Digital Crane Scale User Manual: Model: Ocs-S9

Download as pdf or txt
Download as pdf or txt
You are on page 1of 17

DIGITAL CRANE SCALE

USER MANUAL

MODEL: OCS-S9

USER MANUAL
Manual_OCS-S9_V1
TABLE OF CONTENTS

SAFE OPERATION GUIDE--------------------------------------------------------------------------------2

CHAPTER 1 FEATURES AND SPECIFICATION--------------------------------------------------3


1 Features---------------------------------------------------------------------------------------------------3
2 Main technical data-------------------------------------------------------------------------------------3
3 Specification----------------------------------------------------------------------------------------------4
4 Appearance illustration--------------------------------------------------------------------------------4
5 Power supply------------------------------------------------------------------------------------------ -5

CHAPTER 2 SCALE&INDICATOR DISPLAY--------------------------------------------------------5


1 Scale display----------------------------------------------------------------------------------------------5
2 Mobile display ------------------------------------------------------------------------------------------- 6

CHAPTER 3 OPERATION GUIDE-----------------------------------------------------------------------7


1 Turn on/off---------------------------------------------------------------------------------------------------7
2 Zero-----------------------------------------------------------------------------------------------------------7
3 Tare------------------------------------------------------------------------------------------------------------8
4 Peak----------------------------------------------------------------------------------------------------------8
5 Hold----------------------------------------------------------------------------------------------------------8
6 Accumulation------------------------------------------------------------------------------------------------8
7 Accumulation Search--------------------------------------------------------------------------------------8
8 Accumulation Clear----------------------------------------------------------------------------------------8
8 Battery Voltage----------------------------------------------------------------------------------------------9
10 Set Point-----------------------------------------------------------------------------------------------------9

CHAPTER 4 PARAMETER SETTING & CALIBRATION--------------------------------------------10


1 Parameter setting----------------------------------------------------------------------------------------10
2 Calibration-------------------------------------------------------------------------------------------------12

CHAPTER 5 DISPLAY ILLUSTRATION------------------------------------------------------------------14

CHAPTER 6 TROUBLESHOOTING GUIDE------------------------------------------------------------15

USER MANUAL 1
SAFE OPERATION GUIDE:
1. The capacity of crane scale have been showed on the overlay, Don’t make lifts beyond
rated capacity of the crane scale.

2. Operators should not maneuver a loaded crane scale over personnel.

3.The scale must be routinely checked for operation. Don’t operate if hook, shackle, sling,
cable, etc show any sign of defects or excessive wear.

4. If the scale is not going to be used for long time, recharging the battery every three
months to preserve the life-span of the battery and recharge it before operation.

5. The battery charger is included in the scale package. Please use this charger, It’s
normal that the charger will become warm when the battery is charging.

6. Please note when the digits flash on the display, it means that the battery needs to be
recharged.
7. Check hook, shackle, safety pins, and latches periodically. Contact your dealer for

parts replacement in case defect, deform or wearing is found.

8. Always lift loads vertically.

USER MANUAL 2
CHAPTER 1 FEATURES AND SPECIFICATION
1. Features
◆ Blue-tooth wireless communication and display
Many functions on Mobile(android system). Such as tare, zero, peak, accumulation. Two display,
one display on Mobile,one display on crane scale,
◆ Two set point
Two user programmable Set-Point can be used for safety and warning applications or for limit
weighing.
◆ User-friendly design:
Cast aluminium scale housing for maximum protection; Large red 5 digit LED display (digit height
30mm); Gravity compensation; Safety factor up to 4 times capacity; Battery Operated with low
battery indication; Automatic turn off , power-saving function.

2. Main technical data


Accuracy Standard According to OIML class III
Scale/Indicator Display 30mm (1.2’’) 5digits LED/25mm(1”) 5digits LCD
Zero Range 4% F.S.
Tare Range 100% F.S.
Stable Time ≤10 seconds
Overload Indication 100% F.S. + 9e
Max. Safety Load 125% F.S.
Ultimate Load 400% F.S.
Scale/indicator Battery Life ≥70 hours
Scale Battery Type Fully sealed Lead-Acid battery, 6V/7Ah
Scale Adapter DC9V/1000mA
Nominal operating Temp. - 10℃ ~ + 40℃
Operating Humidity ≤85% RH under 20℃
Wireless Distance 50M without obstacle

USER MANUAL 3
3. Specification
Max. capacity Min. weigh Division Total display counts
MODEL
(kg) (kg) (kg) (n)
OCS-SB-1 1000 10 0.5 2000
OCS-SB-2 2000 20 1 2000
OCS-SB-3 3000 20 1 3000
OCS-SB-5 5000 40 2 2500
OCS-SB-10 10000 100 5 2000
OCS-SB-15 15000 100 5 3000
OCS-SB-20 20000 200 10 2000

4. Appearance illustration

MODEL A(mm) B(mm) C(mm) H(mm) N.W.


OCS-SB-1 80 340 30 420 12kg
OCS-SB-2 80 340 30 420 12kg
OCS-SB-3 100 340 40 600 14kg
OCS-SB-5 130 360 45 720 23kg
OCS-SB-10 145 360 60 850 43kg
OCS-SB-15 170 360 85 900 60kg
OCS-SB-20 170 360 85 900 65kg

USER MANUAL 4
5. Scale power supply
Battery: 6V/7Ah fully sealed Lead-Acid battery
Adapter: 9VDC/1000mA adapter
Current: the crane scale average current is about DC 100mA±10%, if battery charged, it can be used for
about 50hours.Low battery warning: When the digits flash, it means the battery needs to be recharged.
The scale will power off automatically after 1 to 2 hours operation if the battery is not charged. Please
recharge the battery before operating the scale again.
Charging method: Turn off the crane scale, use adapter in the carton, the charge lamp on the back cover
will become orange when charging, and it will become green when the charging finish.
Generally, charging time should not less than 8hours, the crane scale have charging protection, long
time charging can’t damage the battery.

CHAPTER 2 SCALE & MOBILE DISPLAY

1. SCALE DISPLAY
1. LED DISPLAY
30mm (1.2’’) 5digits LED/LCD
2. INDICATOR LAMP
a. TARE: TARE indicator
b. STB: STABLE indicator
3. KEY
a. ON/OFF
1. Press ON/OFF three seconds,the crane scale turn on.
2. When turn on,Press ON/OFF three seconds,the crane scale turn off.
b. TARE
1. Hang sling or container on the hook, when the weight reading stable, press this key, the digits
display “0”, and “TARE” indicator switch on.
2. Put the goods into container, the scale will display goods net weight.
3. If move goods and container, the scale will display minus value of tare weight.

USER MANUAL 5
2. Mobile display
1. Install APP software to mobile
This is android system! Android 2.0
1. Download the APK software(Everight-1-2P) to mobilephone
2. Running the software
3.Search and press button “BT-2015”

The mobilephone will display as follows:

2. Connect Mobilephone with scale through bluetooth


Press CONNECT button, choose Connect a device-Secure. The mobilephone display

First time there are no paired devices inside, press “Scan for device” to search paired device.
After the paired device display, press it.
It will request you input device pin, input pin: 6666,then press Enter, the mobilephone will display

Of course, you must turn on scale at first.


From the right corner, you will see connected: Everight , now the screen display same content with the
scale.

CHAPTER 3 OPERATION GUIDE


Note:( ) means the key on the mobile phone
【 】means the key on the remote control
〖 〗means the display content
1 TURN ON/OFF
◆Turn on scale
OPERATION DISPLAY ILLUSTRATION
Press ON/OFF 〖88888〗 Display two times
on the scale 〖 Ert 〗 Display two times, ERT is abbreviation of Everight
〖u 2.11〗 Display current software version 2.11
〖XXXXX〗 Display capacity
〖U 6.XX〗 Display current scale battery voltage
〖–––––〗 Waiting stable
〖 0〗 “STB” light on, you can operate now

◆Turn off scale


1 Press ON/OFF on the scale 3 seconds.
2
Press【OFF】Key on remote control

2 ZERO
OPERATION DISPLAY When scale turn on, generally scale displays〖 0〗,if
Press (ZERO) 〖 0〗 display small digits when no load, press this key.

Press【ZERO】 〖 0〗
USER MANUAL 7
3 TARE
◆ TARE DIRECTLY
OPERATION DISPLAY After turn on, hanging tare weight such as sling, cable
1 Press (TARE) 〖 0〗 at first, press this key, “TARE” light on, then the scale
2 Press TARE on 〖 0〗 will display net weight of the goods.
the scale
3 Press【TARE】 〖 0〗

4 PEAK
OPERATION ILLUSTRATION
Press (PEAK) Hold the maximum weight reading when load change
Press (PEAK) again Return to normal weighing status

5 HOLD
OPERATION ILLUSTRATION
Press 【HOLD】 Hold the weight reading no matter load change
Press 【HOLD】 again Return to normal weighing status

6 ACCUMULATION
OPERATION DISPLAY ILLUSTRATION
Press【ACCU】 〖No***〗 Current accumulation time
or (ACCU) 〖H****〗 Front four digits of total value
〖L****〗 Rear four digits of total value
After display above contents two times, the scale will return automatically, you also can press
【SHIFT】to return.

7 ACCUMULATION SEARCH
OPERATION DISPLAY ILLUSTRATION
Press 【 SHIFT 】 〖No.***〗 Current accumulation time
【 ACCU 】 or 〖 ***** 〗 Current weight reading
(search ACCU)
〖H****〗 Front four digits of total value
〖L****〗 Rear four digits of total value
【SHIFT】 〖*****〗 Return
Press【TARE】and【ZERO】can check different times weight and accumulation.

8 ACCUMULATION CLEAR
OPERATION DISPLAY ILLUSTRATION
【CLEAR】or (CLEAR) 〖 CLr 〗 Ask if you want to clear or not
【CLEAR】or (CLEAR) 〖noCLr〗 Cancel clear
【SHIFT】 〖8.8.8.8.8.〗 Confirm clear

USER MANUAL 8
9 BATTERY VOLTAGE
OPERATION DISPLAY ILLUSTRATION
【SHIFT】【F1】 〖U 6.**〗 Display current battery voltage
【SHIFT】 Return
The voltage should be between 5.8 and 6.9, if below than 5.8, the digits flash, you need to
recharge immediately.

10 SET POINT
There are two user programmable Set-Point can be used for safety and warning applications
or for limit weighing.
_ 1 LO SP1 1 HI 2 LO SP2 2 HI

OPERATION DISPLAY ILLUSTRATION


Press 【F1】and 【F2】at same 〖P0000〗 Second digit flash,use function key to move digit,
time 3seconds set value.
Press【TARE】 【ZERO】
【CLEAR】〖P8888〗 Input 8888,8888 is parameter setting password

Press【SHIFT】 〖SETUP〗 Confirm, enter parameter setting program


Press【TARE】 〖SP1 〗 Press【TARE】until it display SP1, Set point 1
Press【TARE】 〖1 OFF〗 New dynamometer, it display 1 OFF, ifparameter
set, it display set value.
Press【ZERO】 to choose 〖1 HI〗or There are 3 choice,1 OFF means you don’t set
〖1 LO〗 point, 1 HI means the indicator will alarm when load
exceed the value you set, 1 LO means the indicator
will alarm when load smaller than the value
Press【TARE】to confirm 〖02000〗 New dynamometer it display 02000, if parameter
set, the screen display set value.
Press【 ZERO 】and【ACCU】〖01000〗 Set “1 HI” or “1 LO” Value. e.g. 1000kg
to change value

Press【TARE】 〖 SP2〗 Set point 2


Press【TARE】 〖2 OFF〗 New dynamometer, it display 2 OFF, if parameter
set, it display set value.
Press【ZERO】to 〖2 HI〗or There are 3 choice,2 OFF means you don’t set
choose 〖2 LO〗 point, 2 HI means the indicator will alarm when load
exceed the value you set, 2 LO means the indicator
will alarm when load smaller than the value

Press【TARE】 〖01200〗 New dynamometer it display 01200, if parameter


set, the screen display set value.
Press【 ZERO 】and【ACCU】〖03000〗 Set “2 LO” or “2 HI” Value. e.g.3000kg
to change value
If you needn’t to set point, you only need to press 【TARE】until it display next function 〖g〗

USER MANUAL 9
CHAPTER 4 PARAMETER SETTING &CALIBRATION
You need to use remote controller to operate parameter setting and calibration
Key function:
KEY FUNCTION
【TARE】 Move digit(the digit choose flash)
【ZERO】 Set value up
【CLEAR】 Set value down
【SHIFT】 Confirm, exit

1. PARAMETER SETTING
OPERATION DISPLAY ILLUSTRATION
Press ON/OFF 3seconds 〖 0〗 Turn on crane scale
Press 【F1】and 【F2】 〖P0000〗 Second digit flash,use function key to move
at same time 3seconds digit, set value.
Press 【 TARE 】【 ZERO 】 〖P8088〗 Input 8088,8088 is parameter setting password
【CLEAR】
Press【SHIFT】 〖SETUP〗 Confirm, enter parameter setting program
Press【TARE】 〖FS=06〗 New scale,the screen display FS=06,if scale
calibrated,the screen display capacity.
Press【ZERO】or【CLEAR】 〖FS=05〗 Choose capacity , the screen display
to choose 02/03/05/06/10/15/20/30/50/60/75 circularly ,
take cap:5000kg as example, choose FS=05
Press【TARE】confirm 〖Id=05〗 New scale,the screen display Id=05, if scale
calibrated , the screen display calibrated
division.
Press【ZERO】or【CLEAR】 〖Id=02〗 Choose division , the screen display
to choose 01/02/05/10/20 circularly. 5000kg scale division
is 2kg, so choose Id=02
〖Pt=0〗 New scale,the screen display Pt=0, if scale
calibrated , the screen display calibrated
decimal position
Press【ZERO】or【CLEAR】 〖Pt=0〗 Choose decimal position, the screen display
to choose 0/1/2/3 circularly, 0=xxxxx, 1=xxxx.x, 2=xxx.xx,
3=xx.xxx,Cap.5000kg scale choose Pt=0
Press【TARE】confirm 〖Ab=24〗 Display zero range , A: zero range by hand; B:
zero range automatically, 0~5 total 6 options:
0=0%F. S; 1=2%F.S; 2=4%F.S; 3=10%F.S;
4=20%F.S; 5=50%F.S,Generally,Don’t change
this parameter.
Press【TARE】confirm 〖Cd=11〗 New scale,the screen display Cd=11,if scale
calibrated,the screen display calibrated value.
USER MANUAL 10

Press【ZERO】or【CLEAR】 〖Cd=11〗 C: zero-tracking range, 0~5 total 6 options;


to choose
0=0d; 1=0.5d; 2=1d; 3=1.5d; 4=2d; 5=2.5d
d: display speed,0~2 total 3 options; 0=slow;
1=average; 2=fast; before calibration, set
Cd=00,after calibration, set Cd=11 again,
generally choose Cd==11 when ex-stock..
Press【TARE】confirm 〖LL=2〗 New scale,the screen display LL=2,if scale
calibrated,the screen display calibrated value.
Press【ZERO】or【CLEAR】 〖LL=1〗 LL:filter parameter,0~5 total 6 options, from
to choose smallest to biggest , before calibration, set
LL=0,after calibration, set LL=1 again, generally
choose LL=1 when ex-stock..
Press【TARE】confirm 〖Un=0〗 New scale,the screen display Un=0, if scale
calibrated,the screen display calibrated unit.
Press【ZERO】or【CLEAR】 〖Un=0〗 Choose unit,Un=0:kg, Un=1:lb, Un=2: other
to choose unit。Generally, choose Un=0 when ex-stock.
Press【TARE】confirm 〖OFF 1〗 New scale,the screen display OFF 1,if scale
calibrated,the screen display calibrated value.
Press【ZERO】or【CLEAR】 〖OFF 2〗 Choose turn off method, 0~2 total 3 options; 0:
to choose turn off by hand; 1: if no operation, the screen
display 〖 -〗after 15minutes to save power;2:
if no operation, the screen display 〖 -〗after
15minutes, and turn off automatically after
60minutes. Generally, choose OFF=2 when
ex-stock.
Press【TARE】confirm 〖g 〗 Acceleration of gravity
Press【TARE】,Press 〖9.7930〗 New crane scale , it display 9.7930, it our
【ZERO】and【CLEAR】 company local g, if scale calibrated, it display
to change value calibrated value. user can change it according
to local area g value, the value range is
9.783-9.832,
Press【SHIFT】 〖 End 〗 Confirm above parameter setting , Exit
parameter setting program.
1. If you find mis-operation before end, Press【TARE】, the screen will display parameter
circularly, correct error, and if you find mis-operation after parameter setting program finished,
you have to enter parameter setting program again to correct.
USER MANUAL 11

2. The capacity you choose relate to division, decimal position, below is detail list:
CAPACITY/DIVISION FS (CAP) Id (DIVISION) Pt(DECIMAL POSITION)
1000kg/0.5kg 10 05 1
2000kg/1kg 02 01 0
3000kg/1kg 03 01 0
5000kg/2kg 05 02 0
10000kg/5kg 10 05 0
15000kg/5kg 15 05 0
20000kg/10kg 20 10 0

2. calibration
OPERATION DISPLAY ILLUSTRATION
〖 0〗 Take Cap:5000kg scale as example,Pay attention that
the screen display 0 after hanging tare weight, you can
hang tare weight at first, then turn on.
Press 【F1】and 【F2】 〖P0000〗 Second digit flash,use function key to move digit, set
at same time value.
3seconds
Press 【 TARE 】 〖P1358〗 Input 1358,1358 is calibration password
【ZERO】 【CLEAR】
Press【SHIFT】 〖SCALE〗 Confirm,into calibration program
Press【SHIFT】 〖CALSP〗 Zero point calibration, wait until STB light on
Press【SHIFT】 〖LoAd1〗 First load calibration
Hanging on first load First load must exceed 20% capacity, Waiting until STB
e.g.1000kg light on.
Press【SHIFT】 〖05000〗 Display capacity
Press 【 TARE 】 〖01000〗 Input 1000, wait until STB light on.
【ZERO】 【CLEAR】
Press【SHIFT】 〖LoAd2〗 Second load calibration
Hanging on second Second load must exceed Load 1+20% capacity,
load e.g.3000kg Waiting until STB light on.
Press【SHIFT】 〖02000〗 Mention you Load2 must exceed 2000kg
Press 【 TARE 】 〖03000〗 Input actual weight 3000, waiting stable
【ZERO】 【CLEAR】
Press【SHIFT】 〖LoAd3〗 Third load calibration
Hanging on third load Third load must exceed Load 2+20% capacity, Waiting
e.g.5000kg until STB light on.
Press【SHIFT】 〖04000〗 Mention you Load3 must exceed 4000kg .
Press 【 TARE 】 〖05000〗 Input actual weight 5000kg. wait until STB light on.
【ZERO】 【CLEAR】
Press【SHIFT】 〖------ -〗 Waiting stable
USER MANUAL 12

〖 -oL- 〗
〖 End 〗 Calibration end
〖5000.0〗 Display calibrated capacity , sometimes it display
4999.5, 5000.5
Press【TARE】 〖05000〗 Move decimal position
Put down weight 〖 0〗 Check zero point, it display 0, and STB, ZERO light on.

1. First load: LOAD1 must larger than 20% capacity,and LOAD1<LOAD2<LOAD3, at the same

time, between LOAD1 and LOAD2, LOAD2 and LOAD3, the interval must bigger than 20%

capacity

2. If LOAD1 is bigger than 80% capacity, such as full capacity, after first load calibration, it will

finish automatically, you only need to press【TARE】to move decimal position. Generally, we

strong recommend you to use this one point calibration program if you have 100% capacity

standard weight

3. If you only have 20% standard weight, after first load calibration, when it display 〖LoAd2〗,

press【OFF】 2 times, then press【TARE】to move decimal position. Calibration finish.


USER MANUAL 13

CHAPTER 5 DISPLAY ILLUSTRATION


DISPLAY ILLUSTRATION REMARK
〖CALSP〗 Enter zero point calibration
〖SCALE〗 Enter calibration
〖SEtUP〗 Enter parameter setting
〖UAdJ 〗 Enter voltage calibration
〖LoAd1〗 First calibration point LoAd1<LoAd2<LoAd3
〖LoAd2〗 Second calibration point
〖LoAd3〗 Third calibration point
〖¯¯¯¯¯〗 Exceed high limit Tare weight can’t exceed full capacity
〖_____〗 Exceed low limit Tare weight can’t be negative
〖-----〗 Waiting
〖Err10〗 Weight less than Min. Capacity Can’t accumulate the value
〖Err11〗 Accumulated times overflow Can’t accumulate after 30times
〖Err12〗 Accumulated weight overflow Can’t accumulate after 99999
〖Err13〗 Error in repeat accumulation Can’t accumulate one weight repeatedly
〖no***〗 Current accumulation times
〖H****〗 Front four digit of accumulated Total weight=front four digit + rare four digit
weight
〖L****〗 Rare four digit of accumulated Total weight=front four digit + rare four digit
weight
〖 CLr 〗 Ask if you really want to delete In case error deletion
accumulated weight
〖noCLr〗 Give up deletion
〖88888〗 Confirm deletion
〖SHIFt〗 Switch
〖 ¯¯¯ 〗 Input value is too large When you input tare or weight value
〖 ___ 〗 Input value is too small When you input tare or weight value
〖noACC〗 No any accumulated content when you check accumulation
〖-oL- 〗 Overload warning Tare + Net weight exceed full capacity + 9e
〖-Lb- 〗 Low battery warning Turn off automatically one minute later
〖U*.**〗 The voltage of current battery
〖 End 〗 End when parameter setting or calibration ready
〖 OFF 〗 Turn off
〖Unstb〗 Input value before STB light on
USER MANUAL 14

CHAPTER 6 TROUBLESHOOTING GUIDE


PROBLEM POSSIBLE CAUSE SOUTION
No display when put Discharged battery Recharge
on Defective battery Replace
Defective button Requires authorized service
Power button not properly Press ON/OFF in three
depressed seconds
Digits flash Low battery Recharge the battery
Recharge indicator Battery charger damaged Check the battery charger
light can’t turn on The Battery charger is not Insert plug again
plugged properly
Display doesn’t Faulty load cell or PCB Requires authorized service
respond to weight Load cell cable damaged Check and replace load cell
changes Out of calibration Check calibration

Display experiences Scale electronics do not After turning on, heating 3-5
excessive Zero drift stabilize after turning on minutes.
between weighment Put the scale on ground for Scale should be hung
too long time
Displayed weight Scale not Zeroed before Depress ZERO before
shows large error applying weight applying weight
Requires recalibration See calibration
Kg/lb wrong selection See operation
Battery can’t recharge Battery damaged Replace battery
Outlet damaged Replace the outlet
Remote distance Receiver window is dirty Check and clean
shortened
Remote Control’s battery is Replace Remote Control’s
low battery.
USER MANUAL 15