0% found this document useful (0 votes)
415 views6 pages

LIS Communication Protocol Specification - 20191126 - Rev.0.2

The document describes the structure and fields of common HL7 message segments used in laboratory information system communications including the MSH, PID, OBR, ORC, SPM, and OBX segments. An example message is provided that shows how these segments would be structured in a message from an analyzer to a server with patient information, test details, device details, specimen details, and test results. Each segment and field in the example message is described.

Uploaded by

Johnny Tello
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
415 views6 pages

LIS Communication Protocol Specification - 20191126 - Rev.0.2

The document describes the structure and fields of common HL7 message segments used in laboratory information system communications including the MSH, PID, OBR, ORC, SPM, and OBX segments. An example message is provided that shows how these segments would be structured in a message from an analyzer to a server with patient information, test details, device details, specimen details, and test results. Each segment and field in the example message is described.

Uploaded by

Johnny Tello
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Document ID Name LIS Communication Protocol Specifications Rev. 0.

2
R-A1R-010 Product For POCcelerator Page 1/6

1. MSH segment structure


-. This segment is the HL7 MSH message segment, indicating the Message Header.
-. Segment Field
• MSH-1: Field Separator (ST)
• MSH-2: Encoding Characters (ST)
• MSH-3: Sending Application (HD) optional
• MSH-4: Sending Facility (HD) optional
• MSH-5: Receiving Application (HD) optional
• MSH-6: Receiving Facility (HD) optional
• MSH-7: Date/Time of Message (DTM)
• MSH-8: Security (ST) optional
• MSH-9: Message Type (MSG)
• MSH-10: Message Control ID (ST)
• MSH-11: Processing ID (PT)
• MSH-12: Version ID (VID)
• MSH-13: Sequence Number (NM) optional
• MSH-14: Continuation Pointer (ST) optional
• MSH-15: Accept Acknowledgment Type (ID) optional
• MSH-16: Application Acknowledgment Type (ID) optional
• MSH-17: Country Code (ID) optional
• MSH-18: Character Set (ID) optional repeating
• MSH-19: Principal Language Of Message (CWE) optional
• MSH-20: Alternate Character Set Handling Scheme (ID) optional
• MSH-21: Message Profile Identifier (EI) optional repeating
• MSH-22: Sending Responsible Organization (XON) optional
• MSH-23: Receiving Responsible Organization (XON) optional
• MSH-24: Sending Network Address (HD) optional
• MSH-25: Receiving Network Address (HD) optional

2. PID segment structure


-. This segment is the HL7 PID segment, indicating the Patient Identification
-. Segment Field
• PID-1: Set ID - PID (SI) optional
• PID-2: Patient ID (CX) optional
• PID-3: Patient Identifier List (CX) repeating
• PID-4: Alternate Patient ID - PID (CX) optional repeating
• PID-5: Patient Name (XPN) repeating
• PID-6: Mother's Maiden Name (XPN) optional repeating
• PID-7: Date/Time of Birth (DTM) optional
• PID-8: Administrative Sex (IS) optional
• PID-9: Patient Alias (XPN) optional repeating
• PID-10: Race (CWE) optional repeating

3. OBR segment structure


-. This segment is the HL7 OBR message segment, indicating the Observation Request
-. Segment Field
• OBR-1: Set ID - OBR (SI) optional
• OBR-2: Placer Order Number (EI) optional
• OBR-3: Filler Order Number (EI) optional
• OBR-4: Universal Service Identifier (CWE)
• OBR-5: Priority (ID) optional
• OBR-6: Requested Date/Time (DTM) optional
• OBR-7: Observation Date/Time # (DTM) optional
• OBR-8: Observation End Date/Time # (DTM) optional
• OBR-9: Collection Volume * (CQ) optional
• OBR-10: Collector Identifier * (XCN) optional repeating
• OBR-11: Specimen Action Code * (ID) optional
Document ID Name LIS Communication Protocol Specifications Rev. 0.2
R-A1R-010 Product For POCcelerator Page 2/6

• OBR-12: Danger Code (CWE) optional

4. ORC segment structure


-. This segment is the HL7 ORC message segment, indicating the Common Order
-. Segment Field
• ORC-1: Order Control (ID)
• ORC-2: Placer Order Number (EI) optional
• ORC-3: Filler Order Number (EI) optional
• ORC-4: Placer Group Number (EI) optional
• ORC-5: Order Status (ID) optional
• ORC-6: Response Flag (ID) optional
• ORC-7: Quantity/Timing (TQ) optional repeating
• ORC-8: Parent (EIP) optional
• ORC-9: Date/Time of Transaction (DTM) optional
• ORC-10: Entered By (XCN) optional repeating
• ORC-11: Verified By (XCN) optional repeating
• ORC-12: Ordering Provider (XCN) optional repeating
• ORC-13: Enterer's Location (PL) optional
• ORC-14: Call Back Phone Number (XTN) optional repeating
• ORC-15: Order Effective Date/Time (DTM) optional
• ORC-16: Order Control Code Reason (CWE) optional
• ORC-17: Entering Organization (CWE) optional
• ORC-18: Entering Device (CWE) optional
• ORC-19: Action By (XCN) optional repeating

5. SPM segment structure


-. This segment is the HL7 SPM message segment, indicating the specimen
-. Segment Field

• SPM-1: Set ID - SPM (SI) optional


• SPM-2: Specimen ID (EIP) optional
• SPM-3: Specimen Parent IDs (EIP) optional repeating
• SPM-4: Specimen Type (CWE)
• SPM-5: Specimen Type Modifier (CWE) optional repeating
• SPM-6: Specimen Additives (CWE) optional repeating
• SPM-7: Specimen Collection Method (CWE) optional
• SPM-8: Specimen Source Site (CWE) optional
• SPM-9: Specimen Source Site Modifier (CWE) optional repeating
• SPM-10: Specimen Collection Site (CWE) optional
• SPM-11: Specimen Role (CWE) optional repeating
• SPM-12: Specimen Collection Amount (CQ) optional
• SPM-13: Grouped Specimen Count (NM) optional
• SPM-14: Specimen Description (ST) optional repeating
• SPM-15: Specimen Handling Code (CWE) optional repeating
• SPM-16: Specimen Risk Code (CWE) optional repeating
• SPM-17: Specimen Collection Date/Time (DR) optional
• SPM-18: Specimen Received Date/Time (DTM) optional
• SPM-19: Specimen Expiration Date/Time (DTM) optional

6. OBX segment structure


-. This segment is the HL7 OBX message segment, indicating the Observation/Result
-. Segment Field.
• OBX-1: Set ID - OBX (SI) optional
• OBX-2: Value Type (ID) optional
• OBX-3: Observation Identifier (CWE)
• OBX-4: Observation Sub-ID (ST) optional
• OBX-5: Observation Value (Varies) optional repeating
Document ID Name LIS Communication Protocol Specifications Rev. 0.2
R-A1R-010 Product For POCcelerator Page 3/6

• OBX-6: Units (CWE) optional


• OBX-7: References Range (ST) optional
• OBX-8: Abnormal Flags (IS) optional repeating
• OBX-9: Probability (NM) optional
• OBX-10: Nature of Abnormal Test (ID) optional repeating
• OBX-11: Observation Result Status (ID)

7. Message example
-. Message formulation from Analyzer to server
-. <SB>: 0x0B
-. <EB>: 0x1C
-. <CR>: 0x0D

<SB>
MSH|^~\&|Protocol|ModelName|DeviceNum||Date||OUL^R24^OUL_R24|1|T|2.6<CR>
PID||PatientID1|PatientID2|PatientID3|||Age|Gender<CR>
OBR||TestName|TestType|SampleType|||Date|||UserID<CR>
ORC|OK|||||||||||||||||DeviceNum|SlotNum<CR>
SPM|1|CartridgeLotName|||||||||||||||||CartridgeExpiryDate<CR>
SPM|2|ControlLotName||ControlName|||||||||||||||ControlExpiryDate <CR>
OBX|1|TX|Result1TestName||Result1Value|Unit1|Result1Flag|Result1Type|||R<CR>
OBX|2|TX|Result2TestName||Result2Value|Unit2|Result2Flag|Result2Type|||R<CR>
OBX|3|TX|Result3TestName||Result3Value|Unit3|Result3Flag|Result3Type|||R<CR>
OBX|4|TX|Result4TestName||Result4Value|Unit4|Result4Flag|Result4Type|||R<CR>
OBX|5|TX|Result5TestName||Result5Value|Unit5|Result5Flag|Result5Type|||R<CR>
<EB><CR>

-. The meaning of segment


Each parameter
Segment example
segment meaning

<SB> Start Block character

MSH|^~\&|Protocol|ModelName|DeviceNum||Date||OUL^R24^OUL_R24|1|T|2.6<CR> Message header

PID||PatientID1|PatientID2|PatientID3|||Age|Gender<CR> Patient information

OBR||TestName|TestType|SampleType|||Date|||UserID<CR> Test information

ORC|OK|||||||||||||||||DeviceNum|SlotNum<CR> Device information

Cartridge
SPM|1|CartridgeLotName|||||||||||||||||CartridgeExpiryDate<CR> information
Control
SPM|2|ControlLotName|ControlName||||||||||||||||ControlExpiryDate <CR> information

OBX|1|TX|Result1TestName||Result1Value|Unit1|Result1Flag|Result1Type|||R<CR> Test result 1

OBX|2|TX|Result2TestName||Result2Value|Unit2|Result2Flag|Result2Type|||R<CR> Test result 2

OBX|3|TX|Result3TestName||Result3Value|Unit3|Result3Flag|Result3Type|||R<CR> Test result 3

OBX|4|TX|Result4TestName||Result4Value|Unit4|Result4Flag|Result4Type|||R<CR> Test result 4

OBX|5|TX|Result5TestName||Result5Value|Unit5|Result5Flag|Result5Type|||R<CR> Test result 5

End Block character,


<EB><CR> Carriage return
Document ID Name LIS Communication Protocol Specifications Rev. 0.2
R-A1R-010 Product For POCcelerator Page 4/6

-. each segment location of the results in HL7 field

Number Parameter Description Valid value Segment location

1 Protocol Protocol version Digit MSH-3

2 ModelName Device model name alphanumeric MSH-4

3 DeviceNum Device serial number alphanumeric MSH-5, ORC-18

Test date
4 Date alphanumeric MSH-7, OBR-7
(YYYYMMDDHHMMSS)

5 UserID Operator ID alphanumeric OBR-10

6 Patient ID1 User input Patient ID alphanumeric PID-2

7 Patient ID2 Specimen barcode ID alphanumeric PID-3

8 Patient ID3 Default Patient ID alphanumeric PID-4

9 Age Patient age Digit PID-7

10 Gender Patient gender Male / Female PID-8

Test name
11 TestName alphanumeric OBR-2
Ex) CRP, PCT

0: General mode
12 TestType 1: QC mode Digit OBR-3
2: QC calibration mode
0: Whole blood
1: Serum/Plasma
2: Urine
13 SampleType Digit OBR-4
3: Feces
4: Swab
5: Etc
14 CartridgeLotName Test cartridge lot name alphanumeric 1st SPM-2

Test cartridge expiry date


15 CartridgeExpiry Date alphanumeric 1st SPM-19
(YYYYMMDD)

16 ControlLotName Control lot name alphanumeric 2nd SPM-2

17 ControlName Control name alphanumeric 2nd SPM-4


Control expiry date
18 ControlExpiryDate alphanumeric 2nd SPM-19
(YYYYMM)

Test name of result 1


19 Result1TestName alphanumeric 1st OBX-3
Ex) CRP, PCT

Test name of result 2


20 Result2TestName alphanumeric 2nd OBX-3
Ex) CRP, PCT

Test name of result 3


21 Result3TestName alphanumeric 3rd OBX-3
Ex) CRP, PCT
Document ID Name LIS Communication Protocol Specifications Rev. 0.2
R-A1R-010 Product For POCcelerator Page 5/6

Test name of result 4


22 Result4TestName alphanumeric 4th OBX-3
Ex) CRP, PCT

Test name of result 5


23 Result5TestName alphanumeric 5th OBX-3
Ex) CRP, PCT

24 Result1Value Quantitative result of Result 1 alphanumeric 1st OBX-5


Qualitative result of Result 1
0: Positive
25 Result1Flag Digit 1st OBX-7
1: Negative
2: Indeterminate
Result type of Result 1
0: Quantitative
26 Result1Type Digit 1st OBX-8
1: Qualitative
2: Semi quantitative

27 Result2Value Quantitative result of Result 2 alphanumeric 2nd OBX-5


Qualitative result of Result 2
0: Positive
28 Result2Flag Digit 2nd OBX-7
1: Negative
2: Indeterminate
Result type of Result 2
0: Quantitative
29 Result2Type Digit 2nd OBX-8
1: Qualitative
2: Semi quantitative

30 Result3Value Quantitative result of Result 3 alphanumeric 3rd OBX-5


Qualitative result of Result 3
0: Positive
31 Result3Flag Digit 3rd OBX-7
1: Negative
2: Indeterminate
Result type of Result 3
0: Quantitative
32 Result3Type Digit 3rd OBX-8
1: Qualitative
2: Semi quantitative

33 Result4Value Quantitative result of Result 4 alphanumeric 4th OBX-5


Qualitative result of Result 4
0: Positive
34 Result4Flag Digit 4th OBX-7
1: Negative
2: Indeterminate
Result type of Result 4
0: Quantitative
35 Result4Type Digit 4th OBX-8
1: Qualitative
2: Semi quantitative

36 Result5Value Quantitative result of Result 5 alphanumeric 5th OBX-5


Qualitative result of Result 5
0: Positive
37 Result5Flag Digit 5th OBX-7
1: Negative
2: Indeterminate
Result type of Result 5
0: Quantitative
38 Result5Type Digit 5th OBX-8
1: Qualitative
2: Semi quantitative
Document ID Name LIS Communication Protocol Specifications Rev. 0.2
R-A1R-010 Product For POCcelerator Page 6/6

39 Unit1 Unit of result 1 alphanumeric 1st OBX-6

40 Unit2 Unit of result 2 alphanumeric 2nd OBX-6

41 Unit3 Unit of result 3 alphanumeric 3rd OBX-6

42 Unit4 Unit of result 4 alphanumeric 4th OBX-6

43 Unit5 Unit of result 5 alphanumeric 5th OBX-6

44 SlotNum Device slot number alphanumeric ORC-19

-. Message formulation from Server to analyzer (Only 3 byte)


-. <SB>: 0x0B
-. <EB>: 0x1C
-. <CR>: 0x0D

- E.O.D -