handling or sample handling difficulties.