Hung Ta: Ht-9745 Programmable Temperature and Humidity Walk-In Chamber
Hung Ta: Ht-9745 Programmable Temperature and Humidity Walk-In Chamber
Hung Ta: Ht-9745 Programmable Temperature and Humidity Walk-In Chamber
HUNG TA
QualityTechnologyService
HUNG TA INSTRUMENT CO., LTD.
Add. : No.17 11th Road Industrial Park, Taichung, Taiwan 40707
Tel: 886-4-23590108
Fax: 886-4-23593110, 23588599, 23591149
E-mail: [email protected]
http://www.hungta.com
Remark:
I N D E X
I. RANGE OF APPLICATION
1.1 This constant temperature & humidity chamber is mainly used to test your
specimens by making the specimens exposed under a circumstance with preset
temperature and humidity, imitating how the change of natural environment gives
to the specimens. And so, the R&D people should pay attention about how to
strengthen the characters of atmosphere resistance, and so achieve the target of
products improvement.
1.2 This chamber adopts high sensitivity components and easy-to-operate controllers.
It is very easy and convenient for any user to operate.
II. ILLUSTRATION
2.1 Control
2.2 In-The-Side-Door
2
3
5 4
7
8
10
HUMIDITY
TEMPERATURE
3.1 The above chart is avaiable under the ambient temperature of 25℃, and under
empty load.
3.2 Any requirements beyond the above area are available upon requests and
discussion.
3.3 The the detailed method of operation for the programmable controller please refer to
the separate manual (for FX5461).
1 110V 50Hz
1 220V 60Hz
3 220V 50Hz
3 380V 60Hz
3 415V 50Hz
2.0mm2 8.0mm2
3.5mm2 14mm2
5.5mm2 22mm2
4.7 Even though this chamber adopts stainless steel boards as the shell body, the
atmosphere with gloom and high humidity would cause the controller and some
sensitive parts malfunctional or defective, and which might shorten the lifetime of
some parts. Keep the atmosphere dry, cool and clean.
4.8 If this chamber is running under 0℃, try not to open the chamber door. Because
when it is functioning under low temperature, opening the chamber door would
easily cause the situation of “ice-seal” upon the vaporizer and some other parts.
The lower the temperature, the worse the situation. If it is absolutely necessary to
open the chamber door, try to minimize the time duration.
4.9 The stability of the power is very important to this chamber. If the voltage is lower
than the nominal one, please procure a voltage stabilizer by yourself. Otherwise,
the test won’t start or the motor or the controller could be burnt due to the
insufficient voltage.
4.10 Both the controller and the recorder are of high sensitive instruments made by
precise electronic components, please make sure the wiring for grounding is
well-connected.
4.11 Please make the wiring of grounding well-connected in order to avoid any power
leakage.
4.12 Try not to re-open the freezing system within three minutes after it is shut down.
Otherwise, the compressor would be damaged easily.
V. OPERATION PROCEDURES
5.1 Pre-operation Inspection:
5.1.1 Make sure the power is properly connected, check the voltage value is correct
and the grounding wire is well-connected.
5.1.2 Make sure the RO is installed properly and check the water source for proper
supply.
5.1.3 Make sure the water level control valve is closed when the water supply
bucket is full.
5.1.4 Make sure the drainage valve of the water storage tank is closed.
5.1.5 Make sure the drainage valve of the water return tank is closed.
5.1.6 Check the drainage pipe is well connected.
5.1.7 Is the gauze available in the chamber?
5.3.7 When the temperature and humidity test is going to be performed, be sure to set
the temperature value between 20°C and 95°C
★ If the temperature set is lower than the room temperature, please don’t try to
perform humidity test.
5.3.8 After the program is set, just depress the RUN/STOP button of the program
controller to start testing.
VI. INSPECTION
6.1 Both the water storage tank and water return tank shall be cleaned regular basis
thus maintaining cleanliness of the water in the tanks.
6.2 The water in the humidifying tank shall be drained on regular basis thus preventing
water quality mutation from affecting the function of the heating system.
6.3 When only above room temperature test is performed, the gauze in the wet ball
cup should be removed to prevent the gauze from starching.
6.4 If the gauze has been subjected to the high temperature operation of 85°C or
above, please replace the gauze before next operation; otherwise, the old gauze
cannot absorb the water any more.
6.5 Prior to gauze replacement, the new gauze shall be immersed in hot water to get
rid of the starch on the gauze before use.
6.6 Gauze to be used for testing shall be replaced once in early three months thus
ensuring the accuracy of the experiment.
6.7 Dust on the condenser shall be removed on regular basis and the condenser shall
at least be maintained once a month thus ensuring the function of the condenser.
6.8 During test, if it is necessary to observe inside the chamber, the lighting switch can
be turned on and the changing conditions inside can be observed through the sight
window.
6.9 If this chamber is equipped with recorder, just turn on the button of the recorder, if
necessary.
6.10 Testing hole is provided on the left side of the chamber, it can be connected with
the circuit inside the chamber.
6.11 Once the over temperature light or low water level light comes on, please refer to
Section IV (Trouble-Shooting). If malfunction has been remedied, just press the
reset button to start testing.
6.12 If abnormal condition is found while testing, please refer to Section VII
(Trouble-Shooting).
6.13 Any voltage-withstanding test to the terminal plate (junction box) on the back of the
chamber shall be prohibited absolutely because under high humidity environment,
the terminal plate (junction box) might be short-circuited.
VII. TROUBLE-SHOOTING
Symptom Possible Causes Remedies
End light comes on The test ended. Please depress the RUN/Stop
button on the Program
Controller.
Low water level light 1. Humidifying supply 1. Add water to the supply
bucket is lack of water
comes on bucket
2. Low water level
2. Set low water level protection
protection switch is set
switch at 110°C
too low
3. Contact local Hung Ta
3. Water level controller or
representative
solenoid valve
4. Contact local Hung Ta
malfunctions
representative
4. Low water level
protection switch
malfunctions
Over temperature light 1. Over temperature 1. Set over temperature
comes on protection switch set protection switch at 110°C
too low 2. Contact local Hung Ta
2. Over temperature representative
protection switch
malfunction
Temperature can not rise 1. Wrong temperature 1. Please reset
to the set value setting 2. Contact local Hung Ta
2. Heating system representative
malfunction 3. Close it tightly
3. Door pad leaks or door
is not closed tightly
Water in the humidifying 1. Water valve of the 1. Please open the water inlet
supply bucket can not flow supply bucket does not valve
into the humidifying bucket open 2. Please unplug the water inlet
2. Air pipeline clogging pipe, and let the water flow
3. Water level controller out, then connect it again
or solenoid valve 3. Contact local Hung Ta
malfunction representative
8.1 During the test, the temperature of the machine is over 100 degree, please take off
the wet ball to avoid any hardening which unable to properly absorb water
sensing.
8.2 Please clean the water cooling tower every month regularly.
8.4 Please test the power switch monthly to see if power leakage switch is working
normally or not.
8.5 The material for inner and outer machine is SUS#304 stainless steel, please use
dishcloth with water to clean the machine to remain bright.
8.6 Replace the absorbent gauze per month. Make sure the sensor at the tip of the
holder is totally covered with the absorbent gauze.
8.7 Clean the radiator with air sprayer every six months