AUTOSCAN 4 Installation Guide

Guía para la instalación eficiente del sistema MICROSCAN AUTOSCAN
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
491 views2 pages

AUTOSCAN 4 Installation Guide

Guía para la instalación eficiente del sistema MICROSCAN AUTOSCAN
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Manual Reproduction Specification

Pages:

Total number of originals: 2 (Including reproduction specification)

Finished page size: 8-1/2x11

NOTES:

1. ALL ORIGINAL PRINTING MASTERS AND ARTWORK MUST BE RETURNED TO


DADE MICROSCAN INC., ATTN.: DOCUMENT SERVICES.
2. SEE SHEET 2 FOR ORIGINAL ARTWORK.
3. STORAGE CONDITION: ENCLOSE IN A CONTAINER.

IMP REV ECN/WR NO. APPROVAL DATE APPROVALS DATE


RG A 00-09974 M Harding 10/01/01 R/D G Galvan 09/28/01 Dade MicroScan Inc.
R/A N/A West Sacramento, CA 95691 USA
QA J Hone 09/28/01
MAT D Miles 09/28/01 AS-4 INSTALLATION CHECKLIST
TS N/A
SW G Galvan 09/28/01 SIZE DRAWING NO REV
CKD (CURRENT REV.) D Henley 09/28/01 MKT A Allaire 09/28/01 A 9020-6647 A
REQ. (CURRENT REV.) B Weathersby 09/29/01 SCALE N/A Sheet 1 OF 2
autoSCAN®-4 Installation Checklist
_____ Check parts received using accompanying packing list.

_____ Unpack autoSCAN®-4, computer, report printer, line conditioner and place in desired operating locations. Plug
in the power cords.

_____ Connect the report printer to the computer's parallel port.

_____ Connect the autoSCAN®-4 to serial port 1 on the back of the computer.

_____ Load report printer with paper.

_____ Inspect the autoSCAN®-4 and ensure that the plastic photodiode shield, the glass diffuser plate, and retaining
clips are in place. With the drawer and cover closed, turn the instrument on. The drawer should come out and the
colorwheel should rotate (as evidenced by the changing colors in the front display). Allow the instrument to
warm up for one hour before calibration and panel reading.

_____ Turn the computer ON. Turn the printer ON

_____ Install LabPro, LabPro Diagnostics and LabPro Interface.

_____ Start the LabPro Command Center.

_____ Enter Utilities.

_____ Select Configuration\ autoSCAN®-4\ Port assignment. Select serial port 1 as the autoScan-4 assigned port. Select
OK.

_____ Select Configuration\ autoSCAN®-4\ Water Blank File. Verify the instrument is on then select OK.

_____ Select Ctrl+R to restore the water blank file.

_____ Insert the water blank disk included with the instrument and select OK. Confirm that the file should be restored by
selecting yes.

_____ Close all utilities windows.

_____ Open AS4 Diagnostics.

_____ From the Data\ Calibration tab calibrate the instrument. From the alignment tab verify the power supply
alignment. From the test tab verify all tests pass.

_____ Place a water blank panel (equilibrated for one hour) onto the autoSCAN®-4 drawer.

_____ From the Data\ Repeatibility tab select 5 as the number of reads. Select Read Panel. Verify that there are no
asterisked values on the screen after the test has been completed. Be sure to view all filters.

_____ Exit Diagnostics.

_____ Select Patient Order Entry from LabPro Command Center select Read\ autoScan-4 Calibrate\ QC Diagnostics.
Run the QC Diagnostics. Verify that everything passes.

_____ Perform Instrument Calibration. Check to make certain the drawer is empty and press the READ button on the
front panel of the instrument. Print the Calibration Report. Verify that it is in control. The autoSCAN®-4 is now
ready to read panels.

NOTE: If you encounter any problems, follow the Total System Checklist.

Date____________________ autoSCAN-4 Serial Number____________________ FSR____________________


9020-6647, Rev. A