Chapter 07 Treatment Planning
Chapter 07 Treatment Planning
Version 2012
IAEA
International Atomic Energy Agency
CHAPTER 7. TABLE OF CONTENTS
7.1 Introduction
7.2 Volume Definition
7.3 Dose Specification
7.4 Patient Data Acquisition and Simulation
7.5 Clinical Considerations for Photon Beams
7.6 Treatment Plan Evaluation
7.7 Treatment Time and Monitor Unit Calculations
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.Slide 1
7.1 INTRODUCTION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 1
7.1 INTRODUCTION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 2
7.1 INTRODUCTION
Beam energies
and
Field sizes
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 3
7.1 INTRODUCTION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 4
7.1 INTRODUCTION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 6
7.1 INTRODUCTION
SSD technique
The distance from the source to the surface of the patient
is kept constant for all beams.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 7
7.1 INTRODUCTION
SAD technique
The center of the target volume is placed at the machine
isocenter, i.e. the distance to the target point is kept
constant for all beams.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 8
7.1 INTRODUCTION
Note:
In contrast to SSD technique,
the SAD technique requires
no adjustment of the patient
setup when turning the gantry
to the next field.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.1 Slide 9
7.2 VOLUME DEFINITION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2 Slide 1
7.2 VOLUME DEFINITION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2 Slide 2
7.2 VOLUME DEFINITION
GTV
CTV
GTV
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.2 Slide 1
7.2 VOLUME DEFINITION
7.2.2 Clinical Target Volume (CTV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.2 Slide 2
7.2 VOLUME DEFINITION
7.2.2 Clinical Target Volume (CTV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.2 Slide 3
7.2 VOLUME DEFINITION
7.2.3 Internal Target Volume (ITV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.3 Slide 1
7.2 VOLUME DEFINITION
7.2.3 Internal Target Volume (ITV)
ITV
CTV
CTV
PTV
ITV
CTV
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.4 Slide 1
7.2 VOLUME DEFINITION
7.2.4 Planning Target Volume (PTV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.4 Slide 2
7.2 VOLUME DEFINITION
7.2.4 Planning Target Volume (PTV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.4 Slide 3
7.2 VOLUME DEFINITION
7.2.4 Planning Target Volume (PTV)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.4 Slide 4
7.2 VOLUME DEFINITION
7.2.5 Organ at Risk (OAR)
PTV
ITV
CTV
OAR
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.5 Slide 1
7.2 VOLUME DEFINITION
7.2.5 Organ at Risk (OAR)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.2.5 Slide 2
7.3 DOSE SPECIFICATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 1
7.3 DOSE SPECIFICATION
frequency dose-area
histogram for the PTV
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 2
7.3 DOSE SPECIFICATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 3
7.3 DOSE SPECIFICATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 4
7.3 DOSE SPECIFICATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 5
7.3 DOSE SPECIFICATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.3 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.1 Need for patient data
Example:
CTV: mediastinum (violet)
OAR:
• Both lungs (yellow)
• Spinal cord (green)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.1 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.1 Need for patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.1 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.1 Need for patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.1 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
2D treatment planning
A single patient contour, acquired using lead wire or
plaster strips, is transcribed onto a sheet of graph paper,
with reference points identified.
Simulation radiographs are taken for comparison with port
films during treatment.
For irregular field calculations, points of interest can be
identified on a simulation radiograph, and SSDs and
depths of interest can be determined at simulation.
Organs at risk can be identified and their depths
determined on simulator radiographs.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
3D treatment planning
CT dataset of the region to be treated is required with a
suitable slice spacing (typically 0.5 - 1 cm for thorax,
0.5 cm for pelvis, 0.3 cm for head and neck).
An external contour (representative of the skin or
immobilization mask) must be drawn on every CT slice
used for treatment planning.
Tumor and target volumes are usually drawn on CT
slices.
Organs at risk and other structures should be drawn in
their entirety, if dose-volume histograms are to be
calculated.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.2 Nature of patient data
A digitally reconstructed
radiograph with super-imposed
beam’s eye view for an
irregular field
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.2 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
• There is always heavy demand for the use of treatment units for
actual patient treatment.
• Using them for simulation is therefore considered an inefficient
use of resources.
• These machines operate in the megavoltage range of energies
and therefore do not provide adequate quality radiographs for a
proper treatment simulation.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
Therefore, dedicated
equipment – fluoroscopic
simulator - has been
developed and was widely
used for radiotherapy
simulation.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.3 Treatment simulation
Modern simulation
systems are based on
computed tomography
(CT) or magnetic
resonance (MR) imagers
and are referred to as CT-
simulators or MR-
simulators.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.3 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
Example:
Precision
required in
radiosurgery
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 8
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.4 Patient treatment position and immobilization devices
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.4 Slide 9
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.5 Patient data requirements
Examples:
• Treatment with a direct field.
• Parallel and opposed fields.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.5 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.5 Patient data requirements
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.5 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.5 Patient data requirements
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.5 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.5 Patient data requirements
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.5 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.5 Patient data requirements
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.5 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 8
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.6 Conventional treatment simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.6 Slide 9
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
Scout films
Pilot or scout films are obtained by keeping the x-ray
source in a fixed position and moving the patient
(translational motion) through the stationary slit beam.
The result is a high definition radiograph which is
divergent on the transverse axis, but non-divergent on the
longitudinal axis.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
Scout films
The target position can also be determined through
comparison between the CT scout and pilot films.
Note: A different magnification between simulator film and
scout film must be taken into account.
This procedure allows for a more accurate determination
of tumor extent and therefore more precise field definition
at the time of simulation.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 5
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.7 Computed tomography-based conventional simulation
Scout films
If scanned in treatment position, field limits and
shielding parameters can be directly set with respect to
the target position, similar to conventional treatment
simulation.
The result is that the treatment port more closely conforms
to the target volume, reducing treatment margins around
the target and increasing healthy tissue sparing.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.7 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
Virtual Simulation
Virtual simulation is the treatment simulation of patients
based solely on CT information.
The premise of virtual simulation is that the CT data can
be manipulated to render synthetic radiographs of the
patient for arbitrary geometries.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
CT-Simulator
Dedicated CT scanners for use in radiotherapy treatment
simulation and planning have been developed.
They are known as
CT-simulators.
Example of a modern
CT-simulator
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
Virtual Simulation
Synthetic radiographs can be produced by tracing ray-
lines from a virtual source position through the CT data of
the patient to a virtual film plane and simulating the
attenuation of x-rays.
Synthetic radiographs are called
Digitally Reconstructed Radiographs (DRRs).
Advantage of DRRs is that anatomical information may be
used directly in the determination of treatment field
parameters.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
Example of a DRR
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 6
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 7
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.8 Computed tomography-based virtual simulation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.8 Slide 8
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.9 Conventional simulator vs. CT simulator
Conventional simulator
Advantage Disadvantages
Useful to perform a Limited soft tissue contrast.
fluoroscopic simulation Tumour mostly not visible.
in order to verify Requires knowledge of
isocenter position and tumor position with respect
field limits as well as to to visible landmarks.
mark the patient for
treatment. Restricted to setting field
limits with respect to bony
landmarks or anatomical
structures visible with the
aid of contrast.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.9 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.9 Conventional simulator vs. CT simulator
CT simulator
Advantages Disadvantages
Increased soft tissue Limitation in use for some
contrast. treatment setups where
Axial anatomical patient motion effects are
information available. involved.
Delineation of target and Require additional training
OARs directly on and qualification in 3D
CT slices. planning.
Allows DRRs.
Allows BEV.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.9 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.9 Conventional simulator vs. CT simulator
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.9 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.9 Conventional simulator vs. CT simulator
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.9 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.10 Magnetic resonance imaging for treatment planning
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.10 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.10 Magnetic resonance imaging for treatment planning
Disadvantage of MRI
It cannot be used for radiotherapy simulation and planning for
several reasons:
The physical dimensions of the MRI and its accessories limit the use
of immobilization devices and compromise treatment positions.
Bone signal is absent and therefore digitally reconstructed
radiographs cannot be generated for comparison to portal films.
There is no electron density information available for heterogeneity
corrections on the dose calculations.
MRI is prone to geometrical artifacts and distortions that may affect
the accuracy of the treatment.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.10 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.10 Magnetic resonance imaging for treatment planning
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.10 Slide 3
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.10 Magnetic resonance imaging for treatment planning
MR CT
On the left is an MR image of a patient with a brain tumour. The target has
been outlined and the result was superimposed on the patient’s CT scan.
Note that the particular target is clearly seen on the MR image but only
portions of it are observed on the CT scan.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.10 Slide 4
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.11 Summary of simulation procedures
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.11 Slide 1
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.11 Summary of simulation procedures
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.11 Slide 2
7.4 PATIENT DATA ACQUISITION AND SIMULATION
7.4.11 Summary of simulation procedures
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.11 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.1 Isodose curves
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.1 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.1 Isodose curves
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.1 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.1 Isodose curves
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.1 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.1 Isodose curves
Isodose curves for a fixed SSD beam Isodose curves for an isocentric beam
normalized at depth of dose normalized at the isocenter
maximum
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.1 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Motorized wedge:
It is a similar physical device, integrated into the head of the unit and
controlled remotely.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Physical wedge:
A set of wedges (15°, 30°, 45°, and 60°) is usually provided with the
treatment machine.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Dynamic wedge:
Produces the same wedged intensity gradient by having one jaw
close gradually while the beam is on.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 5
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Example 1:
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 6
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Example 2:
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 7
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 8
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Example:
A wedge pair of 6 MV beams
incident on a patient.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 9
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.2 Wedge filters
Wedge factor:
Wedge factor is defined as the ratio of dose at a specified
depth (usually zmax) on the central axis with the wedge in
the beam to the dose under the same conditions without
the wedge.
This factor is used in monitor unit calculations to
compensate for the reduction in beam transmission
produced by the wedge.
Wedge factor depends on depth and field size.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.2 Slide 10
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.3 Bolus
depth
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.3 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.3 Bolus
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.3 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.3 Bolus
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.3 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.3 Bolus
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.3 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.3 Bolus
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.3 Slide 5
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.4 Compensating filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.4 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.4 Compensating filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.4 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.4 Compensating filters
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.4.9 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.5 Corrections for contour irregularities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.5 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.5 Corrections for contour irregularities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.5 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.5 Corrections for contour irregularities
<1 0.8
60Co –5 0.7
5 – 15 0.6
15 – 30 0.5
> 30 0.4
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.5 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.5 Corrections for contour irregularities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.5 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.5 Corrections for contour irregularities
3. TAR method
Tissue-air ratio (TAR) correction method is also based on
the attenuation law, but takes the depth of the calculation
point and the field size into account.
Generally, the correction factor CF as a function of depth z,
thickness of missing tissue h, and field size f, is given by:
TAR( z h, f )
CF
TAR( z, f )
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.5 Slide 5
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
Bottom:
Taking into account the real
tissue density
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 5
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
TAR method
Dose at each point is corrected by the factor CF:
TAR( z ', rd )
CF
TAR( z, rd )
where z1 1 = 1
z’ = z1 + ρez2 + z3 z2 2 = e
and z3 3 = 1
z = z1 + z2 + z3 Point P
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 6
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
z = z1 + z2 + z3 Point P
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 7
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
and z2 2 = e
z = z1 + z2 + z3 z3 3 = 1
Point P
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 8
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.6 Corrections for tissue inhomogeneities
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.6 Slide 9
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 1
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 2
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 3
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 4
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 5
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 6
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 7
7.5 Clinical considerations for photon beams
7.5.7 Beam combinations and clinical application
Example:
A parallel-opposed beam
pair is incident on a patient.
Note the large rectangular
area of relatively uniform
dose (<15 % variation).
Isodose curves have been
normalized to 100 % at the
isocentre.
This beam combination is well suited to a large variety of
treatment sites (e.g., lung, brain, head and neck).
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 8
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 9
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 10
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 11
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 12
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 13
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 14
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
Rotational techniques
Isodose curves for
two bilateral arcs
of 120° each.
Note:
Isodose curves are
tighter along the
angles avoided by
the arcs (anterior
and posterior).
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 15
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 16
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 17
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 18
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 19
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 20
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 21
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
Field matching
Field matching at the skin is
the easiest field matching
technique.
However, due to beam
divergence, this will lead to
significant overdosing of
tissues at depth and is only
used in regions where tissue
tolerance is not compromised.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 22
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
Field matching
For most clinical situations field matching is performed
at depth rather than at the skin.
To produce a junction dose
similar to that in the center z
of the open fields, beams must
be matched such that their
diverging edges match at the
desired depth z.
50 % isodose lines
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 23
7.5 CLINICAL CONSIDERATIONS FOR PHOTON BEAMS
7.5.7 Beam combinations and clinical application
Field matching
For two adjacent fixed SSD fields of different lengths L1
and L2, the surface gap g required to match the two fields
at a depth z is: z z
g 0.5L1 0.5L2 SSD
SSD
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.5.7 Slide 24
7.6 TREATMENT PLAN EVALUATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6 Slide 1
7.6 TREATMENT PLAN EVALUATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6 Slide 2
7.6 TREATMENT PLAN EVALUATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6 Slide 3
7.6 TREATMENT PLAN EVALUATION
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6 Slide 4
7.6 TREATMENT PLAN EVALUATION
7.6.1 Isodose curves
Example:
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.1 Slide 1
7.6 TREATMENT PLAN EVALUATION
7.6.1 Isodose curves
Same example:
Isodose line through [%]
the ICRU reference -150
point is 152 %.
-140
Maximum dose 154 %.
-130
-120
The 150 % isodose
curve completely -100
covers the PTV. - 70
- 50
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.1 Slide 2
7.6 TREATMENT PLAN EVALUATION
7.6.1 Isodose curves
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.1 Slide 3
7.6 TREATMENT PLAN EVALUATION
7.6.2 Orthogonal planes and isodose surfaces
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.2 Slide 1
7.6 TREATMENT PLAN EVALUATION
7.6.2 Orthogonal planes and isodose surfaces
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.2 Slide 2
7.6 TREATMENT PLAN EVALUATION
7.6.2 Orthogonal planes and isodose surfaces
Prescription isodose:
white wireframe
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.2 Slide 3
7.6 TREATMENT PLAN EVALUATION
7.6.2 Orthogonal planes and isodose surfaces
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.2 Slide 4
7.6 TREATMENT PLAN EVALUATION
7.6.3 Dose statistics
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.3 Slide 1
7.6 TREATMENT PLAN EVALUATION
7.6.3 Dose statistics
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.3 Slide 2
7.6 TREATMENT PLAN EVALUATION
7.6.3 Dose statistics
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.3 Slide 3
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 1
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
• PTV itself
Frequency
• Specific organ
in the vicinity
of the PTV.
Dose value
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 2
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
total volume
cent volume of total
volume” on the
ordinate against the
dose on the abscissa.
Dose value in Gy
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 3
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 4
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 5
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
Differential DVHs
Example: Prostate cancer
Target
Rectum
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 6
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 7
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
Integral DVHs
Example: Prostate cancer
Target
Critical structure:
rectum
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 8
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 9
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 10
7.6 TREATMENT PLAN EVALUATION
7.6.4 Dose-volume histograms
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.4 Slide 11
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
Port films
A port film is usually an
emulsion-type film, often
still in its light-tight paper
envelope, that is placed
in the radiation beam
beyond the patient.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 1
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
Port films
Two port films are available.
Depending on their sensitivity (or speed) port films can be
used for:
• Localization:
A fast film is placed in each beam at the beginning or end of the
treatment to verify that the patient installation is correct for the
given beam.
• Verification:
A slow film is placed in each beam and left there for the duration
of the treatment.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 2
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 4
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
Double exposure:
• Film is irradiated with the treatment field first.
• Then the collimators are opened to a wider setting, all shielding is
removed, and a second exposure is given to the film.
• The resulting image shows the treated field and the surrounding
anatomy that may be useful in verifying the beam position.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 5
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 6
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 7
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
1. Fluoroscopic detectors.
2. Ionisation chamber detectors.
3. Amorphous silicon detectors.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 8
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 9
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 10
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 11
7.6 TREATMENT PLAN EVALUATION
7.6.5 Treatment evaluation
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.6.5 Slide 12
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
Introductory remark:
The process of treatment planning and optimization may be
considered as completed if the calculated relative dose
distribution shows an acceptable agreement with the PTV.
By way of example, the 80 % isodose curve may well
encompasses the PTV.
It remains to determine the most important final parameter
which controls the absolute dose delivery, that is:
Treatment time (for radiation sources).
or
Monitor units (for linacs).
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 1
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
Accidents in radiotherapy
can and do happen.
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 2
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 3
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 4
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 5
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 6
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 7
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 8
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 9
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
Methods used for adding the dose at the ICRU point from
multiple fields:
1. The simplest method (usually not used):
Each field contributes to the total prescribed dose at the ICRU point
using an equal number of MU (or equal treatment time).
2. Each field contributes to the total prescribed dose at the
ICRU point with different weights.
Prescribed weights for individual fields may refer to:
• ICRU point IP (used for the isocentric techniques).
• Point of maximum dose Dmax of each field
(used for fixed SSD techniques).
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 10
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 11
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 12
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 13
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 14
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7 Slide 15
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 1
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 2
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
[%]
-150
Note: -140
ICRU
Prescribed weights -130 point
refer to the point of -120
maximum dose in -100
- 70
each field: - 50
PA W = 1.0
PRPO W = 0.8
PRPO W = 0.8
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 3
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 4
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
Step 2: For each field i, the dose at the ICRU point, Di(IP),
is calculated by (using 100 MU):
. PDD( z, A, f , E )
Di (IP) D( zmax , Aref , f , E ) RDF( A, E ) WF 100
100
where
.
D(zmax , Aref ,f ,E ) is the calibrated output of the machine
PDD( z, A, f , E ) is the percentage depth dose value
WF is the wedge factor
RDF(A,E) is the relative dose factor (see next slide)
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 5
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 6
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
(dose) = 148.96
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 8
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 9
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.1 Calculations for fixed SSD set-up
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.1 Slide 10
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 1
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 2
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
Step 2: For each field i, the dose at the ICRU point, Di(IC), is
calculated by (using 100 MU):
.
Di (IC) D( zmax , Aref , f , E ) TMR( A, z ) ISF RDF( A, E ) W F 100
where:
.
D(zmax, Aref ,f ,E ) is the calibrated output of the machine.
TMR( A, z ) is the tissue-maximum-ratio at depth z.
WF is the wedge factor.
RDF(A,E) is the relative dose factor.
ISF is the inverse-square factor (see next slide).
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 3
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
.
When the calibrated output factor D(zmax , Aref , f , E ) is used in
isocentric calculations, it must be corrected by the inverse-
square factor ISF unless the machine is actually calibrated
at the isocentre:
2
SSD zmax
ISF SSD
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 4
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
Step 3: Rescale the MUs such that the dose contributions at the
IP are proportional to the pre-defined weights and sum
up the total resultant dose using the rescaled MUs.
Field Starting Dose at Weight Weighted dose Rescaled
MU IP at IP MU
Anterior 100 73.1 1.0 100 %= 73.1 100
Left post. 100 28.7 0.7 70 % = 51.2 178
Right post. 100 28.7 0.7 70 % = 51.2 178
(dose) = 175.44
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 5
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 6
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.2 Calculations for isocentric set-ups
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.2 Slide 7
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.3 Normalization of dose distributions
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.3 Slide 1
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.3 Normalization of dose distributions
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.3 Slide 2
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.4 Inclusion of output parameters in dose distribution
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.4 Slide 1
7.7 TREATMENT TIME AND MONITOR UNIT CALCULATIONS
7.7.5 Orthovoltage and cobalt-60 units
IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 7.7.5 Slide 1