The document provides an overview of the Vic-3D software user interface and functionality. It describes the main menus, toolbars, windows and common tasks in Vic-3D for navigating the interface, opening and saving projects, adding image and data files, performing calibration and analysis, and viewing results. Projects in Vic-3D contain all files and information for a test, including speckle images, calibration images, output data files, and calibration information.

Uploaded by

Mew Jonathan
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views94 pages

Vic-3D Online Manual: Navigating The Help System

The document provides an overview of the Vic-3D software user interface and functionality. It describes the main menus, toolbars, windows and common tasks in Vic-3D for navigating the interface, opening and saving projects, adding image and data files, performing calibration and analysis, and viewing results. Projects in Vic-3D contain all files and information for a test, including speckle images, calibration images, output data files, and calibration information.

Uploaded by

Mew Jonathan
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 94

Vic-3D Online Manual

Navigating the Help System

You can navigate the help system by double-clicking on the items displayed on the left. You
can also press the F1 key while using Vic-3D to get context-sensitive help.

Quick Start

For a quick overview of the steps necessary for image analysis, see the Quick Start guide.

Getting More Help

For more help, please contact Technical Support .


Overview
The user inteface of Vic-3D has many of the familiar control elements found in other
applications. The image below illustrates the user interface. The most commonly used
functions can be accessed by clicking on tool buttons on the Tool Bar. The windows, such as
the Reference Window and 3D Plot windows are grouped inside a Workspace. The List
View on the left of the main window provides a quick overview of image and data files.

File Menu

The File Menu provides the following functions:

● New - creates a new project


● Open - open an existing project
● Save - save the current project
● Save As... - save the current project under a new file name
● Install module licenses - use this menu entry to activate software modules you have
purchased
● Edit plugin options - edits the settings for image codec plugins.
● Quit - quit Vic-3D

Edit Menu

The Edit Menu provides the following functions:

● Undo - undo the last editing operation in the reference image


● Redo - redo the last editing operation in the reference image
● Pan/Select - pan around in a zoomed-in image, or select an existing AOI
● Line - create a Line AOI
● Rectangle - create a Rectangle AOI
● Polygon - create a Polygon AOI
● Cut Region - cut a region from a polygon or rectangle AOI
● Delete - removes the selected AOI
● Initial Guess - show the Initial Guess Dialog to select an initial guess for the currently
active AOI

Project Menu

The Project Menu provides the following functions:

● Speckle images- adds speckle images to the project for analysis.


● Speckle image folder- adds all speckle images from a given folder.
● Calibration images - adds calibration images to the project.
● Data files - adds pre-existing output data files to the project
● Analog data - adds analog data files from Vic-Snap
● Video clip - adds generated AVI files

Calibration Menu

The Calibration Menu provides the following functions:

● Calibrate stereo system - calibrates the cameras and stereo rig simultaneously
● Adjust for cropping - adjusts for differently sized calibration vs. speckle images
● Calibrate camera 1 / 2 - calibrates camera parameters separately
● Calibrate camera orientation - calibrates stereo transformation separately
● Import calibration - imports a calibration from an existing project file

Data Menu

The Data Menu provides the following functions:

● Start analysis - shows the Run dialog to begin 3D analysis


● Apply math operation - allows manipulation of output data
● Export statistics - exports group statistics for output data
● Postprocessing options - shows a submenu to choose from various postprocessing
calculations
● Coordinate transformation - shows a submenu to allow various coordinate system
manipulation
● Export data - exports output data for use with external software

Plot Menu

The Plot Menu provides the following functions:

● 2D Plot- displays an overlay contour plot of output data


● 3D Plot- displays a 3D contour plot of output data
● Inspector- allows choice of various data inspection tools

Window Menu

The Window Menu provides the following functions:

● Cascade - organizes all MDI windows in a cascade


● Tile - tiles all MDI windows

Help Menu

The Help Menu provides the following functions:

● Manual - show this manual.


● About - show version information.

Main Toolbar
The buttons on the main toolbar control commonly used Vic-3D functions. From left to right:

● New project
● Open project
● Save project
● Add speckle images
● Add calibration images
● Calibrate stereo system
● Start analysis
● Histogram control
● Switch to 2D plot/3D plot
● Zoom in/out
● Undo/redo

The histogram control displays the gray level distribution for the currently displayed image.
The red bars on the histogram may be used to adjust the image display. Double-click on the
histogram to automatically adjust the balance, or drag the red bars to set the black and white
levels manually. Double click again to remove the balance adjustment.

The balance control is for display only and does not affect image analysis or stored images.

Animation Toolbar

The buttons on the animation toolbar allow stepping through and animating image files or
plots. The controls, from left to right:

● Play - begins automatically stepping through images/plots.


● Stop - stops the animation.
● Step Back / Step Forward - goes to previous or next image/plot.
● Loop - toggles between looping from last image to first, and bouncing from forward to
backward animation.
● Delay - this is the delay between frames, in ms.
Other Functionality

In the right corner of the status bar at the bottom of the main window, the cursor position and
image grey value is displayed when the mouse is moved inside the reference image or a
deformed image. On the left side of the status bar, a short description of tool buttons and menu
items is displayed when the mouse moves over them.

In the list view on the left side of the main window, some functions can be activated by right-
clicking. Details can be found in the appropriate sections of this menu.

Changes from 3.x

Changes from 3.x versions of Vic-3D fall in the areas of:

● Calibration
● Plotting
● Post-processing
● Data extraction

For information on transitioning existing projects and project files, please contact Technical
Support.
The Start Page
The start page in Vic-3D gives convenient access to frequently-used tasks, recent projects, and
project type selection.
Common Tasks

This section duplicates common tasks from the menu bar. Click to open a project, add speckle or
calibration images, or view this user manual.

Recent Projects

This section contains a list of the most recently accessed projects. Click on a filename to open the
project.

New Project

Depending on which modules you have purchased, Vic-3D may allow one or several project types.

● Normal two-camera system: this is the standard Vic-3D project with all speckle and
calibration images in stereo pairs.
● Split-mirror or prism system: this project type is used when only one camera, fitted with a
split-mirror system, is used for 3D correlation. For more information on configuring and
using such a system, please contact Technical Support.
● Shape only measurement: in this project type, only shape data is calculated for each image.
Displacement is not analyzed. This option can be used to track changes in shape for
configurations where the speckle pattern changes/deteriorates over time, making
displacement correlation problematic.
Projects in Vic-3D
In Vic-3D, all the files and information associated with a test are stored in a project.

Initially, projects are blank. Before completing a Vic-3D analysis, the project must contain:

● One or more speckle images , including a reference image


● A set of calibration images
● A calibration
● One or more areas of interest

Note: Adding speckle and calibration images to the project adds them by filename reference
only; they are not copied or moved on the disk.

When you run a Vic-3D analysis, the output files are stored on a disk and added (by reference)
to the project file. If the project file is not saved or if the data files are manually removed, they
will remain on the disk.

In addition to the items above, you can also choose to add auxiliary data references to the
project file:

● Generated video clips


● Analog data files from Vic-Snap

Notes

● In general, it's good practice to save project files often to avoid losing changes.
● Project files are stored in a human-readable XML format. Advanced users may choose
to edit project files directly in some special cases; for more information, contact
Technical Support.
● If all referenced image and data files are stored in the same folder, their references in the
project file will be local (filename only). If files are drawn from different folders, all
references will be absolute (full pathname) - this will make the project more difficult to
move or share. For convenience, it's good practice to keep all associated files in the
same folder.
● Once a calibration is performed and saved in the project file, the calibration images may
be discarded if desired. All the data calculated in the calibration routine is stored
numerically in the project file.
The Project Toolbar
The plot toolbar is displayed at the left side of the work area by default. It contains information
about image files, data, and calibration for the current project.

The Images Tab

This tab shows all speckle and calibration images associated with the project.

To add speckle images, select Images... Speckle images from the menu bar, or click the speckle
images icon on the main toolbar. The small red arrow indicates the reference image; to set an
image as the reference, right click and click Set as reference .
To add calibration images, select Images... Calibration images from the menu bar, or click the
calibration images icon on the main toolbar .

To remove an image or series of images, select them, right click, and click Remove or Remove
selected .

The Data Tab

The Data tab lists all output data, analog data, and video files associated with the project.

All generated output files are added to the Current data list. Output files not associated with
current speckle images are added to Other data. Double-click on a data file to view a 3D plot.
Analog data from Vic-Snap is listed under Analog data. To add an analog data file, select
File... Add Files... Add Data Files from the main menu. To view a spreadsheet of the data,
double-click the filename.

Generated video files from 2D and 3D animations are added to the Video files list. Double-
click on a video to display it in an external viewer.

The Calibration Tab


This tab is a static display of the current calibration information for the project.

Intrinsic parameters are displayed for each camera:

● sensor center
● focal length (in pixels)
● skew
● 1, 2, or 3 radial distortion coefficients, depending on selected order

Extrinsic parameters are displayed for the stereo rig:

● Alpha (rotation about the optical axis)


● Beta (stereo angle)
● Gamma (tilt)
● Tx, Ty, Tz: relative translation of camera 1
● Baseline: distance between cameras (resultant of Tx, Ty, Tz)
Speckle Images
In Vic-3D, speckle images are image or set of images taken of a specimen as it undergoes load
or motion. You may add one or multiple speckle images by selecting the Speckle images entry
from the Images menu, or by clicking the icon on the main tool bar .

If more than 300-400 images are to be added, select Images... Speckle image folder to add all
images from a specified folder. Each valid image file from the selected folder will be added to
the project; extraneous / calibration images may then be removed. Trying to add too many
images through the Speckle images dialog may result in an error due to operating system
limitations.

After adding speckle images to the project, they will be displayed in the workspace and listed
in the Images tab of the project bar as shown in the figure below.

Viewing Images

Deformed images can be displayed in the workspace by double-clicking on an entry in the


image list view. Alternatively, clicking the right mouse button on an entry of the list view will
show a popup menu providing different options, one of which is View .
When viewing deformed images, you can use the zoom in/zoom out entries in the Edit menu or
the corresponding tool buttons to change the scale of the displayed image.

Animating Images

To animate speckle images, display an image and then use the controls on the Animation
Toolbar to animate the sequence.

Changing between cameras

To change between the two cameras of the stereo-system, right-click on the image viewer. A
popup menu will be displayed as shown in Figure 3 allowing to change between the images
taken by the two cameras.
Figure 3:Right-button menu in image viewer.

Removing Images

Calibration images can be removed by selecting one or multiple images in the list view, and
right-clicking on your selection. Select Remove or Remove selected to remove images from the
list.
The Reference Image
The term Reference Image is used in this manual to describe the image of the specimen taken
while no load was applied. All displacement analyses in Vic-3D are with respect to this
reference image, i.e., the displacements are obtained in a Lagrangian coordinate system.

To select a reference image, right-click on it in the Speckle images list, and select Set as
reference.

After the reference image has been selected, it will be indicated with a red arrow in the images
list.
When the reference image is displayed, the AOI tool buttons and menu entries in the Edit menu
become active.

Selecting an Area-of-Interest

Vic-3D supports the following types of AOIs:

● Line: A number of points evenly spaced along a line. Note: data from line AOI's will
not be visible in 2D plots because it does not comprise an area. This data may be viewed
in exported files only.
● Rectangle: Points spaced on an even grid contained in a rectangular area.
● Polygon: Points spaced on an even grid contained in a polygon.

To specify a particular type of AOI, select the corresponding entry in the Edit menu or the
appropriate button on the tool bar. The selected AOI type will be indicated by the mouse
cursor.

After selecting the AOI type, move the cursor to the desired position in the reference window
and click the left mouse button. You can now move the mouse to the next position, e.g. the end
of the line or the second corner of the rectangle. Clicking the left mouse button again will
complete the AOI selection for all AOI types except polygons. For polygon selection, a double-
click is used to specify the last point of the polygon.

During AOI selection, a yellow line is drawn to indicate the outline of the selection; small
green squares indicate nodes. This is illustrated in the figure below.

Choosing the Subset and Step Size

The subset and step size can be selected after an area of interest is created. Both are adjusted
using the spin boxes in the AOI Toolbar.

The subset size controls the area of the image that is used to track the displacement between
images. The subset size has to be large enough to ensure that there is a sufficiently distinctive
pattern contained in the area used for correlation.

The step size controls the spacing of the points that are analyzed during correlation. If a step
size of 1 is chosen, a correlation analysis is performed at every pixel inside the area-of-interest.
A step size of 2 means that a correlation will be carried out at every other pixel in both the
horizontal and vertical direction, etc. Note that analysis time varies inversely with the square of
the step size; i.e., a step size of 1 takes 25 times longer to analyze than a step size of 5.

The Seed Point

The term Seed Point is used to refer to the point in the reference image where the correlation is
started. The correlation algorithms use the results from the seed point to obtain an initial guess
for the second point analyzed and continue in this manner until all points in the AOI are
analyzed.

For rectangular and polygon AOIs, the seed point can be moved to a user-specified location in
the AOI. To move the seed point, select the 'Hand' tool, or select Edit... Pan/Select from the
menu bar. Then, mouse over the seed point; the cursor will change to indicate node movement.
Click and drag to place the seed point.

The placement of the seed point can greatly influence the amount of work required to select an
initial guess . Ideally, the seed point should be placed in the area of the image that underwent
the smallest amount of motion during the test. An example is illustrated in the figure below.
In this example, the seed point should be placed on the side of the image closest to the fixed
grip of the test frame, and on the center-line of the specimen. This way, even as the test
progresses and the specimen becomes more deformed, the start location for the correlation will
experience relatively small motions.

Editing AOIs

To edit an existing AOI, select the Pan/Select tool. Mouse over any of the green nodes in your
AOI; the mouse cursor changes to indicate node movement. Click and drag to move.

Cutouts

For rectangular and polygon AOIs, the scissors tool can be used to cut areas from the AOI.
This feature is most commonly used if the specimen has cracks, holes, or other areas where
correlation is impossible.

To cut an area from an AOI, click the scissors button on the tool bar or select Edit... Cut
region. The selection of the area to be cut works like selecting a polygon AOI, i.e., corner
points of a polygon can be added by single-clicking the left mouse button, and the last point is
specified by a double-click. Once the cut is complete, new nodes are added to your AOI; these
may be moved like other nodes.

Hints

● Use the scroll wheel to adjust the size of the image.


● When using multiple AOIs for one image, click on an AOI with the pan/select tool to
activate it.
● During AOI selection, the image can be scrolled by moving the mouse outside the
reference image window. This will cause the image to autoscroll if the image does not
fit on the display.
● You can use the Undo/Redo buttons to undo AOI selection and other operations. The
Undo/Redo buttons in the Edit menu will indicate what changes can be undone/redone.
Calibration Images
Calibration images can be added by selecting the Calibration images entry from the Images
menu, or by clicking the icon on the main tool bar .

After adding calibration images to the project, they will be listed in the Images tab of the
project bar, as illustrated below.

Viewing Images

Calibration images can be displayed in the workspace by double-clicking on an entry in the list
view on the left. Alternatively, clicking the right mouse button on an entry of the list view will
show a popup menu providing different options, one of which is View.
When viewing calibration images, you can use the zoom in/zoom out entries in the Edit menu
or the corresponding tool buttons to change the scale of the displayed image.

Changing between cameras

To change between the two cameras of the stereo-system, right-click on the image viewer. A
popup menu will be displayed as shown in Figure 3 allowing to change between the images
taken by the two cameras.
Figure 3:Right-button menu in image viewer.

Removing Calibration Images

Calibration images can be removed by selecting one or multiple images in the list view, and
right-clicking on your selection. Select Remove or Remove selected to remove images from the
list.
Calibration
There are two methods to calibrate the stereo system. The normal way to calibrate a stereo
system uses a series of images of the calibration target in different orientations, where the
calibration target is visible in both cameras in all images. This method of calibration works
well for small viewing areas, where a calibration target that cover the entire field of view can
easily be handled.

The second method of calibration uses two separate series of images of a calibration target for
the left and right camera of the stereo system. The internal camera parameters, such as the focal
length and image center, are calibrated separately for each camera. The orientation of the two
cameras is determined from the speckle pattern on the specimen itself. This method requires
that two marker points with a known distance on the specimen surface are present in the
images.

For either method, first add the calibration images to the project file and then select the
appropriate entry from the Calibration menu on the main menu bar.

Method 1: Joint System Calibration


To calibrate the stereo system, select the Calibrate stereo system entry from the Calibration
menu, or click on the icon on the tool bar. The Stereo System Calibration dialog, shown
below, will appear.
Target Selection

Select the target which was used for calibration from the Target dropdown. If the target is not

listed, click the icon, and enter the parameters for your target. The individual parameters
are illustrated below.
If necessary, select or clear the Black dots checkbox to reflect the grid color; typically, this
option is selected automatically by the software.

Camera Selection

If desired, you may select the appropriate camera using the Camera dropdown. If the camera

that was used is not displayed, you may add it using the button, and entering a known
skew and aspect ratio.

Selecting a camera causes the calibration routine to use a fixed skew and aspect ratio; this can
improve accuracy of the remaining parameters, especially if calibration movements were
limited.

Extracting Target Points


After the calibration target and (optionally) camera have been selected, the image positions of
the circles can be extracted from the calibration images. To extract points from all
images, click the Auto button; each image will be analyzed in turn, with colored dots displayed
on the measured point centers.

Manual Adjustments

For some images, manual adjustments will need to be made to accurately extract points.
Typical cases include:

● Poor / uneven lighting


● Poor depth of field / defocused dots
● Highlights / reflections on grid
● Very small or very large grid

When the Threshold Image box is checked, a processed image will be displayed; this allows
evaluation of the effects of each parameter in real time. The manual adjustment options are as
follows:

● Black dots - check to select black dots on a white background; clear to select white dots.
This setting is calculated automatically by Vic-3D but in some cases may need to be set
manually.
● Glare reduction - applies a filter to the image which brings out local variations/edges;
this option is useful when parts of the grid are bright or washed out, but may cause poor
extraction in some circumstances.
● Adaptive threshold - allows local variations in black/white levels; this option can correct
for uneven lighting across the target.
● Threshold slider - sets the proportional threshold between dot features and background
levels. Set lower for darker images and higher for brighter images.
● Smoothing slider - adjusts the level of smoothing applied to the target image.
● Min. size - this controls the minimum size of features that can be considered target dots.
This value is automatically scaled with the image but may need adjustment for very
small grids, or if background features are erroneously extracted.
● Max. size - this controls the maximum size of features that can be considered target
dots. As with the minimum size, this value is set automatically but may need to be
raised if a very large grid is used.

If necessary, use these adjustments until the target dots appear well-defined in the image. After
adjustments are made, you may click Extract to re-extract a single image, or Auto to re-extract
all images. Note that the Auto feature will not overwrite existing points; to redo some or all
images, click Clear or Clear all and then Auto. If no combination of settings produces a good
extraction, confirm that the grid settings are correct (see Target Selection, above).

Note: In most cases, the only adjustment necessary is to raise or lower the Threshold bar.

If these adjustments do not produce successful extraction, please feel free to contact Technical
Support. For the best targeted response, you may email your calibration image pairs for
analysis.

System Options

Select the High-magnification checkbox if high magnification lenses were used and a poor
calibration is obtained.

For lenses with higher distortion, i.e., poor quality or short focal length lenses, select a
Distortion order of 2 or 3. For most lenses, 1 works well.

Calibration
After the features are extracted from a sufficient number of images, the Calibrate button will
be activated. Click this button to calibrate the stereo system.

Depending on the number of images used for calibration, the program may take several
seconds to calibrate the stereo system. After the calibration is completed, a calibration report
will be displayed. An example of such a report is shown in below.
The calibration report provides useful information as to the quality of the calibration. The
overall standard deviation, as well as standard deviations for each grid configuration, are
displayed here. If the normal range of calibration error is exceeded, the standard deviations will
appear in red for easy identification.

If one or more images show a high deviation, you may return to the extraction by clicking the
Start over button. Locate the image using the Image: dropdown, and make desired adjustments.
To remove the image from the calibration, click Clear to clear extracted points. When done,
click Calibrate again to re-calibrate.

After calibration is complete, click Accept to add the calibration to the current project.

Method 2: Separate Camera Calibration


To calibrate the camera parameters separately, begin by adding calibration images to the
project. Note: Vic-3D expects image files for both cameras to be present on the disk, even if
only one of the two images is used.

Select Calibrate... Calibrate Camera 1 from the main menu bar. Select a target and (if
desired) a camera, and calibrate as above. When complete, the intrinsic parameters for Camera
1 will be added to your project. Repeat for camera 2.
At this point, each camera is calibrated but the stereo rig transformation is still undetermined;
this can be confirmed by viewing the Calibration tab of the project toolbar, as below.

To calibrate the camera orientation, add a reference image to the project and select an area-of-
interest. Ideally, the area-of-interest should cover as large an area of the image as possible.

Since the system is not calibrated, you must provide an initial guess before beginning the
calibration. Then, run the calibration. You will be prompted to run a disparity analysis:
Click Yes to continue or No to cancel. Continuing will run a disparity analysis, where the raw
image disparities are calculated but triangulation is not attempted, and no 3D shape will be
calculated.

Next, select Calibrate camera orientation from the Calibration menu. The dialog shown below
will appear.
On the displayed image, click and drag select a line between two marker points that are a
known distance apart. After selecting the line, the following dialog will appear:

Enter the distance between the two marker points and click Ok. You can click Calculate to
finish the calibration, or change the selected line. On completion, a calibration score will be
displayed:

High numbers indicate problems with the underlying data; if necessary, adjust AOI or initial
guess and re-run.

Fixed Baseline Calibration

The Fixed Baseline option can be used when a stereo calibration is present but has been
altered; for example, in a stereo rig where the cameras have been bumped, or disturbed by
a very dynamic test. It can also be used in the general case when a correlation results in a
higher than expected projection error.

The algorithm works by assuming that the stereo baseline is fixed and that only camera rotation
changed. This assumption works well for most systems where the cameras are fixed in space
by tripods or bars, but allowed to rotate on their mounts.

To perform this operation, select Calibrate... Calibrate camera orientation. Select the Fixed
Baseline scale option, and click the Calculate button. A calibration score will be displayed; to
proceed, click Accept .
Adjustments for Cropping
For many applications, particularly in high-speed imaging, the images collected during the test
may not have been acquired at full sensor resolution in order to increase the frame rate or to
save disk space. This can pose serious difficulties for normal calibration procedures. For
instance, if a dog-bone sample is used, it can be advantageous to reduce the size of the image in
the short dimension of the sample. This can lead to images that are only a few tens of pixels in
one dimension, and it is practically impossible to image a calibration target in this case.

For such applications, it is best to calibrate the camera at the full resolution of the sensor and
only crop the images during the test. The crop dialog seen in the figure below permits
adjustment of the calibration in such cases.

The dialog permits the user to enter the top-left corner of the image section with respect to the
calibration images for both cameras.

Note 1: If a reference image is loaded during calibration with a different size than the
calibration images, the adjustment dialog is automatically shown.

Note 2: The default values for the offsets are computed so that they are correct for most
modern high-speed cameras that crop from the center of the sensor.
Importing a Calibration
To import the calibration from an existing V3D project file, select Calibration... Import
calibration from the main menu bar. You will be prompted to select a project file; when
complete, the calibration will be imported and added to your current project.
Initial Guess Selection
If the motion between successive images is larger than a few pixels, the image correlation algorithm
requires an initial guess. This initial guess can be provided to the algorithm using the initial guess dialog.

Placing the seed point

The placement of the seed point can facilitate or eliminate the need for providing an initial guess.
Generally, it is best to place the seed point of the AOI in the area of the image that underwent the least
amount of motion during the test. For instance, if a specimen is tested in a tensile frame, the seed point
should be placed as close to the stationary grip as possible. In many cases, good placement of the seed
point eliminates the need to give an initial guess at all, since the algorithms can cope with small motion
increments. If there is no stationary point or point with small motion in the image, an initial guess has to
be provided. To make this process easier, it is recommended to place the seed point on an easily
recognizable feature in the image.

Selecting an initial guess

The initial guess dialog can be accessed through the Edit menu or the tool button in the AOI toolbar.
The initial guess dialog is shown below.
The area surrounding the seed point in the reference image is displayed in the box labeled Reference AOI.
Next to it, the pattern in the deformed image is displayed in the box labeled Deformed AOI. The image in
the bottom left of the dialog shows the deformed image. The listbox on the bottom right can be used to
select the deformed image for which the initial guess is being provided.

A histogram control is provided for the reference and deformed images. Adjust the red bars to control
image balance; this can be useful for finding detail in very dark images. Double click the histogram to
automatically set/reset the limits.

Selecting corresponding points

To select corresponding points in the reference and deformed image, the red cross in the bottom left
image has to be placed in the area of the deformed image that corresponds to the area around the seed
point in the reference image. This can be done by clicking and moving the mouse in the bottom image.
The scroll wheel can be used to zoom in and out; hold Shift and drag in the bottom image to pan. The
image in the Deformed AOI box is updated to reflect the location in the bottom image. Once the Reference
Aoi and Deformed Aoi images show the same region, common points can be selected.

If a calibration has already been performed, a green line will appear on Camera 2 images; this line
represents all points which could possibly match the seed point, based on camera geometry. Vic-3D will
attempt to find a match along this line; in most cases, no manual intervention is necessary. Simply
confirm that the left and right zoomed images match (as above); if they do, an initial guess has
automatically been found.

Translation and Complete Guess

An initial guess can be specified for the translation components of the displacement only, or for the
complete deformation including the partial derivatives of the displacement. The latter will only be
neccessary in cases where the strains and/or rotations are very high.

Translation Guess

Follow the steps below to specify a translation guess:

● Select a common point in the Reference Aoi and Deformed Aoi by clicking the mouse.
● Click the Add Point button, or click the right mouse button. The point pair turns yellow, and a
guess is calculated. The icon in the file list will change to a green check to indicate that an initial
guess has been selected for the current image.
● Optionally, repeat the first two steps for more corresponding points; the results will be an average
of all selected points. This can give higher accuracy but is usually not necessary.
● If necessary, repeat this procedure for all deformed files shown in the file list.

Complete Guess

Follow the steps below to specify an initial guess that includes estimates for the partial derivatives of the
displacement (only necessary for very large strains and/or large rotations):

● Check the Complete item in the Guess Type check box.


● Select a common point in the Reference Aoi and Deformed Aoi by clicking the mouse.
● Click the Add Point button or click the right mouse button.
● Repeat for at least two more points. Note that the points should not be co-linear, i.e., try to select
three points that make a triangle.
● Once at least three points are added, the icon in the file list will change to a green check to indicate
that an initial guess has been selected for the current image.
● If necessary, repeat this procedure for all deformed files shown in the file list.

Notes

Typically, if a proper seed point is selected and the image sequence is consistent, no initial guesses will be
required. Checking the initial guess dialog will confirm whether Vic-3D has automatically found an initial
guess for a given image; if matching areas are shown in the small zoom boxes, the guess has been found.

Vic-3D uses an epipolar constraint, derived from the calibration, to automatically search for initial
guesses. Therefore, if the calibration is poor or non-existent, manual initial guesses will be required. Also,
this automatic search is a coarse pixel-based search, so the following conditions may necessitate a manual
guess:

● Very fine speckle pattern


● Low-contrast speckle pattern
● Highly-curved surface, such as a cylinder

Important: when selecting initial guesses for deformed images, both camera views must be given explicit
initial guesses.
Running the Correlation
To run the displacement analysis, select the Run Correlation entry from the Data menu, or
press the button on the tool bar.

The File Tab


The tab on the dialog displays the following options:

Selecting Images
The deformed images to use for correlation analysis can be selected from the list box on the
dialog. Selected images are indicated by a check mark. Above the list box, buttons are
available to select/deselect all image files contained in the list box. To select 1 data file from
from every 2, 5, 10, or n , right-click in the file list and choose the desired option.

If no images are selected, only the reference image is analyzed.

Backup copies

When this option is checked, Vic-3D will make backup copies of existing output files by
replacing their file extension with bak.

Output directory

The directory in which the output files are stored can be selected by clicking the folder icon.

The Options Tab


The tab on the dialog displays the following options:
Interpolation

To achieve sub-pixel accuracy, the correlation algorithms use gray value interpolation.
Generally, more accurate displacement information can be obtained with higher-order
interpolation methods. The currently implemented interpolation methods are

● Cubic polynomial.
● Cubic B-spline.
● Quintic B-spline.

Quintic B-spline interpolation offers the best accuracy while being the slowest of the methods.
The use of cubic polynomial interpolation is not recommended and will likely not be supported
in future versions.
Binomial Low-pass Filter

The binomial low-pass filter option can help reduce interpolation bias for images with a bi-
modal gray value distribution.

Correlation Criterion

There are three correlation-criteria to choose from:

● SSD: Sum of Squared Differences.


● NSSD: Normalized Sum of Squared Differences. Insensitive to scale in lighting
● ZNSSD: Zero-Normalized Sum of Squared Differences: Insensitive to offset and scale
in lighting.

The Thresholding Tab


The tab on the dialog displays the following options:
Epipolar Thresholding

This threshold limits the distance between a point match and the epipolar line corresponding to
the point.

Global Thresholding

This threshold limits the highest error that is acceptable for the 'correlation' value for a subset.
Enable this threshold to remove data where cracks or occlusions are present.

Minimum Contrast

This option automatically removes subsets that show a very low contrast. Increase this value to
remove more data; reduce to retain more data, i.e., if lighting conditions were poor.

The Post-Processing Tab


The tab on the dialog displays the following options:

Coordinate Transformation

You can choose to apply no transformation, an automatic best-plane fit transformation or a pre-
defined coordinate transformation.

Strain Calculation
Checking this option performs a strain computation as each image is processed; results can be
viewed in the preview.

Correlation Results

After you begin the correlation, the following window appears.

The window contains an overview of correlation progress and reults.

Points Analyzed - the number of data points in the image.

Time [s] - The amount of time spent on correlation analysis.

Iterations - a measure of how many possible matches were searched for each point.

Error- a measure of correlation accuracy (specificially, epipolar projection error). A high


number indicates possible problems with calibration or camera synchronization.

Output- the name of the data file written.

Progress bar- indicates the progress of each individual file as it is correlated.


For more information on interpreting correlation results and troubleshooting errors, please
contact Technical Support.
Coordinate System Selection
Coordinate system transformations can be selected based on the 3D data contained in the
reference data file. To efficiently use coordinate transformations, make sure to disable
automatic transformation during data analysis; see Coordinate Transformation for details. To
select a coordinate system, select Data... Coordinate Transformation... Select coordinate
system from the main menu bar.

The following dialog will be displayed:

Selection Method

There four choices to select a coordinate system:

Best Plane Fit

Select this option to calculate a transform into the best fit plane of the surface. This is the same
as selecting Best Plane Fit from the coordinate transform menu.

Select Origin

Use this option to designate an origin for the coordinate system, while leaving the orientation
unchanged. Simply click Select Origin, and pick a point in the image. This point becomes 0,0,0
in the x,y,z coordinate system. The orientation of the axes will be determined from the best
plane fit.

This transform is useful when a specific point in the image is a convenient reference origin, for
example, a notch in the specimen.

Origin and x-axis

Selecting this option allows the designation of the coordinate origin as well as the x-axis. After
selecting, click a point in the image to become the coordinate origin. Click a second point to
indicate the direction of the positive x-axis. The direction of the y-axis will be determined from
the best plane fit.

This transform is useful when both a specific origin point and an orientation are desired; for
instance, the origin could be a crack tip and the axis could be the direction of the crack opening.

3 point selection

This option allows selection of an origin and the complete orientation of the axes. Select a
point to become the origin. Select a second point to indicate the direction of the x-axis. Finally,
select a third point to indicate a third point on the xy-plane; this will completely define an
origin and xyz orientation.

Reference Point

If desired, enter a coordinate to be designated by the reference point; this will change the point
designated by the first click from (0,0,0) to the specified coordinates.

Click Accept to calculate the new coordinate transform. The following dialog appears:
Enter a name to add the transform to the current project.

Applying Transformations
Coordinate system transformations can be applied to data files by selecting the Apply
Transformation entry from the Coordinate System Transformation menu. The following dialog
lets you choose which transformation to apply, and which data files to apply the transformation
to.
To invert (undo) the selected transformation, check Invert Transformation.

Each transform contains 3 rotation components and 3 translation components. By default, only
the rotation is applied to the U, V, and W displacement variables; that is, the displacement
vectors are rotated with the surface, but their magnitude is unchanged. To apply the translation
components as as well, altering the displacement vectors' magnitude, click Displacements only.
Postprocessing Data
Vic-3D offers various options for postprocessing calculated displacement data.

The dialog illustrated above is for calculating strain, although the interface is common to the
other postprocessing options.

Selecting Data Files for Processing

The available data files are displayed in the Data Files list box. To select which files to
process, click on the data file you want to select/deselect. This will toggle the check mark
indicating whether the file is selected or not. For convenience, the buttons labeled All and None
select/deselect all files, while Invert reverses your selection.

Preview

To view the effects of the calculation for a single data file, highlight the file and click the
Preview button. You may view the plot in 2D or 3D as with a standard data plot.
Smoothing Data

This postprocessing option smoothes the selected variable using a local filter.

The Filter box allows selection of a smoothing method. The decay filter is a 90% center-
weighted Gaussian filter and works best for most situations; the box filter is a simple
unweighted averaging filter. The Filter size box controls the size of the smoothing window.
Since the filter size is given in terms of data points rather than pixels, the physical size of the
window on the object also depends on the step size used during correlation analysis.

Calculating Strain

This postprocessing option calculates strain on the surface of the test area. The strains are
calculated from the partial derivatives of the displacement using the Lagrange strain tensor
equations.

By default, strain is calculated for x, y, and shear. Checking the Compute principal strains box
adds first and second principal strains as well as the strain angle.

Calculating Curvature

This option calculates local curvature for the analyzed surface.

Calculating Rotation

This option calculates local in-plane rotation for the analyzed surface.
Calculating Velocity
Vic-3D can calculate rates for displacement and strain, using either a specified time interval or
time retrieved from a .CSV log file.

Once the displacement fields have been calculated from the speckle images, velocities may be
calculated by selecting the Calculate Velocity entry on the Data menu. (If strain rates are
desired, strain should be calculated before opening the Calculate Velocity dialog.) This will
display the dialog shown in the figure below.

Data File Selection

The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

Velocity Calculation

If a Vic-Snap .CSV log file exists for the project, you may select "Time From File" from the
dropdown and select the file, if necessary. Otherwise, select "Constant Time Step" and enter
the known time increment, or select "Constant Frame Rate" to enter a known frame rate, for, i.
e., a high-speed camera.

Click Start to begin; the progress bar will indicate completion. For each strain and
displacement variable in the dataset, a derivative in time will be added and can be viewed as a
contour overlay.
Removing Displacement
This tool is used to remove rigid-body displacement from deformed images, leaving only
deformation components of displacement.

Once the displacement fields have been calculated from the speckle images, this tool can be
started by selecting Data... Remove displacements.

Data File Selection

The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

Processing method
To remove all rigid body displacement, select Average transformation. This will calculate the
average transformation for each image, and invert it to obtain an image with an average
displacement/rotation of 0. Only object deformation will be reflected in the transformed U, V,
and W displacements. This is useful for visualizing displacement fields in tests where
deformation is obscured by larger, rigid-body motions.

To keep a single point stationary, select Point displacement and click the desired point in the
image. Only points within the data set (highlighted in blue) may be selected. With this method,
the selected point will become stationary, with all other displacements remaining relative to
this point. Any initial rotation will still be present after the operation.

To keep three points stationary, select Three-point displacement; then, click three desired
points in the image to define the fixed points. With this method, all displacements and rotations
will be relative to the three selected points, which will become stationary. The point- and three-
point displacement options are useful for, i.e., determining deformation of a test subject
relative to fixed mounting points or standoffs.

Click Compute to proceed with the computation.

Strain calculation

Note that these transforms will not affect calculated strain, nor are they necessary in order to
correctly calculate strain; the strain algorithm is, by nature, insensitive to rigid-body
displacements.
Math Operations
The Math operations dialog allows manipulation of output data by basic math operations. Open
this dialog by selecting Data... Math Operations from the main menu bar.

Data File Selection

The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

Operation

Choose Add, Subtract, Multiply, or Divide to perform the specified operation.

Arguments

The Variable box is used to select the variable to operate on. Any variable in the data set may
be selected.

To use a constant argument, select the Constant radio button and enter the value. For example,
the selections below will multiply the W-displacement value from each data file by 10.

To use the data from an output file, click Data and select a data file. For example, the
selections below will subtract the Z shape from the first data file, from all data files.
Click Start to begin; the progress bar will indicate completion. For each strain and
displacement variable in the dataset, a derivative in time will be added and can be viewed as a
contour overlay.
Calculating Statistics
To export statistics for calculated variables and data files, select Data... Statistics from the
main menu bar.

Statistics
Check the desired item to include or exclude the statistic from the output file.

Variables

Check the desired variables to add them to the calculation. By default, all metric variables are
included, while correlation and pixel variables are excluded.

Data File Selection

The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

Exporting

To complete the calculation, click Ok. You will be prompted for a filename, and the data will
be exported as a .CSV file.
Exporting Data
For efficient file access, Vic-3D stores results in a binary data file format. To use the data with
other programs for post-processing and plotting, the data can be exported by selecting the
Export item from the Data menu.

The dialog shown below will be displayed.

Selecting Files for Exporting


The available data files are displayed in the list box. To select which files to export, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files.

File Formats

The data files can be exported to the following formats:

Comma-Separated Variable

Data entries are separated by commas. This format is understood by most


spreadsheet programs and plotting packages. Variable names are stored in the
data file as comma-separated strings in quotation marks. Exported files will have
the extension csv.

Tecplot

Used for plotting the data with Amtec's (www.amtec.com ) plotting program
Tecplot(TM). Exported files will have the extension dat.

Plain ASCII

This format is plain, space-delimited ASCII text data with one data point per
line. Note: There are no variable names in the data file, and data from different
AOIs is concatenated. Exported files will have the extension txt.

If none of the available file formats fit your needs, please contact support@correlatedsolutions.
com. We will gladly implement data exporting to a format that best suits your needs.
Exporting Grid-Based Data
This option can be used to export your data, sampled at regular intervals spatially and for each
data file, to a single text file. To begin, select Export Grid Data from the Data menu.

Data File Selection

The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

Options

To change the sample interval in pixels, adjust the Sample step value. A value of 1 will sample
every pixel; higher values will result in a sparser data set.

To export blank values to the output file, with a value of 0, check the Export blanks box. If this
box is cleared, blank data points will not be present in the output file.

Extracting Data

To begin, click Start. You will be prompted for an output .csv file name. A progress bar will
appear; when extraction is complete, the dialog will close.
Plotting Data
Data can either be visualized as a three-dimensional plot with or without color contour overlay,
or as a two-dimensional contour plot overlayed over the speckle image. The two-dimensional
plot provides facilities for extracting data from sequences, as well as extracting slices of data.
3D Plot 2D Plot
Output Variables

Introduction

During correlation and optional post-processing, Vic-3D presents a wide range of output data
available for 3D and contour plotting, extraction, and export.

Output Variables

Always Present

● X [mm] – metric position along the X-axis (by default, the horizontal axis).
● Y [mm] – metric position along the Y-axis (by default, the vertical axis). Z [mm] –
metric position along the Z-axis (by default, the out-of-plane axis).
● correlation [1] – a measure of the quality of the correlation result at a given point. 0
indicates a perfect match; higher numbers may result from noise, lighting changes,
breakdown of the speckle pattern, etc.
● x / y [pixel] – point location in the reference speckle image
● U [mm] – metric displacement along the X-axis, from the reference image. For the
reference image, displacement values will always be 0.
● V [mm] – metric displacement along the Y-axis.
● W [mm] – metric displacement along the Z-axis.
● u / v / q / r [pixel] – internal disparity measures

Generating in Postprocessing

Strain Variables

● exx [1] – strain in the X-direction. Positive numbers indicate tension; negative numbers
indicate compression.
● eyy [1] – strain in the Y-direction.
● exy [1] – shear strain.
● e1 [1] – the major principal strain.
● e2 [1] – the minor principal strain.
● gamma [1] – the principal strain angle, measure counterclockwise from the positive X-
axis.

Velocity Variables
● dU/dt [1] – the rate of change of the U-displacement; that is, the velocity of a given
point in the X direction.
● dV/dt [1] – velocity in the Y direction.
● dW/dt [1] – velocity in the Z direction.
● dExx/dt [1] – the rate of change of strain in X, or strain rate in X.
● dEyy/dt [1] – the strain rate in Y.
● dExy/dt [1] – the shear strain rate.

Curvature Variables

● cxx [1] – curvature in the X-direction.


● cyy [1] – curvature in the Y-direction.
● c1 [1] – the major principal curvature.
● c2 [1] – the minor principal curvature.
● gamma [1] – the principal curvature angle, measured counterclockwise from the
positive X-axis.

Rotation Variables

● phi [rad] – the in-plane rotation, in radians.

Key Terms

Strain: the change in length, divided by initial length, for a solid material. For example, a strain
of +10% indicates that the material has expanded by 10%; a strain of -10% indicates that the
material has contracted by 10%. Positive strains are referred to as tensile, while negative strains
are compressive.

Principal strain: the strain for a reference frame that is rotated such that shear becomes zero,
leaving only two strain components at 90º angles. The larger strain becomes the major
principal strain, and the smaller becomes the minor principal strain.
The Plot Toolbar
The plot toolbar is displayed at the top left edge of the work area by default. It contains options
and controls for both the 2D and 3D plots.

Axes

This tab allows selection of axis limits and Z-scaling. To automatically scale these values to fit
the data, check the Auto-rescale coordinates box. To manually set the scales, clear this box,
and enter values as desired. To control the scaling of the Z-axis relative to the X- and Y-axes,
change the Z-scale value. Larger numbers will exaggerate the Z-axis more.

Contour
This tab allows control of the contour overlay of 2D or 3D plots. To automatically scale these
values to fit the data, check the Auto-rescale contour box. To manually set the limits, clear this
box and enter the desired values.

To enable or disable the contour overlay in 3D mode, select Show contour plot.

Use the strain unit control to determine how strain values are displayed; the default is unity, i.
e., mm/mm.

Color
Use this tab to control the display of contour overlays. The Color map box chooses the overall
color set for the plot. The Opacity box sets the opacity of the overlay; this option affects 2D
plots only. The Levels box sets the number of discrete contour levels.

Vector

This tab controls display of strain and displacement vectors.

Major and minor strain direction vectors are displayed on 2D plots only, when strain data is
available; displacement vectors may be viewed in 2D or 3D. Skip and scale control the size and
density of the vectors. The use solid color checkbox causes the vectors to be displayed in a
single color rather than the underlying plot color; the color selector button can be used to
choose this color.
3D Plots
A 3D plot of the data can be displayed by double-clicking on a data file in the list view to the
left of the workspace or by selecting a data file from the Data... Plot menu.

A plot will be displayed in the workspace as shown below.

Controlling the Plot

To rotate the plot, click and drag in the plot window. To zoom in and out, rotate the scroll
wheel. To pan around the plot, click and hold the scroll wheel and move the mouse.

Plot Options

Plot options can be accessed by right-clicking in the plot window.


To switch to a 2D plot, select Show 2D Plot from the context menu, or click 2D in the main
toolbar.

Use the View submenu to select a planar view or revert to the default view.

The Contour variable submenu can be used to select the variable to display.

Click Change legend orientation to toggle between a horizontal and vertical legend.

Copy copies the current plot to the clipboard; Save allows saving the plot as an image file.

Select Export video to save an animated video.

Editing Plot Parameters

To edit other plot parameters, use the plot toolbar.


Animating Plots

To animate 3D plots, bring up the plot display and then use the controls on the Animation
Toolbar to animate the sequence.

Saving the Plot

The displayed plot can be saved as a BMP, PNG, or JPG image file by selecting Save from the
context menu. To copy the plot to the clipboard, select Copy.
2D Plots
A 2D plot of the data can be displayed by double-clicking on a data file in the list view to the
left of the workspace, and then clicking 2D in the main toolbar or clicking Show 2D plot in the
plot context menu.

A plot will be displayed in the workspace as shown below.

Plot Options

Plot options can be accessed by right-clicking in the plot window.


To switch to a 3D plot, select Show 3D Plot from the context menu, or click 3 D in the main
toolbar.

Click Show camera 2 to change the underlying speckle image to the other camera view.

Click Statistics to view a summary of data for the current image, for the currently selected
contour variable.

The Contour variable submenu can be used to select the variable to display.

Click Change legend orientation to toggle between a horizontal and vertical legend.

Copy copies the current plot to the clipboard; Save allows saving the plot as an image file.

Select Export video to save an animated video.

Editing Plot Parameters


To edit other plot parameters, use the plot toolbar.

Inspector Tools

Tools for probing and extracting data are located in the Inspector Toolbar, and can also be
selected by clicking Plot... Inspector in the main menu bar.

From left to right, the tools are:

● Pan/Select: Pans around the contour image, when zoomed in; selects existing extract
points. To select an item, click on the small square handle.
● Inspect point: select this tool and click to probe at a single point. The value for the
currently selected contour variable, at the chosen point, will be displayed.
● Inspect line: select this tool and click once to start a line; click again to finish. The value
will be displayed at each node.
● Inspect polyline: select this tool and click to create line nodes; double-click to finish.
The value will be displayed at each node.
● Inspect circle: select this tool and click to define a center; click again to define a disc.
The value at the center will be displayed.
● Inspect rectangle: select this tool and click to define a center; click again to define a
rectangle. The value at the center will be displayed.
● Delete: choose this tool and click on an existing point/line/area to remove it.
● Extract: click to open the Extract Sequence Data or Line Plot dialogs. For the line tool,
data will be extracted alone the line for the current image. For the other tools, data will
be extracted at the selected point or area, for each image, as a time sequence.

Animating Plots

To animate contour plots, bring up the plot display and then use the controls on the Animation
Toolbar to animate the sequence.

Saving the Plot

The displayed plot can be saved as a BMP, PNG, or JPG image file by selecting Save from the
context menu. To copy the plot to the clipboard, select Copy.
Exporting Videos
To export an animation from a 2D or 3D plot, right-click in the plot and select Export Video.

If the auto-rescaling feature is enabled for contours or axes, you will see a warning:

When rescaling is on, the animation may not appear as expected because each frame will be
scaled differently. Click Yes to continue or Cancel to correct the condition. When complete,
the following dialog appears:
File

Click the icon to select a filename for saving.

Encoder

To export a video file, select AVI from the Encoder drop-down; to export individual frames as
separate, numbered files, select Image Sequence.

Format

Select from available compression formats; options will vary based on system configuration
and installed codecs. Indeo(R) Video 5.11 works well for most Windows XP systems. RGB
Uncompressed videos will be high quality but very large.

Data File Selection


The available data files are displayed in the list box. To select which files to process, click on
the data file you want to select/deselect. This will toggle the check mark indicating whether the
file is selected or not. For convenience, the buttons labeled All and None select/deselect all
files; the Invert button inverts the selection.

To begin, click Export; a progress bar will indicate completion.


Line Plots
Line plots can be generated by using the Inspector Tools in a 2D plot. There are two types of
line plots currently available:

● Data extracted along a line


● Data extracted from a sequence of data files

Line Data Extraction and Plotting


The line data plot appears when the Extract button is pushed on the 2D Plot while a line is
selected:

Adjusting the Number of Data Points

The number of data points that are extracted from the data file along the line can be adjusted
using the Number of Points control on the window.
Plot Style

The plot style can be selected to either show data points, lines or data points and lines.

Sequence Data Extraction and Plotting


The sequence data extraction window appears when the Extract button is pushed on the 2D
Plot while a point, circle, or rectangle is selected:

Selecting Data Files

The data files used for extraction can be selected in the listbox in the top left of the window on
the Extract tab.

Selecting Variables

The variables to extract from the files can be selected in the listbox in the bottom left of the
window on the Extract tab.
If an analog data file is present, time and analog data values may be selected for either axis.

Extracting the Data

After selecting the data files and variables to extract, push the Extract button to extract the
data. Depending on the size and number of data files, this might take some time and a progress
indicator will be shown to indicate the extraction progress. After data extraction, the Plot tab
will become active and the window will appear as follows:

Plot Style

The plot style can be selected to either show data points, lines or data points and lines.

Saving Data

The extracted data can be saved as a comma-separated variable file by pressing the Save Data
in the bottom left of the plot window.

Saving Plots
The plot can be saved as a BMP, PNG, or JPG image file by pressing Save Plot in the bottom
left of the plot window. You may also copy the plot to the clipboard by right-clicking on the
plot area and selecting Copy.
Examples
You can find a complete example project in the examples folder in the Vic-3D installation
directory.

The project is in the subfolder membrane and may also be found in the Start Menu folder.

To calibrate the stereo system for this program, choose the 9x9 - 15mm calibration target in the
Target Selection dialog.
Quick Start
There are only a few steps involved in obtaining shape and deformation measurements from
your images. If you are using Vic-3D for the first time, take a look at the example provided
with the program. Then, try to go through the following steps yourself to quickly familiarize
yourself with the program usage:

1. Add your calibration images to the project.

2. Perform the calibration.

3. Add a reference image and select your area of interest.

4. Add more speckle images, if applicable.

5. Run the correlation analysis.


6. Plot the results

If you encounter any difficulties, please do not hesitate to contact our technical support
department .
Technical Support
If you cannot find an answer to your question in this manual, please do not hesitate to contact
our technical support at [email protected]. You can also find contact
information at our web site at www.correlatedsolutions.com.

We will be happy to assist with topics such as:

● Designing digital image correlation experiments


● Calibration
● Troubleshooting errors
● Interpreting test data
● Achieving optimal results

Bug Reports and Feature Requests

If you encounter a bug in Vic-3D, please let us know about it. Send a short description of the
problem to [email protected], along with any project or image files you think
may help us reproduce the bug.

Also, if you think Vic-3D can be improved by adding a particular feature you would find
helpful, let us know about it. We will try to incorporate your requests in our future updates of
the software.