Manual 409 1000 000 LM Universal III Manual English
Manual 409 1000 000 LM Universal III Manual English
Manual 409 1000 000 LM Universal III Manual English
Level Measurement
For Assistance Call 1-800-527-6297
Outside North America + 215-674-1234
Installation and
Operating Instructions
Series 509-7X
Universal III™ Transmitter
with HART® Protocol
using 409-1000 Electronics
Series 509-7X
Universal III™ Transmitter
with HART® Protocol
using 409-1000 Electronics
TERMS OF PAYMENT: Seller’s standard terms of payment for Buyers who qualify for credit SOFTWARE LICENSE: If goods purchased hereunder include software (“Software”), Buyer
are net thirty (30) days from date of invoice. All invoices must be paid in United States dollars. may use the Software only as part of the goods. Buyer may not use, copy, or transfer any of
the Software except as may be permitted under the applicable License Agreement provided
CREDIT: Seller reserves the right at any time to revoke any credit extended to Buyer or with the goods. Buyer’s right to use, copy or transfer the Software shall terminate upon
otherwise modify terms of payment if Buyer fails to pay for any shipments when due or if in termination of Buyer’s right to use the goods.
Seller’s opinion there is a material adverse change in Buyer’s financial condition. Seller may,
at its option, cancel any accepted Order if Buyer fails to pay any invoices when due. PACKAGING/WEIGHTS AND DIMENSIONS: Buyer specified packing or marking may be
subject to additional charges not otherwise included in the price of the goods. Published
DELIVERY: Shipments are F.O.B place of manufacture (“Shipping Point”) and the Buyer shall weights and dimensions are estimates or approximate only and are not warranted.
pay all freight, transportation, shipping, duties, fees, handling, insurance, storage, demurrage,
or similar charges from Shipping Point. Delivery of goods to common carrier shall constitute FORCE MAJEURE: Seller shall not be responsible for delays in delivery or any failure to
delivery and passing of title to the Buyer, and all risk of loss or damage in transit shall be deliver due to causes beyond Seller’s control, including but not limited to the following items:
borne by Buyer. Any claims or losses for damage or destruction after such delivery shall be acts of God, war, terrorism, mobilization, civil commotion, riots, embargoes, domestic or
the responsibility of Buyer. foreign governmental regulations or orders, governmental priorities, port congestion, acts of
the Buyer, its agents or employees, fires, floods, strikes, lockouts and other labor difficulties,
Seller reserves the right to make delivery in installments which shall be separately invoiced shortages of or inability to obtain shipping space or transportation, inability to secure fuel,
and paid for when due, without regard to subsequent deliveries. Delay in delivery of any supplies or power at current prices or on account of shortages thereof, or due to limitations
installment shall not relieve Buyer of its obligation to accept remaining deliveries. imposed by the extent of availability of Seller’s normal manufacturing facilities.
Acknowledged shipping dates are approximate only and based on prompt receipt of all If a delay excused per the above extends for more than ninety (90) days and the parties have
necessary information from Buyer and Buyer’s compliance with terms of payment. not agreed upon a revised basis for continuing providing the goods or services at the end of
the delay, including adjustment of the price, then Buyer, upon thirty (30) days’ prior written
TAXES: All sales, excise and similar taxes which Seller may be required to pay or collect with notice to Seller may terminate the Order with respect to the unexecuted portion of the goods
respect to the goods and/or services covered by any Order, shall be for the account of the or services, whereupon Buyer shall promptly pay Seller its reasonable termination charges
Buyer except as otherwise provided by law or unless specifically stated otherwise by Seller in upon submission of Seller’s invoices thereof.
writing.
LIMITATION OF LIABILITY: Seller’s liability for any claim of any kind, except infringement of
intellectual property rights, shall not exceed the purchase price of any goods or services
TERMINATION AND HOLD ORDERS: No Order may be terminated by Buyer except upon
which give rise to the claim. SELLER SHALL IN NO EVENT BE LIABLE FOR BUYER’S
written request by Buyer and approval by Seller, and if said request is approved by Seller,
MANUFACTURING COSTS, LOST PROFITS, LOSS OF USE OF THE GOODS OR
under the following conditions: (1) Buyer agrees to accept delivery of all of the units
SERVICES, COST OF CAPITAL, COST OF SUBSTITUTE GOODS, FACILITIES,
completed by Seller through the workday on which Seller receives the written termination
SERVICES OR REPLACEMENT POWER, DOWNTIME COSTS, CLAIMS OF BUYER’S
request; (2) Buyer agrees to pay to Seller all direct costs and expenses applicable to the
CUSTOMERS FOR DAMAGES, OR OTHER SPECIAL, PROXIMATE, INCIDENTAL,
portion of the Order that is incomplete.
INDIRECT, EXEMPLARY OR CONSEQUENTIAL DAMAGES. Any action against Seller
must be brought within eighteen (18) months after the cause of action accrues. These
WARRANTY:
disclaimers and limitations of liability shall apply regardless of the form of action, whether in
A. Hardware: Seller warrants its goods against defects in materials and workmanship under
contract, tort or otherwise, and further shall extend to the benefit of Seller’s vendors,
normal use and service for one (1) year from the date of invoice.
appointed distributors and other authorized resellers as third-party beneficiaries.
B. Software and Firmware: Unless otherwise specified, Seller warrants for a period of one (1)
year from date of invoice that standard software or firmware, when used with Seller specified PROHIBITION FOR HAZARDOUS USE: Goods sold hereunder generally are not intended
hardware, shall perform in accordance with Seller’s published specifications. Seller makes no for application in and shall not be used by Buyer in the construction or operation of a nuclear
representation or warranty, expressed or implied, that the operation of the software or installation or in connection with the use or handling of nuclear material, or for any hazardous
firmware shall be uninterrupted or error-free, or that functions contained therein shall meet or activity or critical application, where failure of a single component could cause substantial
satisfy the Buyer’s intended use or requirements. harm to persons or property, unless the goods have been specifically approved for such a use
C. Services: Seller warrants that services, including engineering and custom application, or application. Seller disclaims all liability for any loss or damage resulting from such
whether provided on a fixed cost or time and material basis, shall be performed in accordance unauthorized use and Buyer shall defend, indemnify and hold harmless the Seller against any
with generally accepted industry practices. such liability, whether as a result of breach of contract, warranty, tort (regardless of the
D. Remedies: Seller’s liability under this section is restricted to replacing, repairing, or issuing degree of fault or negligence), strict liability or otherwise.
credit (at Seller’s option) for any returned goods and only under the following conditions: (1)
Seller must be promptly notified, in writing, as soon as possible after the defects have been EXPORT CONTROL: Buyer shall comply with all export control laws and regulations of the
noted by the Buyer, but not later than (1) year from date of invoice from Seller; (2) The United States, and all sales hereunder are subject to those laws and regulations. Seller shall
defective goods are to be returned to the place of manufacture, shipping charges prepaid by not be named as shipper or exporter of record for any goods sold hereunder unless
the Buyer; (3) Seller’s inspection shall disclose to its satisfaction that the goods were specifically agreed to in writing by Seller. At Seller’s request, Buyer shall furnish Seller with
defective in materials or workmanship at the time of shipment; (4) Any warranty service end-use and end-user information to determine export license applicability. Buyer warrants, in
(consisting of time, travel and expenses related to such services) performed other than at accordance with U.S. Export Law, that goods sold hereunder shall not be destined for
Seller’s factory, shall be at Buyer’s expense. facilities or activities involving nuclear, chemical or biological weapons, or related missile
E.Repaired/Reconditioned Goods: As to out-of-warranty goods which Seller has repaired or delivery systems in named prohibited regions or countries.
reconditioned, Seller warrants for a period of sixty (60) days from date of its invoice only new
components replaced in the most recent repair/reconditioning. GOVERNING LAW: Seller intends to comply with all laws applicable to its performance under
F. Returns and Adjustments: No goods may be returned unless authorized in advance by any order. All matters relating to interpretation and effect of these terms and any authorized
Seller and then only upon such conditions to which Seller may agree. Buyer must obtain an changes, modifications or amendments thereto shall be governed by the laws of the
RMA (Return Material Authorization) number from Seller prior to any return shipment and Commonwealth of Pennsylvania. No government contract regulations or clauses shall apply
such RMA number must appear on the shipping label and packing slip. Buyer shall be to the goods or services, this agreement, or act to bind Seller unless specifically agreed to by
responsible for the returned goods until such time as Seller receives the same at its plant and Seller in writing.
for all charges for packing, inspection, shipping, transportation, or insurance associated with
returned goods. In the event that credit for returned goods is granted, it shall be at the lesser NON-WAIVER BY SELLER: Waiver by Seller of a breach of any of these terms and
of the then current prices or the original purchase price. Claims for shortage or incorrect conditions shall not be construed as a waiver of any other breach.
material must be made within five (5) days after receipt of shipment. SEVERABILITY AND ENTIRE AGREEMENT: If any provision of these terms and conditions
is unenforceable, the remaining terms shall nonetheless continue in full force and effect. This
ALL OTHER WARRANTIES, FOR ANY OF SELLER’S GOODS OR SERVICES, WHETHER writing, together with any other terms and conditions Seller specifically agrees to in writing,
ORAL, WRITTEN, EXPRESS, IMPLIED, STATUTORY OR OTHERWISE, INCLUDING constitutes the entire terms and conditions of sale between Buyer and Seller and supercedes
WITHOUT LIMITATION ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS any and all prior discussions, and negotiations on its subject matter.
FOR PURPOSE ARE EXCLUDED.
Contents
Section 1: Introduction...................................................................................................................... 1
1.1 System Description............................................................................................................ 1
1.2 Technology......................................................................................................................... 1
1.3 Model Numbering.............................................................................................................. 2
1.4 Sensing Element Reference Number................................................................................ 3
1.5 Area Classifications ......................................................................................................... 4
Section 2: Installation....................................................................................................................... 5
2.1 Unpacking.......................................................................................................................... 5
2.2 Mounting the Electronic Unit.............................................................................................. 5
2.3 Wiring the Electronic Unit................................................................................................ 10
2.4 Wiring the Sensing Element............................................................................................ 12
2.5 Spark (Static Electricity) Protection................................................................................. 13
2.6 Surge Voltage (Lightning) Protection............................................................................... 15
2.7 RFI (Radio Frequency Interference) Filters...................................................................... 16
2.8 Electrostatic Filters.......................................................................................................... 18
2.9 Digital Integral Meter........................................................................................................ 19
Section 4: Configuration & Calibration Using the Rosemount™ Model 275 Communicator
with Drexelbrook Device Description.......................................................................... 41
4.1 Drexelbrook Device Description....................................................................................... 41
4.2 Start-up............................................................................................................................ 41
4.3 Configuration................................................................................................................... 42
4.4 Calibration........................................................................................................................ 44
Section 5: Configuration and Calibration Using the Rosemount™ Model 275 Communicator
without a Drexelbrook Device Description.................................................................. 49
5.1 Range / Span Control...................................................................................................... 49
5.2 Rules & Conventions of HART Configuration Software (Model 275)............................... 50
5.3 Tag ID.............................................................................................................................. 50
5.4 Set Up Procedures.......................................................................................................... 51
5.5 Reading Input and Output................................................................................................ 54
5.6 Calibration Using Actual Tank Level................................................................................. 55
5.7 Bench Calibration (if needed).......................................................................................... 56
5.8 Point Calibration............................................................................................................... 58
5.9 Handheld Calibrator Error Messages............................................................................... 59
Section 6: Troubleshooting............................................................................................................. 61
6.1 Identifying a Problem/Symptom....................................................................................... 62
6.2 Troubleshooting Loop Connection................................................................................... 62
6.3 Rosemount Mod. 268 or 275 Calibrator cannot identify or find device........................... 63
6.4 Rosemount Mod. 275 with device description cannot identify or find device................... 63
6.5 Transmitter does not communicate with Drexelbrook PC software.................................. 64
6.6 Troubleshooting Transmitter............................................................................................. 65
6.7 Troubleshooting Sensing Element................................................................................... 67
6.8 Troubleshooting Coaxial Cable........................................................................................ 68
6.9 Static Electricity............................................................................................................... 70
6.10 Radio Frequency Interference.......................................................................................... 70
6.11 Factory Assistance........................................................................................................... 70
6.12 Field Service.................................................................................................................... 71
6.13 Customer Training............................................................................................................ 71
6.14 Return Equipment............................................................................................................ 72
6.15 Universal III Troubleshooting Guide................................................................................. 73
Section 7: Specifications................................................................................................................ 75
7.1 Transmitter Specifications................................................................................................ 75
7.2 Coaxial Cable Specifications........................................................................................... 76
Section 9: Drawings......................................................................................................................... 79
9.1 FM / CSA APPROVAL DRAWINGS................................................................................. 79
9.2 KEMA APPROVAL DRAWINGS...................................................................................... 83
Introduction
Section 1: Introduction
1.1 System Description
The instructions in this manual are for the Drexelbrook
509-7X-XXX Series Universal III™ for level measurement
in liquids, slurries, interfaces and granulars.
Each Drexelbrook 509-7X-XXX system consists of a
k air
Universal III™ (409-1000) series two-wire, 4-20 mA
d d electronic unit and a 700 series sensing element (probe). A
A
k air A
k media
380 series connecting cable is also supplied for connection
C C
of the sensing element to the electronic unit.
k media
The 509-7X-XXX is an admittance-to-current transducer.
kA kA A change in level produces a change in admittance which
C= d C = d results in a change of current. It is termed a two-wire
Figure 1-1 Capacitance transmitter because the same two wires that are used to
Sensing Element power the unit also indicate the change in level (4-20 mA).
1.2 Technology
In a simple capacitance probe, when the level rises and
material covers the sensing element, the capacitance within
the circuit between the probe and the medium (conductive
applications) or the probe and the vessel wall (insulating
applications) increases. This is due to the dielectric constant
(k) of the material, which causes a bridge misbalance. The
signal is demodulated (rectified), amplified and the output
is increased. There are drawbacks, however, especially
when there is coating of the probe.
An RF Admittance level transmitter is the next generation.
Although similar to the capacitance concept, Universal
R III™ employs a radio frequency signal and adds the Cote-
Xc
Shield™ circuitry within the Electronics Unit.
Built-in oscillator buffer and chopper drive circuits permit
separate measurement of resistance and capacitance.
Oscillator circuitry through phase shift Since the resistance and the capacitance of any coating are
cancels small amounts of RF current of equal magnitude (by physical laws), the error generated
by a coating can be measured and subtracted from the total
flow (both Resistive and Capacitive)
caused by coating
output.
This patented Cote-Shield™ circuitry is designed into
High Resistance
Xc (More Coating is
R
the effect of buildup or material coating on the sensing
Tank Wall
element. The sensing element is mounted in the vessel and
provides a change in RF admittance indicating presence of
Coating
material.
The Cote-Shield™ element of the sensor prevents the
transmission of RF current through the coating on the
sensing element. The only path to ground available for the
RF current is through the material being measured.
TFE Insulation
Sensing Element Little to No Resistance
The result is an accurate measurement regardless of the
Figure 1-2 amount of coating on the probe, making it by far the most
RF Admittance Sensing versatile technology, good for very wide range conditions
Element with Cote-Shield from cryogenics to high temperature, from vacuum to
10,000psi pressure, and works with all types of materials.
509-7X Series Universal III™ Transmitter
509 - 007 X - X X X
409 - 10 X 0 - X X X
Introduction
509-7X Series Universal III™ Transmitter
Installation
Section 2: Installation
2.1 Unpacking
Carefully remove the contents of the carton and check each
item against the packing list before destroying any packing
material. If there is any shortage or damage, report it
immediately to the factory.
509-7X Series Universal III™ Transmitter
* *
Figure 2-1 WRONG
Packing Gland CONDULET
Recommended Conduit
Assembly Allows Moisture Infiltration Use only cable CORRECT
Connection
Do Not Disturb! supplied by
CONDUIT
Hold here while AMETEK BREATHER
tightening condulet. Drexelbrook DRAIN
WRONG CORRECT
Figure 2-2
Mounting Dimensions
Inches (mm)
Installation
509-7X Series Universal III™ Transmitter
Packing Gland
Assembly Place Wrench Here
Do Not Turn One When Tightening
Relative to The Condulet
Other
Figure 2-3
Installing Sensing Element
Place Wrench
Here to Install or
Remove From
Vessel
NOZZLE DIAMETER
TOO SMALL
OR
LENGTH TOO LONG
DAMAGE DAMAGE
MAY MAY
OCCUR OCCUR
STILLING HERE HERE
WELL
Figure 2-4 LACKS
Common Installation Mistakes VENT
HOLES
FILL LINE
PROBE AVOID
PROBE FILL
CONTACTS MAY STREAM
SIDE OF STILLING FLEX
INSIDE
WELL PIPE
WALL BUILD-UP
TOUCHES SENSOR
SLUDGE
MAY CLOG
PIPE
SLUDGE
Installation
ROD
STYLE
SENSOR FLEXIBLE
CABLE TYPE
SENSOR
Figure 2-5
Installing Sensing Element AVOID WEIGHTED
CUSTOMER CUSTOMER
SUPPLIED SUPPLIED
SUPPORT SUPPORT
BRACKET
B
Figure 2-6 CLAMP GROUND WIRE
Providing Ground
1/4 inch or larger dia. stainless steel rope
that is anchored or weighted can be used.
Reference* C C
*This is a sensing MOUNT SENSOR IN METAL PIPE
Mounting the level sensor inside a metal
element ground
pipe provides an excellent ground
reference.
E
SUBMERGED METAL STRUCTURE
Use any constantly submerged metal
object such as: pumps, agitators, or
WEIGHT
thermowells. Use only if the process material is
greter than 1000 uMHOS/cm such as acids
and caustic.
F F
SENSOR METAL FLANGE
A submerged metal flange or orifice plate
PUMP can be used.
Use only if the process material is greater
METAL
E thatn 1000 uMHOS/Cm. such as acids and
D caustic.
GROUND
CLAMP
509-7X Series Universal III™ Transmitter
Figure 2-7
Wiring Connections Integral Mounting
10
Installation
Figure 2-8
Universal III Wiring Connections
Remote Mounting
11
509-7X Series Universal III™ Transmitter
12
Installation
13
509-7X Series Universal III™ Transmitter
(ORANGE)
Figure 2-11
Spark Protection for Integral Sensing Elements
Figure 2-12
Spark Protection for Remote Sensing Elements
14
Installation
Figure 2-13
Surge Voltage Protection
15
509-7X Series Universal III™ Transmitter
16
Installation
Figure 2-14
Radio Frequency Interference (RFI) Filters
17
509-7X Series Universal III™ Transmitter
Figure 2-15
Electrostatic Filter
(385-0028-004)
18
Installation
Figure 2-16
Digital Meter in Housing with
Viewport
19
Section 3
Configuration & Calibration
21
509-7X Series Universal III™ Transmitter
HART COMPATIBLE
401-0700-004
Figure 3-1
RS232 Modem Assembly & Loop Connection
22
Configuration & Calibration
HART COMPATIBLE
Notes:
1. Modem will operate from 32º F to
122º F (0º C to 50º C). It can be USB
stored from -40º F to +185º F (-40º C
to +85º C). 0% to 95% relative
humidity - non condensing.
401-0700-007
2. Servic Department 1-800-527-6297
or 1-215-674-1234
Figure 3-1a
USB Modem Assembly & Loop Connection
23
509-7X Series Universal III™ Transmitter
24
Configuration & Calibration
Figure 3-3
PC Software Menu Screen
automatically communicates
all "name plate data" from
transmitter
25
509-7X Series Universal III™ Transmitter
3.7 Configuration
26
Configuration & Calibration
27
509-7X Series Universal III™ Transmitter
3.8 Calibration
There are two methods for calibrating the transmitter
using the PC software:
28
Configuration & Calibration
Jumper Position = 1 2 3 4 5 6
Maximum pF = 20 100 450 2000 10000 45000
SYSTEM # SENSOR # MAXIMUM PROBE LENGTH IN FEET
CONDUCTING LIQUIDS:
509-75-X09 700-1-22 N/A N/A 5.4 20 N/A N/A
509-75-X25 700-5-54 N/A N/A 9.2 40 200 920
509-75-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-75-X13 700-5-18 N/A N/A 1.4 6.25 31 140
509-75-X07 700-2-24 N/A N/A 7.8 20 N/A N/A
509-75-X05 700-2-37 N/A N/A N/A N/A 4.1 14
509-77-X06 700-1-24 N/A N/A 5.4 20 N/A N/A
509-75-X30 700-1-62 N/A N/A 5.4 20 N/A N/A
509-75-724 700-5-29 N/A N/A N/A 4.2 21 95
INTERFACE APPLICATION:
509-76-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-76-X02 700-2-27 N/A N/A 1.0 4.8 20 N/A
509-76-X04 700-2-37 N/A N/A N/A N/A 4.0 14
INSULATING K = 1.5-5:
509-77-X06 700-1-24 N/A 2.7 12.5 20 N/A N/A
509-77-X06 700-1-22 N/A 8.3 20 N/A N/A N/A
509-77-X06 700-5-54 N/A 8.0 37.5 166 N/A N/A
509-77-X06 700-2-57 N/A 3.6 16.2 20 N/A N/A
509-77-X06 700-5-18 N/A 3.8 17.1 76 N/A N/A
509-77-X06 700-2-24 N/A 6.0 20 N/A N/A N/A
509-77-X06 700-2-27 N/A 5.0 20 N/A N/A N/A
509-77-X06 700-2-37 N/A 5.0 14 N/A N/A N/A
Table 3-1
Span Range Position for a number of common Sensing Elements
29
509-7X Series Universal III™ Transmitter
Figure 3-11
A. Go to Level Calibration area of the menu.
Level Calibration area of B. Enter Lower Level value. Press Tab or Enter.
Menu Screen C. Enter Lower Capacitance value. Press Tab or Enter.
D. Enter Upper Level value. Press Tab or Enter.
E. Enter Upper Capacitance value. Press Tab or Enter.
F. Click on Write to Transmitter.
3.8.3 Application Example
Example of an application using the PC software.
(Application Data)
• Vertical Tank
• No Damping
• Caustic or Acid Material in Tank Sensing Element:
700-5-54, Model Code: 74 (See Section 1.4)
• Span Range Switch factory set to 4
• Maximum Capacity of Vessel = 1200 gallons
• Maximum Size of Vessel = 20 feet
• 4 mA (LRV) = 0 gallons
• 20 mA (URV) = 1185 gallons [19.5 feet]
• Point Calibration done using two known level points:
• Lo Cal = 3 feet [selected level]
• Hi Cal = 16 feet [current level]
30
Configuration & Calibration
High Point
16 feet
Figure 3-12 Hi Cal
Application Example Diagram
Low Point
3 feet
Lo Cal
LRV Alternate*
[0 f eet]
LRV 4 mA Point
Figure 3-13
PC Software Menu Screen
View of Application Example
31
509-7X Series Universal III™ Transmitter
32
Configuration & Calibration
33
509-7X Series Universal III™ Transmitter
1. 0
2. 5
3. 10
4. 15
5. 20
6. 25
7. 30
Table 3-2 8. 35
Universal III 9. 40
Strapping Table 10. 45
11. 50
12. 55
13. 60
14. 65
15. 70
16. 75
17. 80
18. 85
19. 90
20. 95
21. 100
Figure 3-15
Menu Screen Transforms to
Strapping Table
34
Configuration & Calibration
35
509-7X Series Universal III™ Transmitter
3.14 Validation
36
Configuration & Calibration
37
509-7X Series Universal III™ Transmitter
AMETEK Drexelbrook
205 Keith Valley Road
Horsham, PA 19044
Telephone: 215-674-1234
FAX: 215-674-2731
Service: 800-527-6297
Strapping Table
Vessel Type: Vertical
Number of Points: 21
Level Volume
In Percent Out Percent
0.00 0.00 o.oo 0.00
1.00 5.00 60.00 5.00
2.00 10.00 120.00 10.00
3.00 15.00 180.00 15.00
4.00 20.00 240.00 20.00
5.00 25.00 300.00 25.00
6.00 30.00 360.00 30.00
7.00 35.00 420.00 35.00
8.00 40.00 480.00 40.00
9.00 45.00 540.00 45.00
10.00 50.00 600.00 50.00
11.00 55.00 660.00 55.00
12.00 60.00 720.00 60.00
13.00 65.00 780.00 65.00
14.00 70.00 840.00 70.00
15.00 75.00 900.00 75.00
16.00 80.00 960.00 80.00
17.00 85.00 1020.00 85.00
18.00 90.00 1080.00 90.00
19.00 95.00 1140.00 95.00
20.00 100.00 1200.00 100.00
Figure 3-20
Typical Printout of Transmitter Data
38
CFg.d (Configure Display) Enter See? Enter
Up/Down Up/Down
Enter Min Value Enter Edit 0 - 99900 Min Value Enter View Min Value
Edit
Up/Down Enter Up/Down Enter
Up/Down Max Value Enter Edit 0 - 99900 Up/Down Max Value Enter View Max Value
Enter
No
Enter
Value =
Enter Edit 3.5 - 4.5
Lrng (Lower Range Value) See? Enter View LRV Enter Enter 4.0?
I.AdJ (DAC Trims)
Yes Yes
Up/Down Up/Down
Enter
Value =
Edit 19.0-21.0
Enter Edit 0 - 99900 Enter 20.0?
Edit Up/Down
Up/Down No
Enter
dELy (Damping Time) See? Enter View Damping Enter
Up/Down
Enter
CAP (Capacitance) Enter View Capacitance
Enter
LOPt (Low Point Cal) See? Enter View Low Point Enter
Up/Down
Up/Down Up/Down
Enter
39
Sc.PO (Scale Percent Out) See? Enter View %Out Scale Enter
Cal? Enter Edit 0 - 99900 Enter
Up/Down Up/Down
Up/Down
Edit Enter Edit 1.0 or 100.0 Enter
Up/Down
Enter
HIPt (High Point Cal) See? Enter View High Point Enter Enter
Up/Down C.CAL (Cap Cal) See? Enter
Up/Down Lower Cap Enter Edit 0 - 99900 Lower Cap Enter View Lower Cap
Enter
tyPE (Type) See? Enter View Type Enter Up/Down Enter Up/Down Enter
Up/Down Up/Down
Up/Down Upper Level Enter Edit 0 - 99900 Up/Down Upper Level Enter View Upper Level
Edit Enter Edit 00 or 30 Enter
Up/Down Enter Up/Down Enter
Up/Down
diAg (Diagnostic Menu) Enter
Upper Cap Enter Edit 0 - 99900 Upper Cap Enter View Upper Cap
HCyl - Flat
Enter
Standard
Escape Menu
Up/Down
Exit Main Menu L.FAC (Load Defaults) Enter Sure?
Enter
Up/Down Device
Restarts
Enter
Configuration & Calibration
Section 4
Configuration & Calibration
4.2 Start-up
After the Universal III transmitter is installed and loop
power is applied, per Section 2, do the following:
41
509-7X Series Universal III™ Transmitter
Optional
Safety
4.23
Optional loop
+ Nominal
Barrier powered indicator
HART® e.g. DLM4000 series 24Vdc Supply
17-30 Vdc
Figure 4-1
Typical Transmitter Loop
Compatible
–
Resistance added if
necessary to make total loop
resistance at least 250 ohms
ABC
7 DEF
8
9
GHI
during the communication
Calibrator or PC Modem may be process.
ABC
7
DEF
8
9
GHI
ABC
7
DEF
8 GHI
9
ABC
7
8
DEF
GHI
9
4.3 Configuration
42
UNIV. III: <TAG>
LEVEL TYPE
UNIV. III: <TAG> AAAAA
LEVEL CONFIG
UNIV. III: <TAG>
Configuration 1. LEVEL TYPE AAA LEVEL UNITS
2. LEVEL UNITS AAA AAAAA
3. MAX LEVEL NNNu <select from list>
UNIV. III: <TAG> 4. LRV NNNu
ONLINE 5. URV NNNu
1. LEVEL NNNu 6. DAMP TIME NNNs UNIV. III: <TAG>
2. VESSEL NNNu 7. RANGE POSITION N MAX LEVEL
3. CAP NNNpF 8. CHG ANLG LOOP ASGN NNNNN
4. REF. CAP. NNNpF XXXXX
5. AO NNNmA
6. % OF RNG NNN%
Required only if weight UNIV. III: <TAG>
or volume readings are LRV
desired. NNNNN
XXXXX
UNIV. III: <TAG>
CONFIGURATION UNIV. III: <TAG>
1. LEVEL CONFIG URV
2. VESSEL CONFIG NNNNN
3. METER CONFIG XXXXX
4. STRAPPING TABLE
5. CALIBRATION UNIV. III: <TAG> UNIV. III: <TAG>
Start VESSEL CONFIG DAMP TIME
NNNNNs
1. VESSEL UNITS AAA XXXXX
UNIV. III: <TAG> UNIV. III: <TAG> 2. VESSEL TYPE AAA
ONLINE UNIV. III: <TAG> BASIC SETUP 3. MAX CAPACITY NNNu UNIV. III: <TAG>
1. DEVICE SETUP 1. TAG 4. DAMP TIME NNNs CURRENT ANALOG
2. LEVEL NNNu 1. PROCESS VARIABLES 2. LEVEL UNITS 5. LRV NNNu LOOP ASSIGN IS
2. CONFIGURATION MENU 6. URV NNNu
4.3 Configuration (Continued)
43
STATUS STATUS DETAILED SETUP XXXXXXXXXX 3. EXIT
RANGE POSITION N 1. UNIV III STATUS
1. HART OUTPUT ENTER METER SPAN VALUE
2. SIGNAL CONDITION NNNNNNNNNN
UNIV. III: <TAG>
3. CALIBRATION XXXXXXXXXX
UNIV. III: <TAG> VESSEL UNITS
UNIV. III STATUS AAAAA
<select from list>
SPAN TOO SMALL AAAAA
DECREMENT SPAN AAAAA UNIV. III: <TAG> UNIV. III: <TAG>
INCREMENT SPAN AAAAA REVIEW DEVICE INFORMATION
1. DISTRIBUTOR UNIV. III: <TAG>
OVERRANGE AAAAA
VESSEL TYPE
UNDERRANGE AAAAA MANUFACTURER 2. MODEL
AAAAA
LEVEL UNITS 3. SERIAL # <select from list>
VESSEL UNITS 4. TAG
VESSEL TYPE 5. DATE
LRU 6. DESCRIPTOR
URV 7. SCRATCHPAD UNIV. III: <TAG>
DAMP TIME 8. DEV ID MAX CAPACITY
LEGEND RANGE POSITION 9. REVISION #'S NNNNNu
ANLG LOOP XXXXX
AAA - ALPHABETICAL DISPLAY POLL ADDR
NNN - NUMERIC DISPLAY TYPE UNIV. III: <TAG>
XXX - FIELD MAY BE EDITTED DISTRIBUTOR HART OUTPUT
u - UNITS MODEL
SERIAL # 1. POLL ADDR Required only if weight
pF - PICOFARADS or volume readings are
mA - MILLIAMPS TAG 2. NUM REQ PREAMS
DATE desired.
s - SECONDS
% - PERCENT DESCRIPTOR
SCRATCH PAD
MUST PERFORM THIS FUNCTION DEV ID UNIV. III: <TAG>
UNIVERSAL REV SIGNAL CONDITION
OPTIONAL FUNCTION FLD DEV REV
SOFTWARE REV 1. CHG ANLG LOOP ASGN
OTHER FUNCTION HARDWARE REV 2. LRV NNNu
3. URV NNNu
4. DAMP TIME NNNs
Configuration & Calibration
5. % OF RNG NNN%
509-7X Series Universal III™ Transmitter
4.4 Calibration
Exit—Calibration is complete.
44
Configuration & Calibration
UNIV.III: <Tag>
ENTER LOW POINT CAL
NNNu
UNIV.III: <Tag> XXXu
POINT CAL
1. LOW POINT CAL
2. HIGH POINT CAL
UNIV.III: <Tag>
ENTER HI POINT CAL
NNNu
UNIV.III: <Tag> XXXu
From configuration CALIBRATION
screen 1. STRAPPING TABLE
2. D/A TRIM
3. POINT CAL
4. CAPACITANCE CAL
5. TRUE LEVEL
6. CHNG ANLG LOOP ASGN
UNIV.III: <Tag>
LOWER LEV
NNNu
XXXu
UNIV.III: <Tag>
LOWER CAP
UNIV.III: <Tag> NNNpF
CAPACITANCE CAL XXXpF
1. LOWER LEV NNNu
2. LOWER CAP NNNpF
3. UPPER LEV NNNu
4. UPPER CAP NNNpF UNIV.III: <Tag>
UPPER LEV
NNNu
XXXu
UNIV.III: <Tag>
UPPER CAP
NNNpF
XXXpF
4.4.2 Capacitance Calibration
Capacitance calibration uses zero and span capacitance
values as the calibration data. These values can be
obtained from the Drexelbrook Service department (or a
previous calibration or identical application). Call 1-800-
527-6297 for assistance. Please provide your DE purchase
order number, transmitter serial number, and application
information to the Service Engineer.
Following is the keystroke sequence for Capacitance
Calibration using the Model 275 Calibrator.
Select Device Setup.
Select Configuration Menu.
Select Calibration.
Select Capacitance Cal—select either Lower Level or
Upper Level depending on whether the next value will be
higher or lower for the second point—edit capacitance value
and level as a pair—return to Capacitance Cal screen.
Exit—Calibration is complete.
45
D/A Trim is NOT a calibration! This is a pre-calibrated alignment to precision factory
settings and is rarely in need of change. The procedure is intended only as a slight "meter"
Strapping Tableadjustment to a known external reference.
4.4.3 D/A Trim
From configuration
screen UNIV.III: <Tag> UNIV.III: <Tag>
UNIV.III: <Tag> POINT NUMBER: X
ENTER POINT POINT NUMBER: X
ENTER INPUT AS A ENTER OUTPUT AS A
NUMBER PERCENT OF MAXIMUM
X % OF MAXIMUM LEVEL:
XXX CAPACITY:
UNIV.III: <Tag UNIV.III: <Tag> X XXX
SELECT EDIT METHOD SELECT EDIT XXX
XXX
!.EDIT ACTUAL POINTS METHOD
2.EDIT POINTS IN % 1. EDIT 1 POINT UNIV.III: <Tag> UNIV.III: <Tag>
UNIV.III: <Tag> UNIV.III: <Tag> POINT NUMBER: X POINT NUMBER: X
3. EXIT
509-7X Series Universal III™ Transmitter
46
POINTS ENTER OUTPUT AS A
1. STRAPPING TABLE % OF MAXIMUM % OF MAXIMUM
2. D/A TRIM 1.CHANGE # POINTS LEVEL: CAPACITY:
3. POINT CAL 2. EDIT POINTS XXX XXX
4. CAPACITANCE CAL 3. REVIEW POINTS XXX XXX
5. TRUE LEVEL 4. SEND POINTS
6. CHNG ANLG LOOP 5. EXIT UNIV.III: <Tag>
ASGN POINT #X:
INPUT: XXX (UNIT)
% INPUT: XXX%
OUTPUT: XXX
(UNITS)
% OUTPUT: XXX%
and Strapping Table configuration.
UNIV.III: <Tag>
UNIV.III: <Tag> POINT NUMBER X HAS
SETTING VESSEL TYPE BEEN MODIFIED. WAIT
TO USER-DEFINED AS POINT IS WRITTEN
TO TRANSMITTER.
D/A Trim
UNIV.III: <Tag>
UNIV.III: <Tag> UNIV.III: <Tag> UNIV.III: <Tag> UNIV.III: <Tag> UNIV.III: <Tag> UNIV.III: <Tag> FLD DEV OUTPUT
CONNECT SETTING FLD DEV ENTER METER FLD DEV OUTPUT 4mA SETTING FLD DEV ENTER METER 20mA EQUAL TO UNIV.III: <Tag>
REFERENCE OUTOUT TO 4mA VALUE EQUAL TO REFERENCE OUTPUT TO 20mA VALUE REFERENCE RETURNING FLD
METER XXXX METER? XXXX METER? DEV TO ORIGINAL
1. YES 2. NO OUTPUT
1. YES 2. NO
Refer to the D/A Trim diagram for the D/A Trim sequence
Configuration & Calibration
• Plan your table by filling out form below. You may use
first column which lists every 5% between 0 and 100%,
or you may fill in your own values in column 2.
• Fill out column 3 with output values corresponding to
those listed in column 1 or 2.
1. 0
2. 5
3. 10
4. 15
5. 20
6. 25
7. 30
8. 35
9. 40
10. 45
11. 50
12. 55
13. 60
14. 65
15. 70
16. 75
17. 80
18. 85
19. 90
20. 95
21. 100
47
Section 5
Configuration & Calibration
Table 5-1
Probe Length vs. Span Position Number / maximum pF
Jumper Position = 1 2 3 4 5 6
Maximum pF = 20 100 450 2000 10000 45000
SYSTEM # SENSOR # MAXIMUM PROBE LENGTH IN FEET
CONDUCTING LIQUIDS:
509-75-X09 700-1-22 N/A N/A 5.4 20 N/A N/A
509-75-X25 700-5-54 N/A N/A 9.2 40 200 920
509-75-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-75-X13 700-5-18 N/A N/A 1.4 6.25 31 140
509-75-X07 700-2-24 N/A N/A 7.8 20 N/A N/A
509-75-X05 700-2-37 N/A N/A N/A N/A 4.1 14
509-77-X06 700-1-24 N/A N/A 5.4 20 N/A N/A
509-75-X30 700-1-62 N/A N/A 5.4 20 N/A N/A
509-75-724 700-5-29 N/A N/A N/A 4.2 21 95
INTERFACE APPLICATION:
509-76-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-76-X02 700-2-27 N/A N/A 1.0 4.8 20 N/A
509-76-X04 700-2-37 N/A N/A N/A N/A 4.0 14
INSULATING K = 1.5-5:
509-77-X06 700-1-24 N/A 2.7 12.5 20 N/A N/A
509-77-X06 700-1-22 N/A 8.3 20 N/A N/A N/A
509-77-X06 700-5-54 N/A 8.0 37.5 166 N/A N/A
509-77-X06 700-2-57 N/A 3.6 16.2 20 N/A N/A
509-77-X06 700-5-18 N/A 3.8 17.1 76 N/A N/A
509-77-X06 700-2-24 N/A 6.0 20 N/A N/A N/A
Figure 5-1 509-77-X06 700-2-27 N/A 5.0 20 N/A N/A N/A
Range Span Jumpers 509-77-X06 700-2-37 N/A 5.0 14 N/A N/A N/A
49
509-7X Series Universal III™ Transmitter
50
Configuration & Calibration
51
509-7X Series Universal III™ Transmitter
52
Configuration & Calibration
53
509-7X Series Universal III™ Transmitter
View PV Input:
or:
or:
54
Configuration & Calibration
55
509-7X Series Universal III™ Transmitter
56
Configuration & Calibration
Company
City State
Customer P.O. Number Item DE# Tag No.
Filled out by: Date Phone Fax
Installation Details
NPT Thread Mount
Flange Mount- if flange mount
B= inches
E= B
A= inches
H= inches E
D= inches
H= inches
IL = inches
A IL
Unless specified otherwise, calibration values of
zero and span capacitances will be based on 4-20 H
Calculations by Drexelbrook
Appx-A-smh.pmd
Page 0 of 1 Created 07/31/1997 by ELS
Revised 08/07/2002 12:21 PM
57
509-7X Series Universal III™ Transmitter
58
Configuration & Calibration
59
Section 6
Troubleshooting
Section 6: Troubleshooting
Problem/Symptom Tests in order of probability Reference Section(s) Comments
Rosemount 268 or 275 Check calibrator connections 6.2 and 6.3 Often a result of loop
Calibrator gives error Check for 250Ω resistance (min.) connection problems
message that no device was Check voltage at transmitter
found Check transmitter 6.6
Rosemount 275 Calibrator Check calibrator connections 6.2 and 6.3 Often a result of loop
gives error message Check for 250Ω resistance (min.) connection problems
that device could not be Check voltage at transmitter
identified Check transmitter 6.6
Can’t communicate Check calibrator connections 6.2 and 6.5 Often a result of loop
with transmitter using Check for 250Ω resistance (min.) connection problems
Drexelbrook PC Software Check voltage at transmitter
Check transmitter 6.6
Try another modem
0 mA output all the time Check voltage at transmitter 6.2 (6.3, 6.4, or 6.5) Probable loop
(no measurable output Check polarity of loop problem
current at any time) Test Transmitter 6.6 Faulty connection in
loop
More than 20 mA output Check for moisture in head of sensor
all the time (output current Test Sensing Element 6.7
always exceeds 20 mA) Test Transmitter 6.6
Check Calibration Section(s) 3, 4, and 5
Output drifts (output Test transmitter without sensing element 6.6
accuracy varies slowly over (drift test)
time…e.g. hours or days) Verify proper sensing element ground Fig. 2-6
reference
Output erratic - (output Test Transmitter 6.6 Erratic readings
jumps around noticeably Check process level often show actual
in terms of seconds or Check for Static Discharge 6.9 process conditions.
minutes) Check for radio interference 6.10 Look for bubbles or
stratification, etc.
Output intermittent (output Check Signal Loop Connections 6.7 Intermittent Loop
jumps quickly usually 6.8 Connection
between >0mA and some
"on scale" value
Inaccurate readings (Level Check calibration Section(s) 3, 4, and 5 Have you verified
readings are incorrect Test transmitter 6.6 actual level? (At
compared to actual known Check method of comparison times even sight
level) gauges can be
misleading.)
Reading does not change Check cables 6.8 Be sure that level
with level Check sensing element 6.7 is really changing.
Test Transmitter 6.6 Possible plugged or
unvented stilling well.
Output goes in opposite Check calibration Section(s) 3, 4, and 5 Probable high point
direction from level change cal/low point cal
reversal or inverted
interface application.
Application-related Comments
Problems
Product Bridging When process material fills what was originally airspace between the sensor and a nozzle
or the vessel, it no longer behaves like a coating. It measures like actual level. Contact
Drexelbrook.
Corrosion of metal parts TFE and stainless steel in the sensor's pressure seal have widely different coefficients
of expansion that sometimes permit pressure leaks to occur. Re-torqueing the packing
assembly can usually fix the problem. Contact the Service department for the proper
torque values and procedure.
Table 6-1 Problem / Symptom Chart
61
509-7X Series Universal III™ Transmitter
62
Troubleshooting
6.3 Rosemount Mod. 268 or 275 Calibrator cannot identify or find device
If the Rosemount 268 or 275 calibrator (generic) gives error
message that no device description was found, use the
following flowchart to troubleshoot.
Check:
1. If current greater than 22 mA, disconnect at sensing
element and re-check. If current returns to normal,
check sensing element using checkout procedure in
Section 5.7. If current does not return to normal,
test transmitter with procedure in Section 5.6.
6.4 Rosemount Mod. 275 with device description cannot identify or find device
This condition may be the result of trying to calibrate a
transmitter with software less than version 3.0 (transmitters
shipped prior to approximately January 1997). There are
two options you can use to identify the device and delete
the error message.
63
509-7X Series Universal III™ Transmitter
Check:
1. If current is greater than 22 mA disconnect at sensing
element and re-check. If current returns to normal,
check sensing element using checkout procedure in
Section 5.7. If current does not return to normal,
Start Is loop test transmitter with procedure in Section 5.6.
Modem does current 2. If loop current is 0 mA, check polarity of loop at
not between 3.7 and No transmitter. If it is OK, check for open loop.
communicate 22 mA? 3. If loop current is between 0 and 3 mA transmitter is
with likely bad. Test with procedure in Section 5.6.
transmitter. 4. Check voltage at transmitter it should be at least 12
volts when 20 mA is flowing in the loop
5. Disconnect the three probe cable connections at
Yes transmitter and retry. If it now communicates, check
cable and probe as described in Sections 5.7 & 5.8.
Is there a
1. Check Power Supply and loop source
minmum of 250
No resistance (the 250 ohms may be built in as
ohms loop
with the Drexelbrook 401-500 series or most
resistance?
DCS inputs).
2. If not there add enough resistance for loop to
be at least 250 ohms.
Yes
Check:
1. Is modem non-isolated
Are you using from ground and/or is your
the Drexelbrook laptop plugged into AC
Are you connected No No
supplied modem? power.? If so you may have
to a DCS?
ground problem. Contact
Service department.
2. Some laptops don't provide
enough voltage to drive
Yes modem correctly from
COM Port. Check with
modem supplier or try a
Yes different laptop.
Is It Polling? No 3. Checkout "Yes" response
tests listed below.
Check:
Yes
1. Is correct COM Port selected at startup?
Make it stop polling or power 2. Is there any software running that would re-direct COM Port such as
transmitter from a separate Windows, mouse drivers, terminal emulation software, or TSR's.
source - then re-check (This can be tested by booting from Drexelbrook Calibration software in
operation. the A: drive)
3. Possible bad RS-232 cable or defective modem.
4. Check modem connections to loop per loop drawing on Figure 3-1.
5. Go to Section 5.6 to test transmitter.
64
Troubleshooting
65
509-7X Series Universal III™ Transmitter
Figure 6-3
Span Range Capacitance Values
66
Troubleshooting
67
509-7X Series Universal III™ Transmitter
Figure 6-5
Sensing Element Testing,
Material Below
the Sensing Element
Testing the Sensing Element - Step 2 (Figure 6-6)
Raise the level in the vessel to cover as much of the sensor
as possible. Repeat the measurement made in step 1.
Readings of 1 meg-ohm or less indicate a pinhole or crack
in the sensing element insulation. Failed insulation is not
field repairable. Consult the Service department for further
assistance.
Figure 6-6
Sensing Element Testing,
Material Covering
the Sensing Element
68
Troubleshooting
OHMMETER
0
8
"0"
CENTER - GROUND OHMS
"0"
CENTER - SHIELD OHMS
SHIELD - GROUND "0" OHMS
SHORT OUT
TWO CONDUCTORS
CHECK FOR CONTINUITY
69
509-7X Series Universal III™ Transmitter
70
Troubleshooting
• FAX: + 215-443-5117,
• E-mail: [email protected]
Please provide the following information:
71
509-7X Series Universal III™ Transmitter
72
Troubleshooting
Provide as much of the following information as possible. All of the information is available from the
Drexelbrook Calibration Software, or from a Rosemount Model 275 with Drexelbrook Device Description
(DD) installed. Information with an asterisk is available from a Rosemount 268 or 275 in the Generic mode.
*LRV (4mA)
*URV (20mA)
73
Section 7
Specifications
Section 7: Specifications
7.1 Transmitter Specifications
Power Requirement
• 12 to 50 VDC
• Minimum of 12 VDC at 20 mA
Input Range
• 409-1000: 1.0 to 45,000 pF
Output Range
• 4-20 mA
Accuracy
• ± 0.25% of range. Accuracy includes the combined effects
of linearity, hysteresis, and repeatability. It refers
to the transmitter only and is measured at reference
conditions of 25 degrees C ±1°, 10-55% R.H. and 24
±1.2 Vdc, using an admittance standard (applied to the
transmitter sensor terminals) in place of the sensor.
Load Resistance
• Maximum Load Resistance = 750 ohms
• Minimum Load Resistance = 250 ohms
Temperature Effect
• ±1% of range per 50°F (30°C).
Supply Voltage Effect
• < 0.1% from 12 to 50 VDC.
Effect of Load Resistance
• < 0.1% for full resistance range at 24 VDC supply.
Response to Step Change
• < 1 second standard (to 90% of final value);
0-90 seconds available with delay.
Fail-Safe
• Low-Level Fail-Safe (LLFS) standard.
Also called direct-acting because current increases as
the level increases.
• High-level Fail-Safe (HLFS).
Also called reverse-acting because current decreases as
level increases.
75
509-7X Series Universal III™ Transmitter
Ambient Temperature
• -40°F to +185°F (-40°C to 85°C)
Calibration Adjustments
• Range Span, 6 positions (side panel)
Intrinsic Safety
• Sensing element and cable: Designed to be intrinsically
safe for Class I Groups A, B, C and D; Class II Groups
E, F, and G, (Class III, Div. 1). Electronics and signal
wires: Intrinsically safe for Class I Groups A, B, C, and
D, Class II Groups E, F and G (Div. 1) when powered
by an intrinsically safe power supply. Nonincendive for
Class I Groups A, B, C, and D;Class II Groups E, F, and
G, Class III, (Div. 2).
76
Normal Maintenance
77
Section 9
Drawings
Section 9: Drawings
79
509-7X Series Universal III™ Transmitter
80
Drawings
81
509-7X Series Universal III™ Transmitter
82
Drawings
83
509-7X Series Universal III™ Transmitter
84
205 Keith Valley Road, Horsham, PA 19044
U.S. and Canada: 1-800-553-9092
24-Hour Service: 1-800-527-6297
International: +1 215-674-1234
An ISO 9001 Certified Company Fax: +1 215-674-2731
E-mail: [email protected]
Website: www.drexelbrook.com