Measurement of Resistance
Measurement of Resistance
Measurement of Resistance
Measuring of resistance
1. Introduction
Resistors are used in many places in electrical circuits to perform a variety of useful tasks.
Properties of resistances play an important role in determining performance specifications for
various circuit elements including coils, windings, insulations, etc. It is important in many cases
to have reasonably accurate information of the magnitude of resistance present in the circuit for
analysing its behaviour. Measurement of resistance is thus one of the very basic requirements
in many working circuits, machines, transformers, and meters. Apart from these applications,
resistors are used as standards for the measurement of other unknown resistances and for the
determination of unknown inductance and capacitance. From the point of view of measurement,
resistances can be classified as follows:
Low Resistances
All resistances of the order less than 1 Ω may be classified as low resistances. In
practice, such resistances can be found in the copper winding in armatures,
ammeter shunts, contacts, switches, etc.
Medium Resistances
High Resistances
Resistances higher than 100 kΩ are classified as high resistances. Insulation resistances
in electrical equipment are expected to have resistances above this range.
The above classifications are, however, not rigid, but only form a guideline for the
method of measurement to be adopted, which may be different for different cases.
For zero setting, Rx = 0, where Rm = internal resistance of the basic PMMC meter coil
The current I2 can be adjusted by varying R2 so that the meter current Im can be held at
its calibrated value when the main current I1 changes due to drop in the battery emf E.
If R2 were not present, then it would also have been possible to bring the
pointer to full scale by adjustment of the series resistance R1, but this would
have changed the calibration all along the scale and cause large error.
2. Shunt-type Ohmmeter
Figure 3 shows the schematic diagram of a simple shunt-type ohmmeter. The shunt-
type ohmmeter consists of a battery in series with an adjustable resistance R1 and a
sensitive dc measuring PMMC ammeter. The unknown resistance Rx to be measured
is connected across terminals A–B and parallel with the meter.
Figure 3. Shunt-type ohmmeter
When the terminals A–B are shorted (Rx = 0), the meter current is zero, since all the
current in the circuit passes through the short circuited path A–B, rather than the meter.
This position of the pointer is marked ‘zero-resistance’, i.e., ‘0 Ω’ on the scale. On the
other hand, when Rx is removed, i.e., the terminals A–B open circuited (Rx→∞), entire
current flows through the meter. Selecting proper value of R1, this maximum current
position of the pointer can be made to read full scale of the meter. This position of the
pointer is marked as ‘∞Ω’ on the scale. Shunt type ohmmeters, accordingly, has ‘0
Ω’ at the left most position corresponding to zero current, and ‘∞Ω’ at the rightmost
end of the scale corresponding to FSD current.
When not under measurement, i.e., nothing is connected across the
terminals A–B (Rx →∞) the battery always drives FSD current through the
meter. It is thus essential to disconnect the battery from rest of the circuit
when the meter is idle. A switch S, as shown in Figure 3, is thus needed to
prevent the battery from draining out when the instrument is not in use.
Internal resistance of the ohmmeter
With terminals A–B open, the full-scale current through the meter is
With Rx connected between terminals A-B, the current out of the battery is
Calculate (a) value of the resistor in series with the battery to adjust the FSD, and (b) at
what point (percentage) of full-scale will 100 Ω be marked on the scale?
meter current is
Thus, percentage of full scale at which the meter would read 100 Ω is
Voltmeter–Ammeter Method for Measuring Resistance
The voltmeter–ammeter method is a direct application of ohm’s law in which the
unknown resistance is estimated by measurement of current (I) flowing through it
and the voltage drop (V) across it. Then measured value of the resistance is
This method is very simple and popular since the instruments required for
measurement are usually easily available in the laboratory. Two types of
connections are employed for voltmeter–ammeter method as shown in Figure 4
It is desired that in both the cases shown in Figure 4, the measured resistance Rm would
be equal to the true value Rx of the unknown resistance. This is only possible, as we will
see, if the ammeter resistance is zero and the voltmeter resistance is infinite.
Case A
In this circuit, the ammeter is connected directly with the unknown resistance, but the
voltmeter is connected across the series combination of ammeter and the resistance Rx.
The ammeter measures the true value of current through the resistance but the voltmeter
does not measure the true value of voltage across the resistance. The voltmeter measures
the sum of voltage drops across the ammeter and the unknown resistance Rx.
Let, voltmeter reading = V
And, ammeter reading = I
Error in measurement is
Equation above denotes the fact that error in measurement using connection method
shown in Case A will be negligible only if the ratio . In other words, if the resistance
under measurement is much higher as compared to the ammeter resistance (Ra.), then
the connection method shown in Case A can be employed without involving much error.
Therefore, circuit shown in Case A should be used for measurement of high
resistance values.
Case B
In this circuit, the voltmeter is connected directly across the unknown resistance,
but the ammeter is connected in series with the parallel combination of voltmeter
and the resistance Rx. The voltmeter thus measures the true value of voltage drop
across the resistance but the ammeter does not measure the true value of current
through the resistance. The ammeter measures the summation of current flowing
through the voltmeter and the unknown resistance Rx.
Let, voltmeter reading = V
And, ammeter reading = I
Thus, V=IR×RX=IV×RV
However, I=IV+IR
measured value of resistance
The measured value Rm of the unknown resistance is thus lower than the true value
Rx by a quantity related to internal resistance of the voltmeter. It is also clear from Eq.
above that true value is equal to the measured value only if the quantity , i.e., if
voltmeter resistance is infinite. In other words, if the voltmeter resistance is much
higher as compared to the resistance under measurement (RV >> RA) then the
connection method shown in Case B can be employed without involving much error.
Therefore, circuit shown in Case B should be used for measurement of low
resistance values.
The ammeter reads the current flowing IR through the resistance Rx and also
the current IV through the voltmeter resistance RV.
Thus, I=IV+IR
Now, the voltmeter and the resistance Rx being in parallel, the voltmeter
reading is given by
V=IR×RX=IV×RV
Current through voltmeter
Voltmeter reads the voltage drop Vr across the resistance Rx and also the
voltage drop Va across the ammeter resistance Ra.
Thus, V = Va + Vr
Voltage drop across ammeter
-3
Va = I × Ra = 120 × 10 × 0.8 = 0.096 V
true voltage drop across the resistance
Vr = V - Va = 40 – 0.096 × 39.904 V
Accuracy of this method depends on whether the battery emf remains constant
between the two measurements. Also, other resistances in the circuit excepting R
and S should also not change during the course of measurement. Readings must
be taken fairly quickly so that temperature effects do not change circuit resistances
appreciably. Measurement accuracy also depends on sensitivity of the ammeter
and also on the accuracy of the standard resistance S.
Wheatstone Bridge for Measuring Resistance
The Wheatstone bridge is the most commonly used circuit for measurement of
medium-range resistances. The Wheatstone bridge consists of four resistance
arms, together with a battery (voltage source) and a galvanometer (null detector).
The circuit is shown in Figure 6
In the bridge circuit, R3 and R4 are two fixed known resistances, R2 is a known variable
resistance and RX is the unknown resistance to be measured. Under operating conditions,
current ID through the galvanometer will depend on the difference in potential between
nodes B and C. A bridge balance condition is achieved by varying the resistance R2 and
checking whether the galvanometer pointer is resting at its zero position. At balance, no
current flows through the galvanometer. This means that at balance, potentials at nodes B
and C are equal. In other words, at balance the following conditions are satisfied:
The detector current is zero, i.e., 1D = 0 and thus I1 = I3 and I2 = I4
Potentials at node B and C are same, i.e., VB = VC, or in other words, voltage drop in the arm
AB equals the voltage drop across the arm AC, i.e., VAB = VAC and voltage drop in the arm BD
equals the voltage drop across the arm CD, i.e., VBD = VCD
Hence, voltage difference between the nodes A and C, or the Thevenin equivalent
voltage between nodes A and C is
VTH = VAC = VA – VC = VAB – VCB = 0.798 – 0.8 = –0.002 V
To obtain the Thevenin equivalent resistance between nodes A and C, the 10 V
source need to be shorted. Under this condition, the Thevenin equivalent
resistance between the nodes A and C is thus
Hence, current through the galvanometer is
𝑽𝑺 = 𝑰 × 𝑺
𝑽𝑿
𝑿= ×𝑺
𝑽𝑺
Value of the standard resistor S should be of the same order as the unknown resistance X.
The ammeter inserted in the circuit has no other function rather than simply indicating
whether there is any current is flowing in the circuit is not. Exact value of the current is not
required for final calculations. It is however, desired that the current flowing through the
circuit be so adjusted that the voltage drop across each resistor is of the order of 1 V to be
suitable for accurate measurement by commercially available potentiometers.
Figure 8 Guard circuit for measurement of high resistance: (a) Circuit without guard (b) Circuit with guard
Due to electrostatic effects, stray charges may be induced in the measuring circuit.
Flow of these stray charges can constitute a current that can be comparable in
magnitude with the low value current under measurement in high resistance
circuits. This may thus, cause errors in measurement. External alternating
electromagnetic fields can also affect the measurement considerably. Therefore,
the measuring circuit needs to be carefully screened to protect it against such
external electrostatic or electromagnetic effects.
While measuring insulation resistance, the test object often has considerable amount of
capacitance as well. On switching on the dc power supply, a large charging current
may flow initially through the circuit, which gradually decays down. This initial transient
current may introduce errors in measurement unless considerable time is provided
between application of the voltage supply and reading the measurement, so that the
charging current gets sufficient time to die down.
High resistance measurement results are also affected by changes in
temperature, humidity and applied voltage inaccuracies.
Reasonably high voltages are used for measurement of high resistances in order
to raise the current to substantial values in order to be measured, which are
otherwise extremely low. So, the associated sensitive galvanometers and micro-
ammeters need to be adequately protected against such high voltages.
Taking these factors into account, the most well-known methods of high
resistance measurement is megohmmeter or meggar.
The moving system in such instruments consists of two coils, the control coil CC and the
deflecting coil CD. Both the coils are mounted rigidly on a shaft that carries the pointer as
well. The two coils move in the air gap of a permanent magnet. The two coils are arranged
with such numbers of turns, radii of action, and connected across the generator with such
polarities that, for external magnetic fields of uniform intensity, the torque produced by the
individual coils are in opposition thus giving an astatic combination. The deflecting coil is
connected in series with the unknown resistance RX under measurement, a fixed resistor
RD and then the generator. The current coil or the compensating coil, along with the fixed
resistance RC is connected directly across the generator. For any value of the unknown, the
coils and the pointer take up a final steady position such that the torques of the two coils are
equal and balanced against each other.
For example, when the resistance RX under measurement is removed, i.e., the test
terminals are open-circuited, no current flows through the deflecting coil CD, but maximum
current will flow through the control coil CC. The control coil CC thus sets itself perpendicular
to the magnetic axis with the pointer indicating ‘∞ Ω’ as marked in the scale shown in Figure
9. As the value of RX is brought down from open circuit condition, more and more current
flows through the deflecting coil CD, and the pointer moves away from the ‘∞ Ω’ mark
clockwise (according to Figure 9) on the scale, and ultimately reaches the ‘0 Ω’
mark when the two test terminals are short circuited.
The surface leakage problem is taken care of by the guard-wire arrangement. The
guard ring (GR in Figure 9) and the guard wire diverts the surface leakage current
from reaching the main moving system and interfering with its performance.
Photographs of some commercially available meggars are shown in Figure 10.
Figure 10 Commercial meggars: (a) Analog type (Courtesy, WACO) (b) Digital type (Courtesy, Yokogawa)