Stratus cs200 Operator Manual
Stratus cs200 Operator Manual
Stratus cs200 Operator Manual
i-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
2014/08 i-3
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
i-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Operator’s Guide
Stratus® CS 200 STAT Fluorometric Analyzer
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Using this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Contacting Siemens Technical Solutions Center . . . . . . . . . . . . . . . . . . 1-3
Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
TestPaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
DilPaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
CalPaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Cannulas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Sample Cups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Operation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Processing Whole Blood . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Processing Plasma . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Processing Automated Dilutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
Quality Control and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-14
System Check Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Optical Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Mechanical Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Fluid Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
Software Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Changing Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Using Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-19
Power On/Power Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Controlled Power Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-21
Restarting the System after Power Shutdown . . . . . . . . . . . . . . . . . . . 1-21
2014/08 i-1
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Calibration Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Principles of Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Updating Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Processing Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Evaluating the Calibration Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Addressing Failed Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Checking Calibration Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Determining Expiration of TestPak Lot Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Deleting a Calibrated TestPak Lot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Transmitting Calibration with POCT1-A2 Enabled . . . . . . . . . . . . . . . . . 3-5
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Maintenance Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Accessory Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Standard Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Monthly Maintenance Overview . . . . . . . . . . . . . . . . . . . . . . . 5-3
Monthly Maintenance Procedures . . . . . . . . . . . . . . . . . . . . . . 5-4
Replacing and Cleaning the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Cleaning Instrument Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Cleaning the Touchscreen and Stylus . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
i-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Troubleshooting Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Troubleshooting Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Interpreting Processing Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Printing Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Miscellaneous Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Troubleshooting Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
System Check Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
Adjusting the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
Adjusting the Fluorometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
Auto Aligning System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Cleaning the Carousel Springs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Cleaning the Carousel Reference Disk and Fluorometer Window . . . . 6-30
Cleaning the Pak Manager Lid Reference Disk . . . . . . . . . . . . . . . . . . . 6-33
Cleaning the Probe Sealing Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
Cleaning the Rotor Containment Boot and Bracket . . . . . . . . . . . . . . . 6-37
Replacing AC Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-40
Replacing DC Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-42
Replacing the Cannula Chamber Lower Seal . . . . . . . . . . . . . . . . . . . . 6-43
Replacing the Cannula Chamber O-Ring Seal . . . . . . . . . . . . . . . . . . . . 6-45
Replacing a Corrupted USB Flash Drive . . . . . . . . . . . . . . . . . . . . . . . . . 6-47
Replacing the Fluorometer Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-49
Replacing the Pipettor Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-53
Replace the Sample Manager Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . 6-56
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Configuration Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Basic Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Advanced Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Basic Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Setting Audio Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Configuring Rotor Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Configuring Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Setting the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Advanced Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Changing Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Configuring Collection Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Configuring LIS / Network Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Configuring Manual Dilution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Configuring an Operator ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Configuring Reference Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Configuring Sample Barcode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Configuring Sample Tube Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Storing the Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Training Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Configuring Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Configuring External Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
2014/08 i-3
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
i-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-1
2014/08 i-5
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
i-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
2014/08 1-1
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
System Overview
The Stratus® CS 200 Analyzer is designed to be user-friendly and provide rapid
turnaround of diagnostics tests. It safely processes whole blood and plasma
samples, delivering the first result within 15 minutes of processing initiation. It
can process up to four tests per sample.
Intended Use
The Stratus® CS 200 STAT Fluorometric Analyzer is a microprocessor-controlled
instrument that measures analytes in body fluids for in vitro diagnostic use.
This instrument is for use by trained healthcare professionals in the clinical
laboratory and Point of Care (POC) settings.
1-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
The guide provides the information required to operate and maintain the
Stratus® CS 200 Analyzer and is organized as follows:
Table 1-1: Stratus® CS 200 Analyzer Chapters
Chapter 4: Quality Control Provides procedures for collecting and interpreting quality
control data on the Stratus® CS 200 Analyzer.
Chapter 7: Configuration Provides procedures for changing the system time and
date, and various other specifications for customizing the
instrument.
Appendix D: Maintenance Logs Provides a log sheet for recording maintenance activities.
Siemens recommends using these as the master and
make copies to track the maintenance performed.
2014/08 1-3
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Supplies
The following supplies are designed specifically for use on the
Stratus® CS Analyzers.
TestPaks
TestPaks are configured with 5 wells and a reaction area. The wells contain
sufficient reagent to complete a single assay.
Figure 1-1: TestPak
DilPaks
DilPaks are used when a sample requires dilution. One DilPak contains
sufficient assay-specific diluent to dilute one sample.
Figure 1-2: DilPak
1-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
CalPaks
CalPaks are used for calibrating TestPak lots. One CalPak contains sufficient
calibrator to run a single calibration.
Figure 1-3: CalPak
Cannulas
The cannula holds the stoppered tube in the sample manager. When
processing begins, the point inside the cannula pierces the tube stopper,
allowing the whole blood to flow to the rotor. Cannulas are designed to be
used with certain types of tubes. Always check the labeling for the appropriate
tube type. For more information, see “Appendix C: Qualified Sample Collection
Tubes” for approved tube types.
Figure 1-4: Cannula
2014/08 1-5
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Rotors
The rotors carry the transferred whole blood to the centrifuge. Each rotor
contains separation gel with a density between that of plasma and blood cells.
During centrifugation, the gel rises to the plasma-cell interface, where it forms
a barrier separating plasma from cells. The plasma is isolated in the rotor,
where it is aspirated by the pipettor and transferred to the TestPak for
processing.
CAUTION: Uneven gel distribution around the entire outer edge inside the
rotor may cause instrument malfunction.
1-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
Sample Cups
Sample cups are used to introduce quality control (QC) material or previously
centrifuged plasma for processing.
Tips
The tip manager holds two disposable tip trays of 40 tips each.
2014/08 1-7
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Tubes
Stoppered qualified blood collection tubes containing anticoagulant are used
with the Stratus® CS 200 Analyzer. For the appropriate tube types, see
“Appendix C: Qualified Sample Collection Tubes” .
Figure 1-8: B-D Sample Tube
Waste Container
The waste container holds used rotors, tips, cups, and paks. A red indicator on
the display indicates when the container is full and needs to be replaced.
1-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
System Components
Before operating the Stratus® CS 200 Analyzer, become familiar with the
components used to process tests and report results.
Figure 1-10: System Components
# Description
2 Printer
3 Touchscreen
4 Tip Manager
8 Rotor Table
9 Sample Door
2014/08 1-9
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
# Description
1 Touchscreen Board
2 Pipettor
3 Tip Manager
6 Rotor Table
1 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
2014/08 1 - 11
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Operation Overview
There are six primary areas involved in the operation of the Stratus® CS 200
Analyzer:
• Sample Manager
• Rotor Table/Centrifuge
• Pak Manager
• Tip Manager
• Display
• Waste Container area
1 - 12 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
• A pipettor with a clean tip aspirates sample from the rotor and transfers it
to the TestPak sample well.
• The pipettor then adds sample and reagents from TestPak wells to the glass
fiber paper.
• The tips are disposed of in the waste container after use to prevent fluid
carryover.
• While in the carousel, the TestPak is brought to testing temperature. The
results are displayed on the screen as they complete and print on a report
slip when all paks for the sample have been processed.
• The carousel moves the TestPak over the fluorometer, where the
fluorescent signal is measured.
• When processing is complete, the sample door opens and the operator
removes the tube/cannula assembly.
• The TestPaks are automatically sent to the waste container.
• If the rotor has been used for four tests, it is automatically sent to the
waste container. Otherwise, it can be held in the centrifuge, depending on
the system configuration.
Note: The waste indicator is red when the waste container is full.
The operator removes it, seals it with the provided lid, and
disposes of it according to local procedures for handling
biohazardous waste. A waste container must be present for
the instrument to operate.
Processing Plasma
A sample cup of plasma can be processed from a special position in the rotor
table. Plasma cannot be tested through the tube/cannula assembly.
The following is an overview of processing plasma samples. For the procedure
to process plasma samples, see “Loading and Processing a Sample Cup” on
page 13 of Chapter 2: Processing Samples and Reporting Results.
• The operator either scans in the patient ID/sample ID or manually enters
one on the display.
• The operator places the plasma into a sample cup and places it in the
special position on the rotor table.
• The operator loads the TestPaks in the Pak Manager Well.
• The screen either shows the scanned patient ID/sample ID or prompts the
operator to enter one.
• If the system is configured to require the operator ID/ collection time and
date entry, the operator manually enters or scans this information.
• The screen indicates which supplies need to be loaded. The supply
indicator turns red when the supply is empty; the Tip Tray indicator
highlights when one tip tray is empty. The Start button remains inactive
until all required supplies and data are present.
• The operator selects Start and the Pak Manager moves the first pak under
the pak barcode reader, which identifies the pak type and related
information.
• The pak carousel moves the pak to the next position. All paks are counted
to ensure that they can be used.
2014/08 1 - 13
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
• A pipettor with a clean tip aspirates the plasma sample from the sample
cup and transfers it to the TestPak sample well.
• The pipettor then adds sample and reagents from TestPak wells to the glass
fiber paper.
• The tips are disposed of in the waste container after use to prevent fluid
carryover.
• While in the carousel, the TestPak is brought to testing temperature. The
results are displayed on the screen as they complete and print on a report
slip when all paks for the sample have been processed.
• The carousel moves the TestPak over the fluorometer, where the
fluorescent signal is measured.
• When processing is complete, the sample door opens and the operator
removes the sample cup from the rotor.
• The TestPaks are automatically sent to the waste container.
Note: The waste indicator is red when the waste container is full.
The operator removes it, seals it with the provided lid, and
disposes of it according to local procedures for handling
biohazardous waste. A waste container must be present for
the instrument to operate.
1 - 14 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
Optical Detection
The optical detection system includes a self-correcting, front-surface
fluorometer calibrated during instrument assembly to a NIST-traceable
reference in order to reduce between-instrument variability. A built-in
feedback diode and onboard references measure the signal at two levels of
fluorescence and correct the output to minimize between-run drift. An
additional internal, solid-phase, fluorescent standard confirms the optical
system integrity prior to each run. A detection response and noise factor
calculation for each sample’s enzymatic reaction rate ensures that each result
meets acceptable performance criteria.
Mechanical Alignment
A system of sensors and motion control software automates mechanical
alignment during instrument installation. The daily System Check verifies
critical alignment parameters through a series of similar routines.
Fluid Delivery
Each daily System Check verifies the integrity of the pneumatic fluid delivery
system through a series of in-line pressure determinations. In addition, fluid
transfers are monitored for atypical pressure profiles during sample
processing.
Temperature
The System Check verifies that the temperature of the thermal chamber and
fluorometer are 37°C ±0.2°C.
Test Results
The display screen indicates approximate time to completion of testing and the
test results as each test is completed. When all tests are finished, a report slip is
printed. The report slip includes the following:
• Patient and/or sample ID
• Instrument serial number
• Operator ID, if configured
• Collection date and time, if configured
• Assay names
• TestPak lot numbers
• Result for each assay
• Units of measurement
• Reference interval, if configured
2014/08 1 - 15
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Error messages when applicable are printed next to or beneath the result. The
results can be transmitted to a remote computer system and also to an
optional external printer available from Siemens.
1 - 16 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
Touchscreen
The touchscreen has many uses, including entering data, selecting functions,
and initiating processing.
Table 1-2: Touchscreen Description
Arrows Select the arrows to move the cursor. The up arrow moves the cursor up or to the
left. The down arrow moves the cursor down or to the right.
Back Arrow Select the Back Arrow to return to the previously displayed screen or to initiate a
controlled shutdown. See “Software Overview” on page 19 of this chapter. This
button does not stop processing.
Change Mode Select to process data in a different mode. Only available from the Home screen.
Delete Select Delete, followed by Enter, to delete the data in the current field.
Help Select Help to display information about the current screen, if available. Select
the Back Arrow icon to return to the previous screen.
2014/08 1 - 17
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Menu Buttons Select the appropriate menu button to complete operating procedures.
Paper Advance Select Paper Advance to feed the printer paper through the paper slot.
Power Select Power to power on or off the instrument. See “Controlled Power
Shutdown” on page 21.
Reagent Status Select to check a TestPak lot’s status before processing a sample to determine if
QC or a calibration is required. Select when the message Check Calibration
Status appears on the System Check report to verify assay calibration expiration.
Sample Illuminates when the sample door is open. Sample door must be closed to
process a sample.
Start Select Start to begin processing or to retrieve information about system needs.
Stop Select Stop to halt processing or stop the printing of a long report.
Tips Illuminates when under 8 tips are left in both tip trays.
Tip Tray Illuminates when one of the tip trays is empty. So long as there are at least 8 tips
present, processing can continue. A new tray should be loaded as soon as
possible.
1 - 18 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
Software Overview
The Stratus® CS 200 Analyzer software has two distinct operational sections:
• Modes: Active when the analyzer is processing. During normal operation,
mode screens are the primary display.
• Menus: Used for viewing and printing information, configuring and
maintaining the analyzer, and troubleshooting.
Modes
The Stratus® CS 200 Analyzer processes data in several modes, each of which
has its own screen:
Table 1-3: Process Data Modes
Mode Description
Manual Dilution Mode Select Man Dil Mode to enter a dilution correction
factor when processing a manual dilution.
Changing Modes
1. From the Home screen, select Change Mode.
2. Use the arrows to highlight the desired mode.
3. Select Accept.
Note: Selecting the Back Arrow without selecting Accept reverts to
the previous mode.
Menus
Display the Main Menu by selecting Menus from the Home screen. The menu
screen shows two main functions:
• Reports
• Maintenance
Using Menus
The Reset button displays on Main Menu screen. Reset is used during
troubleshooting. The Reports and Maintenance functions lead to additional
menus and functions. For more information on the about the mapping of
these relationships, see “Appendix A: Menu Navigation” .
2014/08 1 - 19
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
When using these menu options, you can select Home or the Back Arrow any
time to return to the Home screen.
1 - 20 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide System Overview
2014/08 1 - 21
System Overview Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
1 - 22 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
2014/08 2-1
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Daily Setup
Note: Only trained operators should perform these procedures.
Training must include proper blood-handling techniques.
Perform the following procedures before processing patient samples.
CAUTION: Always install a new, empty waste container after removing one;
otherwise the pak counter will be inaccurate and jams could
occur.
1. Open the left front door.
Figure 2-1: Waste Container
2-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
# Description
2 Tip Trays
2014/08 2-3
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
2-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
Reviewing Limits
The limits shown on the System Check report, except for System Fluorometer
Test Level 2, remain constant. Initial limits for the Level 2 test are based on an
internal stored value which is different for each instrument, and are set for
within ±2.5% of that value when the instrument is manufactured. The internal
value and the limits are updated automatically every 60 days. Retain System
Check reports with your QC records in order to track this updating function.
2014/08 2-5
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Preparing to Process
Checking Reagent Status
Check a TestPak lot’s status before processing a sample to determine if QC or a
calibration is required.
1. Select Reagent Status from the Home screen.
2. Select Next Method until the appropriate assay is displayed.
3. Review the displayed QC or calibration status information to determine if
QC or calibration for a TestPak lot is required.
Specimen Collection
Successful processing includes good phlebotomy technique. See the most
current version of CLSI document H3, “Procedures for the Collection of
Diagnostic Blood Specimens by Venipuncture” for guidelines. Use only
collection tubes approved for use on the Stratus® CS 200 Analyzer. Important
information about specimen collection is included in the Specimen Collection
section of the TestPak Information for Use (IFU) and in the tube manufacturer’s
instructions.
2-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
Processing Samples
Loading and Processing a Sample Tube
CAUTION: Do not remove the tube stopper at any time when testing whole
blood. The instrument will not process tubes without vacuum.
To determine if you have enough sample to process up to four tests, compare
the sample volume in the tube against the sample volume gauge on the front
of the instrument.
Note: If using Sarstedt tubes, use the fill line on the tube, not the
sample volume gauge.
If the sample level in the tube is lower than the level shown on the gauge or
the fill line, do not use the tube. Draw a new sample or centrifuge the sample
externally and transfer the plasma to a cup.
Figure 2-4: Verifying Sample Volume with a Becton Dickinson Sample Tube
1. Select Home.
2014/08 2-7
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
2. If the tube has a barcoded label, scan it using the barcode reader located
on the front of the instrument. To scan, move the tube with label facing
the instrument across the front of the barcode reader.
Note: Tube must be vertical to the barcode reader and 4–6 inches
(10.2–15.2 cm) away from the barcode reader.
Figure 2-5: Scan Barcode Label
2-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
2014/08 2-9
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
10. Insert the cannula/tube assembly into the sample door opening.
Figure 2-7: Load Sample
CAUTION: Make certain that the small opening in the rotor and the
embossed text face up. Seat the rotor firmly, flat on the table.
Figure 2-8: Load Rotor
2 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
13. Hold each TestPak by the edges and place flat into the Pak Manager well.
Make sure the arrow on each TestPak label points toward the instrument.
2014/08 2 - 11
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
2 - 12 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
5. Select Collection Time and enter a time and date with comment code, if
required.
6. Select Accept.
7. Place 500 µL of the diluted sample in a sample cup.
8. Place the cup in position on the rotor table.
9. Select Start.
Note: The result displayed on the screen, on the printed report, and
transmitted to the LIS is corrected for the dilution factor
entered.
Note: The results report for a manually diluted sample includes the
message: Corrected for Manual Dilution.
1 200 0.2
2 300 0.3
3 400 0.4
4 500 0.5
2014/08 2 - 13
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
3. Carefully place the sample cup in the cup position on the rotor table.
WARNING: Make certain the sample cup is seated firmly in the cup
position.
Figure 2-11: Load Sample Cup
4. On the Home screen, enter the Patient ID or Sample ID; select Enter.
Note: Limit for Patient ID or Sample ID is 12 characters; limit for
User ID is 10 characters.
If a barcode is scanned containing more characters than the
limit, the message Invalid Barcode displays and no ID data is
entered. Press Continue and follow the facility matching
patient sample procedure to enter a valid ID.
If ID is entered manually, the Stratus® CS 200 Analyzer alarm
beep will sound with each character added in excess of the
limit; characters will not be accepted beyond the limit. Follow
the facility matching patient sample procedure to enter a
valid ID.
5. Enter the Operator ID; select Enter.
6. Select Collection Time and enter a time and date with comment code, if
required.
7. Select Accept.
8. Hold the TestPak by the edges and place flat into the Pak Manager well.
Make sure the arrow on the TestPak label points toward the instrument.
2 - 14 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
2014/08 2 - 15
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Reporting Results
Once all tests are processed, the analyzer prints a Results Report for the
Sample.
Figure 2-12: Sample Results Report
2 - 16 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Processing Samples and Reporting Results
The Results Report displays Asterisks (***) next to any test Result value that is
outside of the defined reference interval. An explanation of the messages
displayed are detailed in Table 2-2.
Table 2-2: Result Messages
Above Cutoff of <value> Displays when the test result value is above the Cutoff Value.
For more information about using Cutoff Values, see
“Configuring Reference Intervals” on page 10 of Chapter 7.
Above Assay Range Displays when the test result value is above the Assay Range,
Asterisks (*** AAR) replace the Result value. If desired, rerun
the assay as a Dilution. For more information, see “Processing
an Automated Dilution” on page 12 of Chapter 2.
Below Assay Range Displays when the test result value is below the Assay Range,
Asterisks (*** BAR) replace the Result value. If desired, rerun
the assay.
2014/08 2 - 17
Processing Samples and Reporting Results Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
2 - 18 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Calibration
Chapter 3: Calibration
2014/08 3-1
Calibration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Calibration Overview
To accurately measure the concentration of analyte in a sample, the assays
must be calibrated on the Stratus® CS 200 Analyzer. Calibration is
accomplished by a calibration update procedure which “fine-tunes” the
calibration, allowing the TestPak lot, CalPak lot, and analyzer combination to
work at optimal performance.
Principles of Calibration
A calibration update uses one CalPak and three TestPaks corresponding to the
method being calibrated. Information for the update is encoded on the CalPak
and TestPak Barcode labels.
When TestPaks are manufactured, calibration coefficients (C terms) are
assigned to each lot. The analyzer reads these coefficients from the TestPak
barcode and stores them in memory.
In Calibration Mode, the instrument automatically processes the three
TestPaks, using the contents of the CalPak. The mean of the three replicates is
used to mathematically update one of the calibration coefficients, the C1 term.
The instrument subsequently uses the updated C1 term, along with the other
coefficients from the TestPak barcode, to calculate test results when the
specific lot of TestPaks is used.
After calibration update, when a patient sample or QC material is processed,
the microprocessor within the instrument compares the test result to its
internal calibration data for that TestPak lot, as based upon the stored
coefficients. The test result is displayed and printed.
Updating Calibration
Routine Calibration is required:
• for any new TestPak lot.
• every 30 to 90 days, depending on the assay calibration stability claim.
• at the start of a new assay on the instrument.
• whenever Quality Control (QC) monitoring indicates.
• if the replacement of a part causes a change in the performance of QC.
Keep the following in mind regarding routine calibration:
• Refer to the TestPak IFU for calibration stability and calibration interval
information.
• The analyzer stores calibration data for three different TestPak lots of an
assay, allowing a new lot to be calibrated as soon as it is received rather
than waiting until it is time to use it.
• The analyzer will not process TestPaks off an uncalibrated lot. If a
calibration has expired or is due to expire within 72 hours, the message
Check Calibration Status displays on the System Check report. See
“Checking Calibration Status” on page 4 of this chapter.
3-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Calibration
Processing Calibration
1. From the Home screen, select Change Mode.
2. Use the arrows on the touchscreen to highlight Calibration Mode; select
Accept.
3. Enter the Operator ID; select Enter.
4. Load one CalPak per the assay being calibrated.
Note: Frozen CalPaks must be completely thawed at room
temperature and properly mixed prior to use. Refer to the
assay specific CalPak IFU for more details.
5. Load three TestPaks of the same lot number for the assay being
calibrated.
6. Close the sample door.
7. Select Start.
Note: When the calibration is complete, a report prints.
Term Definition
Run Time The date and time when Start was selected to begin the calibration.
TestPak Lot Number The lot number of the TestPaks included in the TestPak barcode.
CalPak Lot Number The lot number of the CalPak included in the CalPak barcode.
TestPak Coefficients Mathematical terms used to calculate the internal calibration curve.
C0, C1, C2, C3
2014/08 3-3
Calibration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
3-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Calibration
2014/08 3-5
Calibration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
3-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Quality Control
2014/08 4-1
Quality Control Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
QC Differences Description
Trends A trend is the gradual increase or decrease in the quality control results
over at least five data points per level for an assay.
Shifts A shift is a sudden rise or fall in at least five consecutive values per level.
4-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Quality Control
2014/08 4-3
Quality Control Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Setting Up QC Parameters
4-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Quality Control
12. To enter ID and ranges for a different assay, select Next Method until the
appropriate assay is displayed.
13. Repeat Step 5 through Step 11 for all levels to be entered or modified.
14. When all levels for assays have been entered, select Home.
2014/08 4-5
Quality Control Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
7. Select Accept.
Note: The message The changes have been stored displays at the
top of the touchscreen.
8. Select Home.
4-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Quality Control
Processing QC
Three levels of quality control material can be processed for each assay used
on the Stratus® CS 200 Analyzer.
QC Modes
The Stratus® CS 200 Analyzer provides three QC modes.
• QC Low for running a level of QC material
• QC Hi (high) for running a second level of QC material.
• QC Misc (miscellaneous) for running an optional third level of QC material.
Note: Make sure there are no particulates in the sample and no air
bubbles on the surface.
Table 4-2 shows the minimum volume required:
Table 4-2: Sample Cup Volume
1 200 0.2
2 300 0.3
3 400 0.4
4 500 0.5
2014/08 4-7
Quality Control Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
To process QC:
1. From the Home screen, select Change Mode.
2. Use the arrows on the touchscreen to highlight the required QC Mode.
3. Select Next Method to select an assay.
Note: If multiple assays have the same QC ID, select any one of the
assays.
4. Select Accept.
Note: QC IDs that were previously entered on the QC Ranges screen
automatically populate on this screen. However, if the ID
field is blank, manual entry is required.
5. Select Enter.
6. Enter the Operator ID, if appropriate; select Enter.
7. Place the sample cup in the cup position on the rotor table.
8. Close the sample door.
9. Place one TestPak per assay, maximum 4, in the Pak Manager Well.
10. Select Start.
Note: Retain all QC records according to your facility accrediting
agency.
4-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Quality Control
2014/08 4-9
Quality Control Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
4 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Maintenance
Chapter 5: Maintenance
2014/08 5-1
Maintenance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Maintenance Overview
Note: Only trained operators should perform these procedures.
Training must include proper blood-handling techniques.
Maintenance is required to keep the Stratus® CS 200 Analyzer in working
order. There are standard, monthly, and "as needed" procedures included in
this chapter.
Note: Record all completed tasks on maintenance log sheets.
Accessory Kit
An accessory kit containing tools, spare parts, and a fluorometer reference pak
comes with the instrument. Tools are required for both monthly and periodic
maintenance.
The replaceable parts described in the “Troubleshooting Procedures” on
page 20 of Chapter 6: Troubleshooting of this chapter, excluding the USB flash
drive and the fluorometer lamp, are also included in the kit.
Standard Maintenance
Standard maintenance activities include:
• Cleaning up spills immediately. The instrument and area around it should
be kept clean.
• Routinely cleaning the outside surfaces of the entire instrument following
facility specific biohazard procedures.
5-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Maintenance
2014/08 5-3
Maintenance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
5-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Maintenance
2014/08 5-5
Maintenance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
5-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Maintenance
7. Remove the pak shield and clean it with a damp cloth; set aside.
8. Loosen the captive screw at the front of the tip waste chute.
Figure 5-7: Tip Waste Chute Captive Screw
2014/08 5-7
Maintenance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
9. Slide the tip chute off the mounting pin and remove; clean with a damp
cloth.
Figure 5-8: Cleaning the Tip Chute with a Damp Cloth
5-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Maintenance
2014/08 5-9
Maintenance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Other Maintenance
Replacing the Thermal Printer Paper
1. Lift the printer lid; remove and discard the used paper roll from the
spindle.
Figure 5-10: Thermal Printer Paper Feed
# Description
1 Spindle
2 Green Lever
3 Roller
5 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Chapter 6: Troubleshooting
2014/08 6-1
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Troubleshooting Overview
This chapter describes Stratus® CS 200 Analyzer errors and provides
instructions on how to resolve them.
Note: Only trained operators should perform these procedures.
Training must include proper blood-handling techniques. If
additional assistance is required, contact the Siemens
Technical Solutions Center or refer to the local support team.
6-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
2014/08 6-3
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Miscellaneous Errors
Symptoms Resolution
2. Ensure the power cord is plugged into both the analyzer and the
wall outlet. If it is unplugged, plug the power cord into the analyzer
and then into the wall. Turn the main power switch on.
3. Verify that power is supplied to the wall outlet.
4. Check the AC and DC fuses. Turn off instrument and replace if
necessary. For more information, see “Replacing AC Fuses” on
page 40 and see “Replacing DC Fuses” on page 42.
The system does not 1. Check the DC fuse lights. If a DC fuse indicator light is not lit, this is
respond when the the fuse that requires replacement.
touchscreen buttons are 2. Turn the power switch off. Wait 30 seconds.
selected 3. Replace the fuse.
4. Close the sample door and turn the power switch on. Save the
report slip with maintenance records.
Expected report not 1. Open the printer door; tear off the paper and remove obstructions
printed that prevents the print head from moving to the left.
2. Select Paper Advance; feed the paper through the paper slot and
Printer Jam close the printer door.
3. From the Home screen, navigate to Menus > Reports > Print
Message: Printer Error, Results.
Possible Paper Jam 4. Using the keypad on the display, select the result to be printed.
5. Select Enter > Print.
6-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Symptoms Resolution
QC Shift 1. Ensure you are using the same QC lot and ranges used before the
shift.
Changing to a different QC 2. Ensure you are using the same TestPak and CalPak lots used
lot usually causes a QC before the shift.
shift 3. Repeat QC processing with fresh well-mixed QC material.
4. Clean the carousel reference disk and fluorometer window. See
Changing to a different pak “Cleaning the Carousel Reference Disk and Fluorometer Window” on
lot can cause a slight QC page 30.
shift 5. Align system components. See “Auto Aligning System
Components” on page 26.
6. Run System Check. See “Running System Check – Electronic QC”
on page 4 of Chapter 2: Processing Samples and Reporting Results.
2014/08 6-5
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Symptoms Resolution
6-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Error Codes
WARNING: If any panel, cover, or guard must be removed for
adjustment, be extremely careful to prevent personal injury.
Electrical power can be hazardous. Follow these instructions
carefully to avoid injury.
Table 6-2: Error Codes, Cause, and Resolution
CH1H • Pak Problem Rerun the sample as a dilution. If result is normal, repeat
CH2H • Reaction is above analyzer undiluted.
CH3H read capability
(high sample)
CH1M Analyzer data error 1. Select the Power icon on the Home screen.
2. Select Yes to shutdown the instrument.
3. Turn the power switch off.
4. Close the sample door and turn the power switch on.
5. Be sure to save the report slip with maintenance
records.
6. Rerun with a new Test Pak.
CH5G Previously Used Pak Rerun with a new, unused Test Pak.
CR1A • Sensor problem 1. From the Home screen, navigate to Menus > Reset.
CR1B • Transport mechanism 2. Check the DC fuse lights. If a DC fuse indicator light is
CR1C problem not lit, this is the fuse that requires replacement.
CR1D • Level sense problem 3. Lift the analyzer lid. Look for and remove any
CR1F • I/O board problem obstructions around the rotor table. Close the lid.
CR1I • Rotor mechanism is 4. Power down the analyzer:
CR1N obstructed a. Select the Power icon.
• Rotor mechanism needs b. Select Yes to shutdown the instrument.
lubrication c. Turn the main power switch off.
d. Replace the DC fuse.
e. Wait 30 seconds then turn the power switch on.
2014/08 6-7
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
CR1K • Analyzer lid not latched 1. Make sure the analyzer lid is closed completely.
CR1O • Rotor or cup not seated 2. Check the position of the cup or rotor to make sure it is
CR1R properly positioned and seated correctly. Close the sample door.
CR1S • Rotor upside down (the 3. From the Home screen, navigate to Menus > Reset >
small opening should be Home.
up) 4. Power down the analyzer:
• Rotor or cup missing a. Select the Power icon.
• Sensor problem b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
CR1Q Rotor transport problem Contact the Technical Solutions Center or refer to the local
support team.
CR2D Rotor transport sensor or 1. From the Home screen, navigate to Menus > Reset.
CR2H solenoid problem 2. Close the sample door; select Home.
CR2L 3. Power down the analyzer:
CR2R a. Select the Power icon.
CR2V b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
CR3A • Centrifuge elevator sensor 1. From the Home screen, navigate to Menus > Reset.
CR3B or mechanism problem 2. Close the sample door; select Home.
CR3C • I/O board problem
3. Check the DC fuse lights. If an DC fuse LED is not lit, this
CR3D • Centrifuge elevator is the fuse that requires replacement.
CR3E mechanism is obstructed
4. Lift the analyzer lid and remove the left side panel. Look
CR3F • Centrifuge elevator
for and remove obstructions around the centrifuge
CR3I mechanism needs elevator. Install the side panel and close the lid.
CR3N lubrication
5. Power down the analyzer:
CR3Q a. Select the Power icon.
b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Replace the DC fuse.
e. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
6-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
CS8T • Loose connections to 1. From the Home screen, navigate to Menus > Reset.
display 2. Close the sample door; select Home.
• Display problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
4. Lift the analyzer lid. Ensure the display cables are
connected and undamaged; close the lid.
5. Power down the analyzer:
a. Select the Power icon.
b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
FL1D Software error in chemistry Contact the Technical Solutions Center or refer to the local
profile support team.
2014/08 6-9
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
FL1E • Loose connections to lamp 1. From the Home screen, navigate to Menus > Reset.
FL1H manager or fluorometer 2. Close the sample door; select Home.
FL1K • Component not available 3. Check the DC fuse lights. If a DC fuse LED is not lit, this
FL1T when needed is the fuse that requires replacement.
FL1U 4. Lift the analyzer lid. Ensure the display cables are
connected and undamaged; close the lid.
5. Power down the analyzer:
a. Select the Power icon.
b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Replace the DC fuse.
e. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
6. Run a System Check. See “Running System
Check – Electronic QC” on page 4 of
Chapter 2: Processing Samples and
Reporting Results.
7. Adjust the Fluorometer. See “Adjusting the
Fluorometer” on page 23.
FL1J Alignment problem. Align the system components (auto align). See “Auto
Aligning System Components” on
page 26.
PI1A • Pipettor carriage sensor or 1. From the Home screen, navigate to Menus > Reset.
PI1B mechanism problem 2. Close the sample door; select Home.
PI1C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PI1D • Carriage mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
PI1F obstructed to the pipettor carriage; close the lid.
PI1I • Carriage mechanism needs 5. Power down the analyzer:
PI1N lubrication a. Select the Power icon.
PI1Q • Component not available b. Select Yes to shutdown the instrument.
PI1T when needed c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
PI2A • Pipettor probe sensor or 1. From the Home screen, navigate to Menus > Reset.
PI2B mechanism problem 2. Close the sample door; select Home.
PI2C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PI2D • Pipettor mechanism is 4. Lift the analyzer lid. Look for and remove any
PI2E obstructed obstructions to the pipette mechanism; close the lid.
PI2F • Pipettor mechanism needs 5. Power down the analyzer:
PI2I lubrication a. Select the Power icon.
PI2N b. Select Yes to shutdown the instrument.
PI2Q c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
P12H • Sample not mixed 1. Repeat the test with another tube from the patient.
properly, causing 2. Inspect the sample cup or used rotor for air bubbles,
aspiration problem clots, or other contaminants.
• Wrong type of sample tube
used
6 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
PI2L • Not enough sample to 1. Repeat the test with another tube or cup, ensuring no
complete the test air bubbles are present in the sample.
• Air bubbles in sample 2. Inspect discarded rotor or cup for proper fluid level.
3. Run System Check. See “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples
and Reporting Results.
PI2S • Air bubbles in sample 1. Repeat the test with another tube or cup, ensuring no
• Leak in pipetting system air bubbles are present in the sample.
• Level sense problem 2. Run System Check. See “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples
and Reporting Results.
PM1A • Pak carousel sensor or 1. From the Home screen, navigate to Menus > Reset.
PM1B mechanism problem 2. Close the sample door; select Home.
PM1C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PM1D • Carousel mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
PM1E obstructed to the carousel mechanism.
PM1F • Carousel mechanism needs 5. Clean the carousel springs. See
“Cleaning the
PM1H lubrication
PM1I
Carousel Springs” on page 27.
6. Power down the analyzer:
PM1L
PM1N a. Select the Power icon.
b. Select Yes to shutdown the instrument.
PM1Q
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
PM1X Pak carousel alignment problem 1. From the Home screen, navigate to
PM1Y Menus > Maintenance > Alignment > Auto All.
PM1Z 2. Close the sample door; select Start.
3. Lift the analyzer lid. Remove the pak manager lid. Look
for and remove obstructions or debris around the
carousel; close the lid.
PM2A • Pak shuttle sensor or 1. From the Home screen, navigate to Menus > Reset.
PM2B mechanism problem 2. Close the sample door; select Home.
PM2C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PM2D • Shuttle mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
PM2F obstructed to the pak shuttle mechanism; close the lid.
PM2I • Shuttle mechanism needs 5. Power down the analyzer:
PM2N lubrication a. Select the Power icon.
PM2Q b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
PM4A • Shuttle mechanism is 1. From the Home screen, navigate to Menus > Reset.
PM4B obstructed by paks 2. Close the sample door; select Home.
PM4C • Waste shuttle sensor or 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PM4D mechanism problem 4. Lift the analyzer lid. Look for and remove obstructions
PM4F • I/O board problem to the waste shuttle mechanism; close the lid.
PM4I • Shuttle mechanism is 5. Power down the analyzer:
PM4N obstructed a. Select the Power icon.
PM4Q • Shuttle mechanism needs b. Select Yes to shutdown the instrument.
lubrication c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
2014/08 6 - 11
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
PU1A • Sample syringe sensor or 1. From the Home screen, navigate to Menus > Reset.
PU1B mechanism problem 2. Close the sample door; select Home.
PU1C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PU1F • Syringe mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
PU1H obstructed to the syringe mechanism; install the top cover.
PU1I • Syringe mechanism needs 5. Power down the analyzer:
PU1N lubrication a. Select the Power icon.
PU1Q b.Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
PU2A • Pipettor syringe sensor or 1. From the Home screen, navigate to Menus > Reset.
PU2B mechanism problem 2. Close the sample door; select Home.
PU2C • I/O board problem 3. Check the DC fuses. If an LED is not lit, replace the fuse.
PU2F • Syringe mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
PU2I obstructed to the syringe mechanism; install the top cover.
PU2N • Syringe mechanism needs 5. Power down the analyzer:
PU2Q lubrication a. Select the Power icon.
b. Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
PU3H • Valve problem 1. From the Home screen, navigate to Menus > Reset.
PU3L • I/O board problem 2. Close the sample door; select Home.
PU3P 3. Power down the analyzer:
PU3R a. Select the Power icon.
b.Select Yes to shutdown the instrument.
c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
SM1A • Cannula chamber sensor or 1. From the Home screen, navigate to Menus > Reset.
SM1B mechanism problem 2. Close the sample door; select Home.
SM1C • I/O board problem 3. Check the DC fuses. If a light is not lit, replace the fuse.
SM1D • Chamber mechanism is 4. Lift the analyzer lid. Look for and remove obstructions
SM1F obstructed to the sample manager or rotor containment mechanism;
SM1I • Chamber mechanism close the lid.
SM1N needs lubrication 5. Power down the analyzer:
SM1Q • Rotor containment a. Select the Power icon.
SM2A mechanism is obstructed b.Select Yes to shutdown the instrument.
SM2B • Rotor containment c. Turn the main power switch off.
mechanism needs d. Wait 30 seconds then turn the power switch on. Save
lubrication the report slip with maintenance records.
• Sensor problem
• Containment home flag
mount is crooked
SM1H Low pressure after pump is reset Check that syringe thumbscrews are secure.
6 - 12 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
SM1L • Problems with sample 1. Clean the sample door area and the cannula chamber
• Dirty sample door lower seal. See
“Cleaning Instrument
• Cannula chamber seal
problems
Surfaces” on page 5 of Chapter 5:
• Leak in sample tubing Maintenance.
2. Replace the cannula chamber lower seal and the
cannula chamber o-ring seal and lubricant. See
“Replacing the Cannula Chamber Lower
Seal” on page 43 and See “Replacing the
Cannula Chamber O-Ring Seal” on
page 45.
3. Use the plastic torque wrench to tighten the blue
fittings for the sample tubing.
4. Replace the sample tubing. See
“Replace the
Sample Manager Tubing” on page 56.
SM1P • Sample not mixed 1. Verify that correct tube type is in use and that the
thoroughly sample is mixed thoroughly.
• Wrong type of tube used 2. Rerun the test using a new cannula and stopper.
• Cannula problem 3. Prepare a plasma sample and process it in a sample cup.
SM2D • Sample door open switch Close the sample door. Make sure the sample light on the
SM2H or solenoid problem keypad is not lit.
SM2L • Misaligned sample door
SM2R open switch or solenoid
SM3K
2014/08 6 - 13
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
SM3O Problems with sample Make sure the cannula/tube assembly is properly seated
tube/cannula in the sample door.
TM1A • Tip tray sensor or 1. From the Home screen, navigate to Menus > Reset.
TM1B mechanism problem 2. Close the sample door; select Home.
TM1C • Level sense problem 3. Check the DC fuses. If a light is not lit, replace the fuse.
TM1D • I/O board problem 4. Lift the analyzer lid. Look for and remove obstructions
TM1F • Tip tray transport to the tip tray transport mechanism; close the lid.
TM1I mechanism is obstructed 5. Power down the analyzer:
TM1N • Tip tray transport a. With the Home screen displayed, select the Power icon.
TM1Q mechanism needs b. Select Yes to shutdown the instrument.
lubrication c. Turn the main power switch off.
d. Wait 30 seconds then turn the power switch on. Save
the report slip with maintenance records.
TM1X Tip tray transport alignment 1. From the Home screen, navigate to Menus >
TM1Y problem Maintenance > Alignment > Auto All.
TM1Z 2. Close the sample door; select Start.
3. Lift the analyzer lid. Look for and remove obstructions
or debris around the tip tray transport table; close the lid.
UE1H Process error during calibration Review the error log and resolve process errors that
occurred during calibration.
6 - 14 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Chemistry
Centrifuge/Rotor Transport
CR2D Rotor transport Table latch solenoid energized but sensor disconnected or open
CR2V Rotor transport Table latch sol de-energized but sensor disconnected or open
CR3D Centrifuge elevator Elevator did not get to its full position
2014/08 6 - 15
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Control System
CS1L Power fail detected Power fail signal detected from 24-volt power supply
Fluorometer
Pipettor
6 - 16 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
PI3R Level Sense Level was not detected or detected at wrong position
Pak Manager
PM1H Carousel More than one home sensor transitioned per RPM
2014/08 6 - 17
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
PM4F Waste Waste shuttle lost steps returning from waste slot
Pump System
Note: 5 mL syringe used for cannulation/sample to rotor; 100 µL syringe used for sample/reagent
pipetting to pak.
6 - 18 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
PU2I Pump Pipettor syringe (100 µL) motor driver had a problem
PU3P Pump Valve solenoid energized but pressure did not change
Sample Manager
SM1Z Sample chamber Sample Manager to rotor table offset out of range
alignment
SM2D Sample manager door Door solenoid energized but switch disconnected or open
2014/08 6 - 19
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Temperature
Tip Manager
TM1Y Tip manager alignment Tip tray to pipettor new align not accepted
Other
UE1H Calibration Aborted One or more Paks encountered process error during calibration
6 - 20 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Troubleshooting Procedures
The following procedures can be performed during a troubleshooting
investigation or as directed by the Siemens Technical Solutions Center or the
local support team.
Dark slope mean Electronic “noise” from first level fluorometer No Pass or Fail results; used for
measurement; used to calculate 120 mV Noise. troubleshooting only.
Reference slope Electronic “noise” from second level fluorometer No Pass or Fail results; used for
mean measurement; used to calculate 120 mV Noise. troubleshooting only.
System Fluidics System pressure readings. No Pass or Fail results; used for
troubleshooting only.
2014/08 6 - 21
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
When it has been determined which part of the system has failed, refer to
Table 6-5 for the appropriate troubleshooting procedure and where to find it in
the guide. For troubleshooting, do the first step, then rerun System Check. Do
the same for each step until the issue clears.
Table 6-5: System Check Troubleshooting
Fluorometer 1. Clean the Carousel Reference See “Cleaning the Carousel Reference Disk
Disk and Fluorometer Window. and Fluorometer Window” on page 30,
2. Clean the Carousel Lid “Cleaning the Pak Manager Lid Reference
Reference Disk. Disk” on page 33, “Auto Aligning System
3. Align System Components. Components” on page 26, “Adjusting the
4. Adjust the Fluorometer. Fluorometer” on page 23, and “Replacing
5. Replace the Fluorometer Lamp. the Fluorometer Lamp” on page 49 of
Chapter 6: Troubleshooting.
Fluidic 1. Tighten loose fittings. See “Cleaning the Probe Sealing Pad ” on
System Test 2. Clean Probe Sealing Pad. page 35 of Chapter 6: Troubleshooting.
6 - 22 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
2014/08 6 - 23
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
3. Locate the blue knob on the upper right corner of the display board.
Figure 6-2: Blue Knob on Display Board
# Description
1 Blue Knob
4. To darken the display, turn the knob clockwise; to lighten the display,
turn the knob counterclockwise.
5. Close the instrument lid and left front door, ensuring the door is securely
latched.
6 - 24 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
8. Place the RefPak from the instrument accessory kit in the Pak Manager
Well; select Enter.
9. Select Enter when stabilizing is complete; stabilizing takes approximately
5 minutes.
Note: The Adj. Target Voltage value must match the Voltage
Target Calculated value within 1 mV. If the Fluorometer
RefPak is not ejected to the waste container, the Fluorometer
must be adjusted. If the Fluorometer RefPak was ejected to
the waste container, proceed to Step 11.
10. If an adjustment is required:
a. Open the left front door.
b. Insert a long, slender tool into the lid release; press firmly and lift
the instrument lid.
c. Locate the small hole in the center of the tip manager and remove
the plug to gain access to the lamp intensity adjustment screw; set
the plug aside.
d. Insert the recessed end of the small, blue, plastic tuning wand from
the instrument accessory kit into the adjustment screw, located
below the tip tray carriage.
2014/08 6 - 25
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
# Description
2 Back of Touchscreen
f. Turn the screwdriver until the Adj. Target Voltage value is within
1 mV of the Voltage Target Calculated value.
g. Close instrument lid; select Enter.
h. After a few minutes, the Voltage Target Calculated and Adj.
Target Voltage values display. If these are not within 1 mV, repeat
Step 10.
Note: If target and adjustment values do not match after the steps
have been repeated, contact the Siemens Technical Solutions
Center or refer to the local support team.
11. Select the Back Arrow icon.
12. Open the left front door.
13. Remove the Reference Pak from the waste container and return it to the
instrument accessory kit.
14. Close the left front door, ensuring the door is securely latched.
15. Reinstall the tip manager plug and close the instrument lid.
6 - 26 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
16. Run a System Check. For more information, see “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples and Reporting
Results.
17. Run Quality Control on all assays in use.
2014/08 6 - 27
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6 - 28 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
3. Pull up on the three release knobs holding the pak manager lid; place the
lid on the rotor table.
Figure 6-5: Pak Manager Lid Release Knobs
4. Hold the carousel with one hand while loosening the wing nut in the
center.
Figure 6-6: Carousel Wing Nut
2014/08 6 - 29
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6. Carefully clean the bottom of the metal springs with an alcohol pad; dry
thoroughly with a lint-free cloth.
Figure 6-7: Metal Springs
CAUTION: Make sure fluid does not come into contact with the reference
disk located on the opposite side of the carousel.
Figure 6-8: Reference Disk
7. Position the carousel in the instrument with the spring side up.
8. Hold the carousel and tighten the wing nut just enough to hold the
carousel firmly in position.
6 - 30 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
9. Position the pak manager lid on the carousel and push down on the
release knobs until each snaps into place.
10. Close the instrument lid and the left front door, ensuring the door is
securely latched.
11. Run a System Check. For more information, see “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples and Reporting
Results.
2014/08 6 - 31
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
3. Pull up on the three release knobs holding the pak manager lid; place the
lid on the rotor table.
Figure 6-10: Pak Manager Lid Release Knobs
4. Hold the pak carousel then loosen and remove the wing nut in the
center.
Figure 6-11: Pak Carousel Wing Nut
6 - 32 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
5. Lift the carousel out of the instrument and turn it over; locate the small
glass reference disk on the bottom.
Figure 6-12: Reference Disk
6. Lightly dampen lens paper with deionized water and clean the lens.
WARNING: Be certain water does not leak behind the glass onto the
reference disk.
7. Dry the area thoroughly with clean lens paper.
2014/08 6 - 33
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
8. Locate the fluorometer window in the black square to the right of the
center bolt.
Figure 6-13: Fluorometer Window
6 - 34 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
2. Insert a long, slender tool into the lid release; press firmly and lift the
instrument lid.
Figure 6-14: Lid Release
3. Locate the disk in the pak manager lid; pull up on the two release knobs
and remove the lid disk; set the lid aside.
Figure 6-15: Pak Manager Lid Disk Release Knobs
2014/08 6 - 35
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Note: The lid disk reference is located on the underside of this disk.
Figure 6-16: Lid Disk Reference
4. Lightly dampen lens paper with deionized water and clean the lid disk
reference disk.
5. Dry the area thoroughly with a clean lens paper.
6. Reinstall the pak manager lid disk in the carousel lid.
7. Press down on the two release knobs until they snap into place.
8. Close the instrument lid and the left front door, ensuring the door is
securely latched.
9. Run a System Check. For more information, see “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples and Reporting
Results.
Note: Keep the System Check report with your QC records.
10. Run Quality Control for all assays in use.
6 - 36 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
2. Insert a long, slender tool into the lid release; press firmly and lift the
instrument lid.
Figure 6-17: Lid Release
3. Pull up on the three release knobs holding the pak manager lid; place the
lid on the rotor table.
Figure 6-18: Pak Manager Lid Release Knobs
2014/08 6 - 37
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
4. Locate the gray, rubber sealing pad on the carousel to the left of the
carousel reference disk.
Figure 6-19: Gray Probe Sealing Pad
6 - 38 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
2. Lift the cover and disconnect the ground strap from the clip on the back
of the instrument; set the cover aside.
3. Open the left front door.
4. Insert a long, slender tool into the lid release; press firmly and lift the
instrument lid.
Figure 6-21: Lid Release
5. Loosen the captive screw at the front of the left side panel.
Figure 6-22: Left Side Panel Captive Screw
6. Disconnect the panel’s ground strap from the clip on the top rail of the
instrument.
2014/08 6 - 39
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
7. Lift the panel up off the mounting pins; set the panel aside.
8. Locate the rotor containment boot on the L-shaped bracket behind the
cannula chamber on the left side of the instrument.
Figure 6-23: Rotor Containment Boot
9. Reach into the instrument and loosen the thumbscrew, holding the boot
assembly against the sample manager.
10. Remove the boot assembly and flip it over to locate the rubber gasket.
6 - 40 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
11. Clean rubber gasket of boot assembly with warm, deionized water.
Figure 6-24: Clean Boot Assembly
Replacing AC Fuses
AC fuses are located in the fuse holder above the instrument power switch.
1. Power down the analyzer:
a. Select the Power icon from the Home screen.
b. Select Yes.
2. Turn the instrument power switch off.
3. Disconnect the power cord from both the analyzer and the wall outlet.
2014/08 6 - 41
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
4. Open the latch on the fuse holder using the small flathead screwdriver
from the instrument accessory kit.
Figure 6-25: Lifting Latch on AC Fuse Holder
Note: Use the groove below latch to guide the small screwdriver in
place.
5. Slide the fuse holder out of the instrument.
6. Remove both fuses; discard per facility procedure.
7. Insert the replacement fuses.
Note: Always replace fuses with the same type, 6.3A, 250V AC, from the
instrument accessory kit. Follow facility reordering procedures to
restock instrument accessory kit fuses.
8. Install the fuse holder in the instrument, ensuring it is securely latched.
9. Close the sample manager door.
10. Connect the power cord to the analyzer before plugging it into the wall
outlet.
11. Turn the instrument power switch on.
6 - 42 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
Replacing DC Fuses
Each DC fuse has an indicator light beneath it. If a light is OFF, replace the
fuse. The chart below indicates the types of DC fuses:
Table 6-6: DC Fuse Types
2014/08 6 - 43
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6 - 44 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
# Description
1 Spring Assembly
4. Remove the chamber lower seal from the bottom of the spring assembly
and discard it.
5. Attach the new chamber lower seal to the rim of the spring assembly.
6. Close the instrument lid and the left front door, ensuring the door is
securely latched.
2014/08 6 - 45
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6 - 46 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
3. Grasp the black ring at the top of the spring assembly and unscrew the
assembly from the cannula chamber.
Figure 6-30: Black Ring
4. Remove the rubber O-ring from the top edge of the chamber and discard
it.
5. Lubricate all surfaces of the new O-ring with Krytox® lubricant from the
instrument accessory kit.
2014/08 6 - 47
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6. Install the new O-ring and screw the spring assembly into the cannula
chamber.
7. Close the instrument lid and the left front door, ensuring the door is
latched.
6 - 48 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
5. Locate the existing operating system and the data USB flash drive ports
on the touchscreen board.
Figure 6-33: Operating USB and Data Flash Drive Ports
# Description
1 USB J2 Port
2 USB J1 Port
6. Remove the corrupted operating system USB flash drive; set aside.
7. Insert the spare operating system USB flash drive in the operating system
port.
8. Close instrument lid.
9. Restart the instrument. See “Restarting the System after Power
Shutdown” on page 21 of Chapter 1: System Overview.
10. Align touchscreen when the 4 calibration point touchscreen alignment
displays; select each point and move to the nearest corner of the screen.
Note: Repeat Step 9 and Step 10 if touchscreen is not aligned.
11. Run a System Check. For more information, see “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples and Reporting
Results.
Note: Keep the System Check report with QC records.
2014/08 6 - 49
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
12. Run Quality Control on all assays in use. If Quality Control values are
outside acceptable ranges, refer to see “Addressing an Out of Range QC
Result” on page 9 of Chapter 4: Quality Control.
6 - 50 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
7. Insert a long, slender tool in the right front door release; press firmly and
open the door.
Figure 6-35: Right Front Door Release
2014/08 6 - 51
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
# Description
1 Lamp Plug
10. Follow the lamp cable from the plug to the lamp. Grasp the end of the
lamp and gently pull it straight out. Discard per facility procedure.
6 - 52 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
11. Grasp the replacement lamp at the cable end; insert the new lamp.
2014/08 6 - 53
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6 - 54 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
10. Completely loosen the two captive screws attached to the back wall at
the right end of the pipettor assembly.
Figure 6-39: Right End Pipettor Assembly Captive Screws
11. Move the left side of the pipettor assembly towards the back of the
instrument; pivot the right end of the pipettor assembly towards the
front of the instrument.
12. Remove the pipettor assembly and place it on a work surface.
13. Open the two large plastic clamps.
Figure 6-40: Plastic Clamps
2014/08 6 - 55
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
14. Thread the red tubing through the ribbon cable opening.
15. Disconnect the other tube fitting and thread it through the opening in
the carriage.
Figure 6-41: Red Pipettor Tubing Threaded through Carriage
16. Remove the red tubing from all clamps and discard.
17. Thread the new red replacement tubing through the clamps and
openings.
18. Install and hand-tighten the red tube fitting at the pipettor carriage end.
19. Attach the red tubing into the small plastic clamps in the pipettor
carriage.
20. Close the large plastic clamps.
21. Insert the left end of the pipettor assembly into the instrument behind
the sample manager; hand tighten the two captive screws.
22. Align the right end with the back wall and the left end with the sample
manager; hand tighten the two captive screws.
23. Tighten the four captive screws with a screwdriver.
24. Install and hand-tighten the tube fitting to the back wall of the
instrument.
25. Connect cable J513A then cable P512A.
26. Tighten the two red tube fittings with the torque wrench from the
instrument accessory kit.
27. Close the instrument lid and the left front door; ensuring the door is
securely latched.
28. Close the sample manager door and turn the instrument power switch
on.
6 - 56 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
29. Run a System Check. For more information, see “Running System Check –
Electronic QC” on page 4 of Chapter 2: Processing Samples and Reporting
Results.
2. Lift the cover and disconnect the ground strap from the clip on the back
of the instrument; set the cover aside.
3. Open the left front door.
2014/08 6 - 57
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
4. Insert a long, slender tool into the lid release; press firmly and lift the
instrument lid.
Figure 6-43: Lid Release
5. Loosen the captive screw at the front of the left side panel; disconnect
the panel’s ground strap from the clip on the top rail of the instrument.
6 - 58 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
6. Lift the panel off its mounting pins; set panel aside.
7. Locate the blue plastic tubing attached to the upper cannula chamber.
Note: Before removing the blue tubing, make note of how it
threads from the upper cannula chamber to the T-fitting on
the side of the pump.
8. Unscrew the fitting to disconnect the blue tubing from the chamber.
Figure 6-44: Blue Plastic Tubing Fitting
9. Trace the blue tubing to the back of the instrument, removing the blue
tubing from the clamps.
2014/08 6 - 59
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
10. Disconnect the other end of the blue tubing from the T-fitting on the side
of the pump.
Figure 6-45: Disconnect Tubing from T-fitting
6 - 60 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Troubleshooting
17. Place the left side panel on its two mounting pins.
18. Tighten the captive screw.
19. Connect the ground strap to the clip on top of the instrument.
20. Support the top cover on the instrument to connect the ground strap to
the clip on the back of the instrument.
21. Reinstall the top cover and tighten the captive screws.
22. Close the instrument lid and the left front door, ensuring the door is
securely latched.
2014/08 6 - 61
Troubleshooting Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6 - 62 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
Chapter 7: Configuration
2014/08 7-1
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Configuration Overview
Note: Only trained operators should perform these procedures.
Most of these settings have been configured during installation. However, if
changes are required, follow the procedures in this chapter.
Basic settings are not password-protected. Advanced settings require a
password for access.
Basic Settings
• Audio
• Rotor Usage
• Set Language
• Time and Date
Advanced Settings
• Change Password
• Collection Information
• LIS/Middleware Settings
• Manual Dilution
• Operator ID
• QC
• Reference Intervals
• Sample Barcode
• Sample Tube Type
• Serial Number
• System Check Lockout
• Training Mode
• Units
7-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
Basic Settings
Setting Audio Alarms
Use to turn on/off audible signals for the alarm, key click, and beep. The
following audio components can be set independently:
Table 7-1: Audible Signals
Signal Description
Retrieve Sample Use this option to OFF –The rotor or sample cup is placed in the waste
retrieve a sample for container at the completion of processing.
investigation. ON – The rotor table returns the rotor or sample cup to
its starting position at the end of processing.
Rotor Hold Timeout Use this option to OFF– The rotor is placed in the waste container at the
run additional tests completion of processing.
on a sample. ON– The rotor remains in the testing position until the
amount of time specified in Timeout duration, up to
30 minutes, is surpassed.
Note: If both options are ON, rotor hold takes place before retrieve
sample.
1. From any mode screen, navigate to
Menus > Maintenance > Setup > Basic.
2. Select Rotor Usage.
3. To change the setting for Retrieve Sample select Retrieve On/Off.
2014/08 7-3
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Configuring Language
Use this setting to specify the language for screen and printouts.
Note: When Accept is selected in this procedure, the language
changes immediately. Be sure to verify the language
selection before proceeding.
1. From any mode screen, navigate to Menus > Maintenance > Setup.
2. Select Basic > Set Language.
3. Select Next Language until the appropriate language is displayed.
4. Select Accept.
5. Select No or Yes at the prompt Screens will change to new language.
Change language?
Note: The message The changes have been stored. displays at the
top of the touchscreen.
6. Select Home.
7-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
8. Select Accept.
Note: The message The changes have been stored. displays at the
top of the touchscreen.
9. Select Home.
2014/08 7-5
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Advanced Settings
Changing Password
Use this setting to change the Setup functions password. When entering an
Advanced screen, the system prompts the operator to enter a numeric
password. Only one password is accepted, but it can be changed as needed.
1. From the Home screen, navigate to
Menus > Maintenance > Setup > Advanced.
2. Enter the password; select Enter.
3. Select MORE three times.
4. Select Change Password.
Note: The password can be 1 to 8 numbers long.
5. Use the number pad on the touchscreen to enter a new password; select
Enter.
6. Enter the new password again; select Enter.
7. Select Home.
7-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
Gnd 1 Gnd
TX data 2 RX data
RX data 3 TX data
Gnd 7 Gnd
Note: Consult with your LIS administrator for the proper settings
before changing these parameters.
1. From the Home screen, navigate to
Menus > Maintenance > Setup > Advanced.
2. Enter password; select Enter.
3. Select MORE.
4. Select LIS Settings.
5. Use the arrows on the touchscreen to navigate to the appropriate field.
2014/08 7-7
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6. Ensure that the Interface Mode is OFF before enabling another mode.
7. Select Next Setting until the appropriate setting is displayed.
Table 7-4: LIS/Network Settings
Data Bits 7, 8
Stop Bits 1, 2
7-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
Configuring an Operator ID
Use this setting to indicate if a valid Operator ID must be entered for
processing functions and whether the Operator IDs should be securely
displayed when printed. This option is also used to enter a list of valid operator
IDs. If an ID is required, the red checklist ID light remains illuminated until a
valid ID is entered on the mode screen. If the Mask Operator ID option is ON
only the first 4 digits display followed by asterisks on all printouts displaying
the Operator ID. The Calibration screen is the only screen that displays the
Operator ID as masked when this option is ON.
If Operator ID is required for processing, the system can store up to 500 valid
Operator IDs.
1. From any mode screen, navigate to
Menus > Maintenance > Setup > Advanced.
2. Enter password; select Enter.
3. Select MORE two times.
4. Select Operator ID.
5. Select Oper ID On/Off to display the appropriate setting; select Accept.
Note: The message The changes have been stored. displays at the
top of the touchscreen.
2014/08 7-9
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
6. If Operator ID is required:
a. Select Valid Oper IDs.
b. Use the arrows on the touchscreen to highlight the appropriate
field. Enter or change the ID number.
c. Select Enter.
7. If Operator IDs should be masked to display only the first 4 digits:
a. Select Mask ID On/Off.
b. Ensure option is ON.
c. Select Accept.
8. To print a list of Operator IDs, select Print.
Note: Operator IDs entry fields will print regardless of how many
valid Operator IDs have been entered.
9. Select Accept.
Note: The message The changes have been stored. displays at the
top of the touchscreen.
10. Select Home.
7 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
2014/08 7 - 11
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
BD Clear
SARSTEDT Green
7 - 12 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
Training Mode
Use this setting to allow or disallow the use of Training (TRNG) Paks. The Quick
Reference Guide that comes with the analyzer explains how to activate and
use this mode.
Note: Do not attempt to process samples when this mode is on.
Training Mode is deactivated when the analyzer is powered
down.
Configuring Units
Use this setting to specify the units for each assay. Units are displayed and
printed with test results, QC results and calibration results.
Primary units are the main unit of measure that will be reported for each assay
and are established during analyzer installation.
Secondary units are a second unit of measure that can be reported along with
the primary unit of measure. The secondary reporting unit will display in
brackets under the primary reporting unit. Secondary units cannot be
configured. The feature can be toggle on or off for the following assays:
Table 7-6: Assays Supporting the Reporting of Secondary Units
With the secondary units enabled, the operator has two options:
• The secondary units can be displayed on the result printout, but not
transmitted to the LIS.
• The secondary units can be displayed on the result printout and
transmitted to the LIS.
For more information to configure the Transmit Secondary feature, see
“Configuring LIS / Network Settings” on page 7.
Configuring reporting options for secondary units is as follows:
• To enable printing of secondary units: Set Sec. Unit On to Enabled.
• To disable printing of secondary units: Set Sec. Unit On to Disabled.
Use the following procedure to configure primary units of measure and
enable/disable reporting of secondary units of measure:
1. From any mode screen navigate to:
Menus > Maintenance > Setup > Advanced.
2. Enter password; select Enter.
3. Select Units.
2014/08 7 - 13
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
7 - 14 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Configuration
4. Connect printer cable to the port on the printer; connect the USB end of
cable to the USB port on the Stratus® CS 200 Analyzer located by the
power switch.
Figure 7-1: USB Port to Connect Printer Cable
5. For more information, see “Restarting the System after Power Shutdown”
on page 21 of Chapter 1: System Overview.
Note: The printer will automatically be configured and begin
printing with the next print request.
2014/08 7 - 15
Configuration Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
7 - 16 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Safety and Specifications
2014/08 8-1
Safety and Specifications Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Safety Overview
This chapter includes:
• Safety instructions
• Descriptions of the tasks to be performed by the installers and customers
• The space, power, and environmental requirements for the
Stratus® CS 200 Analyzer
• Software installation instructions
• Qualified sample collection tubes
8-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Safety and Specifications
• Always wear gloves when working with the Stratus® CS 200 Analyzer.
• Ensure all operators are trained in proper blood-handling techniques.
• Observe all warnings and cautions stated in the procedures.
Safety Notes
Warnings and cautions are included in this guide to emphasize important
instructions. The following examples provide definitions for warnings,
cautions, and notes:
2014/08 8-3
Safety and Specifications Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Mechanical
Electrical
If a panel, cover, or guard must be removed for an electrical adjustment or
check, be extremely careful to avoid personal injury. Do not locate the
Stratus® CS 200 Analyzer within 50 feet (15.3 meters) in any direction of a
strong radiation source such as diathermy equipment.
Chemical
Refer to the assay Information for Use (IFU) for related chemistry safety
precautions.
Safety Labels
The labels described in Table 8-1 appear in various places on the instrument
and in this Operator’s Guide. Understand their meanings before opening the
doors and lids or completing procedures.
Table 8-1: Safety Labels
Potential Biohazard Indicates an area of the instrument that may contact biohazards. Do not
handle the contents or touch the area unless properly protected. Refer to
applicable facility procedures.
Pinch/Crush Hazard Indicates an area of the instrument where the operator could be exposed
to moving parts. Use caution when performing procedures in these
areas.
Puncture Hazard Indicates an area of the instrument where the operator could be exposed
to sharp tips, particularly probes, that could puncture the skin. Use
caution when performing procedures in these areas.
8-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Safety and Specifications
Lamp Precaution Do not touch the lamp bulb or the quartz envelope.
Protective Earth Terminal Indicates the terminal connecting power supply external earth (ground)
to the instrument.
Electrostatic Discharge Indicates the potential for damage to electronics from static discharge.
Always use the ground strap when handling static-sensitive instrument
components.
Burn Hazard Indicates a heated area of the instrument that can burn an
operator. Be sure to power off the instrument and let these
areas cool before touching them or performing diagnostic and
maintenance operations around them.
2014/08 8-5
Safety and Specifications Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Understanding Symbols
8-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Safety and Specifications
Installation Responsibilities
2014/08 8-7
Safety and Specifications Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Space Requirements
Table 8-2: Instrument Space Specifications
Power Requirements
Table 8-3: Instrument Power Specifications
8-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Safety and Specifications
Transient Overvoltage
Installation Category II (branch circuit)
Circuit
The instrument must have a separate, dedicated line with hot, neutral, and
isolated ground in its own conduit. The conduit should start at the distribution
panel and be continuous to the receptacle. Three-wire distribution to the
receptacle is required. Connect the third, green ground wire from the
distribution panel to the receptacle in accordance with NEC paragraph 250-74,
exception 4, unless local codes prohibit. The ground wire should not be tied to
grounds from other loads.
Receptacle
Customer must provide a Hospital Grade receptacle, installed by a qualified
electrician before arrival of the instrument. The receptacle must be accessible
to the 9 feet (2.7 meters) power cord furnished with the instrument. North
America requires NEMA #5-20R 20 amp straight blade receptacle (Hubbell
receptacle No. IG-8310 or equivalent).
Electromagnetic Radiation
Do not locate the instrument within 15 meters (50 feet) in any direction of an
electromagnetic radiation source such as diathermy apparatus.
For more information about the computer interface, see “Configuring LIS /
Network Settings” on page 7 of Chapter 7 and the appropriate Interface
Specifications document supplied by Siemens Healthcare Diagnostics.
2014/08 8-9
Safety and Specifications Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Environmental Specifications
Waste Disposal
Temperature
• Maintain room temperature between 15° C to 30° C (59° F to 86° F)
• Do not exceed a maximum fluctuation of 3° C (5.4° F) per hour
Humidity
Relative humidity (non condensing) should range between 20 and 80 percent.
Thermal Output
200 W (682 BTU/hr)
Noise Output
Average noise under 65 dBA.
Noise Rating: NR 40.
Centrifuge Speed
During each processing run, a microprocessor independently verifies that the
centrifuge speed is between 18,000 rpm and 22,000 rpm. An error message is
generated if the speed is outside this range.
8 - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix A: Menu Navigation
2014/08 9-1
Appendix A: Menu Navigation Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Touchscreen
Table 9-1: Touchscreen Icons
Icon Description
Power Select Power icon to power on or off the instrument. See “Controlled
Power Shutdown” on page 21 of Chapter 1: System Overview.
Paper Advance Select Paper Advance icon to feed the printer paper up through the paper
slot.
Home Select Home icon to return to the Home screen at any time.
Back Arrow Select the Back Arrow to return to the previously displayed screen or to
initiate a controlled shutdown. See “Software Overview” on page 19 of
Chapter 1: System Overview. This button does not stop processing.
Help Select Help to display information about the current screen, if available.
Select the Back Arrow icon to return to the previous screen.
Reset
Select Reset to move all components to their home positions. This function is
used primarily for troubleshooting.
9-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix A: Menu Navigation
Reports
Select Reports to access the following functions:
Print Results
Transmit Results
Calibration Status
About SCS
Maintenance
The following tables show the keys to select to access maintenance functions.
Setup functions are used primarily for system configuration.
Alignments functions are used primarily for troubleshooting.
Mfg Tests and Diagnostic should be used only when directed by the Siemens
Technical Solutions Center or the local support team.
2014/08 9-3
Appendix A: Menu Navigation Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Set Language
Audio
Time/Date
Advanced ---> QC
Ref Intervals
Units
LIS Settings
Collection Info
Serial Number
Training Mode
Change Password
MORE (returns to
Advanced menu)
9-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix A: Menu Navigation
Diagnostics
Pak Manager ----> Pak Entry Pak Carousel Pak Waste Misc Diagnostic
Init Init Init Temp
Step Step Step Pak Barcode
Jog Jog Jog Printer Test
Cycle Cycle Cycle MORE
KeyBoard
Test
Exercise
MORE
2014/08 9-5
Appendix A: Menu Navigation Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
9-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix B: Code Compliance
2014/08 10 - 1
Appendix B: Code Compliance Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Code Compliance
Safety Compliance
The Stratus® CS 200 STAT Fluorometric Analyzer complies with safety
standards UL61010-1, CSA C22.2 #1010.1 and EN61010-1 under the following
environmental conditions [subclause 1.4]:
Table 10-1: Environmental Conditions
Emission Compliance
The Stratus® CS 200 Analyzer complies with CISPR 11 Class A. In a domestic
environment it may cause radio interference; if this occurs, take measures to
mitigate the interference by moving the Stratus® CS 200 Analyzer away from
the affected device or operate the Stratus® CS 200 Analyzer in a shielded
location.
Barcode Reader
The barcode reader uses Class I LED’s [light-emitting diodes]. There is no eye
hazard.
10 - 2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix C: Qualified Sample Collection Tubes
2014/08 11 - 1
Appendix C: Qualified Sample Collection Tubes Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Cannula
Manufacturer Tube Type / Description
Color
11 - 2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix D: Maintenance Logs
2014/08 12 - 1
Appendix D: Maintenance Logs Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
12 - 2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix D: Maintenance Logs
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
2014/08 12 - 3
Appendix D: Maintenance Logs Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
12 - 4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix D: Maintenance Logs
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
2014/08 12 - 5
Appendix D: Maintenance Logs Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
12 - 6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix D: Maintenance Logs
2014/08 12 - 7
Appendix D: Maintenance Logs Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
12 - 8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix E: Operator’s Training Checklist
ATTENTION!
Do Not Mark on These Operator’s Training Checklists
Use Them as Masters and Make Copies as Needed
2014/08 13 - 1
Appendix E: Operator’s Training Checklist Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
13 - 2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix E: Operator’s Training Checklist
Department:______________________________Date: _____________________
Review training materials: Online PEP Training, Quick Reference (flip chart), Basic Operation Sheet, Training Paks, and
Operator’s Guide
Review proper sample handling: qualified collection tube with heparin; inverted to mix
Symbol Description
Required
Optional
2014/08 13 - 3
Appendix E: Operator’s Training Checklist Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
Department:______________________________Date: _____________________
Use Training Paks to process Patient Tube; Identify when DilPaks are required
Review procedure to contact facility personnel, biomed personnel, and the Siemens Technical Solutions Center or
local support team
Symbol Description
Required
Optional
13 - 4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Appendix E: Operator’s Training Checklist
Department:______________________________Date: _____________________
2014/08 13 - 5
Appendix E: Operator’s Training Checklist Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
13 - 6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
A
above assay range
message 2-12
accessory kit
instrument 5-2
tools 5-2
adjusting
fluorometer 6-23
touchscreen display 6-22
advance
paper 1-19, 5-10
advanced settings
configure 7-6
after installation
customer responsibilities 8-7
air filter
replace and clean 5-4
alarm
setting sounds 7-3
turning off 7-3
alert
System Check message 4-5
aligning
system components 6-26
alignment
auto align 6-26
apply
barcode label 2-6
arrows
move cursor 1-17
asterisks
result report 2-17
audio alarms
set 7-3
automatic
dilution 1-14
B
back arrow
exit icon 1-17
icon 9-2
2014/08 14 - 1
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
C
C terms 3-2
calibration
coefficients 3-2
deleting TestPak lot 3-5
modes 3-3
overview 3-2
principles 3-2
processing 1-14, 3-3
report 3-3
review 2-5
status 2-6, 3-4
TestPak lot expiration 3-4
troubleshooting 3-4
when to perform 3-2
calibration status
messages 2-5
calibration with POCT1-A2 enabled
transmit 3-5
CalPak 1-5
barcode label 3-2
description 3-2
using 3-3
cannula 1-5
chamber seals 6-43
description 1-12
load 2-10
tube type 7-12, 11-2
cannula chamber lower seal
clean 6-43
cannula chamber o-ring seal
replace 6-45
cannula chamber seal
clean 5-6
carousel
pak 1-12
carousel reference disk
clean 6-30
carousel springs
clean 6-27
i-2 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
caution
chemical 8-4
electrical 8-4
general 8-3
mechanical 8-4
safety 8-3
centrifuge
description 1-12
installation speed 8-10
changing
date and time 7-4
checking
reagent status 2-6
thermal printer paper 2-4
tip supply 2-3
checklist
training 13-1
using 13-1
chemical
warning and cautions 8-4
circuit
power requirements 8-9
cleaning
cannula chamber lower seal 6-43
cannula chamber seal 5-6
carousel reference disk 6-30
carousel springs 6-27
fluorometer 6-30
instrument surfaces 5-2, 5-5
pak manager lid reference disk 6-33
pak shield 5-7
probe sealing pad 6-35
rotor containment boot and bracket 6-37
sample manager 5-6
tip chute 5-8
touchscreen 5-9
waste container area 5-9
code
compliance 10-2
error 6-7
coefficients
on calibration report 3-3
terms 3-2
collection
configure date and time 7-6
sample tubes 11-2
specimen 2-6
compliance
code 10-2
emission 10-2
safety 10-2
computer interface specifications
installation 8-9
configure
2014/08 i-3
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
D
daily setup 2-2
data
delete 1-17
date
change 7-4
format 7-4
DC fuses
replacing 6-42
deleting
data 1-17
TestPak lot 3-5
description
CalPak 3-2
menus 1-19
modes 1-19
pipettor 1-13
rotor 1-12
i-4 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
E
electrical
warnings and cautions 8-4
electromagnetic radiation
location of instrument 8-9
emission
compliance 10-2
enter
Operator ID 2-8
Patient ID 2-8
Sample ID 2-8
store data 1-17
environmental specifications
installation 8-10
error
codes and messages 6-14
log 6-3
message 6-2, 6-4
print 6-3
errors
codes 6-7
interpreting processing 6-2
miscellaneous 6-4
troubleshooting codes 6-2
exit
back arrow 1-17
icon 9-2
expiration
TestPak lot 3-4
external printer
configuring 7-14
F
failed
calibration 3-4
QC 4-9
flash drive
replace corrupted 6-47
flowchart
software 9-3
2014/08 i-5
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
fluid delivery
System Check 1-15
fluorometer
adjusting 6-23
clean 6-30
description 1-13, 1-14
lamp replacement 6-49
troubleshooting 6-9
window 6-30
format date 7-4
functions
menu 1-19
fuses
replace AC 6-40
replace DC 6-42
G
general
warning and cautions 8-3
H
help
icon 1-17, 9-2
hold
rotor option 7-3
home
icon 1-18, 9-2
home position
reset 9-2
humidity
specifications 8-10
I
icon
back arrow 9-2
exit 9-2
help 1-17, 9-2
home 1-18, 9-2
paper advance 1-18, 9-2
power 1-18, 9-2
ID
operator 3-3, 7-9
patient 2-8, 7-11
QC 4-4
sample 2-8, 7-11
System Check 2-4, 4-5
illuminating
tip trays 1-18
inspect
rotor 2-10
installation
centrifuge speed 8-10
computer interface specifications 8-9
i-6 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
K
key
symbols 8-6
L
label
apply barcode 2-6
barcode 1-12, 7-11
safety 8-4
tube 2-6
language
changing 7-4
levels
QC 4-5
lid reference disk
clean 6-33
lights
DC fuses 6-42
limits
review system check 2-5
LIS
configure 7-7
list
Operator ID 7-9
loading
cannula 2-10
rotor 2-10
TestPak 2-12, 3-3, 4-8
lockout
QC 4-5
System Check 4-5
lockout parameters
QC 4-5
log
error 6-3
maintenance 12-1
monthly maintenance 12-5
2014/08 i-7
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
setup 12-3
M
maintenance
daily setup 2-2
instrument log 12-7
logs 12-1
menu functions 9-3
monthly log 12-5
monthly overview 5-3
overview 5-2
standard 5-2
using logs 12-1
manager
pak 1-12
sample 1-12
tip 1-12
manual
dilution 7-9
manual dilution
configure 7-9
processing 2-12
mechanical
warnings and cautions 8-4
mechanical alignment
System Check 1-15
menus
description 1-19
functions 1-19, 9-2
mode 1-19
reports 9-3
message
error 6-4
messages
above assay range 2-12
calibration status 2-5, 3-4
dilution 2-12
error 6-2, 6-4, 6-14
QC alert 4-4
System Check alert 4-5
modes
calibration 3-3
description 1-19
menus 1-19
modifying 1-19
processing 1-19
QC 1-19, 4-7
System Check 1-19
training 7-13
modifying
modes 1-19
mover cursor
arrows 1-17
i-8 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
N
network setting
configure 7-7
noise output
specifications 8-10
North America
wire size 8-9
notes
safety 8-3
number pad 1-18
O
operation
overview 1-12
Operator ID
entering 2-8
list 7-9
required 7-9
optical detection
System Check 1-15
overview
calibration 3-2
configuration 7-2
maintenance 5-2
monthly maintenance 5-3
operation 1-12
QC 4-2
safety and specifications 8-2
software 1-19
system 1-2
troubleshooting 6-2
P
pak
barcode reader 1-12
pak carousel 1-12, 1-13
description 1-12, 1-13
reference disk 6-30, 6-33
pak manager 1-13
description 1-12, 1-13
lid reference disk 6-33
troubleshooting 6-11
pak shield
clean 5-7
paper
advance 1-19, 5-10
replace 5-10
paper advance
icon 1-18, 9-2
printer 5-10
touchscreen 1-19
parameters
setting up QC 4-4
2014/08 i-9
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
password
configure 7-6
patient
result 1-15
Patient ID
configure 7-11
entering 2-8
patient results
report 1-15
troubleshooting 2-16
pipettor 1-14
description 1-13, 1-14
troubleshooting 6-10
tubing 6-53
placing instrument
electromagnetic radiation 8-9
planning
installation 8-8
plasma
in sample cup 1-13, 2-7
processing 1-13
specimen 1-13, 2-7
POCT1-A2
configure 7-7
POST test 1-21
power
controlled shutdown 1-21
icon 1-18, 9-2
ON/OFF 1-21
switch 1-21
power requirements
circuit 8-9
installation 8-8
receptacle 8-9
recommended service outlet 8-8
transient overvoltage 8-9
preparing
plasma 2-7
to process 2-6
primary units of measure
configure 7-13
principles
calibration 3-2
print
errors 6-3
printer
paper advance 5-10
paper replacement 5-10
probe sealing pad
clean 6-35
procedures
troubleshooting 6-20
process
prepare 2-6
i - 10 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
Q
QC
alert message 4-4
ID 4-4
levels 4-5, 4-7
lockout 4-5
lockout parameters 4-5
material 4-2
mode 1-19
modes 1-19, 4-7
overview 4-2
processing 1-14, 4-7
quality management 4-2
range check off 4-4
range check on 4-4
ranges 4-4, 4-9
report 4-9
setting up parameters 4-4
status 2-6
timeout 4-4
troubleshooting 4-9
QC result
OK 4-9
out 4-9
quality management
QC 4-2
R
range check
QC off 4-4
QC on 4-4
ranges
QC 4-4, 4-9
reagent status
check 2-6
review 2-5
2014/08 i - 11
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
receptacle
power requirements 8-9
recommended service outlet
power requirements 8-8
reference
configure cutoff 7-10
disk 6-30, 6-33
intervals 7-10
reference intervals
outside range 2-17
remote computer system
settings 7-7
transmit 1-16
remote transmission
configure 8-7
replacing
AC fuses 6-40
air filter 5-4
cannula chamber o-ring seal 6-45
corrupted USB flash drive 6-47
fluorometer lamp 6-49
pipettor tubing 6-53
sample manager tubing 6-56
thermal paper 5-10
waste container 2-2
report
calibration 3-3
function 9-3
patient results 1-15, 2-16
QC 4-9
reprinting
results 2-17
requiring
Operator ID 7-9
reset
function 9-2
responsibilities
install 8-7
result
patient 1-15
processing time 2-15
QC OK 4-9
report 2-16
reprinting 2-17
units of measure 7-13
result out
QC 4-9
result report
asterisks 2-17
retrieving
sample from instrument 7-3
test sample 2-6
reusing
sample 2-6
i - 12 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
reviewing
calibration 2-5
reagent status 2-5
system check limits 2-5
System Check report 2-5
test needs 2-6
rotor
containment boot 6-37
description 1-12
hold option 7-3
inspection 2-10
load 2-10
table 1-12
timeout 7-3
usage configuration 7-3
rotor containment boot and bracket
clean 6-37
rotors 1-6
running
System Check 2-4
S
safety
barcode reader 10-2
blood collection 2-6
compliance 10-2
labels 8-4
notes 8-3
warnings and cautions 8-3
safety and specifications
overview 8-2
sample
barcode reader 1-12, 2-6
collection tubes 11-2
cup 4-8
manager 1-12
retrieve from instrument 7-3
reusing 2-6
volume 4-7
volume gauge 2-7, 7-12
sample collection
tube 11-2
sample cup 1-7
plasma 1-13, 2-7
processing 1-13, 2-13
Sample ID
entering 2-8
sample ID
configure 7-11
sample manager
troubleshooting 6-12, 6-13
tubing 6-56
sample manager tubing
replace 6-56
2014/08 i - 13
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
i - 14 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
CalPaks 1-5
cannulas 1-5
DilPaks 1-4
rotors 1-6
sample cups 1-7
TestPaks 1-4
thermal paper 2-4
tips 1-7
tubes 1-8
warning 1-4
waste container 1-8
switch
power 1-21
symbols key 8-6
System Check
description 1-15
fluid delivery 1-15
ID 4-5
lockout 4-5
mechanical alignment 1-15
mode 1-19
optical detection 1-15
report review 2-5
run 2-4
temperature 1-15
troubleshooting 6-20
system components 1-9
align 6-26
system overview 1-2
T
table
rotor 1-12
temperature
specifications 8-10
System Check 1-15
troubleshooting 6-13
test
processing 2-7
retrieve sample 2-6
review needs 2-6
TestPak 1-4
barcode label 3-2
delete lot 3-5
load 2-12, 3-3, 4-8
thermal output
specifications 8-10
thermal printer paper
check 2-4
time
change 7-4
remaining to process 2-15
timeout
QC 4-4
2014/08 i - 15
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
rotor 7-3
tip 1-7
disposal 1-13, 1-14
installing tray 2-3
manager 1-12
trays 1-18
waste chute 5-7
tip chute
clean 5-8
tip manager
troubleshooting 6-13
tip supply
check 2-3
tip tray
illuminates 1-18
tools
instrument accessory kit 5-2
reference pak 6-33
torque wrench 6-55
torque wrench 6-55
touchscreen
adjusting display 6-22
clean 5-9
description 1-17
number pad 1-18
paper advance icon 1-19
start key 1-18
stop key 1-18
training
checklist 13-1
mode 7-13
using checklist 13-1
transient overvoltage
power requirements 8-9
transmitting
calibration with POCT1-A2 enabled 3-5
remote computer system 1-16
troubleshooting
calibration 3-4
control system 6-9
error codes 6-2, 6-14
fluorometer 6-9
overview 6-2
pak manager 6-11
patient results 2-16
pipettor 6-10
procedures 6-20
pump system 6-12
QC 4-9
sample manager 6-12, 6-13
start key 1-18, 7-6
System Check 6-20
temperature 6-13
tip manager 6-13
i - 16 2014/08
Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide Index
tube 1-8
barcode label 2-8
configure 7-12
label 2-6
specimen 1-14, 11-2
specimen collection 1-12, 2-6
type 7-12, 11-2
tube type
cannula 7-12, 11-2
tubing
pipettor 6-53
replace pipettor 6-53
replace sample manager 6-56
sample manager 6-56
U
units of measure
configure 7-13
result 7-13
updating
calibration 3-2
USB flash drive
replace corrupted 6-47
using
CalPak 3-3
maintenance logs 12-1
V
volume
sample 4-7
volume gauge
sample 2-7
W
warnings
chemical 8-4
electrical 8-4
general 8-3
mechanical 8-4
safety 8-3
supplies 1-4
waste
area 5-9
container 1-8, 1-13, 1-14
disposal 8-10
replace container 2-2
waste chute
tip 5-7
whole blood
processing 1-12, 2-7
specimen 1-12
wire size
North America only 8-9
2014/08 i - 17
Index Stratus® CS 200 STAT Fluorometric Analyzer Operator’s Guide
i - 18 2014/08