UT501/UT502 OPERATING MANUAL

Table of Contents

Title Page
Introduction 4

Unpacking the Meter 4

Safety Information 6

International Electrical Symbols 11

The Meter Structure 12

Key Functions 14

Rotary Switch 15

Measurement Operation 16

A. Measuring Voltages 16

B. Measuring Insulation Resistance 18


1
UT501/UT502 OPERATING MANUAL

Maintenance 21

A. General Service 21

B. Replacing the Battery 23

Specifications 25

Safety and Compliances 25

Physical Specifications 26

General Specifications 27

Feature Summary 27

Detailed Accuracy Specifications 28

A. Voltage Measurement 28

B. UT501: Insulation Resistance Measurement 29

UT502: Insulation Resistance Measurement 30


2
UT501/UT502 OPERATING MANUAL

List of Tables
Table Title Page
1. Unpacking Inspection 5

2. International Electrical Symbols 11

3. Meter Front Description 13

4. Key Description 14

5. Rotary Switch Description 15

List of Figures
Figure Title Page
1. The Meter Front Structure 12

2. Voltages Measurement 16

3. Insulation Resistance Measurement 18

4. Battery Replacement 23
3
UT501/UT502 OPERATING MANUAL

Introduction
Uni-Trend Model UT501 and UT502 insulation resistance tester (hereafter, "the Meter")

is a handheld instrument designed primarily to make resistance/ insulation resistance,

DC/AC Voltage measurement.

Unpacking the Meter


The Meter includes the following items:

4
UT501/UT502 OPERATING MANUAL

Table 1. Unpacking Inspection

Item Description Qty

1 English Operating Manual 1 piece

2 Test Lead 1 set

3 Test probe 1 set

4 Alligator clip 1 set

5 Carry bag 1 piece

6 Strap 1 piece

5 1.5V Battery (LR14) 6 pieces

In the event you find any missing or damage, please contact your dealer immediately.

5
UT501/UT502 OPERATING MANUAL

Safety Information
This Meter complies with the standards IEC61010 -1 safety measurement requirement:

in pollution degree 2, overvoltage category (CAT. III 600V, CAT.II 1000V) and double

insulation.

CAT II: Local level, appliance, portable equipment etc., with smaller transient voltage

overvoltages than CAT. III

CAT III: Distribution level, fixed installation, with smaller transient over voltages than

CAT. IV

Use the Meter only as specified in this operating manual, otherwise the protection

provided by the Meter may be impaired.


6
UT501/UT502 OPERATING MANUAL

Danger identifies conditions and actions that pose hazard(s) to the user.

Warning identifies avoiding electric shock.

Caution identifies conditions and actions that may damage the Meter and carrying

out accurate measurement.

International electrical symbols used on the Meter and in this Operating Manual are

explained on page 11.

7
UT501/UT502 OPERATING MANUAL

Danger

Use of instrument in a manual not specifed by the manufactuer may impair

safety features/protection provided by the equipment. Read the following safety

information carefully before using or servicing the instrument.

Do not apply more than 1000V DC or 750V AC.

Do not use the Meter around explosive gas, vapor or dust.

Do not use the Meter in a wet environment.

When using the test leads, keep your figures away from the lead contacts.

Keep your figures behind the finger guards on the leads.

Do not use the Meter with any parts or cover removed.

When carrying out insulation measurement, do not contact the circuit

under test.
8
UT501/UT502 OPERATING MANUAL
Warning

Do not use the Meter if it is damaged or metal part is exposed. Look for

cracks or missing plastic.

Be careful when working above 30V rms, 46.7V ac rms and 70V DC. Such

voltages pose a shock hazard.

Do not change battery when the Meter is in wet environment.

Place test leads in proper input terminals. Make sure all the test leads

are firmly connected to the Meterís input terminals.

Make sure the Meter is turned off when opening the battery compartment.

Carefully read the operating manual before operating the Meter.

Follow the operating manual all the time when using the Meter and keep

the operating manual in a safe place.

Wrong operation will cause incident and damage to the Meter.


9
UT501/UT502 OPERATING MANUAL

Caution

When performing resistance tests, remove all power from the circuit to

be measured and discharge all the power.

When servicing the Meter, use only the same model number or identical

electrical specifications of test leads and power adaptor.

Do not use the Meter if the battery indicator ( ) shows a battery

empty condition. Take the battery out from the Meter if it is not used

for a long time.

Do not use or store the Meter in an environment of high temperature,

humidity, explosive, inflammable and strong magnetic field. The

performance of the Meter may deteriorate after dampened.

10
UT501/UT502 OPERATING MANUAL

International Electrical Symbols


International symbols on the Meter and in this manual are explained in Table 2.

Table 2. International Electrical Symbols

Equipment protected by double or


reinforced insulation.
DCV DC Measurement
ACV AC Measurement
Grounding
See Manual
Conforms to Standards of European
Union

11
UT501/UT502 OPERATING MANUAL

The Meter Structure


Below Figure 1 and Table 3 shows the Meter front structure and description.

a.UT501 b.UT502
Figure 1. The Meter Front Structure
12
UT501/UT502 OPERATING MANUAL

Table 3. Meter Front Description

1 EARTH: Resistance input terminal

2 G: Voltage Measurement input negative terminal

3 V: Voltage input terminal

4 LINE: Resistance input terminal

5 LCD

6 Display Backlight button

7 Data Hold button

8 Insulation Resistance Button

9 Rotary switch

13
UT501/UT502 OPERATING MANUAL

Key Functions
Table 4. Key Description

HOLD Press once to turn the data hold on, is shown on the display.

Press again to turn the data hold feature off. is disappeared.

LIGHT Press once to turn the display backlight on.

Press again to turn the display backlight off.

TEST Press to stop or start an insulation resistance test.

14
UT501/UT502 OPERATING MANUAL

Rotary Switch
Table 5. Rotary Switch Description

ON/OFF Turn on or off the Meter.

ACV Turn the rotary switch to ACV to measure AC Voltage

DCV Turn the rotary switch to DCV to measure DC Voltage

250V/500V/1000V Turn the rotary switch to 250V/500V/1000V, select the

(UT501 ) requested output voltage, to carry out insulation

resistance measurement.

500V/1000V/2500V Turn the rotary switch to 500V/1000V/2500V, select

(UT502 ) the requested output voltage, to carry out insulation

resistance measurement.

15
UT501/UT502 OPERATING MANUAL

Measurement Operation
Below section explains how to make measurements.

A. Measuring Voltages

a.UT501 b.UT502
Figure 2. Voltages Measurement
16
UT501/UT502 OPERATING MANUAL

Catuion

To avoid harm to you or damages to the Meter from electric shock, do not

attemptto meaure voltages higher than 1000V DC or 750V rms AC although

readings may be obtained.

When voltage measurement has been completed, disconnect the connection

between the testing leads and the circuit under test and remove testing leads away

from the input terminals of the Meter.

Do not carry out measurement when the battery compartment is open.

To measure voltages, set up the Meter as Figure 4 and do the following:

1. Turn the rotary switch to DCV or ACV.

2. Insert the red test lead into the V terminal and the black test lead into the G terminal.

Then carry out the measurement.


17
UT501/UT502 OPERATING MANUAL

B. Measuring Insulation Resistance

a.UT501 b.UT502

Figure 3. Insulation Resistance Measurement


18
UT501/UT502 OPERATING MANUAL
Caution

When measuring insulation resistance, please must separate the two test leads.

When performing insulation resistance tests, remove all power from the circuit

to be measured and discharge all the power.

Do not short circuit two test leads under high voltage status.

Do not measure insulation resistance after high voltage output.

Do not carry out measurement when the battery compartment is open.

When the measurement is completed, donít touch the circuit as the circuit has

already stored capacitance, which may cause electric shock.

Don't touch the test leads even after it has been removed from the circuit until

voltages are all released.

Do not carry out measurement when the battery compartment is open.

To measure insulation resistance, set up the Meter as Figure 3 and do the following:
19
UT501/UT502 OPERATING MANUAL
Turn the rotary switch to one of these position 250V / 500V / 1000V / 2500V.

1. When performing insulation resistance tests, remove all power from the circuit

to be measured and discharge all the power.

2. Insert the red test lead into the LINE terminal and the black test lead into

EARTH terminal.

3. Connect the red and black alligator clip to the circuit to be measured, positive

voltage output from LINE terminal.

Continuous Measurement

1. Turn the rotary switch to one of these position 250V / 500V / 1000V / 2500V

2. Press TEST to button to carry out continuous measurement. Output insulation

resistance testing voltage, TEST button light up,

3. Press TEST button to close the insulation resistance measurement voltage

when measurement is completed. TEST button lights off.


20
UT501/UT502 OPERATING MANUAL

Maintenance
This section provides basic maintenance information including battery replacement

instruction.

Warning

Do not attempt to repair or service your Meter unless you are qualified to do

so and have the relevant calibration, performance test, and service information.

A. General Service
Periodically wipe the case with a damp cloth and mild detergent. Do not use

abrasives or solvents.

To clean the terminals with cotton bar with detergent, as dirt or moisture in

the terminals can affect readings.


21
UT501/UT502 OPERATING MANUAL

Turn the Meter to OFF when it is not in use.

Take out the battery when it is not using for a long time.

Do not use or store the Meter in a place of humidity, high temperature,

explosive, inflammable and strong magnetic field.

If the Meter is wet, dry it before use.

22
UT501/UT502 OPERATING MANUAL

B. Replacing the Battery

Figure 4. Battery Replacement


23
UT501/UT502 OPERATING MANUAL
Caution

Donít mix to use old and new batteries.

Be careful the polarity is correct when installing batteries.

Danger

Make sure the battery compartment is closed before using the Meter

Follow Figure 4 and proceed as follows to replace the battery:

Turn the Meter to OFF and remove all connections from the terminals.

Remove the screw from the battery compartment, and separate the battery

compartment from the case bottom.

Replace with 6pcs of new 1.5V (LR14) batteries.

Rejoin the case bottom and battery compartment, and reinstall the screw.
24
UT501/UT502 OPERATING MANUAL

Specifications

Safety and Compliances

Certification

IEC 61010 -1 CAT.II 1000V, CAT.III 600V


Compliances overvoltage and double insulation

standard

25
UT501/UT502 OPERATING MANUAL

Physical Specifications

Display (LCD) Digital: 1999 counts

Operating Temperature 0 ~35 (32 ~95 )

Storage Temperature -20 ~60 (-4 ~140 )

Relative Humidity 70% @ 0 ~35 below;

80% @ -20 ~60 :

Battery Type 6pcs of 1.5V (LR14) batteries

Dimensions (H x W x L) 150 x 100 x 71 mm

Weight Approx. 0.5kg (including battery)

26
UT501/UT502 OPERATING MANUAL

General Specifications
Over loading UT501:

Display 1 M on insulation resistance range.

Voltage display > 1999, display 1 V.

Overload voltage testing is not allowed in this Meter.

UT502: Display OL when overload testing on every range.

Battery Indicator Display .

Feature Summary
Display Backlight Bright backlight for clear readings in poorly lighted areas.

Autorange The Meter automatically selects best range

Warning Red light will on when exporting high voltage.

Voltage Auto release voltage


27
UT501/UT502 OPERATING MANUAL

Detailed Accuracy Specifications


Accuracy: ([a% of reading] + [number of least significant digits]), guarantee for 1 year.

Operating temperature: 18 ~28

Relative humidity: 45~75%RH

A. Voltage Measurement

DC Voltage AC Voltage
Measurement Range 30 ~ 1000V 30V~750V (50/60Hz)
Resolution 1V

Accuracy (2%+3)

28
UT501/UT502 OPERATING MANUAL

B. UT501: Insulation Resistance Measurement

Output Voltage 250V 500V 1000V

3M ~2000M

Display Range (When around less than 4.3M , test lead

insulation is bad or invalid, buzzer will beeps)

Open Circuit Voltage DC250V 10% DC500V 10% DC1000V 10%

Test Current 0.9mA~1.1mA @250k 0.9mA~1.1mA @500k 0.9mA~1.1mA @1M

Short Circuit Around less than 1.8 mA

Accuracy 3M to 99M : (3%+5)

100M ~ 2000M : (5%+5)

29
UT501/UT502 OPERATING MANUAL

UT502: Insulation Resistance Measurement

Output Voltage 500V 1000V 2500V

3M ~2G 5M ~4G 25M ~20G

Display Range (less than 3M, buzzer beeps (less than 5M, buzzer beeps (less than 25M, buzzer beeps

and 0M will be shown) and 0M will be shown) and 0M will be shown)

Open Circuit Voltage DC500V 10% DC1000V 10% DC2500V 10%

Test Current 0.9mA~1.1mA @500k 0.9mA~1.1mA @1M 0.9mA~1.1mA @2.5M

Short Circuit Around less than 1.8 mA

3M to 99.9M : (3%+1)

Accuracy 100M ~ 10G : (5%+1)

10G to 20G : (10%+1)

** END **