Lab Practicle Test
Lab Practicle Test
Apparatus
b) Stop watch
Objective
Procedures
Introduction:
Objectives:
Apparatus:
1. COD Reactor
2. Distilled water and water sample
3. Pipet
4. COD Digestion Reagent Vial
5. DR 2800 Spectrophotometer
Procedures:
1. The power reactor was power on and has been preheated to 150 °C.
2. For BLANK sample: The vial cap was removed for the selected range. Then, the vial
was held at an angle of 45 degrees. The clean pipet was used, and 2.00 mL of sample
was added to the vial.
3. Next, the vial was closed tightly and has been shacked for several time to mix it.
During the mixing, the vial was so hot. Make sure to handle it very carefully and
follow all the precaution.
4. For SAMPLE: The cap from the second vial was removed. Then, the vial was held at
an angle of 45 degrees. The clean pipet was used, and 2.00 mL of sample was added
to the vial.
5. Next, the vial was closed tightly and has been shacked for several time to mix the
sample. The vial was very hot during the mixing.
6. Both of Blank Sample and the Water Sample have been Labelled.
7. The vial was put in the preheated reactor and then, the lid was closed.
8. The start button was pressed, and the vials have been heated for 2 hours.
9. After 2 hours, the vials were put in the tube rack and the vials have been there to cool
at room temperature.
10. Then, the program in the DR 2800 spectrophotometer has been selected to 430 COD
LR.
11. The blank sample cell has been cleaned by using the tissue.
12. The blank sample was inserted into the cell holder.
13. Then, zero was pressed.
14. The 0.0 mg/L COD has shown.
15. The prepared sample cell has been cleaned by tissue.
16. The prepared sample cell has been inserted into the cell holder.
17. Next, the read was pressed on the screen and the result was shown in mg/L COD.
18. The data was recorded in the datasheet.
19. The experiment was repeated from step 1 to step 18 for the other water sample.