Smartor manualEN
Smartor manualEN
Smartor manualEN
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
I
contents
Preface………. ..................................................................................................................VII
Operation Cautions and Safety Guide ............................................................................IX
Chapter 1 Basic Information ..................................................................................... 1
1.1 Appearance .......................................................................................................... 1-1
1.2 Keys ...................................................................................................................... 1-3
1.3 Power Supply ....................................................................................................... 1-4
1.3.2 Using the Battery....................................................................................... 1-4
1.3.3 Charging the Battery ................................................................................. 1-4
1.3.4 Battery Maintenance and Safety ............................................................... 1-5
1.3.5 Adapter for Power Supply ......................................................................... 1-6
1.3.6 Power Supply Mark ................................................................................... 1-7
1.4 Power On/ Off ....................................................................................................... 1-7
1.4.1 Power On .................................................................................................. 1-7
1.4.2 Power Off .................................................................................................. 1-7
1.5 Status Symbol ...................................................................................................... 1-7
1.6 Display Layout ...................................................................................................... 1-8
1.7 How to Assemble and Use System Straps ........................................................ 1-10
1.8 How to Use Back Strap ...................................................................................... 1-14
Chapter 2 Conventional Ultrasound (UT) ............................................................. 2-1
2.1 Menu Structure ..................................................................................................... 2-1
2.2 Menu Operation.................................................................................................... 2-5
2.2.1 Main Menu Operation................................................................................ 2-5
2.2.2 Submenu Operation .................................................................................. 2-5
2.2.3 Page Operation ......................................................................................... 2-7
2.2.4 Menu adjustment....................................................................................... 2-7
2.2.5 Custom function keys ................................................................................ 2-8
2.3 Functional Description .......................................................................................... 2-8
2.3.1 Detect ........................................................................................................ 2-8
2.3.2 Weld Wizard ............................................................................................ 2-16
2.3.3 Forging Wizard ........................................................................................ 2-16
2.3.4 Plate Wizard ............................................................................................ 2-16
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
II
2.3.5 CrackMeas .............................................................................................. 2-16
2.3.6 FFT .......................................................................................................... 2-16
2.3.7 B-Scan..................................................................................................... 2-17
2.3.8 AWS......................................................................................................... 2-19
2.3.9 CSC (Curved Surface Correction) .......................................................... 2-19
2.3.10 API 5UE ................................................................................................. 2-20
2.3.11 Manage .................................................................................................. 2-20
2.3.12 Setup ..................................................................................................... 2-21
2.4 Fast Adjustment.................................................................................................. 2-22
2.4.1 Gain ....................................................................................................... 2-22
2.4.2 Gate ....................................................................................................... 2-22
2.5 Meas ................................................................................................................... 2-23
2.5.1 Reading ................................................................................................. 2-23
2.5.2 MeasPnt ................................................................................................ 2-23
2.5.3 Alarm ..................................................................................................... 2-24
2.6 WaveSet ............................................................................................................. 2-25
2.6.1 PeakEnv ................................................................................................ 2-25
2.6.2 PeakEcho .............................................................................................. 2-25
2.7 B-Scan ................................................................................................................ 2-25
2.7.1 Scan ...................................................................................................... 2-25
2.7.2 Analyse .................................................................................................. 2-26
2.8 AutoC (Auto Calibration) .................................................................................... 2-26
2.9 AngleC (Angle Calibration) ................................................................................. 2-31
2.10 Curve _DAC ..................................................................................................... 2-34
2.11 CurveCtrl_DAC ................................................................................................. 2-37
2.11.1 TCG ....................................................................................................... 2-37
2.11.2 ParamLock ............................................................................................ 2-38
2.11.3 DACData ............................................................................................... 2-39
2.11.4 Crit. ........................................................................................................ 2-39
2.12 Curve_AVG....................................................................................................... 2-41
2.13 CurveCtrl_AVG ................................................................................................. 2-43
2.13.1 Comp ..................................................................................................... 2-43
2.13.2 Unit dB................................................................................................... 2-43
2.14 Weld/Forging/Plate Wizard .............................................................................. 2-44
2.15 AWS .................................................................................................................. 2-44
2.16 CSC .................................................................................................................. 2-45
2.17 API 5UE ............................................................................................................ 2-45
2.18 WeldSimu ......................................................................................................... 2-46
2.19 CrackMeas ....................................................................................................... 2-48
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
III
2.20 FFT (FFT spectrum analysis) ........................................................................... 2-48
Chapter 3 Thickness Gauge (TG) (Option) ........................................................... 3-1
3.1 Menu Structure ..................................................................................................... 3-1
3.2 Functional Description .......................................................................................... 3-7
3.2.1 StdMode .................................................................................................... 3-7
3.2.2 CoatMode ................................................................................................ 3-14
3.2.3 ECHOMode ............................................................................................. 3-17
3.2.4 VelMeas................................................................................................... 3-18
3.2.5 MultiMode ................................................................................................ 3-20
3.2.6 B-Scan..................................................................................................... 3-23
3.3 Image Information column and Coupling Symbol .............................................. 3-24
3.3.1 Reading ................................................................................................... 3-25
3.4 Description of WorkMode ................................................................................... 3-26
3.4.1 StdMode .................................................................................................. 3-26
3.4.2 CoatMode ................................................................................................ 3-27
3.4.3 ECHOMode ............................................................................................. 3-28
3.4.4 Issues Related to WorkMode .................................................................. 3-29
3.5 VelMeas .............................................................................................................. 3-30
3.6 MultiMode ........................................................................................................... 3-31
3.7 B-Scan ................................................................................................................ 3-32
3.8 Calibration .......................................................................................................... 3-34
3.8.1 PrbDly Calibration ................................................................................... 3-34
3.8.2 Vel+PrbDly Calibration ............................................................................ 3-34
3.8.3 Fast Calibration ....................................................................................... 3-35
3.9 TG Probe ............................................................................................................ 3-35
3.10 V-PATH ............................................................................................................. 3-36
3.11 Auto Search and Auto Gain Control ................................................................. 3-36
3.11.1 AutoSearch (AutoSrch).......................................................................... 3-36
3.11.2 Auto Gain Control (AGC) ....................................................................... 3-37
3.12 TDG .................................................................................................................. 3-38
3.13 Data Operation ................................................................................................. 3-39
3.13.1 Save ...................................................................................................... 3-39
3.13.2 DataRcl.................................................................................................. 3-40
3.13.3 Other ..................................................................................................... 3-40
3.14 Application of Thickness Gauge ....................................................................... 3-40
3.14.1 General Flow Chart ............................................................................... 3-40
3.14.2 Application Notes .................................................................................. 3-41
Chapter 4 Management Operation ........................................................................ 4-1
4.1 Storage Management ........................................................................................... 4-1
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
IV
4.1.1 File Type .................................................................................................... 4-2
4.1.2 Save .......................................................................................................... 4-2
4.1.3 Open .......................................................................................................... 4-2
4.1.4 Preview...................................................................................................... 4-3
4.1.5 Delete ........................................................................................................ 4-3
4.1.6 Rename ..................................................................................................... 4-3
4.1.7 StoreSet .................................................................................................... 4-4
4.2 Encoder Management .......................................................................................... 4-5
4.2.1 Edit ............................................................................................................ 4-7
4.2.2 New ........................................................................................................... 4-7
4.2.3 Delete ........................................................................................................ 4-8
4.2.4 Rename ..................................................................................................... 4-8
4.2.5 Test ............................................................................................................ 4-8
4.3 VPATH Management ............................................................................................ 4-8
4.3.1 Edit .......................................................................................................... 4-10
4.3.2 New ......................................................................................................... 4-10
4.3.3 Delete ...................................................................................................... 4-11
4.3.4 Rename ................................................................................................... 4-11
4.4 Workpiece Management .................................................................................... 4-11
4.5 Probe Management ............................................................................................ 4-12
Chapter 5 System Setup ......................................................................................... 5-1
5.1 System.................................................................................................................. 5-1
5.2 Reading ................................................................................................................ 5-4
5.3 Activate ................................................................................................................. 5-4
Chapter 6 Operator Training and System Maintenance...................................... 6-1
6.1 Operator Training ................................................................................................. 6-1
6.2 System Maintenance ............................................................................................ 6-1
6.2.1 Accessories and options ........................................................................... 6-1
6.2.2 Cleaning .................................................................................................... 6-1
6.3 Service.................................................................................................................. 6-1
Chapter 7 Common Failures and Troubleshooting ............................................. 7-1
7.1 Software Update................................................................................................... 7-1
7.2 Cannot Power on the System .............................................................................. 7-1
7.3 Cannot Power off the System .............................................................................. 7-2
7.4 Cannot Identify the SD Card ................................................................................ 7-2
7.5 No Echo ................................................................................................................ 7-3
7.6 Cannot make DAC curves .................................................................................... 7-3
7.7 Cannot make AVG curves .................................................................................... 7-3
7.8 Incorrect depth reading when using an angle probe ........................................... 7-4
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
V
7.9 Incorrect horizontal reading when using an angle probe ..................................... 7-4
7.10 Cannot Charge the Battery ................................................................................ 7-5
DCY2.781.Smartor/V1.0/E-E
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
VI
Preface
Preface
The Smartor digital ultrasonic flaw detector is designed for ultrasound testing of surface
and internal defects of metal and non-metallic materials in industrial areas, as well as and
material thickness measurement. The system features in conventional ultrasound and
ultrasonic thickness measurement.
b) Big Screen: 5.7" TFT high brightness LCD, with resolution 640×480 pixels, display
c) Easy operation: only a few buttons, well-defined intuitive interface, support right-
e) T/R front end - adjustable pulse width negative square emission, amplitude up to
f) UT functions: weld, plate and forging wizards; AGC (auto gain control), Peak
memory, DAC, AVG and video recording; Optional functions such as B-scan, TCG,
probe spectral analysis, CSC (curved surface correction), weld simulation, crack
compensation available.
VII
Preface
Tips:
a) The system screenshots in this manual are generated by effective software version,
b) Some of the features described in the manual are optional, and your system may not
have these features. If you have any questions, please contact the distributor.
c) To know system features and operation, as well as how to perform ultrasonic testing
with the system, please read this Operation Manual carefully before using the system.
Contact Information:
WELD NDT
OR
Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI)
Address: 77 Jinsha Road, Shantou, Guangdong 515041, China
Tel: 86-754-88250150 Fax: 86-754-88251499
Website: http://www.siui.com
E-mail: [email protected]
VIII
Operation Cautions and Safety Guide
a) Please keep this Operation Manual and always operate the system by
b) Before operating the system, the operator shall have received formal training
on ultrasonic testing.
c) Do not try to dismantle and repair the system unless specifically described in
d) Do not place the system in the place of moisture or near radiator, strong
e) When the system is not in use for a long time, it should be powered up at least
h) For cautions on using the battery, see 1.3.4 Battery Maintenance and Safety.
i) For cautions on using the adapter, see 1.3.5 Adapter Power Supply.
IX
Basic Information
1.1 Appearance
1-1
Basic Information
1-2
Basic Information
1.2 Keys
1-3
Basic Information
The power of this system can be either supplied by the battery or the adapter.
When installing or removing the battery, loosen the screw for the battery compartment at
the system back, as shown in Fig. 1-4. Open the battery compartment, and then you may
install or remove the battery.
If the battery is connected to the system, in power-on state, the upper left corner of the
screen displays the current status of the battery.
【Note 1】: When battery power is very low, to avoid battery over-discharge, the
system is turned off automatically.
【Note 2】: When the battery is connected with the system, even if the system is
powered off, there is also a little current consumption. It is suggested to take out
the battery when the system is not in use for a long time.
Charging the battery should be in a dry and ventilated place under normal temperature.
Operation:
2) Connect the adapter to the power supply. Now the adapter indicator is
1-4
Basic Information
3) Insert the adapter plug into the DC power socket on the system. The
system power indicator indicates the current power / charging status: Red
for charging (power-off state), Off for full charge (power-off state), or Green
【Tip】: Battery recharge can be done either when the system is powered on or off.
If the system is powered on, the recharge time might be longer. Charging the battery
while operating the system for a long time may result in system temperature rise
and battery life shortened. Try to avoid this situation.
a) The capacity and service life of the rechargeable battery depend on whether the battery
is used properly. Please recharge the battery with the configured adapter.
1) When charging the battery through the adapter, use stable and reliable mains
2) The system battery can only be recharged with the configured adapter and the
3) Please recharge the system battery in a cool and ventilated room, otherwise the
4) If the system is not in use for a long time, the battery should be discharged and
1-5
Basic Information
5) During storage or transportation, the battery should maintain about half of its
capacity.
6) When system battery comes to the end of its service life, in order to avoid
environmental pollution, do not discard, break or crush the battery at will. Please
standards.
toxic chemicals.
8) Do not charge the battery near fire source or in extremely hot environment.
short circuit.
10) Keep the battery from moisture and never throw it into water.
11) Do not store the battery at the temperature higher than 60℃ or lower than -20℃.
12) Do not use the battery if it has peculiar smelling, heat, distortion, changed color
The adapter may supply power for the system or recharge the battery. After connecting the
adapter to AC power supply, if the adapter indicator is green with no load (not connected
to the main unit for power supply or when recharged). Connect the adapter output plug to
the system adapter input socket for proper use. The adapter indicator turns red if it is
loaded.
【Note 1】: Do not use the adapter on a loosen-plug AC socket. Be noted that the
total load of adapter plugging into the socket shall not exceed the rated load.
【Note 2】:Do not try to open the adapter if the adapter makes abnormal sound, or
the indicator is in abnormal state. Contact the distributor in time.
1-6
Basic Information
In power up state, the Power Supply mark is displayed at the upper left of the screen. For
the description of each power supply mark, see Table 1-5.
1.4.1 Power On
Operation:
2) Long press Power key to boot up the system. The power/charging indicator
on the system panel is green, and the system enters work mode after self-
test.
Long press Power for about 1 second, and the system will enter the shutdown state and
turn off.
【Note】If any problem occurs during system power down, see Chapter 7.
In addition to the power supply symbol, the status identification information bar at the top
of the screen also includes the following items.
1-7
Basic Information
For the select buttons on the top-level desktop of UT (Ultrasonic Testing) and TG
(Thickness Gauge), See Fig. 1-5.
1-8
Basic Information
1-9
Basic Information
5 Image area and scale: For displaying ultrasound image data and the
corresponding scales. (The x-axis is the A-scan sound path scale, and y-
axis is the A-scan amplitude scale)
1.7.1 How to assemble system straps (example for left- handed operation)
(1) Preparation: Place the straps on the left of the system, see Fig. 1-10. The
targets are two braided straps tied to the stainless-steel cylinders of the system
corners.
(2) Reverse the two straps (relative position unchanged) 180 degrees, and keep the
1-10
Basic Information
1-11
Basic Information
(3) Pass the upper braided strap through the buckle, tighten it; do it in the same
1-12
Basic Information
1.7.2 How to use system straps (example for left- handed operation)
(1) Hold the left of the system with your hand going through the straps. See Fig.
1-14
1-13
Basic Information
(3) Tighten the belt and fasten it backward (Velcro), see Fig. 1-16
1-14
Basic Information
1-15
Conventional Ultrasound (UT)
【Note】: For the menus with the mark * in the menu structure, press Enter key
repeatedly to toggle between coarse / fine adjustment (- / ^).
Main Detect
Menu
1st Basic TR WpPrb WaveSet AutoC
Submenu
Gain Voltage Wp Coor Calib
MeasPnt -- EchoType -- --
Alarm -- -- -- --
Magnify -- -- -- --
MeasLine -- -- -- --
Main Detect
Menu
1st AngleC Curve CurveCtrl
Submenu
DAC AVG DAC AVG
2-1
Conventional Ultrasound (UT)
Main Detect
Menu
Range Start Start Type Unit dB
-- DelPnt -- Num --
-- -- -- Line1 --
-- -- -- Line2 --
-- -- -- Line3 --
-- -- -- Line4 --
-- -- -- Line5 --
-- -- -- Line6 --
-- -- -- -- --
-- -- -- -- --
-- -- -- -- --
Main Detect
Menu
1st
AboutAVG >> WeldSimu BEA -- --
Submenu
AboutAVG<< WeldSimu BEA -- --
Freq -- End -- --
Shape -- LineInd -- --
CrysLA -- -- -- --
2nd
Submenu CrysLB -- -- -- --
BaseType -- -- -- --
BaseAper -- -- -- --
Atten. -- -- -- --
Mode -- -- -- --
NFL -- -- -- --
2-2
Conventional Ultrasound (UT)
Main Detect
Menu
K1K2Corr -- -- -- --
Rename -- -- -- --
Export -- -- -- --
StoreSet -- -- -- --
-- LargeR -- -- --
Main Menu CrackMeas/FFT/ AWS/CSC/API 5UE
-- CrackMeas FFT AWS CSC API 5UE
2-3
Conventional Ultrasound (UT)
-- RefBW -- -- --
-- -- -- Width* Rectify
-- -- -- Thresh Reject
-- -- -- MeasPnt EchoType
-- -- -- Alarm --
-- -- -- Magnify --
-- -- -- MeasLine --
Main Menu B-Scan(Analyse>>)
1st Play/Pause Stop Forward Backward Setup
Submenu
Frame Step
Gain Param>>
2nd Range --
Submenu
DspDly --
AStart --
2-4
Conventional Ultrasound (UT)
AThresh --
AMeasP. --
BSta --
BWidth --
BThresh --
BMeasP. --
Logic --
Start --
2nd
Submenu Width --
Thresh --
MeasLine --
The system main menus are displayed in desktop menus vertically at the center of the
screen.
Operation:
(1) Up, Down, Left, Right for switching the cursor selected main menu
There are two levels of submenus: 1st submenu (horizontal), 2nd submenu (vertical).
When controlling a submenu, the cursor is on the selected 1st submenu, as shown in
Fig.1-8.. Now the key operations are as follows:
(1) Up, Down, Left, Right: turn pages of 1st submenu, and refresh the 2nd
2-5
Conventional Ultrasound (UT)
submenu list.
(3) HOME: Exit the 1st submenu control and return to the desktop main menu
control.
When the 2nd submenu is controlled, the cursor is on the selected 2nd submenu, with the
corresponding 1st submenu marked □, as shown in Erro! Fonte de referência não
encontrada.. Now the keys are operated as follows:
b) Value type: to toggle the adjust state, such as "Range" when and only
when the menu value has a ^ or - flag (for fine /coarse adjust state).
c) Key type: Go to the next level menu list, for example, "Curve" -> "Adjust".
2-6
Conventional Ultrasound (UT)
(4) HOME: Exit the 2nd submenu control and return to the 1st submenu control.
For the main menu that contains multiple pages of submenus, turning pages as follows.
Operation:
Select the "Page" submenu, press Left/ Right or Up/ Down key to turn pages.
Operation:
2) Press Left/ Right to adjust the value of the 2nd submenu or the options for
【Note1】: When adjusting the value-type submenu, hold the adjust key to
increase the adjust speed.
【Note2】: For the submenus with coarse / fine adjustment, press Enter key
repeatedly to toggle between coarse / fine adjustment.
2-7
Conventional Ultrasound (UT)
There are two custom function keys: F1 and F2, whose functions can be selected in the
Setup menu according to user needs.
The selectable functions for F1 and F2 keys in UT mode are: Freeze, PeakEnv,
PeakEcho, PrintScreen, ParamFile, CineRec, WaveFill, WaveComp, SecColor,
FullScreen, StoreSet and AGC; the selectable functions in TG mode are: CalibZero,
SingleSrch, CntinueSrch, Freeze, PrintScreen, ParamFile and StoreSet.
2.3.1 Detect
a) Basic
◆ FineGain: For fine adjustment of gain. The 4dB fine adjustment range has
40 steps. When the gain is fine tuned, the displayed gain value does not
change.
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen
display.
◆ Gate: for setting the adjusted gate, i.e. GateA , GateB or GateC.
2-8
Conventional Ultrasound (UT)
◆ Magnify: For magnifying echo in the gate to the full wave image range.
◆ MeasLine: For turning on / off the indicating line of gate measure result.
b) TR
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen
display.
c) WpPrb
★Tip: "Angle" and "ValK" are associated; ValK =tan (probe angle).
d) WaveSet
◆ Coor: For setting coordinate type of horizontal axis in the image area (A
scan echo).
echo will serve as the background to compare with the dynamic wave.
the secondary wave area will be marked in the image area (A-scan echo).
◆ FullScreen: enable/ disable the function to enlarge the image area to full
screen.
e) AutoC
◆ Calib: Select the calibration options, including Vel, PrbDly and Vel+PrbDly.
2-10
Conventional Ultrasound (UT)
◆ Ref1: The reference position of the 1st record point when calibrating.
◆ Ref2: The reference position of the 2nd record point when calibrating, which
◆ Clear: To clear the calibration information, and restore to the default value
f) AngleC
◆ ReflDep: To set up the depth of the center point for the calibration reference
hole.
◆ Calib: To calibrate the angle (ValK) based on the peak record point.
◆ Clear: To clear the calibration information and set up the default value.
g) Curve (DAC)
2-11
Conventional Ultrasound (UT)
◆ Record: To record the peak record point as the curve record point.
◆ DelPnt: To enable the function of deleting the specified curve record point.
h) CurveCtrl (DAC)
◆ Comp: To adjust the compensation gain of the relative curve. The receiver
PrbDly.
◆ DACData: If enabled, the table displays the curve record point information.
◆ Crit.: To select the criterion to set up DAC dB value, or use a custom DAC.
◆ Grade: The menu is adjustable when a DAC criterion is selected. You may
2-12
Conventional Ultrasound (UT)
◆ Line1: The menu is adjustable when a custom DAC curve is selected, and
◆ Line2: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥2, and you may set up the dB value of Line2.
◆ Line3: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥3, and you may set up the dB value of Line3.
◆ Line4: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥4, and you may set up the dB value of Line4.
◆ Line5: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥5, and you may set up the dB value of Line5.
◆ Line6: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥6, and you may set up the dB value of Line6.
47013 is selected as DAC criterion. You may select the proper block level
selected as DAC criterion. You may select the level for testing the thickness
of the flaw.
i) Curve (AVG)
2-13
Conventional Ultrasound (UT)
◆ Record: To record the peak record point as the curve record point.
j) AboutAVG
◆ CrysLA: To set up the he probe crystal size A. It refers to the valid diameter
PrbShape is Round.
◆ Mode: To calculate the AVG curve from the position of 3 or 0.1 times near
◆ NFL: To set up Near-Field Length (NFL), which is adjustable only for dual
probes.
2-14
Conventional Ultrasound (UT)
k) AVG CurveCtrl
◆ Comp: To adjust the compensation gain of the relative curve. The receiver
echo on the V1 and V2 test blocks. Input the value of the probe AVG curve
directly.
l) WeldSimu
◆ DisE: To set up the distance from probe flank to the weld center.
m) BEA
2-15
Conventional Ultrasound (UT)
Enter the weld test wizard screen, and finish weld test settings step by step.
Enter the forging test wizard screen, and finish forging test settings step by step.
Enter the plate test wizard screen, and finish plate test settings step by step.
2.3.5 CrackMeas
◆ H1: The reflected echo height of the 1st endpoint recorded by the system.
◆ H2: The reflected echo height of the 2nd endpoint recorded by the system.
2.3.6 FFT
◆ DspDly: For setting the ultrasonic pulse shift (display delay) on screen
display.
2-16
Conventional Ultrasound (UT)
2.3.7 B-Scan
a) Ctrl
◆ Save: With the memory connected, to save scan files to the memory, with
◆ Curr: In TimeBase mode, " Curr " refers to the scan time that the probe is
coupled to the workpiece; In Encoder mode, " Curr " refers to the scan
◆ Loss: To calculate the total missed time or distance due to excessive scan
b) Setup
◆ Encoder: To select the encoder model when Encoder mode is selected and
◆ Step: To set up trigger time for scan when TimeBase is selected; To set up
2-17
Conventional Ultrasound (UT)
◆ Dis: The menu is adjustable when Encoder is selected. It is for setting the
c) Basic
d) TR
e) (Analyse) Setup
◆ Step: To set up the frame number for each move of scan line.
f) (Analyse) Meas
◆ Gate: for setting the adjusted gate, i.e. GateA , GateB or GateC.
2-18
Conventional Ultrasound (UT)
2.3.8 AWS
value.
measure value.
measure value.
◆ Start: To set up the start position of GateA, which can only be coarse adjusted
◆ Dr: For setting the reference indicator for calibration (the depth of a groove
◆ K: Correction factor.
2.3.11 Manage
a) Storage
storage.
2-20
Conventional Ultrasound (UT)
b) Wp
c) Probe
d) Encoder
2.3.12 Setup
a) System
To view and set up the system basic information, such as system S/N, time and
date.
b) Reading
2-21
Conventional Ultrasound (UT)
To set up the display content of reading areas such as the Test screen.
c) Activate
2.4.1 Gain
Operation:
(1) Gain To enter the gain fast adjustment state, now the screen bottom displays
(2) Up/ Down To adjust the value of the selected item, where the gain highlighted
2.4.2 Gate
Operation:
(1) Gate To enter the GAStart/Width fast adjustment state, now the screen bottom
2-22
Conventional Ultrasound (UT)
2.5 Meas
2.5.1 Reading
The reading column below the information column shows 5 readings. See Fig. 2-3.
The first reading is for gain, in the format: the upper one is total gain (with the unit in the
bracket) and the lower one is (base) gain + Comp.
In addition, the 4 readings can be selected by the user, where the last reading is a big
reading, which is highlighted in larger font size. See Table 2-2 for the options of reading.
Symbol Meaning
Hx Amplitude percentage of echo measurement point within gate x.
HxdB dB difference between echo measurement point amplitude and gate
height within gate x.
DAC curve state: dB difference between echo measurement point
amplitude in the gate and curve point height of the same position on the
selected DAC curve.
AVG curve state: dB difference between echo measurement point
amplitude in the gate and curve point height of the same position on the
minimal equivalent AVG curve.
Sx Sound path of echo measurement point within gate x.
Dx Depth of echo measurement point within gate x.
Rx Horizontal distance from the probe (wedge) flank to the echo
measurement point within gate x.
A Ind (Auxiliary AWS computing) the gain value of the amplitude when
recording B Ref
C Att (Auxiliary AWS computing) the attenuation value of flaw echo
D Rating (Auxiliary AWS computing) the rating value of flaw echo
Sab The sound path distance between measurement points in gate A and B
ERS The FBH equivalent of measuring point echo in gate A in AVG curve
state.
2.5.2 MeasPnt
There are 3 methods for gate MeasPoint: Peak, Flank and JFlank, which are for sound
path and amplitude measurement for different modes. See Fig. 2-4.
2-23
Conventional Ultrasound (UT)
a) Peak: To calculate sound path (S3) and amplitude (H2) based on the echo peak
b) Flank: To calculate sound path (S1) based on the first intersection between the
echo and the gate, and to calculate amplitude (H2) based on the peak in the gate
range.
c) JFlank: To calculate sound path (S1) according to the first intersection between
the echo and the gate, and based on which calculate the amplitude (H1) by
searching for the echo peak where the intersection locates in the gate range.
2.5.3 Alarm
Alarm mode includes: off, Pos (the echo amplitude within the gate is higher than gate
thresh) and Neg (the echo amplitude within the gate is lower than gate thresh).
When giving alarm, the alarm symbol at the top of the screen is red and flashes, otherwise
it is hollow and unchanged.
When the system "Buzzer" is on, it supports sound alarm, and the Buzzer sounds when
giving alarm.
2-24
Conventional Ultrasound (UT)
2.6 WaveSet
2.6.1 PeakEnv
If enabled, the echo amplitude peak values for all points in sound path direction are
recorded and displayed. When higher echo for one point appears, the recorded
value for that position will be refreshed, otherwise it will remain unchanged.
Description: It is mainly used in the situations of angle probe testing positioning
and displaying echo envelope.
2.6.2 PeakEcho
If enabled, the screen will record the peak echoes for all echoes (except for initial
wave) within the display range, and display these echoes in different colors. When
higher echoes appear, all echoes recorded will be refreshed, otherwise they remain
unchanged.
2.7 B-Scan
B-Scan is to display A-scan echo in imaging mode, so as to achieve more intuitive test
result for easy observation and analysis.
2.7.1 Scan
Operation:
(1) Set up the system TR parameters according to the probe connected to the
system.
(2) Adjust the scan mode and parameters in Setup according to the scan.
(4) During the scanning process, the "Curr" menu is updated to the time /position of
the scan.
(5) When the scan is finished, enable "Save" function (which is disabled before
(6) Press Right or Enter to go to “Analyse”, and to browse the scan file just finished.
【Note 】
2-25
Conventional Ultrasound (UT)
(1) Due to the limited storage space, B-Scan supports only 1000 frames of
(2) The system will automatically compress/ stretch the scan image to full
2.7.2 Analyse
Entering Analyse mode and the scan file will be played automatically. Now:
(1) The scan lines (marked in red on the B-Scan image) scroll to show the frame
position.
(2) The waveform image area shows the waveform data information for the frames.
(3) The 2nd menu area displays the information such as "Gain", "Range", "Delay"
In the 1st menu area, there are button menus of "Play / Pause", "Stop" "Forward",
"Backward" and "Finish". Select one and press Enter to perform the operation.
AutoC is for calibrating material velocity and probe delay (probe zero, the acoustic time
from the probe crystal to the incident point) for the system, so as to eliminate the impact
of Vel and PrbDly on the measurement values of sound path, horizontal distance and
vertical distance.
The user may select either material velocity calibration or probe zero calibration, or both
at the same time.
b) DelayC: Calculate probe delay by using the difference between the echo
propagation time of one reflector with known sound path, and the echo time
2-26
Conventional Ultrasound (UT)
Example condition:
(1) The calibration block is as shown in Fig. 2-5, the sound path of the first reflector
Operation:
2-27
Conventional Ultrasound (UT)
2) In “WpPrb”, set up probe “Type” as Angle, “Angle “ is 60.0, now the system will
3) In “AutoC”, set up “Calib” to Vel+PrbDly, and set up Ref1 and Ref2 sound path
of the calibration block according to test block parameters setup (set to 50mm
4) Place the probe on the calibration block, as shown in Fig. 2-5, and get the echo
When the "Ref1" value is adjusted, the gate center position is automatically adjusted to its
set value. If the parameter setup and probe placement are correct, the gate may record
the desired echo. If the gate position is different from the echo position, adjust "Start" menu,
or press Gate key to quickly adjust (see 2.3.2) and enable recording the echo.
Move the probe to get peak position of the reference point, press Right or Enter in “Record”
menu to finish recording Ref1.
2-28
Conventional Ultrasound (UT)
【Note】The red circle on the echo indicates the echo peak position, which can be
searched again through "Reset".
5) Move the probe to get Ref2 echo, as shown in Fig. 2-9. Refer to step 4) for the
operation, and the system gets Ref2 peak position, press Right or Enter in
2-29
Conventional Ultrasound (UT)
6) The system updates the values of "Vel" and "PrbDly", and displays them in the
upper right corner of the waveform area. √ Indicates calibration for the item is
finished.
7) At the screen bottom, it prompts to input the actual probe flank value according
2-30
Conventional Ultrasound (UT)
AngleC is for measuring the angle of an angle probe with a reflector (e.g. side drilled
hole) with known size and depth.
【Note】
(2) When performing angle measurement, the sound path between the probe
and the reflection hole shall be greater than 2 times of NFL, so as to avoid
Example condition:
(1) The calibration block is as shown in Fig. 2-11, ReflDep 30mm, ReflApe 50mm.
2-31
Conventional Ultrasound (UT)
Operation:
3) Place the probe on the test block, as shown in Fig. 2-11. The system gets the
When adjusting "ReflDep" value, the gate center position is automatically adjusted to its
set value. If the parameter setting and the probe are placed properly, the gate may record
the desired echo. If the gate position is different from the echo position, adjust "Start" menu,
or use Gate key to quickly adjust (see 2.3.2), enabling the gate to record the echo.
Move the probe to get the peak of the reference point, press Right or Enter in
“Calib” menu to finish recording.
2-32
Conventional Ultrasound (UT)
4) The system updates the values of "Angle" and "ValK ", and displays them in the
upper right corner of the waveform area as shown in Fig. 2-13 Finish angle
2-33
Conventional Ultrasound (UT)
The Distance Amplitude Correction (DAC) curve is used to indicate the amplitude variation
of the reflector signal with the same size but different distance from the probe. The purpose
of the DAC curve is to compensate material attenuation, near field effect, beam
propagation and surface roughness in image format.
DAC Curve is for guiding the user to make DAC curves based on the detection
requirement.
Operation:
2) Adjust “Start” menu, or press Gate key to adjust quickly (see 2.3.2), enabling
3) Adjust “Gain” menu, or press Gain key to adjust quickly (see 2.3.1), to make
2-34
Conventional Ultrasound (UT)
4) Press Right or Enter key in “Record” menu (or press Enter key in “Start” menu
to record quickly), to finish the setup of one DAC curve record point, “Record”
5) Repeat steps 2)~4) to continuously increase DAC curve record points till the
In this process, if any wrong record point is selected, enable “DelPnt” to delete the
specified record point. See the operation steps of DelPnt below. You can also enable
"Clear" to delete the entire DAC curve.
If the record point position is correct, but the amplitude is deviated, you can enable
"AdjPnt" function to adjust the recorded record amplitude. See the AdjPnt operation
below.
DelPnt Operation:
(2) The operating tips are shown at the screen. See Fig. 2-15.
2-35
Conventional Ultrasound (UT)
Press Enter key to remove the selected point from the DAC curve.
(4) After the deletion is finished, press HOME key to disable "DelPnt" function.
AdjPnt Operation:
(2) The operating tips are shown at the screen. See Fig. 2-16.
2-36
Conventional Ultrasound (UT)
Press Up/ Down key to adjust the amplitude of the selected point.
(4) After the adjustment is finished, press HOME key to disable "AdjPnt" function.
2.11 CurveCtrl_DAC
After making the baseline of DAC curves, you may adjust the curves according to the
standard.
2.11.1 TCG
After making the curves, the default is DAC. You may adjust "Kind" menu and switch to
TCG.
Time-Changed Gain (TCG) is to compensate the same factor as DAC, but with different
representation. DAC is drawn with reference to the reflector peak position when beams
attenuate, and TCG is to set up a function based on time (distance) amplified gain, enabling
the reference reflector peak reaches the same screen height (the amplitude of the first DAC
2-37
Conventional Ultrasound (UT)
point), where the curve is displayed as a straight line through the screen, as shown in Fig.
2-17.
2.11.2 ParamLock
When the DAC Curve is finished, "ParamLock" function is turned on by default. In this case,
the adjustment of the parameter menu (see note (1)) which may cause the reference echo
state to change is not allowed.
However, taking into account of probe wear or other situations, and the inspectors may
need to fine tune the restricted parameters, you can turn off the "ParamLock" function and
open the adjustment of parameter menu, so as to avoid remaking the curves.
【Note】
2) To turn off the "ParamLock" function, the inspector should understand the
2-38
Conventional Ultrasound (UT)
2.11.3 DACData
When the "DACData" function is turned on, the gain value and the attenuation value of
the recorded points in the DAC curves (BaseL, Line1, Line2 and Line3) at 80% amplitude
are displayed in a tabular form. See Fig. 2-18.
【Note】
(2) When the DAC criterion is applied, the table will display the selected
criterion and the reflection hole size of the standard test block for the DAC
2.11.4 Crit.
(1) Custom: The built-in criterion is not used. It is up to the user to adjust the number
(2) GB/T 11345: In the standard, the DAC curve requirements the same as GB/T
2-39
Conventional Ultrasound (UT)
29712.
Grade2
Thickness T
8≤T<15 15≤T<100
(mm)
Flaw Length
L≤T L>T L≤0.5T 0.5T<L≤T L>T
L
Line1 (dB) -4 -10 0 -6 -10
Line2 (dB) -8 -14 -4 -10 -14
Line3 (dB) -14 -14 -14 -14 -14
Grade3
Workpiece
Thickness T 8≤T<15 15≤T<100
(mm)
Flaw Length
L≤T L>T L≤0.5T 0.5T<L≤T L>T
L
Line1 (dB) 0 -6 4 -2 -6
Line2 (dB) -4 -10 0 -6 -10
Line3 (dB) -10 -10 -10 -10 -10
Adjust the "ThkBlock" menu value based on the tested thickness.
When the "FlawBlock" menu value is off, this time, the most stringent curve criterion for
the grade and ThkBlock is created (ie, the curve when L> T). When the FlawBlock" menu
value is not off, the curve is created according to the “Grade”, "ThkBlock" and
“FlawBlock”.
2-40
Conventional Ultrasound (UT)
2.12 Curve_AVG
In the actual ultrasonic testing, due to different shapes and nature of defects, it is difficult
to determine the real size of the defect. Therefore using "equivalent method" may quantify
the defects.
The equivalent method is a method of determining the size of the defect, as compared with
the artificial defect of regular shapes, that is, when the perceived defect wave height is
equal to the artificial defect wave height under the same testing condition, the size of the
artificial defect is called the equivalent size of the defect.
The artificial defect usually used to represent the equivalent size is FBH (flat bottom hole).
For the regular reflector (including artificial defects), under certain assumptions, the
relationship between the echo sound path, the amplitude (dB) and the defect equivalent
can be expressed in a regular way that is called the AVG (DGS) curve, which is typically
used for the area equivalent calculation of the defect for forging.
Where A represents the distance in the unit of NFL; V represents the amplitude in the unit
of dB, with the initial wave height 0 dB, and G is the defect equivalent, in the unit of probe
size.
AVG Curve is to guide the user to make AVG curves according to the test requirements.
Example condition:
(1) The test block is a 100-Φ2 forging block, i.e. Thickness 100mm, FBH Φ2.
(2) The probe is a normal (straight-beam) probe with frequency 2.5MHz, diameter
20.0mm.
Operation:
(1) To ensure the accuracy of curve making, auto calibration is required before
2-41
Conventional Ultrasound (UT)
(3) Press Right key to enter the “AboutAVG” menu list, and set AboutAVG
parameters according to the probe and test block. In this example, "Type" is
menu, or Gate key to adjust quickly (see 2.5.2), enabling the gate to record the
echo. Move the probe to get the peak of FBH, then press Right or Enter key in
2-42
Conventional Ultrasound (UT)
2.13 CurveCtrl_AVG
2.13.1 Comp
If the echo amplitude of the test block in use or the workpiece does not comply with the
theoretical value, Adjust “Comp” menu to compensate the difference, which is commonly
used in case that BW echo serves as the base.
Since BW echo amplitude usually cannot reach the theoretical difference with the
FBH echo amplitude of the same distance, taking the BW amplitude as the base
will result in incorrect FBH equivalent calculated in practical testing. Therefore, in
the situation that BW echo is used as the base, to make the calculation more
accurate, it is suggested to use the error between the actual difference [of base
BW echo amplitude and the echo amplitude of FBH test block of the same distance
(different distance theory)] and the theoretical difference as a compensation
value input.
2.13.2 Unit dB
When the "Unit dB" function is turned off, the AVG curve position is set according
to the equivalent value (mm/ inch); when the "Unit dB" function is turned on, the
base line is 0dB. Set up ±dB value of AVG curve relative to the baseline.
2-43
Conventional Ultrasound (UT)
The personal setup wizard will help users configure the professional operation
process, enabling more professional testing. Each wizard includes 2 modes: Basic
and Profession. Basic mode is for primary users, with experience parameter
presets based on field testing; Profession mode is for users of certain test
experience, with rich parameter settings available.
2.15 AWS
Welding grade is calculated as per AWS (American Welding Society) steel structural
welding code (D1.1) and bridge welding code (D1.5). Before operation, select the
appropriate probe and standard test block according to AWS D1.1 / D1.5.
Operation:
(1) Select "AWS" from the desktop and press Enter to enter the function.
(2) Place the probe on the test point of the standard test block, and ensure good
coupling.
Move the probe back and forth to get the highest echo to make the echo amplitude to about
80%. Adjust “Start” menu, or Gate key to adjust quickly (see 2.5.2), enabling the gate to
record the echo.
(3) Press Right or Enter key in “Record” menu, and “B_Ref” menu (B Reference
(4) Place the probe on the tested work piece and ensure good coupling. The
same as step (2), obtain the highest echo of the flow in the workpiece.
The system will automatically refresh “AInd.”, “CAtt.” and “DRating” measure values.
“AInd.”, A Indication Level, which indicates the gain value of the amplitude when the current
echo amplitude reaches the recorded B Reference Level.
【Note】
(1) All calculations are based on AWS D1.1-2006 and AWS D1.5-2002.
2-44
Conventional Ultrasound (UT)
(2) When recording B Reference Level in step 3), the values of A, C, D are reset
to 0.
2.16 CSC
CSC (Curved Surface Correction) is for depth and horizontal distance correction when
testing circumference with an angle probe.
Operation:
(1) To ensure the accuracy of the curve, Vel and auto calibration are required
(2) Select CSC from the desktop and enter the function.
(3) According to the tested workpiece, input "O.D." and "Thickness" values.
(4) Depending on the probe in use, set up the probe parameter menu values.
(5) If the inner surface of the workpiece is tested, set “Correct “ to “I.D.”; If the
outer surface of the workpiece is tested, set “Correct “ to “O.D.” Now Depth
and Hor measure values will be corrected according to the surface settings
before display.
A quantitative method for flaw depth according to API (American Petroleum Institute).
Operation:
(1) In the desktop, select “API 5UE”to enter the function.
(2) “Dr” menu is set up as a reference indicator for calibration, i.e. the depth of
(3) Place the probe on the standard test block and ensure good coupling. Move
the probe to bring on screen a dynamic echo envelope required by API 5UE.
2-45
Conventional Ultrasound (UT)
(5) Enter “K” menu, and the system will calculate and refresh “K” value.
(6) Place the probe on the tested work piece and ensure good coupling. Move
the probe to bring on screen a dynamic echo envelope required by API 5UE.
(8) Enter “Di” menu, and the system will calculate and refresh “Di” value.
2.18 WeldSimu
WeldSimu is to simulate the weld shape and use the dynamic beam tracking function to
assist the user to quickly determine the location of flaws in the weld.
2-46
Conventional Ultrasound (UT)
Operation:
(1) Set up "WpPrb" parameter based on the tested workpiece and the
connected probe.
(3) Turn on "WeldSimu" function, and the waveform image area is displayed
the waveform with high amplitude. Now do not move the probe, and use a
ruler to measure the distance between the probe flank and the weld center,
(5) Finish settings for "WeldSimu", and enter "Basic" for testing.
2-47
Conventional Ultrasound (UT)
Adjust “Gate”to record the echo. Now the end of the beams indicates the flaw in
the workpiece.
Adjust "Gate", trap echo, then the end of the line that defects in the workpiece
position.
【Note】The color of beams is the same as the waveform, where the section
with color coincides with the gate is the gate test range.
2.19 CrackMeas
To measure crack height with endpoint peak echo method of angle probe.
When incident waves enter the crack tip, some of them are reflected along their original
path, which is called endpoint reflected echoes. The endpoint reflected echo method is to
calculate crack height by measuring the sound path of endpoint peak echo produced when
the main beam is incident on the crack endpoint.
Operation:
8) Now the system will automatically calculate the recorded depth difference
between “H1” and “H2” items, with the result displayed in “Result” menu.
FFT is to obtain the spectrum based on received signals, and further acquire the
measure values related to frequency by measuring the spectrum.
Operation:
(1) In “Detect”, adjust “Gate” to record the analysis echo.
2-48
Conventional Ultrasound (UT)
(3) Now "Range" and "Delay" are automatically magnified according to the gate,
(5) Turn on “MeasFrq” function, and you may select “-6dB” “-12dB” or “-20dB”.
Now the measure values of “Peak (Freq)”, “C.Freq (Freq)”, “L.Edge (Freq)”,
“H.Edge (Freq)” and “RefBW” are refreshed real time. See the figure.
2-49
Thickness Gauge (TG)
Note: The Thickness Gauge (TG) feature is optional. Please contact the supplier for
purchase if necessary.
【Note】: For the menus with the mark * in the menu structure, press Enter key
repeatedly to toggle between coarse / fine adjustment (- / ^).
Main StdMode
Menu
1st Probe Gate Meas Data Calib
Submenu
PrbSel GAStart* Rectify Format Calib
-- -- -- ShowWave Vel*
Main StdMode
Menu
1st Display TDG TempComp -- --
Submenu
Mode TDG TempComp -- --
3-1
Thickness Gauge (TG)
Main StdMode
Menu
LostCoupT Slope* CompVal* -- --
DiffRef* -- -- -- --
AvgNum -- -- --
MeasFreq -- -- --
Freeze -- -- --
Main CoatMode
Menu
1st Probe Gate Meas Data Calib
Submenu
2nd
Submenu CalibZero GBThresh Step Delete Range*
3-2
Thickness Gauge (TG)
-- LowLimit* -- ShowWave --
-- UpLimit* -- -- --
-- Buzzer -- -- --
AvgNum -- -- -- --
MeasFreq -- -- -- --
Freeze -- -- -- --
Main ECHOMode
Menu
3-3
Thickness Gauge (TG)
-- LowAlarm -- Freeze --
-- UpAlarm -- ShowWave --
-- LowLimit -- -- --
-- UpLimit -- -- --
-- Buzzer -- -- --
Main ECHOMode
Menu
AvgNum -- -- --
MeasFreq -- -- --
Freeze -- -- --
Main VelMeas
Menu
1st Probe Gate Meas Data Alarm
Submenu
WorkMode GAStart* Rectify Format LowAlarm
-- -- -- ApplyParam --
-- -- -- Freeze --
3-4
Thickness Gauge (TG)
-- -- -- ShowWave --
-- -- -- -- --
Main VelMeas
Menu
1st Calib Display TDG TempComp --
Submenu
Ref1* Mode TDG TempComp --
PrbDly* MeasFrq -- -- --
Vel* Freeze -- -- --
-- -- -- -- --
-- -- -- -- --
Main MultiMode
Menu
1st Probe MeasParam LMeas Data Alarm
Submenu
PrbSel LayerNum CurArea Format LowAlarm
-- -- DspDiy* Clear --
2nd
Submenu -- -- Rectify Save --
-- -- AutoSrch ApplyParam --
-- -- MaxSGain* Freeze --
-- -- AGC ShowWave --
3-5
Thickness Gauge (TG)
Main MultiMode
Menu
1st Probe MeasParam LMeas Data Alarm
Submenu
-- -- Gain -- --
-- -- Step -- --
-- -- MeasPnt -- --
Main MultiMode
Menu
1st Calib Display TDG TempComp --
Submenu
Ref1* Mode TDG TempComp --
PrbDly* Freeze -- -- --
Vel* -- -- -- --
-- -- -- -- --
-- -- -- -- --
Total -- -- Gate --
-- -- -- Logic --
-- -- -- Start* --
3-6
Thickness Gauge (TG)
-- -- -- Thresh --
-- -- -- MeasPnt --
-- -- -- Alarm --
-- -- -- Magnify --
-- -- -- MeasLine --
-- -- -- -- --
Gain
Range* --
Delay* --
GAStart* --
GAWidth* --
2nd
Submenu GAThresh --
GBStart* --
GBWidth* --
GBThresh --
--
--
2nd --
Submenu
--
3.2.1 StdMode
a) Probe
3-7
Thickness Gauge (TG)
◆ VpathSel: Select Dual, and then select a V-PATH curve from the curve option
curve suffix is "* _echo". For Single probe, the VpathSel is fixed to "None”.
◆ CalibZero: A hot key button. “Ref1” and “Vel” in calibration function are used
for probe delay calibration. That is to perform fast calibration of probe delay
according to the valid echo signals in GateA based on the known material
PrbDly is refreshed.
b) Gate
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen
display.
3-8
Thickness Gauge (TG)
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 3】If the Alarm function is not turned on or does not meet the alarm
requirement, the alarm symbol on the right side of system information
column is a hollow circle.
c) Meas
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC is
◆ AGC: to enable/ disable AGC function. This function can not be enabled with
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ MeasPnt: To set up the measure point value position of the GateA, which
supports two modes: JFlank (the first intersection between the echo and the
gate in gate range) and Peak (the intersection between the peak wave
3-9
Thickness Gauge (TG)
d) Data
(press Right to enter the detailed attribute setup screen, press Left to return
increment/decrement.
◆ CurRow: To set up the row location of data Record, Delete, DataRcl for the
◆ CurCol: To set up the column location of data Record, Delete, DataRcl for
◆ CurPnt: To set up the point location of data Record, Delete, DataRcl for the
◆ To save the current MeasPnt to Data CurRow, CurCol, CurPnt location. The
◆ Delete: Delete the MeasPnt data at the CurRow, CurCol, CurPnt position.
◆ Save: In Storage menu of Manage, select the file type as "Data", and you
may set up the "Storage mode". If it is "Auto", click Save. If the file matches
the save condition, it will be named and saved to the SD card based on the
current system time year / month / day / hour / minute / second) automatically;
If it is "Manual", click Save, and it will pop up the name edit screen for the
◆ DataRcl: To recall the saved MeasPnt data from the current data (specific
3-10
Thickness Gauge (TG)
location by CurRow/ CurCol/ CurPnt), and display the saved measure value
apply the saved MeasPnt parameter to the curren use after unfrozen.
Off: Apply the saved point parameter state to the current use after unfrozen.
On: After unfrozen, return to the parameter state before recall operation (that
means the DataRcl operation is only for preview).
◆ Freeze: The Freeze operation on-screen.
【Note】 The maximum points (all rows, columns and points of each Data) that can
e) Calib
◆ Calib: Select the items for calibration, which includes Vel, PrbDly and
Vel+PrbDly.
◆ Ref1: The reference position of the 1st record point when calibrating.
◆ Ref2: The reference position of the 2nd record point when calibrating, which
situation.
◆ Clear: To clear the calibration information, and restore to the default value
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
3-11
Thickness Gauge (TG)
f) Display
◆ Mode: For setting the display layout and combination on the screen, with
A/BVa, A/Ba/ BVa, Ba/ BVa as options. "A" refers to the A-scan display;
“BVa” refers to the reading display screen at the central area; “Ba” refers to
the data.
options available: Std, MinVal, MaxVal, Avg, Diff (See 3.12 for the description
on reading items).
◆ LstRead: To set up whether to keep the last reading or display blank when
of the latest reading cache display on the right of MainVal, which is invalid in
3-12
Thickness Gauge (TG)
calculation restarted if the probe lost coupling time (LostCoupT) reaches the
setup value. For example, when the measurement mode is "MinVal” and the
last valid measure value is 50mm, if the measured value is 100mm this time,
the displayed measure value should still be 50mm. But if the time interval
between the two measurements exceeds the time of "LostCoupT", then the
◆ DiffRef: For Diff measurement, the thickness value is the measured value
◆ AvgNum: For Avg measurement, the thickness value is the average value of
g) TDG
◆ TDG: For turning on/off TDG function. TDG-Time Distance Gain curve. Gain
When TDG is turned on, the screen displays a TDG curve, which
corresponds to the actual signal position of gain compensation in the time
axis direction; while in the amplitude direction, the start gain and the target
gain are compressed in the screen for display, which does not represent the
actual signal amplitude.
◆ StGain: For setting the start dB value of TDG curve.
≤ 110dB.
◆ StTime: The valid start time of TDG curve (the sound time counting from zero)
h) TempComp
calibration test block is different from that of the test part, you may turn on
1.0~1.0.
3.2.2 CoatMode
a) Probe
b) Gate
3-14
Thickness Gauge (TG)
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen
display.
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 3】If the Alarm function is not turned on or does not meet the alarm
requirement, the alarm symbol on the right side of system information
column is a hollow circle.
c) Meas
◆ Rectify: To set up rectification mode of receiving echo. There are four Rectify
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC is
◆ AGC: to enable/ disable AGC function. This function cannot be enabled with
3-15
Thickness Gauge (TG)
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ CoatThk: For turning on/off coat thickness measurement. If turned on, when
known for measuring the thickness of the coat itself. The measure value is
the distance between the first wave and the peak wave in the gate A range.
Peak (the highest wave intersection). This adjustment of this item may be
used to set up the gate value mode of GateA and GateB at the same time.
◆ Coat Status: Thin / Thick. Based on “Coat Status” setting during AutoSrch,
GateB accurate positioning. Thin is for the coat smaller than or equal to
d) Alarm
3-16
Thickness Gauge (TG)
i) TempComp
3.2.3 ECHOMode
a) Probe
b) Gate
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
the measure value of GateA. The target is the positioning of the next echo
after GateA.
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
3-17
Thickness Gauge (TG)
【Note 3】If the Alarm function is not turned on or does not meet the alarm
requirement, the alarm symbol on the right side of system information
column is a hollow circle.
c) Meas
d) Alarm
3.2.4 VelMeas
a) Probe
◆ WorkMode: For selecting the work mode. Three modes are available:
3-18
Thickness Gauge (TG)
◆ PrbSel: For selecting the TG probe connected to the system, Std and
Custom are optional. The TG probe model series manufactured by SIUI are
object.
◆ CalibZero: A hot key button. “Ref1” and “Vel” in custom calibration are used
for probe delay calibration. That is to perform fast calibration of probe delay
according to the valid echo signals in GateA based on the known material
PrbDly is refreshed.
b) Gate
c) Meas
d) Alarm
3-19
Thickness Gauge (TG)
e) Data Save
3.2.5 MultiMode
a) Probe
The same as 3.2.1a) Probe.
b) MeasParam
◆ MeasObj: For setting the object of MultiMode, i.e. for selecting multi-layer
◆ LVel1: For setting MtVel of the 1st layer. When MultiMode object is Vel, this
◆ LVel2: For setting MtVel of the 2nd layer. When MultiMode object is Vel, this
◆ LVel3: For setting MtVel of the 3rd layer. When MultiMode object is Vel, this
◆ LVel4: For setting MtVel of the 4th layer. When MultiMode object is Vel, this
3-20
Thickness Gauge (TG)
c) LMeas
◆ AreaSw: For turning on/off CurArea and the measurement gate for the
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen
display.
rectification mode of receiving echo. There are four Rectify modes for
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
Peak (the highest wave intersection). This adjustment of this item may be
used to set up the gate value mode for each layer at the same time.
3-21
Thickness Gauge (TG)
d) Data
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
is a red solid circle that keeps flashing.
【Note 3】If the Alarm function is not turned on or does not meet the alarm
requirement, the alarm symbol on the right side of system information
column is a hollow circle.
f) Data Save
3-22
Thickness Gauge (TG)
3.2.6 B-Scan
a) Ctrl
◆ Save: With the memory connected, to save scan files to the memory, with
◆ Curr: In TimeBase mode, " Curr " refers to the scan time that the probe is
coupled to the workpiece; In Encoder mode, " Curr " refers to the scan
◆ Loss: To calculate the total missed time or distance due to excessive scan
b) Setup
◆ Encoder: To select the encoder model when Encoder mode is selected and
◆ Step: To set up trigger time for scan when TimeBase is selected; To set up
◆ Dis: The menu is adjustable when Encoder is selected. It is for setting the
c) Basic
◆ LosMode: To set up the scan stops or continues if the probe loses coupling,
◆ FrzMode: For setting the position of cursor line stays in the scanned image
when pressing the Freeze button. CurVal/ MinVal/ MaxVal for option.
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ DspDly: For setting the ultrasonic pulse shift (display delay) on screen
display.
In the reading column there is a coupling symbol with an inverted arrow. In StdMode, a
solid coupling symbol appears when gate A is coupled; In CoatMode or ECHOMode, a
solid coupling symbol appears when gate A and B are coupled at the same time, or a
hollow coupling symbol appears when one of gate A/B is coupled; In MultiMode, a solid
coupling symbol appears when any gate is coupled. There is no such symbol in other
situation.
The image information column of ultrasound images is as shown in Table 3-2, which is
for parameter setup information. See Table 3-2 for the definition.
Symbol Meaning
ASC AutoSrch enabled.
3-24
Thickness Gauge (TG)
Symbol Meaning
AGC AGC for Gate A/B enabled.
The gate is in valid coupling with the echo. (One gate in good
coupling in MultiMode)
Coupling signal display in AutoSrch; One gate in good coupling
in CoatMode/ ECHOMode.
Others See 1.5.
3.3.1 Reading
The reading column below the information column shows 5 readings. See Fig. 3-1.
The first reading is for gain, in the format: the upper one is total gain (with the unit in the
bracket) and the lower one is (base) gain + Comp.
In addition, 3 readings can be selected by the user, where the last reading is a big value
(BVa), which is highlighted in larger font size. See Table 2-2 for the options of reading.
Symbol Meaning
Hx Amplitude percentage of echo measurement point within gate x.
HxdB dB difference between echo measurement point amplitude and gate
height within gate x.
Std Standard thickness/ VelMeas value.
MaxVal The Max thickness/ VelMeas in the valid measure time in non B-Scan;
The maximum thickness of the scanned image in B-Scan;
Not available for MultiMode.
MinVal The minimum thickness/ VelMeas in the valid measure time in non B-
Scan;
The minimum thickness of the scanned image in B-Scan;
Not available for MultiMode.
Diff The thickness measurement difference with respect to DiffRef.
Not available for VelMeas and MultiMode.
Avg The average thickness/ Vel measurement based on AVGNum in non B-
Scan;
The average thickness of the scanned image in B-Scan;
Not available for MultiMode.
3-25
Thickness Gauge (TG)
3.4.1 StdMode
If WorkMode is StdMode, the thickness reading is the measure value between the
calibration zero and the GateA meas point (JFlank: the zero-crossing point of the first peak
exceeding gate thresh. Peak: the zero-crossing point of the peak exceeding gate thresh).
The small inverted triangle identifier automatically tracks gate measurement position.
For measurement in StdMode, see Fig. 3-2.
Operation:
(1) In Probe menu, set up PrbSel, ProbeDly and MtVel. If required, perform
(2) In Gate menu, adjust the gate position, set up measurement parameters and
turn on the over-limit alarm function.
3-26
Thickness Gauge (TG)
3.4.2 CoatMode
If WorkMode is CoatMode, the thickness reading is the measure value between the gate
A MeasPnt and gate B MeasPnt. (MeasPnt can be selected. JFlank: the zero-crossing
point of the first peak exceeding gate thresh. Peak: the zero-crossing point of the peak
exceeding gate thresh). The small inverted triangle identifier on the gate automatically
tracks the position.
For measurement in CoatMode, see Fig. 3-3.
If CoatThk is enabled, the coating thickness is measured when the base below the coating
is measured by CoatMode. While performing CoatMode, the zero-crossing point of the first
peak exceeding aThresh is measured by the software, and the coating thickness is
acquired from calculation.
Operation:
(1) In Probe menu, set up PrbSel, PrbDly and MtVel. Generally, perform CalibZero
for the known material velocity in Probe menu of StdMode. If required, perform
CalibZero in this menu again, and then enter the measurement state.
(2) In Gate menu, adjust the gate position, set up measurement parameters and turn
3-27
Thickness Gauge (TG)
(3) In Meas menu, set up Rectify and AutoSrch. If CoatThk measurement is enabled,
set up the known CoatVel, and the coating thickness can be measured at the
same time.
3.4.3 ECHOMode
If WorkMode is ECHOMode, the thickness reading is the measure value between gate A
MeasPnt and gate B MeasPnt. (MeasPnt can be selected. JFlank: the zero-crossing point
of the first peak exceeding gate thresh. Peak: the zero-crossing point of the peak exceeding
gate thresh). See Fig. 3-4.
The following is an example of "Peak" set in "MeasPnt".
In ECHOMode, Gate A is the trigger gate. Adjust Gate A to record the first bottom wave (by
ensuring the measurement position in the gate is the peak position of the correct first
bottom wave, which can be determined by adjusting aStart and aWidth, and the position of
the inverted triangle in the gate line), According to the peak position in gate A, bStart is
shifted automatically in proportion to record the second bottom wave. The difference of the
peak in gate B and the peak in gate A is the thickness value of Echo to Echo (also by
ensuring the measurement position in gate B is the peak position of the correct second
bottom wave, which can be determined by adjusting bStart and bWidth, and the position of
ECHOMode is suitable for fast measurement of B1-B2 with different thickness. When
adjusting the position 2 gates for manual measurement is required, use CoatMode. By
adjusting the Start and Width of the two gates, the Bm-Bn value (such as B2-B3, or B3-B4)
is measured.
3-28
Thickness Gauge (TG)
Operation:
(1) In Probe menu, set up PrbSel, PrbDly and MtVel. Generally, perform CalibZero
for the known material velocity in Probe menu of StdMode. If required, perform
CalibZero in this menu again, and then enter the measurement state.
(2) In Gate menu, adjust the gate position, set up measurement parameters and turn
(1) The measured waveform amplitude or gate should be adjusted according to the
measurement requirements for each mode, enabling the signal at the measured
or AGC cannot be achieved, please cancel the automatic function and use
3-29
Thickness Gauge (TG)
manual adjustment.
frequency narrow-pulse probe (i.e. less signal pulse cycles, for example the dual
measurement reading. For the general wide-pulse dual probe, ensure the cursor
measurement accuracy. You may manually adjust the gate start, width and
(3) If Rectify is RF, and GateThresh is default to negative, it measures the negative
to the first peak of negative signal in the gate exceeding gate thresh. The zero-
waveform signal (above the baseline). The measurements correspond to the first
peak of positive signal in the gate exceeding gate thresh. Due to the weak peak
signal, the zero-crossing after the peak is not stable, which is not conducive to
is set Neg) or Neg for Rectify, so as to get the most stable measurements
displayed.
3.5 VelMeas
Subject to the selected WorkMode, the material velocity is calculated and displayed based
on the known Thickness. This function is for velocity measurement of a single-layer
workpiece. For the velocity measurement of a multi-layer workpiece, use MultiMode.
Before exiting VelMeas, choose whether to apply the velocity measurement to the current
material.
Operation:
3-30
Thickness Gauge (TG)
(1) In Probe menu, set up WorkMode, PrbSel, PrbDly and the known thickness. If
(2) In Gate menu, adjust the gate position and set up the measurement parameters.
3.6 MultiMode
If the workpiece has multilayer materials which are linked closely, you can use MultMode
to measure, provided that the material velocity for each layer is known. The actual
measurement effect is dependent on the multi-layer material. If the thickness for each layer
is known, the velocity for each layer can also be measured. For MultMode measurement,
choose whether to apply the velocity measurement of the measurement to the current
material.
For MultMode measurement, up to 4 layers can be measured at the same time. There are
4 measurement gates L1, L2, L3, L4 in total. The measurement for the 1st layer is the sound
path difference between the first wave peak in gate L1 to zero; The measurement for the
2nd layer is the sound path difference between the first wave peak in gate L2 and L1; The
measurement for the 3rd layer is the sound path difference between the first wave peak in
gate L3 and L2; and the measurement for the 4th layer is the sound path difference between
(The description example above takes Peak as "MeasPnt". In the practice, you may also
set it as JFlank, the measure value at the intersection of the gate and the first wave. )
The main reading display for MultMode is the sum of layer measurements, with the
3-31
Thickness Gauge (TG)
Operation:
(3) In LMeas, adjust the gate position for the measurement area, and set up the
3.7 B-Scan
B-Scan function is to perform thickness measurement based on the set time interval or the
encoder, and display the measurement reading of scan process in B-mode images, which
is for thickness measurement image analysis and recording for the scanned workpiece.
The scanning record files can be recalled or previewed in Manage module.
3-32
Thickness Gauge (TG)
Operation:
regular parameters and calibrate zero. Return to the desktop and enters B-Scan
screen. B-Scan uses the current work mode for measurement. It does not support
(2) Set up ScanMode, Step, TimeL in Setup menu. Select the encoder, scan
(3) In “Basic”, set up Basic parameters, thickness display range, the mode for lost
(4) In “Ctrl”, you can start or finish B-Scan. After scanning, save the record or preview
Calibration wizard is for MtVel and PrbDly calibration. The user may choose to calibrate
PrbDly (1-point calibration), or PrbDly and MtVel (2-point calibration) at the same time.
3-33
Thickness Gauge (TG)
3.8 Calibration
1-point Calibration is to calibrate ProbeDly only with known velocity. Before calibration,
couple the probe on the test block to be calibrated.
Operation:
(1) Select the probe and the curve to enter the calibration menu (the "Calib" 1st menu
parameters. Select “PrbDly” as the calibration mode, and set up the test block
(2) Couple the probe on the test block 1 to be calibrated. Adjust “GAStart”, “GAWidth”
and “GAThresh” to bring Gate A to record the first wave. Set up the velocity of the
tested object.
(3) Tap “Record” to finish calibration. For 1-point Calibration, only 1 point is recorded
for PrbDly calibration. If not satisfied with the result, use “Clear” to restore to the
【Note】 After the 1-point Calibration, the ProbeDly value is refreshed by the
calibrated value.
2-point Calibration is to calibrate PrbDly and Vel at the same time. Before calibration,
couple the probe on the test block 1 to be calibrated.
Operation:
(1) Select the probe and the curve to enter the calibration menu (the "Calib" 1st menu
parameters. Select “PrbDly” as the calibration mode, and set up the the test block
thickness for Ref1 and Ref2. Adjust to the appropriate Range and DspDly.
3-34
Thickness Gauge (TG)
and “GAThresh” to bring Gate A to record the first wave. Set up the velocity of the
tested object.
(3) Tap “Record”. If the coupling is good, it prompts “Please record echo for Ref2”.
(4) Couple the probe on the test block 2 to be calibrated. Tap “Record”, and it prompts
“Calibration finished”. If not satisfied with the result, use “Clear” to restore to the
【Note】 After the 1-point and 2-point Calibration, the PrbDly and Vel values
are refreshed by the calibrated values.
3.9 TG Probe
TG probes include standard /custom. The standard TG probes are made by SIUI, with the
TG probe series as options. The NotStd TG probes are non-standard model or user defined
probe types. The probe specification series include: DualDelay, SingleDly, and SingleTch.
The TG probe selection screen is as shown in Fig. 3-7.
Operation:
The probe list is on the left, and the probe attributes on the right. Select one and press
Enter key to confirm.
3-35
Thickness Gauge (TG)
3.10 V-PATH
All the original dual probes have a set of default V-PATH calibration curves. Users can
make a set of UserVpath curves for a specific probe. Generally you need to identify whether
it is StdMode, CoatModeor or ECHOMode, and then make the UserVpath for that
WorkMode. CoatMode and ECHOModee may share the same V-PATH curve.
The VPATH curve for the current WorkMode of the probe in use is displayed on the left,
and the Basic information such as Type; PntNum; point attributes of the selected curve is
displayed on the right.
AutoSrch function may search and adjust waves quickly based on the received wave
signals. According to the wave signals of the tested thickness, AutoSrch quickly adjusts
the system automatically, including the adjustment of DspDly, Range, Gain and the
measurement gate. It provides a shortcut for operation. However, AutoSrch cannot
3-36
Thickness Gauge (TG)
If the probe type is Dual, in StdMode, CoatMode or WorkMode, AutoSrch shall include
auto adjustment of searching waves, range, gain and gates. If WorkMode is StdMode, the
software search will bring gate A to record the first wave and adjust AGC of the first peak
to 80%. If the workmode is CoatMode or ECHOMode, the software search will bring gate
A to record the first wave and adjust AGC of the first peak to 80%, and gate B automatically
records the second wave. AutoSrch in MultiMode and B-Scan is equivalent to auto gain A
only.
If the probe type is Single, AutoSrch in all work modes is equivalent to auto gain A only.
AutoSrch can be set Single auto search, i.e. to auto search once if enabled once, or
Continue (continuous) monitoring and search.
The maximum gain auto adjustment of AutoSrch is bound by the set value of “MaxSGain“.
If “MaxSGain“ is set too high, it may result in searching abnormal noise signals; If
“MaxSGain“ is set too low, it may result in the desired signals not measured. The specific
setup value shall be adjusted subject to the probe and workpiece.
【Note】" Coat Status " is suggested to set as below: set to “Thin” if the coat
thickness is less than or equal to 0.5mm, otherwise set to "thick".
Operation:
Select “Continue” or “Single” or “Off” in "AutoSrch" of Meas menu, or press the hot key
F1or F2 (when the hot key is defined as AutoSrch), to enable or disable this function.
If AGC is enabled, select to perform auto gain control for gate A or gate B (for CoatMode
and ECHOMode only). For StdMode, the first peak amplitude in gate A is automatically
adjusted to 80% height; For CoatMode and ECHOMode, the maximum peak amplitude in
the selected gate is automatically adjusted to 80% height.
AGC and AutoSrch functions cannot be enabled at the same time. The maximum gain auto
adjustment of AGC is bound by the set value of MaxSGain.
Operation:
3-37
Thickness Gauge (TG)
Select “GateA”, “GateB” or “Off” in “AGC “of Meas menu to enable or disable this
function.
3.12 TDG
TDG- Time Distance Gain curve. Gain compensation based on time and distance,
which may compensate the loss of echo amplitude due to propagation of sound
path.
When TDG function is turned on, the screen displays a TDG curve, which
corresponds to the actual signal position of gain compensation in the time axis
direction; while in the amplitude direction, the start gain and the target gain are
compressed in the screen for display, which does not represent the actual signal
amplitude.
Set up the attributes:
◆ StGain: For setting the start dB value of TDG curve.
StGain+CmpGain ≤ 110dB.
◆ StTime: The valid start time of TDG curve (the acoustic time counting from
PrbDly)
3-38
Thickness Gauge (TG)
Take the 2D data (4 rows, 5 columns) as an example for the operation. The figure
when the display mode is A/Ba is as shown in Fig. 3-9.
The upper part of the center is the Ba of the data, and the lower part of the center is the A-
scan figure. In this mode, the small data column of the upper part displays an enlarged
reading, i.e. the current measure value. "Std" means the reading is set to Standard (which
is set up in "Reading"), that is the measure value of gate A.
3.13.1 Save
Operation:
(1) Set up DNum, RowNum, ColNum and PntNum in "Format" (total point number
exceeding 1000.) If the storage file has been saved with data, you cannot
modify the content in "Format”, but only set up again after "Clear" operation.
(2) Exit the "Format" set up screen and return to the "Data" operation screen.
Select the desired "CurRow", "CurCol" and "CurPnt." Couple the probe
(3) Press Record key to save, or set up the Store key in “System” as the hotkey
(4) After saving properly, the measure value will be displayed on the
3.13.2 DataRcl
Operation:
(1) Adjust the "CurRow / CurCol / CurPnt" to the position to recall the data, and
run "DataRcl".
(2) If data is correct, the stored measure value is displayed in the reading column.
The A-scan figure is restored to the case where the data is saved, and the
3.13.3 Other
Save: after running the save function, the data is automatically named in the named format
and saved (set up the storage path in "Storage" of “Manage").
If the workmode is StdMode, see Fig. 3- for the flow chart of general thickness gauge
application.
3-40
Thickness Gauge (TG)
Power
on
N
Known
velocity?
Save to the
current record
position of the Y Save
current file measurement
number value?
Y Continue Y
testing?
Finish
For most materials, such as carbon steel, the general flow chart above is practical and
effective. On how to use the system, especially on calibration and auto search, see 3.11.
When testing a thickness close to the critical measurement range, sometimes AutoSrch
may need manual adjustment and judgment of the operator to ensure the measurement
accuracy. For coarse grain, high-attenuation material (such as ductile iron), special
composite materials and multilayer materials, please choose manual adjustment to
3-41
Thickness Gauge (TG)
measure. Now it is very important that the operator be acquainted with certain ultrasound
theories and testing basic knowledge.
3-42
Management Operation
This chapter describes the operation and technique for storage, work piece, probe, encoder
and VPATH management in the Manage menu.
The screen layout of storage management is shown in Fig. 4-1. For functional description,
see Table 4-1.
4-1
Management Operation
In the file list, it displays the selected file type only. Select "Type" menu, press Enter to
toggle between different options, including "Image", "Param", "Record" and "Data" (for TG
only).
◆ Param file: For saving system setup parameters. The formats include: UT
includes: UT CineRec, "*.urec" for B-scan file, and "*.tgrec" for TG B-Scan file.
◆ Data file: For TG mode ohly, which is for saving TG measure date, in the format
“*.tgdat”.
4.1.2 Save
In the storage management mode, only "Save" of "Param" file type is valid (for the storage
of other file types, use the save operation if F1 or F2 is set for related function).
Operation:
(3) Save the current ParamFile under the specified path (see 4.1.7.2).
4.1.3 Open
Method 1:
(1) Select "Open" menu. Press Right or Enter to enter the file list control;
(2) Press Up or Down to select the desired file to open, and press Enter to open it.
Note: If the type is Image, the file cannot be opened with this method.
Method 2:
4-2
Management Operation
(1) Select "Type" menu. Press Enter to select the type. When the selection is
(2) Select a file and press Enter, Follow the screen prompt to select the responsive
◆ Param: Apply the saved parameter information to the system, and exit the
4.1.4 Preview
Turn on "Preview" and enter the preview mode. You may view part of the information for
the file. See 4.1.3 for the operation steps.
4.1.5 Delete
Turn on “Delete”, and you may delete one, multiple or all files.
Operation:
(1) Select "Delete" menu. Press Right or Enter to enter the file list control.
(2) Select one or all files, press Enter, and follow the screen prompt to delete the
selected file(s).
Note:
4.1.6 Rename
4-3
Management Operation
Operation:
(1) Select "Rename" menu. Press Right or Enter to enter the file list control.
(2) Select one file, press Enter, and you may rename the file in the pop-up file
naming window.
4.1.7 StoreSet
When saving a file (not including image type), it requires to input the file name, Now the
system will provide a default file name to improve the user's operating efficiency. There are
2 types of file name and naming method: Default and Incr.
YYMMDD_hhmmss.
◆ Incr: To set up the format, you may specify the incremental start bit, and the
subsequent files to save will be automatically named in that format and increase.
4.1.7.2 StorePath
To set up the path for file storage. The user may choose "Default" or "Custom" for
StorePath.
4-4
Management Operation
(1) Set "StorePath" menu to "Custom" and select the StorePath information below
the menu. Press Right or Enter to enter the file list control.
(2) Press Up, Down or Enter to switch the directory, and set up StorePath by
4.1.7.3 StoreName
To set up whether to bring up the file name input screen when saving the file.
a) If "StoreName set "Off", the file will be saved directly according to the naming
rule for the set stored file and StorePath when saving a file.
b) If "StoreName" is set "On", the user can modify the file name in the pop-up file
Note: If the storage type is "Image", no file naming window is popped up when
storing.
4.1.7.4 NewFLR
Operation:
(1) Select "NewFLR" menu, press Right or Enter to to enter file list control.
(2) Press Up or Down to switch the current directory to the specified one.
(3) Press Enter and finish NewFLR operation by following the on-screen prompt.
Encoder management is for naming, calibrating and deleting the crawler encoders
connected with the system. The screen interface is shown in Fig. 4-2. See Table 4-2 for
functional description.
4-5
Management Operation
4-6
Management Operation
4.2.1 Edit
4.2.2 New
Operation:
(2) Adjust “Precisn” menu and input the known precision of the encoder.
a) Adjust "CalibD." menu to set up the moving distance for encoder calibration.
b) Enable "Calib". At the bottom of the screen it shows the operation tip.
c) Place the encoder behind the start position and press Enter.
d) Move the encoder to the finish position, i.e. the actual calibrated distance,
then press Enter. If the encoder moves forward during this process, the
"Move" menu will refresh the current moving distance in real time (based on
the vale calculated from encoder precision), otherwise, the prompt is invalid.
e) The new encoder precision is acquired based on the calibration distance and
f) After moving the encoder from the finish position backward to the start
position, press Enter. In this process, the “CurrMov” will refresh its moving
distance (the value calculated from the new encoder precision) in real time.
whether the calibration is normal (under normal situation, after calibration the
menu), if it is normal, follow the tip and press Enter, and the calibrated new
encoder precision will be set to the current setting, otherwise press HOME to
4-7
Management Operation
cancel.
(4) After finishing the step, press Right or Enter, input the encoder name and save
4.2.3 Delete
This function enables the user to delete the created encoder. After the "Delete" function is
enabled, operate by following the operation tip at the bottom of the screen.
4.2.4 Rename
This function enables the user to rename the created encoder. After the "Rename" function
is enabled, operate by following the operation tip at the bottom of the screen.
4.2.5 Test
Operation:
(3) Press Enter to start the test. Now the "move" menu is cleared at the same time.
(4) While the user moves the encoder, observe if the "move" menu value complies
(5) When the moving is finished, return to step (2) to continue testing, or press HOME
The VPATH management interface is as shown in Fig. 4-3. The user may create a new
VPATH curve or edit the VPATH curve created.
4-8
Management Operation
The middle list shows all the VPATH curves (including the built-in and user defined
curves) contained in the system.
4-9
Management Operation
4.3.1 Edit
Operation:
(2) Press Up/ Down to select the curve to be edited, then press Enter.
(3) Adjust "CurPnt" menu to specify the point on the curve to be modified. Then
"BlkThk" and "MeasRes" are updated to the stored value of the CurPnt.
(4) To adjust the CurPnt data, adjust "BlkThk" and "MeasRes" menus.
(5) To insert a point, you can turn on "InsertUp" or "InsertDn" function. One point is
automatically created before or after the current point. "BlkThk" and "MeasRes"
(8) If you are not satisfied with most of the point data in the current curve, turn on
(9) Turn on "Finish" to confirm whether to save the modified curve information.
4.3.2 New
Operation:
(2) Press Up / Down key to select a curve as a reference for the new curve, that is
to create a new curve exactly the same as the specified curve for modification,
4-10
Management Operation
(4) Turn on "Finish" to enter the input state. Input the curve name and confirm
4.3.3 Delete
See 4.2.3.
4.3.4 Rename
See 4.2.4.
【Note】 The default V-PATH curve is for browsing the record points only, while
edit and management functions can be performed on the custom curve. The V-
PATH curve has up to 32 measurement location points, and the selection of the
32 points shall focus on an actual measure range that requires smaller measure
error, such as the range lower than 10mm (steel longitudinal wave), or 20-40mm
(steel longitudinal wave) range, but not a random measure point between the
minimum to the maximum range (generally the default V-PATH is fine). For high
UserVpath curves.
The Wp management screen is as shown in Fig. 4-4. The user may create a new work
piece or edit the created work piece.
The user-defined workpiece can be called from "Wp", and the information such as
thickness and velocity of the workpiece after being called up will be applied to the current
parameter settings.
4-11
Management Operation
The probe management interface is as shown in Fig. 4-5. The user may create a new probe
or edit the probe that has been created.
The user-defined probe can be called from “Probe”, and the information such as probe
angle (ValK) and PrbDly after being called up will be applied to the current parameter
settings.
4-12
Management Operation
4-13
System Setup
In "Setup" menu, you can view and set up the basic information of the system, set up the
display content of Reading column in the test interface and view the activated optional
functions.
5.1 System
Operation:
The system setup consists of "System" and "UT/ TG", wherein "System" is for setting
system parameters, shared by UT and TG modules and consistent; UT and TG are valid
for the corresponding modes.
For the operation, you may refer to the prompts at the bottom of the screen.
Enter key is for quickly toggling the option menu value; Gate and Gain keys are for
increasing/ decreasing menu values.
settings, while some parameters that affect the operation (e.g. language,
5-1
System Setup
◆ Update: Connect the memory card containing the update package to the
automatically enter the update process, and shut down when finished.
Operation:
(2) The display after pressing Enter key is subject to the adjust object. 17-01-01
◆ Time: Adjust system hours/ minutes. Refer to "Date" for the operation.
◆ DispDate: Turn on/off the date displayed at the upper right of the screen.
◆ Unit: Set up the units used in the system, including "mm" and "inch".
【Note】 Lower screen brightness can extend the life of the display, and
extend the system's operating time when the system is battery-powered. If
the ambient brightness is high, the higher screen brightness may improve
the visual effect of the screen.
◆ Outdoor: When the user operates the system outdoors, turn on "Outdoor" to
help the user to view the system menu interface and operate more clearly.
◆ VGA: Set up VGA video output. When the user connects VGA video out port
5-2
System Setup
display screen to the external display. Now the "VGA" symbol will be
◆ Buzzer: used with "Alarm" function. If the alarm requirements are met and
the function is enabled, the buzzer in the system will give sound alarm.
◆ L/R Op: As the system function keys are all on one side, to facilitate the use
of either left or right hand easily, turning on and off this function to rotate the
screen display 180 °, and all the display will be rotated 180 °.
◆ LogExport: Enable this function, and the key operation logs during the use of
【Note】 The logs are saved in the SMARTORLOG folder of the root
directory in the memory by time range.
system.
◆ AGC value (UT only): Set up the target amplitude of auto gain.
◆ Storage key (TG only): For selecting the storage method, including QuickSv
➢ QuickSv: Quickly save the point date of dataset. Now the valid
5-3
System Setup
by the "current row / column / point", and the current line / column /
➢ Choose: For the user to select the storage type, including: PrintScreen,
ParamFile, DataRecord.
◆ Precisn (TG only): set up the display precision of the measurement data.
5.2 Reading
Enter the "Reading" function, and you can select the display object of the reading column.
5.3 Activate
Operation:
(1) Connect the memory to the system till the system prompts the detection of
the memory.
5-4
System Setup
【Note】 The file storage path defaults to the directory with the SN as the
directory.
(4) Send the exported file to the vendor.
(5) Copy the license file obtained from the vendor into the memory.
(6) Connect the memory to the system till the system prompts the detection of
the memory.
(8) The screen displays the list identified as a directory or ini file in the memory.
Through the need and prompt to switch the directory, find the target license
(9) If the function is activated successfully, the software function status, the
5-5
Operator Training and System Maintenance
Maintenance
To ensure a safe and proper operation, the operator shall receive appropriate training on
ultrasound testing, especially training on phased-array ultrasound testing, and read this
Operation Manual before operating this system. The lack of relevant knowledge on
ultrasound testing may lead to unpredictable wrong testing results. Therefore, to acquire
information on ultrasonic testing training and qualification, please contact relevant NDT
societies, organizations or SIUI.
Please use the factory configured accessories and options, such as probe cable and
communication cable. Incompatible cables may result in system internal circuit failure or
external connector damage.
6.2.2 Cleaning
DO NOT drop any liquid into the system. The system enclosure, glass panel and other
parts may be cleaned with a piece of cloth soaked with a little water or neutral household
cleanser.
【Note】: DO NOT use any solvent to clean the system. The plastic parts may be
damaged or become fragile if cleaned with solvent.
6.3 Service
The system adopts the latest technologies and is made of high quality components. The
process inspection, the in-process testing and the quality management system are all
certified by ISO9001, ensuring that the system performance is optimized.
In case any failure is found in the system, please shut off the system and take out the
battery. Contact SIUI service representative whenever it is necessary to repair or re-
calibrate the system. The contact information is listed in the Preface of the Operation
Manual. The warranty does not cover any unauthorized repair or disassembly.
Under the following situations, the safe operation of the system is not ensured:
6-1
Operator Training and System Maintenance
b) The system is stored in adverse environment for a long time (eg. abnormal
humidity-proof.
6-2
Common Failures and Troubleshooting
1) Copy the software update package to the root directory of the SD card.
2) In system power-off state, connect the SD card to the system SD card port.
3) Turn on the system. In the desktop button screen, enter the "Setup" menu
screen, select System" 1st menu, press Enter in the "Update" menu, and
a) If the power supply is from the battery, press Power and the system has no
response:
2) Remove the battery from the system, and wait for about 10 seconds before
installing the battery again in the system. Try to turn on the system.
3) Remove the battery from the system, and try to use the adapter for power
supply.
b) If the power supply is from the adapter, press Power and the system has no
response:
7-1
Common Failures and Troubleshooting
1) Check if the system is installed with battery. If yes, remove the battery, and use
the adapter for power supply separately. Try to turn on the system.
4) Disconnect the adapter from the main unit. Wait for about 10 seconds before
5) If the battery is not used on the system, try to use the system with a full-charge
c) The system can be turned on, but cannot enter the application properly.
1) Turn off the system, and disconnect all the peripherals connected to the
system, including USB device, crawler and network cable. Try to turn on the
system.
2) If step 1) does not work, try to update the system with the update package.
d) If the failure still exists after trying the solution above, please contact the distributor
or SIUI.
If normal shutdown cannot power off the system, long press Power for about 10 seconds
to force the system off, or unplug the adapter and/ or remove the battery to cut off power
supply. Restart the system to see whether it is in normal operation. If not, try to contact
the distributor or SIUI for service as soon as possible.
7-2
Common Failures and Troubleshooting
7.5 No Echo
a) Check if the S/DPrb setting is correct, whether it is set Dual, but with a single
b) Check if the probe cable is in bad contact. Try to replace another probe cable.
c) Check if the setup of Gain, Range, DspDly, PrbDly and Reject are correct.
d) Set the system to Default, use a new probe cable to connect a good probe and
scan on a test block. If echoes appear, the failure is due to improper setting.
e) If the failure still exists after trying the solutions above, please contact the
distributor or SIUI.
a) Check if AVG curves are made. DAC curves cannot be created if AVG curves
are made.
c) Check if the echo amplitude recorded by the gate is not within the range of 20%
~100%.
a) Check if DAC curves are made. AVG curves cannot be created if DAC curves are
made.
c) Check if the echo amplitude recorded by the gate is not within the range of 20%
7-3
Common Failures and Troubleshooting
~100%.
c) Check whether the gate records the echo to be tested or there are clutters with
higher echo amplitude. By measuring PeakEcho within gate, only when the echo
being recorded by the gate properly, can accurate calculation result comes out.
refers to the horizontal distance between the incident point of probe and the flaw,
otherwise, it refers to the horizontal distance between the probe flank and the
flaw.
e) Check whether the gate records the echo to be tested or there are clutters with
higher echo amplitude. By measuring PeakEcho within gate, only when the echo
being hitched by the gate properly, can accurate calculation result comes out.
between the flaw (beam reflection point) and the probe, rather than between the
round-hole center and the probe. Otherwise, the bigger the round-hole diameter
7-4
Common Failures and Troubleshooting
loose.
b) Check if the adapter indicates normal operation. The adapter indicator is green
If the failure still exists after trying the solution above, please contact the distributor or
SIUI.
7-5
System Technical Specifications
Display Screen 5.7''TFT high brightness LCD screen, resolution 640 × 480
Operation
Keys
Method
Unit mm/inch
Model DC-A010
Capacity 5.0Ah
Battery
Voltage 7.4V
Quantity 1pc
Operation
≥8 hours
Time
Model CD-92A
Power 36W
A-1
System Technical Specifications
Network Port ——
Encoder
1 pc (4-core)
Connector
TG calibration
1 pc
block
Operation
-10℃~45℃
Operating Temperature
Conditions
Relative humidity 30%~85% (without condensation)
IP Rating ≥IP66
A-2
System Technical Specifications
Max. supporting
2 pcs
elements
A /D sampling
Max. 240MHz /10bit
frequency
Number of sampling
1024 points, 16bit/point
points
11 levels:
1~4MHz/0.5~10MHz/2~20MHz/1MHz/2.5MHz/4MHz/
Operating Frequency
5MHz/10MHz/13MHz/15MHz/20MHz
A-3
System Technical Specifications
Number of gates: 2
Detection sensitivity
≥65dB (P2.5-20 probe)
surplus
A-4
System Technical Specifications
Indication
0.001/0.01/0.1 mm (0.0001/0.001/0.01 inch)
resolution
A-5
System Technical Specifications
0.80~9.99mm ± 0.05mm
Equipped with TGM5-10L
Indication
10.00~99.99mm ± (1‰H + 0.04)mm probe; H is the measured
accuracy error
thickness value
100.0~400.0mm ± 3‰H mm
Pipe wall
Equipped with TGM5-10L probe, it can measure the wall thickness of
thickness
steel pipe with diameter ≥Φ20mm and wall thickness ≥2.0mm.
measurement
Range 5~1000mm
Pulse Type Negative square, negative spike (auto-match according to probe type)
PRF 200Hz
Operating
0.5~20MHz (auto match to the probe type)
Frequency
0~110dB
off/on; If this function is activated, the display range, gain and gate
AutoSrch position can be automatically adjusted properly according to the
measured waveform signals, so as to improve measurement efficiency.
A-6
System Technical Specifications
Measurement
Std, MinVal, MaxVal, Avg, Diff
reading mode
Display
A/BVa, A/Ba/SVa, Ba/BVa
Combination
Refresh rate of
4/8/16/32Hz
measurements
Data post Parameters, data sets, printscreen files can be played back and analyzed
processing with SuporUp software, and a report can be generated.
A-7
Appendix B Compatible TG Probe Specifications
【Note 1】The table above applies to the thickness measurement (TG) module.
【Note 2】When using a dual probe for measurement in CoatMode or ECHO mode, it is recommended to use
a TGM5-10L wideband narrow-pulse probe for high accuracy and stable measurement readings.
【Note 3】If a low frequency (less than 2.5MHz) or high temperature probe is adopted, it is recommended to
select precision as 0.1mm or 0.01 in.
【Note 4】For measuring coarse or high attenuation materials such as cast iron, it is recommended to select
precision as 0.1mm or 0.01 in.
【Note 5】Using other probes (including the user-defined probes) except the thickness probe supported by SIUI,
the measurement range and precision cannot be predictable and guaranteed.
Rua Bartolomeu Velho nº 689 – Hab 63 ; 4150- 124 PORTO - PORTUGAL | VAT No 509956777
Tel.: (351)938736955 | email: [email protected] | www.weldndt.pt
Mod.WNDT 01-11
8