Catalog ............................................................................................................................................................ i
Declaration ..................................................................................................................................................... I
Chapter 1 Introduction ........................................................................................................................... 1
1.1 Characteristics ............................................................................................................................. 1
1.2 Intended use................................................................................................................................. 1
1.3 Instrument characteristics ............................................................................................................ 1
1.3.1 Instrument Features ..................................................................................................... 1
1.3.2 Power and Environment requirements ...................................................................... 2
1.4 Constitution ................................................................................................................................. 2
Chapter 2 Operation principle .............................................................................................................. 5
2.1 Product overview............................................................................................................................ 5
2.2 Pipe system..................................................................................................................................... 5
2.2.1 Function diagram ............................................................................................................. 5
2.2.2 Principle ............................................................................................................................. 5
2.3 User convenient interface ............................................................................................................... 5
Chapter 3 Installation and power on .................................................................................................. 6
3.1 Unpacking ...................................................................................................................................... 6
3.2 Installation ...................................................................................................................................... 6
3.2.1 Environment requirements ............................................................................................. 6
3.2.2 External connection ......................................................................................................... 6
3.2.3 Wash head installation .................................................................................................... 7
3.3 Power on......................................................................................................................................... 7
Chapter 4 Operation.............................................................................................................................. 10
4.1 Program ........................................................................................................................................ 10
4.1.1 Program No. ................................................................................................................... 10
4.1.2 Washing mode................................................................................................................ 10
4.1.3 Aspiration mode ............................................................................................................. 11
4.1.4 Washing cycle ................................................................................................................ 11
4.1.5 Select row ....................................................................................................................... 11
4.1.6 Rinse volume .................................................................................................................. 11
4.1.7 Soaking time ................................................................................................................... 11
4.1.8 Aspiration time................................................................................................................ 12
4.1.9 Program operation ......................................................................................................... 12
4.2 System settings ............................................................................................................................. 13
4.2.1 Dispense location........................................................................................................... 13
4.2.2 Distributing head vertical position................................................................................ 14
4.2.3 Distributing head horizontal position ........................................................................... 14
4.2.4 Bottom rinsing ................................................................................................................ 14
4.2.5 Shaking time ................................................................................................................... 14
4.2.6 Distributing head standard............................................................................................ 14
4.2.7 Keypad sound................................................................................................................. 14
4.2.8 System setting operation .............................................................................................. 15
i
4.2.8.1 Dispense location adjustment ................................................................................... 15
4.3 Whole washing ............................................................................................................................. 16
4.4 Maintenance ................................................................................................................................. 17
4.4.1 Prime................................................................................................................................ 17
4.4.2 Rinse ................................................................................................................................ 17
4.4.3 System reset ................................................................................................................... 17
4.4.4 Equipment self-test ........................................................................................................ 17
4.4.5 System recovery ............................................................................................................ 18
4.4.6 Dispense check .............................................................................................................. 18
4.4.7 Maintenance operation ................................................................................................. 18
4.5 Power off ...................................................................................................................................... 19
Chapter 5 Attentions ............................................................................................................................. 20
5.1 Protection ..................................................................................................................................... 20
5.2 Before operation ........................................................................................................................... 20
5.3 In the operation............................................................................................................................. 20
5.4 Power safety Caution ................................................................................................................... 20
5.5 Sterilization .................................................................................................................................. 21
Chapter 6 Maintenance and storage................................................................................................. 22
6.1 Maintenance of instrument and parts ........................................................................................... 22
6.2 Troubleshooting............................................................................................................................ 22
6.3 Substitutions ................................................................................................................................. 22
Supplement 1:Symbols ............................................................................................................................. 23
ii
Declaration
Copyright© Guilin Urit Medical Electronic Co., Ltd., All Rights Reserved.
Congratulations to be our honourable guest, welcome to use URIT-670 microplate washer,
which will bring to you fresh experience and convenience.
This manual is complied according to legislation and concrete situation about microplate
washer which made by Guilin Urit Medical Electronic Co., Ltd. It can only use in the domestic of people’s
repulic of china (not includes Taiwan, Hongkong, Macao).
This specification including lastest information till printing. Guilin Urit Medical Electronic Co., Ltd.
copyrights this specification in simplified Chinese version’s edition and specification and reserve the right to
change at any time without Caution.
The referred pictures in this specification are figures for indication only. In some cases, the
appearance of the actual product may differ from the product in photo, the appearance of the
actual product is as it is.
All information in this specification are protected by copyright law, no part of this specification
may be reproduced in any form copying, printing and translating to other language, without the
prior written consent of Guilin Urit Medical Electronic Co., Ltd.
Operator must operate it according to the instruction of this specification strictly. Otherwise,
Guilin Urit Medical Electronic Co., Ltd. is irresponsible for any fault and breakdown of this instrument
that caused by unprofessional operation.
I
《URIT-670 Microplate Washer operating Manual》.
Replace fitting not allowed by Guilin Urit Medical Electronic Co., Ltd., or maintain, mend or
modify instrument not via by Guilin Urit Medical Electronic Co., Ltd.
Discreteness is tear open and installation, pull and extend, debugging afresh.
The instrument will be abated if each hospital or institution cannot carry out a suit of normative
maintenance program.
Guilin Urit Medical Electronic Co., Ltd. will provide circuit diagram not freely if user put forwards
requirement, in addition, we will listing to explain adjustment method and other information, which in order to
help user can let eligible technical person to mend the part that Guilin Urit Medical Electronic Co., Ltd.
Consider user can maintain by themselves.
Caution:
Guilin Urit Medical Electronic Co., Ltd. disclaimer instrument is using in special purpose and
make any silent assurement for its saling and using.
We arouse you give us a call if you meet any problems. Our company’s service department
dredge user technique, service hotline to provide technical support and difficulty solution for user.
Please reserve relative sample and inform agent or contact Guilin Urit Medical Electronic Co., Ltd.
service department immediately if there is any fault occurred. Guilin Urit Medical Electronic Co., Ltd. will send
professional engineer or local service representative to provide timely and fully servicing if the fault cannot
solve by telephone guidance.
Manufacturer, instrument registration information
URIT Medical Electronic Co., Ltd
No.D-07 Information Industry District,High-tech Zone,
Guilin, Guangxi 541004, PR China
Tel: +86(773)2288586 Fax: +86(773)2288560
Web: www.urit.com E-mail: [email protected]
Version No.:03/2017
II
Chapter 1 Introduction
1.1 Characteristics
Main target is rinsing for immune test analysis panel, U and V 96-well microplate which used
by medical institution.
1
Vibrating time 0-60 sec. (can be adjusted)
Immersing time (1~90) sec. (can be adjusted)
Aspiration time (1~10)sec. (can be adjusted)
Weight about 5kg
Dimension 400mm×330 mm×170mm
Input power ≤250W
Fuse current 3A
1.4 Constitution
Display
Keyboard
Plastic
cover
2
Figure1-2 Keyboard panel
Keyboard function:
PROG Enter Program menu for setting operation mode about instrument and
parameter.
Cursor glints showing menu choice increasing, select one program means
content plus one.
Cursor glints showing menu choice decreasing, select one program
means content minuses one.
ENTER Enter or exit when item is chosen. Choose current item when cursor is glinting.
Save current parameters when “wash program” is chosen.
MAINT Enters into maintenance menu and performs multiple maintenance functions
CANC Cancels current operation and returns to main menu, meanwhile, the system
will not save the settings. Stops running current program or resets the
system when wash program is running.
SYS Setup system parameters which wash program will run according to it.
RUN Perform current selected mission.
Instrument rear view is shown in Figure1-3:
3
1.Vacuum plug
4.Power switch
7.Sensor
4
Chapter 2 Operation principle
Under the control of microcontroller, the instrument performs the wash program which was set
in advance. Under the drive of stepper motor, detergent is injected equably into microplate-wells
which are chosen by user, and sucks away waste-liquid by sucking needle after reacting between
detergent and waste of microplate-wells wall.
2.2.2 Principle
Under the control of liquid pump, liquid in the detergent bottle is injected into microplate-wells.
Then the aspiration-needle sucks waste-liquid into waste bottle after reacting between detergent
and waste. The instrument completed washing work by cycled performing thus program.
The instrument uses 240*128 dot matrix LCD display screen for easy operation, and all
parameters are listed on a single screen, which implements visual operation.
Interface divides into program, system and maintenance three main menus, program interface
is defaulted for standby interface. In order to make instrument possess intelligence to react
multiple operation, the system adopt pop up dialogue box form, which giving hints to every
operation.
5
Chapter 3 Installation and power on
3.1 Unpacking
a) Please take the instrument and accessories from the packing box carefully, reserve the
packing material for the use of transportation or storage afterwards.
b) Please check the accessories according to the packing list.
c) Please check if there is spilling detergent or water soaking about the instrument.
d) Please check there is any mechanical damage about it
e) Please check all external-expose lead, inserted part and accessory.
If you have any questions, please contact local dealer immediately.
3.2 Installation
a) Take the detergent bottle and waste bottle out from packing box, place them on the back of
instrument, we suggested place detergent bottle on the left side, waste bottle on the right side.
b) Butt the pipe with two colors junction (green and black) joint with the matched colored
bottle on the back of instrument, and screw the nut.
c) Insert the sensor plug on the waste bottle cap into sensor socket on the back of
6
instrument’s plate.
d) Pour detergent into detergent bottle.
Connect the dispense pipe which using for linking distributing head with pipe junction the
working table according to matched colors.
Make the distributing head parallel supporting arm and put distributing head above supporting
arm, adjusting the adjust screw on the supporting arm to let distributing head parallels working
table.
Tray
3.3 Power on
7
China Guilin Urit
Self-test report
2) Aspiration mo de: Du al
3) Washing cycle: 01
8
Caution: The instrument will keep warning and refuse to enter into work
interface due to any system errors such as self-check failure or abnormal operation, shut
the instrument off and restart it after a while, if the problem still exist, cut off power and
contact our company’s service department.
Caution: Do not place instrument on the position that hard to touch power
switch.
9
Chapter 4 Operation
4.1 Program
Instrument enters Program interface directly after power on, which is shown as Figure 4-1.
Program menu contains program No., washing mode, Aspiration mode, wash cycle, selsct row,
rinse volume, soaking time and aspiration time , these parameters can be set according to
customer’s need. Program No.
Washing program 01
1) Washing mode: Whole
3) Washing cycle: 01
Program No. places right “washing program”, which is shown in Figure 4-1. User can set 50
groups different parameters’ wash program, and system will store these. And the instrument will
run according to your settings when you chose one group of programs and press “run” key.
You can choose wash program by pressing “▲”or “▼” key, when you make your choice,
matched parameter will refresh automatically.
10
4.1.3 Aspiration mode
Wash cycles time:Set time freely between 1 to 20, instrument will repeat this operation.
Set washing rows needed, for example, max row is 8 when distributing head’s quantity is 12,
matched code is ABCDEFGH; max row is 12 when distributing head’s quantity is 8, matched code
is ABCDEFGHIJKL. Choose matched code, selected row will be washed. When distributing
head’s quantity is 8, panel washing interface is shown as Figure 4-2:
Washing program 01
1) Washing mode: Whole
3) Washing cycle: 01
Rinse volume can be set freely between 50 to 450uL, 50uL increasing every time.
Soaking time is the waiting time which the beginning of aspiration after dispense of washing
head completed, startup immersing function can increase cleaning effect, which can be set freely
between 0 to 90 seconds.
11
4.1.8 Aspiration time
Aspiration time is the time aspiration needle aspirates liquid in the microwell bottom, which can
be set freely between 1-10 seconds. Aspiration time shorter is, washing speed faster is, but be
sure to protect effect. If residual volume too large, aspiration time can be increased. System
aspiration time is 1 second.
a) Instrument enter standby main interface after self-check, standby main interface will
acquiescence to program menu interface, which is shown as Figure 4-1.
b) Press “▲”or “▼” key, glinting cursor appears on the screen, which is shown as Figure 4-3.
Continue to press “▲” or “▼”key, make the cursor move to the program want to edit.
Washing program 01
1) Washing mode: Whole
3) Washing cycle: 01
pressing “ENTER” key. System won’t store all edited data automatically if it doesn’t press
“ENTER” key; press “RUN” key when “whole ” or “program No.” is chosen, instrument will
store current parameter automatically and begin to operate whole washing.
Caution: Instrument will not store parameters by pressing “ENTER” key when all
Caution: Press “▲”or “▼”key will edit program when cursor glints; press “▲”or
Press “SYS” key on the keyboard, “system settings” menu on the main interface of standby
which is shown as Figure 3-4 will appear.
System setting menu including the setting of dispense location, distributing location HOR and
VER, bottom rinsing, shaking time, distributer spec and keypad sound, which is shown as Figure
4-5
System settings
1) Dispense location +0
2) Distributing location
Hor +0 Ver +0
3) Bottom rinsing ON
4) Shaking time 00s
5) Distributer spec 12
6) Keypad sound ON
Dispense location is the position when distributing head in dispense situation, dispense needle
bottom relative to the microplate’s level. Appropriate position will prevent liquid’s overflow, in
general, user cannot change system’s default setting.
When washing no microplate appears liquid’s overflow, using this function to adjust ±2mm can
meet different requires.
13
4.2.2 Distributing head vertical position
Distributing head vertical position is the altitude that distributing head moving on the vertical
direction, this altitude can be adjusted ±2mm, when the distributing head move downwards to the
bottom of microplate, the altitude of supporting arm should lower than distributing head(it means
distributing head could fall on the bottom of microwell freely, otherwise Aspiration halfway).
Distributing head on the relative place which in microwell of the first row, this place can be
adjusted by ±2mm, suitable position can obtain better effect of aspiration, in addition, liquid will not
spray on the wall of microwell or outside.
Bottom rinsing dispenses detergent when dispense needle approaching the bottom of
microplate well, which can increase washing effect. User can turn it on or off.
Shaking will make washing effect better. Shaking time can be set between 0 to 60 seconds
freely. Shaking will be closed when time is 0.
Caution: System need edit relative parameter setting when switch 8 needles or 12
needles. After switch, Program menu data will recover to default data.
Warning: Make sure the following is right when replace distributing head,:
14
Caution: Only press key doesn’t make sound when keypad sound is set for close,
but some systems’ hints or warnings are still sufferance to produce sound.
a) In system setup interface, press “▲”or” key, glinting cursor appears on the screen which is
shown as Figure4-6
System settings
1) Dispense location +0
2) Distributing location
Hor +0 Ver +0
3) Bottom rinsing ON
4) Shaking time 00s
5) Distributer spec 12
6) Keypad sound ON
In normal, dispense location setting is forbidden, unsuitable adjustment will lead to abnormal
status or unexpected effect when wash program is running. User can adjust dispense position to
match it If microplate is not standard,
Firstly, select this program, press “ENTER” key, stop it after system location, press “▲” or “▼”
key to adjust aspiration needle position, up or down, observing condition when adjusting it as far
as to get satisfied effect, then press “ENTER” key to quit.
Caution: When user adjusting “Dispense location” and “distributing head location”,
press “ENTER” key to select, instrument will drive distributing head and tray to pointed
position. Other operation will inefficacy before this operation has finished.
Press “▲” or “▼” key each time, instrument will reflect the set situation immediately
(only make effect to press key when operation has finished and sound occur), user can
observe this situation to adjust instrument.
Carry out rinsing after program and setting, the details as following:
a) Place the microplate you want to wash on the tray.
b) Press “▲”or“▼” key, “Program No.” is chosen by glinting bar; press “▲”or “▼” key, choose
program No. want to operate, press ”ENTER” key.
c) Press “▲”or“▼”key, make the “washing program” is chosen by glinting bar.
d) Press “run” key on the keyboard, instrument begins to rinsing according to set data,
washing parameter showing on the LCD, lowest two lows on the bottom show current progress,
user can know progress according to it, process interface is shown as Figure 4-7.
Caution: By press “RUN” key will in effect only when “Washing program”or
Caution: Make sure the specification of distributing head correspond with the
16
4.4 Maintenance
Press “MAINT “key on the keyboard, enter “maintenance” menu under the stand-by main
interface shown as Figure 3-4. Maintenance menu including Prime, Rinse, System reset,
Equipment self-test, system recovery and dispense check, maintenance interface is shown as
Figure 4-9.
M ainte nance
1) P rime
2) R inse
3) S ys te m res et
4) E q uipm en t self-te st
5) S ys te m rec overy
6) D ispens e check
4.4.1 Prime
Prime is turgoring and cleaning for system pipeline in high speed, making instrument’s
accuracy in working. Priming every time when it using for the first time or changing detergent.
Priming can solve for well blocking.
4.4.2 Rinse
Rinseis cleaning for system pipeline, distributing head, dispense needle and aspiration needle,
making expedition of airline and liquid line. Execute rinsing one time or more to make the
instrument working normally for the first using or after long keeping.
Instrument can reset correctly when “system reset” about distributing head and tray completed
in any place. Distributing head located above washing groove after reset.
Report of error will pop up if system reset failed to remind user the error type. Instrument can’t
work normally if system reset failed, it needs mending at this time.
Equipment self-test operation carry out every time automatically when instrument turn on.
Instrument has sever fault will lead to instrument working damage. In this case, protection
measure will on, error report popes up at first, warning lasting, instrument cannot work normally.
Turn off power supply and reset after a while if above situation appears; some part of this
17
instrument has already damaged when problems is lasting, please contact Urit company to check.
Use this self-check function can realize the health condition in normal using.
Execute this function will make each parameter recover to factory installation.
Use this function can recover instrument system when it running or adjusting.
Caution: Memorizer may in fault if system recovery failed or messy code appears
Use this function to check dispense volume is exact or not and path of liquid is blockage or
not,or need washing or not. After operation, instrument will stop operation aspiration after inject
detergent to microwell.
Caution: Dispense volume is program No. dispense volume in program menu, user
can set dispense volume according to “washing program”, select “Dispense check” in
maintenance menu to operate.
a) Under the equipment maintenance interface shown as Figure 4-9, press “▲”or“▼” key,
equipment maintenance interface appears glinting cursor shown as Figure 4-10.
b) Press “▲”or “▼” key, move the cursor to the maintenance program need to operate.
c) Press “RUN” key on keyboard, instrument begins to carry out matched maintenance
program.
d) Instrument returns to the equipment maintenance menu interface shown as Figure 4-9
when maintenance program completed.
e) Press “CANC” key, return to standby main interface.
M ainte nance
1) P rim e
2) R inse
3) S ys te m res et
4) E q uipm en t self-te st
5) S ys te m rec overy
6) D ispens e check
Turn off power switch on rear panel, cut off power supply when instrument doesn’t running.
19
Chapter 5 Attentions
To ensure user safety, keep the accuracy of experiment, avoid the damage of instrument in
case of operation’s fault, following fairs should pay attentions:
5.1 Protection
Infectious specimens and corrosive liquid can often be met in experiment, put on antiseptic
experimental clothing and glove in the operation, Deal with dangerous samples according to
laboratory operation handbook.
a) Install instrument according to this manual strictly, ensure the correct connection of
pipeline to avoid leakage.
b) Operate rinsing and priming more than 2 times to ensure expedition of pipeline and airline
when use washer again after long time storage.
c) Use the hose provided by Urit company for connection.
a) Make sure the external equipment such as circuit, socket in accordance with electrical
appliance safety standard.
b) Do not use this instrument in damp environment
c) Do not turn on the instrument if there is any mechanic damage
d) Cut off all the power supply connecting to the instrument when operate any adjustments,
20
displacement, maintenance or mending about instrument.
e) Technician who is familiar with potential danger should deal with any unavoidable electric
situation.
f) Cut off power supply immediately when instrument is wetting by water, user should deal
with this situation under the instruction of technician.
5.5 Sterilization
Instrument should be sterilized regularly. Gather waste water in special vessel, explanation
after sterilization, clean up the pollutant on the surface of washing plate by 70% alcohol, aspiration
orifice should be wiped several times by 200 mg/L clean liquid, wiped by rinsing agent or rinsing
several times for stand-by using.
21
Chapter 6 Maintenance and storage
6.2 Troubleshooting
a) Screen does not display or on reaction when press keyboard, poor connection about LCD
or keyboard may the possible reason, shut the instrument off and restart it after a while.
b) System can identify trouble automatically when the instrument self-checking or running,
system self-protection start-up to protect instrument if system in severe condition, instrument
return to normal status after problems solved.
c) Identification of instrument’s damage:
1) System faults shown on LCD and self-check report pop up when power on or running
wash program; some parts on instrument have already damaged if above situation appears
too many time, contact Urit company workmen to deal with it.
2) Instrument may have partly damage and stop to using under following circumstances
① Severe external damage exists;
② Still in abnormal situation when repeat operation for many times;
③ Storage too long under the unqualified environment;
④ Transportation pressure overloads.
Contact Urit company workman to deal with any abnormal problems.
6.3 Substitutions
Liquid bottle should be changed every 2 years or earlier under sunlight or keep long time. Urit
company provides following parts, and it will irresponsible for any consequence caused by user
themselves.
1 detergent, waste liquid pipe.
2 Distributing head.
22
3 Tray.
Supplement 1:Symbols
Biohazard
Serial Number
Manufacturer
Recovery
23