The Type 699 differential pressure transmitter is available with switchable pressure ranges from -1 to 1 mbar up to 0 to 50 mbar. It has optional displays and customer-specific adjustment capabilities. Sensors ensure accurate long-term measurements for applications in HVAC and industrial and medical environments.

Uploaded by

a
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
327 views4 pages

Diff Pressure Transmitter - Huba-699

The Type 699 differential pressure transmitter is available with switchable pressure ranges from -1 to 1 mbar up to 0 to 50 mbar. It has optional displays and customer-specific adjustment capabilities. Sensors ensure accurate long-term measurements for applications in HVAC and industrial and medical environments.

Uploaded by

a
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Differential pressure transmitter

Relative and differential


pressure transmitter
Type 699

The type 699 transmitters are available in switchable pressure


ranges and with or without display. The full-version includes
customer specific adjustment possibilities. Especially developed
sensors for each pressure range ensure accurate long term
stable measurement and the large variety of options provide the
perfect platform for use in air conditioning technology as well as
for fine measurement in the industrial and medical environment.

Pressure range + Available with or without LCD display


+ Adjustable measurement range
-1 ... 1 mbar / + Switchable output signals
0 ... 0.3 – 50 mbar + Switchable response curve (linear or root-extracted)
+ Resettable Zero Point (Reset button)
+ Full scale adjustable
+ Attractive price / performance ratio
+ Application at over and low pressure range possible
+ Fast, easy mounting. Housing incorporates integral bracket
for wall or ceiling mounting

Pressure and flow Huba Control Type 699 | Technical data subject to change | Edition 09/2020 1/5
Technical overview
Pressure range
Relative and differential -1 ... 1 mbar / 0 ... 0.3 – 50 mbar

Operating conditions
Medium Air and neutral gases
Medium / ambient 0 ... +70 ºC
Temperature Storage -10 ... +70 ºC
No condensation
< 3 mbar P1 = 50 mbar P2 = 4 mbar
Application at over pressure range
> 3 mbar P1 = 100 mbar P2 = 4 mbar
Tolerable overload on one side < 3 mbar P1 = -4 mbar P2 = -50 mbar
Application at under pressure range
> 3 mbar P1 = -4 mbar P2 = -100 mbar
ambient temperature 2 x overload
Rupture pressure
70 ºC 1.5 x overload

Materials in contact with medium


Sensor Ceramic Al2O3 (96%)
Diaphragm Silicone
Housing Polycarbonat PC

Electrical overview
Output 1) Power supply 1) Load Current consumption 2)
2 wire 4 ... 20 mA 8.0 ... 33 VDC < supply voltage
0.02 A
-8V
[Ohm] < 20 mA
0 ... 10 V 13.5 ... 33 VDC / 24 VAC ±15% > 10 kOhm < 10 mA
0 ... 20 mA 13.5 ... 33 VDC / 24 VAC ±15% < 500 Ohm < 30 mA
3 wire 4 ... 20 mA 13.5 ... 33 VDC / 24 VAC ±15% < 500 Ohm < 30 mA
0 ... 5 V 3) 6.5 ... 33 VDC / 24 VAC ±15% > 10 kOhm < 10 mA
Filter Response time switchable by off / 0.2s / 1s / 5s / 20s
Polarity reversal protection Short circuit proof and protected against polarity reversal. Each connection is protected against crossover up to max. supply voltage.

Dynamic response
Response time < 20 ms
Load cycle < 10 Hz

Protection standard
Without cover IP 00
IP 54
With cover
IP 65

Display
Double spaced per 8 digit alphanumeric
LCD Display
At additional backlight LCD-Display 30 mA current consumption
Module MODBUS RTU RS-485

Ranges of adjustment
The zero piont is adjustable by reset button.
The Full scale is adjustable by DIP-Switch and additional by the turbopoti.

Adjustability
Optional version with self configurable parameters (see order code selection table)

Electrical connection
Screw terminals for wire and stranded conductors up to 1.5 mm2
Cable gland with built-in strain relief PG11

Pressure connection
Connection pipe Ø 6.2 mm

Mounting instructions
Installation arrangement Recommendation: Vertical, (Factory calibration) with pressure connections downwards
Mounting Mounting bracket (integrated in case)

Tests / Admissions
UL
Electromagnetic compatibility CE-conformity acc. EN 61326-2-3

Weight
Without display ~ 90 g
With display ~ 100 g

Packaging
Single packaging in cardboard
Multiple packaging 20 / 40 / 120

Accuracy
Parameter Unit ±0.5 mbar 0 ... 1 mbar 0 ... 3 mbar 0 ... 5 mbar 0 ... 10 - 50 mbar
Tolerance zero point max. % fs ±1.0 ±1.0 ±0.7 ±0.7 ±0.7
Tolerance zero full scale max. % fs ±1.0 ±1.0 ±0.7 ±0.7 ±0.7
Resolution % fs 0.2 0.2 0.1 0.1 0.1
Total of linearity, hysteresis and repeatability max. % fs ±1.0 ±1.0 ±1.0 ±1.0 ±0.6
Long therm stability acc. to DIN EN 60770 % fs ±1.0 ±1.0 ±1.0 ±1.0 ±1.0
TC zero point 4) typ. % fs/10K ±0.2 ±0.2 ±0.2 ±0.1 ±0.1
TC zero point 4) max. % fs/10K ±1.0 ±1.0 ±0.5 ±0.4 ±0.4
TC sensitivity 4) typ. % fs/10K ±0.3 ±0.3 ±0.2 ±0.1 ±0.1
TC sensitivity 4) max. % fs/10K ±0.6 ±0.6 ±0.5 ±0.5 ±0.2

- no additional root-extracted errors Test conditions: 25 ºC, 45% rF, Power supply 24 VDC
- For changing diaphragm position, compensable with zero point reset TC z.p. / TC z.p. 0 … 70 ºC

1) 2) 3)
Adjustable by DIP-Switch At nominal pressure Additional adjustable by software (with LCD-Display only)
4)
TC = Temperature coefficient

2/5 Pressure and flow Huba Control Type 699 | Technical data subject to change | Edition 09/2020
1 2 3 4 5 6 7 8 9 10

Order code selection table 699. X X X X X X X X X X


max. range 9
Pre-adjustment middle range (factory setting) B
min. range (factory setting) C
Pressure range of 0 % to 100% fs 1
Signal range selectable Pressure range of -10 % to 90% fs 2
Pressure range of -100 % to 100% fs 3
mbar (hPa) Pa mmWS inH2O under pressure max.
0 ...0.3/0.5 30/50 3/5 0.1/0.2 -50 Pa 0
0 ... 0.3/0.5/1 30/50/100 3/5/10 0.1/0.2/0.3 -100 Pa 1
0 ... 0.5/1/3 50/100/300 5/10/30 0.3/0.5/1 -50 Pa 2
Pressure range selectable 0 ... 1/3/5 100/300/500 10/30/50 0.5/1/2 -50 Pa 3
0 ... 3/5/10 300/500/1000 30/50/100 1/2/3 -50 Pa 4
0 ... 5/10/16 500/1000/1600 50/100/160 2/3/5 -50 Pa 5
0 ... 10/16/25 1000/1600/2500 100/160/250 3/5/10 -50 Pa 6
0 ... 16/25/50 1600/2500/5000 160/250/500 5/10/20 -50 Pa 7
mbar 0
hPa 4
Pa 2
Pressure unit
kPa 5
mmWS 3
inH2O 6
without Filter dual DIP-Switch 1 0
Linear
Output signal / with Filter (transposable) tenfold DIP-Switch 2
adjustment without Filter dual DIP-Switch 4 0
Square root extracted
with Filter (transposable) tenfold DIP-Switch 3
0 ... 10 V 13.5 ... 33 VDC / 24 VAC ± 15 % (3 wire) 1
0 ... 20 mA 13.5 ... 33 VDC / 24 VAC ± 15 % (3 wire) 3
Output / power supply 13.5 ... 33 VDC / 24 VAC ± 15 % (3 wire) 4
4 ... 20 mA
8.0 ... 33 VDC (2 wire) 5
Output signal complimentary selectable, at delivery no pre-adjustment 9 2 6
without display 0
with display in pressure unit chosen above 1
Option
with display in % fs 2
with module MODBUS 2,3 1 3
without pressure orifice 1
Pressure connection / pressure orifice on P1 2
Connection pipe Ø 6.2 mm
Pressure orifices pressure orifice on P2 3
pressure orifice on P1 and P2 4
without 0
IP 54 with connection kit (metal), 90º angled including tube 2 m long (Fig. 1) 1
Accessories / with connection kit (plastic), straight including tube 2 m long (Fig. 2) 2
Connection Kit without 3
IP 65 with connection kit (metal), 90º angled including tube 2 m long (Fig. 1) 4
with connection kit (plastic), straight including tube 2 m long (Fig. 2) 5
Pressure range variation
(optinal) Indicate W and state range on order (e.g.: W0 ... + 8mbar/OUT1...6V) 9 1 W

Range of characteristic line


a) b) c)
Out
Out Out
100%
100% 100%

Zero @ 50%

Zero @ 10%
0%
0% Pressure 100% -10%1) Pressure 90% -100%1) 0% 100%
Pressure

Depending on the version parameters are adjustable by customer


Version Variable parameters
Dual DIP-Switch Pressure ranges in grades

Tenfold DIP-Switch Pressure ranges in grades; stepless adjustable with Turbo-Poti / output signals / filter (off / on) / response curve (linear / root extracted)

Tenfold DIP-Switch with Display Pressure ranges in grades; stepless adjustable with Turbo-Poti / pressure units / pressure range character / output signals; additional
0 ... 5 V / filter (off / 0.2s / 1s / 5s / 20s) / response curve (linear / root extracted) / backlight (off / 5min / on)

Accessories 2) Order number


Connection kit for vent duct (metal), 90º angled including tube 2 m long (Fig. 1) 104312
Connection kit for vent duct (plastic), straight including tube 2 m long (Fig. 2) 100064
DIN-rail mounting adaptor (Fig. 3) 112854
Module MODBUS 117305
Calibration certificate 104551

1) 2)
under pressure max. acc. order code selection table = -50 Pa/ 100 Pa Accessories supplied loose

Pressure and flow Huba Control Type 699 | Technical data subject to change | Edition 09/2020 3/5
Dimensions in mm / Electrical connections

2 wire 3 wire Universal


2 and 3 wire

Fig. 1 Fig. 2 Fig. 3

92 4

6 11.6
51

35.2
28

97

EN 60 715-TH 35-7.5

4/5 Pressure and flow Huba Control Type 699 | Technical data subject to change | Edition 09/2020