0% found this document useful (0 votes)
280 views1 page

Density, Relative Density and API Gravity: According To IP559 and ASTM D7777

This document discusses density measurements of liquid petroleum products according to industry standards. It describes using a Density2Go handheld density meter to measure the density and temperature of samples, and automatically compensating the measurements to standard reference temperatures. The document provides an example using the meter to measure the density of 2,2,4-trimethylpentane at indoor and outdoor temperatures, finding the results were precise and within acceptable error limits.

Uploaded by

AdikoDenzeyAnonh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
280 views1 page

Density, Relative Density and API Gravity: According To IP559 and ASTM D7777

This document discusses density measurements of liquid petroleum products according to industry standards. It describes using a Density2Go handheld density meter to measure the density and temperature of samples, and automatically compensating the measurements to standard reference temperatures. The document provides an example using the meter to measure the density of 2,2,4-trimethylpentane at indoor and outdoor temperatures, finding the results were precise and within acceptable error limits.

Uploaded by

AdikoDenzeyAnonh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Density, Relative Density and API Gravity

Density2Go Application Note


According to IP559 and ASTM D7777
Introduction Measurement procedure
Determination of the density, relative density or API Samples were measured at laboratory and outdoor
gravity of liquid petroleum products are important environmental temperatures. To verify the accuracy of
quality indicators and are used in quantity the Densito, a water standard was measured with the
calculations or to satisfy application, transportation, test method.
storage and regulatory requirements.

Energy Institute IP 559 and ASTM International Measurement parameters


D7777 test methods cover density determinations Fill speed: Medium
of field applications at Measurement reliability: Medium
test temperatures API table: Refined products (B) (IP 200, ASTM D1250)
between 0 and 40 °C,
and results are Results
compared to a
Indoor (24.3 °C) Outdoor (12.9 °C)
reference temperature
3
d measured in g/cm 0.6902 0.7013
of 15 °C or 60 °F.
3
d at 15 °C in g/cm 0.6989 0.6988
With the Density2Go
s 0.0001 0.0002
handheld density
SG(15 °C /15 °C) 0.6994 0.6995
meters, the sample
°API 70.8 70.8
can be analyzed no
matter the geographic Table 1: Mean results and standard deviation (s) for the
location or time of different sample series (n=6).
year. The measured
density and
temperature can automatically be compensated to:
 Density at 15 °C
Density at 60 °F
 Relative density 15 °C/15 °C
Relative density 60 °F /60 °F
 API gravity at 15 °C
API gravity at 60 °F

Samples and test standards


Fig. 1: Compensated density at 15 °C. The black dotted line
2,2,4-trimethylpentane at 24.3 °C (indoor)
represents the reference value of 2,2,4-trimethylpentane.
2,2,4-trimethylpentane at 12.9 °C (outdoor) Red dotted lines indicate the acceptable limits of error
Water standard for verification according to ASTM D7777, while blue error bars show the
standard deviation (s) of experimental values reported in
Table 1. This indicates that measurement results with the
Instruments and accessories Densito are precise and fall within the acceptable limit of
 Densito (30330857) error.

 Protective Cover (30330860)


 Water Standards (51325005) Further information
 Filling tube 600 mm (30330848) www.mt.com/Density2Go

METTLER-TOLEDO Density2Go_30466414