The document provides instructions for operating an EC-TDS analyzer model CM 183 ELICO. It outlines steps for installation, calibration, and taking measurements of samples. Key steps include keeping the analyzer clean and dust-free, soaking new electrodes in distilled water for 24 hours, calibrating using a 0.01N KCL solution that should read 1409μS/cm at 25°C, rinsing electrodes with distilled water between samples, and switching the analyzer off after taking measurements.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
992 views3 pages

Ec-Tds Analyser - CM 183 ELICO.: 1) Works Instructions

The document provides instructions for operating an EC-TDS analyzer model CM 183 ELICO. It outlines steps for installation, calibration, and taking measurements of samples. Key steps include keeping the analyzer clean and dust-free, soaking new electrodes in distilled water for 24 hours, calibrating using a 0.01N KCL solution that should read 1409μS/cm at 25°C, rinsing electrodes with distilled water between samples, and switching the analyzer off after taking measurements.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 3

EC-TDS ANALYSER –CM

183 ELICO.
1) Works Instructions
1.1 Keep all installation requirements as per given in user’s manual.

1.2 Always keep EC-TDS meter clean, dust free, free form vibration & shocks.

1.3 Always keep the electrode in fresh D.M. water, daily change this D.M. water

1.4 Ensure, various containers are absolutely clean, dry &free from alkalinity.

1.5 The quantity of solutions/ samples are sufficient enough to ensure immersion of
sensor part of the cell/ probes.

1.6 Keep the sample-filled-container near the cell /probe stand and lower the holding
clamp to dip the sensor part of the cell in the sample. Insert thermo- probe also into the
sample.

1.7 Always keep a bundle of tissue paper and rinse bottle containing D.M. water .

1.8 A new EC-TDS meter cell should be soaked in distilled water at least 24 hours .

2) Operation &Calibration
2.1 Switch on the EC-TDS meter and wait for a few minute (5 to 10 min.) for warm up .

2.2 Rinse the cell with D.M. water &wiping with tissue paper .

2.3 For operation , operate the EC-TDS meter as per operation given in EC-TDS meter
manual CM 183 (3.5 operation instructions ,peg no. 6 to 14 )

2.4 For calibration prepared 0.01N KCL solution(procedure given in manual in ANNEXURE-
F, it should be 1409μS/ cm ∓ 1 % at 25 0 C .

2.5 Dip the cell in beaker containing standard solution of KCL( 1409μS/cm at 250 C ) and
check the conductance as per 2.3 . If the reading on screen is in between 1409∓ 1 %
then ,the performance of EC-TDS meter is ok . Record the observed reading in In House
Calibration Register of EC-TDS Meter.
2.6 Rinse the cell with D.M. water and dip in to D.M. water.

Note:-Cleaning of the electrode, Immerse the electrodes of the cell in dilute


Potassium Dichromate (K2Cr2O7) in 6 N Sulphuric Acid (H2 SO4) for 24 Hrs ones in
month. Rinse in running water followed by D.M. water.

For Re- Plastinizing, Follow the procedure given in manual of EC-TDS meter in

ANNEXURE – E on peg E-1.

3) Measurement of EC or TDS of Sample.


3.1 Switch on the EC –TDS meter and rinse the electrode with D.M. water.

3.2 Dip the electrode in a beaker containing sample and follow the procedure 2.3.

3.3 After taking reading of EC or TDS, Remove the electrode from sample & rinse with D.M.
water, wipe with tissue paper, then clamp on stand and immersed in D.M. water.

3.4 Switch off the EC-TDS meter.

SPARKLE CLEAN TECH PVT. LTD.


Always moving forward…
EC-TDS ANALYSER –
CM 183 ELICO
Work Instructions
1.1 Keep all installation requirements as per given in user’s manual .

1.2 Always keep EC-TDS meter clean, dust free, free from vibration &
shocks.
1.3 Always keep the electrode in fresh D.M. water, daily change this D.M.
water.
1.4 Ensure, various containers are absolutely clean, dry & free from
alkalinity.
1.5 The quantity of solutions/ samples is sufficient enough to ensure
immersion of sensor part of the cell/probes.
1.6 Keep the sample-filled-container near the cell /probe stand and lower
the holding clamp to dip the sensor part of the cell in the sample. Insert
thermo- probe also into the sample.
1.7 Always keep a bundle of tissue paper and rinse bottle containing D.M.
water.
1.8 A new EC-TDS meter cell should be soaked in distilled water at least 24
hours.