User Manual METEK - Chandler Chanscope and Dew Point Tester
User Manual METEK - Chandler Chanscope and Dew Point Tester
User Manual METEK - Chandler Chanscope and Dew Point Tester
User Manual
Process Instruments
150 Freeport Road
PN 13-2270 REV P Pittsburgh, PA 15238
Offices
USA - HEADQUARTERS GERMANY
AMETEK Process Instruments AMETEK GmbH
150 Freeport Road Postfach 2165
Pittsburgh, PA 15238, USA D-40644 Meerbusch
Phone: 412-828-9040 OR
Fax: 412-826-0399 Rudolf-Diesel Strasse 16
D-40670 Meerbusch
Germany
USA - Delaware Phone: 49-21-59-9136-0
AMETEK Process Instruments Fax: 49-21-59-9136-39
455 Corporate Boulevard
Newark, Delaware 19702, USA
Phone: 302-456-4400 (Main) SINGAPORE
800-537-6044 (Service) AMETEK Singapore PVT. Ltd.
800-222-6789 (Ordering) 10 Ang Mo Kio Street 65
Fax: 302-456-4444 #05-12 Techpoint
Singapore 569059
Republic of Singapore
USA - Texas Phone: 65-484-2388
AMETEK Process Instruments Fax: 65-481-6588
9750 Whithorn Drive
Houston, Texas 77095, USA FRANCE
Phone: 281-463-2820 AMETEK - APIF
Fax: 281-463-2701 Rond Point de l’épine des champs
Buroplus Bat D
USA - Oklahoma 78990 Elancourt, France
Chandler Engineering Phone: 33 1 30 68 89 20
2001 North Indianwood Avenue Fax: 33 1 30 38 89 29
Broken Arrow, OK 74012, USA
Phone: 918-250-7200 CHINA
Fax: 918-459-0165 AMETEK Process Instruments
Room 406, Metro Tower
CANADA No. 30 Tian Yao Qiao Road
AMETEK Process Instruments Shanghai, 200030
2876 Sunridge Way N.E. Phone: 8621 6426 7049
Calgary, Alberta Fax: 8621 6426 7054
Canada T1Y 7H9
Phone: 403-235-8400
Fax: 403-248-3550
© 2005 AMETEK
This manual is a guide for the use of the AMETEK ChanScope and Dew Point Tester models. Data herein has been verified and validated and is
believed adequate for the intended use of this software. If the software or procedures are used for purposes over and above the capabilities specified
herein, confirmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or
liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.
CHAPTER 2 Specifications
Standard Bureau of Mines-type Dew Point Tester.............................................. 2-1
Accuracy............................................................................................. 2-1
Dew Point Temperature Range........................................................... 2-1
Operating Pressure.............................................................................. 2-1
Connections......................................................................................... 2-1
Weight and Dimensions...................................................................... 2-1
ChanScope Digital Dew Point Tester.................................................................. 2-2
Accuracy............................................................................................. 2-2
Dew Point Temperature Range........................................................... 2-2
Operating Pressure.............................................................................. 2-2
Connections......................................................................................... 2-2
Power.................................................................................................. 2-2
Weight and Dimensions...................................................................... 2-2
| iii
CHAPTER 4 Operation
Using Glycol Filters............................................................................................. 4-1
Determining Equilibrium.................................................................... 4-1
Test Precautions................................................................................................... 4-2
Test Procedure..................................................................................................... 4-4
Confirmation of Dew Point.................................................................................. 4-5
Interpretation of Dew Point Temperature........................................... 4-5
Differentiating between Water and Hydrocarbon Dew Points............................ 4-6
Special Test Procedure for Propane..................................................................... 4-7
CHAPTER 6 Drawings
DWG 13-0052 - Glass Window Assy (sour gas)................................................. 6-1
DWG 13-0054 - Glycol Filter Assy..................................................................... 6-2
DWG 13-0072 - Immersion Cooling Rod Assy.................................................. 6-3
DWG 13-0075 - Model 13-0075 Dew Point Tester............................................. 6-4
DWG 13-0090 - Chillers, Regular and Quick..................................................... 6-5
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (°F)..................... 6-6
DWG 13-0095 - Saturated Moisture Content Air or Nitrogen (°C).................... 6-7
DWG 13-0097 - Saturated Moisture Content Natural Gas (°F).......................... 6-8
DWG 13-0097 - Saturated Moisture Content Natural Gas (°C).......................... 6-9
Table - Equilibrium Water Vapor Contents........................................................ 6-10
DWG 13-0100 - Bureau of Mines Dew Point Tester......................................... 6-15
DWG 13-0125 - Conversion Chart, Dew Point to Moisture Content................ 6-16
DWG 13-0150 - Installation, Dew Point Tester................................................ 6-17
DWG 13-2016 - ChanScope I............................................................................ 6-18
DWG 13-0303 - Illuminated Magnifier Assy (Model 13-075).......................... 6-19
DWG 13-0304 - Illuminated Magnifier (Models 13-100, 13-110).................... 6-20
DWG 13-0305 - Illuminator Assy..................................................................... 6-21
DWG 13-2017 - ChanScope II.......................................................................... 6-22
DWG 13-2227 - Eyepiece Assembly................................................................. 6-23
An operating procedure which, if not strictly observed, may result in personal injury or envi-
ronmental contamination.
An operating procedure which, if not strictly observed, may result in damage to the equip-
ment.
Electrical Safety
Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified
electrician should make electrical connections and ground checks.
Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection
originally provided by the equipment.
Grounding
Instrument grounding is mandatory. Performance specifications and safety protection are void if instru-
ment is operated from an improperly grounded power source.
| v
Warning Labels
These symbols may appear on the instrument in order to alert you of existing conditions.
Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste
system (residential trash). The Wheelie Bin marking on this product is a reminder to dispose of the prod-
uct properly after it has completed its useful life and been removed from service. Metals, plastics and
other components are recyclable and you can do your part by one of the following these steps:
• When the equipment is ready to be disposed of, take it to your local or re-
gional waste collection administration for recycling.
• In some cases, your “end-of-life” product may be traded in for credit towards
the purchase of new AMETEK instruments. Contact your dealer to see if this
program is available in your area.
We warrant that any equipment of our own manufacture or manufactured for us pursuant to our specifications which
shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal
use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such
defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18)
months from date of shipment or twelve (12) months from date of installation of permanent equipment, whichever
period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory,
or at such other location as we may designate, transportation prepaid. Such returned equipment shall be examined by
us and if it is found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced
as aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of
such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost of transportation. In
addition, instead of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back
the defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.
Process photometric analyzers, process moisture analyzers, and sampling systems are warranted to perform the
intended measurement, only in the event that the customer has supplied, and AMETEK has accepted, valid sample
stream composition data, process conditions, and electrical area classification prior to order acknowledgment. The
photometric light sources are warranted for ninety (90) days from date of shipment. Resale items warranty is limited to
the transferable portion of the original equipment manufacturer’s warranty to AMETEK. If you are returning equipment
from outside the United States, a statement should appear on the documentation accompanying the equipment being
returned declaring that the goods being returned for repair are American goods, the name of the firm who purchased
the goods, and the shipment date.
The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after leav-
ing our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect, abuse
or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but shall not
be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be
construed to include improper or inadequate protection against shock, vibration, high or low temperature, overpressure,
excess voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible medium, un-
less the equipment is specifically designed for such service, or exposure to any other service or environment of greater
severity than that for which the equipment was designed.
The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser
from us.
THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRANTIES OF FITNESS OR OF MERCHANT-
ABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LI-
ABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER
PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES,
CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL
PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR
OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE.
WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRAN-
TIES OF FITNESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT,
WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANUFACTURE. WE
DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY
IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.
| vii
This page intentionally left blank.
Description
When a gas-water vapor is cooled with no liquid water present, the absolute hu-
midity of the vapor content remains constant but the relative humidity increases
until it reaches 100% saturation, after which moisture begins to condense. The
temperature at which condensation is first detected is known as the dew point.
The Bureau of Mines-type dew point tester duplicates the conditions required
by the definition of dew point. However, the measurement method is modified
so that it is suitable for measuring dew point at high pressures. This is done by
enclosing a mirror upon which moisture or hydrocarbons condense (when chilled
by a refrigerant) in a pressure chamber. A window in the chamber allows viewing
of moisture condensation.
ChanScope
Both the standard (Bureau of Mines type) dew point tester and the ChanScope
include a sample pressure gauge and tripod socket for ease of use in the field.
Overview | 1-1
Theory of Operation
Water contamination in gases can cause corrosion and other equipment problems,
especially in regulation and metering stations. The determination of hydrocarbon
contamination in pipeline gas is necessary to meet hydrocarbon dew point speci-
fications in natural gas transportation contracts. Gases most frequently tested for
water content include natural gas, sour gas, heat treating furnace gas, and indus-
trial gases such as nitrogen, oxygen, hydrogen, and other bottled gases. The most
accurate and widely used instrument available for determining water content is
the Bureau of Mines-type chilled mirror dew point tester. When this instrument is
used properly, measured dew point accuracy is ± 0.2°F.
Each dew point tester incorporates a pressure-tight chamber to contain the gas or
vapor test sample and valves for controlling gas flow. There is a clear plastic win-
dow at one end of the chamber to observe the interior and the mirror. The mirror
itself is highly polished stainless steel and is attached at its center to a copper
thermowell. A chiller attached to the thermowell controls expansions of refriger-
ant and the temperature of the mirror. The greater the flow rate of refrigerant, the
lower the temperature. The mirror’s temperature is indicated by a thermometer
and uses a bulb that is located close to the back surface of the mirror. An elec-
tronic Resistance Temperature Detection (RTD) is used on some model dew point
testers.
Chiller
There are three types of chiller devices available for use with the standard dew
point tester or the ChanScope.
• Standard Chiller
• Quick Chiller
• Liquid Nitrogen Chiller
The Quick Chiller and the Liquid Nitrogen Chiller allow much faster dew point
determination but there is an increased risk of error and temperature overshoot if
components are not allowed to reach thermal equilibrium.
• The Quick Chiller can be used for dew point measurements down to -20°F
(-30°C) with propane, and down to -90°F (-70°C) with liquid CO2 refriger-
ant.
• The Nitrogen Chiller is a dipstick type and can be used with liquid N2 for
dew points down to and below -200°F (-130°C). The Nitrogen Chiller can
also be used with dry ice/acetone for dew points down to -100°F (-75°C).
Certified Thermometers
Certified thermometers that have been tested for accuracy are also available.
They are supplied with calibration traceable to NIST.
Illuminated Magnifier
A four-power, prefocused magnifier is available as an option, and can be attached
to the instrument to illuminate and magnify the mirror. The battery-illuminated
magnifier ensures optimum accuracy of dew point determinations regardless of
field lighting. These illuminated magnifiers are available for the following dew
point tester models: Model 13-075 (large window, 1000 PSIG) and Models 13-
100 (small window, 3000 PSIG), and 13-110 (small window, 5000 PSIG). The
magnifier is a simple design and is easy to operate. It is held on the dew point
tester using an O-ring.
Overview | 1-3
Glycol Filter
Glycol vapor contained in a gas that has passed through a glycol dehydrator will
condense on the tester’s mirror complicating observation of water dew point. A
glycol filter assembly can be installed in the tester to remove the glycol vapor
before the contaminant enters the instrument. The filter is especially useful when
the glycol condensation temperature is above the water dew point.
Tripod
The tripod is lightweight, rigid aluminum. It has two section legs that are extend-
able to 54 1/2” for support in the field.
Propane Cylinder
The propane cylinder is a size 20 DOT-approved cylinder with a delivery control
valve and a 1/4” male NPT thread. The capacity of the cylinder is 3.75 gallons
(14.2 liters) of propane. The cylinder is shipped empty and can be filled at a local
gas supplier.
CO2 Bottle
The CO2 bottle is a size 20 DOT-approved cylinder with a dip tube, delivery con-
trol valve, and a 1/4” male NPT thread. The capacity of the cylinder is 5 lb (2.3
kg) of CO2. The cylinder can be purchased empty, or full of welding grade liquid
CO2.
Chiller Device
C - Standard Chiller
Q - Quick Chiller
L - Liquid Nitrogen Chiller
Application
N - Normal Service
R - Off-Shore Service
S - Sour Gas Service
X - Cleaned for O2 Service
Standard Thermometers
025 -30 to 110°F
027 -90 to +40°F
038 -100 to +10°C
060 -35 to +50°C
096 -200 to 0°F
099 -0 to +220°F
Certified Thermometers
146 -90 to +40°F
148 -200 to 0°F
149 -30 to 110°F
152 -100 to +10°C
157 -35 to +50°C
manual.
Battery Charger Power
1 - 115 VAC 50/60 Hz
2 - 230 VAC 50/60 Hz
13- 1200- C N 1
13-1200-C-N-1: Typical Part Number
Overview | 1-5
Technical Support
AMETEK is committed to providing you with the best technical support in the
industry. If you need service or application assistance, please call Chandler at
918-250-7200, or call your local AMETEK/Chandler representative.
If you need to return equipment, you will be asked to provide the following infor-
mation before obtaining a Return Material Authorization (RMA) number.
• Model number
• Serial number
• Telephone number
Before returning material, you must get an RMA number from the
factory.
NOTE
Accuracy:
± 0.5°F (± 0.25°C)
Operating Pressure:
Model 13-075 Standard Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-100 Standard Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-110 Standard Tester: 0 to 5000 PSIG (34.6 mPa)
Connections:
Sample inlet and outlet: 1/8” NPT
Refrigerant: 1/8” tube
Specifications | 2-1
ChanScope Digital Dew Point Tester
Accuracy:
± 0.2°C (± 0.4°F) at 40°C (104°F) down to -90°C (-130°F)
Operating Pressure:
Model 13-1175 ChanScope Tester: 0 to 1000 PSIG (6.9 mPa)
Model 13-1200 ChanScope Tester: 0 to 3000 PSIG (20.7 mPa)
Model 13-1210 ChanScope Tester: 0 to 5000 PSIG (34.6 mPa)
Connections:
Sample inlet and outlet: 1/8” NPT
Refrigerant: 1/8” tube
Power
Battery-powered (NiMH), 100-240 VAC, 50/60 Hz Smart Charger with three
interchangeable blades
This chapter shows you how to set up your dew point tester. It includes the fol-
lowing:
• Safety Requirements
• Mechanical Installation
- Equipment Required
- Tubing and Fittings
- Refrigerant Connections
- Glycol Filter Assembly
- Temperature Display Units (ChanScope only)
• Start-Up
- Observing the dew point temperature
• Always wear safety glasses when working with refrigerant and high pressure
gases.
Use extreme caution if you are using the instrument with oxygen or
any other explosive gas reactive to oil. The entire instrument and all
sample tubing must be thoroughly cleaned.
No utilities are required for the standard dew point tester. The Chan-
Scope requires 100-265 VAC, 50/60 Hz for the battery charger.
NOTE
Tools/Equipment Required
• 7/16”, 1/2” open-ended wrenches, 10” adjustable wrench
• Air duster
• Teflon tape
Observe the following precautions when setting up the instrument and doing test-
ing in the field.
• Ensure that the temperature of the sampling line and the dew point tester
are above the dew point temperature of the gas in the pipeline. Heat-traced
sample tubing may be required in cold ambient conditions.
• Ensure that the sampling line does not contain any condensed liquids such as
water, hydrocarbons, glycol, etc.
Remove the rear cover of the ChanScope enclosure to change the temperature
display units. Selections are degrees Celsius (°C) or degrees Fahrenheit (°F).
There is a circuit assembly inside the enclosure. Set the miniature switch setting
#1 to ON for Fahrenheit or to OFF for Celsius. Switches 2, 3 and 4 are not used.
3. Partially open the sample outlet valve just enough to allow flow through the
tester body but not enough to cause a drop in the sample pressure.
4. Allow the sample to flow through the tester until sufficient time has passed to
allow system equilibrium. This should be approximately five to ten minutes.
5. Connect the refrigerant tubing from the refrigerant cylinder to the refrigerant
inlet connection valve on the chiller device.
6. Slowly open the refrigerant valve to start the flow of refrigerant through
the chiller device. Once the temperature starts to drop, close the refrigerant
valve.
7. Slowly open and close the refrigerant valve so that the temperature of the
sample gas drops slowly.
8. Observe the mirror and thermometer/digital readout. At the same time, make
note of the temperature at which a spot of moisture first appears at the center
of the mirror.
9. Allow the mirror to warm up and make note of the temperature at which the
moisture spot disappears.
10. Repeat Steps 7 and 8 until the temperatures observed are within 2°F (1°C).
11. Record the average or mean temperature that you observed as the dew point
temperature.
The filtering material in the glycol filter absorbs both water vapor and glycol va-
por. However, though it quickly becomes saturated with water vapor, it continues
to absorb glycol vapor for a somewhat longer time. It is during this latter period
that water dew points can be determined satisfactorily since once the filter is
water saturated, no more water vapor is removed from the gas and the dew points
will be the same if read without the filter.
Determining Equilibrium
When the sample gas is introduced at the side of the pressure chamber, it flows
through the glycol filter before entering the tester. The filter material should not
cause a drop in pressure at normal flow rates. However, you must allow sufficient
time for the filter material to attain equilibrium with the sample gas pressure. The
time required to reach equilibrium depends on the concentration of water and
glycol in the gas and the density of the filter material in the cartridge. Equilib-
rium is determined by running preliminary dew point tests. Two consecutive test
results without glycol masking of the mirror indicate that the tester is operating
normally.
Operation | 4-1
Test Precautions
These precautions must be observed to prevent serious injury and to ensure ac-
curate measurements.
1. Vent the sample gas to a safe area to prevent the possibility of creating an
explosive atmosphere in an enclosed area.
2. Always wear safety goggles when working with refrigerants and high -pres-
sure gases.
5. Take a representative sample using a sample probe inserted into the center
1/3 of the pipeline.
6. Ensure that the sampling line does not contain condensed water, hydrocar-
bons, glycol, etc.
7. Ensure that the temperature of the sampling line and the dew point tester are
above the dew point temperature of the sample gas. When the atmospheric
temperature is below the dew point temperature of the tester or the gas in the
pipeline or tank, the tester and sample line should be heated with a flexible
heat tape to avoid condensation between the gas source and the instrument.
9. If you have installed a filter to remove glycol from the gas, the material in the
filter absorbs both glycol and water. The filter usually becomes saturated with
water within five minutes but continues to absorb glycol. Only after the filter
is saturated with water can you make accurate dew point readings.
Operation | 4-3
Test Procedure
Follow this procedure in the ordered sequence to ensure accurate readings.
1. Open the sample inlet valve and purge the tester by flowing the test gas
through the sample chamber to expel all the air.
2. Regulate the flow of the test sample by partially closing the outlet valve and
then adjust to ensure that the sample flows continuously at full sample pres-
sure through the chamber.
3. Adjust the outlet valve so that the gas sample can just be detected on your
fingertip. An aspirator bulb or vacuum pump can be used to bring atmospher-
ic or low-pressure samples past the mirror.
4. ChanScope
For models with a digital dew point indicator, verify that the decimal point at
the far right side of the display is not blinking.
A blinking decimal point indicates that the tester is in pause or hold mode.
When in this mode, the displayed temperature is “frozen” and will not
change until you press the HOLD button. The HOLD button is a pushbut-
ton or mementary switch; press HOLD at any time to ENTER or EXIT this
mode.
5. Open the refrigerant valve periodically for a few seconds; allow sufficient
time between valve openings to lower the mirror temperature. You can make
a quick observation to determine an approximate dew point. However, for the
final determination, you should lower the temperature at a rate of no more
than one or two degrees per minute.
6. Observe the mirror and thermometer/digital readout at the same time and
note the temperature at which a spot of moisture first appears at the center of
the mirror.
7. Allow the mirror to warm up and note the temperature at which the moisture
spot disappears.
8. Repeat Steps 5 and 6 until the temperatures observed are within 2°F (1°C).
Example:
1. A dew point determination is done in the normal manner at 1000 PSIG and a
dew point of 30°F is obtained. Plotting on the “Saturated Moisture Content
of Natural Gas as a Function of Temperature and Pressure” chart (Drawing
13-0097) obtains a water vapor content of 6.3 lb/million cubic feet.
2. Pressure in the tester is reduced to 300 PSIG by partially closing the inlet
valve and adjusting the outlet valve to obtain the approximate same flow rate.
If a dew point of 9°F is obtained, you can be sure that the original determina-
tion at 1000 PSIG was correct since the water content remained the same (6.3
lb/MMscf) with the change is pressure and dew point. The water dew point is
lowered as the pressure is lowered, but the water content of the gas remains
the same. Had the water content been different at the second pressure and
dew point, then one of the dew points might have been a hydrocarbon, glycol
or alcohol dew point.
Operation | 4-5
Differentiating between Water and Hydrocarbon Dew Points
Sometimes, the hydrocarbon dew point of the gas will be the same or higher
temperature than the water dew point. In this case, the two dew points must be
differentiated.
• The hydrocarbon wets the mirror and the film expands in an iridescent ring to
cover the mirror’s surface, or appears as a waterfall of droplets that appear to
be pouring from the mirror’s center.
You can also verify the substance on the mirror by determining the
dew point at a lower pressure.
NOTE
Operation | 4-7
This page intentionally left blank.
Maintenance
Tools Required
• Phillips screwdriver
• 1/8” hex wrench
• Use a cotton ball, cotton swab, or soft cloth to clean the window and mirror
surfaces. Be careful to avoid scratching the surfaces.
• To clean the high-pressure window, remove the screw and pull the unit out of
the window nut. Be careful to avoid scratching the window.
· Pressurize the unit to 1.5 times the rated pressure etched on the front of the
unit for 5 minutes.
Charger Functionality
Charging begins when the battery pack is connected to the charger. The LED
is orange before the fast charge starts, and then the LED changes to red.
When the cells are fully charged and the voltage drops (because of the -dV
signal from the cells), the charger will go into a top-off charge mode before
switching to trickle charge mode.
During the top-off charge session, the LED will be green with short, intermit-
tent orange flashes. When the top-off charge is complete, the charger will go
into trickle charge mode, changing the LED to green. The charge current
is now reduced to a safe level, permitting the charger to stay connected to
a NiCd battery pack without damaging the cells. Depending on the specific
manufacturer of NiMH cells, trickle charging in excess of 24 hours may not
be permitted.
If the safety timer disconnects before -dV, the top-off charge will not be
engaged. The charger will then go directly to trickle charge mode and the
LED will be green. If the battery voltage is far below normal, the charger will
stop the fast-charge current and switch to trickle charge mode. The LED
will then indicate “error” by flickering green and red.
If the mains is turned off, the charger will reset and start a new charge cycle
when the mains is turned on again.
Recycle the battery pack at an appropriate facility for NiMH batteries. Visit
www.rbrc.org or call 1-800-8-BATTERY for locations.
• Loosen the gland nut on the RTD and gently pull the RTD out. Replace with
the appropriate thermometer.
• Turn the power switch on to illuminate the mirror and proceed as usual.
Calibration of the ChanScope RTD and the electronics is done at the factory.
Return the tester after obtaining an RMA.
PART # DESCRIPTION
13-0005 Window, Models 13-100 & 13-200 and 13-1200
13-0006 Window retaining nut, Model 13-100 & 13-110
13-0007 Mirror assembly, regular chiller
13-0009 Mirror insulator, Models 13-153 & 13-154 kits
13-0018 Chiller body, regular
13-0020 Thermoguard assembly
13-0021 Lock nut, thermoguard
13-0022 Bulb shield, thermoguard
13-0024 Shield, thermoguard
13-0025 Thermometer, range -30 to +110°F
13-0027 Thermometer, range -90 to +40°F
13-0038 Thermometer, range -100 to +10°C
13-0046 Retaining nut, quick chiller
13-0047 Block, with tube, quick chiller
13-0048 Quick chiller assembly
Glass window assembly, Models 13-100, 13-110, 13-200, 13-210, 13-1200,
13-0052
13-1210 (sour gas and 5,000 PSIG)
13-0053 Cooling rod, nitrogen chiller
13-0054 Glycol filter assembly
13-0060 Thermometer, range -35 to +50°C
Window, glass, Models 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210
13-0064
(sour gas and 5,000 PSIG)
Adapter, glass window, Models 13-100, 13-200, 13-210, 13-1200, 13-1210
13-0065
(sour gas and 5,000 PSIG)
13-0068 Regular chiller assembly
13-0071 Extension, tripod adapter, nitrogen chiller
13-0072 Liquid nitrogen chiller assembly
13-0077 Nut, window (Model 13-075)
13-0078 Nut, mirror, retaining
13-0079 Window, plastic, Models 13-075, 13-175, & 13-1175
13-0084 Adapter, gauge
13-0085 Body, Models 13-075, 13-175, & 13-1175
13-0086 Adapter, tripod
13-0101M Body, Model 13-100, 13-110, 13-200, 13-210, 13-1200, 13-1210
13-0102 Mirror, swivel
13-0111 Body assembly, magnifier
13-0116 Window, magnifier, eye guard
13-0117 Lens retainer, Model 13-075
13-0119 Lens, magnifier, objective
13-0122 Eye guard, plastic, magnifier
PART # DESCRIPTION
13-0204 Window nut, Model 13-175
13-0205 Window nut, Model 13-0200, 13-0210
13-0210 L-bracket, battery
13-0214 Optical assembly
13-0215 Beam splitter assembly
13-0217 Prism assembly
13-0219* Logic board assembly
13-0220* LED board assembly
13-2227 Eyepiece assembly (See DWG 13-2227)
13-0230 Gland nut
13-0234 Switch
13-0236* RTD assembly with strain relief (spring-loaded)
13-0241 Battery holder, two cell
13-0253 Wiring harness assembly
13-0254 Cable assembly
13-1238 Black case with foam inserts
187-20535 Adapter, 220 VAC to 115 VAC, 50/60 Hz, use with P-3116 charger
C08846 Kit, 230 VAC plug adapter
C15472 Eyepiece finder/lens (Fits 13-2227 Eyepiece assembly)
P-0066 O-ring, Eyepiece assembly
P-0221 O-ring, RTD gland nut
P-1044 Lamp, screw base
P-2770 Lamp, bayonet base
P-2820 O-ring, tube support
P-2822** Eye Cup
P-2826 Screw, 2-56 x .250, rnd Head, slot
P-2828 Screw, 6-32 x .250, rnd head, Phillips
P-2830 Screw, 6-32 x 1.0, skt head, SST
P-3113 Battery, C cell, NiCad recharge
P-3115 Holder, (1) C cell, #2173
P-3116 Battery charger, NiCad
PART # DESCRIPTION
13-0204 Window nut, Model 13-1175
13-0205 Window nut, Model 13-1200, 13-1210
13-0215 Beam splitter assembly
13-2227 Eyepiece assembly (See DWG 13-2227)
13-0236* RTD assembly with strain relief (spring-loaded)
13-0307* Circuit assembly
13-1214 Optical assembly
13-1217 Prism assembly
13-1238 Case with foam inserts
13-1250 ChanScope II assembly
13-1252 LED & Hold switch harness
13-1274 Battery pack, rechargeable
13-1276 Switch/battery/harness assembly
13-1277 Assembly, battery charger connector wiring harness
13-2019** Battery charger, NiMH, 1600 mAh, 100-265 VAC, 50/60 Hz, with 3 blades
C09867 Pushbutton hold switch
C11003 Charger connection cap assembly
C11468 Eye cup
C11921 Adapter, charger, US blade
C11922 Adapter, charger, Euro blade
C11923 Adapter, charger, UK blade
C15472 Eyepiece lens (Fits 13-2227 Eyepiece assembly)
P-2820 O-ring, tube support
P-2821 Rocker switch
P-2826 Screw, 2-56 x .250 rnd head, slotted
P-2828 Screw, 6-32 x .250, rnd head, Phillips
P-2830 Screw, 6-32 x 1.125”, skt head, SST (4 required for eyepiece)
* RTD and circuit assembly must be purchased together and be factory-calibrated for NIST
Certification.
Consult the factory for price and availability.
Drawings | 6-1
8
2 9 10
Install
this end to
1
dew point
tester
11
ITEM PART # DESCRIPTION
3 4
5
2
1 6
Drawings | 6-3
14 3/4 approx
13
14 Thermometer
12
15 specify temperature range
11 16
10 17
3
10 1/2
approx
21
2 9 19
20 22
18
1/8 NPT
8
1
7
1/8 NPT
Gas Outlet
5
6
23
Drawings | 6-5
Figure 6-6.
Chart.
Drawings | 6-7
Figure 6-8.
Chart.
Drawings | 6-9
Figure 6-10.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 1.
Drawings | 6-11
Figure 6-12.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 3.
Drawings | 6-13
Figure 6-14.
Equilibrium water
vapor contents in
natural gas above the
critical temperature,
Page 5.
Drawings | 6-15
Figure 6-16.
Dew point to moisture
content conversion
chart.
Drawings | 6-17
Figure 6-18.
ChanScope I
DWG 13-2016.
Figure 6-19.
D ECN T4903; ADDED P-0071 10/3/12 TC
DWG 13-0303.
D D
Illuminated magnifier
INSTALLATION INSTRUCTIONS
1. REMOVE 13-0102 OBSERVATION MIRROR
9
USE WITH MODELS 13-075
C DEWPOINT TESTER, WITH 13-0077 C
ROUND WINDOW NUT.
1 6
7
4 10 8
B B
Drawings | 6-19
THIRD ANGLE PROJECTION STRUCT: SHEET 1 OF 1
TITLE BLOCK REV 3
1 2 3 4 5 6
INSTALLATION INSTRUCTIONS
1. Remove 13-0102 observation mirror from the dew point tester body.
2. Lubricate the magnifier O-ring and push the magnifier over the outside of the
round window nut.
3. Screw 13-0305 illuminator assembly into 13-0124 lens retainer.
4. Rotate 13-0305 illuminator counterclockwise to turn it on.
1 13-0305 Illuminator
2 P-0071 O-Ring
3 13-0124 Lens Retainer
4 H-8014 #8-32 x .5 LG, BHP
Figure 6-20. 5 13-0122 Viewer
Illuminated magnifier 6
1/4-20 x .25 LG, Allen
assembly, Models 13- HD set screw
7 13-0116 Window Eyeguard
100 and 13-110. 8 13-0111 Body Assembly
DWG 13-0304. 9 13-0119 Lens
Figure 6-21.
Illuminator Assembly.
DWG 13-0305.
Drawings | 6-21
Figure 6-22.
ChanScope II.
DWG 13-2017.
Drawings | 6-23
C
A C
4
2
A
3
B B
1
SECTION A-A
ITEM NO. PART NUMBER DESCRIPTION QTY.
1 13-2207 EYEPIECE BODY, THREADED 1
A
2 C15472 LENS,EYEPIECE,DK-17 1 A
3 C11468 EYECUP 1
Eyepiece Assembly
4 P-2830 SCREW,SHCS,SS,6-32X1.125 4
THIS DOCUMENT AND THE DRAWINGS AND
TECHNICAL DATA CONTAINED HEREON ARE THE
BREAK EDGES, DEBURR DRAWN: TWC 11/12/2012
CHANDLER ENGINEERING
DWG 13-2227.
PROPERTY OF CHANDLER ENGINEERING COMPANY UON DIMS ARE IN INCHES MFG:
LLC. REPRODUCTION OR DISSEMINATION IN ANY
1 PLC 0.030 2 PLC 0.010 ENGR: EYEPIECE ASSEMBLY
Figure 6-23.
FORM EXCEPT AS EXPRESSLY AUTHORIZED BY THE
OWNER IS FORBIDDEN. THE HOLDER AGREES TO 3 PLC 0.005 ANGL 1/2
RETURN THE DOCUMENT TO THE OWNER ON
SURFACE FINISH 63 RMS TYPE: PN: 13-2227 REV B SIZE A
THIRD ANGLE PROJECTION STRUCT: PROJ: Imported Data Set
DEMAND. COPYRIGHT BY CHANDLER ENGINEERING
COMPANY LLC SHEET 1 OF 1
1 2 3 4 TITLE BLOCK REV 3
This page intentionally left blank.