Starting The System: Keysight I3070 Series 5 In-Circuit Test System

This document describes starting up and shutting down the Keysight i3070 Series 5 In-Circuit Test System. It includes the following steps: 1) Turn on peripherals and the controller and allow it to boot. Log into Windows using an assigned username and password. 2) Log in as "calibrate" to open diagnostics and boot the testhead, which takes 2-5 minutes. 3) To shut down, first unboot the testhead, then shut down Windows, and finally turn off power to the controller.

Uploaded by

Hà Lê
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
228 views5 pages

Starting The System: Keysight I3070 Series 5 In-Circuit Test System

This document describes starting up and shutting down the Keysight i3070 Series 5 In-Circuit Test System. It includes the following steps: 1) Turn on peripherals and the controller and allow it to boot. Log into Windows using an assigned username and password. 2) Log in as "calibrate" to open diagnostics and boot the testhead, which takes 2-5 minutes. 3) To shut down, first unboot the testhead, then shut down Windows, and finally turn off power to the controller.

Uploaded by

Hà Lê
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Keysight i3070 Series 5 In-Circuit Test System

Starting the System

Overview of i3070 Series 5 System 2


Starting the i3070 Series 5 System 3
Shutting Down 5

This guide describes the i3070 Series 5 In-Circuit Test System (i307x, i317x, and
i327x).
Overview of i3070 Series 5 System
Figure 1 i3070 Series test stations
i307x 4-module ICT system
i327x Ultra series ICT system
fixture
connector for
anti-static wrist strap

Emergency
Shutdown Switch

testhead
rotation switch

controller pod
controller
testhead testhead (testhead rotation
switch in pod)

Table 1 Test station controls

Item Description
Emergency Shutdown Press the Emergency Shutdown switch to power down the testhead
Switch and its associated equipment in an emergency situation.
It turns off all AC power to the testhead, and is equivalent to turning off
the main circuit breaker on the rear of the system.
Main Power Switch Controls the incoming line voltage and is located at the rear of the
system.
Testhead Rotation Rotates the testhead into any position from 0 to 90 degrees from the
Switch horizontal.

Table 2 Operator controls

Item Description
Operator Interface Graphical user interface lets the operator load and run testplans.
Operator footswitch Functions like the START button in the operator screen; can be used
when the operator’s hands are busy.
Bar code scanner Used to save time and minimize errors when entering board serial
numbers.
Guided probe Used by the operator for manual probing of circuits.

2 Starting the i3070 Series 5 System


Starting the i3070 Series 5 System

Boot the Testhead Controller


1 Turn on the monitor and any other peripherals.
2 Turn on power to the controller.
3 Allow the controller to boot.
During the boot process, the system:
• tests and initializes hardware components,
• loads the operating system,
• starts log on and other system services, and
• connects to the LAN.
Read the messages displayed on the screen. These messages could be
important in solving system administration problems.
The controller boot process is complete when the Windows log-on screen
appears.

Log On to the System


The i3070 In-Circuit Test Software requires that decimals in the board file use a
period (.) and not any other character such as a comma.
Make use that Windows is set to use a period for decimals:
Select Control Panel > Regional and Language Options > Regional Options tab.
Click Customize; in the Numbers tab ensure that the selected Decimal symbol
is a period.

You will need the following to log on to the system:


• A Windows user name, which can be a personal user name assigned by the
System Administrator, or one of the standard user names supplied with the
test system (below). Names and passwords are case-sensitive.
• A user name to access the Developer Interface (UnMux system only) or
Engineer Test Interface.

Table 3 i3070 system users

User Name Defaul t Password Definition and Permissions


operator Resetme1 For operators who are using the Operator Interface to test
boards.
tcm Resetme1 For operators who are using BT-Basic to test boards.
user Alterme1 For general users or sustaining engineers.
engineer Alterme1 For test developers or sustaining engineers.
calibrate Keysight1 Used for system calibration and maintenance. This user
has administrative rights.

Starting the i3070 Series 5 System 3


To log on at the testhead controller:
1 Log on to Windows using your assigned user name. Type your password and
press Enter.
2 If you have used a user logon, the Operator window is displayed and you can
begin using the test system.
3 If you have used an engineer logon, you can launch the test development
software using the icons on the desktop.
To log on at a test development workstation:
1 At a developer’s workstation, log on to Windows.
2 Launch the test development software using the icons on the desktop.

Boot the Testhead


1 Log on as calibrate.
2 Open a Korn Shell window and type dgn at the command line to run the
Diagnostics program.
3 When the DGN interface appears, select the Testhead Functions softkey.
4 Boot the testhead using the Testhead Power On softkey.
Booting takes about 2 to 5 minutes, depending on the number of testhead
modules in your system.

4 Starting the i3070 Series 5 System


Shutting Down
Shut down the system before removing power to perform tasks such as installing a
new disk or interface card.
Unboot the testhead before shutting down the controller to
prevent damage to the software.

Perform the following procedures in order:


1 Unboot the testhead:
a Log on as calibrate.
b Open a Korn Shell window and type dgn at the command line to run the
Diagnostics program.
c When the DGN interface appears, select the Testhead Functions softkey.
d Unboot the testhead using the Testhead Power Off softkey.
2 Shut down the controller:
Select Shut Down from the Start menu.
3 Disconnect power to the controller:
Push the power switch on the controller.
Perform this step only after following the above instructions
to shut down the controller. Otherwise, software damage can
occur.

Starting the i3070 Series 5 System 5