Mars 40: Installation/Service Manual
Mars 40: Installation/Service Manual
Mars 40
(High Frequency X-Ray Machine)
INSTALLATION/SERVICE MANUAL
Catalogue No. 23
PRODUCT APPROVALS
SYSTEM APPROVALS
SWITCH ON SAFELY
ATTENTION SYMBOL
QUALIFIED SERVICE
WATER RESISTANCE
INTERFERENCE
WARNING MESSAGE
PROTECTIVE EARTH
OFF
Indicate that equipment can be switched OFF when
pressed this switch.
ON
Warnings:
• This device has not been designed to be used in environments where vapours, anesthetic
mixes flammable with air, or oxygen and nitrous oxide can be detected.
• Before cleaning the device, please disconnect it from the AC Mains supply.
• Wherever necessary, use the suitable accessories, such as the Leaded Aprons, Lead
Goggles, and Thyroid Guards to protect the Patient, User & Support Staff of OT from
radiation exposure.
• While performing the radiography, no one, apart from the Operator and the Patient, must
remain in the room.
• Though this unit has been designed with a quite acceptable protection level from
electromagnetic interference, it is advisable to install it at a certain distance from
electrical energy transformation rooms, from static continuity units, from portable
receiving-transmitting units and for Cellular use. Cellular Telephones are only admitted
at a distance of more than 1.5m from any component of the device. Other medical
instruments and devices that must be used in the same installation area of the unit must
comply with the Electromagnetic Compatibility rules in force. Non-complying
instruments, of which the poor immunity from electromagnetic fields is well known,
must be installed at least 3 m away from the equipment and supplied by a different
electrical line.
• The equipment must be OFF while using devices such as electrical scalpels or similar
instruments. The equipment automatically performs some self-test functions and
therefore is able to detect some anomalous conditions. However, whenever there are
doubts on the performance of the equipment (e.g. images too clear, too dark or with
artifacts), the activity must be suspended and an Allengers Qualified and Authorized
Engineer must be contacted.
The equipment contains in some of its parts, materials that at the end of the unit’s life, must be
recycled at an approved location. Particularly the device contains the following materials and/or
components:
• Tube Head Assembly: Tube Head Assembly of this X-Ray Equipment contains High
Voltage Tank, X-Ray Tube and Oil filled alongside, Printed Circuit Board (PCB) and
Lead used for controlling Stray Radiations.
Allengers is not responsible for the disposal of the equipment performed by the user and
for the costs related to this intervention.
It is assumed by the manufacturer and distributors of this equipment that all persons responsible
for the operation of this equipment are aware of danger and excessive exposure to X-Radiation
and this equipment is sold with the understanding that Allengers Medical Systems Limited,
their agents and representatives have no responsibility for injury or damage which may
result from exposure to X-Radiations.
a) The radiation exposure to the patient should be the minimum exposure required to
produce images of good diagnostic quality. Use radiation with care, caution and at a rate
which is ALARA - ‘As Low As Reasonably Achievable’
b) The speed of the film, or screen and film combinations should be of fastest speed
consistent with the diagnostic objective of the radiographic examination.
c) When a patient or film must be provided with auxiliary support during radiation
exposure:
(i) Mechanical holding devices shall be used when the technique permits.
(iii) If a human holder is required, the holder shall be positioned such that no part
of body shall be struck by primary X-Ray beam unless protected by at least
0.5 mm lead equivalent and shall be protected from direct scatter radiation by
protective lead apron of not less then 0.25 mm lead equivalent.
d) Gonadal shielding of not less then 0.25 mm lead equivalent shall be used for patients
who have not passed their reproductive age during radiographic procedures in which
gonads are under primary beam, except, in cases where shielding would interfere with
diagnostic procedures.
e) Doors of the Room/OT where X-ray equipment is installed should be closed before
making exposures.
f) The X-ray beam should not be directed towards doors or windows of the room, or
towards control panel or darkroom walls unless no other geometry is possible.
i) Use Collimator to set minimal useful field size; keep Image Intensifier in close
proximity to the patient.
1. Switch OFF the Power Supply: If this is not possible, protect yourself
with dry insulating material and pull the victim clear off the conductor.
Do not touch the victim with your bare hands until he/she is clear off
the conductor.
3. Keep the air passage clear by turning the head to one side, opening the
patient’s mouth and cleaning it of water, saline, mucus or blood, a lot of
which might have accumulated in the back of the throat.
4. If the jaw is rigid, try to force the mouth open by pressure on the gum
behind the last molar tooth of the lower jaw. When the upper air
passage is cleared, tilt the head backward and force the jaw forward
from the angles of the jaw in front of the ears. This would prevent
mechanical obstructions to the upper air passages.
5. Hold the chin up and forward with one hand and pinch the nostril of the
victim with other.
6. Take a very deep breath and apply your mouth to that of victim and
blow into his mouth until the chest of the victim moves up indicating
filling of the lungs (Never allow the chin to sag).
7. When the chest has moved up, withdraw your mouth and allow the
chest to sink back.
Mars 40 is an X-Ray Machine with high frequency generator and have features for operator's
ease like APR.
The Radiography Parameters defined in this machine are based on Human Anatomy. This is
reason why this machine is known as APR (Anatomical Programmed Radiography Machine).
There are different keys to define the Body Part to be exposed and depending on the defined
Part by operator, machine it-self set the parameters depending on the built of patient and Screen
Speed combination available for Exposure.
Operator can choose manual mode and can select their own parameters fro exposures.
A Variety of tube stands & tables can be chosen depending upon the customer’s requirements.
This machine is best suited for Chest, Extremities, Skulls, Abdomen & Pelvic for all routine &
special Radiographic & Fluoroscopy examines.
Machine is supplied with different combinations and tables depending on the type of operation
needed.
Whole of the equipment offers innumerable benefits like compact size, consistent output and
negligible dose rate.
Mars 40 with Ceiling Free Stand and Floatex Table makes and ideal combination for all
possible radiographic applications offering ease of patient positioning specially in Trauma
cases.
Mars 40 with Floor to Ceiling Stand & Allpose is used for Special Investigation purposes e.g.
HSG, Barium Investigation etc.
a) Block Diagram:
3 PHASE
440 V AC
INPUT
OPERATOR CONSOLE
CONTROL UNIT
HIGH VOLTAGE
X-MER
X-RAY TUBE
COLLIMATOR
X-RAY
GRID
IMAGE RECEPTOR
b) Sequence of Operation: -
Press ON Switch to Wait till battery signal
440 V AC >> Switch ON MCB >> >>
energize equipment on panel show FULL
<<
Fuse Details:
10
3
1 9
4 8
7
c
12 (APR
Keys)
1. MACHINE ON SWITCH
This switch is used to turn OFF the machine. Indicator along this switch glows
when power is available to the machine but machine is in OFF condition.
3. FAULT INDICATORS
When we switch ON the machine Low Capacitor Indicator will glow about
approx. 50 sec. When this indicator is glowing the machine does not exposes.
When full indicator glows then machine is ready for exposure.
This switch is used to Increase and Decrease the KV. KV can be increased or
decrease from 40 to 125 KV in steps of 1 KV each and the display of KV is shown
on the display above these switches.
This switch is used to Increase and Decrease the mAs. mAs can be Increased or
Decrease from 1 to 200 mAs. The display of mAs is shown on the display above
these switches.
8. BUCKY SWITCH
This switch is used to energize the Bucky. When no Indicator is glowing then none
of the Bucky is selected. When we press it one time then H Indicator glows and
Horizontal Bucky is selected. If we again press the switch then V Indicator starts
glowing and Vertical Bucky is selected
This switch is used to Boost the Filament. When this switch is pressed, Indicator
above this switch glows.
This switch is used to make an Exposure. For exposure, first press the Ready
Switch and then Exposure Switch. Exposure Indicator above this switch glows to
indicate that exposure takes place.
In AEC Mode user can select the KV according the patient size and also select the
chambers as per area of exposure. The machine will automatically select the dose
rate to be given to patient.
12. APR KEYS:-
a) APR Key
When we switch ON the machine then by default machine is in the APR Mode, and the
corresponding indicator will glow. By pressing this switch the machine will come out of
APR Mode and vice versa.
These switches are used to select the body size of the patient.
When XL switch is pressed then the factors selected is for extra large patient and
similarly for L, M, S, XS and baby the factors selected are for LARGE, MEDIUM,
SMALL, EXTRA SMALL and BABY.
These switches are used to Increase or Decrease the mAs as per density of film used by
the Operator.
These switches are used to select the Area and the View in which X-Ray is to be taken.
This Switch is used to define the Area to be Exposed. By pressing these two switches
the operator can select the area for exposure and the corresponding LED for selected area will
glow on the panel e.g. if the operator scroll down by pressing down key to Chest then
corresponding LED will glow on the Panel.
By pressing the V1, V2, V3 & V4 key one can select the view for the X-Ray e.g. if operator
wants to take PA view for the chest then he/she has to press key and if he/she
wants to take LAT view then he/she has to press similarly if one has to take the
SUPINE AP or SUPINE LAT, he/she has to press V3 & V4.
Similar procedure will follow for the rest of selection.
Exposure
This is used to press Ready switch and Exposure switch using retractable type cord.
AUDIBLE ALARM
+VE -VE
Control Side:- Connect LBD supply (24V AC) wires to TS3-8 & TS3-9.
Tube Head Side Connections:- Connect 4 Pin Connector in Electromagnet Panel &
Connect 5 Pin Connector of Electromagnet Panel to In LBD. Connect 6 Pin connector
of Stand Supply in Electromagnet Panel.
The Entire surface of the Bushing & Receptacle should be thoroughly cleaned
for any dirt or dust.
Jelly should be applied on the High Voltage Bushings of the HV Cables so as not
to leave any air between receptacles & bushings. (It is strictly advices that
apply the jelly, which is supplied with the tube. Other jelly should not be
used).
The Earth wire on the HV Transformer should be properly tightened.
All connections on the H.T. Tank & Tube Head should be firmly tightened.
All connections on the H.T. Tank & Tube Head should be firmly tightened.
Inverter Section
Fuses (F4 & F5)
63 A Fuse for
Contactor
ON/OFF
Contactor
Motherboard Card
Hand Switch
connections
Buzzer
Connection
+5 V DC Supply Card
Filament (mA)
Stab Card
+ 12 V SMPS
Filament Card
Filament Card
Filter for
MAIN X-MER
LBD X-MER
Factory Setting
don’t Operate
Pulse 21 Card
TS3 (Rotor, Table &
VBS Connections)
SMPS +24V
Rotor Capacitor
SMPS +15V
Time Delay Card
for Rotor
Rotor Plate
RTS
RE3-RE5 Card
ROT1
FUSE
(8A)
ROT6
ROT2
ROT4
RTS
Note:
1. The letters written in curly brackets {} denotes the location of part/ Component/
contact terminal in wiring diagram.
2. Abbreviation
Relay sequence:- ROT1 – R OT5 – ROT2 – ROT3- ROT5 OFF, ROT6 ON- ROT
1. The letters written in curly brackets {} denotes the location of part/ Component/
contact terminal in wiring diagram.
2. Abbreviation
Phase 1 is given to the CON.1-A {A2} & Programmable Relay Terminal-R {B1}
Phase 2 is given to CON.1-B {A2} & Programmable Relay Terminal-Y {B1}
Phase 3 is given on Terminal Strip TS1-1 and then CON.1-C {A2} &
Programmable Relay Terminal-B {B1}
• Neutral is directly given to Filter from Terminal Block and from Filter to TS1-2,
Programmable Relay Terminal-N {B1} and Contactor Terminal CON.1-D {B2}
• Ground of the Machine should be properly earthed such that voltage between GND and
Neutral must be minimum as possible (Maximum allowable voltage is 3 volts).
• Phase-1 is directly given to ON/OFF Contactor CON.1 Coil {B1} and Phase-2 is given
to it through the NO Contact of RE1A-C {B1}.
• 3 Metal Oxide Variastor (MOV) are used to protect circuit against excessive transient
voltages at TS1-1 (Line) & TS1-2 (Neutral), TS1-1 & TS1-3 (Ground) and TS1-2 &
TS1-3.
• Now 230V from TS1-1 and 0VAC from TS1-2 is fed to 15W Bulb mounted near
Filament Card {B1}, SMPS-1 (12V/1A) {C2-B2}, Cooling Fan Supply {B1} and 0-12V
Transformer (TR-1) {C1-C2}. Now Before ON (0-12V DC) Relay Supply is generated
by SMPS-1 given at TS2-8 (12V DC) & TS2-9 (0V DC). Output of 0-12V Transformer
(TR-1) {C1:C2} is given to Power Supply Card at J1-10 & 11 and Output of Power
Supply Card from J2-10 (+12V DC) and J2-11 (0VDC) is given at TS2-1 & 2
respectively.
• Before ON Relay Supply +12V DC from TS2-8 is fed to RE1, RE1A & RE6 Coils, and
Motherboard Card at JP9-1, JP2-4 and at JP1-1, 2 & 4 & to NO Contact of RE1A-A.
Relay Supply 0V DC from TS2-9 is fed to Motherboard Card {C4} at JP9-3, JP2-7 &
JP1-7, 9 & 12, to the Initial Time Delay Card {C6} at J2-2 & Programmable Relay
Terminal C {B1}
NOTE:
Relays Contacts used for only DC supply is mentioned as DC Relay and
Contacts used for AC supply is mentioned as AC Relay
• As the ON Switch from the Panel is pressed, Transistor Q20 on the Motherboard Card
gets biased and pulls down the Relay Coil RE1 from JP9-2 making it to energize through
NO Contact of RE6-A and then from NO Contact of RE1-C {C4}, 0V Relay Supply is
given to RE1A Coil from JP9-3. Hence RE1A operates which further gives + 12V Relay
Supply to all Relays through NO Contact of RE1A-A {D5}. ON/OFF Contactor of
CON.1 gets Phase-2 from NO Contact of RE1A-C {B1} and gets ON.
• CON.1 Contactor’s Contacts give the 3-Phase supply to Delay Contactor CON.2
Terminals CON.2-A {A2}, CON.2-B {A2} and CON.2-C {A2}. Initially 2-Phase
Supply is fed to Inverter via two Wire Wound Resistances (50Ω, 50W) to limit the
charging current of capacitors inside the Inverter until CON.1 (Delay Contactor)
operates.
• After ON supply now comes at TS1-4 & 5 from Contactor Terminal CON.1-C {A2} and
CON.1-D {B2}. From TS1-4 & 5 Supply is fed to SMPS-3 (24V/2A) {A2}, to Filter
Card {B4}, to Rotor Plate {B5} at RTS-5 & 4 respectively, SMPS-2 (15V/3.4A) {B2},
to Cooling Fan, to LBD Transformer (TR-3) {C1}, to Main Transformer (TR-2) {C2}.
TS1-4 is then fed to NO Contact of RE5A-A {B2} (Time Delay Contactor Supply) to
the NO Contact of RE2A-A {D1} (For Bucky Supply).
• From TS1-4 230VAC Supply given at TS1-10 & from TS1-5 0VAC given to TS1-11 for
Machine ON Indicator Supply.
o 15V DC supply is given at the Connector JP5-2 of Motherboard Card for Low
Indication at Panel through NC Contact of RE5-A {C4}.
• SMPS-3 (24V/2A) {A2} gives 24V DC supply to Pulse 18 COM Card at CP1-1 mounted
in Inverter through NO Contact of RE1-A {A3}.
• Main Transformer (TR2) {C2-B2} gives 15V-0-15V, 0-24V, 0-24V output which are
further given to Power Supply Card {C2-B2}. 15-0-15 to J1-3, 4 & 5 and 24-0V at J1-6
& 7 & J1-8 & 9. Power Supply Card converts these AC supplies to DC supplies.
o 12VDC (Relay) output from TS2-8 is given for Rotor Relay Supply at RTS-10
through NO Contact of RE2-A {B5}.
• LBD Transformer (TR3) {C1} gives 24V AC Supply to LBD & Magnets through Fuse
F5 at TS3-8 & 9.
• After 30-40s, 0V (Relay) supply comes from Connector J2-3 of Initial Time Delay Card
and Relays RE5 & RE5A gets energized which further gives 230V AC Supply to Delay
Contactor CON.2 Coil and then CON.2 operates through NO Contact of RE5A-A {B2}.
As CON.2 Contactor operates, it starts giving 3-Phase Supply to Inverter directly and by
pass the Wire Wound Resistances. NO Contact of RE5-A {C4} gives 15V DC to
Motherboard Card for Full indication at Panel to Motherboard Card at JP5-1 and Initial
Time Delay OK Signal at JP8-5 of Motherboard Card to close the safety Interlock
Signal.
• When L/S Switch is pressed, Relay K2 operates mounted on the Motherboard Card,
which changes Set Filament Voltage for Actual mA & gives 0V (Relay) Supply at JP1-
10 of Motherboard Card to energize the RE3 (Large/Small Relay) & following functions
are performed:
1. When Horizontal Bucky is selected by pressing H-Bucky Switch, LED along with this
Switch glows on the Panel & 0V Signal comes from JP1-3 of Motherboard Card, which
energizes the RE4 (Horizontal Bucky Relay) and 230V AC Supply, is routed through
NO Contact of RE4-A {D1} at TS3-10 for Horizontal Bucky.
2. When Vertical Bucky is selected by pressing V-Bucky Switch, LED along with this
Switch glows on the Panel & 0V Signal comes from Motherboard Card at JP1-5 that
energizes the RE4A Relay & 230V AC Supply is routed through NO Contact of RE4A-
A {D1} at TS3-12 for Vertical Bucky.
NOTE: -230V AC Supply is routed through NO Contact of RE2A-A to RE4 and RE4A
Relays so Supply will be available at the time of Ready.
b. Pre-Rx Signal is generated from Motherboard Card at JP6-3 & goes to Pulse 19
Card at CP1-5 through NO Contact of RE2-B {C4}.
c. Tube Filament Boost during Ready for Exposure. Filament is boosted through set
Filament Signal. Pre-Heat Voltage comes from JP6-6 of Motherboard Card at J7-
6 of Filament Card until Ready Switch is pressed but when Ready is pressed Pre
Heat Voltage converts to Set Filament at Filament Card J7-6 (Set Filament
Voltage for actual mA) which is more than Pre-Heat Voltage.
2. NO Contact of RE2-A {B5} sends 12V DC supply to Rotor Plate (RTS-10), which
energizes the ROT1 Relay. Now Tube Rotor works on 230V AC supply. After 0.8
Second Delay ROT3 Relay energizes & then Rotor starts working on 65V AC through
ROT6. If Rotor Circuit is working OK, a Rotor OK Signal of 15V DC goes from RTS-
12 to Motherboard Card at JP8-4 as a Feedback Signal and completes the Rotor
Interlock Signal.
3. 230V AC supply goes to NO Contact of RE4-A {D1} & RE4A-A {D1} for Bucky
Supply. (RE4 for Horizontal Bucky & RE4A Vertical Bucky Relay) through NO
Contact of RE2A-A {D1}.
2. Com-Rx Signal then generates COM1 & COM2 Signals in Pulse-19 Card, which are
further given to Pulse 18 COM Card for IGBT firing through FRC Cable (CF1). The
output of Inverter goes to Pulse-21 Card on HT Tank.
3. HV Tank Step Up this Voltage, and converts into DC according to the KV Selection.
Output of HV Tank is given to Tube through HV Cables. Then calibrated current flows
in Tube, the current flows & KV Feedback is given to Pulse 19 Card at CP2 by HV Tank
through CP1 of Pulse 21 Card for correction of current & KV.
4. A Buzzer {C4} beeps during Exposure. (Buzzer Supply comes from JP7-1 and JP7-2 of
Motherboard Card).
RE4/RE4A – Horizontal/Vertical
Bucky Supply Relays (Optional)
Exposure
• THERMAL FAULT: - This Signal is used for protection of HT Tank from Over
Heating. This Signal is normally 15V DC generated from HT Tank at CP2-2 and given
to Motherboard Card at JP8-6. If This Signal becomes 0V DC then TH-Fault Indicator
on Panel glows.
• FILAMENT FAULT: - This Signal indicates the Filament glow status of Tube, if there
is any problem in Filament Card Ckt / Tank / Tube. Then this Signal becomes 0V DC
and FIL-Fault glows at Panel. This Signal is normally 10-13V DC generated in Filament
Card at J7-9 and given to Motherboard Card at JP8-3.
• ROTOR OK: - This Signal is generated from Rotor Plate to indicate the proper working
of Rotor Ckt. This Signal is normally 0V DC and during Ready 15V DC if Rotor Ckt. is
working OK. This Signal is fed from RTS-12 to Motherboard Card at JP8-4.
• INITIAL TIME DELAY OK: -This Signal is used to indicate the Full Charging of
Capacitors inside the Inverter. Normally this Signal is 0V DC and when Time Delay
Relay gets ON (RE5 & RE5A) this Signal becomes 15V DC. This Signal fed from RE5-
A to Motherboard Card at JP8-5.
Frequency of Supply
• SET KV SIGNAL: - This Signal decides the generation of KVP. This Signal varies
according to display KVP with a relation of 1V=20KVP. If 40 kVp is selected then
Signal Voltage will be 2.0V DC. This Signal is generated in Motherboard Card by ICU7
-LF412 & can be checked at Set KV Test Point or at JP6-5 in Motherboard Card. This
Signal can be adjusted from POT R10 & R15 on the Motherboard Card.
• PRE -HEAT SIGNAL: -This Signal is used for Filament glow of Tube to warm up the
Tube Filament before Exposure. This Signal controls the Filament Glow & can be
checked in Motherboard Card at Pre-Heat Test Point & adjustments can be done from
R39 Pot of Motherboard Card. This Signal is normally 2.0V DC and fed to Filament
Card at J7-6.
• SET. FILAMENT SIGNAL: - This Signal is used to achieve actual mA. The level of
voltage depends upon the machine rating, KV station, & according to tube. Max.
Voltage can be set is 5.0V DC for Large Filament & 3.5VDC for Small Filament.
Overall Pot is R86 & Individual Pot for Large Filament is VR1 to VR11 & for Small
Filament VR17 to VR26 at Motherboard Card.
• mA STAB. SIGNAL: - This Signal is used to stabilize the mA output according to the
Selected KV station. mA Stab Signal is generated in Filament (mA) Stabilization Card
and can be checked at J6-7 at Filament (mA) Stab Card or at J7-4 of Filament Card. The
level of voltage depends upon the KV station & mA equation of Pulse 19 Card. If mA
are 250 & mA equation is 1V=50mA then Stab. Voltage will be 5.0V DC.
• ANODIC mA:-This Signal is generated in Pulse 19 Card during Exposure and Test
Points are TP6 & TP11 or can be checked at CP1-12. It is a reference voltage, which
indicates the Actual mA, flows through Tube. This Signal is given to Filament Card (J7-
5) for comparison of Stab Signal & Actual mA.
mA Equation:-
1V=50mA
• PRE-RX: - This Signal is used to complete an Interlock in Pulse 19 Card for Exposure.
This Signal is generated in Motherboard Card at JP6-3. Pre-Rx Signal is normally 15V
DC & on load 11VDC. This Signal is given to Pulse 19 Card at CP1-5.
• COM-RX: -This Signal is used for generating COM1 & COM2 Signal in Pulse 19 Card
which is used for firing of Inverter’s IGBT. This Signal is generated in Motherboard
Card at JP6-4. Com-Rx Signal is normally 15V DC & on load 10V DC. This Signal is
given to Pulse 19 Card at CP1-9.
• Step 1:- Remove JP6 connector & connect all other connectors of Motherboard Card
and Ensure all connectors should be tight. Calibrate the Following signals:-
1. Set KV:-Check Set KV between JP6-5 & GND or at Set KV test point of
Motherboard Card. Voltage should be 2.0 V DC at 40 KVP & 6.0 V DC at 120
KVP. If required, adjustment can be done from the Pot R-10 or R-15 at
Motherboard Card.
2. Check Pre-Heat voltage between JP6-6 Motherboard Card & Main GND or at Pre-
Heat Test Point, voltage at this point is 2.0 V DC (approx). If required, adjustment
can be done from the R39 in Motherboard Card.
3. Check Set Filament between main GND and SET FIL Test Point, voltage at this
point is 2.5V DC (Approx.) at starting calibration stage (Can be increased to achieve
mA). If required, adjustment can be done from the R86 (Overall POT) or Individual
pot (For 40KVP 10mAS VR1-Individual POT) at Motherboard Card.
5. Check SET FIL. Signal at J7-6 during ready should be same as set on Motherboard
Card SET Filament Test Point.
• Step 2:- Tight All the connectors of Filament Card after Ensuring All supplies:
1. Check Filament glow- Large Filament should glow during Large selection on Panel
& Small Filament should glow during Small Filament selection.
• Step 4:- Tight the mounting screws of Inverter check proper fixing of P1A and P1B
wires, 3-phase supply connections. Connect all connectors properly of Pulse 19 Card
after checking supply at CP1 Connector.
3. Ensure LD1 should glow after Removing CF1 FRC Cable from Pulse-19 Card.
4. Connect CF1 FRC Cable in Pulse 19 Card ensure FRC Cable should be locked in
Male FRC connector of Pulse 18 COM and Pulse-19 Card properly.
5. Press APR Key on Panel to activate the NON-APR (Manual) Mode. All the
Calibration is to be done in NON-APR (Manual) Mode.
R-17 R-1
R-37 (OVERALL)
If required adjustment can be done from overall POT R-37 on all stations.
6. Make an exposure at 40KVP 4mAS and hold the Image, check the voltage in
CRO.
8.0V
9.0V 8.0V
7.0V
Low Set Fil. mA Wave shape With mA Stab. Function Proper mA Wave shape With mA Stab. Function
mA wave shape in
CRO
Limit of Set Fil.:- For large filament 5.0VDC & For Small Filament 3.7VDC
• Select the Small Filament from Control Panel. Check the actual glowing of Small
Filament. If it is OK then calibrate the mA as per following chart at different KVP
stations by adjusting SET Filament Signal in motherboard card with the pots mentioned
in KV mA chart.
If required adjustment can be done from overall POT VR-86 on all stations
Alpha-Numeric
Display
Green indicator
(In case of Voltage OK)
Press UP/ DN buttons for setting the
limit of parameters
1. DURING MACHINE ON LD9 (+15V) AND LD10 (-15V) WILL GLOW INDICATES
THAT SUPPLY TO CARD IS O.K.
2. Technic chart should be fixed near the machine and operator/customer should be
explained the Technic chart.
3. The Service Engineer should demonstrate the equipment to the operator. The
operator should be given training on essential Do’s and Don’ts like ---
a. Use as much of Shielding material as possible, use lead screens, lead aprons, lead
goggles and shields etc.
c. Use collimators/cones and collimate radiation field to the smallest possible area.
d. Before giving an exposure for the first time in the day, try to warm up the
equipment for 10-15 minutes. For this purpose simply switch ON the machine
and leave it as it is.
4. The Equipment should be handed over to the customer. The Installation Report
should be properly filled along with the Warranty Card and required Serial Nos.
Lead Goggles
Protective Barrier
Lead Apron
Lead Gloves
Thyroid Shield
Gonad Shield
For optimum output of the equipment we should start writing the voltage drop during exposure
at Average working KVP Time and mA settings. This record at the time of installation itself can
be mentioned on the Machine History Card. The date and month of Installation would help us
actually know the extent of voltage drop as our experience suggests that during the summer
season; the voltage drop is on the higher side. During P.M the Voltage drop can be checked with
the reference value and is noted. Any increase in Voltage Drop would thus be noticed and line
checked for any loose connections etc. in the mains switch/pole etc.
Inform customer in advance about your visit preferably through a letter/ telephone and agree for
a convenient time and date.
A). GENERAL:
Whenever visiting for any Preventive Maintenance, please ensure to carry all the Tools, Meters,
Grease, Lubricants, Machine History Card and consumables like the dusters, polish etc. Please
make sure that your Multimeter is calibrated with a Standard Meter lying in your Branch Office.
1. On reaching the Customer’s site enquire from Operator/Doctor for any specific
problems they are facing with the equipment.
2. Rectify first the problems as indicated by the customer and then proceed as under for
carrying out the Preventive Maintenance
1. Before attempting to carry out the P.M. switch ON the equipment and check the working
of the Machine electrically and mechanically. Observe for any flaws, which should be
rectified during the course of P.M.
3. Switch OFF the Main Switch, open the cover and tighten any loose contacts/screws.
Also check the fuse wire Rating and it must comply with the Rating of the equipment in
operation.
4. Open the control panel cover. Keeping the Machine in “OFF” condition tighten all the
screws on the Terminal Blocks/Resistance Taps etc.
5. Check the rating of all the Glass Fuses and replace if necessary.
6. Open cover of ON/OFF & Delay Contactors and check that contacts are not worn out or
burnt & should be clean. Check the operation of the X-Ray contactor. It should not be
sticky.
7. Check Calibration as per calibration procedure.
1. Check all the movements of Tube Head, Tube Stand and Tables. Check for functioning
of all locks and ensure that all movements are smooth and noiseless. Specially check for
Vertical Carriage Up/Down Movement. Rectify wherever required. Remove old grease
with cloth and apply fresh grease on all movement areas (wherever applicable).
2. Open the Table Top and brush out all the dust settled on the Grid and on the Bucky
Motor Area. Clean, apply lubricant and check the Bucky play, it’s lock and tighten all
nuts/bolts/screws in the Table.
3. You must ensure that the alignment of the Center Ray of Light Beam diaphragm
matches with the Table center and Bucky center (Adjust if required).
F) DARK ROOM
Check the Dark Room and its accessories for the following:
1. No light leakage in the Dark Room.
2. Safe light is of proper intensity.
3. Films used should not be beyond the expired dates.
4. Cassette locks are not loose.
5. Screen Quality should be OK and no staining of the screens.
6. Developer and the fixer are fresh and their temperature is maintained properly.
7. Wooden sticks are available for stirring of the Developer and fixer.
8. Radiographer/Operator is trained to handle Dark Room Accessories.
G) GENERAL
1. Enquire if radiographer/Operator needs any training for use of equipment & impart if
necessary.
2. Handover Equipment to customer.
3. Explain/show equipment to customer, apprise him of jobs undertaken.
4. Fill Machine History Cards and get the card/report signed from customer.
5. Inform tentative date of next Preventive Maintenance.