Model 4215 Load Cell Indicator Manual PM-4101
Model 4215 Load Cell Indicator Manual PM-4101
3 Accredited
Table of Contents
1.INTRODUCTION.........................................................................................................................................................................................................4
2. QUICK START.............................................................................................................................................................................................................5
3. RUN MODE.................................................................................................................................................................................................................7
3.1 ITEM BUTTON
3.2 TARE BUTTON
3.3 RESET BUTTON
3.4 VIEW BUTTON
3.5 PRINT BUTTON
3.6 UNIT BUTTON
SETUP MODE...............................................................................................................................................................................................................11
SENSOR DELETE.........................................................................................................................................................................................................14
>CALIBRATION.........................................................................................................................................................................................................14
>>Cal Check
>>Cal by mV/volt
>>Cal by Masses
>>Cal by Shunt Val
>>Counts per Inch
>LIMITS......................................................................................................................................................................................................................16
>SYSTEM OPTIONS.................................................................................................................................................................................................17
>>Printer Baud
>>Auto Identify
>>Auto Zero Ch A/B
>>Com Address
>>Com Baudrate
>>Com Line Feed
>>Retain Tare
>DISPLAY OPTIONS.................................................................................................................................................................................................18
>>Filter Select
>>Decimals Ch A/B
>>Count By Ch A/B
>>Second Line
5. REMOTE OPERATION............................................................................................................................................................................................20
DIGITAL INPUTS
RS232 OPERATION
SENSOR CONNECTORS............................................................................................................................................................................................48
SERIAL PORT CONNECTOR.....................................................................................................................................................................................48
DIGITAL I/O CONNECTOR........................................................................................................................................................................................49
CHART REC. CONNECTOR.......................................................................................................................................................................................50
PRECISION SHUNT CALIBRATION RESISTORS..................................................................................................................................................51
1. Introduction
The Model 4215 is a versatile precision instrument intended for the digital readout of strain gage sensors such as
load cells. Optical encoders are also supported.
Features:
* Full bi-polar 6 digit display (±999,999) using two line, 20 character, vacuum fluorescent display for clear, precise
indication of measured quantities and limit status.
* No knobs or dials. No internal jumpers. All adjustment and calibration is done using user-friendly front panel
menus. Keylock option to prevent accidental or unauthorized changes to the unit’s setup.
* Measure load in pounds, kilograms, Newtons, PSI, MegaPascals, kilo-pounds, kiloNewtons, metric tons, mV/V, or
grams. User defined base area for PSI and MPa. Tare at any time.
* Capture the load peak and valley, display in pounds, kilograms, Newtons, PSI, MegaPascals, kilo-pounds,
kilo-Newtons, metric tons, or mV/V. Reset at any time.
* Measure position and velocity with optional encoder input. Display in English units, metric units, or in percent
of a user defined base length. Reset at any time.
* Four contact closure Limit switches. You can set which signal drives the limit (load, peak, valley, position, or
velocity), normally open / normally closed, set point, and manual reset (latched) or automatic reset (not latched).
* Print current readings of load, peak, valley, position, and velocity using the currently selected units at the touch
of one button.
* Calibration for load cells using manufacturer’s milli-volt per volt calibration factor, 5 point milli-volt per volt, 2 or
5 point known mass, or internal precision shunt calibration. A back panel switch allows you to choose between
two values used for the shunt calibration so there is no need to open the unit.
* Two Channels of Analog output of any signal (load, peak, valley, position, velocity) with user defined scale factor
and offset. Large, buffered ± 10.00 volt output.
* Back panel digital inputs for tare, peak and valley reset, position reset, and print. Other functions available as
custom programmed options.
* Optionally measure the average force using a contact closure to begin and end the averaging period.
2. Quick Start
This section will help you get your Model 4215 set up and operating in just a couple of minutes. The back panel is
shown below. You will need to make at least two connections here. First be sure that the power switch is off (the
lower part of the rocker should be pressed in). Connect the AC power line and plug it in.
Next connect a load cell to the 9 pin female connector labeled “Sensor A”. You should use the cabling supplied with
your Model 4215 and load cell. If you do not have a factory supplied cable, see Appendix B for the recommended
load cell connection.
You are now ready to turn on the power. Each time power is applied to the Model 4215 you will see a sequence of
6 messages appear on the front display. For example,
# Message Meaning
Ch A Cal 4.00000 mVv The mV/V constant used with the load cell on channel A.
5 Ch B Cal 4.00000 mVv The mV/V constant used with the sensor on channel B.
Note: If either of these sensors was calibrated by 5 –Point Mass, then 4 mV/V constants will be displayed here.
Ch A Exc 5.00 Volt Excitation voltage used with the load cell on channel A.
6
Ch B Exc 2.50 Volt Excitation voltage used with the sensor on channel B.
Each of these messages will be displayed for about 3 seconds. After that the front display will start showing the
current readings from the load cell and sensor.
To change what item is being displayed press the leftmost button. This will cycle through the list of items (Load A,
Peak A, Valley A, Load B, Peak B, Valley B, Position, Velocity, Limits, optionally: Avg A, Avg B).
The rightmost button will change the units that are being used. Repeated presses of this button will cycle through
the list of units available for the currently displayed item. For example, with Load A showing you would see Lb, kg,
N, PSI, MPa, Klb, kN, t, mV/V, g and then back to Lb.
3. Run Mode
The Model 4215 has three modes of operation, Run mode, Setup mode, and System Calibration mode. At
power-on it will be in Run mode.
The first line of the display show an item label (Load A), a value (0.0000), and a unit label (Lb).
The second line of the display can show the status of the 4 contact closure limit switches, show another display,
show text, or be blank.
To change what item is being displayed press the leftmost button. This will cycle through the list of items (Load
A, Peak A, Vall A, Load B, Peak B, Vall B, Pos, Vel, and Limits, optionally Avg A, Avg B).
The rightmost button will change the units that are being used. Repeated presses of this button will cycle through
the list of units available for the currently displayed item. For example, with Load showing you would see Lb, kg,
N, PSI, MPa, Klb, kN, t, mV/V, g and then back to Lb.
Message Meaning
Load A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Peak A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Valley A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Load B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Peak B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Valley B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Position Inches, Centimeters, Percent of Base Length.
Velocity Inches per Minute, Centimeters per Minute.
Limits Shows state of each switch, 1 for on, 0 for off, dash for disabled, asterisk for source error.
Avg A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Avg B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
All of these measurements are acquired by the Model 4215 at all times. The display just selects which particular
item you wish to see and what units are used to display or print this data.
The row of buttons have the functions indicated by the upper set of labels (Item, Tare, Reset, View, Print, Unit).
Each of these is described below.
Item Button
The Item button cycles through the list of different items that can be displayed (Load A, Peak A, Vall A, Load B, Peak
B, Vall B, Pos, Vel, Limits, and optional Avg A and Avg B). Use the Unit button to change the units that are used for
the currently displayed item.
Tare Button
The Tare button sets the current load to zero for channel A, channel B, or channels A and B. A submenu will allow
you to select what channel you want to tare. It does not matter what is being displayed when you press this
button. The load is always reset to zero. A single channel unit will automatically set Load A to zero.
Reset Button
The Reset button brings up a list of things that you might want to set to zero:
Pressing either the “+” button or the “>“ button will scroll forward through this list. Pressing either the “-” button
or the “<“ button will scroll backward through the list. Pressing “Enter” will reset the currently displayed item.
Pressing “Esc” will return to Run Mode without resetting anything. Note that all menu lists on the Model 4215
start with the “>“ symbol and that the Esc button will always back out of a menu without making any changes.
View Button
The View button switches between two virtual displays stored in the memory. Each of these displays will
remember the units you have selected for each item, and which item is currently being shown. This may save
you from scrolling through the list of items if there are two particular results that you often use. For instance you
might set the display to show Load in Lb, press View, and then set the display for Peak in PSI. Repeated presses of
the View button will now switch back and forth between Load in Lb and Peak in PSI with a single button push. If
you choose to have the second line show a display, then the View button toggles between the display currently
being activated. The activated display will have a “>” symbol to the left of it.
Print Button
T The Print button will send a set of data to the printer. This will contain the current readings of load, peak, valley,
position, velocity, and limit status. The units used for printing will be the units last selected on the numerical
display.
The printer must be a serial printer that supports ESC/P protocol. Many Epson printers with serial ports are known
to be compatible.
Unit Button
The Unit button cycles through the list of units that are available for the item that is currently being displayed.
Setup Mode
Setup mode is used to change the setup of the Model 4215. To enter setup mode press the two end buttons at the
same time ( “<“ and “>“ ). The row of four buttons will now function as indicated by the lower set of labels (Plus,
Minus, Enter, and Escape).
If the optional keylock is in the locked position you will see “>Setup is Locked” when you try to enter setup mode.
This prevents accidental or unauthorized changes to the setup of the unit.
To exit from setup mode and return to run mode press the escape button (Esc). When you are entering numerical
data or selecting options from a sub-menu the escape button will back you out without making any changes.
The main menu in setup mode contains the following 6 entries which will be described in detail in the following
sections. Note that all main menu items begin with the “>“ symbol and that all submenu items begin with the
“>>“ symbol.
• > User Data Entry -- allows you to set any common math options such as the base area used to display load in
PSI or the base length used to display position in percent.
• > Analog Output -- is used to select the analog channel, source, and scaling for the analog output channels.
• > Sensor Select -- lets you manually set which load cell calibration data is used. If you use auto-id load cells you
will never need this menu.
• > Calibration -- provides access to calibration for load cells, extension sensor, or optical encoders.
• > Limits -- allows you to set the source and options for the 4 contact closure switches.
• > System Options -- contains entries to set the printer baud rate, enable or disable automatic sensor identification,
enable or disable automatic load zeroing, and to set the serial communications address, baud rate, and line feed
options.
• > Display Options -- lets you choose the filtering that is applied to the load measurements, the number of
decimal digits that are shown on the display and on the print, the count by values for channels A and B (1, 2, 5,
10, or 20), and whether the second line of the display shows the limit switch status, a blank line, another display,
or text.
The Plus or Right button ( “+” or “>“ ) will scroll forward through this menu. The Minus or Left button ( “-” or “<“ ) will
scroll backward through this menu. The Enter button selects the currently displayed menu item. The Esc button
will return you to Run Mode.
Entering Numerical Data
At times you will need to enter numerical data into the Model 4215. The method is consistent in all cases. A
number is presented on the left side of the display with an underline cursor.
• Left ( “<“ ) moves the cursor left one position. If you are on the leftmost digit a new zero will be added. Thus
10.000 will become 010.00 where the cursor has moved from the tens place to the hundreds place.
• Right ( “>“ ) moves the cursor right one position. If you are at the far right a new decimal will be added if there
is room. Thus 12.34 will become 12.340 where the cursor has moved from the tenths to the hundredths place.
• Plus ( “+” ) increments the digit above the cursor. Thus 12.34 becomes 13.34. Note that 9 will “wrap around” to 0.
• Minus ( “-” ) decrements the digit above the cursor. Thus 12.34 becomes 11.34. Note that 0 will “wrap around” to
9.
• Plus and Minus (both “+” and “-” at the same time). This will change the sign of the number. Thus 12.34 becomes
-12.34. Note that cursor position does not matter when you change the sign.
• Enter accepts the currently displayed number.
• Esc escapes from numerical entry and discards any changes that have been made.
>>Base Area Ch A
This is the area in square inches that is used to calculate load in pounds per square inch (PSI) or Mega Pascals
(MPa).
>>Base Length
This is the length in inches that is used to calculate position in percent.
1. Channel: -- This entry allows you to select the analog output channel A or B. (use either the left/right or plus/
minus buttons) for the selected source data that will drive the analog output. Press enter to select the displayed
Analog channel.
2. Source: -- This entry allows you to scroll through a list (using the left/right or plus/minus buttons) to select the
source of the data that will drive the analog output. The standard list will include load, peak, valley, position, and
velocity. Press enter to select the displayed item.
3. Units: -- This entry allows you to scroll through the list of available units for the data source you have chosen
in the previous step.
4. Full -- This entry specifies the value of source data (in the selected units) that will cause the Model 4215 to
output positive full scale analog voltage (+10.0 volts).
5. Zero -- This entry specifies the value of source data (in the selected units) that will cause the Model 4215 to
output zero analog voltage (0.0 volts).
Once you have entered these 4 selections the analog output is setup.
>Sensor Select
This menu is used to manually select load cell calibration data from a list stored in the Model 4215. When you
select this menu on a 2 channel unit, you are asked to select Channel A or B. The load cell serial number of the
selected channel is displayed. Use either the left/right or plus/minus buttons to page through the list. The escape
button will return to the setup mode main menu without changing the load cell that is selected. The enter button
will select the currently displayed load cell and briefly display the rated load, last calibration date, and excitation
voltage.
To add new load cells to this list use the Calibration entry on the setup mode main menu described below.
If you are using auto-id load cells this menu will allow you to see which load cell is selected but will not allow you
to change the selection. This ensures that the correct calibration information is always used when you use
Auto-ID load cells.
Sensor Delete
The Model 4215 can store calibration data for 4 different load cells / sensors (25 slots optional). To delete a load
cell from this list choose “>Sensor Select” from the setup mode main menu. Press the left/right or plus/minus keys
until the serial number of the cell you wish to delete is displayed. (You may need to shut off auto-id if it is on, see
the System Options menu for details). Then press both the plus and minus keys at the same time. The Model 4215
will ask “Delete Sensor?” to be sure that you intend to delete this calibration data. Pressing Enter will delete the
cell data from the list. Pressing Escape will return to the load cell select without changing the list.
>Calibration
This menu is used to access all methods of calibration that are supported by the Model 4215. Selecting calibration
from the setup mode main menu will bring up a submenu. This submenu is used exactly like the main menu.
Press left/right or plus/minus to change the menu item. Press escape to return to the setup mode main menu.
Press enter to select the displayed menu item.
If you are re-calibrating a load cell, the Model 4215 will recognize it using either the autoid or a manually entered
serial number. Any time you overwrite existing load cell calibration data you will first be warned with the “Overwrite
Data?” message. Pressing enter will overwrite; pressing escape will cancel the calibration. Another method of
overwriting existing data is to delete the sensor first and then calibrate.
If a load cell with a new serial number (or new auto-id) is calibrated the data is added as a new entry in the load
cell list.
>>Cal Check
Selecting this item will run a calibration check on the load cell / sensor attached to Channel A or B. The Model
4215 will measure the shunt value of the load cell and compare it to the shunt value that was recorded when the
load cell was last calibrated. The rated load and last calibration date for the load cell are displayed followed by the
currently measured shunt value and the shunt value that was recorded when the load cell was calibrated.
Note that the Model 4215 has a switch on the back panel that allows you to select either a 30K shunt or a 60K
shunt. Be sure that this switch is set correctly for the load cell that you are using.
Since this shunt check electronically applies approximately half the rated load, it is important that the load cell /
sensor is not loaded when this check is run. If you see the “Cell Overloaded” message, remove any load on the cell
and try the check again.
>>Cal by mV/volt
This item is used to calibrate a cell using the manufacturer’s milli-volt per volt calibration factor or a series of
milli-volt per volt points with corresponding masses (usually listed on the load cell calibration certificate). Before
you enter in the desired information, another menu will ask if you want to calibrate by milli-volt per volt or 5 Point
milli-volt per volt. You will be asked for the channel, the cell serial number, the current date, the calibration unit,
the cell’s maximum rated load, and the desired excitation voltage (5.0 or 10.0 volts). If milli-volt per volt calibration
was selected, you are then prompted to enter the manufacturer’s milli-volt value. If 5 point milli-volt per volt
calibration was selected, you are prompted for each mass and milli-volt per volt value. One of these points should
be zero mass and zero milli-volts per volt. Enter each of these selections or press the escape button at any time
to return to the setup mode main menu without changing the calibration. The cell should not be under load
when this calibration method is used since a shunt value will be measured and recorded for later use with the “Cal
Check” feature described above. The “No Masses Ready” message is intended to remind you of this fact.
>Limits
This menu is used to set up the 4 contact closure switches. Instead of a submenu there are a series of selections.
Pressing Enter will accept the displayed entry. Pressing Escape will return to the setup mode main menu without
changing the limit setup.
1. Select Limit X -- The first selection chooses which of the limits you want to set up (1-4). Use the left/right or
plus/minus buttons to change the number. Press enter to accept the displayed number. Press escape to return
to the setup mode main menu.
2. Normally Open / Closed -- This selection determines if the switch contacts are open or closed when the
limit is de-activated.
3. Enabled / Disabled -- Choosing enable allows the limit to operate as specified by the following selections.
Choosing disable places the limit in its “normal” state and completes the limit setup.
4. Source: -- This selection allows you to scroll through a list (use either the left/right or plus/minus buttons)
to select the source of the data that will drive the limit. The standard list will include load, peak, valley, position,
velocity and, optionally, average force. Press enter to select the displayed item.
5. Units: -- This entry allows you to scroll through the list of available units for the data source you have chosen
in the previous step.
6. Set -- This is the numerical value of the source at which the limit will be activated (set point). It is specified in
the units you chose in the previous step.
7. Trip if source < / > Set -- Choosing source>set means that the limit will be activated when the source
becomes larger than the set point. Choosing source<set means the limit will be activated when the source
becomes less than the set point.
8. Latching on/off -- Selecting latching off means that the limit will be de-activated automatically when the
source passes the reset point (see next selection). Selecting latching on means that once the limit is activated you
must manually reset it using the Reset button. If you select latching on then the limit setup is done.
9. Rst -- This selection determines the numerical value of the source at which the limit will be deactivated (reset
point). If you selected “source>set” in step 7 then the limit is deactivated when the source is less than the reset
point. If you selected “source<set” in step 7 then the limit is deactivated when the source is greater than the reset
point
Notes:
The Set and Reset conditions for each limit are evaluated independently. If both conditions are met then Reset
will take priority over Set.
Limit status may show an asterisk if the limit is enabled but there is an error in the signal source. For example,
if the limit source is load but there are no load cells currently calibrated, or the load cell has a rating of 0.0 Lb, or
a millivolt per volt constant of zero. Another example would be if the limit source is position and zero has been
entered for the encoder counts-per-inch constant. Limits that are in this error condition will be forced to the off
position as if they were disabled.
>System Options
This menu contains some system wide parameters that are seldom changed. These are listed in a submenu that
works exactly like the setup mode main menu.
>>Printer Baud
The front panel print button and the RS232 print commands will use this baud rate for communication with your
printer. This setting has no effect on units that are not equipped with the printing option. 57.6 K baud is commonly
used on older serial printers 230.4 K baud is used on more recent serial printers.
>>Auto Identify
This allows you to turn automatic load cell identification on or off. Usually you can leave auto-id on even if you
never use automatic load cell identification. In order to use this function you must either have an auto-id equipped
load cell or use an in-line auto-id module attached to your load cell cable.
>>Com Address
This item allows you to set the address of this unit used for the RS232 ASCII command set. Valid values are 1
through 254. If the RS232 multi-drop option is installed you must be sure that all the attached units have different
addresses to prevent data collisions.
>>Com Baudrate
This sets the baud rate for the RS232 ASCII command set. Available values are: 300, 600, 1200, 2400, 4800, 9600,
and 19.2k. Note that the “Special” entry is reserved for future use.
>>Retain Tare
This controls whether the most recent tare value is retained through power down of the unit and used again
immediately on power up. Choosing this option will retain tare on channel A in a single channel unit and both
channels A and B in a twochannel unit.
>>Display Options
This menu contains options that effect how load, peak, and valley are displayed and printed. These are listed in a
submenu that works exactly like the setup mode main menu.
The settling time for each of the 4 levels of filtering is shown in the table below
Level 1 2 2 4
Setting Time (seconds) <1 2 10 30
Read Rate (per second) 60 60 30 10
>>Decimals Ch A/B
This lets you set the maximum number of digits to the right of the decimal that are displayed or printed for
load, peak, and valley. When large load cells are selected the number of decimal digits shown may be limited, for
example a 100 Lb cell will allow 3 digits, while a 1000 Lb cell will only allow 2 digits. Also if a number is too large to
fit on the display some decimal digits may be dropped (the display will “auto-range”). The number of digits shown
for position data depends only on the number of counts per inch of the encoder and is not effected by this setting.
>>Second Line
This option allows you to choose what you want to have displayed on the second line. You can select limits, select
a display, select text, or leave the line blank.
5. Remote Operation
Digital Inputs
There are 4 digital inputs available on the Model 4215. They are accessible on the Digital I/O connector on the
back panel (pins 9 - 16). These inputs are individually optoisolated and include current limiting resistors. An input
voltage anywhere from +4 to +22 volts DC may be used to obtain the “on” state. See Appendix B for a full description
of the Digital I/O connector. If you do not require isolation for your application you can use the +5 or +12 volt and
GND available on the Digital I/O connector (pins 24-26).
These 4 inputs are called IN1 through IN4 and have the following standard functions:
Other functions may be assigned to these inputs by custom programming at the factory.
RS232 Operation
All the features and settings of the Model 4215 can be accessed remotely using the RS232 ASCII command set.
These commands fall into two groups: commands that are like Run Mode, and commands are like Setup Mode.
The Model 4215 supports remote operation using a standard RS232 Interface consisting of ASCII characters. The
baud rate is set using the Com Baudrate item on the System Options menu. Supported rates are 300, 600, 1200,
2400, 4800, 9600, and 19.2k baud. There must be 8 data bits, no parity, and 1 stop bit. To support the multi-drop
RS232 option each unit is assigned a communications address which is set using the Com Address item on the
System Options menu. All the commands have the following format:
@123XYZ<CR> The @ symbol initiates the command, all commands must begin with this symbol. The number
123 is the address of the Model 4215 which should respond to this command. This MUST be 3 digits so type 001
for unit one, and 026 for unit twenty-six. The command itself will be a sequence of letters such as the XYZ shown
above. If the command requires a number as one of its parameters the number must be followed by the pound
sign “#”. Numbers that are less than one should include a leading zero so you should type 0.123# to enter
one-hundred twenty three thousandths. Commands are not processed until a carriage return <CR> is sent.
No Model 4215 will respond to address 000. All Model 4215’s will respond to address 255. Throughout the rest of
this section we will assume that you are using address 123.
Whenever an Model 4215 executes a remote command it will issue an acknowledge that includes the address and
any required data.
Each entry will include the name of the command, followed by a short description that includes the command
format, an example that shows what you would type (shown in plain font) and what the Model 4215 would return
(shown in italics).
Example:
@123?
Acknowledge:
@123 These are the Item numbers:
00 - Load A 01 - Peak A 02 - Vall A 03 – Load B
04 - Peak B 05 – Vall B 09 - Pos 10 - Vel
13 – Limits 14 – Avg A 15 - Avg B
These are the units for Load, Peak, and Valley:
00 - Lb 01 - kg 02 - N 03 - PSI
04 - MPa 05 - Klb 06 – kN 07 - t
08 – mVv 09 - g
These are the units for Position:
00 – In 01 – Cm 02 - %
These are the units for Velocity:
00 – I/M 01 – C/M
Example:
@123FS0200
Acknowledge:
@123 Active Display shows Vall A in Lb
Other Display shows Peak A in Lb
Example:
@123FA
Acknowledge:
@123 Active Display shows Peak A in Lb
Other Display shows Vall A in Lb
Example:
@123V01021
Acknowledge:
@123 Peak A 55.676 N
Example 2:
@123V5000002
Acknowledge 2:
@123 Load A 120.45 Lb Load B 99.02 Lb
@123 Load A 125.29 Lb Load B 99.02 Lb
.....
@123 Load A 137.81 Lb Load B 99.02 Lb
Example3:
@123V01021
Acknowledge3:
@123 Peak A 55.676 N
Example 4:
@123V00002
Acknowledge 4:
@123 Load A 120.45 Lb
@123 Load A 125.29 Lb
.....
@123 Load A 137.81 Lb
Example:
@123P1
Acknowledge:
@123 Load A 45.625 Lb
Peak A 82.396 Lb Vall A -0.005 Lb
Load B 22.560 Lb
Peak B 23.000 Lb
Vall B –0.0001 Lb
Pos 3.500 In
Vel 0.0 I/M
Avg A 0.000 Lb
Avg B 0.000 Lb
Limits 0 0 0 1
Example:
@123R1000000
Acknowledge:
@123 Reset – Tare A
Example:
@123TStress Test Ready
Acknowledge:
@123 Text Displayed - Stress Test Ready
If the optional Keylock is installed and locked you will not be to alter the setup until the unit has been unlocked.
The acknowledge will indicate this fact:
@123 Setup is Locked - use Keylock to unlock
CB1(space)(channel)(serial number)#
CB2(space)(month)(day)(year)
CB3(space) (excitation)(calibration unit)
CB4(space)(rated load)#
where channel is A or B, serial number is the serial number of the load cell, not exceeding 6 digits, month, day, and
year are 2 digit numbers to represent the current date, excitation should be a 0 for 5.00 volts or a 1 for 10.00 volts,
and calibration unit is the desired unit for this calibration, and rated load is the maximum load rating for the load
cell.
Example 1:
@123CB1 A123456# 1000.0#
Acknowledge 1:
@123 Calibrate Begin 1 Command – New
Load Cell S/N: 123456 – Channel A
Example 2:
@123CB2 042298
Acknowledge 2:
@123 Calibrate Begin 2 Command – New Cal Date: Apr22-98
Example 3:
@123CB3 100
Acknowledge 3:
@123 Calibrate Begin 3 Command – New
Excitation Voltage: 10.0 V, Calibration Unit: Lb
Example 4:
@123CB4 1000.0#
Acknowledge 4:
@123 Calibrate Begin 4 Command – New Rated Load: 1000.0 Lb
Example:
@123CV4.5002#
Acknowledge:
@123 Calibrate Command - Reading for Shunt Check...
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 1000.0 Lb , 4.50020 mV/v, 10.00 V , Cal on Oct27-99, 553.26 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
Example:
@123CMV5
Acknowledge:
@123 Calibrate by mV/Volt - 5 Point Ready for Mass CMVM1 command
Example:
@123CMVM15.0005#
Acknowledge:
@123 Calibrate Mass 1 Command entered
Ready for mV/V Value CMVV1 or CE command
The CE or Calibration Escape command will, if entered, end this calibration without calibrating the unit. The unit
will remain calibrated according to the most recent calibration or, if this is the first calibration of the unit, the unit
will remain uncalibrated.
At this point the unit is prompting the user for the first milli-volt per volt value of the calibration. The mV/V value
is entered using the Calibration by mV/V Volt Command with the following format: CMVV1(mV/V value)#.
Example:
@123CMVV11.9926#
Acknowledge:
@123 Calibrate mV/V 1 Command entered
Ready for Mass Value CMVM2 or CE command
The unit will continue to prompt the user for mass and milli-volt per volt values until the five points have been
entered. Then the user will see the following acknowledgement:
The user must now enter the CMVM0 command to finish the calibration. Note the load cell should have no masses
on it at this point since a shunt calibration is run.
Example:
@123CMVM0
Acknowledge:
@123 Calibrate Command - Reading for Shunt Check…
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 1000.0 Lb , 4.50020 mV/v,
4.50020 mV/v,
4.50020 mV/V,
4.50020 mV/V
4.50020 mV/V
5.00 V , Cal on Oct27-99, 553.26 Lb Shunt
Example:
@123CM5
Acknowledge:
@123 Calibrate by Mass – 5 Point Ready for CMP1 command
Example:
@123CMP10.0#
Acknowledge:
@123 Calibrate Mass 1 Command - Reading...
[10 seconds delay]
Calibrate Mass 1 Command - Ready for CMP2 or CE command
The Calibrate Mass 2 (CMP2) command has the format CMP2(mass)#. This command can only be issued after the
CMP1 command has completed. The example uses 500 Lb for the second calibration mass. Be sure you have
placed the mass on the load cell before entering this command since reading will begin immediately.
Example:
@123CMP2500#
Acknowledge:
@123 Calibrate Mass 2 Command - Reading...
[10 seconds delay]
Calibrate Mass 2 Command - Ready for CMP3 or CE command
Masses three through five are done in a similar manner. The Calibrate Mass Point 0 (CMP0) command completes
the 2 or 5-point mass calibration. Remove the masses from the load cell before entering this command since it will
perform the shunt calibration measurement that is recorded for use with the calibration check command.
The format is CMP0.
Example:
@123CMP0
Acknowledge:
@123 Calibrate Command - Reading for Shunt Check...
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v,
3.00230 mV/v,
3.00238 mV/v,
3.00353 mV/v,
10.00 V , Cal on Oct27-99, 49.532 Lb Shunt
unused S/N 13368, 10.00 Lb , 3.00121 mV/v, 10.00 V, Cal on Oct27-99, 4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v, 10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v, 5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
The Calibration counts per Inch View (CIV) command displays the current setting of the encoder counts per inch
constant.
Example:
@123CIV
Acknowledge:
@123 Encoder is 32000 counts per inch
The Calibration counts per Inch Set (CIS) command allows you to change the current setting of the encoder counts
per inch constant.
Example:
@123CIS64000#
Acknowledge:
@123 Encoder is 64000 counts per inch
This line indicates that Limit 1 is a normally open switch (NO) rather than a normally closed switch (NC). The limit
is Enabled rather than Disabled. The signal driving this limit is Load in Lb. The limit will be turned on when the
load exceeds 20.00 Lb. It is not a Latching limit (latching requires a manual reset). The limit will automatically reset
when the load is less than 1.00 Lb.
The limit number chooses which limit (1-4) you want to setup. The normal position should be set to 0 for normally
open or 1 for normally closed. The enable should be set to 0 for disabled or 1 for enabled. If the limit is disabled
the rest of this command may be omitted (see example 3). The item and unit numbers indicate which signal is to
drive this limit. The set point number indicates the signal level at which the limit is to be activated. The < or > sign
indicates whether you want the limit to be activated when the signal is greater than the set point (>) or less than
the set point (<). The latching should be set to 0 for latching off or 1 for latching on. If latching on is selected the
rest of the command may be omitted (see example 2). Latching on will require a manual reset once the limit has
been activated. The reset point indicates the signal level at which the limit should be de-activated.
If you want to enable a limit without otherwise changing its settings you can enter L(limit number)SA(space)
(normal position)1# instead of entering the full setup string (see example 4).
Example 1:
@123L1SA 010000
Acknowledge 1:
@123 Limit Setup Command A – Ready for Command B
Example 2:
@123L1SB 50.0#
Acknowledge 1:
@123 Limit Setup Command B – Ready for Command C
Example 3:
@123L1SC >0
Acknowledge 3:
@123 Limit Setup Command C – Ready for Command D
Example 4:
@123L1SD 10.0#
Acknowledge 4:
@123 Lim 1 NO Enabled Load A Lb Set 50.00 Trip>Set Latch Off Reset 10.00
Example 5:
@123L1SA 10
Acknowledge 5:
@123 Lim 1 NO Disabled Load A Lb Set 50.00 Trip>Set Latch Off Reset 10.00
Example 6:
@123L1SA 11#
Acknowledge 6:
@123 Lim 1 NC Enabled Load A Lb Set 50.00 Trip>Set Latch Off Reset 10.00
Code Baudrate
4 4800
5 9600
6 19.2K
8 57.6K
9 230.4K
Example:
@123OP9
Acknowledge:
@123 Printer Baud Rate is 230.4K
Code Baudrate
0 300
1 600
2 1200
3 2400
4 4800
5 9600
6 19.2K
7 Special
Example:
@123OT1
Acknowledge:
@123 Retain tare is on
Code Count By
0 1
1 2
2 5
3 10
4 20
Example:
@123DTESL Sense I
Acknowledge:
@123 Text Message – Model 4215
The following procedure should be followed to calibrate the Model 4215 itself. This calibration requires the use
of a precision millivolt per volt reference which will be attached on the load cell connector. The resulting scale
factors are recorded in the non-volatile memory of the Model 4215.
ATTENTION: System Calibration should ONLY be performed by trained, QUALIFIED TECHNICIANS utilizing the
proper precision devices to ensure the most accurate performance of the unit.
1. To enter system calibration mode, turn off the Model 4215. Hold down the two end buttons ( “<“ and “>“ ) while
turning power back on. As soon as the display comes on you can release the buttons.
2. The unit will go through its regular start up sequence but rather than entering run mode and showing the load
in pounds a password prompt will be displayed. Enter the number 777 as the system calibration password.
3. The unit will now display “Unit mV / V Cal” to indicate that it is ready to be calibrated. A five minute warm-up
period with the millivolt-per-volt reference attached is recommended at this time if the unit has just been turned
on or the millivolt-per-volt reference has just been attached. Press any button to clear this display.
4. You are now prompted to enter the first of two millivolt per volt references. (We recommend using -3.0 mv/volt
for the first and +3.0 mv/volt for the second.) Set your reference and enter the appropriate number. As soon as
you press Enter the Model 4215 will start reading the reference.
5. Next you are prompted for the second mv/volt value. Set your reference and enter the appropriate number.
Again reading will start as soon as Enter is pressed.
6. When the read is complete, the unit will display “Calibration Done”. Pressing any button at this time will place
the unit back into its normal run mode
When first selected after power up, the value displayed for average force will be zero. To use the features, the
user must activate the isolated input 2, IN2, of the Digital I/O port connector (pins 11 and 12 – see Digital I/O
Connector in Appendix B). Once the Digital Input has been activated, the unit will begin to display a running
average of the force measured after an approximately 2 second delay. The item “Avg A” refers to the average force
measured, calculated, and displayed on channel A, and “Avg B” refers to the average force measured, calculated,
and displayed on channel B. Opening the Digital Input circuit will stop the averaging.
The final average value will remain on the display until the input is again activated or the unit is powered down.
If you do not require isolation for your application, you can use the +12 volt source and an available GND signal
located on the Digital I/O connector (pin 26 and pin 24) to create a circuit that only needs contact open/closure to
activate and deactivate the digital input. See the application diagram below.
Sensor Connectors
DB9 Female
Pin Signal
1 EXCITE - HI
2 SENSE - HI 1
3 CELL OUTPUT - HI
4 CELL OUTPUT - LO
5 SENSE - LO 1
6 EXCITE - LO
7 AUTO ID - A
8 AUTO ID - B
9 CHASSIS GND
Notes:
If the sense lines are not used, SENSE - HI must be tied to EXCITE - HI and SENSE - LO must be tied to EXCITE - LO.
Pin Signal
1 TXD
2 RXD
3 No Connect
4 SERIAL GND
5 TXD+ (Printer)
6 TXD- (Printer)
7 RXD- (RS485)
8 RXD+ (RS485)
9 No Connect
10 SERIAL GND
11-13 No Connect
14 TXD+ (RS485)
15 TXD- (RS485)
Notes:
1. Pins 1,2 and 4 are used for RS232 ASCII command set communications. Use 8 data bits, no parity, 1 stop, and set
the baud rate using the Com Baudrate entry on the System Options menu.
2. The printer uses RS422 differential signaling on pins 5, 6 and 10. Baud rate for the printer is selectable using the
Printer Baud entry on the System Options menu.
3. Pins 7, 8, 14 and 15 are used for RS485 ASCII command set communications. Use 8 data bits, no parity, 1 stop,
and set the baud rate using the Com Baudrate entry on the System Options menu.
Pin Signal
1 LIM 1 A 1
2 LIM 1 B 1
3 LIM 2 A
4 LIM 2 B
5 LIM 3 A
6 LIM 3 B
7 LIM 4 A
8 LIM 4 B
9 ISO IN 1 - HI 2
10 ISO IN 1 - LO 2
11 ISO IN 2 - HI
12 ISO IN 2 - LO
13 ISO IN 3 - HI
14 ISO IN 3 - LO
15 ISO IN 4 - HI
16 ISO IN 4 - LO
17 KEYLOCK 3
18 GND
19 ENCODER CHA-L 4
20 ENCODER CHA
21 GND
22 ENCODER CHB-L
23 ENCODER CHB
24 GND
25 GND
26 +12 VDC (+5 VDC optional 4)
Notes:
1. Limits are optically isolated solid state switches that can control AC or DC voltage. Peak blocking voltage is
350V, maximum continuous current is 120 mA.
2. Digital inputs are opto-isolated with internal current limiting resistors. Voltage range is +4 to +22 VDC on the
“HI” input with respect to the “LO” input. Other voltage ranges are possible with added external resistors, contact
Electro Standards Laboratories for details. If isolation is not required these inputs may be operated using the +12
VDC supplied on pin 26 and the GND on pin 24.
3. The keylock input will be in the “unlocked” state if left unconnected. Tie this pin to GND to “lock” it.
When locked the setup of the Model 4215 cannot be changed (you cannot enter Setup mode).
4. The encoder inputs are intended for use with a +12 VDC quadrature encoder. For single ended signaling use the
CHA and CHB pins, the CHA-L and CHB-L pins are provided for differential signals. The encoder can be supplied
with the +12 VDC and GND available on pins 26 and 24.
The Model 4215 will also accept +5 VDC encoder inputs but these MUST be differential. Single ended +5 VDC
inputs are not supported on the standard unit. Options for +5 VDC encoders are available. Pin 26 will supply +5
VDC for this option and must be ordered accordingly.
Pin Signal
2 ANALOG - OUT 1
3 ANALOG - GND
4 HIGH BANDWIDTH - OUT 2 CHANNEL A
5 4-20 mA – OUT 3 CHANNEL A
6 2nd ANALOG-OUT1 (Optional)
7 4-20 mA – OUT 3 CHANNEL B
8 HIGH BANDWIDTH - OUT 2 CHANNEL B
9 No Connect
Note:
1. The source and scaling of ANALOG - OUT is selected through the menus.
2. The HIGH BANDWIDTH - OUT is a buffered analog output permanently connected to the load cell channel.
3. If the 4-20 mA Option is purchased.
The back panel switch (shown to the left of the two resistors) selects between 30 KΩ and 60 KΩ.
The shunt check relay (shown to the right of the two resistors) connects the selected resistor between the Excite
High and the Cell Output High.