GE Pedprotocols 09
GE Pedprotocols 09
DOC 0509391
GE Healthcare
Technical Publication
Direction 5327782-1EN FeatherLight Imaging
GE Medical Systems does business as GE Healthcare Technology for dose optimization
Page Number
Table of Contents 3
Waivers 4
Legal Notes 7
LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Pink Zone) 8
LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Green Zone) 10
LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Grey Zone) 12
Page 3
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391
DAMAGE IN TRANSPORTATION
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and extent
of damage. Complete instructions regarding claim procedure are found in Section S of the Policy And Procedures Bulletins.
IMPORTANT...X-RAY PROTECTION
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Healthcare, will be glad to assist and cooperate in placing this equipment in use.
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to take
adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the International
have no responsibility for injury or damage which may result from improper use of the equipment. Various protective
materials and devices are available. It is urged that such materials or devices be used.
Page 4
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391
1. The radiologist(s) speak to the medical staff and explain that in order to reduce dose from CT
exams they are planning on using these procedure based/ “zoned” protocols.
2. Every order coming to the radiology department for a pediatric or small patient exam should
include an actual or closely estimated weight. The order should also state if this is the first scan
of the patient (if known).
3. The radiologist or designated person within the department responsible for putting the order
in front of the CT technologists would determine the indication type or “Zone” to be used.
4. The technologists select the correct protocol for scanning.
Page 5
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391
1. Correct centering of the patient’s mass to iso-center is the key to correct mA modulation working properly.
2. Each order should be reviewed by appropriate personnel to insure the nature and need of the scan before the
scan is performed. This includes whether the patient is being scanned for the first time vs. a follow-up scan vs an
angiogram, as well as determining if CT is the appropriate modality for this patient’s needs.
3.Scan parameters such as kV, mA, and noise index should be compared against the numbers on the poster or
chart of the “Zones” before each scan
If you have further questions regarding the use of the protocols please call the GE Helpdesk at 1-800-682-5327.
Page 6
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391
Legal Notes
Copyrights
All Material Copyrighted © 2008 by the General Electric Company, All rights reserved.
Revison History
Revision Version Date Reason for change
1 1 10/07/08 Initial Release for CT Procedure-Based
Protocols
1 2 10/22/2008 Changed for use with 64 slice and 16 slice
systems only
Page 7
g LightSpeed VCT Pediatric CT Protocols (Pink Zone)
Pink Zone--Routine R/O 0-20 lbs = 0-9 kg Pink Zone--Routine R/O 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 21 0.5 Helical S 0.0 I 100.0 2.5 55 64 2.5 1 41 0.5 Helical S 0.0 I 200.0 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 PedBody VCT 80 Yes 5 65 130 No No No S0.0 PedBody VCT 100 Yes 7 80 160 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 2.5 2.5 Standard
Pink Zone--Routine R/O 61-100 lbs = 27.3-45.4 kg Pink Zone--Routine R/O 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 61 0.5 Helical S 0.0 I 300.0 3.75 55 64 3.75 1 81 0.5 Helical S 0.0 I 400.0 5 55 64 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 MediumBody VCT 120 Yes 10 95 190 2 No No S0.0 LargeBody VCT 120 Yes 12 110 220 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard
Pink Zone--Routine R/O 61-100 lbs = 27.3-45.4 kg Pink Zone--Routine R/O 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 27.5 16 3.75 1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 120 Yes 10 95 190 2 No No S0.0 Large 120 Yes 12 110 220 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard
Green Zone--Follow-up, Low Dose 61-100 lbs = 27.3-45.4 kg Green Zone--Follow-up, Low Dose 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 55 64 3.75 1 80 0.5 Helical S 0.0 I 395 5 55 64 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 MediumBody VCT 120 Yes 11 70 140 2 No No S0.0 LargeBody VCT 120 Yes 13 80 160 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard
Green Zone--Follow-up, Low Dose 61-100 lbs = 27.3-45.4 kg Green Zone--Follow-up, Low Dose 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 27.5 16 3.75 1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 120 Yes 11 70 140 2 No No S0.0 Large 120 Yes 13 80 160 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard
Gray Zone--CT Angio 61-100 lbs = 27.3-45.4 kg Gray Zone--CT Angio 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 120 0.5 Helical S 0.0 I 295 2.5 55 64 2.5 1 160 0.5 Helical S 0.0 I 395 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 LargeBody VCT 100 Yes 10 120 240 2 No No S0.0 LargeBody VCT 120 Yes 12 120 240 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard
Gray Zone--CT Angio 61-100 lbs = 27.3-45.4 kg Gray Zone--CT Angio 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 119 0.5 Helical S 0.0 I 295 2.5 27.5 16 2.5 1 159 0.5 Helical S 0.0 I 395 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 100 Yes 10 120 240 2 No No S0.0 Large 120 Yes 12 120 240 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard