Sysmex KX21N Service Manual-Mainten
Sysmex KX21N Service Manual-Mainten
Sysmex KX21N Service Manual-Mainten
33 305D036
D *Service*
1: Clog Removal
2: Service Seq.
3: Settings
4: Test Operation
5: Service Data
6: Special Seq.
S *Clog Removal*
Please wait.
S *Service Seq.*
1: Setting Seq.
2: Deprime Seq.
3: Gain Adjustment
4: Control Mode
5: Calibrator Mode
6: Continuous Mode
7: Clog Adjustment
(3) When the sequence is in progress, “Please wait.” is displayed and the progress status will
appear in the system status area.
Note that the sequence cannot be stopped after the sequence started.
(4) When the sequence is completed, the system will return to the Ready Screen.
S *Deprime Seq.*
(2) Press the Start Switch to start the deprime sequence. The liquid will be deprimed.
Press the [SELECT] key to cancel the program execution. The system will return to the Ready
Screen.
(3) When the sequence is in progress, “Please wait.” is displayed and the progress status will
appear in the system status area.
Note that the sequence cannot be stopped after the sequence started.
(4) When the sequence is completed, the system will return to the Ready Screen.
During this sequence, the WBC, RBC, and PLT gain (sensitivity) is adjusted.
The WBC/RBC gain can be adjusted automatically by changing the resistance of the digital control
(potentiometer) on the analog board, PCB No. 2135 (KX-21) or PCB No. 2150 (KX-21N) using standard 3
substances such as CELLCHECK-400. The gain of PLT channel is adjusted by using the analog control
volume.
(1) From the Service Seq. menu, select 3: Gain Adjustment. The Gain Adjustment menu will appear.
S *Gain Adjustment*
1: WBC/RBC
2: PLT
S No. 1 QC
Ready
(2) To exit from control mode, select 4: Control Mode from the Service Seq. menu again.
S No. 1 CL
Ready
(2) To exit from control mode, select 5: Calibrator Mode from the Service Seq. menu again.
(2) Press the [SELECT] key to stop the sequence and the system will enter the Ready status.
If an error that makes the analysis impossible occurs in continuous mode, the continuous sequence
will stop.
S *Clog Adjustment*
WBC CLOG 97
RBC CLOG 97
[SELECT] to exit.
(2) Adjust the voltage. Refer to Section 4: Adjustment for the procedures.
(3) If you press the [SELECT] key, the Ready Screen will appear. Or after 30 seconds have
passed, the Ready Screen will automatically appear.
3
S *Setting* S *Setting*
1: Initialize 1: Initialize
2: Change 2: Change
3: Print Settings 3: Sysmex Support
4: Print Settings
Figure 5-9-a: Settings Menu (KX-21) Figure 5-9-b: Settings Menu (KX-21N)
5.6.1 Initialize
S * Initialize*
1. Cycle Counter
2. Calibration
3. Stored Data
4. QC Data
5. Hardware Limit
6. Others
If you select [Yes], the settings will be initialized and the system will return to the Ready Screen.
If you select [No], or press the [SELECT] key, the system will return to the Ready Screen without
initializing.
5.6.2 Change
1: Calibration 1: Calibration
2: Hardware Limit 2: Hardware Limit
3: Parameters 3: Parameters
4: Calibration Default
Figure 5-12-a: Change Menu (KX-21) Figure 5-12-b: Change Menu (KX-21N)
(2) Using the [↑] and [↓] keys, select the item you wish to change, and press [ENTER] key.
Or use the numeric keys to select item.
(1) Using the [↑], [↓] and [ENTER] keys, move the cursor to the parameter you wish to change.
(2) Using the numeric keys and decimal key, enter the calibration value.
The acceptable range for each parameter is 0.0 to 999.9. If the input range is exceeded, an
alarm will sound and the data entry is ignored.
For each parameter, numerals that are not significant digits will be cut off. For example, if the
WBC calibration is entered as 97.55, the set value will be 97.5.
S * Calibration*
WBC 100.0
RBC 100.0
HGB 100.0
RBC GAIN 100.0
PLT 100.0
A W-SCR 100.0
W-MCR 100.0
RDW-SD 100.0
RDW-CV 100.0
MPV 100.0
(3) When the cursor is on the bottom line, pressing the [↓] key will change the displayed
parameters.
(4) If you press the [SELECT] key, the change confirmation message will appear in the menu
display area.
A By ECR 399L057
(1) Using the [↑], [↓] and [ENTER] keys, move the cursor to the item you wish to change.
(2) Using the numeric keys and decimal key, enter the limit value.
For each parameter, numerals that are not significant digits will be cut off.
[C] key functions as backspace to delete one character.
If a limit value has already been entered, the previous value will be deleted upon any key entry.
6 Figure 5-15-a: Change Hardware Limits Menu (KX-21) Figure 5-15-b: Change Hardware Limits Menu (KX-21N)
(3) When the cursor is on the bottom line, pressing the [↓] key will change the displayed
parameters.
(4) If you press the [SELECT] key, the change confirmation message will appear in the menu
display area.
A By TB99003
(1) Using the [↑], [↓] keys, move the cursor to the item you wish to change.
(2) Pressing the [←], [→] keys alternates “Use” and “Not Use”.
3
S *Parameters* S *Parameters*
13 Figure 5-17-a: Change Parameters Menu (KX-21) Figure 5-17-b: Change Parameters Menu (KX-21N)
(3) If you press the [SELECT] key, the change confirmation message will appear in the menu
display area.
A By TB99003
Figure 5-18-a: Test Operation Menu (KX-21) Figure 5-18-b: Test Operation Menu (KX-21N)
(1) From the Test Operation submenu, select 2: SV Test Operation. The SV Test Screen will
appear.
S * SV Test*
1 O 11 21
2 12 O 22 O
3 13 O 23 O
4 14 24 O
5 15
6 O 16 O
7 17
8 18
9 19
10 O 20 SV No. 1
(1) From the Test Operation submenu, select 3: HC Output Test or 3: Output Test (HC).
The test data will be sent to the host computer.
(2) After the test data is output, Ready Screen will appear.
(1) From the Test Operation submenu, select 4: Output Test (DP).
The DP-510 will print out the specified characters. Check the print quality and format.
(2) After the test data is output, Ready Screen will appear.
(1) From the Test Operation submenu, select 5: Output Test (GP).
The graphic printer will print out the specified characters. Check the print quality and format.
(2) After the test data is output, Ready Screen will appear.
Send the test characters to the built-in printer (IP: Internal Printer).
1: Print
(3) There are 4 screens of Service Data. Pressing the [←], [→] keys change the displayed screens.
On the other three screens, WBC Sampling Data, RBC Sampling Data and PLT Sampling Data are
displayed respectively. Analysis is possible with the Sampling Data displayed on LCD.
S * Service Data* WB
Ready
No. 1 WB 01/30 12: 34
WBC 0 3 3
2 6 14
3 5 3
6 3 4
6 3 7
3 5 12
5 6 9
4 2 16
2 2 17 TOTAL 155
3 1 RATIO 12
1: Print
(4) If the built-in printer is used, the Service Data can be printed out. Press [1] key to print.
If the printing paper runs out or an error occurs during printing, printing will stop and the printer buffer
will be cleared.
Note that Cycle Counter is also printed out as Service Data.
The real-rime ON/OFF status of sensors and solenoid valves can be displayed on Status Display
screen.
(2) Select 5. Status Display from Maintenance submenu. The Status Display screen appears.
* Status* WB
Ready
SEQ.NO. 12
PRESSURE 0.52
VACUUM 240
HGB CONVERT 550
SENSOR 1 2 3 4 5 6 7
SV 1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
1 2 3 4
<Sensor Status>
The displayed status of each sensor is shown in the table below.
<SV Status>
The displayed status of each SV is shown in the table below.
The numbers 1 through 0 on the three lines correspond to each SV as below.
1 2 3 4 5 6 7 8 9 0 SV No. 1 through 10
1 2 3 4 5 6 7 8 9 0 SV No. 11 through 20
1 2 3 4
Compressor
Clog removal voltage control ON/OFF
Compressor Fan
SV No. 21
D *Special Seq.*
Rinse the hydraulic lines using special tools, SCAT, alcohol, PVA and CELLPACK.
Rinse the hydraulic lines using special tools, CELLCLEAN and RO water before shipment.
(1) Start up the instrument in Factory Maintenance mode. The screen as shown below will be
displayed for a second. The program version number is displayed on this screen as [00-XX].
Sysmex KX-21
[00-14]
1: Continue
(3) Press [1] key to continue start-up sequence. The message as below will be displayed until the
Ready Screen appears.
Please wait.
(4) The maintenance instruction message screens as shown below may appear.
*Scheduled Maint.*
(5) Press [3] key to cancel. When the instrument becomes Ready, the message as below will be
displayed.
Ready
(7) Select 6: Special Seq. from Service submenu. The Special Sequence submenu appears.
(8) Select 4: Factory Settings from submenu. Refer to Section 5.9.4 below for the settings.
NOTE: After the Factory Initialize program is executed, the maintenance instruction messages
may be displayed at power ON. Press [3] key to proceed the start-up, and set the current
date/time, then reset the cycle counter. Refer to Section 5.6.1 to reset the counter.
Set up the system automatically depending on the requirement of each shipping destination (USA,
Europe, China, Japan). The display language, units, parameters, etc. for the selected market are set
up.
(2) Select 6: Special Seq. - 4: Factory Settings from submenu. The screen as below will appear.
D * Factory Setting *
Country USA
Refer to the Operator’s Manual Chapter 10 for the details of each setting item. Also the factory
default setting values are listed in the last pages of Chapter 10.
The followings are the different settings between KX-21 and KX-21N.
Outputs raw data which has not been calibrated with coincidence error compensation.
33 305F036