1600041611-KGM-One Unit Can Be Diverted To RGM)
1600041611-KGM-One Unit Can Be Diverted To RGM)
1) Digital Radiography(DR) System with high Frequency X-Ray Machine (600mA) Item
Code: 1820001386.PRNs: 1600045446-KGM &(1600041611-KGM- One unit can be diverted
to RGM).1600045448-RKP & 1600045444-BHPL, MNG-1
A fully Integrated single console controlled Digital Radiography System with 50KW High
Frequency X-Ray Generator for General Radiography with a Dual flat panel detector along with
table bucky and vertical bucky capable of taking the complete range of radiographic examinations
with the following Specifications & Configuration.
Single integrated console system shall have following functions & indications should be
provided. Following features should be available on the console.
• Dual focus Rotating anode X-ray tube with Small Focus .6 mm & Large
focus 1.2mm
• Anode Heat storage capacity of the tube should be 200 KHU or more.
• Collimator with auto shut off facility should be provided.
HV TANK:
A very compact H.V. Tank filled with high dielectric transformer oil should be provided. The H.V.
Tank should contain H.V. transformer, Filament Transformers, H.V. Rectifiers & H.V. Cable
receptacles.
Tube Stand:
• The digital work station should be based on the latest high speed
processor of at least 32bit.
• It should have image storage disc of 1 TB or more and 4 GB RAM or
more
• It should have the ability of multi format printing
• It should have the ability to adjust the contrast ratio and add
annotations
• It should have the ability to print hospital name with patient details.
• It should have the provision of adding patient photo in film or data.
Power Requirement:
Following accessories should be supplied with the machine at the time of installation.
Specifications:
• Principle: High Pressure Liquid Chromatography (HPLC) Method.
• Primary Sampling: No sample preparation is required) whole blood
capillary.
• Dual System: Dual test flexibility both Hb A2 and HbA2/F/A/C without
changing reagents.
• Automatic Sampling: 10 positions including controls & calibrators.
• Results should be displayed on screen and can be transferred to
another computer.
• Storage capacity should be 500 or more chromatograms.
• System should be user friendly.
• Integral printer for Chromatograms.
• The product should be NGSP Certified and also the method is traceable
to diabetes controls and complication trial (DCCT) reference study.
• Two point calibrator and the calibrator should be traceable to IFCC
reference method.
• Should be CE marked & USFDA approved.
• Should provide sire wave 1KVA UPS with 4 hours battery backup.
• LIS soft ware: Should give for HbA1C test along with sample Bar
code and online reporting system.
• 3 years warranty and 5 years CMC.
• Should supply Dual recorder pack, 200T 12 units along with machine.
• Should supply Diabetic bi-level control (6x5ml) 2 Number packs along
with instrument.