User'S Manual Volk Pictor Plus Digital Imaging Device
User'S Manual Volk Pictor Plus Digital Imaging Device
- Pictor Plus Slit Lamp Adapter (VP2SLM) that enables the camera to be used as stationary imaging
tool.
IM-071 Rev C
Page 2 of 77
Quick start guide
1. Remove Volk Optical Pictor Plus from the sales package and check
that all parts are undamaged.
2. Install Battery as instructed in Appendix B, page 42 of this
manual.
3. Place Cradle on a desk next to the PC (Personal Computer).
4. Connect the other end of the USB-cable to the PC.
5. Connect power unit to the wall plug (mains)
6. Place Pictor Plus on the cradle. Battery starts to charge. Charge battery for four hours before the
first use. When device is not used, it may be stored in the Cradle.
IM-071 Rev C
Page 3 of 77
Table of contents
Quick start guide .................................................................................................................................... 3
Table of contents ................................................................................................................................... 4
1. Indications for use .......................................................................................................................... 5
2. Contraindications for use of the eye optics Retina module VP2RET and the eye optics
Anterior module VP2ANT ............................................................................................................... 6
3. Warnings and cautions ................................................................................................................... 7
4. Important symbols ......................................................................................................................... 8
5. Parts of the device .......................................................................................................................... 9
6. Usage environment requirements ............................................................................................... 12
7. Operating instructions.................................................................................................................. 13
7.1. Preparations ......................................................................................................................... 13
7.2. Connection to a PC ............................................................................................................... 13
7.3. Basic use – starting up, shutting down and taking an image ............................................... 13
7.4. Attaching and detaching optics module .............................................................................. 14
7.5. Device Menu ........................................................................................................................ 15
7.6. Adjusting focus and automatic focus ................................................................................... 17
7.7. Patient Information Editor ................................................................................................... 18
7.8. Reset button ......................................................................................................................... 21
8. Retinal imaging using optics module VP2RET .............................................................................. 22
9. Eye imaging using anterior ophthalmic module VP2ANT ............................................................ 26
10. General imaging without optics module ...................................................................................... 29
11. Error messages ............................................................................................................................. 31
12. Cleaning instructions .................................................................................................................... 32
13. Device maintenance ..................................................................................................................... 33
14. Technical description.................................................................................................................... 34
15. Warranty....................................................................................................................................... 38
Appendixes:
IM-071 Rev C
Page 4 of 77
1. Indications for use
Pictor Plus is a medical digital camera that is used with dedicated optics lenses intended to take images of
the eye fundus, and surface of the eye.
Component: Description:
VP2RET Module for fundus imaging. The Pictor Plus
VP2HAND digital camera with VP2RET Optics module
is intended to capture digital images and video of
the fundus of the human eye.
VP2ANT Module for exterior ophthalmic imaging. The Pictor
Plus VP2HAND digital camera with VP2ANT Optics
module is intended to capture images and video of
the surface of the human eye and surrounding areas.
IM-071 Rev C
Page 5 of 77
2. Contraindications for use of the eye optics Retina module VP2RET and the eye
optics Anterior module VP2ANT
Because prolonged intense light exposure can damage the retina, the use of the device for ocular
examination should not be unnecessarily prolonged, and the brightness setting should not exceed what is
needed to provide clear visualization of the target structures.
The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure time.
If the value of radiance were reduced in half, twice the time would be needed to reach the maximum
exposure limit.
While no acute optical radiation hazards have been identified for direct or indirect ophthalmoscopes, it is
recommended that the intensity of light directed into the patient's eye be limited to the minimum level
which is necessary for diagnosis. Infants, aphakes and persons with diseased eyes will be at greater risk.
The risk may also be increased if the person being examined has had any exposure with the same
instrument or any other ophthalmic instrument using a visible light source during the previous 24 hours.
This will apply particularly if the eye has been exposed to retinal photography.
Pictor Plus VP2RET and VP2ANT are classified as Class 2 based on standard ISO 15004-2:2007. The daily
usage time and maximum allowed number of pulses is calculated based on optical classification results
according to standard ISO15004-2:2007
CAUTION:
“The light emitted from this instrument is potentially hazardous. The longer the
duration of exposure and the greater the number of pulses, the greater the risk of
ocular damage. Exposure to light from this instrument when operated at maximum
output will exceed the safety guideline after:
IM-071 Rev C
Page 6 of 77
3. Warnings and cautions
Use only accessories and battery provided by Volk Optical with this product.
Place PC and cradle outside the patient environment (at least 4 feet distance from the
patient).
Connection between camera and workstation is USB and/or WIFI. Any authorization
procedures should be carried out in workstation.
Images and videos can be copied from camera to workstation via USB and/or WIFI and
then viewed in workstation.
USB write protection is on by default. When protection is on this feature will prevent
writing to camera memory card from PC when connected to the cradle. In case device has
WIFI functionality, USB write protection must be turned off.
IM-071 Rev C
Page 7 of 77
4. Important symbols
Symbol Description
The CE mark on this product indicates it has been tested and
conforms to the provisions noted within the 93/42/EEC Medical
Device Directive. CE mark with notified body identification
number indicates a class IIa product.
0086
Read accompanying user documentation indicates that important
operating instructions are included in this User’s and Maintenance
Manual. Failure to follow these instructions could place the
patient or operator at risk.
Indicates the need for the user to consult the instructions for use
for important cautionary information such as warnings and
precautions that cannot, for a variety of reasons, be presented on
the device itself.
Type BF applied parts.
9V, 1.1 A
IM-071 Rev C
Page 8 of 77
5. Parts of the device
IM-071 Rev C
Page 9 of 77
Cradle:
IM-071 Rev C
Page 10 of 77
Soft key indicators:
LED indicators:
The recharging and connection to PC is indicated with green (charging) and blue (connection) LED-lights:
Right LED-indicator Blue Active when device is placed to the cradle and connected to a PC.
Graphical indicators:
These indicators are displayed on the top row of the display during imaging:
IM-071 Rev C
Page 11 of 77
6. Usage environment requirements
CAUTION:
Pictor Plus is not suitable for use in the presence of flammable anaesthetics.
Pictor Plus is intended to be used inside in a normal room temperature and normal
humidity. Do not use Pictor Plus in an environment where there is possibility that water
condenses to or inside the Pictor Plus. Type of power source is indicated in Chapter 16,
Technical description.
CAUTION:
It is only allowed to attach USB cable and Power source provided in the sales package into
the cradle. If you need replacement to the USB cable or Power source please contact
Manufacturer or your own Retailer.
USB Cable must be connected only to the USB port of a PC computer that complies with
the IEC 60950 standard. Avoid using excessive power or twisting the connector when
connecting the USB cord to a PC.
Place PC and cradle outside the patient environment (at least 4 feet distance from the
patient).
To transfer patient image data the device must be used together with a PC (Personal Computer) running
either Microsoft Windows XP®, Window Vista®, Windows 7®. The device does not need any drivers to be
installed on a PC.
Device can be used with all major patient database applications supporting both textual and image data
recording.
The device must be used according to this manual, quick reference guide and / or information found on
Volk Optical’s website http://www.volk.com
Electromagnetic compatibility information and recommended separation distances between portable and
mobile RF communications equipment and the Volk Optical Pictor Plus are given in Appendix A.
IM-071 Rev C
Page 12 of 77
7. Operating instructions
This chapter gives instructions for using the device. More specific instructions for using optics lenses are
given in the optics lens specific chapters.
7.1. Preparations
Pictor Plus is both charged and connected to a PC (Personal Computer) using the provided cradle. When
Pictor Plus is not used, it may be stored in the cradle. Storing the device in the cradle is not harmful for the
battery, because the battery is charged only when the charge has dropped below a certain limit.
To connect Pictor Plus to the cradle gently place it into the connector hole. The device can be connected
to the cradle with optics attached.
7.2. Connection to a PC
When connected to a PC running Microsoft Windows XP, Vista or 7, the operating system displays query
for AutoPlay. It is possible to select appropriate image viewing program or simply open the folder to view
and then store files to the hard disk of the PC.
7.3. Basic use – starting up, shutting down and taking an image
It is possible to capture both still images and video. Image capture mode is changed from camera menu
that opens by pressing right soft key for 1s.
Still image is captured using the dual action shutter to the second position. Video is captured by keeping
the dual action shutter pressed down in the second position.
Taken image will stay on display until it is cleared by pressing right or left soft key. Image can be zoomed
in instant preview by pressing middle key. There are four zoom levels. Pressing middle key activates the
next level. Move around the image by using arrow keys.
To transfer images to a PC, place device to the cradle. The image transfer and charging is indicated with
green and blue LED-lights and text on LCD-screen.
It is recommended that image data storage is always erased with a new patient.
Device is powered off by keeping left soft key pressed down over 1 second.
IM-071 Rev C
Page 13 of 77
7.4. Attaching and detaching optics module
CAUTION:
Optic modules used with Volk Optical Pictor Plus must include text “PICTOR PLUS” or
“PICTOR”. It is not allowed to attach other objects to the bayonet connector.
IM-071 Rev C
Page 14 of 77
7.5. Device Menu
Menu has six tabs. One is for device settings such as language
selection. There is one tab for retinal imaging (RET), anterior
eye imaging (ANT), and general imaging (DF).
Arrow keys are used to move between tabs: use arrow key up
until tab is active and use left and right arrow keys to change
active tab. Light blue color indicates active tab.
IM-071 Rev C
Page 15 of 77
Preview images
Image preview is opened by selecting “Preview files” with a middle key. Images are browsed by using
arrow keys. Display gives information about usage of the image preview.
Image can be zoomed while previewing by pressing middle key. Move around the image by pressing arrow
keys. Change between the four zoom levels by pressing middle key.
Display brightness
The Display brightness selection has three options: low, medium and high. Choose the suitable level of
display brightness according to for example how the examination room is lit up.
Icons
The icons shown on camera screen can be enabled or disabled to the user’s liking. The most essential
icons such as Menu icon are always visible.
Sounds
By default the camera makes a sound during image capture. This sound can be turned off from Device
menu.
Keyboard backlight
By default the camera keys have a backlight lit up when the camera is turned on. This light can be turned
off from Device menu if it disturbs the user while taking images.
Language
The camera has nine different languages the user can choose from. Default language is English and the
language selection is always shown in English in menu.
Date
Date can be set using the left and right arrow keys. Move between day, month and year by pressing the
middle key. Date format is DD MM YYYY for Finnish, German, French, Italian, Spanish, and Portuguese.
Format is MM DD YYYY for Japanese.
Time
Time can be set using the left and right arrow keys. Move between hours and minutes by pressing the
middle key.
Camera SW version
This item in Device menu tells the camera SW version.
Start query
User can choose start-up query between erasing images, creating new patient folder or no query at all.
When erasing images is chosen as the query the camera will prompt user on start-up whether all images
and videos shall be deleted. When new patient folder query is chosen the camera will ask if a new patient
folder shall be created upon start-up to prevent from mixing images with previous patient. If no query is
selected then camera will go directly to live view upon start-up.
Image focus can be adjusted manually by pressing arrow keys up and down. When focus mode is set to
manual and Retina optics is attached a dioptre scale is visible on the screen.
IM-071 Rev C
Page 17 of 77
7.7. Patient Information Editor
What Is Patient Information Entry?
The Pictor Plus has a feature that allows entry and editing of patient information (e.g., patient name,
ID number or other identifier information).
The patient information can contain a maximum of 32 characters, including letters (upper or
lower case), numbers and some special characters: ( ) , . : - _ @ / and space.
Which patient eye being imaged – left or right – can also be set.
Additionally, they type of image saved (IR illuminated, normal color, red-free or color
corrected) is also saved automatically.
IR illuminated = IR
Normal color = Co
Red-free = RF
Color corrected = CC
IM-071 Rev C
Page 18 of 77
How Is Patient Information Entered?
Patient information is entered in the Patient Information screen. The Patient Information screen
allows setting of which eye will be imaged (left or right). It also allows access to the Patient
Information editing screen.
The current Patient Information entered will be displayed to the right of the Patient ID icon. The
default value is “ABC12345.”
Use the right soft key under the Cancel icon to discard any changes and exit the Patient
Information screen.
When in the camera live view screen, using the Menu button (right soft key, long press):
Use the right soft key (long press) to access the Menu screen
Under the Device tab, the Patient Information entry screen can be accessed from either the
Preview images icon or the New patient folder icon.
From the Preview images icon, you can use the Patient Information entry screen to edit
Patient Information on images you have already taken.
You can edit Patient Information for ALL of the images in a selected folder,
OR
You can edit Patient Information for a single, selected image file.
o Arrow to the Preview images icon and select with the middle key to access the Image
Browser screen.
o In the Image Browser screen, use the arrow keys to navigate to a specific folder or a
specific image file.
When only a folder is selected, the folder name will be highlighted with
white text (the filenames will be dark).
When an image file is selected, that filename AND its folder name will be
highlighted with white text.
o When your selected folder or image file is highlighted, press the right soft key under
the Edit icon to access the Patient Information entry screen and edit the Patient
Information per the above instructions.
REMEMBER:
IF YOU ARE EDITING PATIENT INFORMATION FOR A FOLDER, ALL OF THE IMAGE FILES
IN THAT FOLDER WILL BE CHANGED TO THE NEW PATIENT INFORMATION.
IF YOU ARE EDITING A SINGLE IMAGE FILE, ONLY THAT IMAGE FILE WILL BE
CHANGED TO THE NEW PATIENT INFORMATION
From the New patient folder icon, you can use the Patient Information entry screen to set
Patient Information for images that you are going to take next.
o Arrow to the New patient folder icon and select with the middle key.
o The device will prompt “Create a new patient folder for new patient images?”
Select Yes (left soft key) if you also wish to create a new patient folder in
which to save your next images.
Select No (right soft key) if you do not want a new patient folder.
o Next, the device will prompt “Edit current patient information?”
Select Yes (left soft key) to enter the Patient Information entry screen.
Proceed to edit Patient Information per the above instructions.
Your next images will have the new Patient Information.
Select No (right soft key) to return to the Device tab without editing Patient
Information.
IM-071 Rev C
Page 20 of 77
7.8. Reset button
IM-071 Rev C
Page 21 of 77
8. Retinal imaging using optics module VP2RET
Pictor Plus Retina digital ophthalmic camera is intended to capture digital images and video of the fundus
of the human eye. The device set for retinal imaging consists of:
Camera handset
Attachable ophthalmic lens Retina
Eye cup for Retina
Cradle for charging and image transfer
Pictor Plus Retina is intended for non-mydriatic imaging. This means that infrared is used for targeting
image to the eye fundus and visible light is flashed when image is taken. Pupil does not respond to the
infrared light so examination is convenient for the patient. Constant illumination can be selected from
device menu if mydriatic drops are used. Pictures can also be taken using infrared light for both aiming
and capturing.
Pictor Plus Retina has 9 internal fixation targets for the patient to fixate at while imaging. Below section
will guide how to control the fixation lights.
2. Recommended that both the patient and the examiner should be seated while taking the images.
3. Either autofocus or manual focus can be used. Autofocus range is -11 to +3 diopters, manual focus
range is -20 to +20 diopters.
If patient has a refractive error and autofocus is off, focus need to be adjusted:
Hyperopia: camera is focused to distance by pressing arrow key up. One click of the key is
approximately 1 Dioptre.
Myopia: camera is focused closer by pressing arrow key down. One click of the key is
approximately 1 Dioptre.
5. The middle fixation target is lit when pressing left soft key and it provides a macula centred image.
To change the fixation target press Left Soft Key and use arrow keys to navigate between the 9 targets
as shown in the graphics in lower left corner of the display. If fixation target is turned off ask patient to
look at a target in a wall 6-9 feet (2-3 meters) behind the operator.
6. Light is adjusted using left and right arrow key. There are altogether 10 brightness levels. Default value
is 5. Suitable illumination is typically 2 to 8. When using IR/White capture mode changing illumination
brightness affects only the white capturing flash. If using IR/IR or White/White both aiming and
capturing light are changed.
7. Aim help circle on the screen guides user when to take image. When retina is not fully in view the
circle is red. Once the aim is good and retina fully appears on screen the circle turns green indicating a
IM-071 Rev C
Page 22 of 77
good moment for capturing the image.
8. Approaching the eye is started from 4 inch (10 centimeters) distance. If internal fixation target is not
used patient is asked to look at a target in a wall 6-9 feet (2-3 meters) behind the operator (patient’s
eye targets to infinity and stays still). Pupil is approached until the reflection from the eye fundus can
be seen. The right imaging distance is about 0.8 inch (2 cm). Eye cup must be compressed
approximately half way down. Aim help circle on the display guides to take image once it turns from
red to green.
Camera is stabilised by keeping the outer side of the hand against the patient’s forehead. Example of
the correct usage position is shown below:
9. Still image is captured by pressing the shutter key all the way down. Video is captured by keeping
shutter down. Taken image is displayed on screen until user clears the image by pressing shutter, left
or right soft key. Image can be zoomed in instant preview by pressing middle key. There are four zoom
levels. Pressing middle key activates the next level. Move around the image by using arrow keys.
10. If multiple patients are examined during one session, a new file folder is created for each patient by
pressing middle key for over 3 seconds.
11. Transfer images to a PC after capturing images. Images are transferred to the PC when camera is
placed to the cradle. Pictor Plus works as any other digital camera.
12. To erase the image/video manually, choose Menu, then Preview Images. There will be an option to
erase a specific image or a specific folder.
13. When the camera is removed from the cradle it verifies image data storage erase. It is recommended
that image data storage is always erased before images are captured of a new patient.
IM-071 Rev C
Page 23 of 77
The camera keys function as shown in image below when Pictor Plus Retina optics module is attached:
IM-071 Rev C
Page 24 of 77
Table below includes explanations of the EY settings tab for retinal imaging:
Setting Values (default bolded) Purpose
Capture mode Still / Video Use left and right arrow keys to choose between still capture
and video capture.
Mark side On/Off Mark side of the eye to the image data.
IR brightness Low/High Aiming light brightness
Red-free On/Off Saves a copy of an eye image using green color channel only.
Save IR image On/Off Saves an image taken with IR light in addition to white light
image when turned on.
Color correction On/Off Saves a color corrected image in addition to normal fundus
image.
Illumination mode IR/White, IR/IR, Wh/Wh Use left and right arrow keys to select aiming and capture
light source.
Target blinking On / Off Fixation target light can be set to constant if needed.
Capture mode
Both still images and video can be taken with the Pictor Plus camera. When taking video the shutter
button must be held down while taking the video. The video recording will stop once the shutter button is
released.
Marking side
It is possible to mark which eye was imaged. Marking side is enabled from the menu. When On side is
marked to the file name and to the image. For video files, side is always marked only to the file name.
When marking side is enabled, camera verifies side after each captured image.
Identifiers used for eye images are OS for left and OD for right.
IR brightness
IR brightness can be set to high or low depending on the color of the retina. For dark eyes, use high IR
brightness.
Save Red-free
If red-free image is enabled from menu, then camera will save a picture using only green channel at the
same time when saving original picture.
Save IR image
If Save IR image is enabled from menu the camera will save an image taken with IR image in addition to
saving a color image taken with white light.
Color correction
Turn color correction on when fundus is excessively red. The color correction provides a copy of the image
with normalized white balance and brightness.
It makes most interesting features better visible for bright red fundus.
Illumination mode
Illumination mode can be set to IR/Wh, IR/IR or Wh/Wh. In IR/Wh mode aiming is done using infrared
light and image is captured using white light. The pupil does not react to infrared light and thus it is
recommended to use IR/Wh mode whenever possible. In IR/IR mode both aiming and image capture are
done with infrared light. In Wh/Wh mode aiming and image capture are done with white light
Target blinking
By default the fixation target light is on with blinking illumination. If patient cannot keep their eye
targeted well the light can changed to constant to help focus on the target light.
Pictor Plus Anterior optics module has two light sources, white and cobalt blue light. User can choose
between the light sources by pressing left soft key when Anterior module is attached. Cobalt blue light
enables taking fluorescent pictures of the surface of the eye to reveal any cuts or defects.
1. After Anterior optics module is attached the light source is chosen by pressing left soft key.
2. Still or video image is selected. Selection is made in ES menu.
3. It is recommended to use autofocus.
4. Aiming light is turned on by pressing the shutter key (in front of the device) half way down.
5. Light is adjusted using left and right arrow key. There are altogether 10 brightness levels. Default value
is 5. Suitable illumination is typically 2 to 8.
6. Place the optics module cup on patient’s eye. Patient can be asked to look at different directions
depending on what area of the eye needs to be captured.
Camera is stabilised by keeping the outer side of the hand against the patient’s forehead. Example of
the correct usage position is shown below:
7. Still image is captured by pressing the shutter key all the way down. Video is captured by keeping
shutter down. Taken image is displayed on screen until user clears the image by pressing shutter, left
or right soft key. Image can be zoomed in instant preview by pressing middle key. There are four zoom
levels. Pressing middle key activates the next level. Move around the image by using arrow keys.
8. If multiple patients are examined during one session, a new file folder is created for each patient by
pressing middle key for over 3 seconds.
9. Transfer images to a PC after capturing images. Images are transferred to the PC when camera is
placed to the cradle. Pictor Plus works as any other digital camera.
10. Erase manually the image or video by choosing Menu – Preview images. The specific image or the
specific folder can be chosen to be erased.
11. When the camera is removed from the cradle it verifies image data storage erase. It is recommended
that image data storage is always erased before images are captured of a new patient.
IM-071 Rev C
Page 26 of 77
The camera keys function as shown in image below when Anterior optics module is attached:
IM-071 Rev C
Page 27 of 77
Table below includes explanations of the ES settings tab for anterior eye imaging:
Capture mode
Both still images and video can be taken with the Pictor Plus camera. When taking video the shutter
button must be held down while taking the video. The video recording will stop once the shutter button is
released.
Focus Window
Focus Window helps user position the image. The Focus Window can be moved on the screen by first
pressing Middle key until icon “F” appears in up right corner of display and then using the arrow keys
when aiming.
Shoot mode
Shoot mode function can be used when autofocus is enabled. There are two shoot modes to choose from:
normal and auto.
In normal mode the camera will focus when shutter button is pressed half way down and take a picture
when shutter button is pressed all the way down.
In auto mode when shutter button is pressed the camera will focus and take a picture automatically.
Marking side
It is possible to mark which eye was imaged. Marking side is enabled from the menu. When On side is
marked to the file name and to the image. For video files, side is always marked only to the file name.
When marking side is enabled, camera verifies side after each captured image. Identifiers used for eye
images are OS for left and OD for right.
Zoom cropping
With zoom cropping on the camera will save the zoomed area instead of the whole image area. By default
zoom cropping is turned off and camera will save the whole image area instead of the area that is visible
on the camera screen when user has zoomed in.
IM-071 Rev C
Page 28 of 77
10. General imaging without optics module
It is possible to take pictures using Pictor Plus without any optics module. When taking images without
additional optics, integration time that corresponds to the brightness of the image is adjusted using left
and right arrow key. Indicator for integration time is shown at the bottom of the display.
Below picture shows the functions of camera keys when no optics is attached.
IM-071 Rev C
Page 29 of 77
Table below includes explanations of the DF settings tab for general imaging:
Capture mode
Both still images and video can be taken with the Pictor Plus camera. Capture mode can be chosen in DF
menu and left soft key when in live view. When taking video the shutter button must be held down while
taking the video. The video recording will stop once the shutter button is released.
Focus Window
Focus Window helps user position the image. The Focus Window can be moved on the screen by first
pressing Middle key until icon “F” appears in up right corner of display and then using the arrow keys
when aiming.
Shoot mode
Shoot mode function can be used when autofocus is enabled. There are two shoot modes to choose from:
normal and auto.
In normal mode the camera will focus when shutter button is pressed half way down and take a picture
when shutter button is pressed all the way down.
In auto mode when shutter button is pressed the camera will focus and take a picture automatically.
Zoom cropping
With zoom cropping on the camera will save the zoomed area instead of the whole image area. By default
zoom cropping is turned off and camera will save the whole image area instead of the area that is visible
on the camera screen when user has zoomed in.
IM-071 Rev C
Page 30 of 77
11.Error messages
The Pictor Plus will display error messages about the limitations of the usage. An Error message is always
displayed with an explanatory message providing information about possible actions.
IM-071 Rev C
Page 31 of 77
12.Cleaning instructions
Volk Optical Pictor Plus is a precision optics instrument that should be handled with care. Please note
following cleaning instructions:
Clean Eye optics and Eye surface optics, eye cup, before each use on a new patient:
disinfect eye cup with soft cloth moistened with alcohol (e.g. 70% ethyl alcohol), or
soak eye cup in glutaraldehyde based solution, or hydrogen peroxide and peracetic acid solution
such as Erisan OXY+
rinse the eye cup under running water
dry the eye cup (e.g. with clean paper towel) before subsequent use
If replacement for the eye cup is needed, please contact Volk Optical or your own retailer.
CAUTION:
Clean eye cup before each use on a new patient to avoid contamination.
IM-071 Rev C
Page 32 of 77
13. Device maintenance
Pictor Plus contains a rechargeable Ni-MH battery pack. Service life of the battery is approximately 1-2
years. Battery needs to be replaced on planned intervals. When battery is at end of its service life, usage
time of the device goes down.
There are no other maintenance procedures that can be carried out by the user. All servicing and repairs
other than replacing the battery must be carried out by Volk Optical or Volk Optical certified service
facilities and service personnel. Volk Optical will make available work instructions to repair those parts of
medical electrical equipment that Volk Optical has designated as repairable by service personnel.
CAUTION:
If there are breaks in the device covers or other visual defects, contact Volk Optical
or a Volk Optical certified service facility.
IM-071 Rev C
Page 33 of 77
14. Technical description
CAMERA:
Battery: Rechargeable Ni-MH Cylinder cell, HR4U 700 AAA, 4.8 V, 1.0 Ah
User can change battery. Batteries provided by Volk Optical.
CHARGING CRADLE:
CAUTION:
It is only allowed to attach Battery and Power source provided in the sales package into the
Cradle. If you need replacement to the Battery or Power source, please contact Volk
Optical or your own Retailer.
IM-071 Rev C
Page 34 of 77
NON-MYDRIATIC OPHTHALMIC MODULE VP2RET CONNECTED TO
THE PICTOR PLUS HANDSET VP2HAND
Image resolution: 1536x1152 px (total 1,8 Mpix, informational area 1,41 Mpix)
Dimensions: 160x73mm
Weight: 310 g
Dimensions: 79x70 mm
Weight: 90 g
IM-071 Rev C
Page 35 of 77
ENVIRONMENTAL CONDITIONS FOR USE, STORAGE AND TRANSPORTATION:
Temperature, use: + 10 C to 35 C
Relative humidity, use: 10 % to 80 %
Atmospheric pressure: 800 hPa to 1060 hPa
Please note also EMC information given in Annex A.
Temperature, storage - 10 C to 40 C
Relative humidity, storage: 10 % to 95 %
Atmospheric pressure: 500 hPa to 1060 hPa
NOTE: If stored over 1 month, it is recommended to remove battery. Appendix B gives instructions for removing the
battery.
CAUTION:
Do not leave the eye cup in direct sunlight as it may warm up and burn patients
head when taking image.
IM-071 Rev C
Page 36 of 77
SERIAL NUMBERING:
Sticker indicating serial number of the Handset is attached to the Warranty certificate and in the sticker
next to system connector. Warranty certificate is part of this user’s manual
Serial number of the cradle is in a sticker attached to the underside of the cradle.
Serial numbers of the attachable optics modules are attached to the modules and also to the Warranty
certificate.
Year of manufacture is can be found from the serial number (digits 3 to 4).
There are no strict limitations on expected usage life of the Pictor Plus. Usage life is approximately five
years.
Windows XP, Windows Vista, Windows 7 and Windows 8 are trademarks of Microsoft Corporation.
Hama Pro-Optic is a trademark of Hama GmbH & Co KG.
Erisan OXY+ is a trademark of Farmos Ltd
Do not dispose Volk Optical Pictor Plus as unsorted municipal waste. Prepare Volk Optical Pictor Plus for
reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and Council
of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is
contaminated, this directive does not apply.
For more specific information, please contact Volk Optical or your own retailer.
CONTACT
If you wish to contact your local support personnel please call to:
IM-071 Rev C
Page 37 of 77
15. Warranty
Volk Optical gives device a 1 year warranty for the parts and labor. Warranty for battery is 6 months.
Submitting claim:
Any claim under this warranty must be submitted in writing before the end of warranty period to Volk
Optical. The claim must include a written description of the failure that the device have.
IM-071 Rev C
Page 38 of 77
Appendix A - Electromagnetic compatibility information
MEDICAL ELECTRICAL SYSTEM needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL SYSTEM.
Pictor Plus should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal
operation in the configuration in which it will be used.
PICTOR PLUS is intended for use in the electromagnetic environment specified below. The customer or the user of the Pictor
Plus should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment – guidance
Electrostatic discharge (ESD) ±6 kV contact ± 2 kV, ± 4 kV, ± 6 kV Floors should be wood,
IEC 61000-4-2 ±8 kV air indirect contact concrete or ceramic tile. If
± 2 kV, ± 4 kV, ± 6 kV floors are covered with
contact synthetic material, the
± 2 kV, ± 4 kV, ± 8 kV air relative humidity should be
at least 30 %.
Electrical fast transient/burst ±2 kV for power supply lines ±2 kV for AC power supply Mains power quality should
IEC 61000-4-4 ±1 kV for input/output lines ±1 kV for serial cable be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5 ±1 kV line(s) to line(s) ±2 kV ± 1 kV for AC power Mains power quality should
line(s) to earth supply, 1 Phase without be that of a typical
Protective Earth commercial or hospital
environment.
Voltage dips, short <5 % UT (>95 % dip in UT ) for Manufacturer’s test shows Mains power quality should
interruptions and voltage 0.5 cycle conformance to the be that of a typical
variations on power supply 40 % UT (60 % dip in UT ) for 5 requirements of the IEC commercial or hospital
input lines IEC 61000-4-11 cycles 61000-4-11/EN 61000-4-11 environment. If the user of
70 % UT (30 % dip in UT ) for the Pictor Plus requires
25 cycles continued operation during
<5 % UT (>95 % dip in UT ) for power mains interruptions,
5 sec it is recommended that the
Pictor Plus be powered
from an uninterruptible
power supply or a battery.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the Pictor Plus is used
exceeds the applicable RF compliance level above, the Model 006 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Pictor Plus.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IM-071 Rev C
Page 40 of 77
Manufacturer’s declaration – electromagnetic emissions:
PICTOR PLUS is intended for use in the electromagnetic environment specified below. The customer or the
user of the Pictor Plus should assure that it is used in such an environment.
Emissions test Compliance level Electromagnetic environment – guidance
Uses RF energy only for its internal function. Therefore, its
RF emissions
Group 1 RF emissions are very low and are not likely to cause any
CISPR 11
interference in nearby electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
Is suitable for use in all establishments, including domestic
emissions IEC Not applicable
establishments and those directly connected to the public
61000-3-2
low-voltage power supply network that supplies buildings
Voltage
used for domestic purposes.
fluctuations/ flicker
Complies
emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the Volk
Optical Pictor Plus:
The Pictor Plus is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Volk Optical Pictor Plus can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Volk Optical Pictor Plus as recommended below, according to the maximum
output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distanced
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
IM-071 Rev C
Page 41 of 77
Appendix B - Replacing the battery
A battery pack is specially designed and manufactured for this device. Volk Optical and Volk Optical’s
retailers provide suitable battery packs. Labels in the battery and label inside the battery cover display
following information:
NiMH battery
4/HR-4U AAA
4.8V/1000 mAh
4. Put the new battery in the same position and attach the
connector to its socket.
IM-071 Rev C
Page 42 of 77
Appendix C – Wi-Fi Setup
• The Wi-Fi feature of Pictor Plus is provided using the Eye-Fi SD card already
installed in your Pictor Plus.
– The Eye-Fi SD card is an SD memory card with Wi-Fi capabilities.
• The Eye-Fi card requires two companion software called Eye-Fi Center & Eye-
Fi Helper to be installed on your Windows or Mac computer.
– All Eye-Fi card and account settings are managed using Eye-Fi Center.
– Image and video transfers are managed automatically with Eye-Fi Helper.
– A single downloaded file installs both software items.
IM-071 Rev C
Page 43 of 77
Pictor Plus Wi-Fi:
Overview
Below are the main steps to use Pictor Plus Wi-Fi
PC or Mac
PC or Mac
USB
• Connect your Pictor Plus to your +
computer via its USB cradle
IM-071 Rev C
Page 44 of 77
Eye-Fi Center Software:
Overview
Eye-Fi Center:
• Is a user application
• Creates & manages Eye-Fi
user account
• Manages Eye-Fi card settings for your A single Eye-Fi card…
Pictor Plus Eye-Fi card can be on only
one account
Eye-Fi Helper:
• Runs in the background on your An Eye-Fi Card must be removed from a given Eye-Fi account
computer before it can be used with another Eye-Fi account.
• Transfers images and videos
IM-071 Rev C
Page 45 of 77
Important
• The Pictor Plus must be docked in its charging cradle with the USB cable
connected to your computer to complete the Wi-Fi setup instructions using
the Eye-Fi Center software.
IM-071 Rev C
Page 46 of 77
Eye-Fi Center Installation:
Windows
Download Eye-Fi Center for Windows from
http://support.eye.fi/downloads/
• Run Setup.exe
• Select next
IM-071 Rev C
Page 47 of 77
Eye-Fi Center Installation:
Windows
IM-071 Rev C
Page 48 of 77
Eye-Fi Center Installation:
Windows
IM-071 Rev C
Page 49 of 77
Eye-Fi Center Installation:
Windows
IM-071 Rev C
Page 50 of 77
Eye-Fi Center Installation:
Windows
IM-071 Rev C
Page 51 of 77
Eye-Fi Center Installation:
Mac
Download Eye-Fi Center for Mac from
http://support.eye.fi/downloads/
IM-071 Rev C
Page 52 of 77
Eye-Fi Center Installation:
Mac
• Select Continue
IM-071 Rev C
Page 53 of 77
Eye-Fi Center Installation:
Mac
• Select Continue
IM-071 Rev C
Page 54 of 77
Eye-Fi Center Installation:
Mac
• Select Agree
IM-071 Rev C
Page 55 of 77
Eye-Fi Center Installation:
Mac
• Select Install
IM-071 Rev C
Page 56 of 77
Eye-Fi Center Installation:
Mac
• Select Close
IM-071 Rev C
Page 57 of 77
Eye-Fi Center Installation:
Mac
Firewall setup
1
Go to System Preferences…
> Security & Privacy / Firewall
IM-071 Rev C
Page 58 of 77
Eye-Fi Set up: Windows & Mac
After installing Eye-Fi Center:
• Power on the Pictor Plus
– Enter Menu using a long press on the right soft key
– Under the Device tab, arrow down to the USB write-protection icon and
confirm it is OFF
• Connect cradle to AC power via power cable and computer via USB cord
• Wait until the Eye-Fi Helper software establishes connection with the Eye-Fi card and initializes it
– You will see prompts indicating the Pictor Plus Eye-Fi card has been found
IM-071 Rev C
Page 59 of 77
Eye-Fi Set up: Windows & Mac
IM-071 Rev C
Page 60 of 77
Eye-Fi Set up: Windows & Mac
IM-071 Rev C
Page 61 of 77
Eye-Fi Set up: Windows & Mac
IM-071 Rev C
Page 62 of 77
Eye-Fi Set up: Windows & Mac
– Network mode:
The Pictor Plus Eye-Fi card is connected to a known
Wi-Fi network (e.g., is a company or hospital network).
IM-071 Rev C
Page 63 of 77
Wi-Fi Overview: Direct Mode
Images
PC or Mac +
Software
IM-071 Rev C
Page 64 of 77
Wi-Fi Overview: Direct Mode
1
• Connect the cradle power supply to AC power
• Connect the cradle USB cable to your computer
• Place the Pictor Plus in its cradle
2 3
IM-071 Rev C
Page 65 of 77
Direct mode: Windows & Mac
2
1 Select the amount of time you
want the Eye-Fi card to
attempt to connect to your
defined Private Networks.
IM-071 Rev C
Page 66 of 77
Wi-Fi Overview: Network Mode
Images
Images
PC or Mac +
Software
Pictor Plus
IM-071 Rev C
Page 67 of 77
Wi-Fi Overview: Network Mode
1
• Connect the cradle power supply to AC power
• Connect the cradle USB cable to your computer
• Place the Pictor Plus in its cradle
2 3
IM-071 Rev C
Page 68 of 77
Network mode: Windows & Mac
1 Your network
3
2 Your password
4
• Wi-Fi network can be added at the initial setup or later from the
settings.
• Available Wi-Fi networks can be found from the drop down menu.
• Select desired Wi-Fi network and enter password.
IM-071 Rev C
Page 69 of 77
Network mode: Windows & Mac
• Networks can be removed from the list by selecting a network and then
clicking the Remove button.
IM-071 Rev C
Page 70 of 77
Set “Save To” Locations:
Windows & Mac
Images 1 Videos 1
2 2
IM-071 Rev C
Page 71 of 77
Transfer Images via Wi-Fi
PC or Mac
• When you use your Pictor Plus to take images or videos, the Eye-Fi card will connect to the Eye-Fi
Helper software and transfer them to your computer via Wi-Fi.
• As the images are transferred, the Eye-Fi Helper software will display a pop-up window showing
each image being transferred.
• When the Eye-Fi card has completed transferring images, its Wi-Fi will “go to sleep” to conserve
battery power.
• When you take more new images or videos, the Eye-Fi card will recognize the new activity and
“wake up” its Wi-Fi and start transferring the images or videos.
IM-071 Rev C
Page 72 of 77
How To Remove an Eye-Fi Card
From Your Eye-Fi Account
If needed, you can remove your Pictor Plus
Eye-Fi card from your Eye-Fi account.
IM-071 Rev C
Page 73 of 77
Troubleshooting
Problem Solution
Notification saying that Eye-Fi card is write Verify from Pictor Plus Device menu that USB
protected write protection is disabled (OFF)
Eye-Fi card initialization fails Try to use different USB port and reboot PC
Computer cannot connect to Pictor Plus in Direct Verify that Wi-Fi network is enabled in the
Wi-Fi mode Windows firewall. Select either home or work
network.
Eye-Fi account creation does not succeed Verify that your computer is connected to
Internet
Cannot add Wi-Fi network Verify that your computer is connected to
Internet
Computer cannot connect to Internet In direct mode computer cannot connect to
Internet by using Wi-Fi. Connect to Ethernet
cable for Internet connection.
IM-071 Rev C
Page 74 of 77
Troubleshooting
Problem Solution
Eye-Fi center does not detect Pictor Plus when Reboot Pictor Plus and/or reboot computer.
connected from cradle USB cable
Cannot create account or change settings Eye-Fi lift and place Pictor Plus back to cradle
and wait for Eye-Fi Helper to recognize the Eye-Fi
card.
Computer must be connected to Internet.
Deactivating Eye-Fi card from the account does Computer must be connected to Internet.
not succeed
IM-071 Rev C
Page 75 of 77
Ordering Information:
Orders may be placed with the Authorized Volk Optical Distributor in your region. Authorized Distributor contact
information is available directly from Volk Optical.
Volk Optical Inc. Toll free within the United States: 1-800-345-8655
7893 Enterprise Drive Phone: 440 942 6161
Mentor, Ohio 44060 Fax: 440 942 2257
USA Email: [email protected]
Website: www.volk.com
EU REPRESENTATIVE
The Volk authorized representative based
in the European Union (EU) is:
Keeler Limited
Clewer Hill Road
Windsor
Berkshire SL4 4AA U.K.
Tel: +44(0) 1753 857177
0086
Note: This product complies with current required standards for electromagnetic interferences and should not
present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in
close proximity to other equipment.
Members of the European Union should contact their authorized Volk Distributor for disposal of this unit.
Certificate
FM 71461
IM-071 Rev C
Page 76 of 77
Volk Optical Inc.
7893 Enterprise Drive
Mentor, Ohio 44060
USA
Phone: 440-942-6161
Toll free within the United States: 1-800-345-8655
Fax: 440-942-2257
Email: [email protected]
Website: www.volk.com
IM-071
Effective: May 2, 2014
Revision: C
IM-071 Rev C
Page 77 of 77