(Sampling System) Operation and Maintenance Manual
(Sampling System) Operation and Maintenance Manual
FOR
WATER AND STEAM SAMPLING SYSTEM
FOR
Indonesia ISM Power Plant
ENT KOREA
TABLE OF CONTENTS
Maintenance
Items to be inspected Actions
Period
Daily Plant cooling water temperature - Check whether cooling water supply
(<40℃) an) and Pressure (3~ 6 system is normal.
㎏/㎠) In case of the abnormal condition persists,
Check point: Cooling water consult with
supply Press. & Temp. Indicator person(s) in charge to decide whether
(ex: CP501 / CT501) sampling system
have to be shut down until problem clears.
- Check whether instruments work properly.
Daily Final sample temperature - Check plant cooling water temp..
After cooler(s) below the 40℃ - Check for cooler leakage.
Check point: Temp. Indicator for - Confirm instrument work properly.
each sample line (ex:CT501) - Isolate in the sample line for repair in case
temp.
increase above 30℃
Daily Final sample pressure - Check pressure reducing valve and
After cooler(s) should be pressure relief valve
1 ~ 2 ㎏/㎠ for proper functioning.
Check point: Press. Indicator for - Check for any leakage from tube fittings,
each sample line (ex:CP501) valves, cooler, etc
- Confirm instrument work properly.
- Check for filter blockage, clean filter with
air or water.
If found out
Daily Check the cation exchanger In case of resin changed its color,
(resin) change from purple Regenerate the resin or refill by following
to yellow procedure
Regeneration procedure
Check point: Cation exchanger pour resin in column and fill it with
resin for each Cation approx.10% HCL solution
conductivity line When color of the resin is changed back to
purple.
Drain HCL solution and wash resin with
demi. Water
If above regenerating procedure does not
work, discard
Old resin and refill with new one.
Maintenance
Items to be inspected Actions
Period
PH analyzer If necessary, refill or change KCL-
Weekly - Check sample flow solution(3.5 mol)
Weekly - Check level in electrolyte If necessary, open cap of pH electrode
(KCL-solution, 3.5 mol) slightly and let flow-
Check point: KCL-Bottle for each Out 2~3ml of electrolyte. Fasten cap hand-
pH electrode tight.
- Maintenance schedule
- Cleaning of electrode
If necessary, wipe the green tip of the electrode cautiously with a soft clean,
and damp paper tissue. If necessary, remove grease with a tissue, moisture
with alcohol (Caution! Inflammable!). Do not insert into acids.
- Maintenance schedule
- Cleaning of sensor
If the sensor is contaminated. Clean it with soapy water and a pipe cleaner. If
necessary, dip the tip of the sensor into 5% hydrochloric acid for approximately
30 seconds. After cleaned, please, rinsed of sensor with demineralized water
5 to 10 minutes.
- Maintenance schedule
Step3 – Take the screw cap of the replacement kit (A-87.290.050) and fill it
completely with fresh electrolyte.
Step4 – There is a groove in the electrode thread, where excessive air and
electrolyte can escape while mounting the sensor cap. Hold the sensor in
Upright position, sensor tip pointing downwards with the groove on the upper
side.
Step5 – slowly fix the electrode head and the sensor cap to allow the excess
Electrolyte to escape without building up to much pressure inside the electrode.
Fasten sensor cap finger-tight.
Step6 – Clean the electrode thoroughly and dry the sensing membrane with
Soft tissue. And switch on power. Let the sensor run-in at air for at least 1 hour
Afterwards perform an air calibration. And reinstalled the sensor into the flow
cell.
- Longer stop of operation
- Maintenance schedule
- Reagent preparation
Warning:
1. Prepare the reagents with the best silica free DI water quality available.
Otherwise the measuring values are falsified.
2. All bottles are marked with the chemical name, appropriate danger symbol
and international safety code numbers. Always wear safety goggles when
handling strong acids.
- Maintenance schedule
- Cleaning of electrode
- Sodium electrode: Only use the original etching solution from SWAN.
The etching solution is delivered in two bottles, one containing the acidic
solvent, the other containing the fluoride salt. Dissolve the salt in the
solvent before use and mark the date of mixing.
Rinse the electrode with ditilled water if etching the electrode for the first
time, insert the electrode for 2 minutes into the etching solution.
Rinse the electrode with distilled water.
Do not dry the sensing glass bulb!
- Reference electrode:
Step1 – Remove any ion deposits with a soft paper tissue.
Step2 – Slip the grounded glass sleeve upwards with a turning and pushing
movement, let about 1ml KCl flow out, and fix sleeve finger tight with a
gentle turning and pulling movement.
Replace or refill KCl reservoir. Use only original SWAN KCl.
- pH electrode:
Step1 – Remove any ion deposits with a soft paper tissue.
Step2 – Rinse the electrode with distilled water, Insert the electrode into the
etching solution for 2 minutes.
- Longer stop of operation
- Maintenance schedule
- Hydrazine Sensor:
After cleaning, the sensor is very sensitive. Do not correct the value
Let the instrument run in for at least 24 hour.
Disconnecting:
Step1: Shut down the instrument
described see attached the “Stop
operation before maintenance”
Step2: Take off the sensor cable and put it
To the side to prevent wetting it.
Step3: Unscrew the two white knurled
fastening screws.
Step4: Take the hydrazine sensor out of
the flow cell.
Cleaning:
Step1: Take of the rotor.
Step2: Clean the two orifices with a pipe cleaner or a toothpick.
Step3: Clean the rotor with a pipe cleaner. Wipe the sensor with a soft
Tissue, mainly the platinum parts and the whole area which is in contact
With water.
Step4: Rinse all parts thoroughly with clean water.
Step5: Put rotor on the sensor.
Step6: Mount sensor, Fasten screws hand-tight
Step7: Connect sensor cable.
Step8: Open sample flow
Step9: As soon as the rotor in turning, switch on power.
Step1- Shut down the instrument described see attached the “Stop operation
before maintenance”
Step2- Switch off power of all connected devices.
Step3- Remove filter vessel, empty, and remount
Step4- Dismount the reference electrode (please do not dry of the electrode)
Step5- Remove cable from the hydrazine sensor. Dismount sensor, dry it with
A soft, clean tissue, and store dry.