200250 proTIM-C Module

Description
The 200250 dual-input proTIM-c (conduit mount) provides 2 channels of
measurement. All acceleration-to-velocity (A-V) channels condition the signal
from an accelerometer and integrate it to velocity units. The low frequency A-V
channels are better suited for slower speed applications. A-V with Acceleration
Enveloping (AE) channels provide both integrated velocity units and advanced AE
signal conditioning. Temperature channel types include K-type thermocouples
and 100Ω Platinum (Pt) RTDs with electrical isolation. Each channel is
independent and specified when the proTIM is ordered.

Table 1: proTIM Measurement Types and Transducers


Measurement Type Interfaced Transducer
Acceleration-to-Velocity (General Purpose) 200150
Low Frequency Acceleration-to-Velocity 200155
Acceleration-to-Velocity with AE 200157
K-type Thermocouple 200125*
2-, 3-, or 4-Wire Platinum RTD Industry standard*

*The 200125 is the recommended Trendmaster® sensor for temperature


measurements when the range of a K-type thermocouple is adequate. We do
not recommend the use of other K-type thermocouples to the unique electrical
isolation requirements of the Trendmaster system, and highly recommend the
use of only non-grounded RTDs and non-grounded tip thermocouples to prevent
ground loops. Failure to comply may result in Not OK or NO DATA conditions,
inaccurate readings, or proTIM-C damage. Consult the factory for further
information.

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 1 of 11
Accuracy
Specifications
±15% at 80 Hz
All specifications are at +25 ± 5 °C (+77 ± 9 °F) unless
Frequency
otherwise specified.
Response
Operation outside the specified limits will result in
false or inaccurate readings. 3 Hz to 1 kHz (180 cpm to 60,000
cpm) ± 10% (±0.9 dB)*
200250 proTIM-C and 200150, 200155, 200157 or
200125 transducer systems * The A-V circuitry attenuates
frequencies above 1 kHz. Use of
For detailed specification on the transducers, refer to the 200155 transducer to obtain
the individual transducer data sheets. higher frequency information will
be ineffective.
Electrical: A–V (General Purpose) Channels (with 200150)
Measurement Not OK Range
Range Open transducer signal, power, or
0 to 50 mm/s pk (0 to 2 in/s pk) common is “Not OK”. Shorted
leads are “Not OK” except for SIG+
Over Range shorted to SIG- or common
63 mm/s pk (2.5 in/s pk) shorted to shield.

Resolution Electrical: A–V w/ AE Channels (with 200157)


Measurement
0.025 mm/s (0.001 in/s pk)
Range
nominal
0 to 50 mm/s pk (0 to 2 in/s pk)
Accuracy
Over Range
±15% at 80 Hz
63 mm/s pk (2.5 in/s pk)
Frequency
Response Resolution
10 Hz to 1 kHz (600 cpm to 60,000 0.025 mm/s (0.001 in/s pk)
cpm) ±20% (± 2 dB) nominal
Not OK Range Accuracy
Open transducer signal, power, or ±15% at 80 Hz
common is “Not OK”. Shorted
Frequency
leads are “Not OK” except for SIG+
Response
shorted to SIG- or common
shorted to shield. A-V
Electrical: Low Frequency A–V Channels (with 10 Hz to 10 kHz (600 cpm to
200155) 600,000 cpm) ± 20% (± 2.0 dB)**
Measurement AE
Range
500 Hz to 10 kHz (30,000 cpm to
0 to 50 mm/s pk (0 to 2 in/s pk) 600,000 cpm) ± 20% (±2.0 dB)***
Over Range ** The A-V circuitry attenuates frequencies
above 1 kHz. Use of the 200157
63 mm/s pk (2.5 in/s pk) transducer to obtain higher frequency
information will be ineffective. AE signals
Resolution up to 10kHz are processed at the proTIM.
0.025 mm/s (0.001 in/s pk) *** The 500 Hz filter has a 4-pole
nominal attenuation slope.

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 2 of 11
Not OK Range: Environmental Limits
Open transducer signal, power, or Operating
common is “Not OK”. Shorted Temperature
leads are “Not OK” except for SIG+ -40 °C to +85 °C (-40 °F to +185
shorted to SIG- or common °F)
shorted to shield.
Electrical: Temperature Channels
Measurement
Range
-18 °C to +204 °C (0 °F to +400 °F)
Resolution
0.07 °C (0.12 °F)
Accuracy
K-Type TC
±8 °C (±14 °F), including proTIM-C,
thermocouple & lead wire error,
maximum length of 6 meters.
RTD
±4.45 °C (±8 °F), RTD lead wire
error not included
OK Range
-31 °C to +213 °C (-25 °F to +415
°F)
Not OK
Condition
Temperatures outside the OK
Range
Open RTD or thermocouple wires
are "Not OK"
RTD
Compensation
Coefficient
Alpha in Ω/ΩΩ/°°C
European
0.00385
US Industrial
0.00392
Software
Compensation
At host computer

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 3 of 11
Storage Surge
Temperature Capability
-40°C to +100°C (-40 °F to +212 EN 61000-4-5 (1996), Criteria B.
°F)
Humidity
100% condensing on exposed
surfaces.
100% noncondensing on surfaces
inside conduit.
Note: Apply DC4 grease on connecter
contacts to improve environmental
performance and prevent corrosion.

Enclosure Type
Type 4
Mechanical
Housing
Material
Powder-coated Aluminum
Weight
620 g (22 oz) not including conduit
body
Dimensions
See diagram
Electromagnetic Compatibility
Radiated
Emissions
EN 55011 (1998), Class A.
Electrostatic
Discharge
EN 61000-4-2 (1995), Criteria B.
Radiated
Susceptibility
EN 61000-4-3 (1996), Criteria A.
Conducted
Susceptibility
EN 61000-4-6 (1996), Criteria A.
Electrical Fast
Transient
EN 61000-4-4 (1995), Criteria A.

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 4 of 11
Hazardous Area Approvals Brazil
CSA® Standards BR-Ex ia IIC T4
Installed with MC, AEX-8307-X
intrinsically
safe zener T4 @ -40 °C ≤ Ta ≤ +100 °C
barriers per Intrinsically
drawing 164160 safe entity
Class I, Div 1 Groups A, B, C & D parameters

Class II, Div 1 Groups E, F & G Maximum


Number of
Class III, Div 1 ProTIMs Per
Line

A/Ex ia IIC; Class I, Zone 0 T4 32 (all gas groups)

T4 @ -40 °C ≤ Ta ≤ +100 °C Power Supply


Requirements
Installed
without barriers Ui ≤ 15V Ci ≈ 0 (negligible)
per drawing Ii ≤ 150 mA Li ≈ 0 (negligible)
164161
Signal
Class I, Div 2 Groups A, B, C & D Terminals
T4 @ -40 °C ≤ Ta ≤ +100 °C Ui ≤ 12V Ii ≤ 12 mA
Field Wiring
A/Ex nA IIC, Class I, Zone 2 T4 Parameters

T4 @ -40 °C ≤ Ta ≤ +100 °C See Table 2.

CENELEC Table 2: ProTIM-C Field Wiring Parameters


Capacitance Inductance L/R
Installed with Gas Group
(µF) (mH) (µH/ohm)
intrinsically
IIC 0.09 2 16
safe zener
barriers per IIB 0.705 8 64
drawing 164160 IIA 2.23 16 130
II 1 G Non-incendive
(Zone 2 or Div 2)
EEx ia IIC T4
entity
LCIE 03 ATEX 6459 X parameters
T4 @ -40 °C ≤ Ta ≤ +100 °C Power Supply
Requirements
Installed
without barriers Un = 12 V to 15 V (nominal)
per drawing
In = 50 mA (nominal)
164161
Maximum
II 3 G
cable
EEx nA II T4
See Table 3.
LCIE 03 ATEX 6460 X
Maximum
T4 @ -40 °C ≤ Ta ≤ +100 °C Number of

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 5 of 11
ProTIMs per 200151 Transducer Cables
Line
Used to connect the 200150, 200150, and 2000157
See Table 3. transducers to the proTIM-C.
Table 3: Maximum Cable Length and Number of Operating
ProTIMs per Line Temperature

Maximum -20 °C to +100 °C (-4 °F to +212


Maximum Cable
Gas Group
Length
Number of °F).
ProTIMs per line
Note: These cables may be used at lower
IIC 1000 125 temperatures down to –40 °C (-40
°F), if the cable is not allowed to
IIB 3599 127 move or flex. Flexing these cables at
IIA 4876 127 temperatures below –20 °C (-4 °F)
may damage them.

Minimum Bend
Radius
63.5 mm (2.5 in)
Construction
4-conductor (22 AWG) with foil
shield and drain wire (100%
coverage), polyvinyl chloride (PVC)
outer jacket.
Connectors
Screw-on, 5-pin, keyed connector
on the ProTIM-C end and a
PT06F8-4S (or equivalent) on the
transducer end. Connector
coupling nuts consist of 1/2-20
UNF-threaded 2011 T3 aluminum
or UV-stabilized black nylon.
Contact material is gold-plated
nickel-coated brass.
Classifications:
Cable assembly meets UL 2238.
Cable meets IP67 ingress
protection.

85033 Trendmaster SPA/TIM line cable


Use to connect a SPA to the proTIM-C.
For substitutions, reference guide 101206.
Operating
Temperature
-70 °C to +200 °C (-94 °F to +392
°F).
Conductors
4x 18 AWG stranded, tinned
copper
Specifications and Ordering Information
Part Number 163663-01
Rev. G (03/08)

Page 6 of 11
1x 18 AWG stranded copper, B: Channel B Input Option
tinned overcoat uninsulated drain 01 Acceleration to Velocity
wire (200150)
02 K-Type Thermocouple (200125)
Shielding 03 2 or 3 Wire Pt. RTD
100% aluminum mylar foil out 04 4 Wire Pt. RTD
with helically applied drain wire 05 Low Freq Accel-to-Velocity
(200155)
85% braided tinned copper 06 Accel to Velocity w/ AE
Insulation (200157)
Note: BXX option availability is dependent
Conductors on AXX options chosen. Not all BXX
options are available with each AXX
Fluoroethylene propylene (FEP) option.
Teflon® insulation 0.25 mm (0.010 C: Approvals
in) thick 00 CSA: Class 1 DIV 2, No Barriers
Outer ATEX: Zone 2, No Barriers
05 Multi – Approvals
FEP Teflon insulation 0.38 mm D: Conduit Body Style
(0.015 in) thick 00 No conduit body
Classifications 01 Appleton® Style C body,
malleable iron
NEC article 725 class 3 02 Appleton Style E body,
UL Listed malleable iron
03 Appleton Style C body,
Voltage rating aluminum
04 Appleton Style E body,
300 Vrms
aluminum
Capacitance 05 Weatherproof housing mount
Between
Transducer Cable (for use with 200150, 200150, and
Conductors
200157 accelerometers)
131 pF/m (40 pF/ft) 200151-AXX-BXX-CXX
Between Note: Use the 200151 with the 200250 proTIM
Conductor and only with applications using either a 142485
Drain Wire Housing Cable Adapter or a 141887 Conduit
262 pF/m (80 pF/ft) Cable Adapter
A: Cable Length:
Ordering Information 20 2.0 metre (6.6 feet) cable
40 4.0 metre (13.1 feet) cable
(All conduit bodies have 1-inch hubs.) 60 6.0 metre (19.7 feet) cable
proTIM-C
200250-AXX-BXX-CXX-DXX
A: Channel A Input Option
01 Acceleration to Velocity
(200150)
02 K-Type Thermocouple (200125)
03 2 or 3 Wire Pt. RTD
04 4 Wire Pt. RTD
05 Low Freq Accel-to-Velocity
(200155)
06 Accel to Velocity w/ AE
(200157)
Specifications and Ordering Information
Part Number 163663-01
Rev. G (03/08)

Page 7 of 11
B: Armor Option: 03814231
02 Standard cable, unarmored.
Compression Fitting. For 1-inch
03 Stainless steel over braid
(armored) cable conduit body hubs. Seals for
C: Nut Option: cables when installing proTIM-C
00 Standard aluminum coupling modules without conduit.
nut 85033-02-00
02 Nylon coupling nut
10 Knurled aluminum coupling 300 Meter (1000 ft) Cable. For
nut TIM line .
Note: C01, environmental boot option, is
no longer available and is not
85033-01-00
necessary for use with all 3 existing
stainless steel accelerometers, but
150 Meter (500 ft) Cable. For TIM
may be used with previous plastic line
200150 sensors.
162454
Accessories Dust Cap. For temperature
161934-01 channel connectors.
proTIM-C Installation Manual. 04400000
162411 Dust Cap. For accelerometer
Trendmaster System Manual. channel connectors.

149831-01 141888

Trendmaster DSM Datasheet. Hardware Mounting Kit. For


mounting proTIM-C modules to
01620085 weather-proof housings
Extra Terminal Plugs. For SPA 03810116
line connection. 3 terminal plugs
provided with each proTIM-C Red Silicone Rubber Gasket. Use
module between proTIM-C and
weatherproof housing.
149326-01
163996-01
The 200151 Accel Cable
Environmental Boot Field ProTIM-C Conduit Gasket/Cable
Installation Kit. Contains Seal Installation Kit. Consists of a
everything needed to install the black silicone rubber gasket/seal,
boot in the field. It includes 10 three nylon hose clamps,
sets of boots and clamps, silicone electrical insulating compound,
lubricant, easy-to-use installation and instructions for installation.
tool, and an instruction sheet. 162438-01
You can purchase additional
boots and clamps separately (see Thermocouple and RTD
part number below). The kit is not Connector. With smaller cable
compatible with armored cables crimp seal.
and only compatible with older, 02180005
plastic versions of
accelerometers; not compatible Jumpers. For the proTIM-C
with existing stainless steel thermocouple or RTD terminals.
accelerometers.
04500006
Dow Corning® 4, Electrical
Insulating Compound (5.3 Oz).
Specifications and Ordering Information
Part Number 163663-01
Rev. G (03/08)

Page 8 of 11
Dimensional Diagram
Note: All dimensions in millimetres (inches) except as noted.

200250 Trendmaster proTIM-C

1. Top view
2. Front view
3. Bottom view
4. 1/2-20 5-pole female receptacle
5. Powder-coated aluminum housing
6. Allow 127 mm (5 in) for total height with connector and cable bend
Figure 1: Dimensional Diagram and Wiring Connection Details

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 9 of 11
1. Style C conduit body
2. Style E conduit body
Figure 2: Conduit Body Styles Provided With the ProTIM-C

3
2

5
1
6

1. Style C conduit body


2. Transducer cables
3. Transducers
4. ProTIM-C module
5. Rigid conduit
6. SPA line
Figure 3: Installed ProTIM-C in Style C Conduit Body

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 10 of 11
3

1
1. Retention nuts (2 places)
2. Reinforcing plate
3. Gasket
4. ProTIM-C module
Figure 4: Installed ProTIM-C Module With Weatherproof Housing

Bently Nevada and Trendmaster are trademarks of General Electric Company.


Teflon is a trademark of E.I. du Pont de Nemours and Company.
CSA is a trademark of CSA-International.
Dow Corning is a trademark of Dow Corning Corporation.
Appleton is a trademark of Appleton Electric.
Copyright 2002. Bently Nevada LLC.
1631 Bently Parkway South, Minden, Nevada USA 89423
Phone: 775.782.3611 Fax: 775.215.2873
www.ge-energy.com/bently
All rights reserved.

Specifications and Ordering Information


Part Number 163663-01
Rev. G (03/08)

Page 11 of 11