Direction 5327782-1EN CT Procedure-Based Protocols

DOC 0509391

GE Healthcare
Technical Publication
Direction 5327782-1EN FeatherLight Imaging
GE Medical Systems does business as GE Healthcare Technology for dose optimization

Revision 01 CT Procedure-Based Protocols


Useful for patients ranging from:
- Newborns to 21 years old
- Smaller adults

This Manual Supports the Following Product Names:


LightSpeed VCT XT LightSpeed Pro16 BrightSpeed Elite
LightSpeed VCT LightSpeed VFX16 BrightSpeed Elite Select
LightSpeed16

Copyright © by General Electric Company, 2008


Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

This page intentionally left blank


Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

Page Number
Table of Contents 3

Waivers 4

Tips for successful Procedure Based Protocol implementation. 5

Key Points for Procedure Based Protocols 6

Legal Notes 7

LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Pink Zone) 8

LightSpeed/BrightSpeed Pediatric & Small Patient Procedure-Based CT Protocols (Pink Zone) 9

LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Green Zone) 10

LightSpeed/BrightSpeed Pediatric & Small Patient Procedure-Based CT Protocols (Green Zone) 11

LightSpeed VCT Pediatric & Small Patient Procedure-Based CT Protocols (Grey Zone) 12

LightSpeed/BrightSpeed Pediatric & Small Patient Procedure-Based CT Protocols (Grey Zone) 13

Page 3
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

DAMAGE IN TRANSPORTATION
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and extent
of damage. Complete instructions regarding claim procedure are found in Section S of the Policy And Procedures Bulletins.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall
be performed by qualified GE Healthcare personnel. The products involved (and the accompanying electrical installations)
are highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

IMPORTANT...X-RAY PROTECTION
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Healthcare, will be glad to assist and cooperate in placing this equipment in use.
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to take
adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the International
have no responsibility for injury or damage which may result from improper use of the equipment. Various protective
materials and devices are available. It is urged that such materials or devices be used.

OMISSIONS & ERRORS


report all omissions, errors, and defects in this publication.

Page 4
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

Tips for Successful Procedure Based Protocol Implementation.

1.    The radiologist(s) speak to the medical staff and explain that in order to reduce dose from CT
exams they are planning on using these procedure based/ “zoned” protocols.
2.    Every order coming to the radiology department for a pediatric or small patient exam should
include an actual or closely estimated weight. The order should also state if this is the first scan
of the patient (if known).
3.    The radiologist or designated person within the department responsible for putting the order
in front of the CT technologists would determine the indication type or “Zone” to be used.
4.    The technologists select the correct protocol for scanning.

Page 5
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

Key Points for Procedure Based Protocols

1. Correct centering of the patient’s mass to iso-center is the key to correct mA modulation working properly.

2. Each order should be reviewed by appropriate personnel to insure the nature and need of the scan before the
scan is performed. This includes whether the patient is being scanned for the first time vs. a follow-up scan vs an
angiogram, as well as determining if CT is the appropriate modality for this patient’s needs.
3.Scan parameters such as kV, mA, and noise index should be compared against the numbers on the poster or
chart of the “Zones” before each scan

4. Confirm the weight of the patient before initiating the exposure.

If you have further questions regarding the use of the protocols please call the GE Helpdesk at 1-800-682-5327.

Page 6
Direction 5327782-1EN CT Procedure-Based Protocols
DOC 0509391

Legal Notes
Copyrights

All Material Copyrighted © 2008 by the General Electric Company, All rights reserved.

Revison History
Revision Version Date Reason for change
1 1 10/07/08 Initial Release for CT Procedure-Based
Protocols
1 2 10/22/2008 Changed for use with 64 slice and 16 slice
systems only

Page 7
g LightSpeed VCT Pediatric CT Protocols (Pink Zone)
Pink Zone--Routine R/O 0-20 lbs = 0-9 kg Pink Zone--Routine R/O 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 21 0.5 Helical S 0.0 I 100.0 2.5 55 64 2.5 1 41 0.5 Helical S 0.0 I 200.0 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 PedBody VCT 80 Yes 5 65 130 No No No S0.0 PedBody VCT 100 Yes 7 80 160 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 2.5 2.5 Standard

Pink Zone--Routine R/O 61-100 lbs = 27.3-45.4 kg Pink Zone--Routine R/O 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 61 0.5 Helical S 0.0 I 300.0 3.75 55 64 3.75 1 81 0.5 Helical S 0.0 I 400.0 5 55 64 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 MediumBody VCT 120 Yes 10 95 190 2 No No S0.0 LargeBody VCT 120 Yes 12 110 220 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard

Pink Zone--Routine R/O >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 81 0.5 Helical S 0.0 I 400.0 5 55 64 5
© Copyright 2008 General Electric Company. All rights reserved.
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 LargeBody VCT 120 Yes 15 125 300 2 No No
Direction 5327782-1EN
Series 2 Group 1 Recon 1 Settings DOC 0509391
Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No Page 8
Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter
1 2.5 2.5 Standard
g LightSpeed 16 Pediatric CT Protocols (Pink Zone)
Pink Zone--Routine R/O 0-20 lbs = 0-9 kg Pink Zone--Routine R/O 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 20 0.5 Helical S 0.0 I 95 2.5 27.5 16 2.5 1 40 0.5 Helical S 0.0 I 195.0 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Small 80 Yes 5 65 130 No No No S0.0 Small 100 Yes 7 80 160 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 2.5 2.5 Standard

Pink Zone--Routine R/O 61-100 lbs = 27.3-45.4 kg Pink Zone--Routine R/O 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 27.5 16 3.75 1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 120 Yes 10 95 190 2 No No S0.0 Large 120 Yes 12 110 220 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard

Pink Zone--Routine R/O >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 120 Yes 15 125 300 2 No No
© Copyright 2008 General Electric Company. All rights reserved.
Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Direction 5327782-1EN
1 36 A0.0 R0.0 Standard Full No No No DOC0509391
Series 2 Group 1 Recon 2 Settings Page 9
Group Thick Interval Filter
1 2.5 2.5 Standard
g LightSpeed VCT Pediatric CT Protocols (Green Zone)
Green Zone--Follow-up, Low Dose 0-20 lbs = 0-9 kg Green Zone--Follow-up, Low Dose 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 20 0.5 Helical S 0.0 I 100.0 2.5 55 64 2.5 1 40 0.5 Helical S 0.0 I 195.0 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 PedBody VCT 80 Yes 7 50 100 No No No S0.0 PedBody VCT 100 Yes 9 60 120 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 2.5 2.5 Standard

Green Zone--Follow-up, Low Dose 61-100 lbs = 27.3-45.4 kg Green Zone--Follow-up, Low Dose 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 55 64 3.75 1 80 0.5 Helical S 0.0 I 395 5 55 64 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 MediumBody VCT 120 Yes 11 70 140 2 No No S0.0 LargeBody VCT 120 Yes 13 80 160 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard

Green Zone--Follow-up, Low Dose >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 80 0.5 Helical S 0.0 I 395 5 55 64 5
© Copyright 2008 General Electric Company. All rights reserved.
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 LargeBody VCT 120 Yes 16 90 240 2 No No
Direction 5327782-1EN
Series 2 Group 1 Recon 1 Settings DOC 0509391
Group DFOV A/P R/L Filter Type Vari D3D DMPR Page 10
1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter
1 2.5 2.5 Standard
g LightSpeed 16 Pediatric CT Protocols (Green Zone)
Green Zone--Follow-up, Low Dose 0-20 lbs = 0-9 kg Green Zone--Follow-up, Low Dose 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 20 0.5 Helical S 0.0 I 95 2.5 27.5 16 2.5 1 40 0.5 Helical S 0.0 I 195.0 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Small 80 Yes 7 50 100 No No No S0.0 Small 100 Yes 9 60 120 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 2.5 2.5 Standard

Green Zone--Follow-up, Low Dose 61-100 lbs = 27.3-45.4 kg Green Zone--Follow-up, Low Dose 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 60 0.5 Helical S 0.0 I 295 3.75 27.5 16 3.75 1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 120 Yes 11 70 140 2 No No S0.0 Large 120 Yes 13 80 160 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 2.5 2.5 Standard 1 2.5 2.5 Standard

Green Zone--Follow-up, Low Dose >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 80 0.5 Helical S 0.0 I 395 5 27.5 16 5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
© Copyright 2008 General Electric Company. All rights reserved.
S0.0 Large 120 Yes 16 90 240 2 No No
Series 2 Group 1 Recon 1 Settings Direction 5327782-1EN
Group DFOV A/P R/L Filter Type Vari D3D DMPR DOC 0509391
1 36 A0.0 R0.0 Standard Full No No No Page 11
Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter
1 2.5 2.5 Standard
g LightSpeed VCT Pediatric CT Protocols (Gray Zone)
Gray Zone--CT Angio 0-20 lbs = 0-9 kg Gray Zone--CT Angio 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 40 0.5 Helical S 0.0 I 95 2.5 55 64 2.5 1 80 0.5 Helical S 0.0 I 195.0 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 PedBody VCT 80 Yes 5 100 200 No No No S0.0 SmallBody VCT 100 Yes 7 120 240 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 20 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard

Gray Zone--CT Angio 61-100 lbs = 27.3-45.4 kg Gray Zone--CT Angio 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 120 0.5 Helical S 0.0 I 295 2.5 55 64 2.5 1 160 0.5 Helical S 0.0 I 395 2.5 55 64 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 LargeBody VCT 100 Yes 10 120 240 2 No No S0.0 LargeBody VCT 120 Yes 12 120 240 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard

Gray Zone--CT Angio >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 160 0.5 Helical S 0.0 I 395 2.5 55 64 2.5
© Copyright 2008 General Electric Company. All rights reserved.
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 LargeBody VCT 120 Yes 15 120 300 2 No No
Direction 5327782-1EN
Series 2 Group 1 Recon 1 Settings DOC 0509391
Group DFOV A/P R/L Filter Type Vari D3D DMPR Page 12
1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter
1 1.25 1.25 Standard
g LightSpeed 16 Pediatric CT Protocols (Gray Zone)
Gray Zone--CT Angio 0-20 lbs = 0-9 kg Gray Zone--CT Angio 21-60 lbs = 9.1-27.2 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I100 90 No No No 1 80 10 S0.0 I200 90 No No No
2 80 10 S0.0 I100 0 No No No 2 80 10 S0.0 I200 0 No No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 39 0.5 Helical S 0.0 I 95 2.5 27.5 16 2.5 1 79 0.5 Helical S 0.0 I 195.0 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Small 80 Yes 5 100 200 No No No S0.0 Small 100 Yes 7 120 240 No No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 25 A0.0 R0.0 Standard Full No No No 1 25 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard

Gray Zone--CT Angio 61-100 lbs = 27.3-45.4 kg Gray Zone--CT Angio 101-200 lbs = 45.5-90.7 kg
Series 1 Scout FeetFirst Supine Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO Auto Store Y/N Gating NO
Auto Transfer Y/N Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I300 90 2 No No 1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I300 0 2 No No 2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N Auto Transfer Y/N
Series 2 Group 1 Scan Settings Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval Group Images Speed Type Start End Thick Speed Rows Interval
1 119 0.5 Helical S 0.0 I 295 2.5 27.5 16 2.5 1 159 0.5 Helical S 0.0 I 395 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
S0.0 Large 100 Yes 10 120 240 2 No No S0.0 Large 120 Yes 12 120 240 2 No No
Series 2 Group 1 Recon 1 Settings Series 2 Group 1 Recon 1 Settings
Group DFOV A/P R/L Filter Type Vari D3D DMPR Group DFOV A/P R/L Filter Type Vari D3D DMPR
1 36 A0.0 R0.0 Standard Full No No No 1 36 A0.0 R0.0 Standard Full No No No
Series 2 Group 1 Recon 2 Settings Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter Group Thick Interval Filter
1 1.25 1.25 Standard 1 1.25 1.25 Standard

Gray Zone--CT Angio >200 lbs = >90.8 kg


Series 1 Scout FeetFirst Supine
Auto Store Y/N Gating NO
Auto Transfer Y/N
Scan Kv mA Start End Plane Voice Lights Timer
1 80 10 S0.0 I400 90 2 No No
2 80 10 S0.0 I400 0 2 No No
Series 2 Axial FeetFirst Supine
Auto Store Y/N Gating NO Smart Prep NO Biopsy NO
Auto Transfer Y/N
Series 2 Group 1 Scan Settings
Group Images Speed Type Start End Thick Speed Rows Interval
1 159 0.5 Helical S 0.0 I 395 2.5 27.5 16 2.5
Tilt SFOV Kv SmartmA Noise Indx Min mA Max mA Voice Light Timer
© Copyright 2008 General Electric Company. All rights reserved.
S0.0 Large 120 Yes 15 120 300 2 No No
Series 2 Group 1 Recon 1 Settings Direction 5327782-1EN
Group DFOV A/P R/L Filter Type Vari D3D DMPR DOC 0509391
1 36 A0.0 R0.0 Standard Full No No No Page 13
Series 2 Group 1 Recon 2 Settings
Group Thick Interval Filter
1 1.25 1.25 Standard

You might also like