URIT-670operation Manual 03 2017

Download as pdf or txt
Download as pdf or txt
You are on page 1of 27

Catalog

Catalog ............................................................................................................................................................ i 
Declaration ..................................................................................................................................................... I 
Chapter 1 Introduction ........................................................................................................................... 1 
1.1  Characteristics ............................................................................................................................. 1 
1.2  Intended use................................................................................................................................. 1 
1.3  Instrument characteristics ............................................................................................................ 1 
1.3.1  Instrument Features ..................................................................................................... 1 
1.3.2  Power and Environment requirements ...................................................................... 2 
1.4  Constitution ................................................................................................................................. 2 
Chapter 2 Operation principle .............................................................................................................. 5 
2.1 Product overview............................................................................................................................ 5 
2.2 Pipe system..................................................................................................................................... 5 
2.2.1 Function diagram ............................................................................................................. 5 
2.2.2 Principle ............................................................................................................................. 5 
2.3 User convenient interface ............................................................................................................... 5 
Chapter 3 Installation and power on .................................................................................................. 6 
3.1 Unpacking ...................................................................................................................................... 6 
3.2 Installation ...................................................................................................................................... 6 
3.2.1 Environment requirements ............................................................................................. 6 
3.2.2 External connection ......................................................................................................... 6 
3.2.3 Wash head installation .................................................................................................... 7 
3.3 Power on......................................................................................................................................... 7 
Chapter 4 Operation.............................................................................................................................. 10 
4.1 Program ........................................................................................................................................ 10 
4.1.1 Program No. ................................................................................................................... 10 
4.1.2 Washing mode................................................................................................................ 10 
4.1.3 Aspiration mode ............................................................................................................. 11 
4.1.4 Washing cycle ................................................................................................................ 11 
4.1.5 Select row ....................................................................................................................... 11 
4.1.6 Rinse volume .................................................................................................................. 11 
4.1.7 Soaking time ................................................................................................................... 11 
4.1.8 Aspiration time................................................................................................................ 12 
4.1.9 Program operation ......................................................................................................... 12 
4.2 System settings ............................................................................................................................. 13 
4.2.1 Dispense location........................................................................................................... 13 
4.2.2 Distributing head vertical position................................................................................ 14 
4.2.3 Distributing head horizontal position ........................................................................... 14 
4.2.4 Bottom rinsing ................................................................................................................ 14 
4.2.5 Shaking time ................................................................................................................... 14 
4.2.6 Distributing head standard............................................................................................ 14 
4.2.7 Keypad sound................................................................................................................. 14 
4.2.8 System setting operation .............................................................................................. 15 

i
4.2.8.1 Dispense location adjustment ................................................................................... 15 
4.3 Whole washing ............................................................................................................................. 16 
4.4 Maintenance ................................................................................................................................. 17 
4.4.1 Prime................................................................................................................................ 17 
4.4.2 Rinse ................................................................................................................................ 17 
4.4.3 System reset ................................................................................................................... 17 
4.4.4 Equipment self-test ........................................................................................................ 17 
4.4.5 System recovery ............................................................................................................ 18 
4.4.6 Dispense check .............................................................................................................. 18 
4.4.7 Maintenance operation ................................................................................................. 18 
4.5 Power off ...................................................................................................................................... 19 
Chapter 5 Attentions ............................................................................................................................. 20 
5.1 Protection ..................................................................................................................................... 20 
5.2 Before operation ........................................................................................................................... 20 
5.3 In the operation............................................................................................................................. 20 
5.4 Power safety Caution ................................................................................................................... 20 
5.5 Sterilization .................................................................................................................................. 21 
Chapter 6 Maintenance and storage................................................................................................. 22 
6.1 Maintenance of instrument and parts ........................................................................................... 22 
6.2 Troubleshooting............................................................................................................................ 22 
6.3 Substitutions ................................................................................................................................. 22 
Supplement 1:Symbols ............................................................................................................................. 23 

ii
Declaration
Copyright© Guilin Urit Medical Electronic Co., Ltd., All Rights Reserved.
Congratulations to be our honourable guest, welcome to use URIT-670 microplate washer,
which will bring to you fresh experience and convenience.
This manual is complied according to legislation and concrete situation about microplate
washer which made by Guilin Urit Medical Electronic Co., Ltd. It can only use in the domestic of people’s
repulic of china (not includes Taiwan, Hongkong, Macao).
This specification including lastest information till printing. Guilin Urit Medical Electronic Co., Ltd.
copyrights this specification in simplified Chinese version’s edition and specification and reserve the right to
change at any time without Caution.
The referred pictures in this specification are figures for indication only. In some cases, the
appearance of the actual product may differ from the product in photo, the appearance of the
actual product is as it is.
All information in this specification are protected by copyright law, no part of this specification
may be reproduced in any form copying, printing and translating to other language, without the
prior written consent of Guilin Urit Medical Electronic Co., Ltd.
Operator must operate it according to the instruction of this specification strictly. Otherwise,
Guilin Urit Medical Electronic Co., Ltd. is irresponsible for any fault and breakdown of this instrument
that caused by unprofessional operation.

Limited quality responsibility certificate


《URIT-670 microplate washer operating manual》clear the agreement that concerning quality
responsibility certificate, right and duty for after service as well as generating and terminating
between Guilin Urit Medical Electronic Co., Ltd. and customer.
Guilin Urit Medical Electronic Co., Ltd. will provide one year free service for the instrument that saled by it
or competent agent from the day of installation if there is any damage caused by manufacturer or material
Guilin Urit Medical Electronic Co., Ltd.’ s duty is only for maintenance under the followings, not including
the economic loss or additional fee caused by stopping use instrument. For example:
 Freight(including custom fee)
 Economic and time loss caused by incapable using instrument
 Lodgement, repast or other on the way fee.
 The loss caused by inconvenient.
 Other charge.
Guilin Urit Medical Electronic Co., Ltd. is irresponsible for the direct, indirect or ultimate damage
caused by the following situations. Guilin Urit Medical Electronic Co., Ltd. has no responsibility for the
security, reliability and running situation if the following situation occurs. And at the same time, we
consider that the quality guarantee responsibility, right and duty about sale-after service,
assumpsit of engender and stopping, all of that promissory free service right will be seen as
permanence abandon unconditionally.
 Using abnormally, not maintained or damaged instrument.
 Using regent and accessory that not provided or allowed by Guilin Urit Medical Electronic
Co., Ltd.
 Instrument damage caused by miss operation and carelessness which not according to

I
《URIT-670 Microplate Washer operating Manual》.
 Replace fitting not allowed by Guilin Urit Medical Electronic Co., Ltd., or maintain, mend or
modify instrument not via by Guilin Urit Medical Electronic Co., Ltd.
 Discreteness is tear open and installation, pull and extend, debugging afresh.

The instrument will be abated if each hospital or institution cannot carry out a suit of normative
maintenance program.
Guilin Urit Medical Electronic Co., Ltd. will provide circuit diagram not freely if user put forwards
requirement, in addition, we will listing to explain adjustment method and other information, which in order to
help user can let eligible technical person to mend the part that Guilin Urit Medical Electronic Co., Ltd.
Consider user can maintain by themselves.

Caution:

Guilin Urit Medical Electronic Co., Ltd. disclaimer instrument is using in special purpose and
make any silent assurement for its saling and using.

We arouse you give us a call if you meet any problems. Our company’s service department
dredge user technique, service hotline to provide technical support and difficulty solution for user.
Please reserve relative sample and inform agent or contact Guilin Urit Medical Electronic Co., Ltd.
service department immediately if there is any fault occurred. Guilin Urit Medical Electronic Co., Ltd. will send
professional engineer or local service representative to provide timely and fully servicing if the fault cannot
solve by telephone guidance.
Manufacturer, instrument registration information
URIT Medical Electronic Co., Ltd
No.D-07 Information Industry District,High-tech Zone,
Guilin, Guangxi 541004, PR China
Tel: +86(773)2288586 Fax: +86(773)2288560
Web: www.urit.com E-mail: [email protected]

Supplied by: URIT Medical Electronic Co., Ltd

Wellkang Ltd t/a Wellkang Tech Consulting


Suite B, 29 Harley Street, LONDON W1G 9QR, UK

Version No.:03/2017

II
Chapter 1 Introduction

1.1 Characteristics

1) English-operation system, liquid crystal display, graphic interface.


2) Main target is rinsing for the flat, U and V type 96-wells microplate which be used for
immune test analysis by medical institution.
3) Adjustable programmable washing mode, aspirations mode, washing time, Detergent
quantum, immersing time, pause time,and vibrating time. Supports bottom rinsing and compatible
with multiple microplate specification by adjusting distributing head position.
4) All automatic washing panel processing , washing head localization automatically, user
can adjust its localization manually.
5) Independently designs for aspirations needles and dispense needle, which evidently
decreases cross-pollution.
6) Random row washing function can meet different applications.
7) Automatic alarm when waste liquid overflows.
8) Multiform instrument maintenance functions.
9) Counter-overflowing function, which can avoid cross pollution between well and well or
row and row.
10) With system self-check, system failure protection and system recovery function.

1.2 Intended use

Main target is rinsing for immune test analysis panel, U and V 96-well microplate which used
by medical institution.

1.3 Instrument characteristics

1.3.1 Instrument Features

LCD 240*128 dot matrix LCD


Keyboard 8key
Microplate spec. flat, U and V type96-well microplate panel
Distributing head type 8, 12 well
Residual volume ≤2µL/well
Dispense precision CV≤3% (typical)
Wash program 50 groups
Wash row 1-12 row (can be adjusted freely)
Dispense position +2mm (can be adjusted slightly)
Distributing head’s horizontal position +2mm (can be adjusted slightly)
Distributing head’s vertical position +2mm (can be adjusted slightly)
Dispense volume 50~450µL/well, 50uL / step
Washing circle time 1~20 times(can be adjusted )

1
Vibrating time 0-60 sec. (can be adjusted)
Immersing time (1~90) sec. (can be adjusted)
Aspiration time (1~10)sec. (can be adjusted)
Weight about 5kg
Dimension 400mm×330 mm×170mm
Input power ≤250W
Fuse current 3A

1.3.2 Power and Environment requirements

Power supply: AC 100V~240V, 50/60Hz


Work environment: 5℃~40℃,RH≤80%
Storage environment: -10℃--50℃, RH≤80%
Atmospheric pressure: 70.0kPa -106.0kPa

1.4 Constitution

The instrument is composed of circuit, pipe system, display and accessory.


The instrument’s front panel is shown in Figure1-1.

Display
Keyboard

Plastic
cover

Figure1-1 Front view

LCD screen: Indicates work status of the instrument.


Plastic cover:Prevent dust from entering into microplate inside.
Keyboard:User can operate the instrument by it, which is shown in Figure 1-2.

2
Figure1-2 Keyboard panel

Keyboard function:
PROG Enter Program menu for setting operation mode about instrument and
parameter.
Cursor glints showing menu choice increasing, select one program means
content plus one.
Cursor glints showing menu choice decreasing, select one program
means content minuses one.
ENTER Enter or exit when item is chosen. Choose current item when cursor is glinting.
Save current parameters when “wash program” is chosen.
MAINT Enters into maintenance menu and performs multiple maintenance functions
CANC Cancels current operation and returns to main menu, meanwhile, the system
will not save the settings. Stops running current program or resets the
system when wash program is running.
SYS Setup system parameters which wash program will run according to it.
RUN Perform current selected mission.
Instrument rear view is shown in Figure1-3:

3
1.Vacuum plug

4.Power switch

7.Sensor

5.Power socket 6.Fuse


2.Waste plug 3.Detergent plug

Figure1-3 Rear view

1、Vacuum plug: Using for negative pressure;


2、Waste plug: Waste liquid interface port, use for connect external waste bottle
3、Detergent plug: Using for external Detergent bottle;
4、Power switch: Use for turn on or off the power of instrument;
5、Power socket: Use for connect power supply(220V);
6、Fuse: Check the fuse if there is no power when the instrument’s switch is turn on;
7、Sensor: Connecting the sensor, use for check the waste liquid is full or not.

4
Chapter 2 Operation principle

2.1 Product overview

Under the control of microcontroller, the instrument performs the wash program which was set
in advance. Under the drive of stepper motor, detergent is injected equably into microplate-wells
which are chosen by user, and sucks away waste-liquid by sucking needle after reacting between
detergent and waste of microplate-wells wall.

2.2 Pipe system

2.2.1 Function diagram

Function diagram is shown as Figure 2-1

Figure 2-1 Function diagram

2.2.2 Principle

Under the control of liquid pump, liquid in the detergent bottle is injected into microplate-wells.
Then the aspiration-needle sucks waste-liquid into waste bottle after reacting between detergent
and waste. The instrument completed washing work by cycled performing thus program.

2.3 User convenient interface

The instrument uses 240*128 dot matrix LCD display screen for easy operation, and all
parameters are listed on a single screen, which implements visual operation.
Interface divides into program, system and maintenance three main menus, program interface
is defaulted for standby interface. In order to make instrument possess intelligence to react
multiple operation, the system adopt pop up dialogue box form, which giving hints to every
operation.

5
Chapter 3 Installation and power on

3.1 Unpacking

a) Please take the instrument and accessories from the packing box carefully, reserve the
packing material for the use of transportation or storage afterwards.
b) Please check the accessories according to the packing list.
c) Please check if there is spilling detergent or water soaking about the instrument.
d) Please check there is any mechanical damage about it
e) Please check all external-expose lead, inserted part and accessory.
If you have any questions, please contact local dealer immediately.

3.2 Installation

3.2.1 Environment requirements

a) Avoid direct sunlight.


b) Avoid expose in excessive humidity and too high or too low temperature environment.
To ensure instrument’s normal operation, the suitable ambient temperature is between 5℃ to 40℃,
relative humidity should less than 80%.
c) By all means to keep the instrument from dust.
d) Keep the instrument from the equipment can disturb the test result such as centrifuge,
X-ray machine, display, copier, ect.
e) Do not use mobile telephone, radiophone which can produce strong radiation field. Too
strong radiation field will disturb instrument’s normal working.
f) Voltage’s regular fluctuation will lead to the reduction of function and reliability, user should
solve this problem before using this instrument, for example, AC voltage stabilizer(user should
equip it themselves).
g) The regular interruption of power supply will lead to the reduction of function and reliability,
user should solve this problem before using this instrument, for example,Uninterrupted Power
Supply -----UPS(user should equip it themselves).
h) User have duty to make sure the reliability of power protection ground in using place,
ensure the normal connection between plug ground and socket ground.
i) Do not lay instrument on the place which hard to turn off power switch.
j) By all means to ensure good ventilation.

3.2.2 External connection

a) Take the detergent bottle and waste bottle out from packing box, place them on the back of
instrument, we suggested place detergent bottle on the left side, waste bottle on the right side.
b) Butt the pipe with two colors junction (green and black) joint with the matched colored
bottle on the back of instrument, and screw the nut.
c) Insert the sensor plug on the waste bottle cap into sensor socket on the back of

6
instrument’s plate.
d) Pour detergent into detergent bottle.

3.2.3 Wash head installation

Connect the dispense pipe which using for linking distributing head with pipe junction the
working table according to matched colors.
Make the distributing head parallel supporting arm and put distributing head above supporting
arm, adjusting the adjust screw on the supporting arm to let distributing head parallels working
table.

Installed distributing head is shown as Figure 3-1.

Distributing Distirbuting Supporting Microplate


head pipe arm panel

Tray

Figure 3-1 Installed distributing head

3.3 Power on

a) Place the microplate panel on the worktable.


b) Pour the detergent into detergent bottle.
c) Connect power supply cable, one side to the power plug in the back of plate, the other side
to the power supply(100V-240v).
d) Turn on the power supply on the back of panel, Figure 3-2 appears, instrument begins to
initialize system.

7
China Guilin Urit

Medical Co., Ltd.

URIT-670 Microplate Washer

Figure3-2 Power on interface


e) The instrument begins to self-check after its initialization, the self-check has completed,
self-check report shown as Figure3-3.

Self-test report

HOR parts: Normal

VER parts: Normal

EEPROM status: Normal

Bottle status: Unfull

Figure3-3 Self-check report


f) Instrument can only enter standby main menu interface when self-check has completed
and there is no system fault, which is shown as Figure 3-4.

Was hing p rog ram 01


1) Washing m ode: W hole

2) Aspiration mo de: Du al

3) Washing cycle: 01

4) Sele ct row: ABCDEFGH

5) Rins e volu me : 200u L

6) Soaking tim e: 05s

7) Aspiration time: 01s

Figure 3-4 Standby main interface

8
Caution: The instrument will keep warning and refuse to enter into work
interface due to any system errors such as self-check failure or abnormal operation, shut
the instrument off and restart it after a while, if the problem still exist, cut off power and
contact our company’s service department.

Caution: Do not place instrument on the position that hard to touch power
switch.

Caution: Accessory attachments please see 3.2.2, 3.2.3.

9
Chapter 4 Operation

4.1 Program

Instrument enters Program interface directly after power on, which is shown as Figure 4-1.
Program menu contains program No., washing mode, Aspiration mode, wash cycle, selsct row,
rinse volume, soaking time and aspiration time , these parameters can be set according to
customer’s need. Program No.

Washing program 01
1) Washing mode: Whole

2) Aspiration mode: Dual

3) Washing cycle: 01

4) Select row: ABCDEFGH

5) Rinse volume: 200uL

6) Soaking time: 05s

7) Aspiration time: 01s

Figure4-1 Program menu

4.1.1 Program No.

Program No. places right “washing program”, which is shown in Figure 4-1. User can set 50
groups different parameters’ wash program, and system will store these. And the instrument will
run according to your settings when you chose one group of programs and press “run” key.
You can choose wash program by pressing “▲”or “▼” key, when you make your choice,
matched parameter will refresh automatically.

4.1.2 Washing mode

Washing mode divides into whole washing and row washing.


Whole washing: Dispense-liquid row by row, then sucking-liquid row by row. Generally
whole-washing recommended.
Row washing: Dispense row according by row and aspiration secondly, then enter the next
row.

10
4.1.3 Aspiration mode

Aspiration mode divides into 3 types---single, double and median.


Single: Aspiration needle of distributing head only sucks at one point, which used to
plat-bottom microplate.
Dual: Aspiration needle of distributing head only sucks at two points, which used to plat-bottom
microplate.
Median: distributing head carry out detergent in the middle, which used to U and V type
microplate.

4.1.4 Washing cycle

Wash cycles time:Set time freely between 1 to 20, instrument will repeat this operation.

4.1.5 Select row

Set washing rows needed, for example, max row is 8 when distributing head’s quantity is 12,
matched code is ABCDEFGH; max row is 12 when distributing head’s quantity is 8, matched code
is ABCDEFGHIJKL. Choose matched code, selected row will be washed. When distributing
head’s quantity is 8, panel washing interface is shown as Figure 4-2:

Washing program 01
1) Washing mode: Whole

2) Aspiration mode: Dual

3) Washing cycle: 01

4) Select row: ABCDEFGHIJKL

5) Rinse volume: 200uL

6) Soaking time: 05s

7) Aspiration time: 01s

Figure 4-2 8-needle distributing head interface

4.1.6 Rinse volume

Rinse volume can be set freely between 50 to 450uL, 50uL increasing every time.

4.1.7 Soaking time

Soaking time is the waiting time which the beginning of aspiration after dispense of washing
head completed, startup immersing function can increase cleaning effect, which can be set freely
between 0 to 90 seconds.
11
4.1.8 Aspiration time

Aspiration time is the time aspiration needle aspirates liquid in the microwell bottom, which can
be set freely between 1-10 seconds. Aspiration time shorter is, washing speed faster is, but be
sure to protect effect. If residual volume too large, aspiration time can be increased. System
aspiration time is 1 second.

4.1.9 Program operation

a) Instrument enter standby main interface after self-check, standby main interface will
acquiescence to program menu interface, which is shown as Figure 4-1.
b) Press “▲”or “▼” key, glinting cursor appears on the screen, which is shown as Figure 4-3.
Continue to press “▲” or “▼”key, make the cursor move to the program want to edit.

Washing program 01
1) Washing mode: Whole

2) Aspiration mode: Dual

3) Washing cycle: 01

4) Select row: ABCDEFGH

5) Rinse volume: 200uL

6) Soaking time: 05s

7) Aspiration time: 01s

Figure 4-3 Program interface


c) Press ”ENTER” key, now cursor stop glinting, luminance bar appears on program
parameter, press “▲”or “▼”key, edit the content of program.
d) Press ”ENTER” key,Store current program when edit completed, cursor start to glint, user
can edit other programs according to “▲”or “▼”key at the same time.
e) When all programs editing completed, press “▲”or “▼”key to make the washing program
is chosen by luminance bar, then press “ENTER” key, 4-4 interface appears showing edit is
succeed.

Parameters have been stored!

Figure 4-4 Store interface


f) Row choice operation
This instrument can wash random rows, user must set parameters before washing, otherwise
working status will in abnormal.
Operation: Firstly move cursor to this program, press “ENTER” key, cursor will stay on the first
12
row, press “▲”or“▼”key can choose this row or not, when this row has already set, press “ENTER”
key, move to next one, repeat the same operation to set other rows. When last row finished, press
“ENTER” key to quit, cursor will glinting, press “▲”or “▼”key can choose other programs.

Caution: Results only can be stored when “washing program” is chosen by

pressing “ENTER” key. System won’t store all edited data automatically if it doesn’t press
“ENTER” key; press “RUN” key when “whole ” or “program No.” is chosen, instrument will
store current parameter automatically and begin to operate whole washing.

Caution: Instrument will not store parameters by pressing “ENTER” key when all

the parameters are not edited.

Caution: Press “▲”or “▼”key will edit program when cursor glints; press “▲”or

“▼”key will edit program content when program is chosen.

4.2 System settings

Press “SYS” key on the keyboard, “system settings” menu on the main interface of standby
which is shown as Figure 3-4 will appear.
System setting menu including the setting of dispense location, distributing location HOR and
VER, bottom rinsing, shaking time, distributer spec and keypad sound, which is shown as Figure
4-5

System settings
1) Dispense location +0
2) Distributing location
Hor +0 Ver +0
3) Bottom rinsing ON
4) Shaking time 00s
5) Distributer spec 12
6) Keypad sound ON

Figure 4-5system setting

4.2.1 Dispense location

Dispense location is the position when distributing head in dispense situation, dispense needle
bottom relative to the microplate’s level. Appropriate position will prevent liquid’s overflow, in
general, user cannot change system’s default setting.
When washing no microplate appears liquid’s overflow, using this function to adjust ±2mm can
meet different requires.

13
4.2.2 Distributing head vertical position

Distributing head vertical position is the altitude that distributing head moving on the vertical
direction, this altitude can be adjusted ±2mm, when the distributing head move downwards to the
bottom of microplate, the altitude of supporting arm should lower than distributing head(it means
distributing head could fall on the bottom of microwell freely, otherwise Aspiration halfway).

4.2.3 Distributing head horizontal position

Distributing head on the relative place which in microwell of the first row, this place can be
adjusted by ±2mm, suitable position can obtain better effect of aspiration, in addition, liquid will not
spray on the wall of microwell or outside.

4.2.4 Bottom rinsing

Bottom rinsing dispenses detergent when dispense needle approaching the bottom of
microplate well, which can increase washing effect. User can turn it on or off.

4.2.5 Shaking time

Shaking will make washing effect better. Shaking time can be set between 0 to 60 seconds
freely. Shaking will be closed when time is 0.

Caution: We suggest close vibrating to avoid overflow if liquid is full in microwell.

4.2.6 Distributing head standard

Distributing head specification can choose 8 needles or 12 needles, standard is 12 needles.


8-needle is forbidden if user doesn’t equip, otherwise instrument may in abnormal status. Place
the microplate vertically when user choose 12 needles; place the microplate horizontal when user
choose 8 needles.

Caution: System need edit relative parameter setting when switch 8 needles or 12

needles. After switch, Program menu data will recover to default data.

Warning: Make sure the following is right when replace distributing head,:

A. The specification of distributing head should set correctly in system setup.


B. The direction of placing microplate should be right.
C. The displacement of distributing head is OK.
For example, you should set the specification for 8 needles in system setup when you
replace 8 needles distributing head, placing the microplate horizontally on tray, replacing
the distributing for 8 needles and begin to washing panel, otherwise, some parts of
instrument may damaged.

4.2.7 Keypad sound

Keypad sound can close or not according to user’s need.

14
Caution: Only press key doesn’t make sound when keypad sound is set for close,

but some systems’ hints or warnings are still sufferance to produce sound.

4.2.8 System setting operation

a) In system setup interface, press “▲”or” key, glinting cursor appears on the screen which is
shown as Figure4-6

System settings
1) Dispense location +0
2) Distributing location
Hor +0 Ver +0
3) Bottom rinsing ON
4) Shaking time 00s
5) Distributer spec 12
6) Keypad sound ON

Figure 4-6 system setting edit


b) Press “▲”or “▼”key, move cursor to the program want to edit; press ”ENTER” key, cursor
stop glinting, luminance bar appears on program parameter.
c) Press “▲”or “▼”key to edit program parameters.
d) Press “ENTER” key when edit is completed, store the parameter and the cursor begin to
glinting.
e) Repeat the step from a) to d), which can edit other parameters.
f) Press “▲”or “▼” key when all the parameters completed, make luminance bar appears on
“system settings ”, press “ENTER” key, interface of “Programs have been stored” is shown in
Figure4-4, instrument return to system setup interface.
g) Press “CANC” key, return to stand-by main interface.

4.2.8.1 Dispense location adjustment

In normal, dispense location setting is forbidden, unsuitable adjustment will lead to abnormal
status or unexpected effect when wash program is running. User can adjust dispense position to
match it If microplate is not standard,
Firstly, select this program, press “ENTER” key, stop it after system location, press “▲” or “▼”
key to adjust aspiration needle position, up or down, observing condition when adjusting it as far
as to get satisfied effect, then press “ENTER” key to quit.

4.2.8.2 Distributing head adjustment

Distributing head adjustment is similar to dispense adjustment. When distributing head


15
horizontal position adjusted completed, press “ENTER” key to adjust vertical position.

Caution: When user adjusting “Dispense location” and “distributing head location”,

press “ENTER” key to select, instrument will drive distributing head and tray to pointed
position. Other operation will inefficacy before this operation has finished.
Press “▲” or “▼” key each time, instrument will reflect the set situation immediately
(only make effect to press key when operation has finished and sound occur), user can
observe this situation to adjust instrument.

4.3 Whole washing

Carry out rinsing after program and setting, the details as following:
a) Place the microplate you want to wash on the tray.
b) Press “▲”or“▼” key, “Program No.” is chosen by glinting bar; press “▲”or “▼” key, choose
program No. want to operate, press ”ENTER” key.
c) Press “▲”or“▼”key, make the “washing program” is chosen by glinting bar.
d) Press “run” key on the keyboard, instrument begins to rinsing according to set data,
washing parameter showing on the LCD, lowest two lows on the bottom show current progress,
user can know progress according to it, process interface is shown as Figure 4-7.

Program 01# is running


Total:02 time(s), 08 row (s)
200uL, ASP 01S,Whole,dual
SOA 05S,SHK 00S, bottom
Now: 01 th time,01 th row
01 time(s) left

Figure 4-7 Washing process


e) Rinsing process completed after progress bar is full, buzzing from instrument show
operation finished, interface shown as Figure 4-8 appears and after 2 seconds return to main
interface.

Program 01# completed!

Figure4-8 Washing panel completed


f) Press “PROG” key at this time, return to Program interface; press “MAINT” key, enter
system maintenance interface; press ”SYS” key, enter system setup interface.

Caution: By press “RUN” key will in effect only when “Washing program”or

“program ON” is chosen by luminance bar.

Caution: Make sure the specification of distributing head correspond with the

direction of placing microplate.

16
4.4 Maintenance

Press “MAINT “key on the keyboard, enter “maintenance” menu under the stand-by main
interface shown as Figure 3-4. Maintenance menu including Prime, Rinse, System reset,
Equipment self-test, system recovery and dispense check, maintenance interface is shown as
Figure 4-9.

M ainte nance
1) P rime

2) R inse

3) S ys te m res et

4) E q uipm en t self-te st

5) S ys te m rec overy

6) D ispens e check

Figure 4-9 Equipment maintenance menu

4.4.1 Prime

Prime is turgoring and cleaning for system pipeline in high speed, making instrument’s
accuracy in working. Priming every time when it using for the first time or changing detergent.
Priming can solve for well blocking.

4.4.2 Rinse

Rinseis cleaning for system pipeline, distributing head, dispense needle and aspiration needle,
making expedition of airline and liquid line. Execute rinsing one time or more to make the
instrument working normally for the first using or after long keeping.

4.4.3 System reset

Instrument can reset correctly when “system reset” about distributing head and tray completed
in any place. Distributing head located above washing groove after reset.
Report of error will pop up if system reset failed to remind user the error type. Instrument can’t
work normally if system reset failed, it needs mending at this time.

4.4.4 Equipment self-test

Equipment self-test operation carry out every time automatically when instrument turn on.
Instrument has sever fault will lead to instrument working damage. In this case, protection
measure will on, error report popes up at first, warning lasting, instrument cannot work normally.
Turn off power supply and reset after a while if above situation appears; some part of this
17
instrument has already damaged when problems is lasting, please contact Urit company to check.
Use this self-check function can realize the health condition in normal using.

4.4.5 System recovery

Execute this function will make each parameter recover to factory installation.
Use this function can recover instrument system when it running or adjusting.

Caution: Memorizer may in fault if system recovery failed or messy code appears

can’t recover, contact Urit company to check.

4.4.6 Dispense check

Use this function to check dispense volume is exact or not and path of liquid is blockage or
not,or need washing or not. After operation, instrument will stop operation aspiration after inject
detergent to microwell.

Caution: Dispense volume is program No. dispense volume in program menu, user

can set dispense volume according to “washing program”, select “Dispense check” in
maintenance menu to operate.

4.4.7 Maintenance operation

a) Under the equipment maintenance interface shown as Figure 4-9, press “▲”or“▼” key,
equipment maintenance interface appears glinting cursor shown as Figure 4-10.
b) Press “▲”or “▼” key, move the cursor to the maintenance program need to operate.
c) Press “RUN” key on keyboard, instrument begins to carry out matched maintenance
program.
d) Instrument returns to the equipment maintenance menu interface shown as Figure 4-9
when maintenance program completed.
e) Press “CANC” key, return to standby main interface.

M ainte nance
1) P rim e

2) R inse

3) S ys te m res et

4) E q uipm en t self-te st

5) S ys te m rec overy

6) D ispens e check

Figure4-10 Maintenance operation interface


18
4.5 Power off

Turn off power switch on rear panel, cut off power supply when instrument doesn’t running.

19
Chapter 5 Attentions

To ensure user safety, keep the accuracy of experiment, avoid the damage of instrument in
case of operation’s fault, following fairs should pay attentions:

5.1 Protection

Infectious specimens and corrosive liquid can often be met in experiment, put on antiseptic
experimental clothing and glove in the operation, Deal with dangerous samples according to
laboratory operation handbook.

5.2 Before operation

a) Install instrument according to this manual strictly, ensure the correct connection of
pipeline to avoid leakage.
b) Operate rinsing and priming more than 2 times to ensure expedition of pipeline and airline
when use washer again after long time storage.
c) Use the hose provided by Urit company for connection.

5.3 In the operation

a) Confirm the followings before operate rinsing:


1) Distributing head is installed correctly, quantity correspond with the one in “system
settings”, otherwise it may lead to the damage about distributing head or microplate.
2) Microplate lies in right place.
3) Suitable Program and parameter data have already set.
b) Avoid waste liquid dumping in order to prevent liquid flow into pump.
c) Clean up the surface of instrument with neutral sterile pack fluid if there is any danger
biotical material on it.
d) Keep the cleanness and dryness of tray’s surface, avoid the block of aspiration needle and
dispense needle.
e) Press ”CANC” key to stop instrument in the process running , hand or object force
stopping is forbidden..
f) Do not use hand to push and pull tray and distributing head in the process of running or
washing program completed, which can avoid any damages about some parts.
g) Use needle object to dredge if dispensing is blocked.

5.4 Power safety Caution

a) Make sure the external equipment such as circuit, socket in accordance with electrical
appliance safety standard.
b) Do not use this instrument in damp environment
c) Do not turn on the instrument if there is any mechanic damage
d) Cut off all the power supply connecting to the instrument when operate any adjustments,

20
displacement, maintenance or mending about instrument.
e) Technician who is familiar with potential danger should deal with any unavoidable electric
situation.
f) Cut off power supply immediately when instrument is wetting by water, user should deal
with this situation under the instruction of technician.

5.5 Sterilization

Instrument should be sterilized regularly. Gather waste water in special vessel, explanation
after sterilization, clean up the pollutant on the surface of washing plate by 70% alcohol, aspiration
orifice should be wiped several times by 200 mg/L clean liquid, wiped by rinsing agent or rinsing
several times for stand-by using.

21
Chapter 6 Maintenance and storage

6.1 Maintenance of instrument and parts

a) Do not placed instrument under direct sunlight or heat source nearby.


b) Clean the instrument with soft cloth and neutral detergent regularly; wipe off the liquid on
the tray in time.
c) Dredge Dispense needle well when there is obvious nonunion situation in Dispense.
d) Dredge Aspiration needle well if amount of residue is increasing.
e) The storage of liquid bottle.
f) Excessive residual liquid on working table may overflow the edge and leakage to the
chassis, wipe off by soft clothe.
g) Replace liquid bottle every two years, or earlier if it under the direct sunlight or keep long
time. It can be replaced with same material one.
h) Sterilized instrument regularly, Gather waste water in special vessel, explanation after
sterilization, clean up the pollutant on the surface of washing plate by 70% alcohol, Aspiration
orifice should be wiped several times by 200 mg/L clean liquid, wiped by rinsing agent or rinsing
several times for stand-by using.

6.2 Troubleshooting

a) Screen does not display or on reaction when press keyboard, poor connection about LCD
or keyboard may the possible reason, shut the instrument off and restart it after a while.
b) System can identify trouble automatically when the instrument self-checking or running,
system self-protection start-up to protect instrument if system in severe condition, instrument
return to normal status after problems solved.
c) Identification of instrument’s damage:
1) System faults shown on LCD and self-check report pop up when power on or running
wash program; some parts on instrument have already damaged if above situation appears
too many time, contact Urit company workmen to deal with it.
2) Instrument may have partly damage and stop to using under following circumstances
① Severe external damage exists;
② Still in abnormal situation when repeat operation for many times;
③ Storage too long under the unqualified environment;
④ Transportation pressure overloads.
Contact Urit company workman to deal with any abnormal problems.

6.3 Substitutions

Liquid bottle should be changed every 2 years or earlier under sunlight or keep long time. Urit
company provides following parts, and it will irresponsible for any consequence caused by user
themselves.
1 detergent, waste liquid pipe.
2 Distributing head.

22
3 Tray.

Supplement 1:Symbols

Caution consult accompanying documents

Caution, risk of electric shock

For In Vitro Diagnostic Instrument

Biohazard

Serial Number

Manufacturer

Authorized Representative In The European Community

Recovery

23

You might also like