Manual Power Meter
Manual Power Meter
Vega
Laser Power/Energy Meter
User Manual
1
Chapter 5. Operation with Photodiode Type Sensors59
2
Chapter 9. Calibration, Maintenance and
Troubleshooting ............................................. 91
3
Chapter 1.
Introduction: How to Use This Manual
This manual tells you what you need to know to make full
use of the Vega for all your laser measurement needs. It
includes a "Quick Reference", (Chapter 2) to allow you to
perform basic measurements immediately, without
reading the whole manual.
4
Chapter 2.
Quick Reference
Figure 2.1
Vega Rear Panel View
5
Figure 2.2
Briefly press the on/off switch. (Figure 2.2). The unit will
switch on, and the display will appear.
Note:
To work with a sensor, the sensor must be plugged in
before the unit is switched on.
6
To toggle the state of the backlights on and off:
To zero instrument:
7
Adjusting GUI and Keypad Settings:
Language Options:
Warning:
In the event that you mistakably saved the Japanese
language as the Vega startup language, the way to
switch back to English is:
a. Disconnect the sensor from the Vega.
b. Turn the Vega off and on again. You will reach the
instrument setup screen (titled “head disconnected”
in Japanese).
c. Press the down navigation key 4 times.
d. Press the Enter. This will open a window to select
the instrument language.
e. Select "English" and press the Enter key.
f. Press the “save” soft key and English will now be
saved as your startup language.
8
2.2.3. Updating the firmware of the Vega.
9
To set type of laser being used:
10
To select a range for power measurement:
11
6. Select “Energy Range” and choose the default
energy range.
7. Select “Threshold” if you wish to change the energy
threshold. See section 4.5.3 for details.
8. Select “Laser” and select the laser type you wish to
be the default.
9. Press "Save" to save all present settings then "Exit" to
get back to the main measurement screen.
Further details on configuration in section 4.3.
Warning:
Do not exceed maximum sensor limits for power,
energy, power density and energy density as listed in
Table 5. and Table 6. in Chapter 10. Otherwise, there
is a risk of damaging the absorber.
12
To subtract background and set current reading to zero:
To save configuration:
13
2.4 Photodiode Sensors
14
2.4.4. Setting the Startup Configuration
Warning:
Do not exceed maximum sensor limits for power,
energy, power density and energy density as listed in
Table 5. and Table 6. in Chapter 10. Otherwise, there
is a risk of damaging the absorber.
15
1. From the main power measurement screen press
"Offset".
2. Press "Offset" again to cancel. See Section 5.4.5 for
full details.
16
2.5.2. Changing Chosen Wavelengths (metallic type
only)
17
5. Select “Average” to select the average over period
or select the default, "NONE".
6. If this sensor has a diffuser sensor, set Diffuser to IN
or OUT.
7. Now select “Max Pulse Len” and choose the
shortest pulse length setting longer than your laser’s
pulse length. Warning: Incorrect readings will result
if pulse length is not set up correctly.
8. Select “Laser” and select the laser type you wish to
be the default.
9. Press "Save" to save all present settings then "Exit" to
get back to the main measurement screen.
Note:
Some sensors have only the one pulse length setting
for all pulse lengths. In that case N/A appears. For
further details on configuration see Sections 6.2 and
6.3.
Warning:
Do not exceed maximum sensor limits for power,
energy, power density and energy density as listed in
Table 5. and Table 6. in Chapter 10. Otherwise, there
is a risk of damaging the absorber.
Note:
High sensitivity pyroelectric sensors (PE-9, PE-10) are
very sensitive to vibration, and therefore might read a
false trigger when operating in an acoustically non-
stable environment. Set the threshold to a high
enough value that false triggering does not occur.
Ophir also offers a shock absorbing mounting post
18
(P/N 7Z08268) that helps reduce vibration on the
sensor.
Warning:
While measuring pulsing lasers, erroneous energy
reading will result if energy range is not set up
correctly. See Section 6.4 for details.
20
Chapter 3.
The Vega Display Unit
21
3.2 Smart Connectors
Note:
In ROM versions earlier than 1.96, the Vega does not
recognize when a sensor has been disconnected.
Therefore, when changing sensors, the display should
be turned off; the new sensor plugged in; and then the
Vega should be turned on again. From ROM version
1.96, this is no longer necessary, as the Vega will
recognize when a sensor has been disconnected and
will restart itself. See section 3.6.5 for details
22
Key Functions
Navigation Keys:
These buttons move to the next active item
leftwards, rightwards. These are the
navigation keys. The right and left buttons
move to the next active item rightwards and
leftwards. The up and down buttons move to the next
active item upwards and downwards. The round button in
the center is an enter button to enter the screen of the
item selected or when finished, to leave that screen.
Go or Start:
This puts into operation the active item.
Exit:
This gets you to the previous screen and saves changes
made until instrument is turned off.
Save:
This saves changes permanently. The change is
implemented now and is also saved for future use of the
instrument.
23
3.4 Power Up and Shut Down
Figure 3.1
24
To turn the Vega on:
The unit will switch on, and the display will appear. If no
sensor is connected, a display showing "Head
Disconnected" will appear. If a sensor is connected, the
appropriate default measurement screen will appear.
25
3.5.1. Display Mode
Figure 3.1.1
Needle Screen
26
For thermal and photodiode sensors in power mode, the
analog output is continually updated 15 times per second
with the latest power measurement. For thermal sensors
in single shot energy mode, the analog output can be
configured to be held fixed until the next pulse triggers or
to be released when ready to measure the next pulse. For
pyroelectric sensors, the analog output is updated at up
to 10 times per second with the latest pulse energy.
27
Note (for Thermal Sensors only)
For best results with thermal sensors, it may be
necessary to do the procedure once with the sensor
disconnected then afterwards again with the sensor
connected.
3.6.1. Backlights
28
Nevertheless, since it does shorten the time between
charges, it is recommended to operate the instrument
without the backlight when the instrument is operated
without the charger.
3.6.2. Charging
29
The battery charge is indicated by the
icon. The battery charge is shown
approximately by each segment of the icon, e.g. if 2
segments are shown the battery is ½ full. When the
battery is charging, the segments turn on in sequence.
When the battery is low, the charger should be plugged
in. While the unit will charge even while on and even with
the backlight on, it will charge fastest if the charger is left
in while the unit is off.
Figure 3.2
Vega Rear Panel View
30
inserted before switching on the instrument and then the
instrument will automatically either be in RS232 or USB
mode depending on which cable is attached. See section
4.4.2.6.1 for details of communication with PC.
Figure 3.3
Head Disconnect / Connect Warnings
Note:
When working with Ophir’s StarLab application, the
Vega’s restarting itself will be recognized as a Device
Disconnect.
31
3.7 GUI and Keypad Settings
Figure 3.4
Vega GUI Setup Screen
32
Toggle: Set keypad lighting on and off with a short
press of the on/off key.
Power Save: Light all of the keys whenever one is
pressed. If 10 seconds elapse since the last key
press, turn off the keypad lighting.
33
will be pink, modification windows will be white, and help
screens will be yellow/gold. The keypad will take on the
color of the active window.
Figure 3.5
Display of different windows and their colors
3.7.3.2. Monochrome
34
“Changing”, then the color of the bargraph and the digital
readout changes as a function of percentage of full scale.
For example, if the reading is less than 30% of full scale, it
will be displayed green. Between 30% and 60% it will be
displayed black or white. Between 60% and 90% it will be
displayed yellow and above 90% of full scale it will be
displayed red.
Figure 3.6
Bargraph Style set to Changing (full scale is 3.0µW)
35
Chapter 4.
Operation with Thermopile Absorber Sensors
Warning:
Before using the sensor for power or energy
measurement, check that your laser power, energy
and energy density do not exceed the sensor ratings.
See Table 5. and Table 6. in Chapter 10.
Table 1.
Minimum Flow Rates For Water-Cooled Sensors
36
Energy of a single pulse is measured on the Vega by
digitally integrating the pulse power over time.
37
autoranging mode when switched on, save this
configuration before switching off (see Section 4.3).
38
“Configure” and enter again. Now choose the desired
startup parameters as follows:
39
4.4 Power Measurement
4.4.2.1. Zoom
40
indicating that zoom is engaged. Now press the “Zoom”
button again to return to the unexpanded bargraph
display.
4.4.2.2. Offset
41
the middle of the screen. This allows you to devote all
your attention to tuning the laser, without having to worry
about the Vega settings.
You can now use the tune screen to adjust and maximize
your laser power. (See Figure 4.1). Note that
you can not only see the trend, but also see if you have
already reached a higher value before the numbers on
the left side show you the present power reading as well
as the highest one reached till now. Press “Max”
to reset the maximum reading to the present time. Now
measure the power of the laser and adjust the laser until
you determine that the power has reached a maximum.
42
Figure 4.1
Power Tune Screen
4.4.2.4. Average
43
2. Select the average period you want or select “NONE"
if you do not want to average. Save this as the startup
default if you want by pressing “Save” otherwise just
exit the screen.
3. Note:
Average setting only affects the Numeric display.
Graphic readouts are not affected.
Figure 4.2.
Thermopile Power Screen with average set to 3 seconds
4.4.2.5. Normalize
44
3. If you wish to apply a new value, press “Reset” and
then “Apply” again.
4. If you wish to return to the the regular display, select
“Menu”, enter and go to the display you wish.
Figure 4.3
Thermopile Normalize Screen
45
the sampling rate to set the time interval and log size
and enter (Note: for log periods of 5 minutes or less,
this setting has no affect and every measurement will
be stored). To save these values as defaults, press
“Save” after entering them.
3. Now press “Log”.and you will be in the power graph
screen (See Figure 4.4). Press “Go” and the
measurement will begin immediately. The “Go”
button will now change to “Pause”. If you wish to
pause the measurements without losing data
gathered, press “Pause” . Press “Pause” again to
resume measurements. When the data register is full,
“FULL” will flash on the screen. If “Zoom” is pressed
at any time, the screen display will expand fivefold
around the present power value. Press “Zoom” again
to cancel the expansion. If you want to start a new
measurement and not save the present data, press
“Reset”. If you want to save the data before making
a new graph, see the next section.
Note:
For data logging, the instrument has to be in a manual
power range. If the sensor is configured for
autoranging the, Vega will re-configure it to be in the
manual range that is closest to the last set of
measurements that were taken.
Warning:
Pressing “Reset" will erase all previously stored data
and start recording new data. Pressing any of the
navigation Keys will end the log session and return to
the setup screen.
46
Figure 4.4
Laser Power vs Time Graph
Figure 4.5
Data Storage Screen
47
2. When the screen is entered, the data in the latest
power graph will be stored temporarily in register
“Temp”. With the navigation buttons select the
permanent register you wish to use for storing the
data and press “Save”. The data in the “Temp”
register will be transferred to the selected register.
This data storage is permanent and the data is not lost
when the instrument is turned off. If you try to save
into a file which already has data, a warning will
appear that the new save will erase previous data in
the file. If you want to delete the data from a file,
select the file and press “Delete”.
3. At the bottom of the screen is a summary of the data in
the selected file. If you press “Exit” and then select
“View Files” and enter, you will be able to show the
actual data points instead of a graph. Select the data
file you want and press the Enter Key (middle button
of the navigation keys). This will open a screen
showing all the data points collected in this file. Press
“XXX>” to get to the next page of data and “<XXX” to
get to a previous page. Pressing “Exit” will get you
back to the previous screen.
Note:
In version 1.96, the internal file system format has
been changed to enable logging of larger files. To
prevent possible corruption, logs made in versions
earlier than 1.96 should be uploaded to the PC before
upgrading the firmware version to 1.96 or higher.
Data Transmission to PC
48
Power Measurement RS232 USB
Power log period 20 sec to 600 hours for both
Max points stored onboard
27000 27000
per file (1)
Max points direct to PC file
unlimited unlimited
in real time
Energy measurement RS232 USB
Max real time data logging
>30Hz >2000Hz
rate(2)
Max onboard data logging
4000Hz 4000Hz
rate(2)
Data transfer from
~500 points/s ~500 points/s
instrument to PC
Max points stored onboard 250,000 250,000
4.5.1. Setup
Note:
The present range is displayed on the upper left and
bottom right of the energy measurement screen as
shown in Figure 4.6.
49
Figure 4.6
Thermopile Energy Screen
4.5.2. Measurement
51
4.5.5. Measuring Energy of Rapidly Repeating Pulses
The standard Vega will only measure individual pulses
every 5 seconds or so. You can also calculate the
average energy of rapidly repeating pulses by measuring
average power on the power setting and using the
formula:
For rapidly repeating pulses, you can use one of the Ophir
pyroelectric sensors, as long as the pulse energies do not
exceed the ratings of the pyroelectric absorbers. The
pyroelectric sensors are compatible with the Vega and
just have to be plugged in to be used.
For higher pulse energies, you can use the Ophir model
RP to measure both average power and exact individual
pulse energy of the rapidly repeating pulses.
Note:
For Thermopile sensors logging energy, there are no
setup options.
52
4.5.6.1. To use the Vega for graphical display of energy
- regular mode
Warning:
Pressing "Reset" will erase all previously stored data.
Pressing any of the navigation keys will end the log
session and return to the setup screen.
Note:
Zoom can only be pressed after at least one pulse has
been recorded.
53
Figure 4.7
Energy Log Screen replace
54
Press “Save” if you wish to save this value and then
press enter to return to the standard measurement
screen.
Figure 4.8
Attenuation Screen with Beam % set to 20%
Note 1:
The attenuation screen only affects the value
displayed on that screen and nothing else. The
power and energy scales remain the same, the
damage thresholds remain the same etc. The
attenuator screen is only an extra screen to show a
different calculated number introduced by the user.
Note 2:
The attenuation value is stored in the Vega and not
the sensor so if it is set, any sensor connected will
come up with the value chosen.
55
4.6.2. Power and Energy Density Screens
The Vega gives you the option to set upper and lower
limits of power and energy. If these limits have been set
in the limit screen, then when the power or energy is
above the maximum limit, the message “OVER” will
appear on the screen and when under the minimum limit,
the message “UNDER” will appear on the screen. When
56
in one of the full color modes, the warning and the
reading will be shown in red.
57
To enter the needle screen do as follows:
58
Chapter 5.
Operation with Photodiode Type Sensors
Warning:
Before using the sensor for power measurement,
check that your laser power or energy and energy
density does not exceed the sensor ratings. See
Table 5. and Table 6. in Chapter 10.
59
measure to 30mW or more without saturation. When the
additional filter is installed, the maximum power is on the
order of 300mW (or 3W with model PD300-3W). The
PD300 saturates when the output current exceeds 1.3mA
so the exact maximum power depends on the sensitivity
of the detector at the wavelength used. When saturated,
the legend "OVER" will appear on the screen. Table 2
gives the actual maximum power as a function of
wavelength.
Filter Out
Wave PD300 PD300 PD300 PD300 3A-IS Wave- PD300 PD300
Length -TP -3W -UV Length -IR -IRG
250-350nm N.A. N.A. N.A. 3mW N.A. 800nm 12mW 0.8mW
400nm 30mW 3mW 100mW 3mW N.A. 1-1.3µm 30mW 0.8mW
633nm 20mW 2.5mW 100mW 3mW 1W 1.4µm 30mW 0.8mW
670nm 13mW 2mW 100mW 3mW 2W 1.5µm 25mW 0.8mW
800nm 10mW 1.5mW 100mW 2.5mW 3W 1.6µm 30mW 0.8mW
900mm 10mW 1.5mW 100mW 2.5mW 3W 1.8µm 30mW N.A.
1060nm 25mW 3mW 100mW 3mW 3W
Filter In
Wave PD300 PD300-TP PD300- PD300-UV 3A-IS Wave PD300-IR PD300-
Length 3W Length IRG
250-350nm N.A. N.A. N.A. 300mW N.A. 800nm 0.8mW 100mW
400nm 300mW 1W 3W 300mW N.A. 1-1.3µm 0.8mW 30mW
633nm 300mW 1W 3W 300mW 1W 1.4µm 0.8mW 150mW
670nm 200mW 500mW 2W 300mW 2W 1.5µm 0.8mW 150mW
800nm 100mW 300mW 1.2W 150mW 3W 1.6µm 0.8mW 150mW
900mm 150mW 300mW 1.2W 150mW 3W 1.8µm N.A. N.A.
1060nm 250mW 500mW 2.2W 300mW 3W
Table 2
Maximum Measurable Laser Power as a Function of Wavelength
60
1. From the power measurement screen, select
“Laser” and enter.
2. With the up/down navigation keys, select the first
wavelength desired. Press the right navigation key
to enter the change wavelength screen.
3. Enter the wavelength value by pressing the
up/down navigation keys to change each number
and the right/left navigation keys to move to the
next number. After you are finished, press “Save”
and enter.
4. Repeat steps 2 and 3 for other wavelengths desired.
When finished, press Enter. Those values for which
“Save” was pressed will be default startup values,
the others will be saved until the instrument is
turned off.
63
5.4.3. Operation with Filter In or Out
Warning:
If the PD300 is used in the "Filter IN" setting and the
filter is not installed or vice versa the readings will
be completely incorrect.
If the power of your laser exceeds the maximum for
filter in, you can purchase a thermal or integrating
sphere sensor for that wavelength. Consult your
Ophir agent for details.
64
Note:
The filter status can be seen at a glance on the
screen: filter out will be shown by Filter
and filter in by Filter
5.4.4. Zoom
5.4.5. Offset
65
value, and the second will activate a new value. If you
suspect that the Vega has a permanent zero offset, then
disconnect the sensor while it is in the power
measurement mode.
Note:
If there is a zero offset in the reference value, you
cannot subtract this using the dB offset function.
Instead, before the start of the measurement, press
“Offset” and subtract the zero offset. Then follow
steps 1 and 2 above. The zero offset subtracted when
“Offset” was pressed will be saved in the dBm scale
and you can now use the dB-O setting to measure
true ratio without zero offset problems.
Note:
The average function does not work with autorange
or dBm scales.
The Vega also has the option of graphing the laser power
vs. time. While measurements are being taken, the Vega
will record data until the screen is full, or "Reset" or one
of the navigation keys is pressed.
67
Chapter 6.
Operation with Pyroelectric Sensors
Warning:
Before using the sensor for power or energy
measurement, check that your laser power, energy
and energy density do not exceed the sensor ratings
as listed in the table with the sensor specifications.
Otherwise, there is a risk of damaging the absorber.
68
frequency information, the Vega is also able to display
average power.
69
PE05 10235
Range: 20mJ Menu: Energy
Laser: 532 Average: 1 sec
Threshold: 3%
12.34 mJ
142.37Hz TRIG 20mJ
Power 50s Help
Figure 6.1
Pyroelectric Energy Screen
6.2.1. Setting up the sensor to display the user's chosen
wavelengths.
70
“Save” was pressed will be default startup values, the
others will be saved until the instrument is turned off.
Note:
The BB type sensors have less variation with
wavelength, and in those sensors, fixed wavelength
ranges are selected similar to thermal sensors. This
section is not relevant with those sensors.
Warning:
If the pulse width is incorrectly set to a pulse width
shorter than the actual pulse width of the laser, the
reading will be erroneously low. If it is set to a setting
longer than necessary, the reading will be correct but
noisier.
71
Then the user can choose to measure a number of pulses
and display only the average.
Note:
The Pyroelectric sensor is capable of measuring
pulses up to very high repetition rates on the order of
kilohertz or higher. The Vega actually captures each
pulse up to 4000Hz and will sample pulses at up to
25,000 pulses depending on the sensor. The pulses
collected can be stored for later recovery and
transmission. (See section 4.4.2.6.1). However, the
display can only display at rates up to 10Hz. At higher
rates, if the user has chosen “NONE”, the Vega will
display individual pulses
sampled at a rate of 5Hz. If the user has chosen to
average over a time period, the instrument will
display the average of readings over the period from
the time the screen was entered up to the present. For
instance, if the average period is 30 seconds, at 15
seconds, the average is over 15 s., at 30 seconds it is
over 30 seconds, at 5 minutes, it is over the period
from 4.5 to 5 minutes (30 seconds back from the
present etc.).
72
threshold setting is set to 3% of full scale. However, it can
be set to smaller or larger values. If you are measuring
very low energies, for highest accuracy, it is
recommended to set the threshold to the minimum value,
as long as this does not result in false triggering. For
noisy environments with false triggering, it is
recommended to set the threshold to the lowest value
that eliminates the false readings. You can change the
threshold as follows:
73
2. Now select “Display Mode” and select the desired
startup screen. Press “Save”. (Note that this
selection is saved in the Vega and not the sensor, so
will be operative for all sensors). For more details on
the screen types see section 4.6.
3. Now select “Energy Range” and choose the energy
range you wish to be the default. The Vega will now
be in the range selected when switched on with the
sensor in question. Note that there are only manual
ranges in energy.
4. Now select “Average” and select the period you wish
to average readings over. This feature is especially
useful for lasers with unsteady output. See section
4.4.2.4 for details.
5. Now select “Diffuser” if your sensor has a removable
diffuser and select if you wish diffuser in (installed) or
out (not installed) to be the default.
6. Now select “Max Pulse len”, enter and select the
laser pulse width. Select the shortest value longer
than your laser pulse width and press save.
7. Now select “Laser” and select the laser type you want
to be the default. The Vega will now be correctly
calibrated for the absorption of the detector at the
wavelength of the laser type selected and will startup
with the laser wavelength selected.
8. Now press “Save” if you want this to be the startup
default and press “Exit”.
9. Switch the Vega off then on by means of the on/off
switch. (See Figure 3.1). The Vega will enter the mode
of operation last saved with the "Configure" operation.
74
6.4 Energy, Power or Exposure Measurement
Warning:
Before using the sensor for power or energy
measurement, check that your laser energy, energy
density and average power do not exceed the sensor
ratings as listed in the table with the sensor
specifications at the end of the manual, otherwise,
there is a risk of damaging the absorber. With the
pyroelectric sensor, you have been supplied a test
slide with the same coating as on your pyroelectric
detector. You can also obtain this slide from your
dealer. You should use this slide to test the damage
threshold with your laser pulses. If the slide shows
damage, then either enlarge your beam or lower the
laser energy until damage is no longer seen.
Note:
To measure pyroelectric energies properly, it is
important that the sensor is not grounded to the
optical bench. Make sure that the sensor is isolated
electrically from the ground. The PE sensor has been
supplied with an insulating mounting post for this
purpose.
75
against. It is not necessary to do this procedure again
unless the Vega is used with a different sensor. If the
procedure is not done, errors of 2% or so can occur. To
zero the sensor against the Vega, proceed as follows:
Note 1:
Note 2:
77
3. Press “Enter” to return to the measurement screen. If
you want this to be the default range, press “Save”
before pressing the Enter key.
78
Figure 6.2.
Exposure Screen set to stop after summing 500 pulses
Warning:
Pressing "Reset" will terminate the exposure
measurement session.
6.4.3. Normalize
79
reading (dimensionless) on the screen will be 2/1.5 = 1.33.
The screen will also show in small digits the energy at
which the normalization was done and the present
energy reading.
Figure 6.3
Pyroelectric Normalize Screen
80
6.5 Measuring Repeating Pulses of High Energy
Note:
For Pyroelectric sensors logging energy the only
setup option is ‘sample rate’. The user can choose the
number of pulses he wishes to log - from ‘every pulse’
to ‘1 out of 1000’ pulses - but not the time period.
81
Turbo mode
In regular data log mode above, you can only log every
point at up to 180 points per sec. Also you can not
transmit to PC at higher rates than 180 Hz in regular data
log mode.
In order to log data or transmit it to PC at rates >1000Hz,
you must be in turbo mode. This will put you into a
special mode where the Vega can handle data at high
speed while curtailing other functions. To log in turbo
mode, do as follows:
Note1:
The Vega does not measure frequency in turbo
mode. Therefore you have to indicate the frequency
before starting the measurement. This means you
cannot measure pulses with frequencies which vary
during the measurement run.
Note2:
The Vega will not respond to any buttons until the
turbo logging session is over. This applies either for
data logging or data transmission in turbo mode.
See section 4.4.2.6.1 for further information.
82
6.7 Pyro Scope Adapter
1. Plug the adapter into the Vega meter and then plug
the sensor plug into the adapter. Plug the BNC
connector into a standard 1M scope input.
2. Enter power or energy “Menu” and select “Scope”.
The display will now say “In Scope Mode”. Select the
appropriate measurement range and laser
wavelength in the usual way.
3. You can return to the ordinary measurement mode by
pressing “Energy” (or “Power”) from the scope mode
screen.
83
6.8 PE-DIF Diffuser Sensors
84
Chapter 7.
BeamTrack Features
Figure 7.1
Main Track Screen
85
7.2 Menu Screen
To access the various other BeamTrack features, go to
the Menu screen from the main Track Screen.
Figure 7.2
BeamTrack Menu Screen
Figure 7.3
Pointing Stability
86
A circle is drawn around the latest measurement to be
recorded.
Figure 7.4
Statistics Screen
7.5.1. Zeroing.
87
7.5.2. Line Frequency.
Figure 7.5
Configuration Screen
88
Chapter 8.
Circuit Description
The Vega has two circuit boards: the lower board with
analog circuit, digital processor and its support circuit,
and power supplies; and the upper board with LCD and its
driver.
Analog Circuit:
Digital Processor:
89
permanently, even when the instrument is turned off. (See
section 4.4.2.6.1 on data storage). The digital processor
receives data from the A/D converter and translates it
into a measurement of current in Amps. When used with
thermopile sensors, the data is then processed by a
sophisticated digital filter that speeds up the effective
response time of the sensor and rejects noise. In addition,
the MCU converts data from the fast A/D converter and
translates it into a measurement that can be processed
as necessary. The MCU is also responsible for reading
the keypad, driving the LCD display and communicating
via USB or RS232.
Power Supply:
EMI Protection:
90
Chapter 9.
Calibration, Maintenance and Troubleshooting
9.1 Calibration of Thermopile or RP Sensors
1. Surface Absorbers:
BB (broadband) absorber
EX (excimer) absorber
91
2. Volume absorbers:
The models with the P suffix, for use with pulsed lasers,
have a special absorbing glass with an absorbance of 95
±2% over the operating range. Since the surface is a
glass, its reflectivity does not change even if damaged or
melted locally.
Graph 1
Absorption vs. Wavelength of Various Thermal Sensor Absorbers
92
Ophir power/energy meters with the Broadband, P or PF
type absorbers are individually calibrated by laser at
several wavelengths against a NIST calibrated standard
meter. The meter can be switched to give the exact
calibration at the various wavelengths (532nm, 1064nm
and, where applicable, CO2). Since the sensitivity
changes little with wavelength, the user can use the
settings closest to the calibration wavelength with little
error.
9.1.2.1. Linearity
93
Graph 2.
Absorption vs. Wavelength of Various Pyroelectric Sensor Absorbers
94
4. Select “Factor” and enter the “Modify” screen.
Adjust the factor until the power/energy reading on
the screen equals the known power/energy. Press
“Save” and the power/energy calibration factor
under “Updated” will change accordingly. See Note
1.
Note 1:
For units which have different calibration factors, e.g.
CO2, YAG, or VIS, press "Laser" to select correct type
of laser before calibration. When recalibrating using
"Specific: this laser only“ function, only the
calibration of the chosen laser wavelength will
change. If you wish to change the calibration of all
wavelengths proportionately, then use the function
"Overall: all lasers ".
Note 2:
When changing power calibration, this will also
change energy calibration proportionately. However,
changing energy calibration will not change power
calibration.
95
9.2 Calibration of Photodiode type Sensors
96
ignored. The maximum error due to the above factors is
given in Table 3 below.
* Add ±2% to error for filter in (±4% for PD300-UV from 220 to 400nm).
Table 3
Maximum Error as a Function of Wavelength and Filter
Note:
More exact specifications will be found in the latest
Ophir Laser Measurement Instruments Catalog.
Note 1:
The relative readings at different wavelengths are
fixed by the wavelength calibration curve stored in
the sensor EEROM. When changing the calibration at
one wavelength, the calibration at all other
wavelengths will change proportionately.
97
9.3 Calibration of Pyroelectric type Sensors
98
Absorption(%) vs. Wavelength (nm)
BB - Broadband
Metallic Coating
Graph 3.
Absorption of Ophir Pyroelectric Absorbers
9.3.2. Calibration
99
wavelength range of the sensor in a similar fashion to the
PD300 sensors. Then the sensor is calibrated at one
wavelength using a 905nm repetitively pulsed laser
referenced to a NIST traceable photodiode meter. The
average energy is set to the average power of the
standard power meter divided by the laser frequency.
100
1. Set the instrument to “Energy”.
2. Set to the proper energy range.
3. (For broadband type sensors only. See note 1). From
the main screen enter “Menu” and select
“Calibrate”. Enter and select “Adj:” Select “Specific:
this laser only” if you only want to change the
calibration of this wavelength or “Overall: all lasers”
if you want to change all wavelengths together
proportionately.
4. Apply the known laser power/energy.
5. Select “Factor” and enter the “Modify” screen.
Adjust the factor until the energy reading on the
screen equals the known energy. Press “Save” and
the energy calibration factor under “Updated” will
change accordingly. See Note 1.
Note 1:
For metallic and PD10 & PD10-PJ type sensors, when
the calibration is changed at one laser wavelength,
the overall calibration of all other wavelengths
changes proportionately.
101
Sat: When the photodiode current exceeds 1mA, and the
detector starts to saturate, or the pyroelectric voltage
exceeds the maximum, the message "sat” (=saturated) is
displayed.
9.5 Troubleshooting
Problem Cause/Remedy
Instrument will not operate after Check that charger provides
being completely discharged 200mA.
and connected to charger.
Instrument operates with Battery is low. Recharge
charger but not with battery overnight with the Vega turned
alone. off for 7-8 hours. If the Vega still
doesn't work with battery, then
the NiMH battery is probably
dead. Replace battery (see
section 9.6.1.4).
Problem Cause/Remedy
Instrument triggers on Increase threshold level; See
background noise or Section 4.5.3
sometimes fails to catch large
pulse.
Instrument does not show Increase threshold level; See
ready for a long while after a Section 4.5.3
reading is made.
Non-reproducible results when Decrease threshold level: See
measuring very small energy Section 4.5.3
pulses; or no response to
pulses at low energy.
102
9.5.3. Thermal Sensors, Power
Problem Cause/Remedy
Instrument shows zero Check connections between the sensor
reading in both power and the instrument. (See section
and energy modes. 9.6.1.1). Check that the sensor disc is
operative. Resistance between the
pins 1 and 9 of the sensor connector
should be about 11k. If the sensor is
defective, there will be an open or
short circuit.
Instrument responds Replace sensor disc. See Section-
while sensor is cold, 9.6.1.2
but suddenly fails as it
heats up.
Instrument does not If sensor is very hot, allow it to cool.
return completely to Disconnect the sensor from the
zero on power instrument. If readout unit does not
measurement. zero, follow instructions in section
3.5.5. If the offset persists, try zeroing
with the sensor connected as well, as
described in the same section.
Problem Cause/Remedy
Instrument reads Possible electromagnetic
incorrectly or erratically, interference from pulsing laser is
especially on sensitive causing misreading and/or false
scale. triggering
Instrument triggers even Check the following:
without being exposed to Sensor is mounted to stand using
laser pulses. insulated plastic rod provided with
instrument, and not metal rod.
Instrument shows 1. Try keeping cable away from
frequency which is too bench.
high 2. Move sensor/display further away
from EMI.
103
9.6 Maintenance
104
Chapter 10.
Vega Specifications
Input Specifications
Thermal, Photodiode
Input Ranges 15nA - 1.5mA full scale in 16 ranges
A to D Sampling rate 15Hz
A to D resolution and linearity Resolution: 23 bits plus sign. (0.00001%);
Linearity: better than 17 bits (0.0009%)
Electrical accuracy ±0.25% ± 20pA new; ±0.5% ±50pA after 1 year
Electrical input noise level 500nV or 1.5pA + 0.0015% of input range @3Hz.
Dynamic range 9
9 decades (1:10 )
Input Specifications
Pyroelectric Sensors
Input Range 0 - 6V full scale
A to D Sampling rate 4KHz
A to D resolution 12 bits no sign (0.025% resolution)
Electrical accuracy ±0.25% new; ±0.5% after 1 year
Electrical input noise 2mV
General Specifications
Detector Compatibility Thermopile, photodiode and pyroelectric
Analog output 4 scales, user-configurable: 1v, 2v, 5v, 10v full-scale;
0.03% resolution. 100 ohms impedance
Analog output accuracy ±0.2% (of reading) ±0.3% of full scale volts
Dimensions 110W x 40D x 210H
Mass 450g
Display 320x240 pixel Color TFT LCD; Active area 70x52mm approx
Display digit height 17mm
LCD and Keypad lighting LED’s. Operates from charger or battery
Bargraph segments 320
Battery 3 x AA size. 2.4Ah NiMH battery built in.
Charger input DC: 12 -25V, 5W; AC: 9 - 18V RMS, 5W.
Charge time 7-8 hours; automatically switches to trickle
charge when battery is full
Operation between charges With low backlight and keypad unlit: 16 Hrs, thermal and
photodiode, 11 Hrs, pyroelectric sensors;
With high backlight: and keypad lit: 11/8 Hrs respectively
Data Logging and Com.
Power Measurement RS232 USB
Power log period 20 sec to 600 hours for both
Max points stored onboard per file (1)
27,000 27,000
Max points direct to PC file in real unlimited unlimited
time
Energy measurement RS232 USB
Max real time data logging rate (2)
>30Hz >2000Hz
Max onboard data logging rate (2)
4000Hz 4000Hz
Table 4.
Specifications of System/Meter
105
10.2 Sensor Specifications
Table 5.
Max Power Specifications of Sensors
106
PD - Photodiode
P - P type volume absorber for short pulse lasers
PF - volume absorber for short pulses and high average powers
HE/HE1- volume absorber for high energy pulses
EX - Excimer type, volume absorber
PE - pyroelectric metallic or black absorber
BB - broadband surface absorber, high power density
LP1 - broadband surface absorber for highest power density
BF - very high damage threshold, long pulses
FS - Fused silica window close to detector for divergent beams
Note:
For more detailed and exact specifications, see the
latest Ophir Laser Measurement Instruments Catalog.
Graph 4.
Pulsed Laser Damage Threshold for Thermal Sensors
107
108