Diodos Dosimetria in Vivo
Diodos Dosimetria in Vivo
Diodos Dosimetria in Vivo
ISBN 90-804532-3
David I. Thwaites,
Department of Oncology Physics, Clinical Oncology, University of Edinburgh,
Western General Hospital, Edinburgh, Scotland, U.K.
D.P. Huyskens, R. Bogaerts, J. Verstraete, M. Lööf, H. Nyström, C. Fiorino, S. Broggi,
N. Jornet, M. Ribas, D.I. Thwaites
ISBN 90-804532-3
2
ESTRO
Mounierlaan 83/12 – 1200 Brussels (Belgium)
3
SUMMARY
Entrance in vivo dosimetry with diode detectors has been demonstrated to be a valuable
workload generated by in vivo dosimetry is one of the factors that impedes a widespread
implementation. Especially during the initial period of establishing the technique in clinical
routine, the responsible QA person is confronted with variable tasks, such as purchasing
contacting experienced departments in order to gather information and define the sequence
This booklet is set up as a tool to reduce these initial efforts: it is conceived as a step-by-
step guide to implement entrance in vivo dosimetry with diodes in the clinical routine of a
radiotherapy department.
The first chapter about the preparation of the measurements contains information
guidelines for the calibration of the diodes and the determination of correction factors are
given.
The second chapter discusses the actual tasks of the res ponsible QA person during the
initial training period, with the emphasis on the implementation of the measurement
In the third chapter, the interpretation of the measurement in relation to tolerance and
4
centres. In the final chapter, elaborate contributions are given from five centres about
5
TABLE OF CONTENT
Introduction ……………………………………………………………………………….10
1.1 Equipment.......................................................................................................................13
1.1.2 Electrometer.............................................................................................................16
2.2 Defining guidelines for the persons performing the measurements ......................31
3.2.2.1 Errors in data generation and data transfer (human errors) .................................................... 41
6
3.3.1 Actions after the first measurement.....................................................................45
4.1 What equipment do you use to carry out routin e in vivo dose measurements?.50
4.2 Philosophy of your department concerning the use of in vivo dosimetry? .........53
4.3.4 Time period between checks of calibration and correction factors ................65
4.4 What system do you use for the recording of in vivo dose measurements?.......66
7
5.1.2.3 Wedge correction factor (CFwedge )................................................................................... 83
5.2.1 Introduction.............................................................................................................98
5.2.2.2 Methods...................................................................................................................102
5.2.3 Results....................................................................................................................103
5.2.3.4 Influence of beam modifiers: tray and block correction factor C T and CB ................................108
5.2.4.1 Independence of field size and source-to-surface distance correction factors .............................111
5.2.5 Conclusion.............................................................................................................114
8
5.3 Practical implementation of cost-effective approaches to in vivo dosimetry -
5.3.1 Introduction...........................................................................................................115
5.4.1 Introduction...........................................................................................................132
5.4.3 Equipment..............................................................................................................133
5.4.8 Conclusion.............................................................................................................139
5.5.1.1 Equipment................................................................................................................141
9
5.5.1.2 In vivo measured and expected entrance dose...................................................................141
5.5.2 Results....................................................................................................................144
5.5.2.2 Systematic errors detected before in vivo dosimetry by MU calculation/data transfer check .......145
10
ACKNOWLEDGEMENTS
The authors gratefully acknowledge the support of the EU in the framework of the program
“Europe against Cancer”. The contribution from Leuven on the performance of diodes in
high-energy beams is based on work performed for the most part by Dietmar Georg
measurement procedures has been partially financed by FIS project 98/0047-02. We also
acknowledge the collaboration of Alain Noel (Centre Alexis Vautrin, Nancy) and Ben
11
INTRODUCTION
The aim of this booklet is to provide the radiotherapy community with practical guidelines
for the implementation of in vivo dosimetry (IVD) with diodes at a routin e/departmental
level.
Since in vivo dosimetry with diodes is a broad subject, considering the full map of varieties
encountered in radiotherapy, the authors have restricted themselves to guidelines for the
measurements of the entrance dose with diodes in p hoton beams. This technique of in vivo
supplement and complement basic pre -treatment QA methods, such as the independent
check of dose calculation and data transfer, which should be in routine use in the
The information contained in this booklet is a practically usable distillate from other
includes, for the sake of completeness, publications dealing with exit dose measurements
and midline dose calculations, and in vivo dose measurements in electron beams. The way
other ESTRO booklets on in vivo dosimetry - that have appeared and will appear – and to
other review publications. These include the first ESTRO booklet on in vivo dosimetry
“Methods for in vivo dosimetry in external radiotherapy”, written by J. Van Dam and G.
Marinello [Van Dam 1994], and a new ESTRO booklet “In vivo dosimetry in clinical
practice: When and What to measure? How to correct?”, written by E. van der Schueren,
A. Dutreix and C. Weltens [van der Schueren 2001]. A nice general review on in vivo
The latter two publications highlight the more philosophical questions concerning the use
of in vivo dosimetry. These will not be discussed here. Also, the future use of diode
measurements in relation to conformal irradiation techniques and IMRT - for which point
dose verification is obviously inadequate – is a topic outside the scope of this work.
12
CHAPTER 1 "GETTING STARTED”
1.1 EQUIPMENT
1.1.1 DIODES
combination of high sensitivity, immediate readout, simplicity of operation (no external bias
Silicon diodes can be made starting from either n-type or p-type silicon, which behave
differently because their minority carriers are holes or electrons, respectively. Figure 1.1
illustrates the basic operation of a p-type silicon detector diode. In the boundary between
two regions, one of p-type and another of n -type silicon, there is a depletion of free charge
carriers. When the detector is operating with zero external voltage a potential difference of
about 0.7 V exists over this depletion area, causing the charge carriers created by the
radiation to be swept away into the body of the crystal. As the diode is asymmetrically
doped - the n-type region is much more heavily doped than the p-type region - the
irradiation induced charge flow is comprised almost entirely of electrons (holes in an n -type
diode). Due to defects in the crystal lattice some electrons are trapped and will
consequently not contribute to the diode signal. An n-type diode is more influenced by
Figure 1.1 Schematic overview of the basic principal of a p-type silicon diode used
as a radiation detector.
13
The detector sensit ivity depends on the lifetime of the charge carriers and consequently on
the amount of recombination centres in the crystal, which is determined by the diode type,
the doping level and the accumulated dose. As the radiation induces recombination
centres within the lattice, the sensitivity will decrease with accumulated dose.
another. The different influences are taken into account in calibration and
The effect of radiation damage represents the main limitation of silicon diodes.
Furthermore, other effects related to the detector material have to be considered (Figure
1.2):
• The diode signal depends on the photon energy. This is due to the higher atomic
number of silicon (Z = 14) compared to soft tissue (Z = ∼7) and the corresponding
• The diode signal is dose rate dependent. At high instantaneous dose rates the
recombination. This leads to a proportionally higher response at higher dose rates. This
effect is more pronounced for n-type Si-diode detectors than for those made of highly
14
doped p-type Si [Heukelom 1991b]. The dose rate dependence may change with
• The diode signal is influenced by temperature. In general, the sensitivity increases with
increasing temperature. This effect is less pronounced for an un-irradiated diode and
will increase with accumulated dose. However, the rate of change of sensitivity with
parameters, as SSD (source skin distance), field size, and the presence of wedges, trays and
blocks, will be incorrectly reflected by the diode signal variation. For this reason, correction
factors have to be determined (see Section 1.2.3 describing the practical details). Beside the
physical properties of the diode crystal, other factors contribute to the magnitude of these
correction factors (Figure 1.2). First, there is the inevitable fact that the measureme nts are
performed with the diode located outside the patient or the phantom. The photon scatter
conditions experienced by the diode are therefore different from those at the point of
entrance dose definition, i.e. at the depth of dose maximum inside the patient or the
phantom. For instance in high energy beams the diode reading is nearly independent of
phantom scatter, while the entrance dose is clearly not (see [Jornet 2000], [Lööf 2001],
[Wierzbicki 1998]). In addition, the diode may experience a different amount of head-scatter
electrons. As a consequence, the effect of field size, SSD, and presence of wedges, trays
and blocks on the contaminating head-scatter electron contribution may induce large
In relation to this, the construction of the detector, for instance the thickness and the
shape of the build -up cap, is another factor influencing the diode signal. The shape of the
build-up cap will influence the angular response: a cylindrical cap has a different angular
dependence than a hemi -spherical one. The thickness of the build -up cap determines the
scatter conditions seen by the diode. To minimise the correction factors and ensure a
15
build-up cap with a thickness equal to the depth of maximum dose (see Sections 5.2 and
5.3). On the other hand, it should be kept in mind that a thick build -up cap means a larger
perturbation of the treatment field and may jeopardise the dose to the patient if
measurements are performed during many fractions of the treatment course. In addition, if
the same diode is used for different beam qualities (for instance 6 MV and 18 MV), it may
be preferable to use the same build -up cap for both, in order to avoid confusion and
interchange of build -up caps. It follows therefore that the choice of the "optimal" diode
design is a question of each department’s policy for the in vivo dosimetry procedure.
1.1.2 ELECTROMETER
The diode should be connected to a dedicated electrometer with a low input impedance
and low offset voltage. Diode current generated by sources others than radiation is
considered to be leakage current and is not desirable. The leakage current ideally should be
zero. Due to the input offset voltage of the amplifier, however, there is always a small bias
across the diode introducing a small leakage current. An electrometer used together with a
diode requires therefore the offset voltage of the amplifier to be low, 10µV or less. The
leakage current increases with temperature and accumulated dose due to defects in the
diode and it is essential that the electrometer has adequate zero drift compensation and
stabilisation.
1.1.3 SOFTWARE
There is a range of electrometers available for in vivo dosimetry, having greater or lesser
manual adjustment of the input offset and gain for each channel. This type of electrometer
may allow only one gain setting for each channel while more sophisticated ones offer
several separate calibration sets and correction sets with automatic calculation, storage of
factors and zero drift compensations. Thus one detector may be calibrated to be used in
16
several different irradiation conditions. Most of the electrometers offer the possibility to
computer.
More advanced systems are incorporated with the department’s verification system,
simplifying the management of the in vivo dosimetry procedure. Such system provides the
possibility to store all calibration and correction factors for every diode in use. The
measured diode signal is then automatically converted to dose using the treatment field
parameters downloaded from the patient’s data in the verification system. This gives an
immediate "on line" check of the preparation and treatment delivery in the radiotherapy
• Pre-irradiated diodes have in general lower sensitivity. This parameter has to be taken
into account when choosing the electrometer, which must have a sensitivity range that
matches the diode. The manufacturer of the detectors usually also supplies adequate
electrometers.
• The diodes are available in negative and positive polarity and the electrometer has to
be adapted to this.
• The rate of sensitivity degradation will affect the required calibration frequency. In
general, n-type diodes have larger sensitivity degradation but the rate of degra dation
will decrease after a certain amount of pre -irradiation (both for n- and p-type).
Therefore, the pre -irradiation level is of interest and should be stated by the
manufacturer.
• The best choice of diode design i.e. shape and thickness of the build -up cap depends
on the application. A cylindrical cap (uniform directional response around the detector
17
choice (smaller perturbation than the cylindrical cap). If little influence on the
perturbation of the treatment field is desired (measurements in all sessions) a diode with
thin build-up cap is preferable. This however means that larger correction factors have
to be used for accurate measurements, whereas diodes with thicker caps need smaller
• A number of properties of importance for the clinical use of diodes are related to their
dose rate dependence. If diodes are to be used in irradiation situations with large
variations of dose per pulse i.e. wedged fields or treatments at SSD deviating from
calibration SSD, it is advisable to choose diodes with a low dose per pulse dependence,
There are several different types of diodes commercially available having various
properties with regard to pre -irradiation level, doping type, design and thickness of build -
up cap to accommodate a large photon energy range. For accurate in vivo dosimetry it is
essential that each diode characteristic is well understood in order to utilise it properly and
to interpret. Some specifications should be handled with caution: for instance, it is known
that several commercial diodes lack sufficient build -up for the energy range that they are
specified for (see [Jornet 1996], [Georg 1999] and [Meijer 2001]).
Table 1.1 to Table 1.6 present diodes and selected specifications available from leading
companies.
18
Sensitivity 40 40 150-300 150 40
[nC/Gy]
[% / °C]
Linearity < 1%
Table 1.1 Commercially available diodes listed along with the specifications given by
the manufacturers.
19
Model Application / Build-up cap /
60
E5 Co 5 mm
P10 4 - 8 MV 10 mm
P20 8 -16 MV 20 mm
P30 16 - 22 MV tungsten / 30 mm
20
Model Application / Build-up cap /
1114 1 -4 MV aluminium / 10 mm
30-475 6 – 25 MeV
30-471 1 – 4 MV copper / 7 mm
30-472 5 – 11 MV copper / 14 mm
30-474 18 - 25 MV tungsten / 36 mm
21
Model Application / Build-up cap /
1162 1 - 4 MV aluminium / 10 mm
The signal stability of the diode, influenced e.g. by the leakage current without irradiation,
should be checked after adequate warm-up time with the diode connected to the
electrometer and compensated. Compared to the current obtained for the real measurement,
the leakage current should be insignificant. It is advisable to measure the leakage curr ent
for a time period that is at least five times longer than the time period used in the clinical
application. The leakage current should not exceed 1% in one hour [Van Dam 1994].
A general test of the reliability and stability of the equipment, before using it in clinical
routine, can be performed as follows. The diode positioned on top of a calibration phantom
(see Section 1.2.2) is irradiated for 10 to 15 times with the same reference field. The standard
deviation of the resulting signals should be within 0.5 %. The measurements are repeated
on different days during two weeks. The measurement procedure, including the
measurement equipment, the phantom set-up and diode positioning, is reliable and stable,
if all measurements are within 1 % (provided that the beam output of the treatment unit is
stable).
22
Some centres perform more extensive tests before using the diode, for instance a
measurement of the effective water equivalent thickness of the build -up cap. An example of
The diode is calibrated to measure the entrance dose, i.e. when positioned on the skin of
the patient the measured dose should correspond to the dose to tissue at the depth of
maximum dose of the photon quality in use for a particular beam geometry.
The calibration procedure firstly involves the determination of the calibration factor (F cal ).
It is recommended to calibrate the diode for each beam quality with which it is intended to
be used (see Figure 1.2). Due to the variation of the diode signal with accumulated dose,
calibration should be regularly repeated in time. Time intervals typically vary between
weekly and monthly. The temperature dependence of the diode signal can be accounted for
during calibration, if this is performed at the same temperature as the measurements with
that particular diode in the clinical application. Usually, however, a temperature correction
The entrance dose value in a clinical situation is calculated from the diode measurement as
the product of the diode reading, the calibration factor and the correction factors (equation
1). The calibration factor is defined as the ratio of the ion chamber dose and the diode
D
Fcal = ic (2)
Rdiode ref . condition
23
The diode may be calibrated against the dose monitor chamber of the accelerator or against
a secondary reference chamber. To determine the calibration factor, the diode is positioned
on the surface of a suitable (plastic) calibration phantom (e.g. made of polystyrene). The
ion chamber is inside the phantom on the central axis, at reference depth. According to the
definition of entrance dose, this should be the depth of maximum dose, as indicated in
Figure 1.3. The ion chamber is thus probing the depth dose curve at its maximum, and not
at its subsequent fall-off. As a consequence, if the protocol that is used for absolute dose
SSD
Reference distance Diode
Ion chamber
Depth of d max
Reference field size
Solid phantom
Figure 1.3 Diode calibration procedure for entrance dose measurements. The ionisation
chamber is positioned at the reference depth in the phantom and the diode at
24
determination with the ionisation chamber includes the application of a displacement
factor, this factor should be omitted. If a plastic phantom is used that is not completely
dose-to-water should be employed. The reference SSD is usually 100 cm (for linacs) and
The calibration may be performed with one or several diodes placed in a circle around the
central axis provided that variations in the field flatness are insignificant. Field flatness at
d max should therefore be checked, for instance by measuring the ratio of diode readings at
the circle and at the centre of the field. Furthermore, the diodes should be placed at a
distance from the central axis to avoid perturbation of the beam at the reference chamber.
Subsequent to the determination of the calibration factor, a set of correction factors has to
be established to account for variations in diode response in situations deviating from the
reference conditions (see Figure 1.2). The ultimate factors influencing the diode response
are the field size, source-to-skin distance, presence of beam modifiers such as filters or
wedges, presence of tray and blocks and the beam incident angle (equation 3). As
described in Section 1.1.1, the dependence of the diode signal on most of these factors is
not only arising from the intrinsic properties of the diode crystal, but also from elementary
beam physics, i.e. the fact that the detector experiences different scatter contributions than
the ones experienced at the depth of maximum dose. As a consequence, most of the
correction factors are inherent to the use of dose detectors taped to the patient’s skin, and
the patient measurement is assessed before the diode has reached thermal equilibrium (2 -3
minutes) the influence of temperature dependence may be neglected (see Sections 5.3.3
25
and 5.5.1). Diodes used for TBI, for which the low dose rate is achieved by enlarging the
Correction factors accounting for the variations in response are determined as the ratio of
the reading of an ionisation chamber and the reading of the diode for a clinical irradiation
situation normalised to the same ratio for the refe rence situation (equation 4):
The reference conditions are as stated in Section 1.2.2. However, if the value of a particular
parameter (for instance the field size) influences the value of the correction factor for a
second parameter (e.g. the presence of a tray), the ‘reference condition’ for the
determination of this second correction factor is adapted in order to avoid double inclusion
of the first correction factor. This is made clear in the practical recommendations given
below. These may be modified/simplified according to the type of diode (and previous
experience with that particular type of diode), the clinical application, and the beam quality
in use:
• The variation in response due to different beam incident angles is measured for
different gantry and couch angles and normalised to the response measured when the
• Field size correction factors are measured for square fields ranging e.g. from 5 x 5 cm2 to
• SSD correction factors are measured for SSDs within a range determined by local
clinical conditions, for instance from 75 cm to 110 cm, at the reference field of 10 x 10
cm2. Note that SSD correction factors and field size correction factors are assumed to be
independent. This is not always the case for high energies (see remark below and
[Georg 1999]).
26
• Wedge correction factors may depend on the field size. They are measured at reference
SSD, for different square fields (e.g. fields with 5 cm, 10 cm and 20 cm side length). The
ratio of the signal of the ionisation chamber to the diode signal is in this case
normalised to the same ratio for the open beam (with the same field size).
• Tray correction factors may depend on SSD and field size. They can be determined by
repeating all measurements carried out for the SSD and field size correction factors, and
normalising the data to the reference situation of an open beam with the appropriate
• Block correction factors can be measured for different blocks defining square fields at a
fixed collimator opening (for instance a collimator opening of 20 x 20 cm2 for blocks
defining fields of 5 x 5 cm2, 10 x 10 cm2 and 15 x 15 cm2). The reference condition is again
Practical examples of calibration procedures and typical values of correctio n factors for
particular types of diodes are given in Section 5.1. In order to minimise redundant use of
correction factors, minimum values can be set below which factors are discarded (for
instance if a correction factor deviates less than 1 % from 1, the correction is within the
measurement uncertainty). Other ways of limiting the use of correction factors are
If the dosimetric characteristics of a diode are not (well) known, it is recommended to check
its response extensively at different irradiation conditions to establish the range where no
correction factors are needed. As the type of diode is a major determinant of the magnitude
and the behaviour of most of the correction factors, diodes of the same type will require
similar correction factors, showing similar tendencies. However, when a high accuracy is
required, it is advisable to check also the correction factors for every individual diode.
Correction factors associated with increased diode sensitivity due to variation in beam
energy spectrum are of major importance in high-energy photon beams, especially if diodes
with a thin build-up cap are used. One should also bear in mind that, for insufficient build-
27
up, other interdependencies of correction factors than the ones mentioned above may exist
[Georg 1999]. In practice, one can start by considering all correction factors to be
independent and then check the accuracy of the measured d ose when changing more than
one reference condition at the same time (i.e. field size, SSD and wedge). Useful information
It is good practice to keep a record of the change in the calibration factor in order to
estimate how often re -calibration will be required to achieve a certain accuracy. As the
sensitivity degradation may vary with different beam qualities this is especially important
when diodes are used in various beam qualities. It is advisable to start with weekly
calibrations and to adjust the calibration interval after having monitored the accumulated
dose in between calibrations and the corresponding change in calibration factor for a while.
Depending on the diode type in use the correction factors associated with the dose per
pulse dependence may also change with time due to the accumulated dose. A quick and
efficient test of the long-term stability is to perform a linearity check by measuring the
diode response normalised to an ionisation chamber at two different SSDs. If the ratio is as
expected, the diode is working accurately and the correction factors are still valid. The
change of the temperature dependence with time is accounted for if the diode is calibrated
28
CHAPTER 2 IMPLEMENTATION OF THE MEASUREMENT
PROCEDURE IN CLINICAL PRACTICE
Regarding the workload associated with routine in viv o dose measurements, two
categories of work can be distinguished: the calibration procedures and the actual patient
procedures can either be carried out by a small team of qualified personnel, or assigned to
different groups of personnel within the department (for instance calibration procedures
radiographers/nurses at the treatment units). In the latter case, one person or a small group
of persons have the responsibility for the in vivo dosimetry program and train the others.
The person(s) responsible for the in vivo dosimetry program initiates the implementation of
it. First, he/she gets acquainted with the theoretical background of in vivo dosimetry,
available in literature (see appendix), with the equipment used in the department
(electrometer and diodes) and with the calibration techniques using the (plastic) calibration
phantom. The test of the reliability and stability of the equipment (see Section 1.2.1) is
performed. After the reliability test, the electrometer is calibrated following the instructions
in the manual, and the calibration and correction factors of the diodes are determined as
Another task is the practical training for the personnel performing the measurements. The
importance of the accurate positioning of the diode in the centre of the treatment field is
for which the positioning of the diode is critical. Between each irradiation the diode is
removed and repositioned. The readings of the electrometer should be within 1.5 %.
29
If the beam axis of the treatment field is covered by a shielding block or in case of an
asymmetric field, the penumbra region should be avoided by positioning the diode as close
as possible to the field centre and at a similar SSD. If it is a wedged field, the actual
attenuation of the wedge at the off-axis position of the diode should also be considered.
Before starting patient measurements, it is useful to simulate some patient set-ups with a
phantom. The irradiations and diode measurements are performed in identical conditions as
in the clinical situation. The expected signal is calculated, either with an independent
formula or with a treatment planning system (TPS) able to calculate dose at dmax . The
difference between the calculated and the measured signal should not exceed 1 %. These
patient simulations are an excellent test for the whole measuring procedure: calibration of
the diode and determination of correction factors, calculation of the expected dose, and
diode positioning.
Patient measurements should be started for treatment fields where easy fixation of the
diode in the field centre is possible, for instance mediastinal or large head and neck fields
without wedge. When the deviations between measured and expected signal are smaller
than 3% to 5%, measurements for other treatments like breast or pelvic irradiations can be
initiated.
In the course of the training period, tolerance and/or action levels have to be established.
Since a measurement result out of the tolerance window triggers the chain of measurement
regularly (see Section 3.1), and especially during the training period. In this period it is also
measurements on the same patient during consecutive sessions (at least 5). The mean
30
value, the standard deviation and the deviation of each individual measurement are
evaluated. A small standard deviation is a strong argument for considering the value of
the first measurement as being representative for the whole treatment. This evaluation
MEASUREMENTS
immediate actions to be taken when the measured entrance dose is out of the tolerance
and/or action levels. These guidelines will differ among the radiotherapy departments
depending on the choice of the general philosophy for in vivo dosimetry (which patients,
which fraction, which treatment sites etc.), the education level of radiation technologists,
the existence of a Quality Assurance group and/or the involvement of the physics
department.
Practical guidelines towards the radiation technologists, assuming that they are in charge
• if the measured entrance dose exceeds the tolerance or action levels, what
• if there is a difference between the stated SSD and the actual SDD (source-diode
distance), what should be done i) for isocentric techniques or ii) when using bolus or
Other questions regarding staffing and management of personnel should also be clarified:
• who is the contact person for measurements out of tolerance or action levels?
physicist?
31
• who will perform phantom measurements, if needed?
• who is in charge of the calibration and the determination of correction factors of the
diodes?
Examples of guideline flowcharts, including actions undertaken at different levels, are given
in Figure 2.1 and Figure 2.2; more examples are given in the questionnaire of Chapter 4. The
possible origin of errors and the actions undertaken are discussed in more detail in
Sections 3.2 and 3.3. Typically, the investigation of an error is performed in two steps.
Because of the on-line read-out, the first action can be triggered instantaneously by the
technical staff, who performs an immediate check on the spot. If the origin of the error is
START: correct
entrance measurement yes
yes
out-of-tolerance
yes signal for most of the no
patients that day?
check accelerator no
output
stop error on
yes error found? yes
treatments dosimetry chart?
no
no
stop treatments;
discussion
error in MU
yes
calculation?
no
phantom simulation
check diode
correction factors yes is treatment OK?
OK
no
stop treatment;
discussion
32
Figure 2.1 Example of a flowchart (taken from Barcelona), guiding the actions to be
33
START: treatment plan
entrance measurement yes modification
wedged field
yes yes or IMN (critical no
diode position)?
signal within
action level 2? no
yes
second measurement
“a posteriori” check:
signal
no data transfer check;
within tolerance?
recalculation expected signal
yes
treatment
plan/data transfer yes
error?
no
wrong SDD,
yes wrong calculated
signal?
no
new measurement
by QA personnel
no
phantom simulation
Figure 2.2 Example of a flowchart (taken from Leuven), guiding the actions to be
34
2.3 RECORDING OF IN VIVO DOSIMETRY
Recording of the in vivo entrance dose may be done on a treatment chart, on a separate
sheet for QA and/or in a database accessible in a network (possibly linked to the R & V
system1). The results should be easily available (after the first treatment session, during
It is important to record in vivo dosimetry data together with sufficient information, such as
the date of measurement, the type of field, the treatment unit, the anatomical location and
so on. The more complete the database is, the more information can be derived when
1
Some R & V systems offer at present the possibility to record or enter manually the
35
CHAPTER 3 INTERPRETATION OF THE MEASUREMENT
The choice of tolerance/action levels is very important since they will in practice determine
the number of "errors" detected and will influence the associated workload t o implement or
tolerance window is adopted, some causes of erroneous treatment delivery may not be
detected (for instance a wedge 30° instead of a wedge 15°, presence of a tray etc.). If the
tolerance window is too small, a too large number of measurements will have to be repeated
(due to e.g. inherent statistical fluctuation or a too critical positioning of the diode in e.g.
wedged beams). Clearly, the value and the meaning of the levels are related to the
philosophy of the department regarding in vivo dosimetry. Some centres using in vivo
dosimetry as a routine check for every patient, distinguish the first level, the tolerance
level, from higher action levels. A deviation of the diode signal beyond the tolerance level,
but within the action levels, is considered as a warning, linked to a very limited action.
When in vivo dosimetry is used to check particular treatments, the value of the levels can
vary according to the treatment type. Treatments with high dose - high precision
techniques require narrow tolerance windows, while other treatments have less stringent
levels for patients treated with palliative intent in order to minimise the number of second
measurements, paying more attention to the patients treated with curative or adjuvant
intent.
The determination of the actual value of the level is based on different factors, first of all on
the uncertainty of the diode measurement method. According to Essers and Mijnheer
[Essers 1999], the theoretical uncertainty in measuring the entrance dose with diodes,
taking into consideration the uncertainties in the calibration factor, the correction factors
and the positioning of the diode, is 1.6 % (1 standard deviation (SD)). This means that
36
without any additional cause for deviation or error, 68% of the measurements would be
within 1.6 % and 95 % of the measurements would be within 3.2 % (2 SD) of the expected
dose. This seems to be in agreement with other results reported in the literature, although
such a level of accuracy is probably difficult to obtain for all types of irradiation, for
instance for treatments with tangential wedged beams. It also has to be stressed that this
high level of accuracy is attainable only if a very accurate estimate of diode correction
factors, which could be preferable in small and medium centres with a small physics staff,
means a larger uncertainty in dose estimation and, consequently, results in the necessity of
Other sources of uncertainty, which should be taken into account when choosing the
levels, are:
patient during irradiation; the difficulty in firmly attaching the diode in some regions
• the use of ancillary equipment to set-up the patient (i.e.: head masks, head-rest,
immobilisation shells…) [Essers 1994]. In these cases the diode reading has to be
corrected to take the real SDD into account, which can be difficult to assess in some
situations. Moreover, some loss of backscatter may occur in many situations, which is
another source of uncertainty, as this is usually not taken into account by the TPS.
Also, it should be kept in mind that the positioning of the diode on immobilisation
shells (or on the back of the couch when treating with dorsal fields) results in a larger
uncertainty if the temperature dependence of the diode signal is accounted for in the
• the true SSD, if the diode reading is not corrected to take into account the difference
between the true SSD and the planned one. If the correction is performed, in some
37
• the use of asymmetric fields, e.g. for tangential breast treatments (see Section 5.4.2)
As the in vivo measured entrance dose has to be compared with the expected one, which is
calculated by the TPS or by an independent formula, the uncertainty in the entrance dose
calculation is another factor that should be taken into account. This uncertainty depends
on:
• the way in which treatment unit data have been acquired (for instance the precision
The majority of the radiotherapy centres have a 5 % action level for most treatments (see
Chapter 4 and Sections 5.4). The tolerance level usually coincides with this level, according
to the philosophy that any deviation larger than 5 % must be investigated. A procedure
recommended for establishing the narrow tolerance window required for high dose - high
particular technique and take twice the SD of the measurement uncertainty as the
tolerance/action level ([Essers 1993], [Essers 1994], [Lanson 1999]). In this case, however,
entrance dose measurements are usually combined with exit dose measurements to obtain
Once the tolerance and action levels have been established, the range of acceptable
variation of some of the parameters can be determined in order to facilitate the search for
measured SSD (or SDD) for isocentric and fixed SSD techniques can be determined. The
38
It is important to verify during a certain period whether the tolerance/action levels are
adequate for clinical routine. An important indicator is the rate of second measureme nts,
which is strictly related to the action level. A too small rate (for instance less than 2-3%)
should be regarded with caution because it might indicate that the action level is too high.
Inversely a too high second check rate (for instance larger than 15-20 %) could mean that
the level is too small. In particular a high rate of second checks can generate distrust
concerning the real usefulness of in vivo dosimetry among the operators and the medical
staff. An alternative method for adjusting the tolera nce/action levels is to adapt it to the SD
of the measurements. This parameter can be determined by pooling the patients for a
certain period.
It is clear that the continuous monitoring of systematic in vivo dosimetry after its
implementation is mandatory in order to reduce the errors of the control process and
Section 3.3.3). Such a monitoring could also help defining differe nt tolerance/action levels
for different types of patient treatments and/or beam set-up. For instance, it could become
clear whether wedged beams need higher tolerance/action levels with respect to unwedged
It is important to keep in mind that when a deviation is observed out of the tolerance level,
measuring procedure at a departmental level can affect the confidence in in vivo dosimetry.
If, for instance, during a chart round a radiation oncologist finds out that most of the in
39
vivo measurements are out of tolerance (due to a problem in the QC process), it is rather
difficult to yet convince him that his patients are indeed correctly treated, and/or that in
Malfunctioning of the QC process (Figure 3.1), may be present either at the departmental
level, leading to systematic errors (i.e. for all patients), or at the individual level. Systematic
errors are typically errors (or a shift) in the calibration factors of the diodes or an error in
the correction factors (or omitting some necessary correction factors). Systematic errors
may also be induced by erroneous calculation (with or without TPS) of the entrance dose.
may increase the workload since one might start to look for a “real” dosimetric error and/or
one might request a new measurement to exclude other “individual” errors in the QC chain.
The individual errors in the QC chain for entrance dose are mainly the following ones:
• miscalculation of the expected diode signal from the entrance dose (use of wrong
• bad positioning of the diode: not in the centre, too close to shielding blocks, etc. (cfr.
Section 2.1)
systematic
determine determine convert
calibration correction read-out
factor factors signal/dose
• • • •
•
individual
• •
calculate exp.
position read-out
read-out record SSD
diode electrometer
signal/dose
accessories shielding
wedge
in the beam blocks
40
Figure 3.1 Schematic representation of error-prone steps in the quality control process
To facilitate the analysis of possible errors in the treatment chain, dosimetric errors are
equipment breakdown
treatment unit:
simulator TPS R&V
treatment machine + table
• • • • •
simulator treatment unit patient
set-up set-up
R&V
TPS
prescription
human errors
In Figure 3.2 a compact scheme is given of the radiotherapy process from prescription to
delivery. Each arrow in the diagram represents a transfer of data, which is error prone, and
every box may generate erroneous data. Depending on the organisation of the department
41
and on the possibilities of the available equipment, the practical information transferred
between the boxes may be different. At the end of the chain, before treatment delivery,
parameters are either recorded in a Record and Verify system (R & V) or written on the
treatment chart.
An erroneous transfer of prescribed dose from prescription to TPS can only be detected by
in vivo dosimetry if the entrance dose is calculated by hand or with an independent "home
made" system to predict the expected diode signal. If the planning target dose is - by
mistake - different from the prescribed target dose, the entrance dose calculated (manually)
with the prescribed dose will differ from the entrance dose calculated with the TPS, and
hence will be detected. The dark boxes in the last column of Figure 3.3 represent the other
parameters for
gantry angle
collimator
angle
field size
simulation
modality/
parameters
energy
MUs
(dose/fraction) R&V
TPS
and/or
wedge
treatment chart
shielding
prescription
blocks
parameters
table
parameters
SSD
immobilisation
device
bolus
42
Figure 3.3 Overview of the parameters of the treatment preparation process
which errors can be detected by entrance in vivo dosimetry, whether or not the TPS is used
wedge
for heavy wedges can occur). A discrepancy in the choice of the wedge (for
instance 45° instead of 30°) will also be detected by entrance in vivo dosimetry.
However, errors in the orientation of the wedge will not be detected by entrance in
Since the absorption rate of a tray holder is usually a few percent, its presence or
individual compensators
From a theoretical point of view, measuring the entrance dose is not a conclusive
both the response of the diode and the dose rate of the various beams may be
reported for instance by Essers et al. [Essers 1999] for a dual energy linac.
43
Depending on the magnitude of the discrepancy between the prescribed field size and the
actual field size, the measured in vivo entrance dose theoretically shows an error in the
treatment delivery. In practice this error is so small that errors in field size are normally not
calculation) on the treatment chart and/or the R&V system should always be performed
[AAPM 1994]. It has been demonstrated that this simple tool significantly reduces the
incidence of human erro rs. However, even with this systematic check, in vivo dosimetry is
delivery: the simulator, the TPS, the R & V and the treatment unit (encompassing the
treatment unit itself, the treatment couch, fixation devices etc.). While a breakdown or
breakdown of the treatment unit and more specifically large variations in beam output can
easily be detected by in vivo entrance dose measurements. A typical other source of errors
used new software or a software upgrade for the calculation of MU ([Leunens 1993],
[Lanson 1999]). As far as the R & V system is concerned, malfunctioning is potentially very
dangerous if the system is also used for setting up the patient, and can be discovered by in
vivo dosimetry.
44
3.2.2.3 DISCREPANCIES IN PATIENT POSITIONING/GEOMETRY BETWEEN
Deviations in patient set-up at the time of treatment delivery can be due to random human
errors (especially if the treatment couch parameters are not verified by the R & V system
[Leunens 1993]) or to systematic machine-related errors (like a bad resetting of the “zero”
indicator of the simulator couch (Section 5.4 and [Fiorino 2000]), in which cases they
belong to the two previous categories of errors. Other sources of a wrong positioning,
however, are patient motion or a change in patient thickness due to swelling or shrinkage.
While entrance dose measurements give only direct information about the patient set -up
with respect to the beam (in particular an incorrect SSD), they also result in the detection of
patient thickness errors, if patient thickness is reassessed to trace the origin of a large
incorrect SSD.
If the result of the first measurement is outside the tolerance/action level, a number of
controls should be activated (see Figure 2.2). This chain of controls involves checks in
order to reveal either quality control process errors (e.g. the immediate check of the diode
position, or the “a posteriori” recalculation of the expected signal) or real treatment process
errors (e.g. the immediate check of the patient position, or the data transfer control in the “a
posteriori” check).
First, an immediate check (i.e. with the patient on the treatment couch) of the treatment set -
up and treatment parameters must be performed. The most common errors are differences in
SSD and wrong positio ning of the diode, which can both be checked by the radiographers
on the spot. Differences in SSD due to the use of immobilisation devices or bolus should
be corrected by the appropriate inverse square law correction factor; in other cases the
SSD deviation signifies a real treatment process error (see Section 5.4). An evaluation of
45
the correct positioning of the diode in the field centre can still be made afterwards on a
portal image (for treatment geometry verification), if this was taken simultaneously with the
performed, possibly the same day as the in vivo control. The “a posteriori” check should
concern the data transfer control, checking the congruence of all technical and dosimetry
data of the treatment planning/simulator chart to the corresponding data on the treatment
chart (and/or R&V system). It must include the agreement between the effectively delivered
MU against the planned ones, the correctness of MU calculation and the correct use of
wedges and blocks. If the in vivo dose with blocked fields is lower than the expected one,
the diode position should be checked especially if the block is near the irradiation field
centre. In any case, it is a good practice to perform a second in vivo control check, also if
Tolerance levels generally coincide with action levels in most institutions. If t his is not the
case, performing merely a second in vivo dose measurement could be a limited, time -saving
action related to an out-of-tolerance deviation within the action level. A similar approach
can be followed if two different action levels are defined (“low” and “high”, for instance: 5
and 10 %). If the measurement is outside the “low” level, but still within the “high” action
level, the immediate check may for instance be avoided and just an “a posteriori” check
A number of papers report that a lot of deviations exceeding the action level might not be
related to proven human errors or errors due to equipment breakdown ([Cozzi 1998],
[Essers 1999], [Heukelom 1991a], [Leunens 1990a], [Leunens 1991], [Loncol 1996], [Mangili
1999], [Millwater 1998], [Mitine 1991], [Noel 1995], [Voordeckers 1998]). On the other side a
second measurement reduces the probability of some quality control process errors such
46
as bad diode positioning. So, in the chain of control checks following in vivo dosimetry,
some deviations might not be attributed to a treatment process error nor to a quality
control process error. For such a situation the data should be discussed by the
In order to facilitate the search for the cause of the persistent deviation, an entrance dose
measurement with the diode and an ionisation chamber in a solid phantom in the same
clinical treatment conditions (SSD, field size , gantry and collimator rotation, block,
wedge…) may be useful. This is particularly true during the first phases of implementation
of systematic in vivo dosimetry procedures, when the accuracy of in vivo dosimetry in the
various clinical situations may not fully be assessed. If the deviation between the phantom
entrance dose measured with the diode and the calculated one is acceptable, the in vivo
deviation could be attributed to a difficulty in the diode positioning on the patient’s skin. If
the dose meas ured with the ionisation chamber is in agreement with the expected one, but
the diode reading is not correct, the deviation of the diode signal could be explained by a
bad calibration or wrong correction factors. If the phantom entrance dose measured with
the ionisation chamber is not in agreement with the expected dose, dose calculation
• difficulties in setting up the patient: these are more likely to be detected if diode
When the cause of the discrepancy is identified, an action may be required for the single
patient such as checking patient positioning at the simulator and/or checking the patient’s
(for instance wedged fields), more accurate assessment of diode correction factors or
47
further investigations on the accuracy of dose calculation may be required. After reviewing
a large set of data, a high rate of second checks/persistent deviations for a certain
configuration of fields may also suggest a modification of the tolerance/action level for
such a category.
continuously monitor the adequacy and the efficacy of the system. A periodic review of
the database of in vivo dosimetry data with some statistical analysis is very useful to drive
the physicist and the clinician in adjusting the procedures to the real local conditions. An
important goal is the verification of the adequacy of tolerance/action levels: a too high rate
of second checks may have a negative impact on the operators and efforts should be made
Continuous monitoring of in vivo dosimetry data may therefore indicate the need for
Statistical analysis of the deviations between expected and measured entrance dose may
of the quality control process (for instance, more accurate assessment of diode correction
Although some errors or inaccuracies may also be detected on an individual basis, they
will be clearer with a statistical approach because of the existence of fluctuations in the
measuring procedures. Subgroups of patients can be pooled for instance breast, hea d &
neck, brain, etc. The distribution of the deviations has been shown to reveal systematic
examples concerning large cohorts of patients are given in the literature [Noel 1995],
[Leunens 1990a], [Leunens 1990b], [Leunens 1991] and [Fiorino 2000] and in the single
48
interesting and useful type of evaluation also requires considerable manpower, if no
49
CHAPTER 4 TECHNIQUES AND PROCEDURES IN DIFFERENT
RADIOTHERAPY CENTRES
questionnaire that has been sent to centres having experience with routine in vivo dose
measurements. Apart from the institutions which have co-operated for this booklet, the
‘Nederlands Kanker Instituut’ from Amsterdam and the ‘Centre Alexis Vautrin’ from Nancy
have provided information (see also [Essers 1993, 1994, 1995a and 1995b], [Heukelom
1991a, 1991b, 1992, 1994], [Lanson 1999] for Amsterdam and [Noel 1995] for Nancy).
4.1 WHAT EQUIPMENT DO YOU USE TO CARRY OUT ROUTINE IN VIVO DOSE
MEASUREMENTS?
LEUVEN BARCELONA
(Saturne 40, 42, GE, Clinac 2100, Varian) (Clinac 1800, Varian)
(Scanditronix)
- Diodes
- Diodes (Scanditronix) 6 MV EDP-10 (Scanditronix)
6 MV EDP-20, EDP-20+1 mm Cu 18 MV EDP-30 (Scanditronix)
18 MV EDP-20, EDP-20+1 mm Cu QED 1116 (Sun Nuclear)
TBI EDE P30 (Precitron)
Isorad-p 1164 (Sun Nuclear)
electrons EDD-2 (Scanditronix)
TBI EDP-30 (Scanditronix)
50
NANCY COPENHAGEN
(Scanditronix)
- Diodes - Diodes
p-type diodes except for two n -type 4 MV P10 (Precitron),
TBI
51
AMSTERDAM MILANO
EDINBURGH
Irradiation equipment
X-rays 6, 8, 9, 15, 16 MV / e- 5-20 MeV
(Varian 600, 600CD, ABB CH6, CH20, RDL Dynaray 10)
Diodes (Scanditronix)
6 MV: EDP-10, EDP-10 + 0.6 mm brass
8,9 MV: EDP-20
15, 16 MV: EDP-20, EDP-20 + 1.2 mm brass, copper
electrons EDE, EDD-5, EDD-2
Mounting: home -built quick-swing ceiling mounted system; being rolled out to all
52
4.2 PHILOSOPHY OF YOUR DEPARTMENT CONCERNING THE USE OF IN VIVO
DOSIMETRY?
LEUVEN BARCELONA
- for every patient at first treatment session - for every patient treatment at the second
NANCY COPENHAGEN
- for every patient at the second or third - intention of including every patient
AMSTERDAM MILANO
- for two special treatment techniques with - for every patient, within the first week of
measurements are performed during two take place (reduction of field size, blo cks,
53
EDINBURGH
- all treatment machines and treatment techniques are systematically checked in detail on
sufficient patient numbers to give a statistically valid study. From this systematic
deviations are identified and corrected, random deviations are quantified; decisions are
changes in treatment. Currently not on all machines; this is being rolled out to all
technique, which repeat the initial pilot studies on a small group of patients
- electron measurements ju st beginning
LEUVEN BARCELONA
dose
- for electrons: entrance dose measurements
54
NANCY COPENHAGEN
AMSTERDAM MILANO
correction
55
EDINBURGH
- for initial systematic studies, entrance and exit doses, where possible; typically at the
centre of the field. From these values, target absorbed dose deviations are estimated
- for routine use, typically entrance doses only
- for breast patients, combined entrance and exit doses are measured routinely at a point
LEUVEN BARCELONA
Equipment Equipment
- polystyrene phantom - polystyrene phantom
- Plastic WaterTM phantom (CIRS)
- water phantom equipped with thermostat
Calibration Calibration
- ref. conditions: SSD 100 cm, - ref. conditions: SSD 100 cm, 22.5°C
2
field size 10 x 10 cm field size 10 x 10 cm2
- for absolute dose determination with the - for absolute dose determination with the
ionisation chamber, the Dutch (NCS) ionisation chamber, the Spanish protocol
displacement factor (see Section 1.2) factor (for calibration of entrance dose
- TBI: calibration factors in TBI conditions measurements) (see Section 5.1.2)
- TBI: entrance and exit calibration factors in
TBI conditions
directional dependence
- exit: none (< 1%)
57
NANCY COPENHAGEN
Equipment Equipment
- polystyrene phantom - Solid WaterTM phantom ( RMI 457)
Calibration Calibration
- ref. conditions: SSD 100 cm, - initially against NE Farmer chamber
SSD 60Co 80 cm - periodic calibrations: against the Clinac
2
field size 10 x 10 cm monitor chamber in connection with
- TBI: entrance calibration factors in TBI output check of the treatment machine
conditions Correction for
Correction for - field size, SSD, tray, layers of
- variation of response of exit detector with compensation filter, temperature and
dose rate directional dependence (no wedge
- wedge correction fa ctor correction because of dynamic wedges)
AMSTERDAM MILANO
Equipment Equipment
- polystyrene phantom - acrylic phantom
Calibration Calibration
- ref. conditions: SSD 100 cm - ref. conditions: SSD 100 cm,
field size 15 x 15 cm2 field size 10 x 10 cm2
15 cm thick phantom - TBI: treatment conditions
- Dutch (NCS) protocol without
displacement factor
EDINBURGH
Equipment
- Solid WaterTM phantom (RMI 457)
Calibration
- ref. conditions, 100 SSD, 15 x 15 cm2 field, 15 cm thick phantom
- absolute dose determination using calibrated ionisation chamber using UK protocol:
for entrance dose calibration, take out the displacement correction
for exit dose calibration, add in an average ‘build -down’ correction
Correction for
- measured for every parameter: e.g. SSD, field size, phantom/patient thickness, directional
dependence, temperature, wedge, tray, (for both entrance and exit initially)
- build-up caps used on diodes to minimise the range of values of correction factors
- detailed correction factors used for initial systematic studies and for audits
- for routine use, mid-range correction factors for the irradiation parameters used for a
specific technique and treatment machine are combined into ‘generic’ correction factors
LEUVEN BARCELONA
- the expected entrance dose is the dose - the expected doses (entrance, exit and
the TPS;
- for electrons, the expected dose is the
59
NANCY COPENHAGEN
- the expected dose is the dose calculated at - the expected entrance dose is the dose
the ICRU dose specification point, or the calculated either with an independent
distribution available
AMSTERDAM MILANO
- the exp ected dose is the target absorbed - the expected entrance dose is calculated
EDINBURGH
- the expected entrance dose is either that from the TPS, or calculated manually,
depending on treatment
- the expected exit dose is from the TPS, or calculated manually depending on treatment
- the expected target volume dose is that calculated at the specification point, or the
measurement point
- for routine use, expected diode reading ranges (expected reading is expected dose
divided by calibration factor and by generic correction factor) are supplied to the
LEUVEN BARCELONA
Actions Actions
Cfr. Flow chart of Figure 2.2 immediate action: ∆ ≥ 5 %
- radiographer
- immediate check of treatment parameters
movement of diode)
a posteriori action ∆ ≥ 5 %
- check of all parameters
- IVD at the next treatment session
- if ∆ ≥ 5 % persists: simulation of treatment
61
NANCY COPENHAGEN
Actions Actions
session
∆ ≥ 10 %
request for IVD at the next treatment
Actions Actions
one measurement, and the average, is parameters (including SSDs): the operator
- if deviation persists: patient dose (MUs) is radiographer who sets up the patient
calculations with the TPS - check of all parameters (data transfer, dose
and MU calculation)
- IVD at the next treatment session
if ∆ ≥ 5 % persists, possible measurement
Actions
- immediate check of treatment parameters (radiographer and physicist)
- check that not significantly non-standard, such that generic correction may not apply
(physicist)
- before next treatment check plan, MU calculation, treatment record, treatment data,
etc. (physicist)
- carry out repeat diode measurement on next treatment fraction, check diode position,
64
4.3.4 TIME PERIOD BETWEEN CHECKS OF CALIBRATION AND CORRECTION
FACTORS
LEUVEN BARCELONA
Calibration Calibration
- every month - every 50 Gy of accumulated dose
- TBI: every four TBIs
- electrons: in evaluation
NANCY COPENHAGEN
Calibration Calibration
- once a week - every third month
- TBI: before every first session Correction factors (in evaluation)
AMSTERDAM MILANO
Calibration Calibration
- every two weeks, depending on the - every month
EDINBURGH
65
Calibration
test calibration quarterly; but beginning to use diodes as routine daily treatment machine
dose consistency check. In this case, checked versus ion chamber weekly
Correction factors
4.4 WHAT SYSTEM DO YOU USE FOR THE RECORDING OF IN VIVO DOSE
MEASUREMENTS?
LEUVEN BARCELONA
expected doses;
66
NANCY COPENHAGEN
the treatment chart (relevant information and relevant beam data in a database for
- treatment technique
- beam geometry
- wedge
- immobilisation technique
- ratio meas./calc. entrance dose for
each field
- ratio meas./calc. target absorbed
software
AMSTERDAM MILANO
- the diode measurement system prepares a - manual recording of the in vivo dosimetry
diode measurement file for every patient results with a number of relevant
field, containing beam, patient and diode information (see Section 5.5)
parameters (for instance calibration and - periodic update of Excel files for statistical
correction factors are determined for each analysis
67
diode using look-up tables and simple analysis
gaps)
on hard disk
68
EDINBURGH
- for systematic measurements and for full audits, manual recording of results on
group
LEUVEN
NANCY
70
COPENHAGEN
AMSTERDAM
- administration 1 hour
71
- consultation with co-workers 2 hours
MILANO
EDINBURGH
year/diode
-* methodology designed so that this is in parallel with other tasks, i.e. adds minimal
LEUVEN BARCELONA
errors in TPS target dose, since the and of correction factors for midplane
thickness?
- non-symmetric heterogeneities
NANCY COPENHAGEN
73
- near blocks - mounting of equipment:
- small fields (exit diode) convenient handling of the diodes and the
MV)
AMSTERDAM MILANO
all encountered problems turned out to be action level for tangential wedged beams
real problems with the treatment planning modified with time after statistical analysis
accelerator
EDINBURGH
etc.
- what temperature correction should be applied in certain situations
- measurements below bolus
- measurements close to blocks (particularly CRT blocks), asymmetric fields, etc.
- limitation of resolution of electrometer in small reading situations, particularly for small
74
75
CHAPTER 5 EXPERIENCES FROM DIFFERENT RADIOTHERAPY
CENTRES
In order to provide more detailed examples regarding the implementation and the
functioning of in vivo dosimetry in clinical routine, we have selected contributions from the
authors’ institutions about an aspect of in vivo dosimetry that they have worked on
specifically. Some contributions offer reference data concerning basic procedures, from
for improvement or refinement of procedures. More details and data can be found in [Jornet
This section lists detailed examples of methods and results of the calibra tion procedures,
as explained in Section 1.2, performed in the radiotherapy department of the Hospital Santa
Creu I Sant Pau in Barcelona. A recent study [Jornet 2000] concentrates on the
performance of p-type and n-type diodes in high energy beams, which will be elucidated in
Due to the way diodes are made, two diodes even from the same fabrication batch may
before using them in routine. The results of these tests should be compared with the
The tests performed whenever a new diode is received in our centre are:
2. Intrinsic precision
76
4. Verification of the equivalent water depth of the measuring point (water
Some tests (4 and 5) are only performed for the first 3 or 4 diodes of a particular type. All
diodes are connected to the same channel of the electrometer to avoid drifts and loss of
signal, which depends on the channel to which they are connected. All channels of the
electrometer are checked regularly. The measurements corresponding to these tests are
the isocentre). For most of the tests the diode is fixed on the surface of a plastic phantom,
The results of the tests for different diodes are summarised in Table 5.1.
irradiation (5 min)
(SD)
(10 irradiations)
response/dose (r 2) (0.2 - 7 Gy) (0.2 – 7 Gy) (0.2 - 3.5 Gy) (0.2 - 7 Gy) (0.2 – 7 Gy)
measuring point
(water equivalent
depth)
cm depth
1
designed with a cylindrical build -up cap
77
Table 5.1 Overview of the results of the initial tests for different types of diodes
The signal immediately aft er irradiation is compared to the signal five minutes after the end
of the irradiation. Five minutes is considered as the average of the time periods
We verify that the response is linearly proportional to the absorbed dose for clinical
significant doses. As we verify the linearity between MU and dose regularly, we verify the
POINT
The diode is fixed on the surface of the Plastic WaterTM phantom and covered with a
special Plastic WaterTM slab to avoid air gaps (Figure 5.1). Irradiations with X-rays are
performed while adding Plastic WaterTM slabs on top of the diode until the reading reaches
known, the water equivalent thickness of the build -up cap can be calculated (Figure 5.2).
78
1.62
SSD = 100 cm
1.6
1.58
diode reading
1.56
1.54
1.52
1.5
1.48
1.46
0 0.5 1 1.5 2 2.5 3
plastic water thickness on top of diode (cm)
Figure 5.1 Experimental set-up for the Figure 5.2 Diode signal at 18 MV as a
determination of the water equivalent function of the thickness of the Plastic
thickness of the build-up cap WaterTM slabs on top of an EDP-30 diode.
The depth of dose maximum for a 10 x 10
cm2 field and 18 MV X-rays is 3.5 cm, so
the water equivalent thickness of the
build-up cap of EDP30 is 1.4 cm
One X-Omat V Kodak film is placed inside a Plastic WaterTM phantom at 5 cm depth. The
diode is fixed on the surface of the phantom and an irradiation is performed. Another film is
exposed under the same conditions, but without the diode. The dose decrease at 5 cm
depth underneath the diode is calculated by comparing the two beam profiles at this depth.
We use a film scanner (Scanditronix, an option of our field analyser RFA -300) to obtain
Diodes are calibrated against an ionisation chamber placed at the depth of dose maximum
79
chamber (0.6 cm3) (IC) has a calibration factor traceable to the National Standard Dosimetry
Laboratory in Spain. The diodes are taped on the surface of the phantom near the field
centre, in such a way that they do not perturb the response of the ionisation chamber.
The calibration is performed in reference irradiation conditions (field size at the isocentre 10
x 10 cm2, SSD = 100 cm) (see Section 1.2). As the accelerator rooms are equipped with air-
conditioner, the room temperature is always between 21ºC and 22ºC. First, the reading-in-
experimentally determined factor (in the case of Plastic WaterTM, this factor is 1). To
determine absolute dose-to-water, the Spanish dosimetry protocol is used. This includes
the application of a displacement factor for entrance dose. As the measurements are not
performed on the exponential part of the curve but at the depth of dose maximum, this
factor is not applied. The calibration factor Fcal is determined as the ratio of the absorbed
dose determined with the ionisation chamber and the diode reading (see Section 1.2.2).
As the sensitivity of the diodes depends on dose rate, energy and temperature, some
correction factors will have to be applied to the diode reading when measuring conditions
differ from calibration conditions. Some of the correction factors depend, in addition, on
the diode calibration methodology. Since the diode is fixed on the patient’s skin, the scatter
conditions seen by the diode are obviously not the same as the scatter conditions seen by
the ionisation chamber. Therefore a field correction factor, for example, will have to be
applied even if the diode build -up cap is thick enough to guarantee electronic equilibrium.
Furthermore, as the dose rate sensitivity dependence may change with accumulated dose,
the correction factors that account for this dependence, such as the SSD correction factor,
80
3. Wedge correction factor (CFwedge )
In addition, for some types of diodes, we performed tests to assess the importance of the
following issues:
The results of the measurements of the entrance correction factors are given in Table 5.4
OFic ( c )
CFFS = (5)
OFdiode ( c )
where OF is:
R (c )
OF (c ) = (6)
R(10 cm)
with c the side of the square field in cm, and R the reading.
If the measurements of OFdiode are performed at the same time as the measurements of OFic
using a plastic phantom, attention should be paid to OFic because it may differ from OFic
applied. This factor will probably depend on field size. To simplify things, one can measure
OFdiode and compare it with OFic measured in water at the depth of dose maximum.
Field size correction factors obtained for different diodes in different beam qualities are
81
1.005
1.000
EDP10
0.995
CFfield size
0.990
0.985
0.980
0 10 20 30 40 50
side of square field (cm)
Figure 5.3 CFFS for EDP-10 diodes and 6 MV X-rays. The mean and SD of measurements
1.050
1.030
CF field size
1.010
EDP30
P30
0.990 QED
Isorad-p
0.970
0.950
0 10 20 30 40 50
side of square field (cm)
Figure 5.4 CFFS for EDP-30, P30, QED and Isorad-p diodes in an 18 MV X-ray beam. The
mean and SD for three diodes of each type and three measurements are shown.
82
5.1.2.2 TRAY CORRECTION FACTOR (CFtray )
Shielding blocks are positioned on a tray attached to the treatment head. In our hospital,
the tray for the Clinac accelerator is made of PMMA of 0.5 cm thickness. Inserting a tray
between the source and the patient changes the amount of electrons that reaches the
patient’s skin. Therefore, if the diode does not have an appropriate build -up cap, the tray
To determine CFtray, we measure the tray transmission for different field sizes at the depth
of dose maximum, first with an ionisation chamber and then with the diodes taped to the
surface of the plastic phantom. The transmission factors measured with the ionisation
chamber and with the diodes are compared, and CF tray as a function of field size is obtained
transmissionic ( c )
CFtray = (7)
transmissiondiode ( c )
R ( c , tray )
transmission( c ) = (8)
R( c )
with c the side of the square field in cm, and R the reading.
Inserting a wedge in the beam results in a decrease of the dose rate and a hardening of the
spectrum of the beam. Therefore, as the sensitivity of the diode depends on both dose rate
and energy, a correction factor different from 1 is expected when using wedges.
The wedge correction factor is defined as the ratio between the wedge transmission factor
for a 10 x 10 cm2 field, measured with the ionisation chamber placed at the depth of dose
maximum, and the wedge transmission factor for the same field size, meas ured with the
diode placed at the field centre taped on the surface of the phantom.
83
1.01
1.005
0.995
CFtray
0.99
0.985
0.98 EDP10
0.975
0 10 20 30 40 50
side of the square field (cm)
Figure 5.5 CFtray for EDP-10 diodes and 6 MV X-rays. The mean and SD of CF tray
1.010
1.000
CFtray
0.990
EDP30
P30
QED
0.980
isorad-p
0.970
0 10 20 30 40 50
Figure 5.6 CFtray for EDP-30, P30, QED and Isorad-p diodes in an 18 MV X-ray beam.
The mean and SD for three diodes of each type and three measurements are
shown.
84
transmission( w ,10x10 cm 2 , zmax )IC
CFwedge = (9)
transmission( w ,10x10 cm 2 )diode
For 6 MV X-rays CFwedge was determined for 10 EDP-10 diodes, three times each and for
different field sizes. The estimated uncertainties associated with the determination of this
measurements performed with the same diode on different days. For the EDP-10 diodes and
6 MV X-ray beams the dependence on field size of CF wedge is of the same order as the
uncertainty in the factor itself. Therefore, CF wedge is considered independent of field size
In Table 5.2, the correction factors for the EDP-10, EDP-30, P30, QED and Isorad-p diodes
are shown for the different wedges. For the EDP-30, P30 and Isorad-p diodes, the mean of 5
measurements for three different diodes of the same type is given. In the case of the QED
Table 5.2 CFwedge for different wedges for a 10 x 10 cm2 field and for 6 MV X-rays (EDP-
85
5.1.2.4 SSD CORRECTION FACTOR (CFSSD)
When the SSD is changed, the dose per pulse and the electronic contamination change.
First, the sensitivity of the diodes depends on dose per pulse. Secondly, if the build -up cap
of the diode is not thick enough an overestimation of dose at short SSD can be due to
electronic contamination that would be “seen” by the diode but not by the ionisation
chamber placed at the depth of dose maximum. Therefore, a SSD correction factor different
from 1 is expected.
2
R ic (z max , 10x10 cm , SSD)
R ic ( zmax , 10x10 cm 2 , SSD = 100 cm)
CFSSD = (10)
R diodes (10x10 cm 2 , SSD)
R diodes (10x10 cm 2 , SSD = 100 cm)
The diode is taped on the surface of a Plastic WaterTM phantom. The field size is fixed to 10
x 10 cm2 at the isocentre. The reading of the diode measurement at different SSD normalised
to the reading of the diode measurement 100 cm SSD is compared to the same ratio
measured with an ionisation chamber placed inside a Plastic WaterTM phantom at the depth
of dose maximum.
In Figure 5.7 CFSSD for different types of diodes and for an 18 MV X-ray beam is shown.
86
1.060
1.040
1.020
CF SSD
1.000
EDP30
0.980 P30
QED
0.960 Isorad-p
0.940
70 80 90 100 110 120 130
SSD (cm)
Figure 5.7 CFSSD for EDP-30, P30, QED and Isorad-p diodes for an 18 MV X-ray beam.
The mean and standard deviation for three diodes of each type and three
To measure the directional dependence of the diodes they where placed with the
measuring point (considered to be at the b asis of the diode) at the isocentre on the surface
of the Plastic WaterTM phantom. The variation of the diode response with gantry angle for
a 10 x 10 cm2 field was measured with the diode long axis perpendicular (axial symmetry)
and parallel (tilt symmetry) to the gantry rotation (Figure 5.8). The results of the
Figure 5.8 Design of a P30 diode. The plane containing x is the plane of gantry rotatio n
for tilt symmetry. The plane containing y is the plane of gantry rotation for
axial symmetry.
87
88
1.040
EDP10
1.020
CF axial angle
1.000
0.980
-80 -60 -40 -20 0 20 40 60 80
angle (°)
1.070
EDP10
1.050
CFtilt angle
1.030
1.010
0.990
-80 -60 -40 -20 0 20 40 60 80
angle (°)
Figure 5.9 Axial and tilt symmetry for EDP-10 diodes, for 6 MV X-rays.
89
1.080
EDP30
1.060 P30
QED
Isorad-p
1.020
1.000
0.980
-80 -60 -40 -20 0 20 40 60 80
angle (°)
1.060
EDP30
1.040 P30
QED
Isorad-p
angle
1.020
CF tilt
1.000
0.980
-80 -60 -40 -20 0 20 40 60 80
angle (°)
Figure 5.10 Axial and tilt symmetry for EDP-30, P30, QED and Isorad-p diodes for 18 MV
X-rays.
90
5.1.2.6 TEMPERATURE CORRECTION FACTOR (CFtemperature )
To study the influence of temperature on the diode signal a water phantom equipped with a
thermostat is used. The diodes are taped on a thin slab of Perspex, which is in contact with
the water. The temperature, measured with a digital thermistor provided with an immersion
probe, is slowly increased from 22.5 ºC to 32 ºC. The sensitivity of the diodes is determined
at different temperatures and expressed relative to that at 22.5 ºC. Each temperature is
maintained approximately 20 minutes in order to reach full thermal equilibrium between the
surface of the phantom and the diodes. The procedure is performed twice, when the diodes
are received and after some time of use (1kGy of accumulated dose).
if the temperature at which the diodes have been calibrated (T cal ) is 22.5°C.
If Tcal differs from 22.5ºC, the temperature correction factor is defined as:
For 10 EDP-10 diodes the variation of sensitivity (in percentage) per ºC varies from 0.26 to
0.34. Table 5.3 shows the sensitivity variation per ºC for different EDP-30 diodes, new and
after some time of clinical use. In Figure 5.11 the variation of sensitivity of EDP-30, P30,
91
Diode number new variation(%/ºC) post-irradiation variation (%/ºC)
1 0. 290 0. 275
2 0. 293 0. 291
3 0. 320 0. 340
4 0. 216 0. 274
5 0. 281 0. 273
Table 5.3 Variation of sensitivity (SVWT) in % per ºC for EDP-30 diodes, when the diodes
1.05
EDP30 QED
isorad-p
1.04 P30
signal at tºC/signal at 21.5ºC
1.03
1.02
1.01
0.99
20 25 30 35 20 25 30 35
temperature t (ºC) temperature t (ºC)
Figure 5.11 Variation of sensitivity with temperature of (a) three EDP-30 and three P30
diodes, and (b) three QED and three isora d-p diodes. 21.5ºC was chosen as
normalisation temperature.
92
5.1.2.7 INFLUENCE OF THE DOSE RATE ON THE DIODE’S SENSITIVITY
A Clinac 1800 (Varian) accelerator changes the dose rate by varying the number of pulses
per unit of time and not the dose per pulse. Therefo re, to test the intrinisic influence of
dose per pulse on the diode sensitivity, the following experiment was designed. In order to
exclude electron contamination, the diodes are inserted in a Plastic WaterTM phantom at 10
cm depth. Their flat surface is facing the beam. In this way, the measurements are not
affected by differences in build -up caps. The source-surface distance is then varied from 80
cm (0.56 mGy/pulse) to 130 cm (0.23 mGy/pulse). The field size is chosen constant and in
such a way that the phantom is completely irradiated at any distance. The test has been
performed for EDP-30, P30, QED and Isorad-p diodes. Results are shown in Figure 5.12.
In this experiment, no wedges are used to reduce the dose rate as by doing so, not only
dose rate would be modified but also the energy spectrum. The results of such an
1.020
EDP30
P30
1.010 QED
relative diode signal
Isorad-p
1.000
0.990
0.980
0.200 0.250 0.300 0.350 0.400 0.450 0.500 0.550
dose rate (mGy/pulse)
Figure 5.12 Influence of dose rate on diode signal. Relative sensitivity is defined as the
ratio of the response of the diode at any dose per pulse to the response at the
93
5.1.2.8 SENSITIVITY VARIATION WITH ACCUMULATED DOSE (SVWAD)
The loss of sensitivity with accumulated dose has been studied for EDP-30, P30, QED and
Isorad-p diodes. Readings for 177 MU have been obtained after irradiations of 1500 MU (17
Gy) at 240 MU/min (dose rate used in clinical practice) in an 18 MV X-ray beam. A
Figure 5.13 shows the loss of sensitivity with accumulated dose for the different diodes.
For the EDP diodes the SVWAD can vary substantially. Much smaller values have been
reported [Meijer 2001], which are more in agreement with the specifications in Table 1.1.
-1%
-2%
loss of sensitivity
-3%
-4%
-5% EDP30
P30
-6%
QED
-7% Isorad-p
-8%
accumulated dose (Gy)
Figure 5.13 Sensitivity loss with accumulated dose for EDP-30, P30, QED and Isorad-p
diodes
94
Table 5.4 Summary of entrance correction factors for EDP-30, P30, QED and Isorad-p
diodes, for 18 MV X-rays. For EDP-30, P30 and Isorad-p diodes, the mean of
three diodes of each type and three determinations for each diode is given. For
QED diodes the mean of one measurement and three diodes is given, with the
exception of CF wedge and CF angle., for which the results of one diode and one
95
SSD =100cm 30º 1.004 0.998 0.999 0.989
Field size 45º 0.998 0.998 1.012 0.978
2
= 10 x 10 cm 60º 1.009 1.041 1.015 1.010
SVWT EDP-30 P30 QED Isorad-p
0.23-0.30 0.15-0.21 0.29-0.30 0.19-0.25
%/ºC %/ºC %/°C %/ºC
SVWAD EDP-30 P30 QED Isorad-p
3.4%/100Gy 0.2%/100Gy 0.8%/100Gy 0.3%/100Gy
Table 5.5 Summary of entrance correction factors for EDP-10 diodes for 6 MV X-rays. The
96
Open field 45º 1.015
60º 1.023
SVWT EDP-10
0.26-0.34%/ºC
97
5.2 PERFORMANCE OF SOME COMMERCIAL DIODES IN HIGH ENERGY
PHOTON BEAMS – THE LEUVEN EXPERIENCE
5.2.1 INTRODUCTION
For entrance dose measurements, in vivo diodes are covered with a build -up cap to enable
measurements at a certain depth. Firstly, in any case (unless for surface dose
measurements) some build -up is necessary in order to avoid the initial steep dose gradient.
Secondly, the build -up region is influenced by the variation in the amount of electron
field size, or the presence of beam modifiers [Biggs 1979]. Depending on the cap thickness,
the diode will reflect this variation to a certain degree, in particular in high-energy photon
beams, where the influence of the contaminating electrons in the build -up region is larger,
mainly due to their increased range [Sjögren 1996]. Therefore, the build -up cap of the diode
should ideally have the same thickness as the build -up layer covering the ionization
chamber during the calibration, i.e. the build -up cap thickness should be equal to the depth
On the other hand, thicker build -up caps cause a larger perturbation of the treatment field
(which is however of only limited relevance if the diode is applied during only one or two
seesions). Commercially available diodes have different build -up cap thicknesses; some of
them are designed with a thinner cap to minimise field disturbance . With regard to the
consequences that may jeopardise the accuracy of in vivo dosimetry at high energies: i)
The magnitude of the correction factors will be larger. ii) In a clinical situation where it is
preferable to limit the number of factors, the factors have been established independently
and are used together for various combinations of beam setting. This may no longer be
98
This problem is addressed in some recent studies performed in Leuven [Georg 1999],
Barcelona [Jornet 2000], Copenhagen [Lööf 2001], and Amsterdam [Meijer 2001]. Only
limited information is available describing the correction factor variation and/or the
achievable accuracy for in vivo dosimetry methods in the ‘high’ energy range (16 – 25
MV). Therefore, the first aim of these studies was to assess and analyse the variation of
diode correction factors for entrance dose measurements at higher photon beam energies.
type as well as p-type diodes have been included. The dosimetric characteristics of n -type
diodes have been published in [Jornet 2000]. The results of the comparative study of EDP-
30, QED (p-type diodes) and P30 (n -type diode) in Barcelona are shown Section 5.1, and are
substantiated by the results obtained in Copenhagen [Lööf 2001] and Amsterdam [Meijer
2001].
Results of the study performed in Leuven ([Georg 1999]) will be presented in this
contribution. In addition to determining the correction factors for commercial diodes, the
total build-up thickness of the diodes is modified in order to min imise the correction factor
variation.
5.2.2.1 MATERIAL
The Scanditronix p-type diodes that are recommended for in vivo dosimetry in high energy
photon beams have been investigated: two different EDP-20 diodes - “old” and “new” type
- and EDP-30 diodes. The specifications of these diodes are listed in Section 1.1.4, Table
1.3. The main difference between old and new EDP-20 diodes is the higher doping level of
the new type, which was introduced in order to improve the dose rate properties with
accumulated dose. The EDP-30 also has a high doping level. Furthermore the new EDP-20
type is covered with a thin (0.2 - 0.3 mm) plastic layer. The old type EDP-20 diode, which is
99
For all measurements diodes are connected to commercially available or home made
electrometers with a low input impedance [Rikner 1987]. Additional build -up caps have
been manufactured in the hospital’s mechanical work-shop. These build -up caps have
(nominally) thickness equivalent in attenuation to 10, 15, 20 and 30 mm water, and are made
from either copper, lead, or iron. The build -up caps can easily be added to and removed
from the diode without any damage. Whenever modified with an additional build -up cap,
the diode is described by adding as an index the total build -up thickness of the (modified)
diode, and the build -up cap material is indicated by its chemical symbol, e.g. measurements
made with an EDP-20/30(Cu) diode indicate that the EDP-20 diode has been modified by a
copper build-up cap equivalent in thickness to 10 mm water. For EDP-30, the thickness of
equivalent material [Jornet 1996 and Section 5.1.1.4]. Although this value is apperently not
fixed, but depends on the head-scatter spectrum of the accelerator [Meijer 2001, Sjögren
1998], we have assumed it to be around 15 mm. Therefore, e.g. EDP-30/30(Cu) means that
an additional copper build -up cap equivalent in thickness to 15 mm water has been added.
Ionisation chamber measurements are performed using a cylindrical ionisation chamber (NE
2571, volume 0.6 cm3). All diode correction factors are determined in a large polystyrene
provided by different linear accelerators: a Philips SL20 (18 MV, QI = 0.78), a GE Saturne
43/Series 800 (18 MV, QI = 0.77, 25 MV, QI = 0.786), two GE Saturne 42/Series 700 (18 MV,
QI = 0.77 and 0.78), and a GE Sat II+ (23 MV, QI = 0.79). The GE Sat 43/800 is equipped with
an integrated multileaf collimator wh ile all other machines have conventional collimators.
The main difference between the Philips and the GE accelerators is the treatment head
geometry, especially the collimator design and the flattening filter material and position.
The influence of a wedge filter on the different diode types has been investigated in the
beams of the Saturne 43 linac. The internal tungsten alloy wedge (nominal wedge angle
100
60°) is located between the monitor chamber and the upper pair of jaws. The maximum
The influence of a 0.8 cm PMMA block tray is determined for the photon beams of the
same accelerator with a source-to-tray distance of about 62 cm. Block correction factors are
measured for beams of the Philips SL 20, the GE Sat 42 and GE Sat II+ accelerators with
Perturbation effects are determined using film dosimetry. Therefore Kodak X-Omat films are
phantom. All films are evaluated with a Vidar Scanner and film dosimetry software
(Poseidon, Precitron, Sweden) running on a PC. Optical densities are converted into
101
5.2.2.2 METHODS
The entrance dose is measured with the diodes using the methodology described in
Section 1.2 and Section 5.1. Field size correction factors are assumed to be independent of
SSD and SSD correction factors are assumed t o be independent of field size:
In order to verify equation (13) at higher photon energy, where electron contamination can
have a significant influence, several measurements are performed at 80 and 120 cm SSD for
large and small field sizes (5 x 5 cm2 and 30 x 30 cm2), at 18 and 25 MV. If this assumption
would not be valid in the energy range under consideration, a two -dimensional correction
Assuming the validity of equation (13), field size correction factors CFS are measured for
square fields ranging from 5 x 5 cm2 to the maximum field size of 40 x 40 cm2 at reference SSD
(100 cm). Source-surface-distance correction factors CSSD are measured for SSDs varying
Wedge correction factors C wedge are measured at reference SSD for square fields of 5 cm, 10
cm and 20 cm field size. Since the wedge position is critical, measurements with wedge are
performed for collimator orientations of 90° and 270°, and Cwedge is determined as the
average value. A tray correction factor Ctray is determined repeating all CFS and CSSD
measurements for a block tray. Block correction factors Cblock are measured for a fixed
collimator opening of 20 x 20 cm2 for blocks defining square fields of 5 cm, 8 cm, 10 cm, 14.1
cm, and 17.3 cm side length (at the isocenter plane). The block-to-isocentre distance varies
The variation of correction factors for diodes from the same batch is estimated from CFS
and CSSD measurements for three EDP-30 and two new type EDP-20 diodes.
102
5.2.3 RESULTS
Equation (13) has been checked for old and new type EDP-20 and EDP-30 diodes in 18 MV
and 25 MV photon beams provided by GE linacs (SAT 42 and 43). The agreement between
the measured correction factor C(FS,SSD) and the correction factor calculated from C FS and
CSSD is between 1 - 1.5 % for modified and unmodified old type and new type EDP-20
diodes. The larger deviations around 1.5 % are observed for a 5 x 5 cm2 field size either at 80
or at 120 cm SSD. For the modified EDP-30/30(Cu) diodes deviations do not exceed 1.5 %,
while for unmodified EDP-30 diodes maximum deviations around 2 % are observed for a 5 x
5 cm2 field at 80 cm SSD. Total build-up thickness larger than 30 mm (e.g. old type EDP-
Table 5.6 Field size correction factor variation for unmodified old type EDP-20 diodes at
103
5.2.3.2 FIELD SIZE CORRECTION FACTOR CFS WITHOUT TRAY
Field size correction factors for unmodified old type EDP-20 diodes are shown in Table 5.6
as a function of square field size for 18 MV photon beams provided by different linacs. For
field sizes between 5 x 5 cm2 and 30 x 30 cm2 CFS varies by only 1.7 % on the Philips SL 20
Old type EDP-20 diodes without additional build -up show the largest field size correction
factor variation. In a field size range between 5 x 5 cm2 and 40 x 40 cm2 CFS varies by about
When adding build -up caps, the variation could be substantially reduced. For example, for
modified old type EDP-20/30(Cu), for the same field size range and energies, CFS varies
New EDP-20 diodes without additional build -up show a much smaller CFS variation than
the old ones: 2.2 and 2.5 %, respectively, in 18 MV and 25 MV beams provided by the Sat
modified new type EDP-20/30(Cu), but the CFS variation of 2.5 % at 25 MV could not be
improved when modifying the build -up thickness for this type of diode. For both types of
EDP-20 diodes the CFS variation increased when a total water equivalent build -up
EDP-30 diodes without additional build -up show a larger variation of CFS than new type
EDP-20 diodes. When increasing the FS from 5 x 5 to 40 x 40 cm2 the difference between
maximum and minimum CFS values reaches 4 % and 4.4 % at 18 MV from the Sat 42 and 43,
but is only 2.4 % at 25 MV. For the EDP-30/30(Cu) the CFS variation decreases to 0.5 % for
the 18 MV beam of the Sat 42 accelerator and to less than 1.5 % for the 18 MV beam of the
Sat 43 accelerator, but it increases slightly at 25 MV. For the EDP-30/35(Cu) and EDP-
30/45(Cu) the CFS variation increased as compared to the EDP-30/30, this increase is more
pronounced at 25 MV. Figure 5.14 shows the variation of field size correction factors as a
function of field size, dependent on build -up thickness for the different diodes in a 18 MV
104
1.02
(a)
1.01
1.00
0.99
C FS
0.98
0.95
1.02 (b)
1.01
1.00
0.99
CFS
0.98
Figure 5.14 Variation of the field size correction factor C FS for different types of diodes
GE Sat 43 linac: (a) old type EDP-20, (b) new type EDP-20, (c) EDP-30. All
results are obtained with copper build-up caps. The error bars indicate the
105
1.01
1.00
0.99
CFS
0.98
EDP-30
0.97 EDP-30/25
EDP-30/30
0.96 EDP-30/35
EDP-30/40
0.95
0 10 20 30 40 50
square field size (cm)
Table 5.7 shows the variation of CSSD for different linacs for unmodified old type EDP-20
diodes. For the same quality index the difference in SSD correction factors (at a given SSD)
reaches 1 % at maximum at small SSDs. When decreasing the SSD from 100 to 70 cm, C SSD
decreases by about 7 % for 18 MV photon beams with a QI = 0.78, and around 10 % for a
QI = 0.77.
In 18 MV and 25 MV photon energies provided by the GE Sat 43 and/or Sat 42 linacs, SSD
correction factors without tray are almost independent of build -up cap thickness for all
types of diodes. If the total build-up cap thickness does not exceed 30 mm, CSSD for the
same diode with and without additional build -up varies by less than 1 % at a given SSD.
Maximum deviations of more than 1.5 % between C SSD of unmodified and modified diodes
(at a given SSD) are observed only if the total build -up thickness reaches 40 or 45mm.
106
1.05
1.04
1.03
1.02
1.01
C SSD
1.00
OLD EDP-20
0.99 OLD EDP-20/30
0.98 OLD EDP-20/40
NEW EDP-20
0.97 NEW EDP-20/30
0.96 NEW EDP-20/40
0.95
80 90 100 110 120 130
SSD (cm)
Table 5.7 Variation of SSD correction factor for unmodified old type EDP-20 diodes at 18
Figure 5.15 Variation of the source-surface correction factor CSSD with additional
build-up for new and old type EDP-20 diodes in 18 MV photon beams
provided by a GE Sat 43 linac. All results are obtained w ith copper build-
up caps.
The new type of EDP-20 diodes shows a much smaller variation of C SSD with SSD than the
old type. For new type EDP-20 diodes CSSD increases by about 4-5 % at 18 and 25 MV from
GE linacs when increasing the SSD from 80 cm to 120 cm. This variation exceeds 10% for
the old type EDP-20 for almost all build-up cap combinations. Figure 5.15 shows the C SSD
variation as a function of source-surface distance and build -up for new and old type EDP-
107
20 diodes in 18 MV photon beams provided by a GE Sat 43 linac. The error bars indicate the
accuracy of ± 0.5 % in the determination of diode correction factors. All results of modified
diodes displayed in Figure 5.15 are obtained with copper build -up caps.
For EDP-30 diodes with a total build -up less than or equal to 30 mm, CSSD increases by
about 3.5 - 4.5 % at 18 and 25 MV when increasing the SSD from 80 cm to 120 cm. In 25 MV
photon beams this variation is only 2% for the modified EDP-30/45(Cu) diode.
FACTOR CT AND CB
Without additional build -up CFS values for a 40 x 40 cm2 field with and without tray show
variations up to 2% for EDP-20 and 3% for EPD-30 diodes. This difference decreases to
about 1 - 1.5 % for a 30 x 30 cm2 field and to less than 1 % for fields up to 20 x 20 cm2.
For unmodified diodes the SSD correction factor with and without tray varies by about 1.5
–2 % at a SSD of 80 cm. This variation is almost independent of energy and t ype of diode.
For SSDs between 90 and 120 cm the influence of the tray on C SSD is less than 1%.
The influence of the tray at small source-surface distances and for large field sizes can be
decreased by adding build -up. For modified EDP-20/30(Cu) and EDP-30/30(Cu) diodes
correction factors with and without tray, field size correction factors C FS as well as source-
Table 5.8 shows the variation of block correction factors as a function of blocked field size
for different types of diodes and linacs. The collimator setting is kept constant at 20 x 20
cm2 for all blocked fields. The build -up material is indicated by its chemical symbol.
For unmodified old type EDP-20 diodes block correction factors at 18 MV provided by the
Philips SL 20 linac do not differ by more than 1 % from unity, even when reducing the field
size to 5 x 5 cm2 for a fixed collimator setting of 20 x 20 cm2. For the GE Sat 42 linac, CB
reaches 1.032 at 18 MV for the smallest field size. If sufficient build -up material is provided,
108
the influence of blocks can be significantly reduced. For the modified EDP-20/30 at 18 MV
from the GE Sat 42 CB could be reduced to 1.4% for the blocked 5 x 5 cm2.
Wedge correction factors at 18 and 25 MV are of the order of 1.01 - 1.02 for new type EDP-
20/30(Cu) and EDP-30/30(Cu) diodes, and reach 1.06 - 1.07 for the old type EDP-20/30(Cu)
diodes.
The variation of the wedge correction factor C W with field size is less than 1 % for modified
and unmodified new type EDP-20 and EDP-30 diodes at 18 and 25MV. Only for unmodified
old type EDP-20 diodes a small field size dependence could be observed: the C W difference
between a 5 x 5 cm2 and a 20 x 20 cm2 field reaches 2 % at 25 MV and is slightly higher than
1 % at 18 MV. With additional (copper) build -up (old type EDP-20/30 or EDP-20/40) this
109
Table 5.8 Variation of block correction factor for several unmodified and modified
radiotherapy centres. The results of EDP-20 diodes refer to the old type.
The correction factor variation for diodes from the same batch is estimated from
measurements using unmodified diodes, which show the largest CFS variations. The CFS
difference for a given field size is less than 1 % for two EDP-20 diodes and reaches 1.3 % as
a maximum for the three EDP-30 diodes. The difference of SSD correction factors for a fixed
SSD is less than 1 %, for both types of diodes. The small deviation of correction fac tors for
diodes from the same batch is in agreement with observations from other authors [Fontenla
1996b]. It can therefore be concluded that it is sufficient to determine correction factors for
Perturbation effects are determined for old and new type EDP-20 and EPD-30 diodes with
and without additional copper build -up caps in a 10 x 10 cm2 field at 18 and 25 MV. The
relative dose reduction (dose reduction in % with respect to the flat part of the profile at a
specific depth) for unmodified old and new type EDP-20 diodes is about 7-8 % at 18 MV,
and 6-6.5 % at 25 MV, respectively. The corresponding values for the unmodified EDP-30
diode are between 3 % and 3.5 % at 18 and 25 MV, respectively. Depending on energy and
depth, these values are increased by 2-3 % when adding copper build -up caps.
The additional build -up caps have thickness typically around one mm and increase
110
5.2.4 DISCUSSION
CORRECTION FACTORS
For SSDs larger than or equal to 80 cm equation (13) is valid within an uncertainty of 1.5 %
at 18 - 25 MV if the total build-up cap thickness does not exceed 30 mm. It should be noted
that the accuracy in the determination of diode correction factors is better than ± 0.5 %.
The validity of equation (13) within a certain limit has to be considered when defining
tolerance and action levels for entrance dose measurements in higher energy photon
The head-scatter variation of a therapy unit depends strongly on treatment head geometry
([Dutreix 1997], [Nilsson 1998], [Sixel 1996], [Van Gasteren 1991]), and GE linacs are known
to have a pronounced head-scatter variation with field size. Therefore, the independence of
CFS and CSSD was checked on two GE linacs. Since field size correction factors vary much
less for other linac types (see Table 5.7) the findings for the GE linacs are considered to be
For EDP-30 diodes the smallest variation of the field size correction factor at 18 MV is
obtained with an additional water equivalent (copper) build -up thickness of 15 mm. The
results for both types of EDP-20 diodes are in good agreement with this. At 18 MV, as well
as at 25 MV, for old and new EDP-20 diodes the smallest FS correction factor (C FS )
variation is observed when adding 10 mm water equivalent build -up material, which also
corresponds to about 30 mm depth. This optimal depth has been confirmed using different
build-up cap materials: copper, lead and iron. It should be noted that even with a modified
EDP-10/30(Cu) diode a 1.6 % CFS variation is seen at 18MV for the Ph ilips SL 20 linac when
the field size varies between 6 x 6 and 35 x 35 cm2. For EDP-30 diodes in 25 MV photon
111
beams, the CFS variation of unmodified diodes could not be decreased by adding build -up.
The different diode response at 25 MV for EDP-20 and EDP-30 diodes is most probably due
to the difference in build -up material of the unmodified diodes (stainless-steel versus
tantalum).
For the same diode or diodes from the same batch, field size and source-surface distance
correction factors measured at 18 MV on the GE Sat 43, equipped with an MLC, and
measured on the GE Sat 42, having a conventional collimator (CC), agree mostly within 1%.
The main difference between the two linacs are the collimators, while the flattening filter
position and design are similar. Due to the different collimator design they show a
difference in head-scatter variation with field size (12 % for the GE Sat 43+MLC linac versus
16 % for the Sat 42+CC linac when increasing the field size from 4 x 4 to 40 x 40 cm2), but the
depth dose characteristics in the build -up region are almost identical. This finding is in
agreement with a study where two different collimators have been mounted subsequently
on the same accelerator, while all other parts of the linac rema ined the same [Georg 1997].
Here, it has been shown that the depth of maximum dose as well as the skin dose is almost
independent of collimator design while the head-scatter variation is not. Similar doses in
the build-up region but different head-scatter variation may be explained by contaminating
electrons emanating from the flattening filter, which is the main source of contaminating
electrons at higher energies ([Nilsson 1985], [Nilsson 1986], [Sjögren 1996]). The head-
scatter variation, on the other hand, has to be determined at depths large enough to avoid
the unpredictable influence of contaminating electrons in order to describe the field size
The same arguments can be used to explain the difference in diode correction factors for
the same type of diode for the same photon beam quality but provided by linacs from
different manufacturers.
112
5.2.4.4 BEAM MODIFIERS
Even for large field sizes at small SSDs the influence of the tray could be decreased to
negligible values when adding additional build -up caps. It should be noted that the
thickness of the tray holder is only 8 mm; thicker trays may have a larger influence.
Additional build-up material also reduces the influence of blocks to less than 1% for
clinical relevant applications. The influence of trays and blocks can be explained by
electrons produced by trays and blocks is only about 5 – 10 % at depths around 3 - 4 cm,
1998]).
Block correction factors for modified diodes have to be taken into account only when the
field size is substantially reduced by blocks (see Table 5.8). Large field reduction by blocks
can reduce the area of flattening filter seen by the detector which might be accounted for
by a field size correction factor rather than by a block correction factor. However, such
The different doping level for the new type EDP-20 diode was introduced in order to
improve the dose rate linearity of the diode after dose accumulation [Grusell 1993].
Therefore the new type EDP-20 diodes show a smaller variation of SSD correction factors
The relative dose reduction caused by the diode is almost independent of depth, which is
in agreement with investigations performed with diodes from another manufacturer [Alecu
1997]. The attenuation at higher photon energies induced by unmodified EDP-20 diodes is
somewhat larger than previously reported values for Scanditronix diodes ([Leunens 1990b],
[Nilsson 1988], [Rikner 1987]), while it is smaller for the EDP-30. When adding build -up
caps these perturbation effects are increased for all diodes, but they are still smaller than
113
the 13 % attenuation in a 15 MV photon beam reported for a diode with a cylindrical build -
When performing in vivo dosimetry on a weekly basis and assuming a perturbation effect
of 10 %, the dose reduction due to the presence of the diode will be 2 % for treatments with
5 sessions per week. This value can even be reduced by care fully varying the diode
position for each in vivo measurement or by restricting the in vivo procedure to the
beginning of a treatment.
5.2.5 CONCLUSION
When performing entrance dose measurements at the depth of maximum dose, the diode
contaminating electrons, whose dose contributions at shallow depths vary with treatment
geometry and treatment unit. The build -up thickness of commercial diodes is not sufficient
to exclude this influence at higher photon energies. In the energy range between 18 - 25
MV a total diode build -up thickness of 30 mm is found to be the ‘best compromise’ for all
types of diodes and treatment units considered. An additional advantage of a build -up cap
modification when using commercial diodes at higher energies, is the reduction of the
influence of beam modifiers. The additional perturbations caused by the increased build -up
thickness do not have an influence on the practical aspects of in vivo dosimetry (for
114
5.3 PRACTICAL IMPLEMENTATION OF COST-EFFECTIVE APPROACHES TO
IN VIVO DOSIMETRY - THE EDINBURGH EXPERIENCE
5.3.1 INTRODUCTION
doses was begun in Edinburgh in 1992. Prior to that, TLD -based in vivo dosimetry had
been used for some considerable time, but only for critical organ dosimetry and for
verification of complex treatments, such as TBI and TSEI. [Thwaites 1990] This has
continued. In 1991, a Scanditronix DPD-6 electrometer and a set of p-type silicon diodes
(EDE, EDP-10, EDP-20) was purchased with the following initial aims:
• to carry out clinical pilot studies to test each machine, treatment site and treatment
technique to assess the accuracy of the overall radiotherapy process at the point of
dose delivery
• from this, to identify and rectify any systematic errors, and also
• to measure global (as well as specific treatment machine and technique) dosimetric
As the programme progressed, further electrometers (DPD3 and DPD510) and diodes
(EDD2, EDD5, newer-style EDP-10 and EDP-20, all Scanditronix) were acquired and
additional aims were developed linked to routine implementation. From the results of the
clinical pilot studies, a cost-benefit evaluation was carried out to consider how best to
115
At the same time, further work was initiated to attempt to simplify the routine
• the use of additional build -up caps on the standard diodes to minimi se the ranges of
• the use of combined mid -range ‘generic’ correction factors for a specific modality,
• how data is communicated and recorded, to and from the treatment unit
• how the diodes are mounted and handled in the treatment room
level students interested in the work, who carried out dissertation projects at the testing
and development stages under the supervision of the physicist in charge of the diode
project. Trainee physicists have investigated the initial physics testing and workup of the
systems, including calibration and corrections of the dio des ([Kidane 1992], [Brown 1996])
and phantom and clinical pilot studies [Brown 1996]. A part -time (one day per week)
research radiographer, working within the physics department has been involved with the
project since soon after it started and has carrie d out work on initial clinical implementation,
clinical pilot studies, routine implementation and methodology ([Blyth 1997], [Blyth 2001]).
Trainee radiation oncologists have carried out clinical pilot studies ([Millwater 1993],
[Millwater 1998], [Elliott 1999]). At the time of writing, a number of papers are in
Implementation on the linacs in the department has been gradual. Initial detailed studies
were carried out on just one linac, then this has been extended one-by-one to all linacs in
the department. Routine implementation has also been rolled out gradually. Thus different
116
stages of the programme have been in operation on different lin acs at the same time. This
process is not yet complete. At the time of writing, the department is undergoing a major
re-equipping. Allied to this, new diode mounts are being installed in each room, as new
The following sections briefly discuss the approaches which have been taken in
development and those adopted for routine usage in the department, with an emphasis on
chosen are specific to the department. The optimum solutions are not necessarily the same
for every department (nor even for every treatment type within a given department), as
they are influenced by the resources available, the level of implementation, the accuracy
deemed necessary, the other aspects of the department’s quality system and QA
programme in operation, etc. However it is observed that many of the solutions have been
arrived at independently by other departments who have gone through a similar process,
including others contributing to this booklet. Where details are similar and are discussed in
A conventional approach wa s taken to the initial physics testing and workup. Whilst this
is time-consuming, it is necessary to carry this out in full detail at the outset, whatever the
level of routine implementation is to be. Confidence in the diode programme results, in the
evaluation of clinical pilot studies and in the tolerance levels applied can only be based on
a comprehensive commissioning and evaluation of the diode systems. Thus initial testing
included (see Sections 1.2.1, 5.1): stability of diode signal (leakage), reproducibility of
depth of the diodes and measurement of the related perturbation of the radiation field
117
Following this, entrance and exit dose calibrations were carried out (see Section 1.2.2 and
The calibration phantom used wa s of 30 x 30 cm2 epoxy -resin water-equivalent plastic and a
standard thickness of 15 cm was chosen. The phantom was set up over the thin meshed
area of the treatment couch. For entrance dose calibration, the ion chamber was positioned
at the depth of dose maximum and an SSD of 100 cm was set. To minimise subsequent
routine calibration times, the methodology tested and implemented was to position a
number of diodes spaced around the central axis for calibration, sufficiently far out so as
not to perturb the ion chamber reading. A field size of 15 x 15 cm2 was selected, to ensure a
sufficient field margin around this ring of diodes. The dose calculated from the calibrated
ion chamber was corrected for the displacement factor, whilst the diode readings were
corrected for the small measured beam non-flatness at their distance out from the central
axis. For exit dose calibrations, the ion chamber and diodes were left in the same positions,
the gantry rotated to 180o, the SSD was reset to 100 cm and simultaneous irradiations were
again carried out. As the exit diode is to be used to compare measured doses against those
calculated using local planning data, the exit calibration factors were corrected for the
measured lack of scatter to the ion chamber at its exit calibration depth (build -down effect).
Entrance and exit correction factors were determined for each individual diode. Standard
methods were used for this (e.g. Sections 1.2.3 and 5.1; [Van Dam 1994], [Mayles 2000]) and
similar magnitude corrections were obtained to those reported in the literature. Correction
factors were determined for each beam for field size, SSD, tray, wedge, blocks, incident
angle and temperature. All were investigated for both entrance and exit measurements. In
addition for exit measurements, the correction factors for phantom thickness were
determined.
Following all of this initial work, irradiations were carried out on phantoms, comparing
118
of the methodology and the validity of the calibration and correction factors. In addition
detailed phantom studies have been carried out to aid in relating entrance and exit dose
At this stage, a quality control programme was implemented for the diodes. Initially this
approximately yearly [Mayles 2000]. In addition records were begun of the approximate
cumulative doses that each diode had received, as an indication of whether checks should
be more frequent. It was recognised with experience that calibration factors can be
routinely and frequently checked in practice by using the diodes in the linac daily check
weekly basis. A quick check on correction factor validity can be carrie d out using a second
SSD measurement (see Section 1.2.4). If any of these show significant changes then this
Pilot clinical studies were carried out for diffe rent treatment machines, beams, treatment
sites and treatment techniques. Simpler situations were investigated first, e.g. flat surfaces,
shells, perpendicular incidence, etc., where diode positioning problems were expected to be
less, whilst the underlyin g methodology was tested in the clinical setting. Entrance and exit
doses were measured once per week throughout treatment. All relevant correction factors
were applied to the measurements. The only exception to this was the temperature
correction, as the diode was generally positioned after field setup and just before
irradiation, therefore it was felt in most circumstances that the correction required due to
the temperature change was minimal. Temperature corrections were applied if the diode was
in position for a significant time. Diodes were positioned wherever possible on the beam
central axis, but account was taken of block presence, asymmetric field position etc. where
appropriate. The doses were compared to expected entrance and exit doses calculated from
119
the planning data, and the deviations were quantified. These were used to estimate target
volume prescription point dose deviations from the individual fields, aided by the phantom
studies. The values for all fields were combined in proportion to the weight of the field to
provide an estimate of the overall deviation of delivered dose from prescribed dose for the
whole treatment. For tangential field breast treatments, the diodes were placed at a point
midway between the field centre and the medial border of the medial field and used there to
measure entrance dose from the medial field and exit dose from the lateral field. The
combined corrected dose was compared to the dose value from the plan at the appropriate
In all the clinical pilot studies measurements were repeated on a weekly basis throughout
treatment. The observed deviations from all measurements on the same patient were
averaged to provide the best estimate of the overall treatment course deviation betwee n
delivered dose and prescribed dose to the target volume prescription point.
• Typical distributions of individual entrance dose results for various clinical pilot studies
showed mean differences, measured to expected, close to zero. Standard deviations lay
within the range 1.2 % to 4.1 % (typically 1.5 – 3 %), depending on site and linac.
• Exit doses showed mean (systematic) differences varying with site, field and method of
treatment planning. Typic ally mean measured doses were observed to be lower than
4.5 %)
• For total treatment course dose delivery to the target volume (prescription point), some
examples of mean deviations (followed by SD), from the combination of all fields and
120
breast: -4 % (2.5 %)
breast: -2 % (2.7 %)
%)
The SD for the distribution of observed differences of estimated prescription point doses
(from combined field measurements) from the expected values was normally lower than the
SD for individual field data. Similarly the SD for measurements which were repeated on the
same patient and averaged over the treatment course was lower than the s.d for individual
measurement data. Some of the reasons for observed discrepancies between measured and
expected doses, on investigation, were found to be due to the in vivo methods. These
include diode positioning problems such as contact, cable pulling, etc.; positioning
difficulties such as entrance or exit through couch, etc.; wrong correction factor use;
measurements through immobilisation devices, etc.; the limiting resolution of the diode
electrometer for small dose wedged components of fields; diode positioning uncertainties
under large wedges or on steeply angled surfaces. On the other hand some causes were
patient data acquisition, to dose calculation errors (e.g. for tangential field breast
irradiation), to the use of non-CT planning in some situations, to patient set-up variations,
and to incorrect treatment parameters. Some causes were due to changes in the patient at
the time of treatment as compared to the plan, for instance systematic patient size and
shape changes, or random changes such as bowel gas in line with the diode, etc. Some
causes were a combination of factors. The conformal blocked treatments illustrate one such
case. The initial mean deviation was observed to be + 1.5 % (apparent measured dose
greater than expected). On investigation approximately half of this difference was due to an
121
incorrect correction to the diode reading to account for the presence of the conformal
blocks (i.e. diode use and methodology) and approximately half was due to the MU
calculation in these situations (i.e. real change to delivered dose to the patient). On
correcting both of these errors the mean deviation subsequently measured was 0.1 %. This
example nicely illustrates that all discrepancies should be investigated, that diode
dosimetry is precise enough to identify problems at the 1 %, or sub-1 %, level if the system
is implemented carefully and that the diode methodology and use can itself introduce
errors, which should also be suspected and investigated when discrepancies are observed.
department, considering more than 5000 individual entrance dose measurements, the mean
dose ratio (measured to expected) is close to unity (1.001) and the standard deviation is
close to 3%. This, of course, also inherently includes the uncertainties associated with the
diode measurements.
Tolerance levels were chosen on the basis of the pilot studies, at approximately 2 SD, with
the aim of not being too wide that significant problems were missed, but not too narrow
that time-consuming investigations were triggered which were inconclusive or which gave
rise to reduced confidence in the system. 5% was selected for entrance dose measurements
and 8% for exit doses (although tighter tolerances of 3% and 6% may be applied for
Having carried out the detailed studies outlined above, multidisciplinary discussion then
centred on how the department was to utilise diode verification dosimetry in routine
practice. For this the daily positioning and recording was to be passed from the research
radiographer to the normal treatment unit radiographers, unit by unit as the routine use was
122
rolled out to the linacs one-by-one. Possible usage and the procedures involved were
evaluated in terms of cost-effectiveness and with the aim of minimising the time involved at
the treatment units, as the patient workload per linac in the department is high.
The decisions on how and wh en to use diodes routinely, and the rationale for each, were:
department’s main aim for routine use was to identify significant errors in treatment,
which had not been picked up by the other levels of the quality system, which includes
treatment parameters and patient set-up. It was felt that entrance dose checks were the
relatively simple way. To carry out exit dose checks on a routine basis, there is a
significant increase in the time and resources required, without a similar significant pay-
back.
• to initially aim to check all patients, building up to this gradually, linac-by-linac, as the
pilot studies had not shown any situations which were significantly worse or better
than others. The aim is to re -assess this decision periodically, taking into account the
• to normally take just one measurement on each patient, which must be carried out
within the first few days of treatment, ideally within the first two fractions, but in no
circumstances more than one week into treatment, so that any problems are identified
early in treatment and rectified. Frequently, dio de measurements are carried out on each
linac on one particular day and all new patients on that machine are monitored. This
implies that the number of patients monitored on each machine per week is generally in
123
• to apply a general tole rance level of 5 % to individual entrance dose measurements, as
discussed above. The action level is also made equal to the tolerance level, i.e. all
• the actions taken if a deviation is observed over this level have evolved with time.
However this department has come independently to a very similar scheme to that
developed in the Leuven department, as discussed in Section 2.2, Figure 2.2 of this
booklet. On-the-spot checks are carried out on the treatment machine when significant
discrepancy is noted. These checks include patient position, diode position, beam
parameters, etc. In addition, at the earliest opportunity, the treatment unit staff notify a
expected signal from the diode, etc. and looks for possible reasons why there may be a
recognised, a second check would be made on the following fraction, with a member of
the physics department present. If this measurement is within tolerance, then the
treatment is deemed acceptable. If not and the same immediate or afterwards checks do
not identify a valid reason, then the physics department organises a phantom study to
simulate the treatment, comparing ion chamber measurements to the diode to decide
whether the treatment should continue without change or not, i.e. investigating both
In addition to the above, the routine in vivo dose measurement programme aims to carry
out full entrance and exit dose studies (and estimation of target volume dose from these)
major changes in planning systems or planning calculations, to ensure that the whole
124
• full studies on critical patient/treatment groups, for instance dose escalation groups
(with improved tolerances, as discussed above), TBI (but here using TLD), etc.
• and occasional full audit studies on selected groups of limited numbers of patients, as a
repeated overall check on both the diode methods and on the total radiotherapy
process.
All these essentially mirror the clinical pilot studies in operation and are carried out by
physics personnel and/or the research radiographer, in conjunction with the radiographers
minimise the time involved at the treatment unit. This has included:
A full set of correction factors was applied to all the measurements in the clinical pilot
studies. At this stage it is necessary to obtain the best accuracy possible. However for
routine use on routine treatments (although not necessarily so for critical groups, such as
dose escalated patients), it was thought possible that a simpler approach, omitting
methodology and less quality control on the diodes. Initially a comparison was carried out
for patient groups from the clinical pilot studies, where dose estimates made by removing
the correction factors were evaluated. Typically the mean value changed by acceptable
amounts, depending on site, but the standard deviation generally increased significantly.
This implied that an increased tolerance level would have to be used and would have lead
to situations where some significant clinical discrepancies in dose would have not been
recognised. Therefore this approach was deemed not acceptable. Given that the same
125
diode systems are to be used for the more critical groups, it is still necessary to measure
Build-up caps have been constructed for the diodes to match the build -up more closely to
the beams in the department, with the aim of reducing the spread of correction factors
dependent on secondary electron and photon spectrum effects (see also Section 5.2 for
‘high energy’ beams). Unmodified EDP-10 diodes were appropriate for the 4 MV beam and
unmodified EDP-20 diodes were appropriate for the 8 - 9 MV beams. However, additional
caps of 0.6 mm of brass, in combination with EDP-10 diodes, have been investigated for the
6 MV beams. For the 15 and 16 MV beams, additional caps of 1.2 mm of brass, copper and
stainless steel, in combination with EDP-20 diodes, have been studied. These additional
caps bring the build -up thickness to 15 mm water-equivalent for 6 MV beams and 30 mm
for both ‘old-style’ and ‘new-style’ EDP-10 and EDP-20 diodes with these caps. Detailed
In summary:
• the caps produce a reduced range of those entrance correction factors which depend at
least in part on secondary electron and photon spectrum effects, such as field size, tray,
• in general, the caps improve the situation to the stage that the correction factors can be
ignored, i.e. the spread is within ± 0.5 %, for instance field size, trays, etc., or a varying
factor can be replaced by one single factor, e.g. for the motorised wedge for all field
sizes
• the changes were observed to be significantly less for Scanditronix ‘new-style’ diodes
than for ‘old-style’ diodes, in that the newer versions have less variation on some of
126
• the range of values for some correction factors, such as for angle of in cidence, are made
worse
• the shadowing effect is, of course, increased by the use of caps. However as the
number of times that measurements are carried out on an individual patient in the
department is small, even in the full studies where measurements are repeated once per
• the gains were less obvious for the high energy beams (15 and 16 MV) than for the 6
MV beams.
Additional build-up caps of 0.6 mm brass are currently routinely used for all our 6 MV beam
diode measurements. The routine use of build -up caps for 15 and 16 MV beams is still
under discussion.
The use of ‘generic’ correction factors has been investigated for the case of standard
diodes as well as for dio des with build-up caps. For a given treatment modality and for a
specific treatment field, the range of treatment parameters has been investigated for a
representative sample of patients. For each relevant parameter the range of the appropriate
correction factor has been considered and a combined correction factor calculated from the
mid-range values (or the values judged to be most representative). In some cases this has
required a judgement on, for example, the range of beam fractions that are wedged and
unwedged in particular clinical situations (for motorised wedge machines) and a weighted
correction factors around this generic correction factor is small for the norma l range of
treatment parameters used for different patients for a particular field in a particular type of
treatment. Also this is generally better when build -up caps are used, as the range of a
127
Generic correction factors are currently in use for routine entrance dose measurements for
our 6 MV beams using diodes with build -up caps. Tables of generic correction factors are
available, listed by diode, treatment unit and modality (where normally one specific diode
treatment technique and by treatment field. In this way, only one factor is required and is
easily available in any particular situation. This approach implies that if an out-of-tolerance
value is observed, one of the things that the investigating physicist does first is check that
the particular treatment parameters used for that patient are within a tolerable range of the
factors used to produce the generic factor. In no case so far has the use of generic factors,
Full correction factors are still used in any audit studies, critical group studies and new
The application of generic correction factors is also in use in Amsterdam [Meijer 2001]. The
time required for analyzing the patient measurements is hereby substantially reduced, while
keeping the accuracy within acceptable limits. For prostate treatments, the additional
uncertainty for the target absorbed dose as a consequence of the interpatient varience of
the diode correction factors is estimated to be 0.2 % (1 SD). A similar, though more
radically simplifying approach is used by Alecu et al. [Alecu 1998]. They eliminate the
treatment situations that deviate the most from the usual reference calibration conditions.
One of the aims of the department in considering routine diode implementation was to
minimise the time necessary at the treatment unit and also to minimise the duplication of
effort. Therefore for the routine use of diodes fo r monitoring standard treatments, some
consideration was given to the calculation of expected measurement values and to how the
128
Throughout the initial and follow-up studies, all measurements have been recorded
manually onto separate in vivo dosimetry sheets and then entered into a spreadsheet by
the research radiographer (or other research student involved in the project). The
calibration and correction factors are applied in the spreadsheet and the resulting dose
compared to the expected value, also calculated there from data input from the
prescription/treatment sheet and from planning data for the patient and for the treatment
machine.
For routine use, the cost-effective method adopted for calculating the exp ected result was
for physics staff, at the treatment planning stage, to produce an expected diode entrance
reading, taking the expected daily given dose and dividing this by the calibration factor for
the diode/treatment modality and by the generic correction factor for the diode, treatment
modality, treatment technique and field. This is easily done at the time of planning and MU
calculation when the given dose is being recorded. At the same time the ± 5 % tolerance is
applied to this value, so that a range of readings is written on the treatment sheet in an in
vivo dosimetry section. The treatment unit radiographers then simply have to check that
the measured reading is within this range and, at the simplest level, tick one box. If the
measurement is not within range, they must cross another box and place the sheet in a tray
for reference to the physics group for further investigation. In practice this process is
simple to operate. For example all breast patients on any particular treatment unit have the
same range of required diode readings, provided the dose and number of fractions is
standard. For our matched units from the same manufacturer, this same range applies for all
One consequence is that this does not record numerical data. However, if required for
analysis, this is obtainable either by requesting the radiographers to write down the
reading, as well as ticking one or other of the boxes, or by directly grabbing the readings
via the electrometer/PC interface. In general, we have not done either of these things for
quantitatively assess the performance of the systems by limited patient number studies
129
repeating full entrance and exit measurements, as discussed above, and currently feel there
In the initial studies, the diodes were connected via cables following the normal dosimetry
channel route from the tre atment room out to the control area. Following this in the early
cable from the floor area as possible and permanent junction boxes were installed on the
wall in the treatment room. Diode mounts on the couch and on the gantry were considered
and some prototype designs studied. Whilst the dedicated research radiographer, or other
research students, were involved in carrying out the measurements this was not a problem,
as they took care of the cabling, the position of the diode throughout patient setup and the
positioning of the diode. This was independent of what the treatment unit radiographers
were doing and practical and logistics problems were minimal. A number of instances of
connector damage were noted due to excess strain on the connectors from the hanging
cables, or due to damage when connectors were on the floor or were trapped between the
floor and the rotating patient support system. When the measurements were beco ming
routine and were being rolled out to the treatment unit radiographers, this system was not
acceptable. Instead a simple rotating mount was installed on the ceiling above the
treatment unit, directly above and approximately half-way along the horizontal arm of the
rotating gantry. From this an inverted L-shaped cable support was suspended, made from
light-weight cylindrical pipe, with the cable down the centre. The cable goes through the
centre of the rotating mount and above the false ceiling to a cable-way out of the room to
the control area where the electrometer is sited. At the other end it terminates in a
connector at the end of the pipe. The diode connects to this and rests on a quick-remove
hook at the base of the pipe. The swinging cable support is very easily swung completely
out of the way for patient setup and in towards the isocentre for diode positioning. The
system is shimmed to hold position at any point. The height of the lower edge of the cable -
130
supporting pipe was chosen taking into account the range of heights of our radiographers,
such that it is above the head of the tallest, to prevent accidents, but low enough for the
shortest to reach for swinging in or out! This system, designed in conjunction with the
research radiographer and the treatment unit radiographers has been very well accepted
and is very quick and simple in operation. Cable and connector problems have been
negligible since its installation. As new treatment units are installed in our current re -
equipping programme, similar systems are being installed in each treatment room.
Significant simplification and time -saving can be achieved in diode quality control, by
ensuring quick checks are carried out fairly frequently, but requiring little time. These are
used as a warning system, so that major re -evaluations are only carried out when problems
are indicated at this level. This has been discussed above. In addition, a running check on
accumulated diode dose can also give an indication of when re -checks are likely to be
necessary.
131
5.4 LARGE SCALE IN VIVO DOSIMETRY IMPLEMENTATION – THE
COPENHAGEN EXPERIENC E
5.4.1 INTRODUCTION
In 1999 in vivo diode dosimetry was implemented in the Finsen Centre (FC) in Copenhagen,
Denmark. We started with a small group of FC patients, selected by virtue of their relatively
simple treatments. The initial purpose was to develop an effective and reliable quality
has gradually been achieved. In the FC approximately 2250 patients are treated per year.
Seven linear accelerators (linacs) are available with energies ranging from 4 MV to 18 MV.
At this stage diode measurements are carried out on all linacs but there are still specific
types of treatment to be incorporated into the procedure. Our intention was to perform the
protocol where every patient treated should be undergoing in vivo dosimetry commencing
at the start of the treatment course. Taking this into consideration and the capacity of the
centre, our approach has been to utilise an in vivo dosimetry (IVD) system with a relatively
5.4.2 METHODOLOGY
The patient dose measurement is carried out in the beginning of the treatment course,
within the first three fractions in order to make it possible to correct any detected errors.
Expected diode values are derived from an independent spread sheet program containing a
database of correction factors and beam data (depth dose distributions). The program is
not integrated with the Record and Verify system (R&V), hence the prescribed dose and
beam parameters are manually entered into the spreadsheet. Clearly this procedure
increases the number of errors in the quality control (QC) process (errors in input data) but
132
computerised planning, the entrance dose at a point on the central axis calculated with the
As asymmetric field technique (half- or three quarter of the field blocked) is standard
practice at the FC, the position of the diode during measurements has to be determined
during the IVD preparation. Therefore additional corrections of the expected diode reading
at the central axis are added in the spreadsheet to account for various focus-to-diode
distances and off-axis positions in wedged fields and treatment fields modulated wit h a
couch or bolus, the actual focus-to-diode distance is recorded during treatment, followed
5.4.3 EQUIPMENT
chose electrometers easy to handle and diodes with low sensitivity degradation with
accumulated dose, in order to reduce the need for repetitious calibration. Presently we use
the Apollo-5 electrometers and diodes P10 (4 MV), P20 (6-8 MV) and P30 (18 MV) (MDS
Nordion AB). In the tangential treatment technique, measurements are carried out with
cylindrically shaped diodes, Isorad-p diodes (Sun Nuclear Corporation), to minimise the
influence of diode directional dependence thus reducing the number of correction factors.
One treatment unit is equipped with QED diodes (Sun Nuclear Corporation). The use of
different types of diodes will allow us to evaluate statistical fluctuations due to various
diode characteristics.
Currently, the calibration frequency is once every third month. Typically, the sensitivity
has decreased by less then 0.5 % during this period of time (corresponding to around
400Gy). The calibration is performed with a Solid WaterTM phantom in conjunction with the
weekly constancy dosimetry check (ion chamber in plastic phantom) of the treatment unit.
133
Here, some aspects have to be considered regarding the type of errors one aims to detect.
The ultimate choice of calibration method is to calibrate the IVD system to absorbed dose-
to-water using an ionisation chamber with a traceable calibration factor. Any malfunctions
of the treatment unit as well as human errors, such as an erroneous calibration of the
treatment unit, are then detected. However, this calibration procedure is onerous and time -
consuming. If the QC system is calibrated against a constancy check of the linac output
malfunctioning of the treatment unit is most likely to be detected while an erroneous unit
calibration may be veiled. The third, commonly used method would probably not detect
systematic error may be introduced if the diode QC system is calibrated against the mo nitor
chamber of the linac i.e. the diodes are adjusted to the level of that linac’s specific day’s
output. Eventhough modern linear accelerators have a high stability (constancy within ± 2
process (when diode measurements are used to check dose delivery to the target volume).
parameters and it is not likely that any day-to-day variations in linac performance in
Considering the type of deviations that we aimed to detect, we believed that diode
The variation in diode re sponse with temperature was accounted for in the calibration
procedure by means of adding a temperature correction factor to the diode reading during
134
calibration. However, it was considered inappropriate to include the influence of
temperature dependence in the head & neck region where immobilisation devices are used
in most cases.
5.4.5 CORRECTION FACTORS
This section does not deal with diode characteristics or with the variation of different
correction factors, as they are presented and discussed in Chapter 1 and Sections 5.1 and
5.2. Correction factors were applied to account for field size dependence and variations in
response at different SSDs in the energy range of 8 – 18 MV. No wedge correction factor
was necessary as only dynamic wedges are used in the FC. However, a correction factor
accounting for the non-linear dose per pulse dependence of the n -type diodes was applied
in
accounted for in the calibration procedure. In tangential treatment fields the cylindrically
shaped diode has been adopted, consequently no correction of the directional dependence
was required.
In order to reduce the number of correction factors , each diode type (P10, P20) had one set
of correction factors per energy; i.e. the same factors were applied regardless of treatment
Fairly broad tolerance levels were chosen during the initial phase of the implementation
with the intention that they would be gradually minimised with hindsight. Reasonable
levels were established from phantom measurements and selected patient measurements.
The tolerance levels, coinciding with action levels, were related to the complexity of the
treatment delivery according to Table 5.9, rather than the intention of the treatment
(radical/palliative).
135
Treatment Treatment technique Tolerance Diode type
site level
Breast, Tangential fields, 6 - 8 MV ± 8% Isorad-p (Sun Nuclear)
wedge
Table 5.9 Tolerance and action levels established for different treatment sites and the
Since the start in 1999, over 3000 treatment fields have been monitored by means of diode
measurements of entrance dose. Figure 5.16, Figure 5.17 and Figure 5.18 shows of a major
part of all measurements as a percentage deviations from expected values, grouped in bins
of 1 %.
136
90
4 MV 6 - 8 MV 18 MV
80 Average 2.64 0.43 0.67
1 SD 2.9 2.3 2.2
70 Number 173 347 392
60
4 MV 6 – 8 MV 18 MV
Number of fields
50
40
30
20
10
0
<-9.5 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 > 9.5
% Deviation from expected value
Figure 5.16 Deviations from expected value of diode measurements in manually planned
160 4 MV 6 - 8 MV 18 MV
Average 1.64 0.31 1.2
140 1 SD 3.0 2.7 2.2
Number 776 329 699
120
100 4 MV 6 – 8 MV 18 MV
Number of fields
80
60
40
20
0
<-9.5 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 > 9.5
Figure 5.17 Deviations from expected value of diode measurements in computer planned
137
50
Mastectomy Lumpectomy
45 Average 3.51 -1.03
1 SD 3.12 4.71
40 Number 149
289
35
Number of fields Mastectomy Lumpectomy
30
25
20
15
10
0
<-9.5 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 > 9.5
% Deviation from expected value
Figure 5.18 Deviations from expected value of diode measurements in breast treatments
The positioning of the diode becomes more critical in treatment fields incorporating
wedges or compensating filters, reducing the precision in the QC process. This can also be
treatments the level of complexity is low (no wedges, compensation filters or field
asymmetry). Consequently, the variations from the expected value are smaller; i.e. for
energies 6 – 8 MV a standard deviation of 2.3 % (1 SD) was calculated and 2.2 % for 18 MV
(Figure 5.16). In more sophisticated, computer planned treatments (Figure 5.17) the
corresponding figures were slightly higher for 6 - 8 MV: 2.7 %. The largest spread was
This would, to some degree, be expected because of the higher level of complexity. The
positive variation suggesting a systematic error. The distribution had an average of 3.5 %
138
the output factors used to calculate monitor units in quarter fields (¾ blocked field) i.e. the
shift of the average was induced by systematic errors in the treatment process.
At the end of the year 2000, one of our 4 MV energy units had a breakdown and during the
subsequent absolute dose-to-water calibration the output was, by human error, adjusted
nearly 5% higher than intented. The diode system was not recalibrated at the same time
and consequently, the erroneous calibration was detected with patient diode
measurements. However, these measurements (a total of 27) on their own, do not explain
the shift in the mean average of the distribution in 4 MV energy treatment fields. The shift
process rather than the treatment process, considering the fact that the dosimetry accuracy
in this treatment technique has been carefully verified. The shift has to be further
investigated. One aspect may be reconsidered: the diode temperature dependence that is
not accounted for in 4 MV treatment fields. Further actions have to be taken to improve the
Besides the erroneous unit calibration and the systematic errors in the dose calculations
the diode QC process so far has detected two errors in the treatment process: 1) A
discrepancy in beam energy between calculation (manually planned) and treatment delivery
caused by an erroneous input in the R&V system. 2) For a patie nt with two target volumes,
the treatment fields were mixed up; caudal target volume was treated with cranial treatment
5.4.8 CONCLUSION
ongoing process that continuously needs evaluation and refinement. Our goal was to
implement diode measurements for every patient undergoing external radiotherapy in the
FC. To achieve an acceptable workload, a broad tolerance window was defined, accepting
139
The uncertainty in the QC process is a combination of parameters in diode dosimetry
(calibration and correction factors) and the diode measurement (diode positioning). The
latter uncertainty is, according to our experience, of greater influence, especially when
are adopted, and should not be underestimated. The professional training of personnel and
clear guidelines on how to accurately perform measurements are the necessary precursors
To measure every patient once is a major task but not unachievable. Our results show that
even an in vivo dosimetry system with wide tolerance values will provide a reliable QC
process, where human or systematic errors in dose calculations are likely to be detected.
140
5.5 RESULTS OF SYSTEMATIC IN VIVO ENTRANCE DOSIMETRY – THE
MILANO (HSR) EXPERIENCE
Systematic entrance dose in vivo dosimetry was gradually implemented at our Institute
from the end of ’94 in one of the three treatment rooms. It has recently been extended to
another treatment room (November 1999), and the extension to the third room is currently in
In this review we present our experience during the period November ’94 – April 2000 on
5.5.1.1 EQUIPMENT
The two treatment rooms where in vivo dosimetry was implemented are provided with a 6
MV linear accelerator (Linac 6/100); no record and verify systems are available at these
facilities.
For the entrance dose measurements we used p-type silicon diodes (Scanditronix EDP 10)
connected to a multi-channel electrometer (DPD 510, DPD 3 Scanditronix). The diodes are
diode signal in entrance absorbed dose. Procedures for the calibration are similar to those
reported in the ESTRO Booklet n° 1 [Van Dam 1994]. Cadplan (Varian-Dosetek Oy, versions
2.6.2, 2.7, 3.1) is our 3D treatment planning system (TPS); however, during the period
November ’94 - June ’95 we used (2D) older versions of Cadplan (2.5, 2.6.1).
In vivo entrance dosimetry was generally performed at the first session of the therapy
(always within the first three or four days from the start of the treatment). After patient set -
up, just before the start of the irradiation, the diode was positioned at the centre of the
irradiation field, except in the case where the centre is shielded by blocks: in this case the
141
diode was positioned far from the penumbra region. Before fixing the diode, the source-to-
skin distance (SSD) was read and compared with the value defined during the simulator
session and used for treatment planning. In this way possible effects on the prescribed
dose due to small discrepancies between the two conditions (planning/therapy) could be
Diode readings were not corrected by inverse square correction factors due to the choice
of monitoring the “true” accuracy of treatment delivery. Only if the focus-to-diode distance
was different from the SSD because of the use of immobilisation systems or other ancillary
equipment, the diode reading was corrected by an appropriate inverse square correction
factor.
The measured entrance dose was defined as the diode reading corrected by correction
factors. The influence on the diode signal of collimator opening, source-skin distance
(SSD), wedges, tray and obliquit y of the beam with respect to the diode axis was
previously investigated and appropriate correction factors were determined. The diode
signals were initially not corrected to take temperature into account. Only during the last
year we introduced a “clinical” correction factor for temperature. Due to the limited time
when diodes are in contact with the skin in clinical conditions (around 1 – 2 min.), based on
The expected entrance dose was calculated from the prescribed tumour dose by an
independent formula, based on tabulated TPR values and peak scatter factors for the
A 5% action level was applied. However, in the case of opposed beams, a larger error was
accepted in one of the two beams (up to 7-8 %), if the average percentage deviation of the
two opposed fields was within the 5% limit. In this way, we took into account that small
SSD errors may compensate each other in terms o f isocentre dose. The 5% action level was
chosen on the basis of the expected accuracy and reproducibility of our measurement
142
system. This level roughly corresponds to 2 SD where SD is the global standard deviation
of the entrance dose measurement accuracy in most critical conditions (blocked and
wedged fields, tangential beams), excluding patient set-up inaccuracy. The global standard
deviation includes the accuracy and the reproducibility of the entrance dose measurements
with the ionisation chamber and with the diode system and the single SDs of the measured
correction factors.
If the discrepancy between the measured entrance dose and the expected one was below
the action level, in vivo dosimetry data (measured dose, expected dose and the percentage
deviation) was registered by a physicist with the appropriate of information for further
statistical analysis.
When the action level was exceeded, all treatment parameters were verified and the
measurement was generally repeated; a physicist directly performed all these phases (the
repetition of the in vivo dosimetry check may be performed by the technician, often with
the presence of a physicist). Firstly, the agreement between the treatment planning data
and the corresponding simulator data with the control of the data transfer on the treatment
chart was carefully checked. MU calculation/data transfer errors were expected to be rare
because a double check of the treatment chart data, of the treatment planning and of the
MU calculation was always done before tre atment delivery. Further checks concerned the
right use of wedges and blocks, including the presence of a block near the irradiation field
centre. If one of these checks was positive, the percentage deviation between the entrance
measured dose and the expected one could be explained. After correcting the mistake, a
second in vivo dosimetry check was performed to confirm the agreement between the
These checks were not always able to explain the deviation between the measured e ntrance
in vivo dose and the calculated one. Irrespective of this, a second check was always
performed. If a “large” deviation was detected even after the second check, we sometimes
measured the entrance dose on a solid phantom with diode and ionisation chamber in the
same treatment conditions (i.e.: field width, wedge, blocks, SSDs…) and compared it with
143
the expected one. Because of the accumulated experience in assessing the causes of the
From 1991 a double check procedure of MU calculation and data transfer/treatment chart
the ability of this simple tool in strongly reducing the occurrence of systematic errors
before treatment delivery. Every day a “controller” checks the MU and the dose
distribution calculation that has been performed by another operator, together with a check
5.5.2 RESULTS
The in vivo dose measurements on 2001 patients revealed 14 systematic errors. 12 (0.6 %)
were serious (i.e.: leading to an under/over-dosage larger than 5 %) and 6 (0.3 %) were
larger than or equal to 10 %. In Table 5.10 the causes of all errors are reported. If excluding
the rate of serious errors detected by in vivo dosimetry (which “really” escape the MU
calculation/data transfer check) was equal to 0.4 %. A number of minor errors due to an
patient thickness occurring during ‘97, due to a bad resetting of the “zero” indicator of the
simulator couch.
144
5.5.2.2 SYSTEMATIC ERRORS DETECTED BEFORE IN VIVO DOSIMETRY BY MU
In Table 5.10 the systematic errors detected before delivering the treatment by a check of
the treatment chart, of the treatment planning and of the MU calculation are reported. Data
refer to a longer period (‘91-‘99) and to patients treated in all treatment rooms. The rate of
serious errors was found to be equal to 1.53 % with a 0.77 % rate for errors larger than 10
%.
Globally, 156/2001 (7.8 %) patients underwent more than one check; only 6/2001 had more
than 2 checks (see Table 5.11). The rate of second checks was higher for breast patients
In Table 5.12 the values of mean, median and SD of the distribution of the deviations
between measured and expected doses are reported together with the rate of deviations
larger than 5, 7 and 10 %. Globally, we had a mean deviation equal to 0.2 % with a SD equal
to 3.1 %. The rates of deviations larger than 5, 7 and 10 % resulted to be 10.3, 2.6 and 0.2 %
respectively. When averaging the deviations of opposed beams, the standard deviation
was 2.7 % and the rates of deviations larger than 5, 7 and 10 % reduced to 4.9, 0.8 and 0.0 %
respectively.
When considering the presence of a wedge, the rate of deviations larger than 5 and 7 %
was significantly higher than in the group without wedge (p < 0.0001). Similarly, the
presence of a block was related to a higher rate of deviations larger than 5 and 7 %
(respectively p < 0.03) and a systematic deviation from 0 (+ 1.5, p < 0.001).
145
When looking at the different anatomical sites, larger SD were found for breast fields;
smaller SD were found for vertebrae. Most of these results have been discussed elsewhere
[Fiorino 2000].
146
In vivo dosimetry MU calculation and
data transfer check
* Estimated
** Minor errors due to uncorrect assessment of thickness were not considered
*** Large errors due to wrong assessment of thickness
° Excluding 5 large “thickness” errors due to bad resetting of simulator couch(see
text)
Table 5.10 The systematic errors detected by in vivo dosimetry are shown together
147
systematic errors were defined as those which, if undetected, could lead to a
N N pts %
All (averaging opposed beams) 2095 0.2 0.2 2.7 4.9 0.8 0.0
Blocked & unwedged 1156 1.4 1.5 2.9 10.7 2.8 0.3
148
Unblocked 2511 -0.4 -0.5 3.0 9.4 2.2 0.1
Unblocked & unwedged 1660 -0.8 -0.9 2.6 6.7 1.3 0.0
Neck (AP-PA) & 103 1.0 1.0 3.3 10.7 5.8 1.0
Supraclavicular
pelvis, abdomen, thorax AP/PA 1261 0.5 0.3 3.0 10.1 2.0 0.0
Table 5.12 Deviations between measured and expected entrance dose: mean and median
a number of ways. If more than one check was performed, the data of the last
systematic errors might escape the independent check of dose calculation and data
transfer, which should be always performed before treatment delivery. Moreover, in vivo
dosimetry permits detection of a number of minor errors (SSDs and thickness errors) which
would be undetected by the independent check, thus improving the global quality of the
treatment. The high accuracy which can be reached by in vivo dosimetry with diodes, once
appropriate correction factors are applied, can also, by pooling patient data, detect
machine-related problems (for example, the discovered bad resetting of the simulator
couch), wrong configuration of treatment units on TPS and uncorrected procedures during
the chain which precedes treatment delivery, wh ich could lead to systematic errors in dose
149
delivery on a large number of patients. Another important goal of in vivo dosimetry is
maintaining a high level of attention on quality by all the involved staff. For this reason, in
our opinion, it is reasonable to suppose that without in vivo dosimetry, the rate of serious
systematic errors could be higher than the one reported by ourselves and in other similar
studies. In our opinion, any effort in implementing systematic in vivo dosimetry is justified.
above all, the allocation of human resources and a relevant organisational work. This is the
main cause of the limitation of systematic in vivo dosimetry to one of the three treatment
A very important element is the need of precise indications if the action level is exceeded.
Too high a number of second in vivo dosimetry checks may induce a negative impression
concerning the aim of the check and could generate distrust among personnel. Our results
indicate that the 5 % choice for the action level was appropriate. However, it could be
useful to set different action levels depending on the type of beam: in our case it will be
150
APPENDIX 1: LITERATURE OVERVIEW
151
[Aukett 1991] R.J. Aukett
in vivo dosimetry
electrons
Electrons as the cause of the observed dmax shift with field size in high-
152
[Bjärngard 1993] B.E. Bjärngard, P. Vadash and T. Zhu
neutron beam
radiotherapy
153
[Bolla 1995] M. Bolla, H. Bartelink, G. Garavaglia, D. Gonzalez, J.C. Horiot, K.A.
in radiotherapy
and P. Signorotto
practice
P. Signorotto
delivery
154
Radioth. Oncol. 50: 291-300 (1999)
thermoluminescence
dose assessment
cancer irradiation
[Cross 1992] P. Cross, D.J. Joseph, J. Cant, S.G. Cooper and J.W. Denham
155
Radiother. Oncol. 42: 245-248 (1997)
fluoride
[de Boer 1993] S.F. de Boer, A.S. Beddar and J.A. Rawlinson
programme
calculation as a QA procedure
156
[Dutreix 1992] A. Dutreix, G. Leunens, J. Van Dam and E. van der Schueren
[Dutreix 1997] A. Dutreix, B.E. Bjärngard, A. Bridier, B. Mijnheer, J.E. Shaw and H.
Svensson
[Edwards 1997] C.R. Edwards, M.H. Grieveson, P.J. Mountford and P. Rolfe
photons
treatments
[Essers 1995] M. Essers, J.H. Lanson, G. Leunens, T. Schnabel and B.J. Mijnheer
157
[Essers 1996] M. Essers
cohort of patients
radiotherapy
Physical factors affecting absorbed dose to the skin from cobalt -60
158
[Gandola 1990a] L. Gandola, S. Siena, M. Bregni, E. Sverzellati, P. Piotti, C. Stucchi, A.
conventional collimator
[Georg 1999] D. Georg, B. De Ost, M.-T. Hoornaert, P. Pilette, J. Van Dam, M. Van
159
[Graham 1981] W.J. Graham, D.W. Anderson, D.J. Landry and C.R. Bogardus, Jr.
160
[Grusell 1984] E. Grusell and G. Rikner
detector
detectors
[Hazuka 1993] M.B. Hazuka, D.N. Stroud, J. Adams, G.S. Ibbott, and J.J. Kinzie
161
Int. J. Radiation Oncology Biol. Phys. 25: 339-343 (1993)
[Heukelom 1991a] S. Heukelom, J.H. Lanson, G. van Tienhoven and B.J. Mijnheer
Clinical and physical aspects of total body irradiation for bone marrow
transplantation in Nijmegen
[Horiot 1993] J.C. Horiot, J. Bernier, K.-A. Johansson, E. van der Schueren and H.
Bartelink
162
[Howlett 1996] S.J. Howlett and T. Kron
MV X-ray beam
[Hughes 1995] D.B. Hughes, A.R. Smith, R. Hoppe, J.B. Owen, A. Hanlon, M. Wallace
[Indovina 1989] P.L. Indovina, M. Benassi, G.C. Giacco, A. Primavera and A. Rosati
163
[Jornet 1996] N. Jornet, M. Ribas and T. Eudaldo
body irradiation
285 (1993)
164
Uncertainty analysis of absorbed dose calculations from
thermoluminescence dosimeters
compare?
compensating filter
[Kutcher 1994] G.J. Kutcher, L. Coia, M. Gillin, W.F. Hanson, S. Leibel, R.J. Morton, J.R.
Wingfield
165
Comprehensive QA for radiation oncology: report of AAPM Radiation
Brassart
conducteurs
[Lambert 1983] G.D. Lambert, W.E. Liversage, A.M. Hirst and D. Doughty
[Lanson 1999] J.H. Lanson, M. Essers, G.J. Meijer, A.W.H. Minken, G.J. Uiterwaal and
B.J. Mijnheer
Agency (1988)
166
Dosimetry of 106Ru eye applicators with a p -type silicon detector.
system
[Leunens 1990a] G. Leunens, J. Van Dam, A. Dutreix and E. van der Schueren
[Leunens 1990b] G. Leunens, J. Van Dam, A. Dutreix and E. van der Schueren
[Leunens 1991] G. Leunens, J. Verstraete, J. Van Dam, A. Dutreix and E. van der
Schueren
[Leunens 1992a] G. Leunens, J. Verstraete, W. Van Den Bogaert, J. Van Dam, A. Dutreix
radiation treatment affecting the final result. "Garbage in, garbage out".
167
Assessment of dose inhomogeneity at target level by in vivo dosimetry:
[Leunens 1993] G. Leunens, J. Van Dam, A. Dutreix and E. van der Schueren
[Leunens 1994] G. Leunens, J. Verstraete, J. Van Dam, A. Dutreix and E. van der
Schueren
results
168
[Luse 1996] R.W. Luse, J. Eema, T. Kwiatkowski and D. Schumacher
Calandrino
[Meijer 2001] G.J. Meijer, A.W. Minken, K.M. van Ingen, B. Smulders, H. Uiterwaal
irradiation.
in vivo dosimetry
169
Int. J. Radiation Oncology Biol. Phys. 27: 953-958 (1993)
[Mijnheer 1991] B.J. Mijnheer, S. Heukelom, J.H. Lanson, L.J. van Battum, N.A.M. van
Report
patients?
170
Radiology 175: 873-876 (1990)
beams
radiation therapy
171
J. Eur. Radiother. 8 (1987)
in vivo dosimetry
[Ostwald 1995] P.M. Ostwald, T. Kron, C.S. Hamilton and J.W. Denham
dose determination
[Planskoy 1996a] B. Planskoy, A.M. Bedford, F.M. Davis, P.D. Tapper and L.T. Loverock
[Planskoy 1996b] B. Planskoy, P.D. Tapper, A.M. Bedford and F.M. Davis
dosimetry
[Podgorsak 1995] M.B. Podgorsak, J.P. Balog, C.H. Sibata and A.K. Ho
172
In vivo dosimetry using a diode detector system: regarding Lee et al.,
parameters
[Redpath 1992] A.T. Redpath, D.I. Thwaites, A. Rodger, M. Aitken and P.D. Hardman
vivo dosimetry
173
Radiation damage induced dose rate non-linearity in an n-type silicon
detector.
dosimetry
semiconductor diodes
Errazquin
174
[Sánchez 1995] F. Sánchez-Doblado, J.A. Terrón, B. Sánchez-Nieto, R. Arráns, L.
[Sen 1996] A. Sen, E.I. Parsai, S.W. McNeeley and K.M. Ayyangar
175
Acta Radiologica Therapy Physics Biology 15: 340-356 (1976)
thermoluminescence
Errazquin
[Van Dam 1988] J. Van Dam, A. Rijnders, L. Vanuytsel and H.-Z. Zhang
[Van Dam 1992a] J. Van Dam, C. Vaerman, N. Blanckaert, G. Leunens, A. Dutreix and E.
176
Are port films reliable for in vivo exit dose measurements?
[Van Dam 1992b] J. Van Dam, G. Leunens, A. Dutreix and E. van der Schueren
radiotherapy?
to correct?
[Van Dyk 1993] J. Van Dyk, R.B. Barnett, J.E. Cygler and P.C. Shragge
[Van Tienhoven 1997] G. van Tienhoven, B.J. Mijnheer, H. Bartelink and D.G. Gonzalez
177
[Van Gasteren 1991] J.J.M. Van Gasteren, S. Heuke lom, H.J. Van Kleffens, R. van der
department
[Wall 1982] B.F. Wall, C.M.H. Driscoll, J.C. Strong and E.S. Fischer
178
Int. J. Radiation Oncology Biol. Phys. 33: 475-478 (1995)
[Weltens 1993] C. Weltens, G. Leunens, A. Dutreix, J.M. Cosset, F. Eschwege and E. van
der Schueren
[Weltens 1994] C. Weltens, J. Van Dam, G. Leunens, A. Dutreix and E. van der Schueren
dosimetry.
relative dosimetry
Geneva (1988)
[Wolff 1998] T. Wolff, S. Carter, K.A. Langmack, N.I. Twyman and P.P. Dendy
179
Electron contamination in 8 and 18 MV photon beams
180
Entrance in vivo dosimetry with diode detectors has been demonstrated to be a valuable
technique among the standard quality assurance methods used in a radiotherapy department.
Although its usefulness seems to be generally recognized, the additional worklo ad generated
departments in order to gather information and define the sequence of the steps to be
undertaken.
This booklet is set up as a tool to reduce these initial efforts: it is conceived as a step -by-step
guide to implement entrance in vivo dosimetry with diodes in the clinical routine of a
radiotherapy department.
The first chapter about the preparation of the measurements contains information (including
commercial specifications) on diodes, electrometers and software. Practical guidelines for the
calibration of the diodes and the determination of correction factors are given.
The second chapter discusses the actual tasks of the responsible QA person during the initial
training period, with the emphasis on the implementation of the measurement procedure (e.g.
In the third chapter, the interpretation of the measurement in relation to tolerance and action
ISBN 90-804532-3