2700-2790 Manual
2700-2790 Manual
2700-2790 Manual
DIGITAL AC
Title Page
CLAMP METER
Overview...................................................... 1
MODEL - KM 2700 / KM 2790 Unpacking Inspection................................... 2
Rules For Safe Operation............................ 3
International Electrical Symbols................... 5
The multimeter Structure.............................. 6
Functional Buttons....................................... 7
Display Symbols........................................... 8
Measurement Operation.............................. 10
A. DC Voltage Measurement................... 10
B. AC Voltage Measurement.................... 11
C. AC Current Measurement.................... 13
AC D. Measuring Resistance......................... 14
300 DATA HOLD
A E. Measuring Diodes & Continuity............ 16
F. Measuring Frequency........................... 18
CLAMP METER
2700
General Specifications.............................. 20
WDCV ACV ACA Electrical Specifications............................ 21
1000
200 200V
mV
V 750V
200V
A. AC Current........................................... 21
200K
200
20A
200A
B. AC Voltage........................................... 21
300A
C. DC Voltage.......................................... 21
OFF OFF
D. Resistance........................................... 22
E. Diodes and Continuity Test.................. 22
F. Frequency............................................. 22
500V MAX
COM
!
VW Maintenance............................................... 23
1000VDC
750VAC
MAX
A. Replacing the Battery........................... 23
Test Certificate........................................... 24
Warranty...................................................... 25
OPERATION MANUAL
Overview Unpacking Inspection
! Warning Open the package case and take out the
To avoid electric shock or personal injury, Clampmeter. Check the following items
read the “Safety Information” and “Rules for carefully to see any missing or damaged part :
Safe Operation”carefully before using the
Meter.
Item Description Qty.
Digital Clampmeter Model - 2700/2790 1 English Operating Manual 1 piece
(hereafter referred to as “the Meter”) is a 3½ 2 Test Lead 1 pair
digits Clampmeter with steady operations, and
highly reliable hand-held measuring instrument
In the event you find any Part missing or
having different measurement positions. The
damaged, please contact your dealer
Clampmeter not only can measure AC/DC
immediately.
Voltage, AC Current, Resistance, Diodes and
Continuity, Frequency (in model 2790) but also
has Data Hold, Full Icon Display.
1 2
Rules For Safe Operation (1) Rules For Safe Operation (2)
To avoid possible electric shock or personal -voltage capacitors before testing resistance,
injury, and to avoid possible damage to the Meter continuity, diodes, or current.
or to the equipment under test, adhere to the Replace the battery as soon as the battery
n
~ AC (Alternating Current). 1
DC (Direct Current).
Grounding.
2 3
Diode. 7 COM
!
VW
1000VDC
750VAC
MAX
Fuse. 6 ( Figure 1)
! Warning ! Refer to the Operating Manual.
1) TRANSFORMER JAWS :
Caution ! Risk of Electric Shock. Pick up the AC current flowing through the
conductor.
2) TRIGGER :
Press the lever to open the transformer jaws
When the lever is released, the jaws will close
again.
3) DATA HOLD SWITCH :
A push switch (push ON/OFF) Data-Hold switch
will freeze reading when pressed.
4) LCD DISPLAY :
A 3½ digit display (maximum reading 1999)
indicates measured values, and features symbols
indicating function, Data - Hold, Low Battery,
Continuity, Diode.
5) FUNCTION SELECTOR :
To Select ACV, ACA, DCV, RESISTANCE,
FREQUENCY (in model 2790).
5 6
6) INPUT JACKS (VW and COM) : Display Symbols (see figure 2)
Test leads are inserted into these jacks for
Voltage, Resistance, Frequency
(in Model - 2790) measurements ~ Hz
and Continuity & Diode Checks. H A
7) DROP- PROOF WRIST STRAP :
Prevents the instrument from slipping off
the hand while in use. 2 20 200 V
DATA HOLD
the Meter.
Press DATA HOLD once to enter
n
3
~ Indicator for AC voltage or
current, The displayed value
is the mean value.
(Yellow Button) hold mode. 4 Indicates negative reading.
Press DATA HOLD again to exit
n
7 8
Display Symbols (see figure 2) Measurement Operation
Make sure the Low Battery display - + is not on,
n
2700
CLAMP METER
COM
!
1000VDC
750VAC
VW
Red Black
MAX
( Figure 3 )
! Warning
To avoid harm to you or damages to the Meter from
electric shock, please do not attempt to measure
voltages higher than 1000V DC or 750V AC rms
although readings may be obtained.
The DC Voltage ranges are : 2V, 200V and 1000V. To
measure DC voltage, connect the Meter as follows :
1) Insert the red test lead into the VW input
terminal and the black test lead into the COM input
terminal
2) Set the rotary switch to an appropriate measurement
position in V range.
3) Connect the test leads across with the object being
measured.
The measured value shows on the display.
9 10
Caution : ! Warning :
If the value of voltage to be measured is
n To avoid harm to you or damages to the Meter
unknown, use the maximum measurement from electric shock, please do not attempt to
position (1000V) and reduce the range step by measure voltages higher than 1000V or 750V rms
step until a satisfactory reading is obtained. although readings may be obtained.
The LCD displays “1” indicating the existing
n
Caution :
If the value of voltage to be measured is
n
2700
D. Resistance Measurement (see figure 6)
CLAMP METER
COM VW
!
1000VDC
750VAC
MAX
(figure 5)
! Warning
Never attempt an in-circuit current measurement where CLAMP METER
2700
13 14
To measure resistance, connect the meter as follows E. Measuring Diodes & Continuity (See figure 7)
Note :
nThe test leads can add 0.1W to 0.3Wof error to
the Resistance measurement. To obtain
precision readings in low-resistance, that is the 2700
readings of resistance.
(Figure 7)
nWhen there is no input, for example in open
circuit condition, the Meter displays “ 1 “.
When resistance measurement has been
Testing Diodes
completed, disconnect the connection between
the testing leads and the circuit under test. Use the diode test function to check diodes, The
diode test sends a current through the
Caution: semiconductor junction, and then measures the
1) Never connect high voltage to the input voltage drop across the junction. A good silicon
sockets with the switch in Resistance range. junction drops between 0.5V and 0.8V.
2) Using Resistance measurement function in a
live circuit will produce false results and may To test a diode out of a circuit, connect the Meter
damage the instrument. In many cases the as follows :
suspect component must be disconnected from 1) Insert the red test lead into the VW terminal
the circuit to obtain an accurate reading. and the black test lead into the COM terminal.
2) Set the rotary switch to position.
15 16
3) For forward voltage drop reading on any n When continuity testing has been completed,
Semiconductor component, place the red test disconnect the connection between the testing
lead on the component’s anode and place the leads and the circuit under test.
black test lead on the component’s cathode. The
measured value shows on the display. F: Frequency Measurement (in model 2790)
(see figure 8)
Caution : ~
In a circuit, a good diode will produce a forward
n
voltage drop reading of 0.5V to 0.8V; however;
the reverse voltage drop reading can vary
depending on the resistance of other pathways
between the probe tips.
Connect the test leads to the proper terminals as
n
said above, to avoid error display. The LCD will CLAMP METER
2700
GENERAL SPECIFICATIONS :
19 20
ELECTRICAL SPECIFICATION RESISTANCE
n Accuracies are ± (% reading + number of digits). Range Resolution Accuracy
at 23 ± 5°C less than 75% R.H. 200 W0.1 W
± (1% rdg + 1 dgt)
200 KW100 W
AC CURRENT (50~60 Hz) TEST VOLTAGE : 200W RANGE 3.2V MAX.
Range Resolution Accuracy 200K W RANGE 0.3V MAX.
20 A 0.01 A ± (3% rdg + 4 dgts.) OL - PROTECTION : 500VDC/350VAC ON ALL RANGE
200 A 0.1 A
± ( 2% rdg + 4 dgts.)
300 A 1 A DIODE TEST
600 A* 1 A ± ( 1.5% rdg + 4 dgts.)
TEST CURRENT 1.0 ± 0.6mA
TEST VOLTAGE 3.2V MAX.
AC VOLTAGE (50~500 Hz)
OL - PROTECTION : 500VDC/350VAC.
Range Resolution Accuracy
200 V 0.1 V ± (1.2% rdg + 4 dgts.)
CONTINUITY BEEPER
750 V 1 V ± (1.5% rdg + 4 dgts.)
THRESHOLD < 30W
INPUT IMPEDANCE : 10 M
OL - PROTECTION : 1200 VDC/800 VAC ON ALL RANGES. RESPONSE TIME < 100mS
OL - PROTECTION : 500VDC/350VAC.
DC VOLTAGE
Range Resolution Accuracy FREQUENCY (Autoranging)
2 V 1 mV Frequency function in Model - 2790
200 V 0.1 V ± (0.5% rdg + 1 dgt) Range Accuracy
1000 V 1 V 2 KHz
± (0.5% rdg + 3 dgts)
INPUT IMPEDANCE : 10MW 20 KHz
OL - PROTECTION : 500VDC/350VAC ON 200mV RANGE SENSITIVITY : 80Vrms min.
1200VDC/800VAC ON OTHER RANGE. OL - PROTECTION : 500VDC or AC rms
(50Hz ~ 500Hz).
* 600A range in Model - 2790
21 22
Maintenance
MUMBAI
! Warning
To avoid false reading, replace the battery
TEST CERTIFICATE
as soon as the battery indicator - + appears.
DIGITAL AC CLAMPMETER
To replace battery :
This Test Certificate warrantees that
Disconnect the connection between the testing
n the product has been inspected and
leads and the circuit under test, and remove the tested in accordance with the published
testing leads away from the input terminals of the
specifications.
Meter.
Turn the Meter OFF.
n The instrument has been calibrated by
Remove the screws from the battery compartment,
n
using equipment which has already
and separate the battery compartment from the
case bottom.
been calibrated to standards traceable
Remove the battery from the battery compartment.
n to national standards.
Replace the battery with a new Standard 9V
n
Battery.
Rejoin the battery compartment and the case
n
2700
MODEL NO. ___________
bottom, and install the screw.
DATE: ___________
QC
ISO 9001 KUSAM-MECO
REGISTERED PASS
23 24
WARRANTY THIS WARRANTY IS BUYER’S SOLE AND
EXCLUSIVE REMEDY AND IS IN LIEU OF ALL
Each “KUSAM-MECO” product is warranted to be
OTHER WARRANTIES, EXPRESS OR IMPLIED,
free from defects in material and workmanship
INCLUDING BUT NOT LIMITED TO ANY IMPLIED
under normal use & service. The warranty period is
WARRANTY OF MERCHANTABILITY OR
one year (12 months) and begins from the date of
FITNESS FOR A PARTICULAR PURPOSE.
despatch of goods. In case any defect occurs in
“KUSAM-MECO” SHALL NOT BE LIABLE FOR
functioning of the instrument, under proper use,
ANY SPECIAL, INDIRECT, INCIDENTAL OR
within the warranty period, the same will be rectified
CONSEQUENTIAL DAMAGES OR LOSSES,
by us free of charges, provided the to and fro freight
INCLUDING LOSS OF DATA, ARISING FROM
charges are borne by you.
ANY CAUSE WHATSOEVER.
This warranty extends only to the original buyer or
All transaction are subject to Mumbai Jurisdiction.
end-user customer of a “KUSAM-MECO” authorized
dealer.
This warranty does not apply for damaged Ic’s,
fuses, burnt PCB's, disposable batteries, carrying
case, test leads, or to any product which in “KUSAM-
MECO’s” opinion, has been misused, altered,
neglected, contaminated or damaged by accident or
abnormal conditions of operation or handling.
“KUSAM-MECO” authorized dealer shall extend this
warranty on new and unused products to end-user
customers only but have no authority to extend a
greater or different warranty on behalf of “KUSAM-
MECO”. G 17,Bharat Industrial Estate, T. J. Road,
“KUSAM-MECO’s” warranty obligation is limited, at Sewree (W), Mumbai - 400 015. INDIA.
option, free of charge repair, or replacement of a Sales Direct : (022) 24156638
defective product which is returned to a “KUSAM- Tel. : (022) 24124540, 24181649.
Fax : (022) 24149659
MECO” authorized service center within the
Email : kusam_meco@vsnl.net
warranty period.
Website : kusam-meco.co.in, kusamelectrical.com
25 26