Astm D7112 Estabilidad de Asfaltenos
Astm D7112 Estabilidad de Asfaltenos
Astm D7112 Estabilidad de Asfaltenos
FIG. 1 PORLA Heavy and Crude Oil Stability and Compatibility Analyzer
FIG. 2 Schematic Diagram of PORLA Heavy and Crude Oil Stability and Compatibility Analyzer
--`,,`,`,,```,,,``,``,,````````-`-`,,`,,`,`,,`---
FIG. 4 PORLA Step Measurement Screen
12.1.5 Enter the weight of solvent collected in the appro- in the instrument. However, evaporation of xylene does occur when
priate screen. several sample cups sit in the sample carousel during the fully automatic
12.1.5.1 For the xylene pump rate, the instrument calculates mode of operation.
the updated pump rate in g/s and displays it in the appropriate
13. Quality Control Monitoring
field in the parameter screen.
12.1.5.2 For the n-heptane pump rate, the instrument calcu- 13.1 Confirm the performance of the instrument and test
lates the updated pump rate and displays it in mL/s in the procedure by analyzing quality control (QC) samples.
appropriate field in the parameter screen. 13.1.1 The QC sample(s) should be a stable and homoge-
12.2 Evaporation Correction Coeffıcient—Determine the neous residual fuel oil or crude oil similar in composition and
evaporation correction coefficient for the xylene by weighing viscosity to the type of samples routinely tested, and containing
(accurate to two decimal places) a sample cup containing a from 1 to 10 mass % asphaltene. Asphaltene content may be
stock solution after 0.5, 1.0, and 1.5 h of standing in the sample measured by Test Method D6560.
carousel and calculate the average solvent loss by evaporation 13.2 Prior to monitoring the measurement process, deter-
in g/h. Input this information into the PORLA parameter mine the average value and control limits for the QC sample
screen, which is accessed by means of the Parameter button in (see Practice D6299 and Section 3 of Manual 77).
the PORLA Step measurement screen.
12.2.1 Verify the evaporation correction coefficient quar- 13.3 Record QC results and analyze by control charts or
terly when the instrument is in continuous service. other statistically equivalent techniques to ascertain the statis-
12.2.2 If a specific sample contains volatile components, tical control status of the total test process.7,8 Investigate any
such as a whole crude oil, use the semi-automatic Light Ends
mode of operation. In this mode of operation, evaporation is
minimized by loading the stock solution cups one at a time into 7
ASTM MNL 7, Manual on Presentation of Data Control Chart Analysis , 6th
the carousel. See 14.9. edition, Section 3: Control Chart for Individuals. Available from ASTM Interna-
tional Headquarters.
NOTE 3—No significant evaporation of solvent or sample occurs during 8
In the absence of explicit requirements given in the test method, this clause
the titration procedure because the sample cup is sealed in the test position provides guidance on QC testing frequency.
out-of-control data for root cause(s). The results of this high, warm up the oil in an oven before shaking. A temperature
investigation may, but not necessarily, result in instrument of 80°C is normally sufficient.
recalibration. 14.1.3 Weigh out 20 g of the xylene into the sealable
13.4 The frequency of QC testing is dependent on the container (bottle), accurate to two decimal places, and then add
criticality of the quality being measured, the demonstrated 40 g of the oil, also weighed to two decimal places.
stability of the testing process and customer requirements.8 14.1.4 Shake the stock solution container until the oil has
Generally, a QC sample should be analyzed each testing day dissolved in the xylene. Ensure that the oil has completely
for routine samples. The QC frequency should be increased if dissolved before proceeding. A visually homogeneous stock
a large number of samples are routinely analyzed. However, solution mixture usually indicates complete dissolution of oil
when it is demonstrated that the testing is under statistical in the xylene.
control, the QC testing frequency may be reduced. The QC 14.1.5 In the case of oil samples that are very difficult to
sample testing precision should be periodically checked against dissolve, use a refluxing device instead of shaking the bottle.
the ASTM method precision to ensure data quality.7 When using the refluxing device, let the stock solution cool
13.5 It is recommended that, if possible, the type of QC down before pouring it into the sealable container.
sample that is regularly tested be representative of the samples
14.2 Test Information and Instrument Settings:
routinely analyzed. An ample supply of QC sample material
14.2.1 Open the PORLA Step measurement screen and type
should be available for the intended period of use, and shall be
in the operator’s name and sample name in the appropriate
homogeneous and stable under the anticipated storage condi-
--`,,`,`,,```,,,``,``,,````````-`-`,,`,,`,`,,`---
fields.
tions.
13.6 See Practice D6299 and Manual 77 for further guid- 14.2.2 Type in the oil and xylene weights, accurate to two
ance on test method QC and control charting techniques. decimal places, used in the preparation of the stock solution, in
the appropriate fields.
14. Procedure 14.2.3 Type in the density of a crude oil sample, in g/mL at
14.1 Sample Preparation: 15°C, in the appropriate field, if compatibility parameters are
14.1.1 Stock Solution—Prepare a stock solution in a seal- desired. Determine density by a method appropriate to the
able container such as a screw cap glass bottle (volume about sample type, if required.
100 mL) from 40 g of the oil under test and 20 g of xylene, as 14.2.4 Select xylene from the drop down list of aromatic
described below. solvents displayed under Aromatic.
14.1.2 Thoroughly mix the oil by vigorously shaking the 14.2.5 Before each test, ensure that there is an adequate
sample container for at least 2 min. If the viscosity of the oil is volume of solvent in the aromatic (xylene) and paraffinic
NOTE—1/X is the oil to solution ratio, where the solution is the sum of the volumes of the aromatic and paraffinic solvents (xylene and n-heptane).
FIG. 6 Typical Analyzer Plot of Stability Results
--`,,`,`,,```,,,``,``,,````````-`-`,,`,,`,`,,`---
SUMMARY OF CHANGES
Subcommittee D02.14 has identified the location of selected changes to this standard since the last issue
(D7112–05a) that may impact the use of this standard.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the ASTM website (www.astm.org/
COPYRIGHT/).
--`,,`,`,,```,,,``,``,,````````-`-`,,`,,`,`,,`---